Manual Endoscopios EG290Kp EC380LKp
Manual Endoscopios EG290Kp EC380LKp
Notes
Read this manual before operating, and save this book for future reference. Failure to read and thoroughly understand the information presented in this
manual, as well as those developed for ancillary endoscopic equipment and accessories, may result in serious injury including infection by cross
contamination to the patient and/or user. Furthermore, failure to follow the instructions in this manual may result in damage to, and/or malfunction of,
the equipment.
This manual describes the recommended procedures for inspecting and preparing the equipment prior to its use and for the reprocessing and
maintenance of the equipment after its use. It does not describe how an actual procedure is to be performed, nor does it attempt to teach the beginner
the proper technique or any medical aspects regarding the use of the equipment.
It is the responsibility of each medical facility to ensure that only well educated and appropriately trained personnel, who are competent and
knowledgeable about the endoscopic equipment, antimicrobial agents/processes and hospital infection control protocol be involved in the use and the
reprocessing of these medical devices. Known risks and/or potential injuries associated with flexible endoscopic procedures include, but are not limited
to, the following: perforation, infection, hemorrhage, burns and electric shock.
Current infection control guidelines require that G.I.scopes and other semi-critical medical devices, that normally come into contact with intact mucous
membranes, such as in the gastrointestinal tract, must at least be high-level disinfected before patient use. Only the user can determine if an instrument
has undergone appropriate infection control procedures prior to each clinical use. It must be recognized that infection control practices involve many
complex and often controversial issues which are constantly evolving. PENTAX strongly recommends that user remain informed of the latest federal
and local regulations, and encourages users to follow infection control guidelines developed by various organizations for health care professionals.
The text contained in this manual is common for various types/models of PENTAX endoscopes and users must carefully follow only those sections and
instructions pertaining to the specific instrument models appearing on the front cover.
If you have any questions regarding any of the information in this manual or concerns pertaining to the safety and/or use of this equipment, please
contact your local PENTAX representative.
Sterility Statement
The instruments identified in this instructional booklet are reusable medical devices. Since they are packaged non-sterile, they must be high-level
disinfected or sterilized BEFORE initial use. Prior to each subsequent procedure, they must be subjected to an appropriate cleaning and either high-
level disinfection or sterilization process.
Conventions
Throughout this manual, the following conventions will be used to indicate a potentially hazardous situation which, if not avoided;
: may result in property-damage. Also, advises owner/operator about important information on the use of this equipment.
Prescription Statement
Federal (U.S.A) law restricts this device to sale by or on the order of a physician or other appropriately licensed medical professional.
5. POST REPROCESSING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5-1. SERVICING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5-2. CARE AND MAINTENANCE TIPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
1. NOMENCLATURE AND FUNCTION
1-1. VIDEO ENDOSCOPE
U
CONTROL BODY and other accessories
L
MODEL DESIGNATION
F
R
D
RIGHT/LEFT DEFLECTION
CONTROL KNOB
ENHANCE BUTTON
Push to activate the VCR for recording live procedures.
NOTE:
To avoid damaging the endoscope, do NOT twist,
rotate or bend excessively any of the rubber strain
reliefs.
-1-
DISTAL END
(Refer to the inside rear cover of this manual)
RUBBER STRAIN RELIEF
BENDING SECTION
INSERTION TUBE
PVE CONNECTOR
Can be rotated within
a 180 range PVE SOAKING CAP OE-C9
This cap must be securely attached before
immersion.
SUCTION NIPPLE
For attachment to
external suction LIGHT GUIDE
source Transmits light from
light source to distal
end of endoscope.
WARNING: Note:
Immediately after use, the metal light guide prong and the Ensure that the soaking cap has been securely
electrical contacts/pins of the endoscope may be HOT. attached (by properly rotating it) to prevent the cap from
To avoid burns, do not touch these areas immediately coming off during reprocessing. Failure to securely
after use. For safer handling after a procedure, grasp the attach the soaking cap can result in scope damage.
PVE connector housing.
-2-
1-2. ACCESSORIES
(CS6021SN)
( CS-C5S )
3) Biopsy Forceps
CUPS HANDLE
Pink handle defines
autoclavable forceps.
CAUTION:
Because of the effect that accessories used through the instrument channel of the endoscope can
have on the performance of the endoscope itself, it is strongly recommended that PENTAX accesso-
ries be used with PENTAX endocopes. If a unique or highly specialized accessory is available from
another source, please contact PENTAX to arrange a test of its compatibility before using it through
the PENTAX endoscope.
Maximum outer diameter of an endoscopic accessory instrument must be at least 0.2 mm less than
the specified instrument channel diameter in PENTAX endoscopes. Working length of an endoscopic
accessory instrument should be approximately 30 cm longer than the endoscope working length.
NOTE:
Depending upon country and/or local PENTAX distributor, each PENTAX endoscopic accessory may
be an optional accessory.
For patient contact endoscopic accessories, follow the specific and detailed instructions on use, care
and maintenance supplied with each product.
To confirm the exact condition of any new accessory device, check the labeling/packaging accompany-
ing the product. Each label/package should clearly identify the contents as either sterile or non-sterile.
-3-
1-3. VIDEO PROCESSOR
CAUTION:
Please refer to the instruction supplied with the processor.
NOTE:
Do not use the new PENTAX OS-H4 water bottle cap with the older OS-H2 water container/bottle.
Although the cap may appear to fit onto the bottle, air may escape resulting in insufficient pressure and
flow of air and water during the endoscopic procedure. Both the PENTAX water bottle cap and bottle
(container) are identified by their appropriate model designation. Ensure that an OS-H4 cap is used only
with the OS-H4 water container/bottle.
Do not overtighten the bottle cap. Overtighting can cause the bottle cap to break.
EPK-i
LAMP SWITCH
WARNING:
The lifetime of the lamp in EPK-i processor is 500 hours. Prior to use, check the lamp life meter on the rear
panel to ensure the lamp life is less than 500 hours. After 500 hours of use, the image quality will
deteriorate. Excessive use of the lamp beyond 500 hours could break the lamp inside the equipment
resulting in damage to the equipment.
EPK-1000 BRIGHTNESS
CONTROLS
AIR PUMP SWITCH
EPK-1000
POWER SWITCH
LAMP SWITCH I : ON
O: OFF
WATER BOTTLE
INTERFACE
SOCKET
WARNING:
The lifetime of the lamp in EPK-i processor is 500 hours. Prior to use, check the lamp life meter on the front
panel to ensure the lamp life is less than 500 hours. After 500 hours of use, the image quality will
deteriorate. Excessive use of the lamp beyond 500 hours could break the lamp inside the equipment
resulting in damage to the equipment.
NOTE:
The lamp life could be affected by frequency of use. In which case, the lamp life might become shorter than
its respective rated hours (EPK-i: 500 hours, EPK-1000 and EPK-100p: 400 hours)
-4-
2. PREPARATION AND INSPECTION FOR USE
Prior to use, the endoscope, video processor and endoscopic accessory instruments must be carefully inspected for
cleanliness and proper function to determine that they are appropriate for patient use:
NOTE:
PENTAX video endoscopes contained in this manual are only compatible with PENTAX video processor,
model EPK-1000 and EPK-i.
CAUTION:
To avoid discontinuation of endoscopic procedure, have an extra (spare) instrument available as a
standby device. Should any unforeseen event or circumstance render the original instrument inoperable
and/or unsafe for patient.
Please refer to the Owners Manual of the specific model of PENTAX video processor for complete instructions.
1) Attach water bottle assembly, 2/3 filled with sterile water to the appropriate location on the left side of the
video processor.
WARNING:
The addition of defoaming agents to the water supply is NOT recommended. Due to their nature, these
silicone based agents cling tenaciously to surfaces. Unless they are rinsed very thoroughly, a barrier
could be created which could reduce the effectiveness of the disinfection/sterilization process. Addition-
ally, repeated use of such defoamers could eventually lead to residual silicone build up resulting in
equipment malfunction such as clogged air and/or water channels.
2) Set the drain lever on the water bottle assembly to the upright position labeled A/W (air/water).
3) Plug the processor into a properly grounded receptacle with the power switch in the OFF position.
4) Make sure that the PENTAX PVE connector is aligned
with the interface socket on the front panel of the proc-
essor.
5) Connect the endoscope to the interface socket on the
EPK-1000
processor as illustrated. Rotate the lever of the inter-
face socket clockwise after insertion.
CAUTION
Failure to do so may create a miscommunication
WATER BOTTLE
between the scope and processor and may result
air/water feeding tube in the loss of the systems automatic brightness
control and in excessive light emission from the
scope. Excessive light intensity can in turn
increase the amount of delivered thermal energy
causing heating of the scope tip, specifically the
illumination (LCB) cover glass and/or objects in
contact with the distal cover glass.
-5-
9) Prior to each procedure, check the endoscope image quality displayed on the monitor. Confirm that the image
quality, color, automatic brightness (iris) functions are acceptable as per the instructions provided with the
PENTAX video processor.
NOTE:
As indicated elsewhere in PENTAX product labeling, endoscopes particularly the quality of the endo-
scopic image should be checked prior to patient use.
CAUTION:
If the endoscope is intended to be clinically used after testing of individual scope functions (suction, air/
water delivery, water jet, etc.) without further reprocessing, the following precaution should be exercised.
Use fresh distilled or sterile water during individual scope function tests to avoid recontamination of the
previously reprocessed instrument by waterborne microorganisms. Tap water, especially that which may
be left idle and uncovered for a prolonged period of time, should not be used during any inspection/test-
ing of the endoscope.
Before proceeding with inspection of individual functions, PENTAX endoscopes should be tested for the integrity
of their water-tight design (example: tear in the instrument channel). This test is described in another section of
this manual entitled: Leakage Testing.
CAUTION:
Various types of endoscope leakage testers exist including manual, electro-mechanical and automated
versions, some of which are stand alone units and others which may be integrated into Automated Endo-
scope Reprocessors (AERs)/Washer-Disinfectors (WDs). It must be recognized that PENTAX does not
evaluate non-PENTAX leak tester systems to satisfy their specific products claims, for their effectiveness
to accurately detect leaks and/or for their compatibility with PENTAX endoscopes. Insufficient pressures
may adversely affect the endoscope, especially if pressurization occurs during automated reprocessing
at elevated temperatures. PENTAX accepts no responsibility for use of non-PENTAX leakage testers.
