Problems in Particle Size
Problems in Particle Size
KEY POINTS
The following key points are discussed:
• It is important that the results generated by a particle size analysis are thoroughly understood in the
context of their appropriate application.
• The identification and correction of problems encountered in laser diffraction analysis is discussed.
• Examining the sample under a microscope is a critical step in verifying that the results obtained by a
particle size analysis correspond well to the observed particle size.
• Broken particles may occur due to use of excessive ultrasonic energy in liquid sample preparation or
excessive air pressure in dry powder sample preparation.
• Most homogenous solid particle sample types should produce a continuous particle size distribution, so
the occurrence of a distinct disconnected or semi-disconnected peak should be considered suspicious
and thoroughly investigated.
• Distributions with distinctly different modes of occurrence (disconnected peaks) can indicate issues such
as bubble peaks, thermal artifacts, dry dispersion artifacts, opalescent or reflective artifacts, and optical
model artifacts.
• Orthogonal techniques are useful in verifying the accuracy of a particle size distribution.
• When the particle size distribution of a material needs to be measured, steps should be taken to ensure
that the measured data output reflects the test sample rather than an artifact, bubble, or other source of
data contamination.
INTRODUCTION
The results obtained by a particle size analysis can easily be taken for granted. If the particle size
distribution of a material must be quantified and reported, it is important that the full implications of the
analytical technique and generated data are investigated and understood. While complying with designated
testing methods and procedures is of great importance in a quality assurance environment, it can be argued
that the purpose for testing is irrelevant if the results obtained by these methods are inaccurate. Using a
consistent method is necessary for highlighting lot-to-lot variability, but the results only reflect the actual
particle size insofar as they are known to be accurate. Classifying the "true" particle size of a material can
be a difficult task, as the results obtained by an analysis are highly dependent on the technique by which
they are measured. Even so, there are ways that a given particle size result can be verified to ensure greater
confidence in the accuracy of the data.
When reviewing the results of a particle size analysis, there are several common red flags that can lead one
to question the accuracy of a given data output. While many of these problems can apply to a variety of
particle sizing methodologies, this article focuses on those common to laser diffraction (or laser light
scattering), as it represents one of the most widely used techniques. This article aims to briefly discuss a
few of these problems, how to identify them, and how to correct them.
BROKEN PARTICLES
While specific circumstances may necessitate otherwise for certain applications, the goal of a typical
particle size analysis is to measure the particle size distribution of the individual or primary particles in a
given sample. According to the National Institute of Standards and Testing (NIST), a primary particle is
defined to be the smallest identifiable subdivision in a particulate system (1). It is important that the
technique used to measure particle size does not itself result in altering the primary particle size. While this
sounds simple enough, it can at times represent a gray area. The method used to disperse particulate must
be examined closely. Observing the sample under a microscope prior to analysis is a critical step in
verifying that the results obtained by a test correspond well to the expected particle size in a given
dispersion.
Ultrasonic energy can be invaluable in dispersing agglomerated primary particulate in a liquid suspension.
However, applying excessive ultrasonic energy can cause primary particulate to fracture in some cases.
Particles especially susceptible to fracturing can include those which are elongated or acicular, soft or
friable, or those that are platy. It is critical to observe the dispersion under a microscope before and after
ultrasonic energy is applied in order to evaluate the effect of sonication and the progress of the dispersion.
Figure 1 shows the result of typical levels of ultrasonic energy on a particle type that is susceptible to
fracturing. The photo on the left shows a hand-mixed dispersion of particulate up to approximately 250
microns in length. When ultrasonic energy is applied, as in the photo on the right, the particulate is
shattered to less than 100 microns in length. The dispersion quality is otherwise sound. So without
examination prior to the application of ultrasonic energy, it would be easy to unknowingly deliver
significantly undersized data if the particles on the right were to be analyzed.
FIGURE 1. Particles in a liquid dispersion: hand-mixed (left) and with ultrasonic energy applied via
ultrasonic bath (right).
Likewise, the effects of air pressure as a dispersion method for dry powders must be monitored carefully.
