MS3000 User Guide
MS3000 User Guide
www.malvern.com
3000
User manual
Head office:
Mastersizer, Malvern and the green 'hills' logo are registered trademarks in the UK and/or other
countries, and are owned by Malvern Instruments Ltd.
Table of contents
Part 1 - Operators Guide
Software overview
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Powering on and starting the software . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Software user interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Customising the software interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Maintenance utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Making measurements
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Measurement types and process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Measurement prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Making SOP measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Making manual measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Measurement settings reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23
Reporting
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
About reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Viewing and using reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Creating and editing reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
System security
About security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Activating security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Software licensing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Audit history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
Advanced features
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Doing more with SOPs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Editing results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
Exporting data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7
Part 3 - Appendices
Setup reference
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Installing the software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Making connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Specifications
Optical unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Chemical compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
Introduction
to this manual
Introduction
This manual covers the operation of the Mastersizer 3000 particle characterisation
system. The instrument measures the size of particles contained within a sample,
presenting data according to the users needs.
This user manual is a companion to the Mastersizer 3000 Basic Guide, which
gives Health and Safety, maintenance, troubleshooting and other vital information
which all users must read.
Part 3 Appendices
Appendix A gives setup reference material and Appendix B details specification
and regulatory information.
Warning!
The instrument and the samples to be measured may be hazardous if mis-
used. Users must read the Health and Safety information in the Basic
Guide before operating the system.
Warning!
Removal of the covers by unauthorised personnel will invalidate the war-
ranty of the instrument. Unless advised within the content of this manual,
only Malvern Instruments trained personnel are permitted to remove the
main cover of any part.
Supervisor
The supervisor is the person responsible for the management and safety of the
instrument and its operation. The supervisor is responsible for the training of the
operators. The supervisor can perform all user maintenance routines identified in
the Basic Guide and within this manual.
Warning!
Under no circumstances should the supervisor or an operator remove the
main cover of the instrument.
Operator
An operator is a person trained in the use of the system. The operator can perform
all user maintenance routines identified in the Basic Guide.
Warning!
Failure to follow these guidelines could result in the emission of laser radi-
ation. Laser radiation can be harmful to the body and cause permanent eye
damage.
Assumed information
To make full use of this manual, the user should understand the following points.
Naming convention
Within this manual:
The Mastersizer 3000 is referred to as the optical unit or the instrument.
The sample dispersion units are referred to as the Hydro MV or the Aero
S, or as the dispersion unit. Wet dispersion units refers to the Hydro LV,
MV and EV. Dry dispersion units refers to the Aero S.
The combination of the optical unit, one or more dispersion units and the
computer is referred to as the system.
Ribbon commands
The control ribbon in the Mastersizer software is divided into tabs:
The first button on this tab bar is referred to as the Application button and has a
similar function to the File menu on older systems.
Ribbon commands are referred to in this manual in the form: ribbon tab selection,
followed by option. For example, Home-Manual Measurement describes click-
ing the option shown below which is found on the Home tab:
Manuals
The following manuals are available for the Mastersizer 3000:
Basic Guide
A concise yet essential guide that provides an introduction to the system and its
usage covering Site requirements, Health and Safety and an instrument and
accessory overview. Finally a brief guide to making a measurement and view-
ing the results is included.
User Guide
The primary source of information about the Mastersizer 3000 hardware and its
use. See About this manual earlier in the chapter.
Dispersion unit guides
Separate guides are supplied with both the Hydro and Aero dispersion units. These
provide essential information on the connection, operation and maintenance of the
dispersion units.
Help System
Each key software window has a Help button giving information specific to it.
Additional Info Tips are also available whenever is shown - simply move the
mouse pointer over this icon to see a relevant tip.
Technical support
Malvern Instruments offers several methods of technical support, which are
described in this section.
Email support
Note
This is the primary support method only use the other methods if this is
not available at your location.
Additional support
Contact the International Malvern Helpdesk if the local Malvern Instruments
representative is not available:
Telephone: +44 (0)1684 891800 or email: [email protected].
Note
This help line is primarily English speaking.
Remote support
Malvern Instruments offers a remote support service, delivered by an internet con-
nection. Benefits include fast and efficient fault diagnosis, as well as reducing
downtime and costs.
Introducing the
Mastersizer 3000
Introduction
This chapter provides information about the Mastersizer 3000 hardware and iden-
tifies all major features of the system.
The following topics are covered:
A description of the purpose of the instrument and its accessories.
A hardware overview of a typical wet and dry system.
Detailed reference on the physical parts of the optical unit.
Information about the parts of the cell, including how to insert and remove it.
Note
Hardware maintenance for details on how to maintain the hardware,
please refer to the Basic Guide and the separate dispersion unit guides.
Optical unit
The main optical unit, sometimes referred to as the optical bench or just the instru-
ment, is the core of the Mastersizer 3000 system. In its simplest terms, its purpose
is to transmit red laser light and blue light through a sample and then use its detec-
tors to generate data about the light scattering pattern caused by the particles in the
sample, which can be interpreted by the Mastersizer software to provide accurate
particle size information.
Measurement cell
The cell is the interface between the dispersion unit and the optical unit. The sam-
ple is routed between measurement windows in the cell so that the laser can pass
through it in order to make a measurement.
Typical systems
Wet dispersion
A wet system consists of the optical unit and either the Hydro LV/MV or EV dis-
persion unit. These dispersion units are designed to circulate a liquid sample
through the Mastersizer 3000's measurement cell. Many materials need to be meas-
ured as wet dispersions. Typically these are materials that are used or produced wet
in the manufacturing process or need to be dispersed in a liquid in order to achieve
reproducible dispersion.
Hydro LV/MV
The Hydro MV and LV units are designed to create a suspension of particles in
water or other liquid media. The Hydro LV can handle a liquid volume of 600ml,
whereas the Hydro MV unit has a volume of 120ml. Both can be configured to
automatically deliver liquid dispersants and also dispose of the sample following a
measurement. This enables the cleaning and filling of the units as part of an SOP.
The large capacity of the Hydro LV makes it ideally suited to the measurement of
samples which have very broad size distributions or contain large particles. In some
cases, sample materials are not plentiful and the MV's smaller volume sample dis-
persion unit is a sensible choice to minimize sample and dispersant consumption
while retaining the benefits of totally automated sample dispersion and measure-
ment.
Hydro EV
In the mineral, chemical, metal powder refining and cement industries, measure-
ments of larger sample volumes are often most appropriate for QC of the manufac-
turing process and finished product. The large capacity of this unit makes it ideally
suited to the measurement of such samples as well as larger materials having broad
size distributions.
The Hydro EV is a wet dispersion unit designed to circulate a liquid sample held
within a standard laboratory beaker through the Mastersizer 3000's measurement
cell. As such the filling and emptying of this unit is a manual process.
3 1 2 4
ill 8620
This shows:
Optical unit (as described in the next section) measures the sample using
red and blue light wavelengths.
Dispersion unit (in this case the Hydro LV) circulates the sample through
the cell
Wet cell sample flows through the air cell so that it passes through the
instruments laser path.
Computer running the Mastersizer software this controls both the opti-
cal unit and the attached dispersion unit, as well as processing and presenting
the measured data.
Other wet dispersion units include the Hydro MV and Hydro EV.
Dry dispersion
A dry system comprises of the main optical unit plus the Aero S dispersion unit.
The Aero S is designed for providing an even circulation of dry powder throughout
Mastersizer 3000 measurement cell.
Direct software control over the feed rate and air pressure is provided, which ena-
bles flexible control of both the sample throughput and dispersion. The unit can be
controlled either independently of measurement or from within a manual/SOP
measurement. The following illustration shows a typical dry dispersion unit instal-
lation:
4 5 2 3 1
ill 8652
This shows:
Optical unit (as described in the next section) - measures the sample using
laser and blue light detection.
Dry dispersion unit (in this case the Aero S) - circulates the sample through
the cell.
Dry cell - sample flows through the air cell so that it passes through the instru-
ments laser path.
Vacuum unit removes the sample from the cell exhaust once it has been
measured.
3 6 4 1
5 7 8 9 2
ill 8622
The optical unit comprises:
Optical unit
Cell (in this example, the wet cell). The following types of cell are available:
wet - for Hydro units and dry (sometimes known as the air cell) - for use with
the Aero S unit.
End panel provides the communication connectors for the optical unit as
well as the power connection and switch. See Appendix A for further connec-
tion reference.
Protection window stops dust/dirt entering the system. Clean this window
periodically. See Appendix B for more information on cleaning the protection
window.
Tube/cable routing neatly routes all connection pipes and cables under-
neath the instrument.
End panel
The end panel of the instrument provides the communications and power connec-
tions.
1
2
ill 8623
Power switch press once to power on the unit, press again to power off.
USB provides data communication with the PC. For best performance con-
nect to a high speed USB port on the PC:
CAN 1-3 supplies power and data communication to the dispersion unit.
Power input input for 24V power supply unit.
Cell components
The following illustration shows the key parts of a wet cell for a more detailed
description of both the wet and dry cells, refer to the Hydro or Aero Dispersion
Unit Guides respectively:
1 2
6
3
ill 8625
Cell eject button ejects the cell from the cell bay locking mechanism so that
it can be withdrawn from the cell bay.
Cell handle only lift the cell by this handle.
Sample output (blue) connected to the sample in (blue) on the dispersion
unit.
Caution!
Do not try to remove the cell from the cell bay without first pressing the
cell eject button as this could damage the locking mechanism.
2. Withdraw the cell from the instrument using its handle in the direction
shown in the illustration on p.2-7.
Note
If the cell is raised and lowered too many times in a short period, the firm-
ware will lock the cell in place for a period of time before it can be ejected
again; this is to prevent damage to the locking mechanism.
Software overview
Introduction
The Mastersizer software serves several key functions: it controls both the optical
unit and the dispersion unit; it allows users to create and edit measurement files; it
enables users to run manual or SOP measurements on both wet and dry samples,
and finally it provides access to powerful reporting and data analysis features.
This chapter gives details on the following:
Powering on and starting the software.
Description of the software user interface.
Customising the software interface.
Maintenance utilities.
Note
This section provides a description of the key elements of the software
only. For information on how to run a measurement, see Chapter 4.
Some advanced features of the software are also detailed in the Supervisors
section of this guide and reference information for all features is provided
in the Help system.
If the instrument is not connected correctly, the status bar indicates:
If this is the case, check the instruments connection to the USB port on the
PC and the power connections to the instrument.
If the dispersion unit is not detected, the following is shown:
If this is the case, check the connection from the instrument to the dispersion
unit. Also, a blue pulsating power light on the dispersion unit indicates that it is
connected correctly to the instrument.
Note:
Before using the instrument to measure a sample, wait for 30 minutes for
the laser to stabilise.
5
6
Ribbon selector tabs provide quick access to the different control ribbons.
The Application button (leftmost) gives quick access to file, print and soft-
ware options (including security) as well as software version information.
Control ribbon single-click access to key software functions.
Record View panel lists all measurements within the active measurement
file. Multiple measurement files can be opened simultaneously, which are then
accessible by different tabs within the Record View.
Reports tabs provides access to reports on the currently selected record. The
reports shown are a function of the currently selected workspace - this is dis-
cussed later in this chapter.
Data Quality tab provides guidance on the measurement quality and dis-
plays a simple pass/fail data report based on selected models, together with tips
on how to improve the measurement. Described further later in this chapter.
All files tab view all records from all open measurement files in a single tab.
Status bar shows the instrument mode, connected user, 21 CFR status,
notifications (see later in this chapter), connected accessory type and the instru-
ment connection status (as well as its serial number when connected).
Control ribbons
Control ribbons provide quick access to key software functions.
To select an option, click on the appropriate direct-access button, for example
Extract SOP.
If you contract the software window horizontally, the individual ribbon options are
grouped under a single icons instead. The ribbon then operates in a similar manner
to menus in that the group must be selected initially to reveal the sub-items, for
example Result-Extract SOP.
To select a different control ribbon, click the selector tabs above the ribbon
(Home, View Tools or Audit).
Each open measurement file is given a tab in the Record View: click one of
these tabs to show all of the records contained in that measurement file.
Note
You can open as many measurement files as required, and also combine
results, using All Files, to view all records within a single tab, making
graphical data overlay possible. See Merging records in Chapter 5 for
more information.
Tip
To copy the data shown in the record view to the clipboard, select records
from the record view and then choose Clipboard-Copy on the Home
ribbon. This data can then be pasted into any other compatible Windows
application, for example, Word or Excel.
Powerful analysis features also allow records to be sorted, grouped and filtered so
that users can focus on only those records that are of relevance to their application.
See Chapter 5 for more information on grouping and filtering.
Reports tab
Reports are displayed automatically whenever you select a record, or number of
records, from the Record view (assuming that the selected view is showing the
Reports tab).
Note:
For detailed information on the process of viewing, editing or creating new
reports, see Chapter 7 Reporting.
The display is updated to show a quality review of each record selected. Click
to toggle whether items that have passed or failed are displayed.
Workspace Viewer
This section describes the purpose of workspaces and the Workspace Viewer.
Note
See the Help system for information on the Macros tab.
Workspaces are a collection of settings that define the information presented in the
Reports and Record View, as well as which folders the system uses for accessing
SOPs, measurement data and export data.
Each user can access both a Private or Shared workspace, both of which are config-
urable. This makes it easy for an individual to optimise the software for their own
use, with the Private workspace, or to collaborate better across an organisation
using a Shared workspace. The latter is particularly useful within quality con-
trolled environments.
The following settings and data are associated with the workspace:
SOPs and SOP Templates presented to users (i.e. their file locations).
Record parameters shown in the Record View.
Reports listed in the Reports view (see also Chapter 7).
Measurement data file location.
Exported data and Export Templates file locations.
Note
Any of these items that are created when the Shared workspace is active
will be available to all users of that system (providing their security privi-
leges allow them access to these functions). If the Private workspace is
selected, these items are only available to the current user.
Any work subsequently performed on the system will now use the selected work-
space both to save data and to access settings.
5. Once you have added new views, make them the active view by clicking on
them - an arrow is shown next to the selected view name.
The reports and the Record View are updated immediately.
6. To edit or delete a view, first select it and then move the mouse pointer over
the view name again to reveal the Edit and Remove icons.
Shortcuts
The shortcuts listed are links to Mastersizer system folders and provide quick
access to all of your SOPs, measurement files and exported data.
The folder that is linked to from this panel is specific to the selected workspace.
Clicking on a shortcut to a folder will open that folder within Windows Explorer.
If the Private workspace is selected, the folders all stem from:
C:\Users\YOURNAME\Documents\Malvern Instruments\Mastersizer
3000\Workspace\
Creating shortcuts
Shortcuts are an easy mechanism for users to set-up a set up links to folders, e.g. on
a network, that can be shared between all users on that PC.
To create a shortcut: Create new shortcuts to files/folders by dragging them into
the Shortcuts panel from Windows Explorer. Shortcuts to web sites can be added
by dragging a bookmark from Internet Explorer into the Shortcuts panel.
Options
Choose Options from the Application button menu to display the Mas-
tersizer 3000 Options window. Then click the required link on the left to view its
corresponding settings.
General Preferences
Click General from the Options window to choose the following settings:
Show settings when running manual measurements this automatically
displays the measurement settings window whenever a manual measurement is
initiated. This can be useful during method development where the user needs
to switch back into the record view after making each measurement and then
make slight alterations to the measurement settings for the subsequent meas-
urement. If this option is de-selected, the settings are still available to the user
by clicking on the Manual Measurement window.
Show edited records at the same level as original records in the record
view - rather than stagger edited records as sub-items of the records from
which they originate, they are displayed at the same level in the tree. This gives
the software the same layout/behaviour as the original Mastersizer 2000 soft-
ware.
Themes
Several themes are available which modify the colour settings of the user interface.
Click Theme from the Options window and then select a scheme that suits your
preference by clicking Use this theme.
Graphs
Choose Graphs from the Options menu to access further options that relate to
the presentation of graphs:
Line weight specify the line weights (Normal, Light or Heavy) to be used in
all graphs.
Colour scheme choose from one of several colour schemes. The Master-
sizer 2000 theme emulates the appearance of the Mastersizer 2000 graph col-
our scheme, Greyscale is optimised for printing on a monochrome printer,
High Contrast provides better on-screen visibility.
Save/Restore My Layout
Save My Layout is a simple feature that allows you to store the exact position of all
elements of the Mastersizer 3000 interface. This could be useful in situations
where you have optimised the interface for your current analysis process.
This indicates that pre-set positioning options are available for the element. Whilst
still holding down the mouse button, move the mouse pointer over one of the
direction arrows in this tool - a shaded box is shown indicating how the element
will be placed. Release the mouse button to place the element.
Note:
By default the system initially returns to the window layout that was pre-
sent when the software was last closed.
In this example, the Manual measurement, Save and 2-pane tabbed buttons
have been added to the Quick Access Toolbar.
2. To remove the button, right-click it and choose Remove from Quick Access
Toolbar.
Maintenance utilities
The Maintenance window gives access to various background information about
the system that can be useful as part of an ongoing maintenance schedule or in the
process of contact with the Malvern Instruments Help Desk.
Choose Tools-Maintenance from the ribbon to display the Maintenance
window:
Notifications
Notifications are brief notes that indicate any errors, problems or necessary reme-
dial/investigative action that may be required in order to keep the system function-
ing correctly. Follow the advice given in any notifications and then remove them by
clicking Clear All.
The Status bar also provides a visual reminder that notifications are available:
Click the Notifications icon on the status bar to view the Maintenance window.
System information
The main purpose of the system information area is to display details about the
software, operating system and hardware connected. This information must be
passed to the Malvern Instruments Helpdesk whenever you log a support call. To
send this information to Malvern Instruments:
If you have email software on the Mastersizer computer:
Click Contact Malvern, complete the form and then click OK. Your default
email software then starts and an email is generated containing the relevant.
If you do not have email software on the Mastersizer computer:
Click Save text report to save a text file containing this information. Then
copy this information onto a data stick and transfer it to a computer that has an
email connection in order to attach the file to send to Malvern Instruments.
Print the information using the Print option and then refer to this information
if you wish to contact the Helpdesk via telephone.
System usage
The system usage section provides details on the total Running Time of the instru-
ment and any connected accessory. Additionally details are provided on the run-
ning time of the accessory stirrer running time and ultrasound if applicable.
Logs
The Logs section contains links to all the logs generated in the previous 24 hour
period. After 24 hours, if no further log files have been generated, only the last log
file generated is retained. Log files contain no data that is of direct use to users, but
may be requested by the Malvern Helpdesk when trying to resolve any issue. Either
click the links to open the file, enabling the text to be copied into an email, or
choose Open folder and then locate the log file you wish to send.
Making
measurements
Introduction
The Mastersizer 3000 is capable of measuring samples dispersed in a wet or dry
medium, either manually as a one-off (or for developing an SOP method) or by
utilising predefined settings stored within an SOP file.
This chapter is the main reference for the measurement making process and
includes sections on:
Measurement types a brief background on the overall process of making a
Mastersizer measurement.
Measurement file basics how to create and open measurement files.
Making an SOP measurement includes how to create SOPs as well as the
steps involved in running an SOP.
Making a manual measurement the basic steps involved in manual meas-
urement with a wet dispersion unit.
Measurement settings reference detailed reference on all measurement
settings available to both SOP and manual measurements.
Measurement prerequisites
Note
The instrument and accessory must have been connected and set up in
accordance with the guidance in the Mastersizer Basic Guide and the dry
or wet dispersion unit guide. This section assumes all appropriate connec-
tions have been made.
1. Switch on the optical unit. The pulsating blue light on the top of the instru-
ment indicates that it is on. Check also that the blue light on the dispersion unit
is pulsating - power to the dispersion unit is drawn from the instrument.
Note
Leave the instrument powered on for 30 minutes before making a
measurement to allow the laser to stabilise.
Note
If, after successfully starting the Mastersizer software, any of the functions
described in the rest of this section are unavailable to you, this will be due
to your login credentials contact your system Supervisor for assistance.
3. Ensure that the status bar indicates that the instrument is connected correctly:
4. If the instrument is not connected correctly (i.e. if anything else is shown in the
status bar), refer to Chapter 3 for more information on how to fix this.
5. Create a new measurement file by selecting New-Measurement File from
Home section of the control ribbon:
2. Select an SOP template from the list in the New SOP window or click the
From an existing SOP... button to locate an SOP in the file system that fulfils
(or is close to) your required criteria.
3. The SOP Editor window is now displayed as illustrated (the Material set-
tings are shown in this example):
4. Step through the SOP Editor by clicking the section titles on the left (Sample-
Identification, Sample-Material etc.) or clicking the Forward / Back arrows on
its toolbar.
Note
Complete each of the sections detailed in the Measurement Settings
Reference section in later in this Chapter.
5. When all the required settings have been specified, save the SOP by choosing
Documents-Save on the ribbon (or Save from the Application button menu)
and then giving the SOP a logical name.
Click on an SOP to display its properties in the bottom panel of this window.
3. After selecting an appropriate SOP, click OK.
4. The Measurement Display window is shown.
The progress status bar at the top of the window reports both the current status
and what to do next.
Note
To stop both the SOP and the accessory during a measurement, click the
Abort button . This all stop command should only be used when
a problem is suspected - this is because the sample will stop circulating,
which could affect further measurements if the sample is not then refilled.
6. When the Measure Background stage is complete, the SOP pauses. The sys-
tem now requests that you add sample - do this until the Obscuration bar
indicates about 10-20% for a wet dispersion unit. (This is a rough guide only
as this setting is highly sample-dependent - refer to the Help system for more
guidance.)
7. Click Start again. The measurement now commences, measuring first the red
and then the blue light values. The SOP may make a number of measurements
before completion, depending on the Number of measurements specified in
the SOP settings. When complete, the Trend view is updated with the new
measurement figures:
This is an example of the trend view after four measurements have been made. The
currently selected range in this example includes just two records: Record Number
2 and 3 - the figures displayed in the coloured boxes relate only to those two
records.
The data shown relates either to the individual measurement or range of measure-
ments selected by the user. The default view shows particle size statistics. Trend
plots can be configured to show any parameter the statistics shown on a plot line
relates to a single parameter only.
Main figure the actual size measurement (m) for the single base record
that is currently selected (Dv10 - blue, Dv50 - green, Dv90 - red).
Av the average particle size (m) for the selected range.
RSD Relative Standard Deviation (%) for the selected range.
Min the minimum particle size (m) for the selected range.
Max the maximum particle size (m) for the selected range.
To modify which figures are shown in the Trend view:
1. Right-click anywhere on the trend graph and choose Statistics.
2. Select or de-select the items required.
Tip
Click anywhere on the graph line to display data points - click again on the
line to remove them. Move the mouse pointer over a data point to display
its record number and size figure, for example 4, 891.09 indicates that
Record 4 has a measured size of 891.09m.
D
A
ill 8655
2. Empty the sample into the hopper .
The mesh basket and ball bearings may be used on coarser samples to smooth
the flow of sample, breaking up any loose agglomerates before they fall into the
hopper.
3. Close the lid of the Aero.
4. Select Run SOP from the Measurements section of the ribbon.
The SOP Selector window lists all available SOPs.
5. Choose an appropriate SOP for the Aero dispersion unit (ensure also that the
SOP is optimised for the venturi type that is fitted to your Aero unit - either
stainless steel or ceramic venturi) and then click OK.
6. Click Start the SOP automatically aligns the optical system and then
measures the background using red and blue light.
7. After this, the sample tray starts vibrating and the SOP-specified air pressure is
applied. The overall rate at which the sample is fed into the cell is a function of
the combination of air pressure, vibration rate and hopper flow-rate setting.
Note
To stop both the SOP and the accessory during a measurement, click the
Abort button .
8. The SOP selected should be optimised for the dry dispersion unit and will only
make a measurement once the obscuration is within the specified range (1-10%
obscuration is the rough guide for a dry dispersion unit).
9. The results are shown immediately in the Trend view as they are gathered.
When all of the measurements have been made, the system pauses - optionally,
click Clean System on the top progress bar to initiate a cleaning cycle.
10. Complete the measurement by closing the SOP Measurement window.
11. The results are now presented in the Record View.
Note
This process description assumes that a wet dispersion unit is connected to
the instrument.
The basic procedure is similar to that outlined in the SOP measurement section,
but requires the user to set up the measurement settings initially, and then manu-
ally initiate more stages during the live measurement process:
1. From the Home ribbon click Manual measurement from the Measure-
ments group. The Manual Measurement window is displayed, together
with the Manual Measurement Settings window:
Note
This section now provides information on how to enter only some of the
key measurement settings - for complete details on all settings, see Meas-
urement settings reference later in this chapter.
7. From the Initialise Instrument drop down menu, select Auto Alignment.
8. Click the Start button to initialise the instrument - when complete, the
system progresses to the Measure Background stage.
9. Filling and starting the dispersion unit:
Depending on the accessory that is currently connected, the control options
(shown in the panel on the right) will vary. For a wet dispersion, a steady circu-
lation of dispersant needs to be in place before adding the sample. To do this,
ensure the accessory is filled with dispersant - if using an LV/MV, click Fill in
the Dispersant section of the Accessory panel.
Note
See Controlling dispersion units (manual mode) later in this chapter
for more information.
10. When the accessory is full, set the Stirrer speed and start the accessory:
11. Click Start on the measurement progress bar at the top of the window. Both
the red and blue light values for the background are measured.
12. The system now requests the user to Add Sample - add small quantities of
sample to the dispersion unit until the obscuration is within range (10-20% in
the case of a wet dispersion). Refer to the Help system for more guidance. With
a dry dispersion, the system will skip this step.
Note
With dry dispersions, the whole sample is loaded into the hopper in the
Aero before the measurement process is started.
13. Click Start again twice to begin the measurement. First the red and then the
blue light values are measured. When this is complete, the Trend view is
updated with the new measurement figures (see Running an SOP in this
chapter for more detailed information on the Trend view). The particle Size
Distribution graph is shown in the upper right panel. In the example shown
here, just a single measurement was specified in the manual measurement set-
tings.
14. Continue this final Measure Sample process as many times as required by
clicking Start again. To complete the measurement, just close the Manual
Measurement window. The new records are now added to the Record View.
Tip
The same controls are also available, when not making a measurement,
from the Active accessory control feature (choose Tools-Accessories
from the main control ribbon). This option could be useful, for example, if
you wanted to keep a sample in circulation whilst undertaking a report
analysis.
The options available on this panel depend on the connected accessory as follows:
Hydro LV/MV
The accessory controls panel for the Hydro LV/MV is shown as follows (in this
example the dispersant has been filled):
Option Description
Stirrer Use the slider bar to set the RPM of the stirrer. In Manual
Measurement mode it is necessary to start the stirrer
manually before making the measurement. To initiate the
stirrer once the speed has been set, click the Start but-
ton. To stop the stirrer click the Stop button.
The green bar represents the user-requested speed. The
thin grey bar beneath this is the actual stirrer speed. There
will be a small lag as the actual speed adjusts to match
the requested speed.
Option Description
Enable pulsed In this mode, sonication is applied in pulses (with fixed
sonication duration on and off pulses). This option can be used in
combination with both continuous or timed sonication
mode. Specify the duration of on and off pulses by enter-
ing the required values (seconds) into the corresponding
boxes. This can prevent overheating of the sample, and is
also useful on fragile samples that could otherwise be
broken up by more aggressive ultrasonication.
Manual Select this option if you wish to start and stop the stirrer
manually for a period of time.
Timed Select this option to perform a precisely timed period of
ultrasound application. When Timed is selected, the timer
controls become available - click +/- to increase/decrease
the time period for which ultrasound will be applied.
Fill Fill the Hydro unit with dispersant. This function opens the
(dispersant) dispersant valve and then pumps dispersant into the unit
until the correct level is reached (detected by a sensor)
and then closes the dispersant valve. The tank indicator
next to this button shows full and empty respectively:
Empty Empty all dispersant (and sample) from the Hydro unit.
(dispersant) This function opens the drain valve and initiates a pump-
ing sequence - the process stops when the unit is empty
and then closes the drain valve.
Drain valve Opens/closes the drain valve. This will allow any disper-
(dispersant) sant and sample in the system to drain naturally. No pump
is applied, so this process will not completely void the
system to the same extent as the Empty function. This
could be useful if, for example, the system had been man-
ually overfilled and needed some adjustment of the level.
Dispersant Opens/closes the dispersant input valve. It is necessary to
valve open the valve before manually adding dispersant. This
might be necessary if, for example, the system had been
partially drained (using the Drain valve option) and then
needed to be manually topped up.
Hydro EV
The accessory controls panel for the Hydro EV is shown as follows:
Option Description
Stirrer Use the slider bar to set the RPM of the stirrer. In Manual
Measurement mode it is necessary to start the stirrer
manually before making the measurement. To initiate the
stirrer once the speed has been set, click the Start but-
ton.
Ultrasound Choose an appropriate level of ultrasound for the sample
by moving the slider bar.
Enable pulsed In this mode, sonication is applied in pulses (with fixed
sonication duration on and off pulses). This option can be used in
combination with both continuous or timed sonication
mode. Specify the duration of on and off pulses by enter-
ing the required values (seconds) into the corresponding
boxes. This can prevent overheating of the sample, and is
also useful on fragile samples that could otherwise be
broken up by more aggressive ultrasonication.
Manual Select this option if you wish to start and stop the stirrer
manually for a period of time.
Timed Select this option to perform a precisely timed period of
ultrasound application. When Timed is selected, the timer
controls become available - click +/- to increase/decrease
the time period for which ultrasound will be applied.
Aero S
The accessory controls panel for the Aero S is shown as follows (in this example the
Air pressure/Feed rate have been set and the accessory started):
Option Description
Standby By default the Aero is in standby mode when the acces-
sory is connected and powered up, but not active. In this
mode the front LED is pulsating.
Air flow This enables the vacuum cleaner and also the com-
pressed air supply used for sample dispersion.
Feed Start the vibrating feed tray. Selecting this option also
automatically turns on the Air Flow.
Stop Stop all operations and close all valves. The accessory is
returned to the Standby status.
Air pressure Use this slider bar to set the air pressure at which the
sample is circulated (from 0-4 bar, in 0.1 bar increments).
Lower air pressures tend to be better for fine or fragile
particles, higher air pressures for agglomerates or metallic
particle samples.
Feed rate Use the slider bar to set the rate at which the sample is
fed into the system - effectively this controls the vibration
speed of the feed tray. The correct feed rate is one at
which the sample is circulated steadily through the sys-
tem and gives the required obscuration - this rate is best
established as part of a method development process.
The Measurement settings window is displayed whenever you create a new SOP or
manual measurement. This window groups settings into logical sections in which
various parameters for the measurement are defined as follows:
Sample identify the sample and provide more information to users of the
SOP. Define the physical characteristics of the sample and dispersant.
Measurement set both the red and blue light measurement durations, the
number of, and delay duration between, measurements. Specify the obscura-
tion levels for the measurement.
Sample Dispersion control the behaviour of the currently connected acces-
sory.
Data Processing specify the scattering model, latex analysis options and
modify the light detectors used in a measurement. Define any Analysis
Smoothing, Size Range/Resolution options and set size banding for all meas-
urement histograms and tables.
Note
When progressing through the settings, any icon next to a field indicates
that the user must specify a value for the field before continuing to the next
section.
The rest of this section goes through each of the measurement settings in more
detail.
Sample - Identification
The Sample Identification settings section allows you to enter details that will both
identify the sample and provide more information to users. When the measure-
ment is run, the user is prompted to confirm or alter the fields specified here.
Field Description
Sample Name A descriptive name for the sample, such as "Batch 1A" or
(mandatory) "Series 3, Sample 1". This is the only default sample iden-
tification field.
To add further fields for example, Notes, Batch, Lot, Group etc., click the
Add button. The new field is added to the list. Enter a name for the field in the
drop down menu to the left of the field. Check Required field if this is man-
datory for the operator to enter.
To re-order the list of fields, drag the item up or down by clicking and drag-
ging the icon next to the item.
Sample - Material
Use the Material properties section to enter physical characteristics of the mate-
rial being measured.
To select a previously added material, click Browse database, then select an item
from the database and click Use these values. See the Help system for more on
the Materials and Dispersants databases.
Alternatively, manually complete the fields detailed below.
Field Description
Material name Descriptive name for the material.
Refractive Value of between 0 and 5. This value relates to the speed
index of light within the material, which in turn allows the degree
of refraction (light bending) to be predicted when light
passes from one medium to another.
Absorption Value between 0 and 10, which is a measure of the quan-
index tity of light absorbed by the particles. Generally, transpar-
ent samples will have a low or zero absorption while
coloured or black samples will have a higher value.
Density Value of between 0.001 and 25g/cm.
Field Description
Different blue- If required, add the Refractive index (blue-light) and
light properties Absorption index (blue-light). See Help System for more
information.
References and Any further comments that help to describe the material.
notes
Sample - Dispersant
Use the Sample Dispersant section settings to specify the properties of the dis-
persant used in the measurement.
It is possible to specify the dispersant details manually or to choose from the Dis-
persants Database. To select a previously added material from the Dispersants
Database, click Browse database, then select an item from the database and click
Use these values.
Alternatively, manually complete the fields detailed below.
Field Description
Dispersant Descriptive name for the dispersants, for example, Etha-
name nol.
Field Description
Refractive Value of between 0 and 5. This value relates to the speed
index of light within the material, which in turn allows the degree
of refraction (light bending) to be predicted when light
passes from one medium to another.
References and Enter any further comments that help to describe the dis-
notes persant.
If the newly added dispersant will be used in the future, add it to the Dispersants
Database by clicking Add to database.
Sample - Instructions
Use the Sample Instructions settings to specify pre-measurement and post-
measurement instructions that need to be drawn to the attention of the user.
These are then shown in the Sample Documentation panel/window before and fol-
lowing the measurement respectively. In this example pre-measurement instruc-
tions have been specified and are displayed above the Sample Name field in the
Sample Documentation window.
Measurement - Duration
The Measurement Duration settings can be used to separately specify both the red
and blue light measurement durations.
Field Description
Background This facility will make a measurement using only clean
measurement dispersant this is then subtracted from the sample
duration measurement by the system in order to clean the data.
(seconds) This is often set to the same duration as the sample
measurement.
Sample meas- The optimum measurement time depends on the size of
urement dura- the sample, its particle size distribution and the dispersion
tion (seconds) accessory used. If a material is monomodal, its essential
particle size characteristics can be captured in less time
than a material with a broad particle size distribution,
which will need to be measured longer to ensure that the
coarser particles have been properly represented.
Some red and yellow samples (such as pigments) can absorb light in the blue spec-
trum. This may lead to problems with the validity of the data acquired during the
blue light part of the measurement. If you have such a sample, the blue light meas-
urement can be disabled, so limiting the measurement to the main red laser.
If no blue light measurement is required, select Don't perform blue light meas-
urement. This will speed up the measurement time.
Measurement - Sequence
The Measurement Sequence settings window allows you to specify the number of
measurements and the duration of any delay between measurements.
Select either Automatic to have the system measure until stability is reached, or
Use the number I specify and then enter the Number of measurements (up to
999) and Delay between measurements (up to 9999 seconds). Specifying several
measurements, e.g. 5, allows repeatability to be assessed.
If the material is difficult to disperse, requiring ultrasonication, surfactant or both
to achieve a stable dispersion, several measurements of the same sample may be
required in order to determine the ideal dispersion time. In general, repeat meas-
urements are performed until the last measurement obtained agrees closely with
the previous one.
Measurement - Obscuration
The Measurement Obscuration settings enable you to specify the obscuration lev-
els between which the measurement will be conducted. This is critical for ensuring
the correct amount of sample is added to the measurement system.
The optimal obscuration settings for a measurement are both sample and disper-
sion unit dependent. As a rough guide, use a range of 10-20% for a wet dispersion
unit and 1-10% for a dry dispersion unit.
Note
For more advice on obscuration settings see the Help.
Field Description
Obscuration Enter a percentage value from 0.1 to 50 above which the
lower limit (%) measurement should be run.
Obscuration Enter a percentage value from 0.1 to 50 below which the
higher limit (%) measurement should be run.
Field Description
Air pressure Use this slider to set the air pressure at which the sample
is circulated (from 0-4 bar, in 0.1 bar increments). Lower
air pressures tend to be better for fine or fragile particles,
higher air pressures for agglomerates or metallic particle
samples.
Feed rate Use the slider bar to set the rate at which the sample is
fed into the system effectively this controls the vibration
speed of the feed tray. The correct feed rate is one at
which the sample is circulated steadily through the sys-
tem and gives the required obscuration - this rate is best
established as part of a method development process.
Field Description
Venturi type Select the currently attached venturi type. If you do not
know this, select Unknown Venturi.
Tray type Select the currently attached tray type. If you do not know
this, select Unknown Hopper.
Hopper gap Enter the gap as currently set on the hopper hardware.
(mm)
Field Description
Stirrer speed This accessory type has an independent stirrer allowing
flexible control of sample flow and agitation. The stirrer
speed may be controlled manually using the Wet Acces-
sory dialogue or by a predefined SOP for automatic oper-
ation.
Manual control would typically be used to determine the
optimum measurement setting prior to using an SOP pro-
cedure or when doing one-off manual measurements.
Field Description
Stirrer speed To control the stirrer from an SOP, determine the opti-
(continued) mum stirrer speeds initially using manual control, ensuring
that the lowest setting to give satisfactory dispersion is
obtained. Make a note of the settings and enter these into
the SOP Accessory control settings. Click and drag the
slider bar to set the speed.
It is also possible to control the accessory manually by
using the Accessory Controls panel (from the manual
measurement window) or the from the Tools-Accessories
option at any other time.
Tank fill The tank can be filled with dispersant manually or auto-
behaviour matically. When set to Automatic the dispersant is
plumbed into the dispersant inlet port at the rear of the
(Hydro MV/LV accessory and a solenoid valve controls the flow of dis-
only) persant into the tank.
However, if a solvent is to be used as a dispersant, it is
unlikely that the solvent will be plumbed into the acces-
sory. In these situations the filling of the tank can be set to
Manual. Under manual filling the solenoid valve is
switched off (this will mean that the dispersant inlet does
not have to be disconnected during a manual fill) and the
tank can be filled by hand. The software will tell the opera-
tor when to fill the tank during an SOP measurement.
Ultrasonics This sample dispersion unit has an ultrasonic probe which
can assist the dispersion of cohesive samples. The level
and duration of ultrasound may be set manually using the
Wet Accessory dialogue or may be set by a predefined
SOP for automatic operation.
To determine the correct ultrasonic level for an SOP, ini-
tially determine the optimum ultrasound level and duration
using manual control, ensuring that the lowest setting giv-
ing satisfactory dispersion is obtained. Record the set-
tings and transfer these to the SOP in the Sampler
Settings page using the slider controls. Click and drag the
slider bar to set the power level.
Field Description
Ultrasonics Set the Ultrasound Mode to one of the following:
(continued...) Manual - this is typically be used to determine the opti-
mum measurement setting prior to using an SOP proce-
dures or when performing one-off manual measurements.
Pre-Measurement - ultrasound is activated for a set time
prior to measurement. For pre-measurement mode, enter
the required duration in seconds in the Duration box.
Continuous (from Sample Addition) - ultrasound is acti-
vated after sample is added and will run continuously.
Continuous (from Measurement Start) - ultrasound is
active throughout the measurement. Ultrasound will start
after the electrical background is complete. The stabilis-
ing period will add a delay between the electrical back-
ground and the optical alignment - during this delay
ultrasound is active, allowing bubbles to be driven from
the dispersant.
To manually control the ultrasonic level from the Wet
Accessory dialogue, either select and drag the Ultrasound
slider bar to set the level, or select the bar and use the
keyboard control up/down/left/right arrows; this will step
the ultrasonic level up or down in 10% divisions.
Alternatively, for an exact setting, double click on the dis-
played ultrasonic value and type in the required level.
Pulsed Enable pulsed ultrasonication - in this mode, ultrasoni-
Ultrasonics cation is applied in pulses (with fixed duration on and off
pulses). This option can be used in combination with both
continuous or timed ultrasonication mode. Specify the
duration of on and off pulses by entering the required val-
ues (seconds) into the corresponding boxes. This can pre-
vent overheating of the sample, and is also useful on
fragile samples that could otherwise be broken up by
more aggressive ultrasonication.
Note
The Ultrasonics will switch off automatically after 20 minutes continuous
use.
Field Description
Clean Each Clean type specifies a set number of cycles (noth-
Sequence ing further is changed in the cleaning method). Select a
Clean type of either Quick, Normal, Extensive or Cus-
tom. If Custom is selected the user must also manually
specify the number of Clean cycles. If measuring the
same sample type in succession a Quick clean may be
sufficient. If making the last measurement of the session,
it is prudent to perform an Extensive clean to ensure that
the system is as free as possible from contamination in
readiness for the next measurement session.
Ultrasonication Ultrasound can reduce agglomeration, which may help
further with cleaning. Select whether to Use ultrasound
during the clean sequence or No ultrasound during
clean.
Tip:
If you are unsure which analysis settings are best suited to your application,
try reapplying different analysis settings to the measurement records. To
do this, use the Edit Result feature after the first set of measurements has
been made see Chapter 5 for more information.
Field Description
Analysis Model General Purpose this model is appropriate for most
samples, including naturally occurring and milled materi-
als. You should select this model unless you know that
your sample meets more specific requirements stated for
the other analysis modes provided for the system.
Field Description
Analysis Model Narrow Modes this model is appropriate for sample
(continued...) consisting of one or more narrow modes, where each
mode spans much less than a decade in size. It is not
intended for use with broad distributions which exhibit
more than one peak.
Verification Latex this model is designed to enable the
analysis of one or more very narrow Latex size standards,
such as those used during verification of the optical unit
performance.
For all of the analysis models use the Keep a single
result mode option in the Advanced settings to force the
analysis to only report the most significant mode present
within the sample.
Fine Powder This option should be enabled for samples containing a
Mode significant proportion of material below 10 microns in size.
(Aero units only) It can help improve the result reproducibility, and can be
used with any of the Mastersizer 3000 analysis modes.
Field Description
Advanced: Use Keep a single result mode if you know your sample
Analysis Modes has a single mode size distribution. This option eliminates
any small modes produced by noise on the measurement.
Advanced: Set the number of low scatter-angle detectors to be
Disable Detec- ignored by the analysis. Thermal fluctuations in the dis-
tors persant can cause signals in the inner channels to be
interpreted as large particles. Set the number of detectors
to remove the effect of these fluctuations, at the expense
of reducing the sensitivity to the presence of genuinely
large particles.
Advanced: Remove Blue Light from Analysis allows a measure-
Red Light Only ment made with red and blue light to be analysed as if the
blue light was not used.
A practical use of size bands is when using sieves that conform to specific standard
sizes - size bands could be set to corresponding ranges.
The system can either automatically set size bands for histograms and charts, or
these can be user-specified.
To automatically select appropriate high and low sizes based on the measure-
ment result, click Use default sizes.
To manually define size bandings, click Use the sizes that I specify. The
User sizes (m) and Generate user sizes panels become active.
Creating size bands:
1. From the Generate user sizes panel, enter the required Lower size and
Upper size limits (in m).
2. Then enter the Number of bands into which the table or histogram should
be divided.
3. Select a Distribution type:
Linear: Divides the bands into equal sizes.
Logarithmic: Bands are represented by multiples of a constant factor that
accounts for the size range specified.
4. Click Generate sizes to populate the set of figures in the User sizes panel.
5. To add more size bands, enter a figure into the Size (m) field and then click
the button. The new figure is added to the list in the correct numerical posi-
tion.
6. To delete one of the size bands, select the figure you wish to remove and then
click .
