Brivo OEC 715/785/865 Software Release WKS-1.x.x DICOM CONFORMANCE STATEMENT

Uploaded by

sec.ivb
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
249 views123 pages

GEHC DICOM Conformance - BrivoOEC715 785 865 VWKS1 XX - DOC1188156 - Rev3

Brivo OEC 715/785/865 Software Release WKS-1.x.x DICOM CONFORMANCE STATEMENT

Uploaded by

sec.ivb
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 123

Technical

Publications

DOC1188156
Revision 3

Brivo OEC 715/785/865 Software Release WKS-1.x.x


DICOM CONFORMANCE STATEMENT

Copyright 2013 by General Electric Co.

Do not duplicate

g
GE Healthcare
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

REVISION HISTORY
REV DATE REASON FOR CHANGE
1 Nov 12 2012 Initial Release
2 Mar 07 2013 Update document per Support Center Review Comments:
BrivoOec715.785.865DcsReview_HHComments_Mar072013.xls
x
3 Jun 18 2013 Update document per SPR HCSDM00203914
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

THIS PAGE LEFT INTENTIONALLY BLANK


BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
CONFORMANCE STATEMENT OVERVIEW
This DICOM Conformance Statement captures the DICOM capabilities of the Brivo
OEC 715/785/865 identified below.

Table 0.1 provides an overview of the network services supported by Brivo OEC
715/785/865 system.

Table 0.1 NETWORK SERVICES


SOP Classes User of Service Provider of
(SCU) Service (SCP)
Transfer
Computed Radiography Image Storage No Yes
Secondary Capture Image Storage No Yes
X-Ray Angiographic Image Storage Yes Yes
X-Ray Radiofluoroscopic Image Storage No Yes
X-Ray Radiation Dose SR Yes No
Query/Retrieve
Patient Root Query/Retrieve Information Model FIND Yes No
Patient Root Query/Retrieve Information Model MOVE Yes No
Study Root Query/Retrieve Information Model FIND Yes No
Study Root Query/Retrieve Information Model MOVE Yes No
Print Management
Basic Film Session SOP Class Yes No
Basic Film Box SOP Class Yes No
Basic Grayscale Image Box SOP Class Yes No
Basic Grayscale Print Management Meta SOP Class Yes No
Printer SOP Class Yes No
Workflow Management
Storage Commitment Push Model SOP Class Yes No
Modality Performed Procedure Step SOP Class Yes No
Modality Worklist Information Model FIND SOP Class Yes No

Table 0.2 provides an overview of the Media Storage Application Profiles supported by
Brivo OEC 715/785/865 system.

Table 0.2 - MEDIA SERVICES


Media Storage Application Profile Write Files Read Files
(FSC or FSU) (FSR)
Compact Disk - Recordable
General Purpose CD-R Yes Yes

4
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

TABLE OF CONTENTS

1. INTRODUCTION ............................................................................................................... 8
1.1 Overview ...................................................................................................................................................................... 8

1.2 Overall DICOM Conformance Statement Document Structure ............................................................................ 9

1.3 Intended Audience .................................................................................................................................................... 10

1.4 Scope and Field of Application ................................................................................................................................ 10

1.5 Important Remarks .................................................................................................................................................. 10

1.6 References .................................................................................................................................................................. 11

1.7 Definitions .................................................................................................................................................................. 11

1.8 Symbols and Abbreviations...................................................................................................................................... 13

2. NETWORK CONFORMANCE STATEMENT.................................................................. 16


2.1 Introduction ............................................................................................................................................................... 16

2.2 Implementation Model ............................................................................................................................................. 16

2.3 AE Specifications ....................................................................................................................................................... 19

2.4 Communication Profiles ........................................................................................................................................... 39

2.5 Extensions / Specializations/ Privatizations ............................................................................................................ 39

2.6 Configuration ............................................................................................................................................................ 40

2.7 Support of Extended Character Sets ....................................................................................................................... 41

2.8 Codes and Controlled Terminology ........................................................................................................................ 41

2.9 Security Profiles ........................................................................................................................................................ 42

3. MEDIA STORAGE CONFORMANCE STATEMENT ...................................................... 43


3.1 Introduction ............................................................................................................................................................... 43

3.2 AE Specifications ....................................................................................................................................................... 44

3.3 Augmented and Private Application Profiles ......................................................................................................... 47

3.4 Extensions / Specializations / Privatizations ........................................................................................................... 47

5
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
3.5 Configuration ............................................................................................................................................................ 47

3.6 Support of Extended Character Sets ....................................................................................................................... 48

4. X-RAY ANGIOGRAPHIC INFORMATION OBJECT IMPLEMENTATION ..................... 49


4.1 Introduction ............................................................................................................................................................... 49

4.2 Brivo OEC 715/785/865 Mapping of DICOM Entities .......................................................................................... 49

4.3 IOD Module Table .................................................................................................................................................... 49

4.4 Information Module Definitions .............................................................................................................................. 50

4.5 Standard Extended and Private Data Attributes ................................................................................................... 64

4.6 Standard Extended and Private Context Groups .................................................................................................. 64

5. X-RAY RADIATION DOSE STRUCTURED REPORT INFORMATION OBJECT


IMPLEMENTATION.................................................................................................................. 65
5.1 Introduction ............................................................................................................................................................... 65

5.2 Brivo oec 715/785/865 Mapping of DICOM Entities ............................................................................................. 65

5.3 IOD Module Table .................................................................................................................................................... 65

5.4 Information Module Definitions .............................................................................................................................. 66

5.5 Standard Extended and Private Data Attributes ................................................................................................... 75

5.6 Standard Extended and Private Context Groups .................................................................................................. 75

5.7 Standard, Standard Extended and Private Templates .......................................................................................... 75

6. MODALITY WORKLIST QUERY IMPLEMENTATION ................................................... 86


6.1 Introduction ............................................................................................................................................................... 86

6.2 Brivo oec 715/785/865 Mapping of DICOM Entities ............................................................................................. 86

6.3 Worklist Query Module Table ................................................................................................................................. 86

6.4 Worklist Query Module Definitions ........................................................................................................................ 87

7. MODALITY PERFORMED PROCEDURE STEP IMPLEMENTATION .......................... 96


7.1 Introduction ............................................................................................................................................................... 96

7.2 Relationship between Scheduled and Performed Procedure Steps ...................................................................... 96

7.3 Modality Performed Procedure Step Module Table ............................................................................................. 96

6
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
7.4 Modality Performed Procedure Step Module Definitions ..................................................................................... 96

7.5 Standard Extended and Private Data Attributes ................................................................................................. 102

7.6 Standard Extended and Private Context Groups ................................................................................................ 102

8. STORAGE COMMITMENT PUSH MODEL IMPLEMENTATION ................................. 104


8.1 Storage Commitment Push Model Information Object Definition .................................................................... 104

9. BASIC DIRECTORY INFORMATION OBJECT IMPLEMENTATION .......................... 106


9.1 IOD Module Table .................................................................................................................................................. 106

9.2 Information Module Definitions ............................................................................................................................ 106

9.3 Private Data Dictionary .......................................................................................................................................... 111

10. PRINT MANAGEMENT IMPLEMENTATION................................................................ 112


10.1 Introduction ............................................................................................................................................................. 112

10.2 Basic Film Session SOP Class ................................................................................................................................ 112

10.3 Basic Film Box SOP Class ...................................................................................................................................... 112

10.4 Basic Grayscale Image Box SOP Class ................................................................................................................. 114

10.5 Printer SOP Class ................................................................................................................................................... 115

11. QUERY IMPLEMENTATION ......................................................................................... 117


11.1 Brivo oec 715/785/865 Mapping of DICOM Entities ........................................................................................... 117

11.2 Information Model Keys ........................................................................................................................................ 117

7
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

1. INTRODUCTION

1.1 OVERVIEW
This DICOM Conformance Statement is divided into Sections as described below:

Section 1 (Introduction), which describes the overall structure, intent, and


references for this Conformance Statement

Section 2 (Network Conformance Statement), which specifies the GEHC


equipment compliance to the DICOM requirements for the implementation of
Networking features.

Section 3 (Media Storage Conformance Statement), specifies the Brivo OEC


715/785/865 equipment compliance to the DICOM requirements for the
implementation of Media Storage features.

Section 4 (X-Ray Angiographic Information Object Implementations),


specifies the Brivo OEC 715/785/865 Medical Systems equipment compliance
to DICOM requirements for the implementation of an X-Ray Information
Objects.

Section 5 (X-Ray Radiation Dose SR Information Object Implementations),


specifies the Brivo OEC 715/785/865 Medical Systems equipment compliance
to DICOM requirements for the implementation of X-Ray Radiation Dose SR
Information Object Implementation feature.

Section 6 (Modality Worklist Query Information Model), specifies the Brivo


OEC 715/785/865 Medical Systems equipment compliance to DICOM
requirements for the implementation of the Modality Worklist service.

Section 7 (Modality Performed Procedure Step Implementation), specifies


the Brivo OEC 715/785/865 Medical Systems equipment compliance to DICOM
requirements for the implementation of the Modality Performed Procedure Step
SOP Classes.

Section 8 (Storage Commitment Push Model Implementation), specifies the


Brivo OEC 715/785/865 Medical Systems equipment compliance to DICOM
requirements for the implementation of the Storage Commitment Push feature.

Section 9 (Basic Directory Information Object Implementation), specifies


the Brivo OEC 715/785/865 Medical Systems equipment compliance to DICOM
requirements for the implementation of a Basic Directory Information Object.

Section 10 (Print Management Implementation), specifies the Brivo OEC


715/785/865 Medical Systems equipment compliance to DICOM requirements
for the implementation of the Basic Print Meta SOP Classes (Grayscale).

Section 11 (Query/Retrieve Information Model), specifies the Brivo OEC


715/785/865 Medical Systems equipment compliance to DICOM requirements
for the implementation of the Patient, Study Root Query/Retrieve service.

8
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
1.2 OVERALL DICOM CONFORMANCE STATEMENT DOCUMENT STRUCTURE
The Documentation Structure of the GEHC DICOM Conformance Statements is shown
in the Illustration below.

GEHC DICOM Conformance Statements

@
http://www.ge.com/DICOM

APPLICATION ENTITY SPECIFICATION


(SERVICE CLASSES, INFORMATION OBJECTS, MESSAGE EXCHANGES, ETC.)

CT Advantage
C onformanc e
MR Advantage
Statement
Conformance
Direc tion:
Briv o OEC
Statement ......
Product 715/785/865
Direction:
C onformanc e
Conformance
Statement
Implementation: Statement
DOC1188156
Direction:
......

DICOM STANDARD

Standard DICOM

Specification: Part 1
DICOM

DICOM
Part 2
D ICOM
Part 3

Part 4
DICOM

Part 20

9
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
This document specifies the DICOM implementation. It is entitled:

Brivo OEC 715/785/865 Version WKS-1.x..x


Conformance Statement for DICOM
DOC1188156

This DICOM Conformance Statement documents the DICOM Conformance Statement


and Technical Specification required interoperating with the GEHC network interface.

The GEHC Conformance Statement, contained in this document, also specifies the Lower
Layer communications which it supports (e.g., TCP/IP). However, the Technical
Specifications are defined in the DICOM Part 8 standard.

For more information regarding DICOM, copies of the Standard may be obtained on the
Internet at http://medical.nema.org. Comments on the Standard may be addressed to:

DICOM Secretariat
NEMA
1300 N. 17th Street, Suite 1752
Rosslyn, VA 22209
USA
Phone: +1.703.841.3200

1.3 INTENDED AUDIENCE

The reader of this document is concerned with software design and/or system integration
issues. It is assumed that the reader of this document is familiar with the DICOM
Standard and with the terminology and concepts which are used in that Standard.

1.4 SCOPE AND FIELD OF APPLICATION


It is the intent of this document to provide an unambiguous specification for GEHC
implementations. This specification, called a Conformance Statement, includes a
DICOM Conformance Statement and is necessary to ensure proper processing and
interpretation of GEHC medical data exchanged using DICOM. The GEHC
Conformance Statements are available to the public.

The reader of this DICOM Conformance Statement should be aware that different GEHC
devices are capable of using different Information Object Definitions. For example, a
GEHC CT Scanner may send images using the CT Information Object, MR Information
Object, Secondary Capture Object, etc.

Included in this DICOM Conformance Statement are the Module Definitions which
define all data elements used by this GEHC implementation. If the user encounters
unspecified private data elements while parsing a GEHC Data Set, the user is well
advised to ignore those data elements (per the DICOM standard). Unspecified private
data element information is subject to change without notice. If, however, the device is
acting as a "full fidelity storage device", it should retain and re-transmit all of the private
data elements which are sent by GEHC devices.

1.5 IMPORTANT REMARKS


The use of these DICOM Conformance Statements, in conjunction with the DICOM
Standards, is intended to facilitate communication with GE imaging equipment.
However, by itself, it is not sufficient to ensure that inter-operation will be

10
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
successful. The user (or user's agent) needs to proceed with caution and address at
least four issues:

Integration - The integration of any device into an overall system of interconnected


devices goes beyond the scope of standards (DICOM v3.0), and of this introduction
and associated DICOM Conformance Statements when interoperability with non-GE
equipment is desired. The responsibility to analyze the applications requirements and
to design a solution that integrates GE imaging equipment with nonGE systems is
the user's responsibility and should not be underestimated. The user is strongly
advised to ensure that such an integration analysis is correctly performed.
Validation - Testing the complete range of possible interactions between any GE
device and nonGE devices, before the connection is declared operational, should
not be overlooked. Therefore, the user should ensure that any nonGE provider
accepts full responsibility for all validation required for their connection with GE
devices. This includes the accuracy of the image data once it has crossed the
interface between the GE imaging equipment and the nonGE device and the
stability of the image data for the intended applications.
Such a validation is required before any clinical use (diagnosis and/or treatment) is
performed. It applies when images acquired on GE imaging equipment are
processed/displayed on a non-GE device, as well as when images acquired on non-
GE equipment is processed/displayed on a GE console or workstation.

Future Evolution - GE understands that the DICOM Standard will evolve to meet
the user's growing requirements. GE is actively involved in the development of the
DICOM Standard. DICOM will incorporate new features and technologies and GE
may follow the evolution of the Standard. The GEHC protocol is based on DICOM
as specified in each DICOM Conformance Statement. Evolution of the Standard may
require changes to devices which have implemented DICOM. In addition, GE
reserves the right to discontinue or make changes to the support of
communications features (on its products) described by these DICOM
Conformance Statements. The user should ensure that any nonGE provider,
which connects with GE devices, also plans for the future evolution of the DICOM
Standard. Failure to do so will likely result in the loss of function and/or connectivity
as the DICOM Standard changes and GE Products are enhanced to support these
changes.
Interaction - It is the sole responsibility of the nonGE provider to ensure that
communication with the interfaced equipment does not cause degradation of GE
imaging equipment performance and/or function.

1.6 REFERENCES
NEMA PS3 Digital Imaging and Communications in Medicine (DICOM)
Standard, available free at http://medical.nema.org/

1.7 DEFINITIONS
Informal definitions are provided for the following terms used in this Conformance
Statement. The DICOM Standard is the authoritative source for formal definitions of
these terms.

Abstract Syntax the information agreed to be exchanged between applications,


generally equivalent to a Service/Object Pair (SOP) Class. Examples: Verification SOP
Class, Modality Worklist Information Model Find SOP Class, Computed Radiography
Image Storage SOP Class.

11
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Application Entity (AE) an end point of a DICOM information exchange, including
the DICOM network or media interface software; i.e., the software that sends or receives
DICOM information objects or messages. A single device may have multiple
Application Entities.

Application Entity Title the externally known name of an Application Entity, used to
identify a DICOM application to other DICOM applications on the network.

Application Context the specification of the type of communication used between


Application Entities. Example: DICOM network protocol.

Association a network communication channel set up between Application Entities.

Attribute a unit of information in an object definition; a data element identified by a


tag. The information may be a complex data structure (Sequence), itself composed of
lower level data elements. Examples: Patient ID (0010,0020), Accession Number
(0008,0050), Photometric Interpretation (0028,0004), Procedure Code Sequence
(0008,1032).

Information Object Definition (IOD) the specified set of Attributes that comprise a
type of data object; does not represent a specific instance of the data object, but rather a
class of similar data objects that have the same properties. The Attributes may be
specified as Mandatory (Type 1), Required but possibly unknown (Type 2), or Optional
(Type 3), and there may be conditions associated with the use of an Attribute (Types 1C
and 2C). Examples: MR Image IOD, CT Image IOD, Print Job IOD.

Joint Photographic Experts Group (JPEG) a set of standardized image compression


techniques, available for use by DICOM applications.

Media Application Profile the specification of DICOM information objects and


encoding exchanged on removable media (e.g., CDs)

Module a set of Attributes within an Information Object Definition that are logically
related to each other. Example: Patient Module includes Patient Name, Patient ID, Patient
Birth Date, and Patient Sex.

Negotiation first phase of Association establishment that allows Application Entities to


agree on the types of data to be exchanged and how that data will be encoded.

Presentation Context the set of DICOM network services used over an Association, as
negotiated between Application Entities; includes Abstract Syntaxes and Transfer
Syntaxes.

Protocol Data Unit (PDU) a packet (piece) of a DICOM message sent across the
network. Devices must specify the maximum size packet they can receive for DICOM
messages.

Security Profile a set of mechanisms, such as encryption, user authentication, or digital


signatures, used by an Application Entity to ensure confidentiality, integrity, and/or
availability of exchanged DICOM data

Service Class Provider (SCP) role of an Application Entity that provides a DICOM
network service; typically, a server that performs operations requested by another
Application Entity (Service Class User). Examples: Picture Archiving and
Communication System (image storage SCP, and image query/retrieve SCP), Radiology
Information System (modality worklist SCP).

12
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Service Class User (SCU) role of an Application Entity that uses a DICOM network
service; typically, a client. Examples: imaging modality (image storage SCU, and
modality worklist SCU), imaging workstation (image query/retrieve SCU)

Service/Object Pair (SOP) Class the specification of the network or media transfer
(service) of a particular type of data (object); the fundamental unit of DICOM
interoperability specification. Examples: Ultrasound Image Storage Service, Basic
Grayscale Print Management.

Service/Object Pair (SOP) Instance an information object; a specific occurrence of


information exchanged in a SOP Class. Examples: a specific x-ray image.

Tag a 32-bit identifier for a data element, represented as a pair of four digit
hexadecimal numbers, the group and the element. If the group number is odd, the
tag is for a private (manufacturer-specific) data element. Examples: (0010,0020) [Patient
ID], (07FE,0010) [Pixel Data], (0019,0210) [private data element]

Transfer Syntax the encoding used for exchange of DICOM information objects and
messages. Examples: JPEG compressed (images), little endian explicit value
representation.

Unique Identifier (UID) a globally unique dotted decimal string that identifies a
specific object or a class of objects; an ISO-8824 Object Identifier. Examples: Study
Instance UID, SOP Class UID, SOP Instance UID.

Value Representation (VR) the format type of an individual DICOM data element,
such as text, an integer, a persons name, or a code. DICOM information objects can be
transmitted with either explicit identification of the type of each data element (Explicit
VR), or without explicit identification (Implicit VR); with Implicit VR, the receiving
application must use a DICOM data dictionary to look up the format of each data
element.

1.8 SYMBOLS AND ABBREVIATIONS


AE Application Entity

AET Application Entity Title

Workstation Brivo OEC 715/785/865 Workstation AE

Workstation AE Brivo OEC 715/785/865 Workstation AE

CAD Computer Aided Detection

CD-R Compact Disk Recordable

CR Computed Radiography

CT Computed Tomography

DB Database

DHCP Dynamic Host Configuration Protocol

DICOM Digital Imaging and Communications in Medicine

13
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
DX Digital X-ray

FSC File-Set Creator

FSU File-Set Updater

FSR File-Set Reader

HL7 Health Level 7 Standard

HLF High Level Fluoro

IOD Information Object Definition

IPv4 Internet Protocol version 4

IPv6 Internet Protocol version 6

ISO International Organization for Standards

JPEG Joint Photographic Experts Group

LUT Look-up Table

MPPS Modality Performed Procedure Step

MR Magnetic Resonance Imaging

MWL Modality Worklist

O Optional (Key Attribute)

PACS Picture Archiving and Communication System

PDU Protocol Data Unit

R Required (Key Attribute)

RDSR Radiation Dose Structured Reports

RF Radiofluoroscopy

SC Secondary Capture

SCP Service Class Provider

SCU Service Class User

SOP Service-Object Pair

SPS Scheduled Procedure Step

SR Structured Reporting

TCP/IP Transmission Control Protocol/Internet Protocol

14
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
U Unique (Key Attribute)

VR Value Representation

XA X-ray Angiography

15
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

2. NETWORK CONFORMANCE STATEMENT

2.1 INTRODUCTION
This section of the DICOM Conformance Statement specifies the Brivo OEC
715/785/865 compliance to DICOM requirements for Networking features.

Brivo OEC 715/785/865 runs on industrial computer. It provides the following DICOM
functionalities:

Send Echo message to DICOM Verification SCP.

Export DICOM images and RDSR to a DICOM Storage SCP.

Query DICOM Modality Worklist from a DICOM Worklist SCP.

Querying for saved examinations and retrieving saved examinations from a DICOM
Query/Retrieve SCP

Sending exam start and end messages to a DICOM Modality Performed Procedure
Step SCP.

Sending storage commitment requests to a DICOM Storage Commitment SCP.

Print acquired images to a DICOM Printer.

2.2 IMPLEMENTATION MODEL

2.2.1 Application Data Flow Diagram


The network application model for the Brivo OEC 715/785/865 is shown in the following
Illustration :

16
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
ILLUSTRATION 21
BRIVO OEC 715/785/865 NETWORK APPLICATION MODEL AND DATA FLOW DIAGRAM

Remote
Storage Storage
SCP

Remote
Storage
Commit
SCP
Brivo OEC
715/785/865 Remote
Verify Echo
Workstation
AE SCP

Remote
Modality
Modality
Worklist
Worklist
Query
SCP

Remote
Report
MPPS
MPPS
SCP

Query Remote
Retrieve Q/R
SCP

Local
Store
SCP

Remote
DICO Print
M SCP
Print

DICOM
Standard
Interface

The product supports activity Storage which performs images/RDSR storage and
storage commitment functions; The results of that functionalities are communicated to
Remote Store Server for DICOM image/RDSR storage and storage commitment.

The product supports activity Print which performs images print functions; The results
of that functionalities are communicated to Remote Printer/Server for DICOM image
print functions.

The product supports activity Verify which interacts with Remote Verify SCP for
DICOM Echo purpose.

17
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
The product supports activity Modality Worklist Query which performs Modality
worklist query, from remote Modality Worklist SCP.

The product supports activity Query: Initiate an association with a Remote AE to query
for images on the remote host. Once the association has been established, it sends a
sequence of requests (Patient, then Study then Series Level requests) to the Remote AE.
After all responses are received, Workstation AE will issue a Series-Level C-FIND-RQ
request to get the series for a study in the list.

The product supports activity Retrieve: Send a C-MOVE-RQ request to a Remote AE


after successful association establishment. The Workstation AEs Storage SCP will
receive the images over a separate association.

The product supports activity Report MPPS: When the user begins the image
acquisition process and generates the first image, the Workstation AE sends N-CREATE
message to the configured MPPS SCP to indicate that the image acquisition process has
been started for the requested procedure. The operator can close the acquisition session
either by completing the acquisition process or discontinuing the ongoing scan. On
closing the acquisition session, the Workstation AE sends N-SET message to the
configured MPPS SCP to indicate the acquisition state of the requested procedure, with
appropriate MPPS status (COMPLETED/DISCONTINUED).

2.2.2 Functional Definition of AE's


Application Entity Brivo OEC 715/785/865 Workstation AE (i.e. Workstation AE)
initiates the following functionalities :

DICOM Host Verification: Initiate an association and send a C-ECHO-RQ message


to the remote DICOM AE; the remote DICOM server will send back a C-ECHO-
RSP message with a status of success if the operation is successful.

DICOM Image Storage: Initiate an association to a remote AE to send image(s). If


the remote AE accept the presentation context applicable to the image(s), the
WorkstationAE will send the image(s) by invoking C-STORE-RQ operation for each
image on the same association. The remote DICOM server will send back C-
STORE-RSP with status success if the operation is successful.

Storage Commitment: Initiates a DICOM Storage Commitment request once a Store


request has completed successfully. An N-ACTION request specifying all the
successfully transferred images will be issued to the remote AE. The remote AE will
reply accepting or refusing the request on the same association. If accepted the
remote AE will issue an N-EVENT response specifying which images have been
committed and which, if any, have not.

