0% found this document useful (0 votes)
471 views

ServiceManual For Data Communication EU PDF

This document provides instructions for connecting TOMEY products to a PC for data communication. It describes the required equipment, available communication protocols including X-modem and Z-modem, and steps for setting up the connection and transferring data. Products that support X-modem transfer data via a standard RS-232C cable connected to a PC running software like HyperTerminal. Z-modem products similarly transfer data via cable but save it as a CSV file readable by Microsoft Excel. Settings like baud rate must match between the device and PC.

Uploaded by

Abi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
471 views

ServiceManual For Data Communication EU PDF

This document provides instructions for connecting TOMEY products to a PC for data communication. It describes the required equipment, available communication protocols including X-modem and Z-modem, and steps for setting up the connection and transferring data. Products that support X-modem transfer data via a standard RS-232C cable connected to a PC running software like HyperTerminal. Z-modem products similarly transfer data via cable but save it as a CSV file readable by Microsoft Excel. Settings like baud rate must match between the device and PC.

Uploaded by

Abi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

Service

Data Communication v1.1


Manual

1. Information for data communication


Many of TOMEY products can send measurement data to PC.
This is an instruction to do data communication between PC and TOMEY unit.

Required tool
1) RS-232C cable (2m, cross-link, male-female) Art-No. 904-036
2) Software (ex: Hyper terminal)

Communication protocol
X-modem: TL-900, TL-2000, TL-3000, TL-2000A, TL-3000A. TR-1000, RC-1000, TR-3000, RC-3000, TR-4000,
RC-4000, RC-5000, AL-2000, UD-6000 (older software than 1.30/2.20)

Z-modem: AL-100, SP-3000, AL-3000, UD-6000 (new software than and 1.30/2.20)

Other way: TMS-3, TMS-4, RT-6000

<Page 1>
Service
Data Communication V1.1
Manual

2. Communication with X-modem format product.


1. Connect the Computer and TOMEY unit by RS-232C cable.
(If you need, use gender changer)

2. Open software which can read binary code. (For example: HyperTerminal)

3. Adjust communication condition on Tomey unit and communication software of the computer
(Baud rate, Data bit, Parity, Stop bit and so on)

4. Collect some data on the Tomey unit and send data to the computer
(Carriage return should be active by setting of software)

*Caution
If Tomey product cannot change communication condition, PC has to adjust as follows.

Baud Rate 2400 bps


Data bit 7bit
Parity Even
Stop bit 1bit
Data code ASCII
Flow control Xon/Xoff

<Page 2>
Service
Data Communication V1.1
Manual

3. Communication with Z-modem format product.


1. Connect the Computer and TOMEY unit by RS-232C cable.
(If you need, use gender changer)

2. Open software which can receive Z-mode-data. (For example: HyperTerminal)

3. Adjust communication condition on Tomey unit and communication software of the computer.
(Baud rate, Data bit, Parity, Stop bit and so on)

4. Collect some values on the Tomey unit and send data to the computer

Information
A. You will send measurement data with “CSV” format. This file able to be read with Microsoft Excel.
B. File name is saved as “TOMEY. csv”.
C. If you select simple Z-modem for receiving data it will save last as actual TOMEY.CSV, The Computer
automatically rename the older data to TOMEY0.CSV TOMEY1.CSV TOMEY2.CSV ...
D. If you select Z-modem with crash recover, new data is also “TOMEY. CSV” file. It must be removed for the
next measurement

<Page 3>