897N0602G AspireCRm QC Manual PDF
897N0602G AspireCRm QC Manual PDF
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
CONTENTS
FOREWORD ................................................................................................................................. 5
Introduction to Measurements ............................................................................................. 6
S value ............................................................................................................................................ 7
Overview of Testing ................................................................................................................... 8
Section A
0
Section B
1
2
CNR ..................................................................................................................................................16
Phantom Image* ........................................................................................................................20
3
4
Printer QC ......................................................................................................................................28
Monitor QC ...................................................................................................................................30
Section C
5
Section D
6
Section E
7
8
Section F
9
10 Printer QC ......................................................................................................................................51
11 Monitor QC ...................................................................................................................................53
12 S Value Confirmation ................................................................................................................57
13 System Resolution .....................................................................................................................61
14 CR Reader Scanner Performance..........................................................................................64
15 Mammographic Unit Assembly Evaluation* ....................................................................67
16 Collimation Assessment* ........................................................................................................70
17 Automatic Exposure Control (AEC) System Performance Assessment...................76
18 System Artifact Evaluation*....................................................................................................82
19 Image Quality Evaluation* ......................................................................................................85
20 Dynamic Range...........................................................................................................................91
21 Primary Erasure (Additive and Multiplicative Lag Effects)...........................................94
(*) Portions of this test were reprinted from the ACR Mammography Quality Control Manual.
American College of Radiology; 1999
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FOREWORD
Quality assurance (QA) in mammography means the planned and organized actions designed to
provide confidence that mammographic equipment and related components operated in a facility
will reliably and consistently produce diagnostic quality mammograms for the screening and
diagnosis of breast cancer with minimum dose to patients and staff.
This document, FCRm/CRm Quality Control for Mammography Operation Manual, describes how to
confirm and verify that the Fuji Computed Radiography System when used with an MQSA compliant
X-ray system for screen-film is working in a consistent manner according to the specifications for use as
part of a Full Field Digital Mammography (FFDM) system.
This document applies to the CR-IR 348CL (hereinafter referred to as CR Console) and the CR-IR
363AWS (hereinafter referred to as AWS-c).
Always follow all applicable laws and regulations for your jurisdiction. If anything in this
document is in conflict with applicable laws or regulations, the applicable law or
regulation shall take precedence.
Acknowledgement
We would like to thank the American College of Radiology, Reston, Virginia for their permission to
include portions of the 1999 ACR Mammography Quality Control Manual in this document which
have been reprinted with permission of the American College of Radiology.
No other representation of that material is authorized without expressed, written permission from the
American College of Radiology.
DISCLAIMERS
1
No part or all of this manual may be reproduced in any form without prior permission.
The information contained in this manual may be subject to change without prior notice.
FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
installation, relocation, remodeling, maintenance, and repair performance by other than dealers
specified by FUJIFILM Corporation.
FUJIFILM Corporation shall not be liable for malfunctions and damages of FUJIFILM Corporation
products due to products of other manufacturers not supplied by FUJIFILM Corporation.
FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from remodeling,
maintenance, and repair using repair parts other than those specified by FUJIFILM Corporation.
FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
negligence of precautions and operating methods contained in this manual.
FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from use
under environment conditions outside the range of using conditions for this product such as
power supply, installation environment, etc. contained in this manual.
FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from natural
disasters such as fires, earthquakes, floods, lightning, etc.
Caution: Rx Only (Federal law restricts this device to sale by or on the order of a physician.)
Trademark
FCR is a trademark or a registered trademark of FUJIFILM Corporation.
Copyright 2006-2009 FUJIFILM Corporation. All rights reserved.
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Introduction to Measurements
To produce consistently high quality images, each part of the imaging chain must function within the
limits of performance given. Users must therefore be able to evaluate the status of the acquisition
system, including detector, the processing system and the display systems.
Assumption
Many tests in this manual are specific to the Fuji Computed Radiology (FCR) mammography system
and its related components. Some tests are designed to verify use of the FCR mammography
system in conjunction with an X-ray generator and imaging chain components. It is therefore
required that the X-ray generator and the collimator are performing and calibrated in accordance
with MQSA requirements for screen-film systems. The display systems used for final interpretation
(soft copy and/or hard copy) must be calibrated and performing in accordance with the
manufacturer's quality control program for mammography use.
Clearview RU:
Aspire CRm RU:
Cautions:
Perform the S value Conformation test prior to the remaining Acceptance and QC tests, using
only molybdenum (Mo) target and Mo filter at 25kVp.
If the mammography X-ray unit will not allow 20 mR exposure, you may use the lowest mR
exposure available greater than 20. Always use 25 kVp for this test. Do not use any additional
filtration. The use of another filter may cause an error.
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
S value
FCRm/CRm associates an S value with each FCRm/CRm image. Usually the S value is displayed with
other data in the hard copy or soft copy display.
Technically, the S value is defined by the equation S = 4 10(4-Sk), which is derived from the Sk value
determined by the Exposure Data Recognizer (EDR).
The meaning of the S value differs from that of relative sensitivity or relative speed used with screenfilm systems. Nonetheless, the S value can be used as a guide that roughly represents relative
sensitivity.
Note however, that:
With all other factors constant, the S value changes with time as the photomultiplier tube in the
FCRm/CRm equipment ages requiring periodic calibration.
The S value varies if the positioning setup changes even if the X-ray exposure and tube voltage
remain unchanged.
If the exposure voltage (kVp) changes, the histogram changes, varying the S value.
If the X-ray apparatus changes, the S value changes even if the same exposure conditions are
used.
If the FCRm/CRm exposure menu is changed, the EDR conditions change so S value comparisons
would be invalid.
Even when exactly the same conditions are employed, the S value changes if the time interval
between IP exposure and IP insertion changes. When the X-ray generator, tube voltage, patient,
menu, positioning setup, and elapsed time from exposure remain unchanged, the S value is a
relative measure of the X-ray dose and is inversely proportional to it. For example, the S value is
approximately doubled when the dose is reduced by half.
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Overview of Testing
FREQUENCY
SECTION A : Getting Started
TEST DESCRIPTION
Set-up & Baseline Images
RESPONSIBILITY
Technologist and
Medical Physicist
Technologist
Phantom Image
Printer QC*
Technologist
Monitor QC*
SECTION C : Monthly
(and when problems are
suspected)
Visual Checklist
Technologist
SECTION D : Quarterly
(and when problems are
suspected)
Repeat Analysis
Technologist
SECTION E : Semi-Annual
(and when problems are
suspected)
Compression
Technologist
Imaging Plate (IP) Fog
Viewing and Viewing Conditions*
Annual Printer QC*
Annual Monitor QC*
S Value Confirmation
System Resolution
CR Reader Scanner Performance
Mammographic Unit Assembly Evaluation
Collimation Assessment
SECTION F : Annual
(and when problems are
suspected)
Medical Physicist
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Section
8th Edition
897N0602G
Getting Started
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
System Resolution*
Dynamic Range*
Monitor QC (Annual)
Primary Erasure*
Inter-Plate Consistency
Image Quality*
Dose
S Value Confirmation*
Upon successful completion of the FCRm/CRm Reader Unit baseline tests, satisfy any MQSA
obligations prior to clinical use.
STEP 1: SETTING BASELINE TEST CONDITIONS
When testing equipment, it is very important that the test conditions are reproducible and
consistent. This involves setting up a baseline log for the X-ray room, FCRm/CRm system
parameters, and viewing conditions. These must be recorded in the EXPOSURE & IMAGE PROCESSING
SET UP. Always start by using your most consistent X-ray room and equipment. Decide on the X-ray
room and record it in the EXPOSURE & IMAGE PROCESSING SET UP. Select an IP and cassette of each
size that are in good condition to be put aside for the QC tests. Label them as such and write this
information in the EXPOSURE & IMAGE PROCESSING SET UP. Now record the AEC sensor position,
settings, X-ray generator, target, filter, grid and Image processing parameters that are being used
clinically and that you will use for the ACR MAPP.
Ensure that the menus shown in the following table are available. If you cannot find the menus,
create the menus according to the table on the following page.
10
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
MFP
PEM *
GA
GT
GC
GS
MRB
MRT
MRE
MDB
MDT
MDE
PRN
PTE
PTC PRE
CREATED FROM HR
SENSITIVITY
Physics, Sensitivity
1.0
1.2
0.00
Physics, Artifacts
1.0
1.2
0.00
Physics, Collimation
1.3
1.2
0.00
Physics, IP Fog
1.3
1.2
0.00
0.0
0.0
0.4
2.0
1.4
0.6
1.3
NOTE
In order to set up a different image processing parameter, please change MPM code of the menu
created.
8th Edition
897N0602G
11
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
12
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
X-ray Generator
Target
Cassette ID No.
Filter
IP ID No.
KVp
mAs
Grid
MDB
MDT
MDE
NOTE
Refer to User Guide - Mammography for detailed information about Image Processing
Parameters.
8th Edition
897N0602G
13
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
14
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Section
Section
8th Edition
897N0602G
B
A
A
B
15
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test must be performed at installation after relocating the image reader as in a mobile setting
before imaging patients. In routine use, perform at a frequency of no less than weekly.
PROCEDURE
1.
Using the 18x24cm dedicated mammography QC cassette, erase an IP using the secondary
erasure mode on the reader. Place the loaded cassette in the cassette holder assembly.
Set a manual technique using the following factors: large focal spot, Mo/Mo 26 kVp and 125
mAs. If 125 mAs is not available on the X-ray generator, use the next highest mAs setting.
Note: Establish the baseline CNR value by performing this test five times and averaging the
five CNR values. As the positioning of the aluminum plate may have some effect on the
calculated CNR, remove and reposition the aluminum plate after each exposure to
reflect this normal variation in the baseline. Record the CNR value from each exposure
on the CNR BASELINE WORKSHEET. Calculate the average CNR and 20% values
and record them as indicated on the CNR BASELINE WORKSHEET. Repeat this
process whenever a new baseline is required (see Precautions and Caveats).
16
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
2.
Place the 4cm thick acrylic phantom with a 0.2mm thick aluminum object on top. Make sure
that the aluminum object completely covers the area shown in the diagram below.
Calculation Areas
(20mm 20mm)
Aluminum Object
20mm
BG
Al
20mm
60mm
20mm
Acrylic Slabs
6.
Calculate CNR using the QC calculation tool and record the value in the data form.
7.
Perform the Weekly CNR Test on all mammographic exposure units at your facility that use this
CR Reader.
17
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
18
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
0.2mm Al
Cassette - IP Number :
26kVp 125mAs
FCRm/CRm reader ID :
CNR 1 :
CNR 2:
CNR 3 :
CNR 4 :
CNR 5 :
AVE. CNR (BASELINE):
+/-20% :
to
SIGNATURE:
8th Edition
897N0602G
19
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test must be carried out initially at installation, after appropriate calibration of the equipment to
establish a baseline level. Subsequently, it must be carried out at least weekly, after service to any
equipment, (i.e. the X-ray generator, FCRm/CRm reader unit, image laser printer, and workstation) and
whenever changes in image quality are suspected.
With mobile mammographic equipment, the performance of the system must also be confirmed
after arriving at a new location and prior to imaging patients, in addition to the weekly test.
PROCEDURE
20
1.
Using the designated mammography cassette, erase the IP using the secondary erasure mode
on the reader.
2.
3.
Place the phantom on the cassette holder and position it so the edge of the phantom is fully
covering and aligned with the chest wall side of the image receptor. Center the phantom, left
to right.
4.
The 4mm thick acrylic disc required to be placed on the phantom for hard copy QC may be
present during softcopy QC, but is not required for softcopy QC. For QC of softcopy used for
final interpretation, proceed to step 6.
5.
For hardcopy QC, secure the acrylic disk to the top of the phantom within the image area but
position it so it will not obscure details in the phantom. A suitable location is between and
slightly below the first and second largest fibers. Once located, the disc may be permanently
attached to the phantom with super glue.
6.
7.
Verify the location of the AEC detector. It should be in the same location used for previous
phantom images and completely covered by the phantom.
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
8.
Make an exposure using the AEC technique factors (target, filter, kVp, grid, density control
setting, etc.) currently used clinically for a 4.2-cm compressed breast of 50/50 composition.
9.
Optionally, you may plot the mAs on the control chart after making the exposure. Recall that
mAs will change significantly in the AEC mode if a different kVp, target, or filter is selected. This
step, plotting mAs, is optional and not required.
10. After exposure, ensure that a predetermined interval time between 5 and 10 minutes elapses
prior to the reading process. Process the IP using the "Physics, ACR MAPP" menu.
NOTE
It is important to control the time interval between exposure and reading of the IP. To minimize
variability, it is recommended that an interval be chosen somewhere between 5 and 10 minutes
and that this same interval be used consistently.
11. Record the AEC density control setting on the control chart.
12. For printed images as a final interpretation: Measure the following densities. The background
density should be measured at the geometric center of the phantom image. The measured OD
must not be less than 1.20.
The DD measurement is obtained by subtracting the density inside the acrylic disc from the
density directly adjacent to the disc, in the direction perpendicular to the anode-cathode axis.
For consistent results, these measurements must be made at the same location each time.
Plot the background optical density and the DD on the Hard Copy Phantom Control Chart.
For softcopy images as a final interpretation: Plot the S value from the phantom image
exposure on the S value range line on the SOFT COPY PHANTOM CONTROL/CNR CHART .
VIEWING CONDITIONS
Phantom images should be read under optimal viewing conditions. General lighting should be at a
low level and diffused. Viewboxes or workstations should be positioned to avoid light from windows,
other viewboxes, and other sources of bright light, either direct or reflected. Images should be
masked to eliminate extraneous light. Use a magnification of 2X or higher for scoring speck groups
as well as other test objects.
8th Edition
897N0602G
1.
When scoring the phantom image, each object is scored separately. Always count the number
of visible objects from the largest object of a given type (i.e., fiber, speck group or mass)
downward until a score of 0 or 0.5 is reached, then stop counting for that object type.
2.
Count each fiber as one point if the full length of the fiber is visible and the location and
orientation of the fiber are correct.
Count a fiber as 0.5 if not all, but more than half, of the fiber is visible and its location and
orientation are correct. Add each full or partial fiber to the total score, from the largest down to
the smallest visible, until a score of 0 or 0.5 is reached.
3.
After determining the last fiber to be counted, look at the overall background for artifacts.
If a fiber-like artifact appears anywhere in the wax insert area of the image, but not in an
appropriate location or orientation, deduct the artifactual fiber from the last real half or
whole fiber score if the artifactual fiber is equally or more apparent.
Deduct only from the last real fiber.
Record the final score after artifact deduction in the appropriate space on the chart.
21
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
4.
Use large fieldof-view magnification to assist in the visualization of specks. For hardcopy
images, use a 2X or higher magnifying glass. For softcopy images, use the workstations
electronic magnification function. Starting with the largest speck group, count each speck
group as one point if four or more of the six specks in the group are visible in the proper
locations. Count a speck group as 0.5 if two or three of the six specks in the group are visible
in the proper locations. Count a speck group as 0 if none or only one of the six specks in the
group is visible in the proper location.
Add each full or partial speck group to the total speck group score, from the largest down to
smallest visible group, until a score of 0 or 0.5 is reached.
5.
After determining the last speck group to be counted, look at the overall background for
artifacts. If noise or speck-like artifacts are visible in the wrong locations within the area of the
wax insert, and are as apparent as the real specks, deduct them one for one from the
individual specks counted in the last whole or half speck group scored, and adjust the score of
the last group appropriately. Record the final score after artifact deduction in the appropriate
space on the chart.
6.
Count each mass as one point if the minus density object is visible in the correct location, and
the mass appears to be generally circular against the background (i.e., greater than of the
perimeter or circumference is visible). A mass is counted as 0.5 point if the minus density object
is visible in the correct location, but the mass does not have a generally circular appearance.
Add each full or partial mass to the total mass score, from the largest mass down and until a
score of 0 to 0.5 is reached.
Record the raw mass score before artifact deduction.
7.
After determining the last mass to be counted, look at the overall background for artifacts.
If a mass-like artifact is seen in the wrong location within the area of the wax insert, deduct the
artifactual mass from only the last real whole or half mass scored if the artifactual mass is
equally or more apparent. Record the final score after artifact deduction in the appropriate
space on the chart.
8.
Using magnification, carefully examine the image for non-uniform areas, the presence of dirt or
dust artifacts, grid lines, processing artifacts, or any other artifacts and compare the image to
the original and previous images.
9.
Notice any artifacts or grid lines on the image. Investigate the source of any artifacts or grid
lines and record the events on the Maintenance/Corrective Action Log of the Phantom Control
Chart. The medical physicist can provide assistance in identifying the source of the artifacts.
NOTE
Mammography phantom images should always be viewed
22
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
If printed images are used for final interpretation: The optical density (OD) of the film shall be within
control limits of 0.20, from the established operating level.
The acrylic discs used for the measurement of the density difference (DD) may vary in thickness.
Consequently, the density difference is a relative, not absolute measurement and is to be used only
for quality control purposes. It is essential to use the same acrylic disc if comparisons are to be
made between different facilities.
The density difference (disc vs. adjacent density) shall not vary by more than 0.05 OD from the
established operating level. If a new operating level for background optical density is chosen, then a
new operating level for density difference must be established. It is essential that all printers used to
8th Edition
897N0602G
23
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
print mammography images at one facility produce similar film optical densities. It is not acceptable
to have one unit producing a film with an OD of 1.40 and another producing an OD of 1.80.
If softcopy images are used for final interpretation: The S value of the phantom image confirms the
exposure unit output and the FCRm/CRm reader sensitivity setting. The S value result must be plotted
on the SOFT COPY PHANTOM CONTROL/CNR CHART and must not vary by greater than 20%. If the
results exceed this control limit, make sure that the kVp and mAs are set properly and that the correct
menu is selected.
For both softcopy and hardcopy:
If the performance criteria for this test are not met, a second phantom image test should be
performed and evaluated. If the criteria are still not met, the reasons for this failure must be
investigated, corrective action taken, and the results documented before patients are examined with
this system. Any visual difference between the current phantom image and the saved original phantom
image should be investigated. Printer or FCRm/CRm reader unit artifacts, grid lines, or grid artifacts
should not be present since any of these may degrade clinical images.
The appearance of artifacts, grid lines, or grid artifacts; the number of masses, specks and fibers
visualized; or any other change in the visual appearance of the image should be reported
immediately to the medical physicist for further evaluation.
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken on the component(s) that caused the failure
before any further examinations are performed using the failed component(s). If the component(s) that
caused the failure (e.g. exposure unit or CR reader unit) is replaced by an alternative device and the
mammography system passes the re-test, image acquisition and interpretation may continue using
that combination of devices.
If the failure has been determined to be of a diagnostic device used for mammographic image
interpretation (e.g. laser printer, physicians review station) the source of the problem shall be
identified and corrective action shall be taken before that device can be used for mammographic
image interpretation. Clinical imaging may be continued. If available, an alternative approved
diagnostic device may be used for mammographic image interpretation
NOTE
Phantom images must be retained in the QC records for the last full year. (Softcopy images can be
locked on the QA and PACS workstation). The original baseline image must be retained until it is
necessary to establish a new baseline image.
24
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Cassette ID No.:
IP ID No.:
Year:
Month:
Date:
Initials:
AEC SETTING:
CNR Value:
CNR OK:
Density Difference
+0.05
+0.05
Baseline
-0.05
-0.05
Background Density
+0.20
+0.20
+0.15
+0.15
Baseline
-0.15
-0.15
-0.20
-0.20
No.Visible
Fibers
+0.5
Specks
Masses
+0.5
- 0.5
- 0.5
+0.5
+0.5
- 0.5
- 0.5
+0.5
+0.5
- 0.5
- 0.5
mAs (optional)
+15%
+15%
Baseline
-15%
8th Edition
897N0602G
-15%
25
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Cassette ID No.:
IP ID No.:
Year:
Month:
Date:
Initials:
AEC SETTING:
CNR Value:
CNR OK:
No.Visible
Fibers
+0.5
Specks
Masses
+0.5
- 0.5
- 0.5
+0.5
+0.5
- 0.5
- 0.5
+0.5
+0.5
- 0.5
- 0.5
S Value
+20%
+20%
Baseline
-20%
-20%
mAs (optional)
+15%
+15%
Baseline
-15%
26
-15%
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
PROBLEM
SIGNATURE
Prepared by
8th Edition
897N0602G
27
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test must be performed initially before interpreting mammograms. *Perform weekly for dry print
imagers, daily before printing mammographic images for wet print imagers. Perform after any
service or maintenance on the mammographic printer system and whenever problems are
suspected.
PROCEDURE
SET UP
1.
2.
Measure the image to identify the step which has an optical density closest to but not less than
1.20. Designate this step as the mid-density (MD) step.
3.
Measure the image to identify the step with an optical density closest to but not less than 2.20
and the step that has an optical density closest to but not less than 0.45. Designate these
steps as the high-density (HD) and low-density (LD) steps. The difference between these two
steps is designated as the density difference (DD).
4.
Measure the lightest (B+ F or unexposed) portion of the image. This is designated as B+F.
28
Print the QC pattern. Write the date and time on the film. Label the film with the printer ID if
there is more than one printer.
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
2.
3.
Plot the mid-density (MD), the density difference (DD) and the base-plus-fog (B+F) on the
control chart.
4.
Determine if any of the data points exceed the control limits. If not, go to step 6. If so, expose
and process a second QC strip, double checking that the correct procedure is followed. If the
same results are obtained, proceed to step 5.
5.
Circle any out-of-control data points and repeat the test. If any data point is still out of control,
correct the cause of the problem and repeat the test to confirm that the problem has been
corrected. Note the cause of the problem and the corrective action and plot the in-control
point.
6.
Determine if there are any trends, i.e., three or more data points moving in one direction (either
upwards or downwards), in the MD, DD, or B+F. If trends are present but the data points have
not, as yet, exceeded the control limits, clinical mammogram images can be printed. It is
necessary to determine the cause of the trend and to monitor the processor closely to assure
that the control limits are not exceeded.
7.
Retain the test film for at least the MQSA prescribed period, e.g., the last full month for daily
QC and the last 12 weeks for weekly QC.
8th Edition
897N0602G
29
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test must be performed initially before interpreting mammograms, weekly, after any service to
the monitor, and whenever problems are suspected.
PROCEDURE
1.
2.
For primary monitors only: record the maximum and minimum luminance levels as directed by
your monitor or QC software manufacturer. The maximum luminance (Lmax) is typically
specified by the monitor manufacturer as the highest value that can be used without
compromising other performance characteristics, such as lifetime or resolution. The minimum
brightness (Lmin) is influenced by Lmax.
3.
For both primary and secondary monitors: examine the image carefully to determine the
visibility of the following features:
a. Verify that the 5% signal level inset is visible in the larger 0% signal field.
b. Verify that the 95% signal inset is visible in the 100% signal field.
30
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
c. Verify that each gray level step from 0% to 100% can be distinguished individually.
d. Verify that the alphanumeric characters in the image appear sharp and in focus.
e. Verify that the high-contrast bar images in the center and at the corners of the QC pattern
are distinguishable. The low-contrast patterns are not evaluated in this test.
f. Record the results of the evaluations on your own form or one provided by the QC software
manufacturer.
PERFORMANCE
If the following conditions are met, the Monitor QC Weekly check is acceptable:
For primary monitors only: the maximum luminance value (Lmax) is within the range specified by
the monitor manufacturer. For reference, the AAPM report recommends that once set, the
target level (Lmax) is maintained within 10% of the desired value.
For both primary and secondary monitors: the 5% inset in the 0% field and the 95%- inset in the
100% field shall be visible and the 0% to 100% gray level steps shall be distinguishable from the
adjacent steps.
The alphanumeric characters appear sharp and focused.
The high contrast line-pair patterns are distinguishable at the center and the corners of the
display in both the horizontal and vertical orientations.
NOTE
The ability to distinguish the low-contrast line-pair patterns is not a requirement for this test.
Unless required by the monitor manufacturer, there is no action limit specification for Lmin in this
procedure
CORRECTIVE ACTION
For primary monitors (used for final interpretation). If this test produces results that fall outside the
action limits as specified by the manufacturer, the source of the problem shall be identified and
corrective action shall be taken before that device can be used for mammographic image
interpretation. Clinical imaging may be continued. If available, an alternative approved diagnostic
display device may be used for mammographic image interpretation, provided it has passed the
applicable QC tests.
For secondary monitors (used for image acquisition QC). If this test produces results that fall outside
the action limits as specified by the manufacturer, the source of the problem shall be identified and
corrective action shall be taken on the monitor that failed before any further examinations are
performed using that monitor. If the monitor that failed is replaced by an alternative monitor that
passes the test, image acquisition may resume using the alternative monitor.
8th Edition
897N0602G
31
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Room:
Left or Single Monitor
Serial Number:
Right Monitor Serial
Number:
Date:
Date:
Monitor
Luminance
(primary only)
LEFT
RIGHT
Monitor
MIN / MAX
MIN / MAX
Record values:
Lmax Pass/Fail:
0%-5%
contrast:
95%-100%
contrast:
Luminance
(primary only)
LEFT
RIGHT
MIN / MAX
MIN / MAX
Record values:
PASS
FAIL
PASS
FAIL
Lmax Pass/Fail:
0%-5%
contrast:
95%-100%
contrast:
PASS
FAIL
PASS
FAIL
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
Gray Steps:
YES
NO
YES
NO
Gray Steps:
YES
NO
YES
NO
Alphanumeric:
YES
NO
YES
NO
Alphanumeric:
YES
NO
YES
NO
Line-pair
images:
YES
NO
YES
NO
Line-pair
images:
YES
NO
YES
NO
SIGNATURE
SIGNATURE
Date:
Date:
Monitor
Luminance
(primary only)
LEFT
RIGHT
MIN / MAX
Monitor
MIN / MAX
Record values:
Lmax Pass/Fail:
FAIL
PASS
FAIL
YES
NO
YES
NO
YES
NO
YES
NO
Gray Steps:
YES
NO
YES
NO
Alphanumeric:
YES
NO
YES
Line-pair
images:
YES
NO
YES
SIGNATURE
32
(primary only)
LEFT
RIGHT
MIN / MAX
MIN / MAX
Record values:
PASS
0%-5%
contrast:
95%-100%
contrast:
Luminance
Lmax Pass/Fail:
PASS
FAIL
PASS
FAIL
YES
NO
YES
NO
YES
NO
YES
NO
Gray Steps:
YES
NO
YES
NO
NO
Alphanumeric:
YES
NO
YES
NO
NO
Line-pair
images:
YES
NO
YES
NO
0%-5%
contrast:
95%-100%
contrast:
SIGNATURE
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Section
Section
8th Edition
897N0602G
C
A
A
C
33
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test should be carried out monthly or after any service or maintenance on the mammographic
X-ray system.
PROCEDURE
1.
Review all of the items listed on the visual checklist and indicate the status. Be sure to rotate
the C-arm the way you would for patient imaging.
2.
34
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
VISUAL CHECKLIST
Facility:
Room#:
Unit:
Year:
Pass = 9
Fail = F
Does not apply = NA
Month
Day
Initials
C-Arm
SID indicator or marks
Angulation Indicator
Locks (all)
Field Light
High tension cable/other cables
Smoothness of Motion
Cassette Holder
Cassette Lock (small and large)
Compression Device
Compression Scale
Amount of Compression:
Automatic
Manual
Grid
Control Booth
Hand Switch Placement
Window
Panel switches/lights/meters
Technique charts
FCRm/CRm
Condition of cassettes
Condition of IPs
Other
Shields/Aprons/Gloves
Cones or collimators
Cleaning Solution
8th Edition
897N0602G
35
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
36
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Section
Section
8th Edition
897N0602G
D
A
A
D
37
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
Repeated images shall be evaluated quarterly.
In order for the repeat rates to be meaningful, a patient volume of at least 250 patients or 1,000
exposures is needed.
PROCEDURE
1.
Start by removing all existing reject images (hardcopy) in the department taken prior to the
start of the analysis.
2.
Take inventory of the image supply as a starting point to determine the total number of images
consumed during this test.
3.
Start collecting all rejected images. Continue to collect for the length of time needed to
radiograph at least 250 consecutive patients.
4.
Sort the rejected images into categories such as poor positioning, motion, compression, under
exposure, (these might be due to exposure or processing), artifacts (streaks, spots, etc.).
Good images (they appear to be acceptable mammograms when retrospectively evaluated
during the repeat analysis) may have also been repeated.
Some images may not have resulted in an additional exposure of the patient but may have also
been rejected. These include clear and QC images. Although it is appropriate to include wire
localization images as part of the reject analysis, they should not be included in the repeat
analysis because they are taken as part of the wire localization process.
NOTE
Rejected images are all images that are in the reject bin, including repeated images. Repeated images
are images that are retaken for inadequate quality. The reject bin does not include additional views
required to image selected tissue seen on the first image. It also does not include images taken for the
purposes of including tissue that could not be positioned on the image receptor due to the size of the
breast. For facilities using softcopy for final interpretation maintain a list of repeated images using the
REPEAT RATE ANALYSIS form included in this procedure.
38
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
5.
Some facilities place all images (repeated and good images) in the patients film jacket so there
are no repeated images in the department. In this case, the reject/repeat analysis chart should
be completed as patient examinations are carried out.
6.
Tabulate the counts from Steps 4 and 5, determining the total number of repeated images,
rejected images, and the total number of images exposed during the analysis period.
7.
Determine the overall percentage of repeated images by dividing the total number of repeated
images by the total number of images exposed during the analysis period, then multiply by
100. Next, determine the overall percentage of rejected images by dividing the total number of
rejected images by the total number of images exposed during the analysis period, and
multiply by 100.
8.
Determine the percentage of repeats in each reason for repeat category by dividing the
repeats in the category by the total number of repeated images and multiply by 100.
8th Edition
897N0602G
39
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Any corrective action should be recorded on the bottom of the repeat analysis form.
The effectiveness of the corrective actions must be assessed by performing another repeat analysis
after the corrective actions have been implemented.
It is important to study films that are too dark or too light to determine if the underlying cause is the
exposure equipment, image printer, patient positioning, technique or sub-optimal setting of digital
image processing.
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken within thirty days of the test date. Clinical
imaging and mammographic image interpretation may be continued during this period.
40
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FROM:
TO:
1824
2430
OTHER
TOTAL
PROJECTION
REASON FOR REJECT
CC
MLO
ML or LM
AXILLARY
OTHER
TOTAL
REPEATS
% of
REPEATS
POSITIONING
PATIENT MOTION
COMPRESSION
ARTIFACTS
X-RAY EQUIP
MALFUNCTION
SOFTWARE
MALFUNCTION
AEC
MISPLACEMENT
UNDER
EXPOSURE
OVER
EXPOSURE
INCORRECT
PATIENT ID
WASTE
SUB-TOTAL
GRAND TOTAL
REPEAT RATE =
REJECT RATE =
REPEATS
TOTAL FILMS INCLUDING REPEATS
ALL REJECT FILMS
TOTAL FILMS INCLUDING REPEATS
REPEAT RATE =
REJECT RATE =
SIGNATURE:
8th Edition
897N0602G
41
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
42
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Section
Section
8th Edition
897N0602G
EA
A
E
43
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test should be carried out initially, then semi-annually, as part of a Mammography Equipment
Evaluation (MEE), and whenever reduced compression is suspected.
PROCEDURE
1.
Place a towel on the cassette holder (to protect the cassette holder), then place the analog
bathroom scale on the towel, with the dial or read-out positioned for easy reading.
Locate the center of the scale directly under the compression device.
2.
Place several towels on top of the scale to prevent damage to the compression device.
3.
Using the power drive, activate the compression device and allow operation until it stops
automatically.
4.
5.
6.
Using the manual drive, move the compression device downward until it stops.
7.
8.
44
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
8th Edition
897N0602G
45
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
COMPRESSION
DATE:
SCALE I.D.:
MAMMOGRAPHIC UNIT
AUTOMATIC COMPRESSION
MANUAL COMPRESSION
lbs.
lbs.
lbs.
lbs.
lbs.
lbs.
lbs.
lbs.
ACCEPTANCE LIMITS
AUTOMATIC (25 - 45 lbs)
CORRECTIVE ACTION NEEDED?
PASS
FAIL
YES
NO
YES
NO
PASS
FAIL
Comments:
SIGNATURE:
46
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test should be performed initially before any patients are imaged and then semi-annually or
whenever the location of the storage space for the IPs relative to the X-ray exposure unit or the
radiation shielding is changed. In mobile mammography, it is not necessary to perform this test after
each relocation if the storage space for the IPs relative to the X-ray exposure unit or the radiation
shielding is not changed.
PROCEDURE
1.
Select an FCRm/CRm cassette from the exposure room stock. Erase the IP using the secondary
erasure mode on the FCRm/CRm image reader.
2.
Tape the coin to the front (exposure side) of the cassette and place the cassette in the
cassette storage area of the room with the coin facing the X-ray tube (i.e. the coin is between
the X-ray tube and the IP so that any X-ray exposure from the tube would cast a shadow on
the IP.)
3.
Leave the cassette in place for a minimum of 3 clinical studies, or produce 10 exposures of the
accreditation phantom to simulate clinical use.
4.
Remove the coin and process the IP in the FCRm/CRm reader unit using the Physics, IP Fog
menu.
5.
Examine the image to see if the coin is visible and record the results on the IP Fog report form.
8th Edition
897N0602G
47
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
IP FOG
DATE:
ACCEPTANCE LIMITS
COIN VISUALIZED?
CORRECTIVE ACTION NEEDED?
PASS
FAIL
YES
NO
YES
NO
PASS
FAIL
SIGNATURE:
48
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Section
49
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
When interpreting hardcopy FCRm/CRm images, it is suggested that you view the printed FCRm/CRm
images under the same conditions as you would screen-film mammography images. Film masking
devices and hot lights must be available for use by the interpreting physician. The 1999 ACR
Mammography Quality Control Manual includes a section on viewbox luminance, room illuminance,
and masking, which may be used as a reference.
When interpreting softcopy images, follow the monitor manufacturers quality control procedures.
For monitors without a manufacturer-provided QC program, follow the monitor QC procedures
provided in this manual.
8th Edition
897N0602G
50
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test must be performed initially upon installation before interpreting hard copy mammograms,
and after major repairs, such as those requiring a mammography equipment evaluation (MEE), and
then annually.
PROCEDURE
SET UP
1.
2.
Identify lines in each dimension that have a specified relationship (e.g. fixed length or distance
between lines). Use these same lines for subsequent testing.
3.
Measure the length of lines in the step above to establish a baseline. If the printer manufacturer
provides a test image with specified distances, compare your measurement to the provided
specification.
51
1.
2.
Using the ruler, confirm that the lines are straight and undistorted.
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
3.
4.
5.
Inspect the image for banding and jagged edges of the straight lines.
6.
8th Edition
897N0602G
52
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test must be performed initially before interpreting soft copy mammograms, annually, and after
major repairs, such as those requiring a mammography equipment evaluation (MEE), and whenever
problems are suspected.
8th Edition
897N0602G
53
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
PROCEDURE
MONITOR IMAGE QUALITY
1.
2.
Record the maximum and minimum luminance levels as directed by your monitor or QC
software manufacturer. The maximum luminance (Lmax) is typically specified by the monitor
manufacturer as the highest value that can be used without compromising other performance
characteristics, such as lifetime or resolution. The minimum brightness (Lmin) is influenced by
Lmax.
3.
Examine the image carefully to determine the visibility of the following features:
Verify that the 5% signal level inset is visible in the larger 0% signal field.
Verify that the 95% signal inset is visible in the 100% signal field.
Verify that each gray level step from 0% to 100% can be distinguished individually.
Verify that the alphanumeric characters in the image appear sharp and in focus.
Verify that the high-contrast bar images in the center and at the corners of the QC pattern
are distinguishable. The low-contrast patterns are not evaluated in this test.
Record the results of the evaluations on your own form or one provided by the QC software
manufacturer.
2.
Record the results of the evaluation on the Annual Monitor Evaluation report.
PERFORMANCE
If the following conditions are met, the Monitor QC (Annual) check is acceptable:
MONITOR IMAGE QUALITY:
The maximum luminance value (Lmax) is within the range specified by the monitor
manufacturer. For reference, the AAPM report recommends that once set, the target level
(Lmax) is maintained within 10% of the desired value.
The 5% inset in the 0% field and the 95%- inset in the 100% field shall be visible and the
0% to 100% gray level steps shall be distinguishable from the adjacent steps.
The alphanumeric characters appear sharp and focused.
54
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
The high contrast line-pair patterns are distinguishable at the center and the corners of the
display in both the horizontal and vertical orientations.
NOTE
The ability to distinguish the low-contrast line-pair patterns is not a requirement for this test.
Unless required by the monitor manufacturer, there is no action limit specification for Lmin in this
procedure.
CORRECTIVE ACTION
For primary monitors (used for final interpretation). If this test produces results that fall outside the
action limits as specified by the manufacturer, the source of the problem shall be identified and
corrective action shall be taken before that device can be used for mammographic image
interpretation. Clinical imaging may be continued. If available, an alternative approved diagnostic
display device may be used for mammographic image interpretation, provided it has passed the
applicable QC tests.
For secondary monitors (used for image acquisition QC). If this test produces results that fall outside
the action limits as specified by the manufacturer, the source of the problem shall be identified and
corrective action shall be taken on the monitor that failed before any further examinations are
performed using that monitor. If the monitor that failed is replaced by an alternative monitor that
passes the test, image acquisition may resume using the alternative monitor.
8th Edition
897N0602G
55
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Room:
Final Interpretation?
YES
NO
Workstation:
Cleared by FDA?
YES
NO
NA
YES
NO
Right Monitor
YES
NO
Lmax target:
Lmax measured:
(Lmax measured Lmax target)/Lmax target x 100%:
Within 10%:
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
LUMINANCE RESPONSE*
Left or Single Monitor
Right Monitor
* It is not necessary to measure Monitor Luminance Response on monitors that feature DICOM GSDF display
functionality and pass the Monitor Image Quality test.
SIGNATURE:
56
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test must be performed initially before any patients are imaged and then annually as part of
routine service PM (preventive maintenance), and whenever the FCRm/CRm reader unit PMT
(photomultiplier tube) calibration is suspect.
NOTE
This is a test of the FCRm/CRm reader unit and only needs to be conducted using one X-ray generator
per reader.
PROCEDURE
1.
Using the designated QC cassette, erase the IP using the secondary erasure mode on the
reader.
2.
Remove the compression paddle from the field and place the dose meter on the bucky surface
as described in illustration.
3.
Select 25kVp and a Mo target and filter combination. Adjust the mAs to give the dose meter
reading closest to but not less than 20 mR. Record the technique.
NOTE
If the mammography X-ray unit will not allow 20 mR exposure, you may use the lowest mR
exposure available that is greater than 20. Always use 25 kVp for this test.
Do not use any additional filter except the designated Mo filter. The use of another filter may
cause an error.
8th Edition
897N0602G
4.
Once you have established the proper technique for your X-ray generator, make a series of at
least three exposures prior to removing the meter or placing the cassette in the field. Record
the exposure technique and record the measured mR for each exposure.
5.
Verify that the mR readings are within 5% of each other. If they are not, repeat the test on a
more reproducible X-ray unit.
6.
7.
Remove the dose meter from the beam and place the erased Fuji IP / Cassette on the top of the
breast support surface centered in the X-ray field.
57
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
8.
Expose the cassette at the predetermined settings. Wait 10 minutes and process the IP using
the Physics, Sensitivity menu. Repeat this step two additional times creating a total of three
images. Record the S value for each exposure.
9.
Calculate the exposure average(mR) and S value average. Divide the averaged exposure
results by 20(mR) where 20(mR) is a fixed standard value. Multiply by the averaged S value to
obtain the corrected S value.
Example: 25 kVp
Averaged exposure = 30.0mR
Averaged S value = 83
Corrected S value = 30.0mR/20mR 83 = 124.5
58
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Measure the dose using a mammographic dose meter at the location shown below.
1824 cassette tube side view
5-1/4"
132mm
2-1/4
60mm
8th Edition
897N0602G
Chest wall
59
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
S VALUE CONFIRMATION
DATE:
Exposure Unit ID:
Reader Unit ID:
Dose Meter ID:
kVp:
mAs:
Exposure 1:
Exposure 2:
Exposure 3:
Average:
IP Exposure Time 1:
IP Read Time 1:
IP Exposure Time 2:
IP Read Time 2:
IP Exposure Time 3:
IP Read Time 3:
Yes
No
S value 1:
S value 2:
S value 3:
Average:
Corrected S value:
ACCEPTANCE LIMITS:
Corrected S value (must be between 96 and 144):
Corrective Action Needed:
PASS
FAIL
YES
NO
YES
NO
PASS
FAIL
Comments:
SIGNATURE:
60
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
Must be performed initially before any patients are imaged and then annually at routine PM and
whenever the system resolution is suspect.
PROCEDURE
1.
Using the designated cassette, erase the IP using the secondary erasure mode on the reader unit.
2.
3.
4.
Center the resolution pattern laterally, on top of the absorber and along the chest wall edge at
a slight (3 to 5 degree) angle.
In this position, the patterns bars lie parallel to the anode-cathode axis for the first image.
5.
Select the focal spot used for imaging an average breast and set a manual exposure based on
the kVp and mAs established for the phantom image test.
6.
7.
Repeat steps 1-6 with the bars oriented perpendicular to the anode-cathode axis and with the
highest line frequency of the bar phantom positioned against the chest wall at a slight (3 to 5
degree) angle.
8th Edition
897N0602G
61
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
2.
Note the highest frequency pattern where lines are distinctly visible throughout at least half of
the bar length and record.
62
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
SYSTEM RESOLUTION
DATE:
Exposure Unit ID:
Reader Unit ID:
ACCEPTANCE LIMITS
8 lp/mm 2 lpmm
CORRECTIVE ACTION NEEDED?
PASS
FAIL
YES
NO
YES
NO
PASS
FAIL
ACCEPTANCE LIMITS
8 lp/mm 2 lpmm
CORRECTIVE ACTION NEEDED?
PASS
FAIL
YES
NO
YES
NO
PASS
FAIL
SIGNATURE:
8th Edition
897N0602G
63
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
Must be performed initially before any patients are imaged and then annually at routine PM and
whenever the laser optics- related components are serviced.
PROCEDURE
1.
Using the designated cassette, erase the IP using the secondary erasure mode on the reader.
2.
Position the cassette for a non-grid exposure, centered so the X-ray beam covers as much of
the cassette front as possible.
3.
Place the steel rulers on the cassette surface so the rulers form the letter T.
Place the T into the X-ray field centered along the chest wall side of the cassette.
4.
Expose and process the cassette using the same exposure and image processing menu as
used in the S Value Confirmation test.
64
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
8th Edition
897N0602G
65
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
ACCEPTANCE LIMITS
WORKSTATION MONITOR
CORRECTIVE ACTION NEEDED?
LASER FILM PRINTER*
CORRECTIVE ACTION NEEDED?
PASS
FAIL
YES
NO
PASS
FAIL
YES
NO
YES
NO
PASS
FAIL
ACCEPTANCE LIMITS
WORKSTATION MONITOR
CORRECTIVE ACTION NEEDED?
LASER FILM PRINTER*
CORRECTIVE ACTION NEEDED?
PASS
FAIL
YES
NO
PASS
FAIL
YES
NO
YES
NO
PASS
FAIL
SIGNATURE:
66
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test must be performed at installation of the mammography exposure unit and on an annual
basis and as part of a Mammography Equipment Evaluation (MEE).
PROCEDURE
1.
Verify that the freestanding dedicated mammography unit is mechanically stable under normal
operating conditions.
2.
Verify that all moving parts move smoothly, without undue friction, that cushions or bumpers
limit the range of available motions, and that no obstructions hinder the full range of motions
within these limits.
3.
Set and test each lock detent independently to ensure that mechanical motion is prevented
when the lock or detent is set.
4.
Verify that the image receptor holder assembly is free from wobble or vibration during normal
operation.
5.
Verify that the image receptor slides smoothly into the proper position in the image receptor
holder assembly and that the image receptor is held in place securely by the image receptor
compartment for any orientation of the image receptor holder assembly.
6.
If provided, verify that the compressed breast thickness scale (analog or digital) is accurate
to within 0.5 cm under conditions of moderate compression (15 - 20 lbs.) and reproducible
to within 2mm between 1 and 8 cm. The phantom used for this test should be large enough
to simulate a typical breast and be positioned towards the chest-wall side of the bucky. The
compressed breast thickness should be measured at the center of the chest-wall position of
the automatic exposure control (AEC) sensor. This should be done for at least the small and
large image receptors and compression paddles.
7.
Verify that in normal operation, the patient and operator are not exposed to sharp or
rough edges or other hazards including electrical hazards.
8.
Perform the decompression test as described in MQSA final regulations and record the result
on the Mammographic Unit Assembly Evaluation Chart.
9.
Verify that the operator is protected by adequate radiation shielding during exposure.
8th Edition
897N0602G
67
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
68
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Room#:
X-ray Generator:
Pass =
Fail = F
Does not apply = NA
1.
2.
3.
4.
5.
6.
7.
8.
9.
SIGNATURE:
8th Edition
897N0602G
69
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
OBJECTIVE
To assure that the collimator or cone allows full coverage of the image receptor by the X-ray field
but does not allow significant radiation beyond the edges of the image receptor. To assure that the
chest wall edge of the compression paddle aligns with the chest wall edge of the image receptor
without showing on the image.
FREQUENCY
This test must be performed upon installation or major service of the X-ray tube or collimator, when
a problem related to collimation is suspected, and annualy thereafter.
PROCEDURE
70
1.
Place an appropriately sized cassette loaded with an erased IP in the normal orientation in the
image receptor holder.
2.
3.
Place the large cassette on top of the image receptor holder with the front of the cassette
toward the X- ray source and assure that the large cassette extends beyond the image
receptor holder on the chest-wall side by about 1 cm.
4.
5.
Remove the compression paddle. (The compression paddle should be removed before
placement of the coins to assure a sharp demarcation at the edges of the light field.)
6.
Turn on the collimator light and place the four identical smaller coins inside the light field with
one edge of each coin just touching the edge of the light field. The coin on the chest-wall side
should be shifted to the right of center about 2 inches so it does not superimpose the AEC
detector.
7.
Replace the compression paddle and position it 4.2 cm from the breast support.
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
8.
Tape the larger coin underneath the compression paddle shifted about 2 inches to the left so it
does not superimpose the AEC detector. Be sure the coin's outer edge is tangent to the inner
lip of the chest-wall side of the compression paddle. This coin marks the chest-wall edge of
the paddle.
9.
Place a sheet of acrylic attenuating material on top of the paddle, so that all radiation reaching
the cassettes must pass through the attenuator.
10. Make an exposure using AEC at the lowest setting on the X-ray generator and process the IPs
using the Physics, Collimation menu.
11. Repeat steps 1 through 9 for all routinely used collimator/bucky/compression paddle
combinations and target materials. When testing the large image receptor, the top cassette
may be positioned diagonally to capture all four edges of the X-ray field, or two large cassettes
may be used on top of the image receptor holder.
Large cassette
(on the image receptor holder)
Coin
Illuminated area by
collimator light
Coin
Coin
Coin
Chest Wall Side (Top view)
Comperssion paddle
Large coin
8th Edition
897N0602G
71
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Attenuating material
Compression paddle
Coin
Large cassette
Coin
Coin
Coin
72
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Small
coin
Exposure area
Small
coin
Small
coin
Small
coin
Large
coin
Small
coin
Small
coin
Collimated
exposure
area
Large
coin
Small
coin
Small
coin
8th Edition
897N0602G
73
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
74
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
COLLIMATION ASSESSMENT
DATE:
Exposure Unit ID:
Source to Image Receptor Distance (SID) in mm:
Deviation between X-ray field and light field:
Collimator
Target Material
Left Edge deviation (mm):
Right Edge deviation (mm):
Sum of left and right edge deviations:
Sum as % of SID
Anterior Edge deviation (mm)
Chest Edge deviation (mm)
Sum of anterior and chest wall deviations (mm)
Sum as % of SID
Action Limit:
18 24 cm
24 30 cm
If a light field that passes through the X-ray beam limitation device is provided, it shall be aligned with the X-ray field so that the
total of any misalignment of the edges of the light field and the X-ray field along either the length or the width of the visually
defined field at the plane of the breast support surface shall not exceed 2% of the SID.
18 24 cm
24 30 cm
Yes / No
Yes / No
All systems shall have beam-limiting devices that allow the entire chest wall edge of the X-ray field to extend to the chest wall
edge of the image receptor and provide means to assure that the X-ray field does not extend beyond any edge of the receptor
by more than 2% of the SID.
18 24 cm
24 30 cm
Yes / No
Yes / No
Target Material
Difference between compression
paddle edge and image edge
Difference as % of SID
Paddle Edge Visible on Image Receptor?
Action Limit:
If the chest wall edge of compression paddle is within the image receptor or projects beyond the chest wall edge of the image
receptor by more than 1% of SID, seek service correction.
JUDGEMENT:
PASS
FAIL
Comments:
SIGNATURE:
8th Edition
897N0602G
75
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test must be performed prior to the first clinical use with FCRm/CRm on each mammographic
exposure unit, annually thereafter, and as part of a mammography equipment evaluation (MEE).
PROCEDURE
AEC DENSITY CONTROL FUNCTION
76
1.
Select a loaded 18x24cm mammography cassette from the clinical inventory. Erase the IP
using the secondary erasure mode on the image reader. Place the cassette in the bucky.
2.
Position a 4cm thick absorber on the patient support. Lower the compression device to
contact with the absorber. Make sure that the absorber completely covers the AEC detector
area clinically used for a 4cm thick compressed breast.
3.
Select the kVp used to image the ACR MAP phantom and record the value on the chart
labeled DENSITY CONTROL FUNCTION.
4.
Expose the 4 cm absorber at the AEC systems 0 density control setting. Do not process the
IP in the image reader.
5.
Record the mAs delivered and the AEC density control setting on the chart.
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
6.
Repeat steps 4-5 for the other density control settings used clinically.
7.
Erase the IP (which has been exposed multiple times without processing and erasure) using the
primary erase mode on the FCRm/CRm reader unit.
Select a loaded 18x24cm mammography cassette from the clinical inventory; erase the IP using
the secondary erasure mode of the FCRm/CRm reader unit.
2.
3.
Place the 4cm thick absorber in the X- ray field, lower the compression device into contact
with the absorber.
4.
For systems equipped with multiple positions for the AEC sensor, select the position used
most often clinically. Make sure that the absorber completely covers the AEC sensor position
most often used clinically.
5.
Prepare the mammography X-ray unit for operation in the AEC mode, selecting the density
control setting that is most often used clinically (typically 0).
6.
Position the ion chamber in the X-ray beam without covering the AEC sensitive area.
7.
Expose the cassette and IP using a fixed kVp and target/filter combination typically used for a
4cm breast.
8.
Record the resultant mAs and the measured air kerma of each exposure on the data form
titled REPRODUCIBILITY CHART.
9.
Repeat steps 6 and 7 three more times, completing the AEC REPRODUCIBILITY test.
10. Erase the exposed IP using the Primary Erasure mode of the FCRm/CRm reader unit.
11. Using AEC, expose the 4cm thick absorber for each imaging mode that is used clinically on
that X-ray unit (e.g. small image receptor and grid, large image receptor and grid, magnification
mode with small focal spot, and no grid). It is not necessary to measure or record air kerma
during the IMAGE MODE TEST.
12. Enter the results of each exposure on the IMAGE MODE TRACKING form.
13. Erase all exposed IPs using the Primary Erasure mode of the FCRm/CRm reader unit.
CNR PER OBJECT THICKNESS
8th Edition
897N0602G
1.
Using the designated 18x24cm QC mammography cassette and IP, erase the IP using the FCRm/
CRm reader unit's secondary erasure mode. Place the loaded cassette in the 18x24cm bucky.
2.
Place the 2cm thick 24x30cm acrylic absorber on the patient support. The absorber may
extend beyond the chest wall side of the 18x24cm patient support.
3.
4.
Move the AEC sensor to its most forward position (closest to the chest wall).
77
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
5.
Place the 0.2mm Al sheet on the top of the compression device (not directly on top of the
absorber) and position it as shown in the diagram. Use the field indicator light as a guide for
the center of the bucky and to identify the AEC sensor positions. Position the Al sheet as
shown in the diagram so that the Al covers the indicated Calculation Area while positioned as
far as possible from the most forward AEC sensor position.
Calculation Areas
(20mm 20mm)
Aluminum Object
20mm
BG
Al
20mm
60mm
20mm
Acrylic Slabs
Make an AEC exposure using the AEC MODE and kVp normally used clinically for a 2cm thick
compressed breast.
7.
After the predetermined interval, process the IP using the Physics Sensitivity menu.
NOTE
To control the effect of image fading with time on this test, it is important to control the time
interval between X-ray exposure and reading of the IP so that the interval be consistent whenever
this test is performed. The interval may be as short as possible consistent with the testing
workflow. We recommend trying an interval between 5 and 10 minutes, adjusting as necessary,
and then using the same interval consistently.
8.
Calculate CNR of the image using the QC calculation tool function on the CR Console/AWS-c,
and record the kVp, target, filter, mAs, and CNR value on the CNR PER OBJECT THICKNESS
data form.
9.
Repeat steps 6 through 8 using 4cm and 6cm thick 24x30 acrylic absorbers using the kVp
and AEC exposure settings normally used clinically for 4 and 6cm compressed breasts
respectively.
10. Assigning the CNR value from the 4cm exposure a value of 100%, calculate the relative CNR
percentage values for 2cm and 6cm and record them on the data form.
