Troubleshooting Guide System: MODULARIS Variostar
Troubleshooting Guide System: MODULARIS Variostar
Troubleshooting Guide
System
MODULARIS Variostar
Disclaimer
Siemens provides this documentation as is without the assumption of any liability under
any theory of law.
The content described herein shall be used by qualified personnel who are employed by
Siemens or one of its affiliates or who are otherwise authorized by Siemens or its affiliates
to use such documents.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are not entitled to use this documen
tation without prior written authority.
Copyright
Siemens, 2006 refers to the copyright of a Siemens entity such as:
Siemens Healthcare GmbH - Germany
Siemens Aktiengesellschaft - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 2 of 63
CS PS XP
Siemens, 2006
Restricted
Table of Contents
General
1.1
1.2
1.3
1.4
Siemens, 2006
Restricted
Page 3 of 63
CS PS XP
22
22
24
25
26
27
30
31
31
31
32
SPL1-130.840.20.09.02
09.13
10
10
10
11
11
12
13
13
14
15
16
17
18
19
20
20
21
22
5
5
7
8
9
9
10
32
32
33
34
36
37
38
MODULARIS Variostar
Table of Contents
40
Service software
6.1
6.2
6.3
39
40
40
41
42
44
45
46
47
48
49
50
HyperTerminal program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
English Windows 95/98/NT/XP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Working with the HyperTerminal program . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.1 Select option: h . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.2 Select option: p . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.3 Select option: d . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.4 Select option: l. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.5 Select option: v . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.6 Select option: f . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.7 Select option: e . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.8 Select option: g . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.9 Select option: r . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.10 Select option: s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.11 Select option: i. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.12 Select option: r . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.13 Select option: s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.3.14 Software download with Windows 95 / 98 / NT / XP (only for old
D3 board) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
50
51
52
52
52
53
54
54
55
55
56
57
57
58
58
58
60
61
62
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 4 of 63
CS PS XP
Siemens, 2006
Restricted
General 1
CAUTION
When performing service work and tests, observe:
Tests and adjustments performed with radiation on are identified by the radi
ation warning symbol x.
1.1.1
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 5 of 63
CS PS XP
MODULARIS Variostar
1 General
If the on-site line voltage or line frequency differs from the information indicated upon
delivery of the Modularis Uro Plus, the values given are invalid. The values should be
marked invalid (crossed out with the comment "invalid values" and the service engineer
should sign and date this copy).
In this case, the equivalent leakage current must be measured again. The value must not
exceed 1 mA.
The initial value measured must be documented.
Repeat measurements
When service or repair work is performed on the primary power supply circuit (e.g. repairs
to the power-on circuit), the equivalent leakage current test must be repeated. The values
measured in the subsequent test must not exceed the threshold value of 1 mA as speci
fied in VDE 0751, Part 1. In addition, they must not exceed the initial measured value by
more than 50%. If the value exceeds this threshold, the system must be repaired. The val
ue measured must be documented.
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 6 of 63
CS PS XP
Siemens, 2006
Restricted
General 1
The specified articles are listed in the STC (Service Tools Catalog) if nothing else is indica
ted (the STC is a component of the Spare Parts Catalog).
n
PC connection cable, 5m
ESD equipment
97 06 979
99 94 831
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
99 00 440
Page 7 of 63
CS PS XP
MODULARIS Variostar
1 General
SIREMOBIL Compact or
ARCADIS Varic Technical Documentation
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 8 of 63
CS PS XP
Siemens, 2006
Restricted
General 1
1.4.1
MODULARIS Variostar
n
(1) C plus shock wave head with a focus depth of 140 mm including an adapted iso
center phantom.
Fig. 1: Variostar
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 9 of 63
CS PS XP
MODULARIS Variostar
10
2.1.1
Error messages from the SIREMOBIL Compact/ ARCADIS Varic/Orbic have not been listed.
(Refer to the corresponding service documents of the C-arm.)
Error text:
System on.
Possible cause/
action:
E99
2.1.2
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
E11
Error text:
Possible cause/
action:
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 10 of 63
CS PS XP
Siemens, 2006
Restricted
Error text:
Possible cause/
action:
2.1.3
Error text:
Possible cause/
action:
E15
Error text:
Possible cause/
action:
2.1.4
Controls, E20-E21
If the control panel is dark: check fuse F5 on the D3 board.
E20
Error text:
No contact with the control unit (this error is visible only when
reading out the error log with the service PC).
Possible cause/
action:
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 11 of 63
CS PS XP
MODULARIS Variostar
11
12
Error text:
Possible cause/
action:
1. If the F5 fuse on the D3 board (old part no. 5647651) is blown, replace it with a 3.15 A version and cor
rect the old value on the D3 board with a permanent marker.
2.1.5
Controls, E22-E25
E22
E23
Error text:
QRS signal from ECG unit missing, i.e. no ECG triggering possi
ble.
Possible cause/
action:
Error text:
Possible cause/
action:
E24
Error text:
C-arm blocked.
No change in the actual potentiometer value in the angulation
drive of the MODULARIS Variostar.
Possible cause/
action:
E25
Error text:
Possible cause/
action:
Potentiometer connected?
Check the D3.X30 connector!
Connect the cable to the potentiometer.
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 12 of 63
CS PS XP
Siemens, 2006
Restricted
Controls, E26-E29
E26
E27
E28
Error text:
Possible cause/
action:
Error text:
Control unit.
Possible cause/
action:
Key sticking.
Error text:
Possible cause/
action:
2.1.7
Error text:
Possible cause/
action:
Monitoring, E31-E34
E31
Error text:
Possible cause/
action:
E32
Error text:
Possible cause/
action:
Let the shock wave head cool down, but do not switch the sys
tem off!
Check to ensure that the cooling unit is operating.
If the M1 cooling pump is running, the pump may need to be
replaced.
If the error reoccurs, check the function of the cooling unit.
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 13 of 63
CS PS XP
MODULARIS Variostar
13
14
Error text:
Possible cause/
action:
2.1.8
Monitoring, E40-E48
E40
E41
E42
E43
E44
E45
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
M1 motor OK?
Error text:
Possible cause/
action:
M2 motor OK?
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Replace D3 board.
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 14 of 63
CS PS XP
Siemens, 2006
Restricted
Error text:
Possible cause/
action:
E47
Error text:
Possible cause/
action:
Short circuit?
F4 fuse on D3 (old) board?
New D3 board, part no.(10142950) - check internal voltages in
service mode on the hand control unit.
Check valve (coupling circuit).
E48
2.1.9
Error text:
Possible cause/
action:
Monitoring, E49-E54
E49
Error text:
Possible cause/
action:
E50
E52
Siemens, 2006
Restricted
Error text:
Possible cause/
action:
Replace D3 board.
Error text:
Possible cause/
action:
SPL1-130.840.20.09.02
09.13
Page 15 of 63
CS PS XP
MODULARIS Variostar
15
16
Error text:
Possible cause/
action:
E54
2.1.10
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
The line voltage of the power supply is not stable. This will
cause the line voltage to collapse during the charging process.
Check the power supply (wall outlet) for line resistance or volt
age fluctuations during shock wave release. If the line voltage
is correct, the charging / energy unit is defective.
E61
Error text:
Possible cause/
action:
E62
Error text:
Possible cause/
action:
E63
Error text:
Possible cause/
action:
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 16 of 63
CS PS XP
Siemens, 2006
Restricted
E65
E66
E67
2.1.11
Error text:
The requested high voltage nominal value is not the same one
as confirmed by the charging/energy unit.
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
E69
E70
E71
E72
Siemens, 2006
Restricted
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
SPL1-130.840.20.09.02
09.13
Page 17 of 63
CS PS XP
MODULARIS Variostar
17
18
E74
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
E76
E77
1)
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Before replacing the charging / energy unit, check the fiber optic cable (LWL):
n
Disconnect the fiber optic cable at both ends and check by holding it up to the
light
or
n
2.1.12
Place it on the Clk (U5) of the D3 board and check whether light is visible.
Error text:
Possible cause/
action:
The line voltage of the power supply is not stable. This will
cause the line voltage to collapse during the charging process.
Check the power supply (wall outlet) for line resistance or volt
age fluctuations during shock wave release. If the line voltage
is correct, the charging / energy unit is defective.
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 18 of 63
CS PS XP
Siemens, 2006
Restricted
Error text:
Possible cause/
action:
WARNING
When error 79 is displayed, do not try to release further shockwaves.
A short circuit in the shock wave head or in the high voltage cable will destroy the
high voltage switch in the charging/energy unit.
Check the shock wave head and the high voltage cable for short circuits be
fore replacing the charging/energy unit.
2.1.13
E81
E82
E83
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
E85
Error text:
Possible cause/
action:
E86
Siemens, 2006
Restricted
Error text:
Possible cause/
action:
SPL1-130.840.20.09.02
09.13
Page 19 of 63
CS PS XP
MODULARIS Variostar
19
20
Error text:
Watchdog responded.
Possible cause/
action:
2.1.14
Ultrasound, E90-E93
E90
E91
Error text:
Possible cause/
action:
Error text:
Potentiometer skips.
Possible cause/
action:
E92
E93
2.1.15
Error text:
Possible cause/
action:
Error text:
Erroneous US values.
Possible cause/
action:
Possible cause/action
Software download.
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 20 of 63
CS PS XP
Siemens, 2006
Restricted
Possible cause/action
The service switch is set to normal mode - only for boards that still have no
software.
In case the D3 board has been replaced, the baudrate is 9600 data bits.
If a reload of the software is performed, the baudrate has to be 19200 data
bits.
2.1.16
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 21 of 63
CS PS XP
MODULARIS Variostar
21
22
3.1.1
Error messages from the SIREMOBIL Compact/ ARCADIS Varic/Orbic have not been listed.
(Refer to the corresponding service documents for the C-Arm.)
I01
I02
I03
I04
I05
I08
I09
I10
MODULARIS Variostar
Error text:
Possible cause/
action:
none
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Move STW-Arm to 0.
Possible cause/
action:
Error text:
Degas in progress.
Possible cause/
action:
Graphical view
Error text:
Please wait.
Possible cause/
action:
Waiting.
SPL1-130.840.20.09.02
09.13
Page 22 of 63
CS PS XP
Siemens, 2006
Restricted
I12
Error text:
Emergency-Stop pressed.
Possible cause/
action:
Error text:
Possible cause/
action:
I20
I21
I22
I23
I51
I56
I57
I58
I59
Siemens, 2006
Restricted
Error text:
Brake pressed!
Possible cause/
action:
Check brake.
Error text:
Possible cause/
action:
Check S1 switch.
Error text:
Possible cause/
action:
Check S2 switch.
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
n.a.
Error text:
Card is removed.
Possible cause/
action:
Error text:
No card inserted.
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
ACR 30 disconnected.
Possible cause/
action:
SPL1-130.840.20.09.02
09.13
Page 23 of 63
CS PS XP
MODULARIS Variostar
23
24
I91
I92
I93
I99
3.1.2
Error text:
Possible cause/
action:
Error text:
Delete Counter?
Possible cause/
action:
Error text:
Possible cause/
action:
Adjust tubus.
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
E13
Information:
Possible cause/
action:
n.a.
Information:
Possible cause/
action:
Weak connection.
Hose bent.
Hose damaged.
Motor M1 defective.
E14
Error text:
Possible cause/
action:
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 24 of 63
CS PS XP
Siemens, 2006
Restricted
Error text:
Possible cause/
action:
3.1.3
Controls, E22-E29
E22
Error text:
QRS signal from ECG unit missing, i.e. no ECG triggering possi
ble.
