Design of The Real Time ECG Signal Processing and Monitoring System Using Lab VIEW With Data Acquisition System
Design of The Real Time ECG Signal Processing and Monitoring System Using Lab VIEW With Data Acquisition System
ISSN No:-2456-2165
IJISRT18JU244 www.ijisrt.com 96
Volume 3, Issue 6, June – 2018 International Journal of Innovative Science and Research Technology
ISSN No:-2456-2165
Generated ECG signal on Electrocardigraph is shown is The positive input electrode will be placed on the left
below figure 3 inner wrist, the negative input electrode will be placed on the
right inner wrist, and the ground electrode will be connected to
the ankle. This will allow for readings to be taken across the
heart with relative accuracy. This placement of electrodes car-
ried out ECG signal which can be given to the instrumentation
amplifier, low pass filter, notch filter and then DAQ and final-
Fig 3:- Normal Heart Beats ly it displayed on the LabVIEW i.e. Graphical user interface.
B. Instrumentation Amplifier :
The instrumentation amplifier offers low power con-
sumption with higher accuracy and precision. The ECG signal
Fig 4:- Fast Heart Beats carried from ECG probes is of having very small in amplitude.
This signal is amplified in this instrumentation amplifier with
gain of 1000. The gain of this instrumentation amplifier is con-
trolled by the register RG. To construct a multistage instrumen-
tation with a gain of 1000, or 60 dB, the following equation
should be applied.
Fig 5:- Slow Heart Beats Gain=(1+2*R1/ RG)
A. Lead Probes
A standard ECG used in medical practice has twelve
electrodes that are placed in a mild semicircular pattern around
the lower region of the heart. These no of electrodes causes
more noise added into ECG signal. So it becomes difficult to
find abnormalities in ECG signal. For that three lead probes
are used in such a manner that placement of these three elec-
trodes makes Einthoven’s Triangle as shown in below figure
7.
IJISRT18JU244 www.ijisrt.com 97
Volume 3, Issue 6, June – 2018 International Journal of Innovative Science and Research Technology
ISSN No:-2456-2165
order low pass filter. The gain of this filter has chosen to be E. NI myDAQ 6008
equal to one. So all the input signal will arrived at output side. The National Instruments USB-6008 devices provide
eight single-ended analog input (AI) channels, two analog out-
The circuit diagram of this second order low pass filter is put (AO) channels, 12 DIO channels, and a 32-bit counter with
shown in below figure 9. a full-speed USB interface. It consists of one 16 bit ADC and
one 16 bit DAC. Here we used 16 bit ADC where the signal
from notch filter is in analog converted into digital format and
given to LabVIEW for graphical presentation. The block dia-
gram of this NI USB 6008 is shown in below figure 11
F. LabVIEW :
LabVIEW is Laboratory Virtual Instrumentation
Engineering Workbench tool. Here we interfaced NI myDAQ
6008 USB with LabVIEW. The digitised ECG signal is
graphically represented on this LabVIEW software.
IJISRT18JU244 www.ijisrt.com 98
Volume 3, Issue 6, June – 2018 International Journal of Innovative Science and Research Technology
ISSN No:-2456-2165
[5] P O. Bobbie, C Z Arif, H Chaudhari, and S Pujari, Elec-
trocardiogram (EKG) Data Acquisition and Wireless
transmission, Proceedings of WSEAS International Con-
ference on System Engineering, 2004.
[6] Channappa Bhyri*, Kalpana.V, S.T.Hamde, and
L.M.Waghmare, “Estimation of ECG features using Lab-
VIEW ” TECHNIA – International Journal of Computing
Science and Communication Technologies, VOL. 2, NO.
1, July 2009. (ISSN 0974-3375).
[7] V. Bindhu • G. Ranganathan • R. Rangarajan, “Statistical
Analysis of Heart Rate Signal Features Using Lab-
VIEW ”, The National Academy of Sciences, India 2014.
IV. CONCLUSION
REFERENCES
IJISRT18JU244 www.ijisrt.com 99