Modality Worklist Query: Initiate an association with a remote AE to query for


scheduled exams, a C-FIND-RQ request will be sent to the remote AE. C-FIND-RSP
with query results will be received.

Query: Initiate an association with a Remote AE to query for images on the remote
host. Follows the Patient->Study->Series sequence C-FIND-RQ request (the request
more than once) will be sent to the Remote AE once an association has been
established. Once all the responses have been received, the operator needs to select
an exam in the local database browser; on selection of the exam the DICOM Server
AE will issue a Series-Level C-FIND-RQ request to get the series for a study in the

18
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
list. Similarly the Image-Level C-FIND-RQ will be issued for the series selected
from the series list.

Retrieve: Send a C-MOVE-RQ request to a Remote AE for retrieve of images after


successful establishment. At this time the Local Store SCP listener will start, not
until after receiving complete images the Local Store SCP listener will stopped.

DICOM Print: Initiate an association with a remote AE to print grayscale images,


please refer to Section 2.3.1.2.2.2.1 for details

2.2.3 Sequencing of Real-World Activities


This sequence is only applicable for Remote AE where Storage Commitment Option is
allowed in Network Manager.

1. The user selects the images and sends them to a remote host.

2. If the remote DICOM AE is associated with a Storage Commitment Provider AE and if


the images are successfully sent to the DICOM AE, then an N-ACTION-RQ request is
sent automatically to the associated Storage Commitment Provider AE. The Storage
Commitment Provider AE can be configured independently from the remote DICOM AE
with network address, port and AE title.

2.3 AE SPECIFICATIONS

2.3.1 Workstation AE Specification


The Workstation Application Entity provides Standard Conformance to the following
DICOM SOP Classes as an SCU and/or as an SCP:

SOP Class Name SOP Class UID SCU SCP

Verification SOP Class 1.2.840.10008.1.1 Yes No

Computed Radiography Image Storage 1.2.840.10008.5.1.4.1.1.1 No Yes

Secondary Capture Image Storage 1.2.840.10008.5.1.4.1.1.7 No Yes

X-Ray Angiographic Image Storage 1.2.840.10008.5.1.4.1.1.12.1 Yes Yes

X-Ray Radiofluoroscopic Image Storage 1.2.840.10008.5.1.4.1.1.12.2 No Yes

X-Ray Radiation Dose SR 1.2.840.10008.5.1.4.1.1.88.67 Yes No

Patient Root Query/Retrieve Information Model 1.2.840.10008.5.1.4.1.2.1.1 Yes No


FIND

Patient Root Query/Retrieve Information Model - 1.2.840.10008.5.1.4.1.2.1.2 Yes No


MOVE

Study Root Query/Retrieve Information Model - FIND 1.2.840.10008.5.1.4.1.2.2.1 Yes No

Study Root Query/Retrieve Information Model - 1.2.840.10008.5.1.4.1.2.2.2 Yes No


MOVE

Modality Worklist Information Model - FIND 1.2.840.10008.5.1.4.31 Yes No

19
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Modality Performed Procedure Step 1.2.840.10008.3.1.2.3.3 Yes No

Storage Commitment Push Model 1.2.840.10008.1.20.1 Yes No

Basic Film Session SOP Class 1.2.840.10008.5.1.1.1 Yes No

Basic Film Box SOP Class 1.2.840.10008.5.1.1.2 Yes No

Basic Grayscale Image Box SOP Class 1.2.840.10008.5.1.1.4 Yes No

Basic Grayscale Print Management Meta SOP Class 1.2.840.10008.5.1.1.9 Yes No

Printer SOP Class 1.2.840.10008.5.1.1.16 Yes No

2.3.1.1 Association Establishment Policies

2.3.1.1.1 General
The DICOM Application Context Name (ACN), which is always proposed, is:

Application Context Name 1.2.840.10008.3.1.1.1

The maximum length PDU receive size for the Workstation AE is:

Maximum Length PDU 16384

2.3.1.1.2 Number of Associations

The Workstation AE will initiate a maximum of 1 simultaneous association to remote


nodes.

The Workstation AE will support a maximum of 1 simultaneous associations initiated by


remote nodes.

2.3.1.1.3 Asynchronous Nature


Asynchronous mode is not supported. All operations will be performed synchronously.

2.3.1.1.4 Implementation Identifying Information


The Implementation UID for this DICOM Implementation is:

Brivo OEC 715 Implementation UID 1.2.840.113619.6.320


Brivo OEC 715 Implementation Version Name BRIVO_OEC_100

20
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

Brivo OEC 785 Implementation UID 1.2.840.113619.6.321


Brivo OEC 785 Implementation Version Name BRIVO_OEC_100

Brivo OEC 865 Implementation UID 1.2.840.113619.6.333


Brivo OEC 865 Implementation Version Name BRIVO_OEC_100

2.3.1.2 Association Initiation Policy


When the Workstation Application Entity initiates an Association for any Real-World
Activity, it will propose the Presentation Contexts for all Real-World Activities; i.e.,
there is only a single, comprehensive Presentation Context Negotiation proposed for the
AE.

The Workstation AE proposes only a single Transfer Syntax in each Presentation


Context; i.e., for each Abstract Syntax in the following Presentation Context Tables, the
AE proposes one Presentation Context for each specified Transfer Syntax.

2.3.1.2.1 Real-World Activity Storage

2.3.1.2.1.1 Associated Real-World Activity


Upon a request by the operator the selected images will be sent to the configured DICOM
Store SCP.

2.3.1.2.1.2 Proposed Presentation Context Table

Presentation Context Table Proposed by Workstation AE for Activity Storage


Abstract Syntax Transfer Syntax Role
Name UID Name List UID List

X-Ray Angiographic 1.2.840.10008.5.1.4.1.1.12. Implicit VR Little Endian 1.2.840.10008.1.2 SCU


Image Storage 1 Explicit VR Little Endian 1.2.840.10008.1.2.1
Explicit VR Big Endian 1.2.840.10008.1.2.2

X-Ray Radiation Dose 1.2.840.10008.5.1.4.1.1.88. Implicit VR Little Endian 1.2.840.10008.1.2 SCU


SR 67 Explicit VR Little Endian 1.2.840.10008.1.2.1
Explicit VR Big Endian 1.2.840.10008.1.2.2

Storage Commitment 1.2.840.10008.1.20.1 Implicit VR Little Endian 1.2.840.10008.1.2 SCU


Push Model Explicit VR Little Endian 1.2.840.10008.1.2.1
Explicit VR Big Endian 1.2.840.10008.1.2.2

21
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
2.3.1.2.1.2.1 SOP Specific DICOM Conformance Statement for All Storage SOP Classes
The Workstation AE includes optional data elements in the SOP Instances as described in
Sections 4.

Although the Modality Type could be configured to RF/XA/CR/SC when user export
image to interchange media or remote DICOM server, only XA Image IOD is used.

Following are the status codes that are more specifically processed when receiving
messages from a Storage SCP equipment : If store command completed normally, with a
status of success or some warning then the image was accepted.

Service Status Further Meaning Application Behavior When Receiving Status


Status Code Code

Failure A700- Refused: Out of resources Terminate the operation. Logs store failed message
A7FF along with error comment returned from SCP.
Displays store failure message.
A900- Error: Data Set does not match
A9FF SOP Class

C000- Error: Cannot Understand


CFFF

0122 SOP Class Not Supported

Warning B000 Coercion of Data Elements Ignored by application and current operation
continues
B006 Elements Discarded

B007 Data Set does not match SOP


Class

Success 0000 DICOM storage operation is The DICOM operation progress screen
success disappearance indicates current operation is
success.

* * Any other status code. Upon receiving a C-STORE confirmation


containing a Refused status, this implementation
will terminate the association.

2.3.1.2.1.2.2 SOP Specific DICOM Conformance Statement for the X-Ray Radiation Dose Storage SOP Classes
See Section 2.3.1.2.1.2.1 SOP Specific DICOM Conformance Statement for All Storage
SOP Classes for details on general Storage Service SCU processing also applicable to
the X-Ray Radiation Dose Storage SOP Classes.
The Workstation AE supports creation and transmission of X-Ray Radiation Dose SOP
Instances referencing Instances of the following Storage SOP Classes:

22
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
SOP Class Name SOP Class UID

X-Ray Radiation Dose SR 1.2.840.10008.5.1.4.1.1.88.67

2.3.1.2.1.2.3 SOP Specific DICOM Conformance Statement for the Storage Commitment Push Model SOP Class
SCU
The Workstation AE uses DICOM network storage services to transfer SOP Instances
which are to be committed.

The Workstation AE may request Storage Commitment for Instances of any of the
Composite SOP Classes it supports as an SCU (see Section 2.3.1.2.1.2.1).

The Storage Commitment Information Object is described in Section 8.

Upon receiving a Storage Commitment N-EVENT-REPORT (Storage Commitment


Result on the same association), the Workstation AE will return a Success status.

Note: N-EVENT-REPORT is only expected on the same association.

Following are the status codes that are more specifically processed when receiving N-
Action responses from Storage Commitment SCP equipment:

Service Status Further Meaning Application Behavior When Receiving Status


Status Code Code

Failure 0119 Class-instance conflict Terminates the association and operation. Logs
failure message along with error comment returned
0210 Duplicate invocation from SCP and displays failure message.

0115 Invalid argument value

0117 Invalid SOP Instance

0212 Mistyped argument

0123 No such action

0114 No such argument

0118 No such SOP Class

0112 No such SOP Instance

0110 Processing failure

0213 Resource limitation

0211 Unrecognized operation

Success 0000 The progress screen is disappeared.

23
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
* * Any other status code. Deemed a failure.

2.3.1.2.2 Real-World Activity DICOM Print

2.3.1.2.2.1 Associated Real-World Activity

Upon a request by the operator the selected images will be sent to the configured DICOM
Printer (DICOM Print SCP).

2.3.1.2.2.2 Proposed Presentation Context Table

Presentation Context Table Proposed by AE Workstation for Activity DICOM Print


Abstract Syntax Transfer Syntax Role Extended

Name UID Name List UID List Negotiation

Basic Grayscale Print 1.2.840.10008.5.1.1.9 Implicit VR Little Endian 1.2.840.10008.1.2 SCU None
Management Meta SOP Explicit VR Little Endian 1.2.840.10008.1.2.1
Class Explicit VR Big Endian
1.2.840.10008.1.2.2

2.3.1.2.2.2.1 SOP Specific DICOM Conformance Statement for Basic Grayscale Print Management SOP Classes
The Workstation AE uses the following DIMSE services of the supported SOP Classes:

SOP Class SOP Class UID DIMSE Service Element SCU Usage
Basic Film Session 1.2.840.10008.5.1.1.1 N-CREATE Used (Mandatory)
N-SET Not Used
N-DELETE Used
N-ACTION Not Used
Basic Film Box 1.2.840.10008.5.1.1.2 N-CREATE Used (Mandatory)
N-ACTION Used (Mandatory)
N-DELETE Used
N-SET Not Used
Printer 1.2.840.10008.5.1.1.16 N-EVENT-REPORT Used (Mandatory)
N-GET Used
Basic Grayscale 1.2.840.10008.5.1.1.4 N-SET Used (Mandatory)
Image Box

When a manual print operation is initiated, the AE:

1. Initiates a DICOM association and negotiates Presentation Contexts

24
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
2. N-GETs printer status using the Printer SOP Class

3. N-CREATEs a Basic Film Session SOP Instance, and then followed by N-DELETE of
the Basic Film Session SOP Instance.

4. N-CREATEs a Basic Film Box SOP Instance for each film

5. N-SETs the Image Box SOP Instance for each image on the film

6. Prints by N-ACTION on the Basic Film Box SOP Instances, then followed by N-
DELETE of the Basic Film Box SOP Instance

7. Receives N-EVENT-REPORTs of the well know Printer SOP Instance indicating


printer status if the SCP send a request.

8. Releases the DICOM association after printing is successful or failure has been
signaled to the user

9. The DICOM operation progress screen disappearance indicates current operation is


success.

The Print management IODs is defined in Section.

2.3.1.2.2.2.1.1 Basic Film Session SOP Class

Following are the status codes that are more specifically processed when receiving
messages from a Print SCP equipment for the Basic Film Session SOP Class N-
CREATE:

Service Status Further Meaning Application Behavior When receiving Status


Status Codes Codes

Failure 0119 Class-instance conflict Terminate the operation. Logs store failed
message along with error comment returned from
0210 Duplicate invocation SCP. Displays print failure message.

0111 Duplicate SOP Instance

0106 Invalid attribute value

0117 Invalid SOP instance

0120 Missing attribute

0121 Missing attribute value

0212 Mistyped argument

0105 No such attributes

0118 No such SOP Class

0112 No such SOP Instance

25
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
0110 Processing failure

0213 Resource limitation

0211 Unrecognized operation

Warning B600 Memory allocation not supported Ignored by application and current operation
continues

Success 0000 Film session successfully created The progress screen is still displayed.

* * Any other status code. Deemed a failure.

Following are the status codes that are more specifically processed when receiving
messages from a Print SCP equipment for the Basic Film Session SOP Class N-
DELETE:

Service Status Further Meaning Application Behavior When receiving Status


Status Codes Codes

Failure 0119 Class-instance conflict Terminate the operation. Logs store failed
message along with error comment returned from
0210 Duplicate invocation SCP. Displays print failure message.

0117 Invalid SOP instance

0212 Mistyped argument

0118 No such SOP Class

0112 No such SOP Instance

0110 Processing failure

0213 Resource limitation

0211 Unrecognized operation

Success 0000 Film session successfully deleted The progress screen is still displayed.

* * Any other status code. Deemed a failure.

2.3.1.2.2.2.1.2 Basic Film Box SOP Class


Following are the status codes that are more specifically processed when receiving
messages from a Print SCP equipment for the Basic Film Box SOP Class N-CREATE:

Service Status Further Meaning Application Behavior When receiving Status

26
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Status Codes Codes

Failure C616 There is an existing Film Box that has Terminate the operation. Logs store failed
not been printed and N-ACTION at the message along with error comment returned from
Film Session level is not supported. A SCP. Displays print failure message.
new Film Box will not be created when a
previous Film Box has not been printed.

0119 Class-instance conflict

0210 Duplicate invocation

0111 Duplicate SOP Instance

0106 Invalid attribute value

0117 Invalid SOP instance

0120 Missing attribute

0121 Missing attribute value

0212 Mistyped argument

0105 No such attributes

0118 No such SOP Class

0112 No such SOP Instance

0110 Processing failure

0213 Resource limitation

0211 Unrecognized operation

Warning B605 Requested Min Density or Max Density Ignored by application and current operation
outside of printers operating range. The continues.
printer will use its respective minimum
or maximum density value instead.

Success 0000 Film box successfully created The progress screen is still displayed.

* * Any other status code. Deemed a failure.

Following are the status codes that are more specifically processed when receiving
messages from a Print SCP equipment for the Basic Film Box SOP Class N-ACTION:

27
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Service Status Further Meaning Application Behavior When receiving Status
Status Codes Codes
Failure C602 Unable to create Print Job SOP Instance; Terminate the operation. Logs store failed message
print queue is full along with error comment returned from SCP.
C603 Image size is larger than image box size Displays print failure message.
(by using the specified magnification
value)
C604 Image position collision : multiple images
assigned to single image position
C613 Combined Print Image size is larger than
the Image Box size
0119 Class-instance conflict
0210 Duplicate invocation
0115 Invalid argument value
0117 Invalid SOP Instance
0212 Mistyped argument
0123 No such action
0114 No such argument
0118 No such SOP Class
0112 No such SOP Instance
0110 Processing failure
0213 Resource limitation
0211 Unrecognized operation
Warning B603 Film Box SOP Instance hierarchy does Ignored by application and current operation
not contain Image Box SOP Instances continues.
(empty page)
B604 Image size is larger than image box size,
the image has been demagnified.
B609 Image size is larger than the Image Box
size. The Image has been cropped to fit.
B60A Image size or Combined Print Image size
is larger than the Image Box size. Image
or Combined Print Image has been
decimated to fit.
Success 0000 Film accepted for printing; if supported, The progress screen is still displayed.
the Print Job SOP Instance is created
* * Any other status code. Deemed a failure.

Following are the status codes that are more specifically processed when receiving
messages from a Print SCP equipment for the Basic Film Box SOP Class N-DELETE:

Service Status Further Meaning Application Behavior When receiving Status


Status Codes Codes

28
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Failure 0119 Class-instance conflict Terminate the operation. Logs store failed
message along with error comment returned from
0210 Duplicate invocation SCP. Displays print failure message.

0117 Invalid SOP instance

0212 Mistyped argument

0118 No such SOP Class

0112 No such SOP Instance

0110 Processing failure

0213 Resource limitation

0211 Unrecognized operation

Success 0000 Film box successfully deleted The progress screen is still displayed.

* * Any other status code. Deemed a failure.

2.3.1.2.2.2.1.3 Basic Grayscale Image Box SOP Class

Following are the status codes that are more specifically processed when receiving
messages from a Print SCP equipment for the Basic Grayscale Image Box SOP Class N-
SET:

Service Status Further Meaning Application Behavior When receiving Status


Status Codes Codes
Failure C603 Image size is larger than image box size Terminate the operation. Logs store failed
message along with error comment returned from
C605 Insufficient memory in printer to store SCP. Displays print failure message.
the image

C613 Combined Print Image size is larger than


the Image Box size

0119 Class-instance conflict

0210 Duplicate invocation

0106 Invalid attribute value

0212 Mistyped argument

0117 Invalid SOP instance

0121 Missing attribute value

29
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
0105 No such attributes

0118 No such SOP Class

0112 No such SOP Instance

0110 Processing failure

0213 Resource limitation

0211 Unrecognized operation

Warning B604 Image size larger than image box size, Ignored by application and current operation
the image has been demagnified. continues.

B605 Requested Min Density or Max Density


outside of printers operating range. The
printer will use its respective minimum
or maximum density value instead.

B609 Image size is larger than the Image Box


size. The Image has been cropped to fit.

B60A Image size or Combined Print Image


size is larger than the Image Box size.
The Image or Combined Print Image has
been decimated to fit.

Success 0000 Image successfully stored in Image Box The progress screen is still displayed.

* * Any other status code. Deemed a failure.

2.3.1.2.2.2.1.4 Printer SOP Class

The Workstation AE supports the Printer SOP Class to receive information on the status
of the printer.

For the product behavior description when receiving N-Event-Report request, refer to
section 10.5.1 Printer N-Event-Report Attributes.

Also, following are the status codes the Application may send back in the N-Event-
Report response command to the Printer SOP Class SCP Equipment that sent the N-
Event-Report request:

Service Status Further Status Code Explanation Related Fields


Status Code Meaning Sent Back to the
SCP
Success 0000 If an N-EVENT-REPORT DIMSE service is received when None
the association is active, Print SCU handles the relevant states
but the data received is ignored and send back Success.

30
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

For the attribute list requested by this product and for the behavior of this product on each
returned value, refer to Section 10.5.22 Printer N-Get Attributes

Following are the status codes that are more specifically processed when receiving
messages from a Print SCP equipment for the Printer SOP Class N-GET:

Service Status Further Meaning Application Behavior When receiving Status


Status Codes Codes

Failure 0107 Attribute list error Terminate the operation. Logs store failed
message along with error comment returned from
0119 Class-instance conflict SCP. Displays print failure message.

0210 Duplicate invocation

0117 Invalid SOP instance

0212 Mistyped argument

0118 No such SOP Class

0112 No such SOP Instance

0110 Processing failure

0213 Resource limitation

0211 Unrecognized operation

Success 0000 Success The progress screen is still displayed.

* * Any other status code. Deemed a failure.

2.3.1.2.3 Real-World Activity Verify

2.3.1.2.3.1 Associated Real-World Activity


The user may initiate a DICOM Verify Request in the configuration screen for each
respective remote SCP configuration (Print, Store, Worklist Query, MPPS, SC and Q/R ).

A valid response from the SCP will result in a The DICOM server has verified
successfully screen displayed on the screen. In the event that the SCP does not response
for some reason, the operation will timeout and the Workstation AE will close the
association and results in The DICOM Verify Fails screen displayed to user.

Note: The default timeout is 30 seconds for DICOM Echo and this value could be
configured from system setup screen.

31
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
2.3.1.2.3.2 Proposed Presentation Context Table

Presentation Context Table Proposed by Workstation AE for Activity Verify


Abstract Syntax Transfer Syntax Role Extended

Name UID Name List UID List Negotiation

Verification SOP 1.2.840.10008.1.1 Implicit VR Little Endian 1.2.840.10008.1.2 SCU None


Class Explicit VR Little Endian 1.2.840.10008.1.2.1
Explicit VR Big Endian
1.2.840.10008.1.2.2

2.3.1.2.3.2.1 SOP Specific DICOM Conformance Statement for Verify SOP Classes
The message The DICOM server has verified successfully is displayed when the
DICOM Verify is successfully performed, otherwise message The DICOM Verify
Fails is displayed.

2.3.1.2.4 Real-World Activity Modality Worklist Query

2.3.1.2.4.1 Associated Real-World Activity


Upon a request by the operator, the current scheduled exam list should be updated from
the remote Modality Worklist Server.

2.3.1.2.4.2 Proposed Presentation Context Table

Presentation Context Table Proposed by Workstation AE for Activity Modality Worklist


Abstract Syntax Transfer Syntax Role
Name UID Name List UID List

Modality Worklist 1.2.840.10008.5.1.4.31 Implicit VR Little Endian 1.2.840.10008.1.2 SCU


Information Model - Explicit VR Little Endian 1.2.840.10008.1.2.1
FIND Explicit VR Big Endian
1.2.840.10008.1.2.2

2.3.1.2.4.2.1 SOP Specific DICOM Conformance Statement for the Modality Worklist Information Model - FIND
SOP Class
The Workstation AE includes matching keys in the Modality Worklist queries as
described in Section 6.

The results of the Worklist query are displayed to the user in the Scheduled Exam Screen.
Currently, the maximal capability of received scheduled exams is 500 and not
configurable.

The Workstation AE checks the Specific Character Set tag (0008,0005) contained in a C-
FIND RSP, ISO_IR 100 or blank will be accepted, and all other value will be discarded.
Responses without this tag will also be accepted.

32
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
A C-FIND CANCEL will be sent when the maximal capability of received scheduled
exams(500 exams) is reached or user cancels operation by clicking the Cancel button on
the progressing screen.

Following are the status codes that are more specifically processed when receiving
messages from a Modality Worklist SCP equipment:

Service Status Further Meaning Application Behavior When Receiving Status


Status Code Code

Failure A700 Refused: Out of resources Terminates the association and operation. Logs
failure message along with error comment returned
A900 Error: Identifier does not match from SCP and displays failure message.
SOP Class

C000- Error: Unable to process


CFFF

0122 SOP Class Not Supported

Cancel FE00 Matching terminated due to The progress screen is disappeared, the received
cancel scheduled exams will be displayed.

Success 0000 Matching is complete - No final The progress screen is disappeared, all the received
identifier is supplied scheduled exams will be displayed.

Pending FF00 Matches are continuing - The progress screen is still displayed.
Current Match is supplied and
any Optional Keys were
supported in the same manner
as Required Keys.

FF01 Matches are continuing - The progress screen is still displayed.


Warning that one or more
Optional Keys were not
supported for existence for this
Identifier

* * Any other status code. Deemed a failure.

2.3.1.2.5 Real-World Activity Report MPPS

2.3.1.2.5.1 Associated Real-World Activity


The Modality Performed Procedure Step N-CREATE is sent with status of IN
PROGRESS when start exam is initiated by the operator. The Modality Performed
Procedure Step N-SET is sent with status of COMPLETED when the exam is
completed (all images acquired). If the exam is aborted, a status of
DISCONTINUED is sent.

33
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
2.3.1.2.5.2 Proposed Presentation Context Table

Presentation Context Table Proposed by Brivo OEC 715/786/865 AE for Activity Modality Worklist
Abstract Syntax Transfer Syntax Role Extended

Name UID Name List UID List Negotiation

Modality Performed 1.2.840.10008.3.1.2.3.3 Implicit VR Little Endian 1.2.840.10008.1.2 SCU None


Procedure Step Explicit VR Little Endian 1.2.840.10008.1.2.1
Explicit VR Big Endian
1.2.840.10008.1.2.2

2.3.1.2.5.2.1 SOP Specific DICOM Conformance Statement for Modality Performed Procedure Step SOP Class
MPPS will be created at the start of an image acquisition step; an MPPS will be set after
the completion/discontinue of a post-processing step when the user selects
Complete/Discontinue .

