Manual Operacion Elgar
Manual Operacion Elgar
PROGRAM Menu Status and Information Data:
Menu Item Status and Information Line Range/Default
PHASE A [<enter> for next phase] Phases A, B, C
AMPL [program output RMS/DC voltage] * 0156 V or 0312 V
(default: 0.00V)
CURL [program RMS/DC current limit] 013 A or 06.5 A (default: 5.00A
dependent on voltage range)
FREQ
[program output frequency] **
405000 Hz (default: 60 Hz)
ANG [program phase angle offset] *** 0360(default: 0, 120, 240
dependent on phase selected)
FUNC
[program output waveform] ****
Default: Sine
Operation SW 5250ASW 3500ASW 1750A
3-6 Operation Manual
* Reference Section 3.4, Front Panel Programming Conventions.
** There is only one frequency setting for all three phases. Changing the frequency
of one phase changes the frequency for all phases.
*** The SW can accept only positive numbers when programming phase angle
offset. Because phase angle offset represents a phase lead with respect to the
internal reference, the formula below should be used for a selected phase to
represent phase angle in terms of phase lag:
Desired Phase Angle Lag = 360 (Ph.B Angle Ph.A Angle)
The examples below are for a twophase quadrature system:
For phase angles Ph.A = 0, Ph.B = 270, phase lag would be 90.
90= 360 (270 0)
For phase angles Ph.A = 45, Ph.B = 120, phase lag would be 285.
285= 360 (120 45)
**** To operate the SW as a DC supply, select the function as DC+ or DC- for
positive and negative DC. AC+DC Coupling must also be selected from the
INSTRUMENT menu. The AMPL field is programmed to 0V as a precaution
when changing to or from the DC+ or DC- function.
3.3.2.1 Locking Program Fields
For two- or three-phase systems, individual programming fields of amplitude, current
limit, phase angle, and function may be locked so that any change made to phase A
will be made to phases B and C. The Lock function is intended for front panel
operation. When using GPIB, use the SOURCE0 command.
To lock a field:
1. Move the field select cursor to the field to be locked in the Phase A Program
menu.
2. Press the F1 function key. The symbol should appear to the right of the field
indicating the lock mode is active.
3. Enter the value to be locked into the field.
NOTE: The lock will not be active until a value is entered; previously entered
values are not locked.
To unlock a field:
1. Move the field select cursor to the locked field in the Phase A Program menu.
2. Press the F1 function key. The symbol should disappear.
SW 5250ASW 3500ASW 1750A Operation
Operation Manual 3-7
3.3.2.2 Front Panel 8tore/Recall
The current state of the SW can be saved and restored using the Front Panel
Store/Recall feature. The following Program Menu parameters are affected:
Amplitude
Current Limit
Frequency
Phase Angle
Function
The Store/Recall Menu can be accessed by pressing F2 from any Program menu edit
field. Setups are stored/recalled as a number from 0 to 49. The incr./dec. mode is not
available for this menu. Store/Recall functions do not affect the Lock/Unlock status of
the Program menu fields. If a field is locked, it will remain locked. However, the Recall
parameters for phases B and C will be updated
To store a setup:
1. Press F2 from any Program menu edit field.
2. Move the field select cursor to STORE SETUP.
3. Enter a setup number from 0 to 49.
To recall a setup:
1. Press F2 from any Program menu edit field.
2. Move the field select cursor to RECALL SETUP.
3. Enter the previously stored setup number from 0 to 49.
Operation SW 5250ASW 3500ASW 1750A
3-8 Operation Manual
3.3.3 MEA8URE Menu
If equipped with the Test and Measurement option, the MEASURE menu allows you to:
Measure voltage (RMS)
Measure current (RMS)
Measure frequency
Measure phase angle (referenced to a master)
Measure power (W)
Measure apparent power (VA)
Measure power factor (PF)
Measure peak current.