Possible cause/
action:
E23
Error text:
Possible cause/
action:
E24
Error text:
Possible cause/
action:
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 25 of 63
CS PS XP
MODULARIS Variostar
25
26
Error text:
Possible cause/
action:
Potentiometer connected?
Potentiometer ok?
Check the D3.X30 connector!
E28
Error text:
Possible cause/
action:
E29
3.1.4
Error text:
Possible cause/
action:
Monitoring, E31-E52
E31
Error text:
Possible cause/
action:
E32
Error text:
Possible cause/
action:
E34
Error text:
Possible cause/
action:
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 26 of 63
CS PS XP
Siemens, 2006
Restricted
Error text:
Possible cause/
action:
E37
Error text:
Possible cause/
action:
E38
E40
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
M1 motor OK?
Check cooling pump fuse.
D3 defective.
E41
E42
E52
3.1.5
Error text:
Possible cause/
action:
M1 motor OK?
Error text:
Possible cause/
action:
M2 motor OK?
Error text:
Possible cause/
action:
Siemens, 2006
Restricted
Error text:
SPL1-130.840.20.09.02
09.13
Page 27 of 63
CS PS XP
MODULARIS Variostar
27
28
E60
Error text:
Possible cause/
action:
The line voltage of the power supply is not stable. This will
cause the line voltage to collapse during the charging process.
Check the power supply (wall outlet) for line resistance or volt
age fluctuations during shock wave release. If the line voltage
is correct, the charging / energy unit is defective.
E61
Error text:
Possible cause/
action:
E62
Error text:
Possible cause/
action:
Wait until the system cools down, but do not switch off the sys
tem.
Check the ventilation openings.
10 minute pause.
Replace the charging / energy unit!
E63
E64
E65
E67
MODULARIS Variostar
Error text:
Possible cause/
action:
Error text:
The requested high voltage nominal value is not the same one
as confirmed by the charging/energy unit.
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
SPL1-130.840.20.09.02
09.13
Page 28 of 63
CS PS XP
Siemens, 2006
Restricted
E69
E70
E71
E72
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
E73
E75
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
E78
Error text:
Possible cause/
action:
Error text:
Possible cause/
action:
The line voltage of the power supply is not stable. This will
cause the line voltage to collapse during the charging process.
Check the power supply (wall outlet) for line resistance or volt
age fluctuations during shock wave release. If the line voltage
is correct, the charging / energy unit is defective.
Replace the charging / energy unit!
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 29 of 63
CS PS XP
MODULARIS Variostar
29
30
Error text:
Possible cause/
action:
WARNING
When error 79 is displayed, do not try to release further shockwaves.
A short circuit in the shock wave head or in the high voltage cable will destroy the
high voltage switch in the charging/energy unit.
Check the shock wave head and the high voltage cable for short circuits be
fore replacing the charging/energy unit.
1)
Before replacing the charging / energy unit, check the fiber optic cable (LWL):
n
Disconnect the fiber optic cable at both ends and check it by holding it up to the
light
or
n
3.1.6
Place it on the Clk (U5) of the D3 board and check whether light is visible.
Ultrasound, E90-E93
E90
Error text:
Possible cause/
action:
E91
Error text:
Potentiometer skips.
Possible cause/
action:
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 30 of 63
CS PS XP
Siemens, 2006
Restricted
E93
3.1.7
Possible cause/
action:
Error text:
Erroneous US values.
Possible cause/
action:
F20
F53
3.1.8
Error text:
Error text:
Possible cause/
action:
Replace D3 board.
Error text:
Possible cause/
action:
Replace D3 board.
Error text:
Possible cause/
action:
Replace D3 board.
3.1.9
Display
Possible cause/action
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 31 of 63
CS PS XP
MODULARIS Variostar
31
32
Service menu
Old D3, part number: 56 47 651. This D3 board cannot be replaced with the new D3
board (part no.10142950). The boards are not compatible.
Fig. 4: Service_Main
Display
Element
Description
For the service menu to be ac
cessed, the S2 switch on the D3
Board must be in service.
All service functions except a SW
download or error list download
can be done with the hand control
unit.
Pulse count
HV unit val.
Int. Voltages
Error log
MODULARIS Variostar
C-/SH-arm
System setup
SPL1-130.840.20.09.02
09.13
Page 32 of 63
CS PS XP
Siemens, 2006
Restricted
(1)
(2)
(3)
(4)
(5)
4.1.2
Pulse counters
Old D3, part number: 56 47 651
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 33 of 63
CS PS XP
MODULARIS Variostar
33
34
Display
Element
Description
System
Shock Head
Fig. 6: Pulse_Counters
4.1.3
HV unit values
Old D3, part number: 56 47 651
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 34 of 63
CS PS XP
Siemens, 2006
Restricted
Siemens, 2006
Restricted
Display
Element
Description
HV-actual
HV-nominal
U-1/2
I-charge
I-discharge
Surge/Temp
Umax (Last
Pulse)
I-discharge
(Last Pulse)
SPL1-130.840.20.09.02
09.13
Page 35 of 63
CS PS XP
MODULARIS Variostar
35
36
Internal voltages
Display
Element
Description
All internal voltages are displayed.
Cooling pump
Fig. 8: Internal_Voltages
+/-10 %
Tube Drive Fuse
Analog Supply 1
(Nominal 18V).
+/- 10 %
Analog Supply 2
(Nominal 18V).
+/- 10 %
Analog Supply 3
Nominal 5V)
+/- 10 %
Reference Tempera Reference temperature sensor volt
ture Sensor (Nominal age value
10V).
+/- 2 %
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 36 of 63
CS PS XP
Siemens, 2006
Restricted
Error log
Display
Element
Description
Scroll down
(button)
Save errors
(button)
Fig. 9: Error_Log
Delete errors
(button)
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 37 of 63
CS PS XP
MODULARIS Variostar
37
38
Display
Element
Description
C-arm current
End switch
S1 reached
Possible values:
- Off (switch not reached)
- On (switch reached)
- Activated if C-arm reaches end posi
tion left
Brake active
Possible values:
- Off (brake not active)
- On (brake active)
Possible values:
- Off (switch not reached)
- On (switch reached)
- Activated if C-arm reaches end posi
tion right
MODULARIS Variostar
SH-arm decli
nation angle
Actual declination angle of the SHarm. Zero position is UT. Degrees are
counted clockwise.
SH-arm resis
tor value
SPL1-130.840.20.09.02
09.13
Page 38 of 63
CS PS XP
Siemens, 2006
Restricted
System setup
Display
Element
Description
ECG trigger
option
Table orienta
tion
Start potenti
ometer cali
bration (but
ton)
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 39 of 63
CS PS XP
MODULARIS Variostar
39
40
Service menu
New D3 board, part number: 101 42 950 (Variostar serial number 3000 and higher). The
software for the D3 is loaded on the SD card which is inserted in the card slot on the D3
board. When the D3 board is replaced, the SD card from the old board has to be inserted
in the new board. Follow the instructions in the service menu tab card System setup on
the hand control unit.
Display
Element
Description
For the service menu to be accessed,
the S2 switch on the D3 board must
be in service.
Pulse count
HV unit val.
Int. Voltages
Error log
MODULARIS Variostar
C-/SH-arm
System setup
UI
D3
SPL1-130.840.20.09.02
09.13
Page 40 of 63
CS PS XP
Siemens, 2006
Restricted
(1)
(2)
(3)
(4)
(5)
5.1.2
Pulse counters
New D3 board, part number: 101 42 950.
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 41 of 63
CS PS XP
MODULARIS Variostar
41
42
Display
Element
Description
System
Shock head
Charging unit
5.1.3
HV unit values
New D3 board, part number: 101 42 950.