Users should check with the leakage tester manufacturer and confirm their specific product claims,
including compatibility with PENTAX endoscopes at various temperatures and their ability to detect leaks
with/without fluid immersion and with/without flexing of the scopes distal bending section.
CAUTION:
To avoid further damage to the endoscope or the possibility of malfunction during a procedure, do not
use an endoscope with any abnormalities or outward signs of damage.
c) Make sure that the entire endoscope is clean and has been subjected to either a high-level disinfection or
sterilization process before each patient use.
WARNING:
All instruments must be reprocessed prior to first time use, after any repairs/service and before every
patient use. When utilizing chemo-thermal processes for reprocessing the endoscope, the instrument
should be allowed to return to room temperature prior to use and/or further handling.
-6-
NOTE:
The distal end of the endoscope as well as the electrical contacts/pins on the PVE connector must be
protected against damage from impact. Never apply excess force such as twisting, or severe bending
to the flexible portion of the endoscope.
During pre-use inspection, ensure that the distal objective lens and the illumination (LCB) cover glass
are clean and no residues are present on these distal surfaces. If not, crisp images can NOT be dis-
played. Wipe with a gauze or the like moistened with enzymatic detergent solution.
As indicated elsewhere in PENTAX product labelling, endoscopes particularly the quality of the endo-
scopic image should be checked prior to patient use.
Ideally all patients should be prepped well to maximize visualization of the intended areas of interest.
Patient material and sections should be removed from the area of observation to eliminate the potential
to blur the endoscopic image and/or obscure the illumination system.
Prior to a procedure, remove debris or secreta from observation area as much as a possible.
Continuing use of the light guide with sticky debris might cause steam because debris is deprived of
moisture by heat. If steam is found on the light guide during a procedure, stop it immediately and with-
draw the scope carefully from a patient.
NOTE:
Flexible endoscopes and other sophisticated medical instruments are constructed of special materials,
unique parts and intricate components with strict dimensional tolerances. Specialized assembly techniques
and application of specific sealants and/or adhesives are required to ensure the watertight integrity and
maintain the functionality of these devices. It is therefore imperative that endoscopes be routinely checked
to ensure that parts used in their construction are not loose, missing or compromised that could otherwise
negatively affect the functionality of these devices. Compromised or loose components could result in device
failure, scope damage (via fluid invasion) and/or in incomplete decontamination of used instruments.
PENTAX recommends that prior to use endoscopes should be carefully inspected for their integrity and
checked for any looseness in the mating or joining of components including the following parts/areas:
the channel inlet assembly (biopsy inlet port) ( )
the suction nipple/connector ( )
the air/water inlet port ( )
any valve cylinder ( )
basically, any inlet or outlet port associated with an internal channel, an indirect patient contact portion of
the endoscope
rubber strain relief along insertion tube and umbilical cable (rotate clockwise only to tighten)
One method to check for looseness is to lightly grip
the exposed part, and while grasping the component
carefully attempt to move it in various directions. Use
of a lintfree gauze while grasping metal parts is
recommended as a protection for ones fingers.
If any part/component remains loose (after
attempting to tighten) and/or if there is any
indication or suspicion of an abnormality or
outward signs of damage, do NOT use the
endoscope.
Contact your local PENTAX service facility.
-7-
CAUTION:
To avoid damaging the endoscopes, do NOT twist, rotate or bend excessively any of the rubber strain
relief ( , ) during inspection, clinical use, reprocessing or any handling activity. Be particularly cautious for
the insertion tube strain relief ( ). When wiping the insertion tube and the umbilical cable, use a slow back
and forth motion to wipe them along the tube/cable. Never apply excessive force or torque on these strain
reliefs or slim tubes/cables. During ANY handling of the instrument avoid excess force, twisting, rotation and/
or bending of the actual insertion tubes and umbilical cables to prevent inadvertent damage (crush,
compression, deformity, etc.) to these parts as well as to internal components contained within the
endoscope.
U
L
U
F
U
F
F
R
L
D
F
F
R
D
R
NOTE:
The deflection locks are of the friction type and the degree of friction is adjustable. The degree of locking
friction depends on the rotational position of the lever.
CAUTION:
ANY lack of smooth operation of the deflection controls may be an early indication of internal damage to
and/or part(s) failure within the endoscopes angulation system. To avoid the possibility of further endo-
scope damage or the potential for malfunction of the angulation system, do NOT use the endoscope if the
angulation mechanism does not operate properly.
Prior to use ensure that the deflection controls can rotate smoothly, that there is no grinding or excess
friction within the angulation system and that the distal bending section bends freely and smoothly.
NEVER APPLY EXCESSIVE FORCE TO THE DEFLECTION CONTROLS!
When an endoscope exhibits excessive knob play or if angulation is lost in any direction, do NOT use
the instrument. Excessive knob play can be defined as rotating of the angulation control knob(s) in any
one direction for more than 30 degrees without any corresponding distal tip deflection. The examples
above are indications that service is required to avoid more serious problems with the angulation control
system, including angle or pulley cable/wire breakage and/or the possibility of a frozen distal bending
section.
A frozen bending section can make instrument extraction from a patient more difficult.
-8-
3) Inspection of Air/Water Feeding
OF-B121 a) Prior to use, the air/water feeding valve (OF-B121)
should be inspected. Remove the valve from the
control body and make sure that the black rubber
check-valve (OE-C14) in good condition is prop-
Dont let the skirt erly attached to the bottom of the metal valve stem.
fold under
b) If the rubber check-valve is missing or not attached
properly, correctly reposition the check-valve by
turning it several times on the air/water valve stem.
OE-C14
For proper positioning, there should be no clear-
ance (gap) between the rubber check valve and the
metal air/water valve stem
WARNING:
A worn or damaged check-valve should be replaced with a new one which has already been subjected
to a high-level disinfection or sterilization procedure (Check valve set, model OE-C15 is optionally avail-
able). For repeated use, always ensure that the check-valve has already been reprocessed. A dam-
aged, worn or missing check-valve could create continous air flow or excessive air insufflation and result
in potential patient injury such as pneumatic perforation. If the check-valve is not attached properly, it
could fall off during the endoscopic procedure and result in potential injury to the patient.
CAUTION:
If there is any doubt as to the suitability for use of the rubber check-valve, replace with a fully reprocessed
one prior to each procedure. A set of 10 pieces of the OE-C14 check-valve are available in a package as
model OE-C15.
c) To confirm that the check-valve is attached properly, visually inspect the check-valve and install the air/
water valve into the air/water cylinder on the control body. Proceed to test the air delivery function.
d) Connect the scope to the video processor. Turn air pump ON to desired pressure setting.
Place the scope distal tip into sterile water and confirm that no air bubbles exit the distal air nozzle.
WARNING:
If air bubbles are observed during the test, the rubber check-valve MUST BE REPLACED. Repeat the
test procedure with a new check-valve (OE-C14).
-9-
g) Release the air/water valve to determine if the valve
freely returns to its OFF (neutral) position and
OF-B121 delivery of water (and air) ceases.
h) If the air/water valve OF-B121 does not function
properly, does not move smoothly or feels sticky,
remove the valve and apply a very small amount of
silicone oil lubricant onto all the O-rings. Do NOT
use excess oil, avoid blobs, large drops and/or
squirts of oil directly onto the metal valve stems -
O Ring : OF-B143 instead, simply place a small droplet of oil on ones
sterile gloved forefinger and gently swirl between
thumb and forefinger. Next place the valve with
O-ring in-between thumb and finger and gently rotate the valve so that the oil is evenly applied to the outer
edges of each rubber O-ring. Make sure the oil is applied to all O-rings and wipe off all excess.
Do NOT apply excess oil. Doing so can allow for inadvertent migration of the oil inside channels or other
areas not intended to be lubricated.
NOTE:
Excessive silicone oil (lubricant) should be avoided to prevent occlusion of the internal air or water chan-
nels/nozzles and potential impairment of the normally clear endoscopic image.
i) If air and/or water do not flow properly, NEVER attempt to clear the air or water nozzles with a needle or
any other sharp object. Instead, the following steps should be followed.
1 Disconnect the endoscope from the video processor.
2 Remove the air/water feeding valve.
3 Using a cotton tipped applicator and alcohol, clean the valve recess (receptacle) in the control body thor-
oughly to remove any debris. Do NOT attempt to insert the applicator into the small openings within the
valve receptacle as the cotton or applicator could become lodged within these openings and cause channel
blockage.
4 Following the section of this manual for cleaning the air and water channel with detergent, flush deter-
gent through both the air and water channel.
Then flush the air several times to force any residual solution out of the channel.
Remove the adapters and install the air/water feeding valve.
(Alternate) By leaving the air/water valve in the cylinder instead of the adapter, one may direct pressured
fluid (or air) independently to either channel to expel debris from and/or more forcefully flush solution
into either the air or water channel. This should not be attempted on a completely clogged/blocked air or
water channel/nozzle.
NOTE:
If blockage of the line is encountered, avoid use of excessive pressure to prevent scope damage.
5 Test for normal delivery of air and water. It may be necessary to repeat the above procedure if normal air
and water delivery is still not available.
NOTE:
Do NOT apply excessive force in an attempt to unblock a clogged channel as the channel and/or brush
could become damaged. Whenever channel damage is suspected, the endoscope should be leak tested.
If repeated attempts to flush the air/water system are unsuccessful, do not attempt to use the endoscope
on a patient. Contact the PENTAX service department.
NOTE:
Prior to clinical use, it is important that the entire air channel system be dry. Failure to thoroughly dry the
air system could result in an unclear or blurry image caused by very fine droplets of moisture being swept
over and/or onto the objective lens at the distal end of the scope.
- 10 -
CAUTION:
The instrument, Air/Water channel systems are made of stainless steel, poly phenylene oxide and fluo-
rine-contained polymers. When any fluids are used with this scope, please read carefully and follow all
instructions in the manual supplied with the fluids for use and pay special attention to any reactions with
the materials identified in the intended fluid path. Only the user can determine if the fluids are appropriate
for patient use.
WARNING:
Improperly installed valves may not function as
originally intended. Such valves may not return
to their neutral (released) positions and/or they
may provide continuous suction. Continuous
aspiration can cause loss of air/fluid, difficulty in
Correct Incorrect
maintaining proper insufflation and/or inadvertent
suctioning of tissue into the distal instrument
channel opening.
OF-B120
OF-B120
metal
WARNING:
A worn or damaged value and/or O-ring should be replaced with a new one. The entire value mecha-
nism should be subjected to a high-level disinfection or sterilization procedure prior to use (O-ring set,
model OF-B127, is optionally available). Failure to do so could result in continuous aspiration which in
certain clinical situation can suction tissue into the distal channel opening at the scope tip and/or create
a loss of insufflated air via the suction system.
A compromised valve could also result in the potential for reflux or spit-back of patient fluids that may
present infection control risks.
- 11 -
d) Release the suction control valve to determine if the valve freely returns to its OFF position and the aspira-
tion of water ceases.
CAUTION:
If the instrument is to be used immediately after the inspection, use only fresh distilled or sterile water.
To avoid recontamination of a previously reprocessed endoscope, avoid use of idle/uncovered tap water.
NOTE:
A rubber inlet seal in good condition must be on the instrument channel inlet to prevent the loss of suction
and a risk of cross contamination to the end user due to the potential for reflux (spit-back) of patient flu-
ids. Worn seal will result in leakage and should be replaced. To ensure maximum performance of these
sealing mechanisms, consider replacing the rubber inlet seal with a new fully reprocessed one for each
procedure.
e) If the suction valve does not move smoothly or feels sticky, remove the valve from the suction cylinder
on the control body of the endoscope. Apply a small amount of silicone oil lubricant, OF-Z11, onto rubber
part and rubber the O-ring. Place a small droplet of oil on ones sterile gloved forefinger and gently swirl
between thumb and forefinger. Next place the valve with O-ring in-between thumb and finger and gently
rotate the valve so that the oil is evenly applied to the outer edges of the rubber O-ring.
Remove/wipe off excess lubricant with a soft gauze. Do not use excessive silicone oil.
5) Inspection of Biopsy Forceps and Instrument Channel
CLOSE a) Make sure there are no kinks in the flexible shaft of
OPEN the biopsy forceps.
b) The jaws of the forceps must be free of any residual
debris. Any debris must be cleaned from the for-
ceps before they are used. Use only sterile forceps.
c) The handle mechanism on the forceps should be
operated to open and close the jaws. This mecha-
nism should operate freely.
d) Close and inspect the jaws of the forceps to make
sure the cups are in proper alignment. If the forceps
has a spike, the spike must be completely straight
and fully within the cups.
WARNING:
The use of any forceps or accessory that shows any sign of damage or difficulty of operation must be
avoided. Any malfunction of a forceps or accessory during a patient procedure could result in serious
injury to the patient. Also, the use of damaged forceps or accessories may result in serious and costly
damage to the endoscope.
e) Any accessory should be slowly inserted through the instrument channel inlet with the endoscope in a
straight position. There should be no resistance encountered. If resistance is encountered, do not attempt
to introduce the accessory further. The instrument channel may be damaged and the endoscope should not
be used. Contact the PENTAX service department.
WARNING:
All patient contact accessories must be thoroughly cleaned and subjected to an appropriate high-level
disinfection or sterilization process before being used for the first time and subsequently after each clin-
ical use.
CAUTION:
The instrument, Air/Water channel systems are made of stainless steel, poly phenylene oxide and fluo-
rine-contained polymers. When any fluids are used with this scope, please read carefully and follow all
instructions in the manual supplied with the fluids for use and pay special attention to any reactions with
the materials identified in the intended fluid path. Only the user can determine if the fluids are appropriate
for patient use.
- 12 -
NOTE:
Accessories should always be inspected and checked with the particular endoscope prior to each
procedure.
WARNING:
Do NOT clinically use the endoscope if any irregularity or abnormality is suspected. If there is any doubt
as to the suitability of use for any endoscope component, replace it with a new fully reprocessed one. An
instrument irregularity may cause scope damage and/or compromise patient or user safety.
WARNING:
Every endoscope should be properly disinfected or sterilized before being used for the first time. The
endoscope should have been properly cleaned, disinfected or sterilized after any previous use and after
being returned for any repairs/service.
Refer to the companion manual describing in detail PENTAX reprocessing instructions.
WARNING:
Current infection control guidelines require that endoscopes and their patient contact accessories either
be sterilized or at the least be subjected to high-level disinfection. Accessories which ENTER STERILE
TISSUE or THE VASCULAR SYSTEM must be sterilized before patient use. It is recommended that any
accessory intended for use in the biliary tract be subjected to an appropriate sterilization process. Only
the user can determine if an instrument has undergone appropriate infection control procedures prior to
each clinical use.
1) If the endoscope has just recently been reprocessed, has been prepared or stored properly and passed all pre-
procedure inspections, the instrument should be ready to use. If necessary, the scopes insertion tube may be
wiped down with a gauze dampehed with 70-90% ethyl or isopropyl alcohol.
NOTE:
Contact the manufacturer and follow local regulations regarding safe use, appropriate handling and dis-
posal of alcohol products. Material Safety Data Sheets (Health and Safety Data Sheets or similar docu-
ments depending upon country) available from the alcohol manufacturer should provide guidance to end
users about composition, hazards, chemical and physical properties, first aid, handling and storage, sta-
bility, precautions, disposal, etc. associated with alcohol solution.
2) Gently wipe the insertion tube with a gauze moistened with 70-90% ethyl or isopropyl alcohol.
3) Gently clean the objective lens with a cotton-tip applicator moistened with 70-90% ethyl or isopropyl alco-
hol. A lens cleaner (anti-fogging agent) may also be applied via gauze or other applicator.
4) Check the endoscopic image and confirm that it is of acceptable quality for clinical use. Refer also to the
owners manual supplied with the PENTAX video processor for inspection of the image quality.
5) (SCOPES TO BE INTRODUCED TRANSORALLY)
Prior to trans-oral insertion of the endoscope, place a bite-block (mouthpiece) into the patients mouth to pro-
tect the endoscope from damage during the procedure. Failure to do so can result in scratches, tears and/or
crushing of the insertion portion of the endoscope.
6) Apply a medical grade water soluble lubricant to the insertion tube. Do not use petroleum based lubricants.
- 13 -
NOTE:
The objective lens and the light guide must be kept free of the lubricant and excess lens cleaner.
Prior to a procedure, remove debris or secreta from observation area as much as possible.
WARNING:
Never drop this equipment or subject it to severe impact as it could compromise the functionality and/or
safety of the unit. Should this equipment to be mishandled or dropped, do not use it. Return it to an
authorized PENTAX service facility for inspection or repair.
- 14 -
3. DIRECTIONS FOR USE
WARNING:
This instrument should only be used by physicians who have thoroughly studied all the characteristics of
this instrument and who are familiar with the proper techniques of endoscopy. During the procedure,
always wear protective garments such as gloves, gowns, face masks, etc. to minimize the risk of cross
contamination.
3-1. PRETREATMENT
WARNING:
The light emission from the endoscope could cause thermal injury. To minimize the risk, use only the
minimum amount of brightness and avoid close stationary viewing and unnecessary prolonged use.
4) The deflection controls should be used as needed to position the scope. The deflection of the tip should be
done under direct vision in a gentle and deliberate manner.
CAUTION:
ANY lack of smooth operation of the deflection controls may be an early indication of internal damage to
and/or part(s) failure within the endoscopes angulation system. To avoid the potential for malfunction of
the angulation system, do NOT use the endoscope if the angulation mechanism does not operate prop-
erly.
Ensure that the deflection controls can rotate smoothly, that there is no grinding or excess friction within
the angulation system and that the distal bending section bends freely and smoothly.
NEVER APPLY EXCESSIVE FORCE TO THE DEFLECTION CONTROLS!
If during a procedure angulation is lost in any direction such as when cables snap (broken pulley wire,
broken angle wire, etc.), do NOT continue to use the instrument and do NOT rotate the deflection con-
trols. Should the angulation system fail for any reason, stop the procedure, release the lock lever and
carefully withdraw the endoscope under direct visualization.
The examples above are indications that service is required to avoid more serious problems with the
angulation control system, including the possibility of a frozen distal bending section.
A frozen bending section can make instrument extraction from a patient more difficult.
5) Insufflation can be controlled by the combined use of the air valve or CO2 valve to increase the amount of
insufflation and the suction control to decrease the level of insufflation. The user may elect to use CO2 deliv-
ery in lieu of air delivery by using the CO2 gas adapter (OF-G11) in combination with a special CO2/water
valve. (OF-B130)
- 15 -
NOTE:
Be careful not to deliver too much air.
WARNING:
It must be recognized that variations in air flow (pressure and volume) for patient insufflation may exist
from one manufacturers equipment (processor and/or scope type) to another. It is, therefore, important
to closely monitor the patient at all times and to aspirate excessive air to prevent overinsufflation and
potential pneumatic perforation.
6) Procedures involving poorly prepped patients should be avoided as excessive patient material can negatively
affect certain scope channel functions as well as the ability to maintain a clear endoscopic view.
7) Mucous, fluids and/or other patient material should be aspirated via the instrument/suction channel and suc-
tion control valve to improve visualization. Maintain a clear view during aspiration, avoid prolonged suction
time and use the minimum level of negative pressure required to perform the clinical procedure.
WARNING:
Do not apply excessive negative pressures (high suction settings) and/or prolonged contact of the distal
instrument channel opening (scope tip) against mucosal surfaces to avoid suction polyps, bleeding
and/or other trauma to the patient. During aspiration keep as clear as possible an endoscopic view of
patient anatomy and maintain some distance from scope tip to tissue to avoid suctioning of mucosa
onto/into the distal channel opening.
NOTE:
Continuing use of the light guide with sticky debris might cause steam because debris is deprived of
moisture by heat. If steam is found on the light guide during a procedure, stop it immediately, and with-
draw the scope carefully from a patient.
8) The objective lens may be cleaned during the procedure by alternately using the air/water and suction control
valves.
NOTE:
Should debris on the objective lens be difficult to clean, one can temporarily use the HIGH air pressure
setting on the processor and simultaneously press the air/water and suction control valves. Return air
pressure setting to original selection before proceeding.
9) Image capture, hard copy documentation, video recording, etc. may be carried out as necessary.
10) Before withdrawing the scope, trapped air should be suctioned to reduce patient discomfort.
11) When attempting to withdraw the scope, return the deflection locking levers to their free position. Always
withdraw the scope under direct visualization.
- 16 -
3-3. BIOPSY
CAUTION:
For ALL types of endoscopic accessory instruments, always maintain a view of the accessory during
advancement, use and withdrawal of the device.
WARNING:
For safety reasons, always insert and advance the accessory in the standard, non-magnified mode.
Magnified vision reduces the depth of the viewing field making it difficult to maintain a clear view of the
accessory.
1) Insert the forceps through the slit in the rubber inlet seal. Be certain to hold the forceps handle in such a way
to ensure that the jaws of the forceps are in a fully closed position during insertion.
NOTE:
When the cups are first passed through the inlet
m seal, a temporary resistance will be encountered.
5c Hold the shaft tightly at about 5cm from the cups
and push it through.
NOTE:
During insertion, if the forceps are found hard to advance further due to resistance, decrease the deflec-
tion of the bending section to a level suitable for smooth insertion and insert the forceps again.
CAUTION:
Never apply excessive pressure when introducing
any accessory since the instrument channel may
be damaged. Malfunction of the scope as well as
costly repairs may result.
NOTE:
Because of the effect accessories used in the instrument channel of the endoscope can have on the per-
formance of the endoscope itself, it is strongly recommended that only PENTAX accessories be used
with PENTAX endoscopes. If a unique or highly specialized accessory is available from another source,
please contact PENTAX to arrange a test of its compatibility before using it through the PENTAX endo-
scope.
WARNING:
Accessories which ENTER STERILE TISSUE or THE VASCULAR SYSTEM must be sterile.
Accessories intended for use in the biliary tract should be sterilized before patient use.
- 17 -
3-4. LASER
Laser equipment should only be used by physicians who have thoroughly studied all the characteristics of the
equipment and who are familiar with the proper techniques of endoscopic laser therapy. The user must carefully
read and follow all instructions in the operating manual supplied with the Laser equipment. The Laser equipment
should be carefully and thoroughly inspected and calibrated. Only the user can determine if the condition of the
Laser equipment is suitable.
WARNING:
The PENTAX endoscopes identified in this manual are compatible with Nd:YAG laser (wavelength
1064 nm) and Diode Laser (wavelength 800 nm ~ 1000 nm) only. Do not use these endoscopes with
other types of laser such as KTP, He-Cd, or Excimer Laser Systems. It could result serious injury to the
patient.
2) The gas adapter, which can be secured to the air/water port on the PVE connector, has a luer receptacle to
accept tubing from an external source of non-explosive gas. As long as the air/water feeding tube from a
PENTAX water bottle assembly is connected to the gas adapter and the air pump in the video processor is
turned OFF, non-explosive gas can be delivered.
NOTE:
Set the pressure below 49 kPa (0.5 kg/cm2, 7.1 PSI) and the flow rate at about 1 liter/min.
CAUTION:
Open the valve of the CO2 gas cylinder only AFTER turning off the pump switch of the processor. Failure
to do so will apply excessive pressure to the processor and can cause damage to the air pump.
3) Flow of gas from the nozzle at the distal end of the scope can be checked by placing the tip of the scope under
water and covering the hole on the top of the CO2/water valve. The flow rate of gas should be no greater than
the rate of air delivery when the air feeding valve on the control head of the scope is covered.
NOTE:
It should be noted that as long as the valve of the CO2 gas cylinder is OPEN and the hole at the top of the
A/W feeding valve is NOT covered, CO2 gas will constantly be vented through the A/W valve into the
room. To reduce excessive CO2 concentrations, it is, therefore, recommended to close the CO2 gas cylin-
der valve, work in a well ventilated room, and use air delivery whenever possible during examinations
which are lengthy or in very confined quarters. As an alternative, the optionally available gas/water valve,
Model OF-B130, may be used in place of the standards air/water valve. OF-B130 is a closed two-stage
valve mechanism. Pressing the first stage delivers CO2 gas and fully depressing the second stage acti-
vates water delivery.
When using the OF-B130 valve, there will be no venting of CO2 gas into the room.
Replace OF-B130 with the air/water valve OF-B121 after using the CO2 gas.
4) The water delivery system is activated by depressing the CO2/water feeding valve.
- 18 -
5) The operator and assistant(s) should wear surgical gloves to avoid burns during use of Laser equipment.
NOTE:
One may choose to leave the OF-G11 adapter attached to the endoscope during conventional air insuf-
flation using the standard air/water valve. However, the luer sideport of the OF-G11 must be capped.
Similarly, for normal water delivery, the air pump must be turned ON and the plastic luer lock cap must be
secured to the OF-G11 adapter.
6) The laser probe should be introduced through the endoscope in the same manner as described for biopsy for-
ceps in section 3-3.
7) The position of the active portion of the laser probe should always be clearly visualized before Laser equip-
ment is activated.
8) It should be recognized that a variety of factors can effect the quality of the video endoscope image during
laser use. Intensity of the aiming beam, high power setting of the laser, close distance of laser fiber to scope
tip, excessive tissue burning, can each adversely influence image quality. To obtain optimum results, it is rec-
ommended that the power settings of the aiming beam and laser be adjusted to minimal levels capable of
achieving the desired clinical effect.
9) Follow standard hospital protocol regarding safe-use of lasers, including the wearing of safety eyewear.
WARNING:
Prior to activation of the laser, make sure that the laser fiber exits the distal channel opening of the
endoscope. Failure to do so can result in scope damage and possible ignition/burning of the endoscope.
NOTE:
It is normal for the aiming light to appear white in the video scope image.
WARNING:
Activation of the laser at high power settings may
cause patient injury or thermal damage of the
endoscopes tip.
Avoid use of high power.
PROBE
- 19 -
3-5. ELECTRO-SURGERY
WARNING:
Please refer to the operating manual provided with the electrosurgical unit. Electrosurgical systems may
be of the floating type (BF type, CF-type) or non-floating (B type). To avoid patient and user burn, use
only the floating type ESU (such as ERBOTOM ICC 200) /accessory systems. Do not use the non-float-
ing (B type) electrosurgical systems. The electrosurgical generator and any electrosurgical accessory
should be carefully and thoroughly inspected. Only the user can determine if the condition of the electro-
surgical generator and the electro-surgical accessory are suitable.
WARNING:
To avoid patient and user burn, follow the instruction below before electrosurgical energy is delivered.
1) Use only the electrosurgical generator with the floating grounding type (BF or CF Type). Do not use
the non-floating (B type) electrosurgical systems.
2) Wear rubber gloves and face masks.
3) The position of the target area, the insulated distal portion of the electrosurgical accessory and the
active portion of the electrosurgical accessory, should be visible.
4) The active portion of the electrosurgical accessory should not touch the surrounding tissue directly or
via fluids.
5) The head of any lesion such as polyp should not touch the surrounding tissue directly or via fluids.
6) Physicians and assisting personnel should avoid contact with the patient while high frequency
energy is delivered.
7) Electro-surgical energy should be delivered for as short a time period as necessary to accomplish
the desired clinical effect.
8) Select a high frequency output power setting suitable for the particular intended procedure in order to
avoid thermal invasion of the tissue or insufficient coagulation resulting in excessive bleeding.
9) To avoid the risk of thermal injury, use only insulated accessories.
Never use non-insulated devices while performing endoscopic electrosurgical procedures.
CAUTION:
It should be recognized that the use of electro-surgical accessory devices employing high frequency cur-
rent may interfere with the normal endoscopic image and this interference is not indicative of a malfunc-
tion of the video endoscope system. PENTAX has developed an earth cable, model OL-Z3 intended to
reduce potential RF interference and electronic noise that may appear in the endoscope image when
using electro-surgical devices. Ensure to always correctly connect cable OL-Z3 between the endoscope
and video processor as described in the instructions provided with the OL-Z3
to video processor
to endoscope
- 20 -
4. CARE AFTER USE
IMPORTANT INSTRUCTIONS
Cleaning-Disinfection-Sterilization: PENTAX Endoscopes
To maintain maximum performance and a long service life of the endoscope, proper care after each proce-
dure is extremely important. Immediately after the completion of a procedure, the endoscope and its com-
ponents should be thoroughly and carefully cleaned. If the endoscope or its components are left uncleaned
for some time after use, dried blood, mucous, contrast material or other patient debris may cause damage
to the instrument (component) or may interfere with the ability of the user to properly reprocess the endo-
scope or component
NOTE:
This manual has been written having regard to ISO 17664 and national guidelines on reprocessing of medical
products (e.g. RKI, DGS/DHOS).
NOTE:
This owners manual contains detailed recommendations on the manual reprocessing of PENTAX
endoscopes using PENTAX supplied cleaning/disinfecting adapters. Automated endoscope reproces-
sors (AER) may provide a means of reprocessing flexible endoscopes, including PENTAX instruments.
Only those Automated Endoscope Reprocessors (AERs) should be used whose manufacturers provide
device-specific instructions and have validation data that support each AER claim with respect to PEN-
TAX model instruments.
AER manufacturers should be consulted for their specific claims including but not necessarily limited to
a) the ability of the AER to provide a cleaned and high-level disinfected (or sterilized) endoscope and
scope components (ex. valves),
b) the identification of any special feature area (internal channel) or scope component that can not be
reprocessed and therefore requires manual reprocessing,
c) the microbial quality of the rinse water,
d) the inclusion of an automated alcohol rinse cycle.
e) the inclusion of a terminal drying cycle that removes the majority of water/fluid within scope channels,
f) maintenance procedures for water filter replacement and/or decontamination of the filtration system to
ensure the microbial claim of water, etc.
g) compliance with local regulations and/or guidelines
- 21 -
CAUTION:
(Automated Reprocessing) Unless the AER/WD has validated channel monitoring capability for device-
specific scope model.
prior to automated reprocessing check and confirm the patency of any lumens/channels to ensure
that all internal channels are unblocked and/or unclogged
prior to use of any AER/WD confirm the integrity of all channel tubes and recommended channel clean-
ing adapters
once tubing sets are connected to the appropriate channels, check the tubing to confirm that there are
no kinks or crimps that could restrict flow of solution throughout the endoscope
(if the unit allows) during reprocessing check and confirm that reprocessing solutions are exiting
expected channel ports
Failure to adhere to the above recommendations could result in inadequate cleaning, disinfection (or
sterilization) of and/or improper removal of residues from all instrument channels/surfaces.
- 22 -
4-1. ENDSCOPES
- 23 -
(1) INTERNAL CHANNEL
Instrument Channel
Air/Water Nozzle
Water Bottle
Suction Tube
Suction Source
Video Processor
The illustration above shows the actual routes taken by air, water, and suction through single biopsy/instrument
channel PENTAX videoscopes with forward optics.
- 24 -
(2) REPROCESSING SYSTEM
EG-290Kp/EC-380Kp
A/W Instrument Channel Cleaning Adapter OF-B115
Instrument Channel
Air/Water Nozzle
Suction Tube
Suction Nipple
ETO Cap
(REMOVE)
Luer-Slip Syringe
with
Cleaning/Disinfectiing
Solution
CAUTION:
BEFORE IMMERSING:
a) The Red ETO gas sterilization venting cap must be taken OFF.
b) The PVE Soaking cap should be securely ON the electrical contacts.
To reprocess a PENTAX single-channel videoscope, first an enzymatic detergent and then a high-level disinfectant
or sterilant must be exposed to all internal lumens as well as to all external instrument surfaces and scope components
(air/water valve, suction valve, etc.). Exposure times of detergent and disinfectant/sterilant must be strictly adhered
to as recommended by each respective manufacuturer.
Please note that all solution entrance ports and flow pathways are illustrated above.
NOTE:
Prior to exposure of all internal channels to an enzymatic detergent and high-level disinfectant (or steri-
lant), all brushable channels should be manually cleaned with cleaning brushes.
- 25 -
4-1-2 PRE-CLEANING
WARNING:
The importance of meticulous mechanical cleaning of the endoscope and its removable components
cannot be overemphasized.
Prior to disinfection or sterilization, all instruments and components must be scrupulously cleaned.
Failure to do so could result in incomplete or ineffective disinfection and sterilization.
During the reprocessing process, always wear protective garments such as gloves, gowns, face masks,
etc. to minimize the risk of cross contamination.
- 26 -
4-1-3 CLEANING
1) Prepare a basin with warm water and a mild enzymatic detergent per detergent manufacturers instructions.
The solutions must be enzymatic detergents or other cleaning agents specially formulated to clean flexible
endoscopes. For specific brands of compatible solutions, please contact your local PENTAX service facility
or sales representative.
CAUTION:
BEFORE IMMERSING:
PVE a) The Red ETO gas sterilization venting cap
Soaking Cap ETO Cap
must be taken OFF.
b) The PVE Soaking cap should be securely
ON the electrical contacts.
WARNING:
Immediately after use, the metal light guide prong and the electrical contacts/pins of the endoscope may
be HOT. To avoid burns, do not touch these areas immediately after use. For safer handling after a pro-
cedure, grasp the PVE connector housing.
Suction air/water CAUTION:
control feeding valve
valve The use of an enzymatic detergent immediately
Inlet seal after each procedure to dissolve and remove
organic contaminants and proteinaceous debris
is essential to the care and maintenance of the
ETO cap
endoscope from the standpoints of infection
control and functionality.
NOTE:
Detergent refers to an enzymatic detergent or other cleaning solution specifically formulated for use and
compatible with flexible endoscopes.
3) Allow all items to soak in an enzymatic solution for a time period recommended by the manufacturer of the
enzymatic detergent.
NOTE:
Do not squeeze or severely bend the insertion tube.
Do not use any abrasive materials.
Be careful to avoid damage to the distal lenses.
- 27 -
4) Manual Cleaning of Suction System by Brush
CAUTION:
1) It is highly recommended that only PENTAX cleaning brushes specified in our manuals should be used to clean
PENTAX endoscopes.
2) PENTAX cleaning brushes have been specially designed to clean various PENTAX internal channel systems and
valves, ports or cylinders. Validation studies have been performed supporting the use of PENTAX supplied
brushes and cleaning adapters for cleaning PENTAX scopes following PENTAX manual reprocessing instruc-
tions.
3) Over the years some other manufacturers cleaning brushes/devices have been found to damage PENTAX
scopes and/or create the need for servicing as some cleaning devices can become lodged (stuck) inside various
lumens of PENTAX scopes. The likelihood for scope damage or servicing increases if a cleaning device does not
have a protective tip (or contains any sharp-edged surface), if its flexible shaft uses a flimsy plastic material that is
not firm enough to allow for easy accessory advancement and/or if the proper sequence and/or direction of chan-
nel cleaning is not followed as described in PENTAX manuals.
4) The cleaning brushes should be always inserted as described in this manual.
5) It is recommended that cleaning brushes should be grasped about 5 cm (2 inches) from any channel ports/open-
ings while advancing/withdrawing a brush. To avoid scope and brush damage, do not use force.
6) To prevent excess friction between brush and channel, do NOT tightly coil the scope while brushing. Keep the
insertion tube and umbilical cable as straight as possible and never attempt to pass the cleaning brush through a
fully angulated scope. Failure to follow these instructions could result in scope or brush damage.
7) Cleaning brushes/devices, especially those with metal coiled flexible shafts should NOT be inserted into the distal
exit/opening of channels. Doing so can damage the endoscopes.
Suction Valve
Channel Inlet
Instrument Channel
Suction Channel
Suction Tube
Suction Nipple
- 28 -
Cleaning Brush CS6021SN A variety of special brushes have been provided to mechan-
3 2
ically brush clean the entire suction/instrument channels
and tubes. Whenever possible, the entire endoscope should
be immersed in detergent solution during the remainder of
the cleaning procedure.
Brush clean the entire instrument/suction channel system:
a) Using the cleaning brush provided, insert the brush
into the opening of the suction nipple and gently pass
1 the brush until it appears in the suction control valve
receptacle. This will clean the suction tube within the
light guide/umbilical cable.
Then gently withdraw the brush. Repeat at least 3 times
until it is visibly clean.
b) Next, insert the brush into the opening at the bottom of
(CS-C5S)
the suction control valve receptacle (cylinder) on the
control head and gently advance until resistance is felt.
DO NOT USE EXCESS FORCE.
Then gently withdraw the brush. Repeat at least 3 times
until it is visibly clean.
NOTE:
Be sure to inspect the bottom of the suction control
OF-B120
valve receptacle on the control head for any
Depress
debris.
CS-C5S
c) Insert the brush into the instrument channel inlet and
gently advance the brush until it exits the distal end of
the scope. Clean debris off the brush and then gently
withdraw the brush.
Repeat several times ensuring that only a clean brush is
introduced into the channel each time.
d) Using the large bristle of the specially designed clean-
ing brush (CS-C5S), scrub clean the surfaces inside the
suction control valve receptacle on the control head.
Do not insert the brush excessively.
e) Scrub all internal and external surfaces of the suction
valve OF-B120 using the smaller side of the cleaning
brush (CS-C5S).
CAUTION:
A worn or damaged valve and/or O-ring should be replaced with a new one. The entire valve mechanism
should be subjected to a high-level disinfection or sterilization procedure prior to use (O-ring set, model
OF-B127, is optionally available). Failure to do so could result in continuous aspiration which in certain
clinical situations can suction tissue into the distal channel opening at the scope tip and/or create a loss
of insufflated air via the suction system.
A compromised valve could also result in the potential for reflux or spit-back of patient fluids that may
present infection control risks.
NOTE:
Brushing of all internal channels is not a substitute for exposure to an appropriate cleaning solution. Man-
ual cleaning by brush complements and augments the cleaning effectiveness of chemical cleaning (i.e.
enzymatic detergent).
- 29 -
5) Chemical Cleaning by Detergent Solution
CAUTION:
Current PENTAX manual cleaning instructions have been validated using enzymatic detergent solutions.
For non-enzymatic solutions always adhere to the specific manual cleaning instructions developed by
the manufacturer of the detergent product. When using a non-enzymatic detergent during automated
reprocessing refer to the validated cleaning instructions developed by the manufacturer of the Auto-
mated Endoscope Reprocessor (AER) or Washer Disinfector (WD) unit/system.
NOTE:
It is imperative that the OF-B115 channel cleaning adapter be securely attached to their respective valve
cylinders. Failure to properly match and secure the cleaning adapter could result in ineffective and incom-
plete reprocessing.
Air/Water Channel
a) For Air/Water Cleaning
Cleaning Adapter (OF-G17) Air/water channel cleaning adapter, model OF-G17,
OF-B115
has a standard ANSI luer lock connector to which a
syringe or other device should be attached. Connect
OF-G17 to the air/water port on the PVE connector/
plug. Fresh detergent should be flushed through this
connector and will simultaneously flow through
both the air and water channels within the scope.
Please refer to the internal schematics.
Provided the enzymatic detergent is allowed remain in contact with the internal channel surfaces for the recom-
mended exposure time, the enzymatic solution should dissolve and clean any remaining debris within these chan-
nels.
(Alternate) By leaving the air/water valve in the cylinder instead of the OF-B115 adapter, one may direct pressured
fluid (or air) independently to either channel to expel debris from and/or more forcefully flush solution into either
the air or water channel. This should not be attempted on a completely clogged/blocked air or water channel/noz-
zle.
NOTE:
If blockage of the line is encountered, avoid use of excessive pressure to prevent scope damage.
- 30 -
As an alternative, solution can be drawn into the instrument channels by attaching tubing from an aspirator to the
suction nipple, as long as the aspirator is turned on, detergent solution can be suctioned through the scope.
WARNING:
Avoid introduction of air during the flushing process.
Confirm that no air bubbles exit from the channel openings at the scope distal tip. The presences of the
air bubbles could prevent contact of the deterfent with channel surfaces.
WARNING:
The enzymatic detergent solution should remain in contact with ALL internal channels and external
scope surfaces for the time period recommended by the manufacturer of the detergent.
WARNING:
It is important that ALL internal channels (air, water, instrument, etc.), external scope surtaces and com-
ponents be thoroughly rinsed with clean water to remove residual detergent solution.
6) Prior to rinsing, attach the previously removed adapters and components, and purge all internal channels with
air (using a syringe) to expel residual detergent solution out from each channel.
7) Using clean water, immerse the entire endoscope as well as all removed components and thoroughly rinse all
items.
8) With all cleaning adapters, including irrigation tube, still attached to the endoscope, flush all previously air
purged channels with 200 mL clean water. All internal channels must be thoroughly rinsed to remove resid-
ual detergent and debris. Then remove the endoscope and its components from the solution.
9) Rinse water remaining within the channels should be purged using air to prevent dilution and/or adulteration
of antimicrobial agents to be used in the subsequent disinfection or sterilization process.
NOTE:
70-90% ethyl or isopropyl alcohol for disinfection followed by compressed air, not greater than 165 kPa
(1.69 kg/cm2, 24 PSI), may be used to facilitate drying.
10) Gently dry all external surfaces of the endoscope with a soft gauze or the like. Dry the objective lens with a
cotton tip applicator. Do not put tension on the insertion tube on the endoscope while drying since the outer
cover of the bending section may be excessively stretched.
