Filter Integrity Test Machine
Filter Integrity Test Machine
• Easy handling
INDUSTRIES:
• Chemical Industry
• Pharmaceutical Industry
• Biotech
• Environmental Industry
Web: www.donaldson.com
Process Filtration Filter Test Center
PRODUCT DESCRIPTION
The Donaldson Filter Test Center (FTC) serves This instrument is suitable for the
to determine the filter efficiency of sterile and testing of the following filter
elements:
ventilation filters. It contains all the required
components to perform an easy and safe test of • (P)-BE
filter elements. It has room in every lab due to its
• (P)-SRF
compact structure. The installed process
computer takes over all of the important • (P)-SRF N
functions during the measuring process,
• Other sterile air/gas depth filters
resulting in an easy and clear operation for the
user. The FTC is therefore suitable for series • Liquid depth filters
tests, as well as for occasional examinations of
individual filter elements.
The Donaldson FTC also contains the The aerosol generator provides a constant
components that serve the aerosol production as supply of test aerosol. The test aerosol is
well as the measurement of the aerosol checked by the system when the FTC is started
concentration. All functions are monitored and and in fixed intervals.
controlled by a process computer, which is
The photometer measures the concentration of
equipped with real-time capable multi-tasking
the test aerosol as well as the concentration of
software.
the aerosol filtered by the filter element to be
tested. These are forwarded to the process
computer.
-2-
PRODUCT DESCRIPTION
Process Filtration Filter Test Center
PRODUCT SPECIFICATIONS
Product Specifications
Weight: 32 kg
Compressed air supply: At least 5 bar (maximum of 10 bar) at a minimum of 40 l/min, dry air, free of
oil and particles.
Do not expose the equipment to extreme humidity over an extended period. Undesirable dewing
(condensation of humidity on the equipment) may occur, if cold equipment is brought into essentially
warmer surroundings. In this case, acclimatize the equipment for approx. 2 hours to the room
temperature after disconnecting it from the power supply. If the equipment is left switched on, a
moisture effect can be eliminated due to the continuously positive temperature gradient between the
interior of the equipment and the surroundings.
-3-
PRODUCT SPECIFICATIONS & MAINTENANCE
Process Filtration Filter Test Center
TEST DESIGN AND TEST RESULT
DELIVERY SCOPE
The delivery scope of the FTC contains the following single components: