HPLC Method Validation
HPLC Method Validation
Introduction peatability and intermediate precision), specificity, detec-
Methods Validation: Establishing documented evidence tion limit, quantitation limit, linearity, range, and robustness
that provides a high degree of assurance that a specific (Figure 1). In addition, methods validation information
method, and the ancillary instruments included in the should also include stability of analytical solutions and sys-
7
method, will consistently yield results that accurately reflect tem suitability.
the quality characteristics of the product tested.
Health Canada (HC) has also issued guidance on meth-
8
Method validation is an important requirement for any ods validation entitled Acceptable Methods Guidance. HC
package of information submitted to international regula- has been an observer of ICH, and has adopted ICH guide-
tory agencies in support of new product marketing or clini- lines subsequent to its reaching Step Four of the ICH
cal trials applications. Analytical methods should be vali- process. An acceptable method predates ICH, and HC
dated, including methods published in the relevant pharma- plans to revise this guidance to reflect current ICH termi-
copoeia or other recognized standard references. The suit- nology.
ability of all test methods used should always be verified Figure 2 shows the data required for different types of
under the actual conditions of use and should be well docu- analysis for method validation. Where areas of the Accept-
mented. able Methods Guidance are superseded by ICH Guidelines
1 2
Methods should be validated to include consideration of Q2A and Q2B, HC accepts the requirements of either the
characteristics included in the International Conference on ICH or Acceptable Methods Guidance; however, for
1, 2
Harmonization (ICH) guidlines addressing the validation method validation, ICH acceptance criteria are preferred.
of analytical methods. Analytical methods outside the scope HCs Acceptable Methods Guidance provides useful guid-
of the ICH guidance should always be validated. ance on methods not covered by the ICH guidelines (e.g.,
ICH is concerned with harmonization of technical re- dissolution, biological methods), and provides acceptance
quirements for the registration of products among the three criteria for validation parameters and system suitability tests
major geographical markets of the European Community for all methods.
(EC), Japan, and the United States (U.S.) of America. The HC has also issued templates recommended as an ap-
recent U.S. Food and Drug Administration (FDA) methods proach for summarizing analytical methods and validation
3-5
validation guidance document, as well as the United States data ICH terminology was used when developing these tem-
6
Pharmacopoeia (USP), both refer to ICH guidelines. plates.
The most widely applied typical validation characteris- This paper suggests one technique of validating meth-
tics for various types of tests are accuracy, precision (re- ods. There are numerous other ways to validate methods, all
4 I n s t i t u t e o f Va l i d a t i o n Te c h n o l o g y
Ghulam A. Shabir
Figure 1
___________________________________________________________________________
ICH, USP, and FDA Methods Validation Characteristics Requirements for Various Types of Tests
Figure 2
____________________________________________________________________
Health Canada Methods Validation Parameter Requirements for Various Types of Tests
equally acceptable when scientifically justified. required accuracy, and required sensitivity. (Note: Most of
Prepare a Protocol the acceptance criteria come from the characterization
study.) Furthermore, some tests may be omitted, and the
The first step in method validation is to prepare a proto- number of replicates may be reduced or increased based on
col, preferably written, with the instructions in a clear step- scientifically sound judgment.
by-step format, and approved prior to their initiation. This A test method is considered validated when it meets the
approach is discussed in this paper. The suggested accep- acceptance criteria of a validation protocol. This paper is a
tance criteria may be modified depending on method used, step-by-step practical guide for preparing protocols and per-
A n a l y t i c a l M e t h o d s Va l i d a t i o n 5
Ghulam A. Shabir
forming test methods validation with reference to High Methods validation must have a written and approved
Performance Liquid Chromatography (HPLC) (use simi- protocol prior to its initiation. A project controller will se-
lar criteria for all other instrumental test method valida- lect a validation Cross-Functional Team (CFT) from var-
tion) in the quality system compliance industry. ious related departments and functional areas. The project
controller assigns responsibilities. The following tables il-
Analytical Methods Validation Protocol lustrate one suggested way of documenting and preserv-
Approval Cover Page ing a record of the approvals granted at the various phases
Summary Information
Summary Information
Organization name
Site location
Department performing validation
Protocol title
Validation number
Equipment
Revision number
Project Controller
Document Approval
Document Approval
Department /
Functional Area Name Signature Date
Technical Reviewer
End Lab Management
Health & Safety
Quality Assurance
Documentation Control
(reviewed and archived by)
Revision History
Revision History
Revision No. Date Description of change Author
6 I n s t i t u t e o f Va l i d a t i o n Te c h n o l o g y
Ghulam A. Shabir
A n a l y t i c a l M e t h o d s Va l i d a t i o n 7
Ghulam A. Shabir
5 (e.g.) 25
10 50
15 75
20 100
30 150
40 200
Equation for regression line = Correlation coefficient (r2) =
Replicate 1
Replicate 2
Replicate 3
Replicate 4
Replicate 5
Replicate 6
Replicate 7
Replicate 8
Replicate 9
Replicate 10
Mean
SD
RSD%
8 I n s t i t u t e o f Va l i d a t i o n Te c h n o l o g y
Ghulam A. Shabir
The precision data used for this assessment is the preci- Record the retention time, peak area, and peak height on
sion of the three replicate samples analyzed at each level the datasheet. Calculate the mean, standard deviation,
in the accuracy studies. and RSD.
