0% found this document useful (0 votes)
35 views

Portable Helium Oxygen Analyzer: Trimix 4001

The Trimix 4001 is a portable analyzer that measures the percentages of helium, oxygen, and nitrogen in gas mixes for technical diving. It automatically corrects for environmental factors to eliminate errors from manual calculations. The analyzer has a large, backlit LCD display and provides additional metrics such as maximum operating depth. It is powered by a rechargeable battery and is waterproof, with a floatation case and alarms to ensure diver safety.

Uploaded by

entico
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
35 views

Portable Helium Oxygen Analyzer: Trimix 4001

The Trimix 4001 is a portable analyzer that measures the percentages of helium, oxygen, and nitrogen in gas mixes for technical diving. It automatically corrects for environmental factors to eliminate errors from manual calculations. The analyzer has a large, backlit LCD display and provides additional metrics such as maximum operating depth. It is powered by a rechargeable battery and is waterproof, with a floatation case and alarms to ensure diver safety.

Uploaded by

entico
Copyright
© © All Rights Reserved
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Portable Helium Oxygen Analyzer

Trimix 4001
A portable Trimix Analyzer designed to ensure the correct mix of helium, oxygen and nitrogen for technical diving.
This easy-to-operate diving instrument analyzes gas mixes, automatically correcting for environmental conditions and
eliminating the risk of errors from manual calculations. It has a large backlit LCD display that is easily read even in
low light conditions, and which provides additional information such as a Maximum Operating Depth calculation. The
features make the Trimix 4001 one of the best nitrox analyzers on the market for diving oxygen analyzers.

Designed for field use


The AII-4001 will operate for 16 hours of continuous
use, and fully charges in two hours. An optional adapter
is available to charge the unit using the 12 V DC power
outlet from a car or boat, so the unit can be charged
even if there is no easy access to mains power.

The case is watertight and with the lid closed will float
dropped overboard, and being red will be easily spotted.

Highlights
• Displays O2, He and balance gas (0-100%)
• Automatically adjusts for environmental conditions to
avoid risk of nitrogen narcosis during dives down to
46 meters
• Includes temperature, barometric pressure and RH
sensors
• Optional integral pump
• Long life battery with 16 hours of use, with two-hour
charge time
• 15-minute auto-off feature to save battery
• Watertight, IP65 enclosure

www.aii1.com
Technical Specifications
Measurement range 0-100% Helium, 0-100% Oxygen

Accuracy ±2% of range at constant conditions

Response time T90 < 10 seconds

Sensor model He: AII-41-100, O2: AII-11-75D

Sensor life He: 10 years, O2: 60 months in air

Calibration interval Prior to use with ambient air or a known concentration of oxygen

Inlet pressure < 5 psig with 1-2 SCFH. Ambient air can be used for calibration if optional integral pump is fitted

Display Backlight LCD 6.4 x 6.4cm (2.5 x 2.5”)

Dimensions: 221 x 190.5 x 96.5mm (8.7 x 7.5 x 3.8")

Ingress protection: Watertight case (when closed) IP65

Compensation Temperature, barometric pressure, relative humidity

Storage temperature 0°C to 50°C (32°F to 112°F) 50°C intermittently

Operating temperature 0°C to 45°C (32°F to 113°F)

Rechargeable battery, 9 V DC from 110/220 V AC


Power (can charge battery in 2 hours).
Optional 12 V DC car charger.

Automatic Correction - Eliminating operator error


At depths down to 46 meters, divers are at risk of nitrogen narcosis, oxygen toxicity and decompression sickness. To
avoid this, a mixture of helium, oxygen and nitrogen is used for breathing gas instead of air and the ratio of these
gases is calculated precisely for the depth of the dive. Changes in temperature, pressure and humidity affect the results
of calibrations and measurements, with a possible error of up to 6.7%. The AII-4001 automatically compensates for
environmental changes, with built-in pressure, temperature and
RH sensors, eliminating the risk of operator error.

Analytical Industries Inc. 2855 Metropolitan Place, Pomona, CA 91767 USA.


Tel: 909-392-6900, Fax: 909-392-3665, www.aii1.com, e-mail: [email protected]
Please note: Analytical Industries Inc. adopts a continuous development program which sometimes necessitates specification changes without notice.
Please contact us for the latest version. Issue No: Portable Helium Oxygen Analyzer_partnumber_V2_UK_1018

www.aii1.com

You might also like