Liquiline CM442
Liquiline CM442
Technical Information
Liquiline CM442/CM444/CM448
Multiparameter transmitter with a maximum of eight
measuring channels based on digital Memosens
technology
Table of contents
2 Endress+Hauser
Liquiline CM442/CM444/CM448
Environment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Pollution degree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Pressure compensation to environment . . . . . . . . . . . . . . . . . . 36
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . . 36
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Language packages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Product structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Weather protection cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Post mounting kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Measuring cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Additional functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Other accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
3 Endress+Hauser
Liquiline CM442/CM444/CM448
Conductivity
Inductive conductivity
measurement in wastewater
treatment
• Sensor Indumax CLS50D
with fixed cable
Conductive conductivity
measurement in power plant
cooling water
• Sensor Condumax CLS15D
a0012412
Oxygen
Oxygen in aeration basins
• Immersion assembly Dipfit
5 4 3 CYA112
• Holder CYH112
a0012411
• Sensor
Measuring system (e.g. two-channel device)
– COS61D (optical)
1 Liquiline with fixed cable,
2 Weather protection cover CYY101 (optional) – COS51D (amperometric)
3, 5 Sensor cable CYK10 or fixed cable (digital fixed cable
sensors with Memosens protocol) cable CYK10
4 Power supply cable (to be provided by the customer, not Figure: CYA112 with COS61D
part of the scope of supply)
a0012415
a0015030 a0015077
4 Endress+Hauser
Liquiline CM442/CM444/CM448
If mounting outdoors, always use the weather protection cover (see "Accessories") to protect the
transmitter against weather conditions.
Application example Measuring point at wastewater treatment plant outlet (open channel)
• Transmitter CM444-AAM44A0FF with:
4x Memosens, Modbus TCP, 4 x relays for cleaning/limit value, 2 x analog current input
• pH and temperature with CPS11D, item 1, (www.products.endress.com/cps11d)
• Turbidity with CUS51D, item 2, (www.products.endress.com/cus51d)
• Nitrate with CAS51D, item 3, (www.products.endress.com/cas51d)
• Spectral absorption coefficient with CAS51D, item 4, (www.products.endress.com/cas51d)
• Flow from external measurement via current input
• Assembly holder CYH112 with assemblies CYA112 (www.products.endress.com/cyh112)
a0019831
Data retention
• Storage of all measured values, incl. values of external sources, in the non-volatile memory (data
logbook)
• Data called up on site via user-defined measuring menu and load curve display of the data logbook
• Transmission of data by ethernet and storage in a tamper-proof database (Field Data Manager)
• Data export to csv file (for Microsoft Excel)
Endress+Hauser 5
Liquiline CM442/CM444/CM448
Equipment architecture
Slot and port assignment
a0019832
a0019833
Order of the modules Depending on the version ordered, the device is supplied with a number of electronic modules, which
are assigned in a specific sequence in ascending order to slots 0 to 7.
If you do not have a particular module, the next moves up automatically:
• The basic module (which is always present) always occupies slots 0 and 1
• Fieldbus module 485
• Memosens input module 2DS (DS = digital sensor)
• Extension module for digital inputs and outputs DIO (DIO = digital input and output)
• Current input module 2AI (AI = analog input)
• Current output module 4AO or 2AO (AO = analog output)
• Relay modules AOR, 4R or 2R (AOR = analog output + relay, R = relay)
Modules with 4 ports are connected before modules of the same type with 2 ports.
Basic rule for hardware Please note the following if upgrading the device:
upgrades The sum of all current inputs and outputs may not exceed 8!
6 Endress+Hauser
Liquiline CM442/CM444/CM448
Determining the hardware You must be aware of the type of modules and the number of them supplied with the device you have
delivery status ordered to determine the delivery status of your Liquiline.
• Basic module
– One basic module in all versions. Always occupies slots 0 and 1.
• Fieldbus module
– Optional, and only one fieldbus module is possible.
• Input modules
– Must be clearly assigned to the number of optional inputs ordered.
– Examples:
2 current inputs = module 2AI
4 Memosens inputs = 2 inputs with basic module + module 2DS with 2 further inputs
• Current outputs and relays
– Various module combinations can exist.
The following table will help you find out which modules you get depending on the type and
number of outputs.
