Standard Calibration Procedure
Doc. No. Call/SCP/022 Rev. 00 May 01, 2015
Temperature Calibrator
STANDARD CALIBRATION PROCEDURE
TEMPERATURE CABLIBRATOR
Prepared by
Muhammad Ajlan Signature Date : 01.05.2015
Reviewed by
Muhammad Sarwar Signature Date : 01.05.2015
Approved by
Qaswar Mehmood Signature Date : 01.05.2015
Controlled Copy No. Uncontrolled Copy
Issued to :
Page 1 of 4
Standard Calibration Procedure
Doc. No. Call/SCP/022 Rev. 00 May 01, 2015
Temperature Calibrator
This document is the property of Qstar Engineering, and is issued only for persons or entity as
written on this document. It is not to be retransmitted, printed or copied without prior written
permission of company.
Table of Contents
Sr. No. Title Page No(s).
1.0 Scope 03
2.0 Standard Equipment 03
3.0 Preliminary Operations 03
4.0 Calibration Procedure 03
5.0 Related Documents 03
Page 2 of 4
Standard Calibration Procedure
Doc. No. Call/SCP/022 Rev. 00 May 01, 2015
Temperature Calibrator
1.0 Scope
This procedure provides guidelines for calibration of Temperature Calibrator.
2.0 Standard Equipment
2.1 Process Calibrator & Sensor with a valid certificate of calibration traceable to
international standard in accordance with an industrial code or a regulation shall be used
as a standard for calibration.
2.2 The reference standard shall be kept under controlled environmental conditions e.g. dust
free environment and Lab. temperature 22°C ± 3°C. Lab Humidity 60%RH ± 20%RH.
3.0 Preliminary Operations
3.1 The details of Temperature Calibrator received for calibration shall be entered in the
Calibration Job Form (Form: Call/001).
3.2 The Temperature Calibrator shall be inspected and marked () in relevant block of
equipment’s primary inspection section of Calibration Data Sheet (Form: Call/013).
4.0 Calibration Procedure
4.1 The Temperature Calibrator to be calibrated shall be maintained at same place of
standard (Process Calibrator) for 03 hours minimum.
4.2 Insert the Sensor of Process Calibrator into Temperature calibrators Thermo Well and
connect the output of Sensor with Processor Calibrator.
4.2 Select the Temperature Calibrator under test for the desired range and give the sufficient
time for stable reading.
4.3 The Temperature Calibrator shall be checked at approximately 20% 40%, 60% and 80%
of full range.
4.4 The relevant section of Calibration Data Sheet (Form: Call/013) shall be marked.
4.5 The readings of Temperature Calibrator shall be recorded in measured value column
against the standard readings in set value column of Calibration Data Sheet (Form:
Call/013).
4.6 After completion of calibration the Temperature Calibrator and the equipment should be
secured.
Page 3 of 4
Standard Calibration Procedure
Doc. No. Call/SCP/022 Rev. 00 May 01, 2015
Temperature Calibrator
4.7 Calibration Certificate (Form: Call/002) shall be issued for calibrated Temperature
Calibrator as per Calibration Data Sheet Record (Form: Call/013).
4.8 A Calibration Sticker (Form: Call/003) containing following information shall be affixed
on the calibrated Thermocouple.
a) Item code
b) Calibrated on
c) Next calibration date
5.0 Related Documents
All the latest formats are documented in Master List of Document / Record and placed in CQF
file for reference.
All the latest Standard Calibration Procedures (SCP) are controlled and issued to concerned
Calibration Engineers.
Page 4 of 4