MA200 Instruction Manual PDF
MA200 Instruction Manual PDF
1 (2/4)
• No part of this manual may be reproduced or transmitted in any form without prior written permission from
Nikon.
• The contents of this manual are subject to change without notice.
• Although every effort has been made to ensure the accuracy of this manual, errors or inconsistencies may
remain. If you note any points that are unclear or incorrect, please contact your nearest Nikon
representative.
• Some of the equipment described in this manual may not be included in the set you have purchased.
• If you intend to use any other equipment with this product, read the manual for that equipment too.
• If the equipment is used in a manner not specified by the manufacturer, the protection provided by the
equipment may be impaired.
1
Safety Precautions
To ensure correct and safe operation, read this manual before using the product.
Although this product is designed and manufactured to be completely safe during use, incorrect usage or
failure to follow the safety instructions provided may cause personal injury or property damage. To ensure
correct usage, read this manual carefully before using the product. Do not discard this manual and keep it
handy for easy reference.
Safety instructions in this manual are marked with the following symbols to highlight their importance. For your
safety, always follow the instructions marked with these symbols.
Symbol Description
Disregarding instructions marked with this symbol may lead to serious injury or
Warning death.
Disregarding instructions marked with this symbol may lead to injury or property
Caution damage.
Symbol Description
Caution for heat
This marking on the back of the lamphouse calls your attention on the following (the
position of this symbol is shown in Figure 1.1-3):
• The lamphouse becomes extremely hot while the lamp is on and immediately
after it is turned off.
• Do not touch the lamphouse during and immediately after lighting to prevent the
risk of burns.
• Make sure that the lamphouse is sufficiently cool before the lamp replacement.
2
Safety Precautions
WARNING
2. Do not disassemble
Disassembling the microscope or the microscope system may result in electric shock or malfunctions.
Damage or injury that may occur due to mishandling is unwarranted. Never attempt to disassemble any part
other than the parts described in this manual. If you experience problems with the microscope or the
microscope system, contact your nearest Nikon representative.
5. Power cord
Be sure to use the specified power cord for the product. Using a wrong power cord may result in malfunctions
or fire. The product is classified as subject to Class I protection against electrical shock. Make sure it is
connected to an appropriate ground terminal (protective earth terminal). To prevent electrical shock, always
turn off the power switch (set the switch to the “c” position) for the product before connecting or
disconnecting the power cord. For specifications of the power cord, refer to “7. Specifications.”
3
Safety Precautions
8. Air vents
Do not block the air vents on the product and the lamphouse. If the air vents are blocked, the temperature
inside the product will rise. And it may result in damage or fire.
4
Safety Precautions
CAUTION
4. Installation location
This product is a precision optical instrument. The usage or storage in an inappropriate environment may
result in malfunctions or poor performance. Consider the following factors when selecting an installation
location:
• Select an installation location with a temperature from 0 to +40°C and a relative humidity of 85% or less
(there should be no condensation).
Select a storage location with a temperature from -20 to +60°C and a relative humidity of 90% or less
(there should be no condensation).
If installed or stored in a location subject to high temperatures and humidity, mold or condensation may
form on the lens, resulting in lowered performance and possible damage to the microscope.
• Avoid a brightly lit location, such as exposed to direct sunlight or directly under a room light. If there is
excessive ambient light, the image may not clearly be visible.
• Always install the product with a surrounding clear area of 10 cm or more.
• Install the product in a location free from considerable dust or dirt.
• Install the product on a flat surface with little vibration.
• Install the product on a sturdy desk or table for the base of the microscope system.
• Select a layout that allows easy removal of the power cord from the AC inlet of the product in the event of
an emergency.
• Do not install in a narrow space such as a shelf or locker.
• Do not place anything on the product.
• Cover the product to avoid dust when storing.
• For details about the operating environment and storage environment, refer to “7. Specifications.”
5
Safety Precautions
• Do not hold the focus knobs, eyepiece tube, lamphouse, stage, and so on, when carrying the microscope.
They may come off and may cause serious injury or malfunction.
• Be careful not to pinch your hands or fingers during transportation.
8. Cable routing
Make sure the cables are routed properly. Do not bring the cables into contact with the lamphouse. If a cable
comes into contact with the lamphouse, the cable sheath may melt and it may result in an electrical shock or
fire. Additionally, connect the cables by placing them into the cable keeper on the rear of the microscope main
body.
6
Contents
Introduction ....................................................................................................................................................................... 1
Safety Precautions............................................................................................................................................................ 2
Warning and Caution Symbols Used in This Manual ................................................................................................. 2
Meaning of Symbols Used on the Product ................................................................................................................. 2
WARNING .......................................................................................................................................................... 3
CAUTION ........................................................................................................................................................... 5
2. Microscopy ............................................................................................................................................................. 14
2.1 Bright-field Microscopy under the Episcopic Illumination .............................................................................. 16
2.2 Dark-field Microscopy under the Episcopic Illumination ................................................................................ 19
2.3 Polarization Microscopy under the Episcopic Illumination (simplified/sensitive color)................................... 20
2.4 Differential Interference Contrast Microscopy under the Episcopic Illumination............................................ 21
2.5 Epi-fl Microscopy........................................................................................................................................... 22
2.6 Bright-field Microscopy under the Diascopic Illumination .............................................................................. 24
2.7 Polarization Microscopy under the Diascopic Illumination (simplified/sensitive color)................................... 28
3. Operation Details.................................................................................................................................................... 29
3.1 Power ON/OFF ............................................................................................................................................. 29
3.1.1 Power of the microscope............................................................................................................... 29
3.1.2 Power supply of the lamp .............................................................................................................. 29
3.2 Illumination.................................................................................................................................................... 30
3.2.1 Brightness control and illumination ON/OFF ................................................................................. 30
3.2.2 Switching the Internal/External brightness control......................................................................... 30
3.2.3 Displaying the POWER LED ......................................................................................................... 31
3.3 Selecting the Microscopy Method ................................................................................................................. 31
3.4 Eyepiece Tube .............................................................................................................................................. 33
3.4.1 Selecting optical path .................................................................................................................... 33
3.4.2 Adjusting the eyelevel risers.......................................................................................................... 34
3.5 Adjusting the Interpupillary Distance............................................................................................................. 34
3.6 Adjusting the Diopters................................................................................................................................... 35
3.7 Adjusting the Focus (for focus operation) ..................................................................................................... 36
3.7.1 Using the coarse/fine focus knob .................................................................................................. 36
3.7.2 Adjusting the torque for the coarse focus ring ............................................................................... 36
3.8 Placing the Sample and Operating the Stage ............................................................................................... 37
3.8.1 Placing the sample........................................................................................................................ 37
3.8.2 Changing the Observation Position............................................................................................... 38
3.9 Operating the Revolving Nosepiece and the Objective................................................................................. 39
3.9.1 Revolving nosepiece in combination with the objective................................................................. 39
3.9.2 Changing the objectives ................................................................................................................ 39
3.9.3 Displaying the address for the revolving nosepiece ...................................................................... 39
3.10 Filter.............................................................................................................................................................. 40
3.11 Adjusting the Field Diaphragm (for the episcopic Illumination) ..................................................................... 41
3.12 Adjusting the Aperture Diaphragm (for the episcopic illumination) ................................................................ 42
3.13 Using the Polarizer/Analyzer Unit (MA2-PA/MA2-UPA) ................................................................................ 43
3.13.1 Inserting/removing the polarizer/analyzer from the optical path .................................................... 43
7
Contents
4. Assembly ................................................................................................................................................................ 68
4.1 About the System.......................................................................................................................................... 70
4.2 Combination List for the Unit......................................................................................................................... 71
8
Contents
5. Troubleshooting ..................................................................................................................................................... 98
5.1 Viewing Problems and Control Problems...................................................................................................... 98
5.2 Electrical System Problems ........................................................................................................................ 101
7. Specifications....................................................................................................................................................... 104
9
1 Part Name and Function
Eyepiece Sample holder
10X, 12.5X, 15X (and 10X The stage is equipped with the standard
are equipped with the sample holder (with a sample clip). (Refer to
mask eyepiece.) 3.8.)
