Method Validation 2012 PDF
Method Validation 2012 PDF
control procedures
Vladimír Kocourek
Prague, 2012
WHAT IS VALIDATION?
► Validation is a process, within which the method is demonstrated
to be suitable for its purpose. It documents methods performance !
Applicability
Analytical method for decision making
Defining Selection of
the task method
Sample
handling
Analytical
procedure
measuring
► PRECISION
ACCURACY
► TRUENESS
► RUGGEDNESS
TRUENESS AND PRECISION = ACCURACY
ERRORS OF MEASUREMENT
X = μ + ε + Σδ
RELATED TO RELATED TO
TRUENESS PRECISION
X = μ + ε + Σδ
True value is an idealized concept and „true value“ cannot be
known exactly!
Hence the REFERENCE VALUE represents a true value in routine
practice
Reference value usually provided with reference to:
► Certified reference material
► Reference measurement procedure
► Known amount of analyte added into the sample (spike)
TRUENESS
TRUENESS is closeness of agreement between the mean of
an N number of replicate measured values and a
REFERENCE (TRUE) value.
SAMPLE
RESULT
CRM
(matrix) Compare with
certified value
t- test :
texp < tcrit (P, n-1)
Certified Reference Material (CRM, SRM)
http://www.erm-crm.org/html/homepage.htm
TRUENESS
HOW TO ESTIMATE TRUENESS (RECOVERY)
Sample 1 Value 1
ANALYTICAL PROCESS
Sample 2 Value 2
Sample 3 Value 3
CRM Sample 4 Value 4
Y±y Sample 5 Value 5
n≥6
MEAN VALUE
COMPARISON X±x
STATISTICAL TESTING
Sample 1 Value 1
ANALYTICAL PROCESS
Sample 2 Value 2
INCUBATION
Sample 3 Value 3
SAMPLE Sample 4 Value 4
Sample 5 Value 5
BLANK
n≥6
Addition of known
amount of an analyte MEAN VALUE
to each replicate X±x
(SPIKE)
http://www.erm-crm.org/html/homepage.htm
TRUENESS
HOW TO ESTIMATE TRUENESS (RECOVERY)
MEAN VALUE
RECOVERY (%) = . 100
ADDED AMOUNT
Sample 1 Value 1
Sample 2 Value 2
INCUBATION
Sample 3 Value 3
SAMPLE Sample 4
Sample 5
Value 4
Value 5
LU LU
Ar - BaA
23.152 - 1-MePyr
LU
23.152 - 1-MePyr
03
11
88
7.
9.
3.
99
68
10
ea 6
A22.872
60
ea 2
A22.872
:2
:9
Ar 7.7
:2
Ar 5.3
ea
7
4
23.591
25
21
25.004 - B[a]A
200 200
a:
a:
re
re
50
150 150
40
30 100 100
20
50 50
10
0 0
23.75 24 24.25 24.5 24.75 25 25.25 25.5 25.75 m 21.5 22 22.5 23 23.5 24 24.5 m 21.5 22 22.5 23 23.5 24 24.5 m
6.35
6.07
6.84 Same sample, conditions,
%
7.05
operator,...
7.65
7.54
8.10
9.67
0
5.40 5.60 5.80 6.00 6.20 6.40 6.60 6.80 7.00 7.20 7.40 7.60 7.80 8.00 8.20 8.40 8.60 8.80 9.00 9.20 9.40 9.60 9.80
Two different
chromatographs of the
same type
1 1
(xi – x )2 x= xi
σ = n
n
RESULT VALUES
σ
RSD (%) = 100
x
PRECISION
INCREASING NUMBER OF CONSIDERED RANDOM ERROR SOURCES
Two components:
σ2r – intra-laboratory variance (typical)
σ2L – between laboratories variance
„William Horwitz“ …
PRECISION
REPRODUCIBILITY - HORWITZ
HORWITZ CURVE
H o r w it z o v a k ř iv k a
40
30
CV (%)
20
10
0
1 10 0 10 0 0 0 1 000 000 10 0 0 0 0 0 0 0
<1 35 > 45
1 - 10 30 32 -45
10 - 100 20 22 – 32
3500000
3000000
2
R = 0,9977
2500000
Peak area
2000000
1500000
1000000
500000
0
0 1000 2000 3000 4000 5000
-1
DON concentration (µg kg )
LINEARITY, CALIBRATION AND RANGE
GENERAL RECOMENDATIONS FOR LINEAR CALIBRATION
VALIDATED
1·0 CONCENTRATION
LEVEL
VALIDATED
0·8 RANGE
0·6 VALIDATED
Signal
CONCENTRATION
LEVEL
0·4
0·2
Linear range
0
0 0·2 0·4 0·6 0·8 1·0 1·2
Concentration
LIMIT OF DETECTION / QUANTIFICATION
LIMIT OF DETECTION / QUANTIFICATION
Limit of Detection (LOD): the smallest concentration of analyte
in the test sample which can be reliably distinguished from zero.
►LOD is concentration of analyte which induce signal (S) that
is 3 times higher than the background noise level (N). S/N=3
S/N > 6
S/N > 2
signal
noise
DETECTABILITY – how to lower LOD ?
Increase of S/N ratio in GC/MS-NCI analysis of
nitronaphtalenes (parsley sample) due to increasing
electromultiplyier voltage in detector (1-nitroNAP = 0.1
ppb, 2-nitroNap = 0.08 ppb)
72 100
80
70 60
Time--> 11.80 12.00 12.20 12.40 12.60 Time--> 11.80 12.00 12.20 12.40
Vladimir.Kocourek@vscht.
LIMIT OF DETECTION / QUANTIFICATION
HOW TO ESTIMATE LOD & LOQ
MEASUREMENTS
CALCULATION OF S/N VALUES
LOD → LOQ
Signal at LOD vs. background noise
1,64*sblank 1,64*sblank
=5% =5%
s b la n k
y blank YLOD
Spiked samples
Training
Reference material
Inter-laboratory study
REPEATABILITY TRUENESS
REPRODUCIBILITY
LOD/LOQ Selectivity
Performance
characteristics
Range Ruggedness
Fit-for-purpose
USEFUL DOCUMENTS
USEFUL DOCUMENTS
Criteria for validation of methods used in official control of
contaminants and residues in food and feed:
All documents and methods issued to personnel in the laboratory shall be
reviewed and approved for use by authorized personnel prior to issue.
When mistakes occur in records, each mistake shall be crossed out, not
erased, made illegible or deleted, and the correct value entered alongside.
All such alterations to records shall be signed by the person making the
correction.
In the case of records stored electronically, equivalent measures shall be taken to
avoid loss or change of original data.
The laboratory shall use test methods, including methods for sampling,
which meet the needs of the customer and which are appropriate for the
tests it undertakes.
Methods published in international, regional or national standards shall
preferably be used.
Existing stock and working solutions may be tested against newly prepared solutions
by comparing the detector responses obtained from appropriate dilutions of individual
standards or mixtures of standards.
The means from at least three replicate measurements for each of two solutions (old
and new) should not normally differ by more than ±10 %. The mean from the new
solution is taken to be 100%. Differences in apparent concentration between old and
new standards must be investigated.
Calibration
Responses used to quantify residues must be within the dynamic range of the detector.
Extracts containing high-level analytes may be diluted to bring them within the calibrated
range.
Validation of analytical methods should include determination of recovery at the
proposed reporting limit.
Calibration by interpolation between two levels is acceptable providing the difference
between the 2 levels is not greater than a factor of 4, and where the mean response
factors, derived from replicate determinations at each level, indicate acceptable
linearity of response with the higher being not more than 110 % of the lower response
factor.
Single-level calibration may provide more accurate results than multi-level calibration
if the detector response is variable with time. When single-level calibration is
employed, the sample response should be within ± 20 % of the calibration standard
response if the MRL is exceeded.
The potential for matrix effects to occur should be assessed at method validation.
They are variable in occurrence and intensity but some techniques are particularly
prone to them. If the techniques used are not inherently free from such effects,
calibration should be matrix-matched routinely.
Where a calibration standard is a mixture of isomers, etc., of the analyte, detector
response generally may be assumed to be similar, on a molar basis, for each component.
Contamination & interference
Samples must be separated from each other, and from other sources of
potential contamination, during transit to, and storage at, the laboratory.
Pest control in, or near, the laboratory must be restricted to pesticides that
will not be sought as residues.
The method must be tested for sensitivity, mean recovery (as a measure of
trueness or bias), precision, and limit of quantification (LOQ).
This means that spiked recovery experiments to check the accuracy of the
method should be undertaken.
A minimum of 5 replicates is required at both the reporting limit, and at
least another higher level, perhaps an action level, for example the MRL.
The (method) LOQ is defined as the lowest validated spike level meeting
the method performance acceptability criteria:
mean recovery for each pesticide in the range 70-120 %, with a RSDr ≤ 20%.
Negative results for represented analytes are supported only indirectly by the
recovery and LCL data for representative analytes and must be interpreted with
caution.