Rain Sensor (RG11) Interfacing Instructions
Rain Sensor (RG11) Interfacing Instructions
2.78
70.71
!
.25
6.40
.38 - APPLICATION WARNING -
9.53 Do not use the RG-11 in any application where the false
.75
indication of water or a missed valid detection of water
19.05
2.17 could cause damage to life or property. It is the
55.20
responsibility of the system designer / integrator to
design redundancy into the system so that the failure of
.41
.75 4.73
any one component, including the RG-11 or other
10.40
19.16 120.23 sensor, does not result in disaster. The manufacturer of
the RG-11, Hydreon Corporation, will in no way be
liable for consequential damages due to the failure or
DIMENSIONAL VIEW false indication of one of its sensors.
RAIN GAUGE CONNECTIONS
INTERFACE
TERMINAL RG-11 DC OR
PWR1
BLOCK AC IN
PWR1 POWER IN PWR2
PWR2 (AC OR DC)
NC
NC RAIN
RELAY SENSING
NO NO
CONTACTS CIRCUITS
COM COM
SWITCH ON PROGRAM /
COMMUNICATIONS RELAY CONTACTS
CONNECTOR MAX 1 A,
12345678 (NOT USED IN MOST 24V DC OR AC
APPLICATION)
PROGRAMMING EQUIVALENT SCHEMATIC
DIP SWITCH
RAIN GAUGE WIRING EXAMPLES
LINE VOLTAGE CONTROL “IT’S RAINING” APPLICATION
COM - The relay output of the RG-11 is
rated for 24V AC or DC, at 1 A. Thus,
120 VAC OPERATOR NC the RG-11 cannot drive a line voltage
LINE IN COM OPEN (120 VAC) load without an external
SWITCH relay.
CLOSED
24 VAC NO -The RG-11 requires a low voltage AC
TRANSFORMER or DC supply and may not be directly
MOTORIZED powered from the 120 VAC line.
24 VAC
DAMPER
For this example, these
switches are on:
RG-11 SW 5 - Sets the sensor to
“It’s Raining’ mode.
PWR1 SW 3 - Monostable extend
PWR2 output for 15 mintues. TIPPING BUCKET
REPLACEMENT EXAMPLE
NC
NO
COM
RG-11
SWITCH ON LOW VOLTAGE 12V DC
“IT’S RAINING” PWR1 + POWER
12345678 APPLICATION PWR2 - SUPPLY
+ 12V DC NC
POWER NO COUNTER
RG-11 E.G. Red Lion Cub-3
- SUPPLY COM
PWR1 SWITCH ON
PWR2 COM
NC 12345678
NO RUN
COM 12V DC, 1 A In many cases, the RG-11 may directly replace
OR LESS, tipping buckets in existing systems.
SWITCH ON MOTOR THAT The switch settings shown here- all off– will
Shown are RUNS WHEN
switches 1, 2, and
emulate a tipping bucket of 0.01”
IT IS RAINING.
12345678 5 on => only turn
on in a heavy
rainfall.
Specifications J2 Connector
J2 is a pin-field on 0.1” centers, used for programming, development,
Parameter Value
and testing of the RG-11. Most applications do not connect to J2,
Input Voltage 12 – 30VDC or and we make this information available only for special applications.
12 – 26 VAC
50V surge Connector field is 0.025” square pins on 0.1” centers. An example
Reverse polarity protected to 50V compatible connector is Molex part number 22-01-3067. This is
Current Drain 15 mA nominal. (No outputs on, not raining, no available from Digi-Key as part number WM2004-ND. The necessary
heater) crimp-on wire terminals are Molex 08-55-0131 / DigiKey WM4591-
about 1.5 mA in micro-power sleep mode. ND.
50 mA with output on.
55 mA - With heater on, 24V dc input. J2 Pin assignments
Output Relay closure, Normally Open and Normally J2 - 1 GND
Closed contacts. J2 - 2 +5V OUT
Max load 1A, 24 VDC. J2 - 3 SW4
J2 - 4 SW1
Operating Temperature -40 C to +60C
J2 - 5 SW2
range
J2 - 6 RS232 and SW 5.
