EC/TDS/SALT Testr: Nstruction Manual

This document provides instructions for using a waterproof multi-range EC/TDS/Salt tester with temperature display. The tester can be switched on by pressing the ON/OFF key and measures conductivity, TDS or salt concentration based on the selected model. The user can select different measurement ranges by pressing and holding the °C/°F key while turning the tester on. The tester automatically temperature compensates readings to 25°C and displays the temperature in the lower display. The HOLD function allows freezing the measurement reading on the display.

Uploaded by

laili
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
257 views

EC/TDS/SALT Testr: Nstruction Manual

This document provides instructions for using a waterproof multi-range EC/TDS/Salt tester with temperature display. The tester can be switched on by pressing the ON/OFF key and measures conductivity, TDS or salt concentration based on the selected model. The user can select different measurement ranges by pressing and holding the °C/°F key while turning the tester on. The tester automatically temperature compensates readings to 25°C and displays the temperature in the lower display. The HOLD function allows freezing the measurement reading on the display.

Uploaded by

laili
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

This Instruction Manual is also available for download at: www.eutechinst.com or www.4oakton.

com Switching On Measurement


Press ON/OFF key to switch on the tester. The LCD shows the power-up sequence 1. Press the ON/OFF key to switch on the tester. The ‘MEAS’ indicators appears
as illustrated in Figure 2. When the tester is on, if you do not press a key for 8.5 when the tester is in measurement mode.
minutes, the tester automatically switches off to conserve batteries. 2. Dip the electrode into the test solution making sure that it is fully immersed.
Stir to clear any trapped air bubbles from the electrode and let the reading
INSTRUCTION MANUAL CAL ATC ATC stabilize. For plus models, you can opt for the cup style measurement by filling
pp m
MEAS HOLD MEAS
the electrode cup with sample of test solution.
Ec.p 20.00 0.0
EC/TDS/SALT Testr ON/OFF
pH% S R.mV °C °F

°C
°F
ms

Au o
uS

26.8 °C Note: The LCD indicates ‘Or’ (over range) if the reading is outside the selected
Large Screen Waterproof Multi Range range. If this occurs, select an appropriate range to suit the reading.
Model number & Selected range & Measurement
software revision number upper limit of the range mode
Conductivity/TDS/Salt Tester 3. The upper display shows the main reading (conductivity/TDS/Salt) of the
with Temperature Display Figure 2: Power Up Sequence solution, automatically temperature compensated (ATC) to normalized temperature
of 25°C. The lower display shows the temperature of the solution.
Introduction Range Selection
Thank you for selecting microprocessor-based waterproof EC / TDS / SALT tester with Depending on the selected model, you can set the tester to limit its reading to a
large dual line display. You have one of the following models: particular measuring range (PU, LO or HI) or full scale (AUTO). The default setting HOLD Function
is AUTO. When you select a range other than AUTO, the tester can be calibrated This feature lets you freeze the display for a delayed observation.
• ECTestr11 • ECTestr11+ • TDSTestr11 • TDSTestr11+ • SALTTestr11
only for that particular range. If you try to measure a sample which has a higher
conductivity/TDS value than that of the selected measuring range, the LCD shows 1. Press HOLD key to freeze the measurement. The tester goes to hold mode and
Non-plus models (ECTestr11, TDSTestr11 & SALTTestr11) come with user-replaceable
‘OR’ error message. Refer Specif ica
Specifica tions section for available ranges of the
ications ‘HOLD’ indicator is displayed in LCD. The measurements are frozen and the ‘MEAS’
two-pin type sensor and have many user friendly features such as Dual-range
selected model. indicator disappears.
measurement, the Hold function, Automatic Temperature Compensation (ATC) and
Self-Diagnostic Messaging capabilities.
To select a rrang
ange:
ange: 2. Press HOLD key again to release the measurement. The ‘HOLD’ indicator is no
1. Switch off the tester. Press and hold °C/°F key and then switch on the tester longer displayed. The tester goes back to measurement mode.
Plus models (ECTestr11+ & TDSTestr11+) come with the user-replaceable cup type
using ON/OFF key. Release °C/°F key.
sensor and have additional features such as Multi-range measurement, up to 3-point
calibration and higher resolution measurement.
2. The tester goes to range selection mode. The LCD shows the currently selected MEAS
ATC
ppm
HOLD HOLD
ATC
ppm
HOLD MEAS
ATC
ppm