7. To save size band information that you wish to re-use within other SOPs, click
to save the current sizes. To use previously saved sizes, click and then
locate the relevant .siz file.
Viewing
measurement results
Introduction
The Mastersizer 3000 delivers powerful features for manipulating and presenting
the data gathered in measurement files.
This chapter provides the following information:
An overview on working with measurement files.
Features of the Analysis report.
Fundamental concepts to help understand measurement data.
Note
Although Mastersizer 2000 files can be opened for analysis, they cannot be
edited. They are shown with a differently coloured icon and also a lock,
which indicates their read-only status (like the Examples file shown here):
Tip
To view more information about an open measurement file, move the
mouse pointer over its tab - information about the file type, access mode
and location are provided.
2. Click the down arrow to display a list of available parameter values upon which
to filter (these are the values shown in the record).
3. Select any of the values to show only those records with values matching the
checked filters. When a filter has been set, the column header shows the fol-
lowing icon: . As many filters as required can be set up in this way.
4. It is also possible to extend the functionality of this grid by manually typing text
into the sorting header area - filtering criteria can include numerical informa-
tion, date ranges and strings. Click on the column header and then enter text -
the display updates immediately to show only those items that are within the
scope of the filter - see Parameter filtering logic in the Help system.
In the example shown here a filter of >50 has been applied (meaning, show
only those items whose Dv50(m) value is greater than 50).
To remove a filter that has been selected from the predefined list, either de-select
the individual check box from the filter menu or choose the Clear filter option -
this removes all filters for that column. For any filter text that has been manually
entered, either delete the text or click the X button which is shown next to the text
(in the adjacent column).
Note
Check box filters: having all filters either selected or deselected has the
same effect, namely that all of the records are shown.
Grouping records
Read this section for details on how to group records by parameter type.
The Record view allows records with the same parameter values to be grouped
into collapsible sections. Any parameter can be selected on which to group. It is
then possible to add further sub groups. Examples of the use of groups include,
grouping by Operator Name, Measurement Date, Accessory Type etc.
1. To group items by a parameter, drag the required column header into the area
above the columns, titled Drag a column header here... The records are
now presented in collapsible groups based on groupings of the chosen parame-
ter of the same value. For example if Scattering Model was used as the group-
The grouping hierarchy can be edited by dragging and dropping the group
parameter headers left or right as appropriate.
4. To remove groupings, drag the group parameter header back into the records
area.
Merging records
This section provides information on how to use the All files feature of the
Record View.
When working with multiple Mastersizer measurement files, it can be useful to
compare data by merging all of the files into a single (virtual) measurement file so
that Mastersizer's powerful sorting, grouping and reporting features can be
exploited.
1. Open more than one measurement file.
2. Click the All files tab (rightmost):
The records contained in all of the open measurement files are shown combined
into a single file. It is now possible to sort, group, filter and report on the records
displayed in this view. If graphically analysing multiple records in reports, each
record's data is overlaid using a different colour.
Note:
This feature is not available from the All files view.
2. Enter a Sample name and a Reason for creating the new result.
3. Then click the Create Average button.
A new record is created within the current measurement file. To retain the
newly created record, save the measurement file.
Editing records
Each Mastersizer record is the result of the processing of raw measurement data.
By editing a record it is possible to re-apply new values to some of the parameters
used in the original measurement settings in order to generate a completely new
record.
This can be used to explore the effect of using different settings from those chosen
when the measurement was made. Typically adjustments are made to improve the
fit of the calculated result to the data. Use the SOP summary report (see Chap-
ter 9) to show which parameters have been altered.
To edit a record:
1. From the Record View, right click on a record, or group of records, and
choose Edit Result.
The Result Editor window is displayed. This contains all of the information
set up in the original measurement. Some of the options are unavailable
because they relate only to the physical set up of the original measurement, for
example, all of the Measurement Duration options.
2. Edit any available sections of the Result Editor as required. Any edited infor-
mation is clearly highlighted by a different background colour.
3. Click OK when all settings have been altered to your specification. The newly-
set values are now re-applied to the existing raw measurement data. A brief
message is displayed whilst this happens. When the result has been edited, the
edited result is then displayed as the first item in a tree structure.
4. Expand the edited record to show the "contained" original result by clicking the
icon. Note: It is not possible to edit a record without also retaining its origi-
nal settings. Further edits to the record result in more items contained within -
the most recently edited record, which is always shown as the container record
(in this example, edit 3 was the final edit made):
Note:
If you save the measurement file, the new settings will overwrite the origi-
nal file settings. To create a result-edited version of a file and still retain the
original file settings, just save the file with a different name before editing.
Analysis report
The Analysis report provides a useful breakdown of the
measurement data.
To display this report, select the record(s) to analyse in the Record View and
then click the Analysis report from within the Reports tab.
7
1
6 5
9 8
10
3 2
The statistics of the distribution are calculated from the results using the
derived diameters D[m,n] an internationally agreed method of defining the
mean and other moments of particle size. See British standard BS2955:1993 for
more details.
1
m 3 ------------
-
Vi di
D m n = ---------------------------
mn
n3
Vi di
Ls is the light intensity measured in the central detector when a sample is pre-
sent in the cell, Lb is the same but with clean dispersant (i.e. with no sample).
Obscuration is usually expressed as a percentage: 100 x Ob.
An ideal range is between 3 and 20%, depending on the sample and dispersion
unit used.
Distribution shows the type of distribution the analysis has used. Options
include change from volume to surface area, length or number. Remember
that the Mastersizer measurement is fundamentally a measurement of the vol-
ume distribution - transforming the result into a surface, length or number dis-
tribution is a mathematical process that may amplify any error in the original
result, especially at the fine end of the size distribution.
Uniformity a measure of the absolute deviation from the median.
X i d x 0.5 d i
Uniformity = ------------------------------------------------
d x 0.5 X i
Here d(x,0.5) is the median size of the distribution (where x is replaced by v, s,
l or n) and di and xi are respectively the mean diameter o f, and result in, size
class i.
Specific Surface Area (SSA) the total area of the particles divided by the
total weight.
V
6 -----i
d 6
SSA = ----------------i = ---------------------
pD 3 2
p Vi
where Vi is the relative volume in class i with mean class diameter of di and p is
the particle density.
If the SSA is used, it is important that the density of the material is defined (in
the SOPs Material settings). This figure is a mathematical calculation based
on the assumption that the particles are both spherical and non-porous.
Fundamental concepts
To understand the meaning of the results from the Mastersizer, a number of fun-
damental concepts require explanation. These are:
How the concentration is calculated.
The results are volume-based.
The result is expressed in terms of equivalent spheres.
How the distribution parameters are derived.
---I- = e
b
I0
Where:
I is the intensity of light at a distance b in the particle field of absorbance .
I0 is the intensity of the light beam as it enters the particle field.
I / I0 is the relative transmission T of the beam (measured directly by the
instrument). I0 is the intensity of the laser beam measured at the receiver when
no sample is present and I is the intensity with sample in the beam.
Expressing the Beer-Lambert law in terms of relative transmission and re-arranging
to solve for absorbance gives:
1
= ------ Ln T (1)
b
Volume concentration
The term contains information about the concentration and size of the particles.
From scattering theory the light attenuated by a particle i may be described by:
2
a i = Q i r i n i
Where:
Qi is the efficiency of light extinction (by scattering and absorption), calculated
from Mie theory for a particle of radius ri.
The second term is the cross-sectional area of the particle and the final term ni
is the number of particles of radius ri.
4 3
In terms of the volume of particles V i = --- r i n i , the equation above becomes, for
3
an ensemble of particles:
3 Qi Vi
i = --- ----------
-
4 ri
The size of the particles is expressed by the diameter d and the volume terms can be
separated into a relative volume distribution v and a total concentration Cv (total
volume of particles in a unit volume of dispersant). The equation then becomes:
3 Qi Vi
i = --- C v ----------
-
2 di
Substituting the above into equation (1) and solving for the concentration gives:
2Ln T -
C v = -----------------------
Qi vi
3b --------- -
di
2000Ln T
C v ppm = ------------------------------
Qi vi
3b --------- -
di
This equation provides the concentration in parts per million (ppm). To calculate
the value as a percentage volume concentration, the final value is divided by 10,000.
In the above equation:
The Transmission (T) is a value between 0 and 1 and is measured directly by
the instrument.
The particle size distribution vi is the relative volume in size-class i with mean
diameter di.
Qi - the mean extinction term for size-class i - is calculated from scattering the-
ory and is a function of the optical properties of the particle and dispersant
media.
Volume-based results
The first, and probably most important, point to remember in interpreting results
is that the fundamental size distribution derived by this technique is volume-based.
This means that when a result indicates, for example, that 11% of the distribution is
in the size category 6.97-7.75 microns, this means that the volume of all particles
with diameters in this range represents 11% of the total volume of all particles.
It is useful to consider a numerical example to illustrate this point. Suppose, for
simplicity, that a sample consists of only two sizes of particle, 50% by number hav-
ing a diameter of 1 micron and 50% by number a diameter of 10 microns. Assum-
ing that the particles are spherical, the volume of each of the larger particles is 1000
times the volume of one of the smaller ones. Thus, as a volume distribution, the
larger particles represent 99.9% of the total volume.
The graph below illustrates this for a more realistic distribution:
%
30
20
1 1
10
0
0.1 1.0 10.0 100.0 1000.0 10000.0
Particle Diameter (m.)
ill 1874
Equivalent spheres
Mie theory presumes that the particles measured are perfect spheres. In practice
they are very rarely so. This causes a problem in the definition of the term meas-
ure the particles size: if the particle is an irregular shape, which particular dimen-
sion should be measured?
As an example, imagine that I give somebody a matchbox and a ruler and ask them
to tell me the size of it. They may reply by saying that the matchbox is 50mm x
25mm x 10mm. They cannot say that the matchbox is 25mm as this is only one
aspect of its size. It is not possible to describe the three dimensional matchbox with
one unique dimension. Obviously the situation is even more complex for irregular
shaped particles such as grains of sand or the pigment particles in paint.
Most people want a single measurement to describe their sample, for example, they
wish to say that their sample is made up of 50 micron particles. What is required is
a unique number that describes the particle. There is only one shape that can be
described by one unique number and that is a sphere. If we say we have a sphere of
50 microns, this describes it exactly. We cannot do the same even for a cube as 50
microns can refer to its edge or to a diagonal.
One way to get a single unique number to describe an irregular shaped particle is to
compare some feature of the actual particle to an imaginary spherical particle.
Some typical methods of doing this are:
Equivalent surface area calculating the diameter of a theoretical sphere
that has the same surface area of the original particle.
Equivalent maximum length this is where the diameter of a theoretical
sphere is the same as the maximum dimension of the original particle.
Equivalent minimum length this is where the diameter of a theoretical
sphere is the same as the minimum dimension of the original particle.
There are many other methods available to do this. This technique is known as
equivalent spheres.
The Mastersizer uses the volume of the particle to measure its size. In the example
above, the matchbox has a volume of 50 x 25 x 10mm = 12500mm3. If the Master-
sizer was able to measure this size of particle it would take this volume and calcu-
late the diameter of an imaginary sphere that is equivalent in volume - in this case it
will be a sphere of about 30mm diameter.
Obviously the answer will be different if the surface area or maximum dimension
of the matchbox is used to calculate an equivalent sphere. All of these answers are
correct but each is measuring a different aspect of the matchbox. We can therefore
only seriously compare measurements that have been measured using the same
technique.
ill 7973
The sphere of equivalent volume would have a diameter centred on:
6V
3 ------- = 33m
with a spread from 20 to 60m.
It is interesting to compare this with other techniques. Sieving would pass the par-
ticles through a 20m aperture and classify them as 20m. Sedimentation would
give a result related to the total surface area, in this case reporting a diameter of
around 40m.
d i l di
2
X d d
i i 2
= ---------------------------------- = D [K + 2,K] + D [K + 1,K
X i
skewness
X d d D K + 3 K d d + 3
3
i i 3 2 2
kew = ---------------------------------- = -------------------------------------------------------------------
3
X
3
i
kurtosis
X d d
4
i i 4 3 3 2
D K + 4 K d 4D K + 3 K + 3d + 6d
rt = ---------------------------------- 3 = -------------------------------------------------------------------------------------------------------------------------
4
X
4
i
Distribution K Xi
Volume 3 Vi
Surface 2 Vi /di
Length 1 Vi /di 2
Number 0 Vi /di 3
Optical models
This section contrasts the Fraunhofer approximation used in some instruments
with the Mie theory which underpins Mastersizer 3000 operation. Although Mie is
the preferred scattering model, users can also select Fraunhofer if required from
the Data Processing - Analysis measurement settings.
Fraunhofer approximation
Older instruments and some existing instruments rely on the Fraunhofer approxi-
mation only. This assumes that:
The particle is much larger than the wavelength of light employed. ISO 13320
defines this as being greater than 40x wavelength (25m when a He-Ne laser is
used).
All sizes of particle scatter with equal efficiencies.
The particle is opaque and transmits no light.
These assumptions are incorrect for many materials and for small particles they can
give rise to errors approaching 30%, especially when the relative refractive index of
the material and medium is close to unity, or when the particles are transparent.
When the particle size approaches the wavelength of light the scattering efficiency
becomes a complex function with maxima and minima present.
Mie theory
The Mastersizer 3000 uses the full Mie theory which completely solves the equa-
tions for interaction of light with matter. This allows completely accurate results
over a large size range.
Density the density in g/cm3. This is used to calculate the Specific Surface
Area (SSA). Users who want to include the SSA as a derived parameter must
complete this field.
The residual shown in the Parameters report indicates how well the calculated
data fitted the measurement data.
Sample preparation
guidance
Introduction
Preparation of the sample before it is added to the system can be critical. Over half
the problems encountered when measuring a sample are caused by poor sample
preparation. If the sample is sticking together, dissolving or floating on the surface,
or if it is not a representative sample, the result will be incorrect.
There are many techniques available to ensure that the sample is prepared success-
fully. Once a suitable dispersion technique is found for the sample, standardise the
procedure (using an SOP) so that comparisons can be made between samples.
This chapter gives information on:
Sample preparation flow chart summarises the process.
Representative sampling.
Considerations for dry samples.
Considerations for wet samples.
Symptoms of poor sample preparation.
Yes Does it No
disperse in
water?
Ultrasound if
necessary Yes No
Does it
float?
Analyse
Try Surfactant
No Does Yes
Ultrasound
work?
Yes No Analyse
Does this
disperse it?
Ultrasound if
necessary
Analyse
ill 7825
Using liquid-borne particle suspensions, such as latex and emulsions, is usually
straight forward. Use the suspension phase as the dispersant, mix the sample well
and add directly to the Mastersizer beaker. Ultrasound is not usually necessary and
will almost certainly affect the size of liquid particles.
Representative sampling
When taking a sample for a measurement, it is vital to ensure that the sample used
is representative of the whole sample. This section gives guidelines.
If the sample is taken from a bottle or container, take care to ensure that it is thor-
oughly mixed. When the sample is a powder, large particles tend to rise to the top
of the container as smaller ones work their way to the bottom.
In most samples there are some large and some small particles, but the majority fall
between these extremes. If a sample is taken from the top of the container, mainly
large particles will be measured. If this is compared to a measurement with the
sample taken from the centre of the container the results will differ noticeably.
If the sample is stored in a container this should be mixed thoroughly before meas-
urement. Do not shake the container as this often increases the separation of the
particles. Instead, hold the container in both hands and gently roll the container,
continually changing its orientation for about 20 seconds. This method works bet-
ter if the container is only half full.
Spinning rifflers
If the distribution of particles within a sample is particularly broad, representative
sampling can be difficult. If problems continue, using a spinning riffler may help.
This is a vibrating hopper which vibrates the sample down a chute.
The act of vibrating the sample causes the larger particles to separate out and travel
down the chute first. At the end of the chute a set of rotating pans collects the sam-
ple evenly. When all of the sample has passed down the chute each collecting pan
will contain a representative sample.
Liquid samples
Liquid samples can also separate out if stored in containers, with larger particles
sinking to the bottom. Again, its necessary to mix the sample thoroughly to get a
representative sample. Sample splitters/rifflers are also available for liquid samples.
Beware of using a magnetic stirrer to mix a liquid sample. Large particles tend to
move to the outside of the container due to centrifugal separation, and this can lead
to sample biasing.
Note
Application Notes describing how to develop a method for dry sample
analysis are available at www.malvern.com.
Warning!
Do not warm a dispersant to allow re-gassing if the dispersant is volatile.
Never allow dispersants to reach their boiling points.
Warning!
If a volatile solvent has to be used as dispersant, leave the sample area door
open during use to aid ventilation.
The cost of some of the organic dispersants may limit their use to the Hydro MV,
which typically only uses 120ml of dispersant. The problem of the safe disposal of
the sample after measurement must also be considered. Always adopt the correct
procedures for disposing of the sample and dispersant, following any local guide-
lines. Most local regulations forbid hazardous samples and dispersants to be tipped
down the drain, allowing them to enter the water system.
Note
Application Notes describing how to develop a method for wet sample
analysis are available at www.malvern.com.
Surfactants
Adding a surfactant may help in preparation of the sample it will lower the surface
tension, which promotes wetting of the particles.
Surfactants have to be added in minute quantities, typically one drop per litre of
dispersant. If too much surfactant is added the action of stirring and pumping the
sample may cause it to froth, causing bubbles to be introduced into the system.
Bubbles are seen by the system as particles which can bias the measurement. Anti-
foaming agents may be added to prevent the formation of bubbles but, owing to the
fact that these may contain particulates, they should be added to the dispersant
before the background is measured.
Try adding a drop of surfactant to a quantity of sample and dispersant mixed in a
small beaker. If the sample sinks to the bottom of the beaker in large clumps, dis-
pose of the sample and start again. This time add the sample to a dry beaker, add a
drop of surfactant and mix thoroughly to a thick paste. Add the dispersant and mix
well. This usually avoids the agglomeration caused by adding the dispersant first.
A list of recommended surfactants in order of common use is given below:
Surfactant Nature
Igepal Non-ionic
Teepol L Non-ionic
Synperonic N Non-ionic
Aerosol OT Anionic (solid)
Sodium dodecyl sulphate Anionic
Hyamine 2389 Cationic
Admixtures
Admixtures work by causing the particle surface to become charged, causing parti-
cles to repel each other (i.e. providing a zeta-potential to aid dispersion). Admix-
tures are added in larger quantities, typically 1g/l. A list of commonly used
admixtures is given below:
Sodium Hexametaphosphate (e.g. Calgon)
Sodium Pyrophosphate
Trisodium Phosphate
Ammonia
Sodium Oxalate
Calcium Chloride
As many of these are solid materials that are dissolved into the dispersant, the solu-
tion should be filtered after preparation to remove impurities.
Slurries
The act of mixing up a small quantity of concentrated sample, dispersant and addi-
tives before it is added to the dispersion unit tank is known as preparing a slurry.
Once the particles have been successfully dispersed into a slurry, the sample may be
added to the dispersion unit without any further additions of surfactants etc. The
problem of the sample settling out within the beaker can be solved by using a
pipette to continually stir the sample. While stirring, continually fill and empty the
pipette. Use the pipette to add the sample to the dispersion unit tank.
Use of ultrasonics
In addition to the processes described above, ultrasonics can be applied to help the
dispersion, whether or not it contains a surfactant.
When mixing the sample in the suspension medium, a visual inspection often
shows whether ultrasonics are necessary. If large agglomerates of particles sink to
the bottom of the beaker, try applying two minutes of ultrasonics by placing the
slurry and its beaker in an ultrasonic bath. It will be immediately apparent if this has
been effective. Further ultrasonics can be applied when the sample is added to the
tank, if necessary. This often prevents re-agglomeration, but is not always needed.
Note
Be wary of using ultrasonics with fragile particles; the ultrasonic action
may break up the particles themselves. If in any doubt, microscopic obser-
vation before and after the ultrasonics can establish whether it has helped.
The obscuration and its behaviour during the dispersion of the sample can also
warn of other potential problems.
If the obscuration decreases the size of the particles within the sample may be
increasing; either the sample is sticking together or the particles are actually swell-
ing due to the dispersant. Other causes could be the larger particles settling out due
to inadequate pumping and stirring or even the particles dissolving.
If obscuration increases rapidly, particles may be attaching themselves to the cell
windows due to surface charges. This means material is in the laser beam continu-
ously and the obscuration appears to increase. To solve this, use an admixture.
Bubbles
The Mastersizer sees bubbles as particles and measures them. Bubbles vary in size
but are typically around 100 microns in size. In many cases these bubbles can be
seen clearly as a second and separate peak when the measurement data is analysed.
Always be wary of bubbles in the system.
When the dispersant has been added to the dispersion unit and circulated, we
advise stopping the unit briefly (using the Stop all operations button in the
Active Accessory control) to allow trapped air to escape the system. Also, during
installation, ensure there are no twists or loops in the connected sample tubing.
When adjusting the pump and stirrer speed for a particular sample, ensure the
speed is not so fast that it introduces air into the system.
When a surfactant is added to the sample, excessive speed of the stirrer/pump may
cause frothing. This will force bubbles into the system.
Degassing
If dispersant is stored under pressure or at low temperature, consider de-gassing
before use. The pressure release or temperature rise reduces the solubility of gases,
resulting in bubble formation in the pipes and tank etc. This is a problem with
some mains water supplies. The best answer is to store enough dispersant at room
temperature and pressure for several hours before use to allow de-gassing to occur.
For more information see Choice and preparation of the dispersant above.
Pulses of ultrasonics can help to degas the dispersant after the tank has been cleaned
and filled with fresh dispersant. To do this, use the Active accessory control
panel (Tools-Accessories) and then select Enable pulsed sonication.
Note
Ensure that all dispersants are degassed before being added to the system.
Reporting
Introduction
Reports help to make sense of all the data that is generated in a Mastersizer meas-
urement. They enable you to graphically analyse measurement data and also isolate
areas that are of interest using tables and graphs.
This chapter provides the following information:
About reports.
Viewing and using reports.
Creating and editing reports.
About reports
Mastersizer's reports consist of information sections that are displayed together in a
single report for easy on-screen analysis or printing. Reports can be configured so
that only those items that are of interest to the user are displayed. It is also possible
to alter the order in which the report elements are displayed.
Note
It is only possible to edit user-defined reports, not those supplied with the
system. It is possible however to copy the system reports and create new
reports that are based on them.
Malvern reports
In addition to the ability for users to build reports tailored to their own require-
ments, Malvern Instruments supplies a number of standard reports with the soft-
ware. These are:
Analysis a report that provides a mixture of key parameter sections and a particle
size distribution graph. See Chapter 5 for more information on this report.
Data this report mainly focuses on the raw light scattering data gathered by the
instrument during the measurement. It includes a light scattering graph.
Parameters a summary of the key parameters measured by the system.
Note
Although the Malvern Reports cannot be modified directly, it is possible to
open them for editing and then use a different Report Title to effectively
save a copy.
Displaying reports
1. Select a record or number of records from the Record View.
2. If it is not already shown in the current view, select the Reports tab. One of
the reports for the current workspace is then shown (by default this is the Anal-
ysis report).
It can be easier to work in 2-pane Vertical Split view whilst initially selecting
the records. Any selections or de-selections of records are then updated imme-
diately in the visible report.
Graphs
Three types of graph are used within the reports: result, trend and data graphs.
Whenever a graph is displayed, click on the plot lines to display data points. Move
the mouse pointer over the data lines to see numerical information relating to that
data point.
3. This window contains a list of Available Reports on the left and Selected
Reports on the right. Reports shown with are reports that were created by
Malvern Instruments.
4. Move reports from the Available Reports column to the Selected Reports
column by selecting them and then clicking the Add button. Similarly, move
Selected Reports back to the Available Reports list by selecting them and
then clicking the Remove button.
5. To move a report up or down within the Selected list, select it and then click
the up or down arrow icons as required. The order of this list corre-
sponds to the left/right position of the Reports tabs, the leftmost tab being the
top report in the Selected Reports list.
6. Click OK when you have organised the reports as required.
Tip
This feature is also available in the live Trend View, Light Scattering
and Size Distribution panels in the measurement window.
1. From the report editing mode, click the properties button next to the graph
you wish to alter. The Trend Chart Properties window is displayed.
2. Select either the X or Y Axis button on the left. The data currently plotted is
shown in the top panel of the window.
3. To remove the currently plotted data, click next to the parameter descrip-
tion.
4. To add or select a new data item against which to plot, click the button.
A new <Empty Axis> element is added into the top panel.
5. Select the new <Empty Axis> element to display its properties in the lower
panel.
Note
Parameters with numeric values (such as percentiles) are editable and indi-
cated with . To edit the parameter numeric value, click the parameter
and amend the value. Any value entered that is inappropriate for the
parameter is flagged with an exclamation mark icon - amend the value
accordingly if this is the case. It is possible to enter multiple variants of the
same parameter with different values associated (for example Dv10, Dv50
and Dv90). These are then indicated on the graph with different colours.
Printing reports
Before sending a report to the printer, it can be helpful to view a preview initially.
The print preview option shows how a report will appear when printed on paper,
enabling the effect of alterations to a report to be easily established without wasting
paper.
1. Choose Print-Report Print Preview from the ribbon. A preview of the cur-
rently selected report is displayed:
Zoom to zoom in and out of the print preview, click the icons
respectively. When zoomed in it may be necessary to use the scroll bars in
order to see all of the elements on the report.
100% shows the report at 100% zoom.
Page width stretches the page to view the width of the page within the cur-
rent window.
Whole page zooms out to so that an entire report page can be viewed in the
window.
Two pages zooms out so that two pages of the report can be seen side by
side.
Note:
Use of these options does not have any impact on the final report printout.
Note:
Actual data from the selected measurements is still shown whilst in edit
mode. This greatly simplifies the process of visualising the final output.
2. Enter an appropriate name for the report in the top field (New Report is shown
by default). The report name is used to identify the report in the Reports tabs
and can be modified later if required.
The new section is now added to the report in the main report editing window.
In this example, a parameter grid has been added:
6. To undo the last change made to the report, click the Undo button . If
required, continue to click Undo until all edits in that session have been
undone. To re-do a change that has been undone, click the Redo button .
7. Click anywhere in the report section's heading area to activate the cursor, then
edit the heading for the report section as required.
Tip:
To display a thumbnail preview of the complete report move the mouse
pointer over the icon.
8. To move items up or down in the report, click the arrow buttons on the
right of the item. Alternatively, click and hold the icon and then drag
the element up or down to the required position.
Tip:
To move items around a large report it is often easier to zoom out initially
(so that more of the report is visible) and then drag the report section to
the required area of the page. The mouse pointer must not be positioned
over any of the buttons when clicking and dragging.
9. When you have finished editing the report, click the Finish Editing button at
the top of the window.
10. The software asks whether you would like to save (changes to) the report -
click Save, Don't Save or Cancel as required.
System security
About security
By default the Mastersizer software uses Windows user security. That is, only a
user logged into the computer that is physically connected to the instrument can
access the system. For many environments this level of security is sufficient.
For quality controlled environments however, the optional Malvern Access Con-
figurator (or MAC) adds a fully-featured security system to the Mastersizer soft-
ware. The MAC not only provides control over an individual users access to the
various features of the software, but also logs an audit trail of significant system
events, who performed them and when.
Activating security
Note:
This process should only be undertaken after setting up security privileges
for Mastersizer 3000 in the MAC.
3. Once selected, the user is prompted to enter a user name and password:
The user who logs in must be a user with both administrative rights on the
computer and also one who has been set up with Disable/Enable Security
permission in the MAC.
4. Click the Validate credentials button.
5. Select Browse and then locate the security definitions file (permissions.xml).
This file can be located wherever required but, by default, it is located in:
C:\Program Files\Malvern Instruments\Mastersizer 3000
6. Click OK to finalise the process.
Following this, only users with valid permissions that have been set up in the MAC
will be able to access the functions of the system.
Software licensing
Without a valid licence, the Mastersizer software will not function. The software
automatically generates a licence on the computer that is connected to the instru-
ment. After first installation of the software you will be prompted to accept this
license.
Click Accept licence to accept the license file. If the software was set up by
your administrator or Malvern Instruments, you may not see this window -
this means that the software is already activated.
Note:
The licence file contains the identification of the user and the instrument
from which it was generated. Do not share the licence outside of your
authorised user group.
4. Click Yes and then specify a file name and location in which to save the
licence. Then just send a copy of this .licence file to the Mastersizer 3000 soft-
ware user without a licence. After first attempt to run the software, the user is
prompted with the following:
5. Click Install licence and then browse to the location of the .license file.
6. The credentials of the licence file are then displayed - click Install licence.
Audit history
Note:
Auditing is only available when 21CFR Part 11 mode is enabled. This
requires a special feature key, which can be purchased from Malvern
Instruments.
Enabling auditing
After you have purchased a feature key from Malvern Instruments:
3. Click on the column headers to sort the data or use the methods described for
filtering the record view (see Filtering the Record View in Chapter 5).
The log shows:
Time stamp the time in the following format:
YYYY-MM-DDTHH:MM:SS (e.g. 2011-05-05T12:55:35).
Action type a self explanatory name for the action performed, for example
RecordCreated, Record Edited, RecordDeleted, Startup etc.
User ID the unique ID of the individual user logged onto the system.
Computer ID the network name of the computer that accessed the system.
Tip
Right click a record from the Record View and choose View Record
Audit History to view a history of the selected record only.
For edited records, the following properties are shown in the expanded tree view:
Property the record property (for example, Sample Identifiers) that was
altered.
Old Value the previous value recorded for the property.
New value the new value recorded for the property.
Reason for change this is for averaged records only where a Reason was
specified by the user at the time the averaged record was created.
Modified file name the (.rmd) file name of the record that was modified.
Modified record the record number within the measurement file that was
modified.
As the audit history records all significant system events, the log can grow quite
quickly, making it more difficult for a user to visually analyse the data. The
Archive schedule feature allows you to set the frequency for archiving for the
audit trail. The archived data is still available, but is stored separately to the active
archive. You can view archived data by selecting System (including archived
entries).
4. Click [Archive] Schedule from the Audit ribbon.
Advanced features
Introduction
Advanced users can exploit additional features of the Mastersizer software that are
not normally required as part of the basic measurement process. These features
enable a supervisor to set up options that may help operators use the system in a
more standardised way.
This chapter focuses on the following areas:
Doing more with SOPs, including how to set up SOP templates and analyse
the development history of an SOP.
Editing records that have already been made.
Exporting data and working with export templates.
Maintenance utilities.
Alternatively, save the SOP as an SOP Template by choosing Save as tem-
plate. Refer to Creating and Editing an SOP in Chapter 4 for more informa-
tion.
SOP Templates
SOP templates allow you to create SOPs based on predefined settings, improving
both the speed of creation and consistency. Effectively SOP templates are just
SOPs with additional, optional, metadata attached to them; a Name, Description,
Category and image. It is a good idea to create a template whenever the core set-
tings of an SOP are to be used repeatedly.
The process of creating an SOP template is very similar to creating a standard SOP.
You can either create an SOP from scratch and then save it as a template, or open a
previously created SOP/template, make amendments to it and then save it. SOP
templates can be edited, saved as new templates or just saved as standard SOPs.
Note:
SOP Templates are stored as .msot files in the Windows file structure. If
required, use the Workspace shortcut from the Workspace Viewer panel to
locate the SOP Template folder (the templates are organised in folders
based on any categories that users have assigned to the templates). If any
associated image is removed from the SOP Template folder, a default
image is used instead.
4. In the Save SOP Template window, give the new SOP template a: Name (a
descriptive name for the SOP template), a Description and, if required, assign
a Category (previously added categories are available from the drop down
menu). Categories could be of assistance when your environment has multiple
SOP templates.
5. To save over an existing SOP template (i.e. to edit an existing template), select
Allow new template to override any existing template from the Save
SOP Template window. Also, ensure that the new template uses the same
name as the existing template.
Tip:
To view a set of pre-selected SOP templates relevant only to the currently
attached dispersion unit, choose File-New from the Manual Measure-
ment settings window.
1. In the records view, select the record(s) for which you want to review the SOP
summary and then click View SOP summary on the Home ribbon. Alterna-
tively, from the SOP Editor, choose View-Summary Report.
The SOP Summary report window is displayed. This lists all settings applied
by the SOP at the time of measurement. If more than one record is selected
from the records view, each record is shown in its own column.
2. To Save, Copy (to clipboard) or Print the information shown, click the
appropriate option from the SOP Summary Report window toolbar. The date
that the SOP summary was created is appended to all three options.
3. Select the following options from the SOP Summary Report window tool-
bar as required:
Show all - this is the default view and shows all SOP settings from all
items.
Highlight differences - shows all items (as the default view), but high-
lights using different colours, areas where different values or SOP settings
were applied.
Show only differences - shows all items, but only those SOP settings
where differences exist.
