Open Loop Control System
Open Loop Control System
EXPERIMENTAL PROCEDURE
The following procedures were carried out during the experiment
1. The dryer was switched off and the three trays were removed
2. The dryer was turned on, the fan speed and louvre were adjusted to achieve inlet air
velocity to 0.6m/s.
3. The value at T1 was entered as the ambient temperature (20℃ )
4. The PID was set to automatic and left for 15 minutes for the system to reach operating
temperature.
5. The PID was changed to manual, the trays were returned and each load cell was set to
zero by clicking the zero button.
6. The trays were removed after zeroing and the side door closed.
7. The wet solid was distributed on each tray after being drained in excess water.
8. The sampling interval was set to 5mins and new was selected
9. The trays were loaded, the side doors locked and the total weight recorded.
10. The load cells were zeroed to have negative readings demonstrating drying of soil
sample.
11. For close loop: The PID was changed to automatic and Go was clicked after weight
reading was stabilized.
12. For open loop: the PID was changed to manual and Go was clicked.