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 42 of 63
CS PS XP
Siemens, 2006
Restricted
Display
Element
Description
HV-actual
HV-nominal
U-1/2
I-charge
I-discharge
Umax (last
pulse)
I-discharge
(last pulse)
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 43 of 63
CS PS XP
MODULARIS Variostar
43
44
Internal voltages
Volt
Name
Description
Supply for
+3.3V
VCC3.3
generated from
local logic,
P_24_DCDC (dc-dc)
FPGA I/O
generated from
local logic,
P_24_DCDC (dc-dc)
card reader,
USB
local analog
(LC filter)
circuitry
generated from
P_24_DCDC (linear)
temperature
and pressure
sensors
input voltage,
control unit,
magnetic
clutch
5%
Fig. 16: Internal voltages
+5V
VCC5
5%
+5V
P_5V_A
5%
+18V
VCC18
5%
+24V
P_24V_X
+10%
+24V
P_24V_
input voltage,
motor driver
+10%
MC
fused 3AT
angulation
+24V
P_24V_
input voltage,
valves
+10%
VALVE
+3.6V
real time
clock
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 44 of 63
CS PS XP
Siemens, 2006
Restricted
Error log
Display
Element
Description
Scroll down
(button)
Save errors
(button)
Delete errors
(button)
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 45 of 63
CS PS XP
MODULARIS Variostar
45
46
Display
Element
Description
C-arm current
End switch
S1 reached
Possible values:
- Off (switch not reached)
- On (switch reached)
- Activated if C-arm reaches end posi
tion left
Brake active
Possible values:
- Off (brake not active)
- On (brake active)
End switch
S2 reached
Possible values:
- Off (switch not reached)
- On (switch reached)
Actual declination angle of the SHarm. Zero position is UT. Degrees are
counted clockwise.
SH-arm resis
tor value
C-arm
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 46 of 63
CS PS XP
Siemens, 2006
Restricted
System setup
Display
Element
Description
ECG trigger
option
Fmax
Patient ID
Software up
date D3
SW update D3
Insert SD card with new SW.
Select the disk symbol and follow
the instructions on the display.
Software up
date UI
Language
Select language
English, German, French or Span
ish.
(1)
(2)
(3)
(4)
(5)
(6)
ECG selection
Maximum shock frequency
Clear Patient ID
Software update D3
Software update UI (hand control unit)
Language selection
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 47 of 63
CS PS XP
MODULARIS Variostar
47
48
5.1.8
US localization
Display
Element
Description
Released US devices:
Prima / G20 / X300
US - Poti Cali
brate
Poti Status
Possible states:
- Not calibrated
- Calibrated
4
5
Value
X300 Setup
(1)
(2)
(3)
(4)
(5)
- Tilt not OK
Enabled US
US - Poti Calibrate
Poti Status
Poti Value
X300 Setup
- Setup OK
The G20 ( 1/Fig. 20 Page 48) selection can also be used for Sonoline Adara and Prima.
Before connecting it to the Variostar, the required software must be installed in the So
noline.
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 48 of 63
CS PS XP
Siemens, 2006
Restricted
5.1.9.1
Release the SD card by carefully pressing the card into the holder to release the lock.
Insert the SD card into the card reader of the service PC.
5.1.9.2
Download the file update.tar from the CS intranet, and copy it onto the SD card as
well. The file can be found in the software folder of the Modularis VARIOSTAR.
Remove the SD card from the card reader of the service PC.
Insert the SD card back into the D3 board and press it carefully until it is locked.
Siemens, 2006
Restricted
Activate the Service menu by moving the S2 switch of the D3 board to the Service po
sition. The former software version is still displayed.
Select the "system setup" tab ( Fig. 19 Page 47).
Select the "Floppy" icon ( 4/Fig. 19 Page 47) to download the D3 SW or select the
Floppy icon ( 5/Fig. 19 Page 47) to download the UI (user interface) SW and follow
the instructions. When the software update is finished, a system reboot will be per
formed automatically.
SPL1-130.840.20.09.02
09.13
Page 49 of 63
CS PS XP
MODULARIS Variostar
49
50
6 Service software
To connect to the D3 board via Hyperterminal, the baudrate must always be set to
19200 data bits.
For a new D3 board without any software, select a baudrate of 9600 data bits in order to
make an initial download of the software.
If questions arise or problems occur, use the Help function in HyperTerminal.
The "Courier New" font is recommended for reading the data stored on the diskette.
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 50 of 63
CS PS XP
Siemens, 2006
Restricted
Service software 6
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 51 of 63
CS PS XP
MODULARIS Variostar
51
52
6 Service software
6.3.1
Select option: h
n
6.3.2
Select option: p
n
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 52 of 63
CS PS XP
Siemens, 2006
Restricted
Service software 6
Fig. 23: select_P
6.3.3
Select option: d
Press the "d" key to display the following text:
Fig. 24: Select_d
Sensors (C-grade)
Hi Volt. (V)
P-bellows (mV)
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 53 of 63
CS PS XP
MODULARIS Variostar
53
54
6 Service software
US-offset (*10 mm) Acceptable max offset: 50 mm. When US localization is not availa
ble, the value is 9995.
C-arm rot. mV
6.3.4
Select option: l
n
6.3.5
Select option: v
n
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 54 of 63
CS PS XP
Siemens, 2006
Restricted
Service software 6
Fig. 26: select_v
The current actual voltages from the sensors and internal voltages are displayed (nominal
values are in brackets).
6.3.6
Select option: f
Fig. 27: Select_f
6.3.7
Select option: e
n
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 55 of 63
CS PS XP
MODULARIS Variostar
55
56
6 Service software
Fig. 28: select_e
The last 250 errors (maximum) are displayed in blocks of 20. Between these blocks is
the line "press space to continue, "x" to exit".
To provide a better overview, system power-up is shown with the code E00.
The oldest error has the number 1; the most recent error is all the way at the top.
6.3.8
The number of error messages in the error log is displayed (e.g. 35 error messages).
The errors can only be deleted from the system after all error numbers have been
viewed.
Select option: g
n
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 56 of 63
CS PS XP
Siemens, 2006
Restricted
Service software 6
Fig. 29: select_g
The current actual temperature from the sensors and the internal voltages are displayed.
6.3.9
Select option: r
n
6.3.10
Select option: s
CAUTION
The system and service PC must not be turned off during a software download.
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 57 of 63
CS PS XP
MODULARIS Variostar
57
6 Service software
58
After "1" has been entered, the software download must start within 60 seconds. If the 60
seconds elapse without the download starting, reselect "h" from the entry menu, reselect
"s", etc.
6.3.11
Select option: i
n
Fig. 32:
After pressing the "1" key and then the "h" key, the display for "selection option i" now
reads: "installed".
6.3.12
Select option: r
n
6.3.13
After pressing the "h" key, display for "selection option r" shows the pulse rate that has
been set.
Select option: s
CAUTION
The system and service PC must not be turned off during a software download.
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 58 of 63
CS PS XP
Siemens, 2006
Restricted
Service software 6
Fig. 33:
After "1" has been entered, the software download must start within 60 seconds. If the 60
seconds elapse without the download starting, reselect "h" from the entry menu, reselect
"s", etc.
Once the software download is complete, check that the values in the "ultrasound locali
zation" menu for the applicators are identical to those in document SPL1-130.038.01..
(Intranet). If this is not the case, make the appropriate changes.
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 59 of 63
CS PS XP
MODULARIS Variostar
59
60
6 Service software
6.3.14
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 60 of 63
CS PS XP
Siemens, 2006
Restricted
Chapter 2
Chapter 3
Siemens, 2006
Restricted
SPL1-130.840.20.09.02
09.13
Page 61 of 63
CS PS XP
MODULARIS Variostar
61
62
MODULARIS Variostar
SPL1-130.840.20.09.02
09.13
Page 62 of 63
CS PS XP
Siemens, 2006
Restricted
siemens.com/healthcare