Venti Motion 2: Greater Comfort and Stability in Non-Invasive Ventilation
Venti Motion 2: Greater Comfort and Stability in Non-Invasive Ventilation
VENTI-O2 plus
VENTImotion 2 Display
Accessories
1 VENTIclick – WM 24365
Adaptable humidifier without extra
power supply or extra hoses
2 VENTIpower – WM 27630
Optional external battery guarantees several hours of
1 2 device operation.
4 WEINMANNsupport – WM 93305
PC software with converter cable USB-RS 485
(WM 93321) for settings and analyses
3650.3-01-EN-0911-3 © Copyright Weinmann, Hamburg. Duplication of any kind only with the express permission of Weinmann.
Air pressure range: 600 – 1100 hPa Accuracy: ±1%
(for operation at altitudes up to 4000 m ) automatic altitude adjustment
Trigger level: can be set in 6 levels, separate for inspiration
Electrical connections: 115 – 230 V AC, 50 – 60 Hz and expiration, the trigger for expiration can be
Tolerance -20% +10% disabled in ST mode t
Power consumption at 230 V 115 V
Pressure increase/Pressure
Operation: 0.15 A 0.3 A decrease speed: can be set in 6 levels
Standby: 0.050 A 0.108 A
Accuracy
Classification as per EN 60601-1 volume measurement: at 23 °C: ±15%
Type of protection from elec. shock: Protection class II
Level of protection from elec. shock Type BF Flow at max. rotation speed
at 0 hPa: 285 l/min ±15 l/min
Electromagnetic compatibility as per EN 60601-1-2 Flow at max. rotation speed with
Radio interference suppression: EN 55011 bacteria filter at 0 hPa: 270 l/min ±15 l/min
Immunity to interference: EN 61000-3-2, EN 61000-3-3,
Heating of respiratory air
EN 61000-4-2 bis 6, EN 61000-4-8,
as per HMV: 2.5 °C
EN 61000-4-11
Pressure constancy measured at 20 hPa: ∆p ≤ 1 hPa
Mean sound pressure level/ at 20 hPa about 32 dB (A) as per DIN EN ISO 17510 at 14 hPa: ∆p ≤ 1 hPa
operation as per EN ISO 17510 at 15 hPa about 30 dB (A) in CPAP mode: at 10 hPa: ∆p ≤ 1 hPa
All rights to design and specification modifications reserved·
at a distance of 1 m from device at 12 hPa about 28 dB (A) at 7 hPa: ∆p ≤ 0.5 hPa
in patient position: at 10 hPa about 26 dB (A)
at 7 hPa about 24 dB (A) Fine filter degree of separation
up to 2 µm: ≥ 99.7%
Sound pressure level, alarm: about 62 dB (A)
Fine filter service life: 1000 hours in normal ambient air
IPAP pressure range: 6 to 30 hPa
EPAP pressure range: 4 to 20 hPa Allowable humidity for ≤ 95% relative humidity
CPAP pressure range: 4 to 20 hPa storage and operation: (no condensation)
Pressure accuracy: ±0.6 hPa
Increment: 0.2 hPa (1 hPa ≈ 1 cm H2O) System resistance at air flow of
60 l/min at patient connection opening
Minimal stable limit pressure VENTImotion 2 kPa · s
0.20
(PLSmin) (min. presssure in event of fault): ≥ 0 hPa with hose system WM 24130 and Silentflow WM 23600: l
Maximal stable limit pressure VENTImotion 2 with O2 hose system WM 23737, kPa · s
(PLSmax) (max. pressure in event of fault): ≤ 60 hPa 0.31
VENTIclick WM 24365 and bacteria filter WM 24148: l