0% found this document useful (0 votes)
417 views106 pages

897N0602G AspireCRm QC Manual PDF

AspireCRm_QC_Manual

Uploaded by

Jose Quisca
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
417 views106 pages

897N0602G AspireCRm QC Manual PDF

AspireCRm_QC_Manual

Uploaded by

Jose Quisca
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 106

Aspire FCRm/CRm

Fujifilm CR for Mammography


Quality Control Manual
8th Edition: May 2013

Fujifilm Medical Systems U.S.A., Inc.


897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

CONTENTS
FOREWORD ................................................................................................................................. 5
Introduction to Measurements ............................................................................................. 6
S value ............................................................................................................................................ 7
Overview of Testing ................................................................................................................... 8
Section A
0

Section B

Getting Started .................................................................................................. 9


Set-up & Baseline Images ........................................................................................................10

Weekly (QC Technologist) ............................................................................15

1
2

CNR ..................................................................................................................................................16
Phantom Image* ........................................................................................................................20

3
4

Printer QC ......................................................................................................................................28
Monitor QC ...................................................................................................................................30

Section C
5

Section D
6

Section E
7
8

Section F
9

Monthly (QC Technologist) ..........................................................................33


Visual Checklist* .........................................................................................................................34

Quarterly (QC Technologist) ........................................................................37


Repeat Analysis* .........................................................................................................................38

Semi-Annual (QC Technologist) .................................................................43


Compression* ..............................................................................................................................44
Imaging Plate (IP) Fog ..............................................................................................................47

Annual (Medical Physicist) ...........................................................................49


Viewing and Viewing Conditions .........................................................................................50

10 Printer QC ......................................................................................................................................51
11 Monitor QC ...................................................................................................................................53
12 S Value Confirmation ................................................................................................................57
13 System Resolution .....................................................................................................................61
14 CR Reader Scanner Performance..........................................................................................64
15 Mammographic Unit Assembly Evaluation* ....................................................................67
16 Collimation Assessment* ........................................................................................................70
17 Automatic Exposure Control (AEC) System Performance Assessment...................76
18 System Artifact Evaluation*....................................................................................................82
19 Image Quality Evaluation* ......................................................................................................85
20 Dynamic Range...........................................................................................................................91
21 Primary Erasure (Additive and Multiplicative Lag Effects)...........................................94
(*) Portions of this test were reprinted from the ACR Mammography Quality Control Manual.
American College of Radiology; 1999

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

22 Inter-Plate Consistency ............................................................................................................97


23 kVp Accuracy and Reproducibility .................................................................................... 100
24 Dose ................................................................................................................................................ 101
25 Beam Quality Assessment and Half-Value Layer Measurement ................................ 102
26 Radiation Output .................................................................................................................... 103

Revision History ..................................................................................................................... 105

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

FOREWORD
Quality assurance (QA) in mammography means the planned and organized actions designed to
provide confidence that mammographic equipment and related components operated in a facility
will reliably and consistently produce diagnostic quality mammograms for the screening and
diagnosis of breast cancer with minimum dose to patients and staff.
This document, FCRm/CRm Quality Control for Mammography Operation Manual, describes how to
confirm and verify that the Fuji Computed Radiography System when used with an MQSA compliant
X-ray system for screen-film is working in a consistent manner according to the specifications for use as
part of a Full Field Digital Mammography (FFDM) system.
This document applies to the CR-IR 348CL (hereinafter referred to as CR Console) and the CR-IR
363AWS (hereinafter referred to as AWS-c).
Always follow all applicable laws and regulations for your jurisdiction. If anything in this
document is in conflict with applicable laws or regulations, the applicable law or
regulation shall take precedence.

Acknowledgement
We would like to thank the American College of Radiology, Reston, Virginia for their permission to
include portions of the 1999 ACR Mammography Quality Control Manual in this document which
have been reprinted with permission of the American College of Radiology.
No other representation of that material is authorized without expressed, written permission from the
American College of Radiology.
DISCLAIMERS
1

No part or all of this manual may be reproduced in any form without prior permission.

The information contained in this manual may be subject to change without prior notice.

FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
installation, relocation, remodeling, maintenance, and repair performance by other than dealers
specified by FUJIFILM Corporation.

FUJIFILM Corporation shall not be liable for malfunctions and damages of FUJIFILM Corporation
products due to products of other manufacturers not supplied by FUJIFILM Corporation.

FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from remodeling,
maintenance, and repair using repair parts other than those specified by FUJIFILM Corporation.

FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
negligence of precautions and operating methods contained in this manual.

FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from use
under environment conditions outside the range of using conditions for this product such as
power supply, installation environment, etc. contained in this manual.

FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from natural
disasters such as fires, earthquakes, floods, lightning, etc.

Caution: Rx Only (Federal law restricts this device to sale by or on the order of a physician.)
Trademark
FCR is a trademark or a registered trademark of FUJIFILM Corporation.
Copyright 2006-2009 FUJIFILM Corporation. All rights reserved.

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Introduction to Measurements
To produce consistently high quality images, each part of the imaging chain must function within the
limits of performance given. Users must therefore be able to evaluate the status of the acquisition
system, including detector, the processing system and the display systems.

Assumption
Many tests in this manual are specific to the Fuji Computed Radiology (FCR) mammography system
and its related components. Some tests are designed to verify use of the FCR mammography
system in conjunction with an X-ray generator and imaging chain components. It is therefore
required that the X-ray generator and the collimator are performing and calibrated in accordance
with MQSA requirements for screen-film systems. The display systems used for final interpretation
(soft copy and/or hard copy) must be calibrated and performing in accordance with the
manufacturer's quality control program for mammography use.

Processing System Software


The processing system is evaluated by the inspection and scoring of a test set of phantom images
that have been processed in the CR Console/AWS-c with the standard mammography processing
algorithms EDR (Exposure Data Recognizer), and MFP (Multi-Objective Frequency Processing).
Processing parameters should be fixed as a baseline for QC tests, specifically the American College of
Radiology (ACR) mammography accreditation program phantom (MAPP) or other appropriate phantom
(RMI 156 etc.) test.

Dedicated Imaging Plates (IPs) and Cassettes


Some of the tests described in this manual utilize both sizes of the IPs and cassettes that are designed
for the imaging of high resolution mammograms. The customer should reserve one of each size IP and
one of each size cassette for Quality Control testing only, and should clearly identify these items as
intended for QC only. These IPs should not be used for patient imaging; and only these reserved IPs, and
cassettes should be used for QC procedures unless otherwise directed in this manual.
Use only the cassette and IP type designated for use with your Reader Unit (RU for mammographic
imaging and QC testing.

Clearview RU:
Aspire CRm RU:

Cassette Type DM with HR-BD IP.


Cassette Type CH with HR-VI IP.

Cautions:
Perform the S value Conformation test prior to the remaining Acceptance and QC tests, using
only molybdenum (Mo) target and Mo filter at 25kVp.
If the mammography X-ray unit will not allow 20 mR exposure, you may use the lowest mR
exposure available greater than 20. Always use 25 kVp for this test. Do not use any additional
filtration. The use of another filter may cause an error.

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

S value
FCRm/CRm associates an S value with each FCRm/CRm image. Usually the S value is displayed with
other data in the hard copy or soft copy display.
Technically, the S value is defined by the equation S = 4 10(4-Sk), which is derived from the Sk value
determined by the Exposure Data Recognizer (EDR).
The meaning of the S value differs from that of relative sensitivity or relative speed used with screenfilm systems. Nonetheless, the S value can be used as a guide that roughly represents relative
sensitivity.
Note however, that:
With all other factors constant, the S value changes with time as the photomultiplier tube in the
FCRm/CRm equipment ages requiring periodic calibration.
The S value varies if the positioning setup changes even if the X-ray exposure and tube voltage
remain unchanged.
If the exposure voltage (kVp) changes, the histogram changes, varying the S value.
If the X-ray apparatus changes, the S value changes even if the same exposure conditions are
used.
If the FCRm/CRm exposure menu is changed, the EDR conditions change so S value comparisons
would be invalid.
Even when exactly the same conditions are employed, the S value changes if the time interval
between IP exposure and IP insertion changes. When the X-ray generator, tube voltage, patient,
menu, positioning setup, and elapsed time from exposure remain unchanged, the S value is a
relative measure of the X-ray dose and is inversely proportional to it. For example, the S value is
approximately doubled when the dose is reduced by half.

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Overview of Testing
FREQUENCY
SECTION A : Getting Started

TEST DESCRIPTION
Set-up & Baseline Images

RESPONSIBILITY
Technologist and
Medical Physicist

CNR Weekly Check


SECTION B : Weekly
(and when problems are
suspected)

Technologist
Phantom Image
Printer QC*
Technologist
Monitor QC*

SECTION C : Monthly
(and when problems are
suspected)

Visual Checklist

Technologist

SECTION D : Quarterly
(and when problems are
suspected)

Repeat Analysis

Technologist

SECTION E : Semi-Annual
(and when problems are
suspected)

Compression
Technologist
Imaging Plate (IP) Fog
Viewing and Viewing Conditions*
Annual Printer QC*
Annual Monitor QC*
S Value Confirmation
System Resolution
CR Reader Scanner Performance
Mammographic Unit Assembly Evaluation
Collimation Assessment

SECTION F : Annual
(and when problems are
suspected)

Automatic Exposure Control (AEC) System


Performance Assessment

Medical Physicist

System Artifact Evaluation


Image Quality Evaluation
Dynamic Range
Primary Erasure
Inter-Plate Consistency
kVp Accuracy and Reproducibility
Dose
Beam Quality Assessment and Half-Value
Layer Measurement
Radiation Output

(*) Follow the manufacturers approved QC program if available.

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Section

8th Edition
897N0602G

Getting Started

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Set-up & Baseline Images


This section will guide you through the initial set-up and use of this manual. The tests are
performed by conducting a quantitative and visual inspection of test images and the ACR
Mammography Accreditation Program Phantom (MAPP).
Use this QC manual in conjunction with the operation manual for the FCRm/CRm reader unit,
the CR Console/AWS-c, and the FCRm/CRm User Guide. In the FCRm/CRm User Guide, refer to
Chapter 3 Getting Started - Calibrating the AEC for FCRm/CRm.
BASELINE TESTS OF FCRm/CRm SYSTEM
Complete the installation of the X-ray room and Fujifilm Computed Radiography system before
beginning the QC testing. Perform the tests in the table below as the baseline (first measurement)
for each QC test. If you are unsure about the status of your equipment, then ask the Fujifilm imaging
specialist, service engineer, or your physicist. Verify who is responsible for which device: the X-ray
manufacturer for the X-ray unit, Fujifilm service personnel for the FCRm/CRm and related
equipment.
FCRm BASELINE TESTS
Printer QC (Annual)

System Resolution*

Dynamic Range*

Monitor QC (Annual)

CR Reader Scanner Performance*

Primary Erasure*

CNR Weekly Check*

AEC System Performance Assessment

Inter-Plate Consistency

Image Quality*

Imaging Plate Fog

Dose

S Value Confirmation*

System Artifact Evaluation*

* Indicates tests to be preformed when adding or replacing an FCRm/CRm Reader Unit.

Upon successful completion of the FCRm/CRm Reader Unit baseline tests, satisfy any MQSA
obligations prior to clinical use.
STEP 1: SETTING BASELINE TEST CONDITIONS
When testing equipment, it is very important that the test conditions are reproducible and
consistent. This involves setting up a baseline log for the X-ray room, FCRm/CRm system
parameters, and viewing conditions. These must be recorded in the EXPOSURE & IMAGE PROCESSING
SET UP. Always start by using your most consistent X-ray room and equipment. Decide on the X-ray
room and record it in the EXPOSURE & IMAGE PROCESSING SET UP. Select an IP and cassette of each
size that are in good condition to be put aside for the QC tests. Label them as such and write this
information in the EXPOSURE & IMAGE PROCESSING SET UP. Now record the AEC sensor position,
settings, X-ray generator, target, filter, grid and Image processing parameters that are being used
clinically and that you will use for the ACR MAPP.
Ensure that the menus shown in the following table are available. If you cannot find the menus,
create the menus according to the table on the following page.

10

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

FCRm QC Test Menus and Parameters


GP

MFP

PEM *

GA

GT

GC

GS

MRB

MRT

MRE

MDB

MDT

MDE

PRN

PTE

PTC PRE

CREATED FROM HR
SENSITIVITY
Physics, Sensitivity

1.0

1.2

0.00

Physics, Artifacts

1.0

1.2

0.00

Physics, Collimation

1.3

1.2

0.00

Physics, IP Fog

1.3

1.2

0.00

0.0

0.0

0.4

2.0

CREATED FROM CONTRAST

CREATED FROM AVE5C16


Physics, Dynamic Range
1.0
A
1.6
0
C
R
0.0
Before evaluating the Physics, Dynamic Range test image, set the L value to 4.0.
CREATED FROM AVE5CB08
Physics, ACR MAPP

1.4

0.6

1.3

Notes for the Physics ACR MAPP menu:


1) Change these image processing parameters to those parameters used for typical clinical images at your facility, if
necessary.
2) Set the GS value in the BASE LINE TEST so that the background image brightness or film density is acceptable
* PEM is only used with the Clearview RU-based FCRm systems.

NOTE
In order to set up a different image processing parameter, please change MPM code of the menu
created.

Select a viewbox or diagnostic workstation designed for final interpretation of mammographic


images that is located in a low ambient light environment. Also, mask unexposed areas of the
image to reduce glare. Luminance and ambient light measurements are recommended in order
to check that viewing conditions are adequate. These are included as part of a yearly physics
QC tests. The use of a magnifying lens with 2x magnification or greater is recommended for
evaluating tests from film.
STEP 2: THE QC TESTS
Before starting, read through the tests that make-up the QC program. Ensure that you have
access to everything listed under Required Test Equipment before beginning each test. The
relevant test result forms are found at the end of each test section.
It is important that you prepare for the quarterly REPEAT ANALYSIS as soon as you begin using
this QC program. Although you will only be doing the test quarterly, you will need to start
collecting the reject images now.
The annual tests are intended for the medical physicist and are more involved than the other
tests. A dose meter and densitometer will be required. These must be calibrated and have a
calibration certificate. These tests may also be needed if a problem arises during the daily,
weekly, monthly and quarterly testing. For example, if artifacts appear on images, it is
recommended that you do the artifact evaluation test.
Test images should be stored for future reference as directed in the test procedure.

8th Edition
897N0602G

11

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

STEP 3: TAKING THE FIRST ACR MAPP IMAGES


Prior to producing the first phantom image, confirm that the display (image printer and/or
diagnostic workstation) is in control. Confirm that the FCRm reader unit has successfully passed the
S Value Confirmation Test.
Using the baseline setup from Step 1, expose the ACR MAPP or other FDA approved phantom
following the procedure in Phantom Image test (Test 2, Section B) and produce a baseline
image. Repeat this twice so that you now have 3 images with the same level of quality. If
significant variation is seen between the 3 images, it is important that the system is checked before
continuing with the QC program. Consult with the Fujifilm applications specialist. Choose one of the
3 comparable ACR MAPP images to be used as a reference for future tests.
Also, in accordance with the procedures of the Image Quality Evaluation Test, score the phantom
image to determine how many fibers, speck groups and masses are visible. These scores will
form the baseline for the ACR MAPP image tests.
STEP 4: TESTING AND CORRECTIVE ACTION LOG
Once the baseline values have been recorded and ACR MAPP baseline images have been
acquired and logged, it is possible to begin the QC program. During routine QC testing, a result
may exceed the acceptable limits. When this occurs, repeat the test to confirm the results. If the
re-test still exceeds the acceptable limits record the occurrence on the MAINTENANCE &
CORRECTIVE ACTION LOG. Investigate and correct the problem as directed in the test
procedure.

12

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

EXPOSURE & IMAGE PROCESSING SET UP


DATE:

Baseline Phantom Exposure Settings


Room

X-ray Generator
Target

Cassette ID No.

Filter

IP ID No.

KVp

AEC Sensor Position

mAs

AEC Density Setting

Grid

Baseline Image Processing Parameters


Basic
GA
GT
GC
GS
Multi-Frequency MFP
MRB
MRT
MRE

MDB
MDT
MDE

Pattern Enhancement PEM*


PRN
PTE
PTC
PRE

* PEM is only used with the Clearview RU-based FCRm systems.

NOTE
Refer to User Guide - Mammography for detailed information about Image Processing
Parameters.

8th Edition
897N0602G

13

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

14

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Section
Section

8th Edition
897N0602G

B
A

Weekly (QC Technologist)


Getting
(and
whenStarted
problems are suspected)

A
B

15

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

CNR Weekly Check QC Technologist


OBJECTIVE
To establish an operating level of Contrast to Noise Ratio (CNR) at a specific exposure and weekly
confirm that the CNR remains consistent, within limits over time at the same exposure setting.
NOTE
This procedure is designed to measure the performance of the exposure system
FCR mammography system is a digital imaging system. Image density is programmed to a
specific level when using the auto or semi auto modes such as those used in routine
mammographic projections. Unlike screen-film systems, optical density is not directly related to
exposure. Density and contrast are controlled by image processing.

FREQUENCY
This test must be performed at installation after relocating the image reader as in a mobile setting
before imaging patients. In routine use, perform at a frequency of no less than weekly.

REQUIRED TEST EQUIPMENT


A 4cm phantom made of acrylic (PMMA) of fully covering an 18x24cm field size.
A 0.2 mm Al CNR test object.
The 1824cm dedicated QC Mammography cassette and Imaging Plate.
CNR BASELINE WORKSHEET
Watch/Timer

PROCEDURE
1.

Using the 18x24cm dedicated mammography QC cassette, erase an IP using the secondary
erasure mode on the reader. Place the loaded cassette in the cassette holder assembly.
Set a manual technique using the following factors: large focal spot, Mo/Mo 26 kVp and 125
mAs. If 125 mAs is not available on the X-ray generator, use the next highest mAs setting.
Note: Establish the baseline CNR value by performing this test five times and averaging the
five CNR values. As the positioning of the aluminum plate may have some effect on the
calculated CNR, remove and reposition the aluminum plate after each exposure to
reflect this normal variation in the baseline. Record the CNR value from each exposure
on the CNR BASELINE WORKSHEET. Calculate the average CNR and 20% values
and record them as indicated on the CNR BASELINE WORKSHEET. Repeat this
process whenever a new baseline is required (see Precautions and Caveats).

16

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

2.

Place the 4cm thick acrylic phantom with a 0.2mm thick aluminum object on top. Make sure
that the aluminum object completely covers the area shown in the diagram below.

Calculation Areas
(20mm 20mm)
Aluminum Object

20mm

BG

Al

20mm
60mm

20mm

Acrylic Slabs

Chest Wall Side


3.
4.
5.

Bring the compression device into contact with the phantom.


Make an exposure using the technique factors described in Step 1.
Wait the predetermined interval time and process the IP using the Physics ACR MAPP menu.
NOTE
To reduce the influence of IP image fading characteristics on this test, it is important to control
the time interval between X-ray exposure and reading of the IP. It is important that the interval be
consistent whenever this test is performed. We recommend choosing an interval between 5 and
10 minutes, then using the same interval consistently.

6.

Calculate CNR using the QC calculation tool and record the value in the data form.

7.

Perform the Weekly CNR Test on all mammographic exposure units at your facility that use this
CR Reader.

PRECAUTIONS AND CAVEATS


Always be sure to use the same IP and cassette for this test. Always use the same exposure
technique as established in Step 1. Always wait the same amount of time between exposure and
processing of the IP when performing this test. Always erase the IP before performing this test.
Each exposure unit will produce a unique CNR baseline value. Be sure to only compare weekly test
results against the baseline of the same exposure system.
CNR baselines must be re-established after major service or replacement events. These include but
are not limited to:
X-ray tube replacement
Filter replacement
Replacement of the compression paddle
Change in the phantom used
Change of IP and/or cassette used (always use the same IP and cassette)
Change of grid or table surface
Change of X-ray generator calibration
Change of CR Reader calibration
Following any of these events, a new baseline must be set. Calculate the new CNR baseline from 5
times of measured tests as described above and establish a new baseline.
8th Edition
897N0602G

17

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

PERFORMANCE AND CORRECTIVE ACTION


The CNR of the new (weekly) image must be within +/- 20% of the baseline image (CNR) established
for the same exposure room.
If the CNR of the new image exceeds these limits, repeat the test confirming that all technique
factors are correct. If the new test results in a passing score record the results of the new test on
the Phantom Control/CNR Chart. Failure of this test can be a problem with the X-ray exposure unit or
CR reader unit. If the CR reader unit is used with more than one mammography X-ray exposure unit,
continue to perform the Weekly CNR Test on all mammographic exposure units used with the CR reader
unit.
If one exposure system fails and other exposure system(s) pass, it is likely that the cause of failure is
related to a change in the X-ray exposure system. Retest the failed system.
If the system fails the second test, the problem must be identified and corrective action taken before
any further examinations are performed using that exposure system. Record the corrective actions
on the Maintenance & Corrective Action Log of the Phantom Control/CNR Test Report.
Review the PRECAUTIONS AND CAVEATS section to help determine if the change in CNR is
related to a service event or replacement product change that requires a resetting of baseline CNR
values.
If all exposure rooms fail and there has been no service calls or change in equipment in the exposure
room, the problem may be related to the CR reader unit. If this test produces results that fall outside the
action limits as specified by the manufacturer, the source of the problem shall be identified and
corrective action shall be taken on the component(s) that caused the failure before any further
examinations are performed using the failed component(s). If the component(s) that caused the failure
(e.g. exposure unit, CR reader unit, display device) is replaced by an alternative device and the
mammography system passes the re-test, image acquisition and interpretation may continue using
that combination of devices.

18

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

CNR BASELINE TEST


DATE:
Phantom : 4cm

X-ray generator Number :

0.2mm Al

Cassette - IP Number :

26kVp 125mAs

FCRm/CRm reader ID :

CNR 1 :
CNR 2:
CNR 3 :
CNR 4 :
CNR 5 :
AVE. CNR (BASELINE):
+/-20% :

to
SIGNATURE:

8th Edition
897N0602G

19

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Phantom Image (Weekly) QC Technologist


OBJECTIVE
To assure that contrast, uniformity and image quality (due to the X-ray exposure system, CR reader unit,
printer, and workstation) are maintained at optimum levels.

FREQUENCY
This test must be carried out initially at installation, after appropriate calibration of the equipment to
establish a baseline level. Subsequently, it must be carried out at least weekly, after service to any
equipment, (i.e. the X-ray generator, FCRm/CRm reader unit, image laser printer, and workstation) and
whenever changes in image quality are suspected.
With mobile mammographic equipment, the performance of the system must also be confirmed
after arriving at a new location and prior to imaging patients, in addition to the weekly test.

REQUIRED TEST EQUIPMENT


ACR MAPP
Acrylic 4mm 1cm diameter disc (for hardcopy QC only)
Designated cassette and imaging plate used for mammography
Densitometer (for hardcopy QC only)
Phantom Control/CNR Charts
HARD COPY PHANTOM CONTROL/CNR CHART
SOFT COPY PHANTOM CONTROL/CNR CHART
Watch/Timer

PROCEDURE

20

1.

Using the designated mammography cassette, erase the IP using the secondary erasure mode
on the reader.

2.

Place the cassette in the bucky tray.

3.

Place the phantom on the cassette holder and position it so the edge of the phantom is fully
covering and aligned with the chest wall side of the image receptor. Center the phantom, left
to right.

4.

The 4mm thick acrylic disc required to be placed on the phantom for hard copy QC may be
present during softcopy QC, but is not required for softcopy QC. For QC of softcopy used for
final interpretation, proceed to step 6.

5.

For hardcopy QC, secure the acrylic disk to the top of the phantom within the image area but
position it so it will not obscure details in the phantom. A suitable location is between and
slightly below the first and second largest fibers. Once located, the disc may be permanently
attached to the phantom with super glue.

6.

Bring the compression device into contact with the phantom.

7.

Verify the location of the AEC detector. It should be in the same location used for previous
phantom images and completely covered by the phantom.

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

8.

Make an exposure using the AEC technique factors (target, filter, kVp, grid, density control
setting, etc.) currently used clinically for a 4.2-cm compressed breast of 50/50 composition.

9.

Optionally, you may plot the mAs on the control chart after making the exposure. Recall that
mAs will change significantly in the AEC mode if a different kVp, target, or filter is selected. This
step, plotting mAs, is optional and not required.

10. After exposure, ensure that a predetermined interval time between 5 and 10 minutes elapses
prior to the reading process. Process the IP using the "Physics, ACR MAPP" menu.
NOTE
It is important to control the time interval between exposure and reading of the IP. To minimize
variability, it is recommended that an interval be chosen somewhere between 5 and 10 minutes
and that this same interval be used consistently.

11. Record the AEC density control setting on the control chart.
12. For printed images as a final interpretation: Measure the following densities. The background
density should be measured at the geometric center of the phantom image. The measured OD
must not be less than 1.20.
The DD measurement is obtained by subtracting the density inside the acrylic disc from the
density directly adjacent to the disc, in the direction perpendicular to the anode-cathode axis.
For consistent results, these measurements must be made at the same location each time.
Plot the background optical density and the DD on the Hard Copy Phantom Control Chart.
For softcopy images as a final interpretation: Plot the S value from the phantom image
exposure on the S value range line on the SOFT COPY PHANTOM CONTROL/CNR CHART .

VIEWING CONDITIONS
Phantom images should be read under optimal viewing conditions. General lighting should be at a
low level and diffused. Viewboxes or workstations should be positioned to avoid light from windows,
other viewboxes, and other sources of bright light, either direct or reflected. Images should be
masked to eliminate extraneous light. Use a magnification of 2X or higher for scoring speck groups
as well as other test objects.

DATA ANALYSIS AND INTERPRETATION

8th Edition
897N0602G

1.

When scoring the phantom image, each object is scored separately. Always count the number
of visible objects from the largest object of a given type (i.e., fiber, speck group or mass)
downward until a score of 0 or 0.5 is reached, then stop counting for that object type.

2.

Count each fiber as one point if the full length of the fiber is visible and the location and
orientation of the fiber are correct.
Count a fiber as 0.5 if not all, but more than half, of the fiber is visible and its location and
orientation are correct. Add each full or partial fiber to the total score, from the largest down to
the smallest visible, until a score of 0 or 0.5 is reached.

3.

After determining the last fiber to be counted, look at the overall background for artifacts.
If a fiber-like artifact appears anywhere in the wax insert area of the image, but not in an
appropriate location or orientation, deduct the artifactual fiber from the last real half or
whole fiber score if the artifactual fiber is equally or more apparent.
Deduct only from the last real fiber.
Record the final score after artifact deduction in the appropriate space on the chart.
21

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

4.

Use large fieldof-view magnification to assist in the visualization of specks. For hardcopy
images, use a 2X or higher magnifying glass. For softcopy images, use the workstations
electronic magnification function. Starting with the largest speck group, count each speck
group as one point if four or more of the six specks in the group are visible in the proper
locations. Count a speck group as 0.5 if two or three of the six specks in the group are visible
in the proper locations. Count a speck group as 0 if none or only one of the six specks in the
group is visible in the proper location.
Add each full or partial speck group to the total speck group score, from the largest down to
smallest visible group, until a score of 0 or 0.5 is reached.

5.

After determining the last speck group to be counted, look at the overall background for
artifacts. If noise or speck-like artifacts are visible in the wrong locations within the area of the
wax insert, and are as apparent as the real specks, deduct them one for one from the
individual specks counted in the last whole or half speck group scored, and adjust the score of
the last group appropriately. Record the final score after artifact deduction in the appropriate
space on the chart.

6.

Count each mass as one point if the minus density object is visible in the correct location, and
the mass appears to be generally circular against the background (i.e., greater than of the
perimeter or circumference is visible). A mass is counted as 0.5 point if the minus density object
is visible in the correct location, but the mass does not have a generally circular appearance.
Add each full or partial mass to the total mass score, from the largest mass down and until a
score of 0 to 0.5 is reached.
Record the raw mass score before artifact deduction.

7.

After determining the last mass to be counted, look at the overall background for artifacts.
If a mass-like artifact is seen in the wrong location within the area of the wax insert, deduct the
artifactual mass from only the last real whole or half mass scored if the artifactual mass is
equally or more apparent. Record the final score after artifact deduction in the appropriate
space on the chart.

8.

Using magnification, carefully examine the image for non-uniform areas, the presence of dirt or
dust artifacts, grid lines, processing artifacts, or any other artifacts and compare the image to
the original and previous images.

9.

Notice any artifacts or grid lines on the image. Investigate the source of any artifacts or grid
lines and record the events on the Maintenance/Corrective Action Log of the Phantom Control
Chart. The medical physicist can provide assistance in identifying the source of the artifacts.
NOTE
Mammography phantom images should always be viewed

22

By the same person, if possible


On a quality-controlled QA Workstation, Primary Interpretation Workstation, or Viewbox
Under the same viewing conditions as the Radiologist review stations
Using the same type of magnification used for reading mammograms
Soft copy images should be window and leveled to best demonstrate the objects of interest

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

PRECAUTIONS AND CAVEATS


This test measures the output from all components in the imaging chain. Changes in image quality
may be due to film, cassettes, IPs, X-ray generator, added filtration, printer, or workstation.
Consequently, other tests will be necessary to determine the component, or components, causing
the change. It is also necessary to check if settings for the image reading mode or image processing
parameters are changed. These factors will have a large influence on the results of this test.
When any image parameters are changed, it is recommended to perform the Phantom Exposure
test to re-determine the new baselines. Record the imaging parameters used for this test on the
Exposure and Image Processing Set Up located in this manual.
A weekly phantom image-evaluation is required for only the most commonly used cassette size (18
24 cm), however, it may be useful to also image the phantom periodically with a large (24 30
cm) cassette.
A phantom image should be produced and evaluated if clinical images suggest problems with either
size formats. Typical examples of problems that are specific to the bucky size are grid lines, or
compression paddle artifacts.
Non-uniformity in the images can degrade the quality of clinical mammograms. To evaluate uniformity,
expose an Fuji IP Cassette on top of the cassette holder to compare image uniformity with and without
the grid.

PERFORMANCE AND CORRECTIVE ACTION


The total number of simulated masses, speck groups, and fibers visible in the phantom image
should not decrease by more than one-half, assuming the same individual is viewing the images
under identical conditions.
If a change in the number of test objects is noted, then the present image should be compared to
the original image and the previous image to determine if the change is real or if the individual
viewing the image has changed his or her criteria.
NOTE
At a minimum, the four largest fibers, the three largest speck groups and the three largest masses
must be visible. If your baseline image exceeds the above minimum scores, the new image score
should not decrease by more than one half object point from the original baseline level, however
the score must not fall below the four largest fibers, the three largest speck groups and the three
largest masses.

If printed images are used for final interpretation: The optical density (OD) of the film shall be within
control limits of 0.20, from the established operating level.
The acrylic discs used for the measurement of the density difference (DD) may vary in thickness.
Consequently, the density difference is a relative, not absolute measurement and is to be used only
for quality control purposes. It is essential to use the same acrylic disc if comparisons are to be
made between different facilities.
The density difference (disc vs. adjacent density) shall not vary by more than 0.05 OD from the
established operating level. If a new operating level for background optical density is chosen, then a
new operating level for density difference must be established. It is essential that all printers used to

8th Edition
897N0602G

23

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

print mammography images at one facility produce similar film optical densities. It is not acceptable
to have one unit producing a film with an OD of 1.40 and another producing an OD of 1.80.
If softcopy images are used for final interpretation: The S value of the phantom image confirms the
exposure unit output and the FCRm/CRm reader sensitivity setting. The S value result must be plotted
on the SOFT COPY PHANTOM CONTROL/CNR CHART and must not vary by greater than 20%. If the
results exceed this control limit, make sure that the kVp and mAs are set properly and that the correct
menu is selected.
For both softcopy and hardcopy:
If the performance criteria for this test are not met, a second phantom image test should be
performed and evaluated. If the criteria are still not met, the reasons for this failure must be
investigated, corrective action taken, and the results documented before patients are examined with
this system. Any visual difference between the current phantom image and the saved original phantom
image should be investigated. Printer or FCRm/CRm reader unit artifacts, grid lines, or grid artifacts
should not be present since any of these may degrade clinical images.
The appearance of artifacts, grid lines, or grid artifacts; the number of masses, specks and fibers
visualized; or any other change in the visual appearance of the image should be reported
immediately to the medical physicist for further evaluation.
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken on the component(s) that caused the failure
before any further examinations are performed using the failed component(s). If the component(s) that
caused the failure (e.g. exposure unit or CR reader unit) is replaced by an alternative device and the
mammography system passes the re-test, image acquisition and interpretation may continue using
that combination of devices.
If the failure has been determined to be of a diagnostic device used for mammographic image
interpretation (e.g. laser printer, physicians review station) the source of the problem shall be
identified and corrective action shall be taken before that device can be used for mammographic
image interpretation. Clinical imaging may be continued. If available, an alternative approved
diagnostic device may be used for mammographic image interpretation

NOTE
Phantom images must be retained in the QC records for the last full year. (Softcopy images can be
locked on the QA and PACS workstation). The original baseline image must be retained until it is
necessary to establish a new baseline image.

24

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

HARD COPY PHANTOM CONTROL/CNR CHART


Room:

Cassette ID No.:

IP ID No.:

Year:

Month:
Date:
Initials:
AEC SETTING:
CNR Value:
CNR OK:
Density Difference

+0.05

+0.05

Baseline

-0.05

-0.05

Background Density
+0.20

+0.20

+0.15

+0.15

Baseline

-0.15

-0.15

-0.20

-0.20

No.Visible
Fibers

+0.5

Specks

Masses

+0.5

- 0.5

- 0.5

+0.5

+0.5

- 0.5

- 0.5

+0.5

+0.5

- 0.5

- 0.5

mAs (optional)
+15%

+15%

Baseline

-15%

8th Edition
897N0602G

-15%

25

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

SOFT COPY PHANTOM CONTROL/CNR CHART


Room:

Cassette ID No.:

IP ID No.:

Year:

Month:
Date:
Initials:
AEC SETTING:
CNR Value:
CNR OK:

No.Visible
Fibers

+0.5

Specks

Masses

+0.5

- 0.5

- 0.5

+0.5

+0.5

- 0.5

- 0.5

+0.5

+0.5

- 0.5

- 0.5

S Value
+20%

+20%

Baseline

-20%

-20%

mAs (optional)
+15%

+15%

Baseline

-15%

26

-15%

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

MAINTENANCE & CORRECTIVE ACTION LOG


DATE

PROBLEM

CORRECTIVE ACTION TAKEN

SIGNATURE

Prepared by
8th Edition
897N0602G

27

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Printer QC (Weekly*) QC Technologist


OBJECTIVE
The printer used to produce films for final interpretation of mammography images must be cleared
by FDA for that purpose. The objective of this test is to assure that the printer used for final
interpretation is performing according to the manufacturers specifications.

PRINTERS WITH MANUFACTURER-PROVIDED QC PROGRAM


*Follow the printer manufacturers QC program for test frequency, procedure, performance and
corrective action.

PRINTERS WITHOUT MANUFACTURER-PROVIDED QC PROGRAM


If the manufacturer does not provide a mammography printer QC program use the following.

FREQUENCY
This test must be performed initially before interpreting mammograms. *Perform weekly for dry print
imagers, daily before printing mammographic images for wet print imagers. Perform after any
service or maintenance on the mammographic printer system and whenever problems are
suspected.

REQUIRED TEST EQUIPMENT


QC pattern with a fixed digital value grayscale (e.g. TG18-QC or TG18-PQC). Contact your
printer manufacturer for details.
Calibrated densitometer.
Report form, available from your printer manufacturer.

PROCEDURE
SET UP
1.

Print the QC pattern.

2.

Measure the image to identify the step which has an optical density closest to but not less than
1.20. Designate this step as the mid-density (MD) step.

3.

Measure the image to identify the step with an optical density closest to but not less than 2.20
and the step that has an optical density closest to but not less than 0.45. Designate these
steps as the high-density (HD) and low-density (LD) steps. The difference between these two
steps is designated as the density difference (DD).

4.

Measure the lightest (B+ F or unexposed) portion of the image. This is designated as B+F.

RECORDING THE QUALITY CONTROL STEPS


1.

28

Print the QC pattern. Write the date and time on the film. Label the film with the printer ID if
there is more than one printer.

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

2.

Measure the densities of the designated steps.

3.

Plot the mid-density (MD), the density difference (DD) and the base-plus-fog (B+F) on the
control chart.

4.

Determine if any of the data points exceed the control limits. If not, go to step 6. If so, expose
and process a second QC strip, double checking that the correct procedure is followed. If the
same results are obtained, proceed to step 5.

5.

Circle any out-of-control data points and repeat the test. If any data point is still out of control,
correct the cause of the problem and repeat the test to confirm that the problem has been
corrected. Note the cause of the problem and the corrective action and plot the in-control
point.

6.

Determine if there are any trends, i.e., three or more data points moving in one direction (either
upwards or downwards), in the MD, DD, or B+F. If trends are present but the data points have
not, as yet, exceeded the control limits, clinical mammogram images can be printed. It is
necessary to determine the cause of the trend and to monitor the processor closely to assure
that the control limits are not exceeded.

7.

Retain the test film for at least the MQSA prescribed period, e.g., the last full month for daily
QC and the last 12 weeks for weekly QC.

PRECAUTIONS AND CAVEATS


* Perform this test using only the test patterns described above. Test patterns are specific to the
image matrix size and bit-depth. Using other patterns could yield erroneous results.
Film is produced in batches. Consequently, there may be slight variations in the characteristics of
the film between batches. In addition, film aging and storage conditions can also affect the
sensitometric characteristics of the film. Whenever a new emulsion of film is opened, it is necessary
to perform a laser density calibration with the new film. The operating level on the control chart
should be adjusted to the new levels for MD and DD.

PERFORMANCE AND CORRECTIVE ACTION


If the MD and DD are within 0.15 of their respective operating levels, and the B+F is within + 0.03
of its operating level, the printer is in control, and no further action is required. If the MD and DD fall
outside the above control limits, corrective action must be performed and then confirmed by
performing the QC test again. The source of the problem must be corrected before any
mammographic images are printed. If this test produces results that fall outside the action limits as
specified above or by the manufacturer (whichever is applicable), the source of the problem shall be
identified and corrective action shall be taken before that device can be used for mammographic
image interpretation.
Clinical imaging may be continued. If available, an alternative FDA-approved for mammography
diagnostic display device may be used for mammographic image interpretation, provided it has
passed the applicable QC tests.

8th Edition
897N0602G

29

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Monitor QC (Weekly*) QC Technologist


OBJECTIVE
The objective of this test is to assure that monitors are performing according to the monitor
manufacturers specifications. Both the monitors used for final interpretation of mammography
images (primary monitors), and monitors used for acquisition QC (secondary monitors) must be
tested. An additional requirement is that monitors used for final interpretation must have been
cleared by FDA for that purpose.
Commercial calibration software programs supporting the American Association of Physicists in
Medicine (AAPM) Assessment of Display Performance for Medical Imaging Systems (AAPM On-Line
Report No. 03, www.aapm.org) are available for use with this test.

MONITORS WITH MANUFACTURER-PROVIDED QC PROGRAM


*Follow the monitor manufacturers QC program for test frequency, procedure, performance and
corrective action.

MONITORS WITHOUT MANUFACTURER-PROVIDED QC PROGRAM


If the manufacturer does not provide a mammography monitor QC program, use the following.

FREQUENCY
This test must be performed initially before interpreting mammograms, weekly, after any service to
the monitor, and whenever problems are suspected.

REQUIRED TEST EQUIPMENT


AAPM TG18-QC, SMPTE RP133 or other patterns appropriate for the monitor to be tested
(contact your monitor or QC software manufacturer for details*).
A photometer to measure luminance of primary monitors.
Your own or the QC software manufacturer-provided report form. Note: The results of this and
other tests can be stored electronically as permitted by MQSA regulation or guidance

PROCEDURE
1.

Display the QC pattern.

2.

For primary monitors only: record the maximum and minimum luminance levels as directed by
your monitor or QC software manufacturer. The maximum luminance (Lmax) is typically
specified by the monitor manufacturer as the highest value that can be used without
compromising other performance characteristics, such as lifetime or resolution. The minimum
brightness (Lmin) is influenced by Lmax.

3.

For both primary and secondary monitors: examine the image carefully to determine the
visibility of the following features:
a. Verify that the 5% signal level inset is visible in the larger 0% signal field.
b. Verify that the 95% signal inset is visible in the 100% signal field.

30

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

c. Verify that each gray level step from 0% to 100% can be distinguished individually.
d. Verify that the alphanumeric characters in the image appear sharp and in focus.
e. Verify that the high-contrast bar images in the center and at the corners of the QC pattern
are distinguishable. The low-contrast patterns are not evaluated in this test.
f. Record the results of the evaluations on your own form or one provided by the QC software
manufacturer.

PRECAUTIONS AND CAVEATS


*Perform this test using only the test patterns described above. Test patterns are specific to the
image matrix size and bit-depth. Using other patterns could yield erroneous results.

PERFORMANCE
If the following conditions are met, the Monitor QC Weekly check is acceptable:
For primary monitors only: the maximum luminance value (Lmax) is within the range specified by
the monitor manufacturer. For reference, the AAPM report recommends that once set, the
target level (Lmax) is maintained within 10% of the desired value.
For both primary and secondary monitors: the 5% inset in the 0% field and the 95%- inset in the
100% field shall be visible and the 0% to 100% gray level steps shall be distinguishable from the
adjacent steps.
The alphanumeric characters appear sharp and focused.
The high contrast line-pair patterns are distinguishable at the center and the corners of the
display in both the horizontal and vertical orientations.

NOTE
The ability to distinguish the low-contrast line-pair patterns is not a requirement for this test.
Unless required by the monitor manufacturer, there is no action limit specification for Lmin in this
procedure

CORRECTIVE ACTION
For primary monitors (used for final interpretation). If this test produces results that fall outside the
action limits as specified by the manufacturer, the source of the problem shall be identified and
corrective action shall be taken before that device can be used for mammographic image
interpretation. Clinical imaging may be continued. If available, an alternative approved diagnostic
display device may be used for mammographic image interpretation, provided it has passed the
applicable QC tests.
For secondary monitors (used for image acquisition QC). If this test produces results that fall outside
the action limits as specified by the manufacturer, the source of the problem shall be identified and
corrective action shall be taken on the monitor that failed before any further examinations are
performed using that monitor. If the monitor that failed is replaced by an alternative monitor that
passes the test, image acquisition may resume using the alternative monitor.

8th Edition
897N0602G

31

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

MONITOR QC (WEEKLY) REPORT

Room:
Left or Single Monitor
Serial Number:
Right Monitor Serial
Number:

Date:

Date:

Monitor
Luminance
(primary only)

LEFT

RIGHT

Monitor

MIN / MAX

MIN / MAX

Record values:
Lmax Pass/Fail:
0%-5%
contrast:
95%-100%
contrast:

Luminance
(primary only)

LEFT

RIGHT

MIN / MAX

MIN / MAX

Record values:
PASS

FAIL

PASS

FAIL

Lmax Pass/Fail:
0%-5%
contrast:
95%-100%
contrast:

PASS

FAIL

PASS

FAIL

YES

NO

YES

NO

YES

NO

YES

NO

YES

NO

YES

NO

YES

NO

YES

NO

Gray Steps:

YES

NO

YES

NO

Gray Steps:

YES

NO

YES

NO

Alphanumeric:

YES

NO

YES

NO

Alphanumeric:

YES

NO

YES

NO

Line-pair
images:

YES

NO

YES

NO

Line-pair
images:

YES

NO

YES

NO

SIGNATURE

SIGNATURE

Date:

Date:

Monitor
Luminance
(primary only)

LEFT

RIGHT

MIN / MAX

Monitor

MIN / MAX

Record values:
Lmax Pass/Fail:

FAIL

PASS

FAIL

YES

NO

YES

NO

YES

NO

YES

NO

Gray Steps:

YES

NO

YES

NO

Alphanumeric:

YES

NO

YES

Line-pair
images:

YES

NO

YES

SIGNATURE

32

(primary only)

LEFT

RIGHT

MIN / MAX

MIN / MAX

Record values:
PASS

0%-5%
contrast:
95%-100%
contrast:

Luminance

Lmax Pass/Fail:

PASS

FAIL

PASS

FAIL

YES

NO

YES

NO

YES

NO

YES

NO

Gray Steps:

YES

NO

YES

NO

NO

Alphanumeric:

YES

NO

YES

NO

NO

Line-pair
images:

YES

NO

YES

NO

0%-5%
contrast:
95%-100%
contrast:

SIGNATURE

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Section
Section

8th Edition
897N0602G

C
A

Monthly (QC Technologist)


Getting
(and
whenStarted
problems are suspected)

A
C

33

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Visual Checklist (Monthly) QC Technologist


OBJECTIVE
To assure that mammographic X-ray system indicator lights, display, and mechanical locks and
detents are working properly and that the mechanical rigidity and stability of the equipment is
optimum.

FREQUENCY
This test should be carried out monthly or after any service or maintenance on the mammographic
X-ray system.

REQUIRED TEST EQUIPMENT


VISUAL CHECKLIST

PROCEDURE
1.

Review all of the items listed on the visual checklist and indicate the status. Be sure to rotate
the C-arm the way you would for patient imaging.

2.

Date and initial the checklist where indicated.

PRECAUTIONS AND CAVEATS


Some of the items on the visual checklist are operator convenience features. However, many of the
items are essential for patient safety and high quality diagnostic images. It may be necessary to add
additional items to the list that are specific to particular equipment or procedure.
These should be included on the checklist and in each evaluation.

PERFORMANCE AND CORRECTIVE ACTION


If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken within thirty days of the test date. Clinical
imaging and mammographic image interpretation may be continued during this period.

34

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

VISUAL CHECKLIST
Facility:

Room#:

Unit:

Year:

Pass = 9
Fail = F
Does not apply = NA
Month
Day
Initials

C-Arm
SID indicator or marks
Angulation Indicator
Locks (all)
Field Light
High tension cable/other cables
Smoothness of Motion

Cassette Holder
Cassette Lock (small and large)
Compression Device
Compression Scale
Amount of Compression:

Automatic
Manual

Grid

Control Booth
Hand Switch Placement
Window
Panel switches/lights/meters
Technique charts

FCRm/CRm
Condition of cassettes
Condition of IPs

Other
Shields/Aprons/Gloves
Cones or collimators
Cleaning Solution

8th Edition
897N0602G

35

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

36

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Section
Section

8th Edition
897N0602G

D
A

Quarterly (QC Technologist)


Getting
(and
whenStarted
problems are suspected)

A
D

37

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Repeat Analysis (Quarterly) QC Technologist


OBJECTIVE
To determine the number and cause of repeated radiographs. Analysis of this data will help identify
ways to improve efficiency and reduce costs, as well as reduce patient exposure.

FREQUENCY
Repeated images shall be evaluated quarterly.
In order for the repeat rates to be meaningful, a patient volume of at least 250 patients or 1,000
exposures is needed.

REQUIRED TEST EQUIPMENT


All rejected images (including data for repeated images that may have been placed in a patients
film jacket)
The total number of images produced during the test period
Means for sorting images (hardcopy or softcopy) during analysis
REPEAT RATE ANALYSIS data form

PROCEDURE
1.

Start by removing all existing reject images (hardcopy) in the department taken prior to the
start of the analysis.

2.

Take inventory of the image supply as a starting point to determine the total number of images
consumed during this test.

3.

Start collecting all rejected images. Continue to collect for the length of time needed to
radiograph at least 250 consecutive patients.

4.

Sort the rejected images into categories such as poor positioning, motion, compression, under
exposure, (these might be due to exposure or processing), artifacts (streaks, spots, etc.).
Good images (they appear to be acceptable mammograms when retrospectively evaluated
during the repeat analysis) may have also been repeated.
Some images may not have resulted in an additional exposure of the patient but may have also
been rejected. These include clear and QC images. Although it is appropriate to include wire
localization images as part of the reject analysis, they should not be included in the repeat
analysis because they are taken as part of the wire localization process.
NOTE
Rejected images are all images that are in the reject bin, including repeated images. Repeated images
are images that are retaken for inadequate quality. The reject bin does not include additional views
required to image selected tissue seen on the first image. It also does not include images taken for the
purposes of including tissue that could not be positioned on the image receptor due to the size of the
breast. For facilities using softcopy for final interpretation maintain a list of repeated images using the
REPEAT RATE ANALYSIS form included in this procedure.

38

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

5.

Some facilities place all images (repeated and good images) in the patients film jacket so there
are no repeated images in the department. In this case, the reject/repeat analysis chart should
be completed as patient examinations are carried out.

6.

Tabulate the counts from Steps 4 and 5, determining the total number of repeated images,
rejected images, and the total number of images exposed during the analysis period.

7.

Determine the overall percentage of repeated images by dividing the total number of repeated
images by the total number of images exposed during the analysis period, then multiply by
100. Next, determine the overall percentage of rejected images by dividing the total number of
rejected images by the total number of images exposed during the analysis period, and
multiply by 100.

8.

Determine the percentage of repeats in each reason for repeat category by dividing the
repeats in the category by the total number of repeated images and multiply by 100.

PRECAUTIONS AND CAVEATS


All images that are repeated should be included in the repeat analysis, not just those rejected by the
radiologist. Repeated images that have been placed in the patients jacket must be included in the
repeat analysis.
Including examinations on at least 250 patients (approximately 1000 images) allows for a minimum
number of rejected images so that reasonable statistics can be obtained for the analysis.
Collecting rejected images from a larger number of patients is encouraged because it will yield more
reliable data when evaluating causes for repeats. Facilities that do not examine 250 patients in a
quarter must still assess repeat images at least quarterly to determine the primary causes of
repeated images.
There is a danger that technologists may alter their routine procedures or criteria for accepting
images if they know their repeated images will be analyzed. This should be avoided.
NOTE
If the Auto or Semi EDR mode is selected, underexposed images will exhibit normal density,
however they may possess a grainy appearance and a higher than normal S value. Because of
this density adjustment, overexposed images will also appear normal in density but typically
produce a lower than normal S value. Consult with your medical physicist to optimize AEC or
manual exposures to ensure that optimal image quality is maintained at minimal dose.

PERFORMANCE AND CORRECTIVE ACTION


The overall repeat rate ideally should be approximately 2% or less, but a rate of 5% is probably
adequate if the radiologist and medical physicist agree that this is a reasonable level.
These rates should be based on a image volume of at least 250 patients to be meaningful.
A reason for repeat that is significantly higher than the others indicates an area for potential
improvement.
If the repeat rate exceeds the selected acceptance level of either 2% or 5%, or if the repeat or reject
rate changes from the previously measured rate by more than 2%, the change must be
investigated and corrective action taken if necessary.

8th Edition
897N0602G

39

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Any corrective action should be recorded on the bottom of the repeat analysis form.
The effectiveness of the corrective actions must be assessed by performing another repeat analysis
after the corrective actions have been implemented.
It is important to study films that are too dark or too light to determine if the underlying cause is the
exposure equipment, image printer, patient positioning, technique or sub-optimal setting of digital
image processing.
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken within thirty days of the test date. Clinical
imaging and mammographic image interpretation may be continued during this period.

40

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

REPEAT RATE ANALYSIS


DATE:
FACILITY:

FROM:

TO:

TOTAL NUMBER OF EXAMS:

TOTAL NUMBER OF FILMS:

1824

2430

OTHER

TOTAL

PROJECTION
REASON FOR REJECT

CC

MLO

ML or LM

AXILLARY

OTHER

TOTAL
REPEATS

% of
REPEATS

POSITIONING
PATIENT MOTION
COMPRESSION
ARTIFACTS
X-RAY EQUIP
MALFUNCTION
SOFTWARE
MALFUNCTION
AEC
MISPLACEMENT
UNDER
EXPOSURE
OVER
EXPOSURE
INCORRECT
PATIENT ID
WASTE
SUB-TOTAL
GRAND TOTAL
REPEAT RATE =

REJECT RATE =

REPEATS
TOTAL FILMS INCLUDING REPEATS
ALL REJECT FILMS
TOTAL FILMS INCLUDING REPEATS

REPEAT RATE =

REJECT RATE =

COMMENTS FOR CORRECTIVE ACTION AND GOALS:

SIGNATURE:

8th Edition
897N0602G

41

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

42

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Section
Section

8th Edition
897N0602G

EA

Semi-Annual (QC Technologist)


Getting
(and
whenStarted
problems are suspected)

A
E

43

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Compression (Semi-Annual) QC Technologist


OBJECTIVE
To assure that the mammographic system can provide adequate compression in the manual and
powered modes and that the equipment does not allow too much compression to be applied.
Adequate compression is essential for high quality mammography. Compression reduces the
thickness of tissue that must be penetrated by the radiation, thereby reducing scattered radiation
and increasing contrast, while reducing the breast thickness. Compression improves image
sharpness by reducing the breast thickness, thereby minimizing focal spot blurring of structures in
the image, and minimizing patient motion. In addition, compression makes the thickness of the
breast more uniform; resulting in more uniform film densities and a film which is easier to interpret.

FREQUENCY
This test should be carried out initially, then semi-annually, as part of a Mammography Equipment
Evaluation (MEE), and whenever reduced compression is suspected.

REQUIRED TEST EQUIPMENT


Analog Bathroom scale (the scale must be flat)
Several towels
COMPRESSION data form

PROCEDURE
1.

Place a towel on the cassette holder (to protect the cassette holder), then place the analog
bathroom scale on the towel, with the dial or read-out positioned for easy reading.
Locate the center of the scale directly under the compression device.

2.

Place several towels on top of the scale to prevent damage to the compression device.

3.

Using the power drive, activate the compression device and allow operation until it stops
automatically.

4.

Read and record the compression.

5.

Release the compression device.

6.

Using the manual drive, move the compression device downward until it stops.

7.

Read and record the compression.

8.

Release the compression device.

PRECAUTIONS AND CAVEATS


If the compression force exceeds 200 newtons (20.4 kilograms or 20 deca-newtons or 45 pounds)
in the initial power drive mode, immediately release the compression device and ask a qualified
service engineer to make appropriate adjustments.

44

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

PERFORMANCE AND CORRECTIVE ACTION


A compression force of at least 111 newtons (11.3kilograms or 25 pounds) shall be provided. The
maximum compression force for the initial power drive mode must be between 111 newtons
(11.3kilograms or 25 pounds) and 200 newtons (20.4kilograms or 45 pounds).
If these requirements are not met, a qualified service engineer must make the appropriate internal
adjustments and the system must be retested and pass the test before further exams are
performed.

8th Edition
897N0602G

45

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

COMPRESSION
DATE:
SCALE I.D.:
MAMMOGRAPHIC UNIT

AUTOMATIC COMPRESSION

MANUAL COMPRESSION

lbs.

lbs.

lbs.

lbs.

lbs.

lbs.

lbs.

lbs.

ACCEPTANCE LIMITS
AUTOMATIC (25 - 45 lbs)
CORRECTIVE ACTION NEEDED?

PASS

FAIL

YES

NO

YES

NO

PASS

FAIL

CORRECTIVE ACTION DESCRIPTION:


CORRECTIVE ACTION TAKEN?
CORRECTIVE DATE:
RESULTS OF NEW TEST:

Comments:

SIGNATURE:

46

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Imaging Plate (IP) Fog QC Technologist


OBJECTIVE
To verify that IPs stored in the exposure room exhibit no evidence of unintentional exposure (fog).
Unintentional exposure of an IP may result in images with reduced contrast, increased graininess, or
exposure-related artifacts.

FREQUENCY
This test should be performed initially before any patients are imaged and then semi-annually or
whenever the location of the storage space for the IPs relative to the X-ray exposure unit or the
radiation shielding is changed. In mobile mammography, it is not necessary to perform this test after
each relocation if the storage space for the IPs relative to the X-ray exposure unit or the radiation
shielding is not changed.

REQUIRED TEST EQUIPMENT


One coin
Adhesive tape
One cassette with loaded IP
IP FOG data form

PROCEDURE
1.

Select an FCRm/CRm cassette from the exposure room stock. Erase the IP using the secondary
erasure mode on the FCRm/CRm image reader.

2.

Tape the coin to the front (exposure side) of the cassette and place the cassette in the
cassette storage area of the room with the coin facing the X-ray tube (i.e. the coin is between
the X-ray tube and the IP so that any X-ray exposure from the tube would cast a shadow on
the IP.)

3.

Leave the cassette in place for a minimum of 3 clinical studies, or produce 10 exposures of the
accreditation phantom to simulate clinical use.

4.

Remove the coin and process the IP in the FCRm/CRm reader unit using the Physics, IP Fog
menu.

5.

Examine the image to see if the coin is visible and record the results on the IP Fog report form.

PRECAUTIONS AND CAVEATS


Reading the image using the wrong menu may give an erroneous result.

8th Edition
897N0602G

47

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

PERFORMANCE AND CORRECTIVE ACTION


A shadow of the coin should not be visible. If a shadow is visible, move the IPs to a more protected
location and perform the test again.
If this test produces results that fall outside these action limits, corrective action shall be taken
before further studies are performed. Image interpretation may be continued during this period.

IP FOG
DATE:

Reader Unit ID:


S value:
Storage Location:
# of Exposures:

ACCEPTANCE LIMITS
COIN VISUALIZED?
CORRECTIVE ACTION NEEDED?

PASS

FAIL

YES

NO

YES

NO

PASS

FAIL

CORRECTIVE ACTION DESCRIPTION:


CORRECTIVE ACTION TAKEN?
CORRECTIVE DATE:
RESULTS OF NEW TEST:

SIGNATURE:

48

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Section

Annual (Medical Physicist)


(and when problems are suspected)

Some or all of the annual tests may be required as part of a


Mammography Equipment Evaluation (MEE).

49

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Viewing and Viewing Conditions (Annual) Medical


Physicist
NOTE
This evaluation is specific to the viewbox and reading room conditions, for images for final
interpretation.

When interpreting hardcopy FCRm/CRm images, it is suggested that you view the printed FCRm/CRm
images under the same conditions as you would screen-film mammography images. Film masking
devices and hot lights must be available for use by the interpreting physician. The 1999 ACR
Mammography Quality Control Manual includes a section on viewbox luminance, room illuminance,
and masking, which may be used as a reference.
When interpreting softcopy images, follow the monitor manufacturers quality control procedures.
For monitors without a manufacturer-provided QC program, follow the monitor QC procedures
provided in this manual.

PERFORMANCE AND CORRECTIVE ACTION


Softcopy final interpretation
If the Monitor QC (Annual) test performance criteria, other than interpretation environment
illumination, are not met, identify and correct the source of the problem before further final
interpretation of images displayed on those monitors.
Set the final interpretation environment illumination to the lower of a) that recommended in the
monitor manufacturers quality control program, if applicable, or b) 20 lux or lower. If the illumination
is greater than the value set as described above, the source of the problem must be identified and
corrective action taken before interpreting mammograms under these conditions.

8th Edition
897N0602G

50

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

10 Printer QC (Annual*) Medical Physicist


OBJECTIVE
The printer used to produce films for final interpretation of mammography images must be cleared
by FDA for that purpose. The objective of this test is to assure that the printer used for final
interpretation is performing according to the manufacturers specifications.

PRINTERS WITH MANUFACTURER-PROVIDED QC PROGRAM


*Follow the printer manufacturers QC program for test frequency, procedure, performance and
corrective action.

PRINTERS WITHOUT MANUFACTURER-PROVIDED QC PROGRAM


If the manufacturer does not provide a mammography printer QC program use the following.

FREQUENCY
This test must be performed initially upon installation before interpreting hard copy mammograms,
and after major repairs, such as those requiring a mammography equipment evaluation (MEE), and
then annually.

REQUIRED TEST EQUIPMENT


QC pattern with a fixed geometric pattern (e.g. TG18-QC or SMPTE RP133). Contact your
printer manufacturer for details.
Metric Ruler
Report form, available from your printer manufacturer.

PROCEDURE
SET UP
1.

Print the QC pattern.

2.

Identify lines in each dimension that have a specified relationship (e.g. fixed length or distance
between lines). Use these same lines for subsequent testing.

3.

Measure the length of lines in the step above to establish a baseline. If the printer manufacturer
provides a test image with specified distances, compare your measurement to the provided
specification.

RECORDING THE OBSERVATIONS

51

1.

Print the QC pattern.

2.

Using the ruler, confirm that the lines are straight and undistorted.

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

3.

Measure the length of the lines.

4.

Inspect the image for artifacts.

5.

Inspect the image for banding and jagged edges of the straight lines.

6.

Note any unusual mechanical noises or other observations.

PRECAUTIONS AND CAVEATS


*Perform this test using the QC pattern supplied (or recommended) by the printer manufacturer for
that printer model. If none is available, use a QC test pattern recommended by the AAPM or by one
of the approved MQSA accreditation bodies for this purpose. Using other patterns could yield
erroneous results. Unless otherwise specified by the printer manufacturer, the visual observations
should be made without magnification.

PERFORMANCE AND CORRECTIVE ACTION


The tolerance for film printer artifacts should be similar to the tolerance for artifacts in screen-film
mammography. According to the 1999 American College of Radiology Mammography Quality
Control Manual, not all artifacts can be totally eliminated. It may be useful to use the concept of
ALARA (as low as reasonably achievable) when assessing artifacts. If they can be easily eliminated,
they should. If the artifact is difficult or expensive to eliminate and is subtle (not mimicking or
obscuring clinical information), it may be tolerable. The medical physicist should consult with the
interpreting physician as to whether the artifact is tolerable. Tolerances for artifacts should be lower
with new imaging equipment.
If any of the following problems exist:
The lines are bowed, bent or distorted
The lines exhibit jagged borders
The printer exhibits print or processing artifact that is judged intolerable
Measured distances vary by more than 1% from the reference standards
Artifacts are objectionable and can mimic or obscure clinical information
The source of the problem shall be identified and corrective action shall be taken before that device
can be used for mammographic image interpretation.
Clinical imaging may be continued. If available, an alternative approved diagnostic display device
may be used for mammographic image interpretation.

8th Edition
897N0602G

52

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

11 Monitor QC (Annual*) Medical Physicist


OBJECTIVE
The objective of this test is to assure that monitors are performing according to the monitor
manufacturers specifications and that they are calibrated (or configured) to perform the DICOM
Grayscale Standard Display Function (GSDF). Both the monitors used for final interpretation of
mammography images (primary monitors), and monitors used for acquisition QC (secondary
monitors) must be tested. An additional requirement is that monitors used for final interpretation
must have been cleared by FDA for that purpose.
Commercial calibration software programs supporting the American Association of Physicists in
Medicine (AAPM) Assessment of Display Performance for Medical Imaging Systems (AAPM On-Line
Report No. 03, www.aapm.org) are available for use with this test.

MONITORS WITH MANUFACTURER-PROVIDED QC PROGRAM


*Follow the monitor manufacturers QC program for test frequency, procedure, performance and
corrective action.

MONITORS WITHOUT MANUFACTURER-PROVIDED QC PROGRAM


If the manufacturer does not provide a mammography monitor QC program use the following.

FREQUENCY
This test must be performed initially before interpreting soft copy mammograms, annually, and after
major repairs, such as those requiring a mammography equipment evaluation (MEE), and whenever
problems are suspected.

REQUIRED TEST EQUIPMENT


For the monitor image quality pattern: AAPM TG18-QC, SMPTE RP133 or other patterns
appropriate for the monitor to be tested (contact your monitor or QC software manufacturer for
details*).
For the monitor luminance response: gray scale images (supplied by the monitor or QC software
manufacturer or AAPM TG18-LN).
Calibrated photometer
The Monitor QC (Annual) report form, your own, or the QC software manufacturer provided
report form. Note: The results of this and other tests can be stored electronically as permitted
by MQSA regulation or guidance.

8th Edition
897N0602G

53

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

PROCEDURE
MONITOR IMAGE QUALITY
1.

For both primary and secondary monitors: display the QC pattern.

2.

Record the maximum and minimum luminance levels as directed by your monitor or QC
software manufacturer. The maximum luminance (Lmax) is typically specified by the monitor
manufacturer as the highest value that can be used without compromising other performance
characteristics, such as lifetime or resolution. The minimum brightness (Lmin) is influenced by
Lmax.

3.

Examine the image carefully to determine the visibility of the following features:
Verify that the 5% signal level inset is visible in the larger 0% signal field.
Verify that the 95% signal inset is visible in the 100% signal field.
Verify that each gray level step from 0% to 100% can be distinguished individually.
Verify that the alphanumeric characters in the image appear sharp and in focus.
Verify that the high-contrast bar images in the center and at the corners of the QC pattern
are distinguishable. The low-contrast patterns are not evaluated in this test.
Record the results of the evaluations on your own form or one provided by the QC software
manufacturer.

MONITOR LUMINANCE RESPONSE


For monitors featuring DICOM GSDF display functionality, confirm that the monitors are set for
DICOM DISPLAY (GSDF) and that the monitors successfully meet the criteria of Monitor Image
Quality Test Procedure step 3. If so, you have successfully completed the Monitor QC (Annual) test.
For monitors that do not provide DICOM DISPLAY (GSDF), or fail the Monitor Image Quality Test
Procedure step 3, perform the GSDF calibration and follow the steps below.
1.

Perform the luminance response test as described in the AAPM report.

2.

Record the results of the evaluation on the Annual Monitor Evaluation report.

PRECAUTIONS AND CAVEATS


*Perform this test using only the test patterns described above. Some test patterns are specific to
the image matrix size and bit-depth. Using other patterns could yield erroneous results.

PERFORMANCE
If the following conditions are met, the Monitor QC (Annual) check is acceptable:
MONITOR IMAGE QUALITY:
The maximum luminance value (Lmax) is within the range specified by the monitor
manufacturer. For reference, the AAPM report recommends that once set, the target level
(Lmax) is maintained within 10% of the desired value.
The 5% inset in the 0% field and the 95%- inset in the 100% field shall be visible and the
0% to 100% gray level steps shall be distinguishable from the adjacent steps.
The alphanumeric characters appear sharp and focused.
54

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

The high contrast line-pair patterns are distinguishable at the center and the corners of the
display in both the horizontal and vertical orientations.
NOTE
The ability to distinguish the low-contrast line-pair patterns is not a requirement for this test.
Unless required by the monitor manufacturer, there is no action limit specification for Lmin in this
procedure.

MONITOR LUMINANCE RESPONSE


Both primary and secondary monitors must be calibrated to DICOM Grayscale Standard Display
Function (GSDF). The luminance differences between each measured value must agree with the
expected difference associated with the GSDF and the measured contrast response at any given
point must not exceed 10% of the standard for primary monitors and must not exceed 20% for
secondary monitors.

CORRECTIVE ACTION
For primary monitors (used for final interpretation). If this test produces results that fall outside the
action limits as specified by the manufacturer, the source of the problem shall be identified and
corrective action shall be taken before that device can be used for mammographic image
interpretation. Clinical imaging may be continued. If available, an alternative approved diagnostic
display device may be used for mammographic image interpretation, provided it has passed the
applicable QC tests.
For secondary monitors (used for image acquisition QC). If this test produces results that fall outside
the action limits as specified by the manufacturer, the source of the problem shall be identified and
corrective action shall be taken on the monitor that failed before any further examinations are
performed using that monitor. If the monitor that failed is replaced by an alternative monitor that
passes the test, image acquisition may resume using the alternative monitor.

8th Edition
897N0602G

55

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

MONITOR QC (ANNUAL) REPORT FORM


DATE:

Room:

Final Interpretation?

YES

NO

Workstation:

Cleared by FDA?

YES

NO

NA

Left Monitor Serial


Number:
Right Monitor Serial
Number:
mAs:

MONITOR IMAGE QUALITY REPORT


Left or Single Monitor
Monitors have DICOM Display Functionality feature applied

YES

NO

Right Monitor
YES

NO

Lmax target:
Lmax measured:
(Lmax measured Lmax target)/Lmax target x 100%:

Within 10%:

YES

NO

YES

NO

5% patch is visible in 0% patch?

YES

NO

YES

NO

95%-patch is visible in 100% patch?

YES

NO

YES

NO

Gray steps distinguishable?

YES

NO

YES

NO

Alphanumeric characters sharp?

YES

NO

YES

NO

Line-pair images (center), distinguishable?

YES

NO

YES

NO

Line-pair images (corners) distinguishable?

YES

NO

YES

NO

Overall, all tests pass?

YES

NO

YES

NO

LUMINANCE RESPONSE*
Left or Single Monitor

Right Monitor

Measured Result ok?

Measured Result ok?

Measured contrast response at any given point does


not deviate more than a) 10% of the standard for
final interpretation or b) 20% of the standard for
other use.

* It is not necessary to measure Monitor Luminance Response on monitors that feature DICOM GSDF display
functionality and pass the Monitor Image Quality test.
SIGNATURE:

56

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

12 S Value Confirmation (Annual) Medical Physicist


OBJECTIVE
To assure that the amount of exposure reaching the IP is assigned a proper S value.

FREQUENCY
This test must be performed initially before any patients are imaged and then annually as part of
routine service PM (preventive maintenance), and whenever the FCRm/CRm reader unit PMT
(photomultiplier tube) calibration is suspect.
NOTE
This is a test of the FCRm/CRm reader unit and only needs to be conducted using one X-ray generator
per reader.

REQUIRED TEST EQUIPMENT


Calibrated mammography dose meter
Watch/timer
The 18 x 24 cm cassette and IP dedicated for QC
S VALUE CONFIRMATION data form

PROCEDURE
1.

Using the designated QC cassette, erase the IP using the secondary erasure mode on the
reader.

2.

Remove the compression paddle from the field and place the dose meter on the bucky surface
as described in illustration.

3.

Select 25kVp and a Mo target and filter combination. Adjust the mAs to give the dose meter
reading closest to but not less than 20 mR. Record the technique.
NOTE
If the mammography X-ray unit will not allow 20 mR exposure, you may use the lowest mR
exposure available that is greater than 20. Always use 25 kVp for this test.
Do not use any additional filter except the designated Mo filter. The use of another filter may
cause an error.

8th Edition
897N0602G

4.

Once you have established the proper technique for your X-ray generator, make a series of at
least three exposures prior to removing the meter or placing the cassette in the field. Record
the exposure technique and record the measured mR for each exposure.

5.

Verify that the mR readings are within 5% of each other. If they are not, repeat the test on a
more reproducible X-ray unit.

6.

Average the mR readings from the three test exposures.

7.

Remove the dose meter from the beam and place the erased Fuji IP / Cassette on the top of the
breast support surface centered in the X-ray field.

57

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

8.

Expose the cassette at the predetermined settings. Wait 10 minutes and process the IP using
the Physics, Sensitivity menu. Repeat this step two additional times creating a total of three
images. Record the S value for each exposure.

9.

Calculate the exposure average(mR) and S value average. Divide the averaged exposure
results by 20(mR) where 20(mR) is a fixed standard value. Multiply by the averaged S value to
obtain the corrected S value.
Example: 25 kVp
Averaged exposure = 30.0mR
Averaged S value = 83
Corrected S value = 30.0mR/20mR 83 = 124.5

10. Record the corrected S value on the chart

PRECAUTIONS AND CAVEATS


The measurement of 20 mR requires a highly sensitive and accurate dose meter. Contact your
medical physicist or Fuji service engineer and confirm that he is equipped with a calibrated device
for use in this test. If other measuring devices are used, some error in measurement may occur.

PERFORMANCE AND CORRECTIVE ACTION


The corrected S value shall not exceed the range of 120 20% (96 Corrected S value 144).
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken within thirty days of the test date. Clinical
imaging and mammographic image interpretation may be continued during this period.

58

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Measure the dose using a mammographic dose meter at the location shown below.
1824 cassette tube side view
5-1/4"
132mm

2-1/4
60mm

Cassette top view

Green orientation label

8th Edition
897N0602G

Chest wall

59

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

S VALUE CONFIRMATION
DATE:
Exposure Unit ID:
Reader Unit ID:
Dose Meter ID:
kVp:
mAs:
Exposure 1:
Exposure 2:
Exposure 3:
Average:

mR readings are within 5%?

IP Exposure Time 1:

IP Read Time 1:

IP Exposure Time 2:

IP Read Time 2:

IP Exposure Time 3:

IP Read Time 3:

Yes

No

S value 1:
S value 2:
S value 3:
Average:
Corrected S value:

ACCEPTANCE LIMITS:
Corrected S value (must be between 96 and 144):
Corrective Action Needed:

PASS

FAIL

YES

NO

YES

NO

PASS

FAIL

Corrective Action Description:


Corrective Action Taken:
Corrective Date:
Results of New Test:

Comments:

SIGNATURE:
60

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

13 System Resolution (Annual) Medical Physicist


OBJECTIVE
To measure the spatial resolution of the imaging chain.

FREQUENCY
Must be performed initially before any patients are imaged and then annually at routine PM and
whenever the system resolution is suspect.

REQUIRED TEST EQUIPMENT


Line pair test tool measuring up to 10 lp/mm or greater
A 4.0 cm thick uniform sheet of defect free acrylic large enough to cover a 18 x 24 cm
mammographic cassette.
Magnifying lens of at least 7X magnification
The dedicated 18 x 24 cm Fuji IP Cassette and IP designated for QC
SYSTEM RESOLUTION data form

PROCEDURE
1.

Using the designated cassette, erase the IP using the secondary erasure mode on the reader unit.

2.

Insert the cassette in the cassette holder.

3.

Place the acrylic attenuating material on the surface of the bucky.

4.

Center the resolution pattern laterally, on top of the absorber and along the chest wall edge at
a slight (3 to 5 degree) angle.
In this position, the patterns bars lie parallel to the anode-cathode axis for the first image.

5.

Select the focal spot used for imaging an average breast and set a manual exposure based on
the kVp and mAs established for the phantom image test.

6.

Process the IP using the "Physics, ACR MAPP" menu.

7.

Repeat steps 1-6 with the bars oriented perpendicular to the anode-cathode axis and with the
highest line frequency of the bar phantom positioned against the chest wall at a slight (3 to 5
degree) angle.

PRECAUTIONS AND CAVEATS


The image background density will affect the visibility of the test tool. The density is controlled by
image processing. Magnify the image on the workstation monitor so the density of the finer line pairs
can be evaluated. Window/level adjustment (GA/GS) can be used on the CR Console/AWS-c to
fine- tune density prior to printing or viewing on a mammographic diagnostic workstation.

8th Edition
897N0602G

61

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

DATA INTERPRETATION AND ANALYSIS


1.

View the resolution pattern with magnification as needed.

2.

Note the highest frequency pattern where lines are distinctly visible throughout at least half of
the bar length and record.

PERFORMANCE AND CORRECTIVE ACTION


Images displayed for final interpretation shall be capable of demonstrating 8 lp/mm 2 lp/mm of
system resolution in both the IP-scan and IP-subscan directions.
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken on the component(s) that caused the failure
before any further examinations are performed using the failed component(s). If the component(s) that
caused the failure (e.g. exposure unit, CR reader unit, display device) is replaced by an alternative device
and the mammography system passes the re-test, image acquisition and interpretation may continue
using that combination of devices.

62

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

SYSTEM RESOLUTION
DATE:
Exposure Unit ID:
Reader Unit ID:

Parallel to Anode-Cathode Axis


S VALUE:
OPITICAL DENSITY of IMAGE BACKGROUND:
(Hardcopy only)

ACCEPTANCE LIMITS
8 lp/mm 2 lpmm
CORRECTIVE ACTION NEEDED?

PASS

FAIL

YES

NO

YES

NO

PASS

FAIL

CORRECTIVE ACTION DESCRIPTION:


CORRECTIVE ACTION TAKEN?
CORRECTIVE DATE:
RESULTS OF NEW TEST:

Perpendicular to Anode-Cathode Axis


S VALUE:
OPITICAL DENSITY of IMAGE BACKGROUND:
(Hardcopy only)

ACCEPTANCE LIMITS
8 lp/mm 2 lpmm
CORRECTIVE ACTION NEEDED?

PASS

FAIL

YES

NO

YES

NO

PASS

FAIL

CORRECTIVE ACTION DESCRIPTION:


CORRECTIVE ACTION TAKEN?
CORRECTIVE DATE:
RESULTS OF NEW TEST:

SIGNATURE:

8th Edition
897N0602G

63

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

14 CR Reader Unit Scanner Performance (Annual) Medical Physicist


OBJECTIVE
To establish that the IP reader unit and printer optics and transport system do not exhibit scan or print
jitter.

FREQUENCY
Must be performed initially before any patients are imaged and then annually at routine PM and
whenever the laser optics- related components are serviced.

REQUIRED TEST EQUIPMENT


One pair of 6 inch steel rulers
The dedicated 24 x 30cm Fuji IP Cassette and IP designated for QC
LASER PERFORMANCE TEST data form
NOTE
This is a test of the CR Reader and only needs to be conducted using one X-ray generator per
Reader.

PROCEDURE
1.

Using the designated cassette, erase the IP using the secondary erasure mode on the reader.

2.

Position the cassette for a non-grid exposure, centered so the X-ray beam covers as much of
the cassette front as possible.

3.

Place the steel rulers on the cassette surface so the rulers form the letter T.
Place the T into the X-ray field centered along the chest wall side of the cassette.

4.

Expose and process the cassette using the same exposure and image processing menu as
used in the S Value Confirmation test.

PRECAUTIONS AND CAVEATS


Adjust exposure or the window/level settings so that the contrast of the rulers vs. background
density creates a bright white T against the background density. The rulers forming the T must
have smooth borders free from jagged edges or defects.

64

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

PERFORMANCE AND CORRECTIVE ACTION


Upon inspection, the edges of the T should appear smooth and sharp. After reviewing the T,
look at the white frame surrounding the image area of the film. Linear jagged edges visible to the
naked eye indicate scan or print error of the IP reader unit or printer. The workstation or positioning
monitor can be of help in evaluating the error caused by the IP reader unit. If images are well displayed
on those devices, the film printer may be suspect.
If the printer generates excess jitter, mammography images can be saved within the workstation or
network until the printer laser optics and transport is serviced.
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken on the component(s) that caused the failure
before any further examinations are performed using the failed component(s). If the component(s) that
caused the failure (e.g. exposure unit or CR reader unit) is replaced by an alternative device and the
mammography system passes the re-test, image acquisition and interpretation may continue using
that combination of devices.
If the failure has been determined to be of a diagnostic device used for mammographic image
interpretation (e.g. laser printer, physicians review station) the source of the problem shall be
identified and corrective action shall be taken before that device can be used for mammographic
image interpretation. Clinical imaging may be continued. If available, an alternative approved
diagnostic device may be used for mammographic image interpretation.

8th Edition
897N0602G

65

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

CR Reader Unit Scanner PERFORMANCE TEST


DATE:
Reader Unit ID:
Image Reader ID:

Parallel to Chest Wall side

ACCEPTANCE LIMITS
WORKSTATION MONITOR
CORRECTIVE ACTION NEEDED?
LASER FILM PRINTER*
CORRECTIVE ACTION NEEDED?

PASS

FAIL

YES

NO

PASS

FAIL

YES

NO

YES

NO

PASS

FAIL

CORRECTIVE ACTION DESCRIPTION:


CORRECTIVE ACTION TAKEN?
CORRECTIVE DATE:
RESULTS OF NEW TEST:

Perpendicular to Chest Wall side

ACCEPTANCE LIMITS
WORKSTATION MONITOR
CORRECTIVE ACTION NEEDED?
LASER FILM PRINTER*
CORRECTIVE ACTION NEEDED?

PASS

FAIL

YES

NO

PASS

FAIL

YES

NO

YES

NO

PASS

FAIL

CORRECTIVE ACTION DESCRIPTION:


CORRECTIVE ACTION TAKEN?
CORRECTIVE DATE:
RESULTS OF NEW TEST:
* If hard copy is used for final interpretation

SIGNATURE:

66

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

15 Mammographic Unit Assembly Evaluation (Annual)


Medical Physicist
OBJECTIVE
To ensure that all locks, detents, angulation indicators, and mechanical support devices for the
X-ray tube and image receptor holder assembly are operating properly.

FREQUENCY
This test must be performed at installation of the mammography exposure unit and on an annual
basis and as part of a Mammography Equipment Evaluation (MEE).

REQUIRED TEST EQUIPMENT


MAMMOGRAPHIC UNIT ASSEMBLY EVALUATION data form

PROCEDURE
1.

Verify that the freestanding dedicated mammography unit is mechanically stable under normal
operating conditions.

2.

Verify that all moving parts move smoothly, without undue friction, that cushions or bumpers
limit the range of available motions, and that no obstructions hinder the full range of motions
within these limits.

3.

Set and test each lock detent independently to ensure that mechanical motion is prevented
when the lock or detent is set.

4.

Verify that the image receptor holder assembly is free from wobble or vibration during normal
operation.

5.

Verify that the image receptor slides smoothly into the proper position in the image receptor
holder assembly and that the image receptor is held in place securely by the image receptor
compartment for any orientation of the image receptor holder assembly.

6.

If provided, verify that the compressed breast thickness scale (analog or digital) is accurate
to within 0.5 cm under conditions of moderate compression (15 - 20 lbs.) and reproducible
to within 2mm between 1 and 8 cm. The phantom used for this test should be large enough
to simulate a typical breast and be positioned towards the chest-wall side of the bucky. The
compressed breast thickness should be measured at the center of the chest-wall position of
the automatic exposure control (AEC) sensor. This should be done for at least the small and
large image receptors and compression paddles.

7.

Verify that in normal operation, the patient and operator are not exposed to sharp or
rough edges or other hazards including electrical hazards.

8.

Perform the decompression test as described in MQSA final regulations and record the result
on the Mammographic Unit Assembly Evaluation Chart.

9.

Verify that the operator is protected by adequate radiation shielding during exposure.

10. Verify that all indicator lights are working properly.

8th Edition
897N0602G

67

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

PERFORMANACE AND CORRECTIVE ACTION


Items that are hazardous, or operate improperly must be repaired by appropriate service personnel
before further exams are performed with this unit.
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken within thirty days of the test date. Clinical
imaging and mammographic image interpretation may be continued during this period.

68

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

MAMMOGRAPHIC UNIT ASSEMBLY EVALUATION


DATE:
Facility:

Room#:

X-ray Generator:
Pass =
Fail = F
Does not apply = NA
1.

Free-standing unit is mechanically stable

2.

All moving parts move smoothly, without obstructions to motion

3.

All locks and detent work properly

4.

Image receptor holder assembly is free from vibrations

5.

Image receptor slides smoothly into holder assembly

6.

Image receptor is held securely by assembly in any orientation

7.

Compressed breast thickness scale accurate to 0.5 cm, reproducible to 2 mm

8.

Patient or operator is not exposed to sharp or rough edges, or other hazards

9.

Operator technique control charts are posted

10. Operator protected during exposure by adequeate radiation shielding


11. All indicator lights working properly
12. Auto decompression can be overridden to maintain compression with continuous display
of the override status.
13. Manual emergency compression release can be activated in the event of a power failure
Comments:

SIGNATURE:

8th Edition
897N0602G

69

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

16 Collimation Assessment (Annual) Medical Physicist


NOTE
This test is intended to evaluate the mammographic X-ray unit, not the CR reader.

OBJECTIVE
To assure that the collimator or cone allows full coverage of the image receptor by the X-ray field
but does not allow significant radiation beyond the edges of the image receptor. To assure that the
chest wall edge of the compression paddle aligns with the chest wall edge of the image receptor
without showing on the image.

FREQUENCY
This test must be performed upon installation or major service of the X-ray tube or collimator, when
a problem related to collimation is suspected, and annualy thereafter.

REQUIRED TEST EQUIPMENT


Five coins, four of one size and one of a larger size.
Four mammographic cassettes and IP from the clinical inventory: one small and three large.
A uniform 2-cm thickness of defect free acrylic large enough to cover a 24 x 30 cm
mammographic cassette.
Accurate ruler.
COLLIMATION ASSESSMENT data form

PROCEDURE

70

1.

Place an appropriately sized cassette loaded with an erased IP in the normal orientation in the
image receptor holder.

2.

Load an erased IP in the second cassette.

3.

Place the large cassette on top of the image receptor holder with the front of the cassette
toward the X- ray source and assure that the large cassette extends beyond the image
receptor holder on the chest-wall side by about 1 cm.

4.

Place the collimator to be evaluated in position.

5.

Remove the compression paddle. (The compression paddle should be removed before
placement of the coins to assure a sharp demarcation at the edges of the light field.)

6.

Turn on the collimator light and place the four identical smaller coins inside the light field with
one edge of each coin just touching the edge of the light field. The coin on the chest-wall side
should be shifted to the right of center about 2 inches so it does not superimpose the AEC
detector.

7.

Replace the compression paddle and position it 4.2 cm from the breast support.

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

8.

Tape the larger coin underneath the compression paddle shifted about 2 inches to the left so it
does not superimpose the AEC detector. Be sure the coin's outer edge is tangent to the inner
lip of the chest-wall side of the compression paddle. This coin marks the chest-wall edge of
the paddle.

9.

Place a sheet of acrylic attenuating material on top of the paddle, so that all radiation reaching
the cassettes must pass through the attenuator.

10. Make an exposure using AEC at the lowest setting on the X-ray generator and process the IPs
using the Physics, Collimation menu.
11. Repeat steps 1 through 9 for all routinely used collimator/bucky/compression paddle
combinations and target materials. When testing the large image receptor, the top cassette
may be positioned diagonally to capture all four edges of the X-ray field, or two large cassettes
may be used on top of the image receptor holder.

Large cassette
(on the image receptor holder)
Coin

Illuminated area by
collimator light
Coin

Coin

Coin
Chest Wall Side (Top view)

Figure 1 Arrangement of Procedure Step 6

Comperssion paddle

Large coin

Chest Wall Side (Top view)

Figure 2 Arrangement of Procedure Step 8

8th Edition
897N0602G

71

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Attenuating material
Compression paddle
Coin
Large cassette

Coin

Coin

Coin

Image receptor holder


Chest Wall Side
(Side view)

Figure 3 Arrangement of Procedure

DATA INTERPRETATION AND ANALYSIS


From the image created in the top cassette, measure the deviation between the X- ray field (dark
portion of the image) and the edge of the light field (defined by the exterior edges of the four smaller
coins) for all four sides of the field. The magnitudes of the deviations at the left edge and right edge
(ignoring + or - signs) should be entered on the data form and added together. Similarly, the
deviations at the anterior and posterior (chest-wall) edges should be entered (without regard to sign)
and the magnitudes added together. Record the unit source-to-image distance (SID) on the data
form and calculate the % SID by dividing each sum by the SID and multiplying by 100.
Measure the deviations between the edges of the X- ray field and all four sides of the image
receptor. Use the Image that was placed in the image receptor holder for this measurement. This
can be done by individually aligning the outer edges of the smaller coin on both images and
measuring the distance that the X- ray field edge of the top image extends beyond the image in the
image receptor holder.
Note that slight magnification differences between the two images should be taken into account.
Enter the measured deviations between the X-ray field and image receptor holder image on the
data form. If the X- ray field extends beyond the image, it should be given a "+" sign; if it falls within
the image, it should be given a "-" sign. Calculate the % SID for each side, retaining the + or - signs.
Next, measure the deviation between the edge of the compression paddle (delineated by the outer
edge of the large coin) and the edge of the image receptor. When measuring the distance, note the
difference in sizes of the larger coin on the two images. (The coin image will be bigger on the
bottom image.) Distances should be measured on or referred to the image in the image receptor
holder cassette.

72

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Small
coin

Exposure area
Small
coin

Small
coin

Small
coin

Large
coin

(Chest Wall Side)

Figure 4 Image obtained from the cassette


in the receptor holder

Unexposed area (large imaging plate)

Small
coin

Small
coin

Collimated
exposure
area

Large
coin

Small
coin

Small
coin

(Chest Wall Side)


Figure 5

8th Edition
897N0602G

Image obtained from the cassette


on the top of the receptor holder

73

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

PERFORMANACE AND CORRECTIVE ACTION


All systems shall have beam-limiting devices that allow the entire chest wall edge of the X-ray field to
extend to the chest wall edge of the image receptor and provide means to assure that the X-ray
field does not extend beyond any edge of the receptor by more than 2% of the SID.
If a light field that passes through the X-ray beam limitation device is provided, it shall be aligned
with the X-ray field so that the total of any misalignment of the edges of the light field and the X-ray
field along either the length or the width of the visually defined field at the plane of the breast support
surface shall not exceed 2% of the SID.
The chest wall edge of the compression panel shall not extend beyond the chest wall edge of the
image receptor by more than 1 % of the SID when tested with the compression paddle placed
above the breast support surface at a distance equivalent to standard breast thickness. The shadow
of the vertical edge of the compression paddle shall not be visible on the image.
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken within thirty days of the test date. Clinical
imaging and mammographic image interpretation may be continued during this period.

74

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

COLLIMATION ASSESSMENT
DATE:
Exposure Unit ID:
Source to Image Receptor Distance (SID) in mm:
Deviation between X-ray field and light field:
Collimator
Target Material
Left Edge deviation (mm):
Right Edge deviation (mm):
Sum of left and right edge deviations:
Sum as % of SID
Anterior Edge deviation (mm)
Chest Edge deviation (mm)
Sum of anterior and chest wall deviations (mm)
Sum as % of SID
Action Limit:

18 24 cm

24 30 cm

If a light field that passes through the X-ray beam limitation device is provided, it shall be aligned with the X-ray field so that the
total of any misalignment of the edges of the light field and the X-ray field along either the length or the width of the visually
defined field at the plane of the breast support surface shall not exceed 2% of the SID.

Difference between X-ray field and image receptor:


Collimator
Target Material
Left Edge deviation (mm)
Difference as % of SID
Right Edge deviation (mm)
Difference as % of SID
Antererior Edge deviation (mm)
Difference as % of SID
Chest Wall deviation (mm)
Difference as % of SID
X-ray Field Extends to Chest Wall
Action Limit:

18 24 cm

24 30 cm

Yes / No

Yes / No

All systems shall have beam-limiting devices that allow the entire chest wall edge of the X-ray field to extend to the chest wall
edge of the image receptor and provide means to assure that the X-ray field does not extend beyond any edge of the receptor
by more than 2% of the SID.

Alignment of chest wall edges of compression paddles and image receptors:


Collimator

18 24 cm

24 30 cm

Yes / No

Yes / No

Target Material
Difference between compression
paddle edge and image edge
Difference as % of SID
Paddle Edge Visible on Image Receptor?
Action Limit:

If the chest wall edge of compression paddle is within the image receptor or projects beyond the chest wall edge of the image
receptor by more than 1% of SID, seek service correction.

JUDGEMENT:

PASS

FAIL

Comments:

SIGNATURE:
8th Edition
897N0602G

75

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

17 Automatic Exposure Control (AEC) System Performance


Assessment (Annual) Medical Physicist
OBJECTIVE
To assess the performance of the mammography X-ray exposure units AEC system with regard to
image quality, thickness tracking, and reproducibility and to check the density control (see note)
selector function.
NOTE
This procedure assesses the performance of the X-ray exposure unit. FCRm/CRm is a digital imaging
system. FCRm/CRm controls film density or image brightness to a specific level when using the auto or
semi auto modes such as those used in routine mammographic projections. Unlike screen-film systems,
FCRm/CRm image brightness or film density is not directly controlled by exposure.

FREQUENCY
This test must be performed prior to the first clinical use with FCRm/CRm on each mammographic
exposure unit, annually thereafter, and as part of a mammography equipment evaluation (MEE).

REQUIRED TEST EQUIPMENT


Required - Large acrylic (PMMA) slabs capable of covering a 24x30cm field and combined to
produce thicknesses of 2, 4, and 6cm are required to perform the CNR Per Object Thickness
test.
Optional- for the other non-imaging tests (AEC Density Control Function and AEC
Reproducibility), standard size 10x12.5cm acrylic or BR-12 absorbers can be used.
A 0.2mm thick Al sheet approximately 10x10cm.
One 1824cm and one 24x30cm image receptor (cassette with IP) from the clinical inventory
and the 18x24cm dedicated QC receptor.
Ionization chamber and electrometer calibrated at mammographic X-ray beam energies.
AEC System Performance Assessment forms.

PROCEDURE
AEC DENSITY CONTROL FUNCTION

76

1.

Select a loaded 18x24cm mammography cassette from the clinical inventory. Erase the IP
using the secondary erasure mode on the image reader. Place the cassette in the bucky.

2.

Position a 4cm thick absorber on the patient support. Lower the compression device to
contact with the absorber. Make sure that the absorber completely covers the AEC detector
area clinically used for a 4cm thick compressed breast.

3.

Select the kVp used to image the ACR MAP phantom and record the value on the chart
labeled DENSITY CONTROL FUNCTION.

4.

Expose the 4 cm absorber at the AEC systems 0 density control setting. Do not process the
IP in the image reader.

5.

Record the mAs delivered and the AEC density control setting on the chart.

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

6.

Repeat steps 4-5 for the other density control settings used clinically.

7.

Erase the IP (which has been exposed multiple times without processing and erasure) using the
primary erase mode on the FCRm/CRm reader unit.

AEC REPRODUCIBILITY and IMAGE MODE TRACKING


1.

Select a loaded 18x24cm mammography cassette from the clinical inventory; erase the IP using
the secondary erasure mode of the FCRm/CRm reader unit.

2.

Insert the loaded cassette in the Bucky.

3.

Place the 4cm thick absorber in the X- ray field, lower the compression device into contact
with the absorber.

4.

For systems equipped with multiple positions for the AEC sensor, select the position used
most often clinically. Make sure that the absorber completely covers the AEC sensor position
most often used clinically.

5.

Prepare the mammography X-ray unit for operation in the AEC mode, selecting the density
control setting that is most often used clinically (typically 0).

6.

Position the ion chamber in the X-ray beam without covering the AEC sensitive area.

7.

Expose the cassette and IP using a fixed kVp and target/filter combination typically used for a
4cm breast.

8.

Record the resultant mAs and the measured air kerma of each exposure on the data form
titled REPRODUCIBILITY CHART.

9.

Repeat steps 6 and 7 three more times, completing the AEC REPRODUCIBILITY test.

10. Erase the exposed IP using the Primary Erasure mode of the FCRm/CRm reader unit.
11. Using AEC, expose the 4cm thick absorber for each imaging mode that is used clinically on
that X-ray unit (e.g. small image receptor and grid, large image receptor and grid, magnification
mode with small focal spot, and no grid). It is not necessary to measure or record air kerma
during the IMAGE MODE TEST.
12. Enter the results of each exposure on the IMAGE MODE TRACKING form.
13. Erase all exposed IPs using the Primary Erasure mode of the FCRm/CRm reader unit.
CNR PER OBJECT THICKNESS

8th Edition
897N0602G

1.

Using the designated 18x24cm QC mammography cassette and IP, erase the IP using the FCRm/
CRm reader unit's secondary erasure mode. Place the loaded cassette in the 18x24cm bucky.

2.

Place the 2cm thick 24x30cm acrylic absorber on the patient support. The absorber may
extend beyond the chest wall side of the 18x24cm patient support.

3.

Lower the compression device into contact with the absorber.

4.

Move the AEC sensor to its most forward position (closest to the chest wall).

77

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

5.

Place the 0.2mm Al sheet on the top of the compression device (not directly on top of the
absorber) and position it as shown in the diagram. Use the field indicator light as a guide for
the center of the bucky and to identify the AEC sensor positions. Position the Al sheet as
shown in the diagram so that the Al covers the indicated Calculation Area while positioned as
far as possible from the most forward AEC sensor position.

Calculation Areas
(20mm 20mm)
Aluminum Object

20mm

BG

Al

20mm
60mm

20mm

Acrylic Slabs

Chest Wall Side


6.

Make an AEC exposure using the AEC MODE and kVp normally used clinically for a 2cm thick
compressed breast.

7.

After the predetermined interval, process the IP using the Physics Sensitivity menu.
NOTE

To control the effect of image fading with time on this test, it is important to control the time
interval between X-ray exposure and reading of the IP so that the interval be consistent whenever
this test is performed. The interval may be as short as possible consistent with the testing
workflow. We recommend trying an interval between 5 and 10 minutes, adjusting as necessary,
and then using the same interval consistently.

8.

Calculate CNR of the image using the QC calculation tool function on the CR Console/AWS-c,
and record the kVp, target, filter, mAs, and CNR value on the CNR PER OBJECT THICKNESS
data form.

9.

Repeat steps 6 through 8 using 4cm and 6cm thick 24x30 acrylic absorbers using the kVp
and AEC exposure settings normally used clinically for 4 and 6cm compressed breasts
respectively.

10. Assigning the CNR value from the 4cm exposure a value of 100%, calculate the relative CNR
percentage values for 2cm and 6cm and record them on the data form.

PRECAUTIONS AND CAVEATS


Use the 18x24cm QC IP and cassette for the CNR PER OBJECT THICKNESS test.
Always wait the same time between exposure and processing of the IP when performing this test.
CNR will change after major service or replacement events. These include but are not limited to:
X-ray tube replacement
Filter replacement
Replacement of the compression paddle
Change in phantom used
Change of grid or table surface
Change of X-ray generator calibration
Change of CR Reader calibration
78

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

PERFORMANCE AND CORRECTIVE ACTION


DENSITY CONTROL FUNCTION
Each increase or decrease of the AEC DENSITY setting should result in a 5 to 15% change of mAs.
REPRODUCIBILITY
The coefficients of variation for either air kerma and mAs must not exceed 0.05

IMAGE MODE TRACKING


There should not be a significant difference in exposure between the small bucky and the larger
bucky when employing similar grid types.
CNR PER OBJECT THICKNESS
This is an annual test. Establish a new CNR 4cm reference value each time this test is performed. (The
technologist monitors CNR over time by performing the CNR Weekly test.)
When evaluating the performance of the AEC system, the following two trends should be
demonstrated:
1). First, at a given beam quality (kVp, target, and filter), mAs should increase with increasing
phantom thickness.
2). Second, at a given phantom thickness, the mAs should decrease with increasing kVp.
Evaluate CNR by establishing a baseline CNR value from the 4 cm phantom image and assign the 4
cm CNR image a value of 100%.
The relative CNR of the 2cm image must be equal to or greater than 100%.
The relative CNR of the 6cm image must be equal to or greater than 75%.
If this test produces results that fall outside these action limits, corrective action shall be taken within
thirty days of the test date. Clinical imaging and mammographic image interpretation may be
continued during this period.

8th Edition
897N0602G

79

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Automatic Exposure Control (AEC) System Performance Assessment


DATE:
Small Cassette ID:

Large Cassette ID:

Reader Unit ID:

Exposure Unit ID:

DENSITY CONTROL FUNCTION CHART


Imaging mode:

Small Image Receptor with Grid

Focal Spot:

Large Focal Spot

Phantom thickness:

4cm

kVp:
AEC Density
Control Setting
+5

% mAs
change

mAs

Acceptable level
PASS / FAIL

+4

PASS / FAIL

+3

PASS / FAIL

+2

PASS / FAIL

+1

PASS / FAIL

0 (Normal)

0%

5-15% difference per step

PASS / FAIL

-1

PASS / FAIL

-2

PASS / FAIL

-3

PASS / FAIL

-4

PASS / FAIL

-5

PASS / FAIL

REPRODUCIBILITY CHART
Imaging mode:

Small Image Receptor with Grid

Focal Spot:

Large Focal Spot

Phantom thickness:

4cm

kVp:
Exposure Number

Air kerma

mAs

Acceptance

Action Limit

1
2
3
4
Mean values
Standard
Deviations (SD)
Coefficients of
variation (CV)

PASS / FAIL

The coefficients of variation for both air


kerma and mAs must not exceed 0.05

Continued on the following page.

80

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

DATE:
Small Cassette ID:

Large Cassette ID:

Image Reader ID:

Exposure Unit ID:

IMAGE MODE TRACKING


Absorber: 4 CM
Focal Spot: L
Density Control:
Technique

AEC MODE

kVp

Target

Filter

mAs

Comments

18 x 24 Bucky
24 x 30 Bucky
Mag / No Grid

CNR PER OBJECT THICKNESS


Imaging Mode:
Focal Spot:
Density Control:
AEC Mode(s):
Phantom
Thickness

CNR
kVp

Target

Filter

mAs

CNR Measured

(relative to
4cm PMMA)

2 cm

Acceptable
level
>100%

4 cm

100%

6 cm

100%
>75%

PASS / FAIL

SIGNATURE:

8th Edition
897N0602G

81

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

18 System Artifact Evaluation (Annual) Medical Physicist


OBJECTIVE
To assess the degree and source of artifacts visualized on hardcopy mammograms or phantom
images. This procedure allows the source of artifacts to be traced to X- ray equipment, cassette, IP,
FCRm/CRm reader, or printer/processor so that appropriate measures for elimination of artifacts can be
taken.

FREQUENCY
This test must be performed at installation and annually and whenever artifact evaluation is needed.

REQUIRED TEST EQUIPMENT


A uniform 4-cm thick (approximately) sheet of defect free acrylic large enough to cover a 2430
cm mammographic cassette
The previously selected QC cassettes and IPs of each size
For hardcopy only, mask the illuminated portion of the viewbox outside the area of the film
For hardcopy only, an optical densitometer
SYSTEM ARTIFACT EVALUATION data form

PROCEDURE

82

1.

Use technique factors normally used clinically, choosing the lowest kVp setting used clinically (to
be most sensitive to artifacts). If AEC is normally used, use it in this test. Start with the most
commonly used image receptor size (usually 18 24 cm). Record the technique factors on the
data form.

2.

Place a uniform sheet of acrylic that is large enough to cover the mammographic cassette and
thick enough to have an exposure time of 0.5 second or greater on the image receptor holder
assembly.

3.

Erase the IP using Primary Erase and insert the cassette into the cassette holder.

4.

Position a lead marker such as the number 1 or an arrow on the acrylic sheet in a corner of
the radiation field, preferably outside the normal location of the breast, and pointing along the
long axis of the IP and cassette.

5.

Make an exposure.

6.

Process the IP using the Physics, Artifacts menu. Record all pertinent information on the data
form for this test.

7.

Measure the optical density in the center of the printed film, verifying that it is greater than 1.20,
and record on the data form for this test.

8.

Repeat this process for the other image receptor size. Also, repeat for all available focal spot
sizes, targets and filters used clinically.

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

PRECAUTIONS AND CAVEATS


If artifacts are visible on hardcopy but not during workstation review, the cause of the artifact could
be due to the image printer or film processor.
Compare printed images against the same images viewed on the FCRm/CRm workstation using
magnification and rotation if helpful. There is no need to change window/level or image processing.
If the artifacts appear on both hardcopy and softcopy, the cause of the artifact could be the X-ray
generator, FCRm/CRm reader unit, cassette or IP.
X-ray generator:
Potential sources of artifact include irregular density or projection artifacts from the X-ray tube,
filter, collimator, compression paddle, image receptor holder or grid. To help isolate artifacts
caused by the X-ray generator, produce a second image with the cassette rotated 180 degrees
compared to the first image. If the artifact remains the same relative to the position of the marker
it is likely coming from the exposure system. If the artifact reverses relative to the marker it is likely
coming from the FCRm/CRm reader unit, cassette/IP or image printer.
To help isolate grid artifact, the IP cassette can be exposed on top of the bucky. A non grid image
will help identify artifacts caused by uneven absorption of the bucky surface and the quality of the
grid. If the compression paddle is suspect, images can be produced with and without the
compression paddle in the beam. Contact the X-ray generator manufacturer for additional
information.
FCRm/CRm Reader Unit:
Potential causes of artifact in the FCRm/CRm reader unit are misalignment or dirt on the light
collector, improper imaging parameter setup or an error in spatial uniformity of the digitization
process.
Misalignment of the light collector can produce non-uniform density across the long axis of the
image. If required, a qualified service engineer should check the position of the light collector and/
or perform a shading correction. Image processing parameters are set up and confirmed by a
qualified Imaging Specialist during the turnover of the system. Image processing changes should
only be made by qualified personnel. Errors in spatial uniformity are highly visible on any image
and require immediate attention from the service engineer.
Image Printer:
Potential Printer artifacts include non-uniform density across the image, white spots or streaks
and film scratches. Non-uniform density can occur in the printing and/or development cycles of
the printer. White spots can be caused by dirt or other particles that prevent laser light or thermal
interaction with the film. Dirt or defect in the thermal head of dry printers can produce black or
white lines traveling in the direction of film transport. In order to differentiate printer artifact from IP
damage or dirt, reprint the image using Twin Format. If the artifact appears in the same position
on both images it is an X-ray generator, IP or reader artifact. Artifacts that exist outside the image
field borders of the film typically indicate a printer problem exists.
Contact the printer manufacturer for additional information.

PERFORMANCE AND CORRECTIVE ACTION

8th Edition
897N0602G

If significant artifacts are noted, contact the person maintaining that system component. Not all
artifacts can be totally eliminated. It may be useful to use the concept of ALARA (as low as
reasonably achievable) when attacking artifacts. If they can be easily eliminated, they should be.
If the artifact is difficult or expensive to eliminate and is subtle (not mimicking or obscuring clinical
information), it may be tolerable. The medical physicist should consult with the interpreting physician
as to whether the artifact is tolerable. Tolerances for artifacts should be lower with new imaging
equipment. If this test produces results that fall outside the action limits as specified, the source of
the problem shall be identified and corrective action shall be taken within thirty days of the test date.
Clinical imaging and mammographic image interpretation may be continued during this period.

83

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

SYSTEM ARTIFACT EVALUATION


DATE:
Reader Unit ID:
Exposure Unit ID:
Type of Attenuator
Thickness of Attenuator
kVp Setting
All Target and Filter combinations used clinically must be evaluated. All available focal spot sizes and magnification
modes used clinically must be evaluated.

Exposure Conditions
Cassette - IP Number
IP Size
mAs or AEC setting(s)
Optical Density
Target/Filter Type
Focal Spot Size Large or Small
Magnification Factor Used

Exposure Equipment Artifact


Acceptable?
CORRECTIVE ACTION NEEDED?

PASS
YES

FAIL
NO

YES

NO

PASS

FAIL

PASS
YES

FAIL
NO

YES

NO

PASS

FAIL

PASS
YES

FAIL
NO

YES

NO

PASS

FAIL

CORRECTIVE ACTION DESCRIPTION:


CORRECTIVE ACTION TAKEN?
CORRECTIVE DATE:
RESULTS OF NEW TEST:

FCRm/CRm Reader Artifact


Acceptable?
CORRECTIVE ACTION NEEDED?
CORRECTIVE ACTION DESCRIPTION:
CORRECTIVE ACTION TAKEN?
CORRECTIVE DATE:
RESULTS OF NEW TEST:

Printer Artifact
Acceptable?
CORRECTIVE ACTION NEEDED?
CORRECTIVE ACTION DESCRIPTION:
CORRECTIVE ACTION TAKEN?
CORRECTIVE DATE:
RESULTS OF NEW TEST:

If significant artifacts are visible, contact the appropriate person maintaining or servicing the digital mammography equipment, printer or X-ray
generator.

84

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

19 Image Quality Evaluation (Annual) Medical Physicist


OBJECTIVE
To assess mammographic image quality and to detect temporal changes in image quality.

FREQUENCY
This test must be carried out initially, at installation, after appropriate calibration of the equipment to
establish a baseline level. Subsequently, it must be carried out at least annually to compare with the
previous test image.

REQUIRED TEST EQUIPMENT

ACR MAPP
Acrylic 4mm x 1cm diameter disc (for hardcopy QC only)
Designated cassette and Imaging Plate used for mammography
Image Quality Evaluation Control chart
Watch/Timer
Previous Image

PROCEDURE
1.

Using the designated cassette, erase the IP using the secondary erasure mode on the reader.

2.

Place the cassette in the cassette holder.

3.

Place the phantom on the cassette holder and position it so the edge of the phantom fully
covers and is aligned with the chest wall side of the image receptor. Center the phantom, left
to right.

4.

The 4mm acrylic disc is not needed for softcopy QC. If softcopy is used for final interpretation,
proceed to step 6.

5.

For hardcopy QC, secure the acrylic disc to the top of the phantom within the image area but
positioned so it will not obscure details in the phantom. A suitable location is between and
slightly below the first and second largest fibers. Once located, the disc may be permanently
attached to the phantom with super glue.

6.

Bring the compression device into contact with the phantom.

7.

Verify the location of the AEC detector. It should be in the same location used for previous
phantom images and should be completely covered by the phantom.

8.

Make an exposure using the AEC technique factors (target, filter, kVp, grid, density control
setting, etc.) currently used clinically for a 4.2-cm compressed breast of average density.
.

8th Edition
897N0602G

85

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

9.

Record the mAs on the control chart after making the exposure.

10. After exposure, ensure that a predetermined interval time between 5 and 10 minutes elapses
prior to the reading process. Process the IP using the "Physics, ACR MAPP" menu.
NOTE
It is important to control the time interval between exposure and reading of the IP. To minimize
variability, it is recommended that and interval be chosen somewhere between 5 and 10 minutes
and that this same interval be used consistently.

11. Record the AEC density control setting on the control chart.
12. If printed images are used for final interpretation: Measure the following densities. The
background density should be measured in the geometric center of the phantom image. The
DD measurement is the density inside the acrylic disc subtracted from the density directly
adjacent to the disc, perpendicular to the anode-cathode axis. For consistent results, these
measurements must be made at the same location each time the background optical density
and the DD on the control chart.
If softcopy images are used for final interpretation: Record the S value from the phantom
exposure image on the background density line of the control chart.

VIEWING CONDITIONS
Phantom images should be read under controlled viewing conditions. General lighting should be at
a low level and diffused. Viewboxes or workstations should be positioned to avoid light from
windows, other viewboxes, and other sources of bright light, either direct or reflected. Images
should be masked to eliminate extraneous light. Use large fieldof-view magnification to assist in the
visualization of specks as well as other test objects. For hardcopy images, use a 2X or higher
magnifying glass. For softcopy images, use the workstations electronic magnification function.

DATA ANALYSIS AND INTERPRETATION

86

1.

When scoring the phantom image of the approved phantom, each object is scored separately.
Always count the number of visible objects from the largest object of a given type (i.e., fiber,
speck group or mass) downward until a score of 0 or 0.5 is reached, then stop counting for
that object type.

2.

Count each fiber as one point if the full length of the fiber is visible and the location and
orientation for the fiber are correct. Count a fiber as 0.5 if not all, but more than half, of the fiber
is visible and its location and orientation are correct. Add each full or partial fiber to the total
score, from the largest down to the smallest visible, until a score of 0 or 0.5 is reached.

3.

After determining the last fiber to be counted, look at the overall background for artifacts. If a
fiber-like artifact appears anywhere in the wax insert area of the image, but not in an
appropriate location or orientation, deduct the artifactual fiber from the last real half or
whole fiber score if the artifactual fiber is equally or more apparent. Deduct only from the last
real fiber, not from additional deduction. Record the final score after artifact deduction in the
appropriate space on the chart.

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

4.

Use large fieldof-view magnification to assist in the visualization of specks. For hardcopy
images, use a 2X or higher magnifying glass. For softcopy images, use the workstations
electronic magnification function. Starting with the largest speck group, count each speck
group as one point if four or more of the six specks in the group are visible in the proper
locations. Count a speck group as 0.5 if two or three of the six specks in the group are visible
in the proper locations. Count a speck group as 0 if none or only one of the six specks in the
group is visible in the proper location.
Add each full or partial speck group to the total speck group score, from the largest down to
smallest visible group, until a score of 0 or 0.5 is reached.

5.

After determining the last speck group to be counted, look at the overall background for
artifacts. If noise or speck-like artifacts are visible in the wrong locations within the area of the
wax insert, and are as apparent as the real specks, deduct them one for one from the
individual specks counted in the last whole or half speck group scored, and adjust the score of
the last group appropriately. Record the final score after artifact deduction in the appropriate
space on the chart.

6.

Count each mass as one point if the minus density object is visible in the correct location, and
the mass appears to be generally circular against the background (i.e., greater than of the
perimeter is visible). A mass is counted as 0.5 point if the minus density object is visible in the
correct location, but the mass does not have a generally circular appearance. Add each full or
partial mass to the total mass score, from the largest mass down and until a score of 0 to 0.5
is reached. Record the raw mass score before artifact deduction.

7.

After determining the last mass to be counted, look at the overall background for artifacts. If a
mass-like artifact is seen in the wrong location within the area of the wax insert, deduct the
artifactual mass from only the last real whole or half mass scored if the artifactual mass is
equally or more apparent. Record the final score after artifact deduction on the appropriate
space on the chart.

8.

Using magnification, carefully examine the image for non-uniform areas, the presence of dirt or
dust artifacts, grid lines, processing artifacts, or any other artifacts and compare the film image
to the original and previous images.

9.

Circle any artifacts or grid lines on the film image. Investigate the source of any artifacts or grid
lines.
NOTE

Mammography phantom films images should always be viewed


On the same viewbox or viewing device
By the same individual
Under the same viewing conditions
Using the same type of magnification used for reading mammograms
Softcopy images should be window and leveled to best demonstrate the objects of interest

8th Edition
897N0602G

87

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

PRECAUTIONS AND CAVEATS


This test measures the output from all components in the imaging chain. Changes in image quality may
be due to the film, cassettes, IPs, X-ray generator, FCRm/CRm reader unit, added filtration, printer,
viewbox, or workstation. Consequently, other tests will be necessary to determine the component, or
components, causing the change.
It is also necessary to check if settings for the image reading mode or image processing parameters
are changed. These factors will have a large influence on the results of this test.
When any image parameters are changed, it is necessary to perform the Phantom Exposure test to
re-determine the new baselines. Confirm the imaging parameters used for this test on the Exposure
and Image Processing Chart.

PERFORMANCE AND CORRECTIVE ACTION


The total number of simulated masses, speck groups, and fibers visible in the phantom image
should not decrease by more than one-half, assuming the same individual is viewing the images
under identical conditions.
If a different number of objects are noted, the current image should be compared with previous
images and the original image to determine if a change has really occurred. Because the medical
physicist may not have the opportunity to measure phantom image quality as frequently as the QC
technologist, it is important to review a sample of the phantom images acquired by the technologist
since the previous visit, comparing results with your own assessment of image quality. Any apparent
problems in scoring the phantom should be included as a corrective action and lead to a discussion
with the QC technologist about phantom scoring. The acrylic discs used to produce an image for
the measurement of the density difference may vary in thickness. Consequently, the density difference
is a relative, not absolute measurement and is to be used only for quality control purposes. It is
essential to use the same acrylic disc if comparisons are to be made between different facilities.
NOTE
At a minimum, the four largest fibers, the three largest speck groups and the three largest masses
must be visible. If your baseline image exceeds the above minimum scores, the new image score
should not decrease by more than from the baseline levels.

The mAs noted on the generator read-out typically does not change by more than 15%. If your
measurements exceed this amount, it may be due to a change in kVp, mAs, target material, filter
material, generator calibration, or AEC calibration since the previous image.
If printed images are used for final interpretation:
The optical density (OD) of the film shall be within control limits of 0.20, from the established
operating level.
The acrylic discs used to produce an image for the measurement of the density difference may vary
in thickness. Consequently, the density difference is a relative, not absolute measurement and is to
be used only for quality control purposes. It is essential to use the same acrylic disc if comparisons
are to be made between different facilities.
The density difference (disc vs. adjacent density) shall not vary by more than 0.05 from the
established operating level. If a new operating level for background optical density is chosen, then a
new operating level for density difference must be established. It is essential that all printers used to
print mammography images at one facility produce similar film optical densities. It is not acceptable
to have one unit producing a film with an OD of 1.40 and another producing an OD of 1.80.
88

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

If softcopy is used for final interpretation:


The S value of the phantom image confirms the exposure unit output and the CR reader sensitivity
setting. The S value result must be recorded on the QC chart and must not vary by greater than +/20%. If your results exceed this control limit, make sure that the kVp and /or mAs and the menu
selected are correct and repeat the exposure.
For both softcopy and hardcopy:
If the performance criteria and corrective action for this test are not met, a second phantom image
should be taken and evaluated. If the criteria are still not met the reasons for this failure must be
investigated, corrective action taken, and the results documented before patients are examined with
this system. It is essential that all mammography units, CR mammography readers, and printers at
one facility produce similar film optical densities. It is not acceptable to have one unit, reader or dry
laser printer producing a film with an OD of 1.40 and another producing an OD of 1.80.
Any visual differences between the current phantom film image and the saved original phantom film
image should be investigated. Printers or CR reader artifacts, grid lines, or grid artifacts should not
be present since any of these may degrade clinical images.
Document corrective action for future reference on the Phantom Image Maintenance and
Corrective Action Log. All sources of significant artifacts should be eliminated.
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken on the component(s) that caused the failure
before any further examinations are performed using the failed component(s). If the component(s)
that caused the failure (e.g. exposure unit or CR reader) is replaced by an alternative device and the
mammography system passes the re-test, image acquisition and interpretation may continue using
that combination of devices.
If the failure has been determined to be of a diagnostic device used for mammographic image
interpretation (e.g. laser printer, physicians review station) the source of the problem shall be
identified and corrective action shall be taken before that device can be used for mammographic
image interpretation. Clinical imaging may be continued. If available, an alternative approved
diagnostic device may be used for mammographic image interpretation.

NOTE
Phantom Images should be retained in the QC records for the last full year (softcopy images can
be locked on the QA and PACS workstation). The original baseline image should be retained
until it is necessary to establish a new baseline.

8th Edition
897N0602G

89

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

IMAGE QUALITY EVALUATION


DATE:
Exposure Unit ID:
Reader Unit ID:
Phantom ID#:
AEC Detector position:
Cassette size:
IP & Cassette#:
Previous
Image

Current
Image

Comments

kVp setting
Density control
Phototimed mAs
mAs change
% mAs change = mAs change / mAs 100
S value
S value Change
*Background density
*Background density change
*Density outside disc
*Density inside disc
*Density diff = outside-inside
*Density difference change
Number of fibers seen
Fibers seen after deduction
Fiber change
Number of speck groups seen
Speck groups after deduction
Speck groups change
Number of masses seen
Masses seen after deduction
Mass change
* Hardcopy for final interpretation only.
JUDGEMENT
Comments:

PASS / FAIL
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
SIGNATURE:

90

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

20 Dynamic Range (Annual) Medical Physicist


OBJECTIVE
To confirm the dynamic range of the FCRm/CRm reader unit and imaging plate.

FREQUENCY
At installation and then annually, and when decrease in the dynamic range is suspected.

REQUIRED TEST EQUIPMENT


Uniform 2 cm and 4 cm thicknesses of defect free acrylic
Previously-selected 18 x 24 cm CR QC cassette for mammography and IP used for
mammography
Dynamic Range data form

PROCEDURE
NOTE
This is an evaluation of the CR reader unit. Select one exposure unit for this evaluation.

8th Edition
897N0602G

1.

Using the dedicated mammography cassette, erase the IP using the secondary erasure mode
on the reader.

2.

Place the cassette in the bucky tray.

3.

Place the 2 cm and an additional 4 cm thickness of acrylic attenuating material on the surface
of the Bucky tray as shown in Figure 1.

4.

Bring the compression device into contact with the phantom attenuators.

5.

Verify the location of the AEC detector. It should be completely covered by the 2 cm and 4 cm
thicknesses of acrylic.

6.

Make an exposure using the AEC technique factors (target, filter, kVp, grid, density control
setting, etc.) currently used clinically for a 6cm compressed breast of 50/50 composition.

7.

Read the IP using the "Physics Dynamic Range" menu.

8.

Using the QA function on the CR Console/AWS-c, set the L Value to 4.0.

9.

View the image on the display (monitor or hardcopy) used for final interpretation, and record
your findings on the Dynamic Range report form.

91

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

PRECAUTIONS AND CAVEATS


Be sure to process the image using the correct menu.

PERFORMANCE AND CORRECTIVE ACTION


It is required that the three exposure areas (no acrylic, 2 cm acrylic, and 6 cm acrylic) be
discernable.
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken on the component(s) that caused the failure
before any further examinations are performed using the failed component(s). If the component(s) that
caused the failure (e.g. exposure unit or CR reader unit) is replaced by an alternative device and the
mammography system passes the re-test, image acquisition and interpretation may continue using
that combination of devices.
If the failure has been determined to be of a diagnostic device used for mammographic image
interpretation (e.g. laser printer, physicians review station) the source of the problem shall be
identified and corrective action shall be taken before that device can be used for mammographic
image interpretation. Clinical imaging may be continued. If available, an alternative approved
diagnostic device may be used for mammographic image interpretation.

