Exp 4
Exp 4
Ravitej Uppu
Aim
To measure the value of the unknown capacitance using the method of de
Sauty and Schering bridges.
1 Theory
Bridges are the some of the most accurate measuring devices for measuring
impedance, capacitance, resistance,etc. For our purpose, i.e. for measuring
Capacitance, using a de Sauty’s or Schering bridge is best. They are based
on the principle of Wheatstone bridge that they have two arms. One of
which has the unknown parameter. By getting the bridge balanced using
the Null detector we can find this value.
The de Sauty’s bridge is a direct carry over of the Wheatstone bridge
with the DC source replaced by an AC source. The null detector we will
be using also has an amplifier where the gain can be adjusted. This is con-
nected to DMM which is used for getting the null point. The bridge is as
1
AC Bridges - Ravitej Uppu
2 Procedure
Connect the circuit as shown in the figures above. For the de Sauty bridge,
set the frequency to about 1KHz on the function generator and set the gain
on the null detector to max. Now balance the bridge by varying the ca-
pacitance(C) and resistance (R2 ) until we get the most stable and best null
point. Do this for various values of R1 and also for different frequencies.
For Schering bridge, do the same as above and balance the bridge with
the extra fact that use the Capacitance box given in parallel to get a stable
balance. From the read off values we can calculate the unknown capaci-
tance as Cx = RRs CBA .
2
AC Bridges - Ravitej Uppu
S.No R1 Ω R2 Ω C µF Cx µF
1 400 400 0.4493 0.4493
2 300 403 0.5877 0.4375
3 400 298 0.3359 0.4509
4 400 492 0.5473 0.4450
5 500 400 0.3558 0.4447
6 4000 4000 0.4528 0.4528
7 1000 1000 0.4458 0.4458
S.No R1 Ω R2 Ω C µF Cx µF
1 500 500 0.4432 0.4432
2 200 203 0.4423 0.4356
3 400 308 0.3658 0.4751
4 300 204 0.3029 0.4454
S.No R1 Ω R2 Ω C µF Cx µF
1 1000 1000 4.9873 4.9873
2 400 406 4.4996 4.4331
3 400 305 3.2868 4.3106
4 300 398 5.9543 4.4882
5 200 396 9.0243 4.5577
6 400 205 2.2065 4.4054
3
AC Bridges - Ravitej Uppu
Schering Bridge
Frequency =1.003KHz.
S.No Rx Ω RB Ω CA µF
1 1000 483 1 0.4830
2 1200 597 1 0.4875
3 800 367 1 0.4587
4 700 312 1 0.4457
5 600 269 1 0.4479
6 500 223 1 0.4465