GEHC DICOM Conformance - BrivoOEC715 785 865 VWKS1 XX - DOC1188156 - Rev3
GEHC DICOM Conformance - BrivoOEC715 785 865 VWKS1 XX - DOC1188156 - Rev3
Publications
DOC1188156
Revision 3
Do not duplicate
g
GE Healthcare
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
REVISION HISTORY
REV DATE REASON FOR CHANGE
1 Nov 12 2012 Initial Release
2 Mar 07 2013 Update document per Support Center Review Comments:
BrivoOec715.785.865DcsReview_HHComments_Mar072013.xls
x
3 Jun 18 2013 Update document per SPR HCSDM00203914
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Table 0.1 provides an overview of the network services supported by Brivo OEC
715/785/865 system.
Table 0.2 provides an overview of the Media Storage Application Profiles supported by
Brivo OEC 715/785/865 system.
4
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
TABLE OF CONTENTS
1.
INTRODUCTION ............................................................................................................... 8
1.1
Overview ...................................................................................................................................................................... 8
5
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
3.5
Configuration ............................................................................................................................................................ 47
6
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
7.4
Modality Performed Procedure Step Module Definitions ..................................................................................... 96
7
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
1. INTRODUCTION
1.1 OVERVIEW
This DICOM Conformance Statement is divided into Sections as described below:
8
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
1.2 OVERALL DICOM CONFORMANCE STATEMENT DOCUMENT STRUCTURE
The Documentation Structure of the GEHC DICOM Conformance Statements is shown
in the Illustration below.
@
http://www.ge.com/DICOM
CT Advantage
C onformanc e
MR Advantage
Statement
Conformance
Direc tion:
Briv o OEC
Statement ......
Product 715/785/865
Direction:
C onformanc e
Conformance
Statement
Implementation: Statement
DOC1188156
Direction:
......
DICOM STANDARD
Standard DICOM
Specification: Part 1
DICOM
DICOM
Part 2
D ICOM
Part 3
Part 4
DICOM
Part 20
9
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
This document specifies the DICOM implementation. It is entitled:
The GEHC Conformance Statement, contained in this document, also specifies the Lower
Layer communications which it supports (e.g., TCP/IP). However, the Technical
Specifications are defined in the DICOM Part 8 standard.
For more information regarding DICOM, copies of the Standard may be obtained on the
Internet at http://medical.nema.org. Comments on the Standard may be addressed to:
DICOM Secretariat
NEMA
1300 N. 17th Street, Suite 1752
Rosslyn, VA 22209
USA
Phone: +1.703.841.3200
The reader of this document is concerned with software design and/or system integration
issues. It is assumed that the reader of this document is familiar with the DICOM
Standard and with the terminology and concepts which are used in that Standard.
The reader of this DICOM Conformance Statement should be aware that different GEHC
devices are capable of using different Information Object Definitions. For example, a
GEHC CT Scanner may send images using the CT Information Object, MR Information
Object, Secondary Capture Object, etc.
Included in this DICOM Conformance Statement are the Module Definitions which
define all data elements used by this GEHC implementation. If the user encounters
unspecified private data elements while parsing a GEHC Data Set, the user is well
advised to ignore those data elements (per the DICOM standard). Unspecified private
data element information is subject to change without notice. If, however, the device is
acting as a "full fidelity storage device", it should retain and re-transmit all of the private
data elements which are sent by GEHC devices.
10
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
successful. The user (or user's agent) needs to proceed with caution and address at
least four issues:
Future Evolution - GE understands that the DICOM Standard will evolve to meet
the user's growing requirements. GE is actively involved in the development of the
DICOM Standard. DICOM will incorporate new features and technologies and GE
may follow the evolution of the Standard. The GEHC protocol is based on DICOM
as specified in each DICOM Conformance Statement. Evolution of the Standard may
require changes to devices which have implemented DICOM. In addition, GE
reserves the right to discontinue or make changes to the support of
communications features (on its products) described by these DICOM
Conformance Statements. The user should ensure that any nonGE provider,
which connects with GE devices, also plans for the future evolution of the DICOM
Standard. Failure to do so will likely result in the loss of function and/or connectivity
as the DICOM Standard changes and GE Products are enhanced to support these
changes.
Interaction - It is the sole responsibility of the nonGE provider to ensure that
communication with the interfaced equipment does not cause degradation of GE
imaging equipment performance and/or function.
1.6 REFERENCES
NEMA PS3 Digital Imaging and Communications in Medicine (DICOM)
Standard, available free at http://medical.nema.org/
1.7 DEFINITIONS
Informal definitions are provided for the following terms used in this Conformance
Statement. The DICOM Standard is the authoritative source for formal definitions of
these terms.
11
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Application Entity (AE) an end point of a DICOM information exchange, including
the DICOM network or media interface software; i.e., the software that sends or receives
DICOM information objects or messages. A single device may have multiple
Application Entities.
Application Entity Title the externally known name of an Application Entity, used to
identify a DICOM application to other DICOM applications on the network.
Information Object Definition (IOD) the specified set of Attributes that comprise a
type of data object; does not represent a specific instance of the data object, but rather a
class of similar data objects that have the same properties. The Attributes may be
specified as Mandatory (Type 1), Required but possibly unknown (Type 2), or Optional
(Type 3), and there may be conditions associated with the use of an Attribute (Types 1C
and 2C). Examples: MR Image IOD, CT Image IOD, Print Job IOD.
Module a set of Attributes within an Information Object Definition that are logically
related to each other. Example: Patient Module includes Patient Name, Patient ID, Patient
Birth Date, and Patient Sex.
Presentation Context the set of DICOM network services used over an Association, as
negotiated between Application Entities; includes Abstract Syntaxes and Transfer
Syntaxes.
Protocol Data Unit (PDU) a packet (piece) of a DICOM message sent across the
network. Devices must specify the maximum size packet they can receive for DICOM
messages.
Service Class Provider (SCP) role of an Application Entity that provides a DICOM
network service; typically, a server that performs operations requested by another
Application Entity (Service Class User). Examples: Picture Archiving and
Communication System (image storage SCP, and image query/retrieve SCP), Radiology
Information System (modality worklist SCP).
12
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Service Class User (SCU) role of an Application Entity that uses a DICOM network
service; typically, a client. Examples: imaging modality (image storage SCU, and
modality worklist SCU), imaging workstation (image query/retrieve SCU)
Service/Object Pair (SOP) Class the specification of the network or media transfer
(service) of a particular type of data (object); the fundamental unit of DICOM
interoperability specification. Examples: Ultrasound Image Storage Service, Basic
Grayscale Print Management.
Tag a 32-bit identifier for a data element, represented as a pair of four digit
hexadecimal numbers, the group and the element. If the group number is odd, the
tag is for a private (manufacturer-specific) data element. Examples: (0010,0020) [Patient
ID], (07FE,0010) [Pixel Data], (0019,0210) [private data element]
Transfer Syntax the encoding used for exchange of DICOM information objects and
messages. Examples: JPEG compressed (images), little endian explicit value
representation.
Unique Identifier (UID) a globally unique dotted decimal string that identifies a
specific object or a class of objects; an ISO-8824 Object Identifier. Examples: Study
Instance UID, SOP Class UID, SOP Instance UID.
Value Representation (VR) the format type of an individual DICOM data element,
such as text, an integer, a persons name, or a code. DICOM information objects can be
transmitted with either explicit identification of the type of each data element (Explicit
VR), or without explicit identification (Implicit VR); with Implicit VR, the receiving
application must use a DICOM data dictionary to look up the format of each data
element.
CR Computed Radiography
CT Computed Tomography
DB Database
13
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
DX Digital X-ray
RF Radiofluoroscopy
SC Secondary Capture
SR Structured Reporting
14
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
U Unique (Key Attribute)
VR Value Representation
XA X-ray Angiography
15
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
2.1 INTRODUCTION
This section of the DICOM Conformance Statement specifies the Brivo OEC
715/785/865 compliance to DICOM requirements for Networking features.
Brivo OEC 715/785/865 runs on industrial computer. It provides the following DICOM
functionalities:
Querying for saved examinations and retrieving saved examinations from a DICOM
Query/Retrieve SCP
Sending exam start and end messages to a DICOM Modality Performed Procedure
Step SCP.
16
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
ILLUSTRATION 21
BRIVO OEC 715/785/865 NETWORK APPLICATION MODEL AND DATA FLOW DIAGRAM
Remote
Storage Storage
SCP
Remote
Storage
Commit
SCP
Brivo OEC
715/785/865 Remote
Verify Echo
Workstation
AE SCP
Remote
Modality
Modality
Worklist
Worklist
Query
SCP
Remote
Report
MPPS
MPPS
SCP
Query Remote
Retrieve Q/R
SCP
Local
Store
SCP
Remote
DICO Print
M SCP
Print
DICOM
Standard
Interface
The product supports activity Storage which performs images/RDSR storage and
storage commitment functions; The results of that functionalities are communicated to
Remote Store Server for DICOM image/RDSR storage and storage commitment.
The product supports activity Print which performs images print functions; The results
of that functionalities are communicated to Remote Printer/Server for DICOM image
print functions.
The product supports activity Verify which interacts with Remote Verify SCP for
DICOM Echo purpose.
17
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
The product supports activity Modality Worklist Query which performs Modality
worklist query, from remote Modality Worklist SCP.
The product supports activity Query: Initiate an association with a Remote AE to query
for images on the remote host. Once the association has been established, it sends a
sequence of requests (Patient, then Study then Series Level requests) to the Remote AE.
After all responses are received, Workstation AE will issue a Series-Level C-FIND-RQ
request to get the series for a study in the list.
The product supports activity Report MPPS: When the user begins the image
acquisition process and generates the first image, the Workstation AE sends N-CREATE
message to the configured MPPS SCP to indicate that the image acquisition process has
been started for the requested procedure. The operator can close the acquisition session
either by completing the acquisition process or discontinuing the ongoing scan. On
closing the acquisition session, the Workstation AE sends N-SET message to the
configured MPPS SCP to indicate the acquisition state of the requested procedure, with
appropriate MPPS status (COMPLETED/DISCONTINUED).
Query: Initiate an association with a Remote AE to query for images on the remote
host. Follows the Patient->Study->Series sequence C-FIND-RQ request (the request
more than once) will be sent to the Remote AE once an association has been
established. Once all the responses have been received, the operator needs to select
an exam in the local database browser; on selection of the exam the DICOM Server
AE will issue a Series-Level C-FIND-RQ request to get the series for a study in the
18
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
list. Similarly the Image-Level C-FIND-RQ will be issued for the series selected
from the series list.
1. The user selects the images and sends them to a remote host.
2.3 AE SPECIFICATIONS
19
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Modality Performed Procedure Step 1.2.840.10008.3.1.2.3.3 Yes No
2.3.1.1.1 General
The DICOM Application Context Name (ACN), which is always proposed, is:
The maximum length PDU receive size for the Workstation AE is:
20
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
21
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
2.3.1.2.1.2.1 SOP Specific DICOM Conformance Statement for All Storage SOP Classes
The Workstation AE includes optional data elements in the SOP Instances as described in
Sections 4.
Although the Modality Type could be configured to RF/XA/CR/SC when user export
image to interchange media or remote DICOM server, only XA Image IOD is used.
Following are the status codes that are more specifically processed when receiving
messages from a Storage SCP equipment : If store command completed normally, with a
status of success or some warning then the image was accepted.
Failure A700- Refused: Out of resources Terminate the operation. Logs store failed message
A7FF along with error comment returned from SCP.
Displays store failure message.
A900- Error: Data Set does not match
A9FF SOP Class
Warning B000 Coercion of Data Elements Ignored by application and current operation
continues
B006 Elements Discarded
Success 0000 DICOM storage operation is The DICOM operation progress screen
success disappearance indicates current operation is
success.
2.3.1.2.1.2.2 SOP Specific DICOM Conformance Statement for the X-Ray Radiation Dose Storage SOP Classes
See Section 2.3.1.2.1.2.1 SOP Specific DICOM Conformance Statement for All Storage
SOP Classes for details on general Storage Service SCU processing also applicable to
the X-Ray Radiation Dose Storage SOP Classes.
The Workstation AE supports creation and transmission of X-Ray Radiation Dose SOP
Instances referencing Instances of the following Storage SOP Classes:
22
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
SOP Class Name SOP Class UID
2.3.1.2.1.2.3 SOP Specific DICOM Conformance Statement for the Storage Commitment Push Model SOP Class
SCU
The Workstation AE uses DICOM network storage services to transfer SOP Instances
which are to be committed.
The Workstation AE may request Storage Commitment for Instances of any of the
Composite SOP Classes it supports as an SCU (see Section 2.3.1.2.1.2.1).
Following are the status codes that are more specifically processed when receiving N-
Action responses from Storage Commitment SCP equipment:
Failure 0119 Class-instance conflict Terminates the association and operation. Logs
failure message along with error comment returned
0210 Duplicate invocation from SCP and displays failure message.
23
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
* * Any other status code. Deemed a failure.
