Ntsc/Pal Vectorscope: 3-Channel Operation
Ntsc/Pal Vectorscope: 3-Channel Operation
NTSC/PAL VECTORSCOPE
Leader's Model 5212 represents the new
generation of vectorscopes. It is equally at home
in NTSC and PAL operations, and this will give
us an opportunity to compare monitoring aspects
of both systems in this issue. Other operating
features that will be covered include 3-channel
operation,
simplified,
high-resolution
measurements of differential gain and phase,
precise inter-signal phase measurements, Y/C
operation, and the use of the vectorscope in
stereo monitoring. Other aspects include menuselected calibration, the use of presets and
remote control.
3-Channel Operation
Despite a multi-fold increase in operating
functions, the front panel of the 5212 remains
simple and familiar. Look at Figure 2-1. The
INPUT selectors show a choice of three
channels plus an EXTernal reference. The latter
may raise a question because the sync REF key
also refers to INT/EXT. So what's EXT doing in
the INPUT group? The answer is that keying
EXT in the INPUT group causes the burst of the
external reference to appear on the vector
display. This means that you can overlay one to
three channels and also display the burst of the
external reference at the same time. Figure 2-2
Auto Phase
Just as a waveform monitor display needs a
proper vertical position adjustment (blanking put
at zero IRE) before readings can be taken, the
phase in the vectorscope must be set to place
the burst signal on the -(B-Y) axis. A routine
adjustment before the vector dots can be
examined quantitatively. The new button on the
front panel of the 5212 makes that routine job
instantaneous. When AUTO PHASE is pressed,
the burst signal for the selected phase reference
is automatically placed on the -(B-Y) axis. At that
instant, the phase readout at the lower right of
Figure 2-5 Normal vector display for SMPTE bars and the
vectorscope calibrated for signals with 7.5% setup.
Figure 2-6 Vectors fall short of the targets for SMPTE bars if
the vectorscope is calibrated for signals without setup.
Figure 2-7 Normal PAL color bar vector display. Note that
there are two vectors for each of the six colors as well as two
bursts.
Figure 2-8 PAL, vector display for red only. Note the R and r
vectors caused by + and - V (R-Y) components.
Figure 2-9 Simulated phase error with the +V switch on. Note
that the two vectors straddle the correct R target.
Figure 2-10 Normal display for red only with the +V switch on.
Differential Gain
A unique feature of the 5212 is a very easy
method measuring differential gain and phase.
The unit or system under test must be driven with
a modulated staircase or ramp signal, and the
output supplied to the vectorscope. To measure
differential gain, press DG in the MODE group.
You will see the instrument step quickly through
an automatic calibration procedure. That is, burst
expands to fall on the large circle, and in a split
second the display for DG and a horizontal
reference line appears. See Figure 2-11. The
positive and negative spikes are formed by the
transitions at the step edges, and are ignored for
Differential Phase
12.)
Figure 2-16 Presetting PHASE to put the slave burst on the (B-Y) axis show that it leads the master by 27.2.
Line Select
The 5212 will display the polar display for
selected lines in the raster when used in
conjunction with the 5222 waveform monitor. All
that is needed is connection of a cable equipped
with 9-pin D sub-connectors between the
REMOTE connector on the 5212 and the
REMOTE A connector on the 5222. This cable is
supplied with the 5212. When the 5222 is placed
in the line select mode, the 5212 will then follow
to show the vector display for the selected
line(s). Figure 2-17 shows an example where a
line select window of 10 lines was placed in that
area of the SMPTE bar pattern that contains the
extra set of bars used for monitor setup. It
contains the color bars blue, magenta and cyan,
as shown.
Y/C Operation
Older vectorscopes could not handle separate
Y and C signals unless synchronized by a
composite or black-burst signal applied to a
second channel or an external sync input. The
5212 accepts Y/C signals if set up to Y/C
operation on the SYSTEM menu (look back at
Figure 2-4,) and Y is fed to CH1 and C to CH2.
Figure 2-18 shows the result. The Y/C notice with
REF CH1 and CH2 appears in the upper right
hand corner. This ties up two of the instrument
inputs, but CH3 is available for genlocked
composite or C signals.