Relay
Current 0 2 4
outputs
2 - 1 x 2R 1 x 4R
‣ Sum up the number of modules and sort them according to the specified sequence.
This will give you the slot assignment for your device.
Terminal diagram The unique terminal name is derived from the following:
Slot no. : Port no. : Terminal
a0015979
Creating a terminal diagram using the example of the NO contact (terminal 41) of a relay
Endress+Hauser 7
Liquiline CM442/CM444/CM448
a0019808
CM442-**M1A1F0*
Extension options by using an • Ethernet/PROFIBUS DP/Modbus with module 485 incl. activation
extension module in free slot 2 code for the desired communication type:
– PROFIBUS DP (71140888)
– Modbus RS485 (71140889)
– Modbus TCP (71140890)
– EtherNet/IP (71219868)
– Only Ethernet without fieldbus (71135634)
If fieldbus communication is subsequently required, an activation
code is needed for this.
If you retrofit a 485 module, any existing current outputs are
disabled!
• Additional inputs or outputs, relays:
– Module 2AI (71135639): 2 current inputs
– Module 2AO (71135632): 2 current outputs
– Module AOR (71111053): 2 current outputs, 2 relays
– Module 2R (71125375) or 4R (71125376): 2 or 4 relays
– Module DIO (71135638): 2 digital inputs and 2 digital outputs
Device upgrade to CM444 or CM448 • Upgrade kit 71135644 (100 to 230 V AC) or 71211434 (24 V DC)
– Extension power supply unit and backplane
– Basic module BASE-E (Memosens inputs same as for base device)
– 6 slots for extension modules
• Extension options:
– Second Memosens input (71114663), additional modules same
as for CM442
– Up to 8 measuring channels by using an appropriate number of
Memosens input modules 2DS (71135631)
Basic rule for extensions The sum of all current inputs and outputs may not exceed 8!
8 Endress+Hauser
Liquiline CM442/CM444/CM448
a0015791
Endress+Hauser 9
Liquiline CM442/CM444/CM448
a0019809
CM444-**M42A1FA*
Device upgrade to CM448 6 or 8 measuring channels by using one or two Memosens input
modules 2DS (71135631)
Basic rule for extensions The sum of all current inputs and outputs may not exceed 8!
Restrictions if using CUS71D sensors • With the CM444, any combination of maximum 4 Memosens
for interface measurement sensors is possible.
• An extension to CM448 is not advisable as the maximum number
of Memosens inputs remains limited to 4 if a CUS71D is used.
10 Endress+Hauser
Liquiline CM442/CM444/CM448
a0015827
Endress+Hauser 11
Liquiline CM442/CM444/CM448
a0019810
CM448-**26A1*
Extension options by using extension Only the following is possible for the example above:
modules in free slots 5-7 • Module 2R (71125375) or 4R (71125376): 2 or 4 relays
If extending to eight measuring channels:
• Module 2DS (71135631): 2 Memosens inputs
• Use of the 2 current outputs in the basic module by entering
activation code (71140891)
Additional inputs or outputs and relays if fieldbus module 485 is
removed:
• Module 2AO (71135632): 2 current outputs
• Module AOR (71111053): 2 current outputs, 2 relays
• Module 2R (71125375) or 4R (71125376): 2 or 4 relays
• Module DIO (71135638): 2 digital inputs and 2 digital outputs
Basic rule for extensions The sum of all current inputs and outputs may not exceed 8!
Restrictions if using CUS71D sensors The maximum number of Memosens inputs that can be used is limited
for interface measurement to 4! Here, every combination of CUS71D and other Memosens
sensors is then possible.
12 Endress+Hauser
Liquiline CM442/CM444/CM448
a0015828
Endress+Hauser 13
Liquiline CM442/CM444/CM448
Dependability
Reliability Memosens technology
MEMO SENS
14 Endress+Hauser
Liquiline CM442/CM444/CM448
Self-monitoring functions
Current inputs are deactivated in the event of overcurrent and reactivated once the overcurrent stops.
Board voltages are monitored and the board temperature is also measured.