Range (the default is AUTO) in the lower display. The upper display shows the
Before You Begin maximum possible reading for the selected range. Press HOLD key repeatedly
1500 1500 1535
Remove the electrode’s protective cap. Soak the electrode for a few minutes in until you see the required range (PU, LO or HI). 26.8 °C
26.8 °C
26.2 °C

alcohol to remove any oil stains on the electrodes which will affect the accuracy of Measurement HOLD mode Measurement

the tester. Rinse thoroughly with de-ionized water and shake off dry. Note: If no key is pressed within 5 seconds, LCD shows power-up sequence and mode mode

tester goes to measurement mode. Figure 4: HOLD Function


Key Functions 3. The tester automatically confirms the last selection if no key is pressed for 5
K ey Function seconds. Upper display momentarily shows ‘CO’. The LCD shows power-up
- Power on and off the tester sequence and the tester goes to measurement mode.
(The tester automatically switches of, if no button is pressed for 8.5 seconds) Temperature Unit of Measurement Selection
- In measurement mode, temperature reading switches between Celsius &
°C/°F
Fahrenheit This feature lets you set the unit of measurement of temperature to either Celsius (°C)
- In calibration mode, switches the tester to temperature calibration mode or Fahrenheit (°F).
- In temperature calibration mode, exits calibration mode without confirming
2 2 2 2
calibrated values
- In measurement mode, switches to hold mode freezing the display. 1 °C/ °F HOLD
ENT
HOLD
ENT
HOLD
ENT
When the tester is in the measurement mode, press °C/°F button. The temperature
HOLD
- In hold mode, switches back to measurement mode Tester
20.00 200.0 2000 20.00 display toggles between the Celsius and Fahrenheit reading.
- In manual calibration and temperature calibration modes, exits calibration mode Off ms us us ms
without confirming calibrated values
- In range selection mode, selects a range ON/OFF
Au o PU LO HI ATC ATC
- In measurement mode, enters calibration mode Press & hold °C/°F MEAS ppt °C/ °F MEAS ppt
key and switch on
- In calibration mode, adjusts the calibration values
- In hold mode, enters TDS factor setting mode
the tester
5 seconds 3
6.00 6.00
- In TDS factor setting mode, adjusts TDS factor
Power-up °C

