DPI880 Multi Function Calibrator
DPI880 Multi Function Calibrator
Sensing
10
A2
1
A3
2 7 11
3 6
4
5
B1
12
A2
13 15
14
A3
18 17 16 25
A
19 3
20
21
22
23
24
K405 Issue 3
Table of Contents
Introduction ......................................................................... 2 Specification ...................................................................... 17
Safety ..................................................................................... 2 General ............................................................................................. 17
Marks and symbols on the instrument ................................. 2 Electrical (A1 - Item 10) ............................................................. 17
To start .................................................................................. 3 Electrical connectors (A2) ........................................................ 17
Key to figure A1/A2 (Instrument) ............................................. 3 Temperature ranges (RTD) ...................................................... 18
Key to figure A3 (Display) ............................................................. 3 Resistance ranges (Ohms/RTD) ............................................. 18
Prepare the instrument ................................................................ 4 Frequency ....................................................................................... 18
Power on or off ................................................................................ 4 Temperature ranges (TC) ......................................................... 19
Set up the basic operation ......................................................... 4 mV (TC) range ................................................................................ 19
Select a task (Measure and/or supply) .................................. 4 Customer service ................................................Back cover
Set up the settings ......................................................................... 5
Operation .............................................................................. 6
Electrical connections .................................................................. 6
Communications port connections ........................................ 6
Change the output values .......................................................... 6
Measure/supply mA ...................................................................... 7
Measure/supply Volts or mV ..................................................... 8
Measure/supply Hz or pulses .................................................... 8
RTD/Ohms connections ............................................................... 8
Measure/simulate an RTD or Ohms ....................................... 9
Thermocouple (TC) connections .............................................. 9
Measure/simulate a Thermocouple .................................... 10
Transmitter calibration ............................................................. 10
Switch test ....................................................................................... 11
UPM Pressure measurements ............................................... 11
Error indications ........................................................................... 12
Maintenance ..................................................................... 12
Clean the unit ................................................................................ 12
Replace the batteries ................................................................. 12
Calibration ......................................................................... 13
Before you start ............................................................................ 13
Procedures: mA input ................................................................ 14
Procedures: mA output ............................................................. 14
Procedures: mV/Volts input .................................................... 14
Procedures: mV/Volts output ................................................. 15
Procedures: Hz input/output .................................................. 15
Procedures: CJ input .................................................................. 16
Procedures: RTD (Ohms) input ............................................... 16
Procedures: RTD (Ohms) output ............................................ 16
Procedures: TC (mV) input/output ........................................ 17
Procedures: IDOS UMM ............................................................. 17
Maximum: 30V
Edit Settings
Maximum: 30V
Edit Settings
a) Measure b) Supply
0 to ±30V OR 0 to 12V OR
0 to ±120mV 0 to 120mV
a) Input b) Output
Pt100
°C
°C
Settings
Edit Settings
Settings Settings
a) °C or °F (Input) b) mV (Input) Figure 10: Example configuration - Transmitter calibration
with external loop power
Settings
Specification - Frequency
cpm = counts/minute, cph = counts/hour
Range (Measure): Accuracy:
0 to 999.999 Hz For all the ranges:
0 to 50.0000 kHz 0.003% of reading + 2 counts
cpm: 0 to 999 999
cph: 0 to 999 999
ppm = pulses/minute, pph = pulses/hour
Range (Supply): Accuracy:
0 to 999.99 Hz 0.003% of reading + 0.0023 Hz
0 to 50.000 kHz 0.003% of reading + 0.0336 Hz
ppm: 0 to 59 999 0.003% of reading + 0.138 cpm
pph: 0 to 99 999 0.003% of reading + 0.5 cph
Temperature coefficient
14 to 50°F, 86 to 122°F 0.0017% FS / °F
(-10 to 10°C, 30 to 50°C) (0.003% FS / °C)
Output waveform Square, bipolar
Voltage input 0 to 30 V
Trigger level 0 to 12 V, Resolution: 0.1 V
Output amplitude 0.1 to 12 V dc ± 1% (≤ 10 mA)
0.1 to 12 V ac* ± 5% (≤ 10 mA)
* Peak to Peak