Pseudocapacitance Calculation PDF
Pseudocapacitance Calculation PDF
I- Introduction
This circuit is not present in the electrical
It can sometimes be useful to substitute a RQ circuits list already in the ZFit tool, so the user
circuit by a RC circuit (Fig. 1). must add this circuit using to the Edit menu.
This calculation is only available for a circuit
(R/Q). It corresponds to the determination of a
capacitance value C at an angular frequency
c 2f c corresponding to the maximum
imaginary part on the Nyquist circle obtained
by fitting with the circuit (R/Q). Then, this value
is the solution of the following equation
1 1
C Q1 / R (1 / ) 1
2π RQ 1/ 2πRC
II-Experimental part
A lithium-ion battery, with LiFePO4 as the
positive electrode, was studied in this
application note. The impedance diagram was
recorded considering the setting described in
the Fig. 2, i.e. in the 10 kHz – 10 mHz range
with Va = 10 mV, pw = 1, Na = 1, and the drift
Fig. 3: Nyquist impedance diagram (blue) and
correction button ticked. The impedance
ZFit analysis (red) of a Li-ion battery with a
diagram obtained is given in Fig. 3. L1+R1+Q2/R2+Q3 equivalent circuit.
The ZFit tool was used to fit data with the
equivalent circuit L1+R1+Q2/R2+Q3 (Fig. 4).
Bio-Logic Science Instruments, 1 rue de l'Europe, F-38640 Claix - tel: +33 476 98 68 31 – Fax: +33 476 98 69 09 1
Web: www.bio-logic.info
However, even by clicking on the PseudoC
button, it is not possible to calculate a pseudo-
capacitance value as displayed in Fig. 11. The
pattern in the equivalent circuit is Q/(R+Q) and
not the required one (R/Q).
Fig. 4: L1+R1+Q2/R2+Q3 equivalent electrical
circuit.
References
[1] Electrical circuits containing CPEs, Bio-Logic
website, http://www.bio-logic.info/potentiostat/notes
nd
[2] Impedance spectroscopy, 2 Edition, ed.
Barsoukov E., Ross Mac Donald J., 2005
Bio-Logic Science Instruments, 1 rue de l'Europe, F-38640 Claix - tel: +33 476 98 68 31 – Fax: +33 476 98 69 09 3
Web: www.bio-logic.info