Sequence CO 25.5 78.0 °F

+
Note: INC & DEC keys are located inside the Auto confirmation occurs for
battery compartment.. Refer figure 1. the last selected range
if no key is pressed for 5 sec.
HI
Figure 5: Temperature unit of measurement selection
Note: For ECTestr11 & ECTestr11+ models, the -
caption of HOLD key is ‘HOLD/ENT’ Figure 3: Range selection sequence from AUTO to HI for ECTestr11+
Figure 1: Battery compartment
-1- -2- -3-
About Calibration 2 3 4 5 Note: If you have selected a specific measuring range for the tester, the lower display
After
To ensure higher accuracy, the tester must be calibrated on a regular basis. 1
HOLD
ENT 1 sec.
HOLD
ENT shows the corresponding calibration standard value that matches the selected
Calibration can be manual or automatic (only for ECTestr11 & ECTestr11+); calibration
Tester
Off YES CO YES CO measuring range.
can be 1-point or multi-point. You can choose any combination of the above two 4. Rinse the electrode with the calibration standard that you intend to calibrate and
options for calibration. If you calibrate the tester for 1-point, the calibration is applied ON/OFF A.CAL A.CAL
Press & hold INC then dip the electrode in the other beaker with same calibration standard. Swirl
for all the measuring ranges. In applications where you need higher accuracy, and key and switch on
gently to create a homogenous sample and allow time for the reading to stabilize.
the tester
when you intend to measure values in more than one range, it is recommended to
2 4
select multi-point calibration. Power-up Note: For multi-point calibration, the lower display automatically locks at the
Sequence
calibration standard value that closely matches. The tolerance range is ± 50% of
Selection of Automatic or Manual Calibration the calibration standard. The tester shows error message ‘Er.1’ if you try to calibrate
ECTestr11 & ECTestr11+ models support both automatic & manual calibration while A.CAL
HOLD
ENT
HOLD
ENT with a solution whose conductivity is outside the tolerance range.
all other models have to be calibrated manually. In automatic calibration, the tester Note: Press INC or DEC key if you wish to exit from auto calibration, during any of
automatically detects and verifies known conductivity standard solutions (84uS, the above steps.
1413uS & 12.88mS) . In manual calibration, you can use non-standard solutions Figure 6: Selecting auto calibrtaion & 1-point calibration for ECTestr11+
which may be specific for your application. 5. Press HOLD/ENT key to confirm the calibration. LCD shows ‘CO’ for 2 seconds. The
calibration is complete and the tester returns to measurement mode, if this is a 1-
The factory default is Automatic calibration (Auto). You can enable or disable point calibration.
automatic calibration as described below (for only ECTestr11 & ECTestr11+ models). Auto Calibration
6. For multi-point calibration, the tester goes to the next calibration point, lower
Auto calibration feature is available only for conductivity in ECTestr11 & display showing next calibration standard values. Rinse the electrode in de-ionized
Selection of 1-point or Multi-point Calibration ECTestr11+ models. Make sure that ‘auto calibration’ is enabled as described in water and repeat step 4 & 5 to continue calibrating with next calibration standard
the previous section. Auto calibration is suitable if you use standard conductivity solution.
The factory default is 1-point Calibration. For higher accuracy, it is recommended that
solutions for calibration process.
you calibrate the tester for multiple ranges if you intend to measure values in multiple
ranges. You can enable or disable multi-point calibration as described below. Note: The tester shows error message ‘Er.0’ and returns to measuring mode if the
If you selected 1-point calibration, you need to choose a calibration standard temperature of the calibration solution is not within 0°C to 50°C.
corresponds to the selected measuring range of the tester listed below. If you
Note: If you have selected a specific measuring range for the tester, selecting multi-
have selected multi-point calibration & AUTO measuring range, you can choose any Note: The tester shows error message ‘Er.1’ if you press HOLD/ENT key before the
point calibration has no meaning, as the tester can only be calibrated for 1-point
of the calibration standards listed below. During auto calibration, the tester tester recognizes the calibration standard.
for the selected range. Set the tester to ‘AUTO’ measuring range if you wish to
recognizes the calibration standard if its value is within 50% tolerance. For multi-
calibrate multi-points.
point calibration (with AUTO measuring range), the tester automatically scans
2 3 4 5
To ena
enabble/disa
le/disabble auto calibr
calibraation and multi-point calibr
multi-point calibraation: through all possible calibration points until all of them are calibrated. CAL ATC CAL ATC CAL ATC HOLD
CAL ATC

1. Switch off the tester. Press and hold INC key and then switch on the tester using 1 ENT

ON/OFF key. Selected Measuring Range Calibration Standard Measurement


Mode OR
CAL 0.0
uS
1410uS
1410
uS
2. The tester goes to auto calibration selection mode. The lower display shows ‘A.CAL’ PU (0 to 200.0 uS/cm) 84 us
and the upper display blinks the current choice (‘Yes’ or ‘No’). 1413 1413 CO
LO (0 to 2000 uS/cm) 1413 uS
Displayed for Tester dipped in Confirmation of 1st point
HI (0 to 20.00 mS/cm) 12.88 mS 2 seconds and 1413 uS std. solution displayed for
Note: This mode is available only for ECTestr11 & ECTestr11+ models. For other tester goes to
CAL
3
ATC
and automatically detects 2 seconds
calibration mode the calibration std.
models, go directly to 1-point selection mode, described in step 4 below. AUTO 84 us, 1413 uS, 12.88 mS

