Michael G. Gore - Spectrophotometry and Spectrofluorimetry - A Practical Approach-Oxford University Press, USA (2000)
Michael G. Gore - Spectrophotometry and Spectrofluorimetry - A Practical Approach-Oxford University Press, USA (2000)
and Spectrofluorimetry
Practical Approach Series
Edited by
Michael. G. Gore
Division of Biochemistry and Molecular Biology,
School of Biological Sciences, University of
Southampton.
OXPORD
UNIVERSITY PRESS
OXFORD
UNIVERSITY PRESS
Finally, I would like thank the authors for their participation and contributions
to the text.
It was with deep regret that I learned of the demise of Professor Jorge E.
Churchich shortly before the completion of this book. He was a dedicated
scientist who inspired all who worked with him.
Southampton M.G.G.
1999
VI
Contents
vii
CONTENTS
viii
CONTENTS
2 Background 69
Basic photophysics and time dependence of fluorescence and phosphorescence
decays 69
Fluorescence and phosphorescence energy transfer and sensitized luminescence 71
Observed time dependence of fluorescence 72
Decay associated spectra (DAS) and discrete lifetimes versus lifetime
distributions 75
Polarized excitation and emission anisotropy decay 75
Data analysis 78
3 Equipment for time-resolved fluorescence measurements 84
Excitation sources 85
Detectors 86
Recording electronics 86
4 Phosphorescence 89
Phosphorescence of proteins 89
Time-dependent phosphorescence anisotropy 93
Acknowledgements 95
References 95
ix
CONTENTS
xi
CONTENTS
xii
CONTENTS
xiii
CONTENTS
xiv
Protocol list
XV
PROTOCOL LIST
xvi
Abbreviations
A absorbance, absorption
c, C concentration
D attenuance (A)
DSC differential scanning calorimetry
e molar absorbance coefficient
eATP 1, N6-ethenoadenosine diphosphate
FAD(H) Flavin adenine dinucleotide (reduced form)
Gdn-HCl guanidine hydrochloride
HEPES N-2-hydroxyethylpiperazine-N'-2-ethanesulphonic acid
I intensity of transmitted light
I0 intensity of incident light
ITC isothermal titration calorimetry
l pathlength
MES 2-(N-morpholino)ethanesulphonic acid
MOPS 3-(N-morpholino)propanesulphonic acid
NAD(H) Nicotinamide adenine dinucleotide (reduced form)
OD optical density
PIPES piperazine-N,N'-bis(2-ethanesulphonic acid)
PMT Photomultiplier tube
oF Quantum yield of fluorescence
s second
T transmittance, transmission or temperature
TNP-ATP 2'(3')-o-(2,4,6-trinitrophenyl)adenosine 5'-triphosphate
Tris tris(hydroxymethyl)aminomethane
TRP Tryptophan
TYR Tyrosine
UV ultraviolet
xvii
This page intentionally left blank
Chapter 1
Introduction to light
absorption: visible and
ultraviolet spectra
Robert K. Poole* and Uldis Kalnenieks1
* Krebs Institute for Biomolecular Research, Dept. of Molecular Biology
and Biotechnology, The University of Sheffield, Firth Court,
Western Bank, Sheffield S10 2TN
f
Institute of Microbiology and Biotechnology, University of Latvia,
Kronvalda Boulevard 4, Riga, LV-1586, Latvia
1 Introduction
1.1 Radiation and light
light is a form of electromagnetic radiation, usually a mixture of waves having
different wavelengths. The wavelength of light, expressed by the symbol \, is
defined as the distance between two crests (or troughs) of a wave, measured in
the direction of its progression. The unit used is the nanometre (nm, 10-9 m).
Light that the human eye can sense is called visible light. Each colour that we
perceive corresponds to a certain wavelength band in the 400-700 nm region
(Figure 1). Spectrophotometry in its biochemical applications is generally con-
cerned with the ultraviolet (UV, 185-400 nm), visible (400-700 nm) and infrared
(700-15 000 nm) regions of the electromagnetic radiation spectrum, the former
two being most common in laboratory practice.
Figure 1 The visible spectrum. Colours shown are of the light beam at the wavelengths on
the scale so the colour the eye perceives of a reflectant object in 'white' light is that due to
wavelengths that are not absorbed. Thus, the 'green oxidase' (cytochrome d) looks green but
absorbs light in the blue and red regions of the spectrum.
ROBERT K. POOLE AND ULDIS KALNENIEKS
2 Spectrophotometry
2,1 The Beer-Lambert law
The most widespread use of UV and visible spectroscopy in biochemistry is in
the quantitative determination of absorbing species (chromophores), known as
spectrophotometry. All spectrophotometric methods that measure absorption,
including various enzyme assays, detection of proteins, nucleic acids and dif-
ferent metabolites, reside upon two basic rules, which combined are known as
the Beer-Lambert law. Lambert's law states that the fraction of light absorbed by a
transparent medium is independent of the incident light intensity, and each successive layer
of the medium absorbs an equal fraction of the light passing through it. This leads to an
exponential decay of the light intensity along the light path in the sample,
which can be expressed mathematically, as follows:
loglo(I0/I) = kl 2
where I0 is the intensity of the incident light, I is the intensity of transmitted
light, I is the length of the light-path in the spectrophotometer cuvette, and k is a
constant for the medium, which is deciphered by Beer's law. Beer's law claims
that the amount of light absorbed is proportional to the number of molecules of the chromo-
phore through which the light passes. In other words, the constant k is proportional to
the concentration (c) of the chromophore: k = ec, where e is the molar
absorption coefficient, a property of the chromophore itself. e is numerically
INTRODUCTION TO LIGHT ABSORPTION: VISIBLE AND ULTRAVIOLET SPECTRA
appropriate units. It is sometimes said that Klett units are meaningless (because
they cannot be defined as A can) but, in fact, they are more useful: the variation
between Klett meters is likely to be much smaller than the variation obtained
when measuring a turbid sample in different 'biochemical' spectrophotometers.
Arthur Koch has written eloquently (5, 6) of the factors that influence light
scattering by particles in suspension and of our ability to detect sensitively such
light scattering. In the most practically oriented of these papers (6), Koch com-
pares experimentally various laboratory spectrophotometers and wavelengths
for their suitability as 'turbidimeters'; variation is due largely to what portion of
the scattered light is viewed by the detector. Most of the light scattered by a
microbial suspension deviates by only a few degrees (7°) from the incident beam
(Figure 2). Thus, using an instrument in which the detector is placed as far as
possible from the cuvette will greatly enhance sensitivity, since a low concen-
tration of scattering particles will be sufficient to scatter some of the light from
the incident beam. It can be very instructive to measure optical density of a
single sample (perhaps fixed with formaldehyde to prevent further growth and
Figure 2 Properties of ideal turbidimeter and spectrophotometers. For a clear sample, the
position of the detector (PMT, photomultiplier tube) is unimportant (A). If a turbid sample is
examined in a spectrophotometer where the PMT is far removed from the cuvette (B), the
instrument will behave as a sensitive turbidimeter because it will detect little of the
scattered light. If a turbid sample is examined in a spectrophotometer where the PMT is
close to the cuvette (C), the instrument will detect absorbance even though much of the light
is scattered.
ROBERT K. POOLE AND ULDIS KALNENIEKS
3.3 NAD(P)H
Upon reduction of both nicotinamide adenine dinucleotide (NAD) and nicotina-
mide adenine dinucleotide phosphate (NADP) to NADH and NADPH, respect-
ively, a strong absorption peak appears at 340 nm. The millimolar absorption
coefficient at this wavelength for both reduced coenzymes is 6.22 mM-1 cm-1. As
these compounds function as coenzymes of many dehydrogenases, a large
number of enzyme reactions can be followed (either directly, or using coupled
assays) by monitoring the change of absorbance at 340 nm (8).
3.4 Carotenoids
Carotenoids are members of a large class of red, yellow and orange plant pig-
ments, also found in the membranes of some bacteria. Carotenoids contain long
carbon chains with many conjugated double bonds. Delocalization of electrons
is responsible for their strong absorption in the visible region. In general, the
more conjugated double bonds a molecule contains, the stronger and more red-
shifted its light absorption is. Carotenoids are used as intrinsic membrane probes
of transmembrane electric potential in photosynthetic energy-transducing
membranes (9) due to their very rapid electrochromic shift of absorption
maximum in response to transmembrane electric field.
550.5
AA = 0.5
substituent groups attached to the porphyrin ring and thus the particular type
of haem, its redox state, the spin state of the iron and the presence of bound
ligands.
The haem iron of cytochromes, globins, oxygenases and hydroperoxidases
frequently bind ligands resulting in characteristic absorbance changes. For
example, carbon monoxide (CO) binds to the reduced form of certain haem
proteins, acting as a competitive inhibitor with respect to oxygen in respiration.
The most common laboratory application is in CO difference spectroscopy, i.e.
the identification of CO-binding haem proteins by recording the difference
between a reduced sample and another that has been treated with CO. For
example, a terminal oxidase common in bacteria, cytochrome bo', binds CO at
the oxygen-reactive haem O to give an adduct with an absorption maximum at
about 415 nm. Since the reduced but unligated haem O has an absorption maxi-
mum at about 430 nm, a difference spectrum (reduced + CO minus reduced) will
show a peak near 415 nm and a trough near 430 nm. Also, the CO-Fe bond is
photodissociable so that for certain haem proteins a second type of difference
spectrum, the photodissociation spectrum, can be recorded. This is the differ-
ence between the CO-ligated sample after photolysis (i.e. reduced, unligated)
minus the same sample before photolysis (reduced, CO-ligated). For an excellent
INTRODUCTION TO LIGHT ABSORPTION: VISIBLE AND ULTRAVIOLET SPECTRA
4 Spectrophotometer configurations
4.1 Single beam spectrophotometers
Single beam instruments are widely used for routine laboratory measurements
at a single wavelength. As shown in Figure 4, only a single light beam from the
monochromator passes through the sample compartment. The absorbance zero
or 100 per cent transmittance is adjusted with a cuvette containing buffer or
solvent. Then the absorbance (or transmittance) value of a sample cuvette con-
taining sample solution is measured. It is necessary to put the sample and the
reference in the optical beam alternately, by manual operation, typically with
an interval of a few seconds required to change the cuvettes. However, if full
spectra of the sample and reference are required, the interval between the two
measurements is likely to be several minutes, with the consequent risk that
lamp drift or other sources of instability can lead to errors.
It is impossible to give more detailed information or protocols for using the
wide range of single-beam instruments on the market. The advice given here is
that which must apply throughout this chapter: consult the manufacturer's
manual.
Light source Monochromator Sample compartment Detector Amplifier Hardware and software
for display
Reference cell
Sample cell
Light source Monochromator Sample compartment Detector Amplifier Hardware and software
for display
the reference signal 10 and the sample signal I, and the sample absorbance is
obtained as the logarithm of the ratio of the two signals, which, according to
the Beer-Lambert law, is independent of the incident light intensity. In com-
paring a sample and reference at each wavelength, the split beam spectrophoto-
meter produces a 'difference' spectrum (instead of two separately acquired
'absolute' spectra, one for the reference and one for the sample, in a single-
beam device). In the difference spectrum, those features common to sample and
reference, like buffer absorption, moderate turbidity, or optical imperfections,
are cancelled out.
The split beam mode is used to measure either absolute spectra (i.e. the
difference between a sample in one beam and the solvent, or buffer, or even air
in the other beam) or a difference spectrum. For the latter, the intensities of the
two beams after passing through the cuvettes are again compared but, instead
of comparing a sample to a non-absorbing reference, two nearly identical
samples are compared. The difference between the two may be amplified be-
cause any interfering or background absorbance or light scattering is common
to both samples. It should be noted that in difference spectra both peaks
(positive differences) and troughs (negative differences) are anticipated accord-
ing to whether the 'sample' absorbs more or less light, respectively, than the
reference.
The use of a chopper, usually vibrating or rotating with a frequency of about
50 Hz, limits the kinetic resolution of the split-beam instruments. Spectral
changes occurring on a faster timescale cannot be monitored. Recent advances
in component design with improved stability have to some extent limited the
advantages of split-beam methods and there has been a resurgence of interest in
and availability of simpler single-beam designs.
10
INTRODUCTION TO LIGHT ABSORPTION: VISIBLE AND ULTRAVIOLET SPECTRA
in the split beam mode; Section 4.2), the absorbance of a single sample is
compared at two wavelengths—'sample' (A1 in Figure 6) and 'reference' (A2 in
Figure 6). The sample A1 is selected to correspond to the absorbance maximum of
the signal under study (e.g. 550 nm for cytochrome c) and the reference A2 is
chosen to be, ideally, at an isosbestic point (see Section 6.3) or other wavelength
close to A1 where significant absorbance changes are not anticipated. Wave-
length pairs in the fixed dual X mode should be chosen to optimize:
(1) sensitivity, i.e. high signal:noise (choose A for high absorbance coefficient);
(2) specificity (select for freedom from other components); and
(3) freedom from light scattering changes (select pairs as close as possible,
ideally 15-40 nm apart).
In this way, A2 serves as a reference measurement for the observed absorbance
change to be studied at A1. Thus, if there is an undesired change in the absolute
spectrum of the sample during measurement, e.g. due to particles settling out,
the effects on X1 - X2 will be minimized since the undesired change will be
recorded at both wavelengths. Results are presented on a YT recorder, computer
etc.
The dual-wavelength scanning spectrophotometer was developed by Britton
Chance in the Johnson Foundation in the early 1950s and remains immensely
useful and in production (Figure 7). A single light source is used for two
independent monochromators; one is set to a reference wavelength while the
11
ROBERT K. POOLE AND ULDIS KALNENIEKS
other scans the desired spectral range. The two monochromator exit beams
intercept at right angles at a vibrating chopper mirror which transmits only one
of the beams at a time through the cell (the other beam being reflected). The
two beams arrive at the detector, generally a photomultiplier tube, separated, in
time; the signals are separated electronically, compared and recorded. Results
are presented on an XY recorder, computer etc.
In the non-scanning mode, dual-wavelength instruments are generally used
to study kinetic changes at a pair of wavelengths (sample-reference) allowing
the observation of small absorbance changes despite a very high initial absorb-
ance or in the face of light scattering changes arising from particles settling, for
example. In the dual-wavelength scanning mode, spectra are obtained with
reference to an isosbestic point or other suitable reference wavelength. Note
that since only one cuvette is scanned at a time, difference spectra are obtained
only by computing the difference between two scans. In either mode, the dual-
wavelength instrument is valuable for optically isolating components in mix-
tures using wavelength(s) in which the desired component changes absorption
and other components do not.
12
INTRODUCTION TO LIGHT ABSORPTION: VISIBLE AND ULTRAVIOLET SPECTRA
13
ROBERT K. POOLE AND ULDIS KALNENIEKS
14
INTRODUCTION TO LIGHT ABSORPTION: VISIBLE AND ULTRAVIOLET SPECTRA
Wavelength/rim
15
ROBERT K. POOLE AND ULDIS KALNENIEKS
16
INTRODUCTION TO LIGHT ABSORPTION: VISIBLE AND ULTRAVIOLET SPECTRA
17
ROBERT K. POOLE AND ULDIS KALNENIEKS
dependent on spectral bandwidth and natural bandwidth. The former, but not
the latter, can be set by the operator.
To understand spectral bandwidth, it must be appreciated that 'mono-
chromatic' light from a monochromator never consists only of radiation of
precisely one wavelength (as does the light from a laser). Instead, it has a certain
bandwidth, including smaller amounts of light with shorter and with longer
wavelengths than the particular chosen wavelength. The spectral bandwidth is
defined as the band of wavelengths contained in the central half of the entire
band passed by the exit slit of the monochromator (Figure 11). It is the exit slit
(generally a few mm across) whose physical width can be set by the investigator.
To calculate spectral bandwidth from this slit width, we need to know the
reciprocal dispersion of the monochromator, which can be found in manu-
facturers' literature. Typical values of dX/dx, where x is the slit width, are 2 or 4
nm mm"1. Thus, in the latter case, when the exit slit is opened to 0.5 mm, say,
the spectral bandwidth is 2 nm.
In contrast, the natural bandwidth is an intrinsic property of the sample,
independent of the instrument bandwidth, and is defined as the width (in nm)
at half the height of the sample absorption peak, as shown in Figure 11. For
example, the value for the natural bandwidth of the 340 nm peak of NADH is
58 nm, whereas for most cytochromes at room temperature the natural
bandwidths in the a-region are of the order of 10 nm. It is easy to conceive that
having too broad a spectral bandwidth would result in an apparent decrease of
sample absorption. This is because the incident light would contain a large
fraction of radiation with wavelengths poorly absorbed by the sample.
A rarely seen graph, allowing errors in peak height measurements to be
predicted, is shown in Figure 12. This shows the dependence of the absorbance
peak magnitude on the ratio of the spectral bandwidth to the natural band-
Figure 11 The relationship between natural and spectral bandwidths. The natural bandwidth
shown is for NADH (58 nm, curve A). Curve B is a schematic representation of the spectral
bandwidth of the monochromator exit beam. Taken from (18) where the original source is
cited.
18
INTRODUCTION TO LIGHT ABSORPTION: VISIBLE AND ULTRAVIOLET SPECTRA
6 8 10
Figure 12 Graph allowing calculation of the error in measured peak height for a given
monochromator spectral bandwidth and natural bandwidth of an absorption band. Spectral
bandwidth can be obtained from manufacturers' information or by knowing the physical slit
width of the monochromator and the reciprocal dispersion (nm/mm). Since natural
bandwidths of, for example, cytochrome absorption bands are about 10 nm at room
temperature a spectral bandwidth of 2.5 nm (ratio on the abscissa of 0.25) will introduce no
more than a 3% error in measured peak height. However, for low temperature spectra,
spectral bandwidths of about 0.5 nm are required.
19
ROBERT K. POOLE AND ULDIS KALNENIEKS
Figure 13 Resolution of two adjacent bands. In A, the separation between the bands (AX) is
sufficient for them to be resolved using the criterion that the trough between the adjacent
peaks is no higher than 0.8 of the peak height. In B, the separation between the bands
(AX') is just sufficient for the presence of two bands to be seen. In C, the separation
between the bands (AX") is insufficient for resolution and the bands cannot be
distinguished.
obtained is lower than 80% of the peak height. This is sometimes called the
Rayleigh criterion (see Figure 13). Although spectral resolution is a parameter
that greatly concerns the practical spectroscopist, it is inextricably linked to
spectral bandwidth (see above).
5.5 Temperature
The temperature-controlled sample holders available in most commercial
spectrophotometers, when coupled to an external circulating heater or cooler,
can control temperature in the approximate range 0 to 40°C. Specialized
cuvettes are not necessary and the device is useful for kinetic measurements or
observation of a labile sample. However, there are many advantages in kinetics
and wavelength scanning of being able to operate at much lower temperatures.
5.5.1 77 K
Spectroscopy at 77 K (liquid nitrogen temperatures) has been widely used to
detect differences in the absorption spectra of closely related haemoproteins, to
trap unstable intermediates and steady-states of oxidation and reduction, to
slow down rapid rates of reaction, and to detect and measure very low concen-
trations of haemoproteins. The main effects are a sharpening and enhancement
20
INTRODUCTION TO LIGHT ABSORPTION: VISIBLE AND ULTRAVIOLET SPECTRA
of the absorption bands and a blue shift of 1-4 nm. The absorbance changes are
due to:
(1) a true temperature dependence of the absorbance of the chromophore,
resulting in narrowing, sharpening and shifting to shorter wavelengths of
the bands;
(2) light-scattering changes in the medium, which result in an effective in-
crease in pathlength by multiple internal reflections from ice crystals. [It is
worth noting that enhancement effects of related origin may be observed in
highly scattering suspensions (e.g. of intact cells) at room temperature.]
Reference (19) is a classic account of the practical uses of low temperature
spectroscopy. The enhancement effects are strongly dependent on the suspend-
ing medium and the method used to freeze it (20). For example, a devitrified (i.e.
polycrystalline) 1.4 M sucrose solution can give a 25-fold intensification at
— 190°C, compared with 20°C. The presence of an organic solvent also makes
the enhancement factors more reproducible than in dilute buffers. Glycerol has
the added advantage of suppressing the pH changes that occur on freezing, as
does a careful choice of buffer. The pH of Tris buffer changes dramatically
(increasing) on lowering the temperature, while phosphate and acetate buffers,
for example, are relatively temperature-independent. For a detailed account of
the choice of buffers in low temperature work, Douzou (21) should be
consulted.
Accessories for recording spectra at 77 K are available for many but not all
spectrophotometers. The attachments are generally expensive and many workers
have designed and constructed simple devices for their own instruments.
References to examples are given by Jones and Poole (22).
21
ROBERT K. POOLE AND ULDIS KALNENIEKS
6.2 Baselines
A baseline is the wavelength-dependent difference in absorbance either (i)
between two cuvettes in a split beam apparatus when their contents are thought
to be identical or (ii) between two scans of one cuvette in a dual-wavelength
spectrophotometer over a period when the contents are thought not to have
changed. Any irregularities in a baseline will be included in subsequent dif-
ference spectra and may be superimposed on the spectral regions of interest. In
extreme cases, baseline irregularities may be mistaken for spectral peaks and
troughs, and a steeply sloping baseline can alter the apparent position of an
absorption peak. Therefore, baseline flatness should always be checked before
recording a difference spectrum and either presented with the latter or, at least,
reported. The following additional measures can be taken to improve baseline
flatness.
22
INTRODUCTION TO LIGHT ABSORPTION: VISIBLE AND ULTRAVIOLET SPECTRA
wide
Figure 14 Scheme to show choices necessary for obtaining a routine absorbance spectrum.
23
ROBERT K. POOLE AND ULDIS KALNENIEKS
24
INTRODUCTION TO LIGHT ABSORPTION: VISIBLE AND ULTRAVIOLET SPECTRA
6.5.2 Mixing
Cuvette lids or, better, ground stoppers prevent evaporation of solvent and/or
contamination of the cuvette contents. Tapered Teflon stoppers can be drilled to
accommodate the fine needle of a syringe such as those supplied by Hamilton.
Mixing of cuvette contents can be problematic, especially when the cuvette is
almost full and a lid is fitted; a few glass beads (small enough to lie below the
light path) can provide sufficient agitation when a cuvette is inverted. An
alternative is to use disposable mixers or stirring paddles (Kartell). Those with
small horizontal platforms ('plumpers') can be loaded with up to about 100 ul of
an addition, then rotated in the cuvette between thumb and forefinger to
provide simultaneous and fairly rapid addition and mixing. For semi-micro
cuvettes, the mixers and paddles can be trimmed with a blade.
Where the contents of a cuvette must be continuously stirred during the
scanning of a spectrum, a small magnetic bar can be rotated magnetically
within the cuvette. Some (e.g. Spinbars from Bel-Art Products) are circular with
four small vanes and rotate smoothly inside a 10 X 10 mm cuvette. The small
stirrer motor is usually located under the cuvette and the power supply is
outside the spectrophotometer. A mu metal shield protects the photomultiplier
from 'noise'. High quality dc brush motors driven in a feedback loop to maintain
constant speed, not voltage, minimize stirring artefacts, especially important if,
for example, oxygen electrode measurements are to be made in parallel with
25
ROBERT K. POOLE AND ULDIS KALNENIEKS
6.5.5 Cleaning
The importance of clean cuvettes is self-evident. Routinely, all non-disposable
cuvettes should be emptied immediately after use, rinsed repeatedly in the
solvent (e.g. water), then with clean ethanol or acetone and dried with low
pressure air or nitrogen from a cylinder. It is prudent to install a filter (such as
those with pore sizes of 0.45 um used in filter sterilization) in the gas line.
Cuvette washers (e.g. Aldrich) wash, rinse, and dry cuvettes. Cotton wool 'buds'
can also be useful for dislodging interior, stubborn marks and for drying. The
outside optical surfaces should be polished with clean lens tissue. Note that
plastic 'squeezy' bottles generally used for solvents contain plasticizers such as
butyl phthalate, which can interfere with critical UV spectra.
The Perspex windows of low-temperature cuvettes easily become scratched
and will eventually crack. However, provided they are still liquid-tight (at least
for as long as it takes to plunge them and their contents into liquid nitrogen)
they are quite serviceable in this condition. The opacity of the Perspex is
negligible compared with that of the frozen sample.
Neglected cuvettes that cannot be dismantled for cleaning may require soak-
ing overnight in concentrated sulphuric acid containing a few crystals of
dichromate or permanganate, or boiling in distilled water containing a
laboratory detergent.
26
INTRODUCTION TO LIGHT ABSORPTION: VISIBLE AND ULTRAVIOLET SPECTRA
27
ROBERT K. POOLE AND ULDIS KALNENIEKS
A = 0.25
A = 0.025
Wavelength (nm)
Figure 15 Attempts to record absorbance spectra in a very turbid cell suspension. A shows
the 'absolute' spectrum (a scan of a cuvette with the buffer taken as the baseline) of
reduced membranes of the bacterium Zymomonas mobilis, acquired with a single-beam
instrument (Beckman DU 650). B shows A4 acquired with a dual-wavelength instrument
(SDB-4) with 500 nm as the reference wavelength. C is the reduced minus 'as prepared'
difference spectrum obtained with the same dual-wavelength instrument. The concentration
of membrane protein was 10 mg ml-1; sodium dithionite was used as the reductant.
370 and 460 nm (Xmax420 nm), whilst hexachloroiridate solutions have complex
spectra in this region with Amax at 488 nm. Ammonium persulphate is
colourless. Addition of these oxidants to final concentrations of about 2 mm is
usually satisfactory, although it is common practice to add 'a few grains' to 1-4
ml samples. Further details of these procedures and references are given by
Jones and Poole (22).
28
INTRODUCTION TO LIGHT ABSORPTION: VISIBLE AND ULTRAVIOLET SPECTRA
Protocol 1
Recording the absolute UV-vis spectrum of a soluble
protein in solution
Equipment and reagents
• Protein in buffer, same buffer for • Split-beam scanning spectrophotometer
dilution. Impossible to be prescriptive (or almost any other kind!)
• Reagents for treatment of protein (e.g.
reductant for haem protein)
Method
1 Estimate absorption anticipated at A max , either from knowledge of protein con-
centration and anticipated absorbance coefficient, or from preliminary spectral
scan,
2 Decide on dilution required or adjustment of absorbance signals to be recorded by
choice of pathlength.
3 Establish and set wavelength limits. Select light source (e.g. UV and/or vis).
4 Choose cuvette material (e.g. silica/quartz for UV).
5 Select response time, slit width, scan speed. Use guidelines given in this chapter for
'normal' conditions, follow published protocols, or assess result of present scan
and adjust conditions accordingly (see figure 14).
6 Add buffer to two cuvettes, install in instrument, scan baseline, plot or store,
7 Keep buffer in reference cuvette, add protein to sample cuvette or replace contents
of sample cuvette with protein in buffer, install in instrument, scan, plot or store.
8 Compare baseline (step 6) and absolute spectrum of spectrum (step 7) visually or
subtract first from second.
Protocol 2
Recording the visible difference spectrum of a membrane-
bound chromophore in a turbid sample (cells or
membranes)
Equipment and reagents
• Cells or membranes in buffer, same buffer Split-beam or dual-wavelength scanning
for dilution. Impossible to be prescriptive spectrophotometer
• Reagents for treatment of protein (e.g.
reductant and oxidant for haem protein)
29
ROBERT K. POOLE AND ULDIS KALNENIEKS
Protocol 2 continued
Method
1 Estimate absorption anticipated at Xmax, either from knowledge of protein
concentration and anticipated absorbance coefficient, or from preliminary spectral
scan.
2 Decide on dilution required or adjustment of absorbance signals to be recorded by
choice of pathlength."
3 Establish and set wavelength limits. Select light source (i.e. vis).
4 Choose cuvette material (generally glass but cheap even scratched cuvettes may not
degrade signals).
5 Select response time, slit width, scan speed. Use guidelines given in this chapter for
'normal' conditions, follow published protocols, or assess result of present scan
and adjust conditions accordingly (see Figure 14).
6 Add identical, well mixed samples of cells or membranes to two cuvettes,b install in
instrument, scan baseline, plot or store.
7 Keep sample in reference cuvette, add reagent that will generate the required
difference (e.g. reductant) to sample cuvette, install in instrument, scan, plot or
store.
8 Compare baseline (step 6) and difference spectrum of spectrum (step 7) by
subtracting first from second.
" It is strongly recommended that the entire wavelength range is scanned at various sample
concentrations to avoid signal loss by light scattering at lower wavelengths (see Section 2.3).
b
In haem protein work, for example, this baseline may be that of the untreated or oxidized
sample.
portional to AA/AA. However, more often, derivative spectra are now obtained
by computing data sets obtained in a more conventional scanning mode. Plotting
derivatives up to the second is frequently a standard option in modern scanning
spectrophotometers.
Acknowledgements
RKP is grateful to the large family of research students, postdoctoral researchers
and visiting scientists to his laboratory who have put these techniques through
their paces in obtaining spectra of microbial haem proteins. Professor Britton
Chance has been a source of inspiration for 25 years and introduced us to the
Research Instrumentation Shop (RIS) at the University of Pennsylvania School of
Medicine. RIS staff Norman Graham and Bill Pennie and Current Designs
President Ben Dugan have provided us with machines and limitless help, in-
cluding the reading of a draft of this chapter. We thank Hewlett-Packard,
Molecular Devices and Shimadzu for information on their products. We are
very grateful to Mark Johnson for expertly producing many of the figures and
Ron Adams for Figure 7. Work in this laboratory was supported by BBSRC and
The Royal Society/NATO.
References
1. Brown, S. B. (1980). In An introduction to spectroscopy for biochemists (ed. S. B. Brown),
pp. 1-13. Academic Press, London.
2. Brown, S. B. (1980). In An introduction to spectroscopy for biochemists (ed. S. B. Brown),
pp. 14-69. Academic Press, London.
3. Bashford, C. L. (1987). In Spectrophotometry and spectrofluorimetry—A practical approach
(ed. D. A. Harris and C. L. Bashford), pp. 1-22. IRL Press, Oxford.
4. Fewson, C. A., Poole, R. K. and Thurston, C. F. (1984). Society for General Microbiology
Quarterly, 11, 87.
5. Koch, A. L and Ehrenfeld, E. (1968). Biochimico et Biophysica Acta, 165, 262-273.
6. Koch, A. L. (1970). Analytical Biochemistry, 38, 252.
7. Sambrook, J., Fritsch, E. F. and Maniatis, T. (1989). Molecular cloning, a laboratory
manual, 2nd edn. Cold Spring Harbor Laboratory Press, Cold Spring Harbor
Laboratory, NY.
8. Harris, D. A. (1987). In Spectrophotometry and spectrofluorimetry—A practical approach
(ed. D. A. Harris and C. L. Bashford), pp. 49-90. IRL Press, Oxford.
9. Jackson, J. B. and Crofts, A. R. (1969). FEBS letters, 4, 185
10. Wood, P. M. (1984). Biochimica et Biophysica Acta, 768, 293.
11. Chance, B., Legallais, V., Sorge, J. and Graham, N. (1975). Analytical Biochemistry, 66,
498.
31
ROBERT K. POOLE AND ULDIS KALNENIEKS
12. Poole, R. K., Rogers, N. J., D'Mello, R. A. M., Hughes, M. N. and Orii, Y. (1997).
Microbiology, 143,1557.
13. Kavanagh, E. P., Callis, J. B., Edwards, S. E., Poole, R. K. & Hill, S. (1998). Microbiology,
144, 2271.
14. Bashford, C. L. (1987). In Spectrophotometry and spectrofluorimetry—A practical approach
(ed. D. A. Harris and C. L. Bashford), pp. 115-135. IRL Press, Oxford.
15. Chen, S. S., Yoshihara, H., Harada, N., Seiyama, A., Watanabe, M., Kosaka, H.,
Kawano, S., Fusamoto, H., Kamada, T. and Shiga, T. (1993) American Journal of
Physiology, 264, G375.
16. Thibodeau, E. A. and D'Ambrosio, J. A. (1997) European Journal of Oral Science, 105, 373.
17. Brown, S., Colson, A. M., Meunier, B. and Rich, P. R. (1993). European Journal of
Biochemistry, 213, 137.
18. Poole, R. K. and Bashford, C. L. (1987). In Spectrophotometry and spectrofluorimetry—A
practical approach (ed. D. A. Harris and C. L Bashford), pp. 23-48. IRL Press, Oxford.
19. Wilson, D. F. (1967). Archives of Biochemistry and Biophysics, 121, 757.
20. Vincent, J.-C, Kumar, C. and Chance, B. (1982). Analytical Biochemistry, 126, 86.
21. Douzou, P. (1977) Cryobiochemistry. Academic Press, London.
22. Jones, C. W. and Poole, R. K. (1985). In Methods in Microbiology (ed. G. Gottschalk), 18,
285. Academic Press, London.
23. Chance, B. (1978). In Methods in Enzymology (ed. S. Fleischer and L. Packer), 54, 102.
Academic Press, New York.
24. Dutton, P. L. (1978). In Methods in Enzymology (ed. S. Fleischer and L. Packer), 54, 411.
Academic Press, New York.
25. Kalnenieks, U., Galinina, N., Bringer-Meyer, S. and Poole, R. K. (1998) FBMS
Microbiology Letters, 168, 91.
26. Butler, W. L. (1979). In Methods in Enzymology (ed. S. Fleischer and L. Packer), 56, 501.
Academic Press, New York.
32
Chapter 2
Fluorescence principles and
measurement
Arthur G. Szabo
School of Physical Sciences, University of Windsor, 401 Sunset Avenue,
Windsor, Canada N9B 3P4
1 Introduction
Fluorescence spectrometry is the most extensively used optical spectroscopic
method in analytical measurement and scientific investigation. During the past
five years more than 60000 scientific articles have been published in which
fluorescence spectroscopy has been used. The large number of applications
ranges from the analytical determination of trace metals in the environment to
pH measurements in whole cells under physiological conditions. In the scientific
research laboratory, fluorescence spectroscopy is being used or applied to study
the fundamental physical processes of molecules; structure-function relation-
ships and interactions of biomolecules such as proteins and nucleic acids; struc-
tures and activity within whole cells using such instrumentation as confocal
microscopy; and DNA sequencing in genomic characterization. In analytical
applications the use of fluorescence is dominant in clinical laboratories where
fluorescence immunoassays have largely replaced radioimmunoassay techniques.
There are two main reasons for this extensive use of fluorescence spectro-
scopy. Foremost is the high level of sensitivity and wide dynamic range that can
be achieved. There are a large number of laboratories that have reported single
molecule detection. Secondly, the instrumentation required is convenient and
for most purposes can be purchased at a modest cost. While improvements and
advances continue to be reported fluorescence instrumentation has reached a
high level of maturity.
2 Physical principles
2.1 The absorption process
A review of the physical principles of the fluorescence phenomenon permits
one to understand the origins of the information content that fluorescence
measurements can provide. A molecule absorbs electromagnetic radiation
through a quantum mechanical process where the molecule is transformed
from a 'ground' state to an 'excited' state. The energy of the absorbed photon of
33
ARTHUR G. SZABO
light corresponds to the energy difference between these two states. In the case
of light in the ultraviolet and visible spectral range of 200 nm to 800 nm that
corresponds to energies of 143 to 35.8 kcal mol-1. The absorption of light results
in an electronic transition in the atom or molecule. In atoms this involves the
promotion of an electron from an outer shell orbital to an empty orbital of
higher energy. In molecules, an electron is promoted from the highest occupied
molecular orbital (HOMO) to the lowest unoccupied molecular orbital (LUMO),
This process is governed by a number of selection rules, and the transition
probability for the absorption of a photon is reflected in the molecule's molar
absorptivity, e (M-1 cm -1 ), that is determined from the Beer-Lambert law.
Figure 1(a) provides an example of the absorption process in the case of a
carbon-carbon double bond. On interaction with a photon of appropriate
U) (b)
Figure 1 (a) A TT-TT* transition of a C-C double bond, with the lower energy TT state showing
the IT bonding orbital filled with two electrons with antiparallel spin and the unfilled TT
antibonding orbital (shaded), and the higher energy IT* state with the electron spins still
antiparallel but with one electron in the IT bonding orbital and one in the IT antibonding
orbital, {b) A TT-O-* transition of a C-C double bond with the same lower energy electron
configuration as in figure l(a), and the higher energy a* state with one electron in the IT
bonding orbital and one electron in the IT antibonding orbital (shaded).
FLUORESCENCE PRINCIPLES AND MEASUREMENT
energy an electron from the IT bonding orbital of the ground electronic state is
promoted to a IT* antibonding orbital (LUMO). The energy difference between
these two states corresponds to the absorbed photon energy, AE = hc/A, where h
is Planck's constant, c is the velocity of light in a vacuum and A is the
wavelength of the light. This excitation process is termed aIT-IT*transition and
the excited state is referred to as a IT-IT* state. There are other antibonding
orbitals of higher energy, and in this example such an orbital is a O*
antibonding orbital. The excitation of an electron from the IT bonding orbital to
the o* antibonding orbital would occur when the photon energy corresponds to
the energy difference between the ground state and the TT-o* excited state
(Figure 1(b)).
One of the selection rules that governs the transition probability is that the
spin of the electrons must be retained, i.e. the electron spins in the ground state
IT bonding orbital are paired and antiparallel and they retain their antiparallel
spin in the transition to the excited IT-IT* state. The multiplicity, M, of these
states describes the spin states of the electrons. Electrons are designated to have
a spin quantum number, S, of either +1/2 or -1/2 and so the multiplicity of the
state can be calculated according to equation V.
M=2|SS| + 1 1
In the ground state ES = 0, and so M = 1 and this state is designated as a ground
singlet state, S0. In the IT-IT* excited state the original electron spins are con-
served and M = 1. This excited state is then designated as a singlet state. In the
example in Figure 1 the TT-TT* state is the first excited singlet state, S1, and the
TT-O* state is the second excited singlet state, S2.
35
ARTHUR G.
Figure 2 A modified Jablonski diagram showing the different transitions and processes
between the excited states. S0(0), S±(0) and S2(0) being the 0 level vibrational energy level
of ground state, first singlet state and second singlet state, respectively. 71(0) is the lowest
vibrational level of the first triplet state and P represents the energy level of a stable
photoproduct which may be formed. The transitions with the arrows pointing towards the top
of the figure represent absorption transitions and they are described in the text. The dashed
arrows pointing towards the bottom of the figure represent competing pathways by which the
excited molecule deactivates. I represents internal conversion processes; ISC, intersystem
crossing; NR, non-radiative deactivation; PR, photoproduct formation; hv the radiative or
fluorescence process.
Wavelength (nm)
Figure 3 An absorption spectrum representing the three absorption transitions designated
1, 2 and 3, in Figure 2.
36
FLUORESCENCE PRINCIPLES AND MEASUREMENT
and the several vibrational levels of the excited singlet states. Figure 3 is a
representation of a typical absorption spectrum, where absorption at A1 corres-
ponds to the absorption transition marked 1 (figure 2). Another wavelength, X2,
would correspond to transition 2 (Figure 2) to an upper vibrational level of S1,
and A.3 corresponds to absorption transition 3 (Figure 2) to S2.
The light absorption process occurs very rapidly (10-15 s) to form the excited
state, so fast that there is no change in the nuclear coordinates of the molecule
during the absorption process (Frank Condon transition). If the absorption is
to S2, the molecule undergoes a very rapid process (10-14 s) known as internal
conversion from the lowest vibrational level of S2 to an upper vibrational level
of Sj. In solution the molecule rapidly loses the excess vibrational energy
through collision with solvent molecules. The net result is that within 10-13 to
10-14 s of excitation the molecule is in the lowest (0th) vibrational energy level of
s s
1 ( 1(0))- The absorption process has very rapidly created a metastable state that
can undergo a number of competing processes that were not available to the
ground state.
37
ARTHUR G. SZABO
2.2.6 Phosphorescence
Phosphorescence is also a radiative process. It describes the radiative process
originating from T1 to S0. The required spin inversion of one of the electrons
when it moves from the LUMO to the half filled HOMO is a forbidden process.
When observed it occurs on a timescale much longer than the fluorescence
process. In most cases it can only be observed in frozen glass solutions at low
temperature (77 K).
38
FLUORESCENCE PRINCIPLES AND MEASUREMENT
3 Fluorescence parameters
3.1 Fluorescence spectrum
The light emitted by a molecule from Sj(0) corresponds to the energy difference
between St(0) and different vibrational levels of S0, i.e. S0(n). There will be a
spectrum of energies of the fluorescence photons and this energy spectrum
(Figure 4) is what is measured on a fluorescence spectrometer. The wavelength,
X0 corresponds to the radiative transition from S1(O) to S0(0). This is the highest
energy radiative transition and is the energy difference between these two
states. AtXMAX,the wavelength corresponding to the highest fluorescence in-
tensity, the transition from S1(O) to S0(n) represents the most probable radiative
transition between these two electronic states.
KISC
Wavelength (nm)
Figure 4 A typical fluorescence spectrum representing the energy distribution of the photons
from S1(0) to different vibrational levels of the ground state, S0.
39
ARTHUR G. SZABO
molecule. If I0 is the intensity of exciting light in photons s-1 then the integrated
intensity of fluorescence, IF (photons s1) is:
IF (total number of photons absorbed / s, IA). ( o F ) 5
From the Beer-Lambert law:
6
where I is the transmitted light intensity in photons s-1.
7
and
8
Thus the integrated fluorescence signal depends on the incident light intensity,
the absorbance of the sample at the wavelength of excitation, EX, and oF.
where M(SJ(t) is the concentration M(Sl) at any time t; M(S1)(0) is the con-
centration of MfS at time t = 0; and Sk = kR + kNR + kisc + kpR- Since IF x M(Sl)
then one can write the expression:
showing that the singlet excited state has a finite lifetime that is dependent on
the efficiency of the several competing deactivating processes. The radiative
lifetime of S1 is defined according to:
Thus if the only deactivation process of S1 was fluorescence then TS = TR, and
oF = 1. An important relationship is:
When one measures TS and 4F then the intrinsic molecular rate constant,
kR = TR, can be calculated.
40
FLUORESCENCE PRINCIPLES AND MEASUREMENT
4 Fluorescence spectrometers
The above section has outlined the physical parameters that describe the fluor-
escence process. One can measure the fluorescence spectrum, F(h),the singlet
excited state lifetime, and determine the .These parameters can be inter-
preted in terms of the structure, environment and dynamics of the molecule of
interest. In this section, the different optical and electronic components com-
prising an instrument that can measure F(h) and QF will be described. This
instrument is generally known as a steady state fluorescence spectrometer, since
it integrates the fluorescence intensity over a given time period. Time-resolved
fluorescence instrumentation that is used to measure the excited singlet state
decay times is described in Chapter 3.
Figure 5 is a block diagram of the essential components of a fluorescence
spectrometer. All fluorescence spectrometers comprise a light source, an excita-
tion wavelength selector (the wavelength of excitation is determined from the
molecule’sabsorption spectrum), a sample holder, a second wavelength selector
to isolate the fluorescence wavelength, and a photodetector that converts any
incident fluorescence photons into an electronic signal. Another important
41
ARTHUR G. SZABO
Wavelength (nm)
Figure 6 A scan of the spectral output of a 450 W xenon arc lamp measured by placing an
optically dense sample of rhodamine B in the sample holder. The fluorescence passed
through a 610 nm cut-off filter, and EmM was set at 620 nm. Excitation bandpass, 2 nm;
emission bandpass 2 nm.
42
FLUORESCENCE PRINCIPLES AND MEASUREMENT
Wavelength (nm)
Figure 7 A trace of the monochromator spectral bandpass, with an emission slit width of
2 mm, and a reciprocal dispersion, Cof 4 nm mm-1.
that can be used to isolate a particular line of a light source, such as a mercury
arc, a portion of the continuum spectrum of a xenon arc, or a portion of the
sample's fluorescence spectrum. Besides selecting for the maximum trans-
mission wavelength of a filter, one should also consider the half bandwidth that
defines the range of wavelengths that the filter will transmit. In addition one
must be cognisant that the intensity of transmitted light through a filter will be
reduced as the bandwidth is reduced.
The excitation wavelength, AEX, or fluorescence wavelength, AEM, of light is
often selected through a monochromator. The dispersion element in a mono-
chromator may be either a grating or a quartz prism. A grating monochromator
is preferred since the reciprocal dispersion, D-1, is constant over all A, while for
a prism this parameter varies with X and is cumbersome to account for. Hence a
quartz monochromator is rarely used in most standard laboratory instruments.
D-1 is the parameter that relates the effective bandwidth of A passing through a
monochromator to the slit widths (mm) of the monochromator and has units of
nm mm-1 (Figure 7). The value of D-1 depends on the quality of the mono-
chromator, but typically it ranges from 2 nm mm-1 to 8 nm mm-1.
One further point to note is that a grating monochromator disperses light in
a series of orders. Thus a monochromator setting at 500 nm will also pass light
whose wavelength is 250 nm since the second order of dispersion of light at A1
will occur at 2A1.Placing an appropriate cut-off filter after a monochromator,
i.e. a Pyrex filter in this example, can efficiently remove this second-order
dispersion.
43
ARTHUR G. SZABO
44
FLUORESCENCE PRINCIPLES AND MEASUREMENT
Protocol 1
Preparation of reference solution
Equipment and supplies
• High purity rhodamine B Pasteur pipette
• Ethanol 1-cm square quartz stoppered cuvette or 1-
• Ethylene glycol, spectral grade cm triangular quartz stoppered cuvette
• Methanol, spectral grade Sonication bath
• 25-ml scintillation vial with a screw cap Optical tissues
• Polyfilm Latex gloves
A. Solution preparation
1 The rhodamine dyes are considered to be toxic. It is important to wear gloves while
handling the dyes and to avoid all spills. Weigh 15 mg of rhodamine B into the
scintillation vial.
2 Add 0.5 ml of ethanol to aid the solubility of the rhodamine. Add 4.5 ml of ethylene
glycol and cap the vial. Shake the solution and then hold it in a warm sonication
bath until there is no evidence of any particulate matter in the sample. The
sonicator assists in the solution of the rhodamine.
3 Totally fill the cuvette with the rhodamine solution using a Pasteur pipette.
Stopper the cuvette, and wrap the stopper and top of the cuvette with a small piece
of polyfilm. Wipe the outside of the cuvette with a tissue wet with methanol and
then a dry tissue,
4 Place the cuvette into the reference cell holder of the instrument.
5 The solution should be replaced approximately every six months of steady use.
B. Cleaning the optical elements of the reference channel
1 It is a good idea to clean the optical elements of the reference channel. The beam
splitter and red cut-off filter should be carefully wiped with a tissue wet with
methanol followed by wiping with a dry tissue.
and so the cuvette should be contained in a cell holder that can be thermo-
stattcd. The fluorescence photons are then collected with a lens and focused
onto the fluorescence wavelength selector.
A sample holder that can hold four cuvettes in a rotating turret is an advantage.
In this way other samples and blank solutions can be maintained at the same
temperature and one can simply rotate between the four cuvettes.
45
ARTHUR G. SZABO
cut-off filter is used, then light of all X above the cut-off X will be detected. This
will permit the measurement of very low fluorescence light intensities and
correspondingly low concentrations of the analyte. In standard laboratory
fluorimeters where a large variety of different fluorescence measurements will
be performed, a grating monochromator (EmM, Figure 5) is used to select AEM.
The fluorescence from the sample is focused onto the entrance slit of EmM and
after passing through the exit slit falls onto the photocathode of a photo-
multiplier tube (PMT).
46
FLUORESCENCE PRINCIPLES AND MEASUREMENT
5 Fluorescence spectra
5.1 Inner filter effect
After the sample's absorption spectrum has been measured, one can select an ex-
citation wavelength, AEX, that falls within the absorption envelope. There will be
other factors that may influence the selection of AEX. If the sample is homogene-
ous and contains a single molecular chromophore, then any wavelength may be
appropriate. In the case of samples that contain more than a single molecular
chromophore, such as in a protein (vide infra), then the choice of XEX will depend
on whether preferential excitation of one of those chromophores is desired.
An important parameter of the sample that must be considered is the
absorbance of the sample at AEX. Equation 8 shows that the rate of fluorescence
emission, IF, is determined by the rate of light absorbance. In dilute solution
equation 8 can be expanded in terms of a series and where ecl is small then it can
be approximated that:
IF = I0(ecl) oF 15
This approximation can lead to an error known as the inner filter effect. The
solution at the front of the cell is exposed to a higher excitation light intensity
47
ARTHUR G. SZABO
0.01 1.1
0.05 5.5
0.10 10.6
0.20
than the sample near the opposite side, because of the absorbance of the inter-
vening solution. This will introduce an error and non-linearity in the relation
between IF and A. The terms in the series expansion of equation 8 that have been
neglected in equation 15 can be evaluated and the percentage error in the
fluorescence signal owing to the inner filter effect can be calculated (Table 2).
One usually adjusts the concentration of the sample so that A at AEX is <0.1 in
order to minimize inner filter errors. If it is necessary to perform an experiment
at a certain concentration of the sample and this results in a very high absorb-
ance at the absorption maximum, XAMAX. then a XEX can be selected on the low
energy side of the absorption spectrum where A < 0.1.
48
FLUORESCENCE PRINCIPLES AND MEASUREMENT
Wavelength (nm)
Figure 8 A demonstration of the overlap of excitation and emission monochromator
bandpass whenAEXand AEM are close to one another thus explaining the observation of
Rayleigh scattering of AEX.
For example with AEX = 290 nm a Raman scatter peak will be observed at ARA =
321 nm. Figure 9, includes a spectrum of an aqueous blank that shows this
Raman band and the effect of the Raman signal is seen on the sample fluor-
escence spectrum. In the latter case it appears as a weak shoulder on the high
energy side of the fluorescence wavelength maximum, XFMAX- When the sample
fluorescence is intense the contribution of the Raman band will be negligible. A
simple test to determine if a spurious peak is due to a Raman scattering signal is
49
ARTHUR G. SZABO
Wavelength (nm)
Figure 9 The dashed curve is the fluorescence spectrum of a protein sample withAEX= 290
nm. The solid curve with circles is the spectrum of the buffer blank, showing a maximum
signal at 321 nm, corresponding to the Raman scatter of the water. The solid curve is the
signal resulting from subtracting the buffer blank from the raw spectral data.
to shift XEX by 10 nm to a lower wavelength. The Raman band will shift in the
same direction.
50
FLUORESCENCE PRINCIPLES AND MEASUREMENT
600 too no
Wavelength (nm)
Figure 10 The spectral sensitivity curve of a Hamamatsu R955 photomultiplier tube.
