We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF or read online on Scribd
You are on page 1/ 12
1.0 EXPERIMENT ON DETERMINATION OF pH
SI. No.
Contents
Preamble
44
Aim
1.2
Introduction
Environmental Significance
13
Principle
14
Materials Required
1.41
Apparatus Required
1.4.2
Chemicals Required
15
Sample Handling and Preservation
1.51
Precautions
1.6
Procedure
1.6.1
Preparation of Reagents
1.6.2
Calibrating the Instrument
Testing of Water Sample
17
Calculation
174
Table
18
17.2
Data Sheet
Interpretation of Results
19
Inference
1.10
Evaluation1.0 EXPERIMENT ON DETERMINATION OF pH
PREAMBLE:
“How to determine pH in Water and Wastewater’.
Test procedure is in accordance to IS: 3025 (Part 11) - Reatfirmed 2002
In addition to our Indian Standard, we also discuss in brief regarding the procedure
stated in
(1) APHA Standard Methods for the Examination of Water and Wastewater - 20"
Edition. Method 4500-H’ B.
(2) Methods for Chemical Analysis of Water and Wastes, EPA-600/4-79-020, USEPA,
Method 150.1
4.1 AIM
To determine the PH of the given water sample with the stipulations as per IS: 3025
(Part 11) - Reaffirmed 2002
4.2 INTRODUCTION
The term pH refers to the measure of hydrogen ion concentration in a solution and
defined as the negative log of H" ions concentration in water and wastewater. The
values of pH 0 to a little less than 7 are termed as acidic and the values of pH a little
above 7 to 14 are termed as basic. When the concentration of H* and OH” ions are
equal then it is termed as neutral pH
1.2.1 ENVIRONMENTAL SIGNIFICANCE
Determination of pH is one of the important objectives in biological treatment of the
wastewater. In anaerobic treatment, if the pH goes below 5 due to excess
accumulation of acids, the process is severely affected. Shifting of pH beyond 5 to 10
upsets the aerobic treatment of the wastewater. In these circumstances, the pH is
generally adjusted by addition of suitable acid or alkali to optimize the treatment of the
wastewater. pH value or range is of immense importance for any chemical reaction. A
chemical shall be highly effective at a particular pH. Chemical coagulation,
disinfection, water softening and corrosion control are governed by pH adjustment
Dewatering of sludges, oxidation of cyanides and reduction of hexavalent chromium
into trivalent chromium also need a favorable pH range. It is used in the calculation of
carbonate, bicarbonate, CO2 corrosion, stability index and acid base equilibrium
Lower value of pH below 4 will produce sour taste and higher value above 8.5 a bitter
taste. Higher values of pH hasten the scale formation in water heating apparatus andalso reduce the germicidal potential of chlorine. High pH induces the formation of
trinalomethanes, which are causing cancer in human beings.
4.3. PRINCIPLE
The pH electrode used in the pH measurement is a combined glass electrode. It
consists of sensing half cell and reference half cell, together form an electrode system.
The sensing half cell is a thin pH sensitive semi permeable membrane, separating two
solutions, viz., the outer solution, the sample to be analyzed and the internal solution,
enclosed inside the glass membrane and has a known pH value. An electrical
potential is developed inside and another electrical potential is developed outside, the
difference in the potential is measured and is given as the pH of the sample.
1.4 MATERIALS REQUIRED
1.41 APPARATUS REQUIRED
pH meter
Standard flasks
Magnetic Stirrer
eee]
Funnel
Beaker
Wash Bottle
Tissue Paper
PN OARON A
Forceps
1.4.2 CHEMICALS REQUIRED
1. Buffers Solutions of pH 4.01, 7.0 and 9.2
2. Potassium Chloride
3. Distilled Water
Creer te)
ZNPROCEDURE CHART
‘Switch on the pH meter
(Atieast 30 min before the test) Prepare the buffer
solution 4.0, 7.0 and 9.2
Calibrate the pH meter
to 7.0 using the buffer
‘and by adjusting the
‘calibration knob
® Calbrote the pH meter
09.2 using the buffer
— and by adjusting the
calibration knob
Calibrate the pH meter
to 4.0 using the buffer
‘and by adjusting the
calibration knob
Read the pH meter
by inserting
the sample4.5 SAMPLE HANDLING AND PRESERVATION
Preservation of sample is not practical. Because biological activity will continue after a
sample has been taken, changes may occur during handling and storage.
