SONO-VIEW
Stand-alone display for the reliable control and configuration of processes using SONO moisture
probes. A total of up to 4 SONO probes can be monitored and the respective measurement values
presented at the LCD via a serial interface.
Thank you very much for your decision to purchase this IMKO product.
Should you have any questions in respect to this product, please contact our local
distribution partner or IMKO directly.
D-76275 Ettlingen
Germany
Phone: +49-7243-5921-0
Fax: +49-7243-90856
E-mail: [email protected]
Internet: http://www.imko.de
Page 2
2015-02-13 Manual SONO-VIEW
Table of Content
1 General Notices.................................................................................................................................. 5
3.3 Connection............................................................................................................................... 7
4 Operation ......................................................................................................................................... 9
Seite 3
Manual SONO-VIEW 2015-02-13
Page 4
2015-02-13 Manual SONO-VIEW
1 General Notices
Please carefully read and get acquainted with these operating instructions.
Should any queries arise, please contact our service department under the contact data
depicted above. Do not, under no circumstances, attempt to open or repair the device
yourself. In the event of any warranty claims, please refer these to the retailer you purchased
the device.
The device is subject to technical and optical change within the scope of product
improvement.
Seite 5
Manual SONO-VIEW 2015-02-13
2.2 Connections
Page 6
2015-02-13 Manual SONO-VIEW
3 Initial Commissioning
3.3 Connection
For operation, the SONO-VIEW requires a supply voltage of 7...24V (approx. 80..30mA). A
joint ground wire together with the probes is not required. For the connection with the
probes, it is sufficient to merely connect the two bus lines RTand COM.
Notice: The SONO-VIEW is suited for the display and configuration of up to four
probes. Should more than four probes be connected, an error message will
be generated and the device can not operate correctly.
Seite 7
Manual SONO-VIEW 2015-02-13
3.3.1 Exemplary Connections
Example 1: Connection of the SONO-VIEW with two SONO probes and one joint voltage
supply source.
Page 8
2015-02-13 Manual SONO-VIEW
4 Operation
Seite 9
Manual SONO-VIEW 2015-02-13
After completion of the above steps, all probe numbers
will be presented once more together with the
respectively allocated serial numbers in ascending
order.
Complete the installation process with the button .
Notice: The SONO-VIEW is suited for the display and configuration of up to four
probes. Should more than four probes be connected, an error message will
be generated and the device will not be able to operate properly.
After completion of the installation process, the SONO-VIEW will re-start, verify the
connected probes, and will immediately commence to call up the measurement values.
The moisture value in percent and the respective probe number is always presented. If two
or three probes are connected, the temperature measured by the probe is additionally also
presented. In the event that only one probe is connected, the calibrated radar run-time is also
presented.
If more than one probe is connected to the SONO-VIEW, there is an option to change the
display. For this purpose, actuate the buttons / . Subsequently, all connected individual
probes are presented consecutively. For the purpose of allocation, the probe number is
Page 10
2015-02-13 Manual SONO-VIEW
always also presented in the upper section. If this form of display is maintained for a longer
period, the same will be assumed as Standard. In this case, the SONO-VIEW will present
this set measurement display form after a re-start.
Seite 11
Manual SONO-VIEW 2015-02-13
4.4 Settings
Actuate the button while the measurement display is active and you will reach the setup
menu. Here, you are enabled to perform various settings and call up information regarding
the SONO-VIEW.
Setting Description
New Setup Enables a new detection of connected probes
About SONO- Serial number and further information regarding your SONO-VIEW
VIEW
Info Support information
Set the desired setting using the buttons / . By actuating the button , you can
subsequently enter the selected setting. Actuate the button again in order to exit the setup
menu.
4.4.2 Language
The selection of the language is performed with the buttons / . By actuating the button
, the selected language is set as the standard language. To exit the menu item Language,
actuate the button .
Page 12
2015-02-13 Manual SONO-VIEW
4.4.3 Display Contrast
A bar containing a grey colour gradation will appear.