3cm 3cm

4cm thick sheet


2cm thick sheet

bucky
tray

Cassette

anterior side

Chest wall side

Side view
Figure 1. Test Object Arrangement

92

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Dynamic Range
DATE:

X- ray generator Number


Cassette / IP Number
CR Reader Unit ID / Number

Target

Filter

kVp

mAs

Exposure conditions

Dynamic Range evaluation


Acceptable?
CORRECTIVE ACTION NEEDED?

PASS

FAIL

YES

NO

YES

NO

PASS

FAIL

CORRECTIVE ACTION DESCRIPTION:


CORRECTIVE ACTION TAKEN?
CORRECTIVE DATE:
RESULTS OF NEW TEST:

SIGNATURE:

8th Edition
897N0602G

93

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

21 Primary Erasure (Additive and Multiplicative Lag Effects)


(Annual) Medical Physicist
OBJECTIVE
To assess the erasure performance of the FCRm/CRm image reader unit and imaging plate

FREQUENCY
At installation and then annually and when incomplete erasure is suspected.

REQUIRED TEST EQUIPMENT


ACR MAPP
A uniform 4 cm thick plate of defect-free acrylic large enough to cover an 18 24 cm cassette
Previously selected and loaded 18 x 24 cm CR mammography QC cassette and imaging plate
Watch/timer
Primary Erasure data form

PROCEDURE
NOTE
This is an evaluation of the CR image reader. Select one exposure unit for this evaluation.

Additive Lag test

94

1.

Erase the IP using the secondary erasure mode on the reader.

2.

Place the cassette in the Bucky tray.

3.

Place the ACR MAPP on the patient support and position it so the edge of the phantom is
aligned with the chest wall side of the image receptor. Center the phantom, left to right.

4.

Bring the compression device into contact with the phantom.

5.

Verify the location of the AEC detector. It should be in the same location used for the Phantom
Image test and should be completely covered by the phantom.

6.

Make an exposure using the AEC technique factors (target, filter, kVp, grid, density control
setting, etc.) currently used clinically for a 4.2 cm compressed breast of 50/50 composition.

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

7.

Process the IP using the "Physics, ACR MAPP" menu.

8.

Record the S value on the data form.

9.

Wait one minute and re-process the same IP using the "Physics, ACR MAPP" menu without
making additional exposure.

10. On the new image, change the S value to 10 times that of the original S value recorded during
Step 8 using the QA function of the CR Console/AWS-c.
11. Inspect the image for the visibility of the outline of the ACR MAPP.
Multiplicative Lag test
12. Repeat steps 2 through 7 using the same cassette as in the first test.
13. Disregard the image produced. Remove the returned cassette and IP from the reader and
insert it into the bucky tray.
14. Place a uniform 4 cm thick acrylic plate that is large enough to cover the mammographic entire
cassette on top of the bucky tray.
15. Make an exposure using the AEC technique factors (target, filter, kVp, grid, density control
setting, etc.) currently used clinically for a 4 cm compressed breast of 50/50 composition.
16. Process the IP using the "Physics, ACR MAPP" menu.
17. Inspect the image for the visibility of the outline of an ACR phantom.

PRECAUTIONS AND CAVEATS


Be sure to read the image using the "Physics, ACR MAPP" menu.

PERFORMANCE AND CORRECTIVE ACTION


If significant artifacts are noted, contact the person who maintains that system component, the CR
reader. Not all artifacts can be totally eliminated. It may be useful to use the concept of ALARA (as
low as reasonably achievable) when attacking artifacts. If they can be easily eliminated, they should
be. If the artifact is difficult or expensive to eliminate and is subtle (not mimicking or obscuring
clinical information), it may be tolerable. The medical physicist should consult with the interpreting
physician as to whether the artifact is tolerable. Tolerances for artifacts should be lower with new
imaging equipment.
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken on the component(s) that caused the failure
before any further examinations are performed using the failed component(s). If the component(s)
that caused the failure (e.g. exposure unit, CR reader, display device) is replaced by an alternative
device and the mammography system passes the re-test, image acquisition and interpretation may
continue using that combination of devices.

8th Edition
897N0602G

95

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Primary Erasure (Additive and Multiplicative Lag Effects)


DATE:
Exposur Unit ID:
Cassette - IP Number:
CR Reader Unit ID:

Additive Lag test


Exposure conditions

Target

Filter

kVp

mAs

kVp

mAs

ACR MAPP
S value of the ACR MAPP
S value applied to non-exposure image

10
About 10

Artifact evaluation
Acceptable?
CORRECTIVE ACTION NEEDED?

PASS

FAIL

YES

NO

YES

NO

PASS

FAIL

CORRECTIVE ACTION DESCRIPTION:


CORRECTIVE ACTION TAKEN?
CORRECTIVE DATE:
RESULTS OF NEW TEST:

Multiplicative Lag test


Exposure conditions

Target

Filter

ACR MAPP
A uniform 4 cm sheet image

Artifact evaluation
Acceptable?
CORRECTIVE ACTION NEEDED?

PASS

FAIL

YES

NO

YES

NO

PASS

FAIL

CORRECTIVE ACTION DESCRIPTION:


CORRECTIVE ACTION TAKEN?
CORRECTIVE DATE:
RESULTS OF NEW TEST:

The mAs noted on the generator read-out must not change by more than 15% from the previous test.
SIGNATURE:

96

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

22 Inter-Plate Consistency (Annual) Medical Physicist


OBJECTIVE
To confirm the X-ray absorption and SNR consistency of imaging plate/cassette pairs of one type
and size.

FREQUENCY
At installation, annually, and before introducing new imaging plates or cassettes.
NOTE
The inter-plate consistency test must be performed on new imaging plates and imaging plate
cassettes prior to clinical use. When adding or replacing imaging plates and cassettes of the
same type already in use between annual physicists surveys, the QC technologist may perform
this test on the new imaging plates and cassettes under the oversight of the medical physicist
prior to their first clinical use, recording the results on the inter-plate consistency data form for
review by the medical physicist during the next annual survey. These new imaging plates and
cassettes must meet the criteria established for the group to which they will be added. If the new
imaging plates and cassettes do not fit the criteria for an existing group and require more
exposure, then the testing must be performed by the medical physicist.

REQUIRED TEST EQUIPMENT


A uniform 4 cm thick sheet of defect free acrylic large enough to cover a 2430cm cassette
Imaging plates and cassettes to be evaluated
Watch/timer
Inter-Plate Consistency data form

PROCEDURE
1.

Select a cassette used for mammography, and erase the IP using the secondary erasure mode
of the reader.

2.

Place the cassette in the bucky tray.

3.

Place the 4-cm thick acrylic attenuating material (phantom) on the patient support surface.

4.

Lower the compression device into contact with the phantom.

5.

Make an AEC exposure in the auto mAs mode. Fix the other technique factors (target, kVp,
filter, grid, density control setting, etc.) used clinically.

6.

Wait a predetermined time interval and process the IP using the "Physics Sensitivity" menu.
This menu should be set in the Semi mode and all image should be set to zero (MRE, MDE,
PRE).
NOTE

To reduce the influence of IP image fading characteristics on this test, it is important to control
the time interval between X-ray exposure and reading of the IP. It is important that the interval be
consistent whenever this test is performed. Choose an interval between 5 and 10 minutes, and
then use the same interval consistently.

8th Edition
897N0602G

97

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

7.

Record the mAs of the test exposure on the Inter-Plate Consistency data form.

8.

Repeat the procedure from Step1 to Step 7 for all IPs of the same type and size to be
evaluated. All IPs of the same type and size are considered a group.

9.

Repeat steps 1 through 8 for other groups to be evaluated.

10. Calculate the SNR of all images using the QC calculation tool.
11. Record the SNR result on the Inter-Plate Consistency data form.
12. Before using new imaging plates clinically, compare the consistency of the new imaging plates
against the inventory of all imaging plates of the same group by performing steps 1 through 11.

PRECAUTIONS AND CAVEATS


This test is sensitive to changes in beam quality (kVp, target and filter material). Use an AEC mode
that varies mAs only and does not permit changes in kVp, target and filter material.
This test is also sensitive to changes in the patient support and anti-scatter grid. Because the patient
support and grid change as the size of the IP is changed, comparisons of IPs of different sizes will
also reflect the differences in mAs and SNR introduced by differences in the support and grid.

PERFORMANCE AND CORRECTIVE ACTION


The variation of mAs value must be within 10%
The variation of SNR must be within 15%
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken within thirty days of the test date. Clinical
imaging and mammographic image interpretation may be continued during this period.

60mm

20mm

20mm

Chest Wall Side


Figure 1: ROI of the inter-plate sensitivity test

98

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

INTER-PLATE CONSISTENCY
DATE:
Exposure Unit ID:
CR Reader Unit ID:

AEC-mode

Target

Filter

kVp

Exposure conditions

Group:
Cassette ID

mAs

Average
Value

Limiting
Value

SNR

Average
Value

Limiting
Value

Acceptable?
PASS / FAIL
PASS / FAIL
PASS / FAIL

to

to

PASS / FAIL
PASS / FAIL
PASS / FAIL
PASS / FAIL
PASS / FAIL

(10%)

(15%)

PASS / FAIL
PASS / FAIL

Group:
Cassette ID

mAs

Average
Value

Limiting
Value

SNR

Average
Value

Limiting
Value

Acceptable?
PASS / FAIL
PASS / FAIL
PASS / FAIL

to

to

PASS / FAIL
PASS / FAIL
PASS / FAIL
PASS / FAIL
PASS / FAIL

(10%)

(15%)

PASS / FAIL
PASS / FAIL

SIGNATURE:

8th Edition
897N0602G

99

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

23 kVp Accuracy and Reproducibility (Annual) Medical


Physicist
NOTE
This test is specific to the X-ray generator and not the FCRm/CRm system. Follow the procedures in the
ACR Mammography Quality Control Manual for screen-film systems. If a cassette and imaging plate are
used, erase the plate after each exposure using the Primary Erase cycle.

FREQUENCY
This test must be performed annually.

PERFORMANCE AND CORRECTIVE ACTION


A. The kVp shall be accurate within 5% of the indicated or selected kVp at:
1. The lowest clinical kVp that can be measured by a kVp test device
2. The most commonly used kVp;
3. The highest clinical kVp, and
B. At the most commonly used clinical settings of kVp, the coefficient of variation of reproducibility
of the kVp shall be equal to or less than 0.02.
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken within thirty days of the test date. Clinical
imaging and mammographic image interpretation may be continued during this period.

100

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

24 Dose (Annual) Medical Physicist


NOTE
This test is specific to the X-ray generator and not the FCRm/CRm system. Follow the procedures in the
ACR Mammography Quality Control Manual in accordance with MQSA requirements for screen-film
systems. If a cassette and imaging plate are used, erase the plate after each exposure using the Primary
Erase cycle.

FREQUENCY
This test must be performed at installation and annually.

PERFORMANCE AND CORRECTIVE ACTION


The average glandular dose delivered during a single cranio-caudal view of an FDA-accepted
phantom simulating a standard breast shall not exceed 3.0 milligray (mGy) (0.3 rad) per exposure.
The dose shall be determined with technique factors and conditions used clinically for a standard
breast.
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken on the component(s) that caused the failure
before any further examinations are performed using the failed component(s). If the component(s) that
caused the failure (e.g. exposure unit, CR reader unit, display device) is replaced by an alternative device
and the mammography system passes the re-test, image acquisition and interpretation may continue
using that combination of devices.

8th Edition
897N0602G

101

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

25 Beam Quality Assessment and Half-Value Layer Measurement


(Annual) Medical Physicist
NOTE
This test is specific to the X-ray machine and not the FCRm/CRm system. Follow the procedures in the
ACR Mammography Quality Control Manual in accordance with MQSA requirements for screen-film
systems. The X-ray machine will require a cassette to be present in the cassette holder before an
exposure can be made. Either an FCRm/CRm or screen-film cassette may be used for this purpose. As no
images are recorded in this test, it is not necessary to load an IP in the cassette or a film in the screen-film
cassette. However, if an IP in a cassette is used in this test, erase the IP using the image reader primary
erase cycle after completing this test and before performing other imaging tests or clinical use.

FREQUENCY
This test must be performed annually, after service to the X-ray machine which may effect beam
quality, and as part of an mammography equipment evaluation (MEE).

PERFORMANCE AND CORRECTIVE ACTION


The HVL shall meet the specifications of 21 CFR 1020.30(m) for the minimum HVL.
These values, extrapolated to the mammographic range, are shown below.
Values not shown below may be determined by linear interpolation or extrapolation.
Designed Operating Range
(kV)
Below 50

Measured Operating Voltage


(kV)
20
25
30

Minimum HVL
(millimeters of aluminum)
0.20
0.25
0.30

If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken within thirty days of the test date. Clinical
imaging and mammographic image interpretation may be continued during this period.

TYPICAL HVL MEASUREMENTS FOR DIFFERENT TUBE VOLTAGE AND


TARGET FILTER COMBINATIONS
The specified minimum HVL values in the performance and corrective action section above must be
achieved for the Mo/Mo target/filter combination. Typical HVL values for Mo/Mo and other target/
filter combinations as published by EUREF are provided below for reference and are optional.

KV
25
28
31

Mo/Mo
0.330.2
0.360.2
0.390.2

HVL (mm Al) for target filter combinations(1)


Mo/Rh
Rh/Rh
0.400.2
0.380.2
0.420.2
0.430.2
0.440.2
0.480.2

W/Rh
0.520.3
0.540.3
0.560.3

(1) All values include the effects of attenuation by a compression paddle.


From the EUREF, The European Protocol for the Quality Control of the physical and technical aspects of mammography screening,
Addendum on Digital Mammography To chapter 3 of the European Guidelines for Quality Assurance in Mammography Screening Version
1.0, November 2003 See Section 14, References, for additional information about EUREF.

102

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

26 Radiation Output (Annual) Medical Physicist


NOTE
This test is specific to the X-ray generator and not the FCRm/CRm system. Follow the procedures in the
ACR Mammography Quality Control Manual in accordance with MQSA requirements for screen-film
systems. If a cassette and imaging plate are used, erase the plate after each exposure using the Primary
Erase cycle.

FREQUENCY
This test must be performed annually and upon Mammography Equipment Evaluation (MEE) of the
X-ray system.

PERFORMANCE AND CORRECTIVE ACTION


A. The system shall be capable of producing a minimum output of 7.0 mGy air kerma per second
(800 mR per second) when operating at 28 kVp in the standard mammography mode (Mo/Mo)
at any SID where the system is designed to operate and when measured by a detector with its
center located 4.5 cm above the breast support surface with the compression paddle in place
between the source and the detector.
B. The system shall be capable of maintaining the required minimum radiation output averaged over
a 3.0 second period.
If this test produces results that fall outside the action limits as specified, the source of the problem
shall be identified and corrective action shall be taken within thirty days of the test date. Clinical
imaging and mammographic image interpretation may be continued during this period.

8th Edition
897N0602G

103

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

104

8th Edition
897N0602G

Aspire FCRm/CRm
Fujifilm CR for Mammography
Quality Control Manual

Revision History
Revision

Issue Date

Reason

First Edition 897N0602

07.2006

New Release

2nd Edition 897N0602A

10.2006

Change of corporate logo and references to Fuji Photo


were changed to FUJIFILM on pages 1 and 2 only. No
other changes to content. The 2nd Edition was never
released.

3rd Edition 897N0602B

02.2007

1) Revised to reflect 21 CFR 900.12 (e) (8) Use of test


results. The FDA approved alternative requirement
dated August 2, 2006.
2) Revised weekly and annual Monitor QC and Printer
QC tests when the device manufacturer does not
provide a QC program.
3) Added Printer and Monitor QC to Baseline Tests and
indicated which tests must be performed when
adding or replacing an FCRm image reader. Changed
required Baseline Test from Phantom Image to Image
Quality.
4) The Imaging Plate (IP) Fog test was changed from
annual physicist test to semi-annual technologist test.
5) The Automatic Exposure Control (AEC) System
Performance Assessment test was revised.
It is no longer required to measure air kerma during
the image mode tracking portion of this test.
All exposures made during the CNR per object
thickness portion of this test can be made using
AEC exposure.
The corrective action limits have been changed.
6) A revision table is added to the manual to identify
changes to the previous manual.

4th Edition 897N0602C


5th Edition 897N0602D

6th Edition 897N0602E

12.2009

7th Edition 897N0602F

01.2012

Changes were made to incorporate the CRm system consisting


of the Carbon XL-2 Reader Unit with HR-VI imaging plates and
CH-type IP Cassettes.

8th Edition 897N0602G

05.2013

Added Soft Copy Phantom Control / CNR Chart to page 26. This
chart was missing from page 26 in the 01.2012 release.

8th Edition
897N0602G

Never released.
1) The CNR figures on pages 35 and 96 have been
changed to the same as those in 897N100001A.
2) Revised wording for annual Monitor QC test.
3) Changes have been made along with the release of
the AWS-c.

105

FUJIFILM MEDICAL SYSTEMS U.S.A., INC


419 WEST AVENUE, STAMFORD CT 06902, U.S.A.

You might also like