Eyepiece tube
(Figure illustrates Sample clip
the MA2-TI3.)
Binocular eyepiece
may also be used.
(Refer to 3.4, 3.5,
and 3.6.)
Stage (Figure illustrates the MA2-SR.)
(Refer to 3.8.)
Objectives
(Refer to 3.9.)
Stage (forward/backward)
movement knob for the Y
direction
10
Chapter 1 Part Name and Function
POWER LED
Displays ON/OFF for the power switch or the light-status for the illumination
lamp. Additionally, the display status remains the same when an external
PC is in use.
Power switch OFF: LED turns off
Power switch ON; Brightness control dial OFF: Orange LED lights up
Power switch ON; Brightness control dial ON: Green LED lights up
Note: The Power LED Message is only compatible with the lamphouse of
the MA200 main body for the episcopic illumination, and the HG
precentered fiber illuminator (C-HGFIE motorized type).
Address display
Address of the objective in the optical path lights up.
11
Chapter 1 Part Name and Function
Air vents
Cable keeper
Keeps cables inside.
12
Chapter 1 Part Name and Function
Lamphouse connector
[LAMP DC12V50W]
Connects the cord on the epi
illumination lamphouse (12V
50W Precentered Lamphouse).
Internal/External brightness
RS232C connector [HGFIE] control changeover switch
[EXTERNAL ON/OFF]
Connects the C-HGFIE HG
Precentered Fiber Illuminator. (Refer to 3.2.2, 3.22.1, and
3.23.1.)
ON: Adjusts the brightness on
PC;
RS232C connector [NCNT] OFF: Adjusts the brightness on
Connects the LV-NCNT2 the microscope.
control unit for the motorized
nosepiece.
MA2-TI3 Trinocular Eyepiece Tube ESD (erect image) MA2-TI3 Trinocular Eyepiece Tube
MA2-NUI5 Intelligent Universal Quintuple MA2-NUI5 Revolving Nosepiece
Nosepiece ESD
MA2-MC Magnification Module MA2-MC Intermediate Magnification Unit
Fl Filter Block MA2-FL G General name: MA2-FL Fluorescent Unit
(Other types than G are available.)
MA2-GR Grain Size Slider MA2-GR Grain Scale Slider
MA2-PA Polarizing Filter Cube MA2-PA Polarizer/Analyzer Unit
MA2-UPA Polarizing Filter Cube with 1/4λPlate MA2-UPA Polarizer/Analyzer Unit
MA2-SRSH 10 Specimen Holder General name: Sample Holder
(Other types than SRSH 10 are available.)
13
2 Microscopy
This chapter describes the procedure of each microscopy.
See Table 2.1 for the items required for each microscopy.
• Refer to “4. Assembly,” when the product has not been assembled yet.
• For detailed information about operations of parts of the product, refer to “3. Operation Details.”
• Notes on using an external light source:
The procedures described on the episcopic illumination in 2.1 to 2.5 are based on use of the light source of
the MA200 main body with the LV-LH50PC Lamphouse attached. When using an external light source to
perform the microscopy, follow the notes below.
• The procedure on how to turn the power switch to ON: Turn on the power switch on the external device
in advance, then turn on the MA200 main body.
• Setting the Internal/External brightness control changeover switch of the MA200 main body: Sets to ON
(external mode).
• Adjusting the brightness of the light source: Adjusts the brightness on the external light source.
• To perform the epi-fl microscopy using the C-HGFI/C-HGFIE HG Precentered Fiber Illuminator as the
external light source, screw the compensation filter supplied with the fiber adapter into the MA2-FL
fluorescent unit. (Refer to 4.15.1.)
14
Chapter 2 Microscopy
DIC microscopy p.21 to p.22 5 revo., 6 revo. LU EPI objectives or LU BD For 5 revo.:
under the epi objective only • MA2-PA
illumination • For EPI objectives with 5 polarizer/analyzer unit
revo., LU nosepiece • L-DIHC/L-DIC slider
LV-LH 50PC
adapter M32-25 required (single NR method)
Lamphouse or
C-HGFI/C-HGFIE • BD objective unavailable • For the sensitive color
HG Precentered for 6 revo. polarization microscopy,
Fiber Illuminator the MA2-λP λplate
available
For 6 revo.:
• MA2-UPA
polarizer/analyzer unit
• LV-DIHC/LV-DIC slider
(Senarmont method)
• For the sensitive color
polarization microscopy,
the MA2-λP λplate
available
15
Chapter 2 Microscopy
1-1
16
Chapter 2 Microscopy
Set the microscope for the bright-field microscopy under the episcopic illumination.
If the current unit is the one shown in the column at the right edge under “Other items required” in Table 2.1,
pull out each item from the optical path.
1 Push in the optical path changeover lever for the
2-4 eyepiece tube and select 100% for the binocular
eyepiece. (See 3.4.1.)
2-1
2-2
2 Push in the eyepiece tube/back port changeover
lever to select 100% for the eyepiece tube. (See
3.4.1.)
2-5, 6
3 Push in the BD field changeover lever to select the
“BF (bright-field)” position. (See 3.3.)
2-4 4 Turn the revolving nosepiece to place the 10X
objective into the optical path. (See 3.9.2)
The nosepiece addresses (1 to 7) on the
2-6 front-display light up to indicate the position of the
revolving nosepiece. (See 3.9.3.)
5 Place the NCB11 filter in the filter turret into the
2-7 2-3 optical path and compensate color temperature.
(See 3.10)
6 Adjust the brightness roughly with the brightness
control dial and the desired ND filter in the filter
turret. (See 3.2.1 and 3.10.)
7 Turn the field diaphragm dial and the aperture
diaphragm dial counter-clockwise to the limit to
fully open the field diaphragm and the aperture
diaphragm for the episcopic illumination. (See 3.11
and 3.12.)
Place the sample onto the stage and adjust the focus and the brightness.
1 Use the sample holder according to the selected
3-1 sample. Place the sample onto the stage and
move the stage using the stage movement knob
for the X/Y direction so that the observation
position comes to the center of the view-field. (See
3.8.)
2 Turn the coarse/fine focus knobs and focus on the
target. (See 3.7.)
3 Turn the brightness control dial to adjust the
brightness of the episcopic illumination. (See
3-2 3.2.1.)
3-3
3-1
17
Chapter 2 Microscopy
18
Chapter 2 Microscopy
Attach the items required for the dark-field microscopy under the episcopic illumination
to the microscope. (See Table 2.1.)
Focus on the sample with the bright-field microscopy under the episcopic illumination.
(See Pages 16 and 17.)
Set the microscope for the dark-field microscopy under the episcopic illumination.
1 Turn the revolving nosepiece to place the
3-1
objective of a desired magnification into the optical
path. (See 3.9.2.)
2 Pull out the BD field changeover lever to select the
“DF (dark-field)” position. (See 3.3.)
3-4 The aperture diaphragm and the field diaphragm
are fully opened. (The dial positions do not
change.)
When the polarizer/analyzer or the fl filter is
placed into the optical path, it is automatically
excluded.
3-3 3 Turn the brightness control dial to adjust the
brightness of the episcopic illumination. (See
3.2.1.)
4 Adjust the brightness with the desired ND filter in
3-2
the filter turret. (See 3.10.)
19
Chapter 2 Microscopy
Attach the items required for the polarization microscopy under the episcopic
illumination to the microscope. (See Table 2.1.)
Focus on the sample with the bright-field microscopy under the episcopic illumination.
(See Pages 16 to 17.)
Set the microscope for the polarization microscopy under the episcopic illumination.
1 Turn the revolving nosepiece to place the
3-4 3-1 objective of a desired magnification into the optical
path. (See 3.9.2.)
2 Push in the polarizer/analyzer unit to the second
click-stop position and place it into the optical
3-6 path. (See 3.13.1.)
3 Turn the polarizer rotation ring to the crossed
Nicol’s position. (See 3.13.2.)
4 To perform the sensitive color polarization
microscopy, insert the λplate into the
polarizer/analyzer unit and place it into the optical
3-5 path. (See 3.14.1.)
5 Turn the brightness control dial to adjust the
brightness of the episcopic illumination. (See
3-3, 7 3-2
3.2.1.)
6 Adjust the brightness with the desired ND filter in
the filter turret. (See 3.10.)
7 Turn the polarizer rotation ring to adjust the
polarization while observing the image.
20
Chapter 2 Microscopy
Attach the items required for the differential interference contrast (DIC) microscopy
under the episcopic illumination to the microscope. (See Table 2.1.)
Focus on the sample with the bright-field microscopy under the episcopic illumination.
(See Pages 16 and 17.)
Set the microscope for the DIC microscopy under the episcopic illumination.
1 Turn the revolving nosepiece to place the
3-5 3-1 objective of a desired magnification into the optical
3-4 path. (See 3.9.2.)
2 Push in the polarizer/analyzer unit to the second
click-stop position and place it into the optical
3-7 path. (See 3.13.1.)
3 Turn the polarizer rotation ring to the crossed
Nicol’s position (See 3.13.2.)
4 Attach the DIC slider to the slot on the revolving
nosepiece to place the DIC prism into the optical
path. (See 3.15.1 or 3.16.2.)
3-6 5 Set the interference color. (See 3.15.3 or 3.16.3.)
For using the L-DIHC/L-DIC slider, turn the prism
movement knob, whereas for using the
3-3, 5 3-2 LV-DIHC/LV-DIC slider, turn the polarizer rotation
ring of the polarizer/analyzer unit. In either case,
the interference color changes continuously from
dark color, gray to sensitive red-violet.
6 Turn the brightness control dial to adjust the
brightness of the episcopic illumination. (See
3.2.1.)
7 Adjust the brightness with the desired ND filter in
the filter turret. (See 3.10.)
21
Chapter 2 Microscopy
Attach the items required for the epi-fl microscopy to the microscope. (See Table 2.1.)
To perform the epi-fl microscopy using the C-HGFI/C-HGFIE HG Precentered Fiber Illuminator as the
external light source, screw the compensation filter supplied with the fiber adapter into the MA2-FL
fluorescent unit. (Refer to 4.15.1.)
Find the target and focus on the sample by BF/DF microscopy under the episcopic
illumination. (See Pages 16 to 17, and 19.)
3-3 3-2
22
Chapter 2 Microscopy
23
Chapter 2 Microscopy
Attach the items required for the bright-field microscopy under the diascopic illumination.
(See Table 2.1.)
Turn on the power to light up the diascopic illumination lamp. (See 3.23.1.)
1 Set the brightness control dial to OFF on the
MA200 main body. (See 3.2.1.)
2 Turn on the TI-PS100W power supply connected
to the supporting pillar for dia-Illuminator.
3 Set the MA200 main body to ON.
The power LED on the front is lit.
The light intensity control dial for the TI-PS100W
2-4
power supply is enabled.
4 Turn the light intensity control dial on the
TI-PS100W power supply to adjust the
illumination.
You can change the output voltage within 1V to
12V. When the output voltage is set to around 9V,
2-2
the ideal optical color reproduction generates.
2-3 2-1
2-3
24
Chapter 2 Microscopy
Set the microscope for the bright-field microscopy under the diascopic illumination.
When the “Other items required” (See the column at the right edge of Table 2.1.) is used for the episcopic
illumination, remove each attachment from the optical path in advance.
25
Chapter 2 Microscopy
4-1
6-1
6-2
26
Chapter 2 Microscopy
8-1
If accurate color reproduction is not crucial (e.g., for
color photography), change the lamp voltage with
8-5 the light intensity control dial on the TI-PS100W
power supply.
5 Look into the eyepieces. Move the stage to bring
8-6
the observation target into the center of the
view-field.
6 Turn the coarse/fine focus knobs and focus on the
sample.
27
Chapter 2 Microscopy
Attach the items required for the polarization microscopy under the diascopic
illumination to the microscope. (See Table 2.1.)
Focus on the sample with the bright-field microscopy under the diascopic illumination
and center the condenser. (See pages 23 to 26.)
Set the microscope for the polarization microscopy under the diascopic illumination.
1 Turn the revolving nosepiece to place the
objective of a desired magnification into the optical
3-6 path. (See 3.9.2.)
2 Push the polarizer/analyzer unit attached to the
MA200 main body into the optical path. (See
3.13.1.)
3-3, 4
When using the sextuple revolving nosepiece,
insert the D-DA analyzer slider into the slot on the
revolving nosepiece instead of the
3-5
polarizer/analyzer unit, if desired. (See 3.17.1.)
3 Place the T-P2 polarizer into the optical path and
make a crossed Nicol’s position. (See 3.23.8.)
4 To perform the sensitive color polarization
3-1
microscopy, place the TI-DIC λplate into the
optical path. (See 3.23.8.)
When using the sextuple revolving nosepiece, use
the D-LP λplate instead of the TI-DIC λplate, if
3-2 desired. (See 3.18.1.)
5 Turn the light intensity control dial on the TI-PS
100W power supply to adjust the brightness of the
diascopic illumination.
6 Push in or pull out the desired ND field for the
supporting pillar for dia-Illuminator to place it into
the optical path, then adjust the brightness for the
view-field. (See 3.23.5.)
28
3 Operation Details
The product has a built-in power supply circuit for the halogen lamp. When the specified lamphouse
(LV-LH50PC) is attached, the lamp is lit in connection with the power supply for the microscope.
29
Chapter 3 Operation Details
3.2 Illumination
Brightness range
OFF Index
position
30
Chapter 3 Operation Details
POWER LED
LED lights off: POWER switch OFF
Orange LED lights up: POWER switch ON; brightness control dial OFF
Green LED lights up: POWER switch ON; brightness control dial ON
The setting options for the BD field changeover lever or the settings for the optional items required for this
microscope is shown in Table 3.3-1.
31
Chapter 3 Operation Details
32
Chapter 3 Operation Details
• The light distribution for both the binocular and vertical tube parts of the MA2-TI3 trinocular eyepiece tube
can be performed using the optical path changeover lever.
Optical path
changeover lever
• The light distribution for the eyepiece tube part and the back port can be selected with the eyepiece
tube/back port changeover lever.
33
Chapter 3 Operation Details
Eyelevel riser
If you look through the objective that is 10X or
less than 10X when two eyelevel risers are
attached on the microscope, the rim of the
view-filed may look different from the one without
an eyelevel riser due to reduction of light
intensity.
Figure 3.4-3 Adjusting the eyelevel risers
Adjust the interpupillary distance or the distance between the eyepieces according to user’s eye distance.
Before adjusting the interpupillary distance, follow the
procedure described in “2.1 Bright-field Microscopy
under the Episcopic Illumination” or “2.6 Bright-field
Microscopy under the Diascopic Illumination,” and
focus on the sample with the 10X objective.
Look through the eyepieces and adjust the binocular
part until the view fields for the right and left eyes
coincide.
This will facilitate the observation with both eyes.
The binocular part has a scale for interpupillary
distance. It is recommended to memorize or record
your interpupillary distance, so that the distance
between the eyepieces can be readily adjusted next
Figure 3.5-1 Adjusting the interpupillary distance time.
34
Chapter 3 Operation Details
Adjust the diopter according to the user's eyesight using the diopter adjustment rings on the eyepieces.
Diopter adjustment corrects the difference in the left and right eyesight. This adjustment facilitates binocular
observation and minimizes focal deviation when switching objectives. Make sure to adjust the diopter
adjustment rings on both eyepieces.
35
Chapter 3 Operation Details
36
Chapter 3 Operation Details
■ Sample holder
To use the sample holder, place it onto the stage.
The MA2-SR rectangular stage is supplied with the sample holder (outer diameter: φ108 mm, thickness: 2 mm,
teardrop-shaped aperture: φ22 mm). This sample holder is also supplied with the sample clip to hold the
sample set on the stage. Use the clip to prevent the sample from tipping over or being moved.
The sample holder is supplied with a rotation knob. If you turn the ring holding the rotation knob, you can turn
the sample with the observation position unmoved.
Various options are available for the sample holder as shown below. Select an appropriate option according
to the size or observation method of the sample used.
Item Model Remarks
Sample holder Standard Holder Outer diameter φ108 mm; thickness: 2 mm; teardrop-shaped
(attached to stage) aperture: φ22 mm; supplied with the rotation knob and a sample
clip
MA-SRSH 40 φ40 mm, supplied with the rotation knob and a sample clip
MA-SRSH 10 φ10 mm, supplied with the rotation knob and a sample clip
MA-SRSH 25-40 φ25×40 mm, supplied with the rotation knob and a sample clip
MA2-SRSH Holder set (φ23, 30, 36 mm) supplied with the rotation knob and
a sample clip
MA-SRSH1 φ15 mm
37
Chapter 3 Operation Details
38
Chapter 3 Operation Details
For the MA200, you can use various revolving nosepieces including the quintuple, sextuple or septuple
revolving nosepiece to attach the industrial CFI objective lens. The combination varies depending on the
microscopy applied. Refer to “4.2 Combination List for the Unit” to select an appropriate combination.
■ LU nosepiece adapter M32-25
The adapter runs to switch the screw diameter from large to small for the objective hole on the revolving
nosepiece. To attach EPI objectives on the quintuple revolving nosepiece, screw the LU nosepiece adapter
M32-25 into the hole on the revolving nosepiece.
39
Chapter 3 Operation Details
3.10 Filter
Filter turret 2
40
Chapter 3 Operation Details
Field diaphragm
Field diaphragm knob centering holes
41
Chapter 3 Operation Details
42
Chapter 3 Operation Details
Second click-stop position Attach the polarizer/analyzer unit to the MA200 main
body, then use it for the simplified polarization
First click-stop position
microscopy under the episcopic/diascopic
Polarizer rotation ring
illumination, and the differential interference contrast
microscopy under the episcopic illumination.
Polarizer MA2-PA or MA2-UPA is equipped with both the
polarizer and the analyzer with the slot provided to
Analyzer insert the λ plate for the sensitive color microscopy
(see 3.14). A single push or pull of the unit allows you
to insert or remove both the polarizer and analyzer
from the optical path at the same time.
The MA2-UPA is equipped with the 1/4 λ plate. If
Slot for λ plate
using the MA2-UPA for polarization observation, the
Figure 3.13-1 Polarizer/analyzer unit low-contrast sample such as aluminum can be
observed with clear contrast.
For the simplified polarization microscopy under the
diascopic illumination, only analyzer of the unit is
used.
43
Chapter 3 Operation Details
Dark cross
44
Chapter 3 Operation Details
λ plate
λ plate slider
45
Chapter 3 Operation Details
Prism position index Attach the L-DIHC/L-DIC DIC slider to the quintuple
nosepiece (MA2-NUI5 or LV-NU5A) in combination
Prism position with the MA2-PA polarizer/analyzer unit to perform
Sliding limit groove
the DIC microscopy under the episcopic illumination
in single NR method.
The L-DIHC slider is designed for high contrast
observation.
First
click-stop
Prism setting dial position
Prism movement
knob Second click-stop
position
46
Chapter 3 Operation Details
You can change the interference color continuously by turning the prism movement knob.
Interference color Characteristics
Dark color Observations similar to the dark-field microscopy can be performed.
Gray Phase contrast distribution of the whole sample with a bird’s eye view can be
observed.
Sensitive red-violet Observations with the highest color contrast can be performed.
Both the LV-DIHC and the LV-DIC sliders have A and B types to be selected, according to the objective.
The industrial LU objective has a character (A or B) showing the DIC prism position, which is indicated next to
the magnification/N.A. value on the objective barrel (refer to Figure 3.15-3). Select the DIC slider as indicated
on the objective. In Figure 3.15-3, the character “A” is indicated next to the magnification/ N.A. value. Use the
A-type DIC slider as indicated.
Push the slider into the revolving nosepiece to the limit and the DIC prism enters the optical path.
To remove the DIC prism from the optical path, remove the slider from the revolving nosepiece.
47
Chapter 3 Operation Details
You can change the interference color continuously by adjusting the orientation of the polarizer. (Refer to
3.12.3.)
Interference color Characteristics
Dark color Observations similar to the dark-field microscopy can be performed.
Gray Phase contrast distribution of the whole sample with a bird’s eye view can be
observed.
Sensitive red-violet Observations with the highest color contrast can be performed.
Analyzer
Attach the D-DA analyzer slider to the sextuple
revolving nosepiece (D-ND6) to perform the simplified
polarization microscopy under the diascopic
illumination.
First click-stop The D-DA analyzer is not required when the
position polarizer/analyzer unit (MA2-PA/MA-UPA) is attached
to the microscope during the simplified polarization
Second click-stop
microscopy under the diascopic illumination.
position
48
Chapter 3 Operation Details
λ plate slider
49
Chapter 3 Operation Details
Second click-stop position Attach the fluorescent unit to the MA200 main body to
perform the epi-fl microscopy.
First click-stop position
A fluorescent unit consists of two types of optical
Compensation filter components: an excitation light filter (EX filter), a
attachment barrier filter (BA filter). You can use the four types of
combination, "G", "B", "BV" or "V" as shown in the
Excitation table below. Select the fluorescent unit suitable for
light filter your purpose and sample.
Barrier filter
To perform epi-fl microscopy using the
C-HGFI/C-HGFIE HG Precentered Fiber
Illuminator as the external light source, screw the
Figure 3.19-1 Fluorescent unit compensation filter supplied with the fiber
adapter into the filter attachment of the
fluorescent unit. (Refer to 4.15.1.)
50
Chapter 3 Operation Details
51
Chapter 3 Operation Details
A barrier filter transmits only fluorescent lights emitted by the sample but blocks the excitation lights. This filter
makes it possible to observe the fluorescent image without unnecessary light (that is, on a dark background).
There are two types of barrier filters: LP filters (long-pass filters), which block all lights that are shorter than a
certain boundary wavelength and allow all lights to pass that are longer than the boundary wavelength, and
BP filters (band-pass filters), which allow only lights in a certain bandwidth to pass. The LP filter is a barrier
filter that is provided on the MA200.
The boundary to block or transmit lights using an LP filter is called the cut-on wavelength.
An excitation light is a light that is irradiated to the sample. The fluorophore in the sample absorbs the
excitation light energy. As a result, fluorescent lights are emitted from the fluorophore instead. When a sample
is labeled with a fluorophore that emits fluorescent lights of very close wavelengths to the excitation light,
select a barrier filter with the shortest cut-on wavelength permitted by performance requirements for efficient
fluorescent microscopy.
A longer cut-on wavelength tends to result in a more complete separation between an excitation light and
fluorescent lights, rendering a darker background of the fluorescent image. With the recent advancement in
filter performance, however, filters with shorter cut-on wavelengths can be used for this purpose and they are
used more often than before.
52
Chapter 3 Operation Details
Attach the MA2-GR/MA2-MR slider to the MA200 main body to measure steel structures or to photograph
images.
The following two types of scales are provided for a single slider for Austenitic steel measurement.