OUT LED
The LED in the center of the circuit board turns on when OUT is on, as an aid
to debugging.
Condensation
Generally, the RG-11 will sense condensation as if it were rainfall, but this
seldom amounts to a significant accumulation of water. The built-in low power
heater (DIP SW 8 off) will tend to reduce condensation.
Heater Notes
A built-in low power (0.25W) heater extends operation of the device to
freezing (32 F or 0C). This is disabled if the micro-power (SW 8) is enabled.
Note that this is a very modest amount of power; it will tend to drive off a
modest amount of frost, but will not melt ice.
Dark Sensing
Turns output on when it is dusk-- nominally less than 2000 lux. This is for
applications such as retracting sun-shields in the evening, when they are not
needed. (Only in Mode 1, “It's Raining” applications.)
Rain Gauge Model RG-11 Instructions PN 55-005-100 Manual for SW rev 016
Mode 0: Tipping Bucket Mode 1: It's Raining
Rain Gauge emulates a tipping bucket of the Rain Gauge turns on the relay to indicate that it is
specified size. raining when the rainfall has reached a given
Switch intensity.
Behavior Switch
7 65 4 3 2 1
Function
0 0 0 X 0 0 0 Bucket Size = 0.01” 7 6 5 4 3 2 1
0 1 0 Bucket Size = 0.0001” (Very sensitive) 0 1 Sensitive-- turn on with very light rainfall
(0.1” per hour ).
1 0 0 Bucket Size = 0.2 mm
1 0 Medium Sensitivity-- turn on with medium
1 0 1 Bucket Size = 0.01 mm (sensitive) rain (0.25” per hour. You would want your
car's wipers on steady slow)
1 1 0 Bucket Size = 0.001 mm (Very sensitive)
1 1 Low Sensitivity-- turn on in heavy rainfall. (1”
1 1 1 1 Reserved for system test
per hour. You would want your car's wipers
on high)
In tipping bucket mode, the Rain Gauge effectively emulates a tipping X 0 X X Output off when rain stops.
bucket of the specified size. For example, if the DIP switches are set
to a tipping bucket mode with a bucket size of 0.01", then the output 1 Output Monostable Extended by 15 minutes
will pulse ON for 50mS each time 0.01" of water accumulates, just as
a tipping bucket would. This can be externally totalized, and used to 0 X X X No Dark-Detect - Normal operation
measure rainfall rates. Bucket sizes of 0.001" and 0.0001" are 1 Dark Detect
similar, generating pulses at accumulations of one one-thousandth,
and one ten-thousands of an inch, respectively. These emulate what
a tipping bucket would do if it were possible to make one that small. Use this mode to control equipment that should be controlled,
Metric bucket sizes are available as well, or the Inch unit scales may enabled, open, closed, and so forth depending on whether or not it is
be scaled with external equipment. raining. The output turns on when a given rate of rainfall is detected,
and turns off after it has dropped below a threshold.
Accuracy
We do not claim an accuracy spec for the RG-11. For more Each of the sensitivity levels (set by switches 1 and 2) provides
information see the “Tipping Bucket” link on www.rainsensors.com. different trip and release points. There is much hysteresis built in, but
real rain fall rates typically fluctuate, even in what you may perceive
as a “steady rain”, so expect the output to turn on and off. The output
will remain on for between about 30 seconds and 5 minutes after the
last detected rain drop, depending on sensitivity setting and actual
conditions.
Rain Gauge Model RG-11 Instructions PN 55-005-100 Manual for SW rev 016
Mode 2: Condensation Sensor Mode 3: Wiper Control
Rain Gauge detects condensation or frost Rain sensing wiper control from off through
formation on the surface. intermittent and steady slow speeds.