4.4.5 Info
Contains links to the IMKO homepage and the E-mail address to request support.
4.4.6 USB-IMP-Bridge
As soon as this menu item is called up, the SONO-VIEW changes into a transparent data
mode. All data packets will be now redirected from the USB interface directly on to the IMP-
bus and vice versa. This enables a comfortable configuration of the probes via a connected
PC without the necessity of additional hardware. For this purpose, please download the free
software SonoConfigas well as the respective operating instructions available on the
IMKO-homepage.
Connect the SONO-VIEW to the PC using the provided USB cable. The SONO-VIEW will
connect with the PC as a virtual serial interface (COM-Port). The respectively required driver
is usually automatically installed by all currently used Windows versions. Should the driver
Seite 13
Manual SONO-VIEW 2015-02-13
not be automatically installed, please download the driver under:
http://www.ftdichip.com/Drivers/VCP.htm.
Setting Description
Probe Info Presents information regarding the connected probe
To reach the probe settings, select the individual probe display mode for the probe intended
for configuration in the measuring display using the / buttons (also see Item 4.2). By
actuating the button , the probe setting of the currently used probe is called up.
Notice: It is only possible to configure one probe at a time. Should several probes
require to be adjusted, the procedure must be respectively repeated for
these other probes.
Attention: Ensure that the correct probe is set before commencing with adjusting the
parameters.
Attention: The SONO-VIEW offers the option to access the measuring parameters of the
probe. Before adjusting any parameters, please inform yourself precisely in
regard to the function of the same in the respective probe manual. Any
performed changes may affect the measurement value, the accuracy, and the
measuring rate.
Page 14
2015-02-13 Manual SONO-VIEW
4.5.1 Probe Info
If this menu item is selected, various information of the probe is called up and displayed.
You can exit the menu item Probe Info with the
button .
Mode CS: (cyclic-successive) Without averaging function intended for very short measuring
processes in a range limited to seconds (e.g. 520 seconds) at which up to 100 measurements
Seite 15
Manual SONO-VIEW 2015-02-13
are performed internally per second at a cycle time of 250 milliseconds at the analogue
output. The operating mode CS also serves for the collection of raw values without
averaging and filter functions.
Mode CA: (cyclic average filter) Standard averaging for relatively fast but continuous
measurement processes with filtering and an accuracy of up to 0.1%.
Mode CF: (cyclic floating average with filter) Floating averaging for very slow and
continuous measurement processes with filtering and an accuracy of up to 0.1%.
This is suited for applications such as within a fluid bed dryer, on the conveyor belt, etc.
Mode CK: (cyclic with Kalman filter) suitable for complex application
Mode CC: (cyclic cumulated) with automatic summation of the moisture-quantity
measurements in one batch operation.
Notice: Please respectively also read the further information contained in your probe
operator manual.
Page 16
2015-02-13 Manual SONO-VIEW
Average Time
Filter Upper Limit Offset
CF - Cyclic Floating Filter Lower Limit Offset
Upper Limit Keep Time
Lower Limit Keep Time
Moisture Threshold
CC - Cyclic Cumulate
No Material Delay
Notice: Please respectively also read the further information contained in your
probe operator manual.
Attention: Before adjusting a parameter, please precisely inform yourself in regard to the
function of the same. Any performed change may affect the measurement
value, the accuracy, and the measuring rate.
The parameters are dynamically enabled with the set Average Parameter.
Seite 17
Manual SONO-VIEW 2015-02-13
4.5.6 Basic Balancing
At the exchange of a sensor head, due to deviating cable lengths, it may be necessary to
perform a basic balancing in air. Hereby, the moisture measurement value of the probe is re-
aligned to the correct Zero Value.
Attention: In order to exclude the occurrence of a faulty air calibration, the sensor must
be dry and free of any material during basic balancing.
5 Technical Data
Page 18
2015-02-13 Manual SONO-VIEW
Notes:
Seite 19
Manual SONO-VIEW 2015-02-13
Notes:
Page 20
Precise Moisture Measurement