Perawatan Kolom GC
Perawatan Kolom GC
Figure 1: Cutting Fused Silica Tubing Figure 2: Proper and Improper Cut Capillary End
4 Correct X Incorrect
apply force in a
downward direction
Examples of: (a) A Properly Cut Fused Silica Column End
(b) An Improperly Cut End
tubing should break
cleanly
1 www.phenomenex.com
GC Column Installation
B. Capillary GC Installation with the Cool-Lock™ Nut
• Avoid burning your fingers again – cools with the oven ~exclusive~
• N o Wite-Out or Tipp-Ex needed
® ®
patent pending
1 2 3
1. Insert column through column nut and 2. Hand-tighten the column nut into the gauge 3. Unscrew the installation gauge and install
ferrule. Cleanly cut ≥ 1 cm of the column until the ferrule starts to compress. Adjust the the nut and column into the instrument.
that is protruding from the nut and ferrule. length of the column to the desired length.
Insert the end of the column and column Further tighten the nut into the gauge until the 4. Make sure Cool-Lock Nut is in the full
nut into the bottom of the Cool-Lock Nut column cannot slide. If using the Cool-Lock extension position to achieve the correct
Installation Gauge. Nut, tighten its bottom ferrule. Make sure lower premeasured position, and allow for proper
body/ferrule is in the full extension position. leak checking.
Otherwise, mark the column at the base of the
column nut to ensure the correct installation
depth after installation.
www.phenomenex.com 2
GC Column Conditioning and Testing
A. Column Conditioning Zebron’s Text Mix
1. Allow sufficient time for the carrier gas to flow through the column to • Suitable for Phenomenex Zebron™ and equivalent brands
purge any oxygen that may be in the system.
• C onvenient way to check column performance
2. Raise the temperature of the column to the maximum isothermal
• Affordable and easy-to-use
operating temperature that is listed on the individual Zebron™ GC
Column Test Report. Maintain this temperature until a constant baseline
is achieved. Conditioning times will depend on the phase identity and Zebron Phase Test Mix
thickness, with thicker films taking longer to stabilize. If necessary Part Number
in order to minimize the downtime of the instrument, columns can be ZB-1 AG0-5155
conditioned overnight at the maximum isothermal temperature.
ZB-1ms AG0-7805
B. Installation Testing ZB-1HT AG0-5155
Inject a detectable unretained sample, such as methane for an FID, to determine ZB-5 AG0-5155
dead volume time and linear gas velocity at the desired column temperature. ZB-5ms AG0-7578
Adjust gas pressure for optimal flow depending on carrier gas selection. ZB-5MSi AG0-8362
ZB-5HT AG0-5155
1. The non-retained peak must have ideal peak shape or installation is faulty
and needs to be redone. ZB-35 AG0-5156
ZB-35HT AG0-5156
C. Checking for Leaks ZB-1701 AG0-5156
Use a thermoconductivity detector to check for leaks. It is highly sensitive to ZB-1710P AG0-5156
H2, He, and N2 and will not contaminate the instrument or column. Liquid leak ZB-50 AG0-5157
indicators are not recommended for capillary columns. There is the risk of ZB-624 AG0-5159
drawing the liquid into the column or fittings and contaminating the system.
ZB-WAX AG0-5158
NOTE: If Vespel ferrules are being used, leakage can occur after the initial
® ZB-WAXplus™ AG0-7869
heating phase due to ferrule deformation. Be sure that the fitting is re-tightened ZB-FFAP AG0-5158
after this initial heating phase, and then carefully check all corrections for leaks.
ZB-XLB AG0-7578
Match your column phase to the correct standard to set up initial running
conditions. Should you have any questions on the proper use of these Check
Standards, contact your local Phenomenex GC Specialists.
