Ion Chromatography Procedure: Last Updated: 5.29.13
Ion Chromatography Procedure: Last Updated: 5.29.13
A. Introduction
Ion chromatography (IC) is used to separate out various ions based on their charge. This
technique is especially helpful for measuring the concentration of a particular ion in a water
sample. A small volume of the sample (mobile phase) is pumped through a column of packed
particles (stationary phase) and the time it takes for the ion to pass through is recorded—this is
called the retention time. The retention time of the ion depends on how it interacts with the
column both physically and chemically. Known standard concentrations are used to estimate
sample concentrations.
In the Soil and Water lab, our IC machine is made to analyze the concentration of negative ions
(anions). Some of the common analyzed ions are nitrate, sulfate, and phosphate.
B. Set Up
Lab Equipment/Supplies:
o IC machine
o Plunger for vial caps
o Cassettes (vial holders)
o Sharpie marker
o Jump-drive
o Various pipettors and tips
o 1000ppm stock solutions
o Glassware (beakers, 100mL flasks, etc.)
o Extra sterile containers for holding liquid
o Vials (0.5mL or 5 mL), with black caps
o Gloves
o Deionized, degassed water
Note: Additional lab equipment is necessary if 1000 ppm stock solutions have not been made.
C. Preparation Overview
Note: Please reserve machine a few days in advance by writing your last name, number, and
the type of samples you will be analyzing. This helps the lab keep track of how many total
samples have been analyzed if a new machine part is needed.
Appropriate amounts of the desired salt must be calculated, weighed, and added to 100 mL
flask. Deionized, degassed water can then be added to the 100 mL line on the flask. The
mixture should be shaken vigorously until all the salt is dissolved.
1. Throughout this procedure, be sure to wear gloves. Sweat on the hands can offset
2. Use deionized, degassed water for blanks and for each dilution. Each IC Run should
begin with 4 blanks and end with 3 blanks. This will ensure that the column is thoroughly
cleaned for the next user. Blanks are made from 100% deionized, degassed water.
3. For most runs, the following calibration standard concentrations should be adequate (S1
to S5). However, this can easily be modified depending on the experimenter’s
preferences. Shree suggests keeping the concentrations low. For example 1ppm,
2 ppm, 5 ppm, and 10 ppm. If your results are higher than 100 ppm you should dilute
and rerun the sample. The two runs should be within 5% difference (difference over
average * 100).
4. Calculate how much volume of each 1000 ppm stock solution you need to make 100 mL
of the S5 calibration standards. To solve for the unknown volume, use the following
relationship:
volume1*concentration1=unknown volume2*concentration2
Where volume 1 is the final volume of diluted solution you want (i.e. 100 mL),
concentration is the final concentration of solution you want (i.e. 20 ppm), volume 2 is
how much of the stock solution you need (this is unknown), and concentration 2 is the
original concentration of the stock solution (i.e. 1000 ppm).
5. Prepare five 100mL glass flasks by washing them thoroughly with deionized water and
letting them dry. Using a sharpie, label them with the concentration of each anion, the
date, and your initials. If there is already writing on the flask, it can be cleaned off with
some rubbing alcohol.
6. Clean the pipettor tips and beakers with deionized water before use.
Note: for best precision, the scale can be used to measure the accuracy of the pipettors.
Graduated cylinders are not accurate. It is often difficult to tell which pipettors in the lab
are working well.
8. Pour about half of S5 into a clean spare beaker or container. Using the pipettor, draw 50
mL of S5 out and empty into the S4 flask. Fill the S4 flask up to the 100 mL line with
deionized, degassed water. You now have S4 in the beaker.
9. Repeat this dilution technique for the remaining three standards (S3, S2, and S1).
10. It is useful to have single standards for SO4-S, NO3-N, and other any other ions of
interest, as well as the triples calibration standards. Concentrations of these single
standards can be equal to the highest concentration of the ion found in the triple
standards so the peak will be very clear. These single standards allow you to identify
the peak order of sulfate and nitrate and also giving you an independent check of
calibration. It is best to take these from a second stock solution source but this is not
absolutely necessary; especially if it is not available.
1. All samples must be filtered through a 0.45 um filter before running through the IC. See
section J for notes on preservation of samples including frozen samples. For best
precision vials and caps should be rinsed thoroughly with deionized water. Be sure to
wear gloves throughout the rinsing and filling process.
2. With a Sharpie marker, label the vials with sample numbers. For example S1 to S5 for
the standards, BL for each of 7 blanks, S4r for a replicate of one standard. Individual
standards can be labeled with the ion and the concentration, for example, S40 for
40 ppm Sulfate. If doing dilutions, mark more vials for them. For example, ‘03x5’ for a
five-fold dilution of sample ‘03’ (see Part I.).
Note: A five-fold dilution is one part sample and four parts deionized water.
3. Fill vials at least halfway and cap (~0.6 mL). The cap must fit entirely into the vial when
pressed down so that is flesh with the top of the vial. Caps should be placed into the vial
with the white filter end touching the fluid inside the vial. Please use the black capper to
put on the vial caps as using a kim wipe can leave fibers in the filter that will eventually
clog the machine.
4. Load vials into cassettes in this order: four blanks, five triple calibration standards, all
samples, individual ion standards, and three blanks. If you know that you have samples
with high concentrations you might want to put blanks between them two reduce the risk
of carryover from one sample to the next. It is recommended to put a known check
standard (mid-range concentration) after every 10 samples so you can calculate the
instrument’s drift factor.
1. Fill the eluent bottle with fresh deionized, degassed water and be sure to keep the tube
submerged as you pour the fresh water into the bottle. If air gets into this tube the whole
system has to be flushed out and refilled. Empty the waste bottle under the machine if it
is more than half full.
3. If machine screen is not on, go to ‘Control Panel’ on the file browser and click on
‘ICS_2000_system_AS40.pan’.
4. Warm up the machine for a half hour (click on the ‘Start up’ button). If the pump
pressure goes up and down there may be bubbles in the line. Turn off the pump, open
the door to the IC, and turn waste pump valve (bottom left) just a little until it’s open. Let
the pump run until all the bubble are gone, turn the pump off, tighten the valve, and then
try to turn the pump on again. The pump, eluent, and data acquisition should all be
turned on. Watch the conductance display because it should be stabilized to below 0.5
before you can start the run. Warming up the machine also involves priming the pump
to remove air bubbles from the tubing inside the machine. This can be done from the
digital display on the IC machine.
6. Place the cassettes into the auto-sampler with the first one to be tested in front and last
one at the back. The dots on the cassettes go toward the center of auto-sampler.
7. When conductance is stable at 0.5 uS, turn off the acquisition, and start the new
sequence. Pump should be at around 2000 psi if it’s fluctuating frequently there may be
bubbles in the tube. Please let Shree know if this happens.
8. It is okay to tweak some aspects of the sequence on the computer while it is running.
For example, the "nitrate" program can have its standard concentrations adjusted.
However, this really should be done before the sequence has started.
9. Wait overnight...
1. Double click on one of the single standards and record the peak retention time for this
ion. Repeat this for all other ions of interest.
3. Click on the nitrage.pgm program. This is the template that provides concentration
standards, time window center, and time window width. (Is this the “QNT editor” button?)
4. Fill in the concentration standard values for S1 to S5 as well as the retention times for
each ion. Be sure that the windows do not overlap otherwise peaks will not be
distinguished.
6. It is okay to drop one bad standard result from the calibration since there are five
standards and the fit is linear. Note that this may truncate high or low values.
7. After the calibration looks good, check the computed concentrations for all the triple
standards. Do the retention times observed in the single standards match these?
8. Browse results for all samples, looking at the time plots to ensure that peaks are
properly labeled for each ion. One or more peaks can be missing, meaning; the
concentration of that anion is low. Are there any other unlabelled peaks that interfere
with the three? Also see Part I. below if sample concentration amounts are larger than
the highest standard concentration.
9. Click on one of the samples and highlight the row with the ion of interest. Click on the
“Summary” tab and copy the “Amt” column into Excel. Repeat this for all other ions.
11. Be sure to remove the spent vials from the machine before the next user.
I. Further Dilutions
A higher concentration of an anion comes out earlier (left on chart) than a lower concentration of
the same anion. Sulfate and nitrate come out close together in time, which sometimes means
that a high SO4 will overlap with a low NO3. The likelihood of the peaks merging increases as
the column gets older. If there is a single peak left skewed and below the usual range of sulfate,
this can indicate a very high sulfate value masking a low nitrate. Dilution will cut down the high
sulfate placing its peak "later" and giving a separate (subdued) peak for nitrate. Use the first run
results to decide about dilutions in the second and third replicates. However, if the column is
old and you know you have high concentrations of both SO4 and NO3, it might be best to wait
until the IC column has been changed. See below for sample preservation.
Dilutions are highly recommended if a sample’s predicted concentration is larger than the
highest standard concentration. The IC will use the calibration curves to extrapolate the
unknown concentration of the sample, which may or may not be accurate.
Note: A 10:1 (read as “ten to one”) dilution consists of one part sample and nine parts
deionized, degassed water. Multiply the concentration results by 10.
Preserving Samples
Refrigerate samples immediately and analyze them within 48 hours for best results. Freezing
samples can cause ions to precipitate out. This precipitate may be difficult to see but is
detrimental to the IC. Shree recommends not to freeze samples but instead to add
approximately 100 uL of concentrated H2SO4 to each full 50 mL centrifuge tube immediately
after filtering your samples. If you have already frozen your samples, please re-filter the thawed
sample before running them through the IC. Use a 0.45 um filter.