Gs01e20d01 01e PDF
Gs01e20d01 01e PDF
General AXF
Specifications Magnetic Flowmeter
Integral Flowmeter
GS 01E20D01-01E /Remote Flowtube
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 3
Totalizer Display Function (*2): Preset Totalization (*1)(*2):
The flow rate is counted one pulse at a time according to The parameter setting or status input enables a totalized
the setting of totalization pulse weights. For forward and value to be preset to a setting value or zero.
reverse flow measurement functions, the totalized 0% Signal Lock (*1)(*2):
values of the flow direction (forward or reverse) and the Status input forcibly fixes the instantaneous flow rate
flow direction are displayed on the indicator together display, current output, pulse output, and flow rate
with the units. The difference of totalized values between totalization to 0%.
the forward and reverse flow rate can be displayed.
Alarm Selection Function (*2):
Totalization for the reverse flow rate is carried out only
Alarms are classified into the System Alarms (hard
when Forward and reverse flow measurement func-
failures), Process Alarms (such as Empty Pipe, Signal
tions is selected.
Overflow and Adhesion Alarm), Setting Alarms, and
Damping Time Constant (*2): Warnings.
Time constant can be set from 0.1 second to 200.0 Whether alarms should be generated or not can be
seconds (63% response). The default is 3 seconds. selected for each item.
Span Setting Function (*2): The current output generated for an alarm can be
Span flows can be set in units such as volume flow rate, selected from among 2.4 mA or less, fixed to 4 mA, 21.6
mass flow rate, time, or flow rate value. The velocity unit mA or more, or HOLD.
can also be set. Alarm Output (*1)(*2):
Volume Flow Rate Unit: kcf, cf, mcf, Mgal (US), kgal Alarms are generated only for the items selected via the
(US), gal (US), mgal (US), kbbl (US)*, bbl Alarm Selection Function if relevant failures occur.
(US)*, mbbl (US)*, bbl (US)*, Ml
Self Diagnostics Functions (*2):
(megaliter), m3, kl (kiloliter), l (liter), cm3
If alarms are generated, details of the System Alarms,
Mass Flow Rate Unit (Density must be set.): klb (US), lb
Process Alarms, Setting Alarms and Warnings are
(US), t (ton), kg, g
displayed together with concrete descriptions of
Velocity Unit: ft, m (meter)
countermeasures.
Time Unit: s (sec), min, h (hour), d (day)
* US oil or US Beer can be selected.
Flow Upper/Lower Limit Alarms (*1)(*2):
If a flow rate becomes greater or smaller than the set
Pulse Output (*1)(*2):
value, this alarm is generated. In addition, two upper
Scaled pulse can be output by setting a pulse weight.
limits (H, HH) and two lower limits (L, LL) can be set.
Pulse Width: Duty 50% or fixed pulse width (0.05, 0.1,
If a flow rate becomes greater or smaller than any of the
0.5, 1, 20, 33, 50, 100 ms) can be selected.
set values, the status is output.
Output Rate: 0.0001 to 10,000 pps (pulse/second)
Electrode Adhesion Diagnostics Function (*1) (*2):
Multi-range Function (*1)(*2):
This function enables monitoring of the adhesion level of
Range switching via status input
insulating substances to the electrodes. Depending on
Status input enables the switching of up to two ranges.
the status of adhesion, users are notified by a warning or
Automatic range switching
an alarm via status outputs. If replaceable electrodes are
When the flow rate exceeds 100 % of the range,
used, they can be removed and cleaned when adhesion
transition to the next range (up to four ranges) is
occurs.
carried out automatically. Range swiching can be
confirmed by status outputs and indicator.
Forward and Reverse Flow Measurement Functions (*1)(*2):
Flows in both forward and reverse directions can be
measured. The reverse flow measurement can be
confirmed by status output and indicator.
Totalization Switch (*1)(*2):
The status output is carried out when a totalized value
becomes equal to or greater than the set value.
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 4
Flowtubes (Remote Flowtube/Integral Flowmeter)
Size of AXF Flowtubes: AXF Standard (Lay length code 1)
Unit: mm (in.)
High Grade Enhanced Dual Frequency
Use Process Connection Lining Remote Flowtube Integral Flowmeter Replaceable Electrode
Accuracy 0.2% of Excitation
(Electrode structure code 2)
Rate (*3) (Optional code HF1,HF2) (*3)
2.5 (0.1), 5 (0.2), 10 (0.4), 15 (0.5), 25 (1.0), 32 (1.25), 25 (1.0), 32 (1.25), 25 (1.0), 32 (1.25), 40 (1.5),
25 (1.0), 32 (1.25), 40 (1.5), 50 (2.0), 40 (1.5), 50 (2.0), 40 (1.5), 50 (2.0), 65 (2.5), 50 (2.0), 65 (2.5), 80 (3.0),
PFA 65 (2.5), 80 (3.0), 100 (4.0), 125 (5.0), 65 (2.5), 80 (3.0), 80 (3.0), 100 (4.0), 125 (5.0), 100 (4.0), 125 (5.0), 150 (6.0),
150 (6.0), 200 (8.0), 250 (10), 300 (12) 100 (4.0), 125 (5.0), 150 (6.0), 200 (8.0) 200 (8.0), 250 (10), 300 (12)
150 (6.0), 200 (8.0)
25 (1.0), 32 (1.25), 40 (1.5), 50 (2.0), 25 (1.0), 32 (1.25), 40 (1.5), 25 (1.0), 32 (1.25), 40 (1.5),
Polyurethane 65 (2.5), 80 (3.0), 100 (4.0), 125 (5.0), 50 (2.0), 65 (2.5), 80 (3.0), 50 (2.0), 65 (2.5), 80 (3.0),
Rubber
150 (6.0), 200 (8.0), 250 (10), 300 (12) 100 (4.0), 125 (5.0), 150 (6.0), 100 (4.0), 125 (5.0), 150 (6.0),
200 (8.0) 200 (8.0), 250 (10), 300 (12)
50 (2.0), 65 (2.5), 80 (3.0), 100 (4.0), 50 (2.0), 65 (2.5),
Wafer Natural Soft 125 (5.0), 150 (6.0), 200 (8.0), 250 (10), 80 (3.0), 100 (4.0),
Rubber 300 (12) 125 (5.0), 150 (6.0),
200 (8.0)
50 (2.0), 65 (2.5), 80 (3.0), 100 (4.0), 50 (2.0), 65 (2.5),
EPDM 125 (5.0), 150 (6.0), 200 (8.0), 250 (10), 80 (3.0), 100 (4.0),
Rubber 300 (12) 125 (5.0), 150 (6.0),
200 (8.0)
15 (0.5), 25 (1.0), 40 (1.5), 50 (2.0), 25 (1.0), 40 (1.5), 25 (1.0), 40 (1.5), 50 (2.0),
Ceramics 80 (3.0), 100 (4.0), 150 (6.0), 200 (8.0) 50 (2.0), 80 (3.0), 80 (3.0), 100 (4.0), 150 (6.0),
(*1) 100 (4.0), 150 (6.0), 200 (8.0)
General- 200 (8.0)
purpose Use 2.5 (0.1), 5 (0.2), 10 (0.4), 15 (0.5), 25 (1.0), 25 (1.0), 32 (1.25), 25 (1.0), 32 (1.25), 40 (1.5), 25 (1.0), 32 (1.25), 40 (1.5),
32 (1.25), 40 (1.5), 50 (2.0), 65 (2.5), 40 (1.5), 50 (2.0), 50 (2.0), 65 (2.5), 80 (3.0), 50 (2.0), 65 (2.5), 80 (3.0),
PFA 80 (3.0), 100 (4.0), 125 (5.0), 150 (6.0), 65 (2.5), 80 (3.0), 100 (4.0), 125 (5.0), 100 (4.0), 125 (5.0), 150 (6.0),
200 (8.0), 250 (10), 300 (12), 350 (14), 100 (4.0), 125 (5.0), 150 (6.0), 200 (8.0) 200 (8.0), 250 (10), 300 (12),
400 (16) 150 (6.0), 200 (8.0) 350 (14), 400 (16)
25 (1.0), 32 (1.25), 25 (1.0), 32 (1.25), 25 (1.0), 32 (1.25), 40 (1.5), 25 (1.0), 32 (1.25), 40 (1.5),
40 (1.5), 50 (2.0), 40 (1.5), 50 (2.0), 50 (2.0), 65 (2.5), 80 (3.0), 50 (2.0), 65 (2.5), 80 (3.0),
65 (2.5), 80 (3.0), 65 (2.5), 80 (3.0), 100 (4.0), 125 (5.0), 100 (4.0), 125 (5.0), 150 (6.0),
Polyurethane
100 (4.0), 125 (5.0), 100 (4.0), 125 (5.0), 150 (6.0), 200 (8.0) 200 (8.0), 250 (10), 300 (12),
Rubber
150 (6.0), 200 (8.0), 150 (6.0), 200 (8.0), 350 (14), 400 (16)
Flange 250 (10), 300 (12), 250 (10), 300 (12),
350 (14), 400 (16) 350 (14), 400 (16)
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 5
Size of AXF Flowtubes: AXF Standard (Lay length code 1) (continued)
Unit: mm (in.)
High Grade Enhanced Dual Frequency
Use Process Connection Lining Remote Flowtube Integral Flowmeter Replaceable Electrode
Accuracy 0.2% of Excitation
(Electrode structure code 2)
Rate (*3) (Optional code HF1,HF2) (*3)
15 (0.5), 25 (1.0), 25 (1.0), 32 (1.25), 25 (1.0), 32 (1.25), 40 (1.5),
32 (1.25), 40 (1.5) 40 (1.5), 50 (2.0), 50 (2.0), 65 (2.5), 80 (3.0),
50 (2.0), 65 (2.5), 65 (2.5), 80 (3.0), 100 (4.0), 125 (5.0),
PFA 80 (3.0), 100 (4.0), 100 (4.0), 125 (5.0), 150 (6.0), 200 (8.0)
125 (5.0), 150 (6.0), 150 (6.0), 200 (8.0)
200 (8.0), 250 (10),
300 (12)
25 (1.0), 32 (1.25), 25 (1.0), 32 (1.25), 40 (1.5),
40 (1.5), 50 (2.0), 50 (2.0), 65 (2.5), 80 (3.0),
Polyurethane 65 (2.5), 80 (3.0), 100 (4.0), 125 (5.0),
Rubber
100 (4.0), 125 (5.0), 150 (6.0), 200 (8.0)
Wafer 150 (6.0), 200 (8.0),
250 (10), 300 (12)
50 (2.0), 65 (2.5),80 (3.0), 50 (2.0), 65 (2.5),
Natural Soft 100 (4.0),125 (5.0), 80 (3.0), 100 (4.0),
Rubber 150 (6.0), 200 (8.0), 125 (5.0), 150 (6.0),
250 (10), 300 (12) 200 (8.0)
50 (2.0), 65 (2.5),80 (3.0), 50 (2.0), 65 (2.5),
EPDM 100 (4.0), 125 (5.0), 80 (3.0), 100 (4.0),
Rubber 150 (6.0), 200 (8.0), 125 (5.0), 150 (6.0),
250 (10), 300 (12) 200 (8.0)
Submersible 15 (0.5), 25 (1.0), 25 (1.0), 32 (1.25), 25 (1.0), 32 (1.25), 40 (1.5),
Type 32 (1.25), 40 (1.5), 40 (1.5), 50 (2.0), 50 (2.0), 65 (2.5), 80 (3.0),
50 (2.0), 65 (2.5), 65 (2.5), 80 (3.0), 100 (4.0), 125 (5.0),
80 (3.0), 100 (4.0), 100 (4.0), 125 (5.0), 150 (6.0), 200 (8.0)
PFA
125 (5.0), 150 (6.0), 150 (6.0), 200 (8.0)
200 (8.0), 250 (10),
300 (12), 350 (14),
400 (16)
25 (1.0), 32 (1.25), 25 (1.0), 32 (1.25),
40 (1.5), 50 (2.0), 40 (1.5), 50 (2.0), 65 (2.5),
Polyurethane 65 (2.5), 80 (3.0), 80 (3.0), 100 (4.0),
Flange 100 (4.0), 125 (5.0), 125 (5.0), 150 (6.0),
Rubber
150 (6.0), 200 (8.0), 200 (8.0)
250 (10), 300 (12),
350 (14), 400 (16)
*1: AXF standard lay length dimensions for wafer type ceramics linings are the same as those for ADMAG ceramics linings. T21.EPS
*2: AXF standard lay length dimensions for union joint type ceramics linings are the same as those for ADMAG ceramics linings.
*3: Enhanced dual frequency excitation is not available for models with High grade accuracy.
*4: Not available with 32 mm (1.25 in.), 125 mm (5.0 in.)
*5: Not available with 125 mm (5.0 in.)
*6: Not available with 15 mm (0.5 in.), 125 mm (5.0 in.)
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 6
Size of AXF Flowtubes: Replacement model for earlier ADMAG or ADMAG AE (Lay length code 2)
Unit: mm (in.)
High Grade Enhanced Dual Frequency
Use Process Connection Lining Remote Flowtube Integral Flowmeter Replaceable Electrode
Accuracy 0.2% of Excitation
(Electrode structure code 2)
Rate (Optional code HF1,HF2)
2.5 (0.1), 5 (0.2), 10 (0.4), 15 (0.5), 25 (1.0), 40 (1.5), 50 (2.0),
PFA 25 (1.0), 40 (1.5), 50 (2.0), 80 (3.0), 80 (3.0), 100 (4.0),
100 (4.0), 150 (6.0), 200 (8.0) 150 (6.0), 200 (8.0)
Wafer (*6)
25 (1.0), 40 (1.5), 50 (2.0), 80 (3.0), 25 (1.0), 40(1.5), 50 (2.0),
General- Polyurethane
100(4.0), 150 (6.0), 200(8.0) 80(3.0), 100 (4.0),
purpose use rubber
150 (6.0), 200 (8.0)
PFA 150 (6.0), 200 (8.0), 250 (10) 150 (6.0), 200 (8.0) 150 (6.0), 200 (8.0), 250 (10)
Flange (*7) Polyurethane
150 (6.0), 200 (8.0), 250 (10) 150 (6.0), 200 (8.0) 150 (6.0), 200 (8.0), 250 (10)
rubber
15 (0.5), 25 (1.0), 25 (1.0), 40 (1.5),
40 (1.5), 50 (2.0), 50 (2.0), 80 (3.0), 100 (4.0),
PFA
80 (3.0), 100 (4.0), 150 (6.0), 200 (8.0)
150 (6.0), 200 (8.0)
Wafer (*6)
25 (1.0), 40 (1.5), 25 (1.0), 40 (1.5), 50 (2.0),
Submersible Polyurethane 50 (2.0), 80 (3.0), 80 (3.0), 100 (4.0),
Type rubber 100 (4.0), 150 (6.0), 150 (6.0), 200 (8.0)
200 (8.0)
150 (6.0), 200 (8.0), 150 (6.0), 200 (8.0)
PFA
250(10)
Flange (*7)
Polyurethane 150 (6.0), 200 (8.0), 150 (6.0), 200 (8.0)
rubber 250 (10)
2.5 (0.1), 5 (0.2), 10 (0.4), 15 (0.5), 25 (1.0), 40 (1.5), 50 (2.0),
Explosion Wafer (*6) PFA 25 (1.0), 40 (1.5), 50 (2.0), 80 (3.0), 80 (3.0), 100 (4.0),
proof Type 100 (4.0), 150 (6.0), 200 (8.0) 150 (6.0), 200 (8.0)
Flange (*7) PFA 150 (6.0), 200 (8.0), 250 (10) 150 (6.0), 200 (8.0)
T22.EPS
*6: ADMAG lay length dimensions for wafer type of 250 mm (10 in.), and 300 mm (12 in.) are the same as those for AXF Standard.
And, in case of platinum-iridium (grounding ring code P) or tantalum (grounding ring code T) or None (grounding ring code N) in wafer type of 2.5 mm (0.1 in.)
to 15 mm (0.5 in.), the lay lengths of Replacement model are the same as those for AXF Standard.
*7: ADMAG lay length dimensions for flange type of 15 mm (0.5 in.) to 100 mm (4.0 in.), or 300 mm (12 in.) to 400 mm (16 in.) are the same as those for AXF
Standard. However, in case of platinum-iridium (grounding ring code P) or tantalum (grounding ring code T) or None (grounding ring code N) in flange type of 15
mm (0.5 in.) to 100 mm (4.0 in.), the lay length of AXF Standard are longer by approx. 4mm (0.16 in) than those of earlier ADMAG or ADMAG AE.
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 7
Flowtube Material: Size 25 mm (1.0 in.) to 125 mm (5.0 in.)
Size 2.5 mm (0.1 in.) to 15 mm (0.5 in.)
Part Name Material
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 8
Size 150 mm (6.0 in.) to 400 mm (16 in.) Gasket:
T23-2.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 9
Joints:
Lining Ceramics
Union Joints (size 10 mm or less)
weld joint
(GUW)
nut
screw joint
(GUN, GUR)
Standard
Gasket within Flowtube
Materials for Union Joint
Process Connection Code
Stainless steel
GUW: Union Joint (weld joint)
(JIS SUS316L
Process Connection Code (ANSI 316L SS/EN
GUN, GUR: union joint 1.4404
(screw joint) equivalent))
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 10
INNER DIAMETER OF GROUNDING RING, OUTER DIAMETER FOR EFFECTIVE SEALING;
AXF Standard:
Unit : mm (in.)
Lining PFA/Polyurethane Rubber/
Natural Soft Rubber/ Ceramics
EPDM Rubber
Inner diameter of Outer diameter for Inner diameter Outer diameter
Grounding ring [A] effective sealing [B] of for effective
Size Water Flange Water Flange Grounding ring [A] sealing [B]
2.5 (0.1) 15 (0.59) [12 (0.47)] *1 38 (1.50) 30 (1.18) - -
5 (0.2) 15 (0.59) [12 (0.47)] *1 38 (1.50) 30 (1.18) - -
10 (0.4) 15 (0.59) [12 (0.47)] *1 38 (1.50) 30 (1.18) - -
15 (0.5) 15 (0.59) 38 (1.50) 34 (1.34) 15 (0.59) 33 (1.30)
25 (1.0) 28 (1.10) 53 (2.09) 53 (2.09) 27 (1.06) 50 (1.97)
32 (1.25) 34 (1.34) 58 (2.28) 58 (2.28) - -
40 (1.5) 41 (1.61) 71 (2.80) 71 (2.80) 40 (1.57) 68 (2.68)
50 (2.0) 53 (2.09) 84 (3.31) 84 (3.31) 52 (2.05) 82 (3.23)
65 (2.5) 66 (2.60) 103 (4.06) 103 (4.06) - -
80 (3.0) 77 (3.03) 114 (4.49) 114 (4.49) 81 (3.19) 112 (4.41)
100 (4.0) 102 (4.02) 140 (5.51) 140 (5.51) 98 (3.86) 134 (5.28)
125 (5.0) 128 (5.04) 165 (6.50) 165 (6.50) - -
150 (6.0) 146.1 (5.75) 190 (7.48) 190 (7.48) 144 (5.67) 188 (7.40)
200 (8.0) 193.6 (7.62) 240 (9.45) 240 (9.45) 192 (7.56) 240 (9.45)
250 (10) 243.7 (9.60) 243 (9.57) 300 (11.81) 315 (12.40) - -
300 (12) 294.7 (11.60) 291.3 (11.47) 348 (13.70) 360 (14.17) - -
350 (14) - 323.4 (12.73) - 405 (15.94) - -
400 (16) - 373.5 (14.70) - 465 (18.31) - -
T25.EPS
*1: The I.D. of the process connection code: DD4, DJ1, DJ2 is values in brackets [ ].