Press INC or DEC key to select ‘Yes’ (to enable auto calibration) or ‘No’ (to disable
0.0
uS

auto calibration) To pr
preepar
paree calibr
calibraation standar ds:
standards: 12.88
The two calibration stds. values
Note: Press °C/°F key, if you wish to skip this setting without confirming changes. Use fresh calibration standard solutions listed in the above table. Prepare each displayed sequentially until
tester finds a matching solution
solution in two beakers - one for rinsing and the other for calibration. Rinse the within ± 50% tolerance.
Note
Note: Press °C/°F key twice, if you wish to return to measurement mode without electrode in de-ionized water before calibration.
confirming changes. 6
CAL ATC CAL ATC CAL ATC
HOLD
3. Press HOLD/ENT key to confirm the selection. The display shows ‘CO’. To be gin automa
begin tic calibr
automatic calibraation:
ENT

4. The tester goes to 1-point calibration selection mode. The lower display shows 1410mS
12.80 mS
12.80
mS
Measurement
Mode
‘1.Pnt’ and the upper display blinks the current choice (‘Yes’ or ‘No’). 1. Switch on the tester. Make sure the tester is in measuring mode. Press INC or
Press INC or DEC key to select ‘Yes’ (to enable 1-point calibration) or ‘No’ (to 12.88 12.88 CO
DEC key to enter conductivity calibration mode.
disable 1-point calibration, i.e. to enable multi-point calibration). 2. ‘CAL’ indicator appears in LCD. The display briefly shows ‘CAL’ and the number
The next calibration std. Tester dipped in
displayed until tester finds 12.88 mS std. solution
Confirmation of 2nd point
displayed for
a close match within & automatically detects 2 seconds
of points the tester will be calibrated. ± 50% tolerance the calibration std.
Note: Press °C/°F key if you wish to skip this setting without confirming. 3. The upper display shows the conductivity reading and the lower display
5. Press HOLD/ENT key to confirm the selection. The display shows ‘CO’ for few sequentially shows calibration standard values 1413 uS & 12.88 mS (for Figure 7: Two-point automatic calibration sequence for ECTestr11
seconds and then shows power-up sequence The tester goes to measurement ECTestr11) or 84 uS, 1413 uS & 12.88 mS (for ECTestr11+) if the
mode. measuring range of the tester is set to AUTO.
-4- -5- -6-
Manual Calibration 5. For multi-point calibration rinse the electrode in de-ionized water and repeat Temperature Calibration
step the above steps with another standard solution.
In manual calibration, the tester allows 1-point calibration for each measuring range. Temperature calibration need not be performed every time, unless the temperature
You can use customized calibration solutions with known conductivity/TDS values to reading differs from that of an accurate thermometer. If temperature calibration is
calibrate the tester. . The following table shows acceptable conductivity/TDS ranges of 2 3 3 4 performed, Conductivity/TDS/Salt calibration is mandatory.
calibration solutions for each measuring range. Make sure your calibration solutions ATC CAL ATC CAL ATC
After
CAL ATC
1
are within the given ranges. 5 sec.
1. Switch on the tester. Make sure the tester is in measuring mode. If required,
Measurement
Mode OR
CAL 1400 uS
1413uS
1413
uS press °C/°F key to select the desired unit of measurement for temperature
Selected Acceptable Calibration Standard Range
Measuring Range Conductivity TDS/Salt
1410 1410 CO (Celsius or Fahrenheit). Dip the tester into a solution of known temperature
Displayed for Tester dipped in Upper display reading Confirmation and allow time for the temperature reading to stabilize.
PU 2.0 - 200.0 uS/cm 2.0 - 200.0 ppm 2 seconds and 1413 uS std. solution Adjusted to the actual displayed for
Tester goes to Conductivity value of 2 seconds 2. Press INC or DEC key to bring the tester to the calibration mode. CAL’ indicator
LO 200 - 2000 uS/cm 200 - 2000 ppm Calibration mode Calibration std. solution
appears in LCD. Immediately press °C/°F key to switch to the temperature
HI 2.00 - 20.00 mS/cm 1.00 - 10.00 ppt
calibration mode.
AUTO Select a calibration standard nearer to application sample Figure 8: One-point manual calibration sequence for ECTestr11+