The Workstation AE includes attributes in the Modality Performed Procedure Step N-


CREATE as described in Section 7.4

Following are the status codes that are more specifically processed when receiving an N-
Create or N-Set response from an MPPS SCP equipment:

Service Status Further Meaning Application Behavior When Receiving Status


Status Code Code

Failure 0119 Class-instance conflict Terminates the association and operation. Logs
failure message along with error comment returned
0210 Duplicate invocation from SCP and displays failure message.

0106 Invalid attribute value

0117 Invalid SOP instance

0120 Missing attribute

0121 Missing attribute value

0212 Mistyped argument

0105 No such attributes

0118 No such SOP Class

0112 No such SOP Instance

0110 Processing failure

0213 Resource limitation

0211 Unrecognized operation

34
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Success 0000 The progress screen is disappeared.

0111 Duplicate SOP Instance

* * Any other status code. Deemed a failure.

2.3.1.2.6 Real-World Activity Query/Retrieve

2.3.1.2.6.1 Associated Real-World Activity


Upon a request from the user, a C-FIND-RQ command will be issued to the
Query/Retrieve SCP to query for exams that match the user-defined query criteria. The
association will be released upon the receipt of the C-FIND-RSP confirmation. To
retrieve an exam, the user can either use the information returned from a query or the
information stored in the local database when the exam was archived. The C-MOVE-
RQ command is used to retrieve the exam from the Query/Retrieve SCP. The C-
STORE-RQ from the SCP is expected on another association.

After a DICOM file is retrieved, the attributes are extracted and saved onto the local
system.

2.3.1.2.6.2 Proposed Presentation Context Table

Presentation Context Table Proposed by Brivo OEC 715/786/865 AE for Activity Modality Worklist
Abstract Syntax Transfer Syntax Role
Name UID Name List UID List

Study Root 1.2.840.10008.5.1.4.1.2.2. Implicit VR Little Endian 1.2.840.10008.1.2 SCU


Query/Retrieve 1 Explicit VR Little Endian 1.2.840.10008.1.2.1
Information Model - Explicit VR Big Endian
FIND 1.2.840.10008.1.2.2

Study Root 1.2.840.10008.5.1.4.1.2.2. Implicit VR Little Endian 1.2.840.10008.1.2 SCU


Query/Retrieve 2 Explicit VR Little Endian 1.2.840.10008.1.2.1
Information Model - Explicit VR Big Endian
MOVE 1.2.840.10008.1.2.2

Patient Root 1.2.840.10008.5.1.4.1.2.1. Implicit VR Little Endian 1.2.840.10008.1.2 SCU


Query/Retrieve 1 Explicit VR Little Endian 1.2.840.10008.1.2.1
Information Model - Explicit VR Big Endian
FIND 1.2.840.10008.1.2.2

Patient Root 1.2.840.10008.5.1.4.1.2.1. Implicit VR Little Endian 1.2.840.10008.1.2 SCU


Query/Retrieve 2 Explicit VR Little Endian 1.2.840.10008.1.2.1
Information Model - Explicit VR Big Endian
MOVE 1.2.840.10008.1.2.2

35
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
2.3.1.2.6.2.1 SOP Specific DICOM Conformance Statement for the Patient Root Query/Retrieve Information Model -
FIND, Study Root Query/Retrieve Information Model - FIND SOP Classes
The Workstation AE includes matching keys in the queries as described in Sections 11.

The Acquisition AE evaluates a C-FIND RSP containing the Specific Character Set tag
(0008,0005). Values of ISO_IR 100 or blank will be accepted, and all others will be
rejected.

Responses without this tag will also be accepted.

A C-FIND CANCEL will be sent when the display limit for results is reached.

The product does not support Extended Negotiation for C-FIND.

Following are the status codes that are more specifically processed when receiving
messages from a Query SCP equipment:

Service Status Further Meaning Application Behavior When Receiving Status


Status Code Code

Failure A700 Refused: Out of resources Terminates the association and operation. Logs
failure message along with error comment returned
A900 Error: Identifier does not match from SCP and displays failure message.
SOP Class

C000- Error: Unable to process


CFFF

0122 SOP Class Not Supported

Cancel FE00 Matching terminated due to The progress screen is disappeared, the received
cancel scheduled exams will be displayed

Success 0000 Matching is complete - No final The progress screen is disappeared, all the received
identifier is supplied scheduled exams will be displayed.

Pending FF00 Matches are continuing - The progress screen is still displayed.
Current Match is supplied and
any Optional Keys were
supported in the same manner
as Required Keys.

FF01 Matches are continuing - The progress screen is still displayed.


Warning that one or more
Optional Keys were not
supported for existence and/or
matching for this Identifier

* * Any other status code. Deemed a failure.

36
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
2.3.1.2.6.2.2 SOP Specific DICOM Conformance Statement for the Patient Root Query/Retrieve Information Model -
MOVE , Study Root Query/Retrieve Information Model - MOVE SOP Classes
The C-MOVE-RQ will use the AE Title of the Workstation Application Entity as the
Move Destination AE Title.

A C-FIND CANCEL message is sent if the user clicks the Cancel button during the
move.

Following are the status codes that are more specifically processed when receiving
messages from a Retrieve SCP equipment:

Service Status Further Meaning Application Behavior When Receiving Status


Status Code Code

Failure A701 Refused: Out of resources - Terminates the association and operation. Logs
Unable to calculate number of failure message along with error comment returned
matches from SCP and displays failure message.

A702 Refused: Out of resources -


Unable to perform sub-
operations

A801 Refused: Move Destination


Unknown

A900 Error: Identifier does not match


SOP Class

C000- Error: Unable to process


CFFF

0122 SOP Class Not Supported

Cancel FE00 Sub-operations terminated due The progress screen is disappeared, the received
to a Cancel indication scheduled exams will be displayed

Warning B000 Sub-operations Complete - One Ignored by application and current operation
or more Failures. continues.

Success 0000 Sub-operations Complete - No The progress screen is disappeared, all the received
Failure. scheduled exams will be displayed.

Pending FF00 Sub-operations are continuing - The progress screen is still displayed.

* * Any other status code. Deemed a failure.

2.3.1.3 Association Acceptance Policy


If the Remote AE needs to Query/Retrieve Images, to the local system then it has to be
configured in the Local system to do the same.

37
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
When the DICOM Server AE accepts an association for image storage, it will receive any
images transmitted on that association and store the images on disk.

It will also respond to queries from Remote AEs by sending matching entries. Any
Remote AE can request and receive a list of images on the local database. The Remote
AE must be configured in the local config file list of Remote AE for it to be able to
retrieve images from DICOM Server AE.

2.3.1.3.1 Real-World Activity Local Store SCP

2.3.1.3.1.1 Associated Real-World Activity


The Real-World Activity associated with the Receive Images operation is the storage of
the image on the disk drive.

2.3.1.3.1.2 Accepted Presentation Context Table

Presentation Context Table - Accepted by AE Workstation for Activity Receive Images


Abstract Syntax Transfer Syntax Role Extended

Name UID Name List UID List Negotiation

X-Ray 1.2.840.10008.5.1.4.1 Implicit VR Little Endian 1.2.840.10008.1.2 SCP None


Angiographic .1.12.1 Explicit VR Little Endian 1.2.840.10008.1.2.1
Image Storage
Explicit VR Big Endian 1.2.840.10008.1.2.2

X-Ray 1.2.840.10008.5.1.4.1 Implicit VR Little Endian 1.2.840.10008.1.2 SCP None


Radiofluoroscopic .1.12.2 Explicit VR Little Endian 1.2.840.10008.1.2.1
Image Storage
Explicit VR Big Endian 1.2.840.10008.1.2.2

Secondary Capture 1.2.840.10008.5.1.4.1 Implicit VR Little Endian 1.2.840.10008.1.2 SCP None


Image Storage .1.7 Explicit VR Little Endian 1.2.840.10008.1.2.1
Explicit VR Big Endian 1.2.840.10008.1.2.2

Computed 1.2.840.10008.5.1.4.1 Implicit VR Little Endian 1.2.840.10008.1.2 SCP None


Radiography Image .1.1 Explicit VR Little Endian 1.2.840.10008.1.2.1
Storage
Explicit VR Big Endian 1.2.840.10008.1.2.2

2.3.1.3.1.2.1 SOP Specific DICOM Conformance Statement for all Storage SOP Classes
The Workstation AE provides Level 2 (FULL) Conformance, and stores all standard and
private data elements of received SOP Instances.

Following are the status codes the Application may send back to the SCU Equipment
after performing the requested Storage:

Service Status Further Meaning Status Code Explanation Related Fields


Status Code Sent Back to
the SCU
Failure A700 Refused: Out of resources Out of Disk space (0000,0902)

38
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Success 0000 None

2.3.1.3.2 Real-World Activity Verification SCP

2.3.1.3.2.1 Associated Real-World Activity


An incoming Verification request will cause the Workstation AE to accept the association
and respond with a verification response after the Retrieve DICOM Server AE accepts an
sub-association for image storage.

2.3.1.3.2.2 Accepted Presentation Context Table

Presentation Context Table - Accepted by AE Workstation for Activity Receive Images


Abstract Syntax Transfer Syntax Role Extended

Name UID Name List UID List Negotiation

Verification SOP Class 1.2.840.10008.1.1 Implicit VR Little Endian 1.2.840.10008.1.2 SCP None
Explicit VR Little Endian 1.2.840.10008.1.2.1
Explicit VR Big Endian 1.2.840.10008.1.2.2

2.4 COMMUNICATION PROFILES

2.4.1 Supported Communication Stacks

The DICOM Upper Layer Protocol is supported using TCP/IP, as specified in DICOM
PS3.8.

The TCP/IP stack is inherited from the Linux Operating System.

2.4.2 Physical Media Support


The product is provided with a 10/100 Mb/s or 1Gb/s auto-sensing Ethernet interface.
Additional or alternate network interfaces may be available.

2.4.3 Additional Protocol

The Workstation supports DHCP, user could configure DHCP Enable/Disable from
System Setup screen. When DHCP is Enabled, then, the network information will be
gotten automatically.

Default is Enabled.

2.4.4 IPv4 and IPv6 Support


The Workstation only supports IPv4.

2.5 EXTENSIONS / SPECIALIZATIONS/ PRIVATIZATIONS


Not Applicable.
39
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
2.6 CONFIGURATION

2.6.1 AE Title/Presentation Address Mapping

2.6.2 Configurable Parameters


The following fields are configurable:

The default value for these fields is blank unless otherwise specified.

Network

Use DHCP or static IP (Default DHCP).


Local IP Address
Local Subnet mask
Default Gateway

Configuration:

Local AE Title (Default Brivo OEC)


Station Name
Hospital Name

Every Remote DICOM AE SCP: (Store, Storage Commitment, MPPS, Print, Modality
Worklist, Query/Retrieve)

Server Alias
Remote AE Title
Remote IP Address
Remote Port Number
Gateway IP Address.
Modality Type (Default - RF)

Additional fields for Remote DICOM Print

Min Density for Printer


Max Density for Printer
Configuration Information
Border Density
Empty Density
Number of Copies
Print Priority

40
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Film Destination
Medium Type
Film Size
Format
Film Orientation

Additional fields for Remote DICOM Query Worklist Filter

Patient Name
Patient ID
Requested Procedure ID
Accession Number
Exam Date: Today, Next 2 Days, Next 3 Days, Next 7 Days, Date Range
Performing Physician Name
Modality Type (Default - RF)
Additional fields for Remote DICOM Query/Retrieve Filter

Date: Today, Next 2 Days, Next 3 Days


Patient Name
Patient ID
Accession Number
Modality Type (Default - RF)

Note: All configurations must be performed by a GE Field Engineer.

2.7 SUPPORT OF EXTENDED CHARACTER SETS


As a Query SCU, it will similarly accept response items with ISO_IR 100 values of
Specific Character Set or empty value of Specific Character Set (absence of (0008,0005)
which mean usage of ASCII Chars only).

The product user interface will allow the user to enter characters from the console
keyboard that is within ISO 8859-1. Responses with non-compatible values of
(0008,0005) will be discarded.

2.8 CODES AND CONTROLLED TERMINOLOGY


The product uses no coded terminology.

41
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
2.9 SECURITY PROFILES
The product does not conform to any defined DICOM Security Profiles.

It is assumed that the product is used within a secured environment. It is assumed that a
secured environment includes at a minimum:

1. Firewall or router protections to ensure that only approved external hosts have
network access to the product.
2. Firewall or router protections to ensure that the product only has network access to
approved external hosts and services.
3. Any communications with external hosts and services outside the locally secured
environment use appropriate secure network channels (such as a Virtual Private
Network (VPN))

42
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

3. MEDIA STORAGE CONFORMANCE STATEMENT

3.1 INTRODUCTION
This section of the DICOM conformance statement specifies the Workstation AE
compliance to DICOM requirements for Media Interchange. It details the DICOM
Media Storage Application Profiles and roles that are supported by this product.

The Workstation AE is able to export images and RDSR to DICOM media.

3.1.1 Application Data Flow Diagram

The media interchange application model for the Workstation AE is shown in the
following Illustration:

ILLUSTRATION 31
WORKSTATION AE MEDIA INTERCHANGE APPLICATION MODEL AND DATA FLOW DIAGRAM

Storage FSC
Remote
Brivo OEC 715/785/865 Media Storage
Workstation AE Storage SCP
FSU

Export

FSR
Directory Media Viewer AE Media
Listing Storage

The Workstation AE supports the General Purpose CD-R Interchange profiles.

3.1.2 Functional Definition of AEs


The Workstation AE can perform these functions:

Create a new DICOM File-Set on Media

Update a DICOM File-Set by adding new SOP instances to the media which
supports multi-session.

The Media Viewer AE can perform this function:

Reads a DICOM File-Set from media and reviews recorded DICOM instances
(including images and RDSR).

43
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
3.1.3 Sequencing of Real-World Activities
Storage/Export Images are acquired and archived on Brivo OEC 715/785/865; From the
Image Directory screen, images/RDSR can be selected and copied to media;
Images/RDSR can be viewed by any DICOM application that supports Media Exchange
FSR and X-Ray Radiation Dose Structured Reports.

Directory Listing During the Image Copy, user could choice copying image/RDSR with
a Media Viewer, then the images/RDSR will be copied with a Media Viewer instance;
Images/RDSR can be viewed by this media viewer only on Windows OS (Independent of
the Brivo 715/786/865 Workstation AE).

3.1.4 File Meta Information Options (See PS3.10)


The File Meta-Information for this implementation is:

File Meta-Information Version 1


Brivo OEC 715 Implementation UID 1.2.840.113619.6.320
Brivo OEC 715 Implementation Version Name BRIVO_OEC_100

File Meta-Information Version 1


Brivo OEC 785 Implementation UID 1.2.840.113619.6.321
Brivo OEC 785 Implementation Version Name BRIVO_OEC_100

File Meta-Information Version 1


Brivo OEC 865 Implementation UID 1.2.840.113619.6.333
Brivo OEC 865 Implementation Version Name BRIVO_OEC_100

3.2 AE SPECIFICATIONS

3.2.1 Workstation AE Specification


The Workstation Application Entity provides standard conformance to DICOM
Interchange Option of the Media Storage Service Class. The supported Application
Profiles and roles are listed below.

Supported Application Profile Real World Activity Role Option

STD-GEN-CD Storage/ Export FSC,FSU Interchange

44
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
3.2.1.1 File Meta Information for the Workstation Application Entity
Following are the values set in the File Meta Information for this AE Title:

Source Application Entity Title Not used


Brivo OEC 715 Implementation UID 1.2.840.113619.6.320
Brivo OEC 715 Implementation Version Name BRIVO_OEC_100

Source Application Entity Title Not used


Brivo OEC 785 Implementation UID 1.2.840.113619.6.321
Brivo OEC 785 Implementation Version Name BRIVO_OEC_100

Source Application Entity Title Not used


Brivo OEC 865 Implementation UID 1.2.840.113619.6.333
Brivo OEC 865 Implementation Version Name BRIVO_OEC_100

3.2.1.2 Real-World Activities for the Workstation Application Entity

3.2.1.2.1 Real-World Activity Storage


The Workstation AE acts as a FSC/FSU when requested to copy SOP instance from local
database to the interchange media.

When user selects Storage function, the selected images should be saved as DICOM
SOP instances and copied to media. The DICOM File set on the media should be created
or updated accordingly.

3.2.1.2.1.1 Media Storage Application Profile for the RWA Update File Set
For the list of Application Profiles that invoke this AE for the Real-World Activity
Storage, see the Table in Section 3.2.1 Brivo OEC 715/785/865 where the table
describing the Application Profiles and Real-World Activity is defined.

3.2.1.2.1.1.1 Options for STD-GEN-CD Application Profile


Following are the optional SOP Classes supported by this AE. All SOP Instances use
the Explicit VR Little Endian Uncompressed Transfer Syntax, UID 1.2.840.10008.1.2.1.

SOP Class SOP Class UID

X-Ray Angiographic Image Storage 1.2.840.10008.5.1.4.1.1.12.1

45
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

Common DICOMDIR Directory Records created by this AE will include key attributes
as described in Section 9. Following are the Additional DICOMDIR Keys supported for
this profile:

ADDITIONAL DICOMDIR KEYS


Key Attribute Tag Directory Type Notes
Record Type
Image Type (0008,0008) IMAGE 1C FSC/FSU: Present when the DICOMDIR file is
created and this value should be extracted from
referenced image file.

3.2.2 Media Viewer AE Specification


The Media Viewer Application Entity provides standard conformance to DICOM
Interchange of the Media Storage Service Class. The supported Application Profiles and
roles are listed below.

Supported Application Profile Real World Activity Role

STD-GEN-CD Directory Listing FSR

3.2.2.1 File Meta Information for the Media Viewer Application Entity
Following are the values set in the File Meta Information for this AE Title:

Source Application Entity Title Not used


Media Viewer Implementation UID Not used
Implementation Version Name Not used
Since Media Viewer just used to read file-set and display image, so, above information
could not be set by Media Viewer.

3.2.2.2 Real-World Activities for the Media Viewer Application Entity

3.2.2.2.1 Real-World Activity Directory Listing

The Media Viewer AE acts as a FSR when listing the directory information from
interchange media.

When user load the media on Window OS, the Media Viewer will read the DICOM File-
Set and display/list the DICOM directory entries for those instances in the File-Set on the
media.

3.2.2.2.1.1 Media Storage Application Profile for the RWA Directory Listing
For the list of Application Profiles that invoke this AE for the Real-World Activity
Directory Listing, see the Table in Section 3.2.3 Media Viewer AE Specification where
the table describing the Application Profiles and Real-World Activity is defined.

46
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
3.2.2.2.1.1.1 STD-GEN-CD Application Profile
Following are the SOP Classes supported by this AE and only the Explicit VR Little
Endian Uncompressed Transfer Syntax (UID 1.2.840.10008.1.2.1) is supported.

SOP Class SOP Class UID

X-Ray Angiographic Image Storage 1.2.840.10008.5.1.4.1.1.12.1

Common DICOMDIR Directory Records read by this AE will include key attributes as
described in Section 9. Following are the Additional DICOMDIR Keys supported for this
profile:

Key Attribute Tag Directory Type Notes


Record Type
Image Type (0008,0008) IMAGE 1C FSR: Used.

3.2.3 Media Viewer AE Specification


The Media Viewer Application Entity provides standard conformance to DICOM
Interchange of the Media Storage Service Class. The supported Application Profiles and
roles are listed below.

Supported Application Profile Real World Activity Role

STD-GEN-CD Directory Listing FSR

3.3 AUGMENTED AND PRIVATE APPLICATION PROFILES


No augmented/private profile is implemented.

3.4 EXTENSIONS / SPECIALIZATIONS / PRIVATIZATIONS


No Extensions / Specializations / Privatizations are implemented.

3.5 CONFIGURATION
CD Options (Media Exchange)

Modality Type (Image Default RF, Dose Summary Default SC)


Remove Patient Information (Default - No).
Include Viewer (Default - Include)
Include Overlay(Default - Include)

47
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
3.6 SUPPORT OF EXTENDED CHARACTER SETS
The Workstation AE will support fully only the ISO_IR 100 as extended character sets
for media storage.

48
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

4. X-RAY ANGIOGRAPHIC INFORMATION OBJECT


IMPLEMENTATION

4.1 INTRODUCTION

This section specifies the use of the DICOM X-Ray Angiographic Image IOD to
represent the information included in X-Ray Angiographic Images by this
implementation. Corresponding attributes are conveyed using the module construct.

4.2 BRIVO OEC 715/785/865 MAPPING OF DICOM ENTITIES

The Brivo OEC 715/785/865 maps DICOM Information Entities to local Information
Entities in the products database and user interface.

TABLE 4-1
MAPPING OF DICOM ENTITIES TO BRIVO OEC 715/785/865 ENTITIES
DICOM IE Brivo OEC 715/785/865 Entity
Patient Patient
Study Exam
Series Series
Equipment Equipment
Image Image

4.3 IOD MODULE TABLE

The X-Ray Angiographic Image Information Object Definition comprises the modules of
the following table, plus Standard Extended and Private attributes. Standard Extended
and Private attributes are described in Section 4.5.

TABLE 4-2
X-RAY ANGIOGRAPHIC IMAGE IOD MODULES
Entity Name Module Name Usage Reference
Patient Patient Used 4.4.1.1
Clinical Trial Subject Not Used N/A
Study General Study Used 4.4.2.1
Patient Study Not Used N/A
Clinical Trial Study Not Used N/A
Series General Series Used 4.4.3.1
Clinical Trial Series Not Used N/A
Frame of Synchronization Not Used N/A
Reference

49
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Equipment General Equipment Used 4.4.4.1
Image General Image Used 4.4.5.1
Image Pixel Used 4.4.5.2
Contrast/Bolus Used 4.4.5.3
Cine Not Used N/A
Multi-Frame Not Used N/A
Frame Pointers Not Used N/A
Mask Not Used N/A
Display Shutter Not Used N/A
Device Not Used N/A
Intervention Not Used N/A
X-Ray Image Used 4.4.5.4
X-Ray Acquisition Used 4.4.5.5
X-Ray Collimator Not Used N/A
X-Ray Table Not Used N/A
XA Positioner Used 4.4.5.66
DX Detector Not Used N/A
Overlay Plane Used 4.4.5.7
Multi-Frame Overlay Not Used N/A
Modality LUT Not Used N/A
VOI LUT Used 4.4.5.88
SOP Common Used 4.4.5.9
Frame Extraction Not Used N/A

4.4 INFORMATION MODULE DEFINITIONS


Please refer to DICOM Part 3 (Information Object Definitions) for a description of each
of the entities, modules, and attributes contained within the X-Ray Angiographic
Information Object.

The following modules are included to convey Enumerated Values, Defined Terms, and
Optional Attributes supported and/or expected. Type 1 & Type 2 Attributes are also
included for completeness and to define what values they may take and where these
values are obtained from when generating the instance as well as what are the expected
values when loading such instance. It should be noted that they are the same ones as
defined in the DICOM Standard Part 3 (Information Object Definitions). Also note that
Attributes not present in tables are not supported.

50
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
4.4.1 Patient Entity Modules

4.4.1.1 Patient Module


TABLE 4-3
PATIENT MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
Patient's Name (0010,0010) 2 May be entered from user interface or got from
Worklist.
Supports 5 different components delimited by
^. Supports a maximum length of 64 bytes
including the delimiter.
Patient ID (0010,0020) 2 May be entered from user interface or got from
Worklist.
Truncated to 64 bytes, end truncated.
Issuer of Patient ID (0010,0021) 3 May be entered from user interface or got from
Worklist.
Truncated to 64 bytes, end truncated.
Patient's Birth Date (0010,0030) 2 May be entered from user interface or got from
Worklist
Truncated to 8 bytes, end truncated
Patient's Sex (0010,0040) 2 May be entered from user interface or got from
Worklist.
Truncated to 2 bytes, end truncated
Referenced Patient Sequence (0008,1120) 3 Not Used
>Include SOP Instance Reference Macro
Patient's Birth Time (0010,0032) 3 Not Used
Other Patient IDs (0010,1000) 3 Not Used
Other Patient IDs Sequence (0010,1002) 3 Not Used
>Patient ID (0010,0020) 1 Not Used
>Issuer of Patient ID (0010,0021) 1 Not Used
>Type of Patient ID (0010,0022) 1 Not Used
Other Patient Names (0010,1001) 3 Not Used
Ethnic Group (0010,2160) 3 Not Used
Patient Comments (0010,4000) 3 May be entered from user interface or got from
Worklist
From user interface truncated to 512 bytes.
From worklist truncated to 10240 bytes.
Patient Species Description (0010,2201) 1C Not Used
Patient Species Code Sequence (0010,2202) 1C Not Used
>Include Code Sequence Macro
Patient Breed Description (0010,2292) 2C Not Used
Patient Breed Code Sequence (0010,2293) 2C Not Used
>Include Code Sequence Macro

51
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Breed Registration Sequence (0010,2294) 2C Not Used
>Breed Registration Number (0010,2295) 1 Not Used
>Breed Registry Code Sequence (0010,2296) 1 Not Used
>>Include Code Sequence Macro
Responsible Person (0010,2297) 2C Not Used
Responsible Person Role (0010,2298) 1C Not Used
Responsible Organization (0010,2299) 2C Not Used
Patient Identity Removed (0012,0062) 3 Used.
If Patient Information is removed then set to
"YES", otherwise set to "NO"
De-identification Method (0012,0063) 1C Used.
If Patient Information is removed. Set to
"Limited Data Set" in this case
De-identification Method Code Sequence (0012,0064) 1C Not Used
>Include Code Sequence Macro

4.4.2 Study Entity Modules

4.4.2.1 General Study Module


TABLE 4-4
GENERAL STUDY MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
Study Instance UID (0020,000D) 1 The product will adopt the Study Instance UID
which is returned in the MWL response. If
there is no Study Instance UID returned from
MWL server, then new Study Instance UID
will be created.
Study Date (0008,0020) 2 This value is set to the exam created date.
Study Time (0008,0030) 2 This value is set to the exam created time.
Referring Physician's Name (0008,0090) 2 May be entered from the user interface or got
from Worklist.
Referring Physician Identification (0008,0096) 3 Not Used
Sequence
>Include Person Identification Macro
Study ID (0020,0010) 2 Generated by equipment.
Accession Number (0008,0050) 2 May be entered from the user interface or got
from Worklist.
Study Description (0008,1030) 3 May be entered from the user interface or got
from Worklist.
The default value will be set as Requested
Procedure Description when the tag got from
MWL is not empty. Else this value will be set
as Procedure Description on user interface.

52
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Physician(s) of Record (0008,1048) 3 Not Used
Physician(s) of Record Identification (0008,1049) 3 Not Used
Sequence
>Include Person Identification Macro
Name of Physician(s) Reading Study (0008,1060) 3 Not Used
Physician(s) Reading Study Identification (0008,1062) 3 Not Used
Sequence
>Include Person Identification Macro
Referenced Study Sequence (0008,1110) 3 Got from MWL.
>Include SOP Instance Reference Macro
Procedure Code Sequence (0008,1032) 3 Not Used
>Include Code Sequence Macro

4.4.3 Series Entity Modules

4.4.3.1 General Series Module


TABLE 4-5
GENERAL SERIES MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
Modality (0008,0060) 1 This value could be set to RF/XA/CR/OT. The
proper SOP class will be created based on XA
Image IOD.
Series Instance UID (0020,000E) 1 Created by equipment, using prefix:
715: 1.2.840.113619.2.320.
785: 1.2.840.113619.2.321.
865: 1.2.840.113619.2.333.
+MAC+seriesIndicator(20)+TimeStamp_Exam
MAC: Number string of device MAC address.
Series Number (0020,0011) 2 A number that identifies this Series.
Laterality (0020,0060) 2C The value is empty.
Series Date (0008,0021) 3 This value is set to the first shot date.
Series Time (0008,0031) 3 This value is set to the first shot time.
Performing Physicians Name (0008,1050) 3 Get from the Worklist scheduled performing
physician, or the user can enter from the user
interface.
Performing Physician Identification (0008,1052) 3 Not Used
Sequence
>Include Person Identification Macro
Protocol Name (0018,1030) 3
When using MWL, if Scheduled Protocol
Sequence Code Meaning is not empty then this
value is Scheduled Protocol Sequence Code
Meaning, else this value is Scheduled Procedure

53
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Step Description.
It will be set as Unknown Protocol for local
created exam.
Series Description (0008,103E) 3 The value equal protocol Name.
Operators' Name (0008,1070) 3 Not Used
Operator Identification Sequence (0008,1072) 3 Not Used
>Include Person Identification Macro
Referenced Performed Procedure Step (0008,1111) 3 Used.
Sequence
>Include SOP Instance Reference Macro
Related Series Sequence (0008,1250) 3 Not Used
>Study Instance UID (0020,000D) 1 Not Used
>Series Instance UID (0020,000E) 1 Not Used
>Purpose of Reference Code Sequence (0040,A170) 2 Not Used
>>Include Code Sequence Macro
Body Part Examined (0018,0015) 3 Used
Patient Position (0018,5100) 2C Not Used
Smallest Pixel Value in Series (0028,0108) 3 Not Used
Largest Pixel Value in Series (0028,0109) 3 Not Used
Request Attributes Sequence (0040,0275) 3 Used
>Requested Procedure ID (0040,1001) 1C Used
>Accession Number (0008,0050) 3 Not Used
>Study Instance UID (0020,000D) 3 Not Used
>Referenced Study Sequence (0008,1110) 3 Not Used
>> Include SOP Instance Reference Macro
>Requested Procedure Description (0032,1060) 3 Used
>Requested Procedure Code Sequence (0032,1064) 3 Used
>>Include Code Sequence Macro
Reason for the Requested Procedure (0040,1002) 3 Not Used
Reason for Requested Procedure Code (0040,100A) 3 Not Used
Sequence
>>Include Code Sequence Macro
>Scheduled Procedure Step ID (0040,0009) 1C Used
>Scheduled Procedure Step Description (0040,0007) 3 Used
>Scheduled Protocol Code Sequence (0040,0008) 3 Used
>>Include Code Sequence Macro
>>Protocol Context Sequence (0040,0440) 3 Not Used
>>>Include Content Item Macro
>>>Content Item Modifier Sequence (0040,0441) 3 Not Used
>>>>Include Content Item Macro
Performed Procedure Step ID (0040,0253) 3 Used

54
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Performed Procedure Step Start Date (0040,0244) 3 Used
Performed Procedure Step Start Time (0040,0245) 3 Used
Performed Procedure Step Description (0040,0254) 3 Used
Performed Protocol Code Sequence (0040,0260) 3 Not Used
>Include Code Sequence Macro
>>Protocol Context Sequence (0040,0440) 3 Not Used
>>>Include Content Item Macro
>>>Content Item Modifier Sequence (0040,0441) 3 Not Used
>>>>Include Content Item Macro
Comments on the Performed Procedure (0040,0280) 3 Not Used
Step

4.4.4 Equipment Entity Modules

4.4.4.1 General Equipment Module


TABLE 4-6
GENERAL EQUIPMENT MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
Manufacturer (0008,0070) 2 This value is set to GE Hualun Medical
Systems, Co. Ltd.
Institution Name (0008,0080) 3 Configurable by user on setup screen (The field
name is Hospital Name).
Institution Address (0008,0081) 3 Not Used
Station Name (0008,1010) 3 Configurable by user on DICOM setup screen.
Institutional Department Name (0008,1040) 3 Not Used
Manufacturer's Model Name (0008,1090) 3 This value is set to type and model of the
system: Brivo OEC 715/785/865.
Device Serial Number (0018,1000) 3 Configurable during system install.
Software Versions (0018,1020) 3 This value is set to workstation software
version.
Gantry ID (0018,1008) 3 Not Used
Spatial Resolution (0018,1050) 3 Not Used
Date of Last Calibration (0018,1200) 3 Not Used
Time of Last Calibration (0018,1201) 3 Not Used
Pixel Padding Value (0028,0120) 1C Not Used

55
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
4.4.5 Image Entity Modules

4.4.5.1 General Image Module


TABLE 4-7
GENERAL IMAGE MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
Instance Number (0020,0013) 2 This value is set to image number.
Patient Orientation (0020,0020) 2C The value is empty.
Content Date (0008,0023) 2C This value is set to the date image was taken.
Content Time (0008,0033) 2C This value is set to the time image was taken.
Image Type (0008,0008) 3 See 4.4.5.4.1
Acquisition Number (0020,0012) 3 This value is set to shot number.
Acquisition Date (0008,0022) 3 Not Used
Acquisition Time (0008,0032) 3 Not Used
Acquisition DateTime (0008,002A)) 3 This value is set to shot DateTime.
Referenced Image Sequence (0008,1140) 3 Not Used
>Include SOP Instance Reference Macro
>Purpose of Reference Code Sequence (0040,A170) 3 Not Used
>>Include Code Sequence Macro
Derivation Description (0008,2111) 3 Not Used
Derivation Code Sequence (0008,9215) 3 Not Used
>Include Code Sequence Macro
Source Image Sequence (0008,2112) 3 Not Used
>Include SOP Instance Reference Macro
>Purpose of Reference Code Sequence (0040,A170) 3 Not Used
>>Include Code Sequence Macro
>Spatial Locations Preserved (0028,135A) 3 Not Used
>Patient Orientation (0020,0020) 1C Not Used
Referenced Instance Sequence (0008,114A) 3 Not Used
>Include SOP Instance Reference Macro
>Purpose of Reference Code Sequence (0040,A170) 1 Not Used
>>Include Code Sequence Macro
Images in Acquisition (0020,1002) 3 Not Used
Image Comments (0020,4000) 3 Not Used
Quality Control Image (0028,0300) 3 Not Used
Burned In Annotation (0028,0301) 3 Not Used
Lossy Image Compression (0028,2110) 3 Not Used
Lossy Image Compression Ratio (0028,2112) 3 Not Used
Lossy Image Compression Method (0028,2114) 3 Not Used
Icon Image Sequence (0088,0200) 3 Not Used
>Include Image Pixel Macro

56
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Presentation LUT Shape (2050,0020) 3 Not Used

4.4.5.2 Image Pixel Module


TABLE 4-8
IMAGE PIXEL MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
Samples per Pixel (0028,0002) 1 The value is set to 1.
Photometric Interpretation (0028,0004) 1 The value is set to MONOCHROME2.
Rows (0028,0010) 1 This value is set to 1000.
Columns (0028,0011) 1 This value is set to 1000.
Bits Allocated (0028,0100) 1 This value is set to 16.
Bits Stored (0028,0101) 1 This value is set to 12.
High Bit (0028,0102) 1 This value is set to 11.
Pixel Representation (0028,0103) 1 This value is set to 0 for unsigned integer.
Pixel Data (7FE0,0010) 1 Pixel data of the image.
Planar Configuration (0028,0006) 1C Not Used
Pixel Aspect Ratio (0028,0034) 1C Not Used
Smallest Image Pixel Value (0028,0106) 3 Not Used
Largest Image Pixel Value (0028,0107) 3 Not Used
Red Palette Color Lookup Table (0028,1101) 1C Not Used
Descriptor
Green Palette Color Lookup Table (0028,1102) 1C Not Used
Descriptor
Blue Palette Color Lookup Table (0028,1103) 1C Not Used
Descriptor
Red Palette Color Lookup Table Data (0028,1201) 1C Not Used
Green Palette Color Lookup Table Data (0028,1202) 1C Not Used
Blue Palette Color Lookup Table Data (0028,1203) 1C Not Used
ICC Profile (0028,2000) 3 Not Used
Pixel Data Provider URL (0028,7FE0) 1C Not Used
Pixel Padding Range Limit (0028,0121) 1C Not Used

4.4.5.3 Contrast/Bolus Module


TABLE 4-9
CONTRAST/BOLUS MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
Contrast/Bolus Agent (0018,0010) 2 Got from MWL, else, set as empty.
Contrast/Bolus Agent Sequence (0018,0012) 3 Not Used
>Include Code Sequence Macro
Contrast/Bolus Route (0018,1040) 3 Not Used

57
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Contrast/Bolus Administration Route (0018,0014) 3 Not Used
Sequence
>Include Code Sequence Macro
>Additional Drug Sequence (0018,002A) 3 Not Used
>>Include Code Sequence Macro
Contrast/Bolus Volume (0018,1041) 3 Not Used
Contrast/Bolus Start Time (0018,1042) 3 Not Used
Contrast/Bolus Stop Time (0018,1043) 3 Not Used
Contrast/Bolus Total Dose (0018,1044) 3 Not Used
Contrast Flow Rate(s) (0018,1046) 3 Not Used
Contrast Flow Duration(s) (0018,1047) 3 Not Used
Contrast/Bolus Ingredient (0018,1048) 3 Not Used
Contrast/Bolus Ingredient Concentration (0018,1049) 3 Not Used

4.4.5.4 X-Ray Image Module


TABLE 4-10
X-RAY IMAGE MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
Frame Increment Pointer (0028,0009) 1C Not Used
Lossy Image Compression (0028,2110) 1C Not Used
Image Type (0008,0008) 1 See 4.4.5.4.1
Pixel Intensity Relationship (0028,1040) 1 Value is DISP
Samples per Pixel (0028,0002) 1 Value is 1
Photometric Interpretation (0028,0004) 1 Value is MONOCHROME2
Bits Allocated (0028,0100) 1 Value is 16
Bits Stored (0028,0101) 1 Value is 12
High Bit (0028,0102) 1 Value is 11
Pixel Representation (0028, 0103) 1 Value is 0000H
Scan Options (0018,0022) 3 Not Used.
Anatomic Region Sequence (0008,2218) 3 Not Used.

> Include Code Sequence Macro


>Anatomic Region Modifier Sequence (0008,2220) 3 Not Used
>> Include Code Sequence Macro
Primary Anatomic Structure Sequence (0008,2228) 3 Not Used
> Include Code Sequence Macro
>Primary Anatomic Structure Modifier (0008,2230) 3 Not Used
Sequence
>> Include Code Sequence Macro
R Wave Pointer (0028,6040) 3 Not Used

58
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Reference Image Sequence (0008,1140) 1C Not Used
>Referenced SOP Class UID (0008,1150) 1 Not Used
>Referenced SOP Class UID (0008,1155) 1 Not Used
>Referenced Frame Number (0008,1160) 1C Not Used
>Referenced Segment Number (0062,000B) 1C Not Used
>Purpose of Reference Code Sequence (0040,A170) 3 Not Used
>> Include Code Sequence Macro
Derivation Description (0008,2111) 3 Not Used
Acquisition Device Processing (0018,1400) 3 Not Used
Description
Frame Label Vector (0018,2002) 3 Not Used
Frame Dimension Pointer (0028,000A) 3 Not Used
Calibration Image (0050,0004) 3 Not Used

4.4.5.4.1 Image Type

Define the values of Image Type (0008,0008) that may be sent and under what
circumstances, or refer to the more general description in the General Image Module.

Value 1 shall have the following Enumerated Values:


ORIGINAL identifies an Original Image
DERIVED identifies a Derived Image.
Value 2 shall have the following Enumerated Value:
PRIMARY identifies a Primary Image
SECONDARY identifies a Secondary Image.
Specify which Enumerated Values of Value 3 are created/supported:
SINGLE PLANE

4.4.5.5 X-Ray Acquisition Module


TABLE 4-11
X-RAY ACQUISITION MODULE
Attribute Name Tag Type Attribute Description
KV (0018,0060) 2 KVP of the shot.
Radiation Setting (0018,1155) 1 Fixed as GR
X-Ray Tube Current (0018,1151) 2C Used
X-Ray Tube Current in microA (0018,8151) 3 Used
Exposure Time (0018,1150) 2C mSec of the shot.
Exposure Time in microS (0018,8150) 3 Not Used
Exposure (0018,1152) 2C Not Used
Exposure in microAs (0018,1153) 3 Not Used
Grid (0018,1166) 3 Not Used
Average Pulse Width (0018,1154) 3 Only used when 0018,115A is PULSED.
Pulse Width = 49.5ms
Radiation Mode (0018,115A) 3 Continuous: Fluoro/Digital Spot/HLF/Film

59
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Pulsed: All Pulse xxx
Type of Filters (0018,1161) 3 Not Used
Intensifier Size (0018,1162) 3 Not Used
Field of View Shape (0018,1147) 3 Not Used
Field of View Dimension(s) (0018,1149) 3 Not Used
Imager Pixel Spacing (0018,1164) 3 Not Used
Pixel Spacing (0028,0030) 1C Not Used
Pixel Spacing Calibration Type (0028,0A02) 3 Not Used
Pixel Spacing Calibration Description (0028,0A04) 1C Not Used
Focal Spot (0018,1190) 3 Not Used
Image and Fluoroscopy Area Dose (0018,115E) 3 DAP of the exam.
Product

4.4.5.6 XA Positioner Module


TABLE 4-12
XA POSITIONER MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
Distance Source to Patient (0018,1111) 3 Not Used
Distance Source to Detector (0018,1110) 3 1000mm.
Estimated Radiographic Magnification (0018,1114) 3 Not Used
Factor
Positioner Motion (0018,1500) 2C Not Used
Positioner Primary Angle (0018,1510) 2 Send with empty.
Positioner Secondary Angle (0018,1511) 2 Send with empty.
Positioner Primary Angle Increment (0018,1520) 2C Not Used
Positioner Secondary Angle Increment (0018,1521) 2C Not Used
Detector Primary Angle (0018,1530) 3 Not Used
Detector Secondary Angle (0018,1531) 3 Not Used

4.4.5.7 Overlay Plane Module


The product supports up to 1 overlay in an image.

TABLE 4-13
OVERLAY PLANE MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
Overlay Rows (60xx,0010) 1 Value is 1000.
Overlay Columns (60xx,0011) 1 Value is 1000.
Overlay Type (60xx,0040) 1 Value is G.
Overlay Origin (60xx,0050) 1 Value is 0\0.

60
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Overlay Bits Allocated (60xx,0100) 1 1
Overlay Bit Position (60xx,0102) 1 0
Overlay Data (60xx,3000) 1 Overlay pixel data
Overlay Description (60xx,0022) 3 Value is "Annotation/Measurement Graphic
Layer".
Overlay Subtype (60xx,0045) 3 Not Used
Overlay Label (60xx,1500) 3 Value is "Annotation/Measurement Graphic
Layer".
ROI Area (60xx,1301) 3 Not Used
ROI Mean (60xx,1302) 3 Not Used
ROI Standard Deviation (60xx,1303) 3 Not Used

4.4.5.8 VOI LUT module


TABLE 4-14
VOI LUT MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
VOI LUT Sequence (0028,3010) 1C Not Used
>LUT Descriptor (0028,3002) 1 Not Used
>LUT Explanation (0028,3003) 3 Not Used
>LUT Data (0028,3006) 1 Not Used
Window Center (0028,1050) 1C Only one value is present.
Set to 2^bitsStored/2 when export to DICOM
images
Window Width (0028,1051) 1C Only one value is present.
Set to 2^bitsStored 1 when export to DICOM
images
Window Center & Width Explanation (0028,1055) 3 Not Used

4.4.5.9 SOP Common Module


TABLE 4-15
SOP COMMON MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
SOP Class UID (0008,0016) 1 SOP Class UID for the Class that the dataset
contains, just XA UID.
SOP Instance UID (0008,0018) 1 Created by equipment, using prefix:
715: 1.2.840.113619.2.320.
785: 1.2.840.113619.2.321.
865: 1.2.840.113619.2.333.
+MAC+imageIndicator(40)+TimeStamp_Exa
m
MAC: Number string of device MAC address.

61
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Specific Character Set (0008,0005) 1C Always set to ISO_IR 100( = Latin Alphabet
No. 1)
Instance Creation Date (0008,0012) 3 Not Used
Instance Creation Time (0008,0013) 3 Not Used
Instance Creator UID (0008,0014) 3 Not Used
Related General SOP Class UID (0008,001A) 3 Not Used
Original Specialized SOP Class UID (0008,001B) 3 Not Used
Coding Scheme Identification (0008,0110) 3 Not Used
Sequence
>Coding Scheme Designator (0008,0102) 1 Not Used
>Coding Scheme Registry (0008,0112) 1C Not Used
>Coding Scheme UID (0008,010C) 1C Not Used
>Coding Scheme External ID (0008,0114) 2C Not Used
>Coding Scheme Name (0008,0115) 3 Not Used
>Coding Scheme Version (0008,0103) 3 Not Used
>Coding Scheme Responsible (0008,0116) 3 Not Used
Organization
Timezone Offset From UTC (0008,0201) 3 Not Used
Contributing Equipment Sequence (0018,A001) 3 Not Used
>Purpose of Reference Code (0040,A170) 1 Not Used
Sequence
>>Include Code Sequence Macro
>Manufacturer (0008,0070) 1 Not Used
>Institution Name (0008,0080) 3 Not Used
>Institution Address (0008,0081) 3 Not Used
>Station Name (0008,1010) 3 Not Used
>Institutional Department Name (0008,1040) 3 Not Used
>Manufacturers Model Name (0008,1090) 3 Not Used
>Device Serial Number (0018,1000) 3 Not Used
>Software Versions (0018,1020) 3 Not Used
>Spatial Resolution (0018,1050) 3 Not Used
>Date of Last Calibration (0018,1200) 3 Not Used
>Time of Last Calibration (0018,1201) 3 Not Used
>Contribution DateTime (0018,A002) 3 Not Used
>Contribution Description (0018,A003) 3 Not Used
Instance Number (0020,0013) 3 Image number.
SOP Instance Status (0100,0410) 3 Not Used
SOP Authorization Date and Time (0100,0420) 3 Not Used
SOP Authorization Comment (0100,0424) 3 Not Used
Authorization Equipment (0100,0426) 3 Not Used
Certification Number

62
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
MAC Parameters Sequence (4FFE,0001) 3 Not Used
>MAC ID Number (0400,0005) 1 Not Used
>MAC Calculation Transfer Syntax (0400,0010) 1 Not Used
UID
>MAC Algorithm (0400,0015) 1 Not Used
>Data Elements Signed (0400,0020) 1 Not Used
Digital Signatures Sequence (FFFA,FFFA) 3 Not Used
>MAC ID Number (0400,0005) 1 Not Used
>Digital Signature UID (0400,0100) 1 Not Used
>Digital Signature DateTime (0400,0105) 1 Not Used
>Certificate Type (0400,0110) 1 Not Used
>Certificate of Signer (0400,0115) 1 Not Used
>Signature (0400,0120) 1 Not Used
>Certified Timestamp Type (0400,0305) 1C Not Used
>Certified Timestamp (0400,0310) 3 Not Used
>Digital Signature Purpose Code (0400,0401) 3 Not Used
Sequence
>>Include Code Sequence Macro
Encrypted Attributes Sequence (0400,0500) 1C Not Used
>Encrypted Content Transfer Syntax (0400,0510) 1 Not Used
UID
>Encrypted Content (0400,0520) 1 Not Used
Original Attributes Sequence (0400,0561) 3 Not Used
>Source of Previous Values (0400,0564) 2 Not Used
>Attribute Modification DateTime (0400,0562) 1 Not Used
>Modifying System (0400,0563) 1 Not Used
>Reason for the Attribute (0400,0565) 1 Not Used
Modification
>Modified Attributes Sequence (0400,0550) 1 Not Used
>>Any Attribute from the main data set that was modified or removed; may include Sequence Attributes and their
Items.
HL7 Structured Document Reference (0040,A390) 1C Not Used
Sequence
>Referenced SOP Class UID (0008,1150) 1 Not Used
>Referenced SOP Instance UID (0008,1155) 1 Not Used
>HL7 Instance Identifier (0040,E001) 1 Not Used
>Retrieve URI (0040,E010) 3 Not Used
Longitudinal Temporal Information (0028,0303) 3 Used.
Modified If Include Remove Patient Information then fix
to "MODIFIED".

63
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

4.5 STANDARD EXTENDED AND PRIVATE DATA ATTRIBUTES


Not applicable

4.5.1 Standard Attributes


Not applicable

4.6 STANDARD EXTENDED AND PRIVATE CONTEXT GROUPS


Not applicable

64
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

5. X-RAY RADIATION DOSE STRUCTURED REPORT


INFORMATION OBJECT IMPLEMENTATION

5.1 INTRODUCTION

This section specifies the use of the DICOM X-Ray Radiation Dose SR IOD to represent
results produced and/or received by this implementation. Corresponding attributes are
conveyed using the module construct.

5.2 BRIVO OEC 715/785/865 MAPPING OF DICOM ENTITIES

The Brivo OEC 715/785/865 maps DICOM Information Entities to local Information
Entities in the products database and user interface.

TABLE 5-1
MAPPING OF DICOM ENTITIES TO BRIVO OEC 715/785/865 ENTITIES
DICOM IE Brivo OEC 715/785/865 Entity
Patient Patient
Study Exam
Series Series
Equipment Equipment
Document Document

5.3 IOD MODULE TABLE

The X-Ray Radiation Dose Structured Report Information Object Definitions comprise
the modules of the following tables, plus Standard Extended and Private Attributes.
Standard Extended and Private Attributes are described in Section 5.5

The contents of the SR Document Content are constrained by the supported template, as
identified in Section 5.4.5.2.1.1. Standard, Standard Extended and Private Templates
are further described in Section 5.7.