The characteristics of the water sample may change.
To reduce the change in samples taken for the determination of pH, Keep samples at
4°C. Do not allow the samples to freeze.
Analysis should begin as soon as possible.
1.5.1 PRECAUTIONS
The following precautions should be observed while performing the experiment:
v.
Temperature affects the measurement of pH at two points. The first is caused
by the change in electrode output at different temperatures. This interference
can be controlled by the instruments having temperature compensation or by
calibrating the electrode-instrument system at the temperature of the samples.
The second is the change of pH inherent in the sample at different
temperatures. This type of error is sample dependent and cannot be controlled;
hence both the pH and temperature at the time of analysis should be noted.
In general, the glass electrode, is not subject to solution interferences like color,
high salinity, colloidal matter, oxidants, turbidity or reductants.
Oil and grease, if present in the electrode layer, should be removed by gentle
wiping or detergent washing, followed by rinsing with distilled water, because it
could impair the electrode response
Before using, allow the electrode to stand in dilute hydrochloric acid solution
for at least 2 hours,
Electrodes used in the pH meter are highly fragile, hence handle it carefully.
41.6 PROCEDURE
Three major steps are involved in the experiment. They are
1. Preparation of Reagents
2. Calibrating the Instrument
3. Testing of Sample1.6.1 PREPARATION OF REAGENTS
1. Buffer Solution of pH 4.0
* Take 100 mL standard measuring flask and place a funnel over it.
* Using the forceps carefully transfer one buffer tablet of pH 4.0 to the funnel
* Add little amount of distilled water, crush the tablet and dissolved it
+ Make up the volume to 100 mL using distilled water.
2. Buffer Solution of pH 7.0
* Take 100 mL standard measuring flask and place a funnel over it.
* Using the forceps carefully transfer one buffer tablet of pH 7.0 to the funnel
* Add little amount of distilled water, crush the tablet and dissolved it
* Make up the volume to 100 mL using distilled water.
3. Buffer Solution of pH 9.2
+ Take 100 mL standard measuring flask and place a funnel over it.
* Using the forceps carefully transfer one Buffer tablet of pH 9.2 to the funnel
* Add little amount of distilled water, crush the tablet and dissolved it
+ Make up the volume to 100 mL using distilled water.
4.6.2 CALIBRATING THE INSTRUMENT.
Using the buffer solutions calibrate the instrument.
Step 1
In a 100 mL beaker take pH 9.2 buffer solution and place it in a magnetic
stirrer, insert the teflon coated stirring bar and stir well.
Now place the electrode in the beaker containing the stirred buffer and check
for the reading in the pH meter.
If the instrument is not showing pH value of 9.2, using the calibration knob
adjust the reading to 9.2
Take the electrode from the butfer, wash it with distilled water and then wipe
gently with soft tissue
Step 2
In a 100 mL beaker take pH 7.0 buffer solution and place it in a magnetic
stirrer, insert the teflon coated stirring bar and stir well.
Now place the electrode in the beaker containing the stirred buffer and check
for the reading in the pH meter.
If the instrument is not showing pH value of 7.0, using the calibration knob
adjust the reading to 7.0.Take the electrode from the buffer, wash it with distilled water and then wipe
gently with soft tissue
Step 3
In a 100 mL beaker take pH 4.0 buffer solution and place it in a magnetic
stirrer, insert the teflon coated stirring bar and stir well.
Now place the electrode in the beaker containing the stirred buffer and check
for the reading in the pH meter.
If the instrument is not showing pH value of 4.0, using the calibration knob
adjust the reading to 4.0.
Take the electrode from the buffer, wash it with distilled water and then wipe
gently with soft tissue.
Now the instrument is calibrated.