USP and EP
The limit functions for pharmaceutical water in accordance with USP and EP specifications are
implemented in the software for conductivity measurements:
• Water for Injection (WFI) as per USP <645> and EP
• Highly purified water (HPW) as per EP
• Purified water (PW) as per EP
The uncompensated conductivity value and the temperature are measured for the USP/EP limit
functions. The measured values are compared against the tables defined in the standards. An alarm is
triggered if the limit value is exceeded. Furthermore, it is also possible to configure an early warning
alarm that signals undesired operating states before they occur.
ChemocleanPlus
Freely programmable sequence control
• e.g. for automatic sensor cleaning in retractable assemblies for reliable measurement results in
processes with a high risk of contamination
• Individual, time-based activation of 4 outputs e.g. relays
• Starting, stopping or pausing of activities via digital input or fieldbus signals e.g. from limit position
switches
a0016027 a0012440
Endress+Hauser 15
Liquiline CM442/CM444/CM448
a0015032
SD card
The exchangeable storage medium enables:
• Quick and easy software updates and upgrades
• Data storage of internal device memory (e.g. logs)
• Transfer of complete configurations to a device with an identical setup (backup function)
• Transfer of configurations without the TAG and bus address to devices with an identical setup (copy
function)
Endress+Hauser offers industry-approved SD cards as accessories. These memory cards provide
maximum data security and integrity.
Other SD cards can also be used. However, Endress+Hauser does not accept any responsibility for the
data security of such cards.
External signals for device control and for activating external devices
Hardware options, e.g. module "DIO" with 2 digital inputs and 2 digital outputs or fieldbus module "485"
enable the following:
• via a digital input signal
– measuring range switching for conductivity (upgrade code required, see accessories)
– switching between different calibration datasets in the case of optical sensors
– an external hold
– a cleaning interval to be triggered
– switching on and off a PID controller, e.g. via the proximity switch of the CCA250
– the use of the input as an "analog input" for pulse-frequency modulation (PFM)
• via a digital output signal
– the static transmission (similar to a relay) of diagnostic states, point level switch states etc.
– the dynamic transmission (comparable to a non-wearing "analog output") of PFM signals, e.g. to
control dosing pumps.
16 Endress+Hauser
Liquiline CM442/CM444/CM448
a0016009
Endress+Hauser 17
Liquiline CM442/CM444/CM448
Concentration tables
When the device is delivered from the factory, tables are saved in the device that allow inductive
conductivity measurements to be converted to concentrations of certain substances. 4 user-defined
tables are also possible.
The following factory concentration tables are available:
Data security
All settings, logbooks etc. are stored in a non-volatile memory to ensure that the data are retained even
in the event of a disruption to the power supply.
Password protection
Password-protected login
• For remote operation via web server
• For local operation
Process safety
Two independent PID controllers
• One- or two-sided control
• Limit switches
• 4 cleaning programs which can be programmed independently of each other
IT security
We provide a warranty only if the device is installed and used as described in the Operating
Instructions. The device is equipped with security mechanisms to protect it against any inadvertent
changes to the device settings.
IT security measures in line with operators' security standards and designed to provide additional
protection for the device and device data transfer must be implemented by the operators themselves.
Support in the performance of this task can be requested from Endress+Hauser.
18 Endress+Hauser
Liquiline CM442/CM444/CM448
Input
Measured variables --> Documentation of the connected sensor
Types of input • Digital sensor inputs for sensors with Memosens protocol
• Analog current inputs (optional)
• Digital inputs (optional)
Cable length
Max. 100 m (330 ft)
Span High: 11 to 30 V DC
Low: 0 to 5 V DC
Endress+Hauser 19
Liquiline CM442/CM444/CM448
Output
Output signal Depending on version:
• 2 x 0/4 to 20 mA, active, potentially isolated from one another and from the sensor circuits
• 4 x 0/4 to 20 mA, active, potentially isolated from one another and from the sensor circuits
• 6 x 0/4 to 20 mA, active, potentially isolated from one another and from the sensor circuits
• 8 x 0/4 to 20 mA, active, potentially isolated from one another and from the sensor circuits
• Optional HART communication (only via current output 1:1)
HART
PROFIBUS DP/RS485
Data transmission rate 9.6 kBd, 19.2 kBd, 45.45kBd, 93.75 kBd, 187.5 kBd, 500 kBd,
1.5 MBd, 6 MBd, 12 MBd
Modbus RS485
EtherNet/IP
20 Endress+Hauser
Liquiline CM442/CM444/CM448
Linearization/transmission Linear
behavior
Test voltage
500 V
Cross-section
Max. 2.5 mm2 (14 AWG)
Endress+Hauser 21
Liquiline CM442/CM444/CM448
Relay outputs
Electrical specification Relay types
• 1 single-pin changeover contact (alarm relay)
• 2 or 4 single-pin changeover contacts (optional with extension modules)
0.5 A 450,000
0.5 A 650,000
0.5 A 350,000
Extension modules
0.5 A 450,000
2A 120,000
0.5 A 650,000
2A 170,000
0.5 A 350,000
2A 150,000
22 Endress+Hauser
Liquiline CM442/CM444/CM448
Protocol-specific data
HART Manufacturer ID 11h
Output data 16 measured values (value, unit, status), 8 digital values (value,
status)
Input data 4 setpoints (value, unit, status), 8 digital values (value, status),
diagnostic information
Endress+Hauser 23
Liquiline CM442/CM444/CM448
TCP connections 3
Protocol TCP
Function codes 03, 04, 06, 08, 16, 23
Output data 16 measured values (value, unit, status), 8 digital values (value,
status)
Input data 4 setpoints (value, unit, status), 8 digital values (value, status),
diagnostic information
Polarity Auto-MIDI-X
Connections (max.) CIP 12
I/O 6
Explicit 6
message
Multicast 3 consumers
Web server The web server enables full access to the device configuration, measured values, diagnostic messages,
logbooks and service data via standard WiFi/WLAN/LAN/GSM or 3G router with a user-defined IP
address.
TCP port 80
24 Endress+Hauser
Liquiline CM442/CM444/CM448
Power supply
Supply voltage CM442
Depending on version:
100 to 230 V AC ± 15 %, 50/60 Hz
24 V AC/DC +20/-15 %, 50/60 Hz
CM444 and CM448
Depending on version:
100 to 230 V AC ± 15 %, 50/60 Hz
24 V DC +20 / -15 %
NOTICE
The device does not have a mains switch.
‣ The customer must provide a protected circuit breaker in the vicinity of the device.
‣ This must be a switch or a power-circuit breaker and must be labeled as the circuit breaker for the
device.
‣ At the supply point, the power supply for the 24 V versions must be isolated from dangerous live
cables by double or reinforced insulation.
Fuse CM442
5x20 mm, 250 V, 4.0 A, slow-blow (T4.0A)
CM444 and CM448
Fuse not exchangeable
Cable entries Identification of the cable entry on housing base Suitable gland
E -
) M12x1.5 mm
Recommended assignment
a0018025
Endress+Hauser 25
Liquiline CM442/CM444/CM448
a0012405 a0015871
Basic module BASE-H or -L (two-channel device) Basic module BASE-E (four- and eight-channel device)
1 Power supply for digital fixed cable sensors with 1 Power supply for digital fixed cable sensors with
Memosens protocol Memosens protocol
2 SD card slot 2 SD card slot
3 Slot for display cable1) 3 Slot for display cable1)
4 Service interface 4 Service interface
5 Connections for 2 Memosens sensors 5 Connections for 2 Memosens sensors
6 Current outputs 6 Current outputs
7 Power connection 7 LEDs
8 Fuse 8 Socket for internal supply cable 1)
9 Alarm relay connection 9 Alarm relay connection
26 Endress+Hauser
Liquiline CM442/CM444/CM448
a0015825
a0015872 a0015873
Connecting power supply with BASE-E Overall wiring diagram BASE-E and extension power unit
(B)
A Internal supply cable
B Extension power supply unit
Endress+Hauser 27
Liquiline CM442/CM444/CM448
Connecting optional modules With extension modules you can purchase additional functions for your device.
NOTICE
Unacceptable hardware combinations (due to conflicts in power supply)
Incorrect measurements or total failure of the measuring point as a result of heat build-up or
overloading
‣ If you are planning to extend your controller, make sure the resulting hardware combination is
permitted (Configurator on www.products.endress.com/cm442 or .../cm444 or .../CM448).
‣ Please note that if you are extending CM442 to CM444 or CM448, you must additionally fit an
extension power supply unit and an extension backplane. You must then also use basic module
BASE-E.
‣ Remember that the sum of all current inputs and outputs may not exceed 8!
‣ Please contact your Endress+Hauser sales center should you have any questions.
Module name
28 Endress+Hauser
Liquiline CM442/CM444/CM448
Module name
a0012454
Endress+Hauser 29
Liquiline CM442/CM444/CM448
a0012459 a0012460
Sensors without additional supply voltage Sensors with additional supply voltage
30 Endress+Hauser
Liquiline CM442/CM444/CM448
a0016197
Endress+Hauser 31
Liquiline CM442/CM444/CM448
Performance characteristics
Response time Current outputs
t90 = max. 500 ms for an increase from 0 to 20 mA
Current inputs
t90 = max. 330 ms for an increase from 0 to 20 mA
Digital inputs and outputs
t90 = max. 330 ms for an increase from low to high
Installation
Mounting plate
a0012426
32 Endress+Hauser
Liquiline CM442/CM444/CM448
a0012428
Installation instructions You require the post mounting kit (optional) to mount the unit on a pipe, post or railing (square or
circular, span range 20 to 61 mm (0.79 to 2.40")).
Post mounting
a0012665
Post mounting
1 Weather protection cover (optional) 5 Spring washers, nuts (post mounting kit)
2 Post mounting plate (post mounting kit) 6 Pipe or railing (circular/square)
3 Spring washers, nuts (post mounting kit) 7 Mounting plate
4 Pipe clamps (post mounting kit) 8 Threaded rods (post mounting kit)
Endress+Hauser 33
Liquiline CM442/CM444/CM448
Rail mounting
a0012668
Rail mounting
1 Weather protection cover (optional) 6 Pipe or railing (circular/square)
2 Post mounting plate (post mounting kit) 7 Mounting plate
3 Spring washers, nuts (post mounting kit) 8 Threaded rods (post mounting kit)
4 Pipe clamps (post mounting kit) 9 Screws (post mounting kit)
5 Spring washers, nuts (post mounting kit)
Wall mounting
Mount the controller in such a way that the wall support surface is at least the size of the rear housing
panel.
a0012684 a0012685
1) The size of the drill holes depends on the wall plugs used. The wall plugs and screws must be
provided by the customer.
34 Endress+Hauser
Liquiline CM442/CM444/CM448
Environment
Ambient temperature range CM442
–20 to 60 ˚C (0 to 140 ˚F)
CM444
• Generally –20 to 55 ˚C (0 to 130 ˚F), with the exception of packages under the second point in the
list.
• -20 to 50 ˚C (0 to 120 ˚F) for the following packages:
– CM444-**M40A7FI******+...
– CM444-**M40A7FK******+...
– CM444-**N40A7FI******+...
– CM444-**N40A7FK******+...
CM448
• Generally –20 to 55 ˚C (0 to 130 ˚F), with the exception of packages under the second point in the
list.
• -20 to 50 ˚C (0 to 120 ˚F) for the following packages:
– CM448-***6AA******+...
– CM448-***8A4******+...
– CM448-***8A5******+...
– CM448-**28A3******+...
– CM448-**38A3******+...
– CM448-**48A3******+...
– CM448-**58A3******+...
– CM448-**68A3******+...
– CM448-**26A5******+...
– CM448-**36A5******+...
– CM448-**46A5******+...
– CM448-**56A5******+...
– CM448-**66A5******+...
– CM448-**22A7******+...
– CM448-**32A7******+...
– CM448-**42A7******+...
– CM448-**52A7******+...
– CM448-**62A7******+...
Degree of protection IP 66/67, impermeability and corrosion resistance in accordance with NEMA TYPE 4X
Wall mounting
Frequency range 10 to 150 Hz (sinusoidal)
Amplitude 10 to 12.9 Hz: 0.75 mm
12.9 to 150 Hz: 0.5 g 1)
Test duration 10 frequency cycles/ spatial axis, in 3 spatial axes (1 oct./min)
Electromagnetic Interference emission and interference immunity as per EN 61326-1: 2006, class A for industry
compatibility
Endress+Hauser 35
Liquiline CM442/CM444/CM448
Mechanical construction
Dimensions
a0012396
36 Endress+Hauser
Liquiline CM442/CM444/CM448
Cable mounting rail PBT GF30 FR, stainless steel 1.4301 (AISI304)
Operability
Display Graphic display:
• Back light with switch-off function
• Red display background for alarms alerts users to errors
• Transflective display technology for maximum contrast even in bright environments
• User-definable measuring menus mean you can always keep track of the values that are important
for your application.
• Load curve display
a0012437 a0012438
Operating concept The simple and structured operating concept sets new standards:
• Intuitive operation with the navigator and soft keys
• Fast configuration of application-specific measurement options
• Easy configuration and diagnosis thanks to plain-text display
• All languages that can be ordered are available in every device
a0012445 a0012786
Endress+Hauser 37
Liquiline CM442/CM444/CM448
a0011764
Overview of operation
1 Display (with red display background in alarm condition)
2 Navigator (jog/shuttle function)
3 Soft keys (function depends on menu)
Remote operation Via HART (e.g. using HART modem and FieldCare)
a0015608
38 Endress+Hauser
Liquiline CM442/CM444/CM448
Via PROFIBUS DP
a0015874
PROFIBUS DP
T Terminating resistor
a0015875
Modbus RS485
T Terminating resistor
Endress+Hauser 39
Liquiline CM442/CM444/CM448
a0015876
Language packages The language selected in the product structure is the operating language preset at the factory.
All other languages can be selected using the menu.
• English (US)
• German
• French
• Swedish
• Spanish
• Italian
• Dutch
• Portuguese
• Polish
• Russian
• Turkish
• Chinese (Simplified, PR China)
• Japanese
• Czech
40 Endress+Hauser
Liquiline CM442/CM444/CM448
Ordering information
Product structure You can create a valid and complete order code using the Endress + Hauser Configurator tool on the
Internet.
Enter the following addresses in your browser to access the relevant product page:
www.products.endress.com/cm442
www.products.endress.com/cm444
www.products.endress.com/cm448
1. The following options are available on the right-hand side of the page:
cCSAus The product complies with "CLASS 2252 05 - Process Control Equipment" and "CLASS 2252 85 -
Process Control Equipment - Certified to US Standards" requirements.
Endress+Hauser 41
Liquiline CM442/CM444/CM448
Accessories
The most important accessories that could be delivered at the time this document went to print
are listed below. For information on accessories that are not listed here, please contact your local
service or sales representation.
Weather protection cover CYY101 weather protection cover for field devices, absolutely essential if operating the unit outdoors
• Material: stainless steel 1.4301 (AISI 304)
• Order No. CYY101-A
42 Endress+Hauser
Liquiline CM442/CM444/CM448
Enamel pH electrodes
Ceramax CPS341D
• pH electrode with pH-sensitive enamel
• Meets highest demands of measuring accuracy, pressure, temperature, sterility and durability
• Order as per product structure (--> online Configurator, www.products.endress.com/cps341d)
• Technical Information TI00468C/07/EN
ORP sensors
Orbisint CPS12D
• ORP sensor with Memosens technology
• Dirt-repellent PTFE diaphragm;
• Order as per product structure (--> online Configurator, www.products.endress.com/cps12d)
• Technical Information TI367C/07/EN
Ceraliquid CPS42D
• ORP sensor with Memosens technology
• Ceramic diaphragm and KCl liquid electrolyte
• Order as per product structure (--> online Configurator, www.products.endress.com/cps42d)
• Technical Information TI373C/07/EN
Ceragel CPS72D
• ORP sensor with Memosens technology
• Poison-resistant reference with ion trap
• Order as per product structure (--> online Configurator, www.products.endress.com/cps72d)
• Technical Information TI374C/07/EN
Orbipac CPF82D
• Compact ORP sensor for installation or immersion operation in process water and wastewater
• Order as per product structure (--> online Configurator, www.products.endress.com/cpf82d)
• Technical Information TI191C/07/EN
Orbipore CPS92D
• ORP sensor with Memosens technology
• Open aperture diaphragm for media with high dirt load
• Order as per product structure (--> online Configurator, www.products.endress.com/cps92d)
• Technical Information TI435C/07/EN
pH ISFET sensors
Tophit CPS471D
• Sterilizable and autoclavable ISFET sensor with Memosens technology
• For the food and pharmaceutical industries, process engineering, water treatment and
biotechnology
• Order as per product structure (--> online Configurator, www.products.endress.com/cps471d)
• Technical Information TI283C/07/EN
Tophit CPS441D
• Sterilizable ISFET sensor with Memosens technology
• For media with low conductivity, with liquid KCl electrolyte
• Order as per product structure (--> online Configurator, www.products.endress.com/cps441d)
• Technical Information TI352C/07/EN
Tophit CPS491D
• ISFET sensor with Memosens technology
• Open aperture diaphragm for media with high dirt load
• Order as per product structure (--> online Configurator, www.products.endress.com/cps491d)
• Technical Information TI377C/07/EN
Endress+Hauser 43
Liquiline CM442/CM444/CM448
44 Endress+Hauser
Liquiline CM442/CM444/CM448
Oxygen sensors
Oxymax COS51D
• Amperometric sensor for dissolved oxygen, with Memosens technology
• Order as per product structure (--> online Configurator, www.products.endress.com/cos51d)
• Technical Information TI00413C/07/EN
Oxymax COS61D
• Optical oxygen sensor for drinking water and industrial water measurement
• Measuring principle: quenching
• Memosens protocol
• Material: stainless steel 1.4571 (AISI 316Ti)
• Order as per product structure (--> online Configurator, www.products.endress.com/cos61d)
• Technical Information TI387C/07/EN
Oxymax COS22D
• Sterilizable sensor for dissolved oxygen
• Order as per product structure (--> online Configurator, www.products.endress.com/cos22d)
• Technical Information TI446C/07/EN
Chlorine sensors
CCS142D
• Membrane-covered amperometric sensor for free chlorine
• Memosens technology
• Measuring range 0.01 to 20 mg/l
• Order as per product structure (--> online Configurator, www.products.endress.com/ccs142d)
• Technical Information TI419C/07/EN
Turbidity sensors
Turbimax CUS51D
• For nephelometric measurements of turbidity and solids in wastewater
• 4-beam scattered light method
• With Memosens protocol
• Order as per product structure (--> online Configurator, www.products.endress.com/cus51d)
• Technical Information TI461C/07/EN
Interface measurement
Turbimax CUS71D
• Immersion sensor for interface measurement
• Ultrasonic interface sensor
• Order as per product structure (--> online Configurator, www.products.endress.com/cus71d)
• Technical Information TI490C/07/EN
Endress+Hauser 45
Liquiline CM442/CM444/CM448
46 Endress+Hauser
Liquiline CM442/CM444/CM448
Cable glands
Kit CM44x: gland M
• Set, 6 pieces
• Order no. 71101768
Kit CM44x: gland NPT
• Set, 6 pieces
• Order no. 71101770
Kit CM44x: gland G
• Set, 6 pieces
• Order no. 71101771
Kit CM44x: dummy plug for cable gland
• Set, 6 pieces
• Order no. 71104942
Endress+Hauser 47
Liquiline CM442/CM444/CM448
Communication-specific accessories
Commubox FXA195 HART
• Intrinsically safe HART communication with FieldCare via the USB port
• Technical Information TI00404F
Commubox FXA291
• Connects the CDI interface of measuring devices with the USB port of the computer or laptop
• Technical Information TI00405C
WirelessHART adapter SWA70
• Wireless device connection
• Easily integrated, offers data protection and transmission safety, can be operated in parallel with
other wireless networks, minimum cabling complexity
• Technical Information TI00061S
Fieldgate FXA320
• Gateway for the remote interrogation of 4-20 mA measuring devices via a Web browser
• Technical Information TI00025S
Fieldgate FXA520
• Gateway for the remote diagnostics and configuration of connected HART devices via a Web browser
• Technical Information TI00051S
Field Xpert SFX100
• Compact, flexible and robust industrial handheld terminal for remote configuration and for
obtaining measured values via the HART current output
• Operating Instructions BA00060S
System components
RIA14, RIA16
• Field display unit for integration into 4 to 20 mA circuits
• RIA14 in flameproof metal enclosure
• Technical Information TI00143R and TI00144R
RIA15
• Process display unit, digital display unit for integration into 4 to 20 mA circuits
• Panel mounting
• With optional HART communication
• Technical Information TI01043K
48 Endress+Hauser
www.addresses.endress.com