Ideally, the pressure chosen must be sufficient to fully disperse primary particulate, while leaving these
individual (primary) particles intact. However, because these processes happen simultaneously, the goal is
to maximize dispersion while minimizing attrition. When a dry dispersion is necessary, it is useful and
recommended to verify the quality of the results using a liquid dispersion technique. A liquid dispersion can
be directly observed under a microscope in its final state of dispersion. When multiple measurements are
taken during the analysis, any instability in the preparation can be monitored. Many laser diffraction
instruments can be outfitted to employ both wet and dry dispersion techniques, and this is ideal for
comparison purposes. If possible, the dry dispersion pressure should be adjusted in order to best match the
wet dispersion results, as the wet dispersion results can be directly verified to be accurate by microscopy.
Figure 2 illustrates a sample under several air pressure levels, each represented by a different colored line.
This is referred to as a “pressure titration” in the industry. Note the overall decrease in particle size with
increase in air pressure. The peak or mode (most frequently occurring size) shifts from about 500µm to
about 300µm, while the lower end tail grows, indicating an increase in fines. This suggests that the
particulate is breaking due to the dispersion pressure. Comparing these results to a liquid dispersion
distribution (which is confirmed by microscopy) is an important step in verifying the appropriate air
pressure for a proper dry dispersion. Figure 3 shows an overlay of a dry dispersion pressure level that
corresponds well to the liquid dispersion distribution. Figure 4 is a photograph of the sample particulate as
observed by microscopy.
FIGURE 2. Typical particle size reduction with increasing air pressure. Red line = low air pressure, green
line = high air pressure.
FIGURE 3. Overlay of a dry dispersion (lowest air pressure from Figure 2, red line) with a liquid dispersion
(green line) of the same sample.
FIGURE 4. Photograph of particles in Figure 2 and Figure 3.
Dispersion techniques that alter the primary particle size of a material (intentionally or otherwise) can be
problematic because they not only affect the results of a single measurement, but they can also consistently
affect the results of repeated measurements. For this reason, it is imperative to verify the quality and
accuracy of a dispersion at the time of method inception and validation. The repeatability and
reproducibility of a method do not necessarily reflect the accuracy of the measurement. Without thoroughly
establishing accuracy, the method risks producing repeatable yet erroneous data.
DISCONNECTED PEAKS
If little is known about the particle size distribution of a material submitted for testing, the assessment and
interpretation of the data can be daunting for those less familiar with the field of particle characterization.
However, there are a few rules of thumb that can be useful when reviewing the data. If the sample
submitted for testing consists of a single species (such as the neat API) from a single batch or lot of material
that has been processed in a uniform manner (no mixing of milled and unmilled portions), it generally can
be assumed that the particle size distribution will be continuous. Naturally occurring materials tend to
conform to a lognormal or skewed Gaussian distribution function. See Wiebull’s paper for a discussion on a
general statistical distribution function that can be applied to particle size distributions, as well as other
examples (2). Depending on the particle shape, size, and collective nature (whether there is a presence of
fused agglomerates and unfused primary particulate), the resulting distribution may be more or less broad
with the possibility of having multiple minor modes. However, if these modes or peaks are disconnected
from the primary distribution, the data should be thoroughly investigated to verify the credibility of the
results.
While laser diffraction is an excellent tool for assessing lot-to-lot variability, the detection technique that
makes it so useful in characterizing particle distribution can produce artificial components in the distribution
if left unchecked. These can be easy to identify with experience. Examples of bubble peaks, thermal
artifacts, dry dispersion artifacts, opalescent/reflective particle artifacts, and optical model artifacts are
discussed in the following sections.
BUBBLE PEAKS
Bubble peaks are one of the easiest problematic peaks to conceptualize. When a solid particle sample type
is submitted for liquid dispersion analysis, it can be generally assumed that any bubbles or gaseous
occurrence in the dispersion are not meant to be taken as part of the intended sample for analysis. Generally
speaking, most particle size instruments and technologies cannot distinguish between species present in a
distribution. Therefore, if signals are collected from both the sample and bubbles, the instrument will report
both indiscriminately. While the operator of the instrument should be trained to identify and minimize a
bubble problem before collecting data, it is useful to understand the basic identification features of such a
problem should the results be taken under review.