3 www.phenomenex.com
GC Hints and Tips
A. Four Key Steps to Protecting the GC Column and
Extending Lifetime
Figure 3. Some liners that are available pre-packed with glass wool or provide
It is important to protect the column and instrument components from additional column protection:
exposure to dirty samples. Non-volatile or high molecular weight Description Dimensions Material* Quartz Wool
GC Model No. ID x L x OD (deactivated) (Y / N) Mfr. No. Part No. Unit
components can contaminate the stationary phase, causing peak (mm)
resolution, lower accuracy, and poor column lifetime. Cutting off the
damaged portion will usually restore the column performance, but over Split/Splitless
time performance will degrade to a point where the column can no longer 5880/5890/6890 4 x 78.5 x 6.3 B (y) Y 210-4004 AG0-7515 5/pk
be used. If you are experiencing rapid degradation of column performance, AG0-7582 25/pk
there are several simple ways to help protect your column and increase
lifetime: Split/Splitless, Recessed Gooseneck Liner
5880/5890/6890 4 x 78.5 x 6.45 B (y) Y 5183-4691 AG0-4661 5/pk
1. Ensure Proper Sample Preparation 5183-4692 AG0-4662 25/pk
a. Filter your samples prior to injection with Phenex™ syringe filters (see
Appendix A for details). Cup Splitter/Split Liner
5880/5890/6890 4 x 78.5 x 6.3 B (n) N 5183-4699 AG0-4647 5/pk
b. Use Strata® or Strata™-X SPE Cartridges to eliminate contaminants 5183-4700 AG0-4648 25/pk
(see Appendix D for details).
Cup Splitter/Split Liner
2. Use a Guardian™ or Z-Guard AutoSystem™ 3.5 x 100 x 5 B (n) N 0330-5181 AG0-4663 5/pk
a. Guard Columns – Standard Guards. Z-Guard columns are 5 or 10
meter pieces of deactivated tubing that are connected to an analytical Splitless Single Taper/Liner
column using a glass press-fit connector. The tubing acts like a trap 5880/5890/6890 4 x 78.5 x 6.45 B (y) Y 5183-4693 AG0-4657 5/pk
for non-volatile residues that would otherwise damage the stationary 5183-4694 AG0-4658 25/pk
phase of your analytical column. * B = Borosilicate; Deactivated = Yes (y) or No (n)
b. Guardian Integrated Guard Columns. Unlike traditional guard
columns, there is no mechanical connection between the guard
and the analytical column. The result: all the benefits of a guard
column without the possibility of leaks or activity resulting from a
faulty connection. Please contact a Phenomenex GC Specialist for 4. “Baking Out” the Column
information on the Guardian. The easiest way to reduce column contamination is to add a short, high
temperature bake out at the end of the standard GC method. This bake out
3. Use a Wool Liner helps remove high boiling contaminants that would otherwise remain in the
The liner is the first line of defense for the column and the style column and cause damage. To bake out, the final oven temperature needs
chosen can make a big difference in how much contamination to be set high enough to ensure elution of these compounds, but not so high
gets onto the column. The easiest thing to do is to add a small as to cause thermal damage. This can be done either isothermally, or more
amount of silanized glass wool to a liner, which traps the non- commonly, via a gradient or ballistic increase until the last components
volatile compounds and prevents them from entering the column elute from the column.
(Figure 3). See Appendix B for other recommended GC accessories.
Caution: NEVER exceed the upper temperature limits of the column. DO
NOT exceed more than 15 minutes at the upper isothermal temperature
Caution: glass wool can also add activity for acids, bases, and limit specified for the column.
pesticides. Crushing the glass wool can lead to increased activity, so it is
recommended to purchase pre-packed liners, rather than try to pack your
own.
www.phenomenex.com 4
GC Hints and Tips
B. Carrier Gas Selection and Flow Optimization
It is advisable to use the highest purity gas possible. Ultra high purity Optimal velocities uopt Optimal velocities uopt
(99.99 %), ultra pure carrier (99.995 %), or even research grade (99.9999 for low df values: (cm/sec) for high df values: (cm/sec)
%) is preferred to minimize critical impurities, instrument downtime and
H2 40 H2 25
troubleshooting. Three types of gas are commonly used as a carrier gas:
He 25 He 15
1. Hydrogen (H2): Hydrogen will yield maximal number of theoretical
plates for thin film columns and the high efficiency is largely retained N2 10 N2 7
at velocities higher than uopt. Hydrogen is not generally recommended
due to its hazardous nature.