TABLE 5-2
STRUCTURE REPORT IOD MODULES
Entity Name Module Name Usage Reference
Patient Patient Used 5.4.1.1
Specimen Not Used N/A
Identification
Clinical Trial Subject Not Used N/A
Study General Study Used 5.4.2.1
Patient Study Used 5.4.2.2
Clinical Trial Study Not Used N/A

65
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Series SR Document Series Used 5.4.3.1
Clinical Trial Series Not Used N/A
Frame Of Synchronization Not Used N/A
Reference
Equipment General Equipment Used 5.4.4.1
Document SR Document General Used 5.4.5.1
SR Document Content Used 5.4.5.2
SOP Common Used 5.4.5.3

5.4 INFORMATION MODULE DEFINITIONS


Please refer to DICOM Part 3 (Information Object Definitions) for a description of each
of the entities, modules, and attributes contained within the SR Information Objects.

The following modules are included to convey Enumerated Values, Defined Terms, and
Optional Attributes supported and/or expected. Type 1 & Type 2 Attributes are also
included for completeness and to define what values they may take and where these
values are obtained from when generating the instance as well as what are the expected
values when loading such instance. It should be noted that they are the same ones as
defined in the DICOM Standard Part 3 (Information Object Definitions). Also note that
Attributes not present in tables are not supported.

5.4.1 Patient Entity Modules

5.4.1.1 Patient Module


TABLE 5-3
PATIENT MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
Patient's Name (0010,0010) 2 May be entered from user interface or got from
Worklist.
Supports 5 different components delimited by
^. Supports a maximum length of 64 bytes
including the delimiter.
Patient ID (0010,0020) 2 May be entered from user interface or got from
Worklist.
Truncated to 64 bytes, end truncated.
Issuer of Patient ID (0010,0021) 3 Got from Worklist.
Truncated to 64 bytes, end truncated.
Patient's Birth Date (0010,0030) 2 May be entered from user interface or got from
Worklist
Truncated to 8 bytes, end truncated
Patient's Sex (0010,0040) 2 May be entered from user interface or got from
Worklist.
Truncated to 2 bytes, end truncated
Referenced Patient Sequence (0008,1120) 3 Not Used

66
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Include SOP Instance Reference Macro
Patient's Birth Time (0010,0032) 3 Not Used
Other Patient IDs (0010,1000) 3 Not Used
Other Patient IDs Sequence (0010,1002) 3 Not Used
>Patient ID (0010,0020) 1 Not Used
>Issuer of Patient ID (0010,0021) 1 Not Used
>Type of Patient ID (0010,0022) 1 Not Used
Other Patient Names (0010,1001) 3 Not Used
Ethnic Group (0010,2160) 3 Not Used
Patient Comments (0010,4000) 3 May be entered from user interface or got from
Worklist
From user interface truncated to 512 bytes.
From worklist truncated to 10240 bytes.
Patient Species Description (0010,2201) 1C Not Used
Patient Species Code Sequence (0010,2202) 1C Not Used
>Include Code Sequence Macro
Patient Breed Description (0010,2292) 2C Not Used
Patient Breed Code Sequence (0010,2293) 2C Not Used
>Include Code Sequence Macro
Breed Registration Sequence (0010,2294) 2C Not Used
>Breed Registration Number (0010,2295) 1 Not Used
>Breed Registry Code Sequence (0010,2296) 1 Not Used
>>Include Code Sequence Macro
Responsible Person (0010,2297) 2C Not Used
Responsible Person Role (0010,2298) 1C Not Used
Responsible Organization (0010,2299) 2C Not Used
Patient Identity Removed (0012,0062) 3 Used.
If Patient Information is removed then set to
"YES", otherwise set to "NO"
De-identification Method (0012,0063) 1C Used.
If Patient Information is removed. Set to
"Limited Data Set" in this case
De-identification Method Code Sequence (0012,0064) 1C Not Used
>Include Code Sequence Macro

67
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
5.4.2 Study Entity Modules

5.4.2.1 General Study Module


TABLE 5-4
GENERAL STUDY MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
Study Instance UID (0020,000D) 1 The product will adopt the Study Instance UID
which is returned in the MWL response. If
there is no Study Instance UID returned from
MWL server, then new Study Instance UID
will be created.
Study Date (0008,0020) 2 This value is set to the exam created date.
Study Time (0008,0030) 2 This value is set to the exam created time.
Referring Physician's Name (0008,0090) 2 Got from Worklist.
Referring Physician Identification (0008,0096) 3 Not Used
Sequence
>Include Person Identification Macro
Study ID (0020,0010) 2 Generated by equipment.
Accession Number (0008,0050) 2 May be entered from the user interface or got
from Worklist.
Study Description (0008,1030) 3 May be entered from the user interface or got
from Worklist.
The default value will be set as Requested
Procedure Description when the tag got from
MWL is not empty. Else this value will be set
as Procedure Description on user interface.
Physician(s) of Record (0008,1048) 3 Not Used
Physician(s) of Record Identification (0008,1049) 3 Not Used
Sequence
>Include Person Identification Macro
Name of Physician(s) Reading Study (0008,1060) 3 Not Used
Physician(s) Reading Study Identification (0008,1062) 3 Not Used
Sequence
>Include Person Identification Macro
Referenced Study Sequence (0008,1110) 3 Got from Worklist.
>Include SOP Instance Reference Macro
Procedure Code Sequence (0008,1032) 3 Not Used
>Include Code Sequence Macro

5.4.2.2 Patient Study Module


TABLE 5-5
PATIENT STUDY MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description

68
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Admitting Diagnoses Description (0008,1080) 3 Not Used
Admitting Diagnoses Code Sequence (0008,1084) 3 Not Used
>Include Code Sequence Macro
Patient's Age (0010,1010) 3 Not Used
Patient's Size (0010,1020) 3 Used
Patient's Weight (0010,1030) 3 Used
Occupation (0010,2180) 3 Not Used
Additional Patients History (0010,21B0) 3 Not Used
Admission ID (0038,0010) 3 Not Used
Issuer of Admission ID (0038,0011) 3 Not Used
Service Episode ID (0038,0060) 3 Not Used
Issuer of Service Episode ID (0038,0061) 3 Not Used
Service Episode Description (0038,0062) 3 Not Used
Patients Sex Neutered (0010,2203) 2C Not Used

5.4.3 Series Entity Modules

5.4.3.1 SR Document Series Module


TABLE 5-6
SR DOCUMENT SERIES MODULE ATTRIBUTES
Attribute Name Tag Type Use
Modality (0008,0060) 1 Value = SR
Series Instance UID (0020,000E) 1 Created by equipment, using prefix:
715: 1.2.840.113619.2.320.
785: 1.2.840.113619.2.321.
865: 1.2.840.113619.2.333.
+MAC+seriesIndicator(21)+TimeStamp_Exam
MAC: Number string of device MAC address.
Series Number (0020,0011) 1 A number that identifies this Series.
This number will be set to 1 for image series and
set to 2 for RDSR series
Series Date (0008,0021) 3 First shot date.
Series Time (0008,0031) 3 First shot time.
Series Description (0008,103E) 3 The value equal protocol Name.
Referenced Performed Procedure Step (0008,1111) 2 Always send.
Sequence Referenced SOP Class UID = MPPS SOP Class
UID.
Referenced SOP Instance UID = MPPS SOP
Instance UID.
> Referenced SOP Class / Instance UIDs

69
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
5.4.4 Equipment Entity Modules

5.4.4.1 General Equipment Module


TABLE 5-7
GENERAL EQUIPMENT MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
Manufacturer (0008,0070) 2 This value is set to GE Hualun Medical
Systems, Co. Ltd.
Institution Name (0008,0080) 3 Configurable by user on setup screen (The field
name is Hospital Name).
Institution Address (0008,0081) 3 Not Used
Station Name (0008,1010) 3 Configurable by user on DICOM setup screen.
Institutional Department Name (0008,1040) 3 Not Used
Manufacturer's Model Name (0008,1090) 3 This value is set to type and model of the
system: Brivo OEC 715/785/865.
Device Serial Number (0018,1000) 3 Configurable during system installation.
Software Versions (0018,1020) 3 This value is set to workstation software
version.
Gantry ID (0018,1008) 3 Not Used
Spatial Resolution (0018,1050) 3 Not Used
Date of Last Calibration (0018,1200) 3 Not Used
Time of Last Calibration (0018,1201) 3 Not Used
Pixel Padding Value (0028,0120) 1C Not Used

5.4.5 Document Entity Modules

5.4.5.1 SR Document General Module


TABLE 5-8
SR DOCUMENT GENERAL MODULE ATTRIBUTES
Attribute Name Tag Type Use
Instance Number (0020,0013) 1 Value is 1.
Completion Flag (0040,A491) 1 COMPLETE
Completion Flag Description (0040,A492) 3 Not Used
Verification Flag (0040,A493) 1 UNVERIFIED
Content Date (0008,0023) 1 First shot date.
Content Time (0008,0033) 1 First shot time.
Verifying Observer Sequence (0040,A073) 1C Not Used.
>Verifying Observer Name (0040,A075) 1 Not Used
>Verifying Observer Identification Code (0040,A088) 2 Not Used
Sequence
>>Include Code Sequence Macro
>Verifying Organization (0040,A027) 1 Not Used
>Verification DateTime (0040,A030) 1 Not Used

70
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Author Observer Sequence (0040,A078) 3 Used
>Include Identified Person or Device Macro
Participant Sequence (0040,A07A) 3 Not Used
>Participation Type (0040,A080) 1 Not Used
>Participation DateTime (0040,A082) 2 Not Used
>Include Identified Person or Device Macro
Custodial Organization Sequence (0040,A07C) 3 Not Used
>Institution Name (0008,0080) 2 Not Used
>Institution Code Sequence (0008,0082) 2 Not Used
>>Include Code Sequence Macro
Predecessor Documents Sequence (0040,A360) 1C Not Used
>Include Hierarchical SOP Instance Reference Macro
Identical Documents Sequence (0040,A525) 1C Not Used
>Include Hierarchical SOP Instance Reference Macro
Referenced Request Sequence (0040,A370) 1C Used
>Study Instance UID (0020,000D) 1 The product will adopt the Study Instance UID
which is returned in the MWL response. If there
is no Study Instance UID returned from MWL
server, then new Study Instance UID will be
created by system during exam creation.
>Referenced Study Sequence (0008,1110) 2 Set as empty
>>Include SOP Instance Reference Macro
>Accession Number (0008,0050) 2 The product will adopt the Accession Number
which is returned in the MWL response. If there
is no Accession Number returned from MWL
server, then the value will keep empty.
For local created exam, user could input value
from screen during exam creation.
>Placer Order Number/Imaging Service (0040,2016) 2 Value is empty.
Request
>Filler Order Number/Imaging Service (0040,2017) 2 Value is empty.
Request
>Requested Procedure ID (0040,1001) 2 This value could got from MWL response. If not
contained in MWL response, then keep empty as
default.
>Requested Procedure Description (0032,1060) 2 This value could got from MWL response. If not
contained in MWL response, then keep empty as
default.
>Requested Procedure Code Sequence (0032,1064) 2 Value is empty.
>>Include Code Sequence Macro
>Reason for the Requested Procedure (0040,1002) 3 Not Used
>Reason for Requested Procedure Code (0040,100A) 3 Not Used
Sequence
>>Include Code Sequence Macro

71
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Performed Procedure Code Sequence (0040,A372) 2 Empty.
>Include Code Sequence Macro
Current Requested Procedure Evidence (0040,A375) 1C With the value(s) from the system of all saved
Sequence images for the exam.
This is the list of all images whether they have
been stored to PACS or not.
>Include Hierarchical SOP Instance Reference Macro
Pertinent Other Evidence Sequence (0040,A385) 1C Not Used
> Include Hierarchical SOP Instance Reference Macro
Referenced Instance Sequence (0008,114A) 1C Not Used
>Include SOP Instance Reference Macro
>Purpose of Reference Code Sequence (0040,A170) 1 Not Used
>>Include Code Sequence Macro

5.4.5.2 SR Document Content Module


TABLE 5-9
SR DOCUMENT CONTENT MODULE ATTRIBUTES
Attribute Name Tag Type Use
Observation DateTime (0040,A032) 1C Not Used
Content Template Sequence (0040,A504) 1C Not Used
>Mapping Resource (0008,0105) 1 Not Used
>Template Identifier (0040,DB00) 1 Not Used
Value Type (0040,A040) 1 CONTAINER
Continuity of Content (0040,A050) 1C SEPARATE
Concept Name Code Sequence (0040,A043) 1C Used
>Include Code Sequence Macro
Insert Concept Value attribute(s) (113701,DCM, X-Ray Radiation Dose Report)
Content Sequence (0040,A730) 1C Used
> Relationship Type (0040,A010) 1 HAS CONCEPT MOD
> Referenced Content Item Identifier (0040,DB73) 1C Not Used
> Insert SR DocumentContent Module Recursive inclusion to create document content
tree.
See section 5.4.5.2.1 for the list of supported
templates

5.4.5.2.1 SR Document Content Descriptions

5.4.5.2.1.1 Content Template


The product supports the following root Templates for SR SOP Instances created,
processed, or displayed by the product.

72
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
TABLE 5-10
SR ROOT TEMPLATES
SOP Class Template ID Template Name Use
X-Ray Radiation Dose 10001 X-Ray Radiation Dose Create/Display
SR
Refer to section 5.7 for a detailed description of the supported templates.

5.4.5.3 SOP Common Module


TABLE 5-41
SOP COMMON MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
SOP Class UID (0008,0016) 1 SOP Class UID for the Class that the dataset
contains, just RDSR UID.
SOP Instance UID (0008,0018) 1 Created by equipment, using prefix:
715: 1.2.840.113619.2.320.
785: 1.2.840.113619.2.321.
865: 1.2.840.113619.2.333.
+MAC+imageIndicator(40)+TimeStamp_Exa
m
MAC: Number string of device MAC address.
Specific Character Set (0008,0005) 1C Always set to ISO_IR 100( = Latin Alphabet
No. 1)
Instance Creation Date (0008,0012) 3 Used
Instance Creation Time (0008,0013) 3 Used
Instance Creator UID (0008,0014) 3 Used
Related General SOP Class UID (0008,001A) 3 Not Used
Original Specialized SOP Class UID (0008,001B) 3 Not Used
Coding Scheme Identification (0008,0110) 3 Not Used
Sequence
>Coding Scheme Designator (0008,0102) 1 Not Used
>Coding Scheme Registry (0008,0112) 1C Not Used
>Coding Scheme UID (0008,010C) 1C Not Used
>Coding Scheme External ID (0008,0114) 2C Not Used
>Coding Scheme Name (0008,0115) 3 Not Used
>Coding Scheme Version (0008,0103) 3 Not Used
>Coding Scheme Responsible (0008,0116) 3 Not Used
Organization
Timezone Offset From UTC (0008,0201) 3 Not Used
Contributing Equipment Sequence (0018,A001) 3 Not Used
>Purpose of Reference Code (0040,A170) 1 Not Used
Sequence
>>Include Code Sequence Macro
>Manufacturer (0008,0070) 1 Not Used

73
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Institution Name (0008,0080) 3 Not Used
>Institution Address (0008,0081) 3 Not Used
>Station Name (0008,1010) 3 Not Used
>Institutional Department Name (0008,1040) 3 Not Used
>Manufacturers Model Name (0008,1090) 3 Not Used
>Device Serial Number (0018,1000) 3 Not Used
>Software Versions (0018,1020) 3 Not Used
>Spatial Resolution (0018,1050) 3 Not Used
>Date of Last Calibration (0018,1200) 3 Not Used
>Time of Last Calibration (0018,1201) 3 Not Used
>Contribution DateTime (0018,A002) 3 Not Used
>Contribution Description (0018,A003) 3 Not Used
Instance Number (0020,0013) 3 Image number.
SOP Instance Status (0100,0410) 3 Not Used
SOP Authorization Date and Time (0100,0420) 3 Not Used
SOP Authorization Comment (0100,0424) 3 Not Used
Authorization Equipment (0100,0426) 3 Not Used
Certification Number
MAC Parameters Sequence (4FFE,0001) 3 Not Used
>MAC ID Number (0400,0005) 1 Not Used
>MAC Calculation Transfer Syntax (0400,0010) 1 Not Used
UID
>MAC Algorithm (0400,0015) 1 Not Used
>Data Elements Signed (0400,0020) 1 Not Used
Digital Signatures Sequence (FFFA,FFFA) 3 Not Used
>MAC ID Number (0400,0005) 1 Not Used
>Digital Signature UID (0400,0100) 1 Not Used
>Digital Signature DateTime (0400,0105) 1 Not Used
>Certificate Type (0400,0110) 1 Not Used
>Certificate of Signer (0400,0115) 1 Not Used
>Signature (0400,0120) 1 Not Used
>Certified Timestamp Type (0400,0305) 1C Not Used
>Certified Timestamp (0400,0310) 3 Not Used
>Digital Signature Purpose Code (0400,0401) 3 Not Used
Sequence
>>Include Code Sequence Macro
Encrypted Attributes Sequence (0400,0500) 1C Not Used
>Encrypted Content Transfer Syntax (0400,0510) 1 Not Used
UID
>Encrypted Content (0400,0520) 1 Not Used
Original Attributes Sequence (0400,0561) 3 Not Used

74
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Source of Previous Values (0400,0564) 2 Not Used
>Attribute Modification DateTime (0400,0562) 1 Not Used
>Modifying System (0400,0563) 1 Not Used
>Reason for the Attribute (0400,0565) 1 Not Used
Modification
>Modified Attributes Sequence (0400,0550) 1 Not Used
>>Any Attribute from the main data set that was modified or removed; may include Sequence Attributes and their
Items.
HL7 Structured Document Reference (0040,A390) 1C Not Used
Sequence
>Referenced SOP Class UID (0008,1150) 1 Not Used
>Referenced SOP Instance UID (0008,1155) 1 Not Used
>HL7 Instance Identifier (0040,E001) 1 Not Used
>Retrieve URI (0040,E010) 3 Not Used
Longitudinal Temporal Information (0028,0303) 3 Used.
Modified If Include Remove Patient Information then fix
to "MODIFIED".

5.5 STANDARD EXTENDED AND PRIVATE DATA ATTRIBUTES


Not applicable

5.6 STANDARD EXTENDED AND PRIVATE CONTEXT GROUPS

The Product supports coded terminology using Standard Extended, Private, and
Configurable Context Groups defined in the following sections.

5.6.1 Standard Extended Context Groups


The Product supports the following extensions to standard Context Groups for SR SOP
Instances created by this product. Extensions are indicated by bold text.

5.6.2 Configurable Context Groups


The Product supports the following Configurable Context Groups for SR SOP Instances
created by this product.

Context Group Default Value Set Use

N/A N/A N/A

5.7 STANDARD, STANDARD EXTENDED AND PRIVATE TEMPLATES


The Product supports the Standard Extended and Private Templates defined in the
following sections.

75
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
5.7.1 Standard Templates
The Product supports the following standard templates for SOP Instances created by this
product.

5.7.1.1 Template ID 10001 X-Ray Radiation Dose


TID 10001
PROJECTION X-RAY RADIATION DOSE
Type: Extensible Order: Significant
NL Rel with VT Concept Name VM Req Condition Value Set Constraint
Parent Type
1 CONTAIN EV (113701, 1 M Root Node
ER DCM, X-Ray
Radiation Dose
Report)
2 > HAS CODE EV (121058, 1 M DT (113704, DCM,
CONCEPT DCM, Procedure Projection X-Ray)
MOD reported)
3 >> HAS CODE EV (G-C0E8, 1 M (R-
CONCEPT SRT, Has Intent) 408C3,SRT,Diagnostic
MOD Intent)
4 > INCLUDE DTID (1002) 1-n M
Observer Context
5 > HAS OBS CODE EV (113705, 1 M (113016, DCM,
CONTEXT DCM, Scope of "Performed Procedure
Accumulation) Step").
6 >> HAS UIDREF DCID (10001) 1 M (121126, DCM,
PROPERTIE UID Types "Performed Procedure
S Step SOP Instance
UID")
7 > CONTAINS INCLUDE DTID (10002) 1 MC IFF Single Plane $Plane = EV (113622,
Accumulated X- system DCM, Single Plane
Ray Dose
8 > CONTAINS INCLUDE DTID (10002) 1 MC IFF Biplane system Not Used
Accumulated X-
Ray Dose
9 > CONTAINS INCLUDE DTID (10002) 1 MC IFF Biplane system Not Used
Accumulated X-
Ray Dose
10 > CONTAINS INCLUDE DTID (10003) 1-n M
Irradiation Event
X-Ray Data
11 > CONTAINS TEXT EV (121106, 1 U Patient Comment
DCM,
Comment)
12 > CONTAINS IMAGE EV (121342, 1-n U Not Used
DCM, Dose
Image)
13 > CONTAINS INCLUDE DTID (1020) 1 U Not Used
Person Participant

76
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
14 > CONTAINS CODE EV (113854, 1-n M (113856, DCM,
DCM, Source of "Automated Data
Dose Collection")
Information)

5.7.1.2 Template ID 10002 Accumulated X-Ray Dose


TID 10002
ACCUMULATED X-RAY DOSE
Type: Extensible Order: Significant
NL Rel with VT Concept Name VM Req Condition Value Set Constraint
Parent Type
1 CONTAIN EV (113702, 1 M
ER DCM,
Accumulated X-
Ray Dose Data)
2 > HAS CODE EV (113764, 1 M (113622,DCM, Single
CONCEP DCM, Plane)
T MOD Acquisition
Plane
3 > CONTAI CONTAIN EV (122505, 1-n MC IFF Calibration Data is
NS ER DCM, available
Calibration)
4 >> HAS CODE EV (113794, 1 M (A-2C090, SRT,
CONCEP DCM, Dose "Dosimeter")
T MOD Measurement
Device)
5 >> CONTAI DATETIME EV (113723, 1 M The date of DAP
NS DCM, Calibration performed.
Calibration
Date)
6 >> CONTAI NUM EV (122322, 1 M Units = EV (1, UCUM,
NS DCM, no units)
Calibration Value is 1.
Factor)
7 >> CONTAI NUM EV (113763, 1 M Units = EV (%, UCUM,
NS DCM, Percent)
Calibration Value is 35%.
Uncertainty)
8 >> CONTAI TEXT EV (113724, 1 M Regional Sales
NS DCM, Administrator
Calibration (RSA)
Responsible
Party)
9 > CONTAI INCLUDE DTID (10004) 1 MC XOR row 10, IFF TID
NS Accumulated (10001) Row 2 =
Projection X-Ray (113704, DCM,
Dose Projection X-Ray)
10 > CONTAI INCLUDE DTID (10005) 1 MC XOR row 9, IFF TID Not Used
NS Accumulated (10001) Row 2 = (P5-

77
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Mammography 40010, SRT,
X-Ray Dose Mammography)
11 > CONTAI INCLUDE DTID (1021) 1 MC Required if the $DeviceProcedureRole
NS Device irradiating device is = EV (113859, DCM,
Participant not the recording Irradiating Device)
device and the dose
was accumulated on a
single device.

5.7.1.3 Template ID 10003 Irradiation Event X-Ray Data


TID 10003
IRRADIATION EVENT X-RAY DATA
Type: Extensible Order: Significant
NL Rel with VT Concept Name VM Req Condition Value Set Constraint
Parent Type
1 CONTAIN EV (113706, 1 M
ER DCM, Irradiation
Event X-Ray
Data)
2 > HAS CODE EV (113764, 1 M (113622, DCM, "Single Plane")
CONCEP DCM,
T MOD Acquisition
Plane
3 > CONTAI DATETIM DT (111526, 1 M Date and time of first shot
NS E DCM, DateTime
Started)
4 > CONTAI CODE EV (113721, 1 M (P5-06000, SRT,
NS DCM, Irradiation "Fluoroscopy").
Event Type)
5 > CONTAI TEXT EV (125203, 1 U The text value of the IOD tag
NS DCM, (0018,1030) Protocol Name
Acquisition from the system.
Protocol)
6 > CONTAI CODE EV (T-D0005, 1 U Not Used
NS SRT, Anatomical
structure)
7 >> HAS CODE EV (G-C171, 1 UC If anatomy is bi-lateral Come from XA IOD
CONCEP SRT, Laterality") (0020,0060) Laterality.
T MOD
8 > CONTAI TEXT EV (113780, 1 MC IF Row 13 or Row 14 Not Used
NS DCM, Reference is present and Row 9
Point Definition) is not present
9 > CONTAI CODE EV (113780, 1 MC IF Row 13 or Row 14 (113861, DCM, "30cm in Front
NS DCM, Reference is present and Row 8 of Image Input Surface ").
Point Definition) is not present
10 > CONTAI UIDREF EV (113769, 1 M The UID value of the irradiation
NS DCM, Irradiation event from the system.
Event UID)

78
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
11 > CONTAI NUM EV (122130, 1 MC IFF TID (10001) Row Units = EV (Gy.m2, UCUM,
NS DCM, Dose Area 2 = (113704, DCM, "Gy.m2")
Product Projection X-Ray)
12 > CONTAI NUM EV (111631, 1 MC IFF TID (10001) Row Not Used
NS DCM, Average 2 = (P5-40010, SRT,
Glandular Dose) Mammography)
13 > CONTAI NUM EV (113738, 1 MC IFF TID (10001) Row Units = EV (Gy, UCUM, Gy)
NS DCM, Dose 2 = (113704, DCM, Dose applied at the Reference
(RP)) Projection X-Ray) Point (RP)
AND any of the
values of TID (10001)
Row 14 are not
(113858, DCM,
MPPS Content)
14 > CONTAI NUM EV (111636, 1 MC IFF TID (10001) Row Not Used
NS DCM, Entrance 2 = (P5-40010, SRT,
Exposure at RP) Mammography)
15 > CONTAI NUM EV (112011, 1 UC XOR Row 19 Not Used
NS DCM, Positioner
Primary Angle)
16 > CONTAI NUM EV ( 112012, 1 UC XOR Row 19 Not Used
NS DCM, Positioner
Secondary
Angle)
17 > CONTAI NUM EV (113739, 1 UC IFF Row 4 value = Not Used
NS DCM, Positioner (113613, DCM,
Primary End Rotational
Angle Acquisition)
18 > CONTAI NUM EV (113740, 1 UC IFF Row 4 value = Not Used
NS DCM, Positioner (113613, DCM,
Secondary End Rotational
Angle Acquisition)
19 > CONTAI NUM EV (113770, 1 UC XOR Rows 15,16 Not Used
NS DCM, Column
Angulation)
20 > CONTAI NUM EV (113790, 1 U Not Used
NS DCM, Collimated
Field Area
21 > CONTAI CONTAIN EV (113771, 1-n U Not Used
NS ER DCM, X-Ray
Filters)
22 >> CONTAI CODE EV (113772, 1 U Not Used
NS DCM, X-Ray
Filter Type)
23 >> CONTAI CODE EV (113757, 1 U Not Used
NS DCM, X-Ray
Filter Material)
24 >> CONTAI NUM EV (113758, 1 U Not Used
NS DCM, X-Ray
Filter Thickness
Minimum)

79
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
25 >> CONTAI NUM EV (113773, 1 U Not Used
NS DCM, X-Ray
Filter Thickness
Maximum)
26 > CONTAI CODE EV (113732, 1 UC IFF Row 4 value = 1. The value of code (113630,
NS DCM, Fluoro (P5-06000, SRT, DCM, "Continuous") for
Mode) Fluoroscopy) continuous mode irradiation
events.
2. The value of code (113631,
DCM, "Pulsed") for Pulsed
mode irradiation events.
27 > CONTAI NUM EV (113791, 1 MC IFF Row 26 value = Units = EV ({pulse}/s, UCUM,
NS DCM, Pulse (113631, DCM, pulse/s)
Rate) Pulsed) For Pulsed mode irradiation
events.
28 > CONTAI NUM EV (113768, 1 MC IFF Row 26 value = Units = EV (1, UCUM, no
NS DCM, Number of (113631, DCM, units)
Pulses) Pulsed) 1. For Pulsed mode irradiation
events.
2. Estimated:
a. Manual: exposure time/pulse
width
b. Auto: exposure time/pulse
width (margin of error might be
bigger than manual)
c. Pulse Width = 49.5ms
29 >> HAS CODE EV (121401, 1 MC IFF count of pulses in EV (R-10260, SRT,
CONCEP DCM, Row 28 is estimated Estimated)
T MOD Derivation)
30 > CONTAI NUM EV (113733, 1-n U Units = EV (kV, UCUM, kV)
NS DCM, KVP) For Pulsed/Continuous mode
irradiation events.
31 > CONTAI NUM EV (113734, 1-n U Units = EV (mA, UCUM,
NS DCM, X-Ray mA)
Tube Current)
32 > CONTAI NUM EV (113735, 1 U Units = EV (ms, UCUM, ms)
NS DCM, Exposure
Time)
33 > CONTAI NUM EV (113793, 1-n U Units = EV (ms, UCUM, ms)
NS DCM, Pulse Value is 49.5ms
Width)
34 > CONTAI NUM EV (113736, 1-n U Not Used
NS DCM,
Exposure)
35 > CONTAI NUM EV (113766, 1 U Not Used
NS DCM, Focal Spot
Size)
36 > CONTAI NUM EV (113742, 1 U Not Used
NS DCM, Irradiation

80
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Duration)
37 > CONTAI NUM EV (113767, 1 U Units = EV (mA, UCUM,
NS DCM, Average mA)
X-Ray Tube The same with above X-Ray
Current) Tube Current
38 > CONTAI CODE EV (113745, 1 U Not Used
NS DCM, Patient
Table
Relationship)
39 > CONTAI CODE EV (113743, 1 U Not Used
NS DCM, Patient
Orientation)
40 >> HAS CODE EV (113744, 1 M Not Used
CONCEP DCM, Patient
T MOD Orientation
Modifier)
41 > CONTAI NUM DCID (10008) 1-n U Not Used
NS Dose Related
Distance
Measurements
42 > CONTAI NUM EV (113754, 1 U Not Used
NS DCM, Table
Head Tilt Angle)
43 > CONTAI NUM EV (113755, 1 U Not Used
NS DCM, Table
Horizontal
Rotation Angle)
44 > CONTAI NUM EV (113756, 1 U Not Used
NS DCM, Table
Cradle Tilt
Angle)
45 > CONTAI CODE EV (123014 , 1 M 1. the code from the system
NS DCM, (Target using DICOM Part 16 Context
Region) ID 4031 values and supporting
the code (T-D0001, SRT,
"Topography unknown") for an
'Unknown' value which is
SNOMED term.
2. User could set this value in
EM->Additional Screen
Default is (T-D0001, SRT,
"Topography unknown")
46 > CONTAI CODE EV (111632, 1 U Not Used
NS DCM, Anode
Target Material
47 > CONTAI NUM EV (111633, 1 U Not Used
NS DCM,
Compression
Thickness)
48 > CONTAI NUM EV (111634, 1 U Not Used
NS DCM, Half Value

81
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Layer)

49 > CONTAI CODE EV (111635,DCM, 1-n U Not Used


NS X-Ray Grid)
50 > CONTAI INCLUDE DTID (4007) 1 U Not Used
NS Mammography
CAD Breast
Composition
51 > CONTAI TEXT EV (121106, 1 U Patient Comment
NS DCM,
Comment)
52 > CONTAI INCLUDE DTID (1020) 1-n U Not Used
NS Person Participant
53 > CONTAI INCLUDE DTID (1021) 1 MC Required if the Not Used
NS Device Participant irradiating device is
not the recording
device.
54 > CONTAI IMAGE EV (113795, 1-n MC IFF Image Object is All image UIDs.
NS DCM, Acquired created for this
Image) irradiation event
55 > CONTAI NUM EV (113845, 1 U Not Used
NS DCM, Exposure
Index)
56 > CONTAI NUM EV (113846, 1 U Not Used
NS DCM, Target
Exposure Index)
57 > CONTAI NUM EV (113847, 1 U Not Used
NS DCM, Deviation
Index)

5.7.1.4 Template ID 10004 Accumulated Projection X-Ray Dose


TID 10004
ACCUMULATED PROJECTION X-RAY DOSE
Type: Extensible Order: Significant
NL Rel with VT Concept Name VM Req Condition Value Set Constraint
Parent Type
1 NUM EV (113722, 1 M Units = EV (Gy.m2,
DCM, Dose UCUM, Gy.m2)
Area Product The numeric value of
Total) Total Dose Area
Product. (Radioscopic
and Radiographic)
2 NUM EV (113725, 1 MC IF any of the values of Units = EV (Gy,
DCM, Dose TID (10001) Row 14 UCUM, Gy)
(RP) Total) are not (113858, The numeric value of
DCM, MPPS Cumulative Air
Content). May be Kerma.(Radioscopic
present otherwise. and Radiographic )

82
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
3 NUM EV (113726, 1 MC IFF TID(10003) Row Units = EV (Gy.m2,
DCM, Fluoro 4 value = (P5-06000, UCUM, Gy.m2)
Dose Area SRT, Fluoroscopy) The numeric value of
Product Total) for at least one Total Radioscopic Dose
irradiation event Area Product.
4 NUM EV (113728, 1 MC IFF TID(10003) Row Units = EV (Gy,
DCM, Fluoro 4 value = (P5-06000, UCUM, Gy)
Dose (RP) SRT, Fluoroscopy) The numeric value of
Total) for at least one Cumulative Radioscopic
irradiation event AND Air Kerma.
any of the values of
TID (10001) Row 14
are not (113858,
DCM, MPPS
Content).
5 NUM EV (113730, 1 MC IFF TID(10003) Row Units = EV (s, UCUM,
DCM, Total 4 value = (P5-06000, s)
Fluoro Time) SRT, Fluoroscopy) The numeric value of
for at least one Total Radioscopic time
irradiation event. to the second.
6 NUM EV (113727, 1 M Units = EV (Gy.m2,
DCM, UCUM, Gy.m2)
Acquisition The numeric value of
Dose Area Total Radiographic
Product Total) Dose Area Product.
7 NUM EV (113729, 1 MC IF any of the values of Units = EV (Gy,
DCM, TID (10001) Row 14 UCUM, Gy)
Acquisition are not (113858, The numeric value of
Dose (RP) DCM, MPPS Cumulative
Total) Content). May be Radiographic Air
present otherwise. kerma.
8 NUM EV (113855, 1 M Units = EV (s, UCUM,
DCM, Total s)
Acquisition The numeric value of
Time) Total Radiographic
pedal time in
milliseconds.
9 NUM EV (113731, 1 U Units = EV (1, UCUM,
DCM, Total no units)
Number of The numeric value of
Radiographic Total Film and Digital
Frames) Spot Frames.
10 CODE EV (113780, 1 MC IF Row 2, Row 4 or (113861, DCM, "30cm
DCM, Row 7 is present and in Front of Image Input
Reference Point Row 11 is not present. Surface").
Definition)
11 TEXT EV (113780, 1 MC IF Row 2, Row 4 or Not Used
DCM, Row 7 is present and
Reference Point Row 10 is not present.
Definition)

83
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
5.7.1.5 Template ID 1002 Observer Context
TID 1002
OBSERVER CONTEXT
Type: Non-Extensible Order: Significant
NL Rel with VT Concept Name VM Req Condition Value Set Constraint
Parent Type
1 HAS OBS CODE EV (121005,DCM, 1 MC IF Observer type is device (121007, DCM, "Device")
CONTEX Observer Type)
T
2 HAS OBS INCLUDE DTID (1003) Person 1 MC IFF Row 1 value = Not Used
CONTEX observer identifying (121006,DCM, Person)
T attributes or Row 1 is absent
3 HAS OBS INCLUDE DTID (1004) Device 1 MC IFF Row 1 value =
CONTEX observer identifying (121007,DCM, Device)
T attributes

5.7.1.6 Template ID 1004 Device Observer Identifying Attributes

TID 1004
DEVICE OBSERVER IDENTIFYING ATTRIBUTES
Type: Extensible Order: Significant
NL Rel with VT Concept Name VM Req Condition Value Set Constraint
Parent Type
1 UIDREF EV (121012,DCM, 1 M System Generated
Device Observer UID)
2 TEXT EV (121013,DCM, 1 U Defaults to value of
Device Observer Name) Station Name (0008,1010)
in General Equipment
Module.
If 'Station Name' is blank
the value for Local AE
Title shall be used.
3 TEXT EV (121014,DCM, 1 U Defaults to value of
Device Observer Manufacturer (0008,0070)
Manufacturer) in General Equipment
Module
4 TEXT EV (121015,DCM, 1 U Defaults to value of
Device Observer Model Manufacturers Model
Name) Name (0008,1090) in
General Equipment
Module
5 TEXT EV (121016,DCM, 1 U Defaults to value of
Device Observer Serial Device Serial Number
Number) (0018,1000) in General
Equipment Module
6 TEXT EV (121017,DCM, 1 U Hospital Name.
Device Observer
Physical Location during
observation)
7 CODE EV (113876, DCM, 1-n U (113859, DCM,
Device Role in "Irradiating Device")

84
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Procedure)

5.7.2 Standard Extended Templates


The Product supports the following extensions or restrictions to standard templates for
SOP Instances created by this product. Extensions or restrictions are indicated by bold
text.

85
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

6. MODALITY WORKLIST QUERY IMPLEMENTATION

6.1 INTRODUCTION
This section specifies the use of the DICOM Modality Worklist Information Model used
to organize data and against which a Modality Worklist Query will be performed.

6.2 BRIVO OEC 715/785/865 MAPPING OF DICOM ENTITIES


The Brivo OEC 715/785/865 maps DICOM Information Entities to local Information
Entities in the products database and user interface.

TABLE 6-1
MAPPING OF DICOM ENTITIES TO BRIVO OEC 715/785/865 ENTITIES
DICOM Brivo OEC 715/785/865 Entity
Scheduled Procedure Step Exam
Requested Procedure Exam
Imaging Service Request Exam
Visit Not Applicable
Patient Patient

6.3 WORKLIST QUERY MODULE TABLE

See DICOM PS 3.3 and PS 3.4 for a complete definition of the entities, modules, and
attributes.

86
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
TABLE 6-2
MODALITY WORKLIST INFORMATION MODEL MODULES
Entity Name Module Name Reference
Scheduled Procedure Step SOP Common 6.4.1.1
Scheduled Procedure Step 6.4.1.2
Requested Procedure Requested Procedure 6.4.2.1
Imaging Service Request Imaging Service Request 6.4.3.1
Visit Visit Identification 6.4.4.1
Visit Status 6.4.4.2
Visit Relationship 6.4.4.3
Visit Admission 6.4.4.4
Patient Patient Relationship 6.4.5.1
Patient Identification 6.4.5.2
Patient Demographic 6.4.5.3
Patient Medical 6.4.5.4

6.4 WORKLIST QUERY MODULE DEFINITIONS

Please refer to DICOM Standard PS 3.3. (Information Object Definitions) for a


description of each of the query key attributes contained within the Modality Worklist
Information Model.

6.4.1 Common Scheduled Procedure Step Entity Modules

6.4.1.1 SOP Common Module


TABLE 6-3
SOP COMMON MODULE ATTRIBUTES
Attribute Name Tag Expected Expected Mapped into Note
Matching Returned Instance /
Key Type Key Type MPPS
Specific Character Set (0008,0005) O 1C Yes/Yes See Section 6.4.1.1.1 below

6.4.1.1.1 Specific Character Set


The attribute Specific Character Set (0008,0005) will always be sent.

As a Query SCU, it will similarly accept response items with ISO_IR 100 values of
Specific Character Set or empty value of Specific Character Set (absence of (0008,0005)
which mean usage of ASCII Chars only).

The product user interface will allow the user to enter characters from the console
keyboard that is within ISO 8859-1 characters.

Please also refer to Section 2.7.

87
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
6.4.1.2 Scheduled Procedure Step Module
TABLE 6-4
SCHEDULED PROCEDURE STEP MODULE ATTRIBUTES
Attribute Name Tag Expected Expected Mapped into Note
Matching Returned Instance /
Key Type Key Type MPPS
Scheduled Procedure (0040,0100) R 1 No/No
Step Sequence
>Scheduled Station (0040,0001) R 1 No/No No Matching is performed.
AE Title
>Scheduled (0040,0002) R 1 No/No Using the date configured from the DICOM
Procedure Step Start query filter screen, the following matching can
Date be done:
YYYY- year, MM- month, DD- day.
Range matching can also be done
YYYYMMDD-YYYYMMDD - all matches
included between those dates. The dash is
important
-YYYYMMDD - all matches prior to and
including this date. The beginning dash is
important.
YYYYMMDD- all matches after and including
this date. The ending dash is important.
Single matching where the dates entered is
equal.
The data entered on the user interface is passed
to the Server as entered; if it does not work see
the Server DICOM conformance statement.
>Scheduled (0040,0003) R 1 No/No No Matching is performed.
Procedure Step Start
Time
>Scheduled (0040,0004) O 3 No/No Not Requested
Procedure Step End
Date
>Scheduled (0040,0005) O 3 No/No Not Requested
Procedure Step End
Time
>Modality (0008,0060) R 1 No/No Matching is supported and is configured by the
user interface. This value could be set to
RF/XA/CR/OT/ALL. ALL means any
modality.
Note 1: Send the field as empty to request
universal matching when "ALL" is selected
from the UI.
Note 2:User could only select RF,XA,CR or
OT as Modality Type during image archiving.

88
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Scheduled (0040,0006) R 2* Yes/Yes Matching is performed. The data entered on the
Performing user interface is passed to the server as it is.
Physician's Name Single matching is supported for this data
element.
This value is mapped into (0008, 1050) in the
instance.
This value is mapped into (0008, 1050) in the
MPPS.
Supports 5 different components delimited by
^. Supports a maximum length of 64 bytes
including the delimiter.
>Scheduled (0040,0007) O 1C * Yes/Yes No Matching is performed.
Procedure Step This value is mapped into (0040,0275)->(0040,
Description 0007) & Protocol Name(0018,1030) & Series
Description(0008,103E) in the image instance.
This value is mapped into Series
Description(0008,103E) & EV(125203, DCM,
Acquisition Protocol) in the RDSR instance.
This value is mapped into (0040,0270)->(0040,
0007) & Protocol Name(0018,1030) & Series
Description(0008,103E) in the MPPS.
>Scheduled Station (0040,0010) O 2 No/No No Matching is performed.
Name
>Scheduled (0040,0011) O 2 No/No No Matching is performed.
Procedure Step
Location
>Scheduled Protocol (0040,0008) O 1C Yes/Yes No Matching is performed.
Code Sequence This value is mapped into (0040,0275)-> (0040,
0008) in the instance.
This value is mapped into (0040,0270)-> (0040,
0008) in the MPPS.
>>Code Value (0008,0100) O 1 Yes/Yes No Matching is performed.
>>Coding Scheme (0008,0102) O 1 Yes/Yes No Matching is performed.
Designator
>>Coding Scheme (0008,0103) O 3 Yes/Yes No Matching is performed.
Version
>>Code Meaning (0008,0104) O 3* Yes/Yes No Matching is performed.
>Pre-Medication (0040,0012) O 2C No/No No Matching is performed.
>Scheduled (0040,0009) O 1* Yes/Yes No Matching is performed.
Procedure Step ID This value is mapped into (0040,0275)-> (0040,
0009) in the instance.
This value is mapped into (0040,0270)-> (0040,
0009) in the MPPS.
>Requested Contrast (0032,1070) O 2C Yes/No No Matching is performed.
Agent This value is mapped into (0018, 0010) in the
instance.

89
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Scheduled (0040,0020) O 3 No/No Not Requested
Procedure Step Status
>Comments on the (0040,0400) O 3 No/No Not Requested
Scheduled Procedure
Step

Note: * in the Expected Return Key Type column indicates that this information is
displayed on screen, if available

6.4.1.2.1 Scheduled Station AE Title


Always empty in the request, and not used or set for matching.

6.4.2 Common Requested Procedure Entity Modules

6.4.2.1 Requested Procedure Module


TABLE 6-5
REQUESTED PROCEDURE MODULE ATTRIBUTES
Attribute Name Tag Expected Expected Mapped into Note
Matching Returned Instance /
Key Type Key Type MPPS
Requested Procedure (0040,1001) O 1* Yes/Yes Matching is performed. The data entered on the
ID user interface is passed to the server as it is.
Single matching is supported for this data
element.
This value is mapped into(0040,0275)-
>(0040,1001) in the image instance.
This value is mapped into (0040,a370)->(0040,
1001) in the RDSR.
This value is mapped into (0040,0270)->(0040,
1001) in the MPPS.
Requested Procedure (0032,1060) O 1C * Yes/Yes This value is mapped into (0040,0275)-
Description >(0032,1060) & Study Description(0008,1030)
in the image instance.
This value is mapped into (0040,A370)-
>(0032,1060) in the RDSR.
This value is mapped into (0040,0270)->(0032,
1060) & Study Description(0008,1030) in the
MPPS.
Requested Procedure (0032,1064) O 1C Yes/Yes No Matching is performed.
Code Sequence This value is mapped into (0032, 1064) in the
image instance.
This value is mapped into (0008, 1032) in the
MPPS.
>Code Value (0008,0100) O 1 Yes/Yes No Matching is performed.
>Coding Scheme (0008,0102) O 1 Yes/Yes No Matching is performed.
Designator

90
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Coding Scheme (0008,0103) O 3 Yes/Yes No Matching is performed.
Version
>Code Meaning (0008,0104) O 3 Yes/Yes No Matching is performed.
Study Instance UID (0020,000D) O 1 Yes/Yes The product will adopt the Study Instance UID
which is returned in the MWL response.
If there is no Study Instance UID returned from
MWL server, then new Study Instance UID will
be system created.
Study Date (0008,0020) O 3 No/No Not Requested
Study Time (0008,0030) O 3 No/No Not Requested
Referenced Study (0008,1110) O 2 Yes/Yes This value is mapped into (0008,1110) in the
Sequence image instance.
This value is mapped into (0008,1110) in the
RDSR.
This value is mapped into (0008, 1110) in the
MPPS.
>Referenced SOP (0008,1150) O 1C Yes/Yes
Class UID
>Referenced SOP (0008,1155) O 1C Yes/Yes
Instance UID
Requested Procedure (0040,1003) O 2 No/No No Matching is performed.
Priority
Patient Transport (0040,1004) O 2 No/No Not Requested
Arrangements
Requested Procedure (0040,1005) O 3 No/No Not Requested
Location
Confidentiality Code (0040,1008) O 3 No/No Not Requested
Reporting Priority (0040,1009) O 3 No/No Not Requested
Names of Intended (0040,1010) O 3 No/No No Matching is performed.
Recipients of Results
Reason for the (0040,1002) O 3 No/No No Matching is performed.
Requested Procedure
Requested Procedure (0040,1400) O 3 No/No Not Requested
Comments

Note: * in the Expected Return Key Type column indicates that this information is
displayed on screen, if available

6.4.2.1.1 Study Instance UID


The product will adopt the Study Instance UID which is returned in the MWL response.
If there is no Study Instance UID returned from MWL server, then new Study Instance
UID will be created.

91
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
6.4.3 Common Imaging Service Request Entity Modules

6.4.3.1 Imaging Service Request Module


TABLE 6-6
IMAGING SERVICE REQUEST MODULE ATTRIBUTES
Attribute Name Tag Expected Expected Mapped into Note
Matching Returned Instance /
Key Type Key Type MPPS
Accession Number (0008,0050) O 2* Yes/Yes Matching is performed. The data entered on the
user interface is passed to the server as it is.
Single matching is supported for this data
element.
Truncated to 16 bytes, end truncated
This value is mapped into (0008,0050) in the
instance or RDSR.
This value is mapped into (0008, 0050) in the
MPPS.
Requesting Physician (0032,1032) O 2 No/No No Matching is performed.
Referring Physician's (0008,0090) O 2* Yes/No No Matching is performed.
Name This value is mapped into (0008,0090) in the
instance or RDSR.
Requesting Service (0032,1033) O 3 No/No Not Requested
Reason for the (0040,2001) O 3 No/No Not Requested
Imaging Service
Request
Imaging Service (0040,2400) O 3 No/No Not Requested
Request Comments
Issue Date of Imaging (0040,2004) O 3 No/No Not Requested
Service Request
Issue Time of (0040,2005) O 3 No/No Not Requested
Imaging Service
Request
Placer Order Number (0040,2016) O 3 No/No Not Requested
/ Imaging Service
Request
Filler Order Number / (0040,2017) O 3 No/No Not Requested
Imaging Service
Request
Order entered by ... (0040,2008) O 3 No/No Not Requested
Order Enterers (0040,2009) O 3 No/No Not Requested
Location
Order Callback Phone (0040,2010) O 3 No/No Not Requested
Number

Note: * in the Expected Return Key Type column indicates that this information is
displayed on screen, if available

92
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
6.4.4 Common visit Entity Modules

6.4.4.1 Visit Identification


Not applicable

6.4.4.2 Visit Status


Not applicable

6.4.4.3 Visit Relationship


Not applicable

6.4.4.4 Visit Admission


Not applicable

6.4.5 Common Patient Entity Modules

6.4.5.1 Patient Relationship


Not applicable

6.4.5.2 Patient Identification


TABLE 6-7
PATIENT IDENTIFICATION MODULE ATTRIBUTES
Attribute Name Tag Expected Expected Mapped into Note
Matching Returned Instance /
Key Type Key Type MPPS
Patient's Name (0010,0010) R 1* Yes/Yes Matching is performed.
Single or Wildcard matching is supported for
this data element.
Supports 5 different components delimited by
^. Supports a maximum length of 64 bytes
including the delimiter.
Patient ID (0010,0020) R 1* Yes/Yes Matching is performed.
Single matching is supported for this data
element.
Truncated to 64 bytes, end truncated
Issuer of Patient ID (0010,0021) O 3 Yes/Yes No Matching is performed.
Other Patient IDs (0010,1000) O 3 No/No Not Requested
Other Patient Names (0010,1001) O 3 No/No Not Requested
Patient's Birth Name (0010,1005) O 3 No/No Not Requested
Patient's Mother's (0010,1060) O 3 No/No Not Requested
Birth Name
Medical Record (0010,1090) O 3 No/No Not Requested
Locator

93
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Note: * in the Expected Return Key Type column indicates that this information is
displayed on screen, if available

6.4.5.3 Patient Demographic


TABLE 6-8
PATIENT DEMOGRAPHIC MODULE ATTRIBUTES
Attribute Name Tag Expected Expected Mapped into Note
Matching Returned Instance /
Key Type Key Type MPPS
Patients Birth Date (0010,0030) O 2* Yes/Yes No Matching is performed..
Truncated to 64 bytes, end truncated.
Patient's Sex (0010,0040) O 2* Yes/Yes No Matching is performed.
Truncated to 2 bytes, end truncated
Patient's Weight (0010,1030) O 2* Yes/No No Matching is performed.
Only mapped into image instance.
Truncated to 16 bytes, end truncated
Confidentiality (0040,3001) O 2 No/No No Matching is performed.
constraint on patient
data
Patients Size (0010,1020) O 3* Yes/No No Matching is performed.
Only mapped into image instance.
Truncated to 16 bytes, end truncated
Patients Address (0010,1040) O 3 No/No Not Requested
Patients Telephone (0010,2154) O 3 No/No Not Requested
Numbers
Patient's Age (0010,1010) O 3 No/No Not Requested
Occupation (0010,2180) O 3 No/No Not Requested
Patients Birth Time (0010,0032) O 3 No/No Not Requested
Patient's Insurance (0010,0050) O 3 No/No Not Requested
Plan Code Sequence
> Code Sequence
Macro
Patients Primary (0010,0101) O 3 No/No Not Requested
Language Code
Sequence
> Code Sequence
Macro
> Patients Primary (0010,0102) O 3 No/No Not Requested
Language Code
Modifier Sequence
>> Code Sequence
Macro

94
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Military Rank (0010,1080) O 3 No/No Not Requested
Branch of Service (0010,1081) O 3 No/No Not Requested
Country of Residence (0010,2150) O 3 No/No Not Requested
Region of Residence (0010,2152) O 3 No/No Not Requested
Patients Telephone (0010,2154) O 3 No/No Not Requested
Numbers
Ethnic Group (0010,2160) O 3 No/No Not Requested
Patient's Religious (0010,21F0) O 3 No/No Not Requested
Preference
Patient Comments (0010,4000) O 3* Yes/No No Matching is performed.
Mapped into image and RDSR instance .
Truncated to 10240 bytes, end truncated

Note: * in the Expected Return Key Type column indicates that this information is
displayed on screen, if available

6.4.5.4 Patient Medical


Not applicable

95
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

7. MODALITY PERFORMED PROCEDURE STEP


IMPLEMENTATION

7.1 INTRODUCTION

This section specifies the use of the DICOM Modality Performed Procedure Step
information to be communicated to the Hospital/Radiology information system.

This feature works in conjunction with DICOM Modality Worklist feature, if installed.
However the conformance of this feature is independent of Modality Worklist feature.
For information on conformance of Modality Worklist feature to DICOM standard please
refer to the appropriate section in this document.

7.2 RELATIONSHIP BETWEEN SCHEDULED AND PERFORMED PROCEDURE STEPS

Brivo OEC 715/785/865 supports following relationship between Scheduled Procedure


Step and PPS:

1. Supports a one-to-one relationship between Scheduled Procedure Step and PPS

2. Supports one-to-multiple relationship: Append Case with new PPS based on Original
SPS

3. Supports Unscheduled Case or Acquisition without MWL Data

7.3 MODALITY PERFORMED PROCEDURE STEP MODULE TABLE

See DICOM PS 3.3 and PS 3.4 for a complete definition of the entities, modules, and
attributes.

TABLE 7-1
MODALITY PERFORMED PROCEDURE STEP MODULES
Module Name Reference
SOP Common 7.4.1
Performed Procedure Step Relationship 7.4.2
Performed Procedure Step Information 7.4.3
Image Acquisition Results 7.4.4
Radiation Dose 7.4.5
Billing and Material Management Codes 7.4.6

7.4 MODALITY PERFORMED PROCEDURE STEP MODULE DEFINITIONS


Please refer to DICOM Standard PS 3.3. (Information Object Definitions) for a
description of each of the attributes contained within the Modality Performed Procedure
Step Information Object Definition.

96
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
7.4.1 SOP Common Module
TABLE 7-2
SOP COMMON MODULE ATTRIBUTES
Attribute Name Tag Type for Type for Use
SCU SCU
N- N-SET
CREATE
Specific Character Set (0008,0005) 1C 1C ISO_IR 100

7.4.2 Performed Procedure Step Relationship Module


TABLE 7-3
PERFORMED PROCEDURE STEP RELATIONSHIP MODULE ATTRIBUTES
Attribute Name Tag Type for SCU - N-CREATE
Acquisition Acquisition
without MWL Entry with MWL Entry
Scheduled Step Attributes (0040,0270) 1, Only 1 item 1, Only 1 item
Sequence
>Study Instance UID (0020,000D) 1, System generated. 1,Got from MWL.
>Referenced Study (0008,1110) 2, Empty. 2, Got from MWL,
Sequence Only 1 item.
>>Referenced SOP Class (0008,1150) 1, Not sent 1, Got from MWL.
UID
>>Referenced SOP Instance (0008,1155) 1, Not sent 1, Got from MWL.
UID
>Accession Number (0008,0050) 2, User input. If user does 2, Got from MWL.
not provide input values,
then empty
>Placer Order (0040,2016) 3, Not sent 3, Not sent
Number/Imaging Service
Request
>Filler Order (0040,2017) 3, Not sent 3, Not sent
Number/Imaging Service
Request
>Requested Procedure ID (0040,1001) 2, User input. If user does 2, Got from MWL.
not provide input values,
then empty
>Requested Procedure Code (0032,1064) 3, Empty 3, Not sent
Sequence
>>Code Value (0008,0100) 1, Not sent 1, Not sent
>>Coding Scheme (0008,0102) 1, Not sent 1, Not sent
Designator
>>Code Meaning (0008,0104) 1, Not sent 1, Not sent
>Requested Procedure (0032,1060) 2, User input. If user does 2, Got from MWL.
Description not provide input values,
then empty

97
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Scheduled Procedure Step (0040,0009) 2, User input. If user does 2, Got from MWL.
ID not provide input values,
then empty
>Scheduled Procedure Step (0040,0007) 2, User input. If user does 2, Got from MWL.
Description not provide input values,
then empty
>Scheduled Protocol Code (0040,0008) 2, empty 2, Got from MWL
Sequence
>>Code Value (0008,0100) 1, Not sent 1, Got from MWL
>>Coding Scheme (0008,0102) 1, Not sent 1, Got from MWL
Designator
>>Coding Scheme Version (0008,0103) 3, Not sent 3, Got from MWL
>>Code Meaning (0008,0104) 3, Not sent 3, Got from MWL
Patient's Name (0010,0010) 2, User input. If user does 2, Got from MWL.
not provide input values,
then UNNAMED xxxxx
should be assigned to Last
Name (xxxxx is a 5-
digital number)
Patient ID (0010,0020) 2, User input. If user does 2, Got from MWL.
not provide input values,
then empty
Issuer of Patient ID (0010,0021) 3, empty 3, Got from MWL.
Patient's Birth Date (0010,0030) 2, User input. If user does 2, Got from MWL.
not provide input values,
then empty
Patient's Sex (0010,0040) 2, User input. If user does 2, Got from MWL.
not provide input values,
then empty
Referenced Patient Sequence (0008,1120) 2, empty. 2, empty.
>Referenced SOP Class UID (0008,1150) 1, Not sent 1, Not sent
>Referenced SOP Instance (0008,1155) 1, Not sent 1, Not sent
UID
Admission ID (0038,0010) 3, Not sent 3, Not sent
Issuer of Admission ID (0038,0011) 3, Not sent 3, Not sent
Service Episode ID (0038,0060) 3, Not sent 3, Not sent
Issuer of Service Episode ID (0038,0061) 3, Not sent 3, Not sent
Service Episode Description (0038,0062) 3, Not sent 3, Not sent

98
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
7.4.3 Performed Procedure Step Information Module
TABLE 7-4
PERFORMED PROCEDURE STEP INFORMATION MODULE ATTRIBUTES
Attribute Name Tag Type for Type for Use
SCU SCU
N- N-SET
CREATE
Performed Procedure Step ID (0040,0253) 1 - If Scheduled Procedure Step ID is empty then
this value is current timestamp, else this value is
Scheduled Procedure Step ID.
Performed Station AE Title (0040,0241) 1 - Got from configure file.
Performed Station Name (0040,0242) 2 - Got from configure file.
Performed Location (0040,0243) 2 - Got from configure file.
The value is Hospital Name.
Performed Procedure Step (0040,0244) 1 - Same as exam start date
Start Date
Performed Procedure Step (0040,0245) 1 - Same as exam start time
Start Time
Performed Procedure Step (0040,0252) 1 3 When PPS start ( N-CREATE) message is sent,
Status this element will have the value IN
PROGRESS .
When PPS end ( N-SET) message is sent, this
element will have either COMPLETED or
DISCONTINUED based on user selection.
Performed Procedure Step (0040,0254) 2 3 This value is Scheduled Procedure Step
Description Description
Performed Procedure Type (0040,0255) 2 3 From "Requested Procedure Description" on
Description screen which is come from MWL.
Procedure Code Sequence (0008,1032) 2 3 Always set as Empty for local created exam.
When using MWL, this value is Requested
Procedure Code Sequence.
>Code Value (0008,0100) 1 1 Only when using MWL sent.
>Coding Scheme Designator (0008,0102) 1 1 Only when using MWL sent.
>Coding Scheme Version (0008,0103) 3 3 Only when using MWL sent.
>Code Meaning (0008,0104) 3 3 Only when using MWL sent.
Performed Procedure Step (0040,0250) 2 3 Date when PPS end got installed
End Date
Performed Procedure Step (0040,0251) 2 3 Time when PPS end got installed
End Time
Comments on the Performed (0040,0280) 3 3 Not sent
Procedure Step
Performed Procedure Step (0040,0281) 3 3 Used when exam is DISCONTINUED.
Discontinuation Reason Code Provide a way let user to select one
Sequence
>Code Value (0008,0100) 1 1 Follow CID 9301
>Coding Scheme Designator (0008,0102) 1 1 Follow CID 9301

99
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Coding Scheme Version (0008,0103) 3 3 Follow CID 9301
>Code Meaning (0008,0104) 3 3 Follow CID 9301

7.4.4 Image Acquisition Results Module


TABLE 7-5
IMAGE ACQUISITION RESULTS MODULE ATTRIBUTES
Attribute Name Tag Type for Type for Use
SCU SCU
N- N-SET
CREATE
Modality (0008,0060) 1 - When user create a local scheduled exam, then
this value will be set as "RF" as default.
Got from MWL server.
Study ID (0020,0010) 2 - System creates and Read Only on screen.
Performed Protocol Code (0040,0260) 2 3 Always set as Empty.
Sequence
>Code Value (0008,0100) 1 1 Not sent
>Coding Scheme Designator (0008,0102) 1 1 Not sent
>Coding Scheme Version (0008,0103) 3 3 Not sent
>Code Meaning (0008,0104) 3 3 Not sent
Performed Series Sequence (0040,0340) 2 3 N-Set: Will be sent with referenced instances.
>Performing Physician's (0008,1050) 2 2 Got from MWL server.
Name User input.
>Protocol Name (0018,1030) 1 1 When using MWL, if Scheduled Protocol
Sequence Code Meaning is not empty then this
value is Scheduled Protocol Sequence Code
Meaning, else this value is Scheduled Procedure
Step Description.
It will be set as Unknown Protocol for local
created exam.
>Operators Name (0008,1070) 2 2 Always set as Empty.
>Series Instance UID (0020,000E) 1 1 System creates.
>Series Description (0008,103E) 2 2 This value is Protocol Name.
>Retrieve AE Title (0008,0054) 2 2 Always set as Empty.
> Archive Requested (0040,A494) 3 3 Not sent
>Referenced Image Sequence (0008,1140) 2 2 Will be sent with referenced images. One item
for each image created within the series
>>Referenced SOP Class UID (0008,1150) 1 1 Referenced image SOP Class UID.
>>Referenced SOP Instance (0008,1155) 1 1 Referenced image SOP Instance UID.
UID
>Referenced Non-Image (0040,0220) 2 2 Will be sent with referenced instance
Composite SOP Instance
Sequence

100
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>>Referenced SOP Class UID (0008,1150) 1 1 Referenced instance SOP Class UID.
>>Referenced SOP Instance (0008,1155) 1 1 Referenced instance SOP instance UID.
UID

7.4.5 Radiation Dose Module


TABLE 7-6
RADIATION DOSE MODULE ATTRIBUTES
Attribute Name Tag Type for Type for Use
SCU SCU
N- N-SET
CREATE
Anatomic Structure, Space or (0008,2229) 3 3 Always set as empty.
Region Sequence
Total Time of Fluoroscopy (0040,0300) 3 3 Got value from DB.
Total Number of Exposures (0040,0301) 3 3 Got value from DB.
Distance Source to Detector (0018,1110) 3 3 Got value from configure file.
(SID)
Distance Source to Entrance (0040,0306) 3 3 Not sent
Entrance Dose (0040,0302) 3 3 Got value from DB.
Entrance Dose in mGy (0040,8302) 3 3 The same with (0040,0302), but unit is mGy.
Exposed Area (0040,0303) 3 3 Not sent
Image Area Dose Product (0018,115E) 3 3 Got value from DB.
Comments on Radiation Dose (0040,0310) 3 3 Not Used
Exposure Dose Sequence (0040,030E) 3 3 Zero or more items shall be included in this
sequence.
The items count will be the shot count.
>Radiation Mode (0018,115A) 3 3 Radiation Setting:
Continuous: Fluoro/Digital Spot/HLF/Film
Pulsed: All Pulse xxx
>KVp (0018,0060) 3 3 Got value from DB.
>X-ray Tube Current in A (0018,8151) 3 3 Got value from DB.
>Exposure Time (0018,1150) 3 3 Got value from DB.
>Filter Type (0018,1160) 3 3 Not sent
>Filter Material (0018,7050) 3 3 Not sent

101
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
7.4.6 Billing and Material Management Codes Module
TABLE 7-7
BILLING AND MATERIAL MANAGEMENT CODES MODULE ATTRIBUTES
Attribute Name Tag Type for Type for Use
SCU SCU
N- N-SET
CREATE
Billing Procedure Step (0040,0320) 3 3 Not sent
Sequence
> Code Sequence Macro (0020,0010) 3 3
Film Consumption Sequence (0040,0321) 3 3 Not sent
>Number of Films (2100,0170) 3 3 Not sent
>Medium Type (2000,0030) 3 3 Not sent
>Film Size ID (2010,0050) 3 3 Not sent
Billing Supplies and Devices (0040,0324) 3 3 Not sent
Sequence
>Billing Item Sequence (0040,0296) 3 3 Not sent
>> Code Sequence Macro (0040,0303) 3 3 Not sent
>Quantity Sequence (0040,0293) 3 3 Not sent
>>Quantity (0040,0294) 3 3 Not sent
>>Measuring Units Sequence (0040,0295) 3 3 Not sent
>>> Code Sequence Macro (0018,115A) 3 3 Not sent

7.5 STANDARD EXTENDED AND PRIVATE DATA ATTRIBUTES


The Product supports the Standard and Private Attributes defined in the following
sections in Standard Extended MPPS Instances as Type 3 data elements.

7.5.1 Standard Attributes

The Product supports the following attributes, not specified in the MPPS IOD, in SOP
Instances as Type 3 data elements.

TABLE 7-8
STANDARD EXTENDED ATTRIBUTES
Attribute Name Tag Use
N/A N/A N/A

7.6 STANDARD EXTENDED AND PRIVATE CONTEXT GROUPS


The Product supports coded terminology using Standard Extended, Private, and
Configurable Context Groups defined in the following sections.

7.6.1 Standard Extended Context Groups


The Product supports the following extensions to standard Context Groups for SOP
Instances created by this product. Extensions are indicated by bold text.

102
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
7.6.1.1 Context ID 9301 Modality PPS Discontinuation Reasons
Context ID 9301
Modality PPS Discontinuation Reasons
Type: Extensible Version: 20090616
Coding Scheme Code Value Code Meaning
Designator (0008,0100) (0008,0104)
(0008,0102)
DCM 110500 Doctor cancelled procedure
DCM 110501 Equipment failure
DCM 110502 Incorrect procedure ordered
DCM 110503 Patient allergic to media/contrast
DCM 110504 Patient died
DCM 110505 Patient refused to continue procedure
DCM 110506 Patient taken for treatment or surgery
DCM 110507 Patient did not arrive
DCM 110508 Patient pregnant
DCM 110509 Change of procedure for correct charging
DCM 110510 Duplicate order
DCM 110511 Nursing unit cancel
DCM 110512 Incorrect side ordered
DCM 110513 Discontinued for unspecified reason
DCM 110514 Incorrect worklist entry selected
DCM 110515 Patient condition prevented continuing
DCM 110516 Equipment change

103
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

8. STORAGE COMMITMENT PUSH MODEL IMPLEMENTATION

8.1 STORAGE COMMITMENT PUSH MODEL INFORMATION OBJECT DEFINITION


Please refer to DICOM Part 3 (Information Object Definitions) for a description of each
of the attributes contained within the Storage Commitment Information Object.

The Storage Commitment Information Object is used both for N-ACTION Storage
Commitment Requests by the SCU and N-EVENT-REPORT Storage Commitment
Notifications by the SCP.

8.1.1 STORAGE COMMITMENT MODULE FOR N-ACTION


TABLE 8-1
STORAGE COMMITMENT MODULE FOR N-ACTION
Attribute Name Tag SCU Use SCP Use
Transaction UID (0008,1195) Created by equipment, using Not used
prefix:
715: 1.2.840.113619.8.320.
785: 1.2.840.113619.8.321.
865: 1.2.840.113619.8.333.
+MAC+transactionIndicator(7
0)+TimeStamp_Current
MAC: Number string of
device MAC address.
Storage Media File-Set ID (0088,0130) Not used Not used
Storage Media File-Set UID (0088,0140) Not used Not used
Referenced SOP Sequence (0008,1199) Images/RDSR instances Not used
might be included here.
>Referenced SOP Class UID (0008,1150) Not used
>Referenced SOP Instance UID (0008,1155) Not used
>Storage Media File-Set ID (0088,0130) Not used Not used
>Storage Media File-Set UID (0088,0140) Not used Not used

8.1.2 STORAGE COMMITMENT MODULE FOR N-EVENT-REPORT


TABLE 8-2
STORAGE COMMITMENT MODULE FOR N-EVENT-REPORT
Attribute Name Tag SCU Use
Transaction UID (0008,1195) Value received from SCP, if the value is empty,then return failed.
Retrieve AE Title (0008,0054) Not used
Storage Media File-Set ID (0088,0130) Not used

104
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Storage Media File-Set UID (0088,0140) Not used
Referenced SOP Sequence (0008,1199) Success of storage commitment is reported to the user.
This sequence will be used to mark storage status on screen.
All the successed instance will marked as Stored in the system.
>Referenced SOP Class UID (0008,1150)
>Referenced SOP Instance UID (0008,1155)
>Retrieve AE Title (0008,0054) Not Used
>Storage Media File-Set ID (0088,0130) Not used
>Storage Media File-Set UID (0088,0140) Not used
Failed SOP Sequence (0008,1198) All the failed instance will not be marked as Stored in the
system.
>Referenced SOP Class UID (0008,1150)
>Referenced SOP Instance UID (0008,1155)
>Failure Reason (0008,1197) See Section 8.1.2.1 for the list of processed values.

8.1.2.1 Processing of Failure Reason when received in a N-Event-Report

When receiving an N-Event-Report request with an Event Type ID equal to 2, meaning


that Storage Commitment is complete, but failure exists, following is the set of value that
this Storage Commitment SCU AE is able to process:

Failure Meaning Application Behavior When Receiving Reason


Reason Code
0110H Processing failure Failure and reason is logged. A general failure is
0112H No such object instance reported to the user.

0213H Resource limitation


0122H Referenced SOP Class not supported
0119H Class / Instance conflict
0131H Duplicate transaction UID
* Other Failure Reason code values

105
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

9. BASIC DIRECTORY INFORMATION OBJECT


IMPLEMENTATION

9.1 IOD MODULE TABLE

Table 9-1 identifies the defined modules within the entities which comprise the Basic
Directory IOD. Modules are identified by Module Name.

See DICOM Part 3 for a complete definition of the entities, modules, and attributes.

TABLE 9-1
BASIC DIRECTORY IOD MODULES
Entity Name Module Name Reference
File Set Identification File Set Identification 9.2.1
Directory Information Directory Information 9.2.2

The FSC of this implementation creates a Directory Information Module for new media
and creates XA IOD.

9.2 INFORMATION MODULE DEFINITIONS

Please refer to DICOM Standard Part 3 (Information Object Definitions) for a description
of each of the entities and modules contained within the Basic Directory Information
Object.

The following modules are included to convey Enumerated Values, Defined Terms, and
Optional Attributes supported. Type 1 & Type 2 Attributes are also included for
completeness and to define what values they may take and where these values are
obtained from. It should be noted that they are the same ones as defined in the DICOM
Standard Part 3 (Information Object Definitions). Also note that Attributes not present in
tables are not supported.

106
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
9.2.1 File Set identification Module
TABLE 9-2
FILE-SET IDENTIFICATION MODULE
Attribute Name Tag Type Attribute Description
File-set ID (0004,1130) 2 Current DateTime stamp, format
like:YYYYMMDDHHMMSS
YYYY- year, MM- month, DD- day.
HH Hour 00-23, MM minute 00-59, SS
second 00-59
File-set Descriptor File ID (0004,1141) 3 Not Used
Specific Character Set of File-set (0004,1142) 1C Not Used
Descriptor File

9.2.2 Directory Information Module


TABLE 9-3
DIRECTORY INFORMATION MODULE
Attribute Name Tag Type Attribute Description
Offset of the First Directory Record of the (0004,1200) 1 The number of bytes from the beginning of the
Root Directory Entity file to the first byte of the first Directory
Record. When the Root Directory entity
contains no directory record, this offset is set to
00000000H
Offset of the Last Directory Record of the (0004,1202) 1 The number of bytes from the beginning of the
Root Directory Entity file to the first byte of the last Directory
Record. When the Root Directory entity
contains no directory record, this offset is set to
00000000H
File-set Consistency Flag (0004,1212) 1 FSC writes 0000H: no known inconsistencies.
Directory Record Sequence (0004,1220) 2 FSC creates items in this sequence (Directory
Records).
>Offset of the Next Directory Record (0004,1400) 1C
>Record In-use Flag (0004,1410) 1C Set the value to 0xFFFF for record is in use
>Offset of Referenced Lower-Level (0004,1420) 1C
Directory Entity
>Directory Record Type (0004,1430) 1C Enumerated Values are created by an
FSC/FSU:
PATIENT
STUDY
SERIES
IMAGE
SR DOCUMENT
>Private Record UID (0004,1432) 1C Not Used
>Referenced File ID (0004,1500) 1C Filename is generated to be unique on the media
like: DCM<index>\IMG<number of image>
Field is included if Directory Record Type is not
PATIENT/STUDY/SERIES.