In laser light scattering instruments, air bubble peaks can occur at 40-500 microns in water (3), and most
commonly around 100-300 microns in this author’s experience. Depending on the viscosity and surface
tension of the dispersing medium, this size range can vary. Figure 5 is a typical example of a bubble peak
occurring alongside a particle distribution in an aqueous dispersion. The bubble peak occurs at the coarse
end at approximately 200µm in this example. There is a clear demarcation between the particle peak and
the bubble peak. Though the bubble peak does not overlap with the sample particle size distribution peak, it
will clearly influence the cumulative size moments and will render any associated data erroneous.
Even when all necessary precautions are taken, bubble peaks can occasionally appear in liquid dispersion
data. It is important to be able to identify these peaks as distinct from the particulate to be measured, and to
thereby invalidate them in a regulated environment. A simple check under the microscope is often
sufficient to rule out the presence of particulate in the size range of the suspected bubble peak. Note that the
particulate imaged in Figure 6 is conspicuously absent of any material greater than 100µm in size, unlike
what is implied by the distribution in Figure 5. While certain sample types and dispersions, such as those
containing surfactants or highly viscous carriers, might have a greater likelihood of bubble occurrence, in
general, most bubble peaks occur inconsistently, which makes it easy to rule them out as part of the true
particle size distribution.
ARTIFACT PEAKS
Also referred to as "ghost peaks", artifact peaks are well recognized among laser diffraction instrument
operators, however, their technical cause is often poorly understood. Regardless, their occurrence can be
identified and their associated cause can be discussed and should be investigated. Used here, the term
"artifact peak" is meant as a general description to differentiate any outlier peak from the true particle size
distribution that does not correspond to a tangible presence (such as a bubble, an unintended emulsion
droplet, or agglomerated particles). Several common associated causes are discussed below.
Thermal peaks can be avoided in most cases when proper precautions are taken to ensure that the
temperature of a volatile carrier medium has reached thermal equilibrium within the recirculator of an
instrument prior to taking a background measurement. When they do occur, they correlate to a peak at the
coarse end of the distribution. The thermal difference results in “beam steering,” where a larger signal
reaches the inner or low angle detectors in some instruments. Depending on the size of the primary
particulate, the thermal peak is often completely detached from the “true” particle size distribution. It
should be verified by microscopy that primary particulate does not occur within the size range of the
thermal peak in order to render the thermal peak suspect.
FIGURE 9. Photograph of particles in Figure 8. Note the absence of particles in the 500µm-2000µm range
as compared to the distribution in Figure 8.
FIGURE 11. Photograph of particles from Figure 10. Note the absence of particles in the 300µm-1000µm
range as compared to the distribution in Figure 10.
Optical model artifact peaks are associated with the complex refractive index assigned to a particle for laser
diffraction analysis using Mie theory. Mie theory, as opposed to Fraunhofer theory, assumes that the optical
properties of a particle are known, and so they must be entered into the instrument software as the particle’s
complex refractive index. International Organization for Standardization (ISO)13320:2009 defines the
complex refractive index of a particle as consisting of both a real and an imaginary (absorption) component
(4). The particle’s refractive index (real component) can be directly measured using a Becke line test, for
example. The imaginary or absorption component represents the reduction of intensity of a light beam that
is not due to scattering. Particle properties, such as surface roughness, texture, and reflectivity; internal
reflection or refraction; or color and opacity (i.e. actual absorption) can affect the beam intensity, thereby
influencing the absorption value (4). To the author’s knowledge, there is no commercially available method
of directly measuring this absorption component as it specifically applies to laser diffraction; instead, it is
inferred. Approximations can be made through volume concentration experiments in many instruments.
Beekman et. al. provides an excellent discussion on the importance and impact of the imaginary and
absorption component on laser diffraction particle size distribution data (5).
The laser diffraction document from ISO warns that an inappropriate choice of optical model may result in
significant bias of the resulting particle size distribution (4). Figure 12 is an extreme example of an
inappropriate choice of optical model. Generally, when the value selected differs greatly from the actual
optical properties of the particle type, a detached or partially-detached peak may occur in the distribution,
usually at the fine end (typically resulting in a sub-micron mode). While the particle’s real refractive index
value can contribute to this problem, the absorption value often poses a more significant influence.
Simply changing the optical properties used for the particle drastically changes the resulting recalculated
distribution, see Figure 13.
FIGURE 13. Influence of changing absorption values on a particle size distribution.
FIGURE 14. Photograph of particles from Figures 12 and 13. While submicron particulate is difficult to
photograph due to the limits of optical microscopy, it can be observed as a haze or blinking on a live wet
slide when viewed by microscopy. Nevertheless, it is unlikely that the distribution in Figure 12 is a realistic
depiction of the particles in this sample.
As discussed previously, it can be generally assumed that most homogenous solid particle sample types
should produce a continuous particle size distribution, so the occurrence of a distinct disconnected or
partially-disconnected peak should be considered suspicious and thoroughly investigated. However,
scientifically verifying the presence and relative volume of submicron particulate can be difficult as this
pushes the lower limits of optical microscopy. As laser diffraction instruments provide data on a volume-
basis, it is important to keep in mind that any fine particulate observed must be present in sufficient quantity
to command a presence as a secondary peak alongside coarser particulate in the distribution. It is generally
taught that submicron particulate of this concentration would exist in such quantities so as to produce a
visible "haze" when observed under a darkfield microscope. While this haze is a good indication of the
presence of submicron particulate, the human eye is not capable of making the final call as to the relative
volumetric proportions of the observed particulate. It is imperative that orthogonal testing techniques be
utilized to either support or rule out the suspicious peak wherever possible. As a compliment to laser
diffraction, such techniques include (but are not limited to): image analysis (microscopy), laser light
obscuration, and electro-sensing zone, where possible and applicable.
While these techniques are valuable methods of distribution verification, their lower detection limits may
decrease their ability to verify suspect peaks of entirely submicron orientation. When this is the case,
indicators of fit (such as residual or weighted residual) can be useful in optimizing the optical model of best
match to the measured data (6). Laser diffraction instrument software commonly includes an indicator of
this nature and this can be very useful in assessing the quality and legitimacy of the data. It is important to
note that the use of the residual can be used as one indicator when investigating an optical model, but is not
a definitive form of verification.
CONCLUSIONS
When the particle size distribution of a material needs to be measured, steps should be taken to ensure that
the measured data reflects the test sample rather than a bubble, artifact, or other source of data
contamination. It is imperative that the physical nature of the particles is thoroughly investigated at the
onset of method development and creation. Knowledge of the expected particle size is necessary to rule out
any aberrations in the resulting data output. Optical microscopy and other orthogonal techniques are
valuable tools for assessing and verifying the results of a measurement. Steps can be taken to prevent or
minimize the occurrence of many common problems that manifest as erroneous peaks in the particle size
distributions of laser light scattering instruments. Many of these problems are sporadic and otherwise
inconsistent in occurrence, which can make them easier to identify with thorough testing. Problems that
have the potential to cause inaccurate yet repeatable and reproducible results must be considered and
managed with utmost diligence from the time of method inception. Thorough scientific scrutiny of
suspicious results is necessary in order to either accept or reject measured data or a method. While laser
diffraction instrumentation can be very useful for quality control, it is important that the results obtained by
a given method are well understood and, ideally, are an accurate depiction of the intended particulate to be
characterized.
REFERENCES
1) Hackley, Vincent A., and Chiara F. Ferraris, The Use of Nomenclature in Dispersion Science and
Technology, Washington: US Government Printing Office, Special Publication 960-3, 2001.
2) Weibull, Waloddi. "A Statistical Distribution Function of Wide Applicability." Journal of Applied
Mechanics (1951): 293-97.
3) Xu, Renliang. Particle Characterization: Light Scattering Methods. Dordrecht [u.a.: Kluwer, 2001.
4) ISO, “Particle Size Analysis - Laser Diffraction Methods,” International Standards Organization ISO
13320:2009
5) Beekman, Alice, Daxian Shan, Alana Ali, Weiguo Dai, Stephen Ward-Smith, and Merrill Goldenberg.
"Micrometer-Scale Particle Sizing by Laser Diffraction: Critical Impact of the Imaginary Component of
Refractive Index." Pharmaceutical Research 22.4 (2005): 518-22.
6) Mastersizer 2000 User Manual, MAN0384, Issue 1.0, March 2007, Malvern Instruments Ltd., pg 4-20.
ACKNOWLEDGEMENTS
Many thanks to Rebecca Wolfrom, Bill Kopesky, and Eric Olson for their assistance with editing and
technical input.