016-PE and TOFD
016-PE and TOFD
Article 4
Nonmandatory Appendix N
Time of Flight Diffraction (TOFD) Interpretation
N-410 SCOPE
• Shear-shear diffraction
• Longitudinal-shear diffraction
• Single transducer diffraction (called “back diffraction” or the “tip-
echo method”)
• Twin transducer TOFD with both transducers on the same side of
the flaw/weld.
• Complex inspections, e.g. nozzles.
N-420 GENERAL
(c) TOFD grayscale images display phase changes, some signals are
white-black-white; others are black-white-black. This permits identification of the
wave source (flaw top or bottom, etc.), as well as being used for flaw sizing.
Depending on the phase of the incident pulse (usually a negative voltage), the
lateral wave would be positive, then the first diffracted (upper tip) signal negative,
the second diffracted (lower tip) signal positive, and the backwall signal negative.
This is shown schematically in Figure N-421(c). This phase information is very
useful for signal interpretation; consequently, RF signals and unrectified signals
are used for TOFD. The phase information is used for correctly identifying signals
(usually the top and bottom of flaws, if they can be differentiated), and for
determining the correct location for depth measurements.
Transmitter Receiver
Lateral wave
Back-wall reflection
+ +
LW BW
- -
Upper tip Lower tip
(d) An actual TOFD image is shown in Figure N-421(d), with flaws. The
time-base is horizontal and the axis of motion is vertical (the same as the
schematic in Figure N-421(c). The lateral wave is the fairly strong multicycle
pulse at left, and the backwall the strong multicycle pulse at right. The flaws show
as multicycle gray and white reflections between the lateral and backwall signals.
The scan shows several separate flaws (lack of fusion, porosity, incomplete
penetration and slag). The ultrasonic noise usually comes from grain reflections,
which limits the practical frequency that can be used. TOFD scans may only
show the lateral wave (OD) and backwall (ID), with “noise”. There is also
ultrasonic information available past the backwall (typically shear wave
diffractions), but this is generally not used.
Lack of fusion
Porosity
Lack of fusion
Incomplete
penetration
Slag
A-
d1
d1
h
t1 t2
Error! Cursors
Build-in L
Flaws will not always be symmetrically placed between the transmitter and
receiver transducers. Normally, a single pair of transducers is used, centered on
the weld axis. However, multiple TOFD sets can be used, particularly on heavy
wall vessels, and offsets are used to give improved detection. Also, flaws do not
normally occur on the weld centerline. Either way, the flaws will not be positioned
symmetrically, Figure L-452(a) and this will be a source of error in location and
sizing.
S S
Transmitter Receiver
t0 t0
There will be positional and sizing errors associated with a non-centered flaw, as
shown in Figure N-452(b). However, these errors will be small, and generally are
tolerable since the maximum error due to off-axis position is less than 10% and
the error is actually smaller yet since both the top and bottom of the flaw are
offset by similar amounts. The biggest sizing problems occur with small flaws
near the backwall. Exact error values will depend on the inspection parameters.
Flaw Position Uncertainty
S S
Transmitter Receiver
t0
t1 t2
In practice, the maximum error on absolute depth position lies below 10%.
Flaw lengths parallel to the surface can be measured from the TOFD image by
fitting hyperbolic cursors, Figure N-453, to the ends of the flaws.
Flaw height perpendicular to the surface can be measured from the TOFD image
by fitting cursors on the top and bottom tip signals. The following are two
examples of depth measurements of weld flaws in a 1 in. (25mm) thick plate.
Figure N-454(a) is midwall lack of fusion and Figure N-454(b) is a centerline
crack. Note that TOFD signals are not linear, so midwall flaws show in the upper
third region of the image. It is possible to linearize the TOFD scans by computer
software.
0.43 in.
0.59 in.
0.88 in.
(a) Point flaws, Figure N-481(a), like porosity, show up as single multicycle
points between the lateral and backwall signals. Point flaws typically display a
single TOFD signal since flaw heights are smaller than the ring-down of the pulse
(usually a few millimeters, depending on the transducer frequency and damping).
Point flaws usually show parabolic “tails” where the signal drops off towards the
backwall.
A-scan
Indication
Lateral
Backwall
wave
Transmitter Receiver
Lateral
Surface-
breaking
flaw
1
2
If a midwall flaw is shallow, i.e. less than the transducer pulse ring-down (a few
millimeters), the top and bottom tip signals cannot be separated. Under these
circumstances, it is not possible to differentiate the top from the bottom of the
flaw, so the evaluator can only say that the flaw is less than the ringdown
distance (which depends on transducer frequency and damping, etc.).
(e) Lack of root penetration, Figure N-481(e), is similar to an inside (ID)
far-surface-breaking flaw (see N-481 (b)). This flaw gives a strong diffracted
signal (or more correctly, a reflected signal) with a phase inversion from the
backwall signal. Note that whether signals are diffracted or reflected is not
important for TOFD characterization; the analysis and sizing is the same. Also
note even though there is a perturbation of the backwall signal, the backwall is
still visible across the whole flaw. This material also shows small point flaws and
some grain noise, which is quite common. TOFD typically overemphasizes small
point flaws, which are normally undetected by conventional shear wave pulse-
echo techniques.
Error!
2
3
Figure N-481 (e) Flaw location, and TOFD display showing the
image of the lack of root penetration
(f) Concave root flaws, Figure N-481(f), are similar to lack of root
penetration. The top of the flaw is visible in the TOFD image, as well as the
general shape. The backwall signal shows some perturbation as expected.
2
3
Figure N-481(f) Flaw location and TOFD display showing the image the
concave root flaw.
(g) Sidewall lack of fusion, Figure N-481(g) is similar to a midwall flaw (see
N-481(d)) with two differences. First, the flaw is angled along the fusion line, so
TOFD is effectively independent of orientation, which is not a problem for TOFD.
Second, the upper flaw signal is partly buried in the lateral wave for this particular
flaw. In this instance, the upper tip signal is detectable since the lateral wave
signal amplitude is noticeably increased. However, if this were not the case, then
the evaluator would be unable to accurately measure the flaw depth.
1
2
3
2 1
1
2
Transmitter Receiver
Lateral
Reflected
Back wall
L B
Reflection
1
2
(a) Acceptable scan, Figure N-483(a) . The gain and gate setting are
reasonable, and the electrical noise is minimal.
OD surface-breaking flaw
Lateral wave
Backwall
Signals
becoming
invisible
in this
area.
Signals are
becoming
confused
in these
areas.
Figure N-483(c): TOFD image with the gain set too high
(d) Correct gate settings are critical, because TOFD A-scans are not that
easy to interpret since there are multiple visible signals. As a minimum, the gates
should encompass the lateral wave and longitudinal wave backwall signal; the
gate can extend to the shear wave backwall, if required. Typically, the best signal
to use as a guide is the first (longitudinal wave) backwall, since it is strong and
always present (assuming the transducer separation is reasonably correct). The
following figures show examples of incorrect gate positioning, which will
inherently lead to poor flaw detection.
The first example, Figure N-483(d)(1) shows the gate set too early, the lateral
wave is visible and the backwall is not. Any inside (ID) near-backwall flaws will be
missed.
Lateral wave
Figure N-483(d)(1) TOFD image with the gate set too early
The second example, Figure N-483 (d)(2), shows the gate set too late. The
lateral wave is not visible. The first signal is the backwall, and the second signal
is the shear wave backwall. With this setup, all the outside (OD) near-surface
flaws will be missed.
L-wave
backwall
S-wave
backwall
signal
Figure N-483(d)(2) TOFD image with the gate set too late
The third example, Figure N-483(d)(3), is with the gate set too long. Though this
is not technically incorrect, the image will show the diffracted backwall shear-
shear wave signal. These S-S waves may show additional and confirmatory
information. The diffracted shear waves show the porosity more clearly than the
diffracted longitudinal waves and there is a strong mode-converted signal that
occurs just before the shear wave gate, which could cause interpretation
problems. Normally, the gate is set fairly short to enclose only the lateral wave
and the longitudinal wave backwall to clarify interpretation.
Lateral wave
L-wave
backwall
signal
S-wave
backwall
signal
Figure N-483 (d)(3) TOFD image with the gate set too long
(e) Incorrect (too far apart) transducer separation, Figure N-483(e),
results in the backwall signal becoming distorted, the lateral wave becomes
weaker, and some of the diffracted signal amplitudes drop.
Distorted
L-wave
backwall
Figure N-483(e) TOFD image with transducers set too far apart
(f) Incorrect (too close together) transducer separation, Figure N-483(f),
results in the the lateral waves becoming stronger, and the backwall weaker.
Again, the TOFD image of the flaws is poor.
Strong
lateral wave
Weak L-wave
backwall
signal