Upon a request by the operator the selected images will be sent to the configured DICOM
Printer (DICOM Print SCP).
Basic Grayscale Print 1.2.840.10008.5.1.1.9 Implicit VR Little Endian 1.2.840.10008.1.2 SCU None
Management Meta SOP Explicit VR Little Endian 1.2.840.10008.1.2.1
Class Explicit VR Big Endian
1.2.840.10008.1.2.2
2.3.1.2.2.2.1 SOP Specific DICOM Conformance Statement for Basic Grayscale Print Management SOP Classes
The Workstation AE uses the following DIMSE services of the supported SOP Classes:
SOP Class SOP Class UID DIMSE Service Element SCU Usage
Basic Film Session 1.2.840.10008.5.1.1.1 N-CREATE Used (Mandatory)
N-SET Not Used
N-DELETE Used
N-ACTION Not Used
Basic Film Box 1.2.840.10008.5.1.1.2 N-CREATE Used (Mandatory)
N-ACTION Used (Mandatory)
N-DELETE Used
N-SET Not Used
Printer 1.2.840.10008.5.1.1.16 N-EVENT-REPORT Used (Mandatory)
N-GET Used
Basic Grayscale 1.2.840.10008.5.1.1.4 N-SET Used (Mandatory)
Image Box
24
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
2. N-GETs printer status using the Printer SOP Class
3. N-CREATEs a Basic Film Session SOP Instance, and then followed by N-DELETE of
the Basic Film Session SOP Instance.
5. N-SETs the Image Box SOP Instance for each image on the film
6. Prints by N-ACTION on the Basic Film Box SOP Instances, then followed by N-
DELETE of the Basic Film Box SOP Instance
8. Releases the DICOM association after printing is successful or failure has been
signaled to the user
Following are the status codes that are more specifically processed when receiving
messages from a Print SCP equipment for the Basic Film Session SOP Class N-
CREATE:
Failure 0119 Class-instance conflict Terminate the operation. Logs store failed
message along with error comment returned from
0210 Duplicate invocation SCP. Displays print failure message.
25
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
0110 Processing failure
Warning B600 Memory allocation not supported Ignored by application and current operation
continues
Success 0000 Film session successfully created The progress screen is still displayed.
Following are the status codes that are more specifically processed when receiving
messages from a Print SCP equipment for the Basic Film Session SOP Class N-
DELETE:
Failure 0119 Class-instance conflict Terminate the operation. Logs store failed
message along with error comment returned from
0210 Duplicate invocation SCP. Displays print failure message.
Success 0000 Film session successfully deleted The progress screen is still displayed.
26
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Status Codes Codes
Failure C616 There is an existing Film Box that has Terminate the operation. Logs store failed
not been printed and N-ACTION at the message along with error comment returned from
Film Session level is not supported. A SCP. Displays print failure message.
new Film Box will not be created when a
previous Film Box has not been printed.
Warning B605 Requested Min Density or Max Density Ignored by application and current operation
outside of printers operating range. The continues.
printer will use its respective minimum
or maximum density value instead.
Success 0000 Film box successfully created The progress screen is still displayed.
Following are the status codes that are more specifically processed when receiving
messages from a Print SCP equipment for the Basic Film Box SOP Class N-ACTION:
27
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Service Status Further Meaning Application Behavior When receiving Status
Status Codes Codes
Failure C602 Unable to create Print Job SOP Instance; Terminate the operation. Logs store failed message
print queue is full along with error comment returned from SCP.
C603 Image size is larger than image box size Displays print failure message.
(by using the specified magnification
value)
C604 Image position collision : multiple images
assigned to single image position
C613 Combined Print Image size is larger than
the Image Box size
0119 Class-instance conflict
0210 Duplicate invocation
0115 Invalid argument value
0117 Invalid SOP Instance
0212 Mistyped argument
0123 No such action
0114 No such argument
0118 No such SOP Class
0112 No such SOP Instance
0110 Processing failure
0213 Resource limitation
0211 Unrecognized operation
Warning B603 Film Box SOP Instance hierarchy does Ignored by application and current operation
not contain Image Box SOP Instances continues.
(empty page)
B604 Image size is larger than image box size,
the image has been demagnified.
B609 Image size is larger than the Image Box
size. The Image has been cropped to fit.
B60A Image size or Combined Print Image size
is larger than the Image Box size. Image
or Combined Print Image has been
decimated to fit.
Success 0000 Film accepted for printing; if supported, The progress screen is still displayed.
the Print Job SOP Instance is created
* * Any other status code. Deemed a failure.
Following are the status codes that are more specifically processed when receiving
messages from a Print SCP equipment for the Basic Film Box SOP Class N-DELETE:
28
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Failure 0119 Class-instance conflict Terminate the operation. Logs store failed
message along with error comment returned from
0210 Duplicate invocation SCP. Displays print failure message.
Success 0000 Film box successfully deleted The progress screen is still displayed.
Following are the status codes that are more specifically processed when receiving
messages from a Print SCP equipment for the Basic Grayscale Image Box SOP Class N-
SET:
29
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
0105 No such attributes
Warning B604 Image size larger than image box size, Ignored by application and current operation
the image has been demagnified. continues.
Success 0000 Image successfully stored in Image Box The progress screen is still displayed.
The Workstation AE supports the Printer SOP Class to receive information on the status
of the printer.
For the product behavior description when receiving N-Event-Report request, refer to
section 10.5.1 Printer N-Event-Report Attributes.
Also, following are the status codes the Application may send back in the N-Event-
Report response command to the Printer SOP Class SCP Equipment that sent the N-
Event-Report request:
30
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
For the attribute list requested by this product and for the behavior of this product on each
returned value, refer to Section 10.5.22 Printer N-Get Attributes
Following are the status codes that are more specifically processed when receiving
messages from a Print SCP equipment for the Printer SOP Class N-GET:
Failure 0107 Attribute list error Terminate the operation. Logs store failed
message along with error comment returned from
0119 Class-instance conflict SCP. Displays print failure message.
A valid response from the SCP will result in a The DICOM server has verified
successfully screen displayed on the screen. In the event that the SCP does not response
for some reason, the operation will timeout and the Workstation AE will close the
association and results in The DICOM Verify Fails screen displayed to user.
Note: The default timeout is 30 seconds for DICOM Echo and this value could be
configured from system setup screen.
31
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
2.3.1.2.3.2 Proposed Presentation Context Table
2.3.1.2.3.2.1 SOP Specific DICOM Conformance Statement for Verify SOP Classes
The message The DICOM server has verified successfully is displayed when the
DICOM Verify is successfully performed, otherwise message The DICOM Verify
Fails is displayed.
2.3.1.2.4.2.1 SOP Specific DICOM Conformance Statement for the Modality Worklist Information Model - FIND
SOP Class
The Workstation AE includes matching keys in the Modality Worklist queries as
described in Section 6.
The results of the Worklist query are displayed to the user in the Scheduled Exam Screen.
Currently, the maximal capability of received scheduled exams is 500 and not
configurable.
The Workstation AE checks the Specific Character Set tag (0008,0005) contained in a C-
FIND RSP, ISO_IR 100 or blank will be accepted, and all other value will be discarded.
Responses without this tag will also be accepted.
32
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
A C-FIND CANCEL will be sent when the maximal capability of received scheduled
exams(500 exams) is reached or user cancels operation by clicking the Cancel button on
the progressing screen.
Following are the status codes that are more specifically processed when receiving
messages from a Modality Worklist SCP equipment:
Failure A700 Refused: Out of resources Terminates the association and operation. Logs
failure message along with error comment returned
A900 Error: Identifier does not match from SCP and displays failure message.
SOP Class
Cancel FE00 Matching terminated due to The progress screen is disappeared, the received
cancel scheduled exams will be displayed.
Success 0000 Matching is complete - No final The progress screen is disappeared, all the received
identifier is supplied scheduled exams will be displayed.
Pending FF00 Matches are continuing - The progress screen is still displayed.
Current Match is supplied and
any Optional Keys were
supported in the same manner
as Required Keys.
33
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
2.3.1.2.5.2 Proposed Presentation Context Table
Presentation Context Table Proposed by Brivo OEC 715/786/865 AE for Activity Modality Worklist
Abstract Syntax Transfer Syntax Role Extended
2.3.1.2.5.2.1 SOP Specific DICOM Conformance Statement for Modality Performed Procedure Step SOP Class
MPPS will be created at the start of an image acquisition step; an MPPS will be set after
the completion/discontinue of a post-processing step when the user selects
Complete/Discontinue .
Following are the status codes that are more specifically processed when receiving an N-
Create or N-Set response from an MPPS SCP equipment:
Failure 0119 Class-instance conflict Terminates the association and operation. Logs
failure message along with error comment returned
0210 Duplicate invocation from SCP and displays failure message.
34
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Success 0000 The progress screen is disappeared.
After a DICOM file is retrieved, the attributes are extracted and saved onto the local
system.
Presentation Context Table Proposed by Brivo OEC 715/786/865 AE for Activity Modality Worklist
Abstract Syntax Transfer Syntax Role
Name UID Name List UID List
35
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
2.3.1.2.6.2.1 SOP Specific DICOM Conformance Statement for the Patient Root Query/Retrieve Information Model -
FIND, Study Root Query/Retrieve Information Model - FIND SOP Classes
The Workstation AE includes matching keys in the queries as described in Sections 11.
The Acquisition AE evaluates a C-FIND RSP containing the Specific Character Set tag
(0008,0005). Values of ISO_IR 100 or blank will be accepted, and all others will be
rejected.
A C-FIND CANCEL will be sent when the display limit for results is reached.
Following are the status codes that are more specifically processed when receiving
messages from a Query SCP equipment:
Failure A700 Refused: Out of resources Terminates the association and operation. Logs
failure message along with error comment returned
A900 Error: Identifier does not match from SCP and displays failure message.
SOP Class
Cancel FE00 Matching terminated due to The progress screen is disappeared, the received
cancel scheduled exams will be displayed
Success 0000 Matching is complete - No final The progress screen is disappeared, all the received
identifier is supplied scheduled exams will be displayed.
Pending FF00 Matches are continuing - The progress screen is still displayed.
Current Match is supplied and
any Optional Keys were
supported in the same manner
as Required Keys.
36
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
2.3.1.2.6.2.2 SOP Specific DICOM Conformance Statement for the Patient Root Query/Retrieve Information Model -
MOVE , Study Root Query/Retrieve Information Model - MOVE SOP Classes
The C-MOVE-RQ will use the AE Title of the Workstation Application Entity as the
Move Destination AE Title.
A C-FIND CANCEL message is sent if the user clicks the Cancel button during the
move.
Following are the status codes that are more specifically processed when receiving
messages from a Retrieve SCP equipment:
Failure A701 Refused: Out of resources - Terminates the association and operation. Logs
Unable to calculate number of failure message along with error comment returned
matches from SCP and displays failure message.
Cancel FE00 Sub-operations terminated due The progress screen is disappeared, the received
to a Cancel indication scheduled exams will be displayed
Warning B000 Sub-operations Complete - One Ignored by application and current operation
or more Failures. continues.
Success 0000 Sub-operations Complete - No The progress screen is disappeared, all the received
Failure. scheduled exams will be displayed.
Pending FF00 Sub-operations are continuing - The progress screen is still displayed.
37
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
When the DICOM Server AE accepts an association for image storage, it will receive any
images transmitted on that association and store the images on disk.
It will also respond to queries from Remote AEs by sending matching entries. Any
Remote AE can request and receive a list of images on the local database. The Remote
AE must be configured in the local config file list of Remote AE for it to be able to
retrieve images from DICOM Server AE.
2.3.1.3.1.2.1 SOP Specific DICOM Conformance Statement for all Storage SOP Classes
The Workstation AE provides Level 2 (FULL) Conformance, and stores all standard and
private data elements of received SOP Instances.
Following are the status codes the Application may send back to the SCU Equipment
after performing the requested Storage:
38
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Success 0000 None
Verification SOP Class 1.2.840.10008.1.1 Implicit VR Little Endian 1.2.840.10008.1.2 SCP None
Explicit VR Little Endian 1.2.840.10008.1.2.1
Explicit VR Big Endian 1.2.840.10008.1.2.2
The DICOM Upper Layer Protocol is supported using TCP/IP, as specified in DICOM
PS3.8.
The Workstation supports DHCP, user could configure DHCP Enable/Disable from
System Setup screen. When DHCP is Enabled, then, the network information will be
gotten automatically.
Default is Enabled.
The default value for these fields is blank unless otherwise specified.
Network
Configuration:
Every Remote DICOM AE SCP: (Store, Storage Commitment, MPPS, Print, Modality
Worklist, Query/Retrieve)
Server Alias
Remote AE Title
Remote IP Address
Remote Port Number
Gateway IP Address.
Modality Type (Default - RF)
40
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Film Destination
Medium Type
Film Size
Format
Film Orientation
Patient Name
Patient ID
Requested Procedure ID
Accession Number
Exam Date: Today, Next 2 Days, Next 3 Days, Next 7 Days, Date Range
Performing Physician Name
Modality Type (Default - RF)
Additional fields for Remote DICOM Query/Retrieve Filter
The product user interface will allow the user to enter characters from the console
keyboard that is within ISO 8859-1. Responses with non-compatible values of
(0008,0005) will be discarded.
41
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
2.9 SECURITY PROFILES
The product does not conform to any defined DICOM Security Profiles.
It is assumed that the product is used within a secured environment. It is assumed that a
secured environment includes at a minimum:
1. Firewall or router protections to ensure that only approved external hosts have
network access to the product.
2. Firewall or router protections to ensure that the product only has network access to
approved external hosts and services.
3. Any communications with external hosts and services outside the locally secured
environment use appropriate secure network channels (such as a Virtual Private
Network (VPN))
42
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
3.1 INTRODUCTION
This section of the DICOM conformance statement specifies the Workstation AE
compliance to DICOM requirements for Media Interchange. It details the DICOM
Media Storage Application Profiles and roles that are supported by this product.
The media interchange application model for the Workstation AE is shown in the
following Illustration:
ILLUSTRATION 31
WORKSTATION AE MEDIA INTERCHANGE APPLICATION MODEL AND DATA FLOW DIAGRAM
Storage FSC
Remote
Brivo OEC 715/785/865 Media Storage
Workstation AE Storage SCP
FSU
Export
FSR
Directory Media Viewer AE Media
Listing Storage
Update a DICOM File-Set by adding new SOP instances to the media which
supports multi-session.
Reads a DICOM File-Set from media and reviews recorded DICOM instances
(including images and RDSR).
43
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
3.1.3 Sequencing of Real-World Activities
Storage/Export Images are acquired and archived on Brivo OEC 715/785/865; From the
Image Directory screen, images/RDSR can be selected and copied to media;
Images/RDSR can be viewed by any DICOM application that supports Media Exchange
FSR and X-Ray Radiation Dose Structured Reports.
Directory Listing During the Image Copy, user could choice copying image/RDSR with
a Media Viewer, then the images/RDSR will be copied with a Media Viewer instance;
Images/RDSR can be viewed by this media viewer only on Windows OS (Independent of
the Brivo 715/786/865 Workstation AE).
3.2 AE SPECIFICATIONS
44
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
3.2.1.1 File Meta Information for the Workstation Application Entity
Following are the values set in the File Meta Information for this AE Title:
When user selects Storage function, the selected images should be saved as DICOM
SOP instances and copied to media. The DICOM File set on the media should be created
or updated accordingly.
3.2.1.2.1.1 Media Storage Application Profile for the RWA Update File Set
For the list of Application Profiles that invoke this AE for the Real-World Activity
Storage, see the Table in Section 3.2.1 Brivo OEC 715/785/865 where the table
describing the Application Profiles and Real-World Activity is defined.
45
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Common DICOMDIR Directory Records created by this AE will include key attributes
as described in Section 9. Following are the Additional DICOMDIR Keys supported for
this profile:
3.2.2.1 File Meta Information for the Media Viewer Application Entity
Following are the values set in the File Meta Information for this AE Title:
The Media Viewer AE acts as a FSR when listing the directory information from
interchange media.
When user load the media on Window OS, the Media Viewer will read the DICOM File-
Set and display/list the DICOM directory entries for those instances in the File-Set on the
media.
3.2.2.2.1.1 Media Storage Application Profile for the RWA Directory Listing
For the list of Application Profiles that invoke this AE for the Real-World Activity
Directory Listing, see the Table in Section 3.2.3 Media Viewer AE Specification where
the table describing the Application Profiles and Real-World Activity is defined.
46
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
3.2.2.2.1.1.1 STD-GEN-CD Application Profile
Following are the SOP Classes supported by this AE and only the Explicit VR Little
Endian Uncompressed Transfer Syntax (UID 1.2.840.10008.1.2.1) is supported.
Common DICOMDIR Directory Records read by this AE will include key attributes as
described in Section 9. Following are the Additional DICOMDIR Keys supported for this
profile:
3.5 CONFIGURATION
CD Options (Media Exchange)
47
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
3.6 SUPPORT OF EXTENDED CHARACTER SETS
The Workstation AE will support fully only the ISO_IR 100 as extended character sets
for media storage.
48
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
4.1 INTRODUCTION
This section specifies the use of the DICOM X-Ray Angiographic Image IOD to
represent the information included in X-Ray Angiographic Images by this
implementation. Corresponding attributes are conveyed using the module construct.
The Brivo OEC 715/785/865 maps DICOM Information Entities to local Information
Entities in the products database and user interface.
TABLE 4-1
MAPPING OF DICOM ENTITIES TO BRIVO OEC 715/785/865 ENTITIES
DICOM IE Brivo OEC 715/785/865 Entity
Patient Patient
Study Exam
Series Series
Equipment Equipment
Image Image
The X-Ray Angiographic Image Information Object Definition comprises the modules of
the following table, plus Standard Extended and Private attributes. Standard Extended
and Private attributes are described in Section 4.5.
TABLE 4-2
X-RAY ANGIOGRAPHIC IMAGE IOD MODULES
Entity Name Module Name Usage Reference
Patient Patient Used 4.4.1.1
Clinical Trial Subject Not Used N/A
Study General Study Used 4.4.2.1
Patient Study Not Used N/A
Clinical Trial Study Not Used N/A
Series General Series Used 4.4.3.1
Clinical Trial Series Not Used N/A
Frame of Synchronization Not Used N/A
Reference
49
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Equipment General Equipment Used 4.4.4.1
Image General Image Used 4.4.5.1
Image Pixel Used 4.4.5.2
Contrast/Bolus Used 4.4.5.3
Cine Not Used N/A
Multi-Frame Not Used N/A
Frame Pointers Not Used N/A
Mask Not Used N/A
Display Shutter Not Used N/A
Device Not Used N/A
Intervention Not Used N/A
X-Ray Image Used 4.4.5.4
X-Ray Acquisition Used 4.4.5.5
X-Ray Collimator Not Used N/A
X-Ray Table Not Used N/A
XA Positioner Used 4.4.5.66
DX Detector Not Used N/A
Overlay Plane Used 4.4.5.7
Multi-Frame Overlay Not Used N/A
Modality LUT Not Used N/A
VOI LUT Used 4.4.5.88
SOP Common Used 4.4.5.9
Frame Extraction Not Used N/A
The following modules are included to convey Enumerated Values, Defined Terms, and
Optional Attributes supported and/or expected. Type 1 & Type 2 Attributes are also
included for completeness and to define what values they may take and where these
values are obtained from when generating the instance as well as what are the expected
values when loading such instance. It should be noted that they are the same ones as
defined in the DICOM Standard Part 3 (Information Object Definitions). Also note that
Attributes not present in tables are not supported.
50
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
4.4.1 Patient Entity Modules
51
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Breed Registration Sequence (0010,2294) 2C Not Used
>Breed Registration Number (0010,2295) 1 Not Used
>Breed Registry Code Sequence (0010,2296) 1 Not Used
>>Include Code Sequence Macro
Responsible Person (0010,2297) 2C Not Used
Responsible Person Role (0010,2298) 1C Not Used
Responsible Organization (0010,2299) 2C Not Used
Patient Identity Removed (0012,0062) 3 Used.
If Patient Information is removed then set to
"YES", otherwise set to "NO"
De-identification Method (0012,0063) 1C Used.
If Patient Information is removed. Set to
"Limited Data Set" in this case
De-identification Method Code Sequence (0012,0064) 1C Not Used
>Include Code Sequence Macro
52
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Physician(s) of Record (0008,1048) 3 Not Used
Physician(s) of Record Identification (0008,1049) 3 Not Used
Sequence
>Include Person Identification Macro
Name of Physician(s) Reading Study (0008,1060) 3 Not Used
Physician(s) Reading Study Identification (0008,1062) 3 Not Used
Sequence
>Include Person Identification Macro
Referenced Study Sequence (0008,1110) 3 Got from MWL.
>Include SOP Instance Reference Macro
Procedure Code Sequence (0008,1032) 3 Not Used
>Include Code Sequence Macro
53
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Step Description.
It will be set as Unknown Protocol for local
created exam.
Series Description (0008,103E) 3 The value equal protocol Name.
Operators' Name (0008,1070) 3 Not Used
Operator Identification Sequence (0008,1072) 3 Not Used
>Include Person Identification Macro
Referenced Performed Procedure Step (0008,1111) 3 Used.
Sequence
>Include SOP Instance Reference Macro
Related Series Sequence (0008,1250) 3 Not Used
>Study Instance UID (0020,000D) 1 Not Used
>Series Instance UID (0020,000E) 1 Not Used
>Purpose of Reference Code Sequence (0040,A170) 2 Not Used
>>Include Code Sequence Macro
Body Part Examined (0018,0015) 3 Used
Patient Position (0018,5100) 2C Not Used
Smallest Pixel Value in Series (0028,0108) 3 Not Used
Largest Pixel Value in Series (0028,0109) 3 Not Used
Request Attributes Sequence (0040,0275) 3 Used
>Requested Procedure ID (0040,1001) 1C Used
>Accession Number (0008,0050) 3 Not Used
>Study Instance UID (0020,000D) 3 Not Used
>Referenced Study Sequence (0008,1110) 3 Not Used
>> Include SOP Instance Reference Macro
>Requested Procedure Description (0032,1060) 3 Used
>Requested Procedure Code Sequence (0032,1064) 3 Used
>>Include Code Sequence Macro
Reason for the Requested Procedure (0040,1002) 3 Not Used
Reason for Requested Procedure Code (0040,100A) 3 Not Used
Sequence
>>Include Code Sequence Macro
>Scheduled Procedure Step ID (0040,0009) 1C Used
>Scheduled Procedure Step Description (0040,0007) 3 Used
>Scheduled Protocol Code Sequence (0040,0008) 3 Used
>>Include Code Sequence Macro
>>Protocol Context Sequence (0040,0440) 3 Not Used
>>>Include Content Item Macro
>>>Content Item Modifier Sequence (0040,0441) 3 Not Used
>>>>Include Content Item Macro
Performed Procedure Step ID (0040,0253) 3 Used
54
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Performed Procedure Step Start Date (0040,0244) 3 Used
Performed Procedure Step Start Time (0040,0245) 3 Used
Performed Procedure Step Description (0040,0254) 3 Used
Performed Protocol Code Sequence (0040,0260) 3 Not Used
>Include Code Sequence Macro
>>Protocol Context Sequence (0040,0440) 3 Not Used
>>>Include Content Item Macro
>>>Content Item Modifier Sequence (0040,0441) 3 Not Used
>>>>Include Content Item Macro
Comments on the Performed Procedure (0040,0280) 3 Not Used
Step
55
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
4.4.5 Image Entity Modules
56
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Presentation LUT Shape (2050,0020) 3 Not Used
57
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Contrast/Bolus Administration Route (0018,0014) 3 Not Used
Sequence
>Include Code Sequence Macro
>Additional Drug Sequence (0018,002A) 3 Not Used
>>Include Code Sequence Macro
Contrast/Bolus Volume (0018,1041) 3 Not Used
Contrast/Bolus Start Time (0018,1042) 3 Not Used
Contrast/Bolus Stop Time (0018,1043) 3 Not Used
Contrast/Bolus Total Dose (0018,1044) 3 Not Used
Contrast Flow Rate(s) (0018,1046) 3 Not Used
Contrast Flow Duration(s) (0018,1047) 3 Not Used
Contrast/Bolus Ingredient (0018,1048) 3 Not Used
Contrast/Bolus Ingredient Concentration (0018,1049) 3 Not Used
58
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Reference Image Sequence (0008,1140) 1C Not Used
>Referenced SOP Class UID (0008,1150) 1 Not Used
>Referenced SOP Class UID (0008,1155) 1 Not Used
>Referenced Frame Number (0008,1160) 1C Not Used
>Referenced Segment Number (0062,000B) 1C Not Used
>Purpose of Reference Code Sequence (0040,A170) 3 Not Used
>> Include Code Sequence Macro
Derivation Description (0008,2111) 3 Not Used
Acquisition Device Processing (0018,1400) 3 Not Used
Description
Frame Label Vector (0018,2002) 3 Not Used
Frame Dimension Pointer (0028,000A) 3 Not Used
Calibration Image (0050,0004) 3 Not Used
Define the values of Image Type (0008,0008) that may be sent and under what
circumstances, or refer to the more general description in the General Image Module.
