Validation of Analytical Method
Validation of Analytical Method
-
-
THE VALIDATION PROTO
Identification
A generic summary validation protocol is g1ven for identification methods in Tabl.
Specificity Analyse samples which contain the analyte Method must be able to distinguish a
and samples which do not contain the from materials which are structurally
analyte (including samples containing similar or closely related. Positive re:
materials which are structurally similar or obtained for samples containing the
closely related to the analyte) as per analyre and negative results for those
method. containing the analyte.
I
~'
Compare the result obtained by , Spectrum/ chromatogram/ measurerr
performing the method on the analyte etc. must be the same as that due to :
with that obtained for a certified reference certified reference standard.
standard.
Assay
A generic summary validation protocol is g1ven for assay methods in Table 16.
Spiking drug substance/ drug product No peaks interfering with analyte peak
with the potentially interfering observed.
material.
Minimum resolution between analyte
Prepare samples of the potentially peak and neighbouring peaks = 1.5.
interfering materials.
No co-elution detected from peak
Perform stress studies on drug purity investigation.
substance/ drug product.
Chroma tographic:
or
82
© 2009 Mourne Training Services
THE VALIDATION PROTOCOL
Robustness Investigation using an experimental design No factor has a significant effect on the
based on the .critical factors identified assay results, or is controlled by an
during method development. appropriate system suitability test.
Stability Test solution stability evaluated by Results for original analysis and reanalysis
comparing stored solutions against results within 2%.
from original analysis. Typical storage
.
'". •. .~
f:".
.~_ r" conditions include room temperature and
refrigerated.
;
>
.
•.
different dosing options during clinical trials and post approval. The series are usually
analysed using the same assay method, the sample preparation is altered to account for
the differing amounts of active pharmaceutical ingredient present. During method
validation the design of the study needs to ensure that the application of the method
to all relevant dosage strengths is covered.
Other types of methods may be validated using a similar approach as for assay
methods. Examples are methods for the analysis component of dissolution testing,
and methods for testing uniformity of dosage units. Validation of the sample
preparation component of dissolution testing involves consideration of the method in
question, e.g. the withdrawal and filtration of samples. This topic is discussed by C. C.
Chan et al. in 'Analytical Method Validation and Instrument Performance'."
Impurities
A generic summary validation protocol is given for quantitative impurity methods in
Table 17.
,
0'
•..
83
© 2009 Mourne Training Ser!ices
'THE VALIDATION OF ANALYTICAL METHODS
Table 17 Suggestions for inclusion in a validation protocol for a quantitative impurities method
Spiking drug substance/ dmg product No peaks interfering with the analyte
with the potentially interfering peak or the peaks due to the
material. impurities of interest observed.
Chromatographic:
Range The range is defined by the results Linearity, accuracy and precision
obtained for linearity, accuracy and demonstrated over the range.
precision.
.,
'.
-
..
~
"
..
1.•• •
",
i~
t·
q
84 f'
,~
© 2009 Mourne Training Services .(
'.
THE VALIDATION PROTO
Table 17
Suggestions for inclusion in a validation protocol for a quantitative impurities method
or
Detection limit Prepare 6 samples at or near the expected -0.01 to 0.02% nominal
detection limit (calculate level from
required reporting thresholds, i.e..,QLl3) SIN - 3
by spiking.
Quantitation limit Prepare 6 samples at or near the expected -0.05 to 0.1% nominal
quantirarion limit by spiking.
SIN - 10
Robustness Investigation using an experimental design No factor studied has a significant ef[e
based on the critical factors identified on the assay results, or, factor effects 2
during method development. controlled by an appropriate system
suitability test.
Stability Test solution stability evaluated by Results [or original analysis and reanal-
comparing stored solutions against results within 2%.
[rom original analysis. Typical storage
conditions include room temperature and
refrigerated.
The validation of quantitative analytical methods for residual solvents follows a sin
approach as above where the reporting thresholds for individual solvents are defi
in the ICH guideline, 'Impurities: Guideline for Residual solvents'i> Limit tests
residual soivents require investigation of specificity and the detection limit only.
Other examples of applications of limit tests are for 10w-lev<J toxins (highly tr
substances which can be detrimental to the health of the patient at very
concenrrations) and polymorphic impurities. A limit test requires less validation than
a quantification method and thus is often preferred. More information on the
validation of methods for .low level toxins is provided by Skett in 'Analysis of Drug
Impurities." Experimental approaches and acceptance criteria similar to that used for
impurities may also be applied to methods for water determination.
Bioanalytical methods
A generic summary validation protocol is given for bioanalytical methods in Table 18.
This summary is based on the FDA guidelines for bioanalytical method validation)
Selectivity Blank samples of the biological matrix No interference observed for analyte
from at least 6 sources tested. response.
Samples prepared and tested for all Chromatographic:
potentially interfering materials, e.g.
No peaks interfering with analyte peak
metabolites.
observed.
Minimum resolution between analyte
peak and neighbouring peaks =
1.5.
Precision Minimum of five determinations at 3 %RSD should not exceed 15%, except for
concentrations in the expected range (use LLOQ, where it should not exceed 20%.
accuracy samples).
Recovery Compare extracted samples at 3 Recovery need not be 100%, but should
concentrations (low, medium and high) be consistent.
with unextracred standards that represent
100% recovery.
Lower Limit of
standard).
86
© 2009 Mourne Training Services
THE VALIDATION PROTOCOL
Stability Freeze and thaw stability, short-term Stored result within 15% of the control,
temperature stability and long-term
stability: Equivalence shown using statistical test
based on confidence interval.
3 aliquots of each of high and low
concentrations stored and reanalysed
alongside control samples.
Other methods
For other methods which are not discussed above, the experimental procedure should
be based on requirements in the ICH guidelines. The acceptance criteria should be
considered on a case by case basis and Table 15 through to Table 18 may provide a
starting point for defining suitable values.
Automated methods
The automation of analytical methods is used to increase sample throughput, help
ensure data integrity and improve laboratory safety. I [ a method is considered as
being composed of three elements: sample preparation, measurement and data
handling then the second and third elements are routinely automated using
instrumental analysis. The use of autoinjectors is common in most laboratories and
computer controlled systems allow automatic calculation of ~esults. Sample
preparation is usually the most labour intensive part of an analysis and automated
87
© 2009 Mourne Training Services
E VALIDATION OF ANALYTICAL METHODS
The development of the synthetic route for the drug substance is ongoing.
88
)09 Mourne Training Services
THE VALIDATION PROTOC
• Specificity and evaluation of the quantitation limit are the primary characteristics
ensure that assay and impurity methods meet their intended purposes of poter
and safety.
• For precision testing synthetic mixtures of drug substance and placebo may
used, rather than authentic samples, thus enabling the combination of accura.
precision and potentially linearity at the same time.
• Formal intermediate precision experiments are not yet needed but if differs
laboratories need to operate the method then the handover will require suital
validation.
• The evaluation of the detection limit for impurity methods may be delayed.