GC Troubleshooting - GC Troubleshooting - Hints and Tips - Chromatography - Chromservis

gc troubleshootig
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

HOME(/EN/) ABOUTUS(/I/COMPANYPROFILE) NEWS(/S/NEWS) PRODUCTS/ESHOP(/ESHOP/) KNOWMORE(/I/KNOWMORE)

CONTACTS(/I/CONTACTINFO)
(/en/)
EN

(/i/gc

troubleshooting/g
(/i/gc
greatservice,qualityproducts
andtips?
troubleshooting/g
(/i/gc

lang=EN)
and
troubleshooting/g
(/i/gc

CHROMATOGRAPHY tips?
and
troubleshooting/g
(/i/gc
(/)Chromatography(/s/chromatography)Hintsandtips(/g/hintsandtips)GCtroubleshooting(/c/gctroubleshooting)GC
(/S/CHROMATOGRAPHY) troubleshooting
lang=CZ)
tips?
and
troubleshooting/g
GC/MSTOF(/g/timeof
lang=SK)
flightgcms) GCtroubleshooting tips?
and

GC/MS/MS(/g/triple lang=BG)
tips?
quadrupole) Prevention lang=RU)
Glossary(/g/glossaryin
chromatography) ManyGCproblemscanbepreventedifthecolumnisproperlyinstalledandGCismaintainedroutinely.Forexample,replacingseptaor
lineratregularintervalsandkeepingtheinjectoranddetectorcleanandwellmaintainedshoudsolvemanyproblems.Regularpreventive
Hintsandtips(/g/hints
maintenancedependsonparticularmodelofGCandyoushouldconsultrequiredoperationsintheoperator'sandservicemanuals.
andtips)
Fortheproblemidentificationwerecommendtouseelectronicflowmeterandleakdetector.
DynamicHeadspace
(/c/dynamicheadspace)
Baselineproblems
GCcolumnconditioning
(/c/gccolumn
conditioning) Baselineproblemscouldbedividedinto5categories:drift,noise,offset,spikingandwander.

GCcolumninstallation Driftmeansslowbaselinemovementinonedirection
(/c/gccolumninstallation) Noiseisrapidandrandommovementofthebaselineposition
GCliners(/c/gcliners) Offsetissuddenunexplainedchangeofthebaselineposition
GCseptaselection(/c/gc Spikingispresentedbypeakswithnowidth,eitherpositiveornegative
septaselection) Wanderislowfrequencynoise
GCtroubleshooting(/c/gc
troubleshooting) Downwarddrift
Hintsandtipsoverview
(/c/hintsandtips
overview)
Possiblecause Suggestions
Downwarddriftforafewminutesis Increasetheoventemperaturetoclosetothemaximumcontinuousoperatingtemperaturefor
HPLCtroubleshooting
normalafterinstallinganewcolumn thecolumn.Maintainthetemperatureuntilflatbaselineisobserved.Ifthedetectorsignaldoes
(/c/hplctroubleshooting)
notdropin10minutes,immediatelycoolthecolumnandcheckforleaks.
Chiralcolumnhandling
Unequilibrateddetector Allowsufficienttimefortemperatureequilibrationofthedetector.
(/c/chiralcolumnuse)
Kinetexandorg.solvents Downwarddriftisfrequentlyduetothe Cleanoutcontamination.
(/c/kinetexandorg "backout"ofcontaminantsfromthe
detectororotherpartsoftheGC
solvents)
Linearvelocitysettings
(/c/linearvelocitysettings) Upwarddrift
Signal:noiseincreacein
GC(/c/signalnoise Possiblecause Suggestions
increaceingc)
Damagetothestationaryphaseof Determinethecauseofthedamage.Itmaybeduetoimpuritiesinthecarriergasortoexcessive
Siltekdeactivation theGCcolumn temperatures.Replacecolumn.
(/c/siltekdeactivation)
Driftingasflowrates Cleanorreplacefloworpressureregulator(s).Adjustpressure.
Syringecare(/c/syringe
care)
Thermaldesorption Noise
(/c/thermaldesorption)
Thermaldesorption Possiblecause Suggestions
sorbents(/c/TDsorbents)
Thecolumnmaybeinsertedtoofarintotheflameof Reinstallthecolumn.Besuretoinsertthecolumnintothedetectorexactlythe
Troubleshooting anFID,NPDorFPDdetector correctdistancespecifiedintheinstrumentmanual.
(/c/troubleshooting)
AnairleakcanresultinnoiseinECDandTCD Eliminatetheleak.
PreparativeHPLC detectors
(/g/preparativehplc) Incorrectcombustiongasesorflowratescangenerate Besureyuorgasesarethepropergrade,aswellascleananddry.Resetthe
noisinFID,NPDorFPDdetectors. flowratesofthegasestotheirpropervalues.
PreparativeLC
(/c/preparativelc) Contaminatedinjector Cleaninjector.Replaceinletliner,septaandselas.
SMB(/c/smb) Contaminatedcolumn Bakeoutthecolumn.Cutofffirst10cmofcolumn.Ifitdoesnothelp,replace
thecolumn.
Stationaryphases
(/g/stationaryphases) Defectivedetector Cleanand/orreplacepartsasnecessary.

GC(/c/gc) Defectivedetectorboard ConsultGCmanufacturer.

HPLC(/c/hplc)
MEPS(/c/meps) Offset
Moreabout
chromatography(/c/more Possiblecause Suggestions
aboutchromatography) Offline
SPE(/c/spe) Linevoltagechanges Monitorlinevoltageforcorrelationwithoffset.Ifcorrelationisfound,installvoltage
regulatororensurestablepowersupply.
StationaryPhases
Overview(/c/stationary Poorelectricalchanges Checkelectricalconnections.Tightenanylooseconnections.Cleananydirtyorcorroded
phasesoverview) connections.
UHPLC(/c/uhplc) Contaminatedinjector Cleaninjector.Replaceinletliner,septaandselas.

Syringes(/g/syringes) Contaminatedcolumn Bakeoutthecolumn.Cutofffirst10cmofcolumn.Ifitdoesnothelp,replacethe


column.
Agilentsyringes
(/c/agilentsyringes) ColumninsertedtoofatintotheflameofFID, Reinstallthecolumn.Besuretoinsertthecolumnintothedetectorexactlythecorrect
NPDorFPDdetectors distancespecifiedintheinstrumentmanual.
CTCsyringes(/c/ctc
syringes) Contaminateddetector Cleanthedetectorifpossible.

UHPLC(/g/uhplc)
Spiking
UHPLCindetails
(/c/uhplcindetails)
UHPLCreagents
Possiblecause Suggestions
(/c/uhplcreagents) Electricaldisturbancesenteringthe Trytocorrelatespikeswitheventsinequipmentnearthechromatograph.Periodicityis
chromatographthroughpowercables,even oftenaclue.Turnoffequipmentormoveit.Ifnecessary,installavoltageregulator.
Unitconversions(/g/unit shieldedcables
conversions)

Wander

NEWS Possiblecause Suggestions


Baselinewanderingmaybecausedbychangesinenvironmental Trytocorrelatethewanderingwithenvironmentalparameters.Ifa
Chroms conditionssuchastemperatureorlinevoltage correlationisobserved,youwillknowwhattodo.
LikePage InadequatetemperaturecontrolCheckifvariationscanbe Measuredetectortemperature.
correlatedwithchangesinthebaselineposition.

Bethefirstofyour Wanderingwhileusingisothermalconditionsmaybedueto Changethecarriergasorthegaspurificationtraps.


friendstolikethis contaminatedcarriergas

Contaminatedinjector Cleaninjector.Replaceinletliner,glasswoolandseals.

Contaminatedcolumn Bakeoutthecolumn.Cutofffirst10cmofcolumn.Ifitdoesnothelp,
replacethecolumn.
Chro
mserv
Distortedpeakshapes
is
7hrs
Allpeaksredusedinsize
Restekuvedlna
trhSPMEArrow
novtypSPME,
kterpekonv
tradinSPME
vlkna.Podvejte

Possiblecause Suggestions
Samplevalidity Checktheconcentrationandstabilityofthesample.

Flattenedtoppeaks

Possiblecause Suggestions
Detectoroverload.Thebroadpeaksmayhavearoundedtoporevenvalleys Reducesamplevolume,dilutewithsolventorusehigher
inthetop. splitratio.

Overloadofthesignalprocessingelectronics.Thepeaksareclippedwithflat Attenuatedetectoroutputorreducesampleamount(see
tops. above).

Frontingpeaks

Fronting peaks are usually the result of column overloading. In this case, the effect should be a function of injection volume. Solutions
include reducing the injection volume or using a column with greatr capacity. Columns with larger diameter or thicker stationary phase
coatingsgenerallyhavelargersamplecapacities,however,resolutionmaybereduced.

Ghostpeaks

Possiblecause Suggestions

Offline
Remmantsofprevioussamplesintheinletorcolumn.Ghost Increasethefinaltemperatureandlengthentheruntimetoallowfor
peaksduetoremmantsaremostlikelytooccurwhenincreasing completeelutionofprevioussamples.Ifghostpeakscontinueto
inletorcolumntemperatures. occur,cleantheinlet.Conditionthecolumnattemperaturehigher
thanhasbeenusedbutlowerthanthemaximumcontinuous
operationtemperatureforthecolumn.Cut10cmofftheinletendof
thecolumnand/orrevereseitbeforereconditioningit.Ifitdoesnot
help,replacethecolumn.

Backflashmaycauseremmants.Backflashreferstovapours
fromthesamplewhichexpandtoexceedthevolumeofthe Useseptumpurge
injectorliner.Thesevapoursmaycomeincontactwithcolder Lowerinjectionvolume
spots,suchastheseptumandgasinletsoftheinjector.Less
Enlargeinjectorliner
volatilecomponentsmaycondense.Thesecondensatesmay
Optimizeinjectortemperature
vaporizelaterandinterferewithsubsequentanalyses,sometimes
producing"ghostpeaks". Usepressurepulsedprogram

Bleedfromtheseptumorfragmentsoftheseptumlodgedinthe Cleantheinlet.Replacetheinletliner,glasswoolandseals.
inletorliner.

Irreproduciblepeakheightsorareas

Possiblecause Suggestions
Inconsistentinjection Developareproducibleinjectiontechnique.Useautosampler.

Distortedpeakshapescanadverselyaffectquantitative Correctanyproblemsthatresultinthedistortionofpeakshape.SeePeak
determinations shapeproblems.

Baselinedisturbances SeeBaselineproblems.

VariationsinGCoperatingparameters Standardizeoperatingparameters.

Negativepeaks

Possiblecause Suggestions
Incorrectpolarityoftherecorder Reversepolarityofrecorderconnections.

Incorrectsetupinthesoftware Setuprightparametersinyourchromatographysoftware.

Samplecompoundhasgreaterthermalconductivity Ifpossible,changecarriergas.Otherwisethereisnotasolution.
thanthecarriergasandyouareusingaTCDor
TCDdetector

Detectoroverloadinelementspecificdetectors Havethecompoundofinterestarriveatthedetectoratadifferenttimefromthe
suchasECD,NPD,FPD,etc.,canproduceboth solventorothercompoundsinhighconcentration.H2producesnegativepeakswith
positiveandnegativepeaks TCD(TCD)andheliumcarriergas.

DirtyECDdetectorcangivenegativepeakaftera CleanorreplacetheECDdetector.
positiveone

Nopeaksatall

Possiblecause Suggestions
Defectivesyringe Tryaneworprovensyringe.

"Blown"septumormassiveleaksattheinlet Findandfixleaks.

Problemswithcarriergasflow Adjustgasflow.Checkthecolumnflowaththecolumnoutletbyimmersionto
methanol.

Brokencolumnorcolumninstalledinthewrongway Replaceorreinstallthecolumn.

Thedetectorisnotfunctioningornotconnectedtothe Ensurethatdetectorisworkingproperly.E.g.:IstheflameinaFIDon?Check
recorderorintegrator. connectiontotheoutputdevice.

Selectivesensitivityloss

Possiblecause Suggestions
Contaminationofcolumnand/orlinercanleadtolossofsensitivity Cleanliner.Bakeoutthecolumnorreplaceit.
foractivecompounds

Injectorleaksreducethepeakheightofthemostvolatile Findandfixanyleaks.
componentsofasamplemorethanlessvolatile

Initialcolumntemperaturetoohighforsplitlessinjectionwhichcan
Possiblecause Initialcolumntemperatureshouldbebelowtheboilingpointofthe
Suggestions
preventrefocusingofsample.Thisaffectsthemorevolatile solvent.Decreasetheinitialcolumntemperatureoruseless
componentsmost. volatilesolvent.

Inletdescrimination.Injectortemperatureistoolow.Latereluting Increaseinjectiontemperature.
andlessvolatilecompoundshavelowresponse.

Offline
Splitpeaks

Possiblepeaks Suggestions
Fluctuationsincolumntemperature Repairtemperaturecontrolsystem

Mixedsamplesolventforsplitlessoroncolumninjections Usesinglesolvent

Whenusinginjectiontechniquesthatrequire"solventeffect"refocusingsuchas Installingaretentiongap(5metersofuncoated
splitlessinjectiion,thesolventmustformacompact,continuousfloodedzoneinthe butdeactivatedcolumn)aheadofthe
column.Ifthesolventdoesnotwetthestationaryphasesufficientlyasmightbethe crhomatographiccolumnmayreduceor
caseformethanolusedwithanonpolarstationaryphase,thesolventfloodedzonemay eliminatethisproblem.
beseveralmeterslongandnotofuniformthickness.Thiswillresultinbroadand
distortedpeaksbecausethesoluteswillnotberefocusedintoanarrowbandnearthe
beginningofthecolumn.

Tailingpeaks

Possiblecause Suggestions
Contaminatedoractiveinjector Cleanorreplaceinjectorliner.Donotuseglasswoolintheliner.Ifnecessary,replacethecolumn.
liner,sealorcolumn

Deadvolumeduetopoorly Confirmbyinjectinginertpeak(methane).Ifittails,columnisnotproperlyinstalled.Reinstallliner
installedlinerorcolumn. andcolumnasnecessary.

Raggedcolumnend Scorethetubinglightlywithaceramicscoringwaferorsapphirescriberbeforebreakingit.Examine
theendusingmagnifyingglass.Ifthebreakisnotcleanandtheendsquare,cutthecolumnagain.
Pointtheenddownwhilebreakingitandwhileinstallinganutandferruletopreventfragmentsfrom
enteringthecolumn.Reinstallthecolumn.

Abadmatchbetweenthe Changethestationaryphase.Usuallypolaranalytestailonnopolarcolumns,ordirtycolumns.
polaritiesofthestationaryphase
andthesolvent

Acoldregioninthesampleflow Removeanycoldzonesintheflowpath
path

Debrisinthelinerorcolumn Cleanorreplacetheliner.Cut10cmofftheendofthecolumnandreinstallit.

Injectiontakestoolong. Improveinjectiontechnique.

Splitrationistoolow Increasesplitratiotoatleast20:1

Overloadingtheinlet Decreasethesamplevolumeordilutesample

Sometypesofcompoundssuch Tryamorepolarcolumn.Makeaderivativeofthedsample.
asalcoholicamines,primaryand
secondaryaminesandcarboxylic
acidstendtotail.

Retentiontimeshifts

Possiblecause Suggestions
Changeincolumn CheckGCoventemperature
temperature

Changeingasflowrate Injectadetectableunretainedsamplesuchasmethanetodeterminethelineargasvelocity.Adjustgas
(linearvelocity) pressureorflowtoobtainpropervaluesforyouranalyticalmethod.

Leakintheinjector Checktheseptumfirst.Change,ifnecessary.Findtheleakandfixit.

Changeofsolvent Usethesamesolventforstandardsandsamples.

Contaminatedcolumn Bakeoutthecolumn.Cut10cmofftheendofthecolumn.Ifnecessary,replacethecolumn.

Lossofresolution

Possiblecause Suggestions
Damagetostationaryphaseofcolumn Replacethecolumn.Thisisusuallyindicatedbyexcessivecolumnbleedingorpeaktailing.

Injectorproblems Checkforleaks,inapropriatetemperature,splitration,purgetime,dirtyliner,glasswoolinliner.

Offline
Possiblecause Suggestions
Largeincreaseinsampleconcentration
Dilutesample
Injectless
Usehighersplitratio

Broadsolventfront

Possiblecause Suggestions
Badcolumninstallation Reinstallcolumn

Injectorleak Findandfixleak

Injectionvolumetoolarge Decreasesamplesizeordiluteit

Injectiontemperaturetoolow Increaseinjectiontemperaturesotheentiresampleisvaporized"instantly".Aninjectiontemperature
higherthanthetemperaturelimitofthecolumnwillnotdamagethecolumn.

Splitratioistoolow Increasesplitratio.

Columntemperaturetoolow Increasecolumntemperature(bacarefulonmaximumcolumntemperaturelimit).Usealowerboiling
solvent.

Initialcolumntemperaturetoo Decreasetheinitialcolumntemperature.Usealessvolatilesolventsotheinitialcolumntemperatureis
highforsplitlessinjection belowthesolventboilingpoint.

Purgetimetoolong(splitless Useashorterpurgevalveclosetime.
injection)

Rapidcolumnperformancedegradation

Possible Suggestions
cause
Brokencolumn Replacecolumn.Avoiddamagingthepolyimidecoatingonthecolumn.Avoidetemperaturesabovemaximu
columntemperaturelimit.Avoidabrasionofthecolumn.Remember,evenifthecolumndoesnotbreak
immediately,whenprotectivecoatingisdamagedthecolumnmaypossiblybreakspontaneouslylater.

Columntoohotfor Replacethecolumn.Staybelowlimitsspecifiedforthecolumn.
toolong

Exposureto Findandfixanylieaks.Besurecarriergasissufficientlypure.
oxygen,particularly
atelevated
temperatures

Chemicaldamage Keepinorganicacidsorbasesoutofcolumn.Neutralizesamples.
duetoinorganic
acidsorbases

Contaminationof Preventnonvolatilematerialsfromgettingintocolumn.Forexpample,useaguardcolumn.
thecolumnwith
nonvolatile
materials

Columninstallation

DetailedinformationaboutGCcolumninstallationisavailablehere(/i/capillarygccolumninstallation?lang=EN).

Preventivemaintenance

Injectorcleaning

Detectorcleaning

ABOUTUS PRODUCTS KNOWLEDGEBASE CONTACTS

Company certificates (/i/company Productcatalogue/eshop(/eshop/? Chromatography Contact data (/i/contactinfo?


certificates?lang=EN) lang=CZ) (/i/chromatography?lang=CZ) lang=CZ)
Trade conditions (/i/trading Product certificates (/s/downloads? Gas detection (/i/gasanalysisand Emailcontactform(/emailform.php?
conditions?lang=CZ) lang=CZ) detection?lang=CZ) lang=CZ)
News(/s/news?lang=EN) Physical property measurement
(/i/physicalpropertiesmeasurement?
lang=CZ)

Recommend Share Share

Offline
CopyrightCHROMSERVISs.r.o.,Jakobiho327,CZ10900Praha10~Tel:+420274021211~Znalostndatabze(/s/kb?lang=CZ)~eshopodMyWebdesign.cz(http://mywebdesign.cz/)

Offline

You might also like