PROGRAM
MEASURE
WAVEFORM
SEQUENCE
INSTR
SYSTEM
AMPL A 120.00 V
OFF 0.00
OFF 0.00
OFF 0.00
OFF 0.00
OFF 0.00
MAIN MENU OUTPUT [OFF]
The Measure menu contains six display fields that can be set to any of the available
measurements. To set a measurement field:
1. Move the field select cursor to one of the six measurement fields.
2. Press <Enter>. A pop-up window entitled SELECT MEASUREMENT will
appear; OFF will be selected.
3. Press the Up/Down Arrow keys or use the Knob to scroll through the list of
available measurements.
4. With the desired measurement in the selection window (or OFF to disable a
measurement), press <Enter>.
The measured value will be displayed in the display field. Active measurement fields
are continually updated and retained upon cycling power to the unit.
SW 5250ASW 3500ASW 1750A Operation
Operation Manual 3-9
3.3.4 WAVEFORM Menu
The WAVEFORM menu allows you to perform a variety of actions on waveshapes
and waveforms, including the creation of a new waveform based on a waveshape in
memory.
The waveshapes and waveforms are limited to spikes, dropouts, sags and surges
of existing waveforms in the scratchpad area. This is accomplished by loading the
existing waveshape into the waveform scratchpad then editing it. The edited waveform
may then be viewed, output using the SCRATCH function name in the PROGRAM
menu, or saved to the waveform library.
LOAD
EDIT
VIEW
SAVE
DELETE
WAVEFORM OUTPUT [OFF]
Operation SW 5250ASW 3500ASW 1750A
3-10 Operation Manual
3.3.4.1 LOAD 8ub-Menu
The WAVEFORM menu allows you to LOAD an existing waveform from non-volatile or
flash memory to the waveform scratchpad.
Any existing information in the scratchpad is erased when a new waveform is loaded.
LOAD
EDIT
VIEW
SAVE
DELETE
FUNC Name
WAVEFORM OUTPUT [OFF]
EDIT Sub-Menu Status and Information Data:
Menu Item Status and Information Line Range
FREQ [waveform frequency] 405000 Hz
Vrms [output waveform voltage] 0156 or 0312 V
START [starting phase angle] 0360
TIME [surge/sag duration] 01/Freq
STOP [ending phase angle] Start360
AMPL [drop/spike voltage] -220.6 to 220.6 Vpeak (156 Vrms range)
-441.2 to 441.2 Vpeak (312 Vrms range)
NOTE: Vrms is in terms of RMS voltage while AMPL is in terms of peak voltage.
Both of these values depend on the selected voltage range setting in
the INSTRUMENT menu.
See the example in section 3.5.1.
Operation SW 5250ASW 3500ASW 1750A
3-12 Operation Manual
3.3.4.3 VEW 8ub-Menu
The WAVEFORM menu allows you to VIEW a waveform:
LOAD
EDIT
VIEW
SAVE
DELETE
WAVEFORM OUTPUT [OFF]
SAVE Sub-Menu Status and Information Data:
Menu Item Status and Information Line
NAME [> = next char, < = backsp]
SAVE [save waveform to library]
The waveform is saved and scaled in scratch memory
without regard to the calculated RMS value.
SAVE RMS [save waveform to library]
The waveform is scaled so that the calculated RMS
value of the entire waveform is equal to the value of
the Vrms field in the WAVEFORM EDIT menu.
Operation SW 5250ASW 3500ASW 1750A
3-14 Operation Manual
3.3.4.5 DELETE 8ub-Menu
The WAVEFORM menu allows you to DELETE a waveform stored in non-volatile
memory.
ALL USER WAVEFORMS - Erases the 50 user waveform locations in non-volatile
RAM. Any user waveforms currently running will not be deleted.
INITIALIZE MEMORY - Resets the waveform library to the factory settings. All factory
waveforms will be restored and all user waveforms stored in non-volatile RAM will be
erased. You must cycle power to the unit after this function for the changes to take
effect.
NOTE: Under normal operating conditions this function is not required.
Instead, use DeleteAll User Waveforms.
LOAD
EDIT
VIEW
SAVE
DELETE
FUNC
ALL USER WAVEFORMS
INITIALIZE MEMORY
Name
WAVEFORM OUTPUT [OFF]
Press key to
toggle custom waveforms
Right Arrow
DELETE Sub-Menu Status and Information Data:
Menu Item Status and Information Line
FUNC [<enter> to select waveform]
ALL USER WAVEFORMS [<enter> for all user waveforms]
INITIALIZE MEMORY [Resets to factory defaults]
NOTE: If you attempt to delete a waveform in Flash Memory,
the message ! - STANDARD WAVEFORM will appear.
SW 5250ASW 3500ASW 1750A Operation
Operation Manual 3-15
3.3.5 8EOUENCE Menu
The SEQUENCE menu allows you to create, edit, and execute a sequence.
LOAD
EDIT
SAVE
DELETE
EXECUTE
SEQ MENU OUTPUT [OFF]
3.3.5.1 LOAD 8ub-Menu
The SEQUENCE menu allows you to LOAD an existing sequence from non-volatile or
flash memory to the sequence scratchpad.
NEW - Erases all segments in the SEQUENCE scratchpad.
LOAD
EDIT
SAVE
DELETE
EXECUTE
SEQ
NEW
Name
SEQ MENU OUTPUT [OFF]
LOAD Sub-Menu Status and Information Data:
Menu Item Status and Information Line
SEQ [load sequence to scratchpad]
NEW [clear sequence scratchpad]
Operation SW 5250ASW 3500ASW 1750A
3-16 Operation Manual
3.3.5.2 EDT 8ub-Menu
The SEQUENCE menu allows you to EDIT a sequence in the scratchpad. Frequency
or amplitude can be ramped over the segment duration.
SEG
CYC
A
B
C
FREQ
TIME
FUNC
Sine
Sine
Sine
60.00
0.00
0
0
AMPL
0.00
0.00
0.00
ms
ANGL
0.00
120.00
240.00
NOTE: You can save up to 100 sequences, consisting of up to 1000 segments total.
Once you have reached this maximum, you must delete a previously saved
sequence in order to save a new one.
SW 5250ASW 3500ASW 1750A Operation
Operation Manual 3-19
SAVE Sub-Menu Status and Information Data:
Menu Item Status and Information Line
NAME [> = next char, < = backsp]
SAVE [store sequence in library]
3.3.5.4 DELETE 8ub-Menu
The SEQUENCE menu allows you to DELETE a sequence stored in non-volatile
memory.
ALL USER SEQUENCES - Erases all user sequence locations in non-volatile RAM.
INITIALIZE MEMORY - Resets the sequence library to the factory settings. All factory
sequences will be restored and all user sequences stored in non-volatile RAM will be
erased. You must cycle power to the unit after this function for the changes to take
effect.
NOTE: Under normal operating conditions this function is not required.
Instead, use DeleteAll User Sequences.
LOAD
EDIT
SAVE
DELETE
EXECUTE
SEQUENCE
ALL USER SEQUENCES
INITIALIZE MEMORY
SEQ MENU OUTPUT [OFF]
DELETE Sub-Menu Status and Information Data:
Menu Item Status and Information Line
SEQUENCE [delete stored sequence]
ALL USER SEQUENCES [delete all user sequences]
INITIALIZE MEMORY [reset to factory defaults]
Operation SW 5250ASW 3500ASW 1750A
3-20 Operation Manual
3.3.5.5 EXECUTE 8ub-Menu
The SEQUENCE EXECUTE sub-menu allows the selection of the start, run, and stop
modes of a sequence. The front panel operation is defined here, but all modes are also
available via the GPIB. A sequence can be executed in any combination of the
following modes:
A sequence can be run in its entirety, or stepped through one segment at a time.
When in this step mode, the most recent segment can be repeated.
A sequence can be executed only once, or looped until the STOP command is
received.
A sequence can be terminated with the outputs automatically programmed to
0 volts, restored to the waveforms and values before the sequence began, or
remain at the waveforms and values of the last segment in the sequence. The
last two options will occur with no interruption in output power.
The SEQUENCE EXECUTE sub-menu selections are explained in Table 3-1.
LOAD
EDIT
SAVE
DELETE
EXECUTE
LOAD SEQ
SYNC SELECT SEG
CONTROL STOP
RUN MODE LOOP
STOP MODE ZERO
LOOP CNT 1
Name
SEQ MENU OUTPUT [OFF]
EXECUTE Sub-Menu Status and Information Data:
Menu Item Status and Information Line Range
LOAD SEQ [<enter> to select sequence]
SYNC [sync output configuration] Select Seg / Every Seg
CONTROL [start/stop sequence] Run / Step / Stop
RUN MODE [execute seg once or repeat] Repeat / Single / Loop
STOP MODE [restore output on seg stop] Zero / Program/ End Seg
LOOP COUNT [loop execution count] 19999
SW 5250ASW 3500ASW 1750A Operation
Operation Manual 3-21
8ELECTON PURPO8E
LOAD SEQ Loads a sequence for execution from the sequence library or the sequence
scratchpad. When running under GPIB control, the Event Status Register
and serial polling can be used to indicate when the sequence loading is
complete. The front panel display will indicate Processing sequence... while
loading, and Sequence loaded when complete. This menu option should not
be confused with the sequence load option, which loads a sequence to the
scratchpad for modification.
NOTE: If any of these occurs, the sequence must be re-loaded for execution:
Sequence memory is reset
Any of the sequence fields is updated
Sync output is accessed
Error occurs when loading sequence for execution
SYNC Configures the Rear Panel SYNC Trigger output and Front Panel Trigger
output to generate a pulse for every segment or only selected segments.
After changing the SYNC selection, the sequence must be reloaded.
CONTROL Used to start a sequence, stop a sequence, or start a sequence in the STEP
mode. A sequence must be loaded before RUN or STEP can be selected.
All operations have an immediate effect.
RUN Run the previously loaded sequence. A sequence does not need to
be reloaded when switching between RUN, STEP or STOP. The sequence
will begin executing immediately after a RUN or STEP command.
STEP Step through the previously loaded sequence. Each segment will
execute and remain at the value of the last cycle until instructed to execute
the next segment, or repeat the current segment. When stepping through a
sequence under front panel control, the F1 function key is used to execute
the next segment, and the F2 key is used to repeat the current segment.
(Especially useful for ramping segments.)
STOP Stops the active sequence.
RUN MODE Specifies the running mode of the sequence. This parameter takes effect
once a sequence has begun.
REPEAT Repeats sequence until a STOP command is received.
SINGLE Executes the sequence only once, then returns to the operation
specified in the STOP MODE field.
STOP MODE Specifies the operation mode when a sequence is terminated. This can occur
when running in the SINGLE execution mode, or when STOP is selected.
ZERO Programs the outputs to 0 volts when the sequence is terminated.
PROGRAM Outputs are restored to waveforms and values in the Program
Menu. This mode can be used for a continuous output between sequences.
The Program Menu cannot be modified while a sequence is running. When
this mode is selected, set BOOST to OFF (see System menu).
END SEG The outputs will remain at the waveforms and values of the last
segment in the sequence.
Table 31 Sequence Execute Menu
Operation SW 5250ASW 3500ASW 1750A
3-22 Operation Manual
3.3.6 N8TR {nstrument} Menu
The INSTR menu allows you to:
Set the output range (156 or 312 volts).
Select either AC or AC+DC coupling.
Select current limit mode, either shutdown, foldback or time-out mode
(foldback for the time specified in the ITIMO field then shutdown mode).
Foldback mode is unavailable while SYSTEM/EXTERNAL/ LOW FREQ is on.
Select the time for the time-out mode.
Select the peak overvoltage limit.
Select amplifier parallel operation. When the Parallel mode is ON, only Phase A
will appear in the Program menu. Rear panel output power wiring must match
the Parallel mode selection.
CAUTONl Parallel mode must match the rear panel
output wiring before programming output amplitude.
PROGRAM
MEASURE
WAVEFORM
SEQUENCE
INSTR
SYSTEM
RANGE
COUPLING
I MODE
I TIMO
V_LIM
PARALLEL
MAIN MENU OUTPUT [OFF]
INSTR Menu Status and Information Data:
Menu Item Status and Information Line Range
RANGE [156V or 312V range] 156V / 312V
COUPLING [AC or AC+DC coupling] AC/ AC+DC
I MODE [supervisory current mode] Shutdown / Foldback / Time-Out
I TIMO [current limit time-out] 1009999 ms
V_LIM [peak voltage limit] 20255 Vpk/ 40510 Vpk
PARALLEL [parallel output amplifiers] On / Off
SW 5250ASW 3500ASW 1750A Operation
Operation Manual 3-23
3.3.7 8Y8TEM Menu
3.3.7.1 U8ER 8ub-Menu
The SYSTEM USER sub-menu allows you to:
Configure power up values.
Execute the units self-test.
Enable/disable hint messages.
Enable/disable the sync output in the Program mode
(i.e., when a sequence is not running).
USER
COMM
EXTERNAL
FAULTS
CONFIG
SAVE SYSTEM CONFIG
RESTORE DEFAULTS
HINT MESSAGE ON
SELF-TEST 0000
SYNC OUT ON
SYSTEM OUTPUT [OFF]
USER Sub-Menu Status and Information Data:
Menu Item Status and Information Line
SAVE SYSTEM CONFIG [store new power-up setting]
RESTORE DEFAULTS [factory power-up defaults]
HINT MESSAGE [hint messages on/off]
SELF-TEST [execute self-test]
SYNC OUT [sync output on/off/event]
SAVE SYSTEM CONFIG saves the current system configuration of the Program and
Instrument menus and initializes the SW to these settings on power up.
RESTORE DEFAULTS resets the power up configuration to factory defaults.
Operation SW 5250ASW 3500ASW 1750A
3-24 Operation Manual
SYNC OUT selects the output sync signal to ON, OFF or EVENT. In the ON position,
the signal is driven by the output frequency of the waveform, or sequence settings. In
the EVENT position, a pulse will be generated for any change in VOLT, CURR, FREQ,
PHASE or WAVEFORM.
The SYNC OUT setting only affects normal (non-sequence) mode. To set up the SW
so that a sync pulse only occurs for a specified sequence segment, set this parameter
to OFF. That way, before and after a sequence is run, no other sync pulses will occur.
Table 32 provides self-test result definitions.
BT MEANNG BT MEANNG
0001 LCD Test 0100 Battery Test
0002 EPROM CKSUM 0200 QSPI Test
0004 NVRAM BANK 0 CKSUM 0400 DFI Test
0008 NVRAM BANK 1 CKSUM 0800 Speaker Test
0010 RAM Test 1000 DWSB Self-Test 0
0020 VXI Test 2000 DWSB Self-Test 1
0040 QSCI Test 4000 T&MB Self-Test 0
0080 GPIB Test 8000 T&MB Self-Test 1
Table 32 SELF-TEST Results Definitions
SW 5250ASW 3500ASW 1750A Operation
Operation Manual 3-25
3.3.7.2 COMM 8ub-Menu
The SYSTEM COMM (Communications) sub-menu allows you to:
Configure the GPIB address and display remote status.
Display firmware version information.
USER
COMM
EXTERNAL
FAULTS
CONFIG
SYSTEM OUTPUT [OFF]
GPIB ADDR 25
ECDI VER X.XX
DWSB VER X.XX
TMB VER X.XX
BAUD RATE 57600
DWSB ON-LINE
T&MB ON-LINE
The DWSB and T&MB fields indicate the internal communication status of the Digital
Waveform Synthesis board and Test and Measurement board. These should always
indicate ON-LINE unless a failure has occurred (T&MB will indicate OFF-LINE if this
option is not present).
COMM Sub-Menu Status and Information Data:
Menu Item Status and Information Line Range/Default Value
GPIB ADDR [select gpib address] Range 130
ECDI VER X.XX
DWSB VER X.XX
TMB VER X.XX
[installed firmware version]
BAUD RATE [RS-232 Baud Rate]
Range 30057600
Default 57600
DWSB [Digital waveform status] On-Line/Off-Line
T&MB [Test & Measurement status] On-Line/Off-Line
Operation SW 5250ASW 3500ASW 1750A
3-26 Operation Manual
3.3.7.3 EXTERNAL 8ub-Menu
The SYSTEM EXTERNAL sub-menu allows you to:
Select external modes of operation (Direct Input, External Modulation, External
Gain, and Clock/Lock).
Select alternate compensation (X-LOAD).
Select Low Frequency for operation below 40 Hz when using an external input.
CAUTONl Damage to the equipment may occur if input
frequency requirements are violated (DC, 405000 Hz).
Refer to Table 2-3 for external analog inputs (EXT_IN_A, EXT_IN_B, EXT_IN_C).
Table 33 provides additional data on the External sub-menu.
USER
COMM
EXTERNAL
FAULTS
CONFIG
EXT MOD OFF
DIR INPUT OFF
EXT GAIN OFF
XLOAD OFF
LOW FREQ OFF
CLOCK/LOCK OUT
SYSTEM OUTPUT [OFF]
EXTERNAL Sub-Menu Status and Information Data:
Menu Item Status and Information Line Range
EXT MOD [external modulation] On / Off
DIR INPUT [external direct input] On / Off
EXT GAIN [external gain control] On / Off
XLOAD [reactive load stability] On / Off
LOW FREQ [for external freq Below 40Hz] On / Off
CLOCK/LOCK [clock lock enable/disable] In / Out / Off
SW 5250ASW 3500ASW 1750A Operation
Operation Manual 3-27
8ELECTON PURPO8E
EXT MOD
(External
Amplitude
Modulation)
Amplitude modulation of an output waveform is possible via an input
signal from the rear panel. Input of 0-5 Vrms corresponds to a modulation
of 0-20%. To allow for the modulation voltage, the maximum prog-
rammed voltage is 130 Vrms in low range, and 260 Vrms in high range.
DIR INPUT
(External
Direct Input)
Input on rear panel allows reference signal to go directly to amplifiers.
A signal of 05 Vrms (7.07 Vpeak) or 5VDC corresponds to 0 to full
scale programmed voltage output.
CAUTON: Do not exceed the maximum frequency specification.
EXT GAIN
(External Gain)
Input on rear panel is used to scale output waveform. A 0 to 7.07 VDC
input signal corresponds to 0 to full scale programmed voltage output.
X-LOAD
Reactive Load)
This option can be used to reduce overshoot, undershoot and ringing with
unusual reactive loads. This option should not be used for normal loads.
Do not program frequency >1000 Hz.
LOW FREQ
(Low Frequency)
This option should be used for external AC input signals between DC and
40 Hz (the maximum VA rating must be derated). Also, use for IEC-
1000-3-3 testing, or other conditions requiring fast transient response.
The RMS servo is bypassed. As a result, load regulation drops off, and
current limit (voltage foldback) is disabled.
CLOCK/LOCK
Use this option to configure rear panel clock/lock signal as input or
output. When configured as input, the power outputs will attempt to sync
to clock/lock input frequency. Target output frequency is determined by
frequency value entered in Program Menu FREQ field. Target frequency
should be entered before clock/lock is configured as output. Input
configuration process is as follows:
1. Enter target frequency in Program Menu FREQ field.
2. Go to System External Menu and enable CLOCK/LOCK option by
selecting IN.
3. The unit will now attempt to lock to the input signal; this process can
take up to 5 seconds. Phase lock is indicated on the LCD by the .
symbol to the right of the OUTPUT [ON/OFