Ultrasonic Flaw Detector
SX
OmniScan SX
Smaller, Lighter … Still an OmniScan
• Cost-efficient, single-group
• Two-axis encoding and data
archiving capacity
• Conventional UT, TOFD, and
16:64PR PA capabilities
• 8.4 in. (21.3 cm) touch screen
with OmniScan interface
• Compact, lightweight design
The Lightest and Most
User-Friendly OmniScan
OmniScan SX
Olympus is proud to introduce the OmniScan® SX, a flaw detector that benefits from more than 20 years of phased array
experience and shares the OmniScan DNA. For improved ease of use, the OmniScan SX features a new streamlined software
interface displayed on an 8.4 in. (21.3 cm) touch screen. A single-group and non-modular instrument, the OmniScan SX is
easy to operate and cost-effective for less demanding applications.
The OmniScan SX comes in two models: the SX PA and SX UT. The SX PA is a 16:64PR phased array unit, which, like the
UT-only SX UT, is equipped with a conventional UT channel for pulse-echo, pitch-catch or TOFD inspection. Compared to the
OmniScan MX2, the SX is 33% lighter and 50% smaller, offering an unprecedented level of portability for an OmniScan.
Setup Acquisition
Inspection setup can be performed in NDT SetupBuilder, The OmniScan SX enables easy configuration of inspection
and imported directly, via SD card or USB key, to the parameters for either manual, one-line, or raster encoded
OmniScan SX. Then, only a few basic operations are scans. The acquisition is displayed in real time through
required in the instrument, such as setting the gate and user-selectable views and offers the ability to store data on a
range, before acquisition can begin. It is also very easy hot-swappable SD card or USB 2.0 device.
to create a setup right in the OmniScan SX, thanks to the
following features: • Intelligent layouts.
• Full-screen mode for better visualization of defects.
• Automatic probe recognition.
• Synchronization and measurements can be processed using
• One-step, preconfigured application Wizard. different gate combinations.
• Weld Overlay and RayTracing simulation.
Data Analysis and Reporting
Calibration • Data, reference, and measurement cursors for defect sizing.
To achieve a code-compliant inspection, the Calibration • Extensive readings database and predefined lists for trigono
Wizard ensures that every focal law in every group is the metry, flaw statistics on axes, volumetric position information,
direct equivalent of a single-channel conventional flaw code-based acceptance criteria, corrosion mapping statistics,
detector. The user is guided step-by-step through the and more.
required calibrations, including Velocity, Wedge Delay, • Views are linked for interactive analysis and automatically
Sensitivity, TCG, DAC, AWS, and encoder calibrations. Now, updated when performing off-line gate repositioning.
TOFD PCS calibration and lateral wave straightening can be • Optimized preconfigured layouts for quick and simple length,
performed automatically. depth, and height sizing of flaws.
Whether you prefer performing data analysis on a computer
or simply wish to maximize the time your OmniScan is at
work in the field, OmniPC or TomoView are the perfect
software companions for your OmniScan.
2
Affordable and Portable
Go a Long Way….
The Omniscan® SX provides Olympus with a new and versatile tool to add to its arsenal of innovative and creative
complete market solutions aimed at simplifying your workflow and improving overall productivity.
Phased Array Weld Inspection Corrosion Mapping and Composite Inspection
The OmniScan PA is at the heart of the manual and Zero-degree inspection just became even more acces-
semiautomated phased array weld inspection solutions sible with the arrival of the OmniScan SX. For corrosion or
developed by Olympus for the oil and gas industry. These composite inspection, Olympus offers field-proven solutions
systems can be used for inspection in compliance with for detection of anomalies or wall loss.
ASME, API, and other code criteria, while offering high-
speed detection capabilities, and facilitating indication
interpretation.
TOFD Weld Inspection Component Inspection
TOFD is an easy and efficient approach for primary detection Using ultrasonic techniques, inspection of components
of weld defects. It is quick, cost-effective and capable of siz- can detect cracks, wall loss, and other damage. With the
ing defects present in the volume of the weld, a problematic capacity for both angle and linear zero-degree beams, the
area for manufacturing defects. OmniScan SX is a very cost-efficient solution for this type of
single-group inspection.
