An image matrix is a grid of pixels arranged in rows and columns that makes up a digital image. Each pixel contains a CT number or Hounsfield unit (HU) representing the attenuation of x-rays by the corresponding tissue volume. The HU scale quantifies radiodensity on a scale where water is 0 HU and air is -1000 HU. Image reconstruction uses filtered back projection to convert the measured x-ray attenuation data into pixel values that are displayed as different shades of gray representing different tissue types.

Uploaded by

Innoj Maco
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
0% found this document useful (1 vote)
389 views

CTSCAN - Image Characteristics and Reconstruction

An image matrix is a grid of pixels arranged in rows and columns that makes up a digital image. Each pixel contains a CT number or Hounsfield unit (HU) representing the attenuation of x-rays by the corresponding tissue volume. The HU scale quantifies radiodensity on a scale where water is 0 HU and air is -1000 HU. Image reconstruction uses filtered back projection to convert the measured x-ray attenuation data into pixel values that are displayed as different shades of gray representing different tissue types.

Uploaded by

Innoj Maco
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
You are on page 1/ 2

Topic 4: Image Characteristics

Image Matrix
What is an image matrix?
- Matrix Series of rows & columns (of pixels) that give form to the digital image (Bontrager Glossary)
- A layout of cells in rows & columns (Bushong Glossary)
- Rows & columns each has assigned a number which displayed as an optical density or brightness level on
the monitor.
- (Optical) Density the degree of blackening; it has a numeric value which can be present in varying
degrees from completely black to almost clear. (Bushong Chapter 20)
- Original format (EMI) 80 x 80 matrix for a total of(80x80) 6400 individual cells of information, the current
imaging system provide matrices of 512 x 512 results to 262,144 cells of information.

Pixel (picture element)


- Matrix
- The cell (unit) of digital image
- Cell of information
- An individual matrix box
- Two-dimensional representation of a corresponding (equivalent) tissue volume (voxel). (fig. 29-18 Bushong
chapter 29)

Voxel
o Volume element; correspond to a three-dimensional tissue volume (height, width & Depth).
o Each pixel represents a voxel when an image is viewed.

- Where each pixel contains numeric information called CT number or Hounsfield unit (HU).

Hounsfield Unit
- Also know as Hounsfield scale
- named after Sir Godfrey Newbold Hounsfield, is a quantitative scale for describing radiodensity. (Internet)
- The Hounsfield unit (HU) scale is a linear transformation of the original linear attenuation coefficient measurement into
one in which the radiodensity of distilled water at standard pressure and temperature (STP). (Internet)

CT number / Hounsfield Unit


- Number that represents the attenuation value for each pixel. (Bontrager Glossary)
- Optical density & Brightness level as displayed on the monitor.
- Numbers level ranges from -1000 to +1000 for each pixel.
- Table 29-1: CT Number for Various tissues. (Bushong Chapter 29)
- A precise CT number of a pixel is related to the attenuation coefficient (x-ray) of the tissue contained in the
voxel.
-
Coefficient
o Any of the factors of a product considered in relation to a specific factor
o A number that serves as a measure of some characteristics

Linear Attenuation coefficient


o Numeric expression of the decrease in radiation (attenuation) intensity that follows transmission through matter

- Relative probability interaction between the differential absorption remains constant.

Topic 5: Image Reconstruction

Field of View (FOV)


- The diameter of image reconstruction.

Pixel size = FOV


Matrix size
Filtered back Projection
- The process in which image are stored in computer memory when detector acquired projection during scanning.

Filter
- Refers to mathematical function (x-ray is for metal filter)

Filtered back projection


Example Fig. 29-19 Four pixel matrix

Recap:

- CT images display a variety of shades of gray.


- Do not read the book Fifty shades of gray
- Low-density structures such as lungs, air filled structures will attenuate very little x-ray beam.
- High-density structures such as bones & contrast media attenuate all or nearly all of the beam.
- Linear attenuation coefficient (degree of attenuation) are converted into numeric scale called CT numbers.
- Scanners are calibrated with water as zero (0) basis.
- Tissue type & CT numbers

Tissue type CT Numbers Appearance


Cortical bone +1000 (up to +3,000) White
Muscle +50 Gray
White matter +45 Light Gray
Gray Matter +40 Gray
Blood +20 Gray (White if IC+ is present)
CSF +15 Gray
Water 0 (Baseline)
Fat -100 Dark gray to black
Lung -200 Dark gray to black
Air (Produces the least of attenuation) -1000 Black

- (WW) Window Width refers to the range of CT numbers that are displayed as shades of gray.
- (WL) Window Level controls image brightness or determines the CT number that will be the center of the
window width.
- Both WW & WL called Window Center
- Bitch a female dog.
- Pitch a ratio that reflects the relationship between table speed & slice thickness
- Pitch formula = Couch movement per 360 degrees rotation of tube
Collimation