Switch Switch
Behavior Behavior
7 6 5 4 3 2 1 7 6 5 4 3 2 1
0 1 0 0 0 0 0 Very Sensitive- first sign of condensation 0 1 1 0 x 0 0 Normal Wiper Control
0 1 Sensitive 0 0 1 Wipe More
1 0 Medium Low 0 1 0 Wipe a Lot More
1 1 Low 0 1 1 Wipe a Whole Lot More
1 0 0 Wipe Less
The rain sensor senses condensation by detecting a shift from the 1 0 1 Wipe a Lot Less
"clear" condition. The relay closes when the condensation occurs,
and opens when the condensation goes away. The rain sensor is set 1 1 0 Wipe a Whole Lot Less
to very gradually adapt the clear condition, so that very gradual build
up of dirt or other contaminants do not cause a false trip. 1 1 1 Wipe hardly at all
x 0 x x Normal Slow Cycle Time (1.2 to 3 sec.)
Condensation sensing mode disables the heater and micro-power
mode. 1 Long Slow Cycle Time (3 – 8 sec.)
The RG-11 may be used to control a wiper system. The output relay
turns on when the slow motor winding should be engaged. This will
typically be used to drive an external relay, which will, in turn, drive
the wiper motor windings. This may be used for the wipers for a boat,
ship, locomotive, observation window, or many other applications.
The RG-11 does not care what the wipers are wiping.
The nominal wiper control is set so that it properly controls the wipers
of a passenger car. It is optimized for wiper systems that require
between 1.2 and 3 seconds to make a single complete actuation of
the wipers. A long cycle time is provide (Switch 4 on) for systems
with a wiper actuation cycle time between 3 and 8 seconds. In all
cases, the RG-11 provides a pulse to initiate the wiper actuation.
Most wiper system will include some sort of cam feedback
mechanism that causes the wipers to keep running until they reach a
home position.
Mount the RG-11 so that it generally gets the same rainfall as the
surface to be wiped. Usually, this means about a 45 degree angle.
The RG-11 does not need to be within the actual field of view of the
window. Adjust the sensitivity control DIP switches (3, 2, and 1) to
set the system to wipe more or less, depending on the needs of the
installation.
Rain Gauge Model RG-11 Instructions PN 55-005-100 Manual for SW rev 016
Mode 4: Irrigation Control Mode 6: Drop Detector
Rain Gauge output on means inhibit watering. Switch
Switch Behavior
7 6 5 4 3 2 1
Behavior
8 7 6 5 4 3 2 1 1 1 0 0 X 0 0 Normal drop threshold
X 1 0 X X 0 0 0 Typical Water Control. Inhibit watering 0 1 Sensitive Drop threshold. Expect rare false
for up to 5 days. trips.
0 0 1 Water More 1 0 Hi drop threshold. Trip only with large drops.
0 1 0 Water a lot more 0 X X One pulse per drop, longer pulses for bigger
1 0 0 Water Less drops
Evaporation Rate
Normal Evaporation Rate = 0.11 inches per day
Hi Evaporation Rate = 0.22 inches per day
In irrigation mode, if Switch 8 is on, the control assumes a high
evaporation (or transpiration rate). Set this switch to ON for sandy
soil or other conditions where the soil tends to dry out quickly. The
system will re-enable the irrigation sooner.
Rain Gauge Model RG-11 Instructions PN 55-005-100 Manual for SW rev 016
Mode 7: First Flush / Rain Water Harvest SAFETY, LIMITS OF RAIN GAUGE LIABILITY,
Rain Gauge output on means rain water is being AND WARRANTY
Only the rain sensor is covered-- absolutely no consequential
harvested and first flush has occurred. damages. If this policy is unacceptable in your installation, do
Switch Behavior not use the RG-11.
8 7 6 5 4 3 2 1 Level (in) Gallons Clean Time (days) It is the responsibility of the systems integrator and purchaser of the
Rain Gauge to insure a safe installation. Any mechanical system,
0 1 1 1 X 0 0 0 0.02 12 3 including one that incorporates a Rain Gauge, requires appropriate
0 0 1 0.04 25 5 safety interlocks. Hydreon Corporation (Hydreon) warrants only the
actual cost of the sensor, and only that it is free from defects in
0 1 0 0.08 50 9 workmanship.