51
ARTHUR G. SZABO
where FRB(XEX. 620 nm) is the recorded spectrum of rhodamine B with EmM
constant at 620 nm and AEX is varied, and FMM (AEX, AEM) is the recorded spectrum
using a diffuser plate of MgO and both ExM and EmM are varied synchronously.
If FM(X) represents the measured uncorrected sample fluorescence spectrum
then a corrected sample fluorescence spectrum can be obtained by a multi-
plication of C(X) and FM(X) according to
52
FLUORESCENCE PRINCIPLES AND MEASUREMENT
Protocol 2
Generation of spectral correction factors (250-600 nm)
Equipment and supplies
• A quartz diffuser plate mounted in a • A triangular quartz cuvette filled with a
holder so that the plate is oriented at an concentrated ethylene glycol solution of
angle of 45° to the excitation light rhodamine B (see Protocol 1)
propagation direction or a metal plate • Quartz neutral density filters
coated with an MgO film. This plate
• Fine wire mesh screens
should be mounted in a holder so that it
makes an angle of 35° to the excitation • 610-njn glass cut-off filter
light propagation direction.
Protocol 2 continued
D. Rhodamlne sample
1 Replace the diffuser plate by the triangular cuvette of rhodamine. The incident face
should be opposite to the entrance slit of EmM. In this way the only light incident
on EmM will be rhodamine B fluorescence. Reflected light is directed in the
opposite direction.
E. Instrument settings for FT(A)
1 Remove the ND filters and any wire mesh screens and replace them with the 610-
nm cut-off filter.
2 Set ExM to 250 nm and EmM to 620 nm.
3 The wavelength scan range and scan rate should be the same as for the previous
spectrum.
4 Set the instrument so that only ExM will scan,
5 Record the spectral light signal between AEX 250-600 nm. This spectrum is
considered to be the true lamp spectrum, FT(A) = FRB(AEX. 620 nm). Since EmM and
the PMT only 'saw' a single wavelength (620 nm) any A bias was removed from the
emission detector.
F. Data handling and C(A)
1 Using the computer software divide ST(A) by SL(A).
2 Normalize the correction factor curve that is obtained so that the value at 400 nm
is 1. Most software packages have a normalization command. The curve thus
obtained (equation 17) can be used to correct all fluorescence spectra between 250
and 600 nm.
3 A fluorescence spectrum is corrected according to equation 18.
54
FLUORESCENCE PRINCIPLES AND MEASUREMENT
Figure 11 The excitation spectrum of a protein sample with the AEM = 340 nm.
55
ARTHUR G. SZABO
Wavelength (nm)
(a)
Wavelength (nm)
(b)
Figure 12 (a) A fluorescence spectrum showing two spectral maxima. Xr is the fluorescence
wavelength setting selected to determine the excitation spectrum responsible for the low
energy fluorescence maximum; Xb is the fluorescence wavelength setting selected to
determine the excitation spectrum responsible for the high energy fluorescence maximum,
(b) The solid line is the excitation spectrum when EmM was set at Xr; the dashed line is the
excitation spectrum when EmM was set at \b for the sample whose fluorescence is shown
in (a).
56
FLUORESCENCE PRINCIPLES AND MEASUREMENT
where SM(X) and SS(X) are the integrated areas of the corrected fluorescence
spectra of the sample, M, and the standard, S, respectively; and oF (M) and oF (S)
are their respective quantum yields; and AM and As are their absorbancies at AEX-
Thus
f--oF(S) 22
AM
The determination of oF of a molecule by a primary method is a very difficult
task and has been the subject of several reviews (4, 5). The quantum yields of
several molecules under rigorous experimental conditions have been deter-
mined (6, 7). A selection that is generally useful is listed in Table 3. When such a
standard molecule is chosen in order to measure oF of the molecule of interest,
the exact sample conditions for the standard must be followed. Temperature
control is also critical owing to the temperature dependency of kNR. One of the
most often used standards is quinine sulphate in 1 N H2S04 at concentrations
<10-4 M. At 25°C its oF is 0.546 (AEX = 365 nm). For protein fluorescence, where
tryptophan is the fluorescent chromophore, N-acetyltryptophanamide (NATA) at
pH 7, 20 °C is used as a secondary standard. The accepted value of oF of NATA is
0.14 (8).
When the oF of the sample is determined, F(A) of the sample and the stand-
ard should cover a similar X range. Obviously, both the sample and standard
57
ARTHUR G. SZABO
should absorb at the same XEX- Since it is important to minimize inner filter
effects the absorbance of both the sample and the standard should be <0.05 at
XEX. This becomes one of the largest sources of error in measuring oF since
there is a large percentage error in measuring such low absorbancies. A means
of reducing this error is to measure the absorbance of the sample and standard
in a 5 cm cylindrical cuvette whose outside diameter is 7 mm. The absorbances
of these solutions (c. 2 ml volume) are measured in this cuvette, and then the
solutions are transferred to a 1-cm fluorescence cuvette for the fluorescence
measurements. In this way an absorbance that is five times that in a 1 cm
cuvette is measured and the error in the absorbance is. significantly reduced.
The fluorescence spectra of both the sample and standard should be scanned
over an identical scan range, and it is important that FS(X), the fluorescence
spectrum of the standard, should have reached a value close to zero on its red
edge. The solvent blank is subtracted from these spectra. Since it is unlikely that
the sample and the standard will have identical F(A), the spectra should be cor-
rected for the X sensitivity of the PMT. The resulting corrected spectra, F(A)coRR
are then integrated across the same spectral range, i.e. from Xr to AF. This
integration can usually be conducted in the instrument computer. Alternately,
the integration can be accomplished using Simpson's rule or a trapezoidal
approximation over narrow X ranges (1-5 nm). Prior to the availability of com-
puter interfaced instruments, the integration was accomplished by cutting out
the spectral recording on a chart paper and weighing the paper cut-out.
When one is measuring oF of a fluorophore in an aqueous solution and the
standard is in a non-aqueous solvent then a refractive index correction is
recommended. This correction is applied according to:
23
n2(standard)
where oF' is the measured quantum yield without correction for refractive
index difference, oF is the corrected quantum yield value, n2(sample) is the
square of the refractive index of the solvent of the sample solution, and
n2(standard) is the square of the refractive index of the solvent of the standard
solution. If the sample had been dissolved in benzene and the standard was
quinine sulphate the correction would have been 27%.
6 Fluorescence applications
The application of fluorescence measurements to a large variety of chemical,
biochemical, biological and physical problems is extensive. The majority of
fluorescence applications involve the use of extrinsic fluorescence probes. These
are chromophoric molecules that are attached to or adsorbed onto another
molecule and their fluorescence is measured. These molecules probe the prop-
erties of the substances to which they are attached. The catalogue and manual
published by Molecular Probes (9) is an excellent collection of a large number of
different examples of the use of fluorescence, especially as they relate to bio-
58
FLUORESCENCE PRINCIPLES AND MEASUREMENT
Protocol 3
Measurement of OF of a sample
Equipment and reagents
• Purified sample • Four 1-cm quartz cuvettes
• Purified standard • A 5-cm quartz cylindrical cuvette
• Purified solvents (outside diameter = 0.7 cm)
A. Selection of standard
1 A standard material is selected from a list of quantum yield standards (6, 7 and
Table 3). The standard should have an absorption spectrum in the same range as
that of the sample. The fluorescence spectra of the standard and sample should
span a similar A range.
59
ARTHUR G. SZABO
Protocol continue
60
FLUORESCENCE PRINCIPLES AND MEASUREMENT
61
ARTHUR G. SZABO
around the Ca-CB bond (10,11). Recently it has been shown how Tip analogues
when substituted into proteins in place of the normal Trp residue can provide
useful information on molecular details of interacting segments of protein-
protein and protein-nucleic acid complexes (12). These analogues have extended
ultraviolet absorption beyond the normal Trp absorbance and hence can be
selectively excited in the presence of a large number of normal Trp residues.
It is beyond the scope of this chapter to review this literature. The reader is
referred to several review articles that provide a useful summary (13-15).
It is useful, however, to discuss a few practical aspects of measuring Trp
fluorescence and some of the information that can be obtained. One of the first
factors that it is necessary to consider is the number of Tyr residues in a protein
that contains a single Trp residue. In most cases where the number of Tyr
residues is small, XEX = 295 nm will predominantly excite the single Trp residue
and only Trp fluorescence will be observed. This is because Tyr absorbance at
295 nm is very small (E295 = 18 M-1 cm-1) but not zero. When there are a large
number of Tyr residues in a protein, the sum of the absorbance of these residues
makes a modest contribution to the total absorbance and in these cases the
fluorescence generated with AEX = 295 nm will contain Tyr spectral components
(16). Also FRET from Tyr to Trp may occur and affect the integrated fluorescence
intensity. In order to virtually eliminate the Tyr contribution to the fluores-
cence, A EX can be changed to 300 nm. However, this generates another problem,
especially when one measures oF, since at typical protein sample concen-
trations of 1-5 x 10-6 M the absorbance at 300 nm is very low.
There is also another cautionary note to consider when the oF of proteins is
being measured. In order to selectively excite Trp fluorescence AEX is usually
chosen to be 295 nm or greater in order to minimize the Tyr absorbance. An
assumption that is usually made is that the molar absorptivity at 295 nm of the
particular protein under the different conditions of the experiments is the
same. For example if the thermal unfolding or chemical denaturation of a
protein is being studied using fluorescence hardly any attention is given to the
fact that E295 is in fact changing. This A. is on the very steep falling slope of the
protein absorption spectrum. Changes in E295 are not readily apparent on casual
inspection of the absorption spectrum. As a Trp residue that may originally
have been buried in the interior of a protein becomes solvent exposed, E295 will
change. When performing chemical denaturation experiments this can be
accounted for. However, corrections for these changes in thermal unfolding are
difficult to estimate.
62
FLUORESCENCE PRINCIPLES AND MEASUREMENT
provided they are within 6-10 A of the Trp residue. This quenching process is
thought to be due to an electronic exchange process.
Histidine may quench Trp fluorescence through a proton transfer mech-
anism. There is evidence (17) that position 4 of the indole ring acquires more
electron density in the excited singlet state and can accept a proton from a
proton donor such as histidine.
The neutral amide function of glutamine or asparagine, as well as the amide
backbone of the protein, can quench Trp fluorescence. This is another short-
range electron transfer process. The short lifetime of one of the Trp fluor-
escence decay components may be due to the close proximity of the indole ring
to the amide backbone in that particular rotainer state.
When the Trp residue is located in a hydrophobic environment of the
protein its oF is enhanced and importantly the fluorescence spectral maximum
is shifted to a much higher energy than that normally seen for a solvent
exposed residue.
and the second-order rate constant, kq, is the kinetic term that is a measure of
the process. In the absence of external quencher, Q,
where
and is the solute quantum yield in the absence of added quencher. In the
presence of quencher
63
ARTHUR G. SZABO
0.3
[Quencher] M
Figure 13 A set of typical Stern Volmer fluorescence quenching plots. •, a linear Stern
Volmer plot; A, a plot showing upward curvature owing to a high concentration of quencher
surrounding the fluorophore; •. a plot showing downward curvature owing to the presence of
two fluorescent chromophores being quenched with different efficiencies; *, a linear Stern
Volmer plot representing the case where the chromophore is less accessible to the quencher.
and since the unquenched singlet lifetime, TS = l/(kR + ko), equation 29 can be
simplified to
•= 1+ 30
This is the well known Stern Volmer relationship that relates the fluorescence
yield of the sample to the concentration of the added quencher. A plot of o 0 F /o F
or, since oF F, F°/F versus [Q] should yield a straight line with an intercept of 1,
and the Stern Volmer slope, KSV = kQTs (Figure 13).
Quenching by the addition of an external quencher is a diffusion controlled
process. As a first approximation the value of Ksv reveals the degree of exposure
to the solvent of a Trp residue. A high value suggests a solvent exposed residue,
while a low value indicates that the residue is buried inside the protein. How-
ever, this is not strictly correct since this assumes that TS is the same for both
cases. A more accurate estimation of the degree of exposure of Trp is to calculate
kQfrom Ksv = kQTS. In this case the value of TS should be the intensity weighted
decay tune when multi-exponential decay kinetics are observed (that is often the
case even for a single Trp protein). The intensity weighted decay time is
where 04 and TJ are the pre-exponential term and decay time of the ith fluor-
escence decay component, respectively. It is important to use the intensity
64
FLUORESCENCE PRINCIPLES AND MEASUREMENT
Protocol 4
Acrylamide quenching of protein fluorescence
Equipment and supplies
• High purity spectral grade actylamide • Volumetric pipettes
• High purity buffer for solution of the • 1-cm quartzfluorescencecuvettes
protein
A. Solution preparation
1 A stock solution of 5 M acrylamide is prepared in the sample buffer.
2 2.00 ml of a protein sample is prepared to give an absorbance at AEX- 0,05.
B. Fluorescence and tftratlon measurement
1 Record the fluorescence spectrum of the protein and of the buffer blank.
2 Add aliquots of acrylamide to the protein sample and measure the fluorescence
spectrum of the sample after each addition of acrylamide. The first addition should
be 5 fil of stock acrylamide, to give [acrylamide] = 0.0125 M. A second 5-ul aliquot is
65
ARTHUR G. SZABO
Protocol 4
7 Conclusion
The description above is a summary of the important practical aspects of steady
state fluorescence spectroscopic measurements. Further details and applications
can be obtained from a number of reviews and publications, a selection of
which are referenced below (18-21).
References
1. Teale,J. (1969) Photochem. Photobiol. 10: 363-374.
2. Lenz, B. R., Moore, B, M., and Barrow, D. A. (1979) Biophys. J. 25: 489-494.
3. Melhuish, W. H. (1972) J. Res. Nat. Bur. Stand. 76A: 547.
4. Demas.J. N, and Crosby. C. A. (1971) J. Phys. Chem. 75: 991
66
FLUORESCENCE PRINCIPLES AND MEASUREMENT
67
This page intentionally left blank
Chapter 3
Time-resolved fluorescence
and phosphorescence
spectroscopy
Thomas D. Bradrick* and Jorge E. Churchicht
* Optical Spectroscopy Section/LBC/NHLBI, Bldg. 10, Rm 5D14,
MSC1412, National Institutes of Health, Bethesda, MD 20892-1412, USA
T
Dept. of Applied Biology and Chemical Technology, Hong Kong
Polytechnic University, Hong Kong, PRC
1 Introduction
In this chapter, we outline the basic theory and methodology for making time-
resolved fluorescence and phosphorescence measurements. We begin with a
brief discussion of the intrinsic time dependence of fluorescence and phosphor-
escence decays, and also introduce several important photophysical concepts.
Energy transfer measurements, which are important for determining molecular
distances, are then addressed, followed by the convolution integral (which
describes the luminescence decay that is actually observed in pulsed excitation
experiments) and the relationships that are used to determine fluorescence
lifetimes from phase/modulation data. Polarized fluorescence measurements,
which are an important tool for following molecular motions, are then
discussed, and the fluorescence portion of the chapter concludes with an
overview of data analysis and brief descriptions of the instrumentation that is
used in making time-resolved fluorescence measurements. The remainder of
the chapter is then devoted to a discussion of phosphorescence spectroscopy.
2 Background
2.1 Basic photophysics and time dependence of
fluorescence and phosphorescence decays
The intrinsic time dependence of fluorescence decays is derived in Section 3.3 of
Chapter 2. There it is shown that if a population of excited singlet-state
molecules is generated instantaneously, its size decreases exponentially with
time, as does the intensity of the emitted photons. The fluorescence lifetime TS
is typically used to describe the rate of decay, where Ts-1 = kR + kNR + kISC + kPR
(equation 12, Chapter 2) and kR, kNR, kISC and kpR are the rate constants for the
69
THOMAS D. BRADRICK AND JORGE E. CHURCHICH
where the sum is over all the different fluorophore species 'i' in solution, and
we now express as ai their individual pre-exponential factors. The af depend on
the species' respective kR values, concentrations and extinction coefficients, as
well as on the exciting light intensity and geometric factors (among other
things). In carrying out fluorescence lifetime measurements, some of these
variables are typically not quantified; as their product is thus not known before
hand, it is simply expressed as a constant that is determined when the data are
analysed (see Section 2.6).
The steady-state fluorescence intensity I(A) at emission wavelength X is
related to the time-resolved decay parameters through
70
TIME-RESOLVED FLUORESCENCE AND PHOSPHORESCENCE SPECTROSCOPY
The ratio of the quantum yields for phosphorescence and fluorescence is then
(i.e. op = TSkISC = oISC). The intersystem crossing rate constant (kISC) is sensitive
to spin orbit coupling by heavy atoms (2).
71
THOMAS D. BRADRICK AND JORGE E. CHURCHICH
and eA are plotted versus wavenumber (v), which is the reciprocal of the wave-
length (A., in cm).Rovalues for a number of donor-acceptor pairs free in solution
(for which K2 is 2/3) may be found in (7); otherwise, they must be calculated
using equations 8 and 9.
The efficiency of energy transfer is given by
where oD and oDA are the donor's quantum yields and TD and TDA are its decay
times in the absence and presence of the acceptor, respectively. (The inter-
changeability of quantum yield and lifetime follows from equation 14 of Chapter
2). Because of the transfer efficiency's sensitivity to the donor-acceptor distance R
where oPD and oPD are the phosphorescence quantum yields of the donor in the
absence and presence of the acceptor, respectively, and oPA is the phos-
phorescence quantum yield of the acceptor. When the forbidden nature of a
transition in the donor results in a corresponding increase in the lifetime of its
excited state (as is the case for the triplet state of many molecules), energy ex-
change over long distances can be high (provided there is a good overlap of the
spectra concerned) as the distance over which the triplet-state molecule can
diffuse is relatively great.
72
TIME-RESOLVED FLUORESCENCE AND PHOSPHORESCENCE SPECTROSCOPY
Here, L(t1) is the height and At' the width of the excitation pulse's 8-function
component at time t1. F(t) (the intrinsic fluorescence decay) is given by equation 1
and its argument t - t1 accounts for the fact that the elapsed time varies
between when the different subpopulations are created (at times t1) and when
the fluorescence intensity is measured. The (relative) 'height' of L at t1( appro-
priately 'scales' the magnitude of the resulting fluorescence response. The effect
of convolving an exponential decay with an excitation pulse of finite width will
be seen in Section 2.6.4, where we describe the analysis of pulsed-excitation
data. A fuller discussion of the convolution integral may be found in (9).3,4
3
Photomultiplier tubes, which are commonly used to convert light into an electronic signal for
further processing (see Section 3.2), typically distort both the measured excitation pulse profile
and observed fluorescence decay by broadening them. For that reason, the measured L(t') is often
called the instrument response function as it is actually a convolution of the excitation pulse
and the photomultiplier response. It is, in fact, the 'spread' in the signal caused by the photo-
multiplier, rather than that due to a laser pulse, that most distorts the observed fluorescence
decay away from being exponential and requires we use the convolution integral description.
4
On the much longer timescale over which phosphorescence is emitted, nanosecond pulses that
populate S1 'look' like 8-function pulses. In that case, the triplet state population is still formed
virtually instantaneously and the phosphorescence decay remains undistorted and decays expo-
nentially. Usually the phosphorescence decay profile does not have to be described by a convo-
lution integral unless one is using a xenon flashlamp for excitation, for which the pulses are a
few tens of microseconds wide, with a weak 100 p,s tail.
73
THOMAS D. BRADRICK AND JORGE E. CHURCHICH
where
and the sums are over the number of exponential terms 'i' that are being used
to model the fluorescence decay. fi is the fractional intensity of the ith decay
component (equation 3).
Thus, by measuring the phase shift 0 and the demodulation m, one can deter-
mine the fluorescence decay times (T) and amplitudes (a). Such measurements
are typically made sequentially over as broad a frequency range as possible in
order to minimize the error in the extracted parameters, and to enable one
to distinguish between different multi-exponential fluorescence decays (see
Figure 1 Illustration of exciting light intensity L(f) and fluorescence response Fobs(t) for a
phase/modulation experiment. B and b are the respective amplitudes of oscillation of the
two waves, while A and a are their respective average intensities. 6 is the angle by which
Fobs(t) is delayed or phase shifted with respect to L(t').
74
TIME-RESOLVED FLUORESCENCE AND PHOSPHORESCENCE SPECTROSCOPY
Section 2.6.5). (Note: MHz modulation frequencies are required for measuring
nanosecond fluorescence lifetimes.) The reader who desires additional details
about phase/modulation fluorometry may wish to consult (10).
75
THOMAS D. BRADRICK AND JORGE E. CHURCHICH
Because the molecule's motion is restricted, the anisotropy does not decay to
zero but reaches some final constant value, r00. Equation 19 is typically used to
model the anisotropy decay of a rod-shaped fluorophore such as l-(4-
trimethylammoniumphenyl)-6-phenyl-l,3,5-hexatriene (TMA-DPH) embedded in
a phospholipid membrane, where it reports on the order and dynamics of the
latter. The so-called 'order parameter' S of the membrane is defined as
where 9 is the half angle of the cone. A more general model for restricted
motion is
76
TIME-RESOLVED FLUORESCENCE AND PHOSPHORESCENCE SPECTROSCOPY
2.5.4 Examples
A study of the effects of melittin on phospholipid membranes that included
time-resolved fluorescence anisotropy measurements of l,6-diphenyl-l,3,5-
hexatriene (DPH, the parent compound of TMA-DPH) incorporated into the
membranes is described in (20). (Note: TMA-DPH is now the preferred of the two
probes as its cationic tertiary amine helps to anchor one end of the probe
molecule to the bilayer-water interface.)
References (21-23) describe time-resolved fluorescence anisotropy studies
(using TMA-DPH and DPH as fluorescent probes), of the effect of immune com-
plexes on macrophage membrane fluidity, the effect of succinate and phenyl
succinate on the fluidity of the inner mitochondrial membrane, and the role of
cholesterol in modulating the fluidity of renal cortical brush border and
basolateral membranes, respectively.
A time-resolved fluorescence anisotropy study of the base motions in DNA,
made using the intrinsic fluorescence of the bases themselves is described in
(24). Large amplitude motions were found on the picosecond timescale that are
sensitive to solvent viscosity. This provides support for the idea that base
motions contribute substantially to the observed fluorescence depolarization in
DNA and is consistent with a mechanism that involves concerted motions in the
interior of the polymer.
Reference (25) details a study in which time-resolved fluorescence intensity
and anisotropy decay measurements were combined with stopped-flow mixing
to monitor the local and global dynamics of E. coli dihydrofolate reductase
during refolding. A general review of such (so-called) 'double kinetic' experi-
ments is given in (26).
Recent contributions that both time-resolved fluorescence intensity and
anisotropy decay studies have made to our understanding of the structure and
dynamics of proteins and nucleic acids are reviewed in (27).
2.5.5 Caveats
It is very important that the lifetime of the fluorophore (either intrinsic or
extrinsic) being used to measure the anisotropy decay be appropriate to the
timescale of the motions under examination. Too short a lifetime will mean
that the fluorescence intensity will 'die away' before much rotational motion
takes place; the anisotropy will remain 'high' over the course of the fluorescence
intensity decay and the latter's components will contain limited information
about the motional dynamics. On the other hand, too long a lifetime means that
the fluorescence will become too quickly depolarized on the timescale of the
intensity decay, so that again the signal will contain limited information about
the dynamics. Generally, tryptophan fluorescence (with a lifetime of a couple of
nanoseconds) can be reliably used to measure o values of up to about 25 ns.
77
THOMAS D. BRADRICK AND JORGE E. CHURCHICH
(The required accuracy for o depends, of course, on the purpose of the measure-
ment.) Protein motions with o values greater than this will probably require
that a longer-lived extrinsic fluorophore (e.g. pyrene) be attached to the
molecule. Motions on the microsecond to second timescales can be monitored
using phosphorescent probes (see Section 4.2).
In the event that one wishes to measure only the fluorescence decay times of
a sample, but is forced to make the measurement with polarizers present, it is
imperative that the emission polarizer be oriented at an angle of 54.7° — 55°
with respect to the vertical excitation orientation (the so-called 'magic angle') in
order to correct for the presence of the polarizers. Otherwise, as the 'apparent'
decay rates of the horizontal and vertical fluorescence components are due to
both the depletion of the excited state population and to fluorophore motion (as
we mentioned in Section 2.5.1), the measured 'lifetimes' will be incorrect.
Proteins are generally excited at 295 nm in order to avoid (i) energy transfer
from tyrosine to tryptophan residues, which can occur at shorter wavelengths
(e.g. at 280 nm where tryptophan has its maximum absorbance) and which
depolarizes the fluorescence, and (ii) depolarization of tryptophan fluorescence
due to relaxation from its 1Lb to 1La electronic states. (At shorter wavelengths
electrons are excited to both levels, after which electrons in the higher level
relax radiationlessly to the lower, which also depolarizes the fluorescence.) One
would like to maximize the initial anisotropy in order to obtain the most
sensitivity for polarized fluorescence measurements.
Due to a process called photoselection, r0 can have a maximum value of 0.4.
A measured value for r0 that is significantly less than this strongly suggests that
some fast depolarization process is taking place that is beyond the equipment's
temporal resolution. Energy transfer, excimer formation, fast local libration and
radiationless relaxation between two different electronic states (such as that
from 1Lb to 1La in tryptophan) are some examples of processes that result in a
significant decrease in r0.
78
TIME-RESOLVED FLUORESCENCE AND PHOSPHORESCENCE SPECTROSCOPY
below it. These vertical distances or differences are called residuals; mathemati-
cally, the best fit is defined as being that which minimizes the sum of their
squared values. (The residuals are squared as the maximum likelihood of a
normal distribution is then at their centres.) The individual residuals are often
weighted according to their uncertainties (if known), so that the 'fit' is not
unfairly influenced by values that possess a relatively high degree of
uncertainty. If the sum of the squares of the weighted residuals is 'rescaled'
according to the 'number of degrees of freedom,' we have the definition of x2,
which has an optimal value of unity.
A general discussion of the use of least-squares fitting in fluorescence
measurements may be found in (28). The global analysis of fluorescence data is
discussed in (29). Commercially available time-resolved fluorimeters are typically
sold with data analysis software included. Available stand-alone packages in-
clude the Globals Unlimited suite, which is capable of analysing both time- and
frequency-domain data, stopped-flow kinetics, etc. The Center for Fluorescence
Spectroscopy at the University of Maryland (USA) also offers software for
frequency- and time-domain fluorescence lifetime analysis.
79
THOMAS D. BRADRICK AND JORGE E. CHURCHICH
proceeds from some starting point. One begins the fitting procedure using
initial 'guesses' for the parameter values, and the fitting program employs one
of several techniques to try to reach the best-fitting values by 'stepping' towards
them from the initial guesses. The values for the initial guesses need to be
chosen with care as the fitting program can have difficulty reaching the best-
fitting values if the former are too far away from the latter. For a single
exponential fit, one can estimate the lifetime from the rate at which the
fluorescence decays at long times (i.e. far away from the excitation pulse and
the distortion it causes). For tryptophan fluorescence in proteins, a clustering of
values around 1, 3 and 7 ns occurs, which can serve as initial guesses. For other
fluorophores, published values can be used. The fitting procedure is less
sensitive to the initial guesses for the amplitudes, so arbitrary values for them
often work.
Time-
80
TIME-RESOLVED FLUORESCENCE AND PHOSPHORESCENCE SPECTROSCOPY
Rgures 2 and 3 show how the appearance of the residuals and autocorrelation
function for a pulsed excitation experiment typically depend on the appropri-
ateness of the fitting function. In panel A of both figures, L shows an actual
excitation pulse profile (more properly, the instrument response function) that
was generated by an argon ion laser that pumped a dye laser circulating
rhodamine 6G, the tuned output of which was frequency-doubled to 295 nm
(nanometer = 1CT9 m = 10 A) by passage through a B-barium borate crystal (cf.
Section 3.1.1). The pulse was recorded using a photomultiplier after being
scattered by a dilute silica suspension in distilled water. Synthetic data (F in the
figures) were generated by convolving a double exponential decay (x1 = 2.5, T1
= 2.5 ns, a 2 = 0.5, T 2 = 5.0 ns) with the laser pulse profile. Random Gaussian
noise, appropriately scaled for single photon counting (cf. Section 3.3.1), was
then added to this 'signal'. The resulting peak intensity was about 45500
counts.
Figure 2 shows the results of fitting the convolution of a single exponential to
this decay, which yielded best-fitting values of a = 2.8 and T = 3.0 ns. This
lifetime is close to the intensity-weighted average lifetime of 3.2 ns (given by
E/iTi, where the fi are the fractional intensities (equation 3)), which is usually to
Time-
81
THOMAS D. BRADRICK AND JORGE E. CHURCHICH
be expected. The fit can also be seen in panel A as it falls somewhat below the
'data' at the peak of the fluorescence curve. Otherwise, the two seem to agree
fairly well, especially at long times. The weighted residuals (Figure 2B) indicate,
however, that the single exponential fit deviates from the data essentially every-
where along the decay curve, and the autocorrelation of the residuals (Figure 2C)
is also far from being random, which confirms that a definite trend exists in the
residuals. The x2 value of 14.88 also indicates a poor fit. (Recall that the aim of
the fitting procedure is to 'adjust' the model function so that it evenly 'cuts'
through the noise in the data. If the model is fundamentally inadequate, no
amount of adjustment will enable one to obtain a good fit with a random dis-
tribution in the residuals.)
Figure 3 shows the results of the double exponential fit that was attempted
next. In this case the fit appears to completely overlay the data in panel A and
this impression is borne out by the completely random distribution of the
residuals around the zero line (panel B). The random distribution of the
autocorrelation function values (Figure 3C) confirms that there is indeed no
trend in the residuals, while a x2 value of 0.84 (close to unity) further attests to
the goodness of the fit. A fit of a triple exponential decay was also attempted
(results not shown). A x2 value of 0.85 indicated that the addition of another
term was not warranted, and this judgement was reinforced by the fact that
two of the three exponentials had nearly identical lifetimes. (The general rule is
that, for -300 degrees of freedom (i.e. —305 data points), the addition of
another exponential term is not justified if it results in a less than 10%
improvement in x2-) Thus, the decay is taken to be double exponential, the
results for which were x1 = 2.55, T1 = 2.52 ns, x2 = 0.45 and T2 = 4.61 ns. These
values are in excellent agreement with the generating function concerning the
short lifetime component. The amplitude and lifetime values of the long life-
time component are each about 10% in error, presumably because that 'species'
only contributed about 28% of the total fluorescence intensity at the chosen
emission wavelength, and the two lifetimes only differ by a factor of two. Much
greater accuracy can be obtained by carrying out a global analysis that includes
fluorescence decays measured at other emission wavelengths (i.e. DAS, Section
2.4) at which the longer lifetime component would make a different con-
tribution.
82
TIME-RESOLVED FLUORESCENCE AND PHOSPHORESCENCE SPECTROSCOPY
model function for the anisotropy decay and Ij, is an 'association matrix' (made
up of ones and zeros) that combines the appropriate intensity and anisotropy
decay components, depending on whether the decay is associative or non-
associative (30). The expression for r(t) can be modified as required until a good
fit is obtained, where x2 and the form of the residuals and autocorrelation again
serve as the fitting criteria as described previously. Alternatively, I||(t) = F(t)[l +
2r(t)Lij]/3 and J±(t) = F(t)[1 - r(t)Lij]/3g convolved with the lamp pulse profile may
be fitted simultaneously to their respective polarized fluorescence decay
components in a global analysis.
100
83
THOMAS D. BRADRICK AND JORGE E. CHURCHICH
for which a, = 0.33, Ta = 2.0 ns, a2 = 0.67 and T2 = 5.0 ns. Gaussian random
noise with standard deviations of 0.2° and 0.005 (10, 18) for the phase angle and
modulation, respectively, were added to the calculated values. These standard
deviations were also used as the inverse weighting factors in the non-linear
least-squares analysis of the 'data.' (Note: for phase/modulation data the weight-
ing factors, which are the reciprocals of the standard deviations, are estimated
from repeated measurements so that, depending on their accuracy, the optimal
X2 value may not be unity. The appropriate number of decay components is
simply that which minimizes x2.)
The 'data' are shown in Figure 4A, where the closed circles are the phase angle
values and the open circles are the per cent modulation (modulation x 100%)
values, both plotted as functions of the modulation frequency (in MHz). The
solid lines (which clearly do not everywhere overlay the data) show the results
of simultaneous fits of the single-exponential expressions (9 = arctan(2TTfr) and
m = (1 + 4TT2f2T2)-1/2) to the data (x2 = 30.1), which yielded a value of 4.1 ns for
the single lifetime. This value is close to the intensity-weighted average lifetime
(T) of 4.5 ns, which is usually to be expected. The residuals for this fit (closed
circles in panels B and C for the phase and per cent modulation residuals,
respectively), clearly show trends, which reinforces our impression from panel
A that this decay is not well modelled by a single exponential. The dashed lines
in Figure 4A (which do seem to fit the data well) show the results of a double-
exponential fit to the data (x2 = 0.57), which gave values of x1 = 0.345, T1= 2.04
ns, x2 = 0.655 and T2 = 5.02 ns. These are in excellent agreement with the values
used to generate the data, and the goodness of the fit is indicated by the random
distribution in the residuals (solid squares for the phase and per cent modu-
lation residuals in panels B and C, respectively). Usually, the autocorrelation
function values are not calculated for fits to phase/modulation data. The reader
may consult (10) for the equations used in analysing polarized phase/modulation
data.
84
TIME-RESOLVED FLUORESCENCE AND PHOSPHORESCENCE SPECTROSCOPY
85
THOMAS D. BRADRICK AND JORGE E. CHURCHICH
3.2 Detectors
Commercially available time-resolved fluorescence spectrometers typically use
photomultiplier tubes (PMTs) to convert light into an electronic signal for
further processing. As was pointed out in Footnote 3, however, PMTs distort the
fluorescence decay profile by broadening it; this places a limit on how short a
fluorescence decay time a photomultiplier-based time-resolved system can
measure. PMTs are described in Section 4.6 of Chapter 2 ; the reader can also
find additional information in (36).
Other, 'faster' detectors that can be used in place of a conventional PMT
include microchannel plate photomultiplier tubes (MCP-PMTs) (31) and streak
cameras (37). Because of their expense, the use of these devices is usually con-
fined to 'home built' fluorimeters found in dedicated fluorescence laboratories,
and is therefore not discussed here.
86
TIME-RESOLVED FLUORESCENCE AND PHOSPHORESCENCE SPECTROSCOPY
a portion of the beam may be reflected into a photodiode whose output pro-
vides the triggering pulse. Often there has to be a delay between when the
triggering signal arrives at the recording device and when the latter can first
begin to process the incoming fluorescence signal from the detector. The arrival
of the fluorescence signal typically has to be delayed to allow for this, which can
be accomplished by simply increasing the length of cable connecting the
detector and the recorder. (It takes an electronic signal about 1 ns to travel one
foot (—0.3 m).) What happens next depends on the recording technique being
used.
Luminescence decay
Figure 5 Events occurring during excitation of a sample in the case of pulse sampling
(fluorescence) and phosphorescence measurements, te is the width at half height of the
excitation pulse, td is the delay from the arrival of the triggering signal at tt to the beginning
of observation and tg is the gate width of the detector, during which time the system
measures the hatched portion of the decay curve.
87
THOMAS D. BRADRICK AND JORGE E. CHURCHICH
will tend to average out. Once the pre-set number of samples has been reached,
the capacitor is discharged through a resistor, whose voltage is digitized and
stored. The whole process then begins again, with the shutter now opening
after a new delay time of td' = td + tg. In this way, the shutter's 'window' is
scanned across the decay profile, which allows the latter to be recorded in
discrete, consecutive segments.
Alternatively, the voltage being applied to the detector can be sequentially
gated and the signal from the latter stored in a multi-channel analyser (MCA) as
described in (38). (Note: the main disadvantage that pulse-sampling methods
have is that the PMT is operated in the linear or low-gain analogue mode, so
that its output changes if its high-voltage supply 'drifts.')
Boxcar averagers and their use in signal recovery are described in (39). PTI
(Photon Technology International) manufactures a series of fluorescence life-
time spectrometers (models C-70 to C-73) that use a stroboscopic detection
technique (i.e. a gated PMT, as described in the previous paragraph) and either
nanosecond flashlamp or nitrogen/dye laser excitation.
88
TIME-RESOLVED FLUORESCENCE AND PHOSPHORESCENCE SPECTROSCOPY
4 Phosphorescence
4.1 Phosphorescence of proteins
The triplet lifetimes of aromatic amino acid residues in proteins show a strong
dependence on the viscosity of the matrix. In rigid glasses consisting of mix-
tures containing glycerol/water 50:50 v/v at 77 K, the phosphorescence lifetime
of free tryptophan is typically around 6 s (42). Smaller values of the phosphor-
escence lifetime of proteins, excited at 295 nm, imply specific interactions:
e.g. electron transfer with a group in the immediate vicinity (43) or inter-
tryptophanyl energy transfer. The phosphorescence spectra of proteins are
dominated by tryptophanyl groups that show vibronic structures with a 0-0
vibrational band centred at around 412 nm. The well-resolved vibrational struc-
ture of the phosphorescence emission allows one in some cases to distinguish
between the emission spectra of tryptophanyl groups positioned in environ-
ments of different polarity (44).
For phosphorescence measurements, the samples are typically excited using
a pulsed xenon flash lamp. A commercially available instrument (the Perkin-
Elmer LS-50B luminescence spectrophotometer) is suitable for phosphorescence
89
THOMAS D, BRADRICK AND JORGE E. CHURCHICH
Protocol 1
Phosphorescence sample deoxygenation
Reagents
• Glucose oxidase and catalase (Sigma) and • Protein buffer, pH 7
glucose (Fisher)
Method
1 Prepare stock solutions of glucose oxidase (20 mg/ml) and catalase (2 mg/ml) in
protein buffer.a
2 To the desired amount of lyophilized protein, add 3 mg of glucose for every ml of
final sample volume. Add the desired amount of protein buffer and place the
solution in a cuvette.
3 For every ml of protein/glucose solution, add 1 (ul each of the glucose oxidase and
catalase stock solutions, stopper the cuvette and mix.
a These solutions should keep for a couple of weeks when stored at 4 °C.
90
TIME-RESOLVED FLUORESCENCE AND PHOSPHORESCENCE SPECTROSCOPY
91
THOMAS D. BRADRICK AND JORGE E. CHURCHICH
320 340 360 380 400 420 440 460 480 500
WAVELENGTH (nm)
92
TIME-RESOLVED FLUORESCENCE AND PHOSPHORESCENCE SPECTROSCOPY
escence because the critical transfer distance is 3 A. On the other hand, Trp87
and Trp5, separated by 12 and 22 A, respectively, from the metal binding site,
show weaker sensitized luminescence. However, the significant decrease in the
sensitized luminescence of the double mutant as compared to the single mutant
can be attributed to two independent factors: i.e. change in the distance of the
donor-acceptor pair (Trp87-Tb3+) and change in coordination efficiency as
demonstrated by the binding studies (Table 1).
The reader may find (55) and (56) provide additional useful information about
phosphorescence measurements.
spectra of inositol monophosphatase (2 uM) in the presence of 10 uM TbCI3 before (—) and
after addition of 10 mM MgCI2 (...). The samples were dissolved in 10 mM Tris-HCI buffer
(pH 7.5) at room temperature. The excitation wavelength was 295 nm. The delay time was
0.1 ms, gate time was 1 ms and the slit widths for excitation and emission were 5 nm.
93
0.05
2.0 4.0 6.0 8.0 10.0
Milliseconds
Figure 7 Phosphorescence anisotropy decay in ghosts. The anisotropy decay of eosin-
labelled proteins was monitored as a function of time. The data were fitted according to
equation 21, Reprinted in part with permission from Matayoshi, E. D. and Jovin, T. M. (1991)
Biochemistry, 30, 3527-3538, Copyright 1991 American Chemical Society.
94
TIME-RESOLVED FLUORESCENCE AND PHOSPHORESCENCE SPECTROSCOPY
Acknowledgements
T. D. B. would like to express his gratitude to the National Institute of Neuro-
logical Disorders and Stroke, the National Institutes of Health (NIH), for its
continuing support in the form of a postdoctoral fellowship, and to Dr Jay R.
Knutson, Chief, Optical Spectroscopy Section, National Heart, Lung and Blood
Institute, NIH, for welcoming him into the latter's laboratory. In addition, he
would like to thank Dr Knutson and Dr Solon Georghiou for their critical
reading of and helpful comments concerning the fluorescence portion of this
chapter. J. E. C. wishes to express his appreciation for the support given by the
Hong Kong Technical University during his stay there as a visiting professor.
References
1. Lakowicz, J. R. (1983). Principles of fluorescence spectroscopy, p. 383. Plenum Press, NY.
2. McGlynn, S. P., Azumi, T. and Kinoshita, M. (1969). Molecular spectroscopy of the triplet
state, p. 261. Prentice Hall, NJ.
3. Forster, T. B. (1959). Discuss. Faraday Soc., 27, 7.
4. Forster, T. B. (1965). In Modern quantum chemistry, Part III (ed. O. Sinanoglu), p. 93.
Academic Press, NY.
5. Kellog, R. E. (1967). J. Chem. Phys., 47, 3403.
6. Ermolaev, V. L and Sveshnikova, E. B. (1964). Opt. Spectroscopy (English Translation), 14,
320.
7. Berlman, I. B. (1973). Energy transfer parameters of aromatic molecules. Academic Press,
NY.
8. Cheung, H. C. (1991). In Topics in fluorescence spectroscopy, vol. 2: principles (ed. J. R.
Lakowicz), p. 127. Plenum Press, NY.
9. Fernandez, S. M. and Sha'afi, R. I. (1983). In Fast methods in physical biochemistry and cell
biology (ed. S. M. Fernandez and R. I. Sha'afi), p. 1. Elsevier, NY.
10. Lakowicz, J. R. and Gryczynski, I. (1991). In Topics in fluorescence spectroscopy, vol. 1:
techniques (ed. J. R. Lakowicz), p. 293. Plenum Press, NY.
11. Knutson, J. R., Beechem, J. M. and Brand, L. (1983). Chem. Phys. Lett., 102, 501.
12. Royer, C. A., Mann, C. J. and Matthews, C. R. (1993). Protein Sci., 2,1844.
13. Shen, F., Triezenberg, S. J., Hensley, P., Porter, D. and Knutson, J. R. (1996). J. Biol.
Chem., 271, 4819.
14. Green, S. M., Knutson, J. R. and Hensley, P. (1990). Biochemistry, 29, 9159.
15. Alcala, J. R., Gratton, E. and Prendergast, F. G. (1987). Biophys. J., 51, 587.
16. Alcala, J. R., Gratton, E. and Prendergast, F. G. (1987) Biophys./., 51, 597.
17. Alcala, J. R., Gratton, E. and Prendergast, F. G. (1987). Biophys. J., 51, 925.
18. Knutson, J. R. (1992). In Methods in enzymology (ed. L. Brand and M. L Johnson). Vol.
210, p. 357. Academic Press, NY.
19. Steiner, R. F. (1991). In Topics in fluorescence spectroscopy, volume 2: principles (ed. J. R.
Lakowicz), p. 1. Plenum Press, NY.
20. Bradrick, T. D., Dasseux, J.-L, Abdalla, M., Aminzadeh, A. and Georghiou, S. (1987).
Biocm'm. Biophys. Ada, 900, 17.
21. Petty, H. R., Niebylski, C. D. and Francis, J. W. (1987). Biochemistry, 26, 6340.
22. Mutet, C., Duportail, G., Cremel, G. and Waksman, A. (1984). Biochem. Biophys. Res.
Commun., 119, 854.
23. Molitoris, B. A. and Hoilien, C. (1987). J. Membrane Biol, 99, 165.
24. Georghiou, S., Bradrick, T. D., Philippetis, A. and Beechem, J. M. (1996). Biophys. J., 70,
1909.
95
THOMAS D. BRADRICK AND JORGE E. CHURCHICH
25. Jones, B. E., Beechem, J. M. and Matthews, C. R. (1995). Biochemistry, 34, 1867.
26. Beechem, J. M. (1997). In Methods in enzymology (ed. L. Brand and M. L. Johnson). Vol.
278, p. 24. Academic Press, NY.
27. Millar, D. P. (1996). Curr. opin. struct, biol, 6, 637.
28. Straume, M., Frasier-Cadoret, S. G. and Johnson, M. L. (1991). In Topics in fluorescence
spectroscopy, vol. 2: principles (ed. J. R. Lakowicz), p. 177. Plenum Press, NY.
29. Beechem, J. M., Gratton, E., Ameloot, M., Knutson, J. R. and Brand, L. (1991). In
Topics influorescencespectroscopy, vol. 2: principles (ed. J. R. Lakowicz), p. 241. Plenum
Press, NY.
30. Brand, L., Knutson, J. R., Davenport, L, Beechem, J. M., Dale, R. E., Walbridge, D. G.
and Kowalczyk, A. A. (1985). In Spectroscopy and. the dynamics of molecular biological
systems (ed. P. M. Bayley and R. E. Dale), p. 259. Academic Press, London.
31. Small, E. W. (1991). In Topics in fluorescence spectroscopy, vol. 1: techniques (ed. J. R.
Lakowicz), p. 97. Plenum Press, NY.
32. Rabek.J. F. (1982). Experimental methods in photochemistry and photophysics, p. 593. Wiley,
NY.
33. Silfvast, W. T. (1996). Laser fundamentals. Cambridge University Press, NY.
34. Davis, C. C. (1996). Losers and electro-optics: fundamentals and engineering. Cambridge
University Press, Cambridge, UK.
35. Munro, I. H. and Martin, M. M. (1991). In Topics in fluorescence spectroscopy, vol. 1:
techniques (ed. J. R. Lakowicz), p. 261. Plenum Press, NY.
36. Rabek, J. F. (1982). Experimental methods in photochemistry and photophysics, p. 459. Wiley,
NY.
37. Nordlund, T. M. (1991). In Topics in fluorescence spectroscopy, vol. 1: techniques (ed. J. R.
Lakowicz), p. 183. Plenum Press, NY.
38. Lakowicz, J. R. (1983). Principles of fluorescence spectroscopy, p. 62. Plenum Press, NY.
39. Rabek, J. F. (1982). Experimental methods in photochemistry and photophysics, p. 554. Wiley,
NY.
40. O'Connor, D. V. and Phillips, D. (1984). Time-correlated single photon counting. Academic
Press, London.
41. Birch, D. J. S. and Imhof, R. E. (1991). In Topics in fluorescence spectroscopy, vol. V.
techniques (ed. J. R. Lakowicz), p. 1. Plenum Press, NY.
42. Longworth, J. W. (1971). In Excited states of proteins and nucleic adds (ed. R. F. Steiner
and I. Weinryb), p. 319. Plenum Press, NY.
43. Churchich, J. E. (1966). Biochim. Biophys. Acta, 120, 406.
44. Strambini, G. B. and Gonnelli, M. (1990). Biochemistry, 29, 196.
45. Saviotti, M. L. and Galley, W. C. (1974). Proc. Natl. Acad. Sci. USA, 71, 4154.
46. Englander, S. W., Calhoun, D. B. and Englander, J. J. (1987). Anal. Biochem., 161,
300.
47. Strambini, G. B. (1987). Biophys.]., 52, 23.
48. Lakowicz, J. R. and Weber, G. (1973). Biochemistry, 12, 4161.
49. Guzeman, O. L, Kaufman, J. F. and Porter, G. (1973). J. Chem. Soc., Faraday Trans. II, 69,
708.
50. Gregory, R. B. and Lumry, R. (1985). Biopolymers, 24, 301.
51. Bone, R., Spinger, J. P. and Atack, J. R. (1992). Proc. Natl. Acad. Sci. USA, 89, 10031.
52. Gore, M. G., Greasley, P., McAllister, G. and Ragan, C. I. (1993). Biochem. J., 296, 811.
53. Evans, C. H. (1990). Biochemistry of the lanthanides. Plenum Press, NY.
54. Moreno, F., Corrales, S., Garcia Blanco, F., Gore, M. G., Rees-Milton, K. and
Churchich, J. E. (1996). Eur.J. Biochem., 240, 435.
55. Vanderkooi, J. M. (1992). In Topics in fluorescence spectroscopy, vol. 3: biochemical
applications (ed. J. R. Lakowicz), p. 113. Plenum Press, NY.
96
TIME-RESOLVED FLUORESCENCE AND PHOSPHORESCENCE SPECTROSCOPY
56. Schauerte, J. A., Steel, D. G. and Gafhi, A. (1997). In Methods in enzymology (ed. L. Brand
and M. L. Johnson), Vol. 278, p. 49. Academic Press, NY.
57. Brochon, J.-C., Wahl, P. and Auchet, J.-C. (1972). Eur. J. Biochem., 25, 20.
58. Cherry, R. J. and Schneider, G. (1976). Biochemistry, 15, 3657.
59. Matayoshi, E. D. andjovin, T. M. (1991). Biochemistry, 30, 3527.
97
This page intentionally left blank
Chapter 4
Introduction to circular
dichroism
Alison Rodger and Matthew A. Ismail
Department of Chemistry, University of Warwick, Coventry CV4 7AL
1 Introduction
1.1 Circular dichroism
Circular dichroism (CD) is the ideal technique for studying chiral molecules in
solution. It is uniquely sensitive to the asymmetry of the system. These features
make it particularly attractive for biological systems. CD is by definition the
difference in absorption, A, of left and right circularly polarized light (CPL):
CPL has the electric field vector of the electromagnetic radiation retaining con-
stant magnitude in time but tracing out a helix about the propagation direction
(Figure 1). Following the optics convention we take the tip of the electric field
vector of right CPL to trace out a right-handed helix in space at any instant of
time (Figure 2) (1, 2).
CD spectra can in principle be measured with any frequency of electro-
magnetic radiation. In practice, most CD spectroscopy involves the ultraviolet-
visible (UV-visible) regions of the spectrum and electronic transitions, though
increasing progress is being made with measuring the CD spectra of vibrational
transitions using infrared radiation. We shall limit our consideration to elec-
tronic CD spectroscopy since the practical considerations for vibrational CD
differ from those for electronic CD.
For randomly oriented samples, such as solutions, a net CD signal will only
Figure 1 Schematic illustration of CD. (Source: Circular dichroism and linear dichroism, A.
Rodger and B. Norden, 1997, by permission of Oxford University Press.)
99
ALISON RODGER AND MATTHEW A. ISMAIL
Figure 2 (a) Linearly and (b) circularly polarized electromagnetic radiation. The arrows denote
direction of the electric field £. At each point in space or time, the magnetic field, B, is
perpendicular to the electric field such that k (the propagation direction), E and B form a
right-handed axis system. (Source Circular dichroism and linear dichroism, A. Rodger and B.