WARNING:
Prior to disinfection or sterilization, it is imperative that any solutions previously used in the cleaning
process be thoroughly rinsed and dried. Failure to do so, could result in ineffective or incomplete disin-
fection and sterilization.
CAUTION:
Never subject the endoscope to ultrasonic cleaning methods employing high-frequency ultrasound.
- 31 -
4-1-4 HIGH-LEVEL DISINFECTION
Channel Adapters/Components to be attached Disinfecting solution Rinse water
Air/Water Channel OF-B115 50 mL or more 70 mL or more
OF-G17
Biopsy/Suction Channel OF-B115 100 mL or more 150 mL or more
Rubber Inlet Seal
Before any attempt is made to disinfect the endoscope, the complete cleaning procedure described elsewhere in
this manual must have been completed. Prior to high-level disinfection, the end user should confirm the minimum
effective concentration (MEC) of reused disinfectant as per the manufacturers instructions. Before complete im-
mersion in any disinfecting solution the endoscope should have been Leak Tested as described elsewhere in this
manual.
WARNING:
It is imperative that flexible endoscopes and other semi-critical devices be reprocessed using at least
high-level disinfection with a legally marketed sterilant/disinfectant. Only legally marketed endoscope
automated reprocessing devices/systems whose device specific claims have been validated by the
AER/WD manufacturer and/or reprocessing agents which have been tested and found to be compatible
by PENTAX should be used with PENTAX products.
Generally speaking, 2% and 3.2% alkaline glutaraldehyde solutions which have been FDA cleared
with High-Level Disinfection and/or Sterilization claims are recommended. It should be noted that the
actual percentage of active ingredient (glutaraldehyde) in these solutions, as per their product label,
may vary from the generic and traditional terms 2% glutaraldehyde and/or 3.2% glutaraldehyde.
For specific brands of compatible disinfectants/sterilants, please contact your local PENTAX service
facility or sales representative. Please also refer to the inside front cover of this manual regarding infec-
tion control.
WARNING:
Avoid introduction of air during the flushing process. Confirm that no air bubbles exit from the channel
openings at the scope distal tip. The presence of the air bubbles could prevent contact of the disinfect-
ant with channel surfaces.
- 32 -
2) As long as the entire instrument is immersed in disin-
fecting solution, the cleaning adapters, the syringes
DC
WARNING:
It is imperative that ALL internal surfaces of the channels are in contact with the disinfecting solution for
the time period recommended by the manufacturer of the solution.
3) While fully immersed, manipulate valve mechanisms and inject disinfectant via syringe into/through remov-
able scope components. This will remove entrapped air bubbles that could inhibit contact of solution with
component surfaces and provide for better exposure of surfaces to germicide. Make sure disinfectant is
injected into/through the rubber inlet seal. The endoscopes component parts, including all valves should
remain in contact with the disinfecting solution for the time period recommended by the manufacturer of the
solution and accepted by the user as appropriate to accomplish the desired clinical effect.
4) After the endoscope and its component parts have been in contact with the disinfecting solution for an appro-
priate time, flush all channels with air to purge remaining disinfectant, then remove the scope and its compo-
nents from the solution. Thoroughly rinse the entire endoscope and all its components with sterile water.
5) a)A 20 mL syringe filled with sterile water should be attached to adapter OF-G17 to flush disinfecting solu-
tion from the air and water channels of the scope. Repeat it four times. Fill a syringe with air and flush
through the air and water channels several times to force any residual water out of the tubing and nozzles.
Dry thoroughly.
b)With the A/W instrument channel cleaning adapter OF-B115 attached, rinse the entire suction system, in-
cluding both instrument channel (s) with sterilie water (200 mL or more). Flush air through the instrument
channel(s) several times to remove residual water. Dry thoroughly.
NOTE:
Ideally, all final rinses should be made with sterile water or bacteria-free water whose microbial quality
has been confirmed via monitoring. After water rinsing, 70-90% ethyl or isopropyl alcohol should be
flushed through all channels, followed by compressed air, not greater than 165 kPa (24 PSI) to facilitate
drying. Channel cleaning adapters should be used for alcohol flushing/rinsing and forced air drying. The
outer surfaces of the endoscopes can be dried by gently wiping with a sterile gauze or lint-free cloth sat-
urated with alcohol.
Regardless of the liquid chemical germicide (sterilant or high-level disinfectant) and/or the quality of the
rinse water used, a dry instrument accomplished by a final alcohol rinse followed by forced air is essential
to prevent bacteria colonization and/or infections associated with waterborne microorganisms. Such
infections are more likely to occur when wet/contaminated instruments are used on patients whose
immune systems are compromised or suppressed or when these devices are used in anatomical areas
considered sterile and/or susceptible to these organisms.
6) Following an alcohol rinse, the following steps may be performed to aid in the drying process.
a) Remove the air/water/instrument channel cleaning adapters, OF-B115, air/water/channel adapter OF-G17,
reinstall the previously reprocessed scope components.
b) Attach the scope to an external suction source, and aspirate air through the channel of the scope to remove
any residual alcohol and to air dry channel surfaces.
- 33 -
A/W
c) Remove the soaking cap and attach the scope to the
processor with the air pump turned ON to its
DRAIN HIGHEST pressure setting and the drain lever of
the water bottle set in the DRAIN position, depress
the air/water valve of the scope fully until all alco-
hol has been discharged from the scope. Thor-
oughly drain all alcohol from the air channel as
well by covering the hole in the air/water valve.
Repeat until no moisture or alcohol is seen exiting
the scope or distal tip.
70-90% ethyl or isopropyl alcohol should be
flushed through all channels, followed by com-
pressed air, not greater than 165 kPa (1.69 kg/
cm2, 24 PSI), to facilitate drying.
7) Gently dry all external surfaces of the endoscope with a soft gauze or the like. Do not put tension on the
insertion tube while drying since the outer cover of the bending section may be excessively stretched. Dry
the objective lens with a cotton-tip applicator.
WARNING:
If the endoscope is to be stored after reprocessing , detach removable valves, components, etc. All
channels should be completely dry before storage.
CAUTION:
PENTAX recommended rinsing parameters described in this section on high-level disinfection apply to
the use of a 2.4% alkaline Glutaraldehyde solution containing no surfactants.
Always check with the manufacturer of the liquid chemical germicide (or AER for automated reprocess-
ing) to confirm that their specific rinsing instructions are based upon validated test data demonstrating
that no harmful levels of residues remain on any device or lumen surfaces that may pose a risk to patient
and/or user.
- 34 -
4-1-5 STERILIZATION
Before any attempt is made to sterilize the endoscope, the complete cleaning procedure as described elsewhere in
this manual must have been completed.
CAUTION:
NEVER place the endoscope in a steam autoclave!! NEVER subject the endoscope to ultrasonic clean-
ing methods employing high-frequency ultrasound!! Follow provided ETO gas sterilization Parameters.
b)
CAUTION:
Prior to placing these endoscopes in a Gas Sterilizer or Aeration Chamber:
a) The Red ETO gas sterilization venting cap MUST be ON securely.
b) The PVE soaking cap should be OFF the electrical contacts.
This is opposite of the immersion instructions.
WARNING:
Failure to thoroughly dry all surface areas could result in incomplete or ineffective sterilization. Moisture
could prevent contact of the ETO gas with the actual contaminated surfaces.
2) The following parameters for Ethylene Oxide Gas Sterilization are proposed.
Temperature: 55C(131F)
Relative Humidity: 50%RH
EO Concentration: 600-650 mg/L
Gas Exposure Time: 5 Hours
Aeration: 12 Hours at 55C(131F)
3) Following ETO Gas Sterilization, aeration time of 72 hours at room temperature is required.
4) Aeration Chamber: To shorten the aeration time to 12 hours, an aeration chamber may be used, provided the
temperature does not exceed 55C (131F).
CAUTION:
Prior to placing these endoscopes in an aeration chamber the RED ETO Gas Sterilization Venting Cap
MUST be ON securely and the PVE soaking Cap should be OFF the electrical contacts.
- 35 -
CAUTION:
Use appropriate heat process indicators and/or biological monitors as recommended by the manufac-
turer of the sterilizer.
.
WARNING:
In addition to efficacy claims always check with the manufacturer of the sterilizer/disinfector to confirm
their specific product claims including compatibility and confirm that they have test data supporting their
claims and demonstrating that no harmful levels of any residues (active/inert ingredients, their byprod-
ucts or derivatives of the processed devices) remain on any instrument or lumened surfaces that may
pose a risk to patient and users.
4-2-1 CLEANING
WARNING:
Current infection control guidelines require that biopsy forceps and similar endoscopic accessory instru-
ments which enter sterile tissue or the vascular system or break the mucosal barrier must be sterilized
before each patient use. For patient contact endoscopic accessories, follow the specific and detailed
reprocessing instructions supplied with each product.
CAUTION:
Not all manufacturers of Automated Endoscope Reprocessors (AERs) and Washer-Disinfectors (WDs) make
specific claims nor provide special instructions for reprocessing all of the removable scope components that
are integral to the safe and effective operation of flexible endoscopes. Therefore, should the AER/WD manu-
facturers instructions not specifically address reprocessing of any particular scope component (air/water
valve, suction valve, inlet seal, etc.) in the AER/WD, then those components must be reprocessed manually as
described in PENTAX instructions/labeling. Prior to use, check with each AER/WD manufacturer as to their
specific claims with respect to reprocessing individual endoscope components as well as any endoscopic
accessory instrument.
1) Reusable endoscopic accessory instruments (such as biopsy forceps, needle, etc.) and removable scope com-
ponents (such as suction valve) should be cleaned immediately after each use since dried blood, mucous, or
other debris may cause damage to the instrument and render the mechanism inoperable, or may interfere with
the ability of the user to reprocess the device or component.
2) Place the EAIs and/or components in a basin with fresh enzymatic detergent solution for the time period
recommended by the manufacturer of the detergent being careful not to tightly coil or kink the flexible shaft.
3) Clean the handle and flexible shaft by gently wiping with a soft gauze or the like. Removable components
such as suction valve should be manipulated and detergent injected directly into/onto component surfaces
and then brushed clean.
4) Rinse all residual detergent from the EAIs by immersing the entire EAIs under clean water and manipulating
the handle mechanism. Similarly rinse component surfaces.
5) Ultrasonic cleaning of forceps and similar accessories is then recommended, provided the manufacturers
instructions and the parameters below are followed: Heavily soiled and/or difficult-to-manually clean compo-
nents such as valve mechanisms, inlet seals, etc. should be ultrasonically cleaned prior to subsequent high-
level disinfection or sterilization.
Frequency Range 44 kHz 6%
Time 5 minutes
- 36 -
CAUTION:
NEVER use ultrasonic cleaning methods with high-frequency ultrasound on the endoscope itself.
NOTE:
All detergent must be removed from the inner mechanism of the forceps and individual scope components.
Detergent that remains after the water evaporates may cause increased friction that may render the mecha-
nism inoperable. Residual detergent may also interfere in the subsequent biocidal process.
NOTE:
It is imperative that ultrasonic cleaning of the biopsy forceps, other endoscopic accessory instruments and
other components be performed PRIOR to steam sterilization. Only those PENTAX accessories identified by
their pink colored handle, labeled sa being autoclavable or identified below may be subjected to steam auto-
claving.
6) After cleaning and thorough rinsing, the endoscopic accessory instruments and components should be gently
dried using a soft gauze or the like. Avoid tight coiling or kinking and do NOT put tension on the flexible
shaft of the forceps and similar endoscopic accessory instruments.
NOTE:
Other PENTAX reusable accessories (channel cleaning adapters, cleaning brushes, bite block, etc.) and scope
components (rubber inlet seals, air/water and suction control valves, etc. not specifically identified previously
should be cleaned in a similar manner as above. Ultrasonic cleaning methods are recommended for accesso-
ries whose entire surfaces are not easily accessible by manual cleaning.
Heavily soiled components such as valve mechanisms, inlet seals, etc. should be ultrasonically cleaned prior
to subsequent high-level disinfection or sterilization. For these specific items, please refer to their separate
operation manuals or instruction sheets.
Check-valve OE-C14
Biopsy Forceps
* The endoscopic accessory instruments and scope components listed above may be subjected to ultrasonic
cleaning methods and/or steam sterilization.
- 37 -
4-2-2 HIGH-LEVEL DISINFECTION
WARNING:
Current infection control guidelines require that biopsy forceps and similar endoscopic accessory instru-
ments (EAIs) which enter sterile tissue or the vascular system or break the mucosal barrier must be ster-
ilized before each patient use. It is recommended that any endoscopic accessory instruments intended
for use in the biliary tract be subjected to an appropriate sterilization process. For patient contact endo-
scopic accessories, follow the specific and detailed reprocessing instructions supplied with each prod-
uct.
Before any attempt is made to disinfect endoscopic accessory instruments and/or scope components such as bite
block, air/water and suction control valve, brushes, etc. the complete cleaning procedure as described elsewhere
in this manual, must have been completed. Heavily soiled components such as valve mechanisms, inlet seals, etc.
should be ultrasonically cleaned prior to subsequent high-level disinfection.
1) The entire accessory or component should be immersed in disinfecting solution.
2) Accessory and component surfaces should remain in contact with the disinfecting solution for the time period
recommended by the manufacturer of the solution and accepted by the user as appropriate. To ensure better
contact, manipulate components such as valves while injecting disinfectant into/onto components surfaces.
3) After the item has been in contact with the disinfecting solution for the appropriate amount of time, remove it
from the solution.
4) Rinse all residual disinfecting solution from the accessory/component by immersing it under sterile water.
5) After thoroughly rinsing, the items should be gently dried using a soft gauze or the like. Compressed air may
also be used to facilitate drying.
NOTE:
Ideally, all final rinses should be made with sterile water or bacteria-free water whose microbial quality
has been confirmed via monitoring. After water rinsing, 70-90% ethyl or isopropyl alcohol should be
flushed through lumens of the endoscopic instruments as well as any removable scope components
including valve mechanisms, followed by compressed air, not greater than 165 kPa (24 PSI) to facilitate
drying. External instrument and component surfaces can be dried by gently wiping with a sterile gauze or
lint-free cloth saturated with alcohol.
Regardless of the quality of the rinse water used, a dry instrument accomplished by a final alcohol rinse
followed by forced air is essential to prevent bacteria colonization and/or infections associated with
waterborne microorganisms. Such infections are more likely to occur when wet/contaminated instru-
ments are used on patients whose immune systems are compromised or suppressed or when these
devices are used in anatomical areas considered sterile and/or susceptible to these organisms.
WARNING:
The addition of defoaming agents to the water supply and/or automated reprocessing system is NOT
recommended. Due to their nature, these silicone based agents cling tenaciously to surfaces. Unless
they are rinsed very thoroughly, a barrier could be created which could reduce the effectiveness of the
disinfection/sterilization process. Additionally, repeated use of such defoamers could eventually lead to
residual silicone build up resulting in equipment malfunction such as clogged air and water channels.
Similarly, silicone residues can deposit a film onto the distal objective lens causing blurry endo-
scopic images.
- 38 -
4-2-3 STERILIZATION
WARNING:
Current infection control guidelines require that biopsy forceps and similar accessories which enter ster-
ile tissue or vascular system or break the mucosal barrier must be sterilized before each patient use.
It is recommended that any endoscopic accessary instruments intended for use in the biliary tract be
subjected to an appropriate sterilization process. For patient contact endoscopic accessories, follow the
specific and detailed reprocessing instructions supplied with each product.
Before any attempt is made to sterilize the accessories and/or individual scope components, the complete cleaning
procedure as described elsewhere in this manual must have been completed. Heavily soiled components such as
valve mechanisms, inlet seals, etc. should be ultrasonically cleaned prior to subsequent sterilization.
CAUTION:
Use only the type of packaging material and package configuration as recommended by the manufac-
turer of the sterilizer. Use appropriate heat process indicators and/or biological monitors as recom-
mended by the manufacturer of the sterilizer.
NOTE:
The following sterilization parameters are only valid with sterilization equipment that is properly main-
tained and calibrated.
NOTE:
The PENTAX accessories and scope components listed on page 28 may be subjected to Steam Auto-
claving:
1) Prior to steam sterilization, all autoclavable endoscopic accessary instruments and endoscopic components
previously identified should be thoroughly cleaned using manual and ultrasonic cleaning methods as
described elsewhere in this manual.
2) Autoclaving can then be performed under the following conditions:
CAUTION:
Never place the endoscope in a steam autoclave nor subject it to high-frequency ultrasonic cleaning
methods!
NOTE:
For ethylene oxide sterilization of PENTAX accessories and endoscope components, follow the same
ETO parameters as for PENTAX endoscopes.
- 39 -
4-3. WATER BOTTLE ASSEMBLY
Each water bottle assembly should be cleaned and sterilized at least on a daily basis or more frequently depending
upon the patient and/or type of endoscopic procedure. As with all endoscopic accessories, prior to sterilization,
water bottle assemblies must be scrupulously cleaned. Failure to do so could result in incomplete or ineffective
sterilization.
WARNING:
The addition of defoaming agents to the water supply is NOT recommended. Due to their nature, these
silicone based agents cling tenaciously to surfaces. Unless they are rinsed very thoroughly, a barrier
could be created which could reduce the effectiveness of the disinfection/sterilization process. Addition-
ally, repeated use of such defoamers could eventually lead to residual silicone build up resulting in
equipment malfunction such as clogged air and/or water channels.
NOTE:
Take care in handling the water bottle. DO NOT carry the bottle by the A/W Connector or the hose. When
the bottle cap has been removed from the bottle, be careful in handling the water feeding stem.
Reprocessing At-A-Glance
Autoclave
1) After use, the entire bottle assembly (water container, cap and tubing) should be washed with clean
running water and a dampened gauze or scrub brush.
2) Ultrasonic cleaning of the entire water bottle assembly is then recommended to access difficult to
reach areas, provided the operating frequency is 44 kHz 6% for a period of at least 5 minutes.
3) After washing with a cleaning solution, all surfaces of the water bottle assembly should be thoroughly
rinsed with clean water and dried. Gauze or cloth can be used to wipe dry most surfaces. Compressed
air and 70-90% ethyl or isopropyl alcohol should be used to facilitate drying of hard to reach areas.
CAUTION:
To avoid disconnection and/or bursting of the internal tubing, always set the lever to the Air/Water posi-
tion (upright) and use less than 165 KPa (24 psi) air pressure during forcep air drying.
CAUTION:
Use only the type of packaging material and package configuration as recommended by the manufac-
turer of the sterilizer. Use appropriate heat process indicators and/or biological monitors as recom-
mended by the manufacturer of the sterilizer.
NOTE:
The following sterilization parameters are only valid with sterilization equipment that is properly main-
tained and calibrated.
- 40 -
Steam Sterilization (Autoclaving)
PENTAX water bottle assemblies, Model OS-H4 have been designed to withstand high-pressure steam steri-
lization procedures providing the following parameters are adhered to:
Sterilizer Type : Prevacuum Temperature : 132 ~ 135C (270 ~ 275F) Time : 5 - 18 minutes
During steam sterilization, ensure that the cap and tubing section have been removed from the water bottle
container.
Make sure that the drain lever on the water bottle cap has been set on the A/W position (upright).
- 41 -
5. POST REPROCESSING
1) Following reprocessing, the endoscope may either be reused or placed in storage.
CAUTION:
When utilizing chemo-thermal processes for reprocessing PENTAX endoscopes, the instruments should
be allowed to return to room temperature prior to use and/or further handling.
2) Prior to reuse, ensure that instrument has been properly inspected and fully prepared for the next clinical pro-
cedure.
3) Prior to storage, ensure that all internal channels, scope components, instrument surfaces and accessories are
thoroughly dry.
4) A cotton tipped applicator moistened with 70-90% ethyl or isopropyl alcohol for disinfection may be used to
carefully remove any films or residues left upon the lens surfaces, such as the distal objective lens.
5) The endoscope should be hung in a clean, dry, well-ventilated storage cabinet at room temperature. The
insertion tube and light guide cable should be hung and kept as straight as possible during storage.
CAUTION:
Make sure that all removable components such as the air/water valve, suction valve, rubber inlet
seal(s), and rubber distal tip are detached from the scope. This will allow for better air circulation
throughout the internal channels to ensure thorough drying.
Never store the endoscope, its components and accessories in the carrying case, as this type of dark,
humid and unventilated environment is conducive to bacteria colonization, which increases the risk of
cross contamination. These cases are intended for transportation of the instrument, not storage.
Never store the endoscope in areas of high humidity, high temperatures or in direct exposure to
sunlight or X-rays.
Avoid storage of the scope in cabinets, which have any sharp edges, exposed nails/screws, etc.
Contact with sharp objects can puncture, scratch or otherwise damage the endoscope.
- 42 -
5-1. SERVICING
Prior to returning any instrument for repair to PENTAX, the instrument should first undergo appropriate reproc-
essing/decontamination procedures for the purpose of infection control. Check with your local PENTAX service
facility for more details.
1) All instruments requiring repair should be shipped in the original carrying case with appropriate packing
along with comments describing the instrument damage and complaint.
2) A repair purchase order number, contact name and phone number of the individual responsible for authoriz-
ing repairs, as well as shipping address should be included.
3) The Red ETO gas sterilization venting cap should be attached to the instrument if it will be shipped by air
freight.
4) Any accessories and/or scope component potentially related to the scope damage or complaint should also be
returned with the endoscope.
5) Soaking caps should also be returned with the scope to check/confirm the integrity of their watertight seal.
6) After servicing, all endoscopes must be reprocessed prior to patient use.
WARNING:
Your local PENTAX distributor can provide a list of compatible reprocessing agents with PENTAX
endoscopes based upon material compatibility and functionality studies performed by PENTAX, Japan.
These tests of course apply only to genuine PENTAX parts, components and materials including propri-
etary adhesives, sealants, lubricants, etc. specifically selected for use in PENTAX endoscopes to satisfy
their original design criteria. PENTAX manual reprocessing instructions supplied with each product have
been validated for PENTAX endoscopes utilizing exclusive PENTAX parts/materials and assembled
based upon proprietary PENTAX manufacturing technologies and/or servicing techniques.
It must be recognized that PENTAX does not evaluate non-PENTAX parts, components, materials and/
or servicing methods and therefore questions regarding material compatibility and/or functionality of
PENTAX instruments built with these unauthorized, untested and unapproved items, materials, repair/
assembly methods must be referred to the third party service organization and/or device remanufac-
turer. It is unknown to PENTAX if serviced or remanufactured instruments (performed by unauthorized
PENTAX entities) which still bear a Pentax label are within Pentax device specifications and/or if unau-
thorized activities have significantly changed the instruments performance, intended use, safety and/or
effectiveness.
These companies should confirm the ability for these serviced/remanufactured devices to be reproc-
essed safely and effectively with reprocessing agents/systems recognized as compatible by PENTAX
for standard PENTAX products. These third party companies and/or remanufacturers should be con-
sulted to confirm if they have performed reprocessing validation studies on instrument models which
they have serviced (or remanufactured) that support the cleaning, high-level disinfection and/or steriliza-
tion of these endoscopes via the normal scope OEM reprocessing recommendations, standard AER
device-specific instructions and/or their own unique reprocessing recommendations.
Ultimately, owners of these medical devices are responsible for selecting an appropriate service facility
or vendor whose activities render an instrument to the same expectations and quality of a finished
device supplied by the scope OEM.
- 43 -
5-2. CARE AND MAINTENANCE TIPS
Flexible endoscopes have been an invaluable tool in the medical communitys armamentarium to successfully di-
agnose and treat a wide variety of illnesses in patients for several decades. Perhaps due to their longevity and pro-
gressive design changes over the years which have simplified their use, flexible endoscopes have been somewhat
taken for granted and have erroneously not been considered highly technological medical devices.
In fact, current generation flexible endoscopes although easier to clinically use, are much more sophisticated than
ever. Special reprocessing instructions must be followed to ensure the instruments are patient ready and patient
safe. Special care and handling must be exercised and practiced to prevent instrument malfunction and prolong the
reliability of the endoscope.
The burden of responsibility to ensure safe and reliably functioning instruments is left in the hands of the health-
care professionals who actually care for and reprocess flexible endoscopes.
Naturally, equipment manufacturers share in this responsibility and tremendous efforts have been made in design-
ing instruments which could be reprocessed and maintained as easy as possible. However, due to the nature of
their use and application, flexible endoscopes must be subjected to special cleaning procedures, followed by a dis-
infection or sterilization process after each and every patient use.
To highlight and simplify, what may appear to some as being complicated maintenance and reprocessing instruc-
tions, PENTAX strongly recommends the users review the following suggestions and advice on the care and main-
tenance of your PENTAX flexible endoscopes.
These tips, particularly those involving scope reprocessing should not be construed as shortcuts and are not in-
tended as substitute directions for complete instructions found elsewhere in the owners manual.
** When utilizing chemo-thermal processes for reprocessing PENTAX endoscopes, the instruments should be
allowed to return to room temperature prior to use and/or further handling.
** Make sure that the air/water/instrument channel cleaning adapter is securely attached to the top of the air/water
and suction valve cylinders.
** Avoid soaking of the endoscope with accessories (forceps, injection or aspiration needles, etc.) or any sharp
edged objects which could inadvertently scratch or cut the distal bending section sheath. (Subsequent flexing
back and forth of the rubber sheath could eventually stretch the scratched rubber until a pinhole and leak de-
velops.)
** Exposure to a compatible enzymatic detergent is essential to thorough cleaning of all surfaces of the endo-
scope. Rinsing and drying after cleaning is imperative to prevent dilution and/or adulteration of the disinfect-
ant/sterilant.
** Do not reuse disposable endoscope accessory instruments (EAIs) or scope component intended for single pa-
tient or one time use.
** Do not expose the endoscope or endoscope accessory instruments (EAIs) or scope component to harsh chem-
ical solutions. Strictly adhere to exposure times recommended by the manufacturers of compatible solutions.
** Not all manufacturers of automated endoscope reprocessors (AERs) make specific claims nor provide special
instructions for reprocessing all of the removable scope components that are integral to the safe and effective
operation of flexible endoscopes. Therefore, should the AER manufacturers instructions not specifcally ad-
dress reprocessing of any particular scope component (air/water valve, suction valve, inlet seal, etc.) in the
AER, then those components must be reprocessed manually as descrived in PENTAX instructions/labeling.
Prior to use, check with each AER manufacturer as to their specific claims with respect to reprocessing indi-
vidual endoscope components.
- 44 -
** Avoid contact of any flexible portion of the endoscope with any sharp edge objects (bed frames, table top cor-
ners, sink drains, accessories hanging in storage cabinets, etc.) at any time during the handling, reprocessing
or storage of the endoscope.
** Avoid stretching of the bending section rubber sheath at the distal portion of the scope. During mechanical
cleaning with a dampened gauze, do not use excessive force. A gentle back and forth wiping motion should
be sufficient to remove gross debris. Subsequent soaking in an enzymatic detergent will clean the remainder
of debris.
** Disinfectants and sterilants are toxic substances by nature. All residual solution must be thoroughly rinsed
and dried prior to each patient use.
** The key to preventing clogged air water channels/nozzles is to immediately flush the channels with either air
pressure or fluid/detergent right after removal from the patient. This should be followed by brushing of each
accessible channel and then soaking with an enzymatic detergent.
** Avoid attempting to remove or unscrew scope components which should not be removed. Parts such as the
distal portion of the light guide plug, any inlet ports and any rubber strain reliefs on either the insertion tube
or umbilical cable are essential to the watertight integrity of the instrument. Removal or loosening of these
components and subsequent immersion could lead to fluid invasion into the endoscope and/or compromise
effective reprocessing.
** Check for any sharp edges on all surfaces of an automated cleaning/reprocessing unit which may come in con-
tact with an endoscope. Some units may have sharp edged wire mesh filters and baskets or inlet/outlet ports
which could damage your scope.
** Do NOT overtighten the cap to the water bottle assembly. The metal pipe at the top of the PENTAX water
bottle assembly functions as an inlet port for air from the light source. This inlet pipe should not be used as a
leverage tool to tighten the cap to the water container. Overtightening could cause the plastic cap to crack.
** Do NOT introduce air bubbles into the scopes internal channels during flushing of cleaning and/or disinfect-
ing/sterilizing solutions as these bubbles could interfere in the effectiveness of the disinfection/sterilization
process.
** Do not store the endoscope, its components or accessories in the carrying case as this type of dark, humid and
unventilated environment is conducive to bacteria colonization which increases the risk of cross-contamina-
tion.
** Prior to each use, check the condition of all endoscope accessary instruments.
Do NOT use any accessories with kinked or bent flexible shafts.
Do NOT use forceps with misaligned cups and/or bent needles/spikes.
Do NOT use aspiration or injection needles which are not retractable or whose sharp tips can not be protected.
Do NOT use cleaning brushes without smooth or rounded distal tips.
Do NOT use instruments with exceptionally long rigid sections or whose outer diameter restricts pasage
through the instrument channel/channel inlet.
Use of any of the above accessories could result in channel damage and costly repairs.
** Verification of the effective level of glutaraldehyde (via test strips or similar methods) is recommended to en-
sure potency of glutaraldehyde to achieve high-level disinfection and/or sterilization.
WARNING:
Instrument repairs should only be performed by an authorized PENTAX service facility. PENTAX
assumes no liability for any patient/user injury, instrument damage or malfunctions, or REPROCESS-
ING FAILURE due to repairs made by unauthorized personnel.
WARNING:
NEVER drop this equipment or subject it to severe impact as it could compromise the functionality and/
or safety of the unit. Should this equipment be mishandled or dropped, do not use it. Return it to an
authorized PENTAX service facility for inspenction or repair.
- 45 -
SPECIFICATIONS
EG-290Kp EC-380LKp
*There is no guarantee that instruments selected solely using this minimum instrument channel width will be com-
patible in combination.
NOTE: Specifications are subject to change without prior notice and without any obligation on the part of the man-
ufacturer.
DISTAL END
EG-290Kp EC-380LKp
AIR/WATER
AIR/WATER NOZZLE
NOZZLE
INSTRUMENT INSTRUMENT
CHANNEL CHANNEL
- 46 -
MEMO
- 47 -
MEMO
- 48 -
NOTICE
These instruments are used with Class B Medical Equipment (specified EN55011) and are intended for
hospital or health care districts.
Together, these endoscopes and the compatible processor comply with EN 60601-1-2 (2002) for EU, IEC
60601-1-2 (2001) for other countries.
When used in clinical or residential areas near radio and TV receiver units, these instruments may be
subjected to radio interference.
To avoid and resolve adverse electromagnetic effects, do NOT operate these instruments near the RF
energy equipment
HOYA Corporation
2-7-5 Naka-Ochiai, Shinjuku-ku, Tokyo 161-8525 Japan