1.3.2. Documentation 1.5.3. Acceptance criteria
Record the range on the datasheet. The FDA states that the typical RSD should be 1% for
drug substances and drug products, 2% for bulk drugs
1.3.3. Acceptance criteria and finished products. HC states that the RSD should be
The acceptable range will be defined as the concentra- 1% for drug substances and 2% for drug products. For
tion interval over which linearity and accuracy are ob- minor components, it should be 5% but may reach
8
tained per the above criteria, and in addition, that yields 10% at the limit of quantitation.
9
a precision of 3% RSD.
1.6. Intermediate Precision
1.4. Accuracy
1.6.1. Test procedure
1.4.1. Test procedure Intermediate precision (within-laboratory variation) will
Spiked samples will be prepared at three concentrations be demonstrated by two analysts, using two HPLC sys-
over the range of 50 to 150% of the target concentration. tems on different days and evaluating the relative per-
Three individually prepared replicates at each concen- cent purity data across the two HPLC systems at three
tration will be analyzed. When it is impossible or diffi- concentration levels (50%, 100%, 150%) that cover the
cult to prepare known placebos, use a low concentration analyte assay method range 80 to 120%.
of a known standard.
1.6.2. Documentation
1.4.2. Documentation Record the relative % purity (% area) of each concentra-
For each sample, report the theoretical value, assay tion on the datasheet.
value, and percent recovery. Calculate the mean, stan-
dard deviation, RSD, and percent recovery for all sam- Calculate the mean, standard deviation, and RSD for the
ples. Record results on the datasheet. operators and instruments.
A n a l y t i c a l M e t h o d s Va l i d a t i o n 9
Ghulam A. Shabir
of a number of blank samples and then multiply this num- centration that gives an RSD of approximately 10% for
8
ber by two to estimate the signal at the limit of detection. a minimum of six replicate determinations.
10 I n s t i t u t e o f Va l i d a t i o n Te c h n o l o g y
Ghulam A. Shabir
which is potentially interfering (impurity, excipient, and temperature adjusted by 5C. If these changes are within
degradation product). Reproducibility: RSD for peak the limits that produce acceptable chromatography, they
9, 10
area, height, and retention time will be 1% for six injec- will be incorporated in the method procedure.
tions. Tailing factor (T): T should be 2. Theoretical
3
plates (N): 2000.
A n a l y t i c a l M e t h o d s Va l i d a t i o n 11
Ghulam A. Shabir
List of Appendices
Appendix Document Title Total Pages
No.
Appendix 1
______________________________________________________________________________
Method Validation Personnel Signature and Review Worksheet
Comments:
Appendix 2
______________________________________________________________________________
Equipment and Software Used in Method Validation Worksheet
Comments:
12 I n s t i t u t e o f Va l i d a t i o n Te c h n o l o g y
Ghulam A. Shabir
Appendix 3
______________________________________________________________________________
Document and Materials Used in Method Validation Worksheet
Comments:
Appendix 4
_____________________________________________________________________________
Analytical Test Method Procedure
This procedure should include the entire testing method and all procedures associated with it.
This appendix can appear in any format, but it should always be included in the documentation
A n a l y t i c a l M e t h o d s Va l i d a t i o n 13
Ghulam A. Shabir
14 I n s t i t u t e o f Va l i d a t i o n Te c h n o l o g y