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 11
Size of Inner Diameter of Grounding ring and Outer Mounting/Shapes (Remote Flowtube):
Diameter for Effective Sealing : Electrical Connection: ANSI 1/2 NPT female
ISO M20 1.5 female
JIS G1/2 (PF1/2) female
Direction of Electrical Connection: The direction can be
changed even after delivery.
Note: In case of submersible types or an optional code
AXF Flowtube DHC, the direction can not be changed after
delivery.
B A Terminal Connection at Terminal Box: M4 size screw
lining Grounding:
Grounding resistance 100 or less
* In case of explosion proof type except TIIS, follow the
domestic electrical requirements as regulated in each
country.
* In case of TIIS Flameproof type, refer to description of
Gasket for HAZARDOUS AREA CLASSIFCATION.
Grounding ring
Users flange
Combined Converter:
Electrode Construction: A remote flowtube can be combined with the AXFA11
Non-replaceable Electrode Type Converter or the AXFA14 Converter. If a combined
General-Purpose Use/Submersible Type/Explosion converter is changed from AXFA11 to AXFA14 or vice
proof Type: versa, a new meter factor must be adjusted by flow
PFA, Polyurethane Rubber lining: calibrations.
External insertion type In case that size 250 mm (10 in.) or larger is used in low
Natural Soft Rubber, EPDM Rubber lining: conductivity or high concentration slurries, please use
Internal insertion type the AXFA11 Converter.
Ceramics lining: Integral type Maximum Cable Length:
Sanitary Type: Internal insertion type Combination of AXF remote Flowtube and AXFA11:
Replaceable Electrode Type up to 200 m (660 ft)
Electrode parts can be put into unit to facilitate Combination of AXF remote Flowtube and AXFA14:
replacement or mounting at customer site. up to 100 m (330 ft)
The optional dedicated tool (F9807SK) is required.
Replaceable electrodes are available for the
following:
AXF standard:
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 12
HAZARDOUS AREA CLASSIFICATION ATEX:
*AXF002C AXF400C
FM: Applicable Standard:
*AXF002C AXF400C EN 50014, EN 50018, EN 50019,
Applicable Standard: EN 50020, EN 50028, EN 50281-1-1
FM3600, FM3610, FM3615, Certificate: KEMA 03ATEX2435
FM3810, ANSI/NEMA 250
(Integral Flowmeter) (Integral Flowmeter)
Explosion proof for Class I, Division 1, Groups A, B, C ATEX Flameproof Type
& D. Group: II
Dust-ignition proof for Class II/III, Division1, Groups E, Category: 2G
F & G. EEx dme[ia] IIC T6...T3
Intrinsically safe (electrodes) for Class I, Division 1, Electrode Circuit Um: 250 Vac/dc
Groups A, B, C & D. Maximum power supply voltage: 250 Vac/130 Vdc
SEAL ALL CONDUITS WITHIN 18 INCHES Excitation Circuit: 140V max
WHEN INSTALLED IN DIV. 2, SEALS NOT REQUIRED Enclosure: IP66, IP67
Temperature Class:
Electrode Circuit Um: 250 Vac/dc
Maximum power supply voltage: 250 Vac/130 Vdc Temperature Maximum Process Minimum Process
Class Temperature Temperature
Excitation Circuit: 140V max
Enclosure: NEMA 4X T6 +70C (+158F) 40C (40F)
Temperature Code: T6 T5 +85C (+185F) 40C (40F)
Refer to following table;
T4 +120C (+248F) 40C (40F)
Temperature Maximum Process Minimum Process
T3 +130C (+266F) 40C (40F)
Code Temperature Temperature
T27-2.EPS
T6 +70C (+158F) 40C (40F) Ambient Temp.: 40C to +60C (40F to +140F)
T5 +85C (+185F) 40C (40F)
ATEX Type of Protection Dust
T4 +120C (+248F) 40C (40F)
Group: II
T3 +130C (+266F) 40C (40F) Category: 1D
T27-1_1.EPS
Electrode Circuit Um: 250 Vac/dc
Ambient Temp.: 40C to +60C (40F to +140F) Maximum power supply voltage: 250 Vac/130 Vdc
Excitation Circuit: 140V max
(Remote Flowtube) Enclosure: IP66, IP67
Explosion proof for Class I, Division 1, Groups A, B, C Maximum surface temperature:
& D.
Dust-ignition proof for Class II/III, Division1, Groups E, Maximum Surface Maximum Process
Temperature Temperature
F & G.
Intrinsically safe (electrodes) for Class I, Division 1, T75C (+167F) +70C (+158F)
Groups A, B, C & D. T85C (+185F) +85C (+185F)
SEAL ALL CONDUITS WITHIN 18 INCHES
T100C (+212F) +120C (+248F)
WHEN INSTALLED IN DIV. 2, SEALS NOT REQUIRED
T110C (+230F) +130C (+266F)
Electrode Circuit Um: 250 Vac/dc T29.EPS
Excitation Circuit: 170V max Ambient Temp.: 40C to +60C (40F to +140F)
Enclosure: NEMA 4X
Temperature Code: T6
Refer to following table;
Temperature Maximum Process Minimum Process
Code Temperature Temperature
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 13
(Remote Flowtube) CSA:
ATEX Flameproof Type *AXF002C AXF400C
Group: II Applicable Standard:
Category: 2G For CSA C22.2 Series;
EEx dme[ia] IIC T6...T3 C22.2 No 0, C22.2 No 0.4, C22.2 No 0.5,
Electrode Circuit Um: 250 Vac/dc C22.2 No 25, C22.2 No 30, C22.2 No 94,
Excitation Circuit: 170V max C22.2 No 157, C22.2 No.61010-1-12,
Enclosure: IP66, IP67 C22.2 No.61010-2-030-12
Temperature Class: For CSA E79 Series;
CAN/CSA-E79-0, CAN/CSA-E79-1,
Temperature Maximum Process Minimum Process
Class Temperature Temperature CAN/CSA-E79-7, CAN/CSA-E79-11,
CAN/CSA-E79-18
T6 +70C (+158F) 40C (40F)
Certificate: 1481213
T5 +85C (+185F) 40C (40F) Process Sealing Certification:
T4 +120C (+248F) 40C (40F) Dual Seal certified by CSA to the requirements of
ANSI/ISA 12.27.01.
T3 +150C (+302F) 40C (40F)
No additional sealing required.
T28-2.EPS
Primary seal failure annunciation;
Ambient Temp.: 40C to +60C (40F to +140F)
Deterioration of the flowrate output at nonzero flow
point.
ATEX Type of Protection Dust
Unstable flowrate output at zero flow point.
Group: II
Category: 1D
(Integral Flowmeter)
Electrode Circuit Um: 250 Vac/dc
For CSA C22. 2 Series
Excitation Circuit: 170V max
Explosion proof for Class I, Division 1, Groups A, B, C
Enclosure: IP66, IP67
& D.
Maximum surface temperature:
Dust-ignition proof for Class II/III, Division 1, Groups E,
Maximum Surface Maximum Process F & G.
Temperature Temperature Intrinsically safe (electrodes) for Class I, Division 1,
Groups A, B, C & D.
T75C (+167F) +70C (+158F)
SEAL ALL CONDUITS WITHIN 50 cm OF THE
T85C (+185F) +85C (+185F) ENCLOSURE
T100C (+212F) +120C (+248F) WHEN INSTALLED IN DIV. 2, SEALS NOT REQUIRED
T115C (+239F) +150C (+302F) Electrode Circuit Um: 250 Vac/dc
T30.EPS
Maximum power supply voltage: 250 Vac/130 Vdc
Ambient Temp.: 40C to +60C (40F to +140F) Excitation Circuit: 140V max
Enclosure: Type 4X
Temperature Code:
Temperature Maximum Process Minimum Process
Code Temperature Temperature
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 14
For CSA E79 Series IECEx:
Flameproof for Zone 1, Ex dme[ia] IIC T6...T3 *AXF002C AXF400C
Intrinsically safe (electrodes), Ex ia IIC T6...T3 Applicable Standard:
IEC60079-0, IEC60079-1, IEC60079-7,
Electrode Circuit Um: 250 Vac/dc
IEC60079-11, IEC60079-18, IEC61241-0,
Maximum power supply voltage: 250 Vac/130 Vdc
IEC61241-1, IEC60529 + Edition 2.1
Excitation Circuit: 140V max
Certificate: IECEx KEM 05.0018
Enclosure: IP66, IP67
Temperature Code:
(Integral Flowmeter)
Temperature Maximum Process Minimum Process IECEx Flameproof Type
Code Temperature Temperature Ex demb[ia] IIC T6...T3
T6 +70C (+158F) 40C (40F) Electrode Circuit Um: 250 Vac/dc
Maximum power supply voltage: 250 Vac/130 Vdc
T5 +85C (+185F) 40C (40F)
Excitation Circuit: 140V max
T4 +120C (+248F) 40C (40F) Enclosure: IP66, IP67
T3 +130C (+266F) 40C (40F) Temperature Class:
T27-1.EPS
Temperature
Ambient Temp.: 40C to +60C (40F to +140F) Class
Process Temperature
T4 +120C (+248F) 40C (40F) T105C (+221F) 40C to +85C (40F to +185F)
T3 +150C (+302F) 40C (40F) T120C (+248F) 40C to +120C (40F to +248F)
T28-1.EPS T130C (+266F) 40C to +130C (40F to +266F)
Ambient Temp.: 40C to +60C (40F to +140F) T27-4.EPS
Ambient Temp.:
For CSA E79 Series PFA Lining; 40C to +60C (40F to +140F)
Flameproof for Zone 1, Ex dme[ia] IIC T6...T3 Ceramics Lining; 15C to +60C (5F to +140F)
Intrinsically safe (electrodes), Ex ia IIC T6...T3
Electrode Circuit Um: 250 Vac/dc
Excitation Circuit: 170V max
Enclosure: IP66, IP67
Temperature Code:
Temperature Maximum Process Minimum Process
Code Temperature Temperature
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 15
(Remote Flowtube) TIIS:
IECEx Flameproof Type (Integral Flowmeter)
Ex demb[ia] IIC T6...T3 Construction: Ex de[ia] IIC T4
Electrode Circuit Um: 250 Vac/dc : Converter ; Explosion proof
Excitation Circuit: 170V max Flowtube ; Increased Safety and
Enclosure: IP66, IP67 Intrinsically Safety(ia)
Temperature Class: Electrode ; Intrinsically Safety(ia)
Ambient Temperature: 20 to 60C (power supply code 1)
Temperature
Process Temperature : 20 to 50C (power supply code 2)
Class
Fluid Temperature: 120C max
T6 40C to +70C (40F to +158F)
Electrode Circuit: 250V AC/DC
T5 40C to +85C (40F to +185F) Maximum power supply voltage: 250V AC/130V DC
T4 40C to +120C (40F to +248F) Grounding: JIS Class C(grouding resister 10 or
less) or JIS Class A(grounding resister 10
T3 40C to +150C (40F to +302F)
or less)
T27-5.EPS
*In case that ambient temperature exceeds 50C, use
Ambient Temp.: heat-resistant cables with maximum allowable tempera-
PFA Lining; 40C to +60C (40F to +140F) ture of 70C or above.
Ceramics Lining; 15C to +60C (5F to +140F)
(Remote Flowtube)
IECEx Type of Protection Dust Construction: Ex de[ia] IIC T4
Ex tD A21 IP6x T95C, T105C, T120C, T135C : Terminal box ; Explosion proof
Electrode Circuit Um: 250 Vac/dc Flowtube ; Increased Safety and
Excitation Circuit: 170V max Intrinsically Safety(ia)
Enclosure: IP66, IP67 Electrode; Intrinsically Safety(ia)
Maximum surface temperature: Ambient Temperature: 20 to 60C
Fluid Temperature: 120C max
Maximum Surface Electrode Circuit: 250V AC/DC
Process Temperature
Temperature
Grounding: JIS Class C(grouding resistance 10 or
T95C (+203F) 40C to +70C (40F to +158F) less) or JIS Class A(grounding resistance
T105C (+221F) 40C to +85C (40F to +185F) 10 or less)
Note : In case of TIIS Flameproof type, a remote flowtube
T120C (+248F) 40C to +120C (40F to +248F) is available for combined use with the AXFA14
converter only.
T135C (+275F) 40C to +150C (40F to +302F)
*In case that ambient temperature exceeds 50C, use
T27-6.EPS
heat-resistant cables with maximum allowable tempera-
Ambient Temp.:
ture of 70C or above.
PFA Lining; 40C to +60C (40F to +140F)
Ceramics Lining; 15C to +60C (5F to +140F)
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 16
STANDARD PERFORMANCE Polyurethane Rubber /Natural Soft Rubber /
EPDM Rubber Lining:
Accuracy
Flow Velocity Standard Accuracy
Note: The accuracy of a product before shipment is Size mm (in.)
V m/s (ft/s) (Calibration code B)
defined as totalized value at the result of calibration
test in our water actual flow test facility. 25 (1.0) V 0.3 (1.0) 1.0 mm/s
Calibrated conditions in our water actual test facility to 0.3 V 10
are as follows: 400 (16) 0.35% of Rate
(1.0) (33)
Fluid temperature; 20 10C T09.EPS
Ambient temperature; 20 5C Enhanced dual frequency excitation(Option code HF2) :
Length of straight runs; 10 D or more on the Standard accuracy 1 mm/ s
upstream side; 5 D or more
on the downstream side
Reference conditions; Similar to BS EN29104
Size 25 mm (1.0 in.) to 400 mm (16 in.)
% of Rate
(1993); ISO 9104 (1991)
1.2
Pulse Output: 1.0
PFA/Ceramics Lining: 0.8
Pulse 0.6
Standard High Grade Output High
Standard
gradeAccuracy
Accuracy
Size mm Flow Velocity Accuracy Flow Velocity Accuracy Accuracy 0.4
(in.) V m/s (ft/s) (Calibration V m/s (ft/s) (Calibration 0.35
code B) 0.2
code C)
0
2.5 (0.1) V 0.3 (1) 1.0 mm/s
0 1 2 6 8 10
to 0.3 V 10 0.35% of Velocity[m/s]
F14.EPS
15 (0.5) (1) (33) Rate
V 0.15 (0.5) 0.5 mm/s V 0.15 (0.5) 0.5 mm/s Current Output : Pulse output accuracy plus
25 (1.0)
0.15 V 1 0.18% of Rate 0.05% of Span
to 0.15 V 10 0.35% of (0.5) (3.3) 0.2mm/s
200 (8.0) (0.5) Repeatability:
(33) Rate
1 V 10 0.2% of Rate 0.1% of Rate (V 1 m/s (3.3 ft/s))
(3.3) (33) 0.05% of Rate 0.5 mm/s (V < 1 m/s (3.3 ft/s))
250 (10) V 0.15 (0.5) 0.5 mm/s
Maximum Power Consumption:
to 0.15 V 10 0.35% of
400 (16)
Integral Flowmeter: 12W
(0.5) (33) Rate
T08.EPS
Remote Flowtube: Combined with AXFA11: 20W
Enhanced dual frequency excitation(Option code HF2): Combined with AXFA14: 12W
Standard accuracy 1 mm/ s
Insulation Resistance (*1):
Size 2.5 mm (0.1 in.) to 15 mm (0.5 in.) Integral Flowmeter:
% of Rate
1.2 Between power supply terminals and ground
terminalm : 100M at 500V DC
1.0
Between power supply terminals and input /output
0.8
terminals : 100M at 500V DC
Pulse 0.6
Output High
Standard
gradeAccuracy
Accuracy Between ground terminal and input/output
Accuracy 0.4 terminals : 20M at 100V DC
0.35
0.2 Between input/output terminals : 20M at 100V DC
0 Remote Flowtube:
0 1 2 6 8 10 Between excitation current terminal and signal /
Velocity[m/s]
F14.EPS common terminals : 100M at 500V DC
Between signal terminals : 100M at 500V DC
Size 25 mm (1.0 in.) to 400 mm (16 in.)
% of Rate
Between signal terminals and common terminal (C) :
1.2 100M at 500V DC
1.0
0.8
Standard Accuracy
Pulse 0.6
Output High grade Accuracy
Accuracy 0.4
0.35
0.2
0
0 1 2 6 8 10
Velocity[m/s]
F15.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 17
Withstand Voltage (*1): Pressure Equipment Directive:
Integral Flowmeter Module: H
Between power supply terminals and ground terminal : Type of Equipment: Piping
1400V AC for 2 seconds Type of Fluid: Liquid and Gas
Between power supply terminals and input/output Group of Fluid: 1 and 2
terminals : 1400V AC for 2 seconds General-Purpose Use/Submersible Type/Explosion
Remote Flowtube (option code JF3, KF2, CF1, and SF2) proof Type:
Between excitation current terminal and ground
terminal : 1500V AC for 1 minute DN PS PS DN
MODEL CATEGORY(*2)
Between signal terminals and ground terminal : (mm) (*1) (MPa) (*1) (MPa mm)
1500V AC for 1 minute Article 3, (*3)
AXF002G/C 2.5 4 10
Between signal terminals and excitation current paragraph 3
terminal : 2000V AC for 1 minute Article 3, (*3)
AXF005G/C 5 4 20
Remote Flowtube (option code FF1) paragraph 3
Between signal terminals and ground terminal : 500V AXF010G/C 10 4 40
Article 3, (*3)
AC for 1 minute or 600V AC for 1 second paragraph 3
Between signal terminals and excitation current Article 3, (*3)
AXF015G/W/C 15 4 60
terminal : 2000V AC for 1 minute or 2400V AC for paragraph 3
1 second. Article 3, (*3)
AXF025G/W/C 25 4 100
paragraph 3
AXF032G/W/C 32 4 128 II
CAUTION AXF040G/W/C 40 4 160 II
*1: When performing the Insulation Resistance Test or AXF050G/W/C 50 4 200 II
the Withstand Voltage Test, please obey the following AXF065G/W/C 65 2 130 II
caution. AXF080G/W/C 80 2 160 II
Following the relevant test, wait for more than 10
AXF100G/W/C 100 2 200 II
seconds after the power supply has been turned off
before removing the cover. AXF125G/W/C 125 2 250 II
Remove all wires from terminals before testing. AXF150G/W/C 150 2 300 II
When the power terminal has a lighting protector AXF200G/W/C 200 2 400 III
(optional code A), remove the short bar at the ground AXF250G/W/C 250 2 500 III
terminal.