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
8th Edition
897N0602G
79
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Focal Spot:
Phantom thickness:
4cm
kVp:
AEC Density
Control Setting
+5
% mAs
change
mAs
Acceptable level
PASS / FAIL
+4
PASS / FAIL
+3
PASS / FAIL
+2
PASS / FAIL
+1
PASS / FAIL
0 (Normal)
0%
PASS / FAIL
-1
PASS / FAIL
-2
PASS / FAIL
-3
PASS / FAIL
-4
PASS / FAIL
-5
PASS / FAIL
REPRODUCIBILITY CHART
Imaging mode:
Focal Spot:
Phantom thickness:
4cm
kVp:
Exposure Number
Air kerma
mAs
Acceptance
Action Limit
1
2
3
4
Mean values
Standard
Deviations (SD)
Coefficients of
variation (CV)
PASS / FAIL
80
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
DATE:
Small Cassette ID:
AEC MODE
kVp
Target
Filter
mAs
Comments
18 x 24 Bucky
24 x 30 Bucky
Mag / No Grid
CNR
kVp
Target
Filter
mAs
CNR Measured
(relative to
4cm PMMA)
2 cm
Acceptable
level
>100%
4 cm
100%
6 cm
100%
>75%
PASS / FAIL
SIGNATURE:
8th Edition
897N0602G
81
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test must be performed at installation and annually and whenever artifact evaluation is needed.
PROCEDURE
82
1.
Use technique factors normally used clinically, choosing the lowest kVp setting used clinically (to
be most sensitive to artifacts). If AEC is normally used, use it in this test. Start with the most
commonly used image receptor size (usually 18 24 cm). Record the technique factors on the
data form.
2.
Place a uniform sheet of acrylic that is large enough to cover the mammographic cassette and
thick enough to have an exposure time of 0.5 second or greater on the image receptor holder
assembly.
3.
Erase the IP using Primary Erase and insert the cassette into the cassette holder.
4.
Position a lead marker such as the number 1 or an arrow on the acrylic sheet in a corner of
the radiation field, preferably outside the normal location of the breast, and pointing along the
long axis of the IP and cassette.
5.
Make an exposure.
6.
Process the IP using the Physics, Artifacts menu. Record all pertinent information on the data
form for this test.
7.
Measure the optical density in the center of the printed film, verifying that it is greater than 1.20,
and record on the data form for this test.
8.
Repeat this process for the other image receptor size. Also, repeat for all available focal spot
sizes, targets and filters used clinically.
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
8th Edition
897N0602G
If significant artifacts are noted, contact the person maintaining that system component. Not all
artifacts can be totally eliminated. It may be useful to use the concept of ALARA (as low as
reasonably achievable) when attacking artifacts. If they can be easily eliminated, they should be.
If the artifact is difficult or expensive to eliminate and is subtle (not mimicking or obscuring clinical
information), it may be tolerable. The medical physicist should consult with the interpreting physician
as to whether the artifact is tolerable. Tolerances for artifacts should be lower with new imaging
equipment. If this test produces results that fall outside the action limits as specified, the source of
the problem shall be identified and corrective action shall be taken within thirty days of the test date.
Clinical imaging and mammographic image interpretation may be continued during this period.
83
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Exposure Conditions
Cassette - IP Number
IP Size
mAs or AEC setting(s)
Optical Density
Target/Filter Type
Focal Spot Size Large or Small
Magnification Factor Used
PASS
YES
FAIL
NO
YES
NO
PASS
FAIL
PASS
YES
FAIL
NO
YES
NO
PASS
FAIL
PASS
YES
FAIL
NO
YES
NO
PASS
FAIL
Printer Artifact
Acceptable?
CORRECTIVE ACTION NEEDED?
CORRECTIVE ACTION DESCRIPTION:
CORRECTIVE ACTION TAKEN?
CORRECTIVE DATE:
RESULTS OF NEW TEST:
If significant artifacts are visible, contact the appropriate person maintaining or servicing the digital mammography equipment, printer or X-ray
generator.
84
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test must be carried out initially, at installation, after appropriate calibration of the equipment to
establish a baseline level. Subsequently, it must be carried out at least annually to compare with the
previous test image.
ACR MAPP
Acrylic 4mm x 1cm diameter disc (for hardcopy QC only)
Designated cassette and Imaging Plate used for mammography
Image Quality Evaluation Control chart
Watch/Timer
Previous Image
PROCEDURE
1.
Using the designated cassette, erase the IP using the secondary erasure mode on the reader.
2.
3.
Place the phantom on the cassette holder and position it so the edge of the phantom fully
covers and is aligned with the chest wall side of the image receptor. Center the phantom, left
to right.
4.
The 4mm acrylic disc is not needed for softcopy QC. If softcopy is used for final interpretation,
proceed to step 6.
5.
For hardcopy QC, secure the acrylic disc to the top of the phantom within the image area but
positioned so it will not obscure details in the phantom. A suitable location is between and
slightly below the first and second largest fibers. Once located, the disc may be permanently
attached to the phantom with super glue.
6.
7.
Verify the location of the AEC detector. It should be in the same location used for previous
phantom images and should be completely covered by the phantom.
8.
Make an exposure using the AEC technique factors (target, filter, kVp, grid, density control
setting, etc.) currently used clinically for a 4.2-cm compressed breast of average density.
.
8th Edition
897N0602G
85
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
9.
Record the mAs on the control chart after making the exposure.
10. After exposure, ensure that a predetermined interval time between 5 and 10 minutes elapses
prior to the reading process. Process the IP using the "Physics, ACR MAPP" menu.
NOTE
It is important to control the time interval between exposure and reading of the IP. To minimize
variability, it is recommended that and interval be chosen somewhere between 5 and 10 minutes
and that this same interval be used consistently.
11. Record the AEC density control setting on the control chart.
12. If printed images are used for final interpretation: Measure the following densities. The
background density should be measured in the geometric center of the phantom image. The
DD measurement is the density inside the acrylic disc subtracted from the density directly
adjacent to the disc, perpendicular to the anode-cathode axis. For consistent results, these
measurements must be made at the same location each time the background optical density
and the DD on the control chart.
If softcopy images are used for final interpretation: Record the S value from the phantom
exposure image on the background density line of the control chart.
VIEWING CONDITIONS
Phantom images should be read under controlled viewing conditions. General lighting should be at
a low level and diffused. Viewboxes or workstations should be positioned to avoid light from
windows, other viewboxes, and other sources of bright light, either direct or reflected. Images
should be masked to eliminate extraneous light. Use large fieldof-view magnification to assist in the
visualization of specks as well as other test objects. For hardcopy images, use a 2X or higher
magnifying glass. For softcopy images, use the workstations electronic magnification function.
86
1.
When scoring the phantom image of the approved phantom, each object is scored separately.
Always count the number of visible objects from the largest object of a given type (i.e., fiber,
speck group or mass) downward until a score of 0 or 0.5 is reached, then stop counting for
that object type.
2.
Count each fiber as one point if the full length of the fiber is visible and the location and
orientation for the fiber are correct. Count a fiber as 0.5 if not all, but more than half, of the fiber
is visible and its location and orientation are correct. Add each full or partial fiber to the total
score, from the largest down to the smallest visible, until a score of 0 or 0.5 is reached.
3.
After determining the last fiber to be counted, look at the overall background for artifacts. If a
fiber-like artifact appears anywhere in the wax insert area of the image, but not in an
appropriate location or orientation, deduct the artifactual fiber from the last real half or
whole fiber score if the artifactual fiber is equally or more apparent. Deduct only from the last
real fiber, not from additional deduction. Record the final score after artifact deduction in the
appropriate space on the chart.
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
4.
Use large fieldof-view magnification to assist in the visualization of specks. For hardcopy
images, use a 2X or higher magnifying glass. For softcopy images, use the workstations
electronic magnification function. Starting with the largest speck group, count each speck
group as one point if four or more of the six specks in the group are visible in the proper
locations. Count a speck group as 0.5 if two or three of the six specks in the group are visible
in the proper locations. Count a speck group as 0 if none or only one of the six specks in the
group is visible in the proper location.
Add each full or partial speck group to the total speck group score, from the largest down to
smallest visible group, until a score of 0 or 0.5 is reached.
5.
After determining the last speck group to be counted, look at the overall background for
artifacts. If noise or speck-like artifacts are visible in the wrong locations within the area of the
wax insert, and are as apparent as the real specks, deduct them one for one from the
individual specks counted in the last whole or half speck group scored, and adjust the score of
the last group appropriately. Record the final score after artifact deduction in the appropriate
space on the chart.
6.
Count each mass as one point if the minus density object is visible in the correct location, and
the mass appears to be generally circular against the background (i.e., greater than of the
perimeter is visible). A mass is counted as 0.5 point if the minus density object is visible in the
correct location, but the mass does not have a generally circular appearance. Add each full or
partial mass to the total mass score, from the largest mass down and until a score of 0 to 0.5
is reached. Record the raw mass score before artifact deduction.
7.
After determining the last mass to be counted, look at the overall background for artifacts. If a
mass-like artifact is seen in the wrong location within the area of the wax insert, deduct the
artifactual mass from only the last real whole or half mass scored if the artifactual mass is
equally or more apparent. Record the final score after artifact deduction on the appropriate
space on the chart.
8.
Using magnification, carefully examine the image for non-uniform areas, the presence of dirt or
dust artifacts, grid lines, processing artifacts, or any other artifacts and compare the film image
to the original and previous images.
9.
Circle any artifacts or grid lines on the film image. Investigate the source of any artifacts or grid
lines.
NOTE
8th Edition
897N0602G
87
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
The mAs noted on the generator read-out typically does not change by more than 15%. If your
measurements exceed this amount, it may be due to a change in kVp, mAs, target material, filter
material, generator calibration, or AEC calibration since the previous image.
If printed images are used for final interpretation:
The optical density (OD) of the film shall be within control limits of 0.20, from the established
operating level.
The acrylic discs used to produce an image for the measurement of the density difference may vary
in thickness. Consequently, the density difference is a relative, not absolute measurement and is to
be used only for quality control purposes. It is essential to use the same acrylic disc if comparisons
are to be made between different facilities.
The density difference (disc vs. adjacent density) shall not vary by more than 0.05 from the
established operating level. If a new operating level for background optical density is chosen, then a
new operating level for density difference must be established. It is essential that all printers used to
print mammography images at one facility produce similar film optical densities. It is not acceptable
to have one unit producing a film with an OD of 1.40 and another producing an OD of 1.80.
88
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
NOTE
Phantom Images should be retained in the QC records for the last full year (softcopy images can
be locked on the QA and PACS workstation). The original baseline image should be retained
until it is necessary to establish a new baseline.
8th Edition
897N0602G
89
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Current
Image
Comments
kVp setting
Density control
Phototimed mAs
mAs change
% mAs change = mAs change / mAs 100
S value
S value Change
*Background density
*Background density change
*Density outside disc
*Density inside disc
*Density diff = outside-inside
*Density difference change
Number of fibers seen
Fibers seen after deduction
Fiber change
Number of speck groups seen
Speck groups after deduction
Speck groups change
Number of masses seen
Masses seen after deduction
Mass change
* Hardcopy for final interpretation only.
JUDGEMENT
Comments:
PASS / FAIL
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
SIGNATURE:
90
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
At installation and then annually, and when decrease in the dynamic range is suspected.
PROCEDURE
NOTE
This is an evaluation of the CR reader unit. Select one exposure unit for this evaluation.
8th Edition
897N0602G
1.
Using the dedicated mammography cassette, erase the IP using the secondary erasure mode
on the reader.
2.
3.
Place the 2 cm and an additional 4 cm thickness of acrylic attenuating material on the surface
of the Bucky tray as shown in Figure 1.
4.
Bring the compression device into contact with the phantom attenuators.
5.
Verify the location of the AEC detector. It should be completely covered by the 2 cm and 4 cm
thicknesses of acrylic.
6.
Make an exposure using the AEC technique factors (target, filter, kVp, grid, density control
setting, etc.) currently used clinically for a 6cm compressed breast of 50/50 composition.
7.
8.
9.
View the image on the display (monitor or hardcopy) used for final interpretation, and record
your findings on the Dynamic Range report form.
91
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
3cm 3cm
bucky
tray
Cassette
anterior side
Side view
Figure 1. Test Object Arrangement
92
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Dynamic Range
DATE:
Target
Filter
kVp
mAs
Exposure conditions
PASS
FAIL
YES
NO
YES
NO
PASS
FAIL
SIGNATURE:
8th Edition
897N0602G
93
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
At installation and then annually and when incomplete erasure is suspected.
PROCEDURE
NOTE
This is an evaluation of the CR image reader. Select one exposure unit for this evaluation.
94
1.
2.
3.
Place the ACR MAPP on the patient support and position it so the edge of the phantom is
aligned with the chest wall side of the image receptor. Center the phantom, left to right.
4.
5.
Verify the location of the AEC detector. It should be in the same location used for the Phantom
Image test and should be completely covered by the phantom.
6.
Make an exposure using the AEC technique factors (target, filter, kVp, grid, density control
setting, etc.) currently used clinically for a 4.2 cm compressed breast of 50/50 composition.
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
7.
8.
9.
Wait one minute and re-process the same IP using the "Physics, ACR MAPP" menu without
making additional exposure.
10. On the new image, change the S value to 10 times that of the original S value recorded during
Step 8 using the QA function of the CR Console/AWS-c.
11. Inspect the image for the visibility of the outline of the ACR MAPP.
Multiplicative Lag test
12. Repeat steps 2 through 7 using the same cassette as in the first test.
13. Disregard the image produced. Remove the returned cassette and IP from the reader and
insert it into the bucky tray.
14. Place a uniform 4 cm thick acrylic plate that is large enough to cover the mammographic entire
cassette on top of the bucky tray.
15. Make an exposure using the AEC technique factors (target, filter, kVp, grid, density control
setting, etc.) currently used clinically for a 4 cm compressed breast of 50/50 composition.
16. Process the IP using the "Physics, ACR MAPP" menu.
17. Inspect the image for the visibility of the outline of an ACR phantom.
8th Edition
897N0602G
95
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Target
Filter
kVp
mAs
kVp
mAs
ACR MAPP
S value of the ACR MAPP
S value applied to non-exposure image
10
About 10
Artifact evaluation
Acceptable?
CORRECTIVE ACTION NEEDED?
PASS
FAIL
YES
NO
YES
NO
PASS
FAIL
Target
Filter
ACR MAPP
A uniform 4 cm sheet image
Artifact evaluation
Acceptable?
CORRECTIVE ACTION NEEDED?
PASS
FAIL
YES
NO
YES
NO
PASS
FAIL
The mAs noted on the generator read-out must not change by more than 15% from the previous test.
SIGNATURE:
96
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
At installation, annually, and before introducing new imaging plates or cassettes.
NOTE
The inter-plate consistency test must be performed on new imaging plates and imaging plate
cassettes prior to clinical use. When adding or replacing imaging plates and cassettes of the
same type already in use between annual physicists surveys, the QC technologist may perform
this test on the new imaging plates and cassettes under the oversight of the medical physicist
prior to their first clinical use, recording the results on the inter-plate consistency data form for
review by the medical physicist during the next annual survey. These new imaging plates and
cassettes must meet the criteria established for the group to which they will be added. If the new
imaging plates and cassettes do not fit the criteria for an existing group and require more
exposure, then the testing must be performed by the medical physicist.
PROCEDURE
1.
Select a cassette used for mammography, and erase the IP using the secondary erasure mode
of the reader.
2.
3.
Place the 4-cm thick acrylic attenuating material (phantom) on the patient support surface.
4.
5.
Make an AEC exposure in the auto mAs mode. Fix the other technique factors (target, kVp,
filter, grid, density control setting, etc.) used clinically.
6.
Wait a predetermined time interval and process the IP using the "Physics Sensitivity" menu.
This menu should be set in the Semi mode and all image should be set to zero (MRE, MDE,
PRE).
NOTE
To reduce the influence of IP image fading characteristics on this test, it is important to control
the time interval between X-ray exposure and reading of the IP. It is important that the interval be
consistent whenever this test is performed. Choose an interval between 5 and 10 minutes, and
then use the same interval consistently.
8th Edition
897N0602G
97
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
7.
Record the mAs of the test exposure on the Inter-Plate Consistency data form.
8.
Repeat the procedure from Step1 to Step 7 for all IPs of the same type and size to be
evaluated. All IPs of the same type and size are considered a group.
9.
10. Calculate the SNR of all images using the QC calculation tool.
11. Record the SNR result on the Inter-Plate Consistency data form.
12. Before using new imaging plates clinically, compare the consistency of the new imaging plates
against the inventory of all imaging plates of the same group by performing steps 1 through 11.
60mm
20mm
20mm
98
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
INTER-PLATE CONSISTENCY
DATE:
Exposure Unit ID:
CR Reader Unit ID:
AEC-mode
Target
Filter
kVp
Exposure conditions
Group:
Cassette ID
mAs
Average
Value
Limiting
Value
SNR
Average
Value
Limiting
Value
Acceptable?
PASS / FAIL
PASS / FAIL
PASS / FAIL
to
to
PASS / FAIL
PASS / FAIL
PASS / FAIL
PASS / FAIL
PASS / FAIL
(10%)
(15%)
PASS / FAIL
PASS / FAIL
Group:
Cassette ID
mAs
Average
Value
Limiting
Value
SNR
Average
Value
Limiting
Value
Acceptable?
PASS / FAIL
PASS / FAIL
PASS / FAIL
to
to
PASS / FAIL
PASS / FAIL
PASS / FAIL
PASS / FAIL
PASS / FAIL
(10%)
(15%)
PASS / FAIL
PASS / FAIL
SIGNATURE:
8th Edition
897N0602G
99
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test must be performed annually.
100
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test must be performed at installation and annually.
8th Edition
897N0602G
101
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test must be performed annually, after service to the X-ray machine which may effect beam
quality, and as part of an mammography equipment evaluation (MEE).
Minimum HVL
(millimeters of aluminum)
0.20
0.25
0.30
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken within thirty days of the test date. Clinical
imaging and mammographic image interpretation may be continued during this period.
KV
25
28
31
Mo/Mo
0.330.2
0.360.2
0.390.2
W/Rh
0.520.3
0.540.3
0.560.3
102
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
FREQUENCY
This test must be performed annually and upon Mammography Equipment Evaluation (MEE) of the
X-ray system.
8th Edition
897N0602G
103
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
104
8th Edition
897N0602G
Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual
Revision History
Revision
Issue Date
Reason
07.2006
New Release
10.2006
02.2007
12.2009
01.2012
05.2013
Added Soft Copy Phantom Control / CNR Chart to page 26. This
chart was missing from page 26 in the 01.2012 release.
8th Edition
897N0602G
Never released.
1) The CNR figures on pages 35 and 96 have been
changed to the same as those in 897N100001A.
2) Revised wording for annual Monitor QC test.
3) Changes have been made along with the release of
the AWS-c.
105