2. Helium (He): When hydrogen is not used, helium is the best C. Column Storage
alternative for speed and sensitivity.
Important! The column may be left in the instrument for short-
3. Nitrogen (N2): Nitrogen is the last choice for thin film columns. For term storage. Ensure a flow of carrier gas through the column at
thick film columns, nitrogen yields the highest number of theoretical 100-200 °C. For long-term storage, disconnect from the GC and cap or
plates. However, the optimal velocity is fairly low (long analysis seal the capillary ends. Oxygen and moisture can degrade or irreversibly
times), and the loss in efficiency at higher velocities is high. If
damage the column, especially cyanopropyl-based phases. Wax
resolution is sufficient, hydrogen or helium are good alternatives.
(polyethylene glycol) and cyanopropyl-based phases are also susceptible
to UV-induced degradation and should be shielded from light (fluorescent
Figure 4: Carrier Gas Selection and Velocity Optimization Plots
or sunlight). Store the column in the original box. Upon reinstallation, cut
column ends to ensure that septum fragments or other debris have not
Carrier Gas Length
been left in the column.
H(u) curves for different film H(u) curves for different k values
thicknesses
5 www.phenomenex.com
GC Hints and Tips
D. Chemical Compatibility / Solvent Rinsing of Zebron™ Capillary Columns
Important! Water and organic solvents such as those listed in Table 1 will not damage Zebron column stationary phases. However, inorganic acids and
bases should be completely avoided or rapid degradation and permanent damage to the stationary phase will result. In the event chemical damage is
incurred, the removal of 0.5-2 meter of capillary off the front end will often restore column performance. Select the rinse solvent from the Table below.
ZB-XLB,
ZB-XLB-HT
ZB-624
ZB-35, ZB-35HT
OK
ZB-1701,
ZB-1701P
ZB-50
ZB-WAXplus™ AVOID
ZB-WAX AVOID
ZB-FFAP AVOID
MR-1, MR-2
ZB-Drug-1
ZB-BAC1,
ZB-BAC2
ZB-Bioethanol
0.25 5 40
0.32 5 40
0.53 10 20
Important: Rinse from the back to the front of the column to avoid pushing inlet contaminants further into the column.
www.phenomenex.com 6
Appendix A
Phenex™ Syringe Filters
Choose a pore size based on the nature of your sample and chromatographic method
Sample Description Recommended Filter Pore Size
General clarification of GC samples. 0.45 µm
Removal of fine pariculate matter prior to GC analysis. 0.20 µm
0.20 µm
Phenex-RC AF0-3203-12 100/Pk AF0-2203-12 100/Pk AF0-8203-12 1 100/Pk
(Regenerated Cellulose) AF0-3203-52 500/Pk AF0-2203-52 500/Pk AF0-8203-52 1 500/Pk
Please inquire with Phenomenex about other Phenex Syringe filters offerings.
Above syringe filters are non-sterile. Housing is made of medical-grade polypropylene (PP) and offer luer lock inlet/slip outlet connections, unless otherwise indicated.
1. 26 mm diameter.
7 www.phenomenex.com
Appendix B
Other Recommended GC Accessories
www.phenomenex.com 8
Appendix C
Vials
Headspace Vials
Round Bottom Flat Bottom
20 mL 20 mL
Compatible With The Following Autosamplers Headspace Autosampler Vials
Brand Model Round Flat Part No. Description 100/pk
Bottom Bottom AH0-7546 Headspace Vial 20 mL clear, round bottom,
Agilent HS 7694 • 23 x 75 mm
Carlo-Erba HS 850 • AH0-7547 Headspace Vial 20 mL clear, flat bottom,
Dani HS 39.50 / 86.50 • 23 x 75 mm
Fisons HS 850 • AH0-7548 Aluminum Seal, with PTFE/Silicone septum,
PerkinElmer HS 100 / 101 / 40 • 0.130” thickness
Tekmar-Dohrmann 7000HT • AH0-7550 Aluminum Seal, with Pharma-Fix molded
ThermoQuest HS 850, Trace GC • pressure-release septum
2000 + HS 2000 AG0-8350 Aluminum Seal, Magnetic Crimp Cap, 20 mm
Varian Genesis • • TEF/SIL, blue
9 www.phenomenex.com
Kit packs include caps.
Appendix D
Strata® and Strata™-X SPE Products
SPE products can help to eliminate contaminants and matrix effects,
resulting in increased signal-to-noise ratios and longer column lifetimes.
Phenomenex manufactures two types of SPE sorbents:
Product Offerings:
1. Strata - Silica Solid Phase Extraction (SPE) Products 2. Strata-X-Polymeric Solid Phase Extraction (SPE) Sorbents.
(Same functions as Strata but with the additional benefits):
• Quickly and easily cleans samples, concentrates analytes, and
facilitates solvent switching • De-conditioning resistant surface resists dry out and de-activation
and provides worry-free manual and automated processing
• Removes contaminants and eliminates matrix effects for improved
separation and column lifetime • pH stability from 1-14 for flexible method development
• A bility to load up to 66 % more sample per gram in comparison to
silica-based sorbents per gram
Formats: Strata and Strata-X SPE products are available in multiple formats:
Standard Tubes for Teflon®-Coated Tubes for Large Volume/Flash Tab-less Tubes
Traditional SPE Extra Inert Analysis Analysis Cleanup
20 x 2.0 mm
Need help selecting an SPE Sorbent? Please call your Phenomenex Sample Preparation Specialist for help or visit www.phenomenex.com/info/spehelp
www.phenomenex.com 10
Australia Luxembourg
t: 02-9428-6444 t: +31 (0)30-2418700
f: 02-9428-6445 f: +31 (0)30-2383749
[email protected] [email protected]
Austria Mexico
t: 01-319-1301 t: (55) 5018 3791
f: 01-319-1300 f: (310) 328-7768
[email protected] [email protected]
Netherlands
Belgium t: 030-2418700
t: +31 (0)30-2418700 f: 030-2383749
f: +31 (0)30-2383749 [email protected]
[email protected]
New Zealand
Canada t: 09-4780951
t: (800) 543-3681 f: 09-4780952
f: (310) 328-7768 [email protected]
[email protected] Norway
Denmark t: 81 00 20 05
f: +45 4810 6265
t: 4824 8048 [email protected]
f: 4810 6265
[email protected] Puerto Rico
t: (800) 541-HPLC
Finland f: (310) 328-7768
t: (09)4789 0063 [email protected]
f: +45 4810 6265 United Kingdom
[email protected]
t: 01625-501367
France f: 01625-501796
t: 01 30 09 21 10 [email protected]
f: 01 30 09 21 11 All other countries:
[email protected] Corporate Office USA
t: (310) 212-0555
Germany f: (310) 328-7768
t: 06021-58830-0 [email protected]
f: 06021-58830-11
[email protected]
Ireland
t: 01 247 5405
f: +44 1625-501796
[email protected]
GU61770410_L
www.phenomenex.com
Italy
Phenomenex products are available worldwide. For the distributor in your country,
t: 051 6327511 contact Phenomenex USA, International Department at [email protected]
f: 051 6327555
[email protected]