0 1 1 0.16 100 13
The Rain Gauge is warranted to be free from defects for a period of
1 0 0 0.20 125 15 one year from date of purchase. Under no circumstances will
Hydreon be liable for any consequential damages due to failure or
1 0 1 0.24 150 17 any other mishap involving a Rain Gauge. Hydreon's liability in the
1 1 0 0.28 175 19 event of a failure, or inability to sense a condition, is limited to the
actual cost of the particular sensor. Explicitly, if other objects are
1 1 1 0.32 200 21 destroyed due to water damage, or if any object is destroyed
because of a false indication of water, Hydreon is in no way
0 X X X Normal Clean Time whatsoever liable for anything other than the cost of the Rain Gauge,
1 Gets Dirty Fast and then only if the Rain Gauge is shown to have some defect in
materials or workmanship. Limitations and imperfections of the Rain
Note: Switch 8 must be off for this mode. Gauge do not constitute a defect. Further, if some valuable data is
not gathered because an erroneous indication of any sort due to the
The RG-11 can be used in a Rain Water Harvest / First Flush Rain Gauge, Hydreon is liable only for the cost of the Rain Gauge.
application. In this mode, the RG-11 will not change the relay state
until a certain amount of water has fallen. After a certain amount of It is the responsibility of the system designer and purchasers of the
water has been detected, the relay opens, diverting the water to a Rain Gauge to insure that a failure of the Rain Gauge will not cause
proper holding tank. So configured, the output relay acts as an consequential damages. If a failure in Rain Gauge would cause
"Enable Harvest" control, indicating that at least the desired amount disaster, we recommend against deployment of the Rain Gauge, or
of rainfall has accumulated. against the system in which the Rain Gauge is deployed. If a failure
of a Rain Gauge would cause great expense, Hydreon recommends
Level redundant Rain Gauges, and even in that case do not assume any
The amount of water that will accumulate before the relay closes. liability for consequential damages. It is the responsibility of the
system designer and purchasers of the Rain Gauge to be aware of
performance limitations of the device. If a Rain Gauge fails for any
Gallons reason Hydreon will not be responsible for the labor of servicing and
The number of gallons per thousand square feet of collected area or installing and/or removing the Rain Gauge. Labor is NOT
that are diverted. Note that a conventional first flush diverter system COVERED. Hydreon recommends that the system designer perform
totalizes water after it has made it through the collection system. The a Failure-Mode Effects Analysis that includes the possibility of Rain
RG-11 enables the system as soon as the threshold has fallen, which Gauge failure. If a potential purchaser of the Rain Gauge does not
may be substantially sooner. Take this into account when designing agree with these terms, we ask that the potential purchaser not buy
your system. the Rain Gauge. Deployment of the Rain Gauge implies
understanding and agreeing to these limits of liability.
Clean Time
After the rainstorm is over and the relay of the RG-11 is once again Apply engineering judgment: Hydreon does not claim the RG-11 is a
open, it will take some time for the collecting surface to become dirty perfect rain sensor. It is what it is, and senses what it senses.
again, and once again require a complete first flush. Clean time
means how many days, after the water stopped, it takes for the
system to require a full flush. Up to that time, the system will flush CASE and COSMETIC POLICY
less. For example, a system is set to flush 0.08 of accumulation, and Some amount of yellowing or discoloration of the case is considered
thus the clean time is 8 days. If a storm comes after only 4 days, the normal cosmetic aging of the device, and sensors so affect will not be
system will flush only half as much water before enabling harvest replaced under warranty. Tiny cracks or crazing within the lens is
again. also considered cosmetic, and units so affected will be replaced only
if they are deemed by Hydreon corporation to be considered to be of
Gets Dirty Fast = Switch 4 ON a functional nature.
If switch 4 is on, the system remains enabled for only 12 hours after
the rain storm stops. Also, the Clean Time is halved. Use this setting
if the surface becomes contaminated especially quickly.
Rain Gauge Model RG-11 Instructions PN 55-005-100 Manual for SW rev 016