AXF300G/W/C 300 2 600 III
After testing, be sure to discharge by using a resis-
AXF350G/W/C 350 1 350 II
tance and return all wires and the short bar to its
correct position. AXF400G/W/C 400 1 400 III
T10-1.EPS
Screws must be tightened to a torque of 1.18 N-m or
more. Sanitary Type:
After closing the cover, the power supply can be
restored. DN PS PS D
MODEL CATEGORY (*2)
(mm) (*1) (MPa) (*1) (MPa mm)
Article 3, (*3)
Safety Requirement Standards: AXF015H 15 1 15
paragraph 3
EN61010-1
Article 3, (*3)
EN61010-2-030 AXF025H 25 1 25
paragraph 3
Altitude at installation site: Max. 2000 m above sea leve
AXF032H 32 1 32 I
Installation category based on IEC1010:
Overvoltage category II (II applies to electrical AXF040H 40 1 40 I
equipment which is supplied from the fixed installation AXF050H 50 1 50 I
like distribution board.) AXF065H 65 1 65 I
Pollution degree based on IEC1010 80 1 80 I
AXF080H
Pollution degree 2 (Pollution degree describes the
AXF100H 100 1 100 I
degree to which a solid, liquid, or gas which deteriorates
dielectric strength or surface resistivity is adhering. 2 AXF125H 125 1 125 II
T10-2.EPS
applies to a normal indoor atmosphere.) *1: PS: Maximum allowable pressure for Flowtube
DN: Nominal size
EMC Conformity Standards: ,
*2: For details, see Table 6 covered by ANNEX II of EC
EN61326-1 Class A, Table 2 (For use in industrial locations) Directive on Pressure Equipment Directive 97/23/EC.
EN61326-2-3 *3: AXF002G/C to AXF025G/W/C, AXF015H and
EN61000-3-2 Class A AXF025H are outside the scope of CE marking of
PED.
EN61000-3-3
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 18
NORMAL OPERATING CONDITIONS Cable Length and Liquid Conductivity (Remote
Flowtube):
Ambient Temperature: 40 to +60C (40 to +140F)
*1: Minimum temperature should also be limited according Size 2.5 to 10 mm (0.1 to 0.4 in.)
to minimum fluid temperature of flow tubes specifica- m (ft) Add 1% of Rate to the As show in accuracy
tion. accuracy shown in description
Refer to description of Fluid Temperature and accuracy description
Cable length
Pressure.
*2: Indicators operating range (integral flowmeter): 20 to 100 (330) Combined
+60C (4 to +140F) with AXFA11
50 (170) or AXFA14
*3: Maximum temperature should be +50C (+122F) in converter
3 (10)
the case of power supply code 2 (integral flowmeter). 0 (0)
1 5 10 20 50 (S/cm)
Ambient Humidity: 0 to 100% Liquid conductivity
Lengthy continuous operation at 95% or more is not
recommended. Size 15 to 125 mm (0.5 to 5.0 in.)
Power Supply (integral flowmeter): Add 1% of Rate to As show in accuracy
the accuracy shown in description
Power supply code 1: m (ft) accuracy description
AC specifications 200 (660) Combined with
AXFA11
Cable length
Rated power supply: 100 to 240 V AC, 50/60 Hz converter
(Operating voltage range: 80 to 264 V AC) 100 (330) Combined with
DC specifications AXFA14
converter
Rated power supply: 100 to 120 V DC 25 (85) 50
(Operating voltage range: 90 to 130 V DC) 0 (0)
1 3 5 10 20 (S/cm)
Power supply code 2: Liquid conductivity
AC specifications
Rated power supply: 24 V AC, 50/60 Hz Size 150 to 400 mm (6.0 to 16 in.)
(Operating voltage range: 20.4 to 28.8 V AC) Add 1% of Rate to As show in accuracy
the accuracy shown in description
DC specifications m (ft) accuracy description
Rated power supply: 24 V DC 200 (660) Combined with
AXFA11
Cable length
(S/cm)
Liquid conductivity
796 (2610) F03.EPS
600 (1970) Note: In case that size 250 or 300 mm (10 or 12 in.) is
used for high conductivity fluid (ex. caustic soda,
446 (1460) seawater), please use the flange type.
200 ( 660)
0
20.4 22 24 26 28.8
Usable range E (V)
Cable cross section area: 1.25 mm2
Cable cross section area: 2 mm2
F01.EPS
Fluid Conductivity:
Size 2.5 to 10 mm (0.1 to 0.4 in.): 5 S/cm or larger
Size 15 to 125 mm (0.5 to 5 in.): 1 S/cm or larger
Size 150 to 400 mm (6 to 16 in.): 3 S/cm or larger
Note: Fluids with large flow noise (pure water, fluid with
low conductivity and low viscosity such as alcohol)
cause the output fluctuation. Be careful that it affects
the flow rate measurement. It is recommended to
use the CA capacitance magnetic flowmeter.
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 19
Measurable Flow Rate Range: Fluid Temperature and Pressure:
SI Units (Size: mm, Flow rate: m3/h) Note *1 The following figures show maximum allowable
fluid pressure for the flowtube. Further fluid
Size 0 to Min. Span Flow Rate 0 to Max. Span Flow Rate pressure should also be limitted according to
(mm) (0.1 m/s) (10 m/s) flange rating.
2.5 0 to 0.0018 m3/h 0 to 0.1767 m3/h *2 For fluid temperature of the explosion proof
type, refer to descriptions of HAZARDOUS
5 0 to 0.0071 0 to 0.7068 AREA CLASSIFICATION.
10 0 to 0.0283 0 to 2.8274 PFA Lining (*1)
15 0 to 0.0637 0 to 6.361 General-Purpose Use, Submersible Type, Explosion
25 0 to 0.1768 0 to 17.671 proof Type, Remote Flowtube (electrode structure
code 1: Non-replaceable electrode)
32 0 to 0.2896 0 to 28.952
2.5 to 50 mm (0.1 to 2.0 in.) (flange type, wafer type)
40 0 to 0.4524 0 to 45.23
65 to 200 mm (2.5 to 8.0 in.) (flange type, wafer type)
50 0 to 0.7069 0 to 70.68 250, 300 mm (10, 12 in.) (flange type)
65 0 to 1.1946 0 to 119.45 Pressure 250, 300 mm (10, 12 in.) (wafer type)
MPa (psi) 350, 400 mm (14, 16 in.) (flange type)
80 0 to 1.8096 0 to 180.95
4 (580)
100 0 to 2.8275 0 to 282.74 *1
125 0 to 4.418 0 to 441.7
2 (290)
150 0 to 6.362 0 to 636.1
1 (145)
200 0 to 11.310 0 to 1,130.9
250 0 to 17.672 0 to 1,767.1 0.1 (14.5)
40 10 0 40 100 130 150 160
300 0 to 25.447 0 to 2,544.6 (40) (14) (32) (104) (212) (266) (302) (320)
Temperature C (F) F18-1.EPS
350 0 to 34.64 0 to 3,463
400 0 to 45.24 0 to 4,523 General-Purpose Use and Explosion proof Type,
T11.EPS Integral Flowmeter (electrode structure code 1:
Non-replaceable electrode)
English Units (Size: in., Flow rate: GPM)
Size 0 to Min. Span Flow Rate 0 to Max. Span Flow Rate 2.5 to 50 mm (0.1 to 2.0 in.) (flange type, wafer type)
(in.) (0.33ft/s) (33ft/s) 65 to 200 mm (2.5 to 8.0 in.) (flange type, wafer type)
250, 300 mm (10, 12 in.) (flange type)
0.1 0 to 0.0078 GPM 0 to 0.7780 GPM
Pressure 250, 300 mm (10, 12 in.) (wafer type)
0.2 0 to 0.0312 0 to 3.112 MPa (psi) 350, 400 mm (14, 16 in.) (flange type)
3.0 0 to 7.003 0 to 700.2 *1: For wafer types of 32 mm to 300 mm(1.25 to 12 in.)
and carbon steel flange types (process connection
4.0 0 to 12.45 0 to 1,244 code: C**) of 50 to 400 mm (2.0 to 16 in.), the mini-
5.0 0 to 19.46 0 to 1,945 mum temperature is 10C (14F).
*2: For fluid temperature of the explosion proof type,
6.0 0 to 28.01 0 to 2,800 refer to descriptions of HAZARDOUS AREA
CLASSIFICATION.
8.0 0 to 49.80 0 to 4,979
10 0 to 77.81 0 to 7,780
12 0 to 112.1 0 to 11,203
14 0 to 152.5 0 to 15,249
16 0 to 199.2 0 to 19,918
T24.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 20
General-Purpose Use, Remote Flowtube (electrode General-Purpose Use and Explosion proof Type,
structure code 2: replaceable electrode) Integral flowmeter (electrode structure code 1: Non-
replaceable electrode)
25 to 200 mm (1.0 to 8.0 in.) (flange type, wafer type)
250, 300 mm (10, 12 in.) (flange type) Pressure 2.5 to 50 mm (0.1 to 2.0 in.)
250, 300 mm (10, 12 in.) (wafer type) MPa (psi) 80 to 200 mm (3.0 to 8.0 in.)
Pressure 350, 400 mm (14 to 16 in.) (flange type)
4 (580)
MPa (psi)
3.25 (471)
2 (290)
1 (145) 2 (290)
0.1 (14.5)
10 0 40 100 130 160
(14) (32) (104) (212) (266) (320) 0.1(14.5)
10 0 100 130 180
Temperature C (F) F18-3.EPS (14) (32) (212)(266) (356)
Temperature C (F)
General-Purpose Use, Integral Flowmeter (electrode F19-2.EPS
structure code 2: replaceable electrode) *1: For fluid temperature of the explosion proof type,
refer to descriptions of HAZARDOUS AREA
25 to 200 mm (1.0 to 8.0 in.) (flange type, wafer type) CLASSIFICATION.
250, 300 mm (10, 12 in.) (flange type)
Pressure 250, 300 mm (10, 12 in.) (wafer type) Polyurethane Rubber Lining
MPa (psi)
350, 400 mm (14 to 16 in.) (flange type) General-Purpose Use and Submersible Type,
2 (290)
Remote Flowtube (electrode structure code 1: Non-
1 (145) replaceable electrode)
25 to 50 mm (1.0 to 2.0 in.) (flange type, wafer type)
0.1 (14.5)
10 0 40 100 130
65 to 200 mm (2.5 to 8.0 in.) (flange type, wafer type)
(14) (32) (104) (212) (266)
250, 300 mm (10, 12 in.) (flange type)
Temperature C (F)
F18-4.EPS
250, 300 mm (10, 12 in.) (wafer type)
Note: For replaceable electrodes for fluid temperatures of 350, 400 mm (14, 16 in.) (flange type)
Pressure
10C (14F)or less, please contact Yokogawa MPa (psi)
office. 4 (580)
0.1(14.5)
10 0 130 160 General-Purpose Use, Integral Flowmeter (electrode
(14) (32) (266) (320) structure code 2: replaceable electrode)
Temperature C (F) F18-5.EPS
25 to 200 mm (1.0 to 8.0 in.) (flange type, wafer type)
Note: In case of 120 to 160C (248 to 320F) of fluid 250, 300 mm (10, 12 in.) (flange type)
temperature, please select optional code GH.
250, 300 mm (10 , 12 in.) (wafer type)
350, 400 mm (14 , 16 in.) (flange type)
Ceramics Lining Pressure
MPa (psi)
General-Purpose Use and Explosion proof Type,
Remote Flowtube (electrode structure code 1: Non- 2 (290)
replaceable electrode) 1 (145)
2 (290)
0.1 (14.5)
10 0 100 180
(14) (32) (212) (356)
Temperature C (F)
F19-1.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 21
Natural Soft Rubber Lining Reasonable Figure for Thermal Shock of Creamics Lining:
General-Purpose Use and Submersible Type, Size 2.5 to 25 mm (0.1 to 1.0 in.)
Remote Flowtube (electrode structure code 1: Non- TC(F)
replaceable electrode) 120 (248)
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 22
CAUTIONS FOR INSTALLATION Avoiding Air Bubbles.
If air bubbles enter a measurement pipe, flow rate
Mounting of Flowmeters and Required Lengths indications may be affected and measurement errors
of Straight Runs may be caused.
Based on JIS B 7554 and our piping condition test data, In cases where fluids contain air bubbles, piping must
we recommend the piping conditions as shown in the be designed to prevent them from accumulating in the
following figures. measurement pipe of a flowtube.
When installing two or more magnetic flowmeters on a If a valve exists near the flowtube, try to mount the
single pipe, provide a run of at least 10D between them. flowtube on the valves upstream side in order to
prevent a possible reduction of pressure inside the
D: Flowtube Size pipe, thereby avoiding the possibility of air bubbles.
Gate valve Reducer Expander
fully open pipe pipe
(Correct) (Incorrect)
(Correct)
2D
5D or more or more 0 is allowable. 0 is allowable. 10D or more 2D or more
90-degree bent (Incorrect)
Tee Various valves Valve
F10.EPS
2D
5D or more 0 is allowable. 5D or more 0 is allowable.10D or more or more Avoiding Air Bubbles
F08.EPS
Mounting Orientation
Required straight runs
If electrodes are perpendicular to the ground, air
*1: Do not install anything in the vicinity that may interfere bubbles near the top or precipitates at the bottom may
with the magnetic field, induced signal voltages, or flow cause measurement errors.
velocity distributions of the flowmeter.
*2: A straight run may not be required on the downstream
Ensure that the terminal box of a remote flowtube and
side of the flowmeter. However, if a downstream valve converter of an integral flowmeter are mounted above
or other fitting causes irregularity or deviation in flows, the piping to prevent water from entering them.
provide a straight run of 2D to 3D on the downstream
(Correct)
side.
*3: The valves shall be mounted on the downstream side (Incorrect) (Incorrect)
so that deviated flows do not occur in the flowtube and
to avoid startup from an empty condition. Air bubble
Mounting Positions
Pipes must be fully filled with liquids.
It is essential that pipes remain fully filled at all times,
otherwise flow rate indications may be affected and
measurement errors may be caused.
Piping shall be designed so as to maintain the flowtube
filled with fluids.
Vertical mounting is effective in such cases as when
fluids tend to separate or solid matter may be precipi-
tated. When employing vertical mounting, direct the
fluids from the bottom to the top to ensure that the
pipes remain fully filled.
(Correct) (Incorrect)
h
(Correct) (Incorrect) h>0
h
h>0
F09.EPS
Mounting Positions
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 23
ACCESSORIES Recommended Excitation, Power and Output
Centering device (wafer type only): 1 pc.
Cable:
Hexagonal wrench: 2 pcs. JIS C 3401 control cable equivalent
JIS C 3312 power cable equivalent
14 AWG Belden 8720 equivalent
TERMINAL CONFIGURATION AND Outer Diameter:
With no gland option;
TERMINAL WIRING 6.5 to 12 mm (0.26 to 0.47 in.)
With gland options EG,EU and EW;
Integral Flowmeter
10.5 or 11.5 mm (0.41 or 0.45 in.)
With gland options EP;
Terminal configuration
6 to 12 mm (0.24 to 0.47 in.)
Nominal Cross Section:
Single wire: 0.5 to 2.5 mm2
+ L/+ N/ -
SUPPLY
POWER
+-
DO
+-
DIO
Terminal wiring
Terminal
Symbols Description
Functional grounding
N/-
L/+ Power supply
I+ Current output 4 to 20mA DC
I-
DO+ Pulse output/Alarm output/
DO- Status output
DIO+ Alarm output/Status output
DIO- Status input
Protective grounding
(Outside of the terminal) F41.EPS
Remote Flowtube
Terminal configuration
EX1
C
C
EX1
EX2
B
B
EX2
A
A
For other than Explosion proof type For Explosion proof type
F42.EPS
Terminal wiring
Terminal
Symbols Description
A
B Flow signal output
C
EX1
Excitation current input
EX2
Functional grounding
(Only for explosion proof type)
Protective grounding
(Outside of the terminal) F42_2.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 24
MODEL AND SUFFIX CODE
AXF STANDARD (Wafer Type)
General-purpose Use/Submersible Type/Explosion
proof Type, PFA/Polyurethane Rubber/Natural Soft
Rubber/EPDM Rubber Lining
Model Suffix Code Description Applicable Model
AXF002 Size 2.5 mm (0.1 in.) Integral Flowmeter/Remote Flowtube
AXF005 Size 5 mm (0.2 in.) Integral Flowmeter/Remote Flowtube
AXF010 Size 10 mm (0.4 in.) Integral Flowmeter/Remote Flowtube
AXF015 Size 15 mm (0.5 in.) Integral Flowmeter/Remote Flowtube
AXF025 Size 25 mm (1.0 in.) Integral Flowmeter/Remote Flowtube
AXF032 Size 32 mm (1.25 in.) Integral Flowmeter/Remote Flowtube
AXF040 Size 40 mm (1.5 in.) Integral Flowmeter/Remote Flowtube
AXF050 Size 50 mm (2.0 in.) Integral Flowmeter/Remote Flowtube
AXF065 Size 65 mm (2.5 in.) Integral Flowmeter/Remote Flowtube
AXF080 Size 80 mm (3.0 in.) Integral Flowmeter/Remote Flowtube
AXF100 Size 100 mm (4.0 in.) Integral Flowmeter/Remote Flowtube
AXF125 Size 125 mm (5.0 in.) Integral Flowmeter/Remote Flowtube
AXF150 Size 150 mm (6.0 in.) Integral Flowmeter/Remote Flowtube
AXF200 Size 200 mm (8.0 in.) Integral Flowmeter/Remote Flowtube
AXF250 Size 250 mm (10 in.) Integral Flowmeter/Remote Flowtube
AXF300 Size 300 mm (12 in.) Integral Flowmeter/Remote Flowtube
Use G General-Purpose Use
W Submersible Type Size 15 mm (0.5 in.) to 300 mm (12 in.) Remote Flowtube only
C Explosion proof Type (*5) PFA lining only
Converter -D Integral Flowmeter with 4 to 20mA DC Output and BRAIN Communication
Output Signal -E Integral Flowmeter with 4 to 20 mA DC Output and HART Communication
and -F Integral Flowmeter with Digital communication
Communication (FOUNDATION Fieldbus protocol) (*9)
-G Integral Flowmeter with Digital communication
(PROFIBUS PA protocol) (*10)
-N Remote Flowtube for Combined Use with AXFA11
-P Remote Flowtube for Combined Use with AXFA14
Power Supply 1 Integral Flowmeter, 100 V to 240 V AC or 100 to 120 V DC
2 Integral Flowmeter, 24V AC/DC
N Remote Flowtube
Lining (*8) A Fluorocarbon PFA
U Polyurethane Rubber Size 25 mm (1.0 in.) to 300 mm (12 in.)
D Natural Soft Rubber Size 50 mm (2.0 in.) to 300 mm (12 in.)
G EPDM Rubber Size 50 mm (2.0 in.) to 300 mm (12 in.)