• Austenitic steel measurement scale (ASTM E112-63 reticle numbers 1 to 8)
Compare the reticle pattern (numbers 1 to 8) with the sample-structure size in order to measure the grain
size. Normally, the 10X objective is used in measuring the grain size. To use the 20X objective, select the
reticle number two scales up, whereas to use the 5X objective, select the reticle number two scales down.
• Lattice scale (0.5 mm intervals; Vertical/Horizontal: 20 each)
The lattice scale is normally used for the Nonmetallic Inclusion test, and it can also be used for simple
lattice-measurement. The lattice, whose intervals on a reticle indicate 0.5 mm, consists of 20 vertical or
horizontal lines. When the 10X objective is used, the lattice intervals equivalent to those on the sample
become 0.05 mm.
Austenitic steel When inserted, the grain scale slider is pressed down
measurement scale by a leaf spring installed inside of the slot to adjust
the desired inserting position. To place the scale in
the optical path, use the icons to indicate the current
scale type on the slider and the positioning index line.
Lattice scale
53
Chapter 3 Operation Details
* When inserting the slider into the slot, face its side with the indication toward the front of the microscope.
* To capture the scale using the camera, adjust the desired inserting position by checking the actual scale
position or area on the DS-L2 monitor or a PC monitor.
* Reticle patterns may not be clearly visible on a dark background as in the case of dark-field microscopy.
* Do not touch or press hard against the glass. It may cause damage to the scale reticle.
* Remove the dust on the glass with a soft blower or brush.
54
Chapter 3 Operation Details
Magnification
Magnification changeover turret
55
Chapter 3 Operation Details
HG Precentered
Fiber Illuminator
56
Chapter 3 Operation Details
Attach the supporting pillar for dia-Illuminator 100W to the MA200 main body to perform the microscopy under
the diascopic illumination.
To perform the microscopy, mount the D-LH/LC Lamphouse and condenser etc. to the supporting pillar for
dia-Illuminator 100W, and connect the power supply (TI-PS 100W) to the lamphouse.
Lamphouse
Filter slider
Supporting pillar for
dia-Illuminator 100W
Condenser mount
Condenser
Power supply
57
Chapter 3 Operation Details
To perform the diascopic illumination for the first time or to replace the current condenser, focus and center
the condenser so that the light through the condenser is focused on the correct position of the sample surface
(at the center of the optical path).
1 Be sure to focus the 10X objective on the sample
as described in “2.6 Bright-field Microscopy under
the Diascopic Illumination.”
Field
diaphragm 2 Stop down the field diaphragm by turning the field
dial diaphragm dial on the supporting pillar for
dia-Illuminator 100W until the field diaphragm
Condenser image comes into the view-field.
focus knob
3 Turn the condenser focus knob of the supporting
pillar for dia-Illuminator to focus on the field
Condenser
diaphragm image.
focus knob
4 Turn the condenser centering screws (2 for both
Condenser
sides) for the supporting pillar for dia-Illuminator to
centering
screws move the field diaphragm image to the center of
the view-field.
Field diaphragm image
5 Place the 50X objective into the optical path.
View-field
6 Adjust the size of the field diaphragm image to
closely match the size of the view-field, by turning
the field diaphragm dial on the supporting pillar for
dia-Illuminator.
Move the field diaphragm image into the center of the 7 Turn the condenser centering screws on the
view-field. Adjust its size to match the view-field. supporting pillar for dia-Illuminator to move the
Figure 3.23-3 Focusing and centering field diaphragm image to the center of the
the condenser view-field.
* To observe the sample, turn the field diaphragm
dial so that the field diaphragm image
circumscribes the view-field. (Perform this
adjustment for each objective.)
58
Chapter 3 Operation Details
The field diaphragm is used to limit the irradiation area of the lamp to the view-field of the microscope.
View the supporting pillar for dia-Illuminator 100W from above, then turn the field diaphragm dial
counter-clockwise to open the field diaphragm, or turn the field diaphragm dial clockwise to stop down the
diaphragm. Normally, stop down the field diaphragm to a size that circumscribes (or inscribes) the view-field.
An unnecessarily large irradiation area will result in stray light and flare, thereby reducing the contrast of the
optical image. For photomicroscopy, the field diaphragm should be stopped down to a size slightly larger than
the area to be captured. Avoid stopping down the field diaphragm too close to the frame size to be captured,
as it may result in vignetting.
* Perform the field diaphragm adjustment after completing focusing and centering for the condenser lens.
59
Chapter 3 Operation Details
60
Chapter 3 Operation Details
Filter list
Filter Description
ND2 Adjusts the brightness for general microscopy or photomicroscopy. Reduces the light
(Neutral Density intensity to 1/2. Dims the quantity of light to 1/2 (Transmittance: approx. 50%).
filter)
ND16 Adjusts the brightness for general microscopy or photomicroscopy. Dims the quantity
(Neutral Density of light to 1/16 (Transmittance: approx. 6.3%).
filter)
GIF Improves the contrast for microscopy under a monochrome light or when performing
(Green Interference monochrome imaging.
filter)
NCB11 Corrects the color temperature for general microscopy or filming by daylight color TV
(Neutral Color camera photography. Color reproduction is best when this filter is inserted into the
Balancing filter) optical path, and the lamp voltage is set to the same display as the lamp rating.
Remove from optical path when performing monochrome imaging.
Condenser mount
61
Chapter 3 Operation Details
When performing the simplified polarization microscopy under the diascopic illumination, mount the polarizer
slider on the condenser mount for the supporting pillar for dia-Illuminator 100W.
Push in the slider, and the polarizer enters the optical path.
If you loosen the rotation clamp screw and slightly turn the polarizer, the image contrast changes.
To perform the sensitive color microscopy, pull out the dust-proof plate from the polarizer slider to attach the
TI-DIC λ plate.
Clamp
Rotating section
Dummy hole
Polarizer rotation lever
Dial scale (5° intervals) Dust-proof plate/λ plate
White index mark slider
Fixing ring
Condenser turret
White index mark
Aperture diaphragm lever
Condenser lens
62
Chapter 3 Operation Details
63
Chapter 3 Operation Details
The supporting arm is designed to mount Nikon camera control unit DS-L2.
The Supporting Arm for DS-L2 is equipped with a one-touch release. If you pull up the mount arm
attaching/detaching knob, the DS-L2 can be attached or removed from the supporting arm.
Mount arm The rotation torque can be adjusted depending on the
attaching/detaching knob
Rotation
tightness of the docking section of the mount arm to
torque keep the DS-L2 steady, while achieving more
Mount arm nuts (x2) comfortable operation for angle adjustment. Turn the
two nuts on the docking sections with the supplied
tool and control each rotation torque for the
horizontal/vertical direction.
Supporting
arm For using the DS-L2, refer to the instruction manual
supplied with the DS-L2.
DS-L2
64
Chapter 3 Operation Details
Once you connect the DS Camera Control Unit DS-L2 with the DS camera attached to the vertical tube on
MA200 or the back port using the specified cable, you can operate the DS camera. Additionally, by
connecting MA200 with DS-L2 via USB cable, the following functions are available on the DS-L2.
• Displaying the microscope info menu
• Measurement corresponding to the magnification value of the microscope
• Recording the microscope info while saving the image
1 Check that the MA200 main body and the DS-L2 are connected correctly. (Refer to “Cable Connection for
the DS-L2 and MA200” in 4.16.)
2 Set the power switch on the MA200 main body to the “|” side to turn it on.
3 Push in the power switch on the DS-L2 to turn it on.
Right-click on the mouse over the upper-screen of the DS-L2 to display the microscope info menu. (Refer to
Figure 3.25-1.)
Magnification/NA value
of the objective ■ The microscope info menu buttons
: Closes the microscope info menu.
Magnification value of : Resizes the microscope info menu.
the intermediate
magnification unit : Displays the objective info. (Refer to 3.25-2.)
65
Chapter 3 Operation Details
The nosepiece address of the motorized nosepiece that is mounted on the microscope can be controlled
from the microscope info menu on the DS-L2. To adjust the illumination of the lamphouse (LV-LH50PC)
with the DS-L2 connected, set [EXTERNAL ON/OFF] of the Internal/External brightness control
changeover switch on the rear of the MA200 main body to OFF (Internal mode), then adjust the
brightness control dial.
Once you calibrate the reference length, you can have a simple means of measuring distances of an image
on DS-L2.
When MA200 is connected with the DS-L2, you can save calibration values for each nosepiece address in
order to measure the length of a desired image. Additionally, when the intermediate magnification unit is
attached to MA200, the magnification value of the intermediate magnification unit is automatically multiplied
when measuring the length of the image.
For calibrating and measuring the length of an image, refer to the instruction manual supplied with the DS-L2.
When MA200 is connected with the DS-L2, the information on the objective, the intermediate magnification
unit, or the illumination for the episcopic microscopy, can be recorded as text data.
For setting log information or saving the image, refer to the instruction manual supplied with the DS-L2.
66
Chapter 3 Operation Details
3.25.5 Connecting with the DS camera head switcher (DS-SW: Camera switch BOX, optional)
If the DS-L2 is connected with the DS camera head switcher, two camera heads can be connected to the
DS-L2.
If the DS-L2 is connected with the DS-SW, the register tab appears on the DS-L2 microscope info menu. On
the register tab, the “Measurement interlock for the magnification value of the microscope” function can be set
to ON/OFF for each camera head.
For capturing the image and setting the camera, refer to the instruction manual supplied with the DS-L2.
67
4 Assembly
WARNING
• Before assembling the microscope, be sure to read the WARNING and CAUTION at the
beginning of this instruction manual and follow the instructions written therein.
• To prevent electrical shocks and fire, turn off the power switch (set the switch to the “c” side) and
unplug the power cord from the outlet when assembling the microscope.
CAUTION
• Be careful not to pinch your fingers or hands during assembly.
• Scratches or fingerprints on the lenses will adversely affect the microscopy image. Be careful not to
scratch or touch the lens surfaces. If lenses are contaminated with fingerprint or such, clean them
according to the procedure described in “6. Care and Maintenance.”
• This product is a precision optical instrument. Handle it carefully and do not subject it to a strong
physical shock. In particular, objectives may loose accuracy when exposed to even a weak physical
shock.
■ Required tools
• Two hexagonal screwdrivers (2 mm) (supplied with the microscope)
• Hexagonal wrench (3 mm) (supplied with the microscope)
• Hexagonal wrench (5 mm)
• Filter replacing tool (supplied with the microscope)
When these tools are not used, place them in the tool holder at the right-end of the top-surface of the product.
■ Installation location
This product is a precision optical instrument. The usage or storage in an inappropriate environment may
result in malfunctions or poor performance. Consider the following factors when selecting an installation
location:
• Avoid a brightly lit location, such as exposed to direct sunlight or directly under a room light. If there is
excessive ambient light, the image may not clearly visible.
• Always install the microscope with a surrounding clear area of 10 cm or more.
• Install the microscope in a location that is free from considerable dust or dirt.
• Install the microscope on a flat surface with little vibration.
• Install the microscope on a sturdy desk or table that is able to bear the weight of the instrument.
• Do not install the microscope in a hot and humid location.
• Arrange a layout that allows easy removal of the power cord from the inlet of the product in the event of an
emergency.
• For details about the operating environment and storage environment, refer to “7. Specifications.”
68
Chapter 4 Assembly
If you wish to buy these lamps, please contact your nearest Nikon representative.
69
Chapter 4 Assembly
Filter
(GIF, NCB11, ND2, ND16) D-LH/LC
T-DIC λ plate lamphouse
+ 12V 100W
T-P2 polarizer halogen lamp
C-mount
adapter
TI-C-LWD MC TMD2 ELWD
C-mount adapter
condenser condenser lens
LV-TV
lens Eyepiece
adapter DS-L2
monitor MA2-MP
supporting arm
for DS-L2
MA2-TI3 trinocular
eyepiece/Y-TB
MA2-MC
binocular eyepiece
intermediate
tube
magnification unit
C-ER
eyelevel riser
MA2-FL
fluorescent Unit Sample holder
(G, B, BV, V)
MA2-λP MA2-SR
λ plate rectangular stage
MA2-PA/MA2-UPA
polarizer/analyzer
unit MA200 main body
12V 50W
MA2-GR grain scale slider
halogen lamp
MA2-MR scale slider
LV-LH50PC
Lamphouse (for
EPI Objective the episcopic
LU nosepiece
BD objective illumination)
adapter M32-25
C-LHGFI
HG lamp
LV-HGFA
HG fiber
C-HGFIF
adapter
D-ND6 D-NI7 HG fiber
MA2-NUI5 LV-NU5A motorized
revolving revolving revolving nosepiece nosepiece
nosepiece nosepiece
71
Chapter 4 Assembly
1: For EPI objective on 5 revo., “LU nosepiece adapter M32-25” is required.
2: BD objective is unavailable on 6 revo., 7 revo.
3: The CFI L Plan EPI objective lens cannot be used for DIC microscopy under the epi-illumination, epi-fl microscopy.
4: D-DA analyzer is attachable for 6 revo. only.
5: D-LP λ plate is attachable for 6 revo. only.
6: L-DIC/L-DIHC slider is attachable for 5 revo. only.
7: LV-DIC/LV-DIHC slider is attachable for 6 revo. only.
72
Chapter 4 Assembly
The table below shows a unit system capable of establishing network on a PC or the DS-L2 connected to
MA200 via the USB cable.
Power cord
USB
cable
Camera cable
AC adapter/power cord
exclusive for your PC
Intermediate
magnification unit
Power cord PC
LV-LH50PC Lamphouse
DS-L2
Power cord
AC adapter
Power
Fiber
cord
adapter
USB cable
C-HGFIE
HG Precentered Fiber Illuminator
73
Chapter 4 Assembly
4.3.2 Viewing user control options and status displays on a PC or the DS-L2
The table below shows user control options and status displays of a unit that can be controlled on a PC or the
DS-L2 connected to MA200 via the USB cable.
The “/” indicates external control/status display is unavailable.
To control MA200 on a PC or to display the current status of the MA200 on a PC, use the “SDK” in the
supplied “MA200 Setup Tool” software. For details on how to install the “SDK” on a PC, refer to the
software manual supplied with the “MA200 Setup Tool.”
74
Chapter 4 Assembly
Table 4.3-1 User control options for the unit using the PC/DS-L2
PC DS-L2 Brightness
Unit Notes
Control Status Control Status control dial
1: Shutter open/close or ND select can be controlled using the remote controller (C-HGFIE-C HG controller) exclusive for the C-HGFIE.
2: The shutter open/close or ND select can be controlled if the remote controller (C-HGFIE-C HG controller) exclusive for the C-HGFIE
is used.
3: The rotation of the revolving nosepiece can be controlled on the Motorized Nosepiece Controller 2.
75
Chapter 4 Assembly
You can register the information to be checked on a PC or the DS-L2 with the “MA200 Setup” in the supplied
“MA200 Setup Tool” software. For details on how to install the “MA200 Setup” on an appropriate device, refer
to the software manual supplied with “MA200 Setup Tool.”
To establish an appropriate connection between MA200 and an external device or to use the product safely,
turn on the power as follows.
■ For using the C-HGFIE HG Precentered Fiber Illuminator (motorized type) and the Nosepiece Controller 2
1 Set the power switch on the fiber illuminator to the “|” side and wait until it starts up.
2 Set the power switch on the Nosepiece Controller 2 to the “|” side to turn it on.
3 Set the power switch on the MA200 main body to the “|” side to turn it on.
4 To turn off the dia pillar illuminator, set the power switch to the “c” side in the sequence of the fiber
illuminator, the MA200 main body, the Nosepiece Controller 2.
■ For using the MA2-DP Supporting Pillar for Dia-Illuminator 100W under the diascopic illumination
1 Set the brightness control dial on the MA200 main body to OFF.
2 Set the power switch on the power supply (TI-PS100W) to the “|” side to turn it on.
3 Set the power switch on the MA200 main body to the “|” side to turn it on.
76
Chapter 4 Assembly
CAUTION
• To prevent electrical shock and damage to the microscope, always turn off the power switch (set the
switch to the “O” side) and unplug the power cord from the outlet before attaching or detaching the
lamphouse or replacing the lamp.
• To prevent burn injury, allow the lamp and the lamphouse to cool down completely (for at least 30
minutes after the lamp is turned off), before replacing lamps or attaching/detaching the lamphouse.
• Never touch the glass surface of the lamp with bare hands. Doing so will cause fingerprints, grease,
etc., to generate ghost images on the lamp surface, reducing the illumination. If you do get any
fingerprints or dirt on the lamp, wipe them clean.
• Make sure the lamphouse cover is securely fitted to the lamphouse after replacing lamps. Never
turn on the lamp with the lamphouse cover removed.
• When you dispose of the replaced lamp, do not break it. Instead, dispose of the used lamp as
industrial waste or dispose of it according to the local regulations and rules.
77
Chapter 4 Assembly
* The lamp exclusive for this microscope is the LV-HL50W 12V 50W longlife halogen lamp. The 12V 50W
shortlife halogen lamp from other manufacturers (OSRAM HLX 64610, OSRAM HLX64611, PHILIPS
7027) may also be used. If you wish to buy these lamps, please contact your nearest Nikon representative.
The following instructions are applied to all revolving nosepieces to attach the microscope.
* For using the MA2-NUI5, D-ND6, or D-NI7 manual revolving nosepiece, be sure to provide a cable
connection between the revolving nosepiece and the microscope.
* For using the LV-NU5A motorized nosepiece, be sure to provide a cable connection for the LV-NCNT2
Nosepiece Controller 2.
Hole to pull out the 1 Fully loosen the nosepiece clamp screw located
Nosepiece clamp screw nosepiece cable on the front of the mount using the supplied
hexagonal screwdriver.
2 Align the revolving nosepiece with the dovetail at
the mount and insert it from the front on the
microscope to push it into the limit.
3 Fix the revolving nosepiece with the nosepiece
clamp screw.
4 Adhere the address sticker for the revolving
nosepiece supplied with the MA200 main body.
Look down at the revolving nosepiece, then
adhere the address sticker clockwise in ascending
order starting with the revolving address1.
Sticker of the nosepiece The “|” sticker is adhered on the address 1 for the
Nosepiece cable I address 1 on the MA2-NUI5 revolving nosepiece. (See the figure on
MA2-NUI5 revolving the left.)
nosepiece
5 Pass the nosepiece cable through the hole on the
inner-rear of the microscope.
78
Chapter 4 Assembly
4.5.2 Cable connection for the manual revolving nosepiece and the MA200
Connect the cable of the revolving nosepiece to the connector on the microscope.
1 Loosen the rear-panel clamp screw on the
microscope with the supplied hexagonal
screwdriver to remove the rear-panel.
2 Connect the cable of the revolving nosepiece to
the connector on the microscope.
Connect the cable of the MA2-NUI5 revolving
nosepiece to the round connector on the
microscope.
Square connector
Connect the cable of the D-ND6/D-NI7 revolving
nosepiece to the square connector.
Round connector
3 Secure the clamp screw to reattach the
rear-panel.
* The address information for the revolving nosepiece is displayed on the MA200 front display via cable
connection. Additionally, the address information can be output to a device including a PC or the DS-L2 via
USB connection.
Make sure to turn off the motorized Nosepiece Controller 2 to attach or remove the signal cable.
79
Chapter 4 Assembly
4.6 Stage
• To attach or remove the stage, hold the stage securely to prevent from being fallen.
• To remove the stage from the microscope with the supporting pillar for dia-Illuminator 100W
mounted, be sure that the stage does not come into contact with the condenser, etc.
• A cross roller guide is provided to move the stage in the X/Y direction. The Y direction movement
rack is located at the bottom of the stage. Do not touch the rack during your operation, as you may
get injured.
Y direction
movement rack
The MA2-SR rectangular stage is supplied with a sample holder (outer diameter: φ108 mm, Teardrop-shaped
aperture: φ22 mm, with a detachable sample clip). Attach the sample holder onto the mount part on the
center-stage. Other types of sample holders are also available (see 3.8.1). Remove the standard sample
holder from the center-stage to attach a desired sample holder instead. Additionally, the rotation knob is
supplied with the sample holder (except MA-SRSH1) and screw the knob into the hole on the ring.
80
Chapter 4 Assembly
4.7 Objectives
4.9 Eyepieces
81
Chapter 4 Assembly
4.11 Filters
You can attach the φ25-mm, up to 3-mm thick filter on the optional slot for the filter turret 1 or 2.
When handling filters, put on gloves and do not touch the surface of filters with bare hands. And be
careful not to let dust or fingerprints get on them.
82
Chapter 4 Assembly
For using the DIC slider in single NR method, attach the MA2-NUI5 revolving nosepiece or LV-NU5A
motorized nosepiece.
1 Loosen the DIC slider limit screw on the
Sliding limit groove nosepiece using a hexagonal screwdriver.
2 Remove the slot from the revolving nosepiece to
insert the DIC slider.
3 Screw in the DIC slider limit screw.
When removing the DIC slider from the nosepiece,
fully loosen the DIC slider limit screw using a
Prism setting dial hexagon screwdriver, and then pull out the slider.
Prism movement knob
Quintuple nosepiece
83
Chapter 4 Assembly
Attach the DIC slider in Senarmont method to the D-ND6 revolving nosepiece.
Attach the D-DA analyzer and the D-LP λ plate to the D-ND6 revolving nosepiece.
λ plate
Sextuple nosepiece
84
Chapter 4 Assembly
85
Chapter 4 Assembly
86
Chapter 4 Assembly
You can attach the HG precentered fiber illuminator (for the episcopic illumination) or the supporting pillar for
dia-Illuminator 100W as an external light source.
CAUTION
• To use an external light source, carefully read the instruction manual and make sure to follow the
instructions.
• To assemble and connect parts, check that the power supplies for the light source and the product
are turned off and that the power cord is unplugged from the outlet.
WARNING
• An HG precentered fiber illuminator emits very strong light including ultraviolet light that is harmful to
the eyes and skin. Never turn on the power for the light source before completion of assembling and
connecting parts.
• To prevent a fire of flash
To attach the HG precentered fiber illuminator to the microscope with the LV-NU5A motorized
nosepiece mounted, be sure to use the motorized HG precentered fiber illuminator. Because it is
necessary to control the shutter according to the rotation of the motorized nosepiece, connect the
fiber illuminator and [HGFIE] connector on the microscope using the RS232C cross cable supplied
with the C-HGFIE. If you fail to make this connection, a fire of light may enter in the view-field at the
rotation of the motorized nosepiece.
87
Chapter 4 Assembly
When working, hold the dia pillar illuminator to prevent it from falling.
Clamp screw
Fall-stop screw
88
Chapter 4 Assembly
Condenser lens
89
Chapter 4 Assembly
Condenser
Condenser clamp screw
indication
position
ELWD-S condenser
Attach the D-LH/LC Lamphouse to the supporting pillar for dia-Illuminator 100W.
1 Insert the lamphouse into the lamphouse mount at
Insert hexagonal Lamphouse
clamp screw
the top of the supporting pillar for dia-Illuminator.
screwdriver and
tighten clamp screw. Align the positioning pin on the dia pillar
illuminator with the groove on the cylinder of the
lamphouse.
2 Insert a hexagonal screwdriver into the hole on the
top of the supporting pillar for dia-Illuminator.
Secure the lamphouse by tightening the
lamphouse clamp screw.
3 Secure the lamphouse cable with the cable clamp
on the back of the supporting pillar for
dia-Illuminator.
The cable clamp is attached to the supporting
pillar for dia-Illuminator with the hooks on its sides.
To remove the cable clamp, push in the hooks
from the sides.
The cable clamp can hold up to four cables.
Cable clamp
90
Chapter 4 Assembly
WARNING
• When replacing the lamp, turn off the power supply and unplug the power cord.
• The lamp and its surroundings will be hot while the lamp is on and immediately after it is turned off.
When replacing the lamp, wait approximately 30 minutes after turning off the lamp, and make sure
that the lamp has cooled down sufficiently.
• Do not touch the lamp glass with your bare hands. Fingerprints and other dirt on the lamp may result
in uneven illumination and reduce the service life of the lamp. Wear gloves when handling the lamp.
• Securely close the lamphouse cover after replacing the lamp. Never turn on the lamp with the cover
removed.
• When you dispose of the replaced lamp, do not break it. Instead, dispose of the used lamp as
industrial waste or dispose of it according to the local regulations and rules.
Lamp
91
Chapter 4 Assembly
• Do not touch filters or other optical components with your bare hands.
• Latches at both ends of the filter slider are at their end point when sliding. When removing a filter
slider, you can slide it out by pushing up the latch on the opposite side of the tab with your finger to
release the filter slider. Applying undue force on the filter slider can break the latches.
EXTERNAL/CTRL
connector*
92
Chapter 4 Assembly
White
index mark
Mount rotation
clamp screw
93
Chapter 4 Assembly
You can mount the camera control unit DS-L2 on the MA200 main body by attaching the supporting arm.
1 Loosen clamp screws (2) using the supplied
hexagonal screwdriver and remove the cover from
the attaching part of the supporting pillar for
dia-Illuminator 100W/supporting arm on
top-surface of the microscope.
94
Chapter 4 Assembly
(2)
95
Chapter 4 Assembly
96
Chapter 4 Assembly
You can perform photomicroscopy by mounting the photomicrographic equipment such as CCD camera on
the microscope with the LV-TV TV or the C-mount adapter attached.
Apply the camera adapter to the vertical tube port on the MA2-TI3 trinocular eyepiece or the back port on the
upper-surface of the microscope. Be sure to apply the adapter or zoom lens compatible with the camera.
CAUTION
After mounting the camera, be sure to place cables into the cable keeper. If a cable comes into contact
with the lamphouse, the cable sheath may melt resulting in an electrical shock or fire.
WARNING
Make sure to use the specified power cord. Using a wrong power cord may result in malfunctions or
fire. This product is classified as subject to Class I protection against electrical shock. Make sure it is
connected to an appropriate ground terminal (protective earth terminal).
For specifications of the power cord, refer to “7. Specifications.”
Turn off the power switch on the product (set the switch to the “O” side).
Insert the plug of the power cord into the AC inlet on the back of the product. Then, securely plug the power
cord to the outlet.
97
5 Troubleshooting
Improper use of the microscope may adversely affect performance, even if the microscope is not damaged. If
any of the problems listed in the table below arise, take the countermeasures indicated.
If the problem is not listed below, or if the problem cannot be resolved by the suggested countermeasure,
unplug the power cord and contact your nearest Nikon representative.
98
Chapter 5 Troubleshooting
99
Chapter 5 Troubleshooting
100
Chapter 5 Troubleshooting
101
6 Care and Maintenance
Nikon recommends daily care and maintenance for maintaining the performance as long as possible.
Do not let dust, fingerprint, etc., get on the lenses. Dirt on the lenses, filters, and the like will adversely affect
the optical performance of the microscope. If lenses are contaminated, clean them according to the procedure
described in “6.1 Cleaning the Lenses and Filters.” When cleaning, be sure to turn off the power switch (set
the switch to “c” side) to avoid any operation error.
Do not let dust, fingerprint, etc. get on the lenses and filters. Dirt on the lenses, filters, etc. will adversely affect
the view of image. If any lens gets dirty, clean it as described below.
• Either brush away dust with a soft brush, or else gently wipe it off with a piece of gauze.
• Only if there are fingerprints or grease on a lens, dampen lightly a piece of soft, clean cotton cloth, lens
tissue, or gauze with absolute alcohol (ethyl or methyl) and gently wipe off the dirt.
• Absolute alcohol is highly flammable. Be careful when handling it, when around open flames, when turning
the power switch on/off, and so on.
• Follow the instructions provided by the manufacturer when using absolute alcohol.
6.2 Cleaning the Painted Parts, Plastic Parts, and Printed Parts
Do not use organic solvents such as alcohol, ether, or paint thinner on painted components, plastic
components, or printed components. Doing so could result in discoloration or in peeling of the printed
characters. For persistent dirt, dampen a piece of gauze with neutral detergent and wipe gently.
102
Chapter 6 Care and Maintenance
6.3 Storage
• Store this product in a dry place where mold is not likely to form.
• Store the objectives and eyepieces in a desiccator or similar container with a drying agent.
• Put the dust-proof cover over this product to protect it from dust.
• Before putting on the dust-proof cover, turn off the power switch of the product (set the switch to the “Ο”
side) and wait until the lamphouse gets cool sufficiently.
Periodical inspections of this product are recommended in order to maintain peak performance. Contact your
nearest Nikon representative for details.
103
7 Specifications
Type ECLIPSE MA200
Dimension 413 (W) × 337 (D) × 308 (H) mm, including the handle and mount parts
Weight 20 kg
Optical system Objective: CFI60 system (chromatic aberration free infinity optics system)
Eyepiece: Field number: 22, 25
Revo. nosepiece: 5 revo. (MA2-NUI5, LV-NU5A)
6 revo. (D-ND6)
7 revo. (D-NI7)
Illumination Episcopic illumination: Built-in lamp power supply
Illumination: Köhler illumination
Microscopy method: Bright-field, Dark-field, Simplified polarization, Differential interference
contrast and Fluorescence (UV excitation light non-correspondence)
Filter: ND4, ND16, GIF, and NCB11
Lamp ratings: 12 VDC, 50W halogen lamp
Specified lamp: LV-HL50W 12V 50W longlife halogen lamp
Specified lamp house: LV-LH50PC Precentered Lamphouse
Diascopic illumination: TI-PS100W power supply
Power supply specification
Input ratings: 100-240 VAC (±10%), 1.8 A, 50/60 Hz
Built-in fuse ratings: 250V T4A
Output ratings: 12 VDC 100W
Maximum output current: 8.4 A
Electric shock protection class: Class I
Remarks: UL listed product, GS approved
Illumination: Köhler illumination
Microscopy method: Bright-field, Simplified polarization
Filter: ND2, ND16, GIF, and NCB11
Lamp ratings: 12 VDC, 100W halogen lamp
Specified lamp: Halogen lamp (OSRAM HLX64623, PHILIPS77241)
Specified lamp house: D-LH/LC Precentered Lamphouse LC
Lamphouse attachment method: MA2-DP supporting pillar for dia-Illuminator 100W
Focusing Manual operation type single axis focus knob mechanism
mechanism (left side with coarse/fine focus, right side with fine focus knob)
Stroke: 4 mm (±2 mm: top datum surface for stage)
Coarse focus knob: 4.0 mm/revolution (with coarse torque adjustment mechanism)
Fine focus knob: 0.1 mm/revolution (1 μm/marking)
Input ratings Input voltage: 100-240 VAC ±10% 50/60 Hz
Rated current: 1.2 A maximum
104
Chapter 7 Specifications
• The product complies with Australian EMI standard. (AS/NZS CISPR11 Group1 Class A)
• CE marking
• The product meets EU Low Voltage Directive requirements.
• The product meets EU EMC Directive requirements. (EN61326)
105