Nordgn, 1997, by permission of Oxford University Press.)
Wavelength (nm)
100
INTRODUCTION TO CIRCULAR DICHROISM
2 Measuring a CD spectrum
2.1 The instrumentation
Measuring a CD spectrum is a routine procedure assuming one has access to a
CD instrument. CD instruments are usually referred to as circular dichrometers
or spectropolarimeters; however, dichrograph and spectropolarograph better
describe the output of modern instruments. If your sample gives a good UV-
visible absorbance spectrum then it is highly likely that (if it is chiral) you will
get a good CD spectrum. The essential features of a CD spectrometer are a source
of (more-or-less) monochromatic left and right circularly polarized light and a
means of detecting the difference in absorbance of the two polarizations of
light. As the absorbance signal of a typical CD spectrum is a fraction of a per
cent of that observed in the normal absorbance spectrum, the normal method
of achieving these requirements is to implement a polarization phase-
modulation technique.
In most commercial CD spectrometers, a photoelastic modulator (in older
101
ALISON RODGER AND MATTHEW A. ISMAIL
instruments a Pockels cell) produces alternatively right and left circularly polar-
ized light with a switching frequency of typically 50 kHz. The light intensity is
constant, but upon passage through a sample exhibiting CD an intensity fluctu-
ation (corresponding to the different absorptions of left and right circularly
polarized light) that is in phase with the modulator frequency appears. The un-
absorbed photons hit a PMT which produces a current whose magnitude
depends on the number of incident photons. This current is detected by a lock-
in amplifier. Thus, the DC (direct current) component of the PMT current de-
pends on the total absorption of light by the sample (and on lamp intensity and
monochromator characteristics), and the AC (alternating current) component
relates to the CD (a lock-in amplifier is needed to determine the phase of the AC
component which contains information about the sign of the CD). A larger
voltage is applied to the PMT when the number of incident (i.e. non-absorbed)
photons is smaller, thus the high tension (HT) voltage on a CD machine is a rough guide
to the absorbance of a sample. The HT voltage can usually be monitored if one
wishes to do so. It is advisable to take this option even with routine samples, as
it is an easy way to detect sample abnormalities.
The CD spectrometers described above are single beam instruments. A new
double beam instrument from On-Line Instrument Systems (OLIS) has recently
come on the market, which uses sample and reference beams to enable direct
calculation of the CD signal. The instrument utilizes two modulated CPL beams
out of phase by 180° which are passed through the sample and detected simul-
taneously. The absorption of left and right CPL is therefore measured directly.
The technique has the advantage that factors such as lamp fluctuations are
negated since they affect both beams equally, whereas the CD behaviour is
characteristic to each beam. The use of a sample/reference system results in
zero drift over time and nearly flat baselines. However, the two beams pass
through different parts of the sample and cuvette. This may or may not be
important.
The radiation source in UV-visible CD spectrometers is a high energy
(150-450 W) xenon arc lamp. In some instruments the lamp is water cooled.
This can prove the most problematic part of the whole instrument since if the
local water supply is used for a flow through system one is at the mercy of its
purity, hardness etc. Alternatively, if a self-contained system is used, then the
fairly narrow bore of the tubing that the water flows through requires a
reasonably efficient pump, and the water must be kept free from algae etc. and
must be cooled. A narrow bore central heating system pump may be the
cheapest solution to the pumping problem and using car radiator cooling fluid
inhibits the growth of most things if regularly changed. Avoid using clear
tubing if possible. Despite the water consumption required, a water flow heat
exchanger may be the only cooling option that is sufficiently efficient to cool
the water (even for a self contained system) for a 450 W lamp. If at all possible,
an instrument with an air cooled lamp is to be preferred. This no longer seems
to involve accepting a loss in sensitivity.
Another requirement of the high energy lamps used in CD machines is that
102
INTRODUCTION TO CIRCULAR DICHROISM
the optics of the instrument need to be purged with nitrogen gas to avoid ozone
being created and reacting with surfaces of mirrors etc. The nitrogen purging is
also required for the sample compartment when running below 200 nm to
avoid having O2 in the sample compartment absorbing the incident radiation.
In practice this means a moderate nitrogen flow rate (3-5 cm3/min) at all times,
with an increase to 20 or more cm3/min when collecting data below 200 nm.
103
ALISON RODGER AND MATTHEW A. ISMAIL
104
INTRODUCTION TO CIRCULAR DICHROISM
Nuts
Metal plates
Figure 4 Cell holder for cylindrical quartz UV-visibfe short pathlength [less than 0.1 mm)
cuvettes.
with the whole incident light beam passing through the cell windows when
data is being collected.
105
ALISON RODGER AND MATTHEW A. ISMAIL
Wavelength (nm)
Figure 5 Normal absorption spectra of 100 uM calf thymus DNA in (i) 10 mM NaCI and 1 mM
sodium cacodylate buffer (solid line) and (ii) 10 mw NaCI, 1 mw sodium cacodylate and
50 uM spermine (broken line), which is sufficient to condense the DNA and provide light
scattering that may appear as an absorbance signal outside the absorption envelope.
106
INTRODUCTION TO CIRCULAR DICHROISM
107
ALISON RODGER AND MATTHEW A. ISMAIL
4 -
2 -
0 -
Wavelength (nm)
Figure 6 CD spectrum of Hoechst 33258 (20 mM). calf thymus DNA (40 mM), NaCI (20 mM)
and phosphate buffer (1 mM; pH 7) (broken line) where it is clear what the noise reduced
spectrum (solid line) should look like.
Protocol 1
Measuring a routine CD spectrum
Equipment
• UV-visible absorbance spectrophotometer • Quartz cuvette of an appropriate type (see
• Computer-controlled CD spectrometer Section 2.3)
• CD spectrum measurement and data • cm3 volumes of sample and reference
analysis software solutions of uM concentrations
A. Sample preparation
1 Determine an appropriate cuvette type for the sample. Use a 1-mm pathlength
quartz cuvette for strongly absorbing samples or when only a small volume of the
sample is available. Otherwise, use a 10-mm pathlength rectangular quartz cuvette."
2 Run a UV-visible spectrum of the buffer/solvent over the wavelength range of
interest. If using a double beam UV-visible spectrophotometer, ensure that the
spectrum is collected with the reference beam empty.b
108
INTRODUCTION TO CIRCULAR DICHROISM
Protocol 1 continued
3 Prepare a solution of the sample such that the sample absorption at the wave-
length of interest is approximately 1 but not more than 1.5 absorbance units. Run a
UV/visible absorbance spectrum, again with an empty reference beam.
4 Compare the sample and buffer solvent UV-visible spectra and check that the
wavelength maximum of interest is due to sample absorption and not some other
component of the system.
B. CD parameter selection
1 Set the wavelength range for the CD spectrum to begin at least 20 nm beyond the
absorption envelope of interest and ensure that the scan begins in a region of the
spectrum where there is no absorption.
2 Place the quartz cuvette containing the sample solution into the sample compart-
ment and ensure that the light beam is not clipping the meniscus of the solution
(increase the volume of liquid in the cuvette if necessary or perhaps raise the
height of the cuvette) or the walls or base of the cuvette (in which case lower the
cuvette or use a different cuvette type). You may be able to remove the cuvette
holder assembly to view down the light beam path. Alternatively use a mirror.
3 Run a fast, approximate scan of the sample to determine an appropriate CD
sensitivity level. Use T = 0.25 s, b - 0.5 nm and s = 500 nm min-1 accumulated over
one scan only,
4 For the accurate spectrum, set T,.Sand b such that T is as large as possible subject to
T x s - b/2. Use s - 100 nm min-1, T - 1 s, b = 1 nm and d = 0.5 nm as a starting
point for routine work. Specify up to eight accumulated scans depending on the
quality of the spectrum required/ In some instances more scans may be required. If
you are uncertain of what to choose, select a large number and stop the run when
the spectrum 'looks OK'. To ensure proper averaging over the number of scans, it is
advisable to stop the instrument while it is rewinding between scans.
5 If collecting CD data for input into the structure fitting program CDsstr (see
Protocol 2), set d = 1 nm.
109
ALISON RODGER AND MATTHEW A. ISMAIL
Protocol 1 continued
3 Overlay the final spectrum on the original spectrum to ensure no peak shape
distortion or data mishandling has occurred.
" Other cuvette types may be appropriate in certain circumstances. See main text for details.
b
If you put the solvent or buffer in the reference beam of a double beam normal absorption
spectrometer you may get a good spectrum even if the solvent/buffer has a significant
absorbance as the instrument will subtract the absorbance of the reference cuvette; however,
there is no such compensation mechanisms in a single beam CD spectrometer.
'The signal to noise ratio is dependent on a number of factors as detailed in the text.
However, increasing the number of accumulated scans is a convenient practical method of
improving the quality of the CD spectrum.
d
By using the same cuvette for both sample and reference measurements, absorption of CPL
by the cuvette is eliminated from the final spectrum.
3 Equations of CD spectroscopy
An electronic transition occurs because either the electric field or the magnetic
field (or both) of the radiation 'pushes' the electrons to a new stationary state.
The effect of the electric field is to cause a linear rearrangement of the electrons;
the net linear displacement of charge during any transition is therefore called
the electric dipole transition moment (edtm) of the transition and is denoted by the
vector u. The magnetic field induces a circular rearrangement of electron
density. The net circulation of charge is the magnetic dipule transition moment
(mdtm), m. In an achiral molecule the net electron redistribution is always
planar. It is usually linear (having u = 0, m - 0 so being electric dipole allowed
(eda) and magnetic dipole forbidden (mdf)) or circular (having u - 0, m 1= 0 so being
electric dipole forbidden (edf) and magnetic dipole allowed (mda)), but it may also be a
planar spiral in molecules of low symmetry.
In a chiral molecule the electron rearrangement during a transition is always
helical which requires that u and m have a parallel component. The Rosenfeld
equation for the CD intensity (or CD strength or rotational strength, R) of a
transition in a collection of randomly oriented chiral molecules is (4):
where 'Im' denotes 'imaginary part of (the magnetic dipole moment operator
contains a factor of \/-1), u is the edtm for the transition from the final to the
initial state, and m is the mdtm for the reverse transition.
CD may sometimes be used to deduce the structure of a system. This is only
really viable when the system can be considered as a collection of chromo-
phores (spectroscopically well-defined subunits of a molecule) each of which is
only slightly perturbed by the rest of the system. In the rest of this section we
shall consider the coupling of two intrinsically achiral chromophorcs that are
chirally oriented with respect to one another. A range of applications of this
theory is given in (4). We have to treat eda/mdf transitions separately from
110
INTRODUCTION TO CIRCULAR DICHROISM
Figure 7 CD resulting from the coupling of one transition in each of two chromophores
where both transitions are eda and degenerate (occurring at energy E). The characteristic
form of an excitonic spectrum results from cancellation of overlapping positive and negative
bands. (Source: Circular dichroism and linear dichroism, A. Rodger and B. Norden, 1997, by
permission of Oxford University Press.)
111
ALISON RODGER AND MATTHEW A. ISMAIL
180°) the angle between -RAC and uc. and T the angle passed through in going
from the y-z projection of ua to that of uC in an anticlockwise direction as illustrated
in Figure 8. Thus
We shall refer to the system as being right handed if the three vectors uc, ua
and RAC (CAR) form a right-handed coordinate system. The angles relate to the
vectors as follows (4):
CAR right-handed
Figure 8 Diagram illustrating the geometry and coordinates for the A/C system described in
the text. Note that T is the angle taken in the anticlockwise direction between the projections
of the edtms onto the y-z plane when the observer is looking down the x-axis. 0° - T <
180° if uc x ua.RAC > 0 (i.e. if the three vectors form a right-handed parallelepiped) and
180° - T < 360° if uc x (ua'RAc - 0. (Source: Circular dichroism and linear dichroism, A.
Rodger and B. Norden, 1997, by permission of Oxford University Press.)
112
INTRODUCTION TO CIRCULAR DICHROISM
113
ALISON RODGER AND MATTHEW A. ISMAIL
c,ua
-eaec - 3RAC1
X ua.RAC} 11
KAC
(a)
Wavelength - Wavelength •
114
INTRODUCTION TO CIRCULAR DICHROISM
2. It is often the case that a = y - 90°. When this situation arises, the geometry
factor in equation 11 is cosTsinT = 1/2-isinf^T) so the CD is a maximum when ua
and u'' are oriented at 45° to one another.
3. If we wish to know the combined effect of many transitions in C on one
transition in A, we simply introduce a sum over c into equation I1.
4. The CD induced at t-c is determined by simply permuting all the 'a' and 'c'
and 'A' and 'C' labels in equation 11, and we see that K(ec) = -R(eJ).
33 Carbonyl n-TT* CD
Long before any theoretical understanding of the carbonyl n—TT* transition CD
had been achieved, it was found empirically that each atom (or group of atoms)
of the molecule induces a contribution to the observed CD whose sign is
determined by the octant in which the atom lies (6). Any atom that has an
equivalent 'partner' in a neighbouring octant has its contribution to the CD
cancelled so it is not a net perturber and may be ignored. Thus the CD of an
n—TT* transition has its sign determined by the factor (Figure 10)
where (x,y,z) are the coordinates of the unit vector along the line from the
carbonyl to the net perturbcr in the coordinate system of Figure 10. It was later
realized (7) that the relative contributions of two or more net perturbers scale
with their polarizability and their distance from the carbonyl group as indicated
below.
The equation describing the octant rule dependence of the dynamic coupling
induced CD of an n-rr* transition is derived as follows (4, 7). One first divides
the molecule into an achiral chromophorc A composed of the C=0 bond and
the two carbon atoms to which it is joined (Figure 11). The rest of the molecule is
divided into chromophores, C, which are subumits of the molecule that do not
exchange electrons with the rest of the molecule. For a molecule where (apart
Figure 10 The octant rule. (Source: Circular dichroism and linear dichroism, A. Rodger and
B. Norden, 1997, by permission of Oxford University Press.)
115
ALISON RODGER AND MATTHEW A. ISMAIL
Figure 11 Geometry of a carbonyl A/C system (cf. Figure 8). (Source: Circular dichroism and
linear dichroism, A. Rodger and B. Norden, 1997, by permission of Oxford University Press.)
from the carbonyl group) all the valence electrons are in sigma bonds, taking
the C to be bonds is a good approximation (7-10). Alternatively, the C may be
defined to be atoms such as carbons and nitrogens. Although the bond approxi-
mation is the better one, in most cases the difference in xyz for the middle of a
bond and one of the terminal atoms of the bond is not great. If there are TT-bonds
or non-bonding electrons in the molecule, then the unit where the electrons are
localized is often more than a single bond, e.g. if the molecule has a -COOH,
— NH2 or — C6H5 group, these will need to be treated as units.
We may write the CD of the n—*tt* transition in terms of an A optical factor,
f(A), and a C optical/geometry factor, g(C), (which includes the energy terms):
The first non-zero term of the CD strength for the carbonyl n-rr* transition has
an A optical factor
116
INTRODUCTION TO CIRCULAR DICHROISM
117
ALISON RODGER AND MATTHEW A. ISMAIL
118
INTRODUCTION TO CIRCULAR DICHROISM
the observed solution CD is only a small percentage of the total crystal signal.
The obvious solution, namely crystal packing effects, has been eliminated as a
possible explanation. The dilemma is probably resolved by realizing that the
above equations give the CD intensity for a D3d environment about the metal. In
practice, the symmetry is lower and other terms contribute to the CD of each
transition; however, these other contributions are equal and opposite for the A2
and E bands so cancel due to the small energy difference of the two bands (4).
where (n1 — nr) is called the circular Inrefringence and A and d have the same units
(usually dm). OR is usually measured by determining the rotation of linearly
polarized light upon passing through the solution. If the linearly polarized light
is rotated clockwise (viewed looking into the light source), the OR is called a
positive or right (dextro) OR. Optical rotation as a function of \ is called ORD
(see above). ORD is an S-shaped curve centred at the CD maximum (so-called
anomalous ORD). For a positive CD band the long-wavelength side of the ORD
curve is a large positive ORD contribution. The short-wavelength side is less
positive or negative. ORD is also non-zero away from absorption bands. Hence
xD values (the ORD at the sodium D line) may be used to characterize the
enantiomeric excess of a solution.
4 Units of CD spectroscopy
The relationship between CD signal (in absorbance units) and sample concentra-
tion and pathlength is analogous to the Beer-Lambert law for absorbance,
namely:
119
ALISON RODGER AND MATTHEW A. ISMAIL
where er is the molar extinction coefficient for the absorption of right circularly
polarized light.
CD is often given in terms of the ellipticity, 0. 0 is obtained from the ratio of
the minor and major axes of the ellipse traced out by the electric field vector of
the elliptically polarized light that emerges from a circularly dichroic sample
onto which linearly polarized light was incident.
where n1 is the absorption index for left circularly polarized light. The wave-
length, A, must be in the same units as J. For small ellipticities
120
INTRODUCTION TO CIRCULAR DICHROISM
where er is the extinction coefficient for the absorption of right circularly polar-
ized light. A related unit that is often used is the mean residue ellipticity. In this
case, as mentioned above, the sample concentration is given in terms of total
numbers of residues (using the average molecular mass of an amino acid resi-
due) without reference to their identity, hence it is an average over the types of
residue present in the macromolecule.
121
ALISON RODGER AND MATTHEW A, ISMAIL
Figure 14 (a) The peptide bond, (b) the primary structure of a protein, and (c) the commonly
occurring L-amino acids. (Source: Circular dichroism and linear dichroism, A. Rodger and B.
Norden, 1997, by permission of Oxford University Press.)
Figure 15 Typical protein CD spectra for particular secondary structural motifs used in CD
protein structure fitting programs.
by the carbonyl ir-bond and is also affected by the involvement of the amide
nitrogen in the -rr-orbitals; its electric dipole transition moment is polarized
somewhere near the line between oxygen and nitrogen and it is centred at
190 nm. In an a-helix (see below), the electric dipole coupling of the TT—tt*
122
INTRODUCTION TO CIRCULAR DICHROISM
5.4.1 a-Helix CD
The (right handed) a-helix is the dominant secondary structure in many pro-
teins and on average accounts for about one third of the residues in globular
proteins. It is a well-defined structural motif where the nth peptide unit forms
hydrogen bonds between its C-O and the N-H of the (n + 4)th peptide and
between its N-H and the (n - 4)th C-O; there is a 0.15 nm translation and 100°
rotation between two consecutive peptide units, giving 3.6 amino acid residues
per turn. Its helix pitch (number of residues times distance between a-Cs on
neighbouring residues) is 0.54 nm.
The CD spectra of a-helices are characterized by a negative band with
separate maxima of similar magnitude at 222 nm (the n-rr* transition) and 208
nm which is part of the rr-TT* transition (see Section 5.3 and Figure 15). The a-
helix is the only motif where the rr—TT* transition has such a long wavelength
component. The a-helix CD is also larger in magnitude than that due to other
motifs so it is apparent upon the most casual inspection of a spectrum. It is no
123
ALISON RODGER AND MATTHEW A. ISMAIL
surprise therefore that the various empirical CD fitting programs that are
available are fairly successful in determining the percentage of a-helical content
of a protein. It should, however, be noted that the magnitude of the CD signal
does vary with variations in the helix, and the helix length (the signal behaves
as if it were for a helix about four residues shorter than it in fact is, which
corresponds to the number of unanchored hydrogen-bonding groups) and the
interactions with neighbouring structural units.
5.4.4 B-sheet CD
An alternative efficient formation of hydrogen bonds occurs between a sheet of
parallel or antiparallel runs of amino acids; these are known as a p-sheets. The
runs of amino acids face in alternate directions so that alternate amino acids
hydrogen bond to neighbouring runs on each side. The spectroscopic character-
ization of p-sheets has proved more difficult than that of a-helices due to the
practical reason that they are less soluble in solvents with a good UV trans-
mission, and due to the intrinsic reason that they are generally structurally less
well-defined: they may be parallel or antiparallel and of varying lengths and
widths. Furthermore, an extended B-sheet is usually found to show a marked
twist, rather than to be planar. Such tertiary structure influences the overall CD
spectrum.
The general characteristics of B-sheet CD may be taken to be a negative band
at about 216 nm and a positive band of comparable magnitude near 195 nm
(13). This level of characterization of the spectrum might be deemed to be not
much worse than that of a-helices were it not for the fact that what is often
called the 'random coil' but is more correctly referred to as 'other' structural
features (see below) have their CD maxima at similar wavelengths and are often
of opposite sign from those of the B-sheet. This means that an empirical fitting
program may incorrectly weight these spectra (and the B-turn components) in
an attempt to better account for the wavelength and magnitude variations that
occur.
5.4.5 B-Turn CD
Typically the strands of an antiparallel B-sheet are linked by B-turns where the
nth peptide unit forms hydrogen bonds with the (n + 3)rd peptide unit. How-
ever, the label B-turn is usually used to include all possible turns that occur, not
simply the ones that enable a single strand to become an antiparallel B-sheet.
About one quarter of the residues in globular proteins then fall into this
124
INTRODUCTION TO CIRCULAR DICHROISM
125
ALISON RODGER AND MATTHEW A, ISMAIL
Protocol 2
Application of CDsstr to compute protein secondary
structure from CD spectra
Equipment and materials
• A PC running Windows 95/98 or NT4 • CD data for the proteins to be analysed
• A copy of the CDsstr program
126
INTRODUCTION TO CIRCULAR DICHROISM
Protocol 2 continued
from the 'Programs' menu in the 'Start' menu. If'Command Prompt' is not present
in the 'Programs' menu, open a DOS window by selecting 'Run' from the 'Start'
menu and typing 'c:\windows\command.com' in the box then clicking 'OK'.*b
7 At the command prompt type 'c:' then 'CD\CDsstr',
8 Type 'copy proCD.tst proCD.dta'.
9 Type 'CDsstr' to initiate the test,
10 Enter the value of each variable as prompted: NbasCD = 22; Nwave = 83; Npro = 3;
ncomb = 100; icombf = 100000.
11 When the command prompt reappears, compare the values in the files anal.out
and reconCD.out with those in anal.tst and reconCD.tst. If CDsstr is functioning
correctly they should exactly match.
B. Using CDsstr for protein CD analysis
1 Delete, rename, or move any file with an .out filename extension remaining in the
CDsstr folder.
2 Delete any previously used file named proCD.dta unless you wish to use it in the
current run.
3 If it is not already available, prepare an input file containing the CD data of the
protein(s) to be analysed. Enter a title on the initial line then enter the CD data
as is (for molar residue concentration) with each value on a new line. Begin at
260 nm and continue in 1 nm increments down to 178 nm. Ensure that the
numerical values entered comply with the Fortran 1F10.2 format. If multiple
protein data sets are being analysed, begin each new set with a title line. Save the
file as c:\CDsstr\proCD.dta.c
4 If using truncated protein CD data (truncation above 200 nm is not recommended),
the basis set protein CD spectra in the basCD.dta file must also be truncated by
removing data points from each data set to match the experimental data set size,c
5 Determine appropriate values (see above) for the variables NbasCD, Nwave, Npro,
Ncomb and icombf.
6 Begin the analysis by opening a DOS window (see Section A above). Type 'c:'. Then
type 'CD\CDsstr' at the command prompt.
7 Enter values for the program variables as prompted.
8 When the command prompt reappears, view the results of the analysis by in-
specting the output files anal.out and reconCD.out.
a
Do not place this folder within any subdirectory if you wish to follow this protocol exactly.
b
CDsstr must be run in MS-DOS from within Windows, Restarting the computer in MS-DOS
mode will not permit the program to run.
c
A commercial spreadsheet is well suited to editing CD data files. Most CD data analysis
software permits a data file to be exported as text. The text file can then be imported into the
spreadsheet, easily formatted for CDsstr input and saved as undelimited ASCII text. Delete,
rather than cut, data cells to remove unwanted text/data.
127
ALISON RODGER AND MATTHEW A. ISMAIL
Upon execution, CDsstr prompts the user for the value of several program
variables. NbasCD is the number of basis set CD spectra and should be set to 22
if the default basCD.dta and secstr.dta files are being used. Nwave is the number
of wavelengths contained in basCD.dta and is 83 for the default basis set (i.e.
with data down to 178 nm). The number of protein CD spectra contained in the
proCD.dta file, Npro, should then be entered and must lie between 1 and 100
(3 for the test file proCD.tst). The value for the number of successful combina-
tions to be considered, ncomb, should usually be set to 100 except for alpha
helical proteins where the program's maximum value of 400 should be used. The
final requested parameter value is the maximum number of trial combinations
allowed before the program stops the analysis, icombf. A value of 100 000 is
usually appropriate for this parameter. The calculated secondary structure out-
put is written to the file anal.out along with the input data. A second output file,
reconCD.out contains the average reconstructed CD spectrum.
Peptide CD structure analysis differs from that of proteins as peptides are
usually a combination of a particular secondary structure and random coil (i.e.
unordered) residues. Proteins seldom if ever have random coils, so the CD of
the proteins in the basis set of the protein analysis program has no random
coil component. Further, the analysis of the CD of a peptide is not under-
determined, and so does not require a flexible method like Variable Selection. A
Convex Constraint Analysis (22) that extracts the component spectra has been
developed into a peptide structure analysis program by Greenfield (14).
128
INTRODUCTION TO CIRCULAR DICHROISM
129
ALISON RODGER AND MATTHEW A. ISMAIL
Figure 16 la) B-, A- and Z-DNA viewed perpendicular to the helix axis and down the helix
axis, (b) Structural formula of DMA illustrating the definitions of torsion angles. (Source:
Circular dichroism and linear dicrimism, A. Rodger and B. Norden, 1997, by permission of
Oxford University Press.)
no
INTRODUCTION TO CIRCULAR DICHROISM
Table 1 Typical geometric parameters (see Figure 8 for definitions) for standard A-, B- and Z-
forms of DNA (30-34)
RNA differs from DNA in that every T is replaced by U, which has one less methyl group, and the backbone sugar
is ribose instead of deoxyribose. Double helical RNA adopts the A-form geometry. The repeat unit of Z-DNA is a
dinucleotide, necessitating two values of each angular parameter. adopts a range of values in Z-DNA.
131
ALISON RODGER AND MATTHEW A. ISMAIL
132
INTRODUCTION TO CIRCULAR DICHROISM
-20
210 230 250 270 290 310 220 240 260 280 300 320
Wavelength / nm Wavelength / nm
Figure 20 (a) Poly[d(A-T)]2 (44 uM; 1 mM phosphate) in B-form (20 mw NaCI) and condensed
(1 M NaCI and 45% v/v ethanol). (b) Poly[d(G-C)]2 as a function of increasing spermine,
[NH3(CH2)3NH2(CH2(CH2CH2)3NH3], concentration showing the B - Z transition. Spermine
concentrations are indicated in the figure. (Source: Circular dichroism and linear dichroism,
A. Rodger and B. Norden, 1997, by permission of Oxford University Press.)
from 180-190 nm. When B-form DNA is stretched to have 10.2 bases per turn
instead of 10.4, the 275 nm peak essentially disappears. This form of DNA can
be induced by methanol or by wrapping the DNA around histone cores to form
nucleosomes.
If B-DNA is compacted and the bases tilted and radially displaced from the
centre of the helix (thus creating a hole when one looks down the helix axis)
(Figure 16 and Table 1) then A-form DNA results. A-DNA is characterized by a
133
ALISON RODGER AND MATTHEW A. ISMAIL
134
INTRODUCTION TO CIRCULAR DICHROISM
ligand:DNA
0.008:1
0.015:1
0.022:1
0.029:1
0.037:1
20
-60
220 240 260 280 300 240 260 280 300
Wavelength / nm Wavelength / nm
Figure 21 ICD of varying concentrations of anthracene-9-carbonyl-N1-spermine upon
interaction with (a) 160 uM poly[d(G-C)]2, 5 mM NaCI and low ligand:DNA:phosphate mixing
ratios, and (b) 130 mm poly[d(A-T)]2, 5 mM NaCI and a wide range of ligand:DNA:phosphate
mixing ratios. / = 1 cm. The smallest magnitude ICD signal corresponds to the lowest ligand
concentration. (Source: Circular dichroism and linear dichmism, A. Rodger and B. Norden,
1997, by permission of Oxford University Press.)
where Lf is the concentration of free ligands, and Sf is the free site concen-
tration. The total site concentration, Stot is given by
135
ALISON RODGER AND MATTHEW A. ISMAIL
To perform a Scatchard plot (48) to analyse the data to give K directly, re-
arrange equation 31 as follows. If we know a (the proportionality constant, see
equation 30) then from ICD spectra we determine Ib and hence Lf
35
where
36
Thus, a plot of r/Lf versus r has slope -JC and y-intercept K/n. The x-intercept
occurs where r = 1/n. The Scatchard plot is a better way of averaging over a data
set than, for example, an arithmetic mean from direct application of equation 31
for a number of data points. The problem with the Scatchard plot is that one
needs to know the value of a in equation 30 to determine Ib, and hence Lf from
the ICD. Ensuring all the ligand in a solution is bound as required for a direct
determination of a is often impossible with DNA. The 'Intrinsic method' was
developed to solve this problem (Figure 22).
We write
Thus, for two different total ligand concentrations, LJtot and Lk t , but the same
macromolecule concentration, i.e. SJot = Stot,
Figure 22 Graph illustrating application of the intrinsic method for determining a and n using
the data given in Figure 21. K has been determined using an average value from repeated
application of equation 37. Alternatively, the data could have been used to perform a
Scatchard plot, x and y are as defined in equations 40 and 41. (Source: Circular dichroism
and linear dichroism, A. Rodger and B. Norden, 1997, by permission of Oxford University
Press.)
136
INTRODUCTION TO CIRCULAR DICHROISM
137
ALISON RODGER AND MATTHEW A. ISMAIL
the DNA increases. Therefore the above methods for determining K cannot be
used as equation 30 does not hold. We may also be observing changes in the DNA
conformation or to ligand-ligand interactions. A particularly dramatic change
in the ligand ICD is observed if as the DNA concentration is decreased (so one
would expect fewer ligands to bind and so a decrease in ICD), the ligands begin to
stack together and the ICD increases. Under these circumstances, we generally
observe a large excitonic CD. Thus spectra such as those of Figure 23 immedi-
ately tell us that significant ligand-ligand interactions are occurring on the
DNA.
Acknowledgement
This text has been adapted from Circular Dichroism and Linear Dichroism by
Alison Rodger and Bengt Norden (1997) by permission of Oxford University
Press.
References
1. Craig, D. P. and Thirunamachandran, T. (1984). Molecular quantum electrodynamics: An
introduction to radiation-molecule interaction. Academic Press, London.
2. Barren, L. D. (1982). Molecular light scattering and optical activity, p. 1. Cambridge
University Press, Cambridge
3. Grandison, A., McGinley, D., Shearer, T., Knight, L., Summers, E., Lyons, C. and
Forde, C. (ed.) (1994). Collins English dictionary (3rd edn), p. 284. HarperCollins
Publishers, Glasgow.
4. Rodger, A. and Norden, B. In Circular and linear dichroism (ed. R. G. Compton, S. G.
Davies, and J. Evans), p. 1. Oxford University Press, Oxford.
5. Bosnich, B. (1969) Ace. Chem. Res., 2, 266.
6. Moffitt, W., Woodward, R. B., Moscowitz, A., Klyne, W. and Djerassi, C. (1961) J. Am.
Chem. Soc., 83, 4013.
7. Hohn, E. G., and Weigang, O. E. (1968) J. Chem. Phys. 48,1127.
8. Rodger, A. and Rodger, P. M. (1988). J. Am. Chem. Soc., 110, 2361.
9. Lightner, D. A., Bouman, T. D., Wijekoon, W. M. D. and Hansen, A. (1985) J. Am. Chem.
Soc. 89, 5805.
10. Fidler, J., Rodger, P. M., and Rodger, A. (1994) J. Am. Chem. Soc. 116, 7266.
11. Schipper, P. E. and Rodger, A. (1986). Chem. Phys., 109, 173.
12. Fasman, G. D. (ed.) (1996). Circular dichroism and the conformational analysis of
biomolecules, p. 1. Plenum Press, New York.
13. Greenfield, N. and Fasman, G. D. (1969). Biochemistry, 8, 4108.
14. Greenfield, N. J. (1996). Anal. Biochem., 235, 1.
15. Johnson, W. C. Jr. (1999). Proteins: Structure, Function and Genetics, 35, 307.
16. Hennessey, J. P. Jr. and Johnson, W. C. Jr. (1981). Biochem., 20, 1085.
17. Compton, L A. and Johnson, W. C. Jr. (1986). Anal. Biochem., 155, 155.
18. Manavalan, P. and Johnson, W. C. Jr. (1987). Anal. Biochem., 167, 76.
19. Sreerama, N. and Woody, R. W. (1993). Anal. Biochem., 209, 32.
20. van Stokkum, I. H. M., Spoelder, H. J. W., Bloemendal, M., van Grondelle, R. and
Groen, F. C. A. (1990). Anal. Biochem., 191, 110.
21. Dalmas, B. and Bannister, W. H. (1995). Anal. Biochem., 225, 39.
22. Perczel, A, Park, K. and Fasman, G. D. (1992). Anal. Biochem., 203, 83.
23. Shiraki, K., Nishikawa, K. and Goto, Y. (1995). J. Mol. Bio!., 245,180.
138
INTRODUCTION TO CIRCULAR DICHROISM
24. Aggell, A., Bell, M., Boden, N., Keen, J. N., Knowles, P. F., McLeish, T. C. B., Pitkeathly,
M. and Radford, S. E. (1997). Nature, 386, 259.
25. Bustamante, C., Tinoco, I., and Maestre, M. F. (1983). Proc. Natl. Acad. Sci. USA, 80,
3568.
26. Gordon, D. J. and Holzworth, G. (1971). Arch, Biochem. Biophys., 142, 481.
27. Mao, D. and Wallace, B. A. (1984). Biochemistry, 23, 2667.
28. Castiglione, E. and Rodger, A. Unpublished data
29. Rankin, S. E., Watts, A. and Pinheiro, T. J. (1998). Biochemistry, 37,12588.
30. Egli, M., Williams, L. D., Gao, Q. and Rich, A. (1991), Biochemistry, 30,11388.
31. Stryer, L. (1988). In Biochemistry (3rd edn), p. 1. W. H. Freeman and Company, New
York.
32. Calladine, C. R. and Drew, H. R. (1992). Understanding DNA: The molecule and how it
work, p. 1. Academic Press, Cambridge.
33. Beveridge, D. L. and Jorgensen, W. L. (1986). Ann. N.Y. Acad. Sci., 482.
34. Gessner, R. V., Fredrick, C. A., Quigley, G. J., Rich, A. and Wang A. H.-J. (1989). ]. Biol.
Chem., 264, 7921.
35. Zaloudek, F., Novros, J. S. and Clark, L. B. (1985). J. Am. Chem. Soc., 107, 7344.
36. Matsuoka, Y. and Norden, B. (1994). J. Phys. Chem., 86,1378.
37. Clark, L. B. (1994). J. Am. Chem. Soc., 116, 5265.
38. Fulscher, M. P. and Roos, B. O. (1995). J. Am. Chem. Soc., 117, 2089.
39. Holmen, A., Broo, A., Albinsson, B. and Norden, B. (1997)J. Am. Chem. Soc., 119,
12240.
40. Parkinson, A. (1998) PhD Thesis, University of Warwick.
41. Williams, A. L Jr., Cheong, C., Tinoco, I. Jr. and Clark L. B. (1986). Nucleic Adds Res.,
14, 6649.
42. Rizzo, V. and Schellman, J. A. (1984) Biopolymers, 23, 435
43. Lyng, R., Rodger, A. and Norden, B. (1992) Chem. Phys. Lett., 70, 17
44. Norden, B. and Tjerneld, F. (1976). Biophys. Chem., 4,191.
45. Dieber, H., Secco, F. and Venturini, M. (1987). Biophys. Chem., 26, 193.
46. Fronaeus, S. (1950). Acta Chem. Scand., 4, 72.
47. Rodger, A. (1993). In Methods in enzymology, Vol. 226, p. 232. Academic Press, London.
48. Scatchard, G. (1949). Ann. N. Y. Acad. Sci., 51, 660.
General CD references
• Barron, L. D. (1982). Molecular light scattering and optical activity. Cambridge University
Press, Cambridge.
• Craig, D. P. and Thirunamachandran, T. (1984). Molecular quantum electrodynamics: An
introduction to radiation-molecule interaction. Academic Press, London.
• Harada, N. and Nakanishi, K. (1983). Circular dichroic spectroscopy: exciton coupling in organic
stereochemistry. University Science Books, California.
• Mason, S. F. (1982). Molecular optical activity and the chiral discrimination. Cambridge
University Press, Cambridge.
• Michl, J. and Thulstrup, E. W. (1986). Spectroscopy with polarized light. VCH, New York.
• Nakanishi, K., Berova, N. and Woody, R. W. (ed.) (1994). Circular dichroism: Principles and
applications. VCH, New York.
• Richardson, F. S. (1979). Theory of optical activity in the ligand-field transitions of
chiral transition metal complexes. Chem. Rev., 79, 17.
• Rodger, A. and Norden, B. In Circular and linear dichroism (ed. R. G. Compton, S. G.
Davies, and J. Evans). Oxford University Press, Oxford.
139
This page intentionally left blank
Chapter 5
Quantitative determination of
equilibrium binding isotherms
for multiple ligand-
macromolecule interactions
using spectroscopic methods
Wlodzimierz Bujalowski and Maria J. Jezewska
Dept. of Human Biological Chemistry & Genetics and The Sealy Centre for
Structural Biology, The University of Texas Medical Branch at Galveston,
Medical Research Building, 301 University Boulevard, Galveston,
TX 77555-1053, USA
1 Introduction
Thermodynamic studies provide information that is necessary in order to under-
stand the forces that drive the formation of ligand-macromolecule complexes.
Knowledge of the energetics of these interactions is also indispensable for
characterization of functionally important structural changes that occur within
the studied complexes. Quantitative examination of the equilibrium interactions
are designed to provide the answers to the questions: What is the stoichiometry
of the formed complexes? How strong or how specific are the interactions? Are
there any cooperative interactions among the binding sites and/or the bound
ligand molecules? Are the binding sites intrinsically heterogeneous? What are
the molecular forces involved in the formation of the studied complexes, or, in
other words, how do the equilibrium binding and kinetic parameters depend on
solution variables (temperature, pressure, pH, salt concentration, etc.)?
Equilibrium isotherms for the binding of a ligand to a macromolecule repre-
sent the relationship between the degree of ligand binding (moles of ligands
bound per mole of a macromolecule) and the free ligand concentration. A true
thermodynamic binding isotherm is model-independent and reflects only this relationship.
Only then, when such an isotherm is obtained, can one proceed to extract physic-
ally meaningful interaction parameters that characterize the free energies of
interaction. This is accomplished by comparing the experimental isotherms to
theoretical predictions based on specific binding models that incorporate known
molecular aspects, such as intrinsic binding constants, cooperativity parameters,
allosteric equilibrium constants, discrete character of the binding sites or over-
lap of potential binding sites, etc. (see below).
141
WLODZIMIERZ BUJALOWSKI AND MARIA J. JEZEWSKA
142
MACROMOLECULE INTERACTIONS USING SPECTROSCOPIC METHODS
143
WLODZIMIERZ BUJALOWSKI AND MARIA J. JEZEWSKA
The thermodynamic basis for these methods is that the total average degree
of binding, 2v(, of the ligand on a macromolecule, including all different dis-
tributions of ligands bound in possible different states 'f , is a sole function of the
free ligand concentration, LF at equilibrium (12, 33). Therefore, at a given free
ligand concentration the average value of Dv( will be the same, independent of
the concentration of the macromolecule, MT. The unique values of LF and 2vf and
the corresponding total ligand, LT, and total macromolecule, MT, concentrations
must satisfy the mass conservation equation
Therefore, if the set of concentrations (I/n, MTi) can be found for which tv{
and LF are constant, then 2v( and LF can be determined from the slope and
intercept, respectively, of a plot of LT versus MT, based on equation I.
where SF is the molar signal of the free macromolecule and Si is the molar signal
of the complex, Mf , which represents the macromolecule with i bound ligands
(i = 1 to n).
The mass conservation equation, which relates MF and Mj to MT, is given by
The partial degree of binding, vi ('i' moles of ligand bound per mole of macro-
molecule), corresponding to all complexes with a given number 'i' of bound
ligand molecules is given by
144
MACROMOLECULE INTERACTIONS USING SPECTROSCOPIC METHODS
and
145
WLODZIMIERZ BUJALOWSKI AND MARIA J. JEZEWSKA
escence is substantially quenched upon binding TNP-ATP, hence this signal was
used to monitor the interaction. Therefore, in this case, the observed fractional
protein fluorescence quenching, iSnhs (MBDF) at a given T.NP-ATP concentration,
is given by (see equation 7)
where iSimjl = iS,/i is the average degree of protein fluorescence quenching per
bound TNP-ATP, when T nucleotide molecules arc bound per DnaR hcxnmer.
Protocol 1
Fluorescence titration of the DnaB helicase with TNP-ADP
Equipment and reagents
• Stock solution of the DnaB helicase (~5 x - TNP-ATP
10-5 M) * Fluorescence cuvette with an optical path
• Storage buffer (50 mM Tris/HCI,pH 8.1, of 0.5 cm
300 mM NaCl, 5 mM MgCLj, 25% glycerol) • Spectronuorimeter
• Binding buffer (50 mM Tris/HCl, pH 8.1,
100 mM NaCl, 5 mM MgCl2,10% glycerol)
Method
1 Diaiyse the stock solution of the DnaB helicase (~5 x 10-5 M (hexamer)) from the
storage buffer (50 mM Tris/HCi, pH 8.1, 300 mM NaCl, 5 mM MgCl2, 25% glycerol) to
the binding buffer (50 mM Tris/HCl, pH 8.1,100 mM NaCl, 5 mM MgCl2,10% glycerol),
The final concentration of the protein stock in the binding buffer will be
approximately 3 x 10"5 M (hexamer}. Measure the exact DnaB concentration using
the extinction coefficient at 280 nm, e280 = 185 000 M-1 cm -1 (hexamer)(21,22). After
dialysis, always store protein solutions on ice.
2 Prepare two stock solutions of TNP-ATP. One solution having the nucleotide con-
centration 1 x 10-4 M and the second solution having the concentration 1 x 10-3 M.
Using two (or more) solutions of the nucleotide allows you to add accurate volumes
of TNP-ATP to the protein sample and to perform titrations over a large range of
nucleotide concentrations.
3 Use a fluorescence cuvette with an optical path of 0.5 cm. This will allow you to use
a small volume (450 ul) of the protein solution and decrease the corrections for the
inner filter effect, due to the increasing absorption (at both excitation and emission
wavelengths) as a result of the increasing concentration of TNP-ATP as the titration
progresses (see below).
4 Set the excitation wavelength at 300 nm (the red edge of the tryptophan absorption
spectrum). Excitation at 300 nm will additionally decrease the inner filter effect. Set
146
MACROMOLECULE INTERACTIONS USING SPECTROSCOPIC METHODS
Protocol 1 continued
147
WLODZIMIERZ BUJALOWSKI AND MARIA J. JEZEWSKA
Figure 1 Fluorescence titration of the DnaB hexamer with TNP-ATP monitored by the
fractional quenching of the protein fluorescence (macromolecular binding density function,
MBDF) in 50 mw Tris/HCI (pH 8.1, 10 °C) containing 20 ITIM NaCI, 5 mM MgCI2, and 10%
glycerol at two different concentrations of the DnaB protein (•) M1 = 5.38 x 10-7 M
(hexamer); (o) M2 = 3.17 X 10-6 M (hexamer). The solid lines are interpolations between
data points added to separate the two data sets. The dashed horizontal line connects the
points, at the same value of the MBDF, for each protein concentration. These points
determine the set of total TNP-ATP concentrations, (L1. and L2), for which the average binding
density, 2v, and the free protein concentration, Lf, are constant and independent of the total
protein concentrations, M,. Reprinted with permission from Bujalowski, W. and Klonowska,
M. M. (1993). Biochemistry, 32, 5888-5900. Copyright (1993) American Chemical Society.
ing process is complex, i.e. these spectroscopic isotherms indicate the existence
of two binding phases (see below). Also, depicted in Figure 1 is the approach by
which a single set of values of 2vf and LF can be obtained from these data. A
horizontal line which intersects both curves and which defines one constant
value of ASobs is drawn. The point of intersection of this horizontal line with
each titration curve defines the set of values of LTi, MTi for which LF and Evf are
constant, as discussed above. The example in Figure 1 only has two titrations,
hence only two sets of values of LTi, MTjare obtained (i = 1 and 2).
For the two titrations at macromolecular (DnaB hexamer) concentrations,
MT1 and M^ (MT2 > MT1), two mass conservation equations can be written in the
form of equation 1 for each set of LTi, MTj. These two equations can then be solved
for Xvj and LF, with the results given by
148
MACROMOLECULE INTERACTIONS USING SPECTROSCOPIC METHODS
upon the used concentrations of the macromolecule and the ligand. Thus, the
most accurate estimates are obtained in the region of the titration curves where
the concentration of a bound ligand is comparable to its total concentration, IT-
In other words, the concentration of the bound ligand must constitute a
significant fraction of the total ligand concentration in solution. In practice, this
limits the accurate determination of the degree of binding, 2v;, to the region of
the titration curves where the concentration of the bound ligand is at least
-10% of the IT. Therefore, a selection of proper concentrations of the macro-
molecule is crucial for obtaining LF and Svf over the largest possible region of
the titration curves, although the accuracy of the determination of 2v< is mostly
affected in the region of the high concentrations of the ligand approaching the
maximum saturation. Such a selection of macromolecule concentrations is
usually based on preliminary titrations which provide initial estimates of the
expected affinity between the ligand and the macromolecule.
In the first step of the analysis of a multiple ligand binding system, the
obtained values of 2vj, corresponding to given values of ASobs, are used to:
(1) determine the maximum stoichiometry of the macromolecule-ligand com-
plex, and
(2) determine the relationship between the signal change, ASobs, and the
average number of bound ligand molecules, i.e. in the considered case, the
average number of bound TNP-ATP molecules per the DnaB hexamer.
Both objectives can be achieved by plotting ASobs as a function of Svf. The de-
pendence of the quenching of the DnaB fluorescence upon the rigorously
determined number of TNP-ATP molecules bound per hexamer is presented in
figure 2. The selected concentrations of the DnaB helicase allowed us to obtain
an accurate (± 5 %) estimate of the average degree of binding up to 5.2 ± 0.3
TNP-ATP molecules bound. Although the maximum value of 2vf cannot be
directly determined, due to the discussed inaccuracy at the high ligand con-
centration region, the plateau of the fluorescence quenching ASmax, corres-
ponding to the maximum saturation can be determined with the accuracy of
±5% (see Figure 1). Therefore, knowing the maximum extent of the protein
fluorescence (ASmax = 0.78) one can perform a short extrapolation of the plot
(ASobs versus 2vf) to this maximum value. Such extrapolation of the data
presented in Figure 2 shows that, at saturation, the DnaB protein hexamer binds
six molecules of TNP-ATP, thus, establishing the maximum stoichiometry of the
DnaB-nucleotide complexes (14).
Notice, the plot of ASobs, as a function of 2vj, shown in Figure 2 is strongly
non-linear. The largest protein fluorescence quenching, up to —0.21 ± 0.03,
occurs upon binding the first TNP-ATP molecule. Average binding of the first
three molecules causes quenching of -0.55 ± 0.03, which corresponds to the
first step in the binding isotherm (see Figure 1). Binding of the next three
nucleotides, in the lower affinity phase, gives the maximum saturation of six
TNP-ATP molecules per DnaB hexamer and is accompanied by an additional
quenching of only -0.21 ± 0.05. The dashed line represents a hypothetical
149
WLODZIMIERZ BUJALOWSKI AND MARIA J. JEZEWSKA
Figure 2 The dependence of the fractional quenching of the DnaB protein fluorescence upon
the total average number of TNP-ATP molecules bound per DnaB hexamer in 50 mM Tris/HCI
(pH 8.1, 10°C) containing 20 mw NaCI, 5 mM MgCI2, and 10% glycerol (•). The average
number of nucleotides bound, at a particular value of the fluorescence quenching, has been
determined from the two titrations shown in Figure 1, using equations 9a and 9b. The
dashed line represents the theoretical situation where there is strict proportionality between
the degree of nucleotide binding and the quenching of the DnaB protein fluorescence. The
solid line is generated using the values of parameters a, 6 and c (see text) obtained from
the non-linear least-square fit, based on the third degree polynomial described by equation
15, with intrinsic binding constant K = 5.9 x 106 W1 and cooperativity parameter, o- =
0.55. Reprinted with permission from Bujalowski, W. and Klonowska, M. M. (1993).
Biochemistry, 32, 5888-5900. Copyright (1993) American Chemical Society.
150
MACROMOLECULE INTERACTIONS USING SPECTROSCOPIC METHODS
intercept and the slope. If a plot of IT versus MT> determined in this manner, is
not linear, this may indicate one or more inconsistent sets of titration data or
aggregation phenomena associated with the ligand or the macromolecule, and
these data should be viewed with caution.
dlnIF
which provides the expression for Svf as
2i>i= [6x + 6(3 + 2<y)x2 + 6(1 + 6cr + 3o-2)x3 + 12(3cr2 + 2cr3)x4 + 30a4x5 + 6cr6x6]/Z
12
Because the rigorous thermodynamic approach described above allowed us
to obtain both 2vf and IF, we can use equation 12 to directly obtain K and a, from
the thermodynamic isotherm (Sv, as a function of LF), and to perform the optim-
ization of the binding parameters, K andCT,using a non-linear least-square fit.
Figure 3(a) shows the dependence of 2v; upon the logarithm of [TNP-ATP]Free.