Electrode L JIS SUS316L (AISI 316L SS/EN 1.4404 Equivalent)
Material (*8) P Platinum-iridium PFA lining only
H Hastelloy C276 Equivalent
T Tantalum PFA lining only
V Titanium
W Tungsten Carbide Size 15 mm (0.5 in.) to 300 mm (12 in.),
PFA/Polyurethane Rubber lining only
Electrode 1 Non-replaceable
Structure 2 Replaceable General-Purpose use, Size 25 mm (1.0 in.) to 300 mm (12 in.),
PFA/Polyurethane Rubber lining only
Electrode Material: JIS SUS316L only
Grounding Ring and N None
Grounding Electrode S JIS SUS316 (AISI 316 SS/EN 1.4401 Equivalent)
Material (*8) L JIS SUS316L (AISI 316L SS/EN 1.4404 Equivalent)
P Platinum-iridium Size 2.5 mm (0.1 in.) to 200 mm (8.0 in.), PFA lining only
H Hastelloy C276 Equivalent
T Tantalum Size 2.5 mm (0.1 in.) to 200 mm (8.0 in.), PFA lining only
V Titanium
Process Connection -AA1 ANSI Class 150 Wafer (*1) Size 2.5 mm (0.1 in.) to 300 mm (12 in.)
(*3)(*11) -AA2 ANSI Class 300 Wafer (*1) Size 2.5 mm (0.1 in.) to 200 mm (8.0 in.)
-AD1 DIN PN 10 Wafer (*2) Size 200 mm (8.0 in.) to 300 mm (12 in.)
-AD2 DIN PN 16 Wafer (*2) Size 65 mm (2.5 in.) to 300 mm (12 in.)
-AD4 DIN PN 40 Wafer (*1)(*2) Size 2.5 mm (0.1in.) to 50 mm (2.0 in.)
-AJ1 JIS 10K Wafer (*1) Size 2.5 mm (0.1in.) to 300 mm (12 in.)
-AJ2 JIS 20K Wafer (*1) Size 2.5 mm (0.1in.) to 200 mm (8.0 in.)
-AG1 JIS F12 (JIS75M) Wafer Size 80 mm (3.0 in.) to 300 mm (12 in.)
Lay Length 1 Standard
Electrical Connection (*6) -0 JIS G1/2 female
-2 ANSI 1/2 NPT female Not available for Submersible Type
-4 ISO M201.5 female Not available for Submersible Type
Indicator (*4)(*7) 1 Integral Flowmeter with indicator (Horizontal)
2 Integral Flowmeter with indicator (Vertical)
N Integral Flowmeter without indicator /Remote Flowtube
Calibration B Standard
C High Grade (*12) Size 25 mm (1.0 in.) to 200 mm (8.0 in.), PFA lining only
/ Optional code (See the Table of Optional Specifications)
T15.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 25
*1: For a wafer type of 2.5 to 10 mm (0.1 to 0.4 in.), prepare 15 mm (0.5 in.) diameter nominal flanges on the process pipe side.
(Process connection codes: AA1, AA2, AD4, AJ1, and AJ2)
*2: Even when DIN PN10 or 16 is required for a model of size 2.5 to 50 mm (0.1 to 2.0 in.), select PN40 (Process connection code: AD4) because there is no
difference in the dimensions of the mating faces.
Even when DIN PN10 is required for a model of size 65 to 150 mm (2.5 to 6.0 in.), select PN16 (Process connection code: AD2) because there is no difference
in the dimensions of the mating faces. 1 2 N
*3: Mating dimensions are based on standards as follow:
ANSI:ASME B 16.5, DIN: DIN 2501, JIS:JIS B 2220 and JIS G 3443-2
*4: N shall be always selected for remote flowtubes
In the case of an integral flowmeter, select from among the figures at the right:
*5: For explosion proof types, specify types of explosion proof certification using the optional codes. In case of TIIS flameproof type, the remote flowtube is available
only for combined use with the AXFA14. For the TIIS flameproof type with wiring using a flameproof packing adapter, select optional code G12 or G11. Available
only for JIS G1/2 female electrical connections.
*6: For explosion proof types, JIS G1/2 female electrical connection is available only for TIIS flameproof type.
*7: In case of integral flowmeters of the TIIS flameproof type, select with indicator(code 1 or 2).
*8: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is
also possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature
steam (150C [302F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*9: For FOUNDATION Fieldbus protocol, refer to GS 01E20F02-01E
*10: For PROFIBUS PA protocol, refer to GS 01E20F12-01E.
*11: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
*12: In the case of calibration code C in remote type, please order a combination of flowtube and converter.
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 26
AXF STANDARD (Wafer /Union Joint Type)
General-purpose Use/Explosion proof Type, Ceramics Lining
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 27
*1: Even when DIN PN10 or 16 is required for a model of size 2.5 to 50 mm (0.1 to 2.0 in.), select PN40 (Process connection code : AD4)
because there is no difference in the dimensions of the mating faces.
Even when DIN PN10 is required for a model of size 65 to 150 mm (2.5 to 6.0 in.), select PN16 (Process connection code : AD2) because there is no difference
in the dimensions of the mating faces.
*2: Mating dimensions are based on standards as follow: 1 2 N
ANSI:ASME B 16.5, DIN: DIN 2501, JIS:JIS B 2220 and JIS G 3443-2
*3: AXF standard lay length dimension for ceramics linings are the same as
those for ADMAG ceramics linings.
*4: N shall be always selected for remote flowtubes
In the case of an integral flowmeter, select from among the figures at the right:
*5: For explosion proof types, specify types of explosion proof certification using the optional codes. In case of TIIS flameproof type, the remote flowtube is
available only for combined use with the AXFA14. For the TIIS flameproof type with wiring using a flameproof packing adapter, select optional code G12 or G11.
Available only for JIS G1/2 female electrical connections.
*6: For explosion proof types, JIS G1/2 female electrical connection is available only for TIIS flameproof type.
*7: In case of integral flowmeters of the TIIS flameproof type, select with indicator(code 1 or 2).
*8: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is
also possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature
steam (150C [302F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*9: For FOUNDATION Fieldbus protocol, refer to GS 01E20F02-01E
*10: For PROFIBUS PA protocol, refer to GS 01E20F12-01E.
*11: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
*12: In the case of calibration code C in remote type, please order a combination of flowtube and converter.
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 28
AXF STANDARD (Flange Type)
General-purpose Use/Submersible Type/Explosion proof Type, PFA/Polyurethane Rubber /Natural Soft Rubber/
EPDM Rubber Lining
Model Suffix Code Description Applicable Model
AXF002 Size 2.5 mm (0.1 in.) Integral Flowmeter/Remote Flowtube
AXF005 Size 5 mm (0.2 in.) Integral Flowmeter/Remote Flowtube
AXF010 Size 10 mm (0.4 in.) Integral Flowmeter/Remote Flowtube
AXF015 Size 15 mm (0.5 in.) Integral Flowmeter/Remote Flowtube
AXF025 Size 25 mm (1.0 in.) Integral Flowmeter/Remote Flowtube
AXF032 Size 32 mm (1.25 in.) Integral Flowmeter/Remote Flowtube
AXF040 Size 40 mm (1.5 in.) Integral Flowmeter/Remote Flowtube
AXF050 Size 50 mm (2.0 in.) Integral Flowmeter/Remote Flowtube
AXF065 Size 65 mm (2.5 in.) Integral Flowmeter/Remote Flowtube
AXF080 Size 80 mm (3.0 in.) Integral Flowmeter/Remote Flowtube
AXF100 Size 100 mm (4.0 in.) Integral Flowmeter/Remote Flowtube
AXF125 Size 125 mm (5.0 in.) Integral Flowmeter/Remote Flowtube
AXF150 Size 150 mm (6.0 in.) Integral Flowmeter/Remote Flowtube
AXF200 Size 200 mm (8.0 in.) Integral Flowmeter/Remote Flowtube
AXF250 Size 250 mm (10 in.) Integral Flowmeter/Remote Flowtube
AXF300 Size 300 mm (12 in.) Integral Flowmeter/Remote Flowtube
AXF350 Size 350 mm (14 in.) Integral Flowmeter/Remote Flowtube
AXF400 Size 400 mm (16 in.) Integral Flowmeter/Remote Flowtube
Use G General-Purpose Use
W Submersible Type Size 15 mm (0.5 in.) to 400 mm (16 in.), Remote Flowtube only
C Explosion proof Type (*6) PFA lining only
Converter -D Integral Flowmeter with 4 to 20 mA DC Output and BRAIN Communication
Output Signal -E Integral Flowmeter with 4 to 20 mA DC Output and HART Communication
and -F Integral Flowmeter with Digital communication
Communication (FOUNDATION Fieldbus protocol) (*10)
-G Integral Flowmeter with Digital communication
(PROFIBUS PA protocol) (*11)
-N Remote Flowtube for Combined Use with AXFA11
-P Remote Flowtube for Combined Use with AXFA14
Power Supply 1 Integral Flowmeter, 100 V to 240 V AC or 100 to 120 V DC
2 Integral Flowmeter, 24 V AC/DC
N Remote Flowtube
Lining (*9) A Fluorocarbon PFA
U Polyurethane Rubber Size 25 mm (1.0 in.) to 400 mm (16 in.)
D Natural Soft Rubber Size 50 mm (2.0 in.) to 400 mm (16 in.)
G EPDM Rubber Size 50 mm (2.0 in.) to 400 mm (16 in.)
Electrode Material (*9) L JIS SUS316L (AISI 316L SS/EN 1.4404 Equivalent)
P Platinum-iridium PFA lining only
H Hastelloy C276 Equivalent
T Tantalum PFA lining only
V Titanium
W Tungsten Carbide Size 15 mm (0.5 in.) to 400 mm (16 in.),
PFA/Polyurethane Rubber lining only
Electrode Structure 1 Non-replaceable
2 Replaceable General-Purpose use, Size 25 mm (1.0 in.) to 400 mm (16 in.)
PFA/Polyurethane Rubber lining only
Electrode Material: JIS SUS316L only
Grounding Ring and N None
Grounding Electrode S JIS SUS316 (AISI 316 SS/EN 1.4401 Equivalent)
Material (*9) L JIS SUS316L (AISI 316L SS/EN 1.4404 Equivalent)
P Platinum-iridium Size 2.5 mm (0.1 in.) to 200 mm (8.0 in.), PFA lining only
H Hastelloy C276 Equivalent
T Tantalum Size 2.5mm (0.1 in.) to 200 mm (8.0 in.), PFA lining only
V Titanium
Process Connection -BA1 ANSI Class 150 Flange (Stainless Steel)(*1) Size 2.5 mm (0.1 in.) to 400 mm (16 in.)
(*4)(*12) -BA2 ANSI Class 300 Flange (Stainless Steel)(*1) Size 2.5 mm (0.1 in.) to 300 mm (12 in.)
-BD1 DIN PN 10 Flange (Stainless Steel)(*2) Size 200 mm (8.0 in.) to 400 mm (16 in.)
-BD2 DIN PN 16 Flange (Stainless Steel)(*2) Size 65 mm (2.5 in.) to 300 mm (12 in.)
-BD4 DIN PN 40 Flange (Stainless Steel)(*1)(*2) Size 2.5 mm (0.1 in.) to 50 mm (2.0 in.)
-BJ1 JIS 10K Flange (Stainless Steel)(*1) Size 2.5 mm (0.1 in.) to 400 mm (16 in.)
-BJ2 JIS 20K Flange (Stainless Steel)(*1) Size 2.5 mm (0.1 in.) to 300 mm (12 in.)
-BG1 JIS F12 (JIS75M) Flange (Stainless Steel) Size 80 mm (3.0 in.) to 400 mm (16 in.)
-CA1 ANSI Class 150 Flange (Carbon Steel) Size 50 mm (2 in.) to 400 mm (16 in.), Not available for TIIS flameproof type.
-CA2 ANSI Class 300 Flange (Carbon Steel) Size 50 mm (2 in.) to 300 mm (12 in.), Not available for TIIS flameproof type.
-CD1 DIN PN 10 Flange (Carbon Steel)(*2) Size 200 mm (8.0 in.) to 400 mm (16 in.), Not available for TIIS flameproof type.
-CD2 DIN PN 16 Flange (Carbon Steel)(*2) Size 65 mm (2.5 in.) to 300 mm (12 in.), Not available for TIIS flameproof type.
-CD4 DIN PN 40 Flange (Carbon Steel)(*2) Size 50 mm (2.0 in.) only, Not available for TIIS flameproof type.
-CJ1 JIS 10K Flange (Carbon Steel) Size 50 mm (2.0 in.) to 400 mm (16 in.), Not available for TIIS flameproof type.
-CJ2 JIS 20K Flange (Carbon Steel) Size 50 mm (2.0 in.) to 300 mm (12 in.), Not available for TIIS flameproof type.
-CG1 JIS F12 (JIS75M) Flange (Carbon Steel) Size 80 mm (3.0 in.) to 400 mm (16 in.), Not available for TIIS flameproof type.
-DD4 DIN PN 40 Flange (Stainless Steel), DN10(*2)(*3)" Size 2.5 mm (0.1 in.) to 10 mm (0.4 in.)
-DJ1 JIS 10K Flange (Stainless Steel ), 10 mm Diameter Nominal (*3) Size 2.5 mm (0.1 in.) to 10 mm (0.4 in.)
-DJ2 JIS 20K Flange (Stainless Steel ), 10 mm Diameter Nominal (*3) Size 2.5 mm (0.1 in.) to 10 mm (0.4 in.)
Lay Length 1 Standard
Electrical Connection (*7) -0 JIS G1/2 female
-2 ANSI 1/2 NPT female Not available for Submersible Type
-4 ISO M201.5 female Not available for Submersible Type
Indicator (*5)(*8) 1 Integral Flowmeter with indicator (Horizontal)
2 Integral Flowmeter with indicator (Vertical)
N Integral Flowmeter without indicator /Remote Flowtube
Calibration B Standard
C High Grade (*13) Size 25 mm (1.0 in.) to 200 mm (8.0 in.),PFA lining only
/ Optional code (See the Table of Optional Specifications)
T17.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 29
*1: For a flange type of 2.5 to 10 mm (0.1 to 0.4 in.), prepare 15 mm (0.5 in.) diameter nominal flanges on the process pipe side.
(Process connection codes: BA1, BA2, BD4, BJ1, and BJ2)
*2: Even when DIN PN10 or 16 is required for a model of size 2.5 to 50 mm (0.1 to 2.0 in.), select PN40 (Process connection codes: BD4,CD4 and DD4) because
there is no difference in the dimensions of the mating faces.
Even when DIN PN10 is required for a model of size 65 to 150 mm (2.5 to 6.0 in.), select PN16 (Process connection codes: BD2 , CD2) because there is no
difference in the dimensions of the mating faces.
*3: For a flange type of 2.5 to 10 mm (0.1 to 0.4 in)(Process connection codes: DJ1, DJ2, and DD4), prepare 10 mm (0.4 in.) diameter nominal flanges on the
process pipe side.
*4: Mating dimensions are based on standards as follow: 1 2 N
ANSI:ASME B 16.5, DIN: DIN 2501, JIS:JIS B 2220 and JIS G 3443-2
*5: N shall be always selected for remote flowtubes
In the case of an integral flowmeter, select from among the figures at the right:
*6: For explosion proof types, specify types of explosion proof certification using the optional codes. In case of TIIS flameproof type, the remote flowtube is
available only for combined use with the AXFA14. For the TIIS flameproof type with wiring using a flameproof packing adapter, select optional code G12 or
G11. Available only for JIS G1/2 female electrical connections.
*7: For explosion proof types, JIS G1/2 female electrical connection is available only for TIIS flameproof type.
*8: In case of integral flowmeters of the TIIS flameproof type, select with indicator(code 1 or 2).
*9: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is
also possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature
steam (150C [302F] or above).
Contact Yokogawa for detailed information of the wetted parts material.
*10:For FOUNDATION Fieldbus protocol, refer to GS 01E20F02-01E
*11: For PROFIBUS PA protocol, refer to GS 01E20F12-01E.
*12: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
*13: In the case of calibration code C in remote type, please order a combination of flowtube and converter.
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 30
AXF STANDARD (Clamp/Union/Butt Weld Connection)
Sanitary Type , PFA Lining
*4: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is
also possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature
steam (150C [302F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*5: For FOUNDATION Fieldbus protocol, refer to GS 01E20F02-01E
*6: For PROFIBUS PA protocol, refer to GS 01E20F12-01E.
*7: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
*8: In the case of calibration code C in remote type, please order a combination of flowtube and converter.
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 31
REPLACEMENT MODEL FOR EARLIER ADMAG OR ADMAG AE (Wafer Type)
General-purpose Use/Submersible Type/Explosion proof Type, PFA/Polyurethane Rubber Lining
For the Wafer Types of size 250 mm (10 in.), 300 mm (12 in.), AXF Standard shall be selected.
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 32
*1: For a wafer type of 2.5 to 10 mm (0.1 to 0.4 in.), prepare 15 mm (0.5 in.) diameter nominal flanges on the process pipe side.
(Process connection codes: AA1, AA2, AD4, AJ1, and AJ2)
*2: Even when DIN PN10 or 16 is required for a model of size 2.5 to 50 mm (0.1 to 2.0 in.), select PN40 (Process connection code: AD4) because there is no
difference in the dimensions of the mating faces.
Even when DIN PN10 is required for a model of size 65 to 150 mm (2.5 to 6.0 in.), select PN16 (Process connection code: AD2) because there is no difference
in the dimensions of the mating faces. 1 2 N
*3: Mating dimensions are based on standards as follow:
ANSI:ASME B 16.5, DIN: DIN 2501, JIS:JIS B 2220 and JIS G 3443-2
*4: N shall be always selected for remote flowtubes
In the case of an integral flowmeter, select from among the figures at the right:
*5: For explosion proof types, specify types of explosion proof certification using the optional codes. In case of TIIS flameproof type, the remote flowtube is available
only for combined use with the AXFA14. For the TIIS flameproof type with wiring using a flameproof packing adapter, select optional code G12 or G11. Available
only for JIS G1/2 female electrical connections.
*6: For explosion proof types, JIS G1/2 female electrical connection is available only for TIIS flameproof type.
*7: In case of integral flowmeters of the TIIS flameproof type, select with indicator(code 1 or 2).
*8: In case of platinum-iridium (grounding ring code P) or tantalum (grounding ring code T) or None (grounding ring code N) in wafer type of 2.5 mm (0.1 in.) to 15
mm (0.5 in.), the lay lengths of Replacement model are the same as those for AXF Standard.
In this case, AXF Standard shall be selected.
*9: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is
also possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature
steam (150C [302F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*10:For FOUNDATION Fieldbus protocol, refer to GS 01E20F02-01E
*11: For PROFIBUS PA protocol, refer to GS 01E20F12-01E.
*12: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 33
REPLACEMENT MODEL FOR EARLIER ADMAG OR ADMAG AE (Flange Type)
General-purpose Use/Submersible Type/Explosion proof Type, PFA/Polyurethane Rubber Lining
For Flange Types of size 15 mm (0.5 in.) to 100 mm (4.0 in.), 300 mm (12 in.) to 400 mm (16 in.), AXF Standard shall be
selected.
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 34
*1: Even when DIN PN10 is required for a 150 (6.0 in.)-mm model, select PN16 (Process connection code: CD2) because there is no difference in the dimensions of the
mating faces.
*2: Mating dimensions are based on standards as follow:
ANSI: ASME B 16.5, DIN:DIN 2501, JIS:JIS B 2220 and JIS G 3443-2
*3: N shall be always selected for remote flowtubes.
In the case of an integral, select from among the following figures:
1 2 N
*4: For explosion proof types, specify types of explosion proof certification using the optional codes. In case of TIIS flameproof type, the remote flowtube is available only for
combined use with the AXFA14. For the TIIS flameproof type with wiring using a flameproof packing adapter, select optional code G12 or G11. Available only for JIS G1/2
female electrical connections.
*5: For explosion proof types, JIS G1/2 female electrical connection is available only for TIIS flameproof type.
*6: In case of integral flowmeters of the TIIS flameproof type, select with indicator(code 1 or 2).
*7: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is also
possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature steam
(150C [302F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*8: For FOUNDATION Fieldbus protocol, refer to GS 01E20F02-01E
*9: For PROFIBUS PA protocol, refer to GS 01E20F12-01E.
*10: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 35
OPTIONAL SPECIFICATIONS FOR FLOWTUBES
Table of Optional Specifications
G
D
N
E
AXF***G- P
AXF***C- P
AXF***W- P
AXF***H- P
AXF***G- F
AXF***C- F
AXF***H- F
For District Urethane resin potting is applied in the terminal box of a remote flowtube.
Heating and Select JIS G1/2 for the electrical connections. 30-meter signal and excitation
DHC
Cooling or cables are pre-wired and waterproof grands with union joints are attached at
Condensation-proof factory.
User-specified Available for the submersible type and a model with optional code DHC. The
Signal and Excitation cable length is limited up to 200 meters when combined with an AXFA11
Cable Length converter, or 100 meters when combined with an AXFA14 converter.
Following L, specify the cable length in three digits as a multiple of 1 meter L***
(e.g., 001, 002, or 005) for a length up to 5 meters, or as a multiple of 5
meters (i.e., 005, 010, 015, or the like) for a length of 5 meters or more. If this
optional code is not selected, a 30-meter length cable is attached.
Burn Out Down (*9) The output level is set to 0 mA during a CPU failure and is set 2.4 mA (-10 %)
or less during an alarm. Standard products are delivered with a setting 25 mA C1
during a CPU failure and 21.6 mA (110%) or more during an alarm.
NAMUR NE43 Output signal Failure alarm down-scale: The output level is set to 0 mA during a
C2
Compliance (*9) limits: 3.8 to CPU failure and is set 2.4 mA (10%) or less during an alarm.
20.5 mA Failure alarm up-scale: The output level is set to 25 mA during a
C3
CPU failure and is set 21.6 mA (110%) or more during an alarm.
Active Pulse Output Active pulses are output in order to drive an external electromagnetic or
(*9) electronic counter directly using the converters internal power supply.
(Nullifies the standard transistor contact pulse output.)
Output voltage: 24 V DC 20%
EM
Pulse specifications:
The drive current of 150 mA or less
Pulse rate: 0.0001 to 2 pps (pulse/second); Pulse width: 20, 33, 50, or
100 ms
Mass Unit Setting The flow rate span, transmission pulse weight, and totalizer display pulse
weight can be set in terms of mass unit. Specify the density of the process
fluid when ordering in addition to the mass flow rate span, transmission pulse
wight (for mass unit), and totalizer display pulse weight (for mass unit).
When ordering a remote flowtube, parameters for 'Mass Unit Setting' will be
set in the corresponding converter before shipment.
1. Density
a. Available density Numerics:
Specify the numeric within the value of 500 to 2000 kg/m3, 4.2 to
16.7 lb/gal, or 31.2 to 124.8 lb/cf. And it can be up to five digits, to a
maximum of 32000 ignoring the decimal point. A fraction is limited to
the fourth decimal place.
b. Available density units: kg/m3, lb/gal, lb/cf
Example: A water density is about 1000kg/m3. In this case specify
1000kg/m3.
MU
However a density is changed by temperature.
Specify the actual density.
(The 1000kg/m3 is equivalent to 8.345lb/gal and 62.43lb/cf.)
2. The mass flow rate span, transmission pulse weight, and totalizer display
pulse weight
a. Available density Numerics:
Specify the numeric within the value of 0.0001 to 32000. And it can
be up to five digits, to a maximum of 32000 ignoring the decimal
point. A fraction is limited to the fourth decimal place.
b. Mass Units Available mass units: t, kg, g, klb, lb
Available time units: /d, /h, /min, /s
Note1: In case of specifying the mass flow span, calculate the volumetric flow
span by the setting density, and specify the available value in the mass
flow span.
Note2: In case of transmission pulse weight and totalizer display pulse weight,
specify the mass unit which was specified as the flow unit.
T26-1.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 36
Table of Optional Specifications (continued)
: Available : Not available
Applicable Model
General Explosion proof Submersible Sanitary
Integral Remote Integral Remote Remote Integral Remote
Item Specifications Flowmeter Flowtube Flowmeter Flowtube Flowtube Flowmeter Flowtube Code
G
D
D
N
N
E
E
AXF***G- P
AXF***C- P
AXF***W- P
AXF***H- P
AXF***G- F
AXF***C- F
AXF***H- F
G3/4 Female Waterproof glands for G3/4 conduits or flexible tubes are attached to the
EW
Waterproof Glands electrical connections. Available only for JIS G1/2 female electric connections.
Waterproof Glands Waterproof glands are attached to the electrical connections. Available
EG
only for JIS G1/2 female electric connections.
Waterproof Glands Waterproof glands with union joints are attached to the electrical
EU
with Union Joints connections. Available only for JIS G1/2 female electric connections.
Plastic Glands Plastic glands are attached to the electrical connections. Available only for
EP
JIS G1/2 female electric connections.
Mirror Finished PFA Mirror finishing on the PFA lining inside of the tube to the smoothness lining.
Lining Available for 15 mm (0.5 in.) and larger sizes.
The Ra is average of measured values on several point. PM
Size 15 to 200 mm (0.5 to 8.0 in.) : Ra 0.05 to 0.15 m
Size 250 to 400 mm (10 to 16 in.) : Ra 0.05 to 0.25 m
Mirror Finished Mirror finishing on the inside of the ceramics tube to Ra 0.1 m.
Ceramics Available for 5 mm (0.2 in.) and larger sizes. CM
The Ra is average of measured values on several point.
Stainless Steel Tag A pendant tag plate of JIS SUS304 (AISI 304 SS/EN 1.4301 equivalent) is
Plate provided. Choose this option when a pendant tag plate is required in
SCT
addition to the standard nameplate with the tag number inscribed on it.
Dimension (Height Width): Appr. 12.5 (4.92) 40 (15.7) mm (inch)
Direction change of 90 degrees rotated converter (or terminal box) to change the direction of
RA
the electrical the electrical connection.
connection (*1) 180 degrees rotated converter (or terminal box) to change the direction of
RB
the electrical connection.
90 degrees rotated converter (or terminal box) to change the direction of
RC
the electrical connection.
Bolts, Nuts, and Bolts, nuts, and Bolts and nuts: Carbon steel;
BCC
Gaskets (*2) gaskets are Gaskets: Chloroprene rubber (*3)
provided for wafer Bolts and nuts: Carbon steel;
BCF
connections. Gaskets: PTFE-sheathed non-asbestos (*4)
Available only for Bolts: JIS SUS304 (AISI 304 SS stainless steel
ANSI 150, JIS10K, equivalent);
or, JIS20K wafer Nuts: JIS SUS403 (AISI 403SS stainless steel BSC
connections. equivalent) ;
Gaskets: Chloroprene rubber (*3)
Bolts: JIS SUS304 (AISI 304 SS stainless steel
equivalent);
Nuts: JIS SUS403 (AISI 403SS stainless steel BSF
equivalent) ;
Gaskets: PTFE-sheathed non-asbestos (*4)
Special Gaskets (*5) Viton gaskets for use with a PFA or ceramics lining with PVC piping.
(*10) Allowable temperature and pressure are equivalent to Valqua #4010,
GA
special fluororubber not mixed. Available for 2.5 mm (0.1 in.) to 200 mm
(8.0 in.) of PFA lining or 15 to 200 mm (0.5 to 8 in.) sizes of ceramics lining.
Acid-resistant Viton gaskets for use with a PFA or ceramics lining with PVC
piping. Allowable temperature and pressure are equivalent to Valqua #4010,
GC
special fluororubber mixed (mixing #D2470). Available for 2.5 mm (0.1 in.) to 200
mm (8.0 in.) of PFA lining or 15 to 200 mm (0.5 to 8 in.) sizes of ceramics lining.
Alkali-resistant Viton gaskets for use with a PFA or ceramics lining with PVC
piping. Allowable temperature and pressure are equivalent to Valqua #4010,
GD
special fluororubber mixed (mixing #RCD970). Available for 2.5 mm (0.1 in.) to 200
mm (8.0 in.) of PFA lining or 15 to 200 mm (0.5 to 8 in.) sizes of ceramics lining.
Alkali-resistant carbonized fluororesin gaskets for use with a ceramics lining with
GF
metal piping. Allowable temperature and pressure are equivalent to Valqua #7026.
Silicon rubber gaskets for Sanitary Type, provided between the lining and
GH
the adapter. For the condition of fluid temp. 120 to 160C (248 to 320F).
Oil-prohibited Use Electrodes, linings, and grounding rings are assembled and packed with
polyethylene after being cleaned with water and acetone and dried with air. K1
The label OIL FREE is affixed.
T26-2.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 37
Table of Optional Specifications (continued)
: Available : Not available
Applicable Model
General Explosion proof Submersible Sanitary
Integral Remote Integral Remote Remote Integral Remote
Item Specifications Flowmeter Flowtube Flowwmeter Flowtube Flowtube Flowmeter Flowtube Code
G
D
N
E
AXF***G- P
AXF***C- P
AXF***W- P
AXF***H- P
AXF***G- F
AXF***C- F
AXF***H- F
Oil-prohibited Use with Electrodes, linings, and grounding rings are assembled and packed with
Dehydrating Treatment polyethylene including desiccants after being cleaned with water and K5
acetone and dried with air. The label OIL & WATER FREE is affixed.
Painting Color Change Coated in black (Munsell N1.5 or its equivalent.) P1
Coated in jade green (Munsell 7.5 BG 4/1.5 or its equivalent.) P2
Coated in metallic silver. P7
Epoxy Resin Coating Epoxy resin coating which has alkali-resistance instead of standard
X1
polyurethane resin coating. The color is same as standard type.
High Anti-corrosion Three-layer coating (polyurethane coating on two-layer epoxy resin
Coating coating) in the same range as that for the standard coating. The color is X2
same as standard type. Salt/alkali/acid/weather-resistance.
Material Certificate Reproduced material certificates for :
PFA/polyurethane: Pipe, electrodes, grounding rings/grounding
M01
electrodes, flanges or mini flanges, adapters (for sanitary type)
Ceramics: only grounding rings or grounding electrodes or union joint
Hydrostatic Test The test verifies the absence of leaks by applying the following water
pressures (which are determined under process connection conditions) to
linings for ten minutes. Test results are described in the Note column of a
test certificate (QIC).
Process Connection: Water Pressure: T01
ANSI Class 150, DIN PN10, JIS 10K 1.5 MPa
ANSI Class 300, DIN PN16, JIS 20K 3.0 MPa
DIN PN40, Union joint (Ceramics lining) 6.0 MPa
JIS F12 1.25 MPa
Calibration Certificate Level 2: The Declaration and the Calibration Equipment List are issued. L2
Level 3: The Declaration and the Primary Standard List are issued. L3
Level 4: The Declaration and the Yokogawa Measuring Instruments Control
L4
System are issued.
Vent Hole With a vent hole provided for permeable fluids (such as nitric acid,
hydrofluoric acid, or sodium hydroxide at high temperature). H
Available only for a PFA lining flange type tube.
Enhanced Dual Available for 25 to 200 mm (1.0 to 8.0 in.) sizes.
Frequency Excitation Products are delivered with the Standard dual frequency excitation mode
(*6) and the Enhanced dual frequency excitation mode enabled.
Although the meter factors for the Standard dual frequency excitation mode
HF1
and the Enhanced dual frequency excitation mode are inscribed on the
nameplate, the flow calibration of optional code HF1 is not performed in
Enhanced dual frequency excitation.Excitation mode select optional code
HF2 when an accurate flow measurement is required.
Available for 25 to 200 mm (1.0 to 8.0 in.) sizes.
Products are delivered with the Standard dual frequency excitation mode
and the Enhanced dual frequency excitation mode enabled.
The meter factor for the Enhanced dual frequency excitation obtained by HF2
flow calibration is inscribed on the nameplate and set into the combined
converter in addition to the meter factor for the Standard dual frequency
excitation.
T26-3.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 38
Table of Optional Specifications (continued)
: Available : Not available
Applicable Model
General Explosion proof Submersible Sanitary
Integral Remote Integral Remote Remote Integral Remote
Item Specifications Flowmeter Flowtube Flowmeter Flowtube Flowtube Flowmeter Flowtube Code
G
D
N
E
AXF***G- P
AXF***C- P
AXF***W- P
AXF***H- P
AXF***G- F
AXF***C- F
AXF***H- F
Five-point Calibration A flow test near 0, 25, 50, 75, and 100% of the user-specified span is
in User-specified performed instead of the flow test of the standard 2m/s span and a test
Span certificate (QIC) is submitted. Specify the span (100% flow span) whose
corresponding flow velocity lies between 0.5 to 10 m/s (0.8 to 10 m/s for
sizes 32, 65, and 125 mm) and that is less than the maximum line capacity.
Selectable range of flow rate span is showing below.
*1: 90-degree rotation 180-degree rotation 90-degree rotation *2: When specifying the optional code BCC or BSC for a PFA
Standard
Optional Code RA Optional Code RB Optional Code RC or ceramics lining, it is advisable to specify the optional
code GA, GC, or GD at the same time to prevent
Electrical Indicator Electrical Indicator
potential leakage caused by the difference in elasticity
Integral Flowmeter Connection Connection
between the flowtube and chloroprene gaskets. Refer to
description of Gasket in the Wetted Part Material.
Electrical Electrical
Remote Flowtube Connection Connection
*3: Allowable temperature and pressure with the optional code BCC or BSC (only for Gaskets:Chloroprene rubber) are equivalent to Valqua #2010.
*4: Allowable temperature and pressure with the optional code BCF or BSF (only for Gaskets:PTFE-sheathed non-asbestos) are equivalent to Valqua #7030 (S).
*5: Special gaskets are inserted between the flowtube and the grounding ring or grounding electrode.
*6: Enhanced dual frequency excitation is not available for models with calibration code C (High Grade Accuracy).
*7: Select optional code G12 or G11 when TIIS Flameproof type with wiring using a flameproof packing adapter. Available only for JIS G1/2 female electric connection.
*8: The TIIS flameproof type is only available for AXF***C-P (remote flowtube for combined use with AXFA14).
*9: In the case of Fieldbus communication type, optional codes C1, C2, C3, EM and G11 are unable to select.
*10:In the case of size 32 mm, optional code GH is unable to select. T26-4.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 39
EXTERNAL DIMENSIONS
AXF Standard, AXF002-AXF015, Wafer Type, PFA Lining
Unit : mm (approx. inch)
Integral Flowmeter Remote Flowtube Integral Flowmeter Remote Flowtube
197(7.76)*1 Ground Terminal
Ground Terminal (M4)
154(6.06) (M4) 66*1 51.5 51.5 28(1.1)
*4 86(3.38) (2.6) (2.03) (2.03)
D 48 (1.89) 111(4.37)
AXF002
73(2.87) 49(1.93)
128(5.04)
G E 1
AXF005
W F 2 A 1 A 1
AXF010
C G N
70(2.76)
AXF015
N
P
Hi
Hr
*4: D, E, F, G; Integral Flowmeter,
)
(d
63.5(2.50)
N, P; Remote Flowtube
H1
3(0.12)
D
* No infra-red switches are furnished 72 (2.83)
L*2 4- 6.2(0.24) 58 (2.28)
for Fieldbus communication type.
Size code 002 005 010 015 *1: When indicator code N is selected, subtract 12 mm (0.47 inch)
from the value in the figure.
Model Size 2.5(0.1) 5(0.2) 10(0.4) 15(0.5) In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional
Lining code A A A A code, add the following value to L (face-to-face length).
Face-to-face 2
length L* 81(3.19) Grounding Ring Code S, L, H, V P, T N
None +0 +26(1.02) -2(0.08)
Remote Outside dia. D 44(1.73) Option
flowtube Code GA, GC, GD +6(0.24) +28(1.10)
Inner diameter of (Special Gaskets)
Grounding ring d 15(0.59)
Integral
flowmeter H1
*3: When submersible type or option code DHC is selected,
Height 144(5.67)
waterproof glands and a 30m long cable are attached.
Hr 268(10.55) Add 9.5kg(20.9lb) to the weight in the table.
Remote Max. Height
flowtube
Weight kg (lb) *3 2.4(5.3)
AXF Standard, AXF025-AXF125, Wafer Type, PFA /Polyurethane Rubber /Natural Soft Rubber /EPDM
Rubber Lining
Unit : mm (approx. inch)
128(5.04)
AXF050 1
W F 2 D A 1
AXF065 2
C G N G
AXF080
(2.76)
N
Hi
70
AXF100
P
Hr
AXF125
d)
D
*5: D, E, F, G; Integral Flowmeter,
(
N, P; Remote Flowtube
Hr
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 40
AXF Standard, AXF150-AXF300, Wafer Type, PFA /Polyurethane Rubber /Natural Soft Rubber /EPDM
Rubber Lining
Unit : mm (approx. inch)
Integral Flowmeter Remote Flowtube Integral Flowmeter Remote Flowtube
197(7.76)*1 Ground Terminal
Ground Terminal
(M4) (M4)
154(6.06) 66*1 51.5 51.5 28(1.1)
*5 86(3.38) (2.6) (2.03) (2.03)
111(4.37)
AXF150 D A
48 (1.89)
73(2.87) 49(1.93)
G E 1 U
128(5.04)
AXF200 1
W F 2 D A 1
AXF250 2
C G N G
70(2.76)
AXF300
N
P
*5: D, E, F, G; Integral Flowmeter,
d)
(
Hi
N, P; Remote Flowtube * In case of size
Hr
250mm or 300mm,
D
position of grounding
H1
ring is different
from the figure to
45-degree.
Size code 150 200 250 300 *1: When indicator code N is selected, subtract 12 mm (0.47 inch)
from the value in the figure.
Model Size 150(6) 200(8) 250(10) 300(12) In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
A,U A,U A,U A,U *2: Depending on the selection of grounding ring code and optional
Lining code
D,G D,G D,G D,G code, add the following value to L (face-to-face length).
Face-to-face *2
length L 200(7.87) 250(9.84) 300(11.81) 350(13.78) Nominal Size: 150 to 200mm
Grounding Ring Code S, L, H, V P, T N
Remote Outside dia. D 202(7.95) 252(9.92) 310(12.20) 358(14.09) None +0 +32(1.26) -2(0.08)
Flowtube Inner diameter of Option
Grounding ring d 146.1(5.75) 193.6(7.62) 243.7(9.59) 294.7(11.60) Code GA, GC, GD +10(0.39) +38(1.5)
(Special Gaskets)
Integral
*3
Flowmeter Width W 202(7.95) 252(9.92) 310(12.20) 358(14.09) Nominal Size: 250 to 300mm
Height H1 243(9.57) 293(11.54) 354(13.94) 402(15.83) Grounding Ring Code S, L, H, V P, T N
Option Code is None +0 -2(0.08)
Hr
Remote Max. Height 367(14.45) 417(16.42) 478(18.82) 526(20.71)
Flowtube *3: When electrode structure 2 is selected, add the following value to
Weight kg (lb)*4 14.5(32.0) 22.1(48.7) 39.0(86.0) 48.3(106.5)
W(width).
Integral Max. Height Hi 405(15.93) 455(17.89) 516(20.31) 564(22.20) Nominal size 150 200 250 300
Flowmeter W +49(1.93) +50(1.97) +49(1.93) +53(2.09)
Weight kg (lb) 16.2(35.7) 23.8(52.4) 40.7(89.7) 50.0(110.2)
*4: When submersible type or option code DHC is selected,
waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table. F24.EPS
128(5.04)
C
AXF010 G N R
N
70(2.76)
P
Hi
N, P; Remote Flowtube
63.5(2.50)
H1
3(0.12)
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 41
AXF Standard, AXF015, Wafer Type, Ceramics Lining
Unit : mm (approx. inch)
Integral Flowmeter Remote Flowtube Integral Flowmeter Remote Flowtube
197(7.76)*1 Ground Terminal
Ground Terminal (M4)
154(6.06) (M4) 66*1 51.5 51.5 28(1.1)
86(3.38) (2.6) (2.03) (2.03)
*3
48 (1.89) 111(4.37)
D
73(2.87) 49(1.93)
128(5.04)
E 1
G
AXF015 F 2 C E 1 A 1
C
G N
70(2.76)
N
Hi
P
Hr
*3: D, E, F, G; Integral Flowmeter,
)
N, P; Remote Flowtube (d
63.5(2.50)
H1
3(0.12)
D
72 (2.83)
L*2 4- 6.2(0.24) 58 (2.28)
* No infra-red switches are furnished
for Fieldbus communication type.
E 1
73(2.87) 49(1.93)
AXF040
G
AXF050 F 2 C E 1 A 1
C
AXF080 G N
70(2.76)
AXF100 N
Hi
Hr
P
D d)
*3: D, E, F, G; Integral Flowmeter, (
H1
N, P; Remote Flowtube
Hr
Remote Max. Height 216(8.50) 235(9.25) 253(9.96) 283(11.14) 308(12.13)
Flowtube
Weight kg (lb) 2.3(5.1) 3.2(7.0) 4.1(9.0) 6.8(15.0) 9.6(21.1)
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 42
AXF Standard, AXF150, AXF200, Wafer Type, Ceramics Lining
128(5.04)
(2.87) (1.93)
73 49
70(2.76)
Hi
Hr
) D
(d
H1
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the
Size code 150 200 value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
Model Size 150(6) 200(8) *2: Depending on the selection of grounding ring code and optional code,
Lining code add the following value to L (face-to-face length).
C C
Grounding Ring Code S, L, H, V P, T N
Face-to-face 2
Hr
Remote Max. Height 378(14.88) 428(16.85)
Flowtube
Weight kg (lb) 20.2(44.5) 33.5(73.9)
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 43
AXF Standard, AXF002-AXF015, JIS/ANSI/DIN Flange Type, PFA Lininig
Ground Terminal
(M4)
Ground Terminal 197(7.76)*1
B J 1 154(6.06) (M4) 66*1 51.5 51.5 28(1.1)
*4
B J 2 86(3.38) (2.6) (2.03) (2.03)
D
AXF002 B A 1 48 (1.89) 111(4.37)
G E 1
73(2.87) 49(1.93)
AXF005 B A 2
128(5.04)
W F 2 A 1 1
AXF010 B D 4
C G N
AXF015 D J 1
70(2.76)
N
D J 2
P
Hi
D D 4 t *2
*4: D, E, F, G; Integral Flowmeter, N-h
Hr
(H2)
N, P; Remote Flowtube D
C
H1
(d)
(2.38)
60.5
* No infra-red switches are furnished L*2
for Fieldbus communication type.
Process Connection BJ1(JIS10K) BJ2(JIS20K) BA1(ANSI Class 150) BA2(ANSI Class 300) BD4(DIN PN40) DJ1(JIS10K) DJ2(JIS20K) DD4(DIN PN40)
Size code 002 005 010 015 002 005 010 015 002 005 010 015 002 005 010 015 002 005 010 015 002 005 010 002 005 010 002 005 010
Model
2.5 5 10 15 2.5 5 10 15 2.5 5 10 15 2.5 5 10 15 2.5 5 10 15 2.5 5 10 2.5 5 10 2.5 5 10
Size (0.1) (0.2) (0.4) (0.5) (0.1) (0.2) (0.4) (0.5) (0.1) (0.2) (0.4) (0.5) (0.1) (0.2) (0.4) (0.5) (0.1) (0.2) (0.4) (0.5) (0.1) (0.2) (0.4) (0.1) (0.2) (0.4) (0.1) (0.2) (0.4)
Lining code A A A A A A A A A A A A A A A A A A A A A A A A A A A A A
2 200
Face-to-face length L -30 * 150(5.91) 200
(7.87) 150(5.91) 200
(7.87) 150(5.91) 200
(7.87) 150(5.91) 200
(7.87) 150(5.91) (7.87) 150(5.91) 150(5.91) 150(5.91)
Outside dia. D 95(3.74) 95(3.74) 88.9(3.50) 95.3(3.75) 95(3.74) 90(3.54) 90(3.54) 90(3.54)
18 to 22 16 20 to 24 18 17 to 22 15.2 20 to 25 18.2 21 to 25 20 18 to 22 20 to 24 21 to 25
Thickness t *2 (0.71 to 0.87) (0.63) (0.79 to 0.94) (0.71) (0.67 to 0.87) (0.60) (0.79 to 0.98) (0.72) (0.83 to 0.98) (0.79) (0.71 to 0.87) (0.79 to 0.94) (0.83 to 0.98)
Inner diameter of d
Remote Grounding ring 15(0.59) 15(0.59) 15(0.59) 15(0.59) 15(0.59) 12(0.47) 12(0.47) 12(0.47)
Flowtube Pitch circle dia. C 70(2.76) 70(2.76) 60.5(2.38) 66.5(2.62) 65(2.56) 65(2.56) 65(2.56) 60(2.36)
Bolt hole interval 45 45 45 45 45 45 45 45
Integral
Flowmeter
Hole dia. h 15(0.59) 15(0.59) 15.7(0.62) 15.7(0.62) 14(0.55) 15(0.59) 15(0.59) 14(0.55)
Number of holes N 4 4 4 4 4 4 4 4
Height H1 141(5.54) 141(5.54) 141(5.54) 141(5.54) 141(5.54) 141(5.54) 141(5.54) 141(5.54)
Height H2 80(3.15) 80(3.15) 80(3.15) 80(3.15) 80(3.15) 80(3.15) 80(3.15) 80(3.15)
Remote Max. Height Hr 265(10.43) 265(10.43) 265(10.43) 265(10.43) 265(10.43) 265(10.43) 265(10.43) 265(10.43)
Flowtube Weight kg (lb)*3 3.4(7.5) 3.5 3.6(7.9) 3.7 3.2(7.1) 3.3 3.6(7.9) 3.7 3.8(8.4) 3.9 3.3(7.3) 3.4(7.5) 3.6(7.9)
(7.7) (8.2) (7.3) (8.2) (8.6)
Integral Max. Height Hi 303(11.91) 303(11.91) 303(11.91) 303(11.91) 303(11.91) 303(11.91) 303(11.91) 303(11.91)
Flowmeter Weight kg (lb) 5.1(11.2) 5.2 5.3(11.7) 5.4 4.9(10.8) 5.0 5.3(11.7) 5.4 5.5(12.1) 5.6 5.1(11.2)
(11.5) (11.9) (11.0) (11.9) (12.4) 5.0(11.0) 5.3(11.7)
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to L (face-to-face length) and t (thickness
of flange).
L t L t L t
Grounding Ring Code S, L, H, V P, T N
None +0 +0 +26(1.02) +13(0.51) -2(0.08) -1(0.04)
Option
Code GA, GC, GD +8(0.31) +4(0.16) +30(1.18) +15(0.59)
(Special Gaskets)
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table. F29.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 44
AXF Standard, AXF025-AXF050, JIS/ANSI/DIN Flange Type, PFA /Polyurethane Rubber /Natural Soft
Rubber /EPDM Rubber Lining
Ground Terminal
(M4)
Ground Terminal 197(7.76)*1
(M4)
154(6.06) 66*1 51.5 51.5 28(1.1)
*4 B A
86(3.38)
(2.6) (2.03) (2.03)
D 48 (1.89) 111(4.37)
AXF025 A B J
G E 1
73(2.87) 49(1.93)
128(5.04)
AXF032 U 1 B D 4
W F 2 C A
AXF040 D 2
C G N C J
70(2.76)
AXF050 G
N C D 4 t *2
N-h
Hi
P
)
Hr
(d
(H2)
*4: D, E, F, G; Integral Flowmeter,
N, P; Remote Flowtube
H1
D
* No infra-red switches are furnished C
for Fieldbus communication type.
L*2
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to L (face-to-face length) and t (thickness
of flange).
L t L t L t
Grounding Ring Code S, L, H, V P, T N
None +0 +0 +26(1.02) +13(0.51) -2(0.08) -1(0.04)
Option
Code GA, GC, GD +8(0.31) +4(0.16) +30(1.18) +15(0.59)
(Special Gaskets)
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table.
F30.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 45
AXF Standard, AXF065-AXF125, JIS/ANSI/DIN Flange Type, PFA /Polyurethane Rubber /Natural Soft
Rubber /EPDM Rubber Lining
128(5.04)
73(2.87) 49(1.93)
B A
*4
B D 2
D
70(2.76)
AXF065 A B J N-h
G E 1 t *2
AXF080 U 1 B G 1
W F 2 1
AXF100 D 2 C A
D
C G N
Hi
AXF125 G C D 2 C
Hr
(H2)
N (d)
C J
P
C G 1
H1
*4: D, E, F, G; Integral Flowmeter,
N, P; Remote Flowtube
L*2
* No infra-red switches are furnished
for Fieldbus communication type.
Process Connection BJ1/CJ1(JIS10K) BJ2/CJ2(JIS20K) BG1/CG1(JIS F12) BA1/CA1(ANSI Class 150) BA2/CA2(ANSI Class 300) BD2/CD2(DIN PN16)
Size code 065 080 100 125 065 080 100 125 080 100 125 065 080 100 125 065 080 100 125 065 080 100 125
Model 65 80 100 125 65 80 100 125 80 100 125 65 80 100 125 65 80 100 125 65 80 100 125
Size (2.5) (3) (4) (5) (2.5) (3) (4) (5) (3) (4) (5) (2.5) (3) (4) (5) (2.5) (3) (4) (5) (2.5) (3) (4) (5)
A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U
Lining code D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G
2 200 200 250 250 200 200 250 250 200 250 250 200 200 250 250 200 200 250 250 200 200 250 250
Face-to-face length L -30 * (7.87) (7.87) (9.84) (9.84) (7.87) (7.87) (9.84) (9.84) (7.87) (9.84) (9.84) (7.87) (7.87) (9.84) (9.84) (7.87) (7.87) (9.84) (9.84) (7.87) (7.87) (9.84) (9.84)
175 185 210 250 175 200 225 270 211 238 263 177.8 190.5 228.6 254.0 190.5 209.6 254.0 279.4 185 200 220 250
Outside dia. D (6.89) (7.28) (8.27) (9.84) (6.89) (7.87) (8.86) (10.63) (8.31) (9.37) (10.35) (7.00) (7.50) (9.00) (10.00) (7.50) (8.25) (10.00) (11.00) (7.28) (7.87) (8.66) (9.84)
2 22 22 22 24 24 26 28 30 22 22 24 26.4 27.9 27.9 27.9 29.4 32.4 35.8 39.1 22 24 24 26
Thickness t* (0.87) (0.87) (0.87) (0.94) (0.94) (1.02) (1.10) (1.18) (0.87) (0.87) (0.94) (1.04) (1.10) (1.10) (1.10) (1.16) (1.28) (1.41) (1.54) (0.87) (0.94) (0.94) (1.02)
Inner diameter of 66 77 102 128 66 77 102 128 77 102 128 66 77 102 128 66 77 102 128 66 77 102 128
Remote Grounding ring d (2.60) (3.03) (4.02) (5.04) (2.60) (3.03) (4.02) (5.04) (3.03) (4.02) (5.04) (2.60) (3.03) (4.02) (5.04) (2.60) (3.03) (4.02) (5.04) (2.60) (3.03) (4.02) (5.04)
140 150 175 210 140 160 185 225 168 195 220 139.7 152.4 190.5 215.9 149.4 168.1 200.2 235.0 145 160 180 210
flowtube Pitch circle dia. C (5.51) (5.91) (6.89) (8.27) (5.51) (6.30) (7.28) (8.86) (6.61) (7.68) (8.66) (5.50) (6.00) (7.50) (8.50) (5.88) (6.62) (7.88) (9.25) (5.71) (6.30) (7.09) (8.27)
Bolt hole interval 45 22.5 22.5 22.5 22.5 22.5 22.5 22.5 45 45 30 45 45 22.5 22.5 22.5 22.5 22.5 22.5
22.5 45 22.5 22.5
Integral 19 19 19 23 19 23 23 25 19 19 19 19.1 19.1 19.1 22.4 22.4 22.4 22.4 22.4
18 18 18 18
Hole dia. h (0.75) (0.75) (0.75) (0.91) (0.75) (0.91) (0.91) (0.98) (0.75) (0.75) (0.75) (0.75) (0.75) (0.75) (0.88) (0.88) (0.88) (0.88) (0.88)
(0.71) (0.71) (0.71) (0.71)
flowmeter
Number of holes N 4 8 8 8 8 8 8 8 4 4 6 4 4 8 8 8 8 8 8 4 8 8 8
Height H1 176 187 211 245 176 195 218 255 200 225 252 177 190 220 247 184 200 233 260 181 195 216 245
(6.93) (7.36) (8.30) (9.65) (6.93) (7.68) (8.59) (10.04) (7.87) (8.85) (9.90) (6.97) (7.48) (8.66) (9.72) (7.24) (7.87) (9.16) (10.22) (7.13) (7.68) (8.49) (9.65)
Height 89 95 106 120 89 95 106 120 95 106 120 89 95 106 120 89 95 106 120 89 95 106 120
H2 (3.50) (3.74) (4.16) (4.73) (3.50) (3.74) (4.16) (4.73) (3.74) (4.16) (4.73) (3.50) (3.74) (4.16) (4.72) (3.50) (3.74) (4.16) (4.72) (3.50) (3.74) (4.16) (4.72)
300 311 335 369 300 319 342 379 324 349 376 301 314 344 371 308 324 357 384 305 319 340 369
Remote Max. Height Hr (11.81) (12.24) (13.18) (14.53) (11.81) (12.56) (13.47) (14.92) (12.76) (13.73) (14.79) (11.85) (12.36) (13.54) (14.61) (12.13) (12.76) (14.04) (15.11) (12.01) (12.56) (13.37) (14.53)
flowtube Weight kg (lb)*3 9.0 9.6 12.4 17.4 9.3 12.4 16.9 24.7 12.2 15.5 19.5 10.8 12.9 17.7 20.8 12.6 16.6 26.8 34.9 9.8 11.9 14.5 19.3
(19.8) (21.2) (27.3) (38.3) (20.5) (27.3) (37.3) (54.5) (26.9) (34.2) (43.1) (23.7) (28.5) (39.1) (45.9) (27.7) (36.6) (59.1) (76.9) (21.6) (26.2) (32.0) (42.5)
338 349 372 407 338 357 380 417 362 386 413 339 352 382 409 346 362 394 421 343 357 377 407
Integral Max. Height Hi (13.31) (13.74) (14.65) (16.01) (13.31) (14.06) (14.95) (16.40) (14.25) (15.21) (16.26) (13.35) (13.86) (15.02) (16.08) (13.62) (14.25) (15.52) (16.58) (13.50) (14.06) (14.85) (16.00)
flowmeter Weight kg (lb) 10.7 11.3 14.1 19.1 11.0 14.1 18.6 26.4 13.9 17.2 21.2 12.5 14.6 19.4 22.5 14.3 18.3 28.5 36.6 11.5 13.6 16.2 21.0
(23.5) (25.0) (31.0) (42.1) (24.3) (31.0) (41.0) (58.3) (30.7) (37.8) (46.8) (27.5) (32.2) (42.8) (49.6) (31.4) (40.4) (62.8) (80.7) (25.4) (29.9) (35.7) (46.2)
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to L (face-to-face length) and t (thickness of
flange).
L t L t L t
Grounding Ring Code S, L, H, V P, T N
None +0 +0 +26(1.02) +13(0.51) -2(0.08) -1(0.04)
Option
Code GA, GC, GD +8(0.31) +4(0.16) +30(1.18) +15(0.59)
(Special Gaskets)
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table.
F31.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 46
AXF Standard, AXF150, AXF200, JIS/ANSI/DIN Flange Type, PFA /Polyurethane Rubber /Natural Soft
Rubber /EPDM Rubber Lining
128(5.04)
N-h
(2.87)
(2.76)
t *2
73
70
D
Hi
(H2)
Hr
C
H1
(
d)
* No infra-red switches are furnished L*2
for Fieldbus communication type.
Process Connection BJ1/CJ1(JIS10K) BJ2/CJ2(JIS20K) BG1/CG1(JIS F12) BA1/CA1(ANSI Class 150) BA2/CA2(ANSI Class 300) BD1/CD1(DIN PN10) BD2/CD2(DIN PN16)
Size code 150 200 150 200 150 200 150 200 150 200 200 150 200
Model 150 200 150 200 150 200 150 200 150 200 200 150 200
Size (6) (8) (6) (8) (6) (8) (6) (8) (6) (8) (8) (6) (8)
A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U
Lining code D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G
Face-to-face 0 2 300 350 300 350 300 350 300 350 300 350 350 300 350
length L -3 * (11.81) (13.78) (11.81) (13.78) (11.81) (13.78) (11.81) (13.78) (11.81) (13.78) (13.78) (11.81) (13.78)
280 330 305 350 290 342 279.4 342.9 317.5 381.0 340 285 340
Outside dia. D (11.02) (12.99) (12.01) (13.78) (11.42) (13.46) (11.00) (13.50) (12.50) (15.00) (13.39) (11.22) (13.39)
2 27 27 33 35 27 29 30.4 33.4 43.5 46.1 29 27 29
Thickness t* (1.06) (1.06) (1.30) (1.38) (1.06) (1.14) (1.20) (1.31) (1.71) (1.81) (1.14) (1.06) (1.14)
Inner diameter of 146.1 193.6 146.1 193.6 146.1 193.6 146.1 193.6 146.1 193.6 193.6 146.1 193.6
Remote Grounding ring d (5.75) (7.62) (5.75) (7.62) (5.75) (7.62) (5.75) (7.62) (5.75) (7.62) (7.62) (5.75) (7.62)
flowtube Pitch circle dia. 240 290 260 305 247 299 241.3 298.5 269.7 330.2 295 240 295
C (9.45) (11.42) (10.24) (12.01) (9.72) (11.77) (9.50) (11.75) (10.62) (13.00) (11.61) (9.45) (11.61)
Integral Bolt hole interval 22.5 15 15 15 30 22.5 22.5 22.5 15 15 22.5 22.5 15
flowmeter Hole dia. 23 23 25 25 19 19 22.4 22.4 22.4 25.4 22 22 22
h (0.91) (0.91) (0.98) (0.98) (0.75) (0.75) (0.88) (0.88) (0.88) (1.00) (0.87) (0.87) (0.87)
Number of holes N 8 12 12 12 6 8 8 8 12 12 8 8 12
281 331 294 341 286 337 281 337 300 357 336 284 336
Height H1 (11.06) (13.03) (11.56) (13.43) (11.26) (13.27) (11.05) (13.29) (11.80) (14.04) (13.23) (11.16) (13.23)
141 166 141 166 141 166 141 166 141 166 166 141 166
Height H2 (5.55) (6.54) (5.55) (6.54) (5.55) (6.54) (5.55) (6.54) (5.55) (6.54) (6.54) (5.55) (6.54)
405 455 418 465 410 461 405 461 424 481 460 408 460
Remote Max. Height Hr (15.94) (17.91) (16.44) (18.31) (16.14) (18.15) (15.93) (18.17) (16.68) (18.92) (18.11) (16.04) (18.11)
flowtube 27.8 37.3 37.1 51.9 29.9 43.2 30.9 49.2 52.5 78.8 42.5 28.7 41.9
Weight kg (lb) *3 (61.3) (82.2) (81.8) (114.4) (65.9) (95.3) (68.0) (108.4) (115.7) (173.7) (93.7) (63.2) (92.5)
443 493 456 503 448 499 443 499 462 519 498 446 498
Integral Max. Height Hi (17.42) (19.39) (17.95) (19.80) (17.64) (19.65) (17.43) (19.66) (18.18) (20.41) (19.61) (17.54) (19.61)
flowmeter 29.5 39.0 38.8 53.6 31.6 44.9 32.6 50.9 54.2 80.5 44.2 30.4 43.6
Weight kg (lb) (65.0) (86.0) (85.5) (118.2) (69.7) (99.0) (71.8) (112.2) (119.5) (177.5) (97.5) (66.9) (96.2)
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to L (face-to-face length)
and t (thickness of flange).
L t L t L t
Grounding Ring Code S, L, H, V P, T N
None +0 +0 +32(1.26) +16(0.63) -2(0.08) -1(0.04)
Option
Code GA, GC, GD +10(0.39) +5(0.20) +38(1.5) +19(0.75)
(Special Gaskets)
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table. F32.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 47
AXF Standard, AXF250-AXF400, JIS/ANSI/DIN Flange Type, PFA /Polyurethane Rubber /Natural Soft
Rubber /EPDM Rubber Lining
B A Unit : mm (approx. inch)
*4 B D
D
AXF250 A B J
G E 1
AXF300 U 1 B G 1
W F 2 1
AXF350 D 2 C A
C G N
AXF400 G C D
N
C J
P
C G 1
*4: D, E, F, G; Integral Flowmeter,
N, P; Remote Flowtube Integral Flowmeter Remote Flowtube
Integral Flowmeter Remote Flowtube Ground Termonal
(M4)
Ground Terminal 197(7.76)*1
154(6.06) 86 (3.38) (M4)
48(1.89) 66*1 51.5 51.5 28(1.1)
49 (1.93)
128(5.04)
(2.87)
73
70
N- h
D
(H2)
C
Hi
H3
Hr
H1
(d
)
t*2
L*2
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to L (face-to-face length)
and t (thickness of flange).
Nominal Size: 250 mm to 300 mm Nominal Size: 350 mm to 400 mm
L t L t L t L t
Grounding Ring Code S, L, H, V N Grounding Ring Code S, L, H, V N
Option Code is None +0 +0 -6(0.24) -3(0.12) Option Code is None +0 +0 -10(0.39) -5(0.20)
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table. F33.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 48
AXF Standard, AXF015-AXF125, Sanitary for Clamp Connection, PFA Lining
Unit : mm (approx. inch)
Integral Flowmeter Remote Flowtube Integral Flowmeter Remote Flowtube
Ground Terminal
Ground Terminal 197(7.76)*1
(M4)
AXF015 (M4)
154(6.06) 86(3.38) 66*1 51.5 51.5 28(1.1)
AXF025 *3 48 (2.6) (2.03) (2.03)
AXF032 D (1.89) 111(4.37)
128(5.04)
AXF040 E 1 H A B
(2.87) (1.93)
49
AXF050 H F 2 A L 1 N H D B 1
AXF065 G N H K B
70(2.76)
AXF080 N
73
Hi
AXF100 P
Hr
AXF125
D
*3: D, E, F, G; Integral Flowmeter, (d)
H1
N, P; Remote Flowtube Adapter for clamp connection
Process Connection HAB / HDB / HKB *1: When indicator code N is selected, subtract 12 mm
(0.47 inch) from the value in the figure.
Size code 015 025 032 040 050 065 080 100 125 *2: When option code DHC is selected, waterproof
Model 15 25 32 40 50 65 80 100 125 glands and a 30m long cable are attached.
Size (0.5) (1) (1.3) (1.5) (2) (2.6) (3) (4) (5) Add 9.5kg(20.9lb) to the weight in the table.
Lining code A A A A A A A A A
Face-to-face 166 166 166 166 176 196 216 246 316
length L (6.55) (6.55) (6.55) (6.55) (6.94) (7.73) (8.52) (9.70) (12.46)
73 73 73 86 99 117 129 155 183
Outside dia. D (2.87) (2.87) (2.87) (3.39) (3.90) (4.61) (5.08) (6.10) (7.20)
Remote 15.7 22.1 34.8 47.5 60.2 72.9 97.4
Flowtube HAB
(0.62) (0.87) (1.37) (1.87) (2.37) (2.87) (3.83)
16 26 32 38 50 66 81 100 125
Integral Inner dia. d HDB
C
(0.63) (1.02) (1.26) (1.50) (1.97) (2.60) (3.19) (3.94) (4.92)
Flowmeter
15.2 22.6 31.3 35.6 48.6 60.3 72.9 97.6 135.7
HKB (0.60) (0.89) (1.23) (1.40) (1.91) (2.37) (2.87) (3.84) (5.34)
98 98 98 111 129 147 157 183 212
Height H1
A
E
F
B
d
(3.86) (3.86) (3.86) (4.37) (5.08) (5.79) (6.18) (7.20) (8.35)
222 222 222 235 253 271 281 307 336
Max. Height Hr (8.74) (8.74) (8.74) (9.25) (9.96) (10.67) (11.06) (12.09) (13.23)
Remote
Flowtube 2 2.7 2.5 2.6 2.9 3.6 4.8 5.7 8.1 12.1
Weight kg (lb)* (6.0) (5.5) (5.7) (6.4) (7.9) (10.6) (12.6) (17.9) (26.7)
260 260 260 273 291 309 319 345 374
Max. Height Hi G
Integral (10.22) (10.22) (10.22) (10.73) (11.44) (12.15) (12.54) (13.56) (14.70)
Flowmeter 4.4 4.2 4.3 4.6 5.3 6.5 7.4 9.8 13.8
Weight kg (lb) (9.7) (9.3) (9.5) (10.1) (11.7) (14.3) (16.3) (21.6) (30.4) Adapter for clamp connection
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 49
AXF Standard, AXF015-AXF125, Sanitary for Union Connection, PFA Lining
Unit : mm (approx. inch)
Integral Flowmeter Remote Flowtube Integral Flowmeter Remote Flowtube
(2.87) (1.93)
AXF050 H F 2 A L 1 N J D B 1 111(4.37)
49
128(5.04)
AXF065 G N J S B
(2.76)
AXF080 N
70
73
Hi
AXF100 P
AXF125
Hr
*3: D, E, F, G; Integral Flowmeter,
N, P; Remote Flowtube
H1
D
(d)
Adapter for union
L connection
C
A
B
E
JKB:
d
Process Connection JKB / JDB / JSB F G
Size code 015 025 032 040 050 065 080 100 125 H
Model 15 25 32 40 50 65 80 100 125
Size (0.5) (1) (1.3) (1.5) (2) (2.6) (3) (4) (5) G
C
Lining code A A A A A A A A A
166 166 166 166 176 196 216 246
JKB (6.55) (6.55) (6.55) (6.55) (6.94) (7.73) (8.52) (9.70)
Face-to-face 166 166 166 166 176 196 236 266 326
A
B
E
F
d
length L JDB (6.55) (6.55) (6.55) (6.55) (6.94) (7.73) (9.31) (10.49) (12.85) JDB:
166 166 166 176 196 216 276
JSB (6.55) (6.55) (6.55) (6.94) (7.73) (8.52) (10.88)
Remote
73 73 73 86 99 117 129 155 183
Flowtube Outside dia. D (2.87) (2.87) (2.87) (3.39) (3.90) (4.61) (5.08) (6.10) (7.20)
Integral 15.2 22.6 31.3 35.6 48.6 60.3 72.9 97.6 H
JKB (0.60) (0.89) (1.23) (1.40) (1.91) (2.37) (2.87) (3.84)
Flowmeter
16 26 32 38 50 66 81 100 125
Inside dia. d JDB (0.63) (1.02) (1.26) (1.50) (1.97) (2.60) (3.19) (3.94) (4.92)
22.5 29.6 35.5 48.5 60.5 72.9 97.6 G
JSB
C
d
A
Max. Height Hr (8.74) (8.74) (8.74) (9.25) (9.96) (10.67) (11.06) (12.09) (13.23) JSB:
Remote
Flowtube 2 2.6 2.6 2.7 3 3.8 4.9 5.9 8.2 13
Weight kg (lb)* (5.7) (5.7) (6.0) (6.6) (8.4) (10.8) (13.0) (18.1) (28.7)
260 260 260 273 291 309 319 345 374 3
Max. Height Hi (10.24) (10.24) (10.24) (10.73) (11.44) (12.17) (12.54) (13.56) (14.70)
Integral
Flowmeter 4.3 4.3 4.4 4.7 5.5 6.6 7.6 9.9 14.7 F
Weight kg (lb) (9.5) (9.5) (9.7) (10.4) (12.1) (14.6) (16.8) (21.8) (32.4) H
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the
figure. Adapters for union connection
*2: When option code DHC is selected, waterproof glands and a 30m long cable are
attached. Add 9.5kg (20.9lb) to the weight in the table.
Process JKB (ISO2853 Union)
Connection JDB (DIN 11851 Union) JSB (SMS1145 Union)
Nominal 15 25 32 40 50 65 80 100 15 25 32 40 50 65 80 100 125 25 32 40 50 65 80 100
Size
70 70 70 83 96 114 126 152 70 70 70 83 96 114 126 152 180 70 70 83 96 114 126 152
A (2.76) (2.76) (2.76) (3.27) (3.78) (4.49) (4.96) (5.98) (2.76) (2.76) (2.76) (3.27) (3.78) (4.49) (4.96) (5.98) (7.09) (2.76) (2.76) (3.27) (3.78) (4.49) (4.96) (5.98)
16 22.2 29.4 34.6 47.6 59.5 72.3 97 16 22.2 29.4 34.6 47.6 59.5 72.3 97 123 22.2 29.4 34.6 47.6 59.5 72.3 97
B (0.63) (0.87) (1.16) (1.36) (1.87) (2.34) (2.85) (3.82) (0.63) (0.87) (1.16) (1.36) (1.87) (2.34) (2.85) (3.82) (4.84) (0.87) (1.16) (1.36) (1.87) (2.34) (2.85) (3.82)
18 25.6 34.3 38.6 51.6 64.1 76.7 102.5 20 30 36 42 54 70 85 104 129 25.4 32 38.1 51 63.5 76.2 102.5
C (0.71) (1.01) (1.35) (1.52) (2.03) (2.52) (3.02) (4.04) (0.79) (1.18) (1.42) (1.65) (2.13) (2.76) (3.35) (4.09) (5.08) (1.00) (1.26) (1.50) (2.01) (2.50) (3.00) (4.04)
15.2 22.6 31.3 35.6 48.6 60.3 72.9 97.6 16 26 32 38 50 66 81 100 125 22.5 29.6 35.5 48.5 60.5 72.9 97.6
d (0.60) (0.89) (1.23) (1.40) (1.91) (2.37) (2.87) (3.84) (0.63) (1.02) (1.26) (1.50) (1.97) (2.60) (3.19) (3.94) (4.92) (0.89) (1.17) (1.40) (1.91) (2.38) (2.87) (3.84)
21.2 29.2 38.2 42.7 56.2 69.9 82.6 108.7 18 30 36 42 54 71 85 104 130 32 40 48 61 73.5 86 120
E (0.83) (1.15) (1.50) (1.68) (2.21) (2.75) (3.25) (4.28) (0.71) (1.18) (1.42) (1.65) (2.13) (2.80) (3.35) (4.09) (5.12) (1.26) (1.57) (1.89) (2.40) (2.89) (3.39) (4.72)
13.5 13.5 13.5 13.5 13.5 13.5 13.5 13.5 25.8 39.8 45.8 51.8 63.8 80.8 94.8 113.8 141.8 11 13 15 15 19 19 30
F (0.53) (0.53) (0.53) (0.53) (0.53) (0.53) (0.53) (0.53) (1.02) (1.57) (1.80) (2.04) (2.51) (3.18) (3.73) (4.48) (5.58) (0.43) (0.51) (0.59) (0.59) (0.75) (0.75) (1.18)
3 3 3 3 3 3 3 3 RD34 RD52 RD58 RD65 RD78 RD95 RD110 RD130 RD160 RD40 RD48 RD60 RD70 RD85 RD98 RD132
G (0.12) (0.12) (0.12) (0.12) (0.12) (0.12) (0.12) (0.12) 1/8 1/6 1/6 1/6 1/6 1/6 1/4 1/4 1/4 1/6 1/6 1/6 1/6 1/6 1/6 1/6
50 50 50 50 50 50 50 50 50 50 50 50 50 50 60 60 65 50 50 50 50 50 50 65
H (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (2.36) (2.36) (2.56) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (2.56)
F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811
Parts No. LA LB LC LD LE LF LG LH KR KS KT KU KV KW KX KY KZ LK LL LM LN LP LQ LR
F35.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 50
AXF Standard, AXF015-AXF125, Sanitary for Butt Weld, PFA Lining
73(2.87) 49(1.93)
AXF040 E 1
128(5.04)
K K B
AXF050 H F 2 A L 1 N 1
K D B
AXF065 G N
70(2.76)
AXF080 N
Hi
AXF100 P
Hr
AXF125
D
*3: D, E, F, G; Integral Flowmeter, (d)
H1
N, P; Remote Flowtube Adapter for butt weld connection
Process Connection KKB / KDB *1: When indicator code N is selected, subtract 12 mm
(0.47 inch) from the value in the figure.
Size code 015 025 032 040 050 065 080 100 125 *2: When option code DHC is selected, waterproof
Model 15 25 32 40 50 65 80 100 125 glands and a 30m long cable are attached.
Size (0.5) (1) (1.3) (1.5) (2) (2.6) (3) (4) (5) Add 9.5kg (20.9lb) to the weight in the table.
Lining code A A A A A A A A A
Face-to-face 126 126 126 126 136 156 176 206 276
length L (4.98) (4.98) (4.98) (4.98) (5.37) (6.16) (6.94) (8.13) (10.88)
73 73 73 86 99 117 129 155 183
Outside dia. D (2.87) (2.87) (2.87) (3.39) (3.90) (4.61) (5.08) (6.10) (7.20)
Remote 15.2 22.6 31.3 35.6 48.6 60.3 72.9 97.6 135.7
Flowtube KKB (0.60) (0.89) (1.23) (1.40) (1.91) (2.37) (2.87) (3.84) (5.34)
Inner dia. d
16 26 32 38 50 66 81 100 125
Integral KDB (0.63) (1.02) (1.26) (1.50) (1.97) (2.60) (3.19) (3.94) (4.92)
Flowmeter
98 98 98 111 129 147 157 183 212
A
C
d
Height H1 (3.86) (3.86) (3.86) (4.37) (5.08) (5.79) (6.18) (7.20) (8.35)
222 222 222 235 253 271 281 307 336
Max. Height Hr (8.74) (8.74) (8.74) (9.25) (9.96) (10.67) (11.06) (12.09) (13.23)
Remote
Flowtube 2 2.6 2.3 2.5 2.8 3.4 4.5 5.3 7.1 11
Weight kg (lb)* (5.7) (5.1) (5.5) (6.2) (7.5) (9.9) (11.7) (15.7) (24.3)
260 260 260 273 291 309 319 345 374
Max. Height Hi (10.24) (10.24) (10.24) (10.73) (11.44) (12.17) (12.54) (13.56) (14.70)
Integral E
Flowmeter 4.3 4 4.2 4.5 5.1 6.2 7 8.8 12.7
Weight kg (lb) (9.5) (8.8) (9.3) (9.9) (11.2) (13.7) (15.4) (19.4) (28.0)
Adapter for butt weld connection
Process
Connection KKB (ISO2037 Butt Weld) KDB (DIN 1185 Butt Weld)
Nominal 15 25 32 40 50 65 80 100 125
Size 15 25 32 40 50 65 80 100 125
70 70 70 83 96 114 126 152 180 70 70 70 83 96 114 126 152 180
A (2.76) (2.76) (2.76) (3.27) (3.78) (4.49) (4.96) (5.98) (7.09) (2.76) (2.76) (2.76) (3.27) (3.78) (4.49) (4.96) (5.98) (7.09)
16 22.2 29.4 34.6 47.6 59.5 72.3 97 123 16 22.2 29.4 34.6 47.6 59.5 72.3 97 123
B (0.63) (0.87) (1.16) (1.36) (1.87) (2.34) (2.85) (3.82) (4.84) (0.63) (0.87) (1.16) (1.36) (1.87) (2.34) (2.85) (3.82) (4.84)
18 25.6 34.3 38.6 51.6 64.1 76.7 102.5 141.2 20 30 36 42 54 70 85 104 129
C (0.71) (1.01) (1.35) (1.52) (2.03) (2.52) (3.02) (4.04) (5.56) (0.79) (1.18) (1.42) (1.65) (2.13) (2.76) (3.35) (4.09) (5.08)
15.2 22.6 31.3 35.6 48.6 60.3 72.9 97.6 135.7 16 26 32 38 50 66 81 100 125
d (0.60) (0.89) (1.23) (1.40) (1.91) (2.37) (2.87) (3.84) (5.34) (0.63) (1.02) (1.26) (1.50) (1.97) (2.60) (3.19) (3.94) (4.92)
30 30 30 30 30 30 30 30 40 30 30 30 30 30 30 30 30 40
E (1.18) (1.18) (1.18) (1.18) (1.18) (1.18) (1.18) (1.18) (1.57) (1.18) (1.18) (1.18) (1.18) (1.18) (1.18) (1.18) (1.18) (1.57)
F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811
Parts No. NN NP NQ NR NS NT NU NV NW ND NE NF NG NH NJ NK NL NM
F36.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 51
Replacement model for Earlier ADMAG or ADMAG AE, AXF002-AXF015, Wafer Type, PFA Lining
Unit : mm (approx. inch)
Integral Flowmeter Remote Flowtube Integral Flowmeter Remote Flowtube
197(7.76)*1 Ground Terminal
Ground Terminal (M4)
154(6.06) (M4) 66*1 51.5 51.5 28(1.1)
*4 86(3.38) (2.6) (2.03) (2.03)
D 48 (1.89) 111(4.37)
AXF002
G E 1
73(2.87) 49(1.93)
128(5.04)
AXF005
W F 2 A 1 A 2
AXF010
C G N
AXF015
70(2.76)
N
P
Hi
Hr
*4: D, E, F, G; Integral Flowmeter,
N, P; Remote Flowtube )
(d
H1
D
3(0.12)
(2.50)
63.5
72 (2.83)
L*2 4- 6.2(0.24) 58 (2.28)
* No infra-red switches are furnished
for Fieldbus communication type.
Size code 002 005 010 015 *1: When indicator code N is selected, subtract 12 mm (0.47 inch)
from the value in the figure.
Model Size 2.5(0.1) 5(0.2) 10(0.4) 15(0.5) In case of explosion proof type with indicator, add 5 mm (0.2
inch) to it.
Lining code A A A A *2: Depending on the selection of grounding ring code and optional
code, add the following value to L (face-to-face length).
Face-to-face *2 85(3.35)
length L Grounding Ring Code S, L, H, V P, T N
Remote
Flowtube Outside dia. None +0 +22(0.87) -6(0.24)
D 44(1.73) Option
Code GA, GC, GD +2(0.08) +24(0.94)
Integral Inner diameter of (Special Gaskets)
Flowmeter Grounding ring
d 15(0.59)
*3: When submersible type or option code DHC is selected,
Height H1 144(5.67) waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table.
Remote Max. Height Hr 265(10.43)
Flowtube 3
Weight kg (lb)* 2.4(5.3)
Replacement model for Earlier ADMAG or ADMAG AE, AXF025-AXF100, Wafer Type,
PFA /Polyurethane Rubber Lining
Unit : mm (approx. inch)
Integral Flowmeter Remote Flowtube Integral Flowmeter Remote Flowtube
Ground Terminal
(M4)
154(6.06) 86(3.38) 197(7.76)*1 Ground Terminal
*4 (M4)
48(1.89) 66*1 51.5 51.5 28(1.1)
AXF025 D (2.6) (2.03) (2.03)
73(2.87) 49(1.93)
AXF040 G E 1 111(4.37)
A
AXF050 W F 2 1 A 2
128(5.04)
U
AXF080 C G N
(2.76)
AXF100 N
70
Hi
Hr
P
*4: D, E, F, G; Integral Flowmeter,
H1
d)
D
(
N, P; Remote Flowtube
L*2
* No infra-red switches are furnished
for Fieldbus communication type.
Size code 025 040 050 080 100 *1: When indicator code N is selected, subtract 12 mm (0.47 inch)
from the value in the figure.
Model Size 25(1) 40(1.5) 50(2) 80(3) 100(4) In case of explosion proof type with indicator, add 5 mm (0.2
inch) to it.
Lining code A,U A,U A,U A,U A,U *2: Depending on the selection of grounding ring code and optional
Face-to-face Grounding Ring Code S, L, H, V P, T N
*2 93(3.66) 106(4.17) 120(4.72) 160(6.30) 180(7.09)
length L None +0 +22(0.87) -6(0.24)
Remote Option
Flowtube Outside dia. D 67.5(2.66) 86(3.39) 99(3.90) 129(5.08) 155(6.10) Code GA, GC, GD +2(0.08) +24(0.94)
(Special Gaskets)
Inner diameter of
Integral Grounding ring d 27(1.06) 40(1.57) 52(2.05) 81(3.19) 98(3.86) *3: When submersible type or option code DHC is selected,
Flowmeter waterproof glands and a 30m long cable are attached.
Height H1 92(3.62) 111(4.37) 129(5.08) 157(6.18) 183(7.20) Add 9.5kg(20.9lb) to the weight in the table.
216(8.50) 235(9.25) 253(9.96) 281(11.06) 307(12.09)
Remote Max. Height Hr
Flowtube 3
Weight kg (lb)* 3.1(6.7) 3.5(7.7) 4.2(9.3) 6.4(14.1) 8.0(17.6)
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 52
Replacement model for Earlier ADMAG or ADMAG AE, AXF150, AXF200, Wafer Type,
PFA /Polyurethane Rubber Lining
Unit : mm (approx. inch)
Ground Terminal
(M4)
Ground Terminal 197(7.76)*1
(M4)
66*1 51.5 51.5 28(1.1)
154(6.06) 86(3.38)
*4 (2.6) (2.03) (2.03)
D 48(1.89) 111(4.37)
73(2.87) 49(1.93)
128(5.04)
G E 1
AXF150 A
W F 2 1 A 2
AXF200 U
C G N
70(2.76)
N
P
d)
*4: D, E, F, G; Integral Flowmeter,
(
Hi
N, P; Remote Flowtube
Hr
D
H1
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the
Size code 150 200
value in the figure.
Size 150(6) 200(8) In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
Model
*2: Depending on the selection of grounding ring code and optional code, add
Lining code A,U A,U the following value to L (face-to-face length).
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 53
Replacement model for Earlier ADMAG or ADMAG AE, AXF150-AXF250, JIS/ANSI/DIN Flange Type,
PFA /Polyurethane Rubber Lining
*4
*4: D, E, F, G; Integral Flowmeter,
D N, P; Remote Flowtube Unit : mm (approx. inch)
C A
AXF150 G E 1
A 1 C D
AXF200 W F 2 2
U 2 C J
AXF250 C G N
C G 1
N
P Integral Flowmeter Remote Flowtube
Integral Flowmeter Remote Flowtube Ground Terminal
(M4)
197(7.76)*1
Ground Terminal
154(6.06) 86 (3.38) (M4) 66*1 51.5 51.5 28(1.1)
48(1.89) (2.6) (2.03) (2.03)
111(4.37)
128(5.04)
73 49 (1.93)
Eye Bolt
(2.76)
(2.87)
70
N- h
D
(H2)
Hi
Hr
C
H3
H1
(d
)
* No infra-red switches are furnished t *2
for Fieldbus communication type. L*2
Process Connection CJ1(JIS10K) CJ2(JIS20K) CG1(JIS F12) CA1(ANSI Class 150) CA2(ANSI Class 300) CD1(DIN PN10) CD2(DIN PN16)
Size code 150 200 250 150 200 250 150 200 250 150 200 250 150 200 250 200 250 150 200 250
Model
150 200 250 150 200 250 150 200 250 150 200 250 150 200 250 200 250 150 200 250
Size (6) (8) (10) (6) (8) (10) (6) (8) (10) (6) (8) (10) (6) (8) (10) (8) (10) (6) (8) (10)
Lining code A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U
Face-to-face 270 340 430 270 340 430 270 340 430 270 340 430 270 340 430 340 430 270 340 430
length L*2 (10.63) (13.39) (16.93) (10.63) (13.39) (16.93) (10.63) (13.39) (16.93) (10.63) (13.39) (16.93) (10.63) (13.39) (16.93) (13.39) (16.93) (10.63) (13.39) (16.93)
280 330 400 305 350 430 290 342 410 279.4 342.9 406.4 317.5 381.0 444.5 340 395 285 340 405
Outside dia. D (11.02) (12.99) (15.75) (12.01) (13.78) (16.93) (11.42) (13.46) (16.14) (11.00) (13.50) (16.00) (12.50) (15.00) (17.50) (13.39) (15.55) (11.22) (13.39) (15.94)
31 31 33 37 39 43 31 33 33 34.4 37.4 39.2 47.5 50.1 56.7 33 35 31 33 35
Thickness t *2 (1.22) (1.22) (1.30) (1.46) (1.54) (1.69) (1.22) (1.30) (1.30) (1.30) (1.38) (1.22) (1.30) (1.38)
(1.35) (1.47) (1.54) (1.87) (1.97) (2.23)
Inner diameter of 140.7 188.9 243 140.7 188.9 243 140.7 188.9 243 140.7 188.9 243 140.7 188.9 243 188.9 243 140.7 188.9 243
Remote Grounding ring d (5.54) (7.44) (9.57) (5.54) (7.44) (9.57) (5.54) (7.44) (9.57) (5.54) (7.44) (9.57) (5.54) (7.44) (9.57) (7.44) (9.57) (5.54) (7.44) (9.57)
flowtube 240 290 355 260 305 380 247 299 360 241.3 298.5 362.0 269.7 330.2 387.4 295 350 240 295 355
Pitch circle dia. C (9.45) (11.42) (13.98) (10.24) (12.01) (14.96) (9.72) (11.77) (14.17) (9.50) (11.75) (14.25) (10.62) (13.00) (15.25) (11.61) (13.78) (9.45) (11.61) (13.98)
Integral
flowmeter Bolt hole interval
22.5 15 15 15 15 15 30 22.5 22.5 22.5 22.5 15 15 15 11.25 22.5 15 22.5 15 15
23 23 25 25 25 27 19 19 23 22.4 22.4 25.4 22.4 25.4 28.4 22 22 22 22 26
Hole dia. h
(0.91) (0.91) (0.98) (0.98) (0.98) (1.06) (0.75) (0.75) (0.91) (0.88) (0.88) (1.00) (0.88) (1.00) (1.12) (0.87) (0.87) (0.87) (0.87) (1.02)
Number of holes N 8 12 12 12 12 12 6 8 8 8 8 12 12 12 16 8 12 8 12 12
284 334 400 297 344 415 289 340 405 284 340 403 303 360 422 339 397 287 339 402
Height H1
(11.18) (13.15) (15.75) (11.69) (13.54) (16.34) (11.38) (13.39) (15.94) (11.18) (13.39) (15.87) (11.93) (14.17) (16.61) (13.35) (15.63) (11.30) (13.35) (15.83)
141 166 197 141 166 197 141 166 197 141 166 197 141 166 197 166 197 141 166 197
Height H2
(5.55) (6.54) (7.76) (5.55) (6.54) (7.76) (5.55) (6.54) (7.76) (5.55) (6.54) (7.76) (5.55) (6.54) (7.76) (6.54) (7.76) (5.55) (6.54) (7.76)
325 375 454 350 395 484 335 387 464 324 387 460 362 426 499 385 449 330 385 459
Height H3
(12.80) (14.76) (17.87) (13.78) (15.55) (19.06) (13.19) (15.24) (18.27) (12.76) (15.24) (18.11) (14.25) (16.77) (19.65) (15.16) (17.68) (12.99) (15.16) (18.07)
408 458 524 421 468 539 413 464 529 408 464 527 427 484 546 463 521 411 463 526
Max. Height Hr
Remote (16.06) (18.03) (20.63) (16.57) (18.43) (21.22) (16.26) (18.27) (20.83) (16.06) (18.27) (20.75) (16.81) (19.06) (21.50) (18.23) (20.51) (16.18) (18.23) (20.71)
flowtube 29 39 64 38.3 53.6 92.5 31.1 44.6 67.4 32.1 50.9 77.4 53.7 80.5 127.0 44.2 67.0 29.9 43.6 68.8
Weight kg (lb)*3 (63.9) (86.0) (141.1) (84.4) (118.2) (203.9) (68.6) (98.3) (148.6) (70.7) (112.2) (170.6) (118.4) (177.5) (279.9) (97.5) (147.7) (65.8) (96.2) (151.7)
Hi 446 496 562 459 506 577 451 502 567 446 502 565 465 522 584 501 559 449 501 564
Integral Max. Height (17.56) (19.53) (22.13) (18.07) (19.92) (22.72) (17.76) (19.76) (22.32) (17.56) (19.76) (22.24) (18.31) (20.55) (22.99) (19.72) (22.01) (17.68) (19.72) (22.20)
flowmeter 30.7 40.7 65.7 40.0 55.3 94.2 32.8 46.3 69.1 33.8 52.6 79.1 55.4 82.2 128.7 45.9 68.7 31.6 45.3 70.5
Weight kg (lb) (67.7) (89.7) (144.8) (88.2) (121.9) (207.7) (72.3) (102.1) (152.3) (74.4) (115.9) (174.4) (122.1) (181.2) (283.6) (101.2) (151.5) (69.6) (100.0) (155.5)
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to L (face-to-face length) and t (thickness of
flange).
Nominal Size: 150, 200 mm Nominal Size: 250 mm
L t L t L t L t L t
Grounding Ring Code S, L, H, V P, T N Grounding Ring Code S, L, H, V N
None +0 +0 +28(1.1) +14(0.55) -6(0.24) -3(0.12) Option Code is None +0 +0 -6(0.24) -3(0.12)
Option
Code GA, GC, GD +2(0.08) +1(0.04) +30(1.81) +15(0.59)
(Special Gaskets)
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table.
F40.EPS
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 54
Flameproof Packing Adapter for TIIS Flameproof Type (Optional code G12 or G11)
T1
Unit : mm
Adapter body(M. Screw) (Approx. inch)
O-Ring
Packing case
16.5(0.65)
Hexagon socket set screw
Packing *
18(0.71) Hexagon socket set screw
O-Ring
C
F L O-Ring
G Washer
Union nut
Packing gland
*Packing D
Clamp ring
(Choose from the table below
depend on cable outside diameter) Clamp nut
T2 O-Ring
B.coupling
Cable(user's scope) F43.EPS
Unless otherwise specified, difference in the dimensions are refer to the following table.
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 55
SIZING DATA (Measurable flow velocity is from 0 m/s.)
SI Units Size: mm
0
0
5
400
0
0
0
0
5
65
32
25
80
20
40
50
25
2.
35
30
15
10
12
5
15
10
10.0
5.0
Flow velocity
(m/s)
1.0
0.5
0.3
0.1
0.001 0.01 0.1 1 10 100 1,000 10,000
Flowrate (m3 /h)
5
1.2
0.1
1.5
0.4
0.5
0.2
2.5
16
14
12
10
5
2
3
1
6
4
8
33
Flow velocity
(ft/s)
10
5.0
2.0
1.0
0.33
0.01 0.1 1.0 10 100 1,000 10,000 100,000
F44.EPS
Flowrate (GPM)
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
<<Contents>> <<Index>> 56
ORDERING INFORMATION
Note 1: When ordering, the span flow rate, unit, output 5. Output pulse weight
pulse weight, and totalizer display pulse weight can
be specified. These parameters will then be set If specified, a span flow rate must be set first, and then
before shipment. volume per pulse must be set.
In the case of remote type, please specify combina- It can be specified numerically within the value from
tion information of flowtube and converter. 0.0001 to 32000, and its unit must be specified in the
These parameters will be set in the combined
converter.
same unit as the span flow rate.
When ordering of a flowtube alone or a converter (Example: When m3 is selected as Span Flow Rate,
alone, these parameters cannot specify. specify 10 m3/p as pulse weight unit.)
Custom configuration request is necessary for
setting these parameters out of regular setting
Unless specified, the relevant product is delivered with
range. the setting at 0 pulse/second.
Note 2: Some options, if ordered, require the relevant 6. Totalizer display pulse weight
specifications to be input when ordering.
If specified, a span flow rate must be set first, and then
1. Model, specification and option codes. volume per totalization pulse count must be set.
2. Combined use (when ordering a remote flowtube) It can be specified numerically within the value from
0.0001 to 32000, and its unit must be specified in the
Model, suffix code, optional code, and tag number (if
same unit as the span flow rate.
specified) of a converter for combined use.
(Example: when m3 is selected as Span Flow Rate,
For detail, refer to GS 01E20C01-01E or GS 01E20C02-01E.
specify 10 m3/p as pulse weight unit.)
3. Tag number
Unless specified, the relevant product is delivered with
Each tag number can be specified in up to 16 the setting at 0 pulse/second.
characters in a combination of letters (capital or small
7. Fluid name
letters), numbers, - and ..
For HART protocol, up to 8 characters can be specified.
If specified, the tag number is inscribed on the products
RELATED INSTRUMENTS
name plate and tag plate (if optional code SCT is
selected). If the product is an integral flowmeter, the tag Calibrator for Magnetic Flowmeter (AM012):
number is also set into the memory of its converter. GS 1E6K2-E
If the user wishes to change only the tag number to be BT200 BRAIN TERMINAL: GS 01C00A11-00EN
set into a converters memory, specify the software tag.
AXFA11G Magnetic Flowmeter Remote Converter:
If a tag number is not specified, the tag number is set
as a blank. GS 01E20C01-01E
4. Span flow rates and units AXFA14G/C Magnetic Flowmeter Remote Converter:
Span flow rate can be specified the numeric within the GS 01E20C02-01E
value of 0.0001 to 32000. AXF FOUNDATION Fieldbus Communication Type Magnetic
And it can be up to five digits, to a maximum of 32000 Flowmeter: GS 01E20F02-01E
ignoring the decimal point.
AXF PROFIBUS PA Communication Type Magnetic
And a fraction is limited to the fourth decimal place. Flowmeter: GS 01E20F12-01E
Integral flowmeter are set to the first range in the
FieldMate: GS 01R01A01-01E
forward direction. Remote flowtube are set to the first
range in the forward direction of the converter (AXFA11
or AXFA14) with which they are to be combined. REFERENCE
A span flow rate and its unit must be specified when the ADMAG, AXF, ADMAG AXF, BRAIN TERMINAL and
Mass Unit Setting (optional code MU) or Five-point FieldMate are registered trademarks of Yokogawa Electric
Calibration in User-specified Span (optional code SC) is Corporation.
specified. Company names and product names used in this material
If a span flow rate and its unit are not specified, the are registered trademarks or trademarks of their respective
relevant product is delivered with the setting at 1 m/s owners.
(3.3 ft/s).
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation GS 01E20D01-01E Dec.28, 2013-00
Subject to change without notice.