MELODY TECHNICAL MANUAL
CHAPTER 5
♦ GENERAL ELECTRIC/ELECTRONIC
DIAGRAMS
♦ LIST OF FUSES
♦ CABLES
VILLA SISTEMI MEDICALI CAP 5 Page 1 of 52
MELODY TECHNICAL MANUAL
MELODY
VILLA SISTEMI MEDICALI CAP 5 Page 2 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 3 of 52
MELODY TECHNICAL MANUAL
MELODY/B
VILLA SISTEMI MEDICALI CAP 5 Page 4 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 5 of 52
MELODY TECHNICAL MANUAL
92-082-5
CPU mother board
Unit ON/OFF drive
C-ARM up/down motor
Compression motor drive
Magnetic clutch
Collimator light
Rotation brake
Mo/Rh filter
Exposure counter
Calibration:
RV2 compression force
RV1 Inverter power supply control circuit
Replacement:
IC20: version for MELODY
Jumpers:
JP1 short=voltage safety bypass
JP2 short/open = toggle state of stand by lamp
JP3 = short
JP4 short= standard clutch voltage
JP4 open= max clutch voltage
Options:
Deata Plus (no differences)
VILLA SISTEMI MEDICALI CAP 5 Page 6 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 7 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 8 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 9 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 10 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 11 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 12 of 52
MELODY TECHNICAL MANUAL
This page has been intentionally left blank
VILLA SISTEMI MEDICALI CAP 5 Page 13 of 52
MELODY TECHNICAL MANUAL
01-170
CPU board
Calibration:
no calibration is requested
Replacement:
EEPROM contains specific firmware
2 k RAM contains panel settings and CLOCK
512 k RAM contains calibrations
VILLA SISTEMI MEDICALI CAP 5 Page 14 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 15 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 16 of 52
MELODY TECHNICAL MANUAL
02-172
Potter-bucky driver
Calibration:
No calibration requested
Replacement:
No adjustment
Jumpers:
JP1, 2, 3, 4, 5, 6 open
JP7 closed
VILLA SISTEMI MEDICALI CAP 5 Page 17 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 18 of 52
MELODY TECHNICAL MANUAL
92-083-6
Keyboard / control panel
Calibration:
RV1 LCD contrast
Replacement:
short S8 in solder-side to bypass digital inclination display
XTAL= 4.3008 MHz
VILLA SISTEMI MEDICALI CAP 5 Page 19 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 20 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 21 of 52
MELODY TECHNICAL MANUAL
This page has been intentionally left blank
VILLA SISTEMI MEDICALI CAP 5 Page 22 of 52
MELODY TECHNICAL MANUAL
92-073-4
Filament power supply
Calibration:
refer to procedure
RV1, RV2, RV3, RV4, RV9 (factory adjusted)
Replacement:
anode current calibration
filament safety circuit calibration:
voltmetric RV3, RV5, JP2
amperometric RV6, RV7, JP1
Configuration:
92-073-3 for single anode tubes
JP3 = closed
VILLA SISTEMI MEDICALI CAP 5 Page 23 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 24 of 52
MELODY TECHNICAL MANUAL
92-074-3
Rotating anode starter
Calibration:
refer to procedure
Replacement:
safety circuit calibration (RV1, RV2)
voltage selection (220V - 230V - 240V)
Warning:
Circuit ground is floating with respect to other blocks in the
unit having the same symbol.
Be careful making measurements on this PCB, use
only insulated instruments.
