Geiger Test Proc
Geiger Test Proc
Sponsor
Nuclear Instruments and Detectors Committee
of the
IEEE Nuclear and Plasma Sciences Society
Abstract: Test procedures for Geiger-Mueller counters that are used for the detection of ionizing radiation
are presented so that they have the same meaning to both manufacturers and users. Also included is
information on bases (i.e., connections) for the counters.
Keywords: gas counter, Geiger-Mueller, radiation detectors
No part of this publication may be reproduced in any form, in an electronic retrieval system or otherwise, without the prior written per-
mission of the publisher.
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
IEEE Standards documents are developed within the IEEE Societies and the Standards Coordinating Com-
mittees of the IEEE Standards Association (IEEE-SA) Standards Board. Members of the committees serve
voluntarily and without compensation. They are not necessarily members of the Institute. The standards
developed within IEEE represent a consensus of the broad expertise on the subject within the Institute as
well as those activities outside of IEEE that have expressed an interest in participating in the development of
the standard.
Use of an IEEE Standard is wholly voluntary. The existence of an IEEE Standard does not imply that there
are no other ways to produce, test, measure, purchase, market, or provide other goods and services related to
the scope of the IEEE Standard. Furthermore, the viewpoint expressed at the time a standard is approved and
issued is subject to change brought about through developments in the state of the art and comments
received from users of the standard. Every IEEE Standard is subjected to review at least every five years for
revision or reaffirmation. When a document is more than five years old and has not been reaffirmed, it is rea-
sonable to conclude that its contents, although still of some value, do not wholly reflect the present state of
the art. Users are cautioned to check to determine that they have the latest edition of any IEEE Standard.
Comments for revision of IEEE Standards are welcome from any interested party, regardless of membership
affiliation with IEEE. Suggestions for changes in documents should be in the form of a proposed change of
text, together with appropriate supporting comments.
Interpretations: Occasionally questions may arise regarding the meaning of portions of standards as they
relate to specific applications. When the need for interpretations is brought to the attention of IEEE, the
Institute will initiate action to prepare appropriate responses. Since IEEE Standards represent a consensus of
all concerned interests, it is important to ensure that any interpretation has also received the concurrence of a
balance of interests. For this reason, IEEE and the members of its societies and Standards Coordinating
Committees are not able to provide an instant response to interpretation requests except in those cases where
the matter has previously received formal consideration.
Note: Attention is called to the possibility that implementation of this standard may
require use of subject matter covered by patent rights. By publication of this standard,
no position is taken with respect to the existence or validity of any patent rights in
connection therewith. The IEEE shall not be responsible for identifying patents for
which a license may be required by an IEEE standard or for conducting inquiries into
the legal validity or scope of those patents that are brought to its attention.
Authorization to photocopy portions of any individual standard for internal or personal use is granted by the
Institute of Electrical and Electronics Engineers, Inc., provided that the appropriate fee is paid to Copyright
Clearance Center. To arrange for payment of licensing fee, please contact Copyright Clearance Center, Cus-
tomer Service, 222 Rosewood Drive, Danvers, MA 01923 USA; (978) 750-8400. Permission to photocopy
portions of any individual standard for educational classroom use can also be obtained through the Copy-
right Clearance Center.
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
Introduction
(This introduction is not part of IEEE Std 309-1999/ANSI N42.3-1999, IEEE Standard Test Procedures and Bases for
Geiger-Mueller Counters.)
This standard is a revision of the 1970 version. It presents standard test procedures for Geiger-Mueller
counters that are used for the detection of ionizing radiation. Also included is information on standard and
typical bases for the counters. This revision has been approved by the Nuclear Instruments and Detectors
Committee (NIDCom) of the IEEE Nuclear and Plasma Sciences Society and by the Accredited Standards
Committee N42 on Nuclear Instrumentation of the American National Standards Institute (ANSI) for which
the IEEE serves as the Secretariat.
These detectors, originally announced by Geiger and Müller in 1928, represented a major step forward in the
detection and counting of individual radioactive events. The amplitude of the output signal is independent of
the energy of the event. Therefore, when only the counting of radioactive events is required, without regard
for the type of radiation or its energy, Geiger-Mueller counters are useful. The output signal is large enough
to operate a scaler without additional amplification, the regulation requirements of the power supply can be
far less than that required for the solid state detectors, and, unlike germanium detectors, no cryostat is
required.
The principal attributes of the Geiger-Mueller counters are their low cost and simplicity, which accounts for
their wide usage in many applications. Ionization chambers and the subsequently developed scintillation
counter and semiconductor radiation detectors are capable of distinguishing between different types of radi-
ation and their energies and can therefore be used for spectroscopy as well as for counting. The scintillation
and semiconductor detectors, because they are solid rather than a gas, are highly efficient for the detection of
gamma rays. The principal attributes of the scintillation counters are high efficiency and moderate energy
resolution; whereas, the semiconductor detectors have outstanding energy resolution.
Ionization chambers, scintillation counters, and semiconductor detectors are dealt with in other standards
publications.
An excellent description of Geiger-Mueller counters and their operation is provided in Chapter 7 of Glenn F.
Knoll’s book, Radiation Detection and Measurement. (See Knoll [B4] in Annex A.)
