MW-12A Operation Manual v4.0
MW-12A Operation Manual v4.0
Microplate Washer
Operator’s Manual
© 2007-2010 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights
Reserved.
, , , , BeneView,
WATO, BeneHeart, are the trademarks, registered or otherwise, of Mindray in
China and other countries. All other trademarks that appear in this manual are
used only for informational or editorial purposes. They are the property of their
respective owners.
All information contained in this manual is believed to be correct. Mindray shall not
be liable for errors contained herein or for incidental or consequential damages in
connection with the furnishing, performance, or use of this manual.
Mindray is responsible for the effects on safety, reliability and performance of this
product, only if:
the electrical installation of the relevant room complies with the applicable
national and local requirements; and
i
WARNING:
is important for the hospital or organization that employs this equipment
to carry out a reasonable service/maintenance plan. Neglect of this
may result in machine breakdown or personal injury.
NOTE:
his equipment must be operated by skilled/trained clinical professionals.
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY
OR FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any
transportation or other charges or liability for direct, indirect or consequential
damages or delay resulting from the improper use or application of the product or the
use of parts or accessories not approved by Mindray or repairs by people other than
Mindray authorized personnel.
Company Contact
ii
Preface
Before using the MW-12A, please read this operator’s manual thoroughly for
relevant operation instructions.
Intended Audience
This manual is geared for clinical professionals to:
WARNING
The MW-12A Microplate Washer is to be operated only by clinical
professionals, doctors or laboratory experimenters that are trained
by Mindray or Mindray-authorized distributors.
Preface 1
Conventions
Safety Symbols
Safety symbols alert you to potentially dangerous conditions and information
that requires your attention. These safety symbols, together with related text,
apply to specific procedures and appear as needed throughout this manual.
Labels
The labels attached to the surface of the instrument use symbols to clarify the
meaning of the text. If any of the labels peels off, contact Mindray customer
service department or your local distributor for replacement. The following chart
explains the labels used in this manual.
Serial number
Manufacture date
Manufacturer
2 Preface
In Vitro diagnostic equipment
Power ON
Power OFF
Graphics
All graphics, including screens, are for illustration purpose only and must not be
used for any other purposes.
EC Representative
Name: Shanghai International Holding Corp. GmbH (Europe)
Address: Eiffestraβe 80, 20537 Hamburg Germany
Phone: 0049-40-2513175
Fax: 0049-40-255726
Preface 3
Safety Precautions
Observe theses safety precautions when using the MW-12A Microplate Washer.
Ignoring any of the precautions may lead to personal injury or equipment
damage.
WARNING
If the instrument is used in a manner not specified by Mindray, the
protection provided by the system may be impaired.
WARNING
When the instrument is powered on, users must not remove its
enclosure.
Spillage of liquid on the analyzer may cause equipment failure and
even electric shock. In case of spillage, switch off the power
immediately, remove the spillage and contact Mindray’s customer
service department or your local distributor.
WARNING
Do not touch such moving parts as plate carrier, manifold, etc.
when the system is in operation.
Do not put your finger or hand into any open part when the system
is in operation.
Preventing Infection
BIOHAZARD
Inappropriately handling samples may lead to biohazardous
infection. Do not touch the sample or waste with your hands. Wear
gloves and lab coat and, if necessary, goggles.
In case your skin contacts the sample or waste, follow standard
laboratory safety procedures and consult a doctor.
4 Preface
Treating Waste Liquids
BIOHAZARD
Some substances in reagent, sample, wash solution and waste are
subject to regulations of contamination and disposal. Dispose of
them in accordance with local or federal guidelines for
biohazardous waste disposal and consult the manufacturer or
distributor of the reagents for details.
Please observe the following instruction to dispose of the used parts to prevent
from environmental pollution and personal injury.
BIOHAZARD
Materials of the manifold, plastic bottle, tubing or washer are
subject to contamination regulations. Dispose of them in
accordance with local or federal guide lines for biohazardous waste
disposal.
WARNING
Ethanol is flammable substance. Please exercise caution while
using the ethanol. Do not use flammable substance around this
instrument.
Preface 5
Precautions on Use
To use the MW-12A Microplate Washer safely and efficiently, please pay much
attention to the following operation notes.
Intended Use
WARNING
This instrument is intended for ELISA test in a clinical laboratory
and used in conjunction with Microplate Reader.
Please consult Mindray first if you want to use the system for other
purposes.
Operator
WARNING
The MW-12A Microplate Washer is to be operated only by clinical
professionals, doctors or laboratory experimenters that are trained
by Mindray or Mindray-authorized distributors.
Environment
CAUTION
Please install and operate the system in an environment specified
by this manual. Installing and operating the system in other
environment may lead to poor washing effect and even equipment
damage.
To relocate the system, please contact Mindray customer service
department or your local distributor.
6 Preface
Preventing Interference by Electromagnetic Noise
CAUTION
Do not use this device in close proximity to sources of strong
electromagnetic radiation (e.g. unshielded intentional RF sources),
as these may interfere with the proper operation. The
electromagnetic environment should be evaluated prior to
operation of the device. This equipment has been designed and
tested to CISPR 11 Class A, in a domestic environment it may
cause radio interference, in which case, you may need to take
measures to mitigate the interference. Do not install devices
generating excessive electromagnetic noise around the system. Do
not use such devices as mobile phones or radio transmitters in the
room housing the system. Do not use other CRT displays around
the system. Interference by electromagnetic noise may lead to
system failure.
Do not use other medical instruments around the system that may
generate electromagnetic noise to interfere with their operations.
Communication Port
WARNING
This instrument has a serial port that is used only for upgrading the
system software. Users must not operate the instrument via this
serial port; otherwise equipment damage may be caused.
Preface 7
8 Preface
Contents
Preface............................................................................................................... 1
Intended Audience .........................................................................................................1
What Can You Find in This Manual................................................................................1
Conventions...................................................................................................................2
Safety Precautions.........................................................................................................4
Precautions on Use .......................................................................................................6
2 Installation................................................................................................2-1
2.1 Unpacking ....................................................................................................2-3
2.2 Installation Requirements .............................................................................2-3
2.2.1 Space Requirements ......................................................................2-3
2.2.2 Power and Environmental Requirements .......................................2-3
2.3 Installing the Washer ....................................................................................2-4
2.4 Installing the Manifold...................................................................................2-4
2.5 Connecting the Bottles .................................................................................2-5
2.6 Installing the Power Cord .............................................................................2-6
Contents I
3 Basic Operations .................................................................................... 3-1
3.1 Overview ......................................................................................................3-3
3.1.1 Keypad...........................................................................................3-3
3.1.2 Main Screen...................................................................................3-3
3.2 Needle Positions ..........................................................................................3-4
3.2.1 Manifold Selection..........................................................................3-4
3.2.2 Position Adjustment .......................................................................3-5
3.3 Parameters................................................................................................. 3-11
3.3.1 Parameter Configuration Screen.................................................. 3-11
3.3.2 Setting up Programs ....................................................................3-12
3.4 Running a Program ....................................................................................3-16
3.5 Maintenance...............................................................................................3-19
3.6 Waste Full ..................................................................................................3-20
5 Troubleshooting...................................................................................... 5-1
5.1 Manifold Failure............................................................................................5-3
5.2 Movement Error ...........................................................................................5-4
5.3 Dispense Needle Error .................................................................................5-4
5.4 Residue Too Much........................................................................................5-5
5.5 Wash Solution Overflow ...............................................................................5-6
5.6 Collision of Manifold and Microplate.............................................................5-6
II Contents
1 The Basics
1.1 System Description ......................................................................................1-3
1.2 Appearance ..................................................................................................1-3
1.3 Control Mode................................................................................................1-4
1.4 Operation Tips ..............................................................................................1-4
1.5 Storage .........................................................................................................1-4
1.2 Appearance
The MW-12A Microplate Washer applies to 96-/48-well microplates. It is
supplied with a 12-/8-channel manifold, wash bottles, waste bottle, and buffer
bottle, and supports two types of wash solution.
NOTE
The initialization will take about 20 seconds.
1 Fill each wash bottle with appropriate wash solution, then connect the
bottles to the washer.
2 Fill the distilled water bottle with distilled water or deionized water for use
in washing.
3 Connect the buffer bottle and empty waste bottle to the washer.
4 Power on the washer. When startup initialization is finished, configure the
wash positions, parameters and then run the wash programs.
5 When the wash procedure is finished, connect the distilled water bottle to
Wash2 port on rear panel of the washer, then launch a rinse
maintenance procedure to clean the tubing for no less than 15s.
6 Power off the washer.
7 Empty and clean the waste bottle in accordance with your laboratory
safety regulations.
CAUTION
When the instrument is not used, lay down the aerosol cover to
prevent from dust and foreign matters.
1.5 Storage
The washer and packaging materials should be stored in an environment that
meets the following requirements:
CAUTION
The washer should be stored without any residual fluid in it.
Perform the following steps before storing the washer:
1) Before turning off the washer, run a rinse procedure for less
than 20s with distilled water or deionized water.
2) When rinsing is finished, disconnect the two wash bottles from
the washer, and place a used microplate on the plate carrier.
3) On the parameter configuration screen, select Wash1 and run
at least one wash program, then launch a rinse procedure for
over 15 seconds.
4) Turn off the washer.
Installation 2-1
2-2 Installation
2 Installation
WARNING:
The MW-12A should be installed only by Mindray or
Mindray-authorized technicians. You should prepare a place
meeting the requirements specified in 2.2 Installation
Requirements
2.1 Unpacking
When you receive the instrument, carefully inspect the package. If you see any
signs of mishandling or damage, file a claim immediately with Mindray
Customer Service Department or your local distributor.
After opening the package, check the delivered goods against the packing list
as well as the appearance of the instrument. If you find anything missing or
damaged, alert Mindray Customer Service Department or your local distributor
immediately.
Installation 2-3
Atmospheric pressure: 70 kPa-106kPa
The instrument should be far from strong magnetic fields.
The instrument should be shielded from direct sunlight.
The instrument should be properly grounded.
The installation site should be well ventilated.
1 After opening the package, take out the fitting pack and washer. Open
the fitting pack and take out the wash bottles, tubing, etc.
2 Set the washer on the operating platform and keep it at least 15cm away
from the wall. Remove the packaging material from the washer.
3 Remove the protective materials from the manifold support, plate carrier
and tubing.
4 Install the manifold as per section 2.4 Installing the Manifold.
5 Install the bottles and waste sensor as per section 2.5 Connecting the
Bottles and 2.6 Installing the Power Cord.
2-4 Installation
2 One end of the dispensing tubing (thinner) and waste tubing (thicker) has
been connected to corresponding black and red ports on the washer.
Connect the other end of the dispensing tubing and waste tubing
respectively to the smaller and bigger connectors on the manifold.
NOTE
Be sure to insert the tubing into the connector until the tubing
cannot proceed any more.
Installation 2-5
1 Connect one end of wash1 tubing (black) to Wash1 connector on the
washer and the other end to the black fitting on wash1 bottle.
NOTE
Be sure to insert the tubing into the connector until the tubing
cannot proceed any more.
Be sure to tighten the caps of waste bottle and buffer bottles during
washing process. Otherwise good aspirating may not be achieved.
2-6 Installation
WARNING
The instrument uses a three-wire power cord and should be
grounded properly; otherwise electric shock or equipment damage
may be caused.
A fuse ([email protected], slow-blow) is provided on the instrument
and must not be replaced by user.
Ensure that the power supply complies with the power
requirements specified in this manual and an appropriate fuse is
installed.
CAUTION
Ensure that the instrument is switched off before connecting the
power cord.
Installation 2-7
2-8 Installation
3 Basic Operations
3.1 Overview ......................................................................................................3-3
3.1.1 Keypad ...........................................................................................3-3
3.1.2 Main Screen ...................................................................................3-3
3.2 Needle Positions ..........................................................................................3-4
3.2.1 Manifold Selection..........................................................................3-4
3.2.2 Position Adjustment........................................................................3-5
3.3 Parameters................................................................................................. 3-11
3.3.1 Parameter Configuration Screen.................................................. 3-11
3.3.2 Setting up Programs ....................................................................3-12
3.4 Running a Program ....................................................................................3-16
3.5 Maintenance...............................................................................................3-19
3.6 Waste Full ..................................................................................................3-20
3.1 Overview
The MW-12A Microplate Washer can be operated via the keypad on the front
panel.
3.1.1 Keypad
The keypad of the instrument is shown in the figure below.
Parameter Position
Wash Maintenance
12 Channel
Cent Pos Left Pos
Right Pos Disp Pos
Bot Auto Ht Bot Manu Ht
NOTE
When using 48-well plates, you must select 12-channel manifold
and fix it with the positioning plate provided with the instrument.
When changing the plate type, you must reset the parameters on
the needle position setup screen.
1 Press the POS key on the front panel of the washer. The needle position
setup screen shows and the manifold type is highlighted by default
12 Channel
Cent Pos Left Pos
Right Pos Disp Pos
Bot Auto Ht Bot manu Ht
Parameter Position
Wash Maintenance
NOTE
The washer has been aligned with standard microplate before
leaving the factory. However, you need to adjust the needle
position according to your plate type, so that the washer can
operate more effectively.
Different microplates may be different in height. Before using the
plates, be sure to set them level and make sure no microwell is
protruded.
Aspiration height refers to the depth that the aspirate needle moves down to the
well during washing. During wash cycles, the aspirate needle aspirates liquid
from the well and sucks it into the waste bottle. You must set appropriate left
and right positions based on different plate types.
Home
position Home
position
Left Pos
Dispensing
height
(reference)
Aspiration
Right Pos
height
(reference)
Microwell
Asp left
Asp tight
position
position
Central position(reference) Central position(reference)
12 Channel XX
Cent Pos Left Pos
Right Pos Disp Pos
Bot Auto Ht Bot Manu Ht
12 Channel XX
Cent Pos Left Pos
Right Pos Disp Pos
Bot Auto Ht Bot Manu Ht
Home
position
Left Pos
Dispensing
height
(reference)
Right Pos
Microwell
Central position(reference)
12 Channel XX
Cent Pos Left Pos
Right Pos Disp Pos
Bot Auto Ht Bot Manu Ht
Home
position
Left Pos
Dispensing
height
(reference)
Right Pos
Microwell
Central position(reference)
NOTE
The left and right positions, relating to crosswise washing,
must not go beyond the microwell to avoid damaging the
needles. Also the two positions must not be too close to
the well center to achieve best washing effects. It is
recommended that the two positions set to about
0.2-0.5mm away respectively from the left and right walls
of the well.
12 Channel XX
Cent Pos Left Pos
Right Pos Disp Pos
Bot Auto Ht Bot Manu Ht
NOTE
The dispensing position must not be set too high, so that
the aspirate needles are far from the well and overflow is
caused. Also the dispensing position must not be set too
low, so that the aspirate needles move deep into the well
and the well opening cannot be washed thoroughly. You
are recommended to adjust the aspirate needles to be
about ±0.5mm away from the well.
Home
position
Left position
Dispensing
height
(reference)
Right position
Adjustment
range
(reference)
Microwell
Central position(reference)
9 Press OK to save your settings and highlight Bot Auto Ht. The needles
move to the bottom of the first row of the plate.
12 Channel XX
Cent Pos Left Pos
Right Pos Disp Pos
Bot Auto Ht Bot Manu Ht
Home
position
Aspiration
height
(reference)
Asp left
Asp tight
position
position
Central position(reference)
NOTE
If warning appears during bottom auto detection, please
click OK to clear it. But if detection failed, it might be that
the connection of the manifold is contaminated, leading to
poor contact. Please refer to 4.2.2 for details on
maintaining the connection of the manifold. Click EXIT to
return, or click OK to start the Bot Manu Ht.
12 Channel XX
Cent Pos Left Pos
Right Pos Disp Pos
Bot Auto Ht Bot Manu Ht
Home
position
Aspiration
height
(reference)
Asp left
Asp tight
position
position
Central position(reference)
NOTE
You are recommended to use Bot Auto Ht, by which you
can get precise aspiration height. If you want to set the
aspirate height manually, you should adjust the aspirate
needles to be at most 0.4mm away from the microwell
bottom.
Parameter Position
Wash Maintenance
CAUTION
You must adjust all position parameters strictly as instructed above;
otherwise the instrument performance may be compromised.
3.3 Parameters
Press the PARA key on the front panel of the washer. The parameter
configuration screen is displayed as follows.
NOTE
Up to 100 wash programs (No.00-No.99) can be stored in the
washer.
To set up a program:
1 Press the PARA key on the front panel of the washer. The program
setup screen is displayed. Prog xx is highlighted by default.
NOTE
If the selected program number is a new number, the
default parameters are displayed on the screen.
NOTE
You should connect Wash1 and Wash2 to the
corresponding connectors on the rear panel of the washer.
NOTE
You can only select Strip when using 48-well microplates.
NOTE
The dispensing volume for bottom wash cannot be defined
and is about 150μl.
NOTE
You must select Std Asp for UV bottom plates.
NOTE
The dispensing volume range is 50-3000μl with increment
of 50μl. The default dispensing volume is 300μl and the
capacity of each well is about 350μl.
When the dispensed volume is greater than 300μl, the
aspirate needles must not be too far away from the well;
otherwise overflow may be caused. You are
recommended to adjust the aspirate needles to be at most
±0.5mm away from the well.
Dispense needle Aspirate needle
Home
position
Left position
Dispensing
height
(reference)
Right position
Adjustment
range
(reference)
Microwell
Central position(reference)
NOTE
The aspiration time is crucial to washing effects and 1.5s
by default. You can adjust the aspiration time as needed,
and it should be no less than 1s.
NOTE:
If things like removing manifold during wash do not
happen, in the meantime standard micro plate is used with
no conspicuous dimension difference between the
manifold and the micro plate, the default mode—No check
can be used. If check mode is selected, when the
connection of the manifold is contaminated, the following
information will be displayed during washing:
Error!
Please check the manifold !
Press OK to continue or
EXIT to cancel !
Parameter Position
Wash Maintenance
NOTE
After any configuration, you can press EXIT to return to the main
screen or press other keys to start corresponding operations with
your settings saved automatically.
CAUTION
Prior to running a program, check if there is enough wash solution
in the wash bottles and if the waste bottle is emptied.
You must run a pre-rinse before running the first wash program
when,
The instrument is started up; or
The wash solution is replaced.
To run a program:
1 Press the WASH key on the front panel of the washer. The wash screen
is displayed as follows. The program number is highlighted by default.
NOTE
The default program is the one that you’ve just complied or
used recently.
WARNING
You must run a pre-rinse before running the first wash
program when,
The instrument is started up; or
The wash solution is replaced.
NOTE
The default rows number is 8 for 12-channel manifold. If
8-channel manifold is selected, the default rows number
will be 12. You can set the rows number as needed.
NOTE
If the check mode is selected, prior to and during washing
the plate, the system will check the position of the manifold.
If a message “Please check the manifold! Press OK to
continue or EXIT to cancel!” pops up, it might be that the
manifold is not correctly placed or the connector of the
manifold is dirty. Follow instruction of 4.2.2 to clean it. Press
Ok to start the washing, Press EXIT to return to the main
screen. In general, No check is selected.
During the washing process, press EXIT to pause. To
resume the washing, press OK or press EXIT to return to
the main screen.
During washing process, the screen shows the current operation status, cycle
times and rows number. The operation status includes Starting air pump
Pre-rinsing, Aspirating, Dispensing, Soaking, Shaking, etc. Cyc X/X indicates
the current cycle and the total cycles. Rows X/X indicates the current row and
the total rows to be washed.
XXXXXXXXX
Cyc X/X Row X/X
Press EXIT to pause !
CAUTION
If the plate is not loaded properly, do not run wash programs.
Otherwise liquid may be spilled into the instrument and even lead to
electric shock or equipment damage.
Running a program when the wash bottles are lack of wash solution
may compromise the washer’s reliability and cause washing failure.
NOTE
The washer will display a prompt message and pause the operation
to remind you to take actions when the waste in the waste bottle
exceeds the limit line.
Pausing
Cyc X/X Row X/X
Press OK to continue or
EXIT to cancel !
3.5 Maintenance
The fluidic tubing should be rinsed with distilled water or deionized water
periodically in order to prevent crystal deposits, thus eliminating equipment
failure.
NOTE
We recommend that you perform a daily rinse prior to turning off
the washer or a periodical rinse procedure.
1 Press the MAINT key on the front panel. The rinse setup screen is
displayed as follows.
Rinse time XX
Ensure Wash 2 is distilled
water! Press MAINT to Rinse
or EXIT to cancel !
NOTE
You should use Wash2 to rinse the fluidic tubing. Check if
the Wash2 bottle is filled with distilled water (or deionized
water). If not, you’ll not obtain satisfactory rinse effects. The
rinse time can be adjusted from 1s to 300s. The default is
10s. You are recommended to set the rinse time to no less
than 7s.
In rinsing
Press EXIT to pause !
Press the EXIT key during rinsing. The following screen is displayed.
Warning !
Waste full ! Press OK to
continue or EXIT to cancel !
BIOHAZARD
Dispose of the reagent, sample, wash solution and waste in
accordance with local or federal guide lines for biohazardous waste
disposal and consult the manufacturer or distributor of the reagents
for details. While disposing of the waste, wear gloves and lab coat
and, if necessary, goggles.
In case your skin contacts the waste, follow standard laboratory
safety procedures and consult a doctor.
CAUTION
Before running a wash program, be sure to check the waste bottle
for overflow and empty it if necessary; otherwise the instrument
performance may be compromised.
.
Maintenance 4-1
4-2 Maintenance
4 Maintenance
To ensure the washer works normally, you must follow this procedure to
maintain it properly.
BIOHAZARD
When performing maintenance, do not touch any wash solution or
waste fluid with your hands. Wear gloves and lab coat and, if
necessary, goggles.
In case your skin contacts the waste fluid, follow standard
laboratory safety procedure and consult a doctor.
CAUTION
The wash solution contains phosphate and may result in crystal
deposits, thus clogging the tubing or connectors inside the washer.
Don’t leave wash solution in the tubing for a long period of time. A
rinse procedure is required everyday when the wash program is
finished. If the tubing and needles are plugged by crystal deposits,
clog them with the stainless needle provided with the washer. If the
crystal deposits cannot be removed due to no cleaning for a long
time, replace the manifold.
Follow this procedure to maintain the tubing clamped by the solenoid valves:
Maintenance 4-3
1 As for tubing clamped by Solenoid Valve 1, first pull the outer tubing a
little, then keep pressing the black end of Valve 1 and pull the inner
tubing.
2 As for tubing clamped by Solenoid Valve 2 first perform a rinse
maintenance for about 20s, then pull the tubing quickly when the
dispense needle is dispensing liquid.
NOTE
The tubing used together with Solenoid Valve 1, and Solenoid
Valve 2, has a service life of about 18 months if maintained as
required by Mindray. Check the tubing frequently for damage and
replace it if any.
NOTE
If used for a long time, the wash solution might adhere to the
manifold retaining screw or metal plate due to spillage and
evaporation. The connector of the manifold should be cleaned in
time to avoid rust or crystallizing affection washing effect.
4-4 Maintenance
4.3 Monthly Maintenance
When used for a long period, the wash bottle, waste bottle, buffer bottle and
distilled water bottle may have crystal deposits. Follow this procedure to wash
the bottles every month:
1 Fill each bottle with 0.8L deionized water (40℃) and tighten the bottle
cap, then shake the bottle.
2 Remove the bottle cap and use a brush to clean the inside wall and
cap of the bottle. While cleaning, ensure no waste fluid is spilled out.
Maintenance 4-5
WARNING
The needles and leaf spring are edged and can cause scratches to
human body. Exercise caution while performing the cleaning job.
4-6 Maintenance
5 Troubleshooting
5.1 Manifold Failure............................................................................................5-3
5.2 Movement Error............................................................................................5-4
5.3 Dispense Needle Error .................................................................................5-4
5.4 Residue Too Much........................................................................................5-5
5.5 Wash Solution Overflow ...............................................................................5-6
5.6 Collision of Manifold and Microplate.............................................................5-6
Troubleshooting 5-1
5-2 Troubleshooting
5 Troubleshooting
NOTE
When alarm occurs, press any key to cancel the alarm and operate
according to the screen prompts.
Error!
Please check the manifold !
Press OK to continue or
EXIT to cancel !
1 Check if the manifold is placed correctly and not tilted, place it level if
needed. Press OK and check if the instrument works normally; or
2 Lift up the manifold, then wipe the retaining screws and metal plates with
clean cotton cloth dipped with ethanol. Install the manifold properly and
press OK; or
Troubleshooting 5-3
mode—No check( )can be used.
5 If the error remains, please contact Mindray Customer Service
Department or your local distributor.
Error!
XXXX movement error !
Restart or see the user
manual !
5-4 Troubleshooting
4 Check if the tubing end inside the wash bottle is blocked. If it is, remove
the bottle cap and then unclog the tubing; or
5 Check if the manifold inlet and dispense needles are clogged. If they are,
remove the foreign matters from the manifold inlet using tweezers, and
use the supplied stainless needle to unclog the blocked dispense
needles; or
6 Interchange the dispensing tubing and the waste tubing, then press
MAINT to backflush the dispensing tubing with air flow, drawing the
foreign matters into the waste bottle. Reconnect the tubing correctly and
run a rinse procedure again.
7 If the error remains, please contact Mindray Customer Service
Department or your local distributor.
Troubleshooting 5-5
tubing correctly and run a rinse procedure again.
8 If the error remains, please contact Mindray Customer Service
Department or your local distributor.
1 If the microplate is not placed properly on the carrier, reset the plate and
attach it to the carrier surface completely; or
2 Check the manifold position is set properly. Reset the crosswise
aspiration positions (Left Pos and Right Pos), ensuring the aspirate
needles lie inside the wells. Also check if the dispensing height is set
correctly; or
3 Check if the plate carrier or manifold support is blocked. Remove the
foreign matters if any.
4 If the error remains, please contact Mindray Customer Service
Department or your local distributor.
WARNING
Servicing the instrument can be performed only by technical
personnel approved by Mindray.
You must troubleshoot the instrument strictly as instructed by the
operator’s manual.
5-6 Troubleshooting
CAUTION
Use the consumables recommended by Mindray. Other
consumables may degrade the system performance.
NOTE
The service life of consumables like plastic bottles, connectors or
manifold is about 24 months and that of tubing is about 18 months.
Please check the consumables and tubing frequently and replace
them if necessary.
Troubleshooting 5-7
5-8 Troubleshooting
6 Appendix
Appendix A Specifications .................................................................................6-3
Appendix B Supplies .........................................................................................6-4
Appendix C Index ..............................................................................................6-5
Appendix 6-1
6-2 Appendix
Appendix A Specifications
Environment Indoor
Power supply 100-240V~, 50/60Hz
Power consumption: 100VA
Dimensions 400 mm×330 mm×145 mm (L×D×H)
Weight 7.2kg
Display 192×64 dot matrix LCD
Keypad consists of 8 functional keys
Operating 5°C-35°C
temperature
Storage 0°C-40°C
temperature
Relative 15%-85%
humidity
Operating -400m-3,000 m (-1,300 feet-6,500 feet)
altitude
Atmospheric 70kPa-106kPa
pressure
Communication RS-232
interface
EMC The equipment complies with the emission and immunity
requirement description in this part of the EN 61326-1:2006
and EN 61326-2-6: 2006.
Appendix 6-3
Appendix B Supplies
6-4 Appendix
Appendix C Index
aspiration .... 3-3, 3-4, 3-6, 3-7, 3-9, handle.......................... 4, 3-20, 4-3
3-10, 3-11, 3-13, 3-15, 5-5
Bot Manu Ht...................... 3-4, 3-9 maintenance . 1, 1-4, 3-19, 4-1, 4-3,
4-4, 4-5, 5-6
bottle... 3, 5, 1-3, 1-4, 2-5, 2-6, 3-5,
3-16, 3-18, 3-19, 3-20, 4-3, 4-5, manifold..... 4, 5, 1-3, 1-4, 2-4, 2-5,
5-4, 5-5, 5-6 2-6, 3-4, 3-6, 3-8, 3-10, 3-18,
4-3, 4-4, 4-5, 5-3, 5-4, 5-5, 5-6,
bottom shape ................. 3-11, 3-12 5-7
bottom wash ...........3-11, 3-13, 5-5 manifold support .2-4, 4-4, 4-5, 5-6
C N
D R
Disp Pos .....................3-4, 3-8, 5-6 retaining screw........... 2-4, 4-5, 5-3
dispense ...... 1-3, 2-5, 3-3, 3-4, 3-8, Right Pos.................... 3-4, 3-7, 5-6
3-11, 3-13, 3-14, 3-18, 3-20, 4-4,
rinse 1-4, 1-5, 2-5, 3-3, 3-17, 3-19,
4-5, 5-4, 5-5, 5-6
4-3, 4-4, 4-5, 5-4, 5-5, 5-6
Appendix 6-5
rows ..................... 3-16, 3-17, 3-18 W
6-6 Appendix
P/N: 0105-20-63148 (4.0)