VILLA SISTEMI MEDICALI CAP 5 Page 25 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 26 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 27 of 52
MELODY TECHNICAL MANUAL
This page has been intentionally left blank
VILLA SISTEMI MEDICALI CAP 5 Page 28 of 52
MELODY TECHNICAL MANUAL
92-090-2
IR remote control transmitter
Calibration:
no calibration is requested
Replacement:
no adjustment
VILLA SISTEMI MEDICALI CAP 5 Page 29 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 30 of 52
MELODY TECHNICAL MANUAL
92-091-2
IR remote control receiver
Calibration:
no calibration is requested
Replacement:
no adjustment
VILLA SISTEMI MEDICALI CAP 5 Page 31 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 32 of 52
MELODY TECHNICAL MANUAL
96-153-0
Mirror motor driver
Calibration:
no calibration is requested
Replacement:
no adjustment
VILLA SISTEMI MEDICALI CAP 5 Page 33 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 34 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 35 of 52
MELODY TECHNICAL MANUAL
MELODY-MELODY/B FUSES
Numeration is referred to general schematics
F1 = mains power
20A 10×38 WEBER CH10 CEI 32-1-4 IEC239
F2 = mains breaker
0.1A delayed 6.2×30
F3 = c-arm motor3.15 delayed 6.2x30
F4 = toroidal transformer primary TF205 – 3.15A delayed 6.2 X 30
F5 = Power on circuit on PCB 92-082 40mA delayed 5×20; F8 is in series to F10 on
PCB 92-082.
F6 = toroidal transformer secondary TF203 – 10A 10X38 weber CH10 CEI
32-1-4 IEC 239
F7 = toroidal transformer secondary TF203 – 10A 10X38 weber CH10 CEI
32-1-4 IEC 239
F8 = External lamp
500mA delayed 6x30
FUSES ON MELODY PCBs
Numeration is referred to single schematics of PCBs
PCB 92-073 - FILAMENT POWER SUPPLY PCB
F1 = 3.15A delayed 6.2×30
F2 = 3.15A delayed 6.2×30
PCB 92-074 - ROTATING ANODE
F1 = 10A fast 6.2×30
F2 = 4A delayed 6.2×30
F3 = 40mA delayed 5×30
PCB 92-082 - CPU MOTHER BOARD PCB
F1 = 2A delayed 5×20
F2 = 1.6A delayed 5×20
F3 = .5A delayed 5×20
F4 = 8A delayed 6.2×32
F5 = 8A delayed 6.2×32
F6 = .5A delayed 5×20
F7 = 2.5A delayed 5×20
F8 = 2.5A delayed 5×20
F9 = 1A delayed 5×20
F10 = 40mA delayed 5×20; F10 is in series to F8 on electric wiring
PCB 93-109 - INVERTER CONTROL PCB
F2 = delayed 5×20 –1A
F3 = Fuse for semiconductors IR B1000/20
VILLA SISTEMI MEDICALI CAP 5 Page 36 of 52
MELODY TECHNICAL MANUAL
TRANSFORMERS PROTECTION ACCORDING TO NORMATIVE CEI 62-5 IEC601-1
WARNING: modifying fuse type and/or value may create risks of fire!
TF 129 (PCB 92-082) Protected by delayed fuses
TF 213 (PCB 92-074) Protected by delayed fuses
TF 205 Protected by delayed fuses
TF 203 Protected by termoswitch that acts on mains breaker turning off the device
and avoiding self power-on at the termoswitch restoration.
Further protection against transient overload by fuses on secondary
circuits.
VILLA SISTEMI MEDICALI CAP 5 Page 37 of 52
MELODY TECHNICAL MANUAL
CABLES
VILLA SISTEMI MEDICALI CAP 5 Page 38 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 39 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 40 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 41 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 42 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 43 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 44 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 45 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 46 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 47 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 48 of 52
MELODY TECHNICAL MANUAL
UART4 4800 BAUDPCB 01-170
CN 13
UART3 9600 BAUD
CN 14
VILLA SISTEMI MEDICALI CAP 5 Page 49 of 52
MELODY TECHNICAL MANUAL
PCB 01-170
UART2 4800 BAUD
CN15
VILLA SISTEMI MEDICALI CAP 5 Page 50 of 52
MELODY TECHNICAL MANUAL
VILLA SISTEMI MEDICALI CAP 5 Page 51 of 52
MELODY TECHNICAL MANUAL
This page has been intentionally left blank