Participants
At the time this standard was completed, the working group of the Nuclear Instruments and Detectors Com-
mittee had the following membership:
David J. Allard, Project leader
Louis Costrell, Project leader
Carl R. Siebentritt, Project leader
David Barclay M. R. Farukhi Kenneth L. Swinth
Joseph G. Bellian James K. Hesch Al N. Tschaeche
Peter Burgess Joseph C. McDonald Michael Unterweger
Edward Fairstein Donald E. Stilwell Sanford Wagner
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
At the time this standard was completed, the Nuclear Instruments and Detectors Committee had the follow-
ing membership:
Donald E. Stilwell, Chair
Michael Unterweger, Vice Chair
Louis Costrell, Secretary
David J. Allard Edward Fairstein Dennis E. Persyk
Joseph G. Bellian Erik H. M. Heijne Paul L. Phelps
William M. Bugg Ronald M. Keyser Carl R. Siebentritt
Christopher Cox Frederick A. Kirsten Kenneth L. Swinth
Larry Darken Glenn F. Knoll Sanford Wagner
W. Kenneth Dawson G. Laurie Miller John Walter
Klaus D. Mueller
When the IEEE-SA Standards Board approved this standard on 18 March 1999, it had the following
membership:
*Member Emeritus
Subsequently, the accredited American National Standards Committee N42 on Radiation Instrumentation
also reviewed and approved this document. At the time of approval, it had the following membership:
Louis Costrell, Chair Michael Unterweger, Vice Chair
Sue Vogel, Administrative Secretary
Organization Represented Name of Representative
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
NASA/GSFC, U.S.................................................................................................................... Donald E. Stilwell*
Nuclear Standards Unlimited................................................................................................... Al N. Tschaeche
Oak Ridge National Laboratory............................................................................................... Charles L. Britton
Swinth Associates .................................................................................................................... Kenneth L. Swinth
U.S. Army Center for EW/RSDA ............................................................................................ Edward Groeber
U.S. Department of Commerce, National Institute of Standards and
Technology ............................................................................................................................ Louis Costrell
Michael Unterweger (Alt.)
U.S. Department of Energy/PCSRD ........................................................................................ Natesh Varma
U.S. Federal Emergency Management Agency ....................................................................... Carl R. Siebentritt
Members-at-Large.................................................................................................................... Paul L. Phelps
Lee J. Wagner
*Deceased
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
Contents
1. Overview.............................................................................................................................................. 1
1.1 Scope............................................................................................................................................ 1
1.2 Purpose......................................................................................................................................... 1
2. References............................................................................................................................................ 1
3.1 Definitions.................................................................................................................................... 2
3.2 Symbols, abbreviations, and acronyms........................................................................................ 5
4. General................................................................................................................................................. 5
8. Test equipment................................................................................................................................... 14
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
9. Specifications of Geiger-Mueller counters ........................................................................................ 15
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
IEEE Standard Test Procedures and
Bases for Geiger-Mueller Counters
1. Overview
1.1 Scope
This standard applies to Geiger-Mueller (G-M) counters that are used for the detection and counting of indi-
vidual particles.
1.2 Purpose
a) To establish test methods and test procedures so that they have the same meaning to both manufac-
turers and users.
b) To present additional information regarding G-M counters, such as information about the counters’
bases.
2. References
This standard shall be used in conjunction with the following publications. When the following standards are
superseded by an approved revision, the revision shall apply.
1ANSI N42 standards are available from the Institute of Electrical and Electronics Engineers, 445 Hoes Lane, P.O. Box 1331, Piscat-
away, NJ 08855-1331, USA (http://www.standards.ieee.org/).
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
IEEE Std 309-1999/ANSI N42.3-1999 IEEE STANDARD TEST PROCEDURES AND BASES
IEC 60151-25 (1971-01), Measurements of the electrical properties of electronic tubes and valves. Part 25:
Methods of measurement of Geiger-Mueller counter tubes.2
IEEE Std 100-1996, IEEE Standard Dictionary of Electrical and Electronics Terms.3
IEEE Std 301-1988 (Reaff 1993), IEEE Standard Test Procedures for Amplifiers and Preamplifiers Used
with Detectors of Ionizing Radiation.
ISO 4037-1:1996, X and gamma reference radiation for calibrating dosimeters and dose rate meters and for
determining their response as a function of photon energy—Part 1: Radiation characteristics and production
methods.4
ISO 4037-2: 1997, X and gamma reference radiation for calibrating dosimeters and dose rate meters and for
determining their response as a function of photon energy—Part 2: Dosimetry for radiation protection over
the energy ranges from 8 keV to 1,3 MeV and 4 MeV to 9 MeV.
ISO 7503-1: 1988, Evaluation of surface contamination—Part 1: Beta-emitters (maximum beta energy
greater than 0.15 MeV) and alpha-emitters.
U.S. National Bureau of Standards. NBS Special Publication 250-16, Calibration of X-Ray and Gamma-Ray
Measuring Instruments.5
Definitions, symbols, and abbreviations herein are as used in this standard and are applicable to G-M
counters.
3.1 Definitions
3.1.1 avalanche: The cumulative process in which charged particles accelerated by an electric field produce
additional charged particles through collision with neutral gas molecules or atoms. It is therefore a cascade
multiplication of ions. (IEEE Std 100-1996)
3.1.2 background counts: Counts caused by ionizing radiation coming from sources other than that to be
measured and by any electronic disturbance in the circuitry that is used to record the counts. (IEEE Std 100-
1996)
3.1.4 check source: A radioactivity source, not necessarily calibrated, that is used to confirm the continuing
satisfactory operation of an instrument. (IEEE Std 100-1996)
3.1.5 count: A single response of the counting system. See also: tube count. (IEEE Std 100-1996)
2IEC publications are available from the Sales Department of the International Electrotechnical Commission, Case Postale 131, 3, rue
de Varembé, CH-1211, Genève 20, Switzerland/Suisse (http://www.iec.ch/). IEC publications are also available in the United States
from the Sales Department, American National Standards Institute, 11 West 42nd Street, 13th Floor, New York, NY 10036, USA.
3IEEE publications are available from the Institute of Electrical and Electronics Engineers, 445 Hoes Lane, P.O. Box 1331, Piscataway,
NJ 08855-1331, USA (http://www.standards.ieee.org/).
4ISO publications are available from the ISO Central Secretariat, Case Postale 56, 1 rue de Varembé, CH-1211, Genève 20, Switzer-
land/Suisse (http://www.iso.ch/). ISO publications are also available in the United States from the Sales Department, American
National Standards Institute, 11 West 42nd Street, 13th Floor, New York, NY 10036, USA (http://www.ansi.org/).
5NBS publications are available from the Superintendent of Documents, U.S. Government Printing Office, P.O. Box 37082, Washing-
ton, DC 20013-7082, USA.
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
FOR GEIGER-MUELLER COUNTERS IEEE Std 309-1999/ANSI N42.3-1999
3.1.6 counter tube: (1) (externally quenched): A radiation-counter tube that requires the use of an external
quenching circuit to inhibit reignition. (2) (gas-filled): A gas tube used for detection of radiation by means
of gas ionization. (3) (gas-flow): A radiation-counter tube in which an appropriate gas-fill concentration is
maintained by a flow of gas through the tube. (4) (Geiger-Mueller): A radiation-counter tube operated in the
Geiger-Mueller region. (5) (self-quenched): A radiation-counter tube in which reignition of the discharge is
inhibited by internal processes. (IEEE Std 100-1996)
3.1.7 counting efficiency: The ratio of the number of observed counts to the total number of ionizing parti-
cles impinging upon the counter surface when the counting rate is so low that dead-time correction is
unnecessary.
3.1.8 counting-rate-versus-voltage characteristic: The counting rate as a function of applied voltage for a
given constant average intensity of radiation. (IEEE Std 100-1996)
3.1.9 dead time: The time interval after the start of an essentially full-amplitude pulse during which a radia-
tion detector is insensitive to further ionizing events. See also: recovery time. (See Figure 4)
3.1.10 differentiator: A high-pass network, usually comprising a series capacitor and a shunt resistor, for
the purpose of reducing the duration of a signal.
3.1.13 gas multiplication factor: The ratio of the charge collected from the sensitive volume to the charge
produced in that volume by the initial ionizing event. (IEEE Std 100-1996)
3.1.14 Geiger-Mueller counter tube: A radiation-counter tube designed to operate in the Geiger-Mueller
region. (IEEE Std 100-1996) (See 3.17.)
3.1.15 Geiger-Mueller region: The range of applied voltage in which the charge collected per isolated
count is independent of the charge liberated by the initial ionizing event. (IEEE Std 100-1996)
3.1.16 Geiger-Mueller threshold: The lowest applied voltage at which the charge collected per isolated
tube count is substantially independent of the nature of the initial ionizing event. (IEEE Std 100-1996) (See
Figure 1.)
3.1.17 half-amplitude recovery time: The time interval from the start of a full-amplitude pulse to the
instant a succeeding pulse can attain an amplitude of 50% of the maximum amplitude of a full-amplitude
pulse. (IEEE Std 100-1996) (See Figure 4.)
3.1.18 initial ionizing event: An ionizing radiation interaction event that initiates a tube count.
3.1.19 ionizing radiation: Particles or photons of sufficient energy to produce ionization in interactions with
matter. [Note that here the usual meaning of radiation (based on IEEE Std 100-1996) has been extended to
include moving nuclear particles, charged or uncharged.]
3.1.20 multiple tube counts: Spurious counts induced by previous tube counts. (IEEE Std 100-1996)
3.1.21 plateau: The portion of the counting-rate-versus-voltage characteristic in which the counting rate is
substantially independent of the applied voltage. (IEEE Std 100-1996) (See Figure 1.)
3.1.22 plateau length: The range of applied voltage over which the plateau of a radiation-counter tube
extends. (IEEE Std 100-1996) (See Figure 1.)
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
IEEE Std 309-1999/ANSI N42.3-1999 IEEE STANDARD TEST PROCEDURES AND BASES
3.1.23 plateau slope: The slope of the plateau of a gas-filled counter tube expressed as the percentage
change in counting rate per 100 V change in applied voltage. Alternatively, the slope can be expressed in per-
cent counting rate per volt change in applied voltage. (See Figure 1.)
3.1.24 pulse-height discriminator: A circuit that responds to pulses above a designated amplitude and gen-
erates a pulse to feed into a counting circuit but that does not respond to lower amplitude pulses.
3.1.26 radial sensitivity: The counting rate of a Geiger-Mueller counter as a function of radial position
across the window of an end-window or pancake Geiger-Mueller counter. (See Figure 3.)
3.1.28 radioactivity standard source: A radioactivity source that has been certified as to absolute radioac-
tivity either by (a) the laboratory recognized as the National Standardizing Laboratory of the country for
radioactivity measurements, NSLR, (NIST in the case of the U.S.) or (b) by a supplier who participates in
measurement assurance activities with the National Standardizing Laboratory when such standards are avail-
able. In such measurement assurance activities, the radioactivity calibration value of the supplier shall agree
with the National Standardizing Laboratory value within the overall uncertainty stated by the supplier in its
verification of the same batch of sources or in its certification of similar sources. Synonym: calibrated
source.
3.1.29 recovery time: The minimum time from the start of a counted pulse to the instant a succeeding pulse
can attain a specified percentage of the maximum amplitude of the counted pulse. (IEEE Std 100-1996)
See also: half-amplitude recovery time.
3.1.30 reignition: A process by which multiple counts are generated within a counter tube by atoms or mol-
ecules excited or ionized in the discharge accompanying a tube count. (IEEE Std 100-1996)
3.1.31 resolving time: The time from the start of a counted pulse to the instant a succeeding pulse can
assume the minimum strength to be detected by the counting circuit. Note: This quantity pertains to the com-
bination of tube and recording circuit. (IEEE Std 100-1996) See also: half-amplitude recovery time.
3.1.32 rise time: The interval between the instants at which the instantaneous value first reaches specified
lower and upper limits, namely 10 and 90% of the peak pulse value. (IEEE Std 100-1996) (See Figure 4.)
3.1.33 sensitive volume: That portion of the radiation counter gas volume having sufficient potential gradi-
ent to operate in the Geiger-Mueller region.
3.1.37 spurious count: A count caused by any event other than the passage into or through the counter tube
of the ionizing radiation to which it is sensitive. (IEEE Std 100-1996)
3.1.38 tube count: A terminated discharge produced by an ionizing event in a radiation-counter tube. (IEEE
Std 100-1996)
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
FOR GEIGER-MUELLER COUNTERS IEEE Std 309-1999/ANSI N42.3-1999
3.1.39 U.S. National Institute of Standards and Technology (NIST): Formerly the National Bureau of
Standards (NBS), NIST is the U.S. federal agency with legally mandated authority to maintain U.S. national
physical standards for the activity of radionuclides.
4. General
In a G-M counter, the electrons produced by the initial ionizing event gain enough energy in their travel to
the anode wire to produce secondary ionization (avalanches). The avalanches produce photons, some of
which will initiate additional avalanches. At some critical voltage (namely, the transition voltage) avalanche
generation becomes self-sustaining, which can paralyze the counter. To prevent this, the avalanche build-up
needs to be quenched. Depending on the filling mixture and operating parameters, tubes can be self or exter-
nally quenched. External quenching can be obtained by circuits that momentarily reduce the operating volt-
age, but the simplest way is to make the anode load resistor sufficiently large, –2 to 10 MΩ. The optimum
resistor value is a compromise between the effectiveness of quenching and the recovery time.
There are four general kinds of G-M counters in common use, identified by their quenching characteristics
and filling gases. They are as follows:
a) Self-quenched organic
b) Resistor-quenched organic
c) Self-quenched halogen
d) Resistor-quenched halogen
The resistor-quenched halogen tube is self-quenching from just above the G-M threshold to a transition volt-
age, above which the quenching depends on having in place a large anode resistor that is ≥100 MΩ.
The performance of a G-M tube is strongly dependent on its operating parameters. The manufacturer shall
furnish the user with all of the information necessary to duplicate the published performance figures, includ-
ing the type of circuitry connected to the counter output and the value of the anode resistor as well as the test
conditions. (For the resistor-quenched counters it is easier to duplicate the manufacturer’s test results if both
employ test circuits having a current- or charge-sensitive input. This minimizes the effects of differences in
cable and input capacitances.)
The counter characteristic is the output counting rate as a function of applied voltage as shown in Figure 1.
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
IEEE Std 309-1999/ANSI N42.3-1999 IEEE STANDARD TEST PROCEDURES AND BASES
5. Test conditions
During the tests, the test area shall be free of all radiation sources other than the minimal background and the
source being used for the test process.
Unless otherwise stated, all characteristics shall be at an ambient temperature of 22˚C ± 2˚C.
Some measurements on G-M counters require the use of a calibrated source; whereas for other measure-
ments, such as plateau determinations, calibrated sources are not needed. For determination of counting rate
as a function of dose rate or fluence rate (as applicable), a calibrated source shall be used and the accuracy of
the calibration shall be stated. In all cases, the sources used shall be identified.
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
FOR GEIGER-MUELLER COUNTERS IEEE Std 309-1999/ANSI N42.3-1999
specific isotope(s) shall be identified. Similarly, X-ray machines may be used if beam quality parameters
(e.g., kilovoltage, filtration, half-value layer) are specified.
A collimated alpha source (ISO 7503-1: 1988) may be used for probing the thin, (0.3 mg/cm2) typically
mica window (e.g., “pancake” or end-window type) of G-M counters to determine their radial sensitivity.
A collimated beta source, such as 90Sr+90Y, may be used for radial sensitivity of cylindrical, thin-wall beta-
gamma counters. (See also ISO 4037.)
A small source of natural or depleted uranium covered with approximately 20 mg/cm2 of aluminum is useful
for checking counters that have thin mica windows or thin windows of another material. This source is espe-
cially suitable for comparing the operating characteristics (e.g., plateau length and slope) of thin window
counters as no correction needs to be made for source absorption due to variations in window thickness.
When absolute G-M counter detection efficiency is measured, calibrated alpha and beta sources shall be
used (e.g., 230Th and 99Tc). Sources shall have single principal emitters, with any secondary emitters pro-
ducing fewer than 3% of the measured total counts. (See ISO 7503-1: 1988.)
Thin-walled and end-window G-M counters are often used for contamination monitoring. Two measure-
ments are frequently performed. The first measurement is the response to point sources of radiation, i.e., the
detector is positioned at a defined distance (typically 3 mm) from a reference source of alpha or beta radia-
tion that has a diameter or maximum dimension that is small compared to the diameter of the detector. The
response is expressed as either
— The “d-mm efficiency” (where d is the source to detector distance), i.e., the probability of a particle
that is emitted from the source being detected, or
— The ratio of the counting rate per unit time per Bq.
The second measurement is the response to large area sources, i.e., sources that are significantly larger than
the maximum dimension of the detector window. In this case, the response of the detector is expressed as
either
— The counting rate per unit emission per cm2 from the source at the defined source to detector separa-
tion, or
— The counting rate per unit time per Bq. Suitable nuclides are the beta emitters 90Sr+90Y and 14C and
the alpha nuclide 238Pu. (See ISO 7503-1: 1988.)