1.6.3 TESTING OF SAMPLE
+ Inaclean dry 100 mL beaker take the water sample and place it ina
magnetic stirrer, insert the teflon coated stirring bar and stir well
+ Now place the electrode in the beaker containing the water sample and
check for the reading in the pH meter. Wait until you get a stable reading
* The pH of the given water sample is 8.84
+ Take the electrode from the water sample, wash it with distilled water and
then wipe gently with soft tissue.
1.7 CALCULATION
To determine the value of pH of the given water sample the readings obtained
are required to be tabulated
4.74 TABLE
‘Sample | Temperature of
No ‘Sample (°C) pHFor sample 1 the temperature of the measurement is 27° C and as obtained the value
of the pH is 7.84.
For sample 2 the temperature of the measurement is 27° C and as obtained the value
of the pH is 7.43.
For sample 3 the temperature of the measurement is 27° C and as obtained the value
of the pH is 8.84.1.7.2 DATA SHEET
DETERMINATION OF pH
DATA SHEET
Date Tested August 30, 2010
Tested By CEM Class, Group &
Project Name CEM, NITTTR Lab
Sample Number = BH, BH, Bite
Sample Location BHI : Perunguai (Lat 12’ 5%” s1.74 § Long £014" 2.82)
Sample Description: Sterface water
‘Sample Location BH2 : Pallikarmai (Lat 12’ 56” 21.02 § Long 80'12" 47.99)
Sample Description: Ground water
Sample Location BH2 : Thiruvanmiyur (Lat 12’ 59° 0.84 § Long £0’ 16" 8.21)
Sample Description: Sea water
TABULATION
Sample | Temperature
of Sample pH
Ne
° co,
1 27 7.84
2 2 7431.8 INTERPRETATION OF RESULTS
The pH of the given water sample is 8.84.
1.9 INFERENCE
pH is @ measure of the hydrogen ion concentration in water. Values lower than 7
indicate acidity and values higher than 7 indicate alkalinity. Drinking water with a pH
between 6.5 and 8.5 is generally considered satisfactory. Acidic waters tend to be
corrosive to plumbing and faucets, particularly if the pH is below 6. Alkaline waters
are less corrosive. Waters with a pH above 8.5 may tend to have a bitter taste.
The pH of the water samples are well within the limit of the drinking water standards.
The pH of the ground water is slightly towards the alkaline side because of some soil
and rocks chemicals might have dissolved in it. In case of the pH of the fresh water,
aquatic plants uses up hydrogen molecules for photosynthesis, which causes the
concentration of hydrogen ions to decrease and therefore the pH is towards the
alkaline side. The sea water is mostly alkaline in nature because of the presence of
different type of salts
4.10 EVALUATION
1. pHis defined as,
a) Logarithm of Hydrogen ions concentration
b) Negative logarithm of Hydrogen ions concentration
c) Hydrogen ion concentration
d) OH ion concentration
2. pH of neutral water is,
a) less than7
b) more than 7
c) 7.0
d) 00
3. The acceptable value of pH of potable water is
a) 7.0t085
b) 6.51095
) 6108.5
d) 65to104. The inner solution present in the glass electrode of pH meter is.
a) HCl
b) KCI
c) NaCl
d) MgCl
5. The buffer solution can be stored for a minimum period at room temperature
a) True
b) False
6. Possible reasons for a relatively low pH value in a river water sample is due to
a) Organic material decomposition to form acidic substances
b) Running long distances
c) Presence of fishes
d) Presence of aquatic plants
7. Possible reasons for a relatively high pH value in a river water sample is due to
a) Running over clay
b) Running long distances
c) Running of fishes,
d) Presence of aquatic plants
8. Aweak acid is one that ionize incompletely in aqueous solution
a) True
b) False
9. A strong base is one that ionizes incompletely in aqueous solution.
a) True
b) False
10. The measurement of pH made by determining the e.m.f of the.
a) cell constant
b) solution
¢) electrode cell
d) calomel electrodeKEY TO ITEMS:
3 32
7 coo
aocatevéd
Sa ae FoGooOR BS