0% found this document useful (0 votes)
77 views4 pages

Omniport 30

The Omniport 30 is a robust, multifunctional hand-held device with a touch screen display that can simultaneously show up to three measurement values. It can measure 23 different parameters including humidity, temperature, dew point, air velocity, and more using various probes. The device stores measurements internally and includes software for viewing and managing data on a PC. Probes are sold separately to measure parameters in various applications like HVAC systems.

Uploaded by

Bams Arifin
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
77 views4 pages

Omniport 30

The Omniport 30 is a robust, multifunctional hand-held device with a touch screen display that can simultaneously show up to three measurement values. It can measure 23 different parameters including humidity, temperature, dew point, air velocity, and more using various probes. The device stores measurements internally and includes software for viewing and managing data on a PC. Probes are sold separately to measure parameters in various applications like HVAC systems.

Uploaded by

Bams Arifin
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

OMNIPORT 30 Multifunctional Hand-Held

The robust multifunctional hand-held meets the


highest requirements and comes with a wide range
OMNIPORT 30
of accurate probes that fit various applications.
Use touch-screen navigation to show, up to three
measurement values simultaneously on the capacitive
TFT display.

A total of 23 measurands (vary according to probe) is


available, including:
• relative humidity RH
• temperature T
• dew point temperature Td
• absolute humidity dv
• mixing ratio r
• air velocity v
• volumetric flow with RH/T probe with CO2 probe
• air pressure p
• CO2 ppm

Store measurement values in the internal memory of the device and use the free SmartGraph3 software to
manage data on your PC. The optional carrying case accommodates the hand-held and probes, as well as
accessories.

Data-Logging & Data-Management


The Omniport 30 offers both continuous and single-point data-
logging. The measurement values of up to three channels are
stored simultaneously, together with a time and date stamp, in
the internal memory of the device.

Stored measurement data, as well as the minimum, maximum,


average and standard deviation values can be shown directly on
the display. Scroll through a recorded measurement and switch
between graphs by using the control cross.

Data management is easy and intuitive by using the free SmartGraph3 software to create graphs that contain
measurement channels of interest for better data analysis. Data can be exported in .csv format and then
imported into EXCEL for further processing.

Features
Data logging Real-time HOLD / MIN / MAX / AVG readout
Internal memory for 2 million measured values Integrated air pressure sensor
23 physical quantities User friendly operation
Capacitive TFT touch screen Free data management software
Displays measurands simultaneously

62 v1.5 / Modification rights reserved OMNIPORT 30


Technical Data
General
Power supply 4 x Alkaline LR6 AA batteries, 1.5 V (not in the scope of supply)
Optional power supply 5V DC via USB (cable included)
Temperature range operating: handheld and handle of sensing probe: 0...50°C (32...122°F)
storage: -20...60°C (-4...140°F)
Internal memory for approx. 2 million measured values
Housing / protection class ABS / IP40
Dimensions (HxWxD) 170 x 62 x 34 mm (6.69 x 2.44 x 1.34”)
Weight ca. 205g (0.45 lbs)
Display TFT display, 54 x 41 mm (2.13 x 1.61”), illuminated
CE compatibility Hand-held: EN61000-6-2:2005 EN61000-6-3:2007
Logprobe: EN61326-1:2013 EN61326-2-3:2013
Integrated air pressure sensor
Measuring range 800 to 1100 mbar (complete accuracy)
Accuracy max. ± 0.5 mbar (at 25 °C, 1013.25 mbar)
Long term stability typ. -1 mbar/year
HUMIDITY / TEMPERATURE PROBES
LOGPROBE 20 - compact, pluggable HVAC probe
Working range: 0...100% RH / -40...80°C (-40...176°F)
43 (1.7”) 32 (1.3”)
Accuracy: ±2% RH (0...90% RH), ±3% RH (90...100% RH) @20°C / 68°F
±0.2°C / ±0.36°F @20°C / 68°F
max. ±0.6°C / ±1.08°F (-40...80°C / -40...176°F)
Order code: LOGPROBE20-HTPC
Response time τ : ≤ 30 sec.
90

Filter: Membrane filter


(0.47”)
Ø12

Order code: LOGPROBE20-HTPA


Response time τ ≤ 10 sec.
90

Filter: Plastic grid filter


LOGPROBE 16 - HVAC probe
Working range: 0...100% RH / -20...70°C (-4...158°F)
70 (2.8”) 32 (1.3”)
Accuracy: ±2% RH (0...90% RH), ±3% RH (90...100% RH) @20°C / 68°F
±0.2°C / ±0.36°F @20°C / 68°F
max. ±0.5°C / ±0.9°F (-20...70°C / -4...158°F)
Response time τ : ≤ 7 sec.
(0.47”)
∅12

90

Order code: LOGPROBE16

LOGPROBE 31 - high temperature probe


Working range: 0...100% RH / -40...180°C (-40...356°F)
250 (9.9”) 32 (1.3”)
(grip of sensing probe up to 50°C/122°F)
Accuracy: ±2% RH (0...90% RH), ±3% RH (90...100% RH) @20°C / 68°F
±0.2°C / ±0.36°F @20°C / 68°F
∅12
(0.5”)

max. ±0.6°C / ±1.08°F (-40...180°C / -40...356°F)


Response time τ : ≤ 30 sec.
90

Temperature dependency: RH: ±0.03% RH/°C (% RH/1.8°F)


Order code: LOGPROBE31

LOGPROBE 30 - confined space probe

250 (9.9”)
Working range: 0...100% RH / -40...100°C (-40...212°F)
Accuracy: ±2% RH (0...90% RH), ±3% RH (90...100% RH) @20°C / 68°F
±0.2°C / ±0.36°F @20°C / 68°F
∅5
(0.2”)

max. ±0.6°C / ±1.08°F (-40...100°C / -40...212°F)


Response time τ : ≤ 15 sec.
90

Temperature dependency: RH: ±0.03% RH/°C (% RH/1.8°F)


Order code: LOGPROBE30
OMNIPORT 30 v1.5 / Modification rights reserved 63
AIR VELOCITY PROBES

LOGPROBE 61/60 - stainless steel probe


Working range: 0.08...2m/s (15...400ft/min) 0.2...20m/s (40...4000ft/min)
200 (7.9”) 12 (0.5”)
-20...70°C (-4...158°F) -20...70°C (-4...158°F)
Accuracy: ± (0.04m/s / 8ft/min+ 1% of m. v.) ± (0.2m/s / 39ft/min + 2% of m. v.)
± 0.7°C / ±1.26°F (0...50°C / 32...122°F) ± 0.7°C / ±1.26°F (0...50°C / 32...122°F)

(0.2”)
∅6
Response time τ : ≤ 1.5 sec.
90
≤ 1.5 sec.
Order code: LOGPROBE61 LOGPROBE60

LOGPROBE 65 - polycarbonate probe

200 (7.9”) 18 (0.7”)


Working range: 0.2...20m/s (40...4000ft/min) / 0...50°C (32...122°F)
Accuracy: ± (0.2m/s / 39ft/min + 3% of m.v.)
± 1°C / ±1.8°F (0...50°C / 32...122°F)
Response time τ : ≤ 1.5 sec.
(0.5”)
∅12

90

Order code: LOGPROBE65

CO2 PROBE

LOGPROBE 802/805/810 - CO2 probe

96 (3.8”)
Working range: 0...2000 / 5000 / 10000ppm
Technology: dual wave NDIR with autocalibration
Accuracy at 25°C 0...2000ppm: < ± (50ppm +2% of measured value)
and 1013mbar 0...5000ppm: < ± (50ppm +3% of measured value)
18.5
(0.7“)

(77°F and 14.69psi) 0...10000ppm: < ± (100ppm +5% of measured value)


Warm-up time: 3 min.
Temperature dependency: typ. 1ppm CO2/°C (-20...45°C) (-4...113°F)
Order code: 0...2000ppm LOGPROBE802
0...5000ppm LOGPROBE805
0...10000ppm LOGPROBE810

Carrying Case
The optional carrying case protects the handheld, probes and accessories
during transport and storage. It helps to avoid mechanical damage, as well
as the contamination of the sensors, which is essential for the optimal long
term performance of the instrument.

HA040906
Protective Cover
The optional cover protects the handheld device during usage in dirty or
oily environments. The integrated magnetic plate allows an easy temporary
fixing of the handheld onto a metallic surface on site, while the hang-on
straps helps keep your hands free while taking measurements.

HA040907

64 v1.5 / Modification rights reserved OMNIPORT 30


Ordering Guide
Pluggable Probes 1) Remote Probes 2)
POSITION 1 BASIC DEVICE OMNIPORT30 OMNIPORT30

POSITION 2 CABLE 2m (6.6ft) HA010813

POSITION 3 PROBES
Humidity / Temperature
LOGPROBE 20 - Response time ≤ 30 sec. LOGPROBE20-HTPC
LOGPROBE 20 - Response time ≤ 10 sec. LOGPROBE20-HTPA
LOGPROBE 16 - HVAC probe LOGPROBE16
LOGPROBE 31 - high temperature probe LOGPROBE31
LOGPROBE 30 - confined space probe LOGPROBE30
Air Velocity
LOGPROBE 61 - stainless steel probe / 0.08...2m/s (15...400ft/min) LOGPROBE61
LOGPROBE 60 - stainless steel probe / 0.2...20m/s (40...4000ft/min) LOGPROBE60
LOGPROBE 65 - polycarbonate probe / 0.2...20m/s (40...4000ft/min) LOGPROBE65
CO2
LOGPROBE 802 - 0...2000ppm LOGPROBE802
LOGPROBE 805 - 0...5000ppm LOGPROBE805
LOGPROBE 810 - 0...10000ppm LOGPROBE810
POSITION 4 CARRYING CASE For basic device and up to 4 probes HA040906 HA040906

1) Directly connected to device, cable is optional


2) Cable is necessary

Order Example
Pluggable Probes: Remote Probes:

Position 1 - Basic Device OMNIPORT30 Position 1 - Basic Device OMNIPORT30


Position 2 - Probe LOGPROBE805 Position 2 - Cable HA010813
Position 3 - Carrying Case HA040906 Position 3 - Probes LOGPROBE16
LOGPROBE61
Position 4 - Carrying Case HA040906

Accessories
Carrying case for basic device,
2 pluggable and 2 remote probes HA040906

Protective cover HA040907

Membrane filter PC (for ∅12mm RH/T probes) HA010118


Metal grid filter PC (for ∅12mm RH/T probes) HA010119
Stainless steel sintered filter (for ∅12mm RH/T probe) HA010103

Plastic grid filter PC (for Logprobe 20-HTPA) HA010121

Cable for remote probes 2m (6.6ft) HA010813

Humidity standards / Calibration device refer to data sheet Humidity Calibration Set

SmartGraph 3 - data management software free download at www.epluse.com/smartgraph3

OMNIPORT 30 v1.5 / Modification rights reserved 65