Infant Flow Sipap Comprehensive Quick Guide
Infant Flow Sipap Comprehensive Quick Guide
SiPAP comprehensive
quick guide
Table of contents
i
Section 1: Circuit setup
WARNING:
14 50 60
70
5
40 80
12
30 90
4
10
8 21 100
3
6
4
2
2
1
• Follow manufacturer’s
instructions for setup and
operation of humidifier.
OPTIONAL:
1
Open bed or crib Isolette
2
Abdominal sensor
5. Verify correct placement. Transducer LED illuminates on expiration, and front panel
LED on inspiration.
OPTIONAL:
3
Pressure Nomogram for Infant Flow System
The Infant Flow SiPAP system is subject to a direct relationship between the controlled gas flow
and airway pressure. For example, 8 to 9 LPM gas flow provides approximately 5 cmH2O.
15.0
13.0
11.0
9.0
(cmH2O)
7.0
5.0
3.0
1.0
4
Alarm test
WARNING:
Alarm test initial settings
Prior to patient application, ensure that all User Air supply > 30 psig (2.1 bar)
Verification testing and calibration procedures are O2 supply > 30 psig (2.1 bar)
successfully completed. User Verification testing and Infant Flow LP generator/circuit
calibration procedures must be done off the patient. nCPAP/press low 9 LPM
Occlude the opening to the patient. Use settings below for step 9
Rate 30 bpm
2. Power up the driver and allow Power On
Check to complete. Ti 0.3 sec
5
Alarm test (continued)
4. Adjust the nCPAP/pres low flow meter to 11 LPM. The high airway pressure alarm activates.
Decrease flow to 8 LPM. Reset the alarms.
5. Adjust the %O2 control to 35%. The High %O2 alarm activates. Return the O2 setting to
30%. Reset alarms.
6. Adjust the %O2 to 25%. The Low %O2 alarm activates. Return the O2 setting to 30%. Reset
the alarms.
7. Disconnect the AC power cord from the wall outlet. The Loss AC alarm activates. Reconnect
the AC power cord. Reset alarms.
8. Occlude exhalation line and increase nCPAP pressure to 11 cmH2O. The High Circuit pressure
alarm activates. Decrease flow to 8 LPM and reset alarms.
9. Select and confirm BiPhasic mode. Change the mandatory rate control setting to 1. Low
Breath rate alarm activates after the default interval of 20 seconds. Return to a rate of 30.
Reset the alarms.
6
Two point oxygen calibration
7
Two point oxygen calibration (continued)
NOTE:
8
Disable oxygen sensor
Disable O2 button
WARNING
9
Startup menu and leak test
10
Startup menu and leak test (continued)
11
Section 2: Operation
Infant Flow SiPAP front panel A B C
A. Power LED
XDCR
E. %O2 control
Infant Flow®
SiPAP
F. Press High flow meter
NCPAP/
Press Low Press High
50 60 70
12
Soft key operation
Description Example
A button that is enabled.
A selected mode or control pending confirmation that is visually highlighted and intermittently
flashes between yellow and white text.
While a button is pressed, the edges are highlighted to provide a pressed appearance.
13
Soft key operation (continued)
Description Example
When there is an active alarm associated with a measured value, the measured value concerned
is displayed with RED FLASHING text. The associated limit value (if any) is displayed in RED.
When an alarm associated with a measured value is resolved, the device remains in a LOW priority
alarm state, with the measured value displayed in YELLOW FLASHING text and the associated limit
displayed in YELLOW, until the alarms are cleared by the operator.
Manual Breath delivers a single BiPhasic cycle at current settings for T-High, Press High and %O2.
Only one BiPhasic cycle is delivered regardless of button press duration.
If no screen interactions occur for a period of 120 seconds and there are no active alarms, the
screen goes to a “locked” state to prevent inadvertent entries. To lock the screen, press the screen
lock button.
14
Operational modes
15
Operational modes (continued)
16
Alarm set/confirm screen
17
Mode Select screen
NOTE:
• Only the available modes will be
displayed on the menu bar.
• For Apnea and Biphasic trigger (tr)*
modes, attach the transducer assembly
and respiratory abdominal sensor.
18
Parameter Adjust screen
NOTE:
• Parameter changes can be made during initial
setup and normal operation.
19
Incompatible settings
• When one parameter (A) change is incompatible with another parameter (B), the software
will automatically make an adjustment to that parameter (B).
• If the adjusted parameter (A) is restored in less than 15 seconds, the parameter (B) change
is reversed.
• When parameter adjustments cause a reduction in
another parameter to maintain requirements for
minimum breath interval, the reduced parameter
is displayed in red for 15 seconds.
EXAMPLE:
• BiPhasic mode with T-High = 2.0 seconds: As rate is increased above 28 breaths per minute,
the constraint on minimum T-Low requires a reduction in T-High. If rate is increased to 29,
then T-High will automatically reduce to 1.9 seconds. If rate is reduced back to 28 within
15 seconds, the previous T-High setting (2.0 sec) is restored.
NOTE:
• T-High automatically reduces to maintain a minimum T-Low of 100 milliseconds.
20
Main screen
21
Monitored Parameter screen
22
Alarm management
Resetting alarms
Press the Alarm Reset button for 3 seconds
to clear resolved and low priority alarms and
to reset alarm limits.
23
Final check and routine inspection
Inspect the system at least every 3 to 4 hours to:
• Ensure the generator is stable, secure and not pulling upward on the nose.
• Ensure the patient’s eyes are clearly visible and the nares are not blocked, when using masks.
• Inspect the fixation device and straps for proper tension and adjust as needed to maintain
a proper fit.
24
Troubleshooting
%O2 > 104% High O2 CAL required Reset FiO2 below limit, O2 CAL
High %O2 High %O2 setting changed, supply gas Correct FiO2, reset alarm limits
failure, water trap overflow
Low %O2 High Blender setting changed, supply gas Correct FiO2, reset alarm limits
failure, water trap overflow
Over pressure High Flow rate set too high, occlusion of Check exhaust tube/filter, reduce flow
exhalation limb, blocked silencer/filter rate, reset alarm limits
Low battery voltage Medium Battery disconnect, fails to hold charge Connect to AC
High nCPAP/ High Setting change, circuit disconnect Reset alarm limits
press low
Low nCPAP/ High Setting change, circuit disconnect, leak Reset alarm limits, check for leaks
press low
High BiPhasic pressure High Press high setting change, Reset alarm limits
circuit disconnect
25
WARNING—U.S. Federal Law restricts this device to sale by or on the order of a physician.
CareFusion
22745 Savi Ranch Parkway
Yorba Linda, CA 92887
800.231.2466 toll-free
714.283.2228 tel
714.283.8493 fax
carefusion.com
© 2015 CareFusion Corporation or one of its subsidiaries. All rights reserved. AirLife, Infant Flow, CareFusion and the CareFusion logo
are trademarks or registered trademarks of CareFusion Corporation or one of its subsidiaries. RC4684 (0215/500) L3416 Rev. C