Syringe Infusion Pump S300: Technical Specification
Syringe Infusion Pump S300: Technical Specification
2 - 60 ml
All major brands available
Other syringes may be calibrated according
to customer needs
FLOW RATE
SYRINGE INFUSION PUMP 0.1 - 99.99 ml/h with 0.01 ml/h increments
100 - 999.9 ml/h with 0.1 ml/h increments
S300 1000 - 2000 ml/h with 1 ml/h increments
INFUSION VOLUME, TIME
Technical specification 0.1 - 20000 ml; 1 min - 200 h
INFUSION ACCURACY
± 2% according to EN 60601-2-24
BOLUS, INITIAL BOLUS
Rate: 20 - 2000 ml/h
Volume: 0.1 - 50 ml
Automatic and manual bolus
OCCLUSION
12 levels (75 ÷ 900mmHg)
UNITS
ng, µg, mg, g, mL, L, µEq, mEq, Eq, mIU, IU,
kIU, mIE, IE, kIE, mmol, mol, cal, kcal, J, kJ
Automatic syringe detection and installation / --, kg, m² / min, h, 24h
3.2” high contrast and resolution colour display
INFUSION MODES
Touchscreen
Continuous, Intermittent, Profile, TPN,
Mobile phone like alphanumeric keyboard
PCA/PCEA*, TCI/TIVA*
Safe and easy interface * Q2 2018
Fast infusion start
ALARMS
Built-in power supply, handle, rotatable pole clamp
3 levels of alarms priority, reminders,
Intelligent occlusion avoidance algorithm warnings
Bolus supply support system
HISTORY INFUSION
Titration with soft or hard limits
Min. 2000 records of typical infusions
Highly advanced alarm system
2 infusion security levels protected by passwords POWER SUPPLY
Day and night modes 100 ÷ 240 VAC; 12.4 ÷ 15.5 VDC, max 1 A
Automatic pump firmware, configuration and drug library BATTERY
update without infusion interruption Ni-MH. Work time: 30 h @ 5 ml/h
HL7 protocol connectivity to PDMS/HIS Full recharge time: ≤ 5 h
DIMENSIONS, WEIGHT
Drug Library 364 x 115 x 165 (W x H x D), ≤ 2.3 kg
CLASSIFICATION
Safe and ready to use list of drug dosing procedures Class II, CF type, Defibrillation proof, IP22
CCA drug dosing procedures SAFETY STANDARDS
CCA pump configuration EN 60601–1, EN 60601–1–2, EN 60601–1-8,
Up to 40 CCAs, up to 40 drug categories, up to 5000 EN 60601-2-24, EN 1789, MDD93/42/EEC–
dosing procedures IIb
Up to 10 fixed concentrations and 1 user definable for
every drug
Default values for selected parameters
Soft and hard limits of several infusion parameters Medima Sp. Z o.o.
2 levels of advisory notes Al. Jerozolimskie 200, 02-486 Warsaw, Poland
Phone: +48 22 313 22 66, fax. +48 22 313 22 69
1011
Upload of a drug library without infusion interruption e-mail: [email protected]
www.medima.pl
DH-S300-01-EN-02_2018.03.23-Data_sheet