3
OmniScan® SX Specifications*
Housing Ultrasound Specifications (applies to OMNISX-1664PR)
Overall dimensions 267 mm × 208 mm × 94 mm Connectors 1 Phased Array connector: Olympus PA connector
(W x H x D) (10.5 in. × 8.2 in. × 3.7 in.) 2 UT connectors: LEMO 00
Weight 3.4 kg (7.5 lb) with battery Number of focal laws 256
Data Storage Probe recognition Automatic probe recognition
Storage devices SDHC card or most standard USB Pulser/Receiver
storage devices
Aperture 16 elements
Data file size 300 MB
Number of elements 64 elements
I/O Ports
Pulser PA Channels UT Channel
USB ports 2 USB ports, compliant with USB 2.0 specifications
Voltage 40 V, 80 V, and 115 V 95 V, 175 V, and 340 V
Audio alarm Yes
Pulse width Adjustable from 30 ns Adjustable from 30 ns to
Video output Video out (SVGA) to 500 ns; resolution of 1,000 ns; resolution of
I/O Lines 2.5 ns 2.5 ns
Encoder 2-axis encoder line (quadrature, up, down, or clock/ Pulse shape Negative square wave Negative square wave
direction) Output impedance 35 Ω (pulse-echo <30 Ω
Digital input 4 digital TTL inputs, 5 V mode);
Digital output 3 digital outputs TTL, 5 V, 15 mA 30 Ω (pitch- catch
maximum per output mode)
Acquisition on/off switch Yes, through configuration of a digital input Receiver PA Channels UT Channel
Power output line 5 V, 500 mA power output line Gain 0 dB to 80 dB, 0 dB to 120 dB maximum
(short-circuit protected) maximum input signal input signal 34.5 Vp-p
550 mVp-p (full-screen (full-screen height)
Pace input 5 V TTL pace input height)
Display Input impedance 60 Ω (pulse-echo 60 Ω (pulse-echo mode);
Display size 21.3 cm (8.4 in.) (diagonal) mode); 50 Ω (pulse-receive mode)
Resolution 800 pixels x 600 pixels 150 Ω (pitch- catch
mode)
Brightness 600 cd/m2
System bandwidth 0.6 MHz to 18 MHz 0.25 MHz to 28 MHz
Viewing angles Horizontal: –80° to 80° Vertical: –60° to 80° (–3 dB) (–3 dB)
Number of colors 16 million Beamforming
Type TFT LCD Scan type Sectorial or linear
Power Supply Group quantity 1
Battery type Smart Li-ion battery Data Acquisition PA Channels UT Channel
Number of batteries 1 Digitizing frequency 100 MHz 100 MHz
Battery life Minimum 6 hours under normal operating Maximum pulsing rate Up to 6 kHz (C-scan)
conditions
Data Processing PA Channels UT Channel
Environmental Specifications
Number of data points Up to 8,192
Operating temperature -10 °C to 45 °C
range (14 ºF to 113 ºF) Real-time averaging PA: 2, 4, 8, 16 UT: 2, 4, 8, 16, 32, 64
Storage temperature range –20 °C to 60 °C (–4 ºF to 140 ºF) with battery Rectifier RF, full wave, half wave +, half wave –
–20 °C to 70 °C (–4 ºF to 158 ºF) without battery Filtering 3 low-pass, 3 band-pass, 3 low-pass, 6 band-pass,
Relative humidity Max. 70% RH at 45°C noncondensing and 5 high-pass filters and 3 high-pass filters
(8 low-pass filters when
Ingress protection rating Designed to meet requirements of IP66 configured in TOFD)
Shockproof rating Drop-tested according to MIL-STD-810G 516.6 Video filtering Smoothing (adjusted to probe frequency range)
Data Visualization
A-scan refresh rate A-scan: 60 Hz; S-scan: 60 Hz
Data Synchronization
On internal clock 1 Hz to 6 kHz
On encoder On 2 axes: from 1 to 65,536 steps
Programmable Time-Corrected Gain (TCG)
Number of points 16: One TCG (time-corrected gain) curve per focal law
Maximum slope 40 dB/10 ns
Alarms
Number of alarms 3
OmniScan MX2 OmniScan SX
If multigroup inspections (ex: two PA probes or combined PA + UT) are Conditions Any logical combination of gates
required or anticipated, Olympus recommends the OmniScan MX2. This
advanced flaw detector's modular platform facilitates the upgrade path — The OmniScan SX meets or exceeds the minimum instrumentation and software
you can start with the module in your price/performance range and upgrade requirements as specified in ASME, AWS, API, and EN codes.
later to one of the many other modules available.
is certified to ISO 9001, ISO 14001, and OHSAS 18001.
*All specifications are subject to change without notice.
www.olympus-ims.com
All brands are trademarks or registered trademarks of their respective owners and third party entities.
Copyright © 2015 by Olympus.
48 Woerd Avenue, Waltham, MA 02453, USA, Tel.: (1) 781-419-3900
12569 Gulf Freeway, Houston, TX 77034, USA, Tel.: (1) 281-922-9300
505, boul. du Parc-Technologique, Québec (Québec) G1P 4S9, Tel.: (1) 418-872-1155
For enquiries - contact 1109 78 Ave, Edmonton (Alberta) T6P 1L8
www.olympus-ims.com/contact-us
OmniScan_SX_EN_LTR_201506 • Printed in the USA • P/N: 920-269-EN Rev. E