PHV-400 Programmable PMT High Voltage Power Supply Operating Instructions
PHV-400 Programmable PMT High Voltage Power Supply Operating Instructions
PHV-400
Programmable PMT High Voltage Power Supply
Operating Instructions
Ver 2.0
PHV-400.DOC
Description:
The Acton Research Corporation PHV-400 Programmable PMT High Voltage
Power Supply is a negative power supply, operating in the range of 0 to –1250 volts
and designed for use with photomultiplier tubes. It is programmable by an analog
input voltage over the entire voltage range. The programming voltage is compatible
with the Acton Research Corporation SC-1 SpectraCard 16-bit digital readout
system. The PHV-400 PMT power supply is also manually adjustable by a front
panel control and digital display.
Turn off the AC power switch on the left rear of the PHV-400. Connect the high
voltage cable supplied from the MHV connector on the rear of the PHV-400 to the
high voltage MHV connector on the photomultiplier housing. Turn the front panel
control fully counter-clockwise to prevent accidental damage to the photomultiplier
tube. Plug in the supplied AC line cord to the rear of the PHV-400 and the other
end of this line cord to the AC outlet.
Operation:
The PHV-400 PMT power supply can be operated remotely by an analog voltage
from the Acton Research Corporation SC-1 SpectraCard or from the front panel
control. Remote or manual operation is determined by the position of the
REMOTE/LOCAL switch on the rear of the PHV-400. With the switch in the up
or REMOTE position, the output voltage is determined solely by the input
programming voltage where +5 volts DC produces an output voltage –1250 volts
DC with a linear output to 0 volts. Note that with no connector in the rear
programming jack, the output voltage is 0 volts with the REMOTE/LOCAL switch
in REMOTE. The front panel control has no effect with the REMOTE/LOCAL
switch in REMOTE. The front panel digital meter reads the correct output voltage
in both modes, however. With the REMOTE/LOCAL switch in the down position
or LOCAL position, the output voltage is determined by the front panel mounted
2
PHV-400.DOC
control where fully clockwise is –1250 volts DC. The output voltage can be
monitored on the digital meter as it is adjusted. The programming voltage has no
effect in the LOCAL mode. It is advisable to adjust the front panel control fully
counter clockwise before changing the position of the REMOTE/LOCAL switch.
Specifications:
Output voltage 0 to –1250 volts DC
Output current 2 ma. maximum
Programming voltage 0 to +5 volts DC for o to –1250 volts DC output
Accuracy +/- (3% of setting + 0.5% of full scale)
Line Regulation +/- 0.001% for input variation of +/-1%
Ripple 0.001% peak-peak at maximum power output
Stability (after 30 minute warm-up) 0.005% per hour, 0.02% per 8 hours
Input connector Switchcraft #142A mating plug #750
3
PHV-400.DOC