SOP History
Each time an SOP is edited and re-saved, its history is recorded for future audit
purposes. The SOP History feature exposes any differences between versions of
the SOP.
This feature only applies to SOPs that have been revised and re-saved. If the Save
As option is used the SOP's history is not maintained.
1. Select View SOP History from the Home ribbon and then choose the
required file. The SOP Version History window is displayed. This lists all
parameters as applied by the SOP. Each version of the SOP is shown in its own
column.
2. To Save, Copy (to clipboard) or Print the information shown, click the
appropriate option from the SOP Version History window toolbar. The date
that the SOP history summary was created is appended to all three options.
3. Select the following options from the SOP Version History window toolbar as
required:
Show all - this is the default view and shows all SOP settings from all
items.
Highlight differences - shows all items (as the default view), but high-
lights using different colours, areas where different values or SOP settings
were applied.
Show only differences - shows all items, but only those SOP settings
where differences exist.
Compare to - becomes active when either Highlight Differences or Show
only differences is selected. This option then allows you select a 'source'
item against which to compare the others - any field in another item that
differs from the source is highlighted or isolated (as selected).
Editing results
This section provides help on the Edit Result feature of the Records view.
Each Mastersizer record is the result of the processing of raw measurement data.
By editing a record it is possible to re-apply new values to some of the parameters
used in the original measurement settings in order to generate a completely new
record.
1. From the Record View, right click on a record, or group of records, and choose
Edit Result. The Result Editor window is displayed.
This contains all of the information set up in the original measurement. Some
of the options are unavailable because they relate only to the physical set up of
the original measurement, for example, all of the Measurement Duration
options.
2. Edit any available sections of the Result Editor as required. Any edited infor-
mation is clearly highlighted by a different background colour (in the example
shown here the lower field has been edited).
3. Click OK when all settings have been altered to your specification. The newly-
set values are now re-applied to the existing raw measurement data. A brief
message is displayed whilst this happens. When the result has been edited, the
edited result is then displayed as the first item in a tree structure.
4. Expand the edited record to show the "contained" original result by clicking the
icon. It is not possible to edit a record without also retaining its original set-
tings. Further edits to the record result in more items contained within - the
most recently edited record is always shown as the container record (in this
example, edit 3 was the final edit made):
Note:
If you save the measurement file, the new settings will overwrite the origi-
nal settings. To create a result-edited version of a file and still retain the
original settings, first copy the file (using Windows) and then edit the copy.
Exporting data
Data recorded by the system can be exported for analysis in other software applica-
tions, for example Microsoft Excel.
It is possible to export either selected records or all records contained in a single
measurement. Additionally, by selecting the All Files tab when multiple measure-
ment files are open, a selection of, or all records from, multiple measurement files
can be exported.
The ability to create export templates further enhances the flexibility of the system,
allowing you to select only those parameters required for your analysis, as well as
providing control over other data formatting options.
2. Select an appropriate export template from the drop-down list - the preview
pane is updated accordingly.
3. Change the presentation of the data using the Ordering, Delimiter, Format-
ting and Header options (see Export templates for more information).
4. Click Export to file to save the data as a separate (.txt or .csv) file, or select
Export to clipboard to copy the data to the clipboard.
Tip:
Drag and drop record data from the Record View straight into Microsoft
Excel/Word. The parameter values copied are all of those currently shown
in the records view.
Export templates
Export templates provide a means of storing, and quickly accessing, commonly-
used sets of parameters and format settings for exporting data. You can build and
store any number of export templates.
Note
Each export template is associated only with the workspace in which it was
constructed. This means, for example, that a template defined in the Mal-
vern default workspace, will not subsequently be displayed in a custom
workspace.
2. Enter a Name and then a short Description that explains the purpose of the
template. For example, Name: "Quick size results", Description: "Size infor-
mation plus sample ID only". The list of parameters associated with the tem-
plate is displayed in the Parameters panel. Initially this consists of the whole
set of system parameters.
3. Click the Change selection and order button to show the Parameter
selection window, then select the required parameters as follows:
Locate parameters by manually expanding the tree structure in the Availa-
ble list, or search for them by typing their first few characters into the
Search... field.
Select a parameter in the Available list and then click the Add button to
transfer it to the Selected list. Similarly, select an item from the Selected
list and click the Remove button to transfer it to the Available list. It is
also possible to select multiple items using the standard Windows selection
methods (i.e. SHIFT and click, CTRL and click).
When all of the required parameters are shown in the Selected list, choose
an order of presentation so that they are shown in the desired left-to-right
order. Select a parameter and then click the up/down buttons to move its
position in the list. Items at the top of the list are shown on the left, items
at the bottom are shown on the right.
Click Apply to apply the settings without closing the window, or click OK
to apply the settings and also close the Parameter Selection window.
4. Choose either rows or columns from the Ordering selector to present the data
horizontally or vertically respectively.
5. From the Delimiter selection, choose either Comma separated (csv) or
Tab separated. Your choice will depend on the requirements of the target
application.
6. Use the Formatting option to select either Format values as displayed in
software, which displays numeric information to a limited number of decimal
places (this may be preferable for presentation purposes); or choose Unfor-
matted values to show all data as recorded by the system.
7. Select Header on/off to specify whether the export should contain the names
of the fields as a header row (or column).
8. Click Save when complete.
Setup reference
Introduction
This chapter provides the following information:
Software installation reference.
Connection reference.
Testing the system.
Making connections
This section explains how to reconnect the system if it is moved after it has been
commissioned. Systems must be commissioned initially by a Malvern Instruments
representative. Commissioning is the process of ensuring a that a safe physical
installation of the Mastersizer 3000 has taken place together with a formal verifica-
tion process to ensure that measurements are accurate.
Connection prerequisites
Ensure that the new location of the system has the correct services and space
before attempting to move the system see the Basic Guide for site requirements
information.
Before moving the system, ensure that any wet dispersion units are drained, any air
lines for dry dispersion units vented, and the system allowed to cool. All pipes and
cables must be disconnected before moving a dispersion unit.
Post-connection verification
To verify that the system has not been affected by the move, make a measurement
of a standard sample before and after moving the system. If the results differ signif-
icantly, check the connections to see if any mistakes were made. Contact your Mal-
vern representative if the difference persists.
4 3 6 7
ill 8627
3. Plug the power cable into the power supply unit and connect the power
cable from the power supply unit to the optical bench power input but do
not connect to the mains power supply at this point.
Warning!
We recommend that the computer and waste sink are not adjacent.
4. Ensure the dispersion unit tubing is connected correctly, using the following
sections for reference.
To verify that the system is still working correctly, make another measurement
using the standard sample and compare the results to those obtained before moving
the system.
Specifications
Optical unit
All specifications correct at time of publication, but may be subject to alteration.
Chemical compatibility
The Mastersizer and its dispersion units have been manufactured from materials
that are considered to give the widest protection from chemical attack. However, it
is important to check that any sample or dispersant you may use is chemically com-
patible with the materials that they will come in contact with within the system.
In normal operation the sample and dispersant should not come into contact with
any component of the optical bench (the sample path is contained within the dis-
persion unit and the sample cell). However, if a sample pipe or sealing ring should
fail then sample and dispersant may drip into the cell area of the optical bench. A
drain is fitted in the base of the cell area to limit the spread of any leak.
Should a small leak occur the sample/dispersant may come into contact with the
following materials:
Component Materials
Detector lens Glass
Protection window Glass
Cell bay walls Aluminium (hard anodised)
Cell bay base Aluminium (hard anodised)
Drain assembly Stainless steel, Tygon tubing
Cell and dispersion unit See applicable dispersion unit manuals
Caution!
The paint finish of the external covers may be permanently damaged if
samples or dispersants are spilt on them. Clean up any spillage immedi-
ately.
Regulatory Information
Disposal of Electrical & Electronic Equipment
This regulation is applicable in the European Union and other European countries
with separate collection systems.
ill 7610
This symbol on the product or on its packaging indicates that when the last user
wishes to discard this product it must not be treated as general waste. Instead it
shall be handed over to the appropriate facility for the recovery and recycling of
electrical and electronic equipment.
By not discarding this product along with other household-type waste, the volume
of waste sent to incinerators or landfills will be reduced and natural resources will
be conserved. 0
For more detailed information about recycling of this product, please contact your
local city office, your waste disposal service, or Malvern Instruments.
CE Declaration of Conformity
The CE badge on this product signifies conformance to the following European
Directives:
EMC directive 2004/108/EC {BS EN 61326-1: 2006}
Low Voltage Directive 2006/95/EC {BS EN 61010-1: 2010}
Note
This equipment has been tested and found to comply with the limits for a
Class A digital device, pursuant to part 15 of the FCC rules. These limits
are designed to provide reasonable protection against harmful interference
when the equipment is operated in a commercial environment.
This equipment generates, uses, and can radiate radio frequency energy
and, if not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications. Operation of
this equipment in a residential area is likely to cause harmful interference
in which case the user will be required to correct the interference at his
own expense.
Note
Changes or modifications not expressly approved by Malvern Instruments
Limited could void the users authority to operate the equipment.
ill 8708
Translation:
This is a Class B product based on the standard of the Voluntary Control Council
for Interference from Information Technology Equipment (VCCI). If this is used
near a radio or television receiver in a domestic environment, it may cause radio
interference. Install and use the equipment according to the instruction manual.
H P
Hydro EV 2-3 paging layout 7-3
Hydro EV settings 4-21 Parameter filters 5-3
Hydro LV/MV 2-3 Parameters report 7-2
Hydro LV/MV settings 4-19 Power input 2-9
Hygroscopic samples 6-4 Power switch 2-9
Protection window 2-8
I
Q
Installation
guidelines A-2 Quick Access Toolbar 3-13
software A-1
R
K
Record
Kurtosis 5-18 edit 9-6
Record View 3-3
L
filtering 5-3
LED (status) 2-8 Refractive index
licence common dispersants 6-6
about 8-2 Remote support 1-5
sharing 8-3 reports
Logs 3-16 about 7-2
add item 7-9
M Analysis 5-9, 7-2
Maintenance utilities 3-15 creating 7-8
Malvern Access Configurator 8-1 Data 7-2
Malvern personnel 1-2 displaying 7-2
Malvern reports 7-2 editing 7-8
manual measurements 4-14 graph 7-9
Measurement Duration 4-30 Malvern 7-2
Measurement Obscuration 4-32 parameter grid 7-9
Measurement Sequence 4-31 Parameters 7-2
Measurement settings reference 4-23 printing 7-7
Mie theory 5-16, 5-20 table 7-9
Model number 1-5 workspace 7-4
zoom 7-4
O Reports tab 3-7
Reports tabs 3-3
Obscuration 4-32
Representative sampling 6-3
mathematical expression 5-10
Operator tasks 1-3 S
Optical model
Fraunhofer 5-19 Sample
Mie theory 5-20 symptoms of poor preparation 6-10
Optical unit 2-4, 2-5 Sample Dispersant 4-28
about 2-2 Sample Dispersion Accessory 4-34
Optical unit components 2-7 Sample dissolving 6-10
Options 3-11 Sample floating on dispersant 6-10
Sample Identification 4-24
User manual