To pr
preepar
paree calibr
calibraation standar ds:
standards: Note: When you enter calibration mode, if the conductivity/TDS/salt reading is outside
Use fresh calibration solutions. Measure conductivity/TDS values of the solution with a the specified range (Or), the tester shows ‘Er.1’ error message. You can still proceed
meter known to be accurate. Prepare each solution in two beakers - one for rinsing to the temperature calibration mode by pressing °C/°F key immediately. If the °C/°F
and the other for calibration. Rinse the electrode in de-ionized water before key is not pressed within 2 seconds, the tester exits the calibration mode and returns
calibration. TDS Factor Setting to the measurement mode.
TDS factor is only applicable for TDSTestr11 & TDSTestr11+ models. 2 2 3 3
To be gin man
begin ual calibr
manual calibraation: The factory default TDS factor is 0.71. You can adjust the TDS factor to suit MEAS
ATC CAL ATC
°C/ °F
CAL CAL
1
1. Switch on the tester. Make sure the tester is in measuring mode. Rinse the different samples of your applications.
electrode with the calibration standard that you intend to calibrate and then dip
Measurement
Mode
7.50 mS OR
7.50
mS
26.8 °C
OR
27.5 °C
the electrode in the other beaker with same calibration standard. Swirl gently to To cchang
hang actor::
hangee TDS ffactor
actor Set required unit of 26.8 °C
7.50 26.8 °C
26.8 °C

create a homogenous sample and allow time for the reading to stabilize. 1. Switch on the tester. Make sure the tester in measurement mode. Press HOLD
measurement for
temperature
Dipped in a solution Calibration Temperature Calibration mode
2. Press INC or DEC key to enter calibration mode. The ‘CAL’ indicator appears in key to bring the tester to the HOLD mode. of known temperature mode

LCD. The display briefly shows ‘CAL’ and the number of points the tester will be 2. Press INC or DEC key to enter the TDS factor setting mode. 4
calibrated. 3. The upper & lower displays of LCD show the last configured TDS factor. The ATC CAL
MEAS
3. The upper display shows the measured conductivity/TDS reading of the solution upper display is adjustable. Use the INC or DEC key to adjust the TDS factor.
based on previous calibration (if any) and the lower display shows the default The adjustable range is 0.4 to 1.0 7.50 mS
27.5 °C
(uncalibrated) conductivity/TDS reading. °C
CO
27.5
Note: If you do not press INC or DEC key within 5 seconds, the tester shows the Confirmation
Measurement
Note: The tester shows error message ‘Er.1’ : confirmation ‘CO’ and returns to measurement mode. mode displayed for
2 seconds
(a) If the reading is over range (Or) of selected measuring range of the tester ,
or 4. Wait for 5 seconds for the tester to automatically confirm the new setting by
(b) If the default (uncalibrated) reading is not within the acceptable calibration displaying ‘CO’ and return to the measurement mode. Figure 10: Temperature calibration sequence

standard range.
Use INC and DEC keys to adjust the upper display to the correct conductivity/TDS 3. The upper display shows the measured temperature reading based on the last set
2 3 3 4
value of the calibration solution. ATC
offset (if any) and the lower display shows the default (uncalibrated) temperature
1 HOLD HOLD ppt After
5 sec. reading based on factory settings. Use INC and DEC keys to adjust the upper
Note: The calibration adjustment window is ± 50% from the default reading.
Measurement
Mode 6.35 OR
0.71 0.50 0.50 temperature reading to the known temperature value of the solution.

Note: If you do not press INC or DEC key within 5 seconds, the tester shows the 26.8 °C
0.71 0.71 CO
Notes: The temperature adjustment window is ± 5°C (± 9°F) from the default reading.
confirmation ‘CO’ and returns to the measurement mode. However, the tester is HOLD mode TDS Factor setting mode Confirmation
displayed for
not calibrated to new values yet. The old calibration is still active. If this happens, 2 seconds 4. Wait for 5 seconds for the tester to automatically confirm the temperature
press INC or DEC key once again to enter calibration mode. calibration value by displaying ‘CO’ and return to the measurement mode.
Figure 9: TDS Factor setting
4. Wait for 5 seconds for the tester to automatically confirm the calibration by Note: To exit temperature calibration mode without confirming the calibration, press
displaying ‘CO’ and return to the measurement mode. °C/°F key or HOLD/ENT key before the automatic confirmation takes place.
.