107
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Referenced SOP Class UID in File (0004,1510) 1C Field is included if Directory Record Type is
not PATIENT/STUDY/SERIES.
Unique ID for the SOP Class of the Instance
stored in the referenced File.
>Referenced SOP Instance UID in File (0004,1511) 1C Field is included if Directory Record Type is
not PATIENT/STUDY/SERIES.
Unique Identifier for the SOP Instance stored in
the referenced file.
>Referenced Transfer Syntax UID in File (0004,1512) 1C Field is included if Directory Record Type is
not PATIENT/STUDY/SERIES.
Unique Identifier for the Transfer Syntax used to
encode the Instance stored in the referenced file.
> Referenced Related General SOP Class (0004,151A) 1C Not Used
UID in File
>Record Selection Keys See 9.2.3

9.2.3 Definition of Specific Directory Records

9.2.3.1 Patient Directory Record Definition


TABLE 9-4
PATIENT KEYS
Key Tag Type Attribute Description
Specific Character Set (0008,0005) 1C ISO_IR 100
Patient's Name (0010,0010) 2 FSC/FSU will fill this attribute with the value
in the referenced composite object instance.
Patient ID (0010,0020) 1 FSC/FSU will fill this attribute with the value
in the referenced composite object instance.
FSC/FSU will fill this mandatory attribute with
a value DCMTKPAT + 6 digits if it is empty
in the referenced composite object instance.

9.2.3.2 Study Directory Record Definition


TABLE 9-5
STUDY KEYS
Key Tag Type Attribute Description
Specific Character Set (0008,0005) 1C ISO_IR 100
Study Date (0008,0020) 1 FSC/FSU will fill this mandatory attribute with
the value in the referenced composite object
instance.
This value is mandatory for Brivo OEC
715/785/865, cant be empty.
Study Time (0008,0030) 1 FSC/FSU will fill this mandatory attribute with
the value in the referenced composite object
instance.
This value is mandatory for Brivo OEC
715/785/865, cant be empty.

108
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Study Description (0008,1030) 2 FSC/FSU will fill this attribute with the value
in the referenced composite object instance.
Study Instance UID (0020,000D) 1C FSC/FSU will fill this mandatory attribute with
the value in the referenced composite object
instance.
This value is mandatory for Brivo OEC
715/785/865, cant be empty.
Study ID (0020,0010) 1 FSC/FSU will fill this mandatory attribute with
the value in the referenced composite object
instance.
FSC/FSU will fill this mandatory attribute with
a value DCMTKSTUDY+ 6 digits if it is
empty in the referenced composite object
instance.
Accession Number (0008,0050) 2 FSC/FSU will fill this attribute using the value
in the referenced composite object instance.

9.2.3.3 Series Directory Record Definition


TABLE 9-6
SERIES KEYS
Key Tag Type Attribute Description
Specific Character Set (0008,0005) 1C ISO_IR 100
Modality (0008,0060) 1 This will be consistent with Image Content
Series Instance UID (0020,000E) 1 FSC/FSU will fill this mandatory attribute with
the value in the referenced composite object
instance.
This value is mandatory for Brivo OEC
715/785/865, cant be empty.
Series Number (0020,0011) 1 FSC/FSU will fill this mandatory attribute with
the value in the referenced composite object
instance.
FSC/FSU will fill this mandatory attribute with
a value start from 1 (+1 for each series) if it is
empty in the referenced composite object
instance.
Icon Image Sequence (0088,0200) 3 Not Used
>Include Image Pixel Macro See 9.2.3.3.1

9.2.3.3.1 Icon Image Key Definition


Not applicable

109
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
9.2.3.4 Image Directory Record Definition
TABLE 9-7
IMAGE KEYS
Key Tag Type Attribute Description
Specific Character Set (0008,0005) 1C ISO_IR 100
Image Type (0008,0008) 3 FSC/FSU will fill this mandatory attribute with
the value in the referenced composite object
instance.
Instance Number (0020,0013) 1 FSC/FSU will fill this mandatory attribute with
the value in the referenced composite object
instance.
Icon Image Sequence (0088,0200) 3 FSC/FSU will create Icon Images at the image
level.
>Include Image Pixel Macro See section 9.2.3.4.1

Note: Please also refer to Section 3.2.1.2.1.1.1 for information about Additional
Keys sent in DICOMDIR.

9.2.3.4.1 Icon Image Key Definition

The FSC creates the icon image sequence from the referenced composite object
instance. The FSR displays image icons to the user without reading the actual image
pixel from the media.
Samples Per pixel (0028,0002) is set to 1.
Photometric Interpretations (0028,0004) is set to MONOCHROME2.
Row (0028,0010) and Column (0028,0011) are set to 128,128.
Bits Allocated (0028,0100) and Bits Stored (0028,0101) are created and set to 8.
High Bit (0028,0102) is created and set to 7.
Pixel Representation (0028,0103) is created and set to 0.
Pixel Data (7EF0,0010) is set with the 128x128 icon image data.

110
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
9.2.3.5 SR Document Directory Record Definition
TABLE 9-8
SR DOCUMENT KEYS
Key Tag Type Attribute Description
Specific Character Set (0008,0005) 1C ISO_IR 100
Instance Number (0020,0013) 1 FSC/FSU will fill this mandatory attribute with
the value in the referenced composite object
instance.
Completion Flag (0040,A491) 1 "COMPLETE"
Verification Flag (0040,A493) 1 "Unverified".
Content Date (0008,0023) 1 First Shot Date
Content Time (0008,0033) 1 First Shot Time
Verification DateTime (0040,A030) 1C Not Used
Concept Name Code Sequence (0040,A043) 1 One item shall be included in this sequence.
>Code Value (0008,0100) 1 "113701"
>Coding Scheme Designator (0008,0102) 1 "DCM"
>Code Meaning (0008,0104) 1 "X-Ray Radiation Dose Report"
Content Sequence (0040,A730) 1C
>Relationship Type (0040,A010) 1 All, and only, Content Items with the HAS
CONCEPT MOD relationship from the root
Content Item shall be included in this
Sequence.
>Include Document Content Macro

9.3 PRIVATE DATA DICTIONARY


Not applicable.

111
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

10. PRINT MANAGEMENT IMPLEMENTATION

10.1 INTRODUCTION
This section of the DICOM Conformance Statement specifies the implementation for the
specific SOP Classes supported in the Basic Grayscale Print Management Meta SOP
Class, the attributes supported for both IODs and services, and the valid range of values
for mandatory and optional attributes.

10.2 BASIC FILM SESSION SOP CLASS

10.2.1 Basic Film Session N-Create Attributes


This table lists the attributes that are sent in the Basic Film Session N-Create Request:

Attribute name Tag Use


Specific Character Set (0008,0005) Not Used
Number of Copies (2000,0010) Range of this value is 1-10, default is 1.
Print Priority (2000,0020) Enumerated Terms can be sent:
HIGH or MED or LOW
Medium Type (2000,0030) Enumerated Terms can be sent:
PAPER, CLEAR FILM, BLUE FILM,
MAMMO CLEAR FILM, MAMMO BLUE FILM
Film Destination (2000,0040) Enumerated Terms can be sent:
MAGAZINE or PROCESSOR
Film Session Label (2000,0050) Set to Brivo OEC Print SCU
Memory Allocation (2000,0060) Not Used
Owner ID (2100,0160) Set to Brivo OEC

10.2.2 Basic Film Session N-Delete

The N-DELETE is used to delete the complete Basic Film Session SOP Instance
hierarchy. As a result, all references to Image SOP Instances within the film session are
deleted.

10.3 BASIC FILM BOX SOP CLASS

10.3.1 Basic Film Box N-Create Attributes


This table lists the attributes that are sent to the SCP in the Basic Film Box N-Create
Request, and that are received in the Basic Film Box N-Create Response from the SCP

112
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Attribute Name Tag Use
Image Display Format (2010,0010) User configurable.
User values =:
STANDARD\1,1
STANDARD\1,2
STANDARD\2,1
STANDARD\2,2
STANDARD\2,3
STANDARD\3,3
STANDARD\3,4
STANDARD\4,4
STANDARD\4,5
STANDARD\5,4
Default value: First selection when declaring printer.
Referenced Film Session Sequence (2010,0500) Used
>Referenced SOP Class UID (0008,1150) Set to 1.2.840.10008.5.1.1.1
>Referenced SOP Instance UID (0008,1155) This value should be returned from SCP in N-CREATE-
RSP Basic Film Session SOP Class; Then this value is
sent to SCP in N-CREATE-RQ Basic Film Box SOP
Class requesting for Basic Grayscale Image Box SOP
Class Instance UID which contained in Referenced Image
Box Sequence from SCP.
Referenced Image Box Sequence (2010,0510) Used
>Referenced SOP Class UID (0008,1150) Set to 1.2.840.10008.5.1.1.4
>Referenced SOP Instance UID (0008,1155) Received in the N-Create Response from SCP, this value
should be sent to SCP in N-SET-RQ Basic Grayscale
Image Box SOP Class
Referenced Basic Annotation Box (2010,0520) Not Used
Sequence
>Referenced SOP Class UID (0008,1150) Not Used
>Referenced SOP Instance UID (0008,1155) Not Used
Film Orientation (2010,0040) Enumerated Terms can be sent:
PORTRAITor LANDSCAPE
Film Size ID (2010,0050) Enumerated Terms can be sent:
8INX10IN
10INX12IN
10INX14IN
11INX14IN
14INX14IN
14INX17IN
24CMX24CM
24CMX30CM
Magnification Type (2010,0060) Set to BILINEAR
Max Density (2010,0130) User configurable. The default is 300
Configuration Information (2010,0150) User configurable. This value specifies the DICOM

113
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
printer parameters. It is defined in the DICOM Printer's
Conformance Statement file.
Referenced Presentation LUT Sequence (2050,0500) Not Used
>Referenced SOP Class UID (0008,1150) Not Used
>Referenced SOP Instance UID (0008,1155) Not Used
Annotation Display Format ID (2010,0030) Not Used
Smoothing Type (2010,0080) Not Used
Border Density (2010,0100) Enumerated Terms can be sent:
BLACK or WHITE
Empty Image Density (2010,0110) Enumerated Terms can be sent:
BLACK or WHITE
Min Density (2010,0120) User configurable. The default is 20
Trim (2010,0140) Set to NO
Illumination (2010,015E) Not Used
Reflected Ambient Light (2010,0160) Not Used
Requested Resolution ID (2020,0050) Not Used
ICC Profile (0028,2000) Not Used

10.3.2 Basic Film Box N-Action Attributes


Following are the Action Reply arguments that are supported if present in the N-Action
response of the Basic Film Box SOP Class

Action Type Name Action Attribute Tag Usage


Type ID SCU

Print 1 Referenced Print Job (2100,0500) Not Used


Sequence
>Referenced SOP (0008,1150) Not Used
Class UID
>Referenced SOP (0008,1155) Not Used
Instance UID
>Print Job ID (2100,0010) Not Used

10.3.3 Basic Film Box N-Delete


The N-DELETE is used to delete the last created Basic Film Box SOP Instance
hierarchy. As a result all the information describing the last film is deleted.

10.4 BASIC GRAYSCALE IMAGE BOX SOP CLASS

10.4.1 Basic Grayscale Image Box Pixel N-Set Attributes


This table lists the attributes that are sent in the Basic Grayscale Image Box N-Set
Request:

114
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Attribute Name Tag Use
Image Position (2020,0010) Based on Image Display Format (2010,0010), range of
values sent is [1 ~ 20].
Basic Grayscale Image Sequence (2020,0110) A sequence which provides the content of the grayscale
image pixel data to be printed
>Samples Per Pixel (0028,0002) 1
>Photometric Interpretation (0028,0004) MONOCHROME2
>Rows (0028,0010) 1000
>Columns (0028,0011) 1000
>Pixel Aspect Ratio (0028,0034) Not Used
>Bits Allocated (0028,0100) 16
>Bits Stored (0028,0101) 12
>High Bit (0028,0102) 11
>Pixel Representation (0028,0103) 0
>Pixel Data (7FE0,0010) Image Pixels
Polarity (2020,0020) Not Used
Magnification Type (2010,0060) Not Used
Smoothing Type (2010,0080) Not Used
Min Density (2010,0120) Not Used
Max Density (2010,0130) Not Used
Configuration Information (2010,0150) Not Used
Requested Image Size (2020,0030) Not Used
Requested Decimate/Crop Behavior (2020,0040) Not Used
Referenced Presentation LUT Sequence (2050,0500) Not Used
> Referenced SOP Class UID (0008,1150) Not Used
> Referenced SOP Instance UID (0008,1155) Not Used

10.5 PRINTER SOP CLASS

10.5.1 Printer N-Event-Report Attributes


The following table describes the product behavior when receiving an N-Event-Report
request from the Printer SCP depending on the Event Type ID value.

115
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Event Type Event Attribute Tag Use
Name Type ID
Normal 1 Ignored
Warning 2 Printer Status Info (2110,0020) Ignored
Film Destination (2000,0040) Ignored
Printer Name (2110,0030) Ignored
Failure 3 Printer Status Info (2110,0020) Ignored
Film Destination (2000,0040) Ignored
Printer Name (2110,0030) Ignored

10.5.2 Printer N-Get Attributes

Attribute Name Tag Use


Printer Status (2110,0010) When status is FAILURE, we'll not execute print any
more and inform user print fail and the printer status is
abnormal
Printer Status Info (2110,0020) Ignored
Printer Name (2110,0030) Ignored
Manufacturer (0008,0070) Ignored
Manufacturer Model Name (0008,1090) Ignored
Device Serial Number (0018,1000) Ignored
Software Versions (0018,1020) Ignored
Date Of Last Calibration (0018,1200) Ignored
Time Of Last Calibration (0018,1201) Ignored

116
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2

11. QUERY IMPLEMENTATION

11.1 BRIVO OEC 715/785/865 MAPPING OF DICOM ENTITIES


The Brivo OEC 715/785/865 maps DICOM Information Entities to local Information
Entities in the products database and user interface.

TABLE 11-1
MAPPING OF DICOM ENTITIES TO BRIVO OEC 715/785/865 ENTITIES
DICOM Brivo OEC 715/785/865 Entity
Patient Exam
Study Exam
Series Exam
Image Image

11.2 INFORMATION MODEL KEYS


Please refer to DICOM Standard PS 3.4 (Service Class Specifications) for a description
of each of the levels contained within the Query/Retrieve Information Model.

11.2.1 Common Query Keys

The query key attributes specified in this section are used at all levels and in all classes of
query.

TABLE 11-2
Q/R PATIENT LEVEL COMMON RETRIEVE ATTRIBUTES
Attribute Name Tag Type SCU Use
Specific Character Set (0008,0005) - See 11.2.1.1.1
Query Retrieve Level (0008,0052) - Set to level of query:
PATIENT
STUDY
SERIES
IMAGE
Retrieve AE Title (0008,0054) - Returned value will be used for Retrieve AE Title of
retrieval of query response item.
Storage Media File-set ID (0088,0130) - Attribute is not requested.
Storage Media File-set UID (0088,0140) - Attribute is not requested.

117
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
11.2.1.1 Q/R Common Attribute Descriptions

11.2.1.1.1 Specific Character Set


In Query requests the character set tag (0008,0005) is set to ISO_IR 100 for all
requests.

In processing of query responses the character set tag (0008,0005) is checked: absence of
the tag is defined as ISO_IR 6 (7 bit ASCII) or the tag is present ISO_IR 100 (ISO-8859-
1 Latin 1) will be accepted.

11.2.2 Patient Level


This section defines the keys at the Patient Level of the Patient Root Query/Retrieve
Information Models that are supported by this implementation.

TABLE 11-3
PATIENT LEVEL ATTRIBUTES FOR THE PATIENT ROOT
QUERY/RETRIEVE INFORMATION MODEL
Attribute Name Tag Type SCU Use
Patient's Name (0010,0010) R* Single value and wildcard matching are supported.
User could enter anything (including wildcard) except
"^","=" and "\".
The value entered by the user will be used as matching
key, but the response from server depends on server's
capability.
During the DICOM archive/sending, the "^" and "="
will be used as delimiter (added by system) in Patient
Name components.
Patient ID (0010,0020) U* Single value and wildcard matching are supported.
User could enter anything (including wildcard) except
"\".
The value entered by the user will be used as matching
key, but the response from server depends on server's
capability.
Referenced Patient (0008,1120) O Not Requested
Sequence
>Referenced SOP Class (0008,1150) O Not Requested
UID
>Referenced SOP Instance (0008,1155) O Not Requested
UID
Patient's Birth Date (0010,0030) O Universal Matching
Patient's Birth Time (0010,0032) O Not Requested
Patient's Sex (0010,0040) O Universal Matching
Other Patient IDs (0010,1000) O Not Requested
Other Patient Names (0010,1001) O Not Requested
Ethnic Group (0010,2160) O Not Requested
Patient Comments (0010,4000) O Not Requested

118
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Number of Patient Related (0020,1200) O Universal Matching
Studies
Number of Patient Related (0020,1202) O Universal Matching
Series
Number of Patient Related (0020,1204) O Universal Matching
Instances

Note: * in the Type column indicates that this information is displayed on screen, if
available

11.2.3 Study Level Patient Root


This section defines the keys at the Study Level of the Patient Root Query/Retrieve
Information Models that are supported by this implementation.

TABLE 11-4
STUDY LEVEL ATTRIBUTES FOR THE PATIENT ROOT
QUERY/RETRIEVE INFORMATION MODEL
Attribute Name Tag Type SCU Use
Study Date (0008,0020) R* Range value matching is supported.
User could select one of "All", "Today", "Since
Yesterday", "Last 3 days", "Last 7 days" and "Last 30
days" from the screen, and then date information will
be sent as "yyyymmdd-yyyymmdd" format to the
server for matching.
Study Time (0008,0030) R* Universal Matching
Accession Number (0008,0050) R* Single value and wildcard matching are supported.
Leading and trailing wildcard matching for value
entered.
The value entered by the user will be used as matching
key.
Study ID (0020,0010) R Universal Matching
Study Instance UID (0020,000D) U Universal Matching
Modalities in Study (0008,0061) O Universal Matching
SOP Classes in Study (0008,0062) O Not Requested
Referring Physician's Name (0008,0090) O Not Requested
Study Description (0008,1030) O* Universal Matching
Procedure Code Sequence (0008,1032) O Not Requested
>Code Value (0008,0100) O* Not Requested
>Coding Scheme (0008,0102) O Not Requested
Designator
>Coding Scheme Version (0008,0103) O Not Requested
>Code Meaning (0008,0104) O Not Requested
Name of Physician(s) (0008,1060) O Not Requested
Reading Study
Admitting Diagnoses (0008,1080) O Not Requested
Description

119
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Referenced Study Sequence (0008,1110) O Not Requested
>Referenced SOP Class (0008,1150) O Not Requested
UID
>Referenced SOP Instance (0008,1155) O Not Requested
UID
Issuer of Patient ID (0010,0021) O Not Requested
Patients Age (0010,1010) O Not Requested
Patients Size (0010,1020) O Not Requested
Patients Weight (0010,1030) O Not Requested
Occupation (0010,2180) O Not Requested
Additional Patient History (0010,21B0) O Not Requested
Other Study Numbers (0020,1070) O Not Requested
Number of Study Related (0020,1206) O Universal Matching
Series
Number of Study related (0020,1208) O Universal Matching
Instances
Interpretation Author (4008,010C) O Not Requested

Note: * in the Type column indicates that this information is displayed on screen, if
available

11.2.4 Study Level Study Root


This section defines the keys at the Study Level of the Study Root Query/Retrieve
Information Model that are supported by this implementation.

TABLE 11-5
STUDY LEVEL ATTRIBUTES FOR THE STUDY ROOT
QUERY/RETRIEVE INFORMATION MODEL
Attribute Name Tag Type SCU Use
Study Date (0008,0020) R* Range value matching is supported.
User could select one of "All", "Today", "Since
Yesterday", "Last 3 days", "Last 7 days" and "Last 30
days" from the screen, and then date information will
be sent as "yyyymmdd-yyyymmdd" format to the
server for matching.
Study Time (0008,0030) R* Universal Matching
Accession Number (0008,0050) R* Single value and wildcard matching are supported.
The value entered by the user will be used as matching
key.
Study ID (0020,0010) R Universal Matching

120
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Patient's Name (0010,0010) R* Single value and wildcard matching are supported.
User could enter anything (including wildcard) except
"^","=" and "\".
The value entered by the user will be used as matching
key, but the response from server depends on server's
capability.
During the DICOM archive/sending, the "^" and "="
will be used as delimiter (added by system) in Patient
Name components.
Patient ID (0010,0020) R* Single value and wildcard matching are supported.
User could enter anything (including wildcard) except
"\".
The value entered by the user will be used as matching
key, but the response from server depends on server's
capability.
Study Instance UID (0020,000D) U Universal Matching
Modalities in Study (0008,0061) O Universal Matching

SOP Classes in Study (0008,0062) O Not Requested


Referring Physician's Name (0008,0090) O Not Requested
Study Description (0008,1030) O* Universal Matching
Procedure Code Sequence (0008,1032) O Not Requested
>Code Value (0008,0100) O Not Requested
>Coding Scheme (0008,0102) O Not Requested
Designator
>Coding Scheme Version (0008,0103) O Not Requested
>Code Meaning (0008,0104) O Not Requested
Name of Physician(s) (0008,1060) O Not Requested
Reading Study
Admitting Diagnoses (0008,1080) O Not Requested
Description
Referenced Study Sequence (0008,1110) O Not Requested
>Referenced SOP Class (0008,1150) O Not Requested
UID
>Referenced SOP Instance (0008,1155) O Not Requested
UID
Issuer of Patient ID (0010,0021) O Not Requested
Patients Age (0010,1010) O Not Requested
Patients Size (0010,1020) O Not Requested
Patients Weight (0010,1030) O Not Requested
Occupation (0010,2180) O Not Requested
Additional Patient History (0010,21B0) O Not Requested
Other Study Numbers (0020,1070) O Not Requested

121
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Number of Study Related (0020,1206) O Universal Matching
Series
Number of Study related (0020,1208) O Universal Matching
Instances
Interpretation Author (4008,010C) O Not Requested
Referenced Patient (0008,1120) O Not Requested
Sequence
>Referenced SOP Class (0008,1150) O Not Requested
UID
>Referenced SOP Instance (0008,1155) O Not Requested
UID
Patient's Birth Date (0010,0030) O Universal Matching
Patient's Birth Time (0010,0032) O Not Requested
Patient's Sex (0010,0040) O Universal Matching
Other Patient IDs (0010,1000) O Not Requested
Other Patient Names (0010,1001) O Not Requested
Ethnic Group (0010,2160) O Not Requested
Patient Comments (0010,4000) O Not Requested
Number of Patient Related (0020,1200) O Not Requested
Studies
Number of Patient Related (0020,1202) O Not Requested
Series
Number of Patient Related (0020,1204) O Not Requested
Instances

Note: * in the Type column indicates that this information is displayed on screen, if
available

11.2.5 Series Level

This section defines the keys at the Series Level of the Patient Root and Study Root
Query/Retrieve Information Models that are supported by this implementation.

TABLE 11-6
SERIES LEVEL ATTRIBUTES FOR THE
QUERY/RETRIEVE INFORMATION MODEL
Attribute Name Tag Type SCU Use
Modality (0008,0060) R* Single value matching is supported.
User could select one of "XA", "RF", "CR", "SC" and
ALL for Modality Type from screen, then the
corresponding string "XA", "RF", "CR", "OT" and
will be sent to server for matching.
Note: "ALL" means ignore the Modality Type, the
Modality Type will leave empty during sending.
Series Number (0020,0011) R Universal Matching
Series Instance UID (0020,000E) U Universal Matching

122
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Number of Series Related (0020,1209) O Universal Matching
Instances
Series Description (0008,103E) O Universal Matching
Performed Procedure Step (0040, 0253) O Not Requested
ID
Referenced Performed (0008,1111) O Not Requested
Procedure Step Sequence
>Referenced SOP Class (0008,1150) O Not Requested
UID
>Referenced SOP Instance (0008,1155) O Not Requested
UID
Request Attribute Sequence (0040, 0275) O Not Requested
>Requested Procedure ID (0040,1001) O Not Requested
>Scheduled Procedure Step (0040,0009) O Not Requested
ID
Performed Procedure Step (0040,0244) O Not Requested
Start Date
Performed Procedure Step (0040,0245) O Not Requested
Start Time

Note: * in the Type column indicates that this information is displayed on screen, if
available

11.2.6 Image Level


Currently no Image level query requests are made.

123