Manual - Return Loss Measurements
Manual - Return Loss Measurements
Sam Wetterlin
[email protected]
10/5/08
This paper describes the procedure for using a reflection bridge (a/k/a return loss bridge) to
manually measure return loss (easily converted to SWR if desired). Manual means we dont
have all the automation of a vector network analyzer, and we also cant utilize elaborate
calibration to improve accuracy.
We need a signal source, whose frequency we must set manually, and we need a way to measure
the output level in dbm, which we will have to record manually. It is possible to get some
automation by using a sweep generator; more on this later.
The bridge routes the input signal to the DUT and outputs a signal proportional to the reflection
received from the DUT. The bridge itself can take various forms. One such bridge is decribed
later. For the moment, suffice it to say that it is not difficult to build a reflection bridge which
can isolate the reflection well enough to measure return losses in the range 0-30 db with as much
accuracy as we generally need.
To get a reference level representing zero return loss, we can just measure the output with the
open standard attached. An open circuit has a return loss of zero. A short circuit also has a return
loss of zero (meaning 100% reflection), but the phase of the reflection is 180 degrees different
between an open and a short. Because we are not measuring phase (return loss is a scalar
quantity), the output with the open and short attached should look identical, in theory. In
practice, there will generally be some difference between the open and short reflection outputs,
though it is often less than 0.5 db. To reduce error, we can use as the reference level the average
between the open and short outputs. But if we have verified that the open/short difference is less
than 1 db, this averaging will give us a reference only 0.5 db from what we would get using the
open reflection as the reference level, so the averaging may not be worth the bother.
That ratio is what we will use to measure the reflection coefficient. Another formula, useful
when you know the impedance of the DUT, is as follows:
Z Z0
Reflection Coefficient = =
Z + Z0
Z=DUT impedance;
Z0=Reference Impedance (here, 50 ohms)
Return loss is the magnitude of that reflection coefficient expressed in decibels as a non-negative
number:
In our case, the conversion into decibels is done automatically by the instrument measuring the
bridge output, as that instrument will measure in dbm.
Those are the technical formulas, but what does a given level of return loss actually mean?
Return loss can be translated into SWR, or into the percent of power reflected by the DUT, or,
roughly, into impedance. Table 1 shows those values for various levels of return loss.
2
Table 1Translating return loss to reflection coefficient, SWR,
percentage of power reflected, and equivalent
resistance, one greater than and one less than 50 ohms.
The correspondence of return loss and impedance deserves a little more attention. Any given
return loss can be generated by a variety of impedances; each such impedance generates the
same magnitude of reflection coefficient, but different phases. Since return loss is a scalar
quantity, it does not include the phase, so all those impedances generate the same return loss.
Figure 2 shows the impedances that generate a reflection coefficient whose magnitude (denoted
by [rho]) is 0.33, corresponding to a return loss of 9.6 db.
There are two pure resistance values which create any finite return loss. As shown in Figure 2, if
the lower one is 50/N (N=2 in this case), then the higher one is 50*N. As the reflection
coefficient decreases (return loss increases), the circle gets smaller and smaller and its center
moves closer and closer to 50 ohms. Therefore, return loss can be thought of as a measure of
how close the impedance is to 50 ohms.
3
As a preliminary matter, we should explore how much accuracy we need when measuring return
loss. One way of looking at return loss is that it measures the amount of power reflected by the
DUT. Generally, if the DUT has return loss of 20 db or better (higher), it is reflecting so little
power (1% of the incident power) that it does not really matter exactly how little it is. On the
other hand, if a DUT reflects half or more of the power reaching it (return loss of 0-3 db), we
may just consider it to be unacceptable, without having to know exactly how much it reflects. So
if we know the return loss is higher than 20 db or lower than 3 db, that may be all we need to
know. And even for values in the 3-20 db range, errors of several db may not be meaningful.
If we look at return loss as an indirect measure of impedance, high return losses mean the DUT
has an impedance close to 50 ohms. For example, as shown in Table 1, resistances between 46.9
and 53.3 ohms have return losses of 30 db or higher. If we consider such a DUT equal to 50
ohms, we have a possible 6% error. For many purposes, that is all the accuracy we need, so in
many cases verifying that the return loss exceeds 30 db is all we need to know, and there is no
point to trying to determine whether it is actually 40 db, 50 db, or infinity. Return losses that are
near zero indicate impedances that are far from 50 ohms. For example, resistances of 3 ohms and
870 ohms have return losses of 1 db. Getting the return loss exactly right in that situation does
not help us much, since we are left with two wildly different resistances that may have that return
loss, and we have no way to choose which is correct, since we lack phase information. In fact, as
noted earlier, the impedance may have a reactive component as well, and there is an entire circle
on the impedance plane of possible DUTs with a return loss of 1. When you look at return loss as
a measure of impedance, it tells you whether you are close or far from 50 ohms; if you are close
you have a good idea what the impedance is; if you are far away you dont even know what
direction you are from 50 ohms, so you know you have a problem, but youre not sure exactly
what it is.
Because return loss is a scalar value, we cannot determine the exact impedance even if we knew
the exact return loss. Phase information would be required. Our manual use of the bridge
provides only scalar information, and is effectively a scalar network analyzer, as opposed to
the fancier vector network analyzers, which provide phase information (magnitude combined
with phase = vector). The absence of phase information in return loss generally makes it
pointless to try to determine return loss precisely.
Figure 3 shows the performance of the Parallel Line Bridge over a wide frequency range. Both
directivity and open/short are graphed; they will be discussed momentarily.
Figure 4 shows the performance over the more limited range where the Parallel Line Bridge is
well suited for manual measurements.
5
Lets first consider directivity. Figure 4 shows that the bridge has directivity better than 45 db
from 1 MHz to 400 MHz, and maintains directivity better than 40 db to 500 MHz. Directivity is
the amount of reflection measured for a 50-ohm DUT (but with the negative sign dropped). In
that case, there is no actual reflection, so the directivity represents a bogus signal which is being
added to the actual reflection. It therefore imposes a fundamental limit on our measurement
accuracy, especially when the actual reflection is not large enough to swamp the bogus signal.
If we have directivity of 40 db, then the bogus signal is 40 db below the signal sent to the DUT.
If the actual return loss is also 40 db, then the voltage levels of the reflection and the bogus
signal are equal. Depending on phase, the combined voltage could be as high as double that of
the actual reflection, or as low as zero. That means our measurement may come out anywhere
from 34 db (i.e. 6 db higher than actual, representing doubling of the voltage) to infinity
(representing complete canceling of the reflection by the bogus signal). On the other hand, if the
actual reflection is much stronger than the bogus signal, as with a return loss of 10 db, then the
possible effects of the bogus signal are much diminished.
Figure 5 shows the relationship between our return loss reading and the possible range of values
of the actual return loss, assuming bridge directivity of 40 db.
Figure 5Possible actual return loss values for a given return loss reading, for
directivity of 40 db, including only the error arising from imperfect directivity.
Figure 5 shows that we can determine return losses with reasonable accuracy if they are below
30 db. Once the return loss reading rises to 40 db and higher, we can say with confidence that the
return loss is at least some minimum value (34 db for a reading of 40 db), but there is no limit to
what the maximum might be. That infinite range may seem like unacceptable error, but
remember that for large return losses we frequently only need to know that the return loss
exceeds some threshold, without caring how far it exceeds the threshold. If we get a very large
6
reading, like 60 db, we can be sure the true return loss is at least 40 db, which may be all we are
looking for. Large return losses means the DUT impedance is near 50 ohms, and it is often not
necessary to pin down exactly how close it is.
The other performance measurement shown in Figures 3 and 4 is the relative level of the
measurements for open and shorted DUTs. This is referred to as open/short because it
represents the ratio of the measured reflection with open and shorted DUTS, though we actually
calculate it by subtracting the short output from the open output, since we are operating in dbm.
Ideally, the open and short should have reflections of equal magnitude (and opposite phase), so
the open/short would be zero. Figure 4 showed that the magnitude of the open/short for the
Parallel Line Bridge is generally less than 1, though between 400 and 500 MHz it rises to 1.7.
If we use the average of those two to establish the reference level for zero return loss, then there
is a possible error of half their difference, because for large impedances it would be more correct
to use the open reflection level as the reference, and for small impedances it would be more
correct to use the short reflection level. Since we dont know the actual impedance, we are not
sure which reference to use. By using the average, we get a possible error equal to half the
difference between the open and short levels. In our case, that error is generally less than 0.5 db,
but reaches 0.85 db at 500 MHz.
Large differences between the open and short may indicate poor return loss at the DUT port of
the bridge. This would mean that part of the reflection received from the DUT would be re-
reflected, causing measurement error. This error is relatively small for open/short magnitudes
under 1, but for higher values with very low return loss DUTs, it might reach 1 db (because they
reflect most of the re-reflection, perhaps doubling the error).
Finally, there is some error in the actual measurement of the bridge output. This error is reduced
to some extent by our reference calibration. For example, if the measurement device is
consistently 1 db low, both our reference level and DUT reflection levels will be low by that
amount, so when we calculate the difference between the two levels (i.e. the measured return
loss), the error will disappear. Particularly when the two measurements are very close, there is
very likely the same error in each measurement, and there will be complete error cancellation.
However, when the reference and DUT reflection levels differ by a large amount (that is, the
measured return loss is large), it is possible that the error in the two measurements is different.
For example, a power meter may have little error measuring a level of 0 dbm, but might have 1
db of error measuring a level of 30 db. That means that when we subtract the DUT reflection
from the open reflection (in dbm), we would end up with an error of 1 db.
Altogether, lets assume that there is an additional 1.5 db of error in each of the lines shown in
Figure 5. Figure 6 shows that increased error.
7
Figure 6Total error for directivity=40 db
Allowing 1.5 db for other error
The total error for directivity of 40 db shown in Figure 6 is acceptable for decent measurements
of return losses up to 30 db. For higher return losses, a bridge with directivity of 40 db is
sufficient to verify that the return loss exceeds a certain threshold. For example, a reading of 35
db verifies that the actual return loss exceeds 30 db. This accuracy is sufficient for most practical
return loss measurements.
8
Figure 7Total error for directivity=50 db
Allowing 0.5 db for other error
Figure 7 shows that the parallel line bridge, with an accurate output measurement, gives us
outstanding accuracy for any return loss we would normally want to measure.
Semi-Automatic Measurement
So far, we have described a point-by-point measurement process where the DUT output is
compared to the open output. With a sweep generator and spectrum analyzer, we can automate
the process a bit. Even better would be to use a tracking generator with the spectrum analyzer,
but I dont have one.
Figure 8 shows a scan of the return loss of a multi-stage LC filter. The sweep generator was set
to sweep at its fastest rate, and the spectrum analyzer was set to sweep at a rate of 5 sec/sweep.
The result is a filled-in area whose envelope is a graph of the bridge output.
9
Figure 8Swept return loss of an LC lowpass filter. The spectrum analyzer reference
level was adjusted so the bridge output with an open DUT was at the top of the grid.
Return loss is the distance between the top and a point on the graph, based on 5 db per
vertical division. The dip measures return loss of 33 db.
In this case, I ignored the difference between the open and short. To be precise, I could have
adjusted the reference level so the open and short outputs straddled the top line, but such
precision is often not necessary. Even without precise measurements, this scan shows where the
return loss dip (53 MHz), and also shows that there is an area of modest return loss (15-20 db)
near 25 MHz. The dip is deep enough that with directivity of 50 db the true return loss could be
in the range 31-35 db, per Figure 6. As a practical matter, if a filter has return loss better than 30
db, it is probably of no concern exactly what the value is.
For wider sweeps, the output level for the open will not be perfectly flat because the insertion
loss of the bridge increases with frequency. That makes it harder to use the top of the grid as a
reference point. For highest accuracy, it would then be necessary to save the scan of the open and
display it on-screen when the DUT is scanned. There is a minor amount of confusion because the
DUT graph will appear within the filled-in area of the open graph, but it is still quite easy to
compare the two levels.
Instead of using a spectrum analyzer, an oscilloscope could be used, but the bridge output would
first have to be converted to dbm, perhaps by a log detector like the AD8310. The log detector
output would then control the Y-axis of the oscilloscope, and the signal generator sweep voltage
would control the X-axis. Note that the oscilloscope would be operating at a fairly low
frequency. Even a PC-based oscilloscope utilizing the sound card could be used.
In reality, no adapter or coax cable has the ideal characteristic impedance. Especially for long
cables and/or high frequencies, the method of attaching the DUT may alter the return loss
measurement. To be safe, check the difference between an open and short attached at the end of
the cable, and measure the return loss of a good 50-ohm termination at the end of the cable. Then
proceed as though the bridge open/short and the bridge directivity are those values you just
obtained. For example, if the termination measures 30 db return loss, then 30 db becomes your
directivity, and you have to analyze the error accordingly.
This is not likely to be an issue for simple adapters or cables under 1 ft. until you exceed several
hundred MHz. One way to compensate is to connect the REF load through the same
adapters/coax used for the DUT. This improves directivity by maintaining symmetry.
If the cable leading from the signal source to the bridge input, or the cable leading from the
bridge output to the spectrum analyzer is not high quality, or if the signal source or spectrum
analyzer do not present good 50-ohm impedances, measurement accuracy may suffer. For that
reason, it is a good idea to use attenuators at the bridge input and output. There are built-in
attenuators of 3 db. If necessary, those can be increased or external attenuators can be used. If in
doubt, measure with and without attenuators. If there is a difference and the attenuators are
decent, then the measurement with the attenuators is the more accurate. However, you cant use
an attenuator on the DUT output, because the attenuator effectively becomes part of the DUT
and raises its return loss.
An Active Bridge
With operational amplifiers it is possible to make a bridge that is extremely precise. Appendices
C and D show such a bridge, that is quite simple to make. Figure 9 shows the performance of
that bridge.
11
Figure 9Performance of Active Bridge
The bridge of Figure 9 has at least 40 db directivity and good open/short up to 300 MHz, and in
more limited ranges can achieve directivities of 50 and 60 db. Two different builds had similar
performance. Lets look at the accuracy that can be achieved in the range below 30 MHz, where
directivity exceeds 60 db and the open/short is zero. Figure 10 shows the error profile, again
allowing 0.5 db of error in the device used to measure the bridge output.
Conclusion
We have examined techniques for making and using reflection bridges to measure return losses
with as much accuracy as we would normally need. For use in the world of transmitters and
antennas, those return loss values can readily be converted to SWR or percent of reflected power.
In other circuitry these bridges can verify that amplifiers or filters have acceptable return loss,
and identify problem areas. High return losses indicate an impedance near 50-ohms; lower return
losses can represent a wide range of impedances (in fact, a circle of impedances) and cannot be
converted to a specific impedance value, because of the lack of phase information. Some specific
uses of return loss measurements will be covered in a separate paper, as will a presentation of
other bridge designs.