Note: To exit calibration mode without confirming the calibration, press HOLD/ENT Note: The tester shows error message ‘Er.0’ and returns to measuring mode if the
key before the automatic confirmation takes place. temperature of the solution is not within 0°C to 50°C.

Note: The tester shows error message ‘Er.0’ and returns to measuring mode if the
temperature of the calibration solution is not within 0°C to 50°C.

-7- -8- -9-


Reset Electrode Maintenance Accessories
Reset option allows you to restore the calibration and other parameters back to 1. Always keep the sensor electrodes clean. Rinse the electrodes with de-ionized Item Order Code
factory default settings. water and wipe them dry with clean cloth before storing with its protective cap. ECTESTR, TDSTESTR & SALTESTR TDSSENSOR
For cup type electrodes, remove the white plastic cup and insert to thoroughly replacement sensor
1. Switch off the tester. Press and hold the HOLD key and then switch on the tester clean viscous solutions. Never scratch electrodes with a hard substance. ECTESTR+ & TDSTESTR+ TDSSENSORPLUS
using ON/OFF key. Release HOLD key. replacement sensor
2. The lower display shows ‘rSt’ (reset) and the upper display blinks ‘No’. Use INC 2. For better performance, soak the electrode in alcohol for 10 to 15 minutes
or DEC key to select ‘Yes’ (to proceed with resting) or ‘No’ (to quit without and rinse with de-ionized water before starting any measurement process.
resetting). This is to remove dirt and oil stains on the electrode which may affect the Warranty
accuracy of the measurements. The waterproof testers are warranted to be free from manufacturing defects for 1 year
Note: Press °C/°F key if you wish to skip to measurement mode without making any and electrode module for 6 months, unless otherwise specified. If repair, adjustment or
selection. Electrode Replacement replacement is necessary and has not been the result of abuse or misuse within the
3. Press HOLD key to confirm your selection. LCD shows ‘CO’. If ‘Yes’ is selected, the When the tester fails to calibrate or gives fluctuating readings in calibration time period specified, please return the tester - freight prepaid - and correction will be
tester resets to its factory default values as listed below. LCD shows power-up standards, you need to change the electrode module. You can replace the made without charge. Out of warranty products will be repaired on a charge basis.
sequence and tester goes to measurement mode. electrode module at a fraction of the cost of a new tester.
Return of Items
2 3 1. With dry hands, grip the ribbed tester collar with electrode facing you. Twist the Authorization must be obtained from your distributor before returning items for any
1
HOLD HOLD collar counter clockwise (see Figure13-A). Save the ribbed tester collar and reason.
Tester
CO Power-up O-ring for later use.
Off Sequence
2. Pull the old electrode module away from the tester.
When applying for authorization, please include information regarding the reason the
ON/OFF 3. Align the four tabs of the new electrode module so that they match the four item(s) are to be returned.
Press & hold HOLD slots on the tester (see Figure 13-B).
key and switch on
the tester 4. Gently push the module into the slots to sit it in position. Push the smaller
Note: Eutech Instruments/Oakton Instruments reserve the right to make improvements
2 O-ring fully onto the new electrode module. Push the collar over the module
in design, construction and appearance of products without notice. Prices are subject
and thread it into place by firmly twisting clockwise.
to change without notice.
YES Note: It is necessary that you recalibrate the tester prior to measurement after
HOLD

an electrode replacement.

A B Insert Electrode Module


Large
Specifications
Rotate ribbed collar O-Ring
Figure 11: Resetting sequence away from you
Model ECTestr11
ECTestr11 ECTestr11+
ECTestr11+ TDSTestr11
TDSTestr11 TDSTestr11+
TDSTestr11+ SALTT
SALTTestr11
TTestr11
Parameter Factory Default Range: PU - 0 to 200.0 uS/cm - 0 to 200.0 ppm -
LO 0 to 2000 uS/cm 0 to 2000 uS/cm 0 to 2000 ppm 0 to 2000 ppm -
User calibration (conductivity/TDS/salt) (Reset) HI 0 to 20.00 mS/cm 0 to 20.00 mS/cm 0 to 10.00 ppt 0 to 10.00 ppt 0 to 10.00 ppt
Ribbed collar Small Large Tab Resolution: PU - 0.1 uS/cm - 0.1 ppm -
Temperature unit of measurement Celsius (°C) Ribbed collar
O-Ring
Small Tab
LO 10 uS/cm 1 uS/cm 10 ppm 1 ppm -
HI 0.10 mS/cm 0.01 mS/cm 0.10 ppt 0.01 ppm 0.10 ppt
Temperature offset 0 Figure 13: Removing collar & inserting electrode Accuracy ± 1% of Full Scale
Auto calibration (for ECTestr11 & ECTester11+) Enable Calibration Type Auto or Manual Auto or Manual Manual Manual Manual
Calibration Points 1 or 2 points 1, 2 or 3 points 1 or 2 points 1, 2 or 3 points 1 point
1-point calibration Enable
Self-Diagnostic Messages Calibration Window ± 50% from each point
Conductivity calibration factor (for ECTestr11 & ECTester11+) 1.0 Calibration PU - 2.0 - 200.0 uS/cm - 2.0 - 200.0 ppm -
Standard Range:LO 200 - 2000 uS/cm 200 - 2000 uS/cm 200 - 2000 ppm 200 - 2000 ppm -
TDS factor (for TDSTestr11 & TDSTester11+) 0.71 3 Bars indicates Battery is full (100%) (Manual) HI 2.00 - 20.00 mS/cm 2.00 - 20.00 mS/cm 1.00 - 10.00 ppt 1.00 - 10.00 ppt 1.00 - 10.00 ppt
Sensor Type Two-pin Cup Two-pin Cup Two-pin
2 Bars indicates 50% of the battery life is left
Low battery TDS Factor - - 0.4 to 1.0 (Default 0.71)-
indicator 1 Bar indicates 25% of the battery life is left Temperature
Changing Batteries Blinking battery casing indicates the need to replace
Range in °C
Range in °F
0.0 to 50.0°C
32.0 to 122°F
batteries with fresh ones as specified by manufacturer Resolution 0.1°C (0.1°F)
Replace the batteries when the low battery indicator Accuracy ±0.5°C (±0.9°F)
The sensor electrodes short circuited
starts blinking. Calibration point 1 point
Or / Ur Replacement sensor is not connected properly to the Calibration Window ± 5°C (± 9°F)from factory default
1. Open the battery compartment lid (with attached + Over range /
(Still) tester during sensor replacement ATC 0 to 50°C
lanyard loop). Under range Temp Coefficient 2% per °C
signal Measured value or temperature value exceeds the specified
Normalization Temp 25.0°C
2. Remove old batteries by pulling plastic ribbon. maximum or minimum value
Auto Off 8.5 minutes after last key press
Replace with fresh ones. Note polarity as shown in ATC / Or / Blinking ‘ATC’, ‘Or’ or ‘Ur’ indicates that there is a shor t Operating Temp 0 to 50°C
Ur (Blinking)
figure 11. - Plastic
Ribbon
or open circuit at the built in temperature sensor
Power Battery 4 X 1.5V”A76" micro alkaline battery
Calibration error due to temperature value not within the Battery Life >150 hrs
Er.0
Er.0
Error specified range LCD Display Custom Dual Display27mm(H)X21mm(W)
Figure 11: Battery compartment Message Calibration error due to Conductivity/TDS/Salt value not
Er.1
Er.1 Dimensions Tester: 16.5 cm X 3.8 cm; 90g
within the specified calibration standard range Weight Boxed: 22cm X 6cmX 5cm; 170 g
-10- -11-
68X068056 Rev 2 May 06