What Is An SOP ?
What Is An SOP ?
water.
Q. Water for pharmaceutical use shall be free heavy metals why ?
A. Heavy metals like lead and arsenic are highly cumulative neurotoxic metals, heavy
metals are not eliminated out of our body easily like other drugs and molecules but
heavy metals bind with proteins and tend to get accumulated in fatty tissues, nerve
tissue is most likely to get damaged by heavy metals, heavy metal causes nervous
tissue damage there for water must be free from heavy metals.
Q. Brazil falls under which climatic zone ?
A. Zone IVB (30 degree celsius and 75% relative humidity)
Q. Change in the size or shape of the original container requires any
stability study?
A. Change in the size or shape of the original container may not necessitate the
initiation of new stability study.
Q. Forced degradation (stress testing) and accelerated stability testing are
same?
A. Forced degradation and stress testing are not same. Stress testing is likely to be
carried out on a single batch of the drug substance. The testing should include the
effect of temperatures (in 10C increments (e.g., 50C, 60C) above that for
accelerated testing), humidity (e.g., 75 percent relative humidity or greater) where
appropriate, oxidation, and photolysis on the drug substance. The testing should also
evaluate the susceptibility of the drug substance to hydrolysis across a wide range of
pH values when in solution or suspension. Photo stability testing should be an
integral part of stress testing.
Q. According to WHO guidelines what is the storage condition of climatic
zone IVa and zone IVb?
A. Zone IV a: 30C and 65% RH (hot and humid countries)
Zone IV b: 30C and 75% RH (hot and very humid countries
Q. Countries comes under climatic zone IVb?
A.Brazil,Cuba,China,Brunei,Cambodia,Indonesia,Malaysia,Myanmar,Philippines,Singa
pore,Thailand
Q. What is the purpose of stress testing in stability studies?
A. Stress testing of the drug substance can help identify the likely degradation
products, which can in turn help establish the degradation pathways and the intrinsic
stability of the molecule and validate the stability indicating power of the analytical
procedures used. The nature of the stress testing will depend on the individual drug
substance and the type of drug product involved.
Q. What is the formula for calculating number of air changes in an area?
A. Number of air changes/hour in an area is
= Total Room Airflow In CFM x 60
Total Volume of room in cubic feet
For calculating Total Room Airflow in CFM, first calculate air flow of individual filter.
Formula is given below.
Air flow (in cfm) = Avg.air velocity in feet/Minute x Effective area of filter
Page 3 of 10
Limit
10%
7.5%
5%
USP
130mg or less
130mg to 324mg
More than 324mg
Limit
10%
7.5%
When a construction is carried out, that calls for reconstruction of AHU system.
Q. What needs to be checked during AHU validation?
A. During AHU validation, following tests shall be carried out
Page 7 of 10
Phase -1
Phase -2
Phase -3
Phase -1
A test period of 2-4 weeks should be spent for monitoring the system
intensively. During this period the system should operate continuously without failure
or performance deviation.Water cannot be used for pharmaceutical manufacturing in
this phase.The following should be included in testing approach.
The sample locations, sampling frequencies and test should be reduced to the
normal routine pattern based on established procedures proven during Phase -1 &
phase - 2.
Q. What are the recommended environmental monitoring limits for
microbial contamination?
Q. What is the difference between calibration and Validation?
Calibration is a demonstration that, a particular
Instrument or device produces results with in specified limits by comparisons
with those produced by a reference or traceable standard over an appropriate
range of measurements.
Whereas Validation is a documented program that provides high degree of
assurance that a specific process, method or system consistently produces a
result meeting pre-determined acceptance criteria.
In calibration performance of an instrument or device is comparing against a
reference standard. But in validation such reference standard is not using.
Calibration ensures that instrument or measuring devices producing accurate
results. Whereas validation demonstrates that a process, equipment, method
or system produces consistent results (in other words, it ensures that uniforms
batches are produced).
Q. Briefly explain about ICH climatic zones for stability testing & long term
storage conditions?
A.ICH STABILITY ZONES
Zone
Type of Climate
Zone I
Temperate zone
Zone II
Mediterranean/subtropical zone
Zone III
Hot dry zone
Zone IVa
Hot humid/tropical zone
Zone IVb
ASEAN testing conditions hot/higher humidity
Long term Storage condition
Climatic Zone
Temperature
Humidity
Minimum Duration
Zone I
21C 2C
45% rH 5% rH
12 Months
Zone II
25C 2C
60% rH 5% rH
12 Months
Zone III
30C 2C
35% rH 5% rH
12 Months
Zone IV
30C 2C
65% rH 5% rH
12 Months
Zone IVb
30C 2C
75% rH 5% rH
12 Months
Refrigerated
5C 3C
No Humidity
12 Months
Frozen
-15C 5C
No Humidity
12 Months
Q. What is bracketing & matrixing in stability testing?
A.Both Matrixing & Bracketings are reduced stability testing designs
Bracketing
The design of a stability schedule, such that only samples of extremes of certain
design factors (ex:strength,package size) are tested at all time points as in full
design.The designs assumes that the stability of any intermediate level is
represented by the stability of extremes tested.
Matrixing
The design of a stability schedule, such that a selected subset of possible samples for
all factor combinations is tested at a specified time point.At a subsequent time point
another subset of samples for all factor combination is tested.The design assumes
Page 9 of 10
that the stability of each subset samples tested represents the stability of all samples
at a given time point.
There for a given time point other than initial & final ones not every batch on stability
needs to be tested.
Q.What are the common variables in the manufacturing of tablets?
A.
Milling conditions
Tablet hardness
Page 10 of 10