59
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Pulsed: All Pulse xxx
Type of Filters (0018,1161) 3 Not Used
Intensifier Size (0018,1162) 3 Not Used
Field of View Shape (0018,1147) 3 Not Used
Field of View Dimension(s) (0018,1149) 3 Not Used
Imager Pixel Spacing (0018,1164) 3 Not Used
Pixel Spacing (0028,0030) 1C Not Used
Pixel Spacing Calibration Type (0028,0A02) 3 Not Used
Pixel Spacing Calibration Description (0028,0A04) 1C Not Used
Focal Spot (0018,1190) 3 Not Used
Image and Fluoroscopy Area Dose (0018,115E) 3 DAP of the exam.
Product
TABLE 4-13
OVERLAY PLANE MODULE ATTRIBUTES
Attribute Name Tag Type Attribute Description
Overlay Rows (60xx,0010) 1 Value is 1000.
Overlay Columns (60xx,0011) 1 Value is 1000.
Overlay Type (60xx,0040) 1 Value is G.
Overlay Origin (60xx,0050) 1 Value is 0\0.
60
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Overlay Bits Allocated (60xx,0100) 1 1
Overlay Bit Position (60xx,0102) 1 0
Overlay Data (60xx,3000) 1 Overlay pixel data
Overlay Description (60xx,0022) 3 Value is "Annotation/Measurement Graphic
Layer".
Overlay Subtype (60xx,0045) 3 Not Used
Overlay Label (60xx,1500) 3 Value is "Annotation/Measurement Graphic
Layer".
ROI Area (60xx,1301) 3 Not Used
ROI Mean (60xx,1302) 3 Not Used
ROI Standard Deviation (60xx,1303) 3 Not Used
61
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Specific Character Set (0008,0005) 1C Always set to ISO_IR 100( = Latin Alphabet
No. 1)
Instance Creation Date (0008,0012) 3 Not Used
Instance Creation Time (0008,0013) 3 Not Used
Instance Creator UID (0008,0014) 3 Not Used
Related General SOP Class UID (0008,001A) 3 Not Used
Original Specialized SOP Class UID (0008,001B) 3 Not Used
Coding Scheme Identification (0008,0110) 3 Not Used
Sequence
>Coding Scheme Designator (0008,0102) 1 Not Used
>Coding Scheme Registry (0008,0112) 1C Not Used
>Coding Scheme UID (0008,010C) 1C Not Used
>Coding Scheme External ID (0008,0114) 2C Not Used
>Coding Scheme Name (0008,0115) 3 Not Used
>Coding Scheme Version (0008,0103) 3 Not Used
>Coding Scheme Responsible (0008,0116) 3 Not Used
Organization
Timezone Offset From UTC (0008,0201) 3 Not Used
Contributing Equipment Sequence (0018,A001) 3 Not Used
>Purpose of Reference Code (0040,A170) 1 Not Used
Sequence
>>Include Code Sequence Macro
>Manufacturer (0008,0070) 1 Not Used
>Institution Name (0008,0080) 3 Not Used
>Institution Address (0008,0081) 3 Not Used
>Station Name (0008,1010) 3 Not Used
>Institutional Department Name (0008,1040) 3 Not Used
>Manufacturers Model Name (0008,1090) 3 Not Used
>Device Serial Number (0018,1000) 3 Not Used
>Software Versions (0018,1020) 3 Not Used
>Spatial Resolution (0018,1050) 3 Not Used
>Date of Last Calibration (0018,1200) 3 Not Used
>Time of Last Calibration (0018,1201) 3 Not Used
>Contribution DateTime (0018,A002) 3 Not Used
>Contribution Description (0018,A003) 3 Not Used
Instance Number (0020,0013) 3 Image number.
SOP Instance Status (0100,0410) 3 Not Used
SOP Authorization Date and Time (0100,0420) 3 Not Used
SOP Authorization Comment (0100,0424) 3 Not Used
Authorization Equipment (0100,0426) 3 Not Used
Certification Number
62
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
MAC Parameters Sequence (4FFE,0001) 3 Not Used
>MAC ID Number (0400,0005) 1 Not Used
>MAC Calculation Transfer Syntax (0400,0010) 1 Not Used
UID
>MAC Algorithm (0400,0015) 1 Not Used
>Data Elements Signed (0400,0020) 1 Not Used
Digital Signatures Sequence (FFFA,FFFA) 3 Not Used
>MAC ID Number (0400,0005) 1 Not Used
>Digital Signature UID (0400,0100) 1 Not Used
>Digital Signature DateTime (0400,0105) 1 Not Used
>Certificate Type (0400,0110) 1 Not Used
>Certificate of Signer (0400,0115) 1 Not Used
>Signature (0400,0120) 1 Not Used
>Certified Timestamp Type (0400,0305) 1C Not Used
>Certified Timestamp (0400,0310) 3 Not Used
>Digital Signature Purpose Code (0400,0401) 3 Not Used
Sequence
>>Include Code Sequence Macro
Encrypted Attributes Sequence (0400,0500) 1C Not Used
>Encrypted Content Transfer Syntax (0400,0510) 1 Not Used
UID
>Encrypted Content (0400,0520) 1 Not Used
Original Attributes Sequence (0400,0561) 3 Not Used
>Source of Previous Values (0400,0564) 2 Not Used
>Attribute Modification DateTime (0400,0562) 1 Not Used
>Modifying System (0400,0563) 1 Not Used
>Reason for the Attribute (0400,0565) 1 Not Used
Modification
>Modified Attributes Sequence (0400,0550) 1 Not Used
>>Any Attribute from the main data set that was modified or removed; may include Sequence Attributes and their
Items.
HL7 Structured Document Reference (0040,A390) 1C Not Used
Sequence
>Referenced SOP Class UID (0008,1150) 1 Not Used
>Referenced SOP Instance UID (0008,1155) 1 Not Used
>HL7 Instance Identifier (0040,E001) 1 Not Used
>Retrieve URI (0040,E010) 3 Not Used
Longitudinal Temporal Information (0028,0303) 3 Used.
Modified If Include Remove Patient Information then fix
to "MODIFIED".
63
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
64
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
5.1 INTRODUCTION
This section specifies the use of the DICOM X-Ray Radiation Dose SR IOD to represent
results produced and/or received by this implementation. Corresponding attributes are
conveyed using the module construct.
The Brivo OEC 715/785/865 maps DICOM Information Entities to local Information
Entities in the products database and user interface.
TABLE 5-1
MAPPING OF DICOM ENTITIES TO BRIVO OEC 715/785/865 ENTITIES
DICOM IE Brivo OEC 715/785/865 Entity
Patient Patient
Study Exam
Series Series
Equipment Equipment
Document Document
The X-Ray Radiation Dose Structured Report Information Object Definitions comprise
the modules of the following tables, plus Standard Extended and Private Attributes.
Standard Extended and Private Attributes are described in Section 5.5
The contents of the SR Document Content are constrained by the supported template, as
identified in Section 5.4.5.2.1.1. Standard, Standard Extended and Private Templates
are further described in Section 5.7.
TABLE 5-2
STRUCTURE REPORT IOD MODULES
Entity Name Module Name Usage Reference
Patient Patient Used 5.4.1.1
Specimen Not Used N/A
Identification
Clinical Trial Subject Not Used N/A
Study General Study Used 5.4.2.1
Patient Study Used 5.4.2.2
Clinical Trial Study Not Used N/A
65
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Series SR Document Series Used 5.4.3.1
Clinical Trial Series Not Used N/A
Frame Of Synchronization Not Used N/A
Reference
Equipment General Equipment Used 5.4.4.1
Document SR Document General Used 5.4.5.1
SR Document Content Used 5.4.5.2
SOP Common Used 5.4.5.3
The following modules are included to convey Enumerated Values, Defined Terms, and
Optional Attributes supported and/or expected. Type 1 & Type 2 Attributes are also
included for completeness and to define what values they may take and where these
values are obtained from when generating the instance as well as what are the expected
values when loading such instance. It should be noted that they are the same ones as
defined in the DICOM Standard Part 3 (Information Object Definitions). Also note that
Attributes not present in tables are not supported.
66
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Include SOP Instance Reference Macro
Patient's Birth Time (0010,0032) 3 Not Used
Other Patient IDs (0010,1000) 3 Not Used
Other Patient IDs Sequence (0010,1002) 3 Not Used
>Patient ID (0010,0020) 1 Not Used
>Issuer of Patient ID (0010,0021) 1 Not Used
>Type of Patient ID (0010,0022) 1 Not Used
Other Patient Names (0010,1001) 3 Not Used
Ethnic Group (0010,2160) 3 Not Used
Patient Comments (0010,4000) 3 May be entered from user interface or got from
Worklist
From user interface truncated to 512 bytes.
From worklist truncated to 10240 bytes.
Patient Species Description (0010,2201) 1C Not Used
Patient Species Code Sequence (0010,2202) 1C Not Used
>Include Code Sequence Macro
Patient Breed Description (0010,2292) 2C Not Used
Patient Breed Code Sequence (0010,2293) 2C Not Used
>Include Code Sequence Macro
Breed Registration Sequence (0010,2294) 2C Not Used
>Breed Registration Number (0010,2295) 1 Not Used
>Breed Registry Code Sequence (0010,2296) 1 Not Used
>>Include Code Sequence Macro
Responsible Person (0010,2297) 2C Not Used
Responsible Person Role (0010,2298) 1C Not Used
Responsible Organization (0010,2299) 2C Not Used
Patient Identity Removed (0012,0062) 3 Used.
If Patient Information is removed then set to
"YES", otherwise set to "NO"
De-identification Method (0012,0063) 1C Used.
If Patient Information is removed. Set to
"Limited Data Set" in this case
De-identification Method Code Sequence (0012,0064) 1C Not Used
>Include Code Sequence Macro
67
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
5.4.2 Study Entity Modules
68
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Admitting Diagnoses Description (0008,1080) 3 Not Used
Admitting Diagnoses Code Sequence (0008,1084) 3 Not Used
>Include Code Sequence Macro
Patient's Age (0010,1010) 3 Not Used
Patient's Size (0010,1020) 3 Used
Patient's Weight (0010,1030) 3 Used
Occupation (0010,2180) 3 Not Used
Additional Patients History (0010,21B0) 3 Not Used
Admission ID (0038,0010) 3 Not Used
Issuer of Admission ID (0038,0011) 3 Not Used
Service Episode ID (0038,0060) 3 Not Used
Issuer of Service Episode ID (0038,0061) 3 Not Used
Service Episode Description (0038,0062) 3 Not Used
Patients Sex Neutered (0010,2203) 2C Not Used
69
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
5.4.4 Equipment Entity Modules
70
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Author Observer Sequence (0040,A078) 3 Used
>Include Identified Person or Device Macro
Participant Sequence (0040,A07A) 3 Not Used
>Participation Type (0040,A080) 1 Not Used
>Participation DateTime (0040,A082) 2 Not Used
>Include Identified Person or Device Macro
Custodial Organization Sequence (0040,A07C) 3 Not Used
>Institution Name (0008,0080) 2 Not Used
>Institution Code Sequence (0008,0082) 2 Not Used
>>Include Code Sequence Macro
Predecessor Documents Sequence (0040,A360) 1C Not Used
>Include Hierarchical SOP Instance Reference Macro
Identical Documents Sequence (0040,A525) 1C Not Used
>Include Hierarchical SOP Instance Reference Macro
Referenced Request Sequence (0040,A370) 1C Used
>Study Instance UID (0020,000D) 1 The product will adopt the Study Instance UID
which is returned in the MWL response. If there
is no Study Instance UID returned from MWL
server, then new Study Instance UID will be
created by system during exam creation.
>Referenced Study Sequence (0008,1110) 2 Set as empty
>>Include SOP Instance Reference Macro
>Accession Number (0008,0050) 2 The product will adopt the Accession Number
which is returned in the MWL response. If there
is no Accession Number returned from MWL
server, then the value will keep empty.
For local created exam, user could input value
from screen during exam creation.
>Placer Order Number/Imaging Service (0040,2016) 2 Value is empty.
Request
>Filler Order Number/Imaging Service (0040,2017) 2 Value is empty.
Request
>Requested Procedure ID (0040,1001) 2 This value could got from MWL response. If not
contained in MWL response, then keep empty as
default.
>Requested Procedure Description (0032,1060) 2 This value could got from MWL response. If not
contained in MWL response, then keep empty as
default.
>Requested Procedure Code Sequence (0032,1064) 2 Value is empty.
>>Include Code Sequence Macro
>Reason for the Requested Procedure (0040,1002) 3 Not Used
>Reason for Requested Procedure Code (0040,100A) 3 Not Used
Sequence
>>Include Code Sequence Macro
71
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Performed Procedure Code Sequence (0040,A372) 2 Empty.
>Include Code Sequence Macro
Current Requested Procedure Evidence (0040,A375) 1C With the value(s) from the system of all saved
Sequence images for the exam.
This is the list of all images whether they have
been stored to PACS or not.
>Include Hierarchical SOP Instance Reference Macro
Pertinent Other Evidence Sequence (0040,A385) 1C Not Used
> Include Hierarchical SOP Instance Reference Macro
Referenced Instance Sequence (0008,114A) 1C Not Used
>Include SOP Instance Reference Macro
>Purpose of Reference Code Sequence (0040,A170) 1 Not Used
>>Include Code Sequence Macro
72
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
TABLE 5-10
SR ROOT TEMPLATES
SOP Class Template ID Template Name Use
X-Ray Radiation Dose 10001 X-Ray Radiation Dose Create/Display
SR
Refer to section 5.7 for a detailed description of the supported templates.
73
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Institution Name (0008,0080) 3 Not Used
>Institution Address (0008,0081) 3 Not Used
>Station Name (0008,1010) 3 Not Used
>Institutional Department Name (0008,1040) 3 Not Used
>Manufacturers Model Name (0008,1090) 3 Not Used
>Device Serial Number (0018,1000) 3 Not Used
>Software Versions (0018,1020) 3 Not Used
>Spatial Resolution (0018,1050) 3 Not Used
>Date of Last Calibration (0018,1200) 3 Not Used
>Time of Last Calibration (0018,1201) 3 Not Used
>Contribution DateTime (0018,A002) 3 Not Used
>Contribution Description (0018,A003) 3 Not Used
Instance Number (0020,0013) 3 Image number.
SOP Instance Status (0100,0410) 3 Not Used
SOP Authorization Date and Time (0100,0420) 3 Not Used
SOP Authorization Comment (0100,0424) 3 Not Used
Authorization Equipment (0100,0426) 3 Not Used
Certification Number
MAC Parameters Sequence (4FFE,0001) 3 Not Used
>MAC ID Number (0400,0005) 1 Not Used
>MAC Calculation Transfer Syntax (0400,0010) 1 Not Used
UID
>MAC Algorithm (0400,0015) 1 Not Used
>Data Elements Signed (0400,0020) 1 Not Used
Digital Signatures Sequence (FFFA,FFFA) 3 Not Used
>MAC ID Number (0400,0005) 1 Not Used
>Digital Signature UID (0400,0100) 1 Not Used
>Digital Signature DateTime (0400,0105) 1 Not Used
>Certificate Type (0400,0110) 1 Not Used
>Certificate of Signer (0400,0115) 1 Not Used
>Signature (0400,0120) 1 Not Used
>Certified Timestamp Type (0400,0305) 1C Not Used
>Certified Timestamp (0400,0310) 3 Not Used
>Digital Signature Purpose Code (0400,0401) 3 Not Used
Sequence
>>Include Code Sequence Macro
Encrypted Attributes Sequence (0400,0500) 1C Not Used
>Encrypted Content Transfer Syntax (0400,0510) 1 Not Used
UID
>Encrypted Content (0400,0520) 1 Not Used
Original Attributes Sequence (0400,0561) 3 Not Used
74
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Source of Previous Values (0400,0564) 2 Not Used
>Attribute Modification DateTime (0400,0562) 1 Not Used
>Modifying System (0400,0563) 1 Not Used
>Reason for the Attribute (0400,0565) 1 Not Used
Modification
>Modified Attributes Sequence (0400,0550) 1 Not Used
>>Any Attribute from the main data set that was modified or removed; may include Sequence Attributes and their
Items.
HL7 Structured Document Reference (0040,A390) 1C Not Used
Sequence
>Referenced SOP Class UID (0008,1150) 1 Not Used
>Referenced SOP Instance UID (0008,1155) 1 Not Used
>HL7 Instance Identifier (0040,E001) 1 Not Used
>Retrieve URI (0040,E010) 3 Not Used
Longitudinal Temporal Information (0028,0303) 3 Used.
Modified If Include Remove Patient Information then fix
to "MODIFIED".
The Product supports coded terminology using Standard Extended, Private, and
Configurable Context Groups defined in the following sections.
75
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
5.7.1 Standard Templates
The Product supports the following standard templates for SOP Instances created by this
product.
76
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
14 > CONTAINS CODE EV (113854, 1-n M (113856, DCM,
DCM, Source of "Automated Data
Dose Collection")
Information)
77
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Mammography 40010, SRT,
X-Ray Dose Mammography)
11 > CONTAI INCLUDE DTID (1021) 1 MC Required if the $DeviceProcedureRole
NS Device irradiating device is = EV (113859, DCM,
Participant not the recording Irradiating Device)
device and the dose
was accumulated on a
single device.
78
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
11 > CONTAI NUM EV (122130, 1 MC IFF TID (10001) Row Units = EV (Gy.m2, UCUM,
NS DCM, Dose Area 2 = (113704, DCM, "Gy.m2")
Product Projection X-Ray)
12 > CONTAI NUM EV (111631, 1 MC IFF TID (10001) Row Not Used
NS DCM, Average 2 = (P5-40010, SRT,
Glandular Dose) Mammography)
13 > CONTAI NUM EV (113738, 1 MC IFF TID (10001) Row Units = EV (Gy, UCUM, Gy)
NS DCM, Dose 2 = (113704, DCM, Dose applied at the Reference
(RP)) Projection X-Ray) Point (RP)
AND any of the
values of TID (10001)
Row 14 are not
(113858, DCM,
MPPS Content)
14 > CONTAI NUM EV (111636, 1 MC IFF TID (10001) Row Not Used
NS DCM, Entrance 2 = (P5-40010, SRT,
Exposure at RP) Mammography)
15 > CONTAI NUM EV (112011, 1 UC XOR Row 19 Not Used
NS DCM, Positioner
Primary Angle)
16 > CONTAI NUM EV ( 112012, 1 UC XOR Row 19 Not Used
NS DCM, Positioner
Secondary
Angle)
17 > CONTAI NUM EV (113739, 1 UC IFF Row 4 value = Not Used
NS DCM, Positioner (113613, DCM,
Primary End Rotational
Angle Acquisition)
18 > CONTAI NUM EV (113740, 1 UC IFF Row 4 value = Not Used
NS DCM, Positioner (113613, DCM,
Secondary End Rotational
Angle Acquisition)
19 > CONTAI NUM EV (113770, 1 UC XOR Rows 15,16 Not Used
NS DCM, Column
Angulation)
20 > CONTAI NUM EV (113790, 1 U Not Used
NS DCM, Collimated
Field Area
21 > CONTAI CONTAIN EV (113771, 1-n U Not Used
NS ER DCM, X-Ray
Filters)
22 >> CONTAI CODE EV (113772, 1 U Not Used
NS DCM, X-Ray
Filter Type)
23 >> CONTAI CODE EV (113757, 1 U Not Used
NS DCM, X-Ray
Filter Material)
24 >> CONTAI NUM EV (113758, 1 U Not Used
NS DCM, X-Ray
Filter Thickness
Minimum)
79
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
25 >> CONTAI NUM EV (113773, 1 U Not Used
NS DCM, X-Ray
Filter Thickness
Maximum)
26 > CONTAI CODE EV (113732, 1 UC IFF Row 4 value = 1. The value of code (113630,
NS DCM, Fluoro (P5-06000, SRT, DCM, "Continuous") for
Mode) Fluoroscopy) continuous mode irradiation
events.
2. The value of code (113631,
DCM, "Pulsed") for Pulsed
mode irradiation events.
27 > CONTAI NUM EV (113791, 1 MC IFF Row 26 value = Units = EV ({pulse}/s, UCUM,
NS DCM, Pulse (113631, DCM, pulse/s)
Rate) Pulsed) For Pulsed mode irradiation
events.
28 > CONTAI NUM EV (113768, 1 MC IFF Row 26 value = Units = EV (1, UCUM, no
NS DCM, Number of (113631, DCM, units)
Pulses) Pulsed) 1. For Pulsed mode irradiation
events.
2. Estimated:
a. Manual: exposure time/pulse
width
b. Auto: exposure time/pulse
width (margin of error might be
bigger than manual)
c. Pulse Width = 49.5ms
29 >> HAS CODE EV (121401, 1 MC IFF count of pulses in EV (R-10260, SRT,
CONCEP DCM, Row 28 is estimated Estimated)
T MOD Derivation)
30 > CONTAI NUM EV (113733, 1-n U Units = EV (kV, UCUM, kV)
NS DCM, KVP) For Pulsed/Continuous mode
irradiation events.
31 > CONTAI NUM EV (113734, 1-n U Units = EV (mA, UCUM,
NS DCM, X-Ray mA)
Tube Current)
32 > CONTAI NUM EV (113735, 1 U Units = EV (ms, UCUM, ms)
NS DCM, Exposure
Time)
33 > CONTAI NUM EV (113793, 1-n U Units = EV (ms, UCUM, ms)
NS DCM, Pulse Value is 49.5ms
Width)
34 > CONTAI NUM EV (113736, 1-n U Not Used
NS DCM,
Exposure)
35 > CONTAI NUM EV (113766, 1 U Not Used
NS DCM, Focal Spot
Size)
36 > CONTAI NUM EV (113742, 1 U Not Used
NS DCM, Irradiation
80
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Duration)
37 > CONTAI NUM EV (113767, 1 U Units = EV (mA, UCUM,
NS DCM, Average mA)
X-Ray Tube The same with above X-Ray
Current) Tube Current
38 > CONTAI CODE EV (113745, 1 U Not Used
NS DCM, Patient
Table
Relationship)
39 > CONTAI CODE EV (113743, 1 U Not Used
NS DCM, Patient
Orientation)
40 >> HAS CODE EV (113744, 1 M Not Used
CONCEP DCM, Patient
T MOD Orientation
Modifier)
41 > CONTAI NUM DCID (10008) 1-n U Not Used
NS Dose Related
Distance
Measurements
42 > CONTAI NUM EV (113754, 1 U Not Used
NS DCM, Table
Head Tilt Angle)
43 > CONTAI NUM EV (113755, 1 U Not Used
NS DCM, Table
Horizontal
Rotation Angle)
44 > CONTAI NUM EV (113756, 1 U Not Used
NS DCM, Table
Cradle Tilt
Angle)
45 > CONTAI CODE EV (123014 , 1 M 1. the code from the system
NS DCM, (Target using DICOM Part 16 Context
Region) ID 4031 values and supporting
the code (T-D0001, SRT,
"Topography unknown") for an
'Unknown' value which is
SNOMED term.
2. User could set this value in
EM->Additional Screen
Default is (T-D0001, SRT,
"Topography unknown")
46 > CONTAI CODE EV (111632, 1 U Not Used
NS DCM, Anode
Target Material
47 > CONTAI NUM EV (111633, 1 U Not Used
NS DCM,
Compression
Thickness)
48 > CONTAI NUM EV (111634, 1 U Not Used
NS DCM, Half Value
81
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Layer)
82
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
3 NUM EV (113726, 1 MC IFF TID(10003) Row Units = EV (Gy.m2,
DCM, Fluoro 4 value = (P5-06000, UCUM, Gy.m2)
Dose Area SRT, Fluoroscopy) The numeric value of
Product Total) for at least one Total Radioscopic Dose
irradiation event Area Product.
4 NUM EV (113728, 1 MC IFF TID(10003) Row Units = EV (Gy,
DCM, Fluoro 4 value = (P5-06000, UCUM, Gy)
Dose (RP) SRT, Fluoroscopy) The numeric value of
Total) for at least one Cumulative Radioscopic
irradiation event AND Air Kerma.
any of the values of
TID (10001) Row 14
are not (113858,
DCM, MPPS
Content).
5 NUM EV (113730, 1 MC IFF TID(10003) Row Units = EV (s, UCUM,
DCM, Total 4 value = (P5-06000, s)
Fluoro Time) SRT, Fluoroscopy) The numeric value of
for at least one Total Radioscopic time
irradiation event. to the second.
6 NUM EV (113727, 1 M Units = EV (Gy.m2,
DCM, UCUM, Gy.m2)
Acquisition The numeric value of
Dose Area Total Radiographic
Product Total) Dose Area Product.
7 NUM EV (113729, 1 MC IF any of the values of Units = EV (Gy,
DCM, TID (10001) Row 14 UCUM, Gy)
Acquisition are not (113858, The numeric value of
Dose (RP) DCM, MPPS Cumulative
Total) Content). May be Radiographic Air
present otherwise. kerma.
8 NUM EV (113855, 1 M Units = EV (s, UCUM,
DCM, Total s)
Acquisition The numeric value of
Time) Total Radiographic
pedal time in
milliseconds.
9 NUM EV (113731, 1 U Units = EV (1, UCUM,
DCM, Total no units)
Number of The numeric value of
Radiographic Total Film and Digital
Frames) Spot Frames.
10 CODE EV (113780, 1 MC IF Row 2, Row 4 or (113861, DCM, "30cm
DCM, Row 7 is present and in Front of Image Input
Reference Point Row 11 is not present. Surface").
Definition)
11 TEXT EV (113780, 1 MC IF Row 2, Row 4 or Not Used
DCM, Row 7 is present and
Reference Point Row 10 is not present.
Definition)
83
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
5.7.1.5 Template ID 1002 Observer Context
TID 1002
OBSERVER CONTEXT
Type: Non-Extensible Order: Significant
NL Rel with VT Concept Name VM Req Condition Value Set Constraint
Parent Type
1 HAS OBS CODE EV (121005,DCM, 1 MC IF Observer type is device (121007, DCM, "Device")
CONTEX Observer Type)
T
2 HAS OBS INCLUDE DTID (1003) Person 1 MC IFF Row 1 value = Not Used
CONTEX observer identifying (121006,DCM, Person)
T attributes or Row 1 is absent
3 HAS OBS INCLUDE DTID (1004) Device 1 MC IFF Row 1 value =
CONTEX observer identifying (121007,DCM, Device)
T attributes
TID 1004
DEVICE OBSERVER IDENTIFYING ATTRIBUTES
Type: Extensible Order: Significant
NL Rel with VT Concept Name VM Req Condition Value Set Constraint
Parent Type
1 UIDREF EV (121012,DCM, 1 M System Generated
Device Observer UID)
2 TEXT EV (121013,DCM, 1 U Defaults to value of
Device Observer Name) Station Name (0008,1010)
in General Equipment
Module.
If 'Station Name' is blank
the value for Local AE
Title shall be used.
3 TEXT EV (121014,DCM, 1 U Defaults to value of
Device Observer Manufacturer (0008,0070)
Manufacturer) in General Equipment
Module
4 TEXT EV (121015,DCM, 1 U Defaults to value of
Device Observer Model Manufacturers Model
Name) Name (0008,1090) in
General Equipment
Module
5 TEXT EV (121016,DCM, 1 U Defaults to value of
Device Observer Serial Device Serial Number
Number) (0018,1000) in General
Equipment Module
6 TEXT EV (121017,DCM, 1 U Hospital Name.
Device Observer
Physical Location during
observation)
7 CODE EV (113876, DCM, 1-n U (113859, DCM,
Device Role in "Irradiating Device")
84
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Procedure)
85
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
6.1 INTRODUCTION
This section specifies the use of the DICOM Modality Worklist Information Model used
to organize data and against which a Modality Worklist Query will be performed.
TABLE 6-1
MAPPING OF DICOM ENTITIES TO BRIVO OEC 715/785/865 ENTITIES
DICOM Brivo OEC 715/785/865 Entity
Scheduled Procedure Step Exam
Requested Procedure Exam
Imaging Service Request Exam
Visit Not Applicable
Patient Patient
See DICOM PS 3.3 and PS 3.4 for a complete definition of the entities, modules, and
attributes.
86
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
TABLE 6-2
MODALITY WORKLIST INFORMATION MODEL MODULES
Entity Name Module Name Reference
Scheduled Procedure Step SOP Common 6.4.1.1
Scheduled Procedure Step 6.4.1.2
Requested Procedure Requested Procedure 6.4.2.1
Imaging Service Request Imaging Service Request 6.4.3.1
Visit Visit Identification 6.4.4.1
Visit Status 6.4.4.2
Visit Relationship 6.4.4.3
Visit Admission 6.4.4.4
Patient Patient Relationship 6.4.5.1
Patient Identification 6.4.5.2
Patient Demographic 6.4.5.3
Patient Medical 6.4.5.4
As a Query SCU, it will similarly accept response items with ISO_IR 100 values of
Specific Character Set or empty value of Specific Character Set (absence of (0008,0005)
which mean usage of ASCII Chars only).
The product user interface will allow the user to enter characters from the console
keyboard that is within ISO 8859-1 characters.
87
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
6.4.1.2 Scheduled Procedure Step Module
TABLE 6-4
SCHEDULED PROCEDURE STEP MODULE ATTRIBUTES
Attribute Name Tag Expected Expected Mapped into Note
Matching Returned Instance /
Key Type Key Type MPPS
Scheduled Procedure (0040,0100) R 1 No/No
Step Sequence
>Scheduled Station (0040,0001) R 1 No/No No Matching is performed.
AE Title
>Scheduled (0040,0002) R 1 No/No Using the date configured from the DICOM
Procedure Step Start query filter screen, the following matching can
Date be done:
YYYY- year, MM- month, DD- day.
Range matching can also be done
YYYYMMDD-YYYYMMDD - all matches
included between those dates. The dash is
important
-YYYYMMDD - all matches prior to and
including this date. The beginning dash is
important.
YYYYMMDD- all matches after and including
this date. The ending dash is important.
Single matching where the dates entered is
equal.
The data entered on the user interface is passed
to the Server as entered; if it does not work see
the Server DICOM conformance statement.
>Scheduled (0040,0003) R 1 No/No No Matching is performed.
Procedure Step Start
Time
>Scheduled (0040,0004) O 3 No/No Not Requested
Procedure Step End
Date
>Scheduled (0040,0005) O 3 No/No Not Requested
Procedure Step End
Time
>Modality (0008,0060) R 1 No/No Matching is supported and is configured by the
user interface. This value could be set to
RF/XA/CR/OT/ALL. ALL means any
modality.
Note 1: Send the field as empty to request
universal matching when "ALL" is selected
from the UI.
Note 2:User could only select RF,XA,CR or
OT as Modality Type during image archiving.
88
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Scheduled (0040,0006) R 2* Yes/Yes Matching is performed. The data entered on the
Performing user interface is passed to the server as it is.
Physician's Name Single matching is supported for this data
element.
This value is mapped into (0008, 1050) in the
instance.
This value is mapped into (0008, 1050) in the
MPPS.
Supports 5 different components delimited by
^. Supports a maximum length of 64 bytes
including the delimiter.
>Scheduled (0040,0007) O 1C * Yes/Yes No Matching is performed.
Procedure Step This value is mapped into (0040,0275)->(0040,
Description 0007) & Protocol Name(0018,1030) & Series
Description(0008,103E) in the image instance.
This value is mapped into Series
Description(0008,103E) & EV(125203, DCM,
Acquisition Protocol) in the RDSR instance.
This value is mapped into (0040,0270)->(0040,
0007) & Protocol Name(0018,1030) & Series
Description(0008,103E) in the MPPS.
>Scheduled Station (0040,0010) O 2 No/No No Matching is performed.
Name
>Scheduled (0040,0011) O 2 No/No No Matching is performed.
Procedure Step
Location
>Scheduled Protocol (0040,0008) O 1C Yes/Yes No Matching is performed.
Code Sequence This value is mapped into (0040,0275)-> (0040,
0008) in the instance.
This value is mapped into (0040,0270)-> (0040,
0008) in the MPPS.
>>Code Value (0008,0100) O 1 Yes/Yes No Matching is performed.
>>Coding Scheme (0008,0102) O 1 Yes/Yes No Matching is performed.
Designator
>>Coding Scheme (0008,0103) O 3 Yes/Yes No Matching is performed.
Version
>>Code Meaning (0008,0104) O 3* Yes/Yes No Matching is performed.
>Pre-Medication (0040,0012) O 2C No/No No Matching is performed.
>Scheduled (0040,0009) O 1* Yes/Yes No Matching is performed.
Procedure Step ID This value is mapped into (0040,0275)-> (0040,
0009) in the instance.
This value is mapped into (0040,0270)-> (0040,
0009) in the MPPS.
>Requested Contrast (0032,1070) O 2C Yes/No No Matching is performed.
Agent This value is mapped into (0018, 0010) in the
instance.
89
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Scheduled (0040,0020) O 3 No/No Not Requested
Procedure Step Status
>Comments on the (0040,0400) O 3 No/No Not Requested
Scheduled Procedure
Step
Note: * in the Expected Return Key Type column indicates that this information is
displayed on screen, if available
90
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Coding Scheme (0008,0103) O 3 Yes/Yes No Matching is performed.
Version
>Code Meaning (0008,0104) O 3 Yes/Yes No Matching is performed.
Study Instance UID (0020,000D) O 1 Yes/Yes The product will adopt the Study Instance UID
which is returned in the MWL response.
If there is no Study Instance UID returned from
MWL server, then new Study Instance UID will
be system created.
Study Date (0008,0020) O 3 No/No Not Requested
Study Time (0008,0030) O 3 No/No Not Requested
Referenced Study (0008,1110) O 2 Yes/Yes This value is mapped into (0008,1110) in the
Sequence image instance.
This value is mapped into (0008,1110) in the
RDSR.
This value is mapped into (0008, 1110) in the
MPPS.
>Referenced SOP (0008,1150) O 1C Yes/Yes
Class UID
>Referenced SOP (0008,1155) O 1C Yes/Yes
Instance UID
Requested Procedure (0040,1003) O 2 No/No No Matching is performed.
Priority
Patient Transport (0040,1004) O 2 No/No Not Requested
Arrangements
Requested Procedure (0040,1005) O 3 No/No Not Requested
Location
Confidentiality Code (0040,1008) O 3 No/No Not Requested
Reporting Priority (0040,1009) O 3 No/No Not Requested
Names of Intended (0040,1010) O 3 No/No No Matching is performed.
Recipients of Results
Reason for the (0040,1002) O 3 No/No No Matching is performed.
Requested Procedure
Requested Procedure (0040,1400) O 3 No/No Not Requested
Comments
Note: * in the Expected Return Key Type column indicates that this information is
displayed on screen, if available
91
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
6.4.3 Common Imaging Service Request Entity Modules
Note: * in the Expected Return Key Type column indicates that this information is
displayed on screen, if available
92
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
6.4.4 Common visit Entity Modules
93
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Note: * in the Expected Return Key Type column indicates that this information is
displayed on screen, if available
94
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Military Rank (0010,1080) O 3 No/No Not Requested
Branch of Service (0010,1081) O 3 No/No Not Requested
Country of Residence (0010,2150) O 3 No/No Not Requested
Region of Residence (0010,2152) O 3 No/No Not Requested
Patients Telephone (0010,2154) O 3 No/No Not Requested
Numbers
Ethnic Group (0010,2160) O 3 No/No Not Requested
Patient's Religious (0010,21F0) O 3 No/No Not Requested
Preference
Patient Comments (0010,4000) O 3* Yes/No No Matching is performed.
Mapped into image and RDSR instance .
Truncated to 10240 bytes, end truncated
Note: * in the Expected Return Key Type column indicates that this information is
displayed on screen, if available
95
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
7.1 INTRODUCTION
This section specifies the use of the DICOM Modality Performed Procedure Step
information to be communicated to the Hospital/Radiology information system.
This feature works in conjunction with DICOM Modality Worklist feature, if installed.
However the conformance of this feature is independent of Modality Worklist feature.
For information on conformance of Modality Worklist feature to DICOM standard please
refer to the appropriate section in this document.
2. Supports one-to-multiple relationship: Append Case with new PPS based on Original
SPS
See DICOM PS 3.3 and PS 3.4 for a complete definition of the entities, modules, and
attributes.
TABLE 7-1
MODALITY PERFORMED PROCEDURE STEP MODULES
Module Name Reference
SOP Common 7.4.1
Performed Procedure Step Relationship 7.4.2
Performed Procedure Step Information 7.4.3
Image Acquisition Results 7.4.4
Radiation Dose 7.4.5
Billing and Material Management Codes 7.4.6
96
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
7.4.1 SOP Common Module
TABLE 7-2
SOP COMMON MODULE ATTRIBUTES
Attribute Name Tag Type for Type for Use
SCU SCU
N- N-SET
CREATE
Specific Character Set (0008,0005) 1C 1C ISO_IR 100
97
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Scheduled Procedure Step (0040,0009) 2, User input. If user does 2, Got from MWL.
ID not provide input values,
then empty
>Scheduled Procedure Step (0040,0007) 2, User input. If user does 2, Got from MWL.
Description not provide input values,
then empty
>Scheduled Protocol Code (0040,0008) 2, empty 2, Got from MWL
Sequence
>>Code Value (0008,0100) 1, Not sent 1, Got from MWL
>>Coding Scheme (0008,0102) 1, Not sent 1, Got from MWL
Designator
>>Coding Scheme Version (0008,0103) 3, Not sent 3, Got from MWL
>>Code Meaning (0008,0104) 3, Not sent 3, Got from MWL
Patient's Name (0010,0010) 2, User input. If user does 2, Got from MWL.
not provide input values,
then UNNAMED xxxxx
should be assigned to Last
Name (xxxxx is a 5-
digital number)
Patient ID (0010,0020) 2, User input. If user does 2, Got from MWL.
not provide input values,
then empty
Issuer of Patient ID (0010,0021) 3, empty 3, Got from MWL.
Patient's Birth Date (0010,0030) 2, User input. If user does 2, Got from MWL.
not provide input values,
then empty
Patient's Sex (0010,0040) 2, User input. If user does 2, Got from MWL.
not provide input values,
then empty
Referenced Patient Sequence (0008,1120) 2, empty. 2, empty.
>Referenced SOP Class UID (0008,1150) 1, Not sent 1, Not sent
>Referenced SOP Instance (0008,1155) 1, Not sent 1, Not sent
UID
Admission ID (0038,0010) 3, Not sent 3, Not sent
Issuer of Admission ID (0038,0011) 3, Not sent 3, Not sent
Service Episode ID (0038,0060) 3, Not sent 3, Not sent
Issuer of Service Episode ID (0038,0061) 3, Not sent 3, Not sent
Service Episode Description (0038,0062) 3, Not sent 3, Not sent
98
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
7.4.3 Performed Procedure Step Information Module
TABLE 7-4
PERFORMED PROCEDURE STEP INFORMATION MODULE ATTRIBUTES
Attribute Name Tag Type for Type for Use
SCU SCU
N- N-SET
CREATE
Performed Procedure Step ID (0040,0253) 1 - If Scheduled Procedure Step ID is empty then
this value is current timestamp, else this value is
Scheduled Procedure Step ID.
Performed Station AE Title (0040,0241) 1 - Got from configure file.
Performed Station Name (0040,0242) 2 - Got from configure file.
Performed Location (0040,0243) 2 - Got from configure file.
The value is Hospital Name.
Performed Procedure Step (0040,0244) 1 - Same as exam start date
Start Date
Performed Procedure Step (0040,0245) 1 - Same as exam start time
Start Time
Performed Procedure Step (0040,0252) 1 3 When PPS start ( N-CREATE) message is sent,
Status this element will have the value IN
PROGRESS .
When PPS end ( N-SET) message is sent, this
element will have either COMPLETED or
DISCONTINUED based on user selection.
Performed Procedure Step (0040,0254) 2 3 This value is Scheduled Procedure Step
Description Description
Performed Procedure Type (0040,0255) 2 3 From "Requested Procedure Description" on
Description screen which is come from MWL.
Procedure Code Sequence (0008,1032) 2 3 Always set as Empty for local created exam.
When using MWL, this value is Requested
Procedure Code Sequence.
>Code Value (0008,0100) 1 1 Only when using MWL sent.
>Coding Scheme Designator (0008,0102) 1 1 Only when using MWL sent.
>Coding Scheme Version (0008,0103) 3 3 Only when using MWL sent.
>Code Meaning (0008,0104) 3 3 Only when using MWL sent.
Performed Procedure Step (0040,0250) 2 3 Date when PPS end got installed
End Date
Performed Procedure Step (0040,0251) 2 3 Time when PPS end got installed
End Time
Comments on the Performed (0040,0280) 3 3 Not sent
Procedure Step
Performed Procedure Step (0040,0281) 3 3 Used when exam is DISCONTINUED.
Discontinuation Reason Code Provide a way let user to select one
Sequence
>Code Value (0008,0100) 1 1 Follow CID 9301
>Coding Scheme Designator (0008,0102) 1 1 Follow CID 9301
99
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Coding Scheme Version (0008,0103) 3 3 Follow CID 9301
>Code Meaning (0008,0104) 3 3 Follow CID 9301
100
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>>Referenced SOP Class UID (0008,1150) 1 1 Referenced instance SOP Class UID.
>>Referenced SOP Instance (0008,1155) 1 1 Referenced instance SOP instance UID.
UID
101
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
7.4.6 Billing and Material Management Codes Module
TABLE 7-7
BILLING AND MATERIAL MANAGEMENT CODES MODULE ATTRIBUTES
Attribute Name Tag Type for Type for Use
SCU SCU
N- N-SET
CREATE
Billing Procedure Step (0040,0320) 3 3 Not sent
Sequence
> Code Sequence Macro (0020,0010) 3 3
Film Consumption Sequence (0040,0321) 3 3 Not sent
>Number of Films (2100,0170) 3 3 Not sent
>Medium Type (2000,0030) 3 3 Not sent
>Film Size ID (2010,0050) 3 3 Not sent
Billing Supplies and Devices (0040,0324) 3 3 Not sent
Sequence
>Billing Item Sequence (0040,0296) 3 3 Not sent
>> Code Sequence Macro (0040,0303) 3 3 Not sent
>Quantity Sequence (0040,0293) 3 3 Not sent
>>Quantity (0040,0294) 3 3 Not sent
>>Measuring Units Sequence (0040,0295) 3 3 Not sent
>>> Code Sequence Macro (0018,115A) 3 3 Not sent
The Product supports the following attributes, not specified in the MPPS IOD, in SOP
Instances as Type 3 data elements.
TABLE 7-8
STANDARD EXTENDED ATTRIBUTES
Attribute Name Tag Use
N/A N/A N/A
102
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
7.6.1.1 Context ID 9301 Modality PPS Discontinuation Reasons
Context ID 9301
Modality PPS Discontinuation Reasons
Type: Extensible Version: 20090616
Coding Scheme Code Value Code Meaning
Designator (0008,0100) (0008,0104)
(0008,0102)
DCM 110500 Doctor cancelled procedure
DCM 110501 Equipment failure
DCM 110502 Incorrect procedure ordered
DCM 110503 Patient allergic to media/contrast
DCM 110504 Patient died
DCM 110505 Patient refused to continue procedure
DCM 110506 Patient taken for treatment or surgery
DCM 110507 Patient did not arrive
DCM 110508 Patient pregnant
DCM 110509 Change of procedure for correct charging
DCM 110510 Duplicate order
DCM 110511 Nursing unit cancel
DCM 110512 Incorrect side ordered
DCM 110513 Discontinued for unspecified reason
DCM 110514 Incorrect worklist entry selected
DCM 110515 Patient condition prevented continuing
DCM 110516 Equipment change
103
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
The Storage Commitment Information Object is used both for N-ACTION Storage
Commitment Requests by the SCU and N-EVENT-REPORT Storage Commitment
Notifications by the SCP.
104
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Storage Media File-Set UID (0088,0140) Not used
Referenced SOP Sequence (0008,1199) Success of storage commitment is reported to the user.
This sequence will be used to mark storage status on screen.
All the successed instance will marked as Stored in the system.
>Referenced SOP Class UID (0008,1150)
>Referenced SOP Instance UID (0008,1155)
>Retrieve AE Title (0008,0054) Not Used
>Storage Media File-Set ID (0088,0130) Not used
>Storage Media File-Set UID (0088,0140) Not used
Failed SOP Sequence (0008,1198) All the failed instance will not be marked as Stored in the
system.
>Referenced SOP Class UID (0008,1150)
>Referenced SOP Instance UID (0008,1155)
>Failure Reason (0008,1197) See Section 8.1.2.1 for the list of processed values.
105
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Table 9-1 identifies the defined modules within the entities which comprise the Basic
Directory IOD. Modules are identified by Module Name.
See DICOM Part 3 for a complete definition of the entities, modules, and attributes.
TABLE 9-1
BASIC DIRECTORY IOD MODULES
Entity Name Module Name Reference
File Set Identification File Set Identification 9.2.1
Directory Information Directory Information 9.2.2
The FSC of this implementation creates a Directory Information Module for new media
and creates XA IOD.
Please refer to DICOM Standard Part 3 (Information Object Definitions) for a description
of each of the entities and modules contained within the Basic Directory Information
Object.
The following modules are included to convey Enumerated Values, Defined Terms, and
Optional Attributes supported. Type 1 & Type 2 Attributes are also included for
completeness and to define what values they may take and where these values are
obtained from. It should be noted that they are the same ones as defined in the DICOM
Standard Part 3 (Information Object Definitions). Also note that Attributes not present in
tables are not supported.
106
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
9.2.1 File Set identification Module
TABLE 9-2
FILE-SET IDENTIFICATION MODULE
Attribute Name Tag Type Attribute Description
File-set ID (0004,1130) 2 Current DateTime stamp, format
like:YYYYMMDDHHMMSS
YYYY- year, MM- month, DD- day.
HH Hour 00-23, MM minute 00-59, SS
second 00-59
File-set Descriptor File ID (0004,1141) 3 Not Used
Specific Character Set of File-set (0004,1142) 1C Not Used
Descriptor File
107
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
>Referenced SOP Class UID in File (0004,1510) 1C Field is included if Directory Record Type is
not PATIENT/STUDY/SERIES.
Unique ID for the SOP Class of the Instance
stored in the referenced File.
>Referenced SOP Instance UID in File (0004,1511) 1C Field is included if Directory Record Type is
not PATIENT/STUDY/SERIES.
Unique Identifier for the SOP Instance stored in
the referenced file.
>Referenced Transfer Syntax UID in File (0004,1512) 1C Field is included if Directory Record Type is
not PATIENT/STUDY/SERIES.
Unique Identifier for the Transfer Syntax used to
encode the Instance stored in the referenced file.
> Referenced Related General SOP Class (0004,151A) 1C Not Used
UID in File
>Record Selection Keys See 9.2.3
108
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Study Description (0008,1030) 2 FSC/FSU will fill this attribute with the value
in the referenced composite object instance.
Study Instance UID (0020,000D) 1C FSC/FSU will fill this mandatory attribute with
the value in the referenced composite object
instance.
This value is mandatory for Brivo OEC
715/785/865, cant be empty.
Study ID (0020,0010) 1 FSC/FSU will fill this mandatory attribute with
the value in the referenced composite object
instance.
FSC/FSU will fill this mandatory attribute with
a value DCMTKSTUDY+ 6 digits if it is
empty in the referenced composite object
instance.
Accession Number (0008,0050) 2 FSC/FSU will fill this attribute using the value
in the referenced composite object instance.
109
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
9.2.3.4 Image Directory Record Definition
TABLE 9-7
IMAGE KEYS
Key Tag Type Attribute Description
Specific Character Set (0008,0005) 1C ISO_IR 100
Image Type (0008,0008) 3 FSC/FSU will fill this mandatory attribute with
the value in the referenced composite object
instance.
Instance Number (0020,0013) 1 FSC/FSU will fill this mandatory attribute with
the value in the referenced composite object
instance.
Icon Image Sequence (0088,0200) 3 FSC/FSU will create Icon Images at the image
level.
>Include Image Pixel Macro See section 9.2.3.4.1
Note: Please also refer to Section 3.2.1.2.1.1.1 for information about Additional
Keys sent in DICOMDIR.
The FSC creates the icon image sequence from the referenced composite object
instance. The FSR displays image icons to the user without reading the actual image
pixel from the media.
Samples Per pixel (0028,0002) is set to 1.
Photometric Interpretations (0028,0004) is set to MONOCHROME2.
Row (0028,0010) and Column (0028,0011) are set to 128,128.
Bits Allocated (0028,0100) and Bits Stored (0028,0101) are created and set to 8.
High Bit (0028,0102) is created and set to 7.
Pixel Representation (0028,0103) is created and set to 0.
Pixel Data (7EF0,0010) is set with the 128x128 icon image data.
110
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
9.2.3.5 SR Document Directory Record Definition
TABLE 9-8
SR DOCUMENT KEYS
Key Tag Type Attribute Description
Specific Character Set (0008,0005) 1C ISO_IR 100
Instance Number (0020,0013) 1 FSC/FSU will fill this mandatory attribute with
the value in the referenced composite object
instance.
Completion Flag (0040,A491) 1 "COMPLETE"
Verification Flag (0040,A493) 1 "Unverified".
Content Date (0008,0023) 1 First Shot Date
Content Time (0008,0033) 1 First Shot Time
Verification DateTime (0040,A030) 1C Not Used
Concept Name Code Sequence (0040,A043) 1 One item shall be included in this sequence.
>Code Value (0008,0100) 1 "113701"
>Coding Scheme Designator (0008,0102) 1 "DCM"
>Code Meaning (0008,0104) 1 "X-Ray Radiation Dose Report"
Content Sequence (0040,A730) 1C
>Relationship Type (0040,A010) 1 All, and only, Content Items with the HAS
CONCEPT MOD relationship from the root
Content Item shall be included in this
Sequence.
>Include Document Content Macro
111
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
10.1 INTRODUCTION
This section of the DICOM Conformance Statement specifies the implementation for the
specific SOP Classes supported in the Basic Grayscale Print Management Meta SOP
Class, the attributes supported for both IODs and services, and the valid range of values
for mandatory and optional attributes.
The N-DELETE is used to delete the complete Basic Film Session SOP Instance
hierarchy. As a result, all references to Image SOP Instances within the film session are
deleted.
112
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Attribute Name Tag Use
Image Display Format (2010,0010) User configurable.
User values =:
STANDARD\1,1
STANDARD\1,2
STANDARD\2,1
STANDARD\2,2
STANDARD\2,3
STANDARD\3,3
STANDARD\3,4
STANDARD\4,4
STANDARD\4,5
STANDARD\5,4
Default value: First selection when declaring printer.
Referenced Film Session Sequence (2010,0500) Used
>Referenced SOP Class UID (0008,1150) Set to 1.2.840.10008.5.1.1.1
>Referenced SOP Instance UID (0008,1155) This value should be returned from SCP in N-CREATE-
RSP Basic Film Session SOP Class; Then this value is
sent to SCP in N-CREATE-RQ Basic Film Box SOP
Class requesting for Basic Grayscale Image Box SOP
Class Instance UID which contained in Referenced Image
Box Sequence from SCP.
Referenced Image Box Sequence (2010,0510) Used
>Referenced SOP Class UID (0008,1150) Set to 1.2.840.10008.5.1.1.4
>Referenced SOP Instance UID (0008,1155) Received in the N-Create Response from SCP, this value
should be sent to SCP in N-SET-RQ Basic Grayscale
Image Box SOP Class
Referenced Basic Annotation Box (2010,0520) Not Used
Sequence
>Referenced SOP Class UID (0008,1150) Not Used
>Referenced SOP Instance UID (0008,1155) Not Used
Film Orientation (2010,0040) Enumerated Terms can be sent:
PORTRAITor LANDSCAPE
Film Size ID (2010,0050) Enumerated Terms can be sent:
8INX10IN
10INX12IN
10INX14IN
11INX14IN
14INX14IN
14INX17IN
24CMX24CM
24CMX30CM
Magnification Type (2010,0060) Set to BILINEAR
Max Density (2010,0130) User configurable. The default is 300
Configuration Information (2010,0150) User configurable. This value specifies the DICOM
113
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
printer parameters. It is defined in the DICOM Printer's
Conformance Statement file.
Referenced Presentation LUT Sequence (2050,0500) Not Used
>Referenced SOP Class UID (0008,1150) Not Used
>Referenced SOP Instance UID (0008,1155) Not Used
Annotation Display Format ID (2010,0030) Not Used
Smoothing Type (2010,0080) Not Used
Border Density (2010,0100) Enumerated Terms can be sent:
BLACK or WHITE
Empty Image Density (2010,0110) Enumerated Terms can be sent:
BLACK or WHITE
Min Density (2010,0120) User configurable. The default is 20
Trim (2010,0140) Set to NO
Illumination (2010,015E) Not Used
Reflected Ambient Light (2010,0160) Not Used
Requested Resolution ID (2020,0050) Not Used
ICC Profile (0028,2000) Not Used
114
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Attribute Name Tag Use
Image Position (2020,0010) Based on Image Display Format (2010,0010), range of
values sent is [1 ~ 20].
Basic Grayscale Image Sequence (2020,0110) A sequence which provides the content of the grayscale
image pixel data to be printed
>Samples Per Pixel (0028,0002) 1
>Photometric Interpretation (0028,0004) MONOCHROME2
>Rows (0028,0010) 1000
>Columns (0028,0011) 1000
>Pixel Aspect Ratio (0028,0034) Not Used
>Bits Allocated (0028,0100) 16
>Bits Stored (0028,0101) 12
>High Bit (0028,0102) 11
>Pixel Representation (0028,0103) 0
>Pixel Data (7FE0,0010) Image Pixels
Polarity (2020,0020) Not Used
Magnification Type (2010,0060) Not Used
Smoothing Type (2010,0080) Not Used
Min Density (2010,0120) Not Used
Max Density (2010,0130) Not Used
Configuration Information (2010,0150) Not Used
Requested Image Size (2020,0030) Not Used
Requested Decimate/Crop Behavior (2020,0040) Not Used
Referenced Presentation LUT Sequence (2050,0500) Not Used
> Referenced SOP Class UID (0008,1150) Not Used
> Referenced SOP Instance UID (0008,1155) Not Used
115
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Event Type Event Attribute Tag Use
Name Type ID
Normal 1 Ignored
Warning 2 Printer Status Info (2110,0020) Ignored
Film Destination (2000,0040) Ignored
Printer Name (2110,0030) Ignored
Failure 3 Printer Status Info (2110,0020) Ignored
Film Destination (2000,0040) Ignored
Printer Name (2110,0030) Ignored
116
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
TABLE 11-1
MAPPING OF DICOM ENTITIES TO BRIVO OEC 715/785/865 ENTITIES
DICOM Brivo OEC 715/785/865 Entity
Patient Exam
Study Exam
Series Exam
Image Image
The query key attributes specified in this section are used at all levels and in all classes of
query.
TABLE 11-2
Q/R PATIENT LEVEL COMMON RETRIEVE ATTRIBUTES
Attribute Name Tag Type SCU Use
Specific Character Set (0008,0005) - See 11.2.1.1.1
Query Retrieve Level (0008,0052) - Set to level of query:
PATIENT
STUDY
SERIES
IMAGE
Retrieve AE Title (0008,0054) - Returned value will be used for Retrieve AE Title of
retrieval of query response item.
Storage Media File-set ID (0088,0130) - Attribute is not requested.
Storage Media File-set UID (0088,0140) - Attribute is not requested.
117
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
11.2.1.1 Q/R Common Attribute Descriptions
In processing of query responses the character set tag (0008,0005) is checked: absence of
the tag is defined as ISO_IR 6 (7 bit ASCII) or the tag is present ISO_IR 100 (ISO-8859-
1 Latin 1) will be accepted.
TABLE 11-3
PATIENT LEVEL ATTRIBUTES FOR THE PATIENT ROOT
QUERY/RETRIEVE INFORMATION MODEL
Attribute Name Tag Type SCU Use
Patient's Name (0010,0010) R* Single value and wildcard matching are supported.
User could enter anything (including wildcard) except
"^","=" and "\".
The value entered by the user will be used as matching
key, but the response from server depends on server's
capability.
During the DICOM archive/sending, the "^" and "="
will be used as delimiter (added by system) in Patient
Name components.
Patient ID (0010,0020) U* Single value and wildcard matching are supported.
User could enter anything (including wildcard) except
"\".
The value entered by the user will be used as matching
key, but the response from server depends on server's
capability.
Referenced Patient (0008,1120) O Not Requested
Sequence
>Referenced SOP Class (0008,1150) O Not Requested
UID
>Referenced SOP Instance (0008,1155) O Not Requested
UID
Patient's Birth Date (0010,0030) O Universal Matching
Patient's Birth Time (0010,0032) O Not Requested
Patient's Sex (0010,0040) O Universal Matching
Other Patient IDs (0010,1000) O Not Requested
Other Patient Names (0010,1001) O Not Requested
Ethnic Group (0010,2160) O Not Requested
Patient Comments (0010,4000) O Not Requested
118
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Number of Patient Related (0020,1200) O Universal Matching
Studies
Number of Patient Related (0020,1202) O Universal Matching
Series
Number of Patient Related (0020,1204) O Universal Matching
Instances
Note: * in the Type column indicates that this information is displayed on screen, if
available
TABLE 11-4
STUDY LEVEL ATTRIBUTES FOR THE PATIENT ROOT
QUERY/RETRIEVE INFORMATION MODEL
Attribute Name Tag Type SCU Use
Study Date (0008,0020) R* Range value matching is supported.
User could select one of "All", "Today", "Since
Yesterday", "Last 3 days", "Last 7 days" and "Last 30
days" from the screen, and then date information will
be sent as "yyyymmdd-yyyymmdd" format to the
server for matching.
Study Time (0008,0030) R* Universal Matching
Accession Number (0008,0050) R* Single value and wildcard matching are supported.
Leading and trailing wildcard matching for value
entered.
The value entered by the user will be used as matching
key.
Study ID (0020,0010) R Universal Matching
Study Instance UID (0020,000D) U Universal Matching
Modalities in Study (0008,0061) O Universal Matching
SOP Classes in Study (0008,0062) O Not Requested
Referring Physician's Name (0008,0090) O Not Requested
Study Description (0008,1030) O* Universal Matching
Procedure Code Sequence (0008,1032) O Not Requested
>Code Value (0008,0100) O* Not Requested
>Coding Scheme (0008,0102) O Not Requested
Designator
>Coding Scheme Version (0008,0103) O Not Requested
>Code Meaning (0008,0104) O Not Requested
Name of Physician(s) (0008,1060) O Not Requested
Reading Study
Admitting Diagnoses (0008,1080) O Not Requested
Description
119
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Referenced Study Sequence (0008,1110) O Not Requested
>Referenced SOP Class (0008,1150) O Not Requested
UID
>Referenced SOP Instance (0008,1155) O Not Requested
UID
Issuer of Patient ID (0010,0021) O Not Requested
Patients Age (0010,1010) O Not Requested
Patients Size (0010,1020) O Not Requested
Patients Weight (0010,1030) O Not Requested
Occupation (0010,2180) O Not Requested
Additional Patient History (0010,21B0) O Not Requested
Other Study Numbers (0020,1070) O Not Requested
Number of Study Related (0020,1206) O Universal Matching
Series
Number of Study related (0020,1208) O Universal Matching
Instances
Interpretation Author (4008,010C) O Not Requested
Note: * in the Type column indicates that this information is displayed on screen, if
available
TABLE 11-5
STUDY LEVEL ATTRIBUTES FOR THE STUDY ROOT
QUERY/RETRIEVE INFORMATION MODEL
Attribute Name Tag Type SCU Use
Study Date (0008,0020) R* Range value matching is supported.
User could select one of "All", "Today", "Since
Yesterday", "Last 3 days", "Last 7 days" and "Last 30
days" from the screen, and then date information will
be sent as "yyyymmdd-yyyymmdd" format to the
server for matching.
Study Time (0008,0030) R* Universal Matching
Accession Number (0008,0050) R* Single value and wildcard matching are supported.
The value entered by the user will be used as matching
key.
Study ID (0020,0010) R Universal Matching
120
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Patient's Name (0010,0010) R* Single value and wildcard matching are supported.
User could enter anything (including wildcard) except
"^","=" and "\".
The value entered by the user will be used as matching
key, but the response from server depends on server's
capability.
During the DICOM archive/sending, the "^" and "="
will be used as delimiter (added by system) in Patient
Name components.
Patient ID (0010,0020) R* Single value and wildcard matching are supported.
User could enter anything (including wildcard) except
"\".
The value entered by the user will be used as matching
key, but the response from server depends on server's
capability.
Study Instance UID (0020,000D) U Universal Matching
Modalities in Study (0008,0061) O Universal Matching
121
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Number of Study Related (0020,1206) O Universal Matching
Series
Number of Study related (0020,1208) O Universal Matching
Instances
Interpretation Author (4008,010C) O Not Requested
Referenced Patient (0008,1120) O Not Requested
Sequence
>Referenced SOP Class (0008,1150) O Not Requested
UID
>Referenced SOP Instance (0008,1155) O Not Requested
UID
Patient's Birth Date (0010,0030) O Universal Matching
Patient's Birth Time (0010,0032) O Not Requested
Patient's Sex (0010,0040) O Universal Matching
Other Patient IDs (0010,1000) O Not Requested
Other Patient Names (0010,1001) O Not Requested
Ethnic Group (0010,2160) O Not Requested
Patient Comments (0010,4000) O Not Requested
Number of Patient Related (0020,1200) O Not Requested
Studies
Number of Patient Related (0020,1202) O Not Requested
Series
Number of Patient Related (0020,1204) O Not Requested
Instances
Note: * in the Type column indicates that this information is displayed on screen, if
available
This section defines the keys at the Series Level of the Patient Root and Study Root
Query/Retrieve Information Models that are supported by this implementation.
TABLE 11-6
SERIES LEVEL ATTRIBUTES FOR THE
QUERY/RETRIEVE INFORMATION MODEL
Attribute Name Tag Type SCU Use
Modality (0008,0060) R* Single value matching is supported.
User could select one of "XA", "RF", "CR", "SC" and
ALL for Modality Type from screen, then the
corresponding string "XA", "RF", "CR", "OT" and
will be sent to server for matching.
Note: "ALL" means ignore the Modality Type, the
Modality Type will leave empty during sending.
Series Number (0020,0011) R Universal Matching
Series Instance UID (0020,000E) U Universal Matching
122
BRIVO OEC 715/785/865 SOFTWARE RELEASE WKS-
1.X.X
GE Healthcare DICOM CONFORMANCE STATEMENT
DOC1188156 REV 2
Number of Series Related (0020,1209) O Universal Matching
Instances
Series Description (0008,103E) O Universal Matching
Performed Procedure Step (0040, 0253) O Not Requested
ID
Referenced Performed (0008,1111) O Not Requested
Procedure Step Sequence
>Referenced SOP Class (0008,1150) O Not Requested
UID
>Referenced SOP Instance (0008,1155) O Not Requested
UID
Request Attribute Sequence (0040, 0275) O Not Requested
>Requested Procedure ID (0040,1001) O Not Requested
>Scheduled Procedure Step (0040,0009) O Not Requested
ID
Performed Procedure Step (0040,0244) O Not Requested
Start Date
Performed Procedure Step (0040,0245) O Not Requested
Start Time
Note: * in the Type column indicates that this information is displayed on screen, if
available
123