The output of the G-M tube is coupled to a signal sensor. In most cases this will be a voltage-sensitive
preamplifier or a univibrator. Either the anode or the cathode may be used for signal output, but the anode
has the advantage of lower stray capacitance to surroundings. Also, either terminal may be at high voltage
(HV). If the output terminal is at HV and the preamplifier is at ground potential, a coupling capacitor, Cc,
shall be used.
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
IEEE Std 309-1999/ANSI N42.3-1999 IEEE STANDARD TEST PROCEDURES AND BASES
In part (a) of Figure 2, the anode is at positive HV and is coupled to the power supply through the load resis-
tor, RL (2 to 10 MΩ as recommended by the manufacturer). As shown, the load resistor is on the sensor side
of the connecting cable. It is also possible to incorporate RL with the G-M tube, but this requires an external
terminal for the package—one for the signal and one for the power supply. Also in part (a) of Figure 2, an
ammeter is shown connected in series with the nominally grounded cathode. A bypass capacitor, Cbp, shall
be mounted at the G-M tube to provide a short path to ground for the signal current.
In part (b) of Figure 2, the anode is at dc ground and the cathode is at negative HV. This arrangement allows
the preamplifier to be dc coupled to the anode, eliminating the need for a coupling capacitor. Note that the
ammeter and the power have been interchanged with respect to their locations in part (a) of Figure 2.
In both configurations, Ct represents the capacitance across the G-M tube (usually the stray capacitance).
To permit duplication of test conditions for comparing test measurements, the following information shall be
provided:
a) The values of the components and their layout. These shall include all resistors and capacitors con-
nected to the G-M tube, either directly or through a connecting wire. If a wire is used, its length shall
be given.
b) A statement as to which components are integral to the package and which are external to it.
c) Input resistance and capacitance of the sensor used for the tests.
d) Cable length and type (RG number) used in the tests, and whether the cable is part of the assembly
or is external to it.
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
FOR GEIGER-MUELLER COUNTERS IEEE Std 309-1999/ANSI N42.3-1999
e) Information about the electronic instrumentation connected to the G-M counter output. The follow-
ing items should be included: gain rise time, dynamic range, test input capacitance and its terminat-
ing resistance, types of connectors, power requirements, and circuit diagram. If it is a commercially
available unit, include the manufacturer’s model number.
f) Operating voltage. The direct-current operating voltage shall be stated with the specification of
counter characteristics. The term operating voltage, as used here, is the supply voltage, and thus is
the voltage across the load-resistor/counter combination.
Measurements shall be based on counts of at least 10 000 total accumulated counts and shall be corrected for
background. The accumulated counts in the background test shall not be less than the square root of the total
counts accumulated with the sample in place.
The background measurement shall be made with the counter totally enclosed in a nuclear radiation shield.
This shield shall consist of an outer layer of lead at least 50 mm thick and an inner liner of at least 3 mm of
aluminum. The maximum background counting rate within the shield shall be specified. Typical
“unshielded” background may also be specified for the test location. (See ISO 4037.)
The counter response to ionizing radiation will depend on the cathode and/or window design, the type and
energy of the ionizing radiation, and geometric factors that are discussed in this subclause. The sources will
depend upon the applications for which the counter is designed (e.g., radiation protection surveys, industrial
gauging, etc.). The radioisotopes used for evaluation shall be stated. For counters designed to detect the
intermediate range of gamma-ray energies, one of the sources shall be 137Cs or 60Co.
For a stated source-detector geometry and each value of ionizing radiation dose or photon fluence, the cur-
rent and/or counting rate shall be recorded or plotted at a stated operating voltage. This shall be done at a
counting rate for which the product of the dead time (see 7.2.2) and counting rate is less than 0.05.
Gamma-ray measurements should be performed using facilities that conform to the recommendations in
ISO 4037 (i.e., facilities that generate a collimated beam in a low scatter environment). Alternatively and
equivalently, a convenient test arrangement that may be used for gamma-ray measurements is one that
employs an unshielded and collimated source placed at a fixed distance from the counter; the distance shall
be at least five times greater than the largest dimension of the sensitive area of the counter. The counter
should be mounted on a rigid jig made of low Z material for test and calibration and should be isolated (>5.0
m) from adjoining materials (e.g., walls, shield blocks, electronics, etc.) to minimize the effect of scatter on
measurements. Counter orientation with respect to the source-detector axis shall be stated. Gamma-ray
measurements on a G-M counter should be performed over a range of dose rates (µGy/h) to observe any
non-linearity in response at higher dose rates [e.g., cpm/(µGy h -1)].
G-M counters are often calibrated by end-users for response (e.g., cpm) to a known photon dose rate in air
(µGy/h) with a 137Cs gamma source (i.e., 662 keV). However, since measurements at other photon energies
are most likely to be made, it is important to know the photon energy response of both bare and compensated
type G-M tubes. Depending on the construction of the tube and the intended use, it may be useful to know
the energy response characteristics ranging from 10 keV to several MeV. All data for relative photon energy
response measurements shall be normalized to 137Cs. All data or plotted curves shall be stated as a relative
energy response ratio, i.e., the product of the observed alternate energy response in, for example,
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
IEEE Std 309-1999/ANSI N42.3-1999 IEEE STANDARD TEST PROCEDURES AND BASES
cpm/(µGy h -1), divided by the 137Cs response with the same units. Similarly, polar response at a stated
energy should be normalized to a recommended calibration geometry, with the G-M counter then rotated
360˚ in an incident-parallel photon beam to obtain the relative response. Calibrated sources or ion chambers
shall be used in these tests.
