43 Opacity
43 Opacity
Introduction
reverse side, or on sheets underneath. In other words, the greater the opacity of a paper,
If all else remains constant, the opacity of a sheet will increase with higher basis
weight, addition of a mineral filler, lower degree of bleaching, lower density, less beating,
less wet pressing, and coating of the sheet. The amount and type of filler added to the
furnish has a profound effect on opacity. Opacity is generally measured with opacimeters
which use the socalled "contrast-ratio method." This method is based on measurement
not of transmitted light, but of reflected light. Opacity values range from a maximum of
100% for perfectly opaque papers (board) to nearly 0% for transparent ones (glassine).
Significance
The determination of opacity is of vital importance to both the manufacturer and the
consumer. Lack of opacity means too much show-through of the print underneath, and
this interferes with readability. Complete hiding of print underneath is usually impractical
and too expensive, and so a small amount of show-through must be tolerated. The
consumer and manufacturer must cooperate in specifying and supplying papers with
9
practical and reasonable opacity limits.
Definitions
Two slightly different measures of opacity have been defined and are widely used
in the paper trade: TAPPI Opacity and the generally more useful Printing Opacity. Both
subscript zero refers to the fact that the black cavity absorbs all light
entering it; thus, the cavity reflects no light and has a reflectance of zero.
R0.89 = light reflectance of a single sheet backed by the Tappi standard backing of
reflectance 0.89.
TAPPI Opacity and Printing Opacity are defined through these reflectances as follows:
10
Tests
1. On the Thwing-Albert Digital Opacimeter (Model 628), turn the switch from
POWER OFF to STANDBY, and allow the instrument to warm up and stabilize for
2. Place the black cavity over the measurement port. Set the digital readout to +0.0
with the ZERO ADJUST control. Check this zero adjustment periodically throughout
the day.
3. Place an opaque pad of paper over the measurement port. Set the digital readout
4. Then place a single sheet, backed by the black cavity, over the measurement port.
6. Place a single sheet backed by the white body (TAPPI standard backing of
reflectance 0.89) over the measurement port. Adjust the digital readout to +100.0
7. Rotate the black cavity behind the sample and read the TAPPI Opacity directly on
11
the digital readout.
How can you be sure that your stack of sheets is thick enough to qualify as an
opaque stack, that is a stack so thick that for all practical purposes no light is transmitted
through it? Follow this procedure: Start out with a stack you think might qualify. Place it
against the specimen port of the opacimeter, back your stack with the black cavity, and
adjust the digital readout to 100. Next, double the thickness of your stack and place this
thicker stack, backed by the black cavity, against the specimen port. If the digital readout
remains unchanged at 100, you can safely assume that your original stack of sheets was
With samples of low brightness and reflectivity, e.g. unbleached kraft, the
adjustment of the digital readout to 100, as required in steps (3) and (6) above, may not
be possible. In such a case, adjust to the highest digital readout obtainable and note this
value. Continue with the procedure and obtain a readout for the single sheet backed by
the black cavity. Divide this value by the first readout noted above and multiply by 100.
12
Report
Report the means determined from five measurements of Printing Opacity and of
References
ASTM D 589
TAPPI T 425
13
14