151
WLODZIMIERZ BUJALOWSKI AND MARIA J. JEZEWSKA
Figure 3. (a) The total average number of nucleotide molecules bound per DnaB hexamer,
2v,, as a function of the free concentration of TNP-ATP in 50 mM Tris/HCI (pH 8.1, 10 °C)
containing 20 mM NaCI, 5 mM MgCI2 and 10% glycerol (•). The solid line is a non-linear
least-square fit, according to the hexagon model (equations 10-12) which provides the
intrinsic binding constant K = 5.9 x 105 M'1 and cooperativity parameter,CT= 0.55 (data
from 14). (b) Ruorescence titrations of the DnaB hexamer with TNP-ATP monitored by the
fractional quenching of the protein fluorescence (macromolecular binding density function,
MBDF) in 50 mM Tris/HCI (pH 8.1, 10°C) containing 20 mM NaCI, 5 mM MgCI2 and 10%
glycerol at two different concentrations of the DnaB protein: (•) 5.38 x 10~7 M; (o) 3.17 x
10-6 M (hexamer). The solid lines are computer fits of the experimental binding isotherms,
according to the hexagon model, obtained by the empirical function approach. First, the
fractional protein fluorescence change (MBDF) has been calculated using equation 15, and
using the values of a = 0.286, b = - 0.0484 and c = 3.609 x 10-3. Subsequently, the
fluorescence titration curves were subjected to the non-linear fit which provided the set of
interaction parameters, intrinsic binding constant, K = 5.9 x 105 M-1, and cooperativity
parameter, o= 0.55. Reprinted with permission from Bujalowski, W. and Klonowska, M. M.
(1993). Biochemistry, 32, 5888-5900. Copyright (1993) American Chemical Society.
Notice, the isotherm is more scattered than the original titration curve, and the
isotherm covers —75% of the total binding curve due to the inaccuracy of deter-
mining the degree of binding in the high concentration range of the nucleotide.
The solid line is a non-linear fit of the binding isotherm which provides 1C = 5.9
X 105 M-1 and o = 0.55.
In practice, higher accuracy in estimating the binding parameters could be
obtained if one analyses the entire original fluorescence titration curves (Figure 1),
152
MACROMOLECULE INTERACTIONS USING SPECTROSCOPIC METHODS
without being limited to the region where the degree of binding can be
accurately estimated. A general method which allows such an analysis of the
entire fluorescence titration curves, for this very complex binding system, is
discussed in the next section.
153
WLODZIMIERZ BUJALOWSKI AND MARIA J. JEZEWSKA
curve, all ten optical constants, ASj, must be known. It should be pointed out
that, at the saturating concentration of the ligand, the observed experimental
maximum quenching, is ASobs = AQ6.
The problem of finding all optical parameters can be avoided by using the
following empirical function approach. This approach can be used for any
ligand-macromolecule systems, where the determination of all optical constants
in an analytical equation is practically impossible. An empirical function, usually
polynomial, is found, which relates the experimentally determined dependence
of the average quenching, ASobs upon 2vj, as shown in Figure 2. This can be
achieved by using a non-linear least-square fit of ASobs as a function of 2vf. In the
case of the DnaB-nucleotide interactions, a minimum third degree polynomial
is necessary to describe this function as defined by (14)
ASobs = a(2Vi) + b(Evi,)2 + c(ESvi)3 15
where a, b and c are fitting constants. This function is then used to generate a
theoretical isotherm for a binding model and to extract the true binding
parameters from the experimentally obtained single fluorescence isotherm. To
generate the theoretical isotherm, the value of the degree of binding, Evi, is
calculated using equation 12 for a given free nucleotide concentration and initial
estimates of the intrinsic binding constant, K, and cooperativity parameter, a.
The obtained Evi is then introduced into equation 15 and the fluorescence
quenching, corresponding to this value of the degree of binding, is calculated.
These calculations are performed for the entire fluorescence titration curve.
Using equation 15, we performed the fit of the empirical plot of ASobs as a
function of Evi shown in Figure 2 (solid line) which provided parameters a =
0.286, b = - 0.0484 and c = 3.609 X 10~3. These parameters define the empirical
dependence of ASobs upon Evi. Subsequently, equation 15 with the obtained a, b
and c, combined with equation 12, which defines the selected model of the
nucleotide binding to the DnaB hexamer, is used in the non-linear fit of the
original fluorescence titration curves with only two fitted parameters, K and a.
The solid lines in Figure 3(b) are the non-linear least-square fits of the entire
fluorescence titration curves, according to the procedure described above.
154
MACROMOLECULE INTERACTIONS USING SPECTROSCOPIC METHODS
bound to the macromolecule in any of its 'i' possible bound states and can be
expressed by
ASobs = SFIF + ESi,Li 16
where SF and IF are the molar signal and concentration of the free ligand,
respectively, and Si and Li are the molar signal and concentration of the ligand
bound in state '?, respectively.
Equation 16 is valid when the molar signal of each species is independent of
concentration (i.e. in the absence of ligand and macromolecule aggregation).
The concentrations of the free and bound ligand are related to the total ligand
concentration by the conservation of mass equation
IT = IF + ELi 17a
where
Ii = viMT 17b
where vi is the partial degree of ligand binding for the ith state. Substituting
equations 17a and 17b into equation 16 and rearranging provides
ASobs - SFIT = MT E(S, - SF)vi 18
Notice, SFlT is the initial signal from the ligand before titration with the macro-
molecule. Dividing both sides by SFLT and next multiplying by (IT/MT) yields
(Sobs - SFIT)1/Ir (Si ~ Sp)
=r =2 v, 19
SFIr \\MT
which can be rewritten as
(Sobs "" SFIr) —I
IT I = Vi(AS)jVj
/AO\ ™
20a
and
ASobs = 2(AS),v, 20b
155
WLODZIMIERZ BUJALOWSKI AND MARIA J. JEZEWSKA
At equilibrium, the values of L,F and Ev1, arc constant for a given value of
ASobs(L,/MT), independent of the macromolecule concentration, Mr. Hence,
under identical solution conditions, one can obtain thermodynamically rigorous
measurements of Evi and 1F from plots of ASohs(LT/MT) versus MT for two or more
titrations performed at different total ligand concentrations. This is accom-
plished by obtaining the set of concentrations (L r; . MTl), from each titration, for
which the quantity ASobs(r,T/MT) is constant, and solving for Ev; and l.F. The
procedure is therefore analogous to the case in which a signal from the
macromolecule is monitored during the titration. The quantity, AS(lhs(lT/MT), is
referred to as the Ligand Binding Density Function (LBDF) (11, 24).
Protocol 2
Fluorescence titration of eADP with the DnaB helicase
Equipment and reagents
• Stock solution of the DnaB helicase (-7 x • Binding buffer (50 mM Tris/HCl, pHS.l,
10- 5 M(hexarner)) 1OOmw NaCI, 5 mM MgCl2, 100 mm acry-
• Storage buffer (50 mM Tris/HCl, pH 8.1, lamide, 10% glycerol)
300 mm NaCI, 5 mM MgCl2, 25% glycerol • eADP
• Fluorescence cuvette with an optical path • Spectrofiuorimeter
of 0.5 cm
156
MACROMOLECULE INTERACTIONS USING SPECTROSCOPIC METHODS
Protocol 2 continued
Method
1 Dialyse the stock solution of the DnaB helicase (~7 x 10'5 M (hexamer)) from the
storage buffer (50 mM Tris/HCl, pH 8.1, 300 mM NaCl, 5 mM MgCl2, 25% glycerol) to
the binding buffer (50 mM Tris/HCl, pH 8.1, 100 mM NaCl, 5 mM MgCl2, 100 mM
acrylamide, 10% glycerol). The final concentration of the protein stock in the
binding buffer will be approximately 5 x 10~5 M (hexamer). Measure the exact DnaB
concentration using the extinction coefficient at 280 nm, em, - 185 000 M-1 cm"1
(hexamer} (21), After dialysis, always store protein solutions on ice,
2 Prepare two stock solutions of the protein. One solution having the DnaB con-
centration 1 x 10~6 M and the second solution having the concentration ~5 x 10"5 M.
Using two solutions of the helicase allows you to add accurate volumes of the
protein to the nucleotide sample and to perform titrations over a large range of the
protein concentration,
3 Prepare stock solutions of eADP, from 3 x 10"5 to 3 x 10"4 M. Determine the nucleo-
tide concentration using the absorption coefficient at 294 nm, t-294 = 2900 M"1 cm"1
(37, 38). Using different stock solutions will allow you to change the concentration
of the eADP in the sample by adding the same volume of different nucleotide stock
solutions.
4 Use a fluorescence cuvette with an optical path of 0.5 cm. This will allow you to use
a small volume (450 ^1) of the fluorescent nucleotide solution.
5 Set the excitation wavelength at 325 nm (the red edge of the eADP absorption
spectrum). Excitation at 325 nm. far from the protein absorption maximum, will
eliminate, to a great extent, the correction for the background of the protein
fluorescence. Set the emission wavelength at 410 nm (the maximum of the eADP
emission spectrum).
6 Set the spectrofluorimeter in the ratio mode. The signal coming from the emission
PMT is now divided by the signal from the reference PMT. This mode allows you to
eliminate the artefacts resulting from any fluctuations of the lamp and the possible,
although slow, drift of the lamp intensity during the titration. Add 450 p,l of the
binding buffer to the cuvette and adjust the signal on the emission PMT to read
approximately 0.01 in the ratio mode. This means that the intensity measured by
the emission PMT is approximately 1% of the intensity read by the reference PMT,
7 Set the polarizer in the excitation channel at the vertical position and the polarizer
in the emission channel at 54.7° (magic angle). Performing titrations using 'magic
angle' conditions will eliminate the artefacts in the fluorescence intensity measure-
ments, due to the possible changes in the fluorescence anisotropy of the sample.
These artefacts, which commonly occur, result from the different sensitivities of the
emission PMT for both the vertically and horizontally polarized light. Although this
may not be a problem when the fluorescence intensity of a large DnaB hexamer is
studied, it could induce artefacts when the emission of the relatively small t-ADP
(affected by the binding to the large DnaB hexamer) is examined, as in this case
(32. 42).
157
WLODZIMIERZ BUJALOWSKI AND MARIA J. JEZEWSKA
Protocof 2 continued
8 Take an aliquot of 20 n-1 of the buffer from the cuvette and add a 20-u.1 aliquot of the
eADP stock solution (e.g. 1 x 10~4 M). Mix gently with a 200 u1 automatic
micropipette. The concentration of eADP in the cuvette is now ~4.5 x 10"6 M. Wait
15 min for the temperature to equilibrate to the temperature of the measurement,
e.g.20°C
9 Titrate the EADP solution with the DnaB protein using aliquots of 1 to 20 (ul. The
final volume of the added protein solution at the end of the titration should not
exceed 15% of the total volume of the nucleotide sample.
10 In an independent titration, titrate the buffer with the DnaB protein solution and
determine the contribution of the protein fluorescence to the observed fluorescence
of the eADP sample for each titration point. At high protein concentrations such a
contribution will become significant, in spite of the fact that the excitation is set at
325 nm (far from the protein absorption spectrum), due to the impurities in the
protein sample and the discrete band pass of the excitation and emission
monochromator. This is your background. B.
11 For each titration point calculate the fractional fluorescence increase, ASobs, using
the formula ASobs = (Sobs - SulT}/SEJ,T. For each titration point 'i' correct the value of
the fluorescence intensity, Si, for the dilution and background using the formula,
Sicor = (si - Bi)(Vi /Vo) where Sicor is the corrected value of the fluorescence intensity at
a given point of titration 'i', Si is the experimentally measured fluorescence
intensity, B( is the background, V, is the volume of the sample at a given titration
point, V0 is the initial volume of the sample.
12 Plot the fractional ligand (eADP) fluorescence increase, ASob!i, as a function of the
logarithm of the total DnaB protein concentration.
£o
-7 -6 -5
Log [DnaB (hexamer)]Tola|
Figure 4 Fluorescence titrations ('reverse titrations') of eADP, at different concentrations
of the nucleotide, with the DnaB helicase in 50 mm Tris/HCI (pH 8.1, 10 °C) containing
100 mM NaCI, 5 mm MgCI2 and 100 mm acrylamide (\ex = 325 nm, \em = 410 nm).
The concentrations of the nucleotide are: (•) 4 x 10"6 M; (*) 6 x 10"6 M; (o) 1 x 10~5 M;
(O) 2 x 10~5 M; (D) 3 x 10"5; (•) 4 x 10~5 M. The solid lines are computer fits of the
experimental binding isotherms, according to the hexagon model, using a single set of
binding parameters, K = 4 x 10s M"1, o- = 0.4 and ASmax = 2.35. Reprinted from Jezewska,
M. J. and Bujalowski, W. (1997) 'Quantitative analysis of ligand-macromolecule interactions
using differential dynamic quenching of the ligand fluorescence to monitor the binding.'
Biophysical Chemistry, 64, 253-269. Copyright (1997) with permission from Elsevier Science.
density function, the total degree of binding, Evi, and the free eADP concen-
trations, [eADP]Free' must be the same, thus, allowing for their determination
(see equation 20). As mentioned above, even though all of the 'reverse' titrations
shown in Figure 4 span the same full range of eADP fluorescence increase, they
do not span the same range of the degree of binding as seen from the LBDF plot
in Figure 5. As a result, multiple titrations at different values of LT are required to
span the full binding density range. A horizontal line, which intersects the LBDF
curves is drawn, defining a constant value of the LBDF, ASobs(LT/MT) (dashed line
in figure 5). The points of intersection of the horizontal line with each binding
density function curve determine the set of values (MTi, LTi) for which IF and Evi
are constant, as shown in Figure 5 for one constant value of the LBDF. Based on
equation 1, the average degree of binding, Evi, and Lf, can then be determined
from the slope and intercept of a plot of LT versus MT at each constant value of
ASobs(Lr/Mp). By repeating this procedure for a series of horizontal lines that span
the range of values of ASobs(lr/Mt) as a function of IF, the values of Evi can be
obtained and a binding isotherm can be constructed (11, 24).
In the construction of a series of LBDF plots, as shown in Figure 5, one should
cover as wide a range of ligand concentrations as possible; however, care should
be taken to avoid large changes in the total ligand concentration between two
successive titrations, since this may bias the determination of IF and Evi. In our
experience, six to eight titrations, using successive total ligand concentrations
that differ by a factor of 1.5-2, will generate an accurate set of data.
159
WLODZIMIERZ BUJALOWSKI AND MARIA J. JEZEWSKA
12
Log[DnaB (hexamer)]Tota|
Figure 5 Dependence of the ligand binding density function (LBDF), ASobs(LT/M,T), on the
logarithm of the total DnaB protein concentration, at different total concentrations of the
EADP (LT) in 50 mm Tris/HCI (pH 8.1, 10 °C) containing 100 mM NaCI, 5 mm MgCI2 and 100
mM acrylamide: (•) L1 = 4 x 10-6 M; (o) L2 = 6 x 10-6 M; (•) L3 = 1 x 10~5 M; (n) LA = 2
x ID'5 M; (A) Lg = 3 x 10~5 M; (*) Lg = 4 X 10"5 M. The solid lines are interpolations
between the data points which separate different data sets and have no theoretical basis.
The horizontal dashed line connects points at the same value of the LBDF, at different total
EADP concentrations (L 1 , L2, L3, L4, L5 and L6) and total DnaB protein concentrations (M1
M2, M3, M4, M5 and M6), at which [eADP]Free and the total degree of nucleotide binding, Ev,,
on the DnaB hexamer are the same. Reprinted from, Jezewska, M. J. and Bujalowski, W.
(1997) 'Quantitative analysis of ligand-macromolecule interactions using differential dynamic
quenching of the ligand fluorescence to monitor the binding.' Biophysical Chemistry, 64,
253-269. Copyright (1997) with permission from Elsevier Science.
160
MACROMOLECULE INTERACTIONS USING SPECTROSCOPIC METHODS
-7
Log[eADP] Free
Figure 6 The dependence of the total average degree of binding of eADP on the DnaB
helicase hexamer, as a function of the logarithm of the free nucleotide concentrations
[eADP]Free in 50 mM Tris/HCI (pH 8.1, 10 °C) containing 100 mM NaCI, 5 mm MgCI2 and 100
mM acrylamide. The solid line is the non-linear least-square fit of the binding isotherm,
according to the hexagon model (equations 10-12), which provides intrinsic binding constant
K = 4 X 105 M-1 and cooperativity parameter o- = 0.4. Reprinted from Jezewska, M. J. and
Bujalowski, W. (1997) 'Quantitative analysis of ligand-macromolecule interactions using
differential dynamic quenching of the ligand fluorescence to monitor the binding.' Biophysical
Chemistry, 64, 253-269. Copyright (1997) with permission from Elsevier Science.
bound ligand (LB/lT), then binding isotherms can be constructed with much
greater ease from a titration at a single ligand concentration (see below). The
dependence of the fractional fluorescence increase of AADP as a function of the
fraction of nucleotide molecules bound to the DnaB protein, LB/IT = Ev((MT/IT), is
shown in Figure 7. The value of LB/Lr has been determined using the binding
density function plots shown in Figure 5.
It is clear that, in the studied binding of eADP to the DnaB hexamer using the
fluorescence of eADP to monitor the binding, there is a linear correspondence
between the relative increase of the nucleotide fluorescence, AS, and the
fraction of the ligand bound, IB/IT- in the examined range of the degree of
binding. However, it is important to check the relationship between the
observed fluorescence change (or any signal used to monitor the binding) and
IB/IT over a wide range of binding densities, since the signal change can, in
general, be dependent upon the degree of binding (see Figure 2). In the case of
the E. coli DnaB protein-eADP interactions, this direct proportionality holds for
the degree of binding up to -5.3 (24). Therefore, under these conditions, the
fractional fluorescence increase of eADP is equal to the fraction of the bound
nucleotide, i.e. ASobs/ASmax = Ig/Lj.. When ASobs is directly proportional to IB/IT,
one can determine the maximum extent of protein fluorescence quenching,
ASmax, from a linear extrapolation of a plot of ASobs versus IB/Lr to IB/IT = 1, as
shown in Figure 7. This short extrapolation (dashed line) to LB/LT = 1 gives the
161
WLODZIMIERZ BUJALOWSKI AND MARIA J. JEZEWSKA
2.5
2.0
1.5
1.0
0.50
0.0
0.2 0.4 0.6 0.8
Figure 7 Relationship between the relative fluorescence increase and the fractional
saturation of the nucleotide for sADP binding to the DnaB helicase in 50 mm Tris/HCI (pH
8.1, 10 °C) containing 100 mm NaCI, 5 mM MgCI2 and 100 mM acrylamide. The selected
concentration of sADP is 2 x 10"5 M. The concentration of the bound nucleotide has been
calculated from [eADP]Bound = (Ev,)[DnaB]Totai, where the total average degree of binding of
eADP on the DnaB protein, Ev,, has been determined by using the LBDF approach described
in the text. Reprinted from Jezewska, M. J. and Bujalowski, W. (1997) 'Quantitative analysis
of ligand-macromolecule interactions using differential dynamic quenching of the ligand
fluorescence to monitor the binding.' Biophysical Chemistry, 64, 253-269. Copyright (1997)
with permission from Elsevier Science.
value of the maximum relative increase, ASmax = 2.35, of the eADP fluorescence
upon saturation with the DnaB protein in studied solution conditions (24).
21a
AS
162
MACROMOLECULE INTERACTIONS USING SPECTROSCOPIC METHODS
and
A,S»u_\/L T \
~\ 21c
ASm
Thus, a single titration can be used to obtain Ev; as a function of IF. However, we
stress that one should not simply assume that the fractional signal change is
equal to the fraction of the bound ligand, since, if it is not true, this could lead
to significant errors. On the other hand, if direct proportionality does not exist
between the signal change and the fraction of the bound ligand over a wide
range of binding densities, the true binding isotherm can still be constructed
without any assumptions by using the LBDF analysis (see Figure 6).
3 Summary
Understanding macromolecular interactions, such as those involving ligands
and macromolecules, requires detailed knowledge of the energetics and kinetics
of the formed complexes. Spectroscopic methods are widely used in character-
izing the energetics (thermodynamics) and kinetics of ligand-macromolecule
interactions in solution. These methods are very convenient to use, do not
require large quantities of material, and, most importantly, do not perturb the
studied processes. However, spectroscopic methods are indirect, i.e., the inter-
actions are measured through monitoring changes of some physico-chemical
parameter (e.g. fluorescence, absorbance, circular dichroism) accompanying the
studied complex formation. In such studies, it is indispensable to determine the
relationship between the observed signal and the degree of binding in order to
obtain thermodynamically meaningful interaction parameters. The approaches
presented in this chapter describe general quantitative methods of the analyses
of macromolecular binding through spectroscopic measurements which allow
an experimenter to determine the thermodynamically rigorous degree of bind-
ing or the degree of macromolecule saturation and the free ligand concentra-
tion. The method enables one to obtain the total degree of binding, Evi and the
free ligand concentration, IF, hence, to construct the entire model-independent
binding isotherm. Only when the thermodynamically rigorous isotherm is
obtained, can it be analysed by using the thermodynamic models which incor-
porate the known molecular aspects of the ligand-macromolecule interactions,
like cooperativity, allosteric conformational changes, overlap of potential
binding sites, etc.
We have illustrated the application of the methods to situations where the
signal used to monitor the interactions originates from a macromolecule or a
ligand, using as examples the E. coli replicative helicase DnaB protein inter-
actions with nucleotide cofactors (14, 24).
163
WLODZIMIERZ BUJALOWSKI AND MARIA J. JEZEWSKA
Acknowledgements
We thank Gloria Drennan Davis for help in preparing the manuscript.
References
1. Riggs, A. D., Bourgeois, S., and Cohn, M. (1970). J. AM. Biol., 53, 401.
2. Jensen, D. E. and von Hippel, P. H. (1977). Anal. Biochem., 80, 267.
3. Draper, D. E. and von Hippel, P. H. (1978) J. Mo!. Biol, 122, 321.
4. Garner, M. M. and Revzin, A. (1981). Nucl Acids Res., 9, 3047.
5. Fried, M. and Crothers, D. M. (1981). Nucl. Acids Res., 9, 6505.
6. Cassel, J. M. and Steinhardt, J. (1969). Biochemistry, 8, 2603.
7. Holbrook, J. J. (1973). Biochem. J., 128, 921.
8. Boschelli, F. (1982). J. Mol. Biol, 162, 267.
9. Kelly, R. C, Jensen, D. E., and von Hippel, P. H. (1976). J. Biol. Chem., 251, 7240.
10. Porschke, D. and Rauh, H. (1983). Biochemistry, 22, 4737.
11. Bujalowski, W. and Lohman, T. M. (1987). Biochemistry, 26, 3099.
12. Lohman, T. M. and Bujalowski, W. (1991). Methods in Enzymology, 208, 258.
13. Heyduk, T. and Lee, J. C. (1990). Proc. Notl. Acad. Sci., 87,1744.
14. Bujalowski, W. and Klonowska, M. M. (1993). Biochemistry, 32, 5888.
15. Bujalowski, W. and Klonowska, M. M. (1994). Biochemistry, 33,4682.
16. Bujalowski, W. and Klonowska, M. M. (1994). J. Biol. Chem., 269, 31359.
17. Bujalowski, W. and Jezewska, M. J. (1995). Biochemistry, 34, 8513.
18. Jezewska, M. J. and Bujalowski, W. (1996). J. Biol. Chem., 271, 4261.
19. Jezewska, M. J., Kim, U.-S., and Bujalowski, W. (1996). Biochemistry, 35, 2129.
20. Bujalowski, W. and Porschke, D. (1988). Biophysical Chem., 30,151.
21. Bujalowski, W., Klonowska, M. M., and Jezewska, M. J. (1994). J. Biol. Chem., 269,
31350.
22. Jezewska, M. J., Kim, U.-S., and Bujalowski, W. (1996). Biophysical Journal, 71, 2075.
23. Jezewska, M. J. and Bujalowski, W. (1996). Biochemistry, 35, 2117.
24. Jezewska, M. J. and Bujalowski, W. (1997). Biophysical Chem., 64, 253.
25. Jezewska, M. J., Rajendran, S., and Bujalowski, W. (1997). Biochemistry, 36,10320.
26. Jezewska, M. J., Rajendran, S., and Bujalowski, W. (1998). Biochemistry, 37, 3116.
27. Jezewska, M. J., Rajendran, S., and Bujalowski, W. (1998). J. Biol. Chem., 273, 9058.
28. Jezewska, M. J., Rajendran, S., Bujalowska, and Bujalowski, W. (1998). J. Biol.
Chem., 273,10515.
29. Holbrook, J. J. and Gutfreund, H. (1973) FEBS Lett., 31,157.
30. Witting, P., Norden, B., Kim, S. K., and Takahashi, M., (1994). J. Biol. Chem., 269, 5700.
31. Halfman, C. J. and Nishida, T. (1972) Biochemistry, 11, 3493
32. Lakowicz, J. R. (1983). Principles of fluorescence spectroscopy. Plenum Press, New York.
33. Hill, T. L. (1985). Cooperativity theory in biochemistry. Springer-Verlag, New York.
34. Parker, C. A. (1968). Photoluminescence of solutions. Elsevier, Amsterdam.
35. Bujalowski, W. and Lohman, T. M. (1989). J. Mol. Biol., 207, 249.
164
MACROMOLECULE INTERACTIONS USING SPECTROSCOPIC METHODS
165
This page intentionally left blank
6
Steady-state kinetics
Athel Cornish-Bowden
Institut Federatif 'Biologie Structurale et Microbiologie', CNRS-BIP,
31 chemin Joseph-Aiguier, B.P. 71,13402 Marseille Cedex 20, France
167
ATHEL CORNISH-BOWDEN
regarding the rate as a quantity in its own right it is not usual in steady-state
kinetics to represent it as a derivative such as -da/dt.
A third case that is useful to consider in catalysed reactions is the zero-order
reaction, in which v apparently does not depend on any concentration:
v=k 3
Although this behaviour is apparent rather than real it is useful to consider it
because some of the constants in enzyme kinetics are zero-order rate constants.
In each of the equations above the k is an overall rate constant. However, it is
often convenient to consider the rate with respect to one reactant, the con-
centrations of the others being maintained constant. For example, if the second-
order reaction above is considered in relation to a only, b being regarded as a
constant:
v = kb a 4
then it is evident that the rate is proportional to a, and it is said to be first-order
with respect to a. The 'constant' kb, which is only constant of course if b is
constant, is called a pseudo-first-order rate constant.
Although the individual steps in an enzyme mechanism may all be either
first-order or second-order reactions, the mechanism normally includes several
steps, with the result that the complete reaction rarely has a simple order. The
fundamental problem, therefore, is to set up conditions in which the rate is
easily measurable and then to determine how it varies with the conditions.
2 Units
All of the quantities required in steady-state kinetics are either concentrations,
normally measured in mol litre"1 or a submultiple, rates, measured in mol litre"1
s-1, or rate constants, with units that vary according to the type of rate constant:
a first-order rate constant has dimensions of reciprocal time, and is typically
measured therefore in s"1, and a second-order rate constant has dimensions of re-
ciprocal time multiplied by reciprocal concentration, and is typically measured
therefore in mol-1 litre s-1. It is obvious from elementary dimensional consider-
ations that a pseudo-first-order rate constant has the same dimensions and units
as a first-order rate constant, and that constants of different order cannot be
meaningfully compared.
By far the most common practice is to regard the rate as an intensive
quantity, i.e. to normalize it to a standard volume, so for any given conditions
the rate is the same whether it refers to a 0.5 ml cuvette or to a 100 litre fer-
menter. However, in some specialized applications it is more usual to define the
rate as a rate of conversion, i.e. as an extensive quantity measured in mol s-1. If
this is done, then the dimensions and units of all the rate constants must
change correspondingly. If all the processes occur in a single compartment,
there is not usually any reason to use extensive quantities, but it becomes much
more convenient, if not essential, when one is concerned with transport
168
STEADY-STATE KINETICS
169
ATHEL CORNISH-BOWDEN
over number), and Km, the Michaelis constant. The form shown in the middle is
more fundamental than that on the right, but the second form, in which koe0 is
written as the limiting rate V, is often used because the enzyme concentration is
not always known in meaningful units. V is the limit that v approaches as the
enzyme becomes saturated, i.e. when a becomes very large, and Km is the value
of a at which v = 0.5V, i.e. at which the rate is half-maximal. Notice that ko is a
first-order rate constant, Km is a concentration and V is a zero-order rate con-
stant, or in other words a rate. The ratio ko/Km is a second-order rate constant,
and is called the specificity constant and given the symbol kA: it is a more
fundamental constant than Km in the analysis of enzyme mechanisms (i.e. it has
a simpler mechanistic meaning), but the equation is usually written in terms of
V (or ko) and Km nonetheless.
One may derive equation 5 from the following model,
(C-j K2 k3
A+E - E A - E P - E +P 6
k-, k-2
which assumes that the reaction passes through an enzyme-substrate complex,
EA, that undergoes catalytic transformation to an enzyme-product complex, EP,
which then breaks down to form products. A simpler mechanism containing
only one intermediate also generates the Michaelis-Menten equation, but this is
too simple to be realistic. Conversely, there are infinitely many more complex
mechanisms that also generate it, so although the three-step mechanism pro-
vides a useful basis for discussion one can never be certain from steady-state
measurements alone that the true mechanism is not more complex.
The Michaelis-Menten parameters can be defined as follows in terms of the
rate constants shown in the above reaction:
7
k_2 + k2 + k3
k1,k2k3
8
_
111
Mk.2 + kj + ks)
Note that none of the three parameters has a simple transparent meaning. The
interpretations commonly attributed to them depend on additional simplifying
assumptions that are not always correct. For example, lCm is often said to be
equal to the equilibrium constant k-1/kj for dissociation of A from EA, but the
expression in equation 9 does not take this form unless k2 is very small, and it is
not safe to regard Km as a measure of the equilibrium dissociation constant.
170
STEADY-STATE KINETICS
171
ATHEL CORNISH-BOWDEN
10
V a
11
v
12
a
The double-reciprocal plot, often attributed to Lineweaver and Burk (2), is illus-
trated in Figure 1 and is based on equation 10; it is the mostly widely used
straight-line plot, but it is also the least satisfactory, because it distorts the effect
of experimental error to such an extent that it is difficult to form any visual
judgement of where the best line should be drawn. Either of the other two plots
is better. The plot of a/v against a (equation 11) is associated with Woolf (3) or
Hanes (4) and is illustrated in Figure 2; notice that the definitions of the slope
and intercept on the ordinate are the opposite way round from those that apply
to the double-reciprocal plot. The plot of v against v/a (equation 12) is variously
attributed to Eadie (5) or to Hofstee (6) and is illustrated in Figure 3. It has the
particular advantage that the entire observable range of v values, from 0 to V, is
i/v
+C/
Slope = Km/V
I/a
Figure 1 The double-reciprocal plot. The dashed lines labelled +C and +U illustrate how the
line is displaced by the presence of a competitive or uncompetitive inhibitor, respectively.
172
STEADY-STATE KINETICS
a/v
Slope = 1/V
Figure 2 The plot of a/v against a. Notice that the definitions of the slope and the intercept
on the ordinate are the opposite way round from those in Figure 1. The dashed lines labelled
+C and +U illustrate how the line is displaced by the presence of a competitive or
uncompetitive inhibitor, respectively.
V-
Figure 3 The plot of v against v/a. Notice that this plot maps the entire range of observable
rates (from 0 to V) onto a finite range of paper. The dashed lines labelled +C and +U
illustrate how the line is displaced by the presence of a competitive or uncompetitive
inhibitor, respectively.
plotted in a finite range; this makes it easy to judge by eye if an experiment has
been well designed. On the other hand, it has the disadvantage that v, normally
the less reliable measurement, contributes to both coordinates, and errors in v
cause deviations along lines through the origin rather than parallel with one or
other axis.
These plots are valuable for illustration purposes, but for actual estimation of
kinetic parameters it is more objective to use appropriate computer programs.
For this purpose it is not sufficient just to apply unweighted linear regression to
the straight-line plots, as this suffers from the same statistical distortions as the
plots themselves. Full treatment would require more space than is available
here, but may be found elsewhere (7). The following two equations for calculat-
ing best-fit values of Km and V give satisfactory results if the standard deviation
173
ATHEL CORNISH-BOWDEN
S(v2/a2)2v2 - [S(v2/a)]2
S(v2/a2)Sv - S(v2/a)2(v/a)
15
Km(l + i/Kic) + a
in which i is the concentration of the inhibitor and Kic is the competitive inhibition
constant. (The qualification 'competitive' and the second subscript c are usually
omitted if only this simplest kind of inhibition is being considered.) Competi-
tive inhibition arises when a molecule binds reversibly to the free enzyme in
such a way as to prevent the substrate from binding. Although this occurs most
obviously when the inhibitor is similar enough to the substrate in structure to
replace it in the substrate binding site, but not similar enough to undergo
reaction, competitive inhibition can also arise in more complex ways.
The opposite extreme from competitive inhibition is uncompetitive inhibition,
and is defined by the presence of a factor that affects the variable term in the
denominator of the Michaelis-Menten equation instead of the constant term:
16
Km + o(l + i/Kiu)
The inhibition constant Kiu is an uncompetitive inhibition constant. This type of in-
hibition is not common, but it is important as a component of mixed inhibition,
when both competitive and uncompetitive effects occur simultaneously:
_
V=
K m ( l + i / K i c ) + a(l+t/K i u )
There is no particular reason for the two inhibition constants Kic and Kiu to be
equal, and most of the mechanisms for mixed inhibition suggest that they
ought to be different. However, the case where Kic = Kiu is often given an un-
deserved prominence in discussions of inhibition, largely because experiments
done many years ago suggested that it was a more common phenomenon than
it is. It is called non-competitive inhibition and its rate equation is the same as
equation 17, but with both Kic and Kiu written simply as K1.
All of these kinds of inhibition are conveniently discussed in terms of apparent
Michaelis-Menten parameters, i.e. the parameters that replace the ordinary para-
174
STEADY-STATE KINETICS
m -
i+
Note that the first two expressions have the same form, and both simplify to
independence of i in the event that one or other inhibition term is negligible.
The expression for the apparent value of Km is more complicated, especially
when one considers how it varies with the different types of inhibition: it
increases with the concentration of a competitive inhibitor, it decreases as the
concentration of an uncompetitive inhibitor increases, it may change in either
direction as the concentration of a mixed inhibitor increases, or it is independ-
ent of inhibitor concentration if the inhibition is non-competitive. In general it
is simplest to regard kA as the parameter affected by competitive inhibition,
negligibly so when the competitive component is negligible; k0 as the parameter
affected by uncompetitive inhibition, negligibly so when the uncompetitive
component is negligible; and Km just as the ratio of the two, i.e. Km = Ko/kA.
The effects of the different kinds of inhibition on the common plots are illus-
trated in Figures 1-3 and follow naturally from equations 18-20. Any competitive
effect alters the apparent value of kA: hence it increases the slope of the plot of
1/v against I/a (figure 1), it increases the ordinate intercept of the plot of a/v
against a (Figure 2), and it decreases the abscissa intercept of the plot of v against
v/a (Figure 3). Conversely, any uncompetitive effect increases the ordinate inter-
cept of the plot of 1/v against I/a, increases the slope of the plot of a/v against a,
and decreases the ordinate intercept of the plot of v against v/a. When both
components of the inhibition are present both kinds of effects occur.
7 Specificity
Specificity is a fundamental property of enzymes, but it is often assessed in an
unsatisfactory way, by comparing the kinetic parameters for different reactions
measured in isolation from one another. In the cell, specificity must clearly
refer to the capacity of an enzyme to react selectively with one substrate when
others are present simultaneously, and any satisfactory measure of specificity
should take account of this (8). The equation for reaction of one substrate A in
the presence of a competing substrate A' follows a form similar to that for
competitive inhibition, equation 15:
v= _ _ 21
Km(l + a'/JC'J + a
175
ATHEL CORNISH-BOWDEN
with the inhibitor concentration replaced by a', the concentration of A', and the
inhibition constant by K'm, the Michaelis constant for the reaction of A' con-
sidered in isolation. The rate of reaction of A' is given by the same equation
with an obvious transposition of symbols:
v= _ _ 22
K'm(l + a/KJ + a'
Dividing one equation by the other, the ratio of rates is the ratio of substrate
concentrations multiplied by the ratio of specificity constants:
y_ k0/Km a_ _ kAa
V k'o/K'm'a' k'Ao'
This result explains the choice of the name specificity constant. Equation 23
applies at all concentrations, and the specificity constant measures specificity at
all concentrations. This point is worth emphasizing, because at very low sub-
strate concentrations the rate of an enzyme-catalysed reaction follows second-
order kinetics approximately, and the specificity constant is the second-order
rate constant in these conditions. As this property was well known long before
the relation to specificity was generally recognized it has sometimes generated
an incorrect idea that there is a similar concentration restriction for specificity.
8 Activators
Activation is the opposite effect to inhibition, and an activator is, naturally
enough, a species that increases the rate of an enzyme-catalysed reaction. In
practice, however, activation has been studied much less than inhibition, and
this section will be correspondingly brief. The most important point to make is
that various other phenomena are also sometimes called activation, and these
should not be confused with true activation. For example, certain enzymes, such
as the digestive enzyme pepsin, are initially synthesized as inactive 'zymogens',
pepsinogen in this case, which are transformed into active enzyme by a post-
translational partial proteolysis, a process that is sometimes called zymogen
activation. Another case is really a confusion: for some enzymes, most notably
ATP-dependent kinases, the true substrate is a complex ion containing a metal,
typically MgATP2", and such enzymes are sometimes said to be activated by
Mg2+ ions.
True activation is, as noted, the converse of inhibition, and can be classified
and analysed in similar ways. In general, whenever a term of the form i/Ki occurs
in an inhibition equation one can conceive of a corresponding activation type
where a term Kx/x defines the effect of an activator X. It is important to realize,
however, that the activation analogue of competitive inhibition cannot be
explained by any mechanism in which it makes sense to talk about competition
between substrate and activator, and thus a term like 'competitive activation'
makes no sense even though it is the inverse of competitive inhibition. Such
activation may occur, for example, if the substrate binds only to the enzyme-
176
STEADY-STATE KINETICS
activator complex and not to the free enzyme, so a more meaningful term might
be compulsory activation.
9,1 pH
Mechanistically, most pH effects are inhibition and activation by one particular
inhibitor and activator, the proton, and the equations that describe them are
the same as those for inhibition and activation. There are, however, some im-
portant differences. First, the fact that the proton concentration can be varied
and controlled over a very wide range makes it possible to observe and measure
properties that would be difficult to observe if they occurred with other species.
Second, unlike most inhibitors and activators the proton usually displays both
properties, acting simultaneously as an inhibitor and an activator. This gives
rise to the familiar bell-shaped curve illustrated in Figure 4, which describes the
response of an enzyme to variation in the pH:
k
K
h K2 24
K, + ~h
In this equation k represents some kinetic constant for the reaction, such as
the catalytic constant or the specificity constant, k is a notional pH-independent
0.5 k
5 7 9 11
pH
Figure 4 Bell-shaped pH profile. The parameter k plotted in the ordinate represents either
the catalytic constant ko, or the specificity constant k0/Km, and the limit k is the value it
would have if all of the enzyme existed in the catalytically active state of protonation.
177
ATHEL CORNISH-BOWDEN
form of the same constant that defines the value it would have if all the enzyme
could be fixed in the active state of protonation, h is the hydrogen-ion con-
centration and K1 and K2 are the acid dissociation constants for two successive
deprotonation steps. This is, of course, the simplest possible simultaneous
activation and inhibition, and much more complex cases also occur. It should
be obvious from inspection of the equation that k approaches zero at both low
pH, where h/Kj is large, and at high pH, where K2/h is large, but is larger at
intermediate pH values. If Kt and K2 are well separated it may approximate to k
in this intermediate region, but may be much less.
We may expect both the free enzyme and all intermediate complexes that
occur in an enzyme-catalysed reaction to exist in multiple states of protonation,
and so we can expect any parameter of the reaction to vary with pH. In general
the pH dependence of the specificity constant kA or ko/Km provides information
about the protonation states of the free enzyme, and that of the catalytic
constant k0, provides information about the protonation states of the enzyme-
substrate complex(es). The pH dependence of Km is usually more complicated, as
it involves contributions from all of the different forms of enzyme; it is most
easily analysed by regarding Km as k0/(K0/Km).
9.2 Temperature
Temperature effects on enzymes follow the same principles as those that affect
simple chemical reactions, based on the Arrhenius treatment (9), but in practice
they are very complicated. This is because all enzyme mechanisms involve
multiple steps, and all of the rate constants describing the different steps vary
separately with temperature. It is unlikely, therefore, that useful information
will emerge from studies of temperature dependence unless the experimenters
are willing to go rather deeply into the theory. If one is working with an en-
zyme for which the mechanism is known in considerable detail, and for which
the rate constants of individual steps can be measured, then the temperature
dependence of any of these steps can be analysed just as if it were obtained in a
chemical kinetic experiment. Apart from this, temperature dependencies are
most likely to be useful when a break (abrupt change in slope) in an Arrhenius
plot coincides with a known physical phenomenon; for example, if one is
studying a lipid substrate that is known to undergo a phase transition at a
particular temperature, then a break in the Arrhenius plot at that temperature
may be helpful for identifying which form of the lipid reacts with the enzyme.
It is important to realize, however, that most of the abrupt changes in slope
that are reported are not real. The reasons for this are primarily psychological:
once a line or set of lines have been drawn over the experimental points in a
graph it is easy to be convinced that they fit the points rather better than they
do in reality, and because of this it is not difficult to 'fit' the data in an
Arrhenius plot so that the change in slope occurs in a place that is easy to
interpret. Consider the data in Figure 5. Is the break at (a) 9 °C or (b) 17.5 °C? Or is
there in reality no break at all but a gradual change of slope over the whole
range plotted, as in (c)?
178
STEADY-STATE KINETICS
(a)
log
2
4
(b)
log A
J I
2
4
(c)
log k
10 Cooperativity
All of the cases considered so far can be regarded as generalizations of the
Michaelis-Menten equation, (equation 5). However, although many enzymes do
behave in this way, at least as a first approximation, there are some important
exceptions. It is simple to calculate from equation 5 that if a = Km/9 then v = 0.1 V
and if a = 9Km then v = 0.9V; in other words, spanning the 10-90% range of
available rates requires an 81-fold range of substrate concentrations, almost two
orders of magnitude. Similar calculations may be done with any of the equations
of the Michaelis-Menten type, whether for additional substrates, inhibitors or
activators, and they imply that as long as enzymes follow Michaelis-Menten
kinetics, their rates vary very little with the sort of changes in substrate and
other concentrations that are likely to occur in the cell. As effective regulation
of metabolism requires that at least some enzymes respond sensitively to
179
ATHEL CORNISH-BOWDEN
M 0
-1
-1
log!
Figure 6 Hill plot for inhibition. As noted by the dashed lines at the two extremes, any Hill
plot should in principle approach a slope of unity at the extremes, even though in practice
the curvature may often be hard to detect and the points may appear to fit a straight line
over the whole experimental range.
180
STEADY-STATE KINETICS
181
ATHEL CORNISH-BOWDEN
12 Concluding remarks
Steady-state analysis of initial rates has formed the basis of most kinetic studies
of enzymes for almost a century, and as such it includes much more detail than
one can hope to cover in a single chapter. Some important topics, such as the
analysis of reactions with more than one substrate, have not been discussed at
all, and nearly all have been dealt with very briefly. A considerably more com-
plete account of all of the material in this chapter may be found elsewhere (11).
References
1. Michaelis, L. and Menten, M. L. (1913) Biochem. Z. 49, 333-369.
2. Lineweaver, H. A. and Burk, D. (1934) J, Am. Chem. Soc. 56, 658-666.
3. Woolf, B. (1932) cited by J. B. S. Haldane and K. G. Stern in Allgemdne chemie tier
enzyme, pp. 119-120. Steinkopff, Dresden and Leipzig.
4. Hanes, C. S. (1932) Biochem.]. 26,1406-1421.
5. Eadie, G. S. (1942) J. Biol. Chem. 146, 85-93.
6. Hofstee, B. H. J. (1952) J. Biol. Chem. 199, 357-364.
7. Cornish-Bowden, A. (1995) Analysis of enzyme kinetic data. Oxford University Press,
Oxford.
8. Fersht, A. (1985) Enzyme structure and mechanism, pp. 150-154. Freeman, New York.
9. Arrhenius, S. (1889) Z. Physik. Chem. 4, 226-248.
10. Hill, A. V. (1910) J. Physiol. 40, iv-vii.
11. Cornish-Bowden, A. (1995) Fundamentals of enzyme kinetics (2nd edn). Portland Press,
London.
182
Chapter 7
Spectrophotometric assays
T. J. Mantle* and D. A. Harris1"
* Department of Biochemistry, Trinity College, Dublin 2, Ireland
f
Department of Biochemistry, University of Oxford, South Parks Road,
Oxford 0X1 3QU
1 Introduction
Chapter 1 describes the principles and practice of spectrophotometry, and in this
chapter we will consider how this technique can be used to measure the amount
of analyte in solution, when the product of a reaction between the analyte and a
'very useful reagent' produces a measurable change in absorbance. Some good
examples of this type of reaction are given together with various protocols for
routine applications such as the measurement of protein concentration.
If there are no contaminating species absorbing at the wavelength of interest,
it may be possible to measure the analyte directly simply by taking absorbance
measurements. Spectrophotometry has also proved to be fundamental for rate
measurements using a wide variety of enzyme-catalysed and non-catalysed
reactions. We will also see that it is often the case that we will use an enzyme
reaction that has gone to completion to measure the amount of an analyte that
is limiting in the reaction. Furthermore, in some cases we can get the reaction
to cycle so that the amplified rates can provide a highly sensitive assay, see, for
example, the Tietze method (1) for measuring glutathione (Section 3.3). We will
also briefly cover the use of so-called 'plate readers' as these are being used
increasingly as 'multi-cuvette spectrophotometers'. Finally, 'plate readers' will
be compared with 'centrifugal analysers', much beloved in clinical chemistry
labs. The reader should note that some methods are dealt with in detail here to
allow selected protocols to be used directly. Also, a number of useful reference
works for enzyme assays are given, including (2-4), which should be consulted.
With a little thought novel assays can often be developed based on the various
'coupling' methodologies described below.
1.1 Spectrophotometers
The design of standard spectrophotometers has been covered in Chapter 1 and
we should simply mention that most modern 'plate readers' work in essentially
the same way. The major difference is in their beam geometry. In spectro-
photometers, samples are read through cuvettes with a horizontal light path
which is normally 1 cm in length allowing the Beer-Lambert Law (see Section
183
T. J. MANTLE AND D. A. HARRIS
1.2) to be applied in its simplest form (constant light path). In 'plate readers' the
vertical light beam results in a pathlength that depends on the volume of fluid
in each well. The major drawback to this configuration is that the pathlength is
not maintained by the dimensions of the cuvette (as in a conventional
spectrophotometer) and small variations in the volume pipetted into the well
are a potential source of error. For most ELISA uses these errors are acceptable;
however, to use these instruments to read absorbances it is advisable to use a
series of dilutions of a standard analyte to set up a calibration curve using a
conventional spectrophotometer, and then to compare the readout for the same
solutions in a 'plate reader'. It is then possible to calculate the pathlength in the
'plate reader' and to make the appropriate corrections. This problem has re-
cently been addressed by Molecular Devices (www.moldev.com) where the path-
length is automatically corrected using near infrared measurements. The same
problem does not apply to 'centrifugal analysers' as the pathlength in this case
is determined by the dimensions of the cuvette. Another downside with earlier
'plate readers' was the restriction placed on the wavelengths that could be used
due to the limited range of filters supplied; however, the introduction of
interference filters has effectively overcome this problem.
184
SPECTROPHOTOMETRIC ASSAYS
185
T. J. MANTLE AND D. A. HARRIS
coefficient for NADH at 340 nm is 6220 M-1 cm-1 which means that a solution of
1 uM has an absorbance of 0.006. This reflects the lowest amount of NADH that
can be measured with confidence using spectrophotometry. If greater sensi-
tivity is required for measuring NADH, then fluorimetry (excitation wavelength
340 nm, emission wavelength 460 nm) can be used.
2.1.1 Volume
Assay mixtures contain, in principle, three components, i.e. the sample contain-
ing the material to be assayed, a diluent and a 'colour' producing reagent
(which of course may mediate an absorbance change in the UV region of the
spectrum). If the volume of the sample is changed to obtain a larger or smaller
colorimetric response, this change is balanced by using a correspondingly
adjusted volume of diluent to keep the sum of the two components constant. It
186
SPECTROPHOTOMETRIC ASSAYS
2.1.2 Diluent
The diluent contains buffers ions and other reagents to allow colour develop-
ment. For example, the Biuret and Lowry assays for protein consists of alkali,
Cu2+ ions (to chelate to the protein) and tartrate to keep the Cu2+ in solution
under alkaline conditions. The sample volume is rarely allowed to exceed 25%
of the volume of diluent. It is especially important to consider that there may be
compounds in the sample that interact with key components of the diluent.
Common interactions that need to be watched for are chelators in the sample
that may lower the free concentration of metals required in the diluent. Useful
lists of compounds that interfere with the Lowry assay for proteins (see Section
4.3) can be found in (5) and (6).
187
T. J. MANTLE AND D. A. HARRIS
enzyme etc.) will not affect the final measurement, since this is made when the
reaction has ended. It follows therefore, that these reactions must be essentially
irreversible, i.e. AG° must be large and negative. If the equilibrium is poised it
may be possible to pull the equilibrium over to the product side by including a
trapping reagent in the assay. For example, in the estimation of ethanol by
alcohol dehydrogenase, the inclusion of semicarbazide in the assay traps the
aldehyde produced as the stable semicarbazone and so drives the reaction to
completion. Even if the reaction is complete, the appropriate absorbance may
not be measured if, as is often the case, the colour reagent produces a product
that is stable for a limited period of time (and hence the colour is only stable for
a limited time). In these cases, it is important to establish a maximum period for
the assay so that decay of the coloured product is not a significant problem. An
additional problem can occur with an enzyme-generated signal if the enzyme-
catalysed reaction is 'leaky'. For example, the assay for ATP involves a 'double
couple' of hexokinase and glucose 6-phosphate dehydrogenase (plus NADP+, see
also Section 6.3.2) in the presence of a high concentration of glucose. Under
these conditions, the reaction is expected to be limited by ATP; however, when
the enzymatically produced glucose 6-phosphate has been consumed there is a
slow increase in A340 (due to NADH) as glucose itself is a poor substrate for G6PDH.
In this case, it is important to extrapolate the progress curve to zero time.
188
SPECTROPHOTOMETRIC ASSAYS
NADPH
189
T. J. MANTLE AND D. A. HARRIS
Protocol 1
A Bluret assay to determine protein concentration
Reagents
• Dissolve 1.5 g of CuS04.5H2O and 6 g • To this solution add 300 ml 10% (w/v)
sodium potassium tartrate (containing sodium hydroxide, mixing during the
four waters of crystallization) in 500 ml addition.
boiled and then cooled distilled water. * Finally add 200 ml distilled water.
This solution is stable for several months at room temperature.
Method
NB. The sample should contain 0.5-4 mg protein in 0.2 ml water or buffer.
1 Construct a standard curve for the assay by adding 2.3 ml of the reagent (see above)
to each of a number of samples (0.2 ml) containing between 0.5 and 5 mg BSA.
2 Add 2.3 ml of the reagent to 0.2 ml of the protein solution being assayed.
3 Leave the mixtures at room temperature for 2 h (alternatively heat to 80°C for 5
min, cool rapidly and stand at room temperature for 10 min).
4 Read the A540 of each sample against a blank of water or buffer plus reagent. Note
that 1 mg of protein gives an A 540 of approximately 0.1.
190
SPECTROPHOTOMETRIC ASSAYS
Protocol 2
A Lowry assay to determine protein concentration
Reagents
• 1% CuSO4.5H2O (A) • 2% sodium carbonate in 0.1 M NaOH (C)
• 1% sodium potassium tartrate.4H2O.(B) • Folin-Ciocalteau reagent (D)
These reagents are stable for several months at room temperature with the exception of
the Folin-Ciocalteau reagent which must he stored at 4°C, The working solutions must
be freshly prepared by mixing equal volumes of reagents (A) and (B) and then diluting
this mixture with 98 volumes of reagent (C) to give the alkaline copper solution. Reagent
(D) must be diluted 1:1 v/v with distilled water before use.
Method
NB, The sample should contain 5-40 ug protein in 0.2 ml water or buffer.
1 Construct a standard curve by adding 2.1 ml alkaline copper solution to each of
several samples (0.2 ml) of solution containing between 1 and 40 ug of BSA,
191
T. J. MANTLE AND D. A. HARRIS
Protocol 2 continued
2 Add 2.1 ml of the alkaline copper solution to 0.2 ml of the protein solution to be
assayed.
3 Leave the mixtures for 10 min.
4 Add 0.2 ml of the diluted Folin-Ciocalteau reagent to each sample and vortex
immediately.
5 Leave the mixtures at room temperature for 1 h,
6 Read the A750 of each sample against a blank of water or buffer plus reagents. Note
that 10 ug of protein gives an A750 of approximately 0.1.
Protocol 3
A bicinchoninic assay to determine protein concentration
Reagents
• 1% bicinchoninic acid, 0.16% sodium tar- • A solution of 4% CuS04.5H2O in distilled /
trate, 2% sodium carbonate, 0.95% sodium deionized water (A)
bicarbonate in 0.1 M NaOH (pH 11.25) (B)
Solutions (A) and {B) are stable for several months at room temperature. To use, mix reagents
(A) and (B) in a ratio of 1:50 to form the bicinchoninic reagent.
Method
NB. The sample should contain 5-40 ug protein in 0.2 ml water or buffer.
1 Construct a standard curve by adding 2.3 ml bicinchoninic reagent to samples of
0.2 ml of BSA (1-40 ug) in water.
2 Add 2.3 ml of the reagent to 0.2 ml of the protein solution to be assayed.
3 Leave the mixtures at room temperature for 2 h (or heat at 60°C for 30 min).
4 Read the A562 of each sample against a blank of water or buffer plus reagents. Note
that 10 ug of protein gives an A562 of approximately 0.1.
192
SPECTROPHOTOMETRIC ASSAYS
Protocol 4
A dye-binding assay for the estimation of protein
concentrations
Reagents
• lOO mg of Serva Blue G dissolved in 50 ml • Concentrated phosphoric acid (85%)
ethanol), then diluted to 500 ml with diluted 1:5 v/v with water (B)
water (A)
Solutions (A) and (B) are mixed in a 1:1 ratio on the day of use to produce the working reagent.
Method
NB. The sample should contain 2-20 ug protein in 0.2 ml water or buffer.
1 Construct a standard curve by adding 2.3 ml of the working reagent to each of
several samples (0.2 ml) of BSA containing between 1 and 20 ug of protein.
2 Add 2.3 ml of the reagent to 0.2 ml of the protein solution being assayed.
3 Mix the solutions and read the A595 of each one immediately. Do not leave the
mixture for periods longer than 30 min. Note that 5 ug protein gives an A595 of
approximately 0.1.
193
T. J. MANTLE AND D. A. HARRIS
Protocol 5
A fluorescence assay for the estimation of protein
concentrations
Reagents
• 7.5 mg of fluorescamine dissolved in 25 • Borate buffer pH 8.5 (3 g boric add + 4.8g
ml dry acetone (A) sodium tetraborate in 200 ml water) (B)
Solution A should be prepared on the day of use, in a dry test tube, and contact with moisture
(e.g. wet pipette tips) should be avoided,
Method
NB. The sample should contain 0.1-2 ug protein in 0,2 ml water or buffer.
1 Construct a standard curve by adding 2.1 ml borate buffer to each of several
samples (0.2 ml) of BSA containing between 0.1 and 2 ug of protein.
2 Add 2.1 ml of buffer to 0.2 ml of the protein solution being assayed.
3 Mix on a vortex mixer and, while still mixing, add 0.2 ml fluorescamine solution.
4 Read the fluorescence of each sample after 2 min, using Xex = 390 nm and \sm =
465 nm
Amines interfere with this assay; aliphatic amines yield fluorescent products
with these compounds, while ammonia, although not giving a fluorescent
product, will combine with (and hence use up) the reagent. Unlike the other
methods for protein determination, these reagents will give responses with
peptides in addition to proteins. Whether or not this is advantageous depends
on the requirements of the investigator,
194
SPECTROPHOTOMETRIC ASSAYS
195
T. J. MANTLE AND D. A. HARRIS
~~dt = K^
Integration gives us the total time for given change in s
so $ Km o
or
or
V/Km = 0.46 min-1 8
or
V = 0.46Km min-1 9
This equation tells us that, for a reaction to go to 'completion' in 10 min, the
amount of enzyme added should be an amount able to change the concen-
tration of substrate in the assay by 0.46Km per minute. If Km = 1 mM, for
example, for 99% completion in 10 min
V s 0.46 mM min-1 (> 0.46 mmol litre0-1 min-1) 10
The concentration of enzyme is conventionally given in U/ml, where 1 U is
the amount of enzyme that will convert 1 umol substrate/min at saturation. If
we add 1 U of enzyme to an assay volume of 3 ml, for example, then V = 1/3
umol/ml/min = 0.33 mmol litre-1 min-1. Thus, if Km = 1 mM, for V >: 0.46 mmol
litre-1 min-1, we need to add s 0.46/0.33 Units of enzyme if the assay volume is
3 ml.
These calculations are useful 'rangefinders', but a number of points should
be borne in mind. It is important to determine experimentally the activity of
the enzyme in the assay medium used. The units of activity that accompany
commercial preparations are those measured under optimal conditions (as
described by the manufacturer), and these conditions may not be those used in
the assay medium being used. Also note (as set out above) that the amount of
enzyme required will increase with increased assay volume. The reaction will,
196
SPECTROPHOTOMETRIC ASSAYS
of course, proceed faster if you add more than the calculated amount and this
may be convenient if the cost is not too great.
Commercial preparations of many enzymes are supplied as suspensions in
high concentrations of ammonium sulphate or may be stabilized with glycerol.
It is often, though not always, necessary to remove these 'protectants' before
using the enzyme. This can be achieved by dialysis or gel fitration (this latter
method will not work with ammonium sulphate suspensions). It is important to
check the stability of the resultant enzyme stock solution. Some enzymes can
be stored frozen as aliquots that may be used once and then discarded. Some
may stand repeated freezing and thawing (this is unusual), while others may
retain considerable activity at 4°C. It is worth investing some time to establish
the best conditions for storing stock enzyme solutions and it is also advisable to
check the activity of stocks on a regular basis. This is as useful a check on the
substrates and buffers as on the stability of the enzyme solutions.
Remember, more enzyme, more activity, so that if the activity of a stock
solution is declining on storage it may still be possible to add sufficient units of
activity simply by adding more from the stock solution. This, of course, rather
depends on the use of small volumes of concentrated stock enzyme solutions in
the final assay mix.
197
T, J. MANTLE AND D. A. HARRIS
Protocol 6
An assay for glucose using glucose oxidase
Reagents
• 1 mM ABTS in 0.1 M sodium phosphate pH 7 • Peroxidase (20 U/ml) in 0.1 M sodium
• Glucose oxidase (100 U/ml) in 0.1 M sodium phosphate pH 7
phosphate pH 7
These reagents are stable for one month at 4°C.
Method
NB, The sample should contain 1-10 nmol glucose in 0,2 ml.
1 Construct a standard curve by adding 2.1 ml ABTS to each of 0,2 ml samples of
known concentrations of glucose (1-10 nmol glucose in 0.2 ml water).
2 Add 2.1 ml of ABTS reagent to 0.2 ml of the sample of glucose to be assayed.
3 Add 0.1 ml peroxidase to all of the above samples.
4 Preincubate for 5 min at 37°C.
5 Start the reaction by the addition of glucose oxidase (0.1 ml),
6 Read the A420 of each sample after 10 min against a blank of reagents with no
glucose added.
Protocol 7
An assay for glucose using glucose 6-phosphate
dehydrogenase
Reagents
• 200 mM Tris/HCl pH 8, containing 2 mm • 100 mM NADP'. In contrast to NADH and
MgCl2 and 0.2 mM EDTA NADPH which should not be kept in acid
* 200 mM MgATP. This is prepared by solutions, NADP" should be stored in
dissolving ATP (1 mmol) in 4 ml MgCl2 buffer below pH 7. As phosphate inhibits
(250 mM) and titrating this solution to pH G6PDH Tris/HCl pH 6.5 is suitable.
7 using 1 M KOH. The final volume should • G6PDH (20 U/ml)
be brought to 5 ml. • Hexokinase (100 U/ml)
Method
NB. The sample should contain 40-400 nmol glucose in 0.2 ml.
1 Add 2.1 ml Tris buffer to 0.2 ml samples of glucose (either containing known
amounts between 40-400 nmol of glucose for setting up a standard curve or the
unknown sample being assayed).
2 Add 0.05 ml MgATP and 0.025 ml NADP" to each sample.
3 Add 0.1 ml G6PDH enzyme, mix the components and preincubate at 37°C for 5 min.
198
SPECTROPHOTOMETRIC ASSAYS
Protocol 7 continued
7 Luminescence-based assays
Measurement of luminescence provides the basis of highly sensitive assays
for ATP, NADPH and H2O2. The production or disappearance of ATP can be
measured by following light emission in the presence of fire-fly lucifernse which
catalyses the following reaction:
ATP + luciferin - O2 -» oxyluciferin PPi + CO2 + AMP + light 12
Likewise, NADPH can be measured using the bacterial luciferasc which
catalyses the following reactions:
NADPH + H' - FMN -> NADPH + FMNH2 13
FMNH2 + RCHO + 02 - H20 + RCOOH + light 14
In which RCHO is a long chain aldehyde (C8 to C12), Finally, earthworm lucifer-
ase can be used for determining H202. Further details can be found in (18).
Conventional spectrophotometers (which are configured to measure the dif-
ference between incident and transmitted light) do not normally show sufficient
sensitivity for determination of bioluminescence. Digital fluorimeters and scin-
tillation counters can often be adapted for measurement of luminescence but.
199
T. J. MANTLE AND D. A. HARRIS
200
SPECTROPHOTOMETRIC ASSAYS
201
T. J. MANTLE AND D. A. HARRIS
addition of alkali performs two functions; one to terminate the reaction at the
required time and the second to produce the chromophure.
Another common procedure for stopped assays is the use of molybdate under
acid conditions to complex and spectrophotometrically assay inorganic phos-
phate. This provides a useful alternative to p-nitrophenyl phosphate for phos-
phatase assays, as physiologically relevant phosphate esters can be used. For
example, this assay also provides a useful method for measuring ATPase
activity. Details of the phosphomolybdate assay arc given here (however, see
Section 8.5 for methodologies that utilize plate readers).
Protocol 8
A phosphomolybdate assay for inorganic phosphate
released by ATPase activity
Reagents
200 mM MgATP. This is prepared by ' 100 mM Tris/HCl pH 8, containing 50 mM
dissolving ATP (1 mmol) in 4 ml MgCl2 KC1. 2 mm MgCl2 and 0.2 mM EDTA (A)
(250 mw) and titrating this solution to pH • 40% trichloroacetic acid (TCA) (C)
7 using 1 M KOH. The final volume should • 1% ammonium molybdate in 1 M sulphuric-
be brought to 5 ml. (B) acid (D)
Solutions (C) and (D) are stable for months at room temperature. The MgATP (B) can be
stored at —20 °C in aliquots, that should be used once and discarded. The colour reagent
is prepared on the day of use by dissolving 1 g FeSO4 VH2O in solution (D).
Method
1 Mix 0.2 ml of sample with 0.8 ml of buffer (A).
2 Pre-incubate the mixture at 30 °C for 5 min.
3 Add 0.02 ml MgATP (B) to initiate the reaction.
4 Terminate the reaction after 10 min by the addition of 0.1 ml TCA (C).
5 Add 1.9 ml of colour reagent and read the A700 after 10 min.
NB. The A700 of a blank solution made up of reagents with no enzyme added should be
measured and subtracted from the reading given in step 5 above. A standard curve may be
constructed by using known amounts of inorganic phosphate in 0.2 ml of water and treated as
above. 1 umol phosphate released should give an A700 of approximately 1,
It should he stressed for any stopped assay that, after demonstrating linear
product-time curves (i.e. to show genuine initial rates), in a preliminary experi-
ment, half-time points should be incorporated into the protocol for The highest
and lowest substrate concentration used, to demonstrate initial rate conditions
are maintained. For example, if the rate is linear up to 20 min then 'half-time'
points of 10 min should be included routinely.
202
SPECTROPHOTOMETRIC ASSAYS
203
T. J. MANTLE AND D. A. HARRIS
disc. The contents of an inner well (e.g. sample) are moved radially outwards
passing into other wells holding reagents. After mixing, the sample/reagent
solutions continue to move outwards to cuvettes located towards the perimeter
of the disc. The disc rotates rapidly presenting each cuvette sequentially to an
optical detector. The reactions develop with time and the absorption (or
fluorescence) of each cuvette is measured each time it passes through the
detector to allow the rate of reaction or final value to be determined.
Since the earlier reviews on centrifugal analysers (28, 29) there has been
significant progress in design and formatting. Artuch et al (30) have recently
described an adaptation of some previously reported methods for the Cobas
Fara II centrifugal analyser, using the supernatant of a unique deproteinized
blood sample for the determination of four key analytes (lactate, pyruvate, p-
hydroxybutyrate and acetoacetate) in bioenergetic studies.
The advent of molecular biology and biotechnology have realized and facili-
tated analyses for DNA or RNA sequences and these are now being married to
spectrophotometric assays using centrifugal analysers. New technological ad-
vances have led to the automation of major parts of the assay process.
Automated systems have been developed for amplification and detection of
nucleic acid sequence for infectious agents, using the polymerase chain reaction,
ligase chain reaction, strand displacement amplification, transcription-associated
amplification, and nucleic acid sequence-based amplification. Development of
such automated systems are based on accumulation and integration of new
molecular biotechnology. There has appeared a fully automated PCR system
(COBAS AMPLICOR), which amplifies target nucleic acid sequences, captures the
biotinylated and amplified products on oligonucleotide-coated paramagnetic
microparticles, and detects the products with an avidin-horseradish peroxidase
conjugate. Automated systems provide improvement not only in terms of labour
efficiency but also assay accuracy. Recently, the extraction of a specific sequence
for hepatitis C virus RNA has been automated, and the RNA can be specifically
extracted from serum by hybridization with probe-coated paramagnetic micro-
particles, and then subjected to in vitro amplification (31).
9,1 Cysteine
The classic reagent is 5,5'-dithiobis-(2-nitrobenzoic acid) or DTNB, sometimes
known as Ellmans reagent (32), which is absolutely specific for the thiol side
204
SPECTROPHOTOMETRIC ASSAYS
chain of cysteine residues. By using a molar excess of DTNB over the protein
thiol concentration in the presence of 1 to 10 mM EDTA at pH 8, the thiol
content of a protein can be conveniently monitored spectrophotometrically by
the release of a thionitrobenzoate ion (TNB~) which absorbs strongly at 412 nm
(s412 = 13600 M-1 cm-1). If all of the cysteine residues are accessible to the
reagent and all have a normal pKa (approximately pH 8.7), then the reaction is
rapid as long as the concentration of reagent is in excess. From the extinction
coefficient of the TNB~ ion released, it is possible to calculate the concentration
of thiol modified. Often a fraction of the protein's thiol groups are not available
for reaction because of steric hindrance to the reagent by side chains of neigh-
bouring residues. For monitoring total content, therefore, the protein is com-
monly unfolded before the reaction by the addition of SDS. However, useful
information on protein conformation/folding can be achieved in the absence of
SDS, where kinetic sets of 'fast' and 'slow' reacting thiols (reflecting cysteine
side chains in distinct environments) can be monitored. Other reagents include
4,4'-dithiopyridine that reacts with thiol groups to release 4-thiopyridone that
absorbs light at 324 nm with an e324 =19 800 M-1 cm-1.
9.2 Lysine
Another useful reagent is trinitrobenzene sulphonic acid (TNBS) which is almost
totally specific for amino groups and hence reacts well with the e-amino groups
of reactive lysine residues. It will, however, also react with the thiol group of
cysteine residues, but the S-TNP conjugate is not stable in mildly alkaline
solutions. Typically, 1 to 10 mM TNBS in buffer (pH 7.5 to 9) is used to modify
proteins. Reaction times are normally 1 to 4 h and the reaction is conveniently
monitored at 340 nm (s340 = 14000 M-1 cm-1). The large absorbance change
coupled with the fairly high degree of specificity means that this is a useful
reagent for kinetic analysis of reactive amine groups (33).
An alternative reagent for lysine residues is pyridoxal phosphate which reacts
with unprotonated amine groups to form a Schiff base. The reagent therefore
modifies 'reactive' lysyl amino groups at pH values around 7.5. The unstable
imine can be reduced using sodium borohydride (added in aliquots to a final
concentration of 1 to 10 mM after having adjusted the pH of the medium to 4.5
to 6) to form a stable adduct which absorbs at 325 nm (e325 = 9710 M-1 cm-1) and
emits fluorescence at 425 nm. Further details can be found in (34).
9.3 Tyrosine
Tetranitromethane is a useful (but extremely hazardous) reagent for modification
of tyrosine side chains to give a coloured product that is easily monitored by
spectrophotometry (35). There are a number of side reactions, including the
oxidation of cysteine and the modification of histidine, tryptophan and
methionine residues at higher pH, coupled with an inherent problem of cross-
linking via tyrosine residues. However, if a protein concentration of 2 to 10 mg/
ml can be achieved then this reaction can be carried out at room temperature
205
T. J. MANTLE AND D. A. HARRIS
9.4 Histidine
A similar spectrophotometric determination of histidine residues is based upon
the reactivity of the imidazole side chain with diethylpyrocarbonate (sometimes
termed ethoxyformic anhydride) at pH 6.0 to yield an N-carbethoxyhistidinyl
derivative (36). The reagent will also react to varying degrees (depending upon
pH) with other residues such as tyrosine, lysine or cysteine. The addition of
hydroxylamine to the modified protein will release the carbethoxy group from
modified histidines and tyrosines but not from cysteines or lysines. The reaction
with the imidazole side chain of histidine may be monitored and quantified by
an increase in absorbance at 240 nm (e240 = 3200 M-1 cm-1 (36)). Again caution
must be used as the use of excess reagent can give rise to disubstituted de-
rivatives of histidine which have higher molar extinctions than the mono-
substituted group and this may lead to over-estimation of the equivalents of
histidine in a protein (37). The carbethoxy-imidazole derivative is unstable and
is readily hydrolysed by water. The rate of the reaction increases with pH above
pH 6.0: at pH 7.0 and 25 °C the half-life of the group is 55 h and at pH 10.0 it is
18 min (38). Furthermore, the diethylpyrocarbonate reagent itself is also
hydrolysed by water and the type of buffer used affects the rate of hydrolysis of
the reagent. The half-time for the breakdown of the reagent in 60 mM sodium
phosphate buffer, pH 6.0 at 25 °C, is 24 min (36). Therefore, the rate of modifica-
tion of histidines in a protein will decrease with time due to the loss of the
functional reagent. The concentration of the reagent will need to be 'topped up'
during a modification experiment to ensure that all available residues are
reacted. Care must be taken to keep the stock reagent anhydrous and stock
solutions should be made up in dry ethanol.
9.5 Tryptophan
There are a number of reagents (e.g. 2-nitrophenyl-sulphenyl chloride and 2-
hydroxy-5-nitrobenzylbromide) which react with tryptophan side chains to give
products which can be measured spectrophotometrically. However, these require
an acid pH to allow the reaction to proceed with any degree of specificity (39,40).
10 Concluding remarks
The continuing improvement of both the available chemical reagents and
instrumentation for analytical purposes is leading to increases in sensitivity and
206
SPECTROPHOTOMETRIC ASSAYS
References
1. Tietze, F. (1969) Anal. Biochem. 27, 502.
2. Bergmeyer, H. U. (1974) Methods of enzymatic analysis. Academic Press, New York,
London.
3. Barman, T. E. (1969 and 1974) Enzyme handbook. Springer Verlag, Heidelberg, New
York.
4. Guilbault, G. G. (1970) Enzymatic methods of analysis. Pergamon, Oxford, London, New
York.
5. Ji, T. H. (1973) Anal. Biochem. 52, 517.
6. Peters, M. A. and Fouts, J. R. (1969) Anal. Biochem. 30, 299.
7. Lowry, O. H., Rosebrough, N. J., Farr, A. L and Randall, R. J. (1951) J. Biol. Chem. 193,
265.
8. Legler, G., Muller-Platz, C. M., Mentges-Hettkamp, M., Pflieger, G. and Julich, E.
(1985) Anal. Biochem. 150, 278.
9. Rodriguez-Vico, P., Martinez-Cayuela, M., Garcia-Peregrin, E. and Ramirez , H. (1989)
Anal. Biochem. 183, 275.
10. Bradford, M. M. (1976) Anal. Biochem. 82, 327.
11. Hernandez, L, Marquina, R, Escalona, J. and Guzman, N. A. (1990) J. Chromatog. 502,
247.
12. Liu, J. P., Hsieh, Y. Z., Wiesler, D, and Novotny, M. (1991) Analyt. Chem. 63, 408.
13. Bridges, M. A., McErlane, K. M., Kwong, E., Katz, S. and Applegarth, D. A. (1986) Clin.
Chem. Acta , 157, 73.
14. Labarca, C. and Paigen, K. (1979) Anal. Biochem. 102, 344.
15. McClure, W. R. (1969) Biochemistry 8, 2782.
16. Tipton, K. F. (1985) In Techniques in the life sciences (ed. K. F. Tipton) Vol. Bl/II,
Supplement BS113, p. 1. Elsevier, Limerick.
17. Tipton, K. F. (1992) Principles of enzyme assay and kinetic studies, in Enzyme assays: A
practical approach (ed. R. Eisenthal and M. J. Danson). IRL Press, Oxford.
18. Harris, D. A. (1987) In Spectrophotometry and spectrofluorimetry: A practical approach (ed.
C. L. Basford and D. A. Harris), pp. 49-90. IRL Press, Oxford, Washington.
19. Selwyn, M. J. (1965) Biochim. Biophys. Acta 105, 193.
20. Hulme, E. C. and Tipton, K. F. (1971) FEES Lett 12,197.
21. Hoffer, H. W. (1971) J. Physiol. Chem. 352, 997.
22. Cohen, G., Kim, M. and Ogwu, V. (1996) J. Neurosci. Methods 67, 53.
23. Frost, G. I. and Stern, R. (1997) Anal. Biochem. 251, 263.
24. Doctor, B. P., Toker, L., Roth, E. and Silman, I. (1987) Anal. Biochem. 166, 399.
25. Fisher, D. K. and Higgins, T. J. (1994) Pharm. Res. 11, 759.
26. Sadrzadeh, S. M., Vincenzi, F. F. and Hinds, T. R. (1993) J. Pharmacol. Toxicol. Methods
30, 103.
27. Drueckes, P., Schinzel, R. and Palm, D. (1995) Anal. Biochem. 230,173.
28. Burtis, C. A., Tiffany, T. 0. and Scott, T. D. (1976) Methods Biochem. Anal. 23,189.
29. Savory, J. (1977) Ann. Biol. Clin. (Paris) 35, 261.
207
T. J. MANTLE AND D. A. HARRIS
30. Artuch, R., Vilaseca, M. A., Farre, C. and Ramon, F. (1995) Eur.}. Can. Chem. din.
Biochem. 33, 529.
31. Albadalejo, J., Alonso, R., Antinozzi, R., Bogard, M., Bourgault, A. M., Colucci, G.,
Fenner, T., Petersen, H., Sala, E., Vincelette, J. and Young, C. (1998) J. Gin. Micrdbiol.
36, 862.
32. Ellman, G. L. (1959) Arch. Biochem. Biophys. 82, 70.
33. Freedman, R. B. and Radda, G. K. (1968) Biochem.]. 108, 383.
34. Lilley, K. S. and Engel, P. C. (1992) Eur.j. Biochem. 207, 533.
35. Sokolovsky, M., Riordan, J. F. and Vallee, B. L. (1966) Biochemistry 5, 3582.
36. Miles, E. W. (1977) Methods in Enzymology, 47, 431.
37. Avaeva, S. M. and Krasnova, V. I. (1975) Bioorg. Khim. 1,1600.
38. Means, G. E. and Feeney, R. E. (1971) Chemical modification of proteins. Holden-Day, San
Francisco.
39. Fontana, A. and Scoffone, E. (1972) Methods in Enzymology 25, 419.
40. Horton, H. R. and Koshland, D. E. (1965) J. Am. Chem. Soc. 87, 1126.
208
Chapter 8
Stopped-flow spectroscopy
M. T. Wilson* and J. Torres
* Department of Biological Sciences, Central Campus, University of
Essex, Wivenhoe Park, Colchester, Essex C04 3SQ
T
Department of Biochemistry, University of Cambridge,
Tennis Court Road, Cambridge CB2 1QT
1 Introduction
There was a time, fortunately some years ago now, when to undertake rapid
kinetic measurements using a stopped-flow spectrophotometer verged on the
heroic. One needed to be armed with knowledge of amplifiers, light sources,
oscilloscopes etc. and ideally one's credibility was greatly enhanced were one to
build one's own instrument. Analysis of the data was similarly difficult. To obtain
a single rate constant might involve a wide range of skills in addition to those
required for the chemical/biochemical manipulation of the system and could
easily include photography, developing prints and considerable mathematical
agility.
Now all this has changed and, from the point of view of the scientist attempt-
ing to solve problems through transient kinetic studies, a good thing too! Very
high quality data can readily be obtained by anyone with a few hours training
and the ability to use a mouse and 'point and click' programs. Excellent stopped
-flow spectrophotometers can be bought which are reliable, stable, sensitive and
which are controlled by computers able to signal-average and to analyse, in
seconds, kinetic progress curves in a number of ways yielding rate constants,
amplitudes, residuals and statistics. Because it is now so easy, from the technical
point of view, to make measurement and to do so without an apprenticeship in
kinetic methods, it becomes important to make sure that one collects data that
are meaningful and open to sensible interpretation. There are a number of
pitfalls to avoid. The emphasis of this article is, therefore, somewhat different
to that written by Eccleston (1) in an earlier volume of this series. Less time will
be spent on consideration of the hardware, although the general principles
are given, but the focus will be on making sure that the data collected means
what one thinks it means and then how to be sure one is extracting kinetic
parameters from this in a sensible way.
With the advent of powerful, fast computers it has now become possible to
process very large data sets quickly and this has paved the way for the applica-
tion of 'rapid scan' devices (usually, but not exclusively, diode arrays), which
allow complete spectra to be collected at very short time intervals during a
reaction. In these circumstances there is a danger of being swamped by data.
209
M. T. WILSON AND J. TORRES
Whereas some years ago a single mixing experiment in a stopped-f low apparatus
would deliver some tens of absorbance values collected at a given wavelength,
the same experiment may now yield thousands of absorbance values at
hundreds of wavelengths and hundreds of time intervals. Analysis of these large
data sets now uses matrix algebra methods (Singular Value Decomposition,
SVD) to filter signal from noise and to assess, objectively, the number of linearly
independent spectral components which contribute to the total spectral change
accompanying the reaction, and to extract the pseudo-spectra (see later) and
their time courses. Inherent to these methods is great data compression, thus
permitting economic and efficient data storage. Once the number of compo-
nents and their time courses are known global analysis of the data allows one to
fit the kinetic processes to a chemical model by solving, iteratively, the differ-
ential equations which describe it. By so doing, not only can the rate constants
be extracted but also the true (model-dependent) spectra of any intermediates
on the reaction pathway. These analytical techniques are very powerful and
easy to implement; however, they can mislead and it is important to retain
one's chemica1/biochemical common sense when examining the output. It is on
these new techniques of rapid scan spectroscopy and the analysis of data that
we also spend some time.
Light
source
Filter or
monochromator
Syringes filled
with reagents
* Stopping syringe Trigger
r •--
Observation point
in flow capillary
Mixing
Push barrier
for syringe movement Detector
To computer To computer
210
STOPPED-FLOW SPECTROSCOPY
ution within the optical cell (pathlength, 1) now ages and the time course of the
reaction may be followed by monitoring the absorbance at a suitable wavelength
(A.). The advantage in following the change in absorbance (AA) is that it is directly
related to the concentration (c) of the absorbing chromophore(s) through the
appropriate molar difference extinction coefficient Ae^, i.e. ex pr0duct - e\ reactant
211
M. T. WILSON AND J. TORRES
Protocol 1
To test the stability of the detection system
Method
1 Fill working syringes with water or buffer,
2 Select a wavelength where the lamp output is high and the PMT is most sensitive,
usually somewhere between 500-600 nm.
3 Set the sensitivity on the recording apparatus to a medium range, say full scale 0.2A
units and set the PMT voltage (usually done automatically).
4 Wash the optical cell through up to 10 times by operating the pneumatic drive
system. This should ensure that the system is bubble-free.
5 Once a reasonably flat baseline is obtained set the sensitivity to maximum, collect
five traces and average these with the software provided.
6 Finally collect and average traces using the internal trigger, i.e. traces which do not
involve mechanical mixing, flow and the arrest of flow. These traces allow one to
monitor lamp fluctuations and other electronic noise.
Figure 2 illustrates the results of such a stability test. The electronic/lamp noise
is essentially zero while the noise generated by the mechanical mixing amounts
to -001A If the noise is above this, then one should clean the apparatus by
purging with detergent solutions or as recommended by the manufacturer.
Bubbles that can occasionally prove difficult to flush out may be removed by
passing degassed water through the instrument. Lamp 'ripples' (--2 Hz) can
sometimes be improved by slight lamp realignment.
212
STOPPED-FLOW SPECTROSCOPY
0.00
generally the one where the absorbance change is greatest on going from
reactants to products. This may not always be the case, and one must sometimes
be prepared to work at wavelengths away from the peak in the difference
spectrum in order to work in regions of the spectrum where the apparatus is
more sensitive.
213
M. T. WILSON AND J. TORRES
Protocol 2
To determine the time between initiation of the signal
recording and the stop of the flow of solution
Reagents and materials
• Prepare a solution of ferric (met) buffer to several cycles of exposure to
myoglobin by dissolving ~2 mg of the vacuum and equilibration with CO (at 101
freeze-dried protein, Sigma Chemical Co., kP pressure) accompanied by vigorous
in 10 ml of buffer (say 0.1 M sodium shaking. Gas equilibrates with a liquid
phosphate pH 7.0}, Add a few crystals (-2 phase faster the greater the interface area
mg) of solid dithionite (Na2S204, BDH) to between them, thus many small bubbles
remove oxygen from solution and to are better than a few larger ones. Draw
reduce the myoglobin to form the ferrous into a syringe a few ml of the CO solution
deoxy species. and add a little sodium dithionite (-2
• Prepare the stock CO solution, 1mM,by mg/10 ml). CO gas is toxic and these
equilibrating water (or dilute buffer) at operations should be performed in a
20 °C with an atmosphere of pure CO at 1 'fume' cupboard.
atmosphere pressure [101 kP). Subject the
Method
1 Load the Mb and the CO solution into the working syringes of the stopped-flow
apparatus. Flush the apparatus through with some five shots to ensure the
solutions have fully replaced water in the mixing chamber and optical cell.
2 Ensure there is no electronic noise filter and collect a further three to five traces,
each of some 400 time points, and average these.
214
STOPPED-FLOW SPECTROSCOPY
0.15
-0.1
0.00 0.02 0.04 0.06 0.08 0.10 0.12 0.14 0.16 0.18 0.20
Time (s)
Figure 3 Reaction between deoxy myoglobin and carbon monoxide. The time courses show
the binding of CO to deoxy myoglobin (10 uM) as a function of CO concentration (41, 56, 88,
150, 275 or 525 uM). The monitoring wavelength was 421 nm and the temperature 20°C.
The reaction was carried out in sodium phosphate buffer, 0.1 M, pH 7.0. The initial flat
portion of the progress curve (seen better in Figure 4) illustrates the flow period. The inset
gives the dependence of the pseudo-first-order rate constant, Kobs on CO concentration.
0.02
-0.1
0.000 0.005 0.010
Time (s)
Figure 4 Analysis of time course to measure the 'flow time' and the 'dead time'. The first
10 ms of the traces in Figure 3 are shown as individual data points. The solid lines are the
exponential fits to the data extrapolated back to true time zero. The flow period is the flat
initial portion of the curve and, in principle, may have any length and depends on the time
between triggering data capture and the flow stopping. The dead time is the time between t
= 0, given by the point of intersection of the curves, and when flow stops. In this experiment
the dead time and flow period are approximately the same, but need not be so.
215
M. T, WILSON AND J. TORRES
Protocol 3
Measurement of the 'dead time' of the instrument
Reagents
The myoglobin and CO solutions prepared as described in Protocol 2 may be used.
Methods
1 Mix the deoxy Mb with 1 mw CO and record at least three traces (at least 200 points
over 50 ms) and average,
2 Dilute the CO solution with buffer which has been degassed and equilibrated with
N2 and to which a small quantity of solid sodium dithionite has been added. To do
this, without exposure to air, take a known volume of the CO solution into a glass
graduated syringe (partially filling it) and connect this by its nozzle to a similar
syringe containing the anaerobic buffer by using a short plastic tube. Push a known
volume of the buffer into the CO solution. It is usually convenient to mix equal
volumes thereby diluting the CO by a factor of two.
3 Record a further set of traces and average.
4 Repeat the CO dilution step and collect data over the appropriate time range a
further three to four times. After each change of [CO) flush through some five shots
to ensure the new solution has completely replaced the previous solution.
216
STOPPED-FLOW SPECTROSCOPY
By extrapolating each curve back into the dead time there should be a time
point where all the curves intersect, giving the true time zero of the reaction.
Figure 4 shows the extrapolation of a set of exponential curves fitted to the data
presented in Figure 3. This is typical for such an experiment. The majority of the
time courses intersect while one is to one side of this point. Nevertheless, the
intersection point for the majority of the lines is a good estimate of time zero.
The average time between this intersection point and the end of the flow period,
i.e. where observation truly starts is approx. 1.3 ms and this is the dead time of
the instrument. This may be made slightly shorter by cleaning the instrument,
making sure the syringes are running smoothly or by slightly increasing the
pressure of the pneumatic drive.
217
M. T. WILSON AND J. TORRES
218
STOPPED-FLOW SPECTROSCOPY
Protocol 4
Determination of a kinetic difference spectrum
Reagents
• Prepare in 50 mM potassium phosphate Equilibrate NO gas at one atmosphere
buffer, pH 7.4, an anaerobic solution of a (101 kP) pressure and 20°C with anaerobic
ruthenate chelate, here potassium water (see Protocol 2 above for CO). This
chloro [hydrogen(ethylenedinitrilo)rathen- yields a solution ~2 mM in NO.
ate], and dilute this in the same anaerobic Dilute, to a final concentration of 50 mM,
buffer to a final concentration of 50 uM of the stock NO solution with the anaerobic
the complex (Solution A), buffer, using the procedure described for
CO above (Protocol 3),
Method
i Fill two optical cuvettes with the anaerobic solution of the ruthenium complex, and
using one as a sample and the other as a reference take the baseline between 220
and 400 um. Gently bubble a little NO gas through the sample solution and take the
spectrum again. This is the difference spectrum between the Ru-NO complex and
the Ru-chelate (see Figure 5),
219
M. T. WILSON AND j. TORRES
Protocol 4 continued
Transfer the remainder of the solutions to the stopped-flow apparatus that has
been previously flushed through with anaerobic buffer.
Set the monochromator to the wavelength of the maximum absorbance difference
in the spectrum (here -290 nm) and record a kinetic trace (Figure 6).
Record time courses at 5-10 nm intervals, being sure to pass through the isosbestic
point indicated in Figure 5.
Plot the absorbance change (corrected at each wavelength) against wavelength, as in
Figure 5. It is clear that the spectral distribution of the absorbance changes follows
the absorption spectrum (there is a factor of two between the amplitude that is
accounted for by the twofold dilution that occurs in the stopped-flow mixing).
Wavelength (nm)
220
STOPPED-FLOW SPECTROSCOPY
221
M. T. WILSON AND J. TORRES
Protocol 5
An experiment to determine the effect of a 'pH jump' on
the absorption of carboxymethylated horse heart
cytochrome c
Reagents
• Prepare a solution of ~5 ml of the protein Prepare a stronger buffer at the desired
(—10 uM) in a weak buffer (say 10 mM pH value, e.g. 0.1 M sodium phosphate
sodium acetate pH 5.5), pH7.
Method
i Fill one drive syringe with the protein solution and fill the other drive syringe with
the strong buffer solution.
Set the wavelength to give the maximum absorbance difference for the pH-
dependent change, determined separately as described above.
Discharge several 'shots' to flush out old reagents/buffer and then record and
average the data from several reactions.
0.25
0.2
0)
0.15
o
c
n
€o
0.05 -
222
STOPPED-FLOW SPECTROSCOPY
A+B- or A+B C+ D
0.12
-0.04
Time(ms)
Figure 8 The same experiment as shown in Figure 7 except the final pH was 9. The inset
table illustrates that with biphasic traces, here two exponentials, the value of rate constants
(k s-1) determined by the fit depend critically on the fit range chosen.
223
M. T. WILSON AND J. TORRES
Integration of the rate equations for the above yield complex and unwieldy
expressions and if true second-order reversible reactions are to be analysed it is
often best to use global analysis to retrieve the rate constants, see below. How-
ever, in many instances the back reaction may be neglected and the reaction
considered as quasi-irreversible. Under these circumstances the rate equation
may be written:
dx
— = k, ([Ao] - x) ([B0] - x) 10
where x is the concentration of the product(s) and [Ao] and [B0] are the initial
concentrations of A and B mixed in the stopped-flow spectrophotometer. The
second-order rate constant (k1; units M-1 s-1) has the form
fct1- n
~ *)
([Ac] - [Bo]) [Ao]([B0] - x)
There are a number of ways to use this equation in the analysis of data but the
simplest way to proceed is to use this equation directly with the fitting program
provided with the apparatus. There is one problem to bear in mind, however, if
such a fitting procedure is used, namely, the routine will be executed using the
measured absorbance values and not the concentration of the reactants. This means
that the second-order rate constants will have the units of AA-1 s-1 and thus
must be converted by use of the appropriate Aex to M-1 s-1.
This equation may be considerably simplified if one conducts the experiment
under conditions in which one of the reactants, say A, is in large excess (>10-
fold over the other). Under these conditions the concentration of the com-
ponent in excess can be considered to remain constant during the reaction, i.e.
[Ao] » x. Under these circumstances x collapses to
x = [B0](l - exp(-kobst)) 12
that has the same form as that describing a first-order process. Here kobs, the
pseudo-first-order rate constant has the form Jcobs = k1[Ao], i.e. the second-order rate
constant multiplied by the concentration of the component in excess.
Under pseudo-first-order conditions ([Ao] » [B0]) reversible second-order
reactions also yield the same exponential forms as above but now
=
^obs M[AO] + k-1 13
Thus under these conditions both k1 and k-1 may be found from experiments in
which the pseudo-first-order rate constant (kobs) is determined at a number of
concentrations of the reactant in excess. The data presented in Figure 3, which
was used to measure the dead time, are the results of such an experiment. In
the inset to Figure 3 the pseudo-first-order rate constants (kobs) derived from
exponential fits to the data are plotted against the CO concentration. The slope
gives the second-order rate constant k1, ~4 x 105 M-1 s-1, and the intercept
k-1, 0.6 s-1. The latter value is rather high compared with the literature value of
0.04 s-1 and illustrates the limitation of the method when k-1 is very small (5).
224
STOPPED-FLOW SPECTROSCOPY
The rate constant of any first-order process is related to the half-time of the
reaction (i.e. the time taken to complete half the reaction) by the expression
Kobs = 1n2/t1/2y. ~ 0.693/t,1/2.
Essentially irreversible second-order processes may also be analysed easily if
experiments are performed under condition in which the starting reactants are
at equal concentration, i.e. [Ao] = [B0]. Under these circumstances the solution of
the rate equation gives
This simple analytical equation may be used in fitting procedures (see Figure 6).
The rate constant is related to the half-time of such a reaction by the expression
k1, = l/[Ao]t1/2,
It can readily be appreciated that if k1[S] is much larger than k2 then on mixing
E and S there is an initial rapid production of p1, the 'burst', the concentration
of P1 being stoichiometric with the initial concentration of the enzyme [E0].
Indeed as it is often easier to measure [P1]than [E0] this provides an excellent
way to measure the latter and is the basis of active site titrations. Once the
enzyme has been acylated it decays to form P2, and in the presence of excess
substrate the cycle is repeated, but in the second and subsequent cycles the
enzyme cannot react with S faster than the enzyme deacylates. Provided [S] »
[E0] this leads to a linear time course for Pt formation, with rate dependent on
the value of k2.
Where k1[S] and K2 are comparable the situation is more complex and analysis
of the mechanism leads to the conclusions depicted in Figure 9. This diagram
gives the basis of active site titrations in which the enzyme is mixed with
225
M. T. WILSON AND J. TORRES
[p.;
Time
figure 9 Schematic diagram illustrating the 'burst' seen in the concentration of p1 on mixing a
protease with its substrate. The amplitude of the 'burst' and the rates are determined by the
kinetic parameters given in the text as are the concentrations of the enzyme and substrate.
0.09
-0.01
Time (seconds)
Figure 10 The reaction between rat tissue Kallikrein (5 (J.M) and p-nitrophenylguanidino
benzoate. Kallikrein (5 uM) was reacted with 6.25, 12.5, 25, 50, 100, 250 and 500 (uM p-
nitrophenylguanidino benzoate. The monitoring wavelength was 402 nm where p-nitrophenol,
P! absorbs. The temperature was 25°C.
226
STOPPED-FLOW SPECTROSCOPY
227
M. T. WILSON AND J. TORRES
obtained at a given time point and each row represents a time course at a given
wavelength. Capital bold letters denotes matrices. Thus A = a,
ti t2 t3 tm
•a11
a21 a22 16
X,
It is often the case that the reaction analysed is not simple and may possibly
contain more than one substrate (or product) and one or more intermediates on
the reaction pathway. This increased number of unknowns in the system can be
successfully determined with the additional information provided by a multi-
wavelength experiment. Obviously, the accuracy of this determination will de-
pend on the amount of information available, and it will be more precise the
more data points are taken in the spectral region or time period of interest.
Thus, a whole process is composed of partial reactions, each with its own rate
constant and spectral components. This information is contained in matrix A,
together with random noise inherent in any such experiment, and can be
extracted using a deconvolution procedure, based on matrix algebra methods,
called singular value decomposition (SVD), followed by global fitting. What is required
is that we analyse the matrix in such a way that we can determine the mini-
mum number of processes contributing to the overall spectral change, i.e. the
number of independent time courses and spectral components which together
can combine to form the whole data set. Then, using this information one fits
the data set to a mechanistic model of the reaction thereby yielding the rate
constants and spectra of all species.
In what follows we give a theoretical outline of the methods and then take
the reader through the steps using a practical example. For further information
and examples the reader may consult references (7-9).
228
STOPPED-FLOW SPECTROSCOPY
20
[C] = [A] 1 + 21
The absorbance at any time point is expressed as the sum eA[A] + eB[B] + ec[C],
and the linearly independent vectors will be eA, EB and ec. As before, certain
combinations of these vectors, e.g. eA, eB and (ec - eB) or (eA - eB) and ss and ec,
are equally valid. In any case, the minimum number of linearly independent
vectors in this system is three.
The number of linearly independent vectors is, however, not necessarily the
same as the number of spectrally distinct species. This is made clearer by
consideration of the next system, which can be defined by three linearly
independent vectors, even though it contains four spectrally different species:
22
C^D
As in the first example,
[A] = [AJ e^'
[B] = [A] (1 - e-k1t*)
and for the second reaction,
[C] = [C] e-k2T
[D] = [C] (1 - e-K2T)
Hence, the absorbance at any time point is represented by eA[A] + eB[B] + ec[C]
+ cD[D], and substituting for the expressions above:
„ ijiQi
fc fi 1 ca~klt -L. ,, FA 1 11
i^ &DI/IQJ It — cf*-k~lt\i ~i ^CL -^2f -4- o Dl\C oJ1 M
„ [p oj1 0*" \ — fi~^2t\I oq
***J
A
229
M. T. WILSON AND J. TORRES
230
STOPPED-FLOW SPECTROSCOPY
231
M. T. WILSON AND J. TORRES
232
STOPPED-FLOW SPECTROSCOPY
233
M. T. WILSON AND J. TORRES
Protocol 6
Reduction of the metmyoglobin-cyanide complex by
sodium dithionite
Reagents
• Horse skeletal muscle myoglobin, • Phosphate buffer, 0.2 M, pH 6.4
supplied as the ferric form MbFe(III) • Potassium cyanide (Sigma)
(Sigma) • Sodium dithionite (BDH)
Method
1 Dissolve MbFe(III) in buffer. To measure the Mb concentration, add sodium dithio-
nite to an aliquot of the solution of MbFe(III) to obtain deoxy Fe(II} (s560 = 13800
M-' cm-1).
2 Prepare a fresh solution of KCN (1 mM) in phosphate buffer at pH 6,4.
3 Add KCN to the MbFe(M) solution to obtain the MetMb-CN complex. Final con-
centrations of 200 uM KCN and 100 uM Mb, approximately. After allowing 1 min to
allow complex formation gently deoxygenate the solution by passage of N2.
4 Prepare fresh sodium dithionite solutions adding the solid to degassed phosphate
buffer at concentrations 100, 25 and 6.2 mw.
5 Mix the MetMb-CN complex with the sodium dithionite solution in a stopped-flow
apparatus.
Syringe A: 100 uM MetMb-CN complex in phosphate buffer at pH 6.4.
Syringe B: Sodium dithionite at concentrations 100, 25 and 6.2 mM in phosphate
buffer at pH 6.4.
6 Collect the spectra in the range 400-600 nm using a diode array.
234
STOPPED-FLOW SPECTROSCOPY
0.7
0.6
0.5-
s°
c
CO
•fi 0.3-
8
_D
< 0.2-
0.1 -
0.0-
Wavelength (nm)
Figure 12 The reaction between the cyanide complex of ferric myoglobin (horse heart) and
sodium dithionite. A subset of the total number of spectra collected is shown. An initial
increase in absorbance is seen at 435 nm and in the visible region, spectra displayed every
20 ms, followed by a slower decrease, only the final spectrum (thicker line) is shown for
clarity.
235
M. T. WILSON AND J. TORRES
Protocol 7
The SVD procedure
Method
1 Decompose the matrix A into the matrices U, V and S using the option provided in
the stopped-flow software, Matlab or other suitable package.
2 Determine the rank of the matrix using any of the methods described above.
Depending on the rank of the matrix a mechanism is chosen to fit the data,
3 Plot the first six columns of the V matrix versus time and of the U column versus
wavelength.
Plots of the U columns (figure l3) and the V columns (Figure 14) show that the
rank is either 3 or 4. The same conclusion is reached by looking at the logs plot
(Figure 15) or at the s values and autocorrelation coefficients for the columns in
V and U (Table 2). For the remaining analysis the rank was taken as three, the
fourth s value being much smaller than the third. A very small fourth com-
ponent is in fact present and was identified as being due to a small fraction of
denatured myoglobin reacting differently from the bulk protein.
As the rank is three, mechanisms compatible with this were chosen for global
fitting to the truncated (reconstructed from the first three columns) matrix A.
Only the sequential mechanism A -- B * C gave a sensible solution, i.e. ail
kinetic and spectral profiles fitted well, and non-negative absolute spectra for all
236
STOPPED-FLOW SPECTROSCOPY
0,40 0,4
Ucol 1
0,2
0,20
0,0
0,00 -0,2
0,4
Ucol3 Ucol 4 0,4
0,0 0,0
-0,4
-0,4
0,00 0,0
-0,50 -0,5
VcoM
0,0
-0,1
-0,2
0,18 Vcol3 Vcol4
0,12 0,0
0,06
-0,2
0,00
-0,06 -0,4
0,30
VcolS vcoie 0,3
0,00 0,0
-0,30 -0,3
-0,60 -0,6
0,01 0,1 1 10 0,01 0,1 1 10
Log10 Time(s) Log10 Time(s)
Figure 14 The first six columns, time courses, of the V matrix following SVD of the total
data set; matrix A.
components. The results of the fitting process are shown together with the
dithionite concentration dependencies of the two processes in Figure 16 and 17.
Thus the results of this analysis may be summarized as follows:
Mb Fe(III)-CN + dithionite -* Mb Fe(II)-CN -» Mb Fe(II) + CN 40
237
M. T. WILSON AND J. TORRES
Figure 15 Plot of logs, against /, showing a break where signal gives way to noise.
s ACV ACU
47.9 0.973 0.998
2.74 0.913 0.982
1.14 0.898 0.978
0.14 0.888 0.605
0.06 0.193 -0.02
0.05 -0.596 -0.02
14
12
10
j 8
6
4
2
0
[Na2S2O4]1/2 (mM)1/2
Figure 16 The dithionite concentration dependence of the pseudo-first-order rate constants
of the fast and slow processes seen in the reduction of Mb-CN.
238
STOPPED-FLOW SPECTROSCOPY
0.6-
0.4-|
0.2-
0.0
400 450 500 550 600
Wavelength (nm)
Figure 17 Global analysis of the reduced (rank = 3) matrix A to a simple sequential model
generates the spectra of the chemical species A, B and C, see text.
239
This page intentionally left blank
Chapter 9
Stopped-flow fluorescence
spectroscopy
Michael G. Gore* and Stephen P. Bottomleyf
* Institute of Biomolecular Sciences, Division of Biochemistry,
School of Biological Sciences, University of Southampton, Bassett Crescent East,
Southampton, S016 7PX
f
Department of Biochemistry and Molecular Biology, Monash University, Clayton,
Victoria 3168, Australia
1 Introduction
Biochemical reactions, such as substrate or coenzyme binding to enzymes are
usually completed in no more than 50-100 ms and thus require rapid reaction
techniques such as stopped-flow instrumentation for their study. Fortunately,
many such reactions can be followed by changes in the absorption properties of
the substrate, product or coenzyme, and examples of these have been described
in Chapters 1, 7 and 8. An alternative possibility is that during the reaction
there is a change in the fluorescence properties of the substrate, coenzyme or
the protein itself. Some reactions, particularly those involving the oxidation/
reduction of coenzymes, involve both changes in absorption and changes in
fluorescence emission intensity. In many cases, the fluorescence properties of
the ligand or protein itself may change when a complex is formed, even in
the absence of a full catalytic reaction occurring, e.g. the protein fluorescence
emission of most pyridine or flavin nucleotide-dependent dehydrogenases
is quenched when NAD(P)H or FADH (respectively) binds to them, due to
resonance energy transfer from the aromatic amino acids of the protein to the
coenzyme. Conversely, the fluorescence emission from the reduced-coenzymes
is usually enhanced on formation of the complex with these enzymes (1-3).
The principles behind both fluorescence and stopped-flow techniques have
been described in preceding chapters (2 and 8, respectively) and therefore
readers should familiarize themselves with these chapters for some of the back-
ground information. In this chapter, we discuss the use of stopped-flow fluor-
escence spectroscopy and its application to a number of biochemical problems.
2 Instrumentation
A typical stopped-flow system is assembled from modular components of a
conventional spectrophotometer/fluorimeter, a device permitting rapid mixing
of the components of a reaction and a data recording system with a fast re-
sponse. Commercially available instruments offer facilities for the observation
of changes in absorption and/or fluorescence emission after rapid mixing of the
241
MICHAEL G. GORE AND STEPHEN P. BOTTOMLEY
PMT1
reagents. These measurements can often be made simultaneously due to the dif-
ferent optical requirements of the two spectroscopic techniques. Figure 1 gives a
generalized diagram of the geometry of a stopped-flow system able to simultane-
ously measure changes in absorption and fluorescence intensity of a reaction.
Note that the flow-cell is illuminated by light selected by monochromator 1
(MCM1) and that the change in light absorption at this wavelength is detected at
180° relative to the light beam. Changes in fluorescence intensity are detected at
90° relative to the excitation light beam via a second monochromator (MCM2), a
narrow band pass filter or a cut-off filter to select the emission wavelength.
242
STOPPED-FLOW FLUORESCENCE SPECTROSCOPY
converter. For example, if 1000 data points are used to describe an observation
period of 100 ms and the effective sampling rate permitted by the A/D converter
is 100 kHz, then 10 samples of the signal will be averaged for each datum point.
The signalnoisc ratio will improve by the square root of the number of samples
in the average, and thus it is clear that a reaction progress curve will have less
noise the slower the reaction and, therefore, the greater the number of signal
sampling events for the same number of data points (1000 in this example}. For
very fast reactions the result must usually be improved by averaging the data
from several reactions. However, the amount of noise observed on a reaction
profile will also depend upon the amplitude of the signal change upon reaction
and how careful the experimenter has been in selecting the conditions for the
experiment, the selling up of the instrument and the preparation of the re-
agents. All of the recommendations made in previous chapters regarding the preparation
of solutions arc particularly important in stopped-flow fluorescence studies.
Protocol 1
To check the mixing efficiency of the stopped-flow
instrument
Equipment and reagents
• 0,1 Msodium phosphate buffer pH 6.5 1 ml of a stock solution of 10 mM 4-
(buffer A) methylumbelliferone in water
• 0.1 M sodium phosphate buffer pH 8.5 Disposable 5-ml syringes and a few glass
(buffer B) beakers
• Pipettes and measuring cylinders
243
MICHAEL G. GORE AND STEPHEN P. BOTTOMLEY
Protocol 1 continued
Method
1 Mix 50 ml of buffers A and B to give 100 ml of buffer, approximately pH 7.1 to give
C.
2 Add 0,1 ml of the stock solution of 4-methylumbelliferone to 19.9 ml of the mixed
buffer C to give D.
3 Add 0.2 ml of the stock solution of 4-methylumbelliferone to 19.8 ml of buffer A to
give E.
4 Equilibrate all solutions at the operating temperature of the stopped-fiow
observation cell.
5 Use solution C to wash out the contents of the two drive syringes of the stopped-
flow apparatus. Refill the two syringes with C and discharge five or more reactions
to wash out and fill the observation cell,
6 Set the excitation wavelength to 388 nm and use a cut-off filter to select emitted
light at 410 nm and above.
7 Flush out the system with solution D and adjust the sensitivity of the instrument to
give a working fluorescence signal. The working signal should be the same as that
obtained in the following procedure.
8 Replace the contents of one drive syringe with solution E and that of the other
drive syringe with solution B.
9 Discharge five or more 'shots' to replace the solutions in the observation cell.
10 Then discharge several more and record any change in signal intensity with time.
Results
On mixing the two solutions E and B, the pH of the reaction mixture will change to that
of solutions C and D (pH 7.1) within the dead time of the instrument, and the degree of
ionization of the fluorophore (originally at pH 6,5) will alter. Observation of the fluor-
escence intensity of the solution at 410 nm and above should yield an 'instantaneous'
increase in fluorescence intensity (within the dead time of the instrument) followed by a
stable signal intensity. Any change in fluorescence intensity over the first 10-50 ms after
mixing suggests that poor mixing is occurring.
244
STOPPED-FLOW FLUORESCENCE SPECTROSCOPY
use. Degas buffers to remove dissolved air that may otherwise come out of
solution following rapid decompression of the reaction solutions and filter all
solutions using an appropriate filter, e.g. Millipore 0.45 um. Sometimes sample
loss may occur during filtration in which case centrifugation of the sample may
help eliminate suspended particles without compromising the sample.
2.5 Artefacts
It is important to verify that the signal change being monitored is due to the
reaction under study and not due to artefacts. Comparison of signal changes
noted under equilibrium conditions should be compared with the kinetic differ-
ence spectrum obtained in the stopped-flow (see Chapter 8). If deviations are
noted then the observed change in fluorescence may well be due to an artefact
and not reflect the reaction under study.
245
MICHAEL G. GORE AND STEPHEN P. BOTTOMLEY
246
STOPPED-FLOW FLUORESCENCE SPECTROSCOPY
247
MICHAEL G. GORE AND STEPHEN P. BOTTOMLEY
0.5-
248
STOPPED-FLOW FLUORESCENCE SPECTROSCOPY
BlsSrPr^
1
E ° Time(s)
Figure 4 Signal intensity and noise level; the effect of slit width and voltage to PMT. A plot
showing the recorded traces from an Applied Photophysics SX.17MV stopped-flow fluorimeter
when 20 uM warfarin (final concentration) in 10 mM potassium phosphate buffer at pH 7.4,
20°C, was mixed with buffer. The fluorescence intensity at 360 nm was measured using
an excitation wavelength of 330 nm. The sensitivity of the instrument was adjusted to give
a similar signal intensity by varying the slit width and the high voltage to the PMT (see
Figure 3).
low dynode voltage (350 V). This is generally the case and hence the amount of
light available and the optical alignment of the instrument are very important
considerations for achieving high sensitivity.
249
MICHAEL G. GORE AND STEPHEN P. BOTTOMLEY
residues, although proteins lacking Trp residues or with many Tyr will show
fluorescence emission from Tyr residues.
The quantum yield and emission maximum of the Trp residue is highly de-
pendent upon the microenvironment of the residue. Trp residues in the buried,
apolar interior of a protein can have emission maxima as low as 308 nm and
those exposed to aqueous solvent have an emission maximum close to 350 nm
(see Section 4.3 and references 37, 38, and 40 in Chapter 12). It is this sensitivity
to environmental changes that makes the Trp residue such an ideal reporter
group. Furthermore, the judicious use of site-directed mutagenesis, particularly
with the aid of structural knowledge of the protein, allows the removal of
unwanted Trp residues or the addition of unique Trp residues to a protein or to
a domain of a protein (6-8). Recent developments in molecular biology now
allow modified Trp residues (e.g. 5-hydroxy-L-tryptophan (9)) with red-shifted
excitation and emission spectra to be located at key positions in proteins. These
residues can be selectively excited, permitting spectral changes to be identified
from specific parts of the protein without having to risk some destabilization of
the protein by the removal of other Trp residues by mutagenesis. Thus Trp
residues can be used as extremely sensitive conformational reporter probes.
250
STOPPED-FLOW FLUORESCENCE SPECTROSCOPY
4.5
4.0
3.5
3
'°
2.5
2.0
1.5
0 1 2 3 4 5 6 7
GdnHCI [M]
Figure 5 The dependence of the rates of folding and unfolding of non-reduced short
consensus repeat (SCR) domains from human complement receptor 1 on the concentration
of Gdn-HCI. Unfolding reactions were achieved by mixing 1:10 v/v 50 uM of a protein
consisting of the first three domains, SCR1-3 (•), or a protein consisting of the single third
domain, SCR3 (•) of complement receptor 1 with various concentrations of Gdn-HCI in
20 mM potassium phosphate buffer, pH 7.4 at 25°C. Refolding reactions were achieved by
mixing the proteins in 3-7 M Gdn-HCI 1:10 v/v with buffer containing various concentrations
of the denaturant to give the final concentrations of Gdn-HCI shown above. This figure has
been reproduced from Clark, N. S., Dodd, I., Mossakowska, D. E., Smith, R. A. G. and Gore,
M. G. Protein Engineering (1996) 9, 10, 877 by permission of Oxford University Press.
domain of complement receptor 1. The minima of the curves occur at the con-
centration of denaturant where the rate of refolding is the same as the rate of
unfolding. That is the concentration of denaturant that gives the midpoint of an
unfolding/folding transition where half of the protein is in each of the unfolded
and folded forms (see Figure 4, Chapter 12). These rates should be compared,
whenever possible, to rates of unfolding determined by stopped-flow CD
measurements (far UV) which reflect changes in the backbone structure of the
polypeptide chain (see Chapters 4,10 and 12).
251
MICHAEL G. GORE AND STEPHEN P. BOTTOMLEY
252
STOPPED-FLOW FLUORESCENCE SPECTROSCOPY
Figure 6 The binding curves obtained on mixing K chains with a mutated immunoglobulin-
binding domain of protein L (SpL) from Peptostreptococcus magnus. The curves show the
changes in protein fluorescence intensity from unique tryptophan residues (W39 or W64) as
each domain binds to K chains at pH 8.0 in potassium phosphate buffer, 20°C. The
concentration of K chain was 3 (AM and the concentration of each mutant of SpL was 30 uM.
These data were kindly provided by Dr J. Beckingham and N. Housden.
chain kept constant). In each case, the rapid phase of change in fluorescence is
followed by a slower, single exponential process with a rate that is independent
of the total concentration of SpL used. This is usually indicative of a structural
change occurring in the complex or in either reactant.
There are several possible models that may, in principle, be able to explain
these results. In order to analyse what is happening in this or any reaction we
must first identify the step or steps in the reaction that lead to signal changes;
in this case the stages at which the rapid and slow changes in fluorescence
occur.
One possible model (a) is given below;
253
MICHAEL G. GORE AND STEPHEN P. BOTTOMLEY
some SpL will isomerize to SpL*. This gives rise to a further change in
fluorescence intensity, an increase for the reaction involving W39 mutant and a
decrease for the reaction involving the W64 mutant. If kf is not much smaller
than k1 [K] then more complex analysis is needed. Figure 7 gives an example of
reaction conditions that permit observation of the second phase of the reaction
in virtual isolation from the initial phase. In this experiment a high concen-
tration of SpL (a non-mutated domain containing no Trp residues) was reacted
with K chain and the change in fluorescence from Tyr residues was monitored
at 305 nm (X.ex = 280 nm). The high rate of the initial phase of the reaction
(shown by a decrease in fluorescence intensity noted when this domain
binds to K chain) is virtually completed at the point indicated by the arrow
(approximately 200 ms). These conditions effectively separate out the two parts
of the reaction, and allow the rate of the slower second phase to be deter-
mined without much contribution from the fast phase of fluorescence change,
thus facilitating curve fitting algorithms (see Chapter 8, Section 4.1.1 and
Figure 8).
The exact form of binding curves for multi-phase reactions such as this will
be complex. Amongst other considerations, the relative amplitudes of the two
phases of signal change will depend upon the position of the equilibrium be-
tween SpL and SpL*. If SpL predominates then the amplitude of the slow change
in fluorescence will be larger than if SpL* predominates. Furthermore, the
1.96
Time (s)
Figure 7 The reaction progress curve obtained on mixing K chains with a non-mutated SpL
domain lacking a Trp residue. The curve show the changes in protein fluorescence intensity
at 305 nm (excited at 280 nm) from tyrosine residues as the domain binds to K chains at
pH 8.0 in potassium phosphate buffer, 20°C. The concentration of K chain was 4 uM and
the concentration of SpL was 70 uM. These data were kindly provided by N. Housden.
254
STOPPED-FLOW FLUORESCENCE SPECTROSCOPY
Here, k1 (M-1 s-1) is the second order rate constant for the formation of the binary
complex and k-1 (s-1) is the first order rate of dissociation of the complex. The
rate of the fast change of fluorescence intensity is proportional to the con-
centration of K chain and represents the rate of formation of SpL.K. Providing
that the rate of formation of the encounter complex SpL.K is rapid compared to
the unimolecular isomerization step (i.e. the two parts of the process can be
considered as two separate reactions), k1 and k-1 may be derived from the slope
and intercept, respectively, of a plot of kobs against the final concentration of K
chain. The Kd for the encounter complex is approximated by k -1 /k 1 -
The observed forward rate for the conformational change is the sum of
the forward and reverse rates of the conformational change between SpL.K
and SpL*K. This is given by fkf + kr in which f is the fraction of the SpL present
as SpL.K. This is readily calculated by considering the initial part of the reaction
i.e.
SpL + K ^ SpL.K
k-i
= [SpL][K] = k_! 3
d
[SpLK] k,
In which [SpL], [K] and [SpL.K] are the concentrations of these components at
equilibrium. Hence [SpL.K] = [SpL][K]/Kd and the fraction of SpL present in the
complex
[SpLK]
[SpL] + [SpL.K]
= [K] ^
+ Kd 5
255
MICHAEL G. GORE AND STEPHEN P. BOTTOMLEY
k1 1+
Which is another form of the equation (3)
+ k r /kf
Note that if kf is very large compared to k,. then these equations simplify to
256
STOPPED-FLOW FLUORESCENCE SPECTROSCOPY
Resolving whether scheme (a) or (b) is the correct model for such a reaction is
often difficult because one can rarely be certain where the changes in fluor-
escence actually occur. Other supporting experimental data are often required.
It should be emphasized that as the rates of a biphasic process become closer
then the method of analysis is more complex and readers should refer to other
texts (see Section 4.3 and references therein).
Diffusion controlled rates of bimolecular association reactions are usually
around 107-108 M-1 s-1 in agreement with theoretical modelling studies. These
high rates are often not obtained experimentally which implies that something
else, such as an isomerization in either the reactant or the complex, is limiting
the rate. However, there are other possible reasons for association rates lower
than expected and these have been reviewed in (17,18).
Finally, a cautionary note. Reaction progress curves exhibiting more than one
rate may not be due to complex kinetic behaviour in the reaction under in-
vestigation, but may be caused by a combination of real spectral changes due to
the chemical reaction under study plus spectral changes due to another un-
known process. For example, it is very common for a protein to become photo-
degraded when illuminated by intense UV radiation such as that given out by a
xenon lamp. This is particularly evident in fluorescence experiments where
high intensity excitation light is used to increase the sensitivity of the system;
thus, slow decreases in fluorescence intensity may be noted over long observa-
tion periods. Alternatively, of course, the ligand may be similarly unstable.
Control experiments in which each reagent is mixed with buffer in the stopped-
flow should always be carried out to check for such degradation processes and
to determine the expected fluorescence intensities of the control at the start
and end point of the reaction.
257
MICHAEL G. GORE AND STEPHEN P. BOTTOMLEY
The critical differences between this reaction and model (b) for the reaction
between SpL and K chain (described above) is that the change in fluorescence is
limited to the unimolecular step (5) and the forward rate of the conformational
change (kf) is relatively fast. The inset to Figure 8 shows that as the concentration
of warfarin is increased, the observed rate of change of fluorescence also
increases. Over this range of concentration of ligand, the bimolecular reaction
is slower than the unimolecular conformational change and is therefore
rate limiting. However, at high concentrations of warfarin, the rate of the
bimolecular step exceeds that of the conformational change and the observed
rate of reaction becomes limited by the latter step.
In a two step reaction such as this with only one source of signal change
at the isomerization step the observed rate of the reaction is as described by
equation
0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0
Figure 8 A reaction progress curve obtained when 20 JJLM warfarin was mixed with 1 um
HSA. In this reaction, 1 um albumin was reacted with 10 uM warfarin (final concentrations) in
100 mM potassium phosphate buffer at pH 7.4 also containing 0.8% NaCI, at 25 °C. The
fluorescence of the warfarin was excited at 330 nm, and the emission intensity at 360 nm
and above was selected by a cut-off filter. The inset shows a plot of the pseudo-first-order
rates, kobs, for the reactions between various concentrations of warfarin and 1umHSA
against the final concentration of warfarin used. We are indebted to Miss Sue Twine for the
provision and use of these data.
258
STOPPED-FLOW FLUORESCENCE SPECTROSCOPY
in which Kd, = f -1 /k, and [W] is the concentration of warfarin. From equation 11 it
can be seen that a plot of fe,,^ against |W] gives a hyperbolic curve as shown in
figure K. The maximum value of kobs at high concentrations of warfarin can be
obtained from the plot shown in the inset to Figure 8 and equals Kf + k,.,
Rearranging equation 11 shows that a plot of l/(k obs , - Kr) against 1/|W] gives a
straight line of slope K d /k f and an intercept on the ordinate of 1/k,-. The value of kr
can be determined by a displacement reaction using a non-fluorescent com-
pound that binds at the same site on albumin as warfarin, e.g. phenylbutazone,
For a detailed kinetic analysis of this reaction see (5).
Protocol 2
Determination of kobs for the binding of warfarin to human
serum albumin
Equipment and reagents
• A stock solution of 100 um human serum * A 360 nm cut-off filter to select the
albumin (HSA) in 100 mM sodium phos- emission wavelength
phate buffer pH 7.4 containing 0,8% NaCl • Some disposable 5-ml syringes and some
• A stock solution of 3 mM warfarin in the small beakers/tubes
same buffer
Method
1 DiJute some of the warfarin solution to 50 umusing the phosphate buffer.
2 Dilute the HSA to 2 um in phosphate buffer.
3 Set the excitation and emission wavelengths to 330 nm and 360 nm (and above),
respectively.
4 Fill both of the stopped-flow drive syringes with buffer and discharge 5-10 'shots'
of these to wash the observation cell.
5 Then replace the buffer in one drive syringe with the 50 uM warfarin. After dis-
charging five 'shots' to flush out the buffer in that side of the system, discharge five
more. After these adjust the slit width of the monochromator selecting the
excitation wavelength and the high voltage on the PMT to obtain a reasonable
working signal. Record the signal voltage (unbound HSA). Note that this will be
lower than that obtained when warfarin is in the presence of HSA.
6 Refill the syringes with phosphate buffer and discharge 10 'shots' and record the
signal voltage (buffer baseline).
7 Now fill one drive syringe with warfarin and the other with 2 uM HSA, discharge
five to six 'shots' to flush out the system. During the last one or two reactions you
should notice that the signal voltage increases as the HSA and warfarin mixture
reaches the observation cell.
8 Discharge five reactions and record the reaction progress curves. These should be
single exponential curves with kobs about 35 s-1 at 25°C.
259
MICHAEL G. GORE AND STEPHEN P. BOTTOMLEY
Protocol 2 continued
9 The warfarin solution tan then be replaced with another at a higher concentration
and the process repeated to obtain a curve such as that shown in the inset to Figure 8.
Remember that if the system has a 1:1 v:v mixing ratio then the final concen-
trations of the reagents will be half that used to fill the syringes.
4 4 Extrinsic probes
Extrinsic probes must be used when the system under study has no useful
intrinsic reporter groups or a reaction produces no fluorescence change from
intrinsic reporter groups. In such cases one may use an appropriate fluorescent
group which is added to the system to report the reaction. Extrinsic reporter
groups come in many forms such as non-covalently and covalently bound
fluorescent labels. Many of the latter can be covalently attached to the protein
of interest at a specific site.
260
STOPPED-FLOW FLUORESCENCE SPECTROSCOPY
261
MICHAEL G. GORE AND STEPHEN P. BOTTOMLEY
reaction), may lead to spurious signal changes during experiments. It has been
found in this laboratory that passage of a labelled protein down a short gel-
filtration column to remove excess reagent is preferable to dialysis, as many of
the fluorescent reagents will adsorb to the column matrix and leave the protein.
The presence of fluorescent groups such as pyrene, fluorescein, eosin etc.
may also give characteristics to the protein or ligand that were previously not
present. For example, such groups on ligands generally decrease water
solubility and sometimes cause the compound to adsorb to the walls of plastic
containers and cuvettes.
262
STOPPED-FLOW FLUORESCENCE SPECTROSCOPY
500 -
430 -
7 360 -
290
220 -
150 -
Mg2+ [UM]
Figure 9 The effect of pH on k1 and k-1 for the binding of Mg2+ to pyrene-labelled
recombinant inositol monophosphatase. Enzyme (3 uM final concentration) was reacted with
various concentrations of Mg2+ ions in 50 mM Tris-HCI buffer at pH 9.0 (•), pH 8.0 (•),
pH 7.3 (n) and pH 6.5 (O) at 20°C. The kabs for each reaction is plotted against the final
concentration of Mg2+ ions used. Reproduced with permission from Thorne, M. R., Greasley,
P. J. and Gore, M. G. (1996) Biochemical Journal, 315, 989-994. © Biochemical Society.
The advantage of this method is that the protein is kept at pH 8.0 and only
exposed to low or high pH solution for a few ms.
It is interesting to note that in this case it is the high rate of dissociation of
the metal ion (k-1) that limits the ability of the stopped-flow technique to deter-
mine k1 at high concentrations of Mg2+ ions at pH 7.0 or below. The data in
Figure 9 suggest that k-1 is approximately 360 s-1 and k1 is approximately 3.7 x
105 M-1 s-1 at pH 6.5. Therefore in the presence of 300 uM Mg2+ ions the forward
rate (k1[Mg2+]) will be 111 s-1 and kobs will be approximately 461 s-1. Thus the
half-time of the reaction will be only 1.5 ms, and much of the reaction will take
place within the dead time of the instrument.
References
1. Holbrook, J. J. (1972) Biochem.j. 128, 921.
2. Holbrook, J. J. Yates, D. W., Reynolds, S. J., Evans, R. W., Greenwood, C. and Gore, M.
G. (1972) Biochem.j. 128, 933.
3. Chaiyen, P., Brissette, P., Ballou, D. P., and Massay, V. (1997) Biochem. 36, 2612.
4. Ikushiro, H., Hayashi, H., Kawata, Y., and Kagamiyama, H. (1998) Biochem. 37, 3043.
5. Wilting, J., Kremer, J. M. H., Ijzerman, A. D. P. and Schulman, S. (1982) Biochim.
Biophys. Acta 706, 96.
6. Gore, M. G., Greasley, P. J., Knowles, M. R., Gee, N., McAllister, G. and Ragan, C. I.
(1993) Biochem.j, 236, 3, 811.
7. Waldman, A. D. B, Clarke, A. R., Wigley, D. B., Hart, K. W., Chia, W. N., Barstow, D.,
Atkinson, T., Munro, I. and Holbrook, J. J. (1987) Biochim. Biophys. Acta 913, 66.
8. Waldman, A. D. B, Hart, K. W., Clarke, A. R., Wigley, D. B., Barstow, D., Atkinson, T.,
Chia, W. N. and Holbrook, J. J. (1988) Biochem. Biophys. Res. Comm. 150, 752.
263
MICHAEL G. GORE AND STEPHEN P. BOTTOMLEY
9. Ross, J. B. A., Senear, D. F., Waxman, E., Kombo, B. B., Rusinova, E., Huang, Y. T.,
Laws, W. R. and Hasselbacher, C. A. (1992) P.N.A.S. USA 89, 24, 12023.
10. Clark, N. S., Dodd, I., Mossakowska, D. E., Smith, R. A. G., and Gore, M. G. (1996) Prot.
Eng. 9, 10, 877
11. Bottomley, S. P., Popplewell, A. G., Scawen, M., Wan, T., Sutton, B. J. and Gore, M. G.
(1994) Prot Eng. 7,12, 1463.
12. James, E., Whisstock, J., Gore, M. G. and Bottomley, S. P. (1999) J. Biol. Chem. 274,14,
9482.
13. Beckingham, J., Bottomley, S. P., Hinton, R., Sutton, B. J. and Gore, M. G. (1999)
Biochem.]. 340,143.
14. Wallis, R., Moore, G. R., James, R. and Kleanthous, C. (1995) Biochem. 34,13743.
15. Lee, A. G., Baker, K., Khan, Y. M. and East, J. M. (1995) Biochem. J. 305,1, 225.
16. Benson, S. W. (1960) The foundation of chemical kinetics. McGraw-Hill, New York.
17. Pontius, B. W. (1993) Trends Biochem. Sci. 18,181.
18. Schreiber, G. and Fersht, A. R. (1993) Biochemistry 32, 5145.
19. Piersma, S. R., Visser, A. J. W. G., de Vries, S. and Duine, J. A. (1998) Biochem. 37, 3068.
20. Spencer, P. D., Slade, A., Atkinson, A. and Gore M. G. (1990) Biochem. Biophys. Acta.
1040,130.
21. Clarke, A. R., Wigley, D. B., Chia, W. N., Barstow, D., Atkinson, T. and Holbrook, J. J.
(1986) Nature 324, No. 6098, 699.
22. Strickland, S., Palmer, G. and Massey, V. (1975) J. Biol Chem. 250,11, 4408.
23. Gutfreund, H. (1975) Enzymes: physical principles. Wiley, London.
24. Fersht, A. (1984) Enzyme structure and mechanism (2nd Edn). W. H. Freeman, New York.
25. Slavik, J. (1982) Biochim. Biophys. Acta. 694,1,1.
26. Jones, B. E., Beecham, J. M. and Mathews, C. R. (1995) Biochem. 34,1867.
27. Teschke, C. and King, J. (1995) Biochem. 34, 6915.
28. Walmsley, A. R., Martin, G. E. M. and Henderson, P. J. F. (1994) J. Biol. Chem. 269,
17009.
29. Gibbons, D. L. and Horowitz, P. M. (1995) J. Biol. Chem. 270, 3, 7335.
30. Lin, 7., Scharz, F. P. and Eisenstein, E. (1995) J. Biol. Chem. 270, 3,1011.
31. Wilton, D. C. (1990) Biochem.j. 266, 435.
32. Muller, N., Lapique, F., Drelon, E. and Netter, P. (1994) J. pharm. Pharmacol 46, 300.
33. Sudlow, E., Birkett, D. J. and Wade, D. N. (1975) Mol. Pharm. 11, 824.
34. Epps, D. E., Raub, T. J. and Kezdy, F. J. (1995) Anal. Biochem. 227, 342.
35. Henderson, I. M. J., Starling, A. P., Wictome, M., East, J. M. E. and Lee, A. G. (1994)
BtochemJ. 297, 3, 625.
36. Dalton, K. A., East, J. M., Mall, S., Oliver, S., Starling, A. P. and Lee, A. G. (1998)
Biochem.j. 329, 3, 637.
37. Greasley, P. J., Hunt, L. G. and Gore, M. G. (1994) Eur.J. Biochem. 222, 453.
38. Thorne, M. R.., Greasley, P. J., and Gore, M. G. (1996) Biochem. ]. 315, 989.
264
Chapter 10
Stopped-flow circular
dichroism
Alison Rodger* and Michael J. Careyf
* Department of Chemistry, University of Warwick, Coventry CV4 7AL
Applied Photophysics Ltd., 203/205 Kingston Road, Leatherhead,
Surrey KT22 7PB
± Introduction
As is apparent from previous chapters (Chapters 6, 8 and 9), understanding the
kinetics of chemical and biological processes is extremely important. Questions
we often consider, explicitly or implicitly, include: Has something happened
'instantaneously' or will it take 20 years? Does changing the conditions or
available reagents affect either the end product or the rate of a process? What
intermediates are produced during a reaction? Can we characterize any inter-
mediates? Do we need to remove them to prevent side reactions?
If some or all of the reactants or products are chiral, then circular dichroism
(CD) detection may be the ideal tool for following the kinetics of a reaction,
and if the half-life of the reaction is of the order of milliseconds to seconds or
even minutes then stopped-flow mixing of the reagents will almost certainly
be the appropriate choice of sample handling method. For reactions with half-
lives of a few minutes to tens of minutes the reagents can be mixed by hand in
a normal cuvette and the signal monitored at an appropriate wavelength. CD is
not well suited to kinetics on timescales of hours due to the baseline drift that
does occur (see Chapter 4, Section 2.5). Some CD spectropolarimeters have the
useful facility of being able to perform a wavelength scan at pre-set intervals as
well as monitoring continuously (except during the wavelength scan) at a
chosen wavelength, thus facilitating the characterization of any intermediates.
In this chapter we shall highlight some of the considerations of the stopped-
flow technique that are particularly relevant to CD experiments. Particular
problems may be encountered when performing CD (as opposed to other
detection methods) stopped-flow experiments. The measured signals are very
small (typical CD intensities are 0.1% or less of the absorbance signal), and the
noise level observed is particularly sensitive to any inhomogeneities or turbu-
lence in the samples. Also, as one of the main applications of stopped-flow CD is
in the study of protein folding and unfolding, samples are often very viscous
and/or corrosive, have significant absorbances due to buffers etc., and the
experiments often require wide and variable mixing ratios.
265
ALISON RODGER AND MICHAEL J. CAREY
2 Instrumentation considerations
In establishing a stopped-flow CD system for a particular laboratory a number of
different factors need to be taken into consideration. The most important of
these are:
(1) available budget;
(2) sensitivity required—the ease of averaging over multiple scans is an
important factor here;
(3) minimum dead time required;
(4) corrosiveness or adsorbance of the samples to be studied;
(5) type of flow system: stepper motor or pneumatic driven;
(6) size and design of optical cell;
(7) number of syringes and mixing stages required;
(8) mixing ratios required and whether these need to be varied from
experiment to experiment;
(9) sample viscosity;
(10) wavelength range for detection, wavelength scanning and bandwidth; and
(11) software.
In addition to the above, there is also the important question of whether the
instrument will be a stand-alone stopped-flow CD system, or a wavelength scan-
ning CD spectropolarimeter with stopped-flow attachment, or a stopped-flow
instrument where one of the detectors is a CD detector. Recent instrument
advances have complicated rather than simplified the decision about instru-
mentation. The importance of the above factors will be summarized before a
brief discussion of particular manufacturers and their current instrumentation
is given below.
266
STOPPED-FLOW CIRCULAR DICHROISM
267
ALISON RODGER AND MICHAEL J. CAREY
stances. The flowlines in some stopped-flow systems are made of stainless steel.
Although medical grade stainless steel is relatively inert, nonetheless, one
should avoid using corrosive or reactive reagents. Some biochemical measure-
ments can be compromised by the presence of ferrous ions. Whilst considering
the interaction between reagents and stopped-flow components, one should
also think about the material used in the loading and drive syringes. Plastic may
adsorb many analytes and there is also the possibility of plasticizers and other
species leaching out, especially in acidic media. Glass has similar problems and
is particularly prone to leaching sodium.
268
STOPPED-FLOW CIRCULAR DICHROISM
269
ALISON RODGER AND MICHAEL J. CAREY
the kinetic record. Mixer options or even premixers can be employed to counter
such difficulties but are likely to compromise the instrumental dead time.
2.11 Software
The software question relates to ease of use, degree of control over the various
instrument functions, whether the required extent of data analysis is possible,
ease of access to data, and ease of transfer of data in a format that can be readily
used on other computer platforms and software packages. For kinetics in par-
ticular, automation of repetitive procedures such as multi-shot averaging is
desirable. Flexibility in the control of data acquisition allows kinetic records to
be optimized according to the event being studied. For example, a full range of
linear and logarithmic time bases covers the requirements of simple kinetic
changes to more complicated multi-step reactions. In addition, by making full
270
STOPPED-FLOW CIRCULAR DICHROISM
Table 1 Approximate costs of instrumentation discussed in the text. Systems with an entry
in the 'attachment required' column must be fitted to another instrument
$ = less than 5k USD; $$ = 5k to 25k USD; $$$ = 25k to 50k USD; $$$$ = 50k to 100k USD; $$$$$ =
100k to 125k USD.
271
ALISON RODGER AND MICHAEL J. CAREY
272
STOPPED-FLOW CIRCULAR DICHROISM
averaging, CD spectra are seen to 'come out' of the background noise. For CD
stopped-flow, OLIS attaches its stopped-flow module to the RSM instrument.
Although this combination will perform adequately where CD changes are of
good intensity, one should be cautious if the need is to measure fast reactions
accompanied by small changes in CD.
In addition to the RX.2000 stopped-flow accessory referred to above, Applied
Photophysics provides a CD detection option, CD.2C, as an upgrade to its
SX.18MV research stopped-flow spectrometer. In this instrument, the standard
optical light guide coupling between the monochromator and the sample hand-
ling unit (used in the absorbance and fluorescence configurations) is replaced by
a purpose built optical module comprising lens coupling, calcite polarizer,
photo-elastic modulator and sample housing. A second monochromator is
added to the spectrometer in order to optimize performance in the 225 nm
region. The stopped-flow CD cuvette is at the end of a short umbilical which is
attached to the SX.18MV stopped-flow sample handling unit. The umbilical
mounted cell slides in or out of the sample holder thus allowing ready inter-
change between the stopped-flow cell and rectangular spectrometer cuvettes. A
mercury-xenon lamp can be substituted for the standard xenon lamp used in
the SX.18MV spectrometer, so as to improve CD kinetic performance at selected
wavelengths. The stopped-flow sample handling unit uses a pneumatic drive.
Ratio mixing thus requires selection of appropriately sized drive syringes. The
viewport on the CD stopped-flow cell has the dimensions 1 0 X 3 X 2 mm (i.e. 60
ul). The dead time for the instrument varies according to the drive ratio (about 4
ms for a 1:1 mix and 8 ms for a 1:10 mix). For a research group already using an
SX.18MV spectrometer, the CD.2C provides a convenient upgrade path for the
measurement of CD kinetics.
273
ALISON RODGER AND MICHAEL J. CAREY
optical interface. The principal difference between the above mentioned CD.2C
and the n*-180 instrument is that the n*-180 is purpose designed both for high-
performance steady-state CD scanning and quality CD kinetics as well. The
optical design allows CD measurements to be made down to 180 nm. It uses
the same stopped-flow cell as the SX.18MV instrument (10 X 2 X 1 mm; 20 ul
volume) which is mounted in a thermostatted, rotatable cell block so that either
2 mm or 10 mm optical paths are readily accessible (changing the optical path in
this way conveniently allows the same protein concentration to be used for
kinetic measurements in the near UV as for the far UV). Fluorescence may be
directly observed from the stopped-flow cell via a suitable cut-off filter and both
fluorescence and CD detectors can be mounted at the same time. Research
groups interested in protein studies will find switching between CD and fluor-
escence, by software alone, particularly convenient. The dead time on this
instrument is approximately 1 ms for a 1:1 mix and 3 ms for a 10:1 mix. The base
instrument also provides high quality absorbance detection and a range of inter-
esting upgrade paths, including fluorescence emission scanning and anisotropy.
The software suite provides complete instrument control as well as many novel
digital optimizations to improve data quality and acquisition efficiency. The
software is multi-tasking so that activities such as data manipulation and
analysis can be pursued whilst the next set of data is being collected.
274
STOPPED-FLOW CIRCULAR DICHROISM
4.3 Baseline
The acquisition of blank (baseline) runs for comparison with the kinetic data set
is to be recommended. Data obtained from suitable blank runs should be sub-
tracted from the main kinetic data in order to correctly define the CD change of
interest (i.e. to correct for any CD offset together with any pressure induced
artefact present immediately after mixing).
275
ALISON RODGER AND MICHAEL J. CAREY
0.2 0.3
Time/s
Figure 1 Kinetics of the hydrolysis of glucuronolactone (Sigma G-8875, 1 mg/ml in water) by
potassium hydroxide (1 M) in 1:1 mixing ratio. Data shown is from a single run measured on
a n*-180. The pathlength is 2 mm, bandwidth 4 nm, the light source is a Xe lamp and the
detection wavelength is 225 nm. A blank has been subtracted. Data can be fitted to a single
exponential: (89.7 ± 0.6)exp[(-8.41 ± 0.13)x] - (0.37 ± 0.34).
276
1.5 -1.5 -1
Time/s Time/log(s)
10
-10
3 -20 O
o -a
-a
-30 m
o
(d)
-40
0 0.5 1 1.5
Time/s Time/s o
Figure 2 Kinetics of the folding of lysozyme from hen egg white (of Boehringer 1243004, 2.2 mg/ml in 6 M Gdn-HCI and 10 mM
o
phosphate buffer pH 6.85) diluted 1:10 by phosphate buffer (10 mm phosphate buffer pH 6.85). Pathlength is 2 mm, bandwidth 4 nm, c
the light source is an HgXe lamp and the detection wavelength is 225 nm. Data shown are from a n*-180. (a) Single shot, (b) single
shot with log time base, (c) average of eight shots, (d) eight shots also showing blank run. Data from the eight-shot run can be fitted
o
to a double exponential: (16.0 ± 0.3)exp[(-42.2 ± 1.5}x] - (11.03 ± 0.12)exp[(-2.54 ± 0.05)x] - (23.39 ± 0.033). o
I
3D
O
C/5
ALISON RODGER AND MICHAEL J. CAREY
Figure 3 Kinetics of the folding of cytochrome c from horse heart (Sigma C-2506,
2.2 mg/ml in 6 M Gdn-HCI and 10 HIM phosphate buffer pH 6.85) diluted 1:10 by phosphate
buffer (10 mM phosphate buffer pH 6.85). Pathlength is 2 mm, bandwidth 4 nm, the light
source is an HgXe lamp and the detection wavelength is 225 nm. Data shown are from a
7r*-180 averaged over 16 repeat iterations. Data can be fitted to a double exponential:
(14.48 ± 0.19)exp[(-41.4 ± 0.9)x] + (7.08 ± 0.05)exp[(-1.233 ± 0.03)x] - (28.80 ±
0.05).
5 Examples
A literature search with key words such as <stopped-flow> or <kinetics> and
<CD> or <circular dichroism> and <millisecond> together with words for
particular types of systems, such a protein, will reveal many examples of
stopped-flow experiments. One area where stopped-flow CD has been a key
technique is that of protein folding and unfolding. In this section we shall
briefly review the role of stopped-flow CD in elucidating folding pathways for
278
STOPPED-FLOW CIRCULAR DICHROISM
Manometer
Tubing
adapter
Reduction
adapter Three-neck round-
bottomed flash
Expansion
adapter
Ultrasonic bath
Conical flask
Figure 5 A simple system for degassing solutions.
lysozyme. The chapter concludes with a final example where the kinetics are
due to the simultaneous interaction of three species one of which is a catalyst.
Varying the concentration of the catalyst in the experiments takes them from
the time domain where manual mixing is feasible to millisecond and faster
timescales.
279
ALISON RODGER AND MICHAEL J. CAREY
coupling with the folded protein in their neighbourhood. Therefore, the near-
UV CD is a more direct probe of tertiary structure.
14 uMCo(lll)
•o
28 uMCo(lll), k=8.1 S~1
I -10 ,
O 56 uMCo(lll), fc=5.6s-1
83 uMCo(lll), *=4.3s~ 1
100 200
Time/s
1
The racemization of [Fe(l,10-phenanthroline)3]2+ is temperature dependent and the enan-
tiomers can be resolved with antimonyl tartrate at 0°C. Antimonyl d-tartrate is commercially
available and most resolutions of metal complexes using it involve collecting the precipitate as
one enantiomer, and precipitating what remains in solution as the perchlorate to give the other
enantiomer. The enantiomeric purity of the more soluble isomer is usually much lower than
that of the less soluble salt. Potassium antimonyl i-tartrate is not commercially available but it
may be synthesized by heating an aqueous slurry of Sb2O3 + KOH + 1-tartaric acid (in molar ratio
1:1:2) under reflux for 2-3 h. Slow evaporation in a fumehood gives the required compound in
crystalline form. [Co(l,10-phenanthroline)3]3+ may also be resolved using antimonyl tartrate
isomers.
280
STOPPED-FLOW CIRCULAR DICHROISM
References
1. Mo, J, Holtzer, M. E. and Holtzer, A (1991) Biopolymers 31, 1417-1427.
2. Radford, S. and Rodger, A. (1988) Unpublished data.
3. Kulkarni, S. K., Ashcroft, A. E., Carey, M., Masselos, D., Robinson, C. V. and Radford, S.
E. (1999) Protein Science, 8, 35.
4. Weissman, J. S. and Kim, P. S. (1991) Science, 253,1386.
5. Evans, P. A. and Radford, E. S. (1994) Current Opinion in Structural Biology 4,100.
6. Lee C. S. (1966) Inorg. Chem. 5,1397.
7. Rehmann, J. P. and Barton, J. K. (1990) Biochemistry 29,1701.
8. Bates, P. J. and Rodger, A. (1992) Unpublished results.
281
This page intentionally left blank
Chapter 11
Spectrophotometry and
fluorimetry of cellular
compartments and intracellular
processes
C. LINDSAY BASHFORD
Department of Biochemistry and Immunology, Cellular and Molecular Sciences
Group, St George's Hospital Medical School, Cranmer Terrace, London SW17 ORE
1 Introduction
Optical spectroscopy, Spectrophotometry and fluorimetry can be used to monitor
processes occurring in living cells provided that suitable chromophores are
present which 'report' on the events in which they participate. The advantages
of optical techniques are manifold. Firstly they can be fast—with appropriate
apparatus events in the pico- and nano-second domains can be studied by fluor-
escence spectroscopy. Secondly they are continuous—instant feedback from the
experimental system can guide the most complex of experimental protocols,
and allow the experimenter to adjust system parameters as necessary. Thirdly
they are convenient, and most laboratories have access to equipment that can
provide quantitative analysis of optical signals; examples include conventional
spectrophotometers/fluorimeters, dedicated instruments (e.g. for fluorescence
lifetime and polarization measurements), cameras, microscopes and plate
readers. Significantly detectors from one apparatus can often be used on others
to open up new experimental protocols. Fortunately the principles underlying
the use of such a diverse array of optical devices are straightforward and
universal—they apply just as much to laboratory 'work-horse' instruments as
they do to the most specialized, laser-illuminated fluorescence microscope. The
availability of fast laboratory computers with large storage capacities means that
most modern spectrometers are microprocessor controlled and digitization of
signals opens up the full range of possibilities of data accumulation, storage,
analysis and interpretation.
The main problem with optical measurements is not the acquisition but
rather the interpretation of the data obtained. Straightforward analysis of the
results depends on the clarity of the experimental design and the appropriate
choice of chromophore. This chapter describes some of the problems that can
be addressed by spectroscopic techniques and attempts to give guidance on
good experimental design.
283
C. LINDSAY BASHFORD
2 Experimental design
2,1 Apparatus
Optical spectroscopy requires either spectrophotometers, to measure absorb-
ance, fluorimeters, to measure fluorescence, or microscopes, which can measure
fluorescence or absorbance of single cells or small groups of cells. Fluorimeters
and spectrophotometers usually require solutions or suspensions of material in
conventional cuvettes; microscopes provide two-dimensional images from
smears, slices or surfaces. Other devices that record signals resolved in two-
dimensions include gel scanners and microplate readers. Essentially these de-
vices sample the 'object' in an organized manner (detectors can be set up to
record absorbance or fluorescence) and information is stored in an electronic
array that maps precisely the physical layout of the original object.
Fluorimeters and spectrophotometers may have attachments that permit
recording of signals from surfaces, or can be adapted to record such signals
using fibre optic 'light guides'. In this case, the guide (a bundle of optical fibres)
conducts the appropriate illumination to the surface and both reflected and
fluorescent light are collected by other fibres and conducted to a detector,
usually a photomultiplier (1). The reflected light has two components: specularly
reflected light, this is the mirror-like reflection, and diffusely reflected light.
Only the latter contains information (analogous to absorbance) concerning the
chromophores present at the surface.
Light output from the image plane of an optical microscope can be assessed
with the light metering system of a camera attached to the microscope. The
time taken correctly to expose a film of given speed (ASA/DIN) is inversely
related to the brightness of the image. The exposure time, that is the interval
between the opening and the closing of the shutter, is measured with a stop-
watch. For accurate measurements the exposure time should lie in the 10-100 s
range; shorter exposures are difficult to time with appropriate precision and
longer exposures are compromised by stray light entering the objective (see
Section 2.1.2 for discussion of stray light) and by photobleaching. The correla-
tion between intensity and reciprocal exposure time is linear over a very wide
range (1), so this procedure provides a handy 'rule-of-thumb' comparison of the
brightness of images. The main disadvantage of this simple approach is that no
allowance is made for the differing brightness of objects in the field of view,
indeed most of the background may be black. Most conventional camera meters
weight light in the centre of the field of view more heavily than that in the
periphery.
A digital approach can resolve the difficulty of signal variation across the
image and CCD (or similar 'digital') cameras are now routinely found in the
camera port of optical microscopes. They provide digital images that can be
analysed pixel by pixel. A wide range of image analysing software, not all of it
specific to particular commercial packages, is available. In such processing, the
key decisions that have to be taken are the zero level, sometimes called the
'black level', and the gain employed. If zero and gain are not chosen correctly a
284
FLUORIMETRY OF CELLULAR COMPARTMENTS AND INTRACELLULAR PROCESSES
2.1.1 Artefacts
Optical investigations of biological systems are beset with hazards and it cannot
be too strongly emphasized that very rigorous criteria must be applied before
interpreting the experimental data. Artefacts are more common in fluorescence
experiments and a few of the most easily avoided ones are considered here.
285
C. LINDSAY BASHFORD
response at one wavelength and a small (or oppositely directed) one at a neigh-
bouring wavelength; this is most likely to occur in compounds which have
intense, narrow ('sharp') absorbance bands.
286
FLUORIMETRY OF CELLULAR COMPARTMENTS AND INTRACELLULAR PROCESSES
287
C. LINDSAY BASHFORD
288
FLUORIMETRY OF CELLULAR COMPARTMENTS AND INTRACELLULAR PROCESSES
membrane (10). This property is useful for studying the mechanisms of energy
transduction in photosynthesis, and stimulated an extensive search for exo-
genous pigments with similar properties which could be used for studies of
membrane potential in a wide variety of systems (11).
In addition to exogenous pigments which register membrane potential, dyes
have been employed to monitor many cellular properties; for example, cyto-
plasmic and organelle pH, cytoplasmic Ca2+ and Mg2+ concentration, membrane
microviscosity, fate of ingested particles. More recently probes have been
developed to report on processes such as apoptosis, critical to cell survival. In
these cases the molecule in question is specifically designed: (i) sensitively to
register the parameter of interest; and (ii) simply to reach its ultimate destina-
tion. Probes of cytoplasmic Ca2+ or pH exhibit characteristic fluorescence or
absorbance depending on the cations in their environment. Membrane-permeant
esters of the indicators diffuse to the compartment where endogenous enzymes
catalyse their hydrolysis to generate the active species in situ. More specific
delivery may be accomplished by exploiting natural endocytotic and/or bio-
synthetic processes. The range of possible probe experiments is so great that it
is impractical to describe each in detail. The publications of Molecular Probes
(their catalogue, the series of leaflets called BioProbes and their very user-
friendly web site http://www.probes.com) provide the most extensive intro-
duction to the types of probe experiments available and the literature from
which those approaches were drawn. It is not always appreciated how simple
many of the experiments are to perform using equipment available in most
laboratories.
289
C. LINDSAY BASHFORD
phore there are a number of factors that always have to be taken into account
(6,12).
290
FLUORIMETRY OF CELLULAR COMPARTMENTS AND INTRACELLULAR PROCESSES
291
C. LINDSAY BASHFORD
292
FLUORIMETRY OF CELLULAR COMPARTMENTS AND INTRACELLULAR PROCESSES
3 Examples
31 Oxidation-reduction state of tissue mitochondria
The properties of mitochonctrial suspensions have long been the centre of
intense study and the characteristics of the pigments of the respiratory chain
have been resolved by dual-wavelength spectroscopy (4; Section 2.4.1). However,
mitochondrial preparations may not reflect all the properties found m vivo.
Direct recording of the respiratory pigments in situ, by absorbance and/or
fluorescence spectroscopy. provides one way of resolving this dilemma. In
anaesthetized animals spectroscopy of exposed tissues (24, 25) reports the
oxidation-reduction state of The electron transport chain and the oxygenation
of haemoglobin/myoglobin. Unfortunately the rapid fluctuations of 02 delivery
and consumption make it difficult to establish convenient steady states. An
alternative procedure is to freeze the tissue rapidly and then to obtain spectra
more leisurely at low temperatures.
Protocol 1
Reflectance spectra of freeze-trapped rat kidney
Equipment and reagents
• liquid nitrogen Wavelength scanning, dual-wave length
• Aluminium tongs for freeze-damping spectrophotometer
('Wollenberger' tongs (26)) Anaesthetized rat
Method
1 Dissect the anaesthetized rat to reveal the kidneys.
2 Chill the Wollenberger tongs in liquid nitrogen. A polystyrene ice bucket is a con-
venient vessel in which to keep the liquid nitrogen; it is easy to break dewar vessels
with aluminium tongs.
3 Bring the cooled tongs close to the kidney.
4 Remove the kidney, place on the cooled tongs and squeeze the tongs together,
5 Keeping the tongs closed return them to the ice bucket containing liquid nitrogen.
6 Open the tongs, remove the crushed, frozen kidney with plastic forceps and store
it in suitably labelled aluminium foil in liquid nitrogen,
7 To record reflectance spectra unwrap the foil surrounding the kidney—in liquid
nitrogen using plastic forceps.
8 Chill the common terminal of the optical fibre connected to the spectrophoto-
meter(l),
9 Under liquid nitrogen press the common terminal of the optical fibre against the
smooth frozen face of the kidney.
293
C. LINDSAY BASHFORD
Protocol 1 cor
10 Record the reflectance spectrum—repeat steps 8-10 for several different regions of
the frozen tissue.
11 Use the white surface of the polystyrene ice bucket as a 'baseline' for computing
the absolute reflectance spectra of the kidney.
12 Use different areas of the frozen kidney surface for computing difference spectra.
0.50r
Figure 1 Reflectance spectra of freeze-trapped rat kidney. Kidneys from anaesthetized rats
were freeze-trapped (26) before, 1 min and 10 min after occluding the renal blood supply.
Reflectance spectra were recorded with a bifurcated light guide attached to a scanning, dual-
wavelength spectra photo meter (1, 24) using 700 nm as the reference wavelength. The
baseline was the spectrum of white polystyrene. Spectra labelled a and b were recorded
from the normally perfused and 1 min ischaemic kidneys, respectively. Spectrum c is the
difference between the 1 min and 10 min ischaemic kidneys. Reproduced with permission
from C. L. Bashford and M. Stubbs (1986) Biochemical Society Transactions, 14,
1213-1214. © The Biochemical Society,
294
FLUORIMETRY OF CELLULAR COMPARTMENTS AND INTRACELLULAR PROCESSES
found after 1 min and the difference between these two states (Figure 1, trace c) is
featureless and gives an indication of the baseline for this experiment. Inter-
estingly, oxyhaemoglobin is not absent from the ischaemic tissue suggesting
that there are metabolic compartments where oxygen is delivered but not
consumed (27, 28).
Similar experiments with freeze-trapped gerbil cerebral cortex (24, 29; and
see Figure 2) indicate that only in extreme anoxia does reduction of the mito-
chondrial pigments occur. Figure 2 shows the difference reflectance spectrum
obtained when the reflectance spectrum from a sample obtained from an
animal after breathing 5% oxygen for 1 min is subtracted from that taken from
an animal which had been breathing 0% oxygen for 1 min. In both situations
the haemoglobin was largely deoxygenated and the difference spectrum of the
whole tissue closely resembles that of reduced minus oxidized mitochondria.
This shows that the respiratory chain still has enough oxygen under conditions
where tissue haemoglobin is almost completely deoxygenated; or put another
way the large difference in affinity for oxygen exhibited by mitochondria and
haemoglobin in vitro is still apparent in vivo.
The experiments illustrated in Figures 1 and 2 employed a microprocessor-
controlled, scanning, dual-wavelength spectrophotometer which was capable of
storing and manipulating spectra.
0.02
0.01
AA
-0.01
295
C. LINDSAY BASHFORD
usually called the membrane potential, can play a significant part in membrane
activities; it is, for example, critical for signalling in excitable cells and makes a
contribution to the accumulation of nutrients driven by the electrochemical
Na+ gradient across the plasma membrane of many cells. In subcellular organ-
elles, such as chloroplasts and mitochondria, the membrane potential is an
important 'high energy' intermediate in the synthesis of ATP. In only a limited
number of circumstances can the potential be recorded directly and other
methods such as the use of dyes with specific sensitivity to membrane potential
are available (30).
Many dyes change either their absorbance or their fluorescence in response
to a change in membrane potential (11). However, it is often difficult accurately
to calibrate the signal with known potentials. In one procedure, the membrane
permeability to a particular cation, usually K+ or H+, is increased by adding a
reagent called an ionophore. In the presence of sufficient ionophore, such that
all other permeabilities are relatively small, the membrane potential (V) will
approach that predicted by the Nernst equation:
V=(-RT/f)ln([M + ] i /[M + ] 0 ) 1
0.02
i) Val [K+]mM ii) Val [K+]mM
1 40
1 U 1
/ 33
0.01 / 20 /**"
^630-590
UUP 9J
0
[
-^d *^J
-0.01 —H 5 minutes p~
Figure 3 Measurement of Lettre cell membrane potential using oxonol-V. (i) 5 x 106 Lettre
cells/ml were suspended in 5 mM HEPES, 1 mw MgCI2, 2.2 x 10-6 M oxonol-V, pH adjusted
to 7.4 with NaOH at 33°C and 155 mM NaCI plus KCI with the final K+ concentration
indicated. Valinomycin was added as indicated to give a final concentration of 1 x 10-6
g/ml. Traces from four separate experiments are superimposed such that the signal before
the addition of valinomycin is identical. The extracellular K+ concentration at which
valinomycin would give no change in absorbance was estimated to be 7.1 mM by
interpolation. The intracellular K+ concentration was 73.3 mM giving a membrane potential
(see equation 1) under these conditions of -62 mV. (ii) 5 x 106 Lettre cells/ml were
suspended in 150 mM NaCI, 5 mM KCI, 5 mM HEPES, 1 mM MgCI2, 2.2 x 10-6 M oxonol-V,
pH adjusted to 7.4 with NaOH at 33°C. Valinomycin (1 x 10-6 g/ml) and KCI to give the
concentration indicated were added as shown by the arrows. The extracellular K+
concentration at which valinomycin would give no change in absorbance was estimated to be
7.1 mM by interpolation, the intracellular K+ concentration was 63.3 mM giving a membrane
potential (see equation 1) under these conditions of -60 mV. Reproduced with permission
from Bashford, C. L, Alder, G. M., Gray, M. A., Micklem, K. J., Taylor, C. C., Turek, P. J. and
Pasternak, C. A. (1985) J. Cell. Physiology, 123, 326.
296
FLUORIMETRY OF CELLULAR COMPARTMENTS AND INTRACELLULAR PROCESSES
where M" is the relevant ion, R the gas constant, T the absolute temperature
and F Faraday's constant. At 37°C this relation simplifies to:
V(mV)= -61.5log(|M'y[M-| 0 ) 2
which shows that, for a 10-fold change in cation gradient, you obtain a 61,5 mV
change in potential. The cation gradient can be manipulated by altering the
medium concentration of M' and hence a calibration curve of optical signal and
potential can be obtained in the presence of ionophore.
Protocol 2
Measurement of plasma membrane potential using
oxonol-V and cells in suspension
Equipment and reagents
• Isotonic saline; e.g. 0.15 M NaCl, 0.005 M Stock solutions of oxonol-V, valinomycin
KC1, 0.005 M HEPES, 0.001 M MgCl2 pH and FCCP, 0.001 M in DMSO or ethanol
adjusted to 7.4 with NaOH. Do not include Stirrer
serum or other albumin-containing media
1-cm square cuvettes
because albumin/protein will sequester
the oxonol and none of the dye will Recording spectrophotometer with a
interact with the cells thermos tatted cell holder
« Pipettes
Method
1 Wash the cells in the saline and resuspend them at a cytocrit of about 0,5-1% v/v.
2 Transfer 1 ml of the cell suspension to a cuvette and place this in the spectrometer,
select the appropriate wavelength(s) and then zero the absorbance/fluorescence
reading.
297
C. LINDSAY BASHFORD
Protocol 2 continued
Stir in oxonol-V to give a final concentration of 1-5 x 10-6 M; record the signal until
it reaches a steady value—this takes up to 10 min.
Then stir in valinomycin or FCCP to give a final concentration of 1-5 x 10-6 M;
record the change in signal.
Then stir in KC1 (valinomycin-treated cells) or acid/base (FCCP) treated cells to
double/halve the relevant cation concentration; record the change in signal.
Use the changes in signal induced by valinomycin/FCCP to estimate the K + / H k
concentration at which addition of valinomycin/FCCP would no give no change in
signal (the 'null-point').
Determine the internal (cytosolic} K ' / H ' concentration and calculate the mem-
brane potential using equation 1.
The first protocol is to incubate the cells in media of differing K' content and
to assess the effect of valinomycin addition (Figure 3(i)). The value of the K null
point is obtained by interpolation. This procedure works best with cells that
have rather low K' permeability; cells more permeable to K + depolarize in high
K' media and the subsequent addition of valinomycin has little effect.
In the second protocol, after valinomycin has been added to the cells and a
new steady reading obtained, increase the concentration of K1 in the medium
by adding successively (Figure 3(ii)) small aliquots of concentrated KC1, about 3 M
is the highest manageable concentration. After taking readings at a few differ-
ent extracellular K' concentrations, obtain the value of the null point by inter-
polation (as in method 1), In both methods it is sufficient to interpolate by
joining data points rather than by using sophisticated least-squares protocols,
especially as the theoretical form of the dye absorbance versus potential plot is
unknown,
In either procedure, once the 'null-point' is known it is necessary to measure
cytoplasmic K' concentration to calculate the potential according to equation 7,
This can be achieved by pelleting the cells through a mixture of di-n-butyl-
phthalate (2 parts) and dinonylphthalale (1 part) with a density of 1.02; cells
pellet through such an oil within 10 s in a Beckman microcentrifuge B. Cell
cations can be determined (after the water content; wet - dry weight) by atomic
absorption spectrometry. The spillover of medium into the cell pellet is usually
less than 0.05%, but this can be verified by including an extracellular space
marker, such as [' 4 C]inulin, in the incubation medium.
There is a danger that the lipophilic valinomycin/K' complex may precipitate
lipophilic anions such as the oxonol dyes and it is advisable TO check on the
calibration described above with another ionophore. The uncoupling agent
FCCP is useful in this respect as it selectively increases the H* permeability of
membranes. The experimental protocol for measuring the potential using FCCP
298
FLUORIMETRY OF CELLULAR COMPARTMENTS AND INTRACELLULAR PROCESSES
299
C. LINDSAY BASHFORD
Protocol 3
Labelling of endosomes with fluorescein conjugates
Equipment and reagents
• Culture medium • Fluorimeter or fluorescence microscope
• Fluorescein conjugate • Media for mounting cells—slides and
cover slips
Method
1 Incubate the cells in culture medium containing the fltiorescein conjugate (present
at ~l mg/ml) at 37°C. The conjugate enters cells rapidly (within 10 min) and is
subsequently transferred to secondary lysosomes (20). Endocytosis and intracellular
membrane traffic are markedly reduced at lower than physiological temperatures.
At 4°C only binding of ligands to cell surface receptors occurs, so receptors can be
'loaded' at low temperature and endocytosis subsequently 'triggered' by raising the
temperature.
2 Remove the extracellular conjugate by washing the cells with cold (4°C), unlabelled
medium. The residual fluorescence arises from within the cells.
3 Place the labelled cells in a fmorimeter/fluorescence microscope and record the
fluorescence intensity. Cells attached to a solid substratum can be monitored with
light guides (1), or by fitting the substrate diagonally in a fluorescence cuvette and
using a conventional fluorimeter (the final optical arrangement resembles the
'front face' system used to record fluorescence from opaque or strongly absorbing
samples). It is important not to move the substrate during signal calibration.
4 Calibrate the signal with extracellular pH using monensin and the protocol
illustrated in Figure 4(a).
normally pH(m, < pK - 1 so that [AH*] = [AH'j + |Aj. 9-Aminoacridine (10 mM)
exhibits a yellow (1) and acridine orange a red rather than the blue fluorescence
seen at low dye concentrations. Cells incubated with 0.01 mM 9-aminoacridine
may have many yellow vesicles when examined by fluorescence microscopy
using the fluorescein settings; this indicates that such vesicles have accumu-
lated the dye by at least two orders of magnitude and that the internal pH must
be in the range 5-6.
500
FLUORIMETRY OF CELLULAR COMPARTMENTS AND INTRACELLULAR PROCESSES
io.2A EG.TA
PH7.3
Sa^o.i Jti
50 HCI HCI , —J {Ca2* — -
I
.PH.4 i
T
t '
k— ^ t pH*?! Triton [
HCI i
Triton i <1 nM
+ EGTA Ca2+
jPj^sl i^pH6
+ EDTA
0 •
Monensin PH 4
Figure 4 Calibration of intracellular probes of pH and Ca2+. (a) Cells labelled with
fluorescein-dextran (Protocol 3; 20} in Hanks balanced salts solution buffered with 20 ITIM
HEPES, pH 7.0 (NaOH) were excited at 493 nm and fluorescence monitored at 520 nm. The
initial signal (6 units) was calibrated by adding as indicated: 1 ug/ml monensin (final
concentration) to equalize the pH of all compartments in the system (20) followed by HCI to
bring the solution pH to the values indicated. The original signal from the cells corresponds
to that found at pH 4.8 in the presence of monensin indicating that the fluorescein-dextran
occupied a compartment whose pH was 4.8. Records should be corrected for cellular
autofluorescence (fluorescence of unlabelled cells) and, after calibration, the supernatant
fluorescence should be measured and found to be <10% of that found in the presence of
cells, (b) Thymocytes labelled with quene-1 (19) were suspended at 5 x 106 cells/ml in a
medium containing the inorganic salts of RPMI 1640 (without phenol red) and supplemented
with 11 mM glucose and 10 mw HEPES, pH 7.3 (NaOH) at 37°C. Fluorescence at 530 nm
was excited at 390 nm; the signal (49 units) was calibrated by permeabilizing the cells with
0.05% Triton X-100 (Triton) in the presence of 0.5 mM EGTA and 0.5 mM EDTA. More than
90% of the dye is released by the Triton and it is titrated by the addition of Tris and HCI to
give the pH values indicated. The signal before permeabilization corresponds to that found
for pH 7.1 in the presence of Triton indicating that cytoplasmic pH was 7.1 when the
medium pH was 7.3. (c) Lymphocytes labelled with quin-2 (18) were suspended at 5 x 106
cells/ml in 130 mM KCI, 20 mM NaCI, 1 mM CaCI2, 1 mM MgCI2, 5 mM glucose, 10 mM
MOPS pH 7.4 (NaOH). Fluorescence at 500 nm was excited at 339 nm. The signal was
calibrated by permeabilizing the cells with 0.05% Triton X-100 (Triton) and subsequent
titration of the medium with EGTA, MgCI2 and KOH to give 0.2 and 0.1 x lO"6 M Ca2+ as
indicated, with 1 mM free Mg2+, pH 7.05 (18). 'Zero' Ca2+ (<1 nM) was achieved by adding
2 mM EGTA and Tris to give a pH > 8.3. Autofluorescence from unlabelled cells has been
subtracted from each reading and the fluorescence of unpermeabilized lymphocytes
corresponds to a Ca2+ level of 150 nm. All traces in this figure are redrawn from data
originally published in (18-20).
301
C. LINDSAY BASHFORD
Protocol 4
Measurement of cytoplasmic Ca2 concentration or pH
using acetoxymethylester-linked indicators
Equipment and reagents
• Isotonic saline; e.g. 0.15 M NaCl, 0.005 M • Stock solutions of acetoxymethylester-
KC1, 0,005 M HBPES, 0.001 M MgCl, pH linked indicator, 0.01 M in DMSO, and
adjusted to 7,4 with NaOH. It helps to Triton X-100, Ca2i. EDTA, EGTA, acid and
include up to 0.025 M NaHCO3 in the base, in water, for calibration titrations
medium (replacing some of the NaCl) so • 1-cm square cuvettes
that intracellular pH remains slightly
• Recording spectra photo meter with a
alkaline chiring hydrolysis of the
thermostatted cell holder
acetoxymethyl ester
• Pipettes • Water bath thermostatted to 37°C
• Stirrer
Method
1 Wash the cells in the saline and resuspend them at acytocrit of about 0.5-1% v/v.
2 Transfer 1 ml of the cell suspension to a cuvette and place this in the spectrometer,
select the appropriate wavelength(s) and then zero the absorbance/fluorescence
reading.
3 Stir in acetoxymethylester-linked indicator to give a final concentration of 1-5 x
10'° M.
4 Follow the hydrolysis of the ester by monitoring the change in absorbance/
fluorescence.
5 When ester hydrolysis is more than 75% complete, this may take 30-90 min, pellet
the cells, resuspend them in unlabelled isotonic saline and incubate them for a
further 30 min at 37 °C.
6 Then pellet the cells once more and resuspend them in isotonic saline for
absorbance/fluorescence measurements.
7 Calibrate the optical signals with extracellular pH or Ca2" after permeabilizing the
cells with Triton X-100 (see Figure 4),
A word of caution about the cleaning of cuvettes is appropriate here; the use
of chromic acid may introduce fluorescence quenchers. These can usually be
removed by soaking the cuvettes in solutions of dictating agents such as EDTA.
Unfortunately some cells seem not to retain the acid form of these pH and
Ca J ~ indicating dyes very long even though the free acid is supposed to be
membrane impermeant. It is very important, therefore, to check that the signal
observed does indeed originate from within cells; this is best done by pelleting a
sample of the suspension and verifying that the supernatant lacks dye (check
the supernatant fluorescence). If leakage is particularly troublesome, reasonable
302
FLUORIMETRY OF CELLULAR COMPARTMENTS AND INTRACELLULAR PROCESSES
data can be obtained by loading the cells for only a short time (much less than
the time taken for complete hydrolysis of the ester), rapidly washing the cells
and immediately making the pH or Ca2+ determination. Unfortunately this
leads to an inevitable wastage of rather expensive starting material.
It is also possible to record pH or [Ca2+] directly using the appropriate
indicators and quantitative confocal microscopy in a similar manner to that
used to measure plasma and mitochondrial membrane potential directly (see
Section 3.2.1). The great advantage of the 'direct' technique is that it removes
the requirement for the destructive titration/calibration procedures used by the
other methods. This opens up the possibility of following dynamic changes in
pH and cell calcium directly.
303
C. LINDSAY BASHFORD
fused with a viral glycoprotein so that the synthesis and processing of the
glycoprotein can be monitored in real time and space (40). These results show
how newly synthesized material moves between the endoplasmic reticulum
and the Golgi apparatus. More recently the fluorophore of GFP has been
mutated to make it sensitive, in one study, to the putative pH values of cellular
compartments (41) and, in another, to cytoplasmic Ca2+ levels (42). In each case,
the advantages of the system are that the photosensitive material is guided
directly to the arena of interest by the normal biosynthetic activities of the cell.
Astonishingly it appears that adding the GFP to endogenous material does not
significantly alter its pattern of processing.
4 Future prospects
The elegance and simplicity of optical experiments ensures that considerable
efforts will continue to be expended to develop photometric and fluorometric
assays of cellular membrane function. The power to direct sensitive chromo-
phores precisely to interesting destinations and the ability to assess them
quantitatively at the single cell level promises to open up the dynamics of
membrane biology to the rigorous analyses of chemistry and physics. The new
indicators of pH and cell calcium will help to unravel the role of these cations in
cell stimulation and differentiation, particularly at the single cell level. Finally,
existing procedures will be used, on account of the wide applicability of the
techniques, on systems inaccessible to more elaborate experiments permitting
a proper investigation of the full range of biological diversity.
Acknowledgements
I would like to thank Mrs B. J. Bashford for her patient preparation of the
diagrams and the Royal Society, Biotechnology and Biological Sciences Research
Council and the Cell Surface Research Fund for financial support. This chapter
is an extensive revision of Chapter 7 in Biological membranes: A practical approach
(ed. J. B. C. Findlay and W. H. Evans), published by IRL Press in 1987; any material
reproduced is by permission of Oxford University Press.
References
1. Harris, D. A. and Bashford, C. L (eds) (1987) Spectrophotometry and spectrofluorimetry. A
practical approach. IRL Press, Oxford and Washington, DC.
2. Miller, J. N. (1981) Standards influorescencespectrometry. Chapman and Hall, London.
3. Bangham, A. D., De Gier, J. and Greville, G. D. (1967) Chem. Phys. Lipids, 1, 225.
4. Chance, B. and Williams, G. R. (1955) J. BioJ. Chem., 217, 395.
5. Chance, B., Schoener, B., Oshino, R., Itshak, F. and Nakase, Y. (1979) J. BioL Chem., 254,
4764.
6. Harbig, K., Chance, B., Kovach, A. G. B. and Reivich, M. (1976) J. Appl. Physiol, 41. 480.
7. Johnson, L. V., Walsh, M. L., Bockus, B. J. and Chen, L. B. (1981) J. Cell Biol, 88, 526.
8. Waggoner, A. S. (1979) Methods Enzymol, 55, 689.
9. Keilin, D. (1925) Proc. R. Soc. Lond. B, 98, 312.
304
FLUORIMETRY OF CELLULAR COMPARTMENTS AND INTRACELLULAR PROCESSES
305
This page intentionally left blank
Chapter 12
Use of optical spectroscopic
methods to study the
thermodynamic stability of
proteins
Maurice R. Eftink and Haripada Maity
Department of Chemistry, 112 Coulter Hall, The University of Mississippi,
MS 38677, USA
1 Introduction
The biophysical characterization of globular proteins will almost always include
some type of study of the unfolding of protein to obtain thermodynamic para-
meters. The basic idea is that a transition between a native and unfolded state,
induced by temperature, pH, or denaturant concentration, can serve as a stand-
ard reaction for obtaining a thermodynamic measure of the stability of the
native state. For example, the free energy change for the unfolding reaction can
be used to compare the stability of a set of mutant forms of a protein (1-4).
This type of analysis is based both on assumptions of the thermodynamic
model for the unfolding process and on assumptions in the way the data are
analysed; some of these assumptions and their limitations will be discussed
below.
There are a variety of methods that can be used to monitor an unfolding
process. A common method is differential scanning calorimetry, DSC, which
measures the variation in the specific heat of a protein-containing solution as a
protein is thermally unfolded (5-7). DSC is a popular method for this purpose, but
optical methods can also provide suitable information for tracking the unfolding
of a protein The spectroscopic signals for the native and unfolded states of a
protein can give some insight regarding the structure of the states, and often can
provide advantages of economy, ease of measurement and amenability to a wide
range of sample concentration. The optical spectroscopic methods that have
been used most often for this purpose are absorption spectroscopy, circular
dichroism and fluorescence, which will be discussed in this chapter. A key to each
of these methods and their use in protein unfolding studies is that the signal is a
mole fraction weighted average of the signals of each thermodynamic state. That
is, the observed signal, S, can be expressed as
307
MAURICE R. EFTINK AND HARIPADA MAITY
where Xj is the mole fraction of species i and si is the intrinsic signal of species i.
In order for a particular spectroscopic signal to be useful for tracking a N <—» U
transition of a protein, the signal must be sufficiently different for the N and U
states.
In this chapter we will first discuss thermodynamic models, and their
assumptions, for protein unfolding reactions. Then we will discuss advantages
and limitations of these three individual optical methods. Since most of the
fundamentals of optical spectroscopy are covered in other chapters in this
volume, emphasis will be placed on thermodynamic models and practical
matters related to the application of the spectroscopic methods.
Kun = 23 3
d
[N]
308
USE OF OPTICAL SPECTROSCOPIC METHODS
5
6
The value of Kun (or AG^) will depend on the extent of progress along the
perturbation axis (temperature, T, pressure, P, pH, etc.). The following sections
give various functions that are generally accepted as describing the dependence
of AG°un on these perturbation axes. One of the equations below, when com-
bined with those above and equation I can describe spectral data as a function of
temperature or chemical denaturant, in terms of the two-state model.
In this case, AHG,un (the value of AH°un at T = TG), ACp and TG would be fitting
parameters, whereas in equation 7b the fitting parameters would be AH°oun,
AS°0,un and ACp.
The heat capacity change for unfolding of proteins has been found to be
positive and to be related to the increase in solvent exposure of apolar side
chains (1-3). In other words, a positive ACp is a result of the hydrophobic effect
and a consequence is that the AG^fTJ for unfolding of a protein will have a
non-linear dependence on temperature, reaching a maximum at some
temperature and showing both high temperature and low temperature induced
unfolding.
309
MAURICE R. EFTINK AND HARIPADA MAITY
model, LEM, is the most widely used relationship for describing denaturant
induced unfolding (10,11). Admittedly, this is an empirical relationship, but it
appears to adequately describe the pattern for denaturant induced unfolding of
a number of proteins. There are other models (and equations) that have been
suggested to describe denaturant induced unfolding, including a binding model
(12,13) and a solvent partitioning model (12), and a combination of the two (14),
but the advantage and popularity of the LEM model is that it has only two
fitting parameters, AGo,un and m, for describing data. The AG° un is the para-
meter of prime interest, since it is a direct measure of the stability of a protein
in the absence of denaturant, and this parameter applies to the ambient solvent
conditions, which can be moderate temperature and pH (e.g. 20 °C and pH 7). By
contrast, it is more difficult to determine AGo,un at 20°C from thermal unfolding
studies, since it requires precise determination of AH° un , ASo,un and ACp and
use of these parameters to extrapolate back from the unfolding temperature.
(which is usually above 50 °C) to 20 °C.
Whereas the AGoiUI1 is a measure of thermodynamic stability of a protein, the
m value also provides structural insights, m values have been suggested to cor-
relate with the change in solvent-accessible apolar surface area upon unfolding
of the protein (12,15). That is, a relatively large m value (which corresponds to a
high susceptibility of the unfolding reaction to denaturant and a steep plot of Xu
or XN versus [d]) has been taken to indicate that there is a large change in the
exposure of apolar side chains on unfolding. This might apply to a protein that
has an extensive core of apolar side chains that are exposed upon denaturation.
Conversely, a small m value (and a less steep dependence of Xu on [d]) would
indicate that there is a lesser change in the exposure of apolar side chains,
which might apply to a protein that has few buried apolar side chains in the N
state (or a case in which the apolar side chains are already on the surface in the
N state, so that there is little change in their exposure on unfolding).
310
USE OF OPTICAL SPECTROSCOPIC METHODS
2.6 Simulations
Shown in Figure 1 are simulated plots of XN versus perturbant axis for various
types of perturbations. Shown in Figure 2 are corresponding simulated plots of
AG^ versus perturbant axis, all for the two-state unfolding reaction and for
thermodynamic parameters given in the legend to Figure 1. The symbols
represent simulated data points for which XN is 0.05 to 0.95, the range over
which this value can be determined with reasonable accuracy. The plots illus-
trate the length of the extrapolation needed to determine AG°0,un from typical
unfolding data.
10 20 30 40 60 6
Temperature (°C) [urea] (M) pH
Figure 1 Simulated plots of the mole fraction of the native state, XN, versus perturbation
axis for a two-state transition described by the following thermodynamic parameters. The
circles indicate the anticipated range of data that is usable for calculating the unfolding
equilibrium constant (or AGSn). The curves were simulated for a two-state model using the
following parameters and equations 3-9 and 11. AHS,un = -20000 cal/mol at T0 = 0°C,
AS° un = -90 cal/mol-K, ACP = 200 cal/mol-K, m = 4.0 kcal/mol-M, pKa,u = 4.5,pKa,N =
2.5, and n = 4.
311
MAURICE R. EFTINK AND HARIPADA MAITY
0 10 20 30 40 50 60 70 1 2 3
Temperature (°C) [urea] (M)
Figure 2 Simulated plots of AGJn versus perturbation axis for the simulations in Figure 1.
The circles are the anticipated range of data over which reliable values of AGun can be
determined.
13
XTI=-
4[P]0
where [P]0 is the total protein concentration (expressed as monomeric form), St
is the relative CD signal of species i and where Kun will depend on the per-
312
USE OF OPTICAL SPECTROSCOPiC METHODS
-10 -
J -100
2 3
[Urea], M
Figure 3 Urea induced unfolding of Cm represser as a function of total protein
concentration, [P]0, and as measured by changes in CD signal (at 222 nm for the less
concentrated protein sample and at 232 nm for the more concentrated sample). The solid
curves are a global fit of the D ^ 2U model with the parameters given in the text.
turbant as given by one of the above equations. That is, the transition should
depend not only on the value of Kun (or AG° il]n ) and m (in the case of urea or Gdn-
HCl induced unfolding), but also on the total subunit concentration, [PI,,.
The data in Figure 3 are for total monomer concentrations of [P]0 - 3.5 H.M and
27 u.M. As protein concentration is lowered, the dimer becomes destabilized. The
solid lines through the two data sets are global fits of equation 13 to the multiple
data sets with a single value of AG°Mm - 11.2 kcal/mol and m= 1.6 kcal/mol-M.
The following paragraphs give a detailed experimental description of the
unfolding study with Cro.
Protocol 1
Observation of the unfolding of a protein
Equipment and supplies
• A fresh, filtered solution of 20 mM A stock solution of the protein dialysed
NaH2P04, pH 7,0 against (or, if lyophilized, dissolved in} the
• Since urea slowly decomposes to form same buffer. The concentration will
cyanate. stock solutions should be depend upon the optical techniques being
prepared fresh every day or two used. For CD studies, a 50-200 uM protein
• Access to a CD instrument, such as an Aviv stock solution should allow a minimal
62DS spectropolarimeter, equipped with a volume to be added to the denaturant
thermoelectric (or thermojacketed. with a solution to give a working concentration
circulating water bath) cell holder to in the 5-20 uM range
maintain constant temperature Pipettes
• Quartz cuvettes
MAURICE R. EFTINK AND HARIPADA MAITY
Protocol 1 continued
Method
1 Set up a series of solutions (5 ml) of buffer containing urea concentrations between
0-8 M urea.
2 Take 2 ml of each of these solutions and add to each an aliquot of 100-200 u1 of the
stock solution of protein, thus forming a series of solutions having the same pro-
tein concentration and varying denaturant concentration,
3 Incubate the solutions overnight.
4 Calculate the concentration of urea in each solution by taking a small sample of
the different urea solutions, measuring the difference between the refractive index
of the urea solution and the buffer, AN. and using the following equation (10):
[urea] = 117.66AN + 29.753AN2 + 185.56AN3
(It is usually best to measure the refractive index of the solutions to which protein
has not been added and then make a small correction of dilution. However, in cases
where the protein concentration is low, it is acceptable to make the refractive
index measurement with that solution.}
5 Measure the CD signal at a chosen wavelength for each sample in the series. The far
UV region of 220-230 nm is sensitive to secondary structure changes. Lower
wavelengths can be used, but one must make certain that light absorbance by the
denaturant does not become a problem.
6 Plot the data as described in this chapter.
See Pace and coworkers for a thorough description of this general protocol (10).
Protocol 2
Observation of the unfolding of a protein, e.g. lambda Cro
Equipment and supplies
• Afresh, filtered solution of 20 mM A solution of 8 M urea in 20 mM NaH2PO4,
NaH2PO4, pH 7.0 containing the desired pH 7.0 also containing the same final
final concentration of the protein (e.g. protein concentration (from addition of
prepared by addition of an aliquot from a an aliquot of concentrated stock). This is
concentrated protein stock solution). the concentrated denaturant solution.
Prepare approximately 2,0 ml of this zero- Depending on the midpoint urea
denaturant solution concentration tor 50% unfolding, one will
• Access to a CD instrument, such as an Aviv need from 2 ml (for a low midpoint) to 6
62DS spectropolarimeter, equipped with a ml of this solution (for a high midpoint)
thermoelectric {or thermojacketed, with a Quartz cuvettes
circulating water bath) cell holder to Pipettes
maintain constant temperature and a
magnetic stirrer and flea for stirring
314
USE OF OPTICAL SPECTROSCOPiC METHODS
Protocol 2 continued
Method
1 Incubate both the zero denaturant and the concentrated denaturant solutions
containing protein at the desired temperature to achieve thermal equilibration.
2 Place 1.8 ml of the zero denaturant solution into the cuvette and then into the
spectropolarimeter, the cell block of which is controlled to the desired tempera-
ture.
3 Measure the CD signal of this first solution as in Protocol 1.
4 Remove an aliquot (e.g. 50-200 ul) of the solution from the cuvette and then add
the same volume of the concentrated denaturant solution to the cuvette.
5 Mix and allow 1-5 min for chemical equilibration. (If after this incubation period
the CD signal is still changing with time, then a longer incubation is needed.)
6 Measure the CD signal of the solution.
7 Repeat steps 4-6 until the concentration of urea in the cuvette has reached a point
where the signal no longer changes.
8 Plot the data as described in this chapter.
If a computer interfaced syringe pump is available, this titration procedure can be
automated, as described below and in (19). This reference also discusses an acquisition
routine for determining when the signal has equilibrated.
See Figure 3 for an example of data.
15 s and digital storage. An aliquot of the 'with denaturant' sample was then
delivered (following the removal of an equal volume of sample from the cuvette
by the second channel of the syringe pump, in order to maintain a constant total
volume in the cuvette; we find that this strategy of ensuring a constant solution
height in the cuvette works best for good mixing of the solutions by the stirring
flea, and the removal of a volume is also needed to avoid overfilling the cuvette
in cases where large total aliquot volumes are required). Following a waiting
period for mixing and chemical equilibration (e.g. 1-5 min), the signal is again
recorded. (See (19) for more details and for a description of a program for
controlling the data acquisiting and for determining when the signal has
equilibrated.)
The raw signal versus denaturant concentration data was then analyzed by
fitting a combination of equations 1, 9, and 13 to the data using a nonlinear
least-squares program (19). (In this particular case, the total protein con-
centration is also needed for equations 13, since this is a dimeric protein; for the
unfolding of a monomeric protein, the fitting equation for a simpler N ^ U
model would be a combination of equations 1, 5, 6, and 9.) In this fitting
procedure, the baseline slopes for the native signal and unfolded signal regions
were assumed to be linear, with these slopes being allowed to vary as fitting
parameters. The two other fitting parameters were the AGo,un (or Kun) and m in
equation 9.
316
USE OF OPTICAL SPECTROSCOPIC METHODS
3.3 Irreversibility
What if the unfolding process is irreversible? A common problem and concern
in protein unfolding studies involves irreversibility of the process. This may be
observed as an aggregation under denaturing conditions and/or an inability to
recover biological activity or characteristic spectroscopic or other physical
properties of the native state when the perturbing conditions are removed. For
example, a common way to observe such irreversibility is to notice a hysteresis
in plots of a signal versus temperature during upward, followed by downward
scans.
Such irreversibility may be modelled as
kd
N ^ U -» D
where D is a permanently denatured form of the protein and kd is the apparent
rate constant for the irreversible process. If, for example, kd has a steep depend-
ence on temperature, acid or base, or if kD depends on protein concentration,
then the irreversibility of the process will increase as the perturbing condition is
made more extreme. That is, it still may be possible to characterize the N ^ U
process by avoiding extreme perturbing conditions. If the U —» D reaction
cannot be avoided, the result will be a skewing of the shape of the transition,
making the recovery of valid thermodynamic data unreliable (18).
Another model for an irreversible process is one in which there is an equili-
brium intermediate, N ^ I ^ U, and the I species reacts in an irreversible
manner. There has been some discussion lately of this mechanism with the idea
that an intermediate may have a greater exposure of apolar side chains and that
this might lead to an enhanced tendency of the intermediate species to self-
aggregate (20).
317
MAURICE R. EFTINK AND HARIPADA MAITY
318
USE OF OPTICAL SPECTROSCOPIC METHODS
319
MAURICE R. EFTINK AND HARIPADA MAITY
-20 a>
Nuclease A I
-40
-60 o
LL
320
USE OF OPTICAL SPECTROSCOPIC METHODS
321
MAURICE R. EFTINK AND HARIPADA MAITY
322
USE OF OPTICAL SPECTROSCOPIC METHODS
the CD signals are much smaller, thus having lower signal-to-noise as compared
to far-UV CD or fluorescence data.
The advantage of CD measurements over the other optical methods is that
the far-UV signals observe changes over the entire protein (i.e. its secondary
structure). Also, the nature of the signal change can be related to structural
changes (e.g. loss of ellipticity at 222 nm can be attributed to loss of a-helix). A
disadvantage is that CD instruments are moderately expensive and less avail-
able than standard absorbance and fluorescence instruments. However, CD
instruments are marketed primarily for the biochemistry-protein-peptide re-
search community and an excellent selection of accessories for protein
unfolding studies is available.
4.3 Fluorescence
When using steady-state fluorescence to study protein unfolding, the most
commonly used fluorophores are the intrinsic tryptophan and tyrosine residues.
Since tryptophan has a larger molar extinction coefficient, a redder absorbance,
and is found in most proteins, this fluorophore is the one most frequently em-
ployed in fluorescence studies. The sensitivity of fluorescence detection is such
that concentrations as low as 10-8 M can be studied with commercial fluori-
meters, and, by selecting smaller cell pathlengths or longer excitation wave-
lengths, concentrations as high as 10-4 M can be studied. A very important
property of tryptophan fluorescence is that it is very dependent on the
environment of the indole side chain, making tryptophan fluorescence
responsive to the structure of a protein.
The emission maximum of tryptophan in proteins can range from 308 nm
for a residue buried in a completely apolar environment (as in the single trypto-
phan variant of azurin (37)) to 350 nm for a fully solvent-exposed residue
(38-40). Due to their apolar character, tryptophan residues are usually either
fully or partially buried in the three-dimensional structures of proteins and are
expected to have emission maxima from 320 to 340 nm. Unfolding of a protein,
and concomitant increase in the solvent exposure of tryptophan residues will
invariably lead to a red-shift in the fluorescence of a tryptophan-containing
protein.
The fluorescence intensity, or quantum yield <I>F, of a tryptophan residue in a
protein also varies over a wide range. The $F of single-tryptophan-containing
proteins can range from 0.01 to 0.4 in globular proteins (40). The wide range
appears to be due to a combination of factors, including the closeness of the
indole side chain to intramolecular quenching groups, such as peptide bonds
and certain amino acid side chains (41-43). There does not appear to be such a
wide dispersion of <£F values for unfolded proteins. The consequence of this
variation in $F, along with the above mentioned red-shift, is that the fluor-
escence intensity of tryptophan-containing proteins will almost always change
at either the red or blue edge of the emission spectrum when a protein unfolds.
The fluorescence intensity (either at a single wavelength, or integrated over
323
MAURICE R. EFTINK AND HARIPADA MAITY
the entire emission envelope) is a signal that follows equation 1 and can be used
to extract thermodynamic information. The apparent emission maximum of a
protein sample does not follow equation 1, so one must exercise caution in its
use. As we have shown by simulations elsewhere (25), if there is a significant
increase or decrease in 4>F of a protein upon unfolding (as well as a red shift),
the measurement of the apparent emission maximum will not give a true
reflection of the change in population of native and unfolded states, so that any
recovered thermodynamic parameters will be under- or over-estimates of the
actual values. While this problem is probably recognized by most people using
fluorescence, there is still a strong desire to be able to use emission maxima in
studying protein unfolding, since the maxima are not dependent on protein
concentration (except possibly for associating systems) or lamp intensity and
thus would make it more convenient in the measurement of a series of solu-
tions. A way to use emission maxima in thermodynamic studies is to do curve
fitting of the composite spectra. That is, it should be possible to fit the fluor-
escence spectra at intermediate points along a transition (e.g. when XN = 0.1 to
0.9) as a linear combination of the basis spectra of the native and unfolded
species times their relative population.
Another steady-state fluorescence signal that can be measured fairly routinely
is the fluorescence anisotropy, r. The anisotropy value for the fluorescence of a
tryptophan residue will depend on its rotational freedom (as expressed by its
rotational correlation time, 4>) and its fluorescence decay time, T, with im-
mobilized and short-lived excited states having largest r values. Since there is a
range of fluorescence lifetime and rotational correlation times for tryptophan
residues in native proteins, and since these 4> and T values will usually have a
narrower range for the unfolded state, it follows that r values will usually
change upon unfolding a protein. However, as we have discussed elsewhere
(25), r values do not follow equation 1 , but instead follow
15
That is, the observed r value depends not only on the rt value and mole fraction,
Xj, for the N and U states, but r also depends on the quantum yield of each state,
<J>i. If one of the states has a larger quantum yield than the other, the result will
be a skewing of the plot of r versus perturbation axis toward the more domi-
nant fluorescing state. In principle, if the <5F (or relative intensity) of the native
and unfolded states are known, one can still use r values to obtain thermo-
dynamic information. We have shown elsewhere that plots of fluorescence
anisotropy versus denaturant concentration for the unfolding of a protein can
give the same thermodynamic parameters as fluorescence versus intensity data,
if the above equation is used (17). Also, it must also be realized that anisotropy
measurements have more noise than simple intensity measurements, so that it
is preferable to just make intensity measurements, both because of the better
quality of the latter data and the more straightforward analysis.
If a protein contains tyrosine but not tryptophan residues, the fluorescence of
324
USE OF OPTICAL SPECTROSCOPIC METHODS
tyrosine can also be used for unfolding studies. The emission maximum of
tyrosine is very insensitive to environment (and protein conformational state)
and its <t>F is also relatively insensitive to environment. These factors, coupled
with the lower extinction coefficient of tyrosines, make it less valuable as a
reporter group. The advantage of tyrosine fluorescence is that there are usually
several such residues in a protein, which partially overcomes the lower extinc-
tion coefficient. Besides tyrosine and tryptophan, extrinsic fluorescence probes
can also be used. For example, dansyl groups can be covalently attached to
cysteine or lysine residues. (See (44) for a detailed description of several extrinsic
probes.)
In cases where there is not a significant change in fluorescence intensity
upon unfolding of a protein, it may be possible to enhance the signal change
by adding a solute quencher or by changing the solvent to D2O. A solute
quencher, such as acrylamide or iodide, will usually be able to quench the
fluorescence of exposed fluorophores in the unfolded state of a protein to a
much greater extent that for the same fluorophores in a native protein.
Alternatively, heavy water is known to increase the fluorescence intensity of
some fluorophores and may have different degrees of enhancement for buried
and exposed fluorophores.
Fluorescence instrumentation is relatively widely available and can be
purchased with various configurations and accessories over a wide range of
prices. As with CD instruments, some fluorescence instruments are designed
to allow automated thermal scans and/or titrations, usually with digital
acquisition.
To summarize, the advantages of fluorescence methods for studying protein
unfolding reactions are the wide concentration range that can be measured and
the sensitivity of the signal to the microenvironment of the fluorophore. Also,
fluorescence signals of the native and unfolded state can provide some low
resolution structural information about these states (at least with respect to the
microenvironment of the fluorophores).
5 Concluding remarks
Optical spectroscopic techniques provide several advantages in studies of the
kinetics and thermodynamics of protein unfolding. This chapter has attempted
to summarize many of these advantages, as they apply to equilibrium studies,
and to also highlight the assumptions involved and to point out some practical
considerations.
Acknowledgements
We acknowledge the contributions of recent graduate students and post-
doctoral associates to our work in this area. In particular, the contributions of
Glen Ramsay and Roxana lonescu are noted. Some of the unpublished work
presented in this chapter was supported by NSF grant MCB 9808635.
325
MAURICE R. EFTINK AND HARIPADA MAITY
References
1. Schellman, J. A. (1987( Annu. Rev. Biophys. Biophys. Chem. 16,115-137; Becktel, W. J.
and Schellman, J. A. (1987) Biopolymers 26,1859-1877.
2. Privalov, P. (1989) Annu. Rev. Biophys. Biophys. Chem. 18, 47-69.
3. Robertson, A. D. and Murphy, K. P. (1997) Chemical Physics 97,1251-1267.
4. Shortle, D. A., Meeker, A. K. and Freire, E. (1988) Biochemistry 27, 4761-4768.
5. Lumry, R., Biltonen, R. and Brandts, J. F. (1966) Biopolymers 4, 917-944.
6. Sturtevant, J. M. (1987) Annu. Rev. Phys. Chem. 38, 463-488.
7. Grikko, Y. V., Privalov, P. L, Sturtevant, J. M. and Venyaminov, S. Y. (1988) Proc. Natl
Acad. Sci. USA 85, 3343-3347.
8. Dill, K. A. and Shortle, D. (1991) Annu. Rev. Biochem. 60, 795-825.
9. Eftink, M. R. and lonescu, R. (1997) Biophys. Chem. 64,175-197.
10. Pace, C. N. (1986) Methods Enzymol. 131, 266-280; Pace, C. N., Shirley, B. A. and
Thomson, J. A. (1989) In Protein structure and function: A practical approach (ed. T. E.
Creighton) pp. 311-330. IRL Press, Oxford.
11. Santoro, M. M. and Bolen, D. W. (1988) Biochemistry 27, 8063-8068; Min, Y. and Bolen,
D. W. (1995) Biochemistry 34, 3771-3781.
12. Tanford, C. (1970) Adv. Protein Chem. 23,1-95; Aune, K. and Tanford, C. (1969)
Biochemistry 8, 4586-4590.
13. Makhatdze, G. I. and Privalov, P. L (1992) J. Mol. Biol. 226, 491-505.
14. Staniforth, R. A., Burston, S. G., Smith, C. J., Jackson, G. S., Badcoe, I. G., Atkinson, T.,
Holbrook, J. J. and Clarke, A. R. (1993) Biochemistry 32, 3842-3851.
15. Myers, J. K., Pace, C. N. and Scholtz, J. M. (1995) Protein Sci. 4, 2138-2148.
16. Barrick, D. and Baldwin, R. L. (1993) Biochemistry 32, 3790-3796.
17. Eftink, M. R. (1998) Biochemistry (Moscow] 63, 276-284.
18. Lepcock, J. R., Ritchie, K. P., Kolios, M. C., Rodahl, A. M., Heinz, K. A. and Kruuv, J.
(1992) Biochemistry 31.12706-12712.
19. Ramsay, G. D., lonescu, R. and Eftink, M. R. (1995) Biophys. J. 69, 801-707.
20. Safar, J., Roller, P. P., Gajdusek, D. C. and Gibbs, C. J., Jr (1993) J. Biol. Chem. 268,
20276-20284.
21. Inoescu, R. M. and Eftink, M. R. (1997) Biochemistry 36,1129-1140.
22. Kuhlman, B. and Raleigh, D. P. (1998) Protein Sci. 7, 2405-2412.
23. DeKoster, G. T. and Robertson, A. D. (1995) J. Mol. Biol. 249, 529-534.
24. Schmid, F. X. (1989) In Protein structure. A practical approach (ed. T. E. Creighton), pp.
251-285. IRL Press, Oxford.
25. Eftink, M. R. (1994) Biophys. J. 66, 482-501.
26. Kirby, E. P. and Steiner, R. F. (1970)J. Phys. Chem. 74, 4480-4490.
27. Robbins, R. J., Fleming, G. R., Beddard, G. S., Robinson, G. W., Thistlethwaite, P. J.
and Woolfe, G. J. (1980) J. Am. Chem. Soc. 102, 6271-6279.
28. Burstein, E. A., Vedenkina, N. S. and Ivkova, M. N. (1973) Photochem. Photobiol. 18,
263-279.
29. Ramsay, G. D. and Eftink, M. R. (1994) Methods Enzymol. 240, 615-645.
30. Jaenicke, R. (1991) Biochemistry 30, 3147-3161.
31. Gittleman, M. S. and Matthews, C. R. (1990) Biochemistry 29, 7011-7020.
32. Eftink, M. R., Helton, K. J., Beavers, A. and Ramsay, G. D. (1994) Biochemistry 33,
10220-10228.
33. Bowie, J. U. and Sauer, R. T. (1989) Biochemistry 28, 7139-7143.
34. Pakula, A. A. and Sauer, R. T. (1989) Proteins: Struct. Funct. genet. 5, 202-210.
35. Herold, M. and Kirschner, K. (1990) Biochemistry 29,1907-1913.
36. Bolotina, I. A., Kurochkin, A. V. and Kircpichnikov, M. P. (1983) KEBS Letters 155,
291-294.
326
USE OF OPTICAL SPECTROSCOPIC METHODS
327
This page intentionally left blank
Chapter 13
The use of spectroscopic
techniques in the study of DNA
stability
John Santalucia, Jr
Department of Chemistry, Wayne State University, Detroit, Ml 48202, USA
1 Introduction
Accurate determination of nucleic acid thermodynamics has become increas-
ingly important in understanding biological function as well as applications in
biotechnology and pharmaceuticals. Knowledge of the thermodynamics of DNA
hybridization and secondary structure formation is necessary for understanding
DNA replication fidelity (1), mismatch repair efficiency (2) and the mechanism
of DNA triplet repeat diseases (3). In addition, RNA folding thermodynamics are
an important aspect of understanding ribozyme catalysis, as well as understand-
ing the regulation of protein expression, mRNA stability and the mechanism of
protein synthesis by the ribosome (4). With the genome sequencing era upon us
(5), it will increasingly become important to predict the folding and hybrid-
ization thermodynamics of DNA and RNA, so that accurate diagnostic tests for
genetic and infectious diseases can be developed. Thus, there is a need to
develop a database of accurate thermodynamic parameters for different nucleic
acid folding motifs (4).
This chapter describes practical aspects of the application of UV absorbance
temperature profiles to determine the thermodynamics of nucleic acid struc-
tural transitions. Protocols and practical advice are presented for issues not
normally addressed in the primary literature but that are crucial for the
determination of reliable thermodynamics, such as sequence design, sample
preparation, choice of buffer, protocols for determining strand concentrations
and mixing strands, design of microvolume cuvettes and cell holder, instru-
mental requirements, data analysis methods, and sources of error. References to
the primary literature and reviews are also provided where appropriate. Sections
of this chapter have been adapted from previous reviews and are reprinted with permission
from the Annual Review of Biochemistry, Volume 62 © 1993, by Annual Reviews
wwwAnnualReviews.org (6) and with permission from Biopolymers © 1997, by John Wiley
& Sons, Inc. (4).
329
JOHN SANTALUCIA, JR
2 Overview of UV melting
The temperature-induced transition between native and random coil states of a
nucleic acid can be conveniently monitored by ultraviolet (UV) absorbance (see
(4, 9, 10) for reviews). The reason for this is that stacked bases have a smaller
absorption per base than unstacked bases; this is called hypochromicity (11,12)
which is defined as:
where Adenatured and Anative are the absorbances at high and low temperature,
respectively. The absorbance versus temperature profile is commonly referred
to as a UV absorbance melting curve (Figure 1). As the temperature increases, the
ratio of molecules in the single-stranded versus native states increases, resulting
in an increase in the UV absorbance. The melting temperature, TM, is defined as
the temperature at which half of the strands are in the native state and half are
in the 'random coil' state. Whereas many methods such as circular dichroism
and NMR can be used to monitor thermal denaturation, UV absorbance is the
most sensitive due to the high molar absorption of the bases (6).
The simplest way to derive thermodynamic parameters from UV melting data
is to apply a van't Hoff analysis of the data by assuming a two-state model (i.e.
native and denatured states) and that the difference in heat capacities of the
native and denatured states, ACP°, is zero (13-16) (more complex models are de-
scribed in Section 5). At each temperature the absorbance can be used to
calculate the fraction of strands in the native and denatured states, thereby
allowing the calculation of an equilibrium constant (10). Thus, the absorbance
versus temperature profile is used to determine the temperature dependence of
Denatured
Upper baseline__..
Lower baseline
Native Structure
20 30 40 50 60 70 90
Temperature (°C)
330
THE USE OF SPECTROSCOPIC TECHNIQUES IN THE STUDY OF DNA STABILITY
the equilibrium constant, allowing the calculation of AH° and AS0 for the tran-
sition (10) from a van't Hoff plot (i.e. InK versus 1/T plot). AG°T is then calculated
using equation 2 (discussed further in Section 5).
AG^ = - RT In K = AH° - TAS° 2
In practice, however, to obtain accurate thermodynamics from the shape of a
melting curve it is necessary to subtract upper and lower temperature baselines
because the extinction coefficients of the denatured and folded states are
temperature dependent (9, 14, 16). The details of one commonly used program
for fitting UV melting curves were published recently (17).
For sequences that form self-complementary duplexes, the melting tempera-
ture, TM, is calculated using equation 3.
_ AH°
M 3
~ AS° + RlnC T
where R is the gas constant (1.987 cal/K mol) and CT is the total oligonucleotide
strand concentration. For non-self-complementary molecules, CT in equation 2.3
is replaced by CT/4 if the strands are in equal concentration, or by (CA - C B /2) if
the strands are in different concentration, where CA and CB are the concentra-
tions of the more concentrated and less concentrated strands, respectively.
Observing the transition with a different technique (e.g. calorimetry, NMR, or
circular dichroism) can test the two-state assumption. The observation of
isosbestic or isodichroic points in the absorbance or circular dichroism spectra,
respectively, is diagnostic of a two-state transition (6). Alternatively, the melting
profile can be monitored at a different wavelength, so that contributions from
different nucleotides are emphasized. Deriving thermodynamics from the
concentration dependence of the TM (see below) can also test the two-state
assumption. If all methods give the same results, the two-state approximation is
validated since the methods have different sensitivities to each species.
The dependence of the TM on the oligonucleotide strand concentration re-
veals the molecularity of the transition (6). For example, formation of a hairpin
structure is a unimolecular process and therefore does not depend on concen-
tration, whereas formation of a duplex is bimolecular and is concentration
dependent for oligonucleotides. Polynucleotides show little concentration de-
pendence of the TM because double-strand initiation, which is the event depend-
ent on concentration, is only a small fraction of the total free energy involved in
the transition. Higher order complexes such as triplexes and quadruplexes
show stronger concentration dependencies (10, 18), but are usually non-two-
state. The concentration dependence of the TM provides an alternative van't Hoff
method for calculating folding thermodynamics. For bimolecular reactions,
thermodynamic parameters can be derived by rearranging equation 3 to give
equation 4 and plotting reciprocal melting temperature (in Kelvins) versus
logarithm of CT (19) (Figure 2).
1 R . „ AS0
331
JOHN SANTALUCIA, JR
334
332
,-> 330
t-i
t4 328
Wj
'0 326
X 324
I
SB 322
320
318
316
-14 -13 -12 -11 -10 -9
Figure 2 Typical 1/TM versus lnCT/4 plot. The data are for the duplex
CGTCTCTCOGGAGJCACG at strand concentrations from 2.62 x 10~" M to 4.00 x 10-6 M in
1.0 M NaCI, pH 7 solution (2). The thermodynamic parameters derived from the plot are:
AG°37 = -6.90 ± 0.21 kcal/mol, AH° = -56.8 ± 0.9 kcal/mol and AS0 = -160.9 ± 2.1
e.u. The thermodynamic parameters derived from the average of the fits are: AG°37,= -6.89
± 0.10 kcal/mol, AH" = -63.7 ± 2.1 kcal/mol and AS" = -183.2 ± 6.4 e.u.
The concentration should be varied more than a factor of fifty for reliable
measurements of AH° and AS0. The general equations for two-state melting of
complexes of any molecularity are provided in an elegant paper by Marky and
Breslauer (18). If the molecularity is known from UV absorbance mixing curves,
where absorbance is measured while varying the stoichiometry of the reacting
strands (20), or from another technique such as NMR, size exclusion chromato-
graphy or non-denaturing gel electrophoresis, then equation 4 can be used to test
the two-state approximation. For example, if significant concentrations of
intermediates are present, then the AH° from equation 2 (from fitting the shape
of the curves) will be expected to be smaller than the AH° from equation 4 (21).
On the other hand, if the transition is known to be two state, differences in AH°
from equation 2 and equation 4 can reveal the molecularity. For example, a triplex
would give a AH° value from equation 2 which is 50% larger than that from
equation 4, and would indicate that equations appropriate for triplexes not
duplexes need to be used (18, 21).
It is important to note that the AH° and AS0 determined by van't Hoff analysis
are determined at the TM. If large temperature extrapolation is required (greater
than -20°C away from TM), then heat capacity effects should be accounted for.
Previous data have indicated that ACp is usually small for nucleic acids (14, 15).
Based on these data, an approximate estimate of AC° is in the range of 0-120
entropy units (cal K-1 mol) per base pair and AC£ is assumed to be temperature
independent. Given AC;, the AHj, AS0-, and AGj- can be calculated more
accurately than equation 2 with the equations 5 (14).
AH* = AH°Tm + AC; x (T - Tm) 5a
AS£ = AS°Tm + AC; X In (T/TJ 5b
AG^ = AH°Tm + AC; x (T - Tm) - TAS°Tm - TAG; X In (T/TJ 5c
332
THE USE OF SPECTROSCOPIC TECHNIQUES IN THE STUDY OF DNA STABILITY
333
JOHN SANTALUCIA, JR
3 Sample
3,1 Sequence design
Design of optimal nucleic acid sequences to address specific scientific questions
or to determine specific incremental contributions to overall thermodynamics
is something of an art. There are, however, a number of simple guidelines that
we routinely use that minimize the number of samples that give artefactual
data. In general, a little time spent on careful experimental design results in a
large saving in time and confusing results in the long run. The two most basic
principles are (i) careful design of individual sequences that do not have the
potential to form undesired structures and (ii) to make multiple measurements
for each unknown so that sequences with anomalous thermodynamics are
readily identified. On-line servers for secondary structure prediction of single-
stranded RNA and DNA are very useful for deducing potential alternative
structures (see http://www.ibc.wustl/~zuker and http://JSLl.chem.wayne.edu/).
Algorithms for the prediction of duplexes and higher molecularity complexes
are currently under development in my laboratory. No matter how carefully
one designs sequences, however, there are always occasional examples where
new motifs with surprising stability are discovered, for example, the excep-
tional stability of tandem GA mismatches in DNA and RNA was unanticipated
(35, 36). Sequences with runs of three or more guanines should be avoided since
they form G-quartet structures and consistently yield lower than expected AH°
parameters (34). Sequences should also be designed to have TM values close to
physiological temperature (37 °C) to minimize temperature extrapolation errors
(particularly if ACP° is large) (13). Sequences should also be designed to have TM
values between 20 to 75 °C to allow upper and lower baselines to be adequately
defined.
For hairpins, sodium concentration below 0.1 M is recommended to minimize
bimolecular self-complementary internal loop formation (37). The predicted
thermodynamics from RNA-MFOLD (28, 38) or DNA-MFOLD (39) programs are
accurate enough to deduce if the putative internal loop will form in high
population (to use MFOLD the two strands should be connected by the dummy
sequence LLL (M. Zuker and J. SantaLucia, unpublished). Simulating the pre-
dicted populations of the hairpin and duplex forms by solving the simultaneous
equilibria is also recommended (31, 36).
For bimolecular structures (duplexes), it is generally recommended that self-
complementary sequences be avoided, since they have high propensity to form
competing hairpin structures. Even non-self-complementary duplexes can some
334
THE USE OF SPECTROSCOPIC TECHNIQUES IN THE STUDY OF DNA STABILITY
335
JOHN SANTALUCIA, JR
combined measurements from our laboratory and others in the literature (108
sequences) to determine the 10 Watson-Crick nearest-neighbour propagation
AG°37 parameters and two initiation parameters (31). The average standard error
in the AG°37 of each of the 108 sequences is estimated as 0.4 kcal/mol; thus, the
average standard error in the 12 unknown parameters is estimated as 0.4
kcal/mol/(108 - 12)1/2 = 0.04 kcal/mol.
In our actual study (31), we rigorously propagated individual errors of experi-
mental measurements to the derived parameters using the variance-covariance
matrix during singular value decomposition (52) and the results are in good
agreement with the average 0.04 kcal/mol error calculated above. This example
dramatically illustrates the value of performing a large number of measure-
ments to reduce statistical uncertainties in derived parameters. It should be
noted, however, that the propagated errors reflect precision of the determined
parameters and not the accuracy, because of the possible presence of systematic
errors in measurements. Another important point is that the derived parameters
are also inaccurate because the imposed model itself is limited (e.g. due to next-
nearest-neighbour interactions). The issue of error propagation during regression
analysis from the measured thermodynamics to the calculated parameters is
discussed in detail in our previous work (4, 31, 53). We have also extended this
type of analysis (i.e. multiple measurements for each unknown) to the deter-
mination of mismatch contributions (31) and this methodology is readily
extended to the determination of other motifs.
336
THE USE OF SPECTROSCOPIC TECHNIQUES IN THE STUDY OF DNA STABILITY
Protocol 1
SEP-PAK desalting (7)
Equipment and reagents
• Solution A: 10 mM ammonium Solution B: 30% acetonitrile (HPLC
bicarbonate (or ammonium acetate) grade)/70 % deionized water with no added
adjusted to pH 4.5 with 1 M HC1. pH 4.5 is buffer.
critical for efficient sample loading of Nucleic acid sample
most sequences. It is important that the
buffer used is volatile so that it does not SEP-PAK column
carry over into your final sample 'Speed-Vac' evaporator or equivalent
preparation. Occasionally, we find that 1.5 ml centrifuge tubes
adjusting the pH to 7 helps sequences that 0.2 micron Millipore filters
are rich in A, C or U (T) to stick to the
column. Quartz cuvettes
Method
1 (Optional) Filter Solutions A and B through a 0.2 micron Millipore filter. This will
remove dust and sterilize the buffers. Do not autoclave the buffers because
acetonitrile and ammonium bicarbonate are volatile.
2 Add 5 ml of Solution A to the dried nucleic acid sample,
3 Attach the SEP-PAK column (C-18 reverse phase column) to a 10 ml syringe barrel
and clamp the syringe barrel to a ring stand. Equilibrate the SEP-PAK column: (i)
rinse with 10 ml of 100% HPLC grade acetonitrile and (ii) rinse with 5 ml of Solution
A. Air bubbles that can impede the solvent flow are easily removed by using a glass
Pasteur pipette to 'swish' solvent in the syringe barrel. Gravity is sufficient force for
all steps if air bubbles are removed. If desired, the flow rate can be increased by
repeatedly pushing the syringe plunger -0.5 cm into the barrel and then removing
the plunger.lt is important to never let the SHP-PAK go dry.
4 Set up a series of 1.5 ml centrifuge tubes to collect fractions (including the sample
load and wash). Directly load the sample onto the column (without letting the SEP-
PAK go dry). Wash with 5 ml Solution A (this desalts the oligomer and also removes
other small molecule impurities). Wash with 5 ml Solution B (30% acetonitrile).
5 Quantify number of A260 units (see Protocol 2) of all fractions including the sample
load and wash. Combine only the fractions eluted with Solution B that contain
large amounts of ougonucleotide. Do not combine the load and wash fractions
since they are not desalted.
6 Dry the sample either by lyophilization or by centrifugal evaporation (i.e. use the
'speed-vac' apparatus).
337
JOHN SANTALUCIA, JR
Protocol 2
Measurement of optical density units
The concept of absorbance units at 260 nm, AMO (also known as optical density units,
ODU) often causes confusion for students. By definition, one A260 unit is the amount of
sample that gives a UV absorbance at X = 260 nm of 1.0 if the sample is dissolved in 1 ml
volume and measured in a 1-cm pathlength cuvette. Since aborbance is proportional to
concentration (Beer's Law} and the volume is 1 ml, the number of A260 units of a sample
is proportional to the number of moles of strands. Knowledge of the number of A2b0 units
is important for planning UV melting experiments to optimize sensitivity.
Method
1 'Zero' the spectrophotometer by placing deionized water or buffer in a cuvette and
use the UV spectrophotometer's calibration routine,
2 Dissolve the DNA (or RNA) in 1 ml of deionized water, put into a 1-cm pathlength
raicrovolume cuvette, and measure the absorbance at 260 nm. If the absorbance is
less than 2, then the absorbance reading is the number of A260 units. If the sample
is dissolved in a volume other than 1 ml. then use the following equation:
Total A260 units = A x V/l
where A is the measured absorbance, V is the sample volume in ml and 1 is the
pathlength in cm.
3 If the absorbance is greater than 2, then the sample must be diluted (usually 10-fold
is sufficients or the absorbance measured in a short pathlength cuvette. Take 0.1 ml
of the original sample and add 0,9 ml of water and put in the 1 cm cuvette.
Remeasure the absorbance. Multiply the absorbance reading by the dilution factor,
D (a factor of 10 in this example), according to the following equation:
Total A260 units = A x D/l
4 To determine the number of moles of sample, we simply rearrange Beer's Law and
use the molar extinction coefficient at 260 nm, c, calculated from the nearest-
neighbour model (10, 63):
Moles = (Total A260 units)/e
Note that if the oligonucleotide folds at room temperature, the extinction
coefficient must be corrected for hypochromicity or the absorbance measured at
-85 °C.
5 Aliquot the volume of solution that contains the desired number of absorbance
units (typically 3.0 A260 units) of oligonucleotide for the melting experiment, and
evaporate to dryness. See Protocol 4 for a description of a procedure to accurately
mix non-self-complementary strands in equal concentration.
338
THE USE OF SPECTROSCOPIC TECHNIQUES IN THE STUDY OF DNA STABILITY
339
JOHN SANTALUCIA, JR
Protocol 3
Sample degassing
Equipment
• Speed-Vac evaporator or equivalent • Quartz microcuvettes
Method
1 The major dissolved gas in aqueous solutions is 02. To remove dissolved oxygen, we
recommend gently bubbling argon through the melting buffer. The buffer can then
be added to the dried nucleic acid sample without introducing gas by gentle
pipetting. Samples also spontaneously degas during the annealing step of the UV
melting experiment (Protocol 6).
2 An alternative procedure is to fill the quartz microcuvette with the desired sample
and to place the cuvette in a centrifuge tube. The sample is then briefly
(approximately 1 min) centrifuged at low speed under vacuum (a 'speed-vac' works
well for this). The only drawback to this procedure is that occasionally it is difficult
to remove the cuvette from the centrifuge tube.
340
THE USE OF SPECTROSCOPIC TECHNIQUES IN THE STUDY OF DNA STABILITY
Protocol 4
Mixing equal concentrations of non-self-complementary
strands* (8)
Equipment and reagents
• Desalted strand A 1.5 nil centrifuge tubes
• Desalted strand B Quartz cuvettes (0.1 cm pathlength)
• Melting buffer Speed-Vac evaporator or equivalent
Method
1 Aliquot -2.5 ODU of desalted strands A and B in separate tubes. Evaporate both
samples to dryness and dissolve each sample in 100 ul melting buffer (not water).
2 Determine the concentration of each strand by placing 60 ul of each sample in
separate 0.1 cm pathlength microcuvettes and measuring their absorbances at 260
nm and at high temperature (we usually use 90 °C). The concentrations, CA and CB,
are then determined from their respective absorbances, single-stranded extinction
coefficients (calculated from the nearest neighbour model (10, 63)). and using
Beer's Law. It is important to measure the absorbances at high temperature since
strands often form self-structure at low temperatures resulting in a hypochromic
absorbance.
3 Use the following equations to calculate the volumes of strands A and B to be
mixed:
VA - 160 X CB - (CA + CB)
VB = 160 X CA - (CA + Cu)
4 Mix together volumes VA and VB of strands A and B, respectively (note VA + VB =
160 ul). The resulting solution has equal concentrations of the two strands within
about 10%. It is important to use gentle pipetting back and forth to achieve good
mixing.
5 To calculate the average e for the mixed sample (which will be used later to
calculate the total strand concentration, CT). use the equation:
"This procedure yields 160 u1 of DNA with strands in equal concentration that can be used for
the UV melting dilution series (Protocol 6) experiment. If a different volume is desired, then
simply substitute the desired volume for 160 in all of the equations below.
341
JOHN SANTALUCIA, JR
Protocol 5
Sample dilution scheme (8)
The following protocol provides 10 concentrations over a 100-fold range and utilizes
custom manufactured microcuvettes" (see text).
Equipment
• Quartz microcuvettes (0.1 cm pathlength)
Method
1 Start with 4 A260 units of dry sample. Dissolve in 160 u1 of the desired melting
buffer. If a non-self-complementary duplex is melted, then use the solution result-
ing from Protocol 4. We usually perform the melting experiment at 280 nm for
Dilution series 1 and at 260 nm for Dilution series 2 so that a larger concentration
range can be studied,
2 Dilution series 1:
(i) Take 60 ul (out of the 160 ul total) and place in cell 1 (0,1 cm). Check that the
absorbance of this sample is below 1.6 at the desired wavelength, dilute with
melting buffer if necessary.
(ii) To the 100 ul remaining solution, add 70 ul of melting buffer, take 60 ul (out of
the 170 ul total) and place in cell 2 (0.1 cm).
(iii) To the 110 u1 remaining solution, add 77 u1 of melting buffer, take 60 u1 (out of
the 187 ul total) and place in cell 3 (0,1 cm).
(iv) To the 127 u1 remaining solution, add 89 u1 of melting buffer, take 120 ul (out
of the 216 u1 total) and place in cell 4 (0.2 cm),
(v) To the 96 ul remaining solution, add 67 ul of melting buffer, take 120 ul (out of
the 163 u1 total) and place in cell 5 (0.2 cm). 43 ul is leftover for the next
dilution series.
3 Dilution series 2: Combine the contents of cells 2, 3 and 4 (300 ul total) with the left
over solution from Dilution series 1 (43 (il) to give a total of 343 ul. If you suspect
sample degradation from the previous melt (as might occur for RNA in the
presence of divalent metals), then prepare a fresh sample for the second dilution
series. If sample evaporation occurred in one of the cuvettes in Dilution series 1 (as
evidenced by a large air bubble in the cuvette}, then do not combine this sample
since the salt concentration in that sample will be more concentrated than desired,
(i) To the combined volume of 343 u1 add 600 ul of melting buffer. Take 300 u1
(out of the 943 ul total) and place in cell 1 (0.5 cm). Check that the absorbance
of this sample is below 1.6 at the desired wavelength, dilute with melting
buffer if necessary.
(ii) To the 643 uI remaining solution, add 400 ul of melting buffer, take 300 ul (out
of the 1043 ul total) and place in cell 2 (0,5 cm).
(iii) To the 743 ul remaining solution, add 470 ul of melting buffer, take 600 ul (out
of the 1213 uI total) and place in cell 3 (1.0 cm).
342
THE USE OF SPECTROSCOPIC TECHNIQUES IN THE STUDY OF DNA STABILITY
Protocol 5 continued
(iv) To the 613 ul remaining solution, add 390 u1 of melting buffer, take 600 ul(out
of the 1003 ul total) and place in cell 4 (1,0 cm).
(v) To the 403 ul remaining solution, add 230 u1 of melting buffer, take 600 u1 (out
of the 633 ul total) and place in cell 5 (1.0 cm).
"Important notes on filling cuvettes: Before filling cuvettes, visually inspect them for cleanliness
(see Protocol 8). When filling cuvettes, be sure to leave a small air space between the top of the sample
and the stopper, to allow for the ~4% volume expansion that occurs for water between 0 to
95 °C If you do not do this, then either the cap will pop off during the melt or the cuvette will
crack. It is also important to use gentle pipetting back and forth to achieve good mixing for
each, of the dilutions. Also be sure to insert a 1 cm x 1 cm strip of Teflon tape between the
stopper and the cuvette to provide a tight seal. Warning: The stopper must be inserted with
sufficient force so that it will not pop off during the melting curve. On the other hand, if the
stopper is forced in too tightly you will crack the cuvette. There is no need to run a buffer
control for each melting curve, since subtracting the buffer data from the sample data adds
noise to the data; thus, all five cuvettes are filled with sample.
4 Instrumentation
4. 1. Microvolume cuvettes and aluminium cuvette adapters
Standard 1 cm x i cm cuvettes are not appropriate for UV melting studies of
nucleic acids because they require too much sample and they are difficult to hcat
evenly. A much preferred option is to have microvolume cuvettes custom manu-
factured (both Precision Glass Inc. and Helma Cells produce high quality cuvettes).
We use microvolume cuvettes with the following pathlengths and volumes: 0.1
cm path, 60 ul; 0.2 cm path, 120 ul; 0.5 cm path, 300 ul; 0.8 cm path, 480 ul; 1.0
cm path, 600 ul With this set of cuvettes, we routinely investigate an 80-100-fold
range in oligonuclcotide concentration (Protocol 5). If millimolar concentrations are
desired (e.g. from an NMR sample), then the 0.1 cm path cells can have a 0.09 cm
thick quartz window inserted to give a net path of 0,01 cm (10). In designing the
cuvettes it is important to minimize the sample volume, minimize the amount of
quartz (since quartz is a thermal insulator), and provide a design that lasts at least
200 melting curves before stress leads to cracking. The cuvettes also need to have- a
tightly fitted Teflon stopper or screw cap, or be covered with Dow silicone oil
(Corning, 200-fluid 20 centipoise viscosity) (do not overfill to prevent spillage
during the volume expansion that occurs with temp-erature increases). Detailed
drawings of our microvolume cuvettes can be ob-tained from the following web
address: http://JSLl .chem.wayne.edu or upon request from the author.
In order to fit the microvolume cuvettes into the standard 1,25 cm x 1.25 cm
sample chamber, it is necessary to have custom adapters machined from
aluminium which properly position the microcuvette in the light beam and
provide optimal thermal contact with the thermoelectric controller. Since the
microcuvettes are different sizes, separate aluminium adapters are required for
the different pathlength cuvettes. An important consideration in the design of
343
JOHN SANTALUCIA, JR
the aluminium adapter is the exact position of the light beam and that the
adapter be snug enough to provide good thermal contact and yet have some
tolerance to allow for volume expansion during the melting curve. The adapter
should also allow for the slight variance in stopper heights. Detailed drawings of
the aluminium adapters can also be obtained from the web address given above.
4.2 Spectrophotometer
There are several instrumental requirements for precision UV melting curve
analysis. Accuracy and stability of the absorbance reading at least in the forth
decimal place is required. This is routinely achieved with double-beam instru-
ments (e.g. our Aviv 14DS is stable in the fifth decimal place) as well as well-
designed single-beam instruments such as the Gilford-250/260 or Beckman-DU
instruments. The instrument also needs to be interfaced to a temperature
controller (see below) and the absorbance and temperature readings need to be
continuously output to a computer file for later data processing. It is important
that the raw absorbance data are output, since many instruments apply 'box-
car' or other averaging methods which make melting curves look smoother, but
distort the shape of the melting curve and decrease the temperature resolution
so that subtle transitions are missed; both of these effects negatively affect the
quality of the derived thermodynamic parameters. The sample chamber should
be connected to a compressed nitrogen source with flow controller. This allows
the sample compartment to be purged with nitrogen so that water conden-
sation is prevented at low temperatures.
The decisive factors in choice of commercial instruments for UV melting
curves are the method of temperature control, method of temperature measure-
ment and the number of cuvettes in the cell compartment. Peltier effect heating/
cooling is the preferred method since circulating water baths with jacketed
cuvettes usually do not change temperature quickly enough to be convenient
(cooling is particularly slow, often taking several hours). All instruments with
Peltier-effect heaters are not equivalent; many of the commercially available
spectrophotometers can cool only to 10°C and heat to ~80°C. The Peltier-effect
heaters in the Aviv 14DS allow routine heating/cooling in the entire 0-100°C
range. The Aviv temperature controller can actually go to significantly higher or
lower temperatures than this range if appropriate solvent mixtures are used to
prevent sample freezing and/or boiling. An important maintenance tip is to
periodically (every 3 months or so) blow out the sample compartment and
thermoelectric controller with compressed nitrogen. This removes dust which
can significantly decrease the performance of the thermoelectric controller. A
thermoelectric controller with multiple cuvettes is essential if many melting
curves need to be recorded. The Aviv 14DS is equipped with a five-cuvette
thermoelectric controller, that allows us to routinely record 15 melting curves
per day. This makes studying the concentration dependence of multi-molecular
complexes routine. A design feature of the Aviv thermoelectric controller is that
the samples are rotated in a circle rather than the linear configuration used in
most commercially available instruments. This ensures uniform heating of the
344
THE USE OF SPECTROSCOPIC TECHNIQUES IN THE STUDY OF DNA STABILITY
Protocol 6
Typical melting protocol
Equipment
• UV spectrophotometer with thermostated • Quartz cuvettes/microcuvettes
cell holding
Method
1 Turn on the instrument at least half an hour before making measurements to allow
the D2 lamp to warm-up to minimize instrument drift. Purge the sample compart-
ment with nitrogen at least 20 min before a melting curve is recorded. Place buffer
in one cuvette and 'zero' the absorbance reading.
2 Load filled microcuvettes (Protocols 4 and 5} into the aluminium cell holders and
place into the thermoelectric controller.
3 Set the wavelength to 260 mm and raise the temperature to 90°C for 5 min. This
'annealing' helps equilibrate the structure and to degas the sample. Record the
absorbances in a laboratory notebook for future calculation of strand concentra-
tions. If optimum hypochromicity is desired, then record a high-temperature
wavelength scan while at high temperature.
4 Ramp the temperature down to 0°C (we actually go to -3'C for solutions in 1 M
NaCl) over —10 min. If the wavelength for optimum hypochromicity needs to be
JOHN SANTALUCIA, JR
Protocol 6 continued
Protocol 7
Sample recovery
UV melting studies typically use only 3 A260 units for a full set of curves and thus samples
are routinely discarded or archived at -70°C in case future analysis is required. If a
sample is precious, then it may be desirable to recover the sample:
1 Generally, it is recommended that the sample purity be checked by analytical HPLC
or gel electrophoresis to ensure that the sample has not degraded.
2 Typically, sample desalting by the methods above (see text and Protocol 1) is
sufficient for further experiments to be performed.
3 Treatment with Chelex-100 is crucial if the sample was previously dissolved in a
buffer containing multivalent metals.
346
THE USE OF SPECTROSCOPIC TECHNIQUES IN THE STUDY OF DNA STABILITY
Protocol 7 continued
In the event that Dow silicone oil (Corning, 200-fluid 20 centipoise viscosity) was
used to prevent sample evaporation, sample recovery can be readily obtained by
pipetting out the entire sample (including oil) onto a Teflon dish. The oil will stick
to the Teflon as the bead of sample is rolled around the dish (10).
Protocol 8
Cleaning of quartz cuvettes
Equipment and reagents
• Plastic gel loading pipette tip • Methanol (HPLC or spectroscopic grade)
• Plastic scissor clamps • Compressed nitrogen
• 50% nitric acid
Safety precaution: Wear safety goggles and latex gloves, since nitric acid is caustic and
mutagenic,
Method
1 After a melting curve is complete, remove the sample using a plastic gel loading
pipette tip.
2 Clamp the cuvette with locking plastic scissor clamps (the scissor clamp should
grip the frosted sides of the cuvette to prevent scratching the optical surfaces).
Remove traces of your sample by rinsing the cuvettes with distilled water in a
squirt bottle. The water is most easily removed from the cuvette with a syringe that
has the needle capped with a short length of narrow gauge plastic tubing (to
prevent the syringe needle from scratching the quartz). We do not recommend
shaking the cuvette upside down to remove the liquid contents since this often
results in the cuvette slipping out of the clamps and breaking the cuvette.
3 Fill the cuvettes with 50% nitric acid (if the nitric"acid is brown it has degraded and
a fresh solution should be made). Place the filled cuvette into a 50% nitric acid bath
for~10 min. Cover the acid bath with a watchglass to prevent dust from getting in.
4 Remove the sample from the nitric acid using the plastic scissor clamps and re-
move the nitric acid using a pipette (not the syringe, since the nitric acid will
corrode the metal syringe needle).
5 Rinse several times with water and only then rinse several times with methanol
(HPLC or spectroscopic grade), Never allow methanol to come into contact with the nitric
acid. Dry the cuvette with compressed nitrogen (be sure to hold the cuvette firmly
as the nitrogen pressure can cause it to be dropped).
6 Store the cuvettes on a benchtop covered with dust-free cloth. Cover the cuvettes
with an inverted beaker to prevent dust accumulation. Cuvettes stored this way are
ready to use for the next melting experiment without further cleaning (unless you
do not trust the cleaning technique of your labmates).
347
JOHN SANTALUCIA, JR
5 Data analysis
5.1 Curve fitting to calculate thermodynamic parameters
The methods for fitting UV melting curves and determining thermodynamics
have been extensively reviewed (4, 9, 10, 14, 16-18). The optimal choice of fit-
ting method depends on a number of factors including molecularity of the
transition, melting temperature of the transition (i.e. whether sufficient upper
or lower baseline is available) and which approximations are introduced (e.g.
two-state model with AC° equal to zero). This chapter presents the methods for
fitting unimolecular and bimolecular transitions since they form the majority
of measurements made on nucleic acids. For higher molecularities the reader is
referred to the primary literature (9, 18, 40). It is appropriate to reiterate here
that careful design of oligonucleotide sequences (see Sections 3.1 and 3.2) which
minimize the potential for the formation of undesired structures significantly
simplifies data analysis and interpretation.
348
THE USE OF SPECTROSCOPIC TECHNIQUES IN THE STUDY OF DNA STABILITY
a = —— 10
a+b
where a and b are the respective vertical distances from the upper and lower
baselines to the experimental melting curve (Figure 1}. The temperature at
which a is equal to 0.5 is denned as the melting temperature, TM. The fraction of
strands in the random coil state is given by 1 — a. Thus the equilibrium
constant at a particular temperature is given by:
By using sloping baselines, one is effectively removing from the shape analysis
those molecules in the sample with intermediate states. Hence, only those
molecules in the sample that actually melt two-state are used to calculate the
thermodynamics. Whilst technically this is somewhat alarming, there is general
agreement that the use of sloping baselines provides the most reliable thermo-
dynamic data (10, 14, 16). Assuming a two-state model with ACp° equal to zero
(i.e. the van't Hoff approximation), the AH° and AS0 can be derived from the
slope and intercept of a InK versus 1/T plot by rearranging equation 2:
In practice, only the data points for 0.15 < a < 0.85 are used due to experi-
mental uncertainty in the 1C values outside of this range (10).
An alternative and preferred method to determine thermodynamics is to
directly fit the entire absorbance versus temperature curve (9, 10, 14, 17). The
total absorbance at temperature T, A(T), is given by the sum of Beer's law
contributions of each component in the solution:
A(T) = eRC(T) X I X [ARC] + eF(T) X I X [AF] 13
where I is the pathlength. Substituting the linear approximations for the base-
lines (i.e. equations 7 and 8) and equation 9 gives:
A(T) = CT X I X [(mRC X T + bRC) X a(T) + (mF X T + bF) X (1 - a(T))] 14
According to equations 11 and 12, a(T) is determined by AH° and AS0. Thus the
experimental absorbance versus temperature curve is fit using multiple non-
349
JOHN SANTALUCIA, JR
[ARC]2 2 X (1 - a)2CT
For non-self-complementary duplexes the relevant equilibrium equations are:
ARC + BRC <— » AB 18
2[AB]
- -
uT
19
= [AB] = 2a
[ARC][BRC] (1 - «)2CT
Note the fourfold difference in equations 17 and 20 which is ultimately reflected
in the different concentration dependence of the TM for self-complementary
versus non-self-complementary sequences (equation 3).
As described in Section 2, the concentration dependence of the TM provides
an alternative van't Hoff method for determining thermodynamics. In this
method, the transition AH° and AS° are obtained by equation 4 from the slope
and intercept of a 1/TM against lnCT/4 plot. Since the TM is relatively insensitive
to non-two-state behaviour and to the choice of baselines, 1/TM versus logCT
plots provide very reliable AH°, AS0, and AG°37 parameters (16), particularly if
350
THE USE OF SPECTROSCOPIC TECHNIQUES IN THE STUDY OF DNA STABILITY
the TM is measured over a large concentration range. It is worth noting that the
TM cannot be accurately determined from an inflection point of the melting
curve determined from the maximum of the temperature derivative of the
absorbance (9, 10). It is necessary to subtract the upper and lower temperature
baselines before performing the differentiation, since they affect the position of
the apparent TM (10, 14). The preferred method for calculating the TM is to use
equation 3 using the AH° and AS° determined from the fit of the shape of the
curve with sloped baselines, as described above (14,16).
Since curve fitting and 1/TM against lnCT/4 plot methods depend differently
on the two-state approximation, agreement of AH° obtained from both methods
provides a test of the validity of the two-state approximation (6,10). Agreement
within 10-15% of AH° parameters from a 1/TM versus lnCT plot and from fits of
the shapes of melting curves is generally regarded as indicative of a two-state
transition (4), though caution is recommended. This criterion suggests the
standard error in the van't Hoff AH° parameters from optical melting is about
5-8%. In most cases that have been studied by both optical melting and
calorimetry, the AH° values are in agreement if the two ways of analysing the
optical data give the same value within 10% (15, 16, 34). Agreement between
enthalpy changes determined by different methods is a necessary, but not a
sufficient criterion to definitively establish two-state behaviour (4, 18). For ex-
ample, the self-complementary DNA sequence (CGTTG,CGXAACG)2 yielded AH°
parameters from a 1/TM versus lnCT plot and from fits of the shapes of optical
melting curves that agree within 10% for the single strand to duplex transition.
The temperature dependence of NMR spectra and comparison of that sequence
with thermodynamics for other sequences, however, revealed that it actually
melts through a hairpin intermediate (31).
An alternative criterion for two-state thermodynamics is to compare the
enthalpy changes from optical melting and calorimetry (15, 16, 34). For tran-
sitions with large ACp°, it appears that van't Hoff analyses and calorimetric data
provide systematically different AH° parameters and the possible origins of
these differences have been recently reviewed (32, 73). Even agreement
between van't Hoff analysis of optical melting and calorimetry does not
guarantee a two-state transition. The AH° values for the DNA duplex,
(GCGTACGCATGCG'CGCATGTGTACGC), are in agreement, but the transition is
not two-state as evidenced by the melting of the individual single strands (31,
40). Known exceptions of this type are rare, however.
For two-state transitions, the AH° parameters obtained by both methods are
equally reliable, so it is best to report the average of parameters from the 1/TM
versus lnCT plot and from the fits of the shapes of curves (41, 74). Obtaining AH°
from melting curves measured at different wavelengths can be used to check
the two-state approximation (6). Importantly, agreement between enthalpy
changes determined by different methods is a necessary, but not a sufficient
criterion to establish two-state behaviour (16, 27, 38). Additional methods for
validation of the two-state model are described in Section 2.
A van't Hoff analysis of UV melting data cannot be used to reliably measure
351
JOHN SANTALUCIA, JR
=
-^Norm ~7 21
A
HT
where ANorm is the normalized absorbance at high temperature and A is the raw
absorbance data. Equation 21 preserves the shape of the melting curve without
distorting the hypochromicity. Alternatively, the data can be 'double normal-
ized' to both high and low temperatures by applying equation 22 to each curve
and then plotting the data on the same graph.
_ A/A LT -1
A ~~
Norm -
HT/ALT
where ALT is the low temperature absorbance. Equation 22 makes all curves start
at 0 and finish at 1 and preserves the shape of the melting curve, but inform-
ation about hypochromicity is lost. Equation 22 is useful for comparing the
melting curves of different sequences (at the same concentration) or comparing
the melting curves of one sequence measured at different wavelengths.
352
THE USE OF SPECTROSCOPIC TECHNIQUES IN THE STUDY OF DNA STABILITY
353
JOHN SANTALUCIA, JR
26
AH° AH° (AH0)2
This equation assumes that there is negligible error in the lnCT term. This is
reasonable because a 10% error in CT propagates according to equation 23 to a 1%
error in lnCT at oligonucleotide concentrations in the range of 10-5 M (7). The
enthalpy-entropy compensation effect is evident in the high quality of pre-
dictions made for AG17 and TM (31,41, 53).
Another way to evaluate error propagation with minimal assumptions about
the experimental errors or how these errors propagate to the derived NN
(nearest neighbour) parameters is to use a resampling analysis of the data (79).
Consider the 'unified data set' of 108 sequences with only Watson-Crick base
pairs (31). First, SVD analysis is used to determine the linear least-squares fit of
the data to obtain the 10 NN parameters and the two initiation parameters as
described (31, 41, 53). Since the unified data set contains 108 equations with 12
unknowns, the problem is overdetermined. A total of 30 resampling trails were
used. For each trial, a different set of 68 randomly selected sequences was used
in the SVD analysis to calculate the 12 unknowns. It is important to check the
column rank of the stacking matrix for each trial. The results from the 30 trials
were averaged and the standard deviations calculated for each NN parameter.
This resampling analysis was performed for AG°,7, AH° and AS0. We also per-
formed controls in which certain classes of sequences (e.g. all sequences from a
particular lab or all sequences greater than 10 bp, etc.) were systematically omitted
from the SVD analysis and found that in all cases the parameters agreed within
the propagated experimental error of the NN from the total dataset.
References
1. Petruska, J., Goodman, M. F., Boosalis, M. S., Sowers, L. C, Cheong, C. and Tinoco, I.,
Jr (1988). Proc. Natl. Acad. Sri. USA., 85, 6252-6256.
2. Peyret, N., Seneviratne, P. A., Allawi, H. T. and SantaLucia, J., Jr (1999). Biochemistry,
38, 3468-3477.
3. Gacy, A. M. and McMurray. (1998). Biochemistry, 37, 9420-9434.
4. SantaLucia, J., Jr and Turner, D. H. (1997). Biopolymers, 44, 309-319.
5. Koonin, S. E. (1998). Science, 279, 36-37.
6. Jaeger, J. A., SantaLucia, J., Jr., and Tinoco, I., Jr. (1993). Annu. Rev. Biochem., 62,
255-287.
7. SantaLucia, J., Jr (1991) in Department of Chemistry, Ph.D. Thesis. University of Rochester,
Rochester, NY.
354
THE USE OF SPECTROSCOPIC TECHNIQUES IN THE STUDY OF DNA STABILITY
355
JOHN SANTALUCIA, JR
47. Hirao, I., Nishimura, Y., Tagawa, Y., Watanabe, K. and Miura, K. (1992). Nucleic Acid
Research, 20, 3891-3896.
48. Antao, V. P., Lai, S. Y. and Tinoco, I., Jr (1991). Nucleic. Adds Res., 19, 5901-5905.
49. Walter, A. E., Turner, D. H., Kim, J., Lyttle, M. H., Muller, P., Mathews, D. H. and
Zuker, M. (1994). Proc. Natl. Acad. Sci. USA, 91, 9218-9222.
50. Peyret, N. and SantaLucia, J., Jr (1999). (In preparation).
51. Bevington, P. R. (1969). Data reduction and error analysis for the physical sciences. McGraw-
Hill, New York. pp. 58-60
52. Press, W. H., Flannery, B. P., Teukolsky, S. A. and Vetterling, W. T. (1989) pp 52-64,
498-520. Cambridge University Press, New York.
53. Xia, T., SantaLucia, J., Jr, Burkard, M. E., Kierzek, R., Schroeder, S. J., Jiao, X., Cox, C.
and Turner, D. H. (1998). Biochemistry, 37,14719-14735.
54. Milligan, J. F., Groebe, D. R., Witherell, G. W. and Uhlenbeck, 0. C. (1987). Nucleic
Acids Res., 15, 8783-8798.
55. Capaldi, D. and Reese, C. (1994). Nucleic Acids Res., 22, 2209-2216.
56. Brown, T. and Brown, D. J. S. (1991). In Oligonudeotides and analogues (ed. F. Eckstein),
pp. 1-24. IRL Press, Oxford.
57. Zimmer, D. and Crothers, D. (1995). Proc. Natl. Acad. Sci. USA, 92, 3091-3095.
58. Arghavani, M. B. and Romano, L. J. (1995). Anal. Biochem., 231, 210-209.
59. McLaughlin, L W. (1989). Chem. Rev., 89, 309-319.
60. Chou, S.-H., Flynn, P. and Reid, B. (1989). Biochemistry, 28, 2422-2435.
61. Zarrinkar, P. P. and Williamson, J. R. (1994). Science, 265, 918-924.
62. Primrose, W. U. (1993). In NMR of macromokcules: A practical approach (ed. G. C. K.
Roberts), pp. 7-34. IRL Press, Oxford.
63. Richards, E. G. (1975). In Handbook of biochemistry and molecular biology: Nucleic acids (ed.
G. D. Fasman), Vol. I, p. 597. CRC Press, Cleveland, OH.
64. Yu, Q., Kandegedara, A., Xu, Y. and Rorabacher, D. B. (1997). Anal. Biochem., 253,
50-56.
65. Allawi, H. T. and SantaLucia, J., Jr (1998). Biochemistry, 37, 9435-9444.
66. Perrin, D. D. and Dempsey, B. (1979). Buffers for pH and metal ion control. Halsted Press,
New York. pp. 6-20
67. Lewis, E. A., Hansen, L. D., Baca, E. J. and Temer, D. J. (1976). J. Chem. Soc. Perkin II
125-128.
68. Delcourt, S. G. and Blake, R. D. (1991). J. Biol. Chem., 266,15160-15169.
69. Blake, R. D., Vosman, F. and Tarr, C. (1981). In Biomolecular stereodynamics (ed. R. H.
Sarma), Vol. I, pp. 439-458. Adenine Press, New York.
70. Yen, S.-W. W. and Blake, R. D. (1980). Biopolymers, 19, 681-700.
71. Freier, S. M., Hill, K. O., Dewey, T. G., Marky, L. A., Breslauer, K. J. and Turner, D. H.
(1981). Biochemistry, 20,1419-1426.
72. Gralla, J. and Crothers, D. M. (1973). J. Mol. Biol., 73, 497.
73. Chaires, J. B. (1997). Biophys. Chem., 64,15-23.
74. Sugimoto, N., Nakano, S., Yoneyama, M. and Honda, K. (1996). Nucleic Acids Res., 24,
4501-4505.
75. Wartell, R. M. and Benight, A. S. (1985). Phys. Rep., 126, 67-107.
76. Vologodskii, A. V., Amirikyan, B. R., Lyubchenko, Y. L. and Frank-Kamenetskii, M. D.
(1984). J. Biomol. Struct. Dyn., 2,131-148.
77. Meyer, S. L. (1975). Data analysis for scientists and engineers. Wiley, New York. Chapter 19
78. Snedecor, G. W. and Cochran, W. G. (1971). Statistical methods. The Iowa State
University Press, Ames, IA. p. 190
79. Efron, B. and Tibshirani, R. (1993). An introduction to the bootstrap. Chapman & Hall,
London.
356
List of suppliers
This core list of suppliers appears in all books in the Practical Approach series.
357
LIST OF SUPPLIERS
Dupont (UK) Ltd, Industrial Products Division, Wedgwood Way, Stevenage, Herts,
SGI4QN
Tel: 01438 734000 Fax: 01438 734382 Web site: www.dupont.com
Dupont Co, (Biotechnology Systems Division), PO Box 80024, Wilmington, DE 19880-
002, USA
Tel: (+)1 302 774 1000 Fax: (+)1 302 774 7321 Web site: www.dupont.com
Eastman Chemical Company, 100 North Eastman Road, PO Box 511, Kingsport,
TN 37662-5075, USA
Tel: (+)1 423 229 2000 Web site: www.eastman.com
Edinburgh Instruments Ltd, Riccarton, Currie, Edinburgh EH14 4AP, UK.
E-mail: [email protected]. URL: http://www.edinst.com.
358
LIST OF SUPPLIERS
Imaging Research Inc., Brock Univesiry, 500 Glenridge Avenue, St. Catherines,
Ontario L2S 3A1, Canada
Instruments S.A., Inc. (SPEX), 3880 Park Avenue, Edison, NJ 08820 USA.
URL: http://www.instrumentssa.com/.(See their web site for international offices.)
Intelligent Automation Systems, 149 Sidney Street, Cambridge, MA 02139, USA
IVEE Development AB, Forsta Langgata 26, SE-413 28 Goteborg, Sweden
Invitrogen BV, PO Box 2312, 9704 CH Groningen, The Netherlands
Tel: (+)800 5345 5345 Fax: (+)800 7890 7890 Web site: www.invitrogen.com
Invitrogen Corporation, 1600 Faraday Avenue, Carlsbad, CA 92008, USA
Tel: (+)1 760 603 7200 Fax: (+)1 760 603 7201 Web site: www.invitrogen.com
359
LIST OF SUPPLIERS
Olis (On-line Instrument Services), Inc., 130 Conway Drive, Suites A and B, Bogart,
GA 30622-1724, USA (http://www.olisweb.com/).
360
LIST OF SUPPLIERS
Omega Optical Inc., P.O. Box 573, Brattleboro, VT, 05302 - 0573, USA.
Email: [email protected].
Qiagen UK Ltd, Boundary Court, Gatwick Road, Crawley, West Sussex, RH10 2AX
Tel: 01293 422911 Fax: 01293 422922 Web site: www.qiagen.com
Qiagen Inc, 28159 Avenue Stanford, Valencia, CA 91355, USA
Tel: (+)1 800 426 8157 Fax: (+)1 800 718 2056 Web site: www.qiagen.com
Quantum Northwest, 9723 W. Sunset Highway, Spokane, WA 99224-9426 USA.
E-mail: [email protected]. URL: www.qnw.com.
Research Genetics Inc., 2130 Memorial Pkwy SW, Huntsville, AL 35801, USA
Research Instrumentation Shop, University of Pennsylvania School of Medicine,
79E John Morgan Building, 3620 Hamilton Walk, Philadelphia, PA 19104-6059, USA
Roche Diagnostics Ltd, Bell Lane, Lewes, East Sussex, BN7 1LG
Tel: 01273 484644 Fax: 01273 480266 Web site: www.roche.com
Roche Diagnostics Corporation, 9115 Hague Road, PO Box 50457, Indianapolis, IN 46256,
USA
Tel: 001 317 845 2358 Fax: 001 317 576 2126 Web site: www.roche.com
Roche Diagnostics GmbH, Sandhoferstrasse 116, 68305 Mannheim, Germany
Tel: 0049 621 759 4747 Fax: 0049 621 759 4002 Web site: www.roche.com
361
LIST OF SUPPLIERS
Tel-Test, Inc., 1511 County Road 129 Po Box 1421, Friendswood, TX 77546, USA
Varian (Head Office), Varian Australia Pty Ltd, 679 Springvale Road, Mulgrave,
Victoria 3170
Tel: (+)61 1300 658 274 Fax: (+)61 1300 658 274 Web site: www.varianinc.com
Vysis Inc., 3100 Woodcreek Drive, Downers Grove, IL 60515, USA
362
Index
363
INDEX
degassing systems 276, 279 dansyl chloride 321 enzyme activators 176-7
flow systems 268 data analysis 30-1, 209-10, enzyme activity 170-1
instrumentation 266-74 227-39, 284, 348-54 andpH 177-8
instrument manufacturers d-d transitions 117-19 and temperature 178-9
271 deactivation 35-8 units 171
lysozyme folding 279-80 dead time 214-17, 267, 275 enzyme assays 200-4
mixing 269-70 decay associated spectra (DAS) enzyme-based assays 195-9
pneumatic drives 268 75 enzyme cooperativity 179-81
protein folding 278-80 degassing 276, 279, 340 enzyme inhibition 174-5
sample properties 267-8, 8-function pulses 73 enzyme kinetics 167-82, 220-7
269-70 deoxygenation 90-1 units 168-9, 171
sensitivity 267 derivative spectra 30-1 enzyme specificity 175-6
stand-alone systems 273-4 desalting 336, 337 eosine labelling 94
stepper motors 268, 269 diethylpyrocarbonate 206 equilibrium binding isotherms
syringe numbers 269 differential scanning 141-65
time constant 274, 275 calorimetry 307, 333-4 equipment suppliers 357-62
wavelength range 270 dilution schemes 342-3 ethoxyformic anhydride 206
wavelength scanning 270 dimerization 4 excitation spectra 54-5, 56
wavelength selection 274 diode array spectrophotometers exocytosis 303
zero-time 275 13-16
circularly polarized light 99 dissymmetry factor 119
CO difference spectroscopy 8 DNA
computer controlled as a catalytic template 280-1 FAD 321
instruments 17, 31, 47, non-self-complementary first-order kinetics 167
270-1 strand mixing 341 first-order rate constant 168,
computer programs, protein stability and pH 340 221-3
structure analysis 125-8 DnaB helicase/sADP flash lamps 42, 85, 89-90, 102,
constant fraction discriminator fluorescence titration 287
88 156-8 flavine adenine dinucleotide
continuous assays 201 DnaB helicase/TNP-ADP (FAD) 321
continuous flow analysers 188 fluorescence titration flow period 213-14, 215
convolution integral 73 146-7 fluorescein derivatives 321
Coomassie Brilliant Blue 193 DNA circular dichroism 129-34 fluorescence 37-8, 69-84
cooperativity of enzymes DNA fluorescence anisotropy fluorescence anisotropy 60-1,
179-81 77 75-8, 324
coumarin derivatives 321 DNA/ligand interactions 134-8 fluorescence assay 193-4
coupled assays 203 double beam fluorescence decay 70
Cro 312 spectrophotometers 9-10 fluorescence life time
curve fitting 348-52 double monochromator distribution 75
cuvettes 25-6 methods 16, 44 fluorescence quantum yield (4>F)
for circular dichroism 103-5, DTNB 204-5 39-40, 55, 57-8, 59-60,
268-9 dual-wavelength 245, 323-5
cleaning 26, 347 spectrophotometers 10-12 fluorescence quenching 62-6
disposable 26 dye-binding assays 193 acrylamide 65-6
for fluorescence spectrometry dyes 193, 289, 292 fluorescence resonance energy
44-5 transfer (FRET) 60
microvolume 343 fluorescence spectra 39, 47-58
mixing contents of 25-6 fluorescence spectrometry,
semi-micro 25 elastic scattering 48-9 steady state 33-67
for UV melting 343 electronic transitions 110-19 applications 58-66
volume determination 25 ellipticity 120-1 computer interfacing 47
cyanine dyes 292 Ellmans reagent 204-5 cuvettes 44-5
cysteine 204-5 endocytosis 303 electronic circuitry 46-7
cytochromes 7-9, 27-8 endosome pH 299-301 excitation spectra 54-5, 56
cytoplasm pH/calcium ion end point assays 187-8 inner filter effect 47-8
concentration 301-3 energy transfer 71-2 instrumentation 41-7
364
INDEX
365
INDEX
366
INDEX
367
INDEX
368