For alpha-, beta-, and X-ray measurements, any geometric factors that can affect the evaluation of the
counter response by more than 3% shall be described. These factors include
The G-M counter shall be connected to an oscilloscope and documented test circuit, and shall be placed in a
radiation field such that the dead-time counting-rate product is 0.05 or less at its rated operating voltage. The
voltage is gradually raised from well below the G-M threshold until pulses appear on the oscilloscope. The
G-M threshold is that voltage at which nearly all pulses are within 95% of their average amplitude. See
Figure 1.
With the counter in a radiation field, a curve shall be plotted of counting rate vs. counter-tube voltage as
shown in Figure 1. The circuit parameters and the input sensitivity shall be stated.
The operating plateau is that portion of the curve that is substantially flat or uniform in slope. The plateau
length from V1 to V2 of Figure 1 is expressed in volts and is the region over which the slope is calculated. It
is quantitatively determined by drawing two lines parallel to the plateau, but separated from the plateau by
±5% of the counting rate at the center of the region (Nm). The ends of the plateau are defined by the intersec-
tions of the lines with the curve (V1 and V2 in Figure 1).
The plateau slope S in units of percent change per 100 V shall be calculated from the following formula
where V1 and V2 are as in Figure 1:
6.5 Life
The life of a G-M counter is defined in terms of the total accumulated counts before specified parameters
change by a specified amount. The counter shall be operated at its specified operating voltage in a radiation
field. The circuit and test conditions (i.e., temperature) used shall be documented (see 5.4). To conserve test
time, it is recommended that the counting rate be high, but not more than 1/10 of the reciprocal of the dead
time (see 7.2.2). When specifying tube life, the extent of change of the following specified parameters shall
be stated:
a) G-M threshold
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
FOR GEIGER-MUELLER COUNTERS IEEE Std 309-1999/ANSI N42.3-1999
b) Plateau length
c) Plateau slope
d) Background counting rate
A collimated alpha or beta source may be used to determine the sensitivity along a diameter of an end-
window counter (typically a mica-window counter). The better the collimation, the more accurate is the
measurement. If a beta source is used, the source holder shall be such that radiation cannot escape from its
sides. The diameter of the collimator hole shall be not larger than 1/5 of its length nor 1/10 of the diameter of
the counter window. The source shall be placed as close to the window as practical and then moved across
the window along a diameter, then rotated 90˚ and the measurements repeated. (See Figure 3.) Counting rate
shall be plotted as a function of radial position. The source and collimator geometry shall be stated.
In cylindrical counters, the electrical field is distorted near each end. The resulting longitudinal dependence
of sensitivity can be investigated in a manner similar to that described in 6.6. Counting rate shall be plotted
as a function of longitudinal position. The source and collimator geometry shall be stated.
Varying the ambient temperature from the normal room temperature will affect the various parameters of a
G-M counter. The amount change will depend on the type of counter and the temperature difference.
The manufacturer shall specify the operating temperature limits and the change of each affected parameter.
If a counter parameter change is approximately linear over the temperature range specified, the change may
be expressed as a temperature coefficient. However, if that is not the case, a plot of the change vs. tempera-
ture shall be provided.
6.9 Hysteresis
Some G-M counters exhibit a voltage-hysteresis effect. Usually this is encountered when the HV is initially
applied and when determining a counting-rate-versus-voltage characteristic (see 6.4). There can be a differ-
ence between the plateau obtained with an increasing-voltage traverse and that obtained with a decreasing-
voltage traverse. Hysteresis may also appear if a plateau is traversed and the voltage is then set at the operat-
ing voltage. This effect is attributed to the charging of insulating surfaces in the counter. Large area G-M
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
IEEE Std 309-1999/ANSI N42.3-1999 IEEE STANDARD TEST PROCEDURES AND BASES
counters with windows made of a dielectric such as mica or thin plastic film are especially vulnerable to hys-
teresis. Repeated counting-rate measurements of a source at a fixed-tube voltage should be made to deter-
mine if the counter has stabilized. To test for hysteresis, the plateau shall be traversed in an ascending and
then in a descending direction with a total elapsed time of less than one hour (much shorter than the charge-
leakage time constants of the insulating surfaces). The ascending and descending plateau curves shall be
shown. Since hysteresis may reduce the G-M counter’s sensitive volume, counter efficiency for alpha and
beta radiation shall be determined once repetitive counts differ by no more than the expected statistical vari-
ation at that counting rate.
6.10 Photosensitivity
Photosensitivity is indicated by a change in the background counting rate of the counter to light (the win-
dows are usually made opaque). Halogen-quenched counters normally do not exhibit such sensitivity. Photo-
sensitivity shall be tested by measuring the background counting rate, first with the counter operating in the
dark and then with the counter exposed to a stated illumination.
The counter is connected in a documented circuit. (See Figure 2.) (Note that a circuit without a coupling
capacitor between the G-M counter and the electronic input may be useful.) The manufacturer shall specify
the value of the anode resistor to be used, and all tests shall be made with an anode resistor of that value. The
characteristics to be measured are listed in this clause. These characteristics are dependent on circuitry,
counter-tube voltage, and counting rate.
G-M counters can be operated in the current mode using a microammeter. This standard does not address
that mode of operation.
7.2 Pulse-shape
Figure 4 illustrates the defined terms that follow in 7.2.1, 7.2.2, 7.2.3, and 7.2.4. Part (a) of Figure 4 is a
drawing made from an oscilloscope picture of the voltage pulse from a typical G-M counter. Since, in gen-
eral, the rise time is much shorter than the dead time, it is shown on a different time scale in part (b) of Fig-
ure 4.
The rise time of a G-M counter pulse is defined as the time interval between 10% and 90% of peak ampli-
tude. The actual circuit used (with component values), voltage on the G-M counter, and the circuit rise time
shall be recorded.
The dead time is defined as the time interval after the start of an essentially full-amplitude pulse during
which a radiation detector is insensitive to further ionizing events [see part (a) of Figure 4].
Circuit parameters, counter-tube voltage, and counting rate shall be recorded along with the dead time. A
curve of dead time vs. voltage is very useful A differentiator shall be used in the signal path. The time con-
stant of the network shall be such that the pulse decays to less than 5% of its maximum value before the suc-
ceeding pulse appears as observed on the oscilloscope screen (see also 7.2.4). It is usual to use a radioactive
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
FOR GEIGER-MUELLER COUNTERS IEEE Std 309-1999/ANSI N42.3-1999
check source for this purpose, but an alternative is a double-pulse X-ray source with adjustable spacing
between the pulses of a pair [see IEC 60151-25 (1971-01)].
The half-amplitude recovery time (see Figure 4) is defined as the time interval from the start of a full-
ampltude pulse to the instant a succeeding pulse can attain an amplitude of 50% of the maximum amplitude
of a full-amplitude pulse. Circuit parameters, counter voltage, and counting rate shall be stated along with
the half-amplitude recovery time. (See also 7.2.4.)
An alternative method for determining dead time or recovery time is the “two-source method” (see ANSI
N42.12-1994 and Knoll [B4]7). In this method, three counting rates are measured using two sources from
the same radioisotope—the first from Source 1, the second from Source 2, and the third from Sources 1 and
2 simultaneously. Additionally, the background counting rate is determined with no sources present.
The counting system discriminator is set at 10% of the full-pulse amplitude for the dead-time measurement
and at 50% for the 50% recovery-time measurement. From the measurements, the respective counting rates
N1, N2, N12 and Nb shall be obtained and the dead time and recovery time determined using the following
equations:
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
IEEE Std 309-1999/ANSI N42.3-1999 IEEE STANDARD TEST PROCEDURES AND BASES
When making these measurements, it is important not to disturb Source 1 when introducing Source 2 and,
similarly, when removing Source 1, not to disturb Source 2. For multichannel analyzer (MCA) systems, the
“live-time” feature is designed to compensate for counting time lost during pulse processing, and further
correction for dead-time and recovery-time losses is usually not required. (At the counting rates achieved
with G-M counters, any loss due to modern MCAs is none-the-less essentially insignificant.)
When either Source 1 or Source 2 is not in use, the source that is not in use shall be replaced by a blank so
that the scattering geometry is unchanged.
In the two-source method, the dead time is computed from the difference between two large quantities. The
statistical fluctuations of each of the counts add quadratically when computing the difference. Therefore,
very long counting times may be necessary for each of the counts if the error in the difference count is to be
kept to a reasonable level.
8. Test equipment
The equipment listed in this clause is typically used in G-M counter tests. It is neither all-inclusive nor
restrictive and does not preclude the use of alternative test instrumentation.
8.1 Voltmeter
To separately measure the voltage across the counter tube and that across the anode resistor, a high imped-
ance meter, such as a very high input-resistance digital voltmeter (DVM), is required. However, in general,
the voltage of interest, such as the operating voltage as defined in this standard (see 5.6), is the voltage across
the anode-resistor/counter combination, (i.e., the voltage provided by the power supply) and thus any accu-
rate voltmeter with reasonable input resistance will suffice.
8.2 Ammeter
To measure the current through the G-M counter tube, a microammeter shall be placed in series with it,
preferably on the grounded side. Alternatively, a suitably isolated meter may be placed in series with the
connection to the high-voltage supply. In either case, a capacitor should be connected across the meter to
provide a low-impedance path for the signal. A battery-operated DVM can be used in ammeter mode, if
available, or by measuring the voltage drop across a known resistance, with the drop not exceeding one volt.
The make and model of the meter shall be stated.
8.3 Oscilloscope
The oscilloscope that is used to view waveforms shall have a rise time of 0.1 µs or less. Where feasible, dc
coupling shall be used. The make and model of the oscilloscope and its loading on the circuit shall be stated
if such loading affects the accuracy of the measurements.
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
FOR GEIGER-MUELLER COUNTERS IEEE Std 309-1999/ANSI N42.3-1999
The pulse pair resolution of the counting system shall be less than the dead time of the G-M counter that is
being tested. The polarity of the input shall be considered as well as the sensitivity, and the system shall be
capable of a wide dynamic range of input pulse amplitudes to obtain satisfactory plateau data. The system
shall be capable of clearly resolving the individual input pulses and shall have a discriminating range suffi-
cient to permit determination of the dead time and the 50% resolving time. The input impedance shall be
defined and shall not be less than the load resistance, RL.
8.5 Scaler
A digital scaler with a pulse resolving time less than that of the G-M tube and with a digital display that has
a capacity of at least 100 000 counts shall be used for measurement of the G-M tube characteristics. It is con-
venient for the scaler to have selectable preset-time or total-count features.
A high-voltage supply that will cover the voltage range of interest for the counter tested is necessary. In
some cases, the supply may be an integral part of the scaler or counting-rate meter. The supply has to be
capable of delivering the maximum tube current demanded and do so without overloading or loss of regula-
tion. The regulation against line and load variations should be 1% or better. The ripple should be less than
100 mV. The output resistance shall not exceed 2% of that of the anode resistor.
A general-purpose laboratory capacitance bridge can be used to measure the capacitance of G-M tubes.
Most G-M tubes have a capacitance of 10 pF or less and the parasitic capacitance to ground of other compo-
nents connected to the output terminal can be expected to add 5 to 15 pF (including connectors). The signal
cable will add to the parasitic capacitance to the extent of about 50 pF/m for a cable with a characteristic
impedance of 100 Ω, increasing to about 100 pF/m for a 50-ohm cable.
A single-channel analyzer shall be capable of clearly resolving the individual input pulses and shall have a
discriminating range sufficient to permit determination of the dead time and the 50% resolving time.
The materials (cathode, cathode plating thickness, liner, anode, window, gas-filling, etc.) of which a G-M
counter is constructed shall be accurately specified. Accurate mechanical drawings shall be made available
by the manufacturer. The thickness of the window or wall (as applicable), together with the window or wall
material, shall be stated by the manufacturer. (Window thickness is typically given in mg/cm2.) (See 6.6 and
6.7.)
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
IEEE Std 309-1999/ANSI N42.3-1999 IEEE STANDARD TEST PROCEDURES AND BASES
To facilitate interchangeability of G-M counter tubes and probes, standard bases and electrode connections
shall be as follows:
1 Cathode
2 Anode
3 No connection
1 No connection
2 Anode
3 No connection
4 Cathode
Single-pin connectors (e.g., BNC, C, MHV, SHV, and others) on G-M counter tubes or probes shall be as
specified by the manufacturer or purchaser.
8JETEC-A3-1 is included in Electron Industries Association Standard EIA 209A, “Standards for Electron Tubes.” The Joint Electron
Tube Engineering Council (JETEC) has been superseded by the Joint Electron Devices Engineering Council (JEDEC).
9See Footnote 8.
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
FOR GEIGER-MUELLER COUNTERS IEEE Std 309-1999/ANSI N42.3-1999
Annex A
(informative)
Bibliography
[B1] Centronic Databook, Geiger-Mueller Tubes, Centronic Ltd., New Addingdon, Croydon, Surrey,
England, Issue No. 1 (1997).
[B2] Dmitriev, A.B., Filatov, A.I., and Eig, L.S., “Performance of halogen counters and some points on their
operation, instruments, and experimental techniques” (in Russian), Prib. Techn. Eksp, pp. 157–161, Sept.–
Oct. 1967; also English translation in Consultants Bureau, New York.
[B3] Fava, L. and Sanguist, V. “Dead time measurements of G-M counters,” Nuclear Instruments and Meth-
ods, vol 42, pp. 325–326, 1966.
[B4] Knoll, G.F., Radiation Detection and Measurement, 2nd ed., New York:Wiley, 1989.10
[B5] Polly, P., “A method for the evaluation of the dose-sensitivity of G-M counters as a function of the
energy,” Nuclear Instruments and Methods, vol 50, pp. 197–208, 1967.
[B6] Price, W.J., Nuclear Radiation Detectors, 2nd ed. New York: McGraw Hill, 1964.
[B7] Radiological Health Handbook, U.S. Dept. of Commerce, Office of Technical Services, PB 121784R,
Sept. 1960.11
[B8] Radiological Health Handbook, U.S. Dept. of Health, Education, and Welfare, pub. no. 2016, Jan.
1970.
[B9] Sharpe, J., Nuclear Radiation Detectors, 2nd ed. New York: Wiley, 1963.
[B10] van Duren, K., “Electrical characteristics of halogen filled Geiger counters,” Vide, no. 95, pp. 235–
248, Sept.–Oct. 1961.
The following publications are listed here because of their historical significance. Such a listing is obviously
incomplete, but the listed publications contain references to additional and earlier publications.
[B11] Corson, D.R. and Wilson, R.R., “Particle and quantum counters,” Review of Scientific Instruments,
vol. 19, pp. 207–233, April 1948.
[B12] Curran, S.C. and Craggs, J.D., Counting Tubes, New York: Academic Press, 1949.
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.
IEEE Std 309-1999/ANSI N42.3-1999
[B13] Curtiss, L.F. “The Geiger-Mueller counter,” National Bureau of Standards, Washington, DC, Circ.
490 Jan. 1950.
[B14] van Duuren, K., Jaspers, A.J.M., and Hermsen, J., “G-M counters,” Nucleonics, vol. 17, pp. 86–94,
June 1959.
[B15] Evans, R.D., “Radioactivity units and standards,” Nucleonics, vol. 1, pp. 32–43, Oct. 1947.
[B17] Friedman, H., “Geiger counter tubes,” Proc. IRE, vol. 37, pp. 791–808, July 1949.
[B19] Korff, S.A., Electron and Nuclear Counters, Van Nostrand, 1946.
[B20] Korff, S.A., Electron and Nuclear Counters, Van Nostrand, 1955.
[B21] Kelley, G.G., Jordon, W.H., and Bell, P.R., “Electron component in Geiger discharge,” Review of Sci-
entific Instruments, vol. 21, pp. 330–332, April 1950.
[B22] Rossi, B.B. and Staub, H.H., “Ionization chambers and counters,” National Nuclear Energy Series,
div. 5, vol. 2, New York: McGraw Hill, 1949.
[B23] Wilkinson, D.H., Ionization chambers and counters, New York: Cambridge University Press, 1950.
[B24] “Gas discharge mechanism of halogen-quenched counters,” Applied Science Research, sec. B, vol. 4,
pp. 237–248, 1955.
[B25] “Double pulses in gas-halogen Geiger counters,” Applied Science Research, sec. B, vol. 5, pp. 368–
386, 1956.
[B26] van Zoonen, D. and Prast, G. Jr., “Properties of argon-bromide counters,” Applied Science Research,
sec. B, vol. 3, pp. 1–17, 1952.
[B27] van Zoonen, D., “The delay in the build-up of halogen-quenched counters,” Applied Science Research,
sec. B, vol. 3, pp. 377–389, 1953.
Authorized licensed use limited to: IEEE Xplore. Downloaded on April 18, 2011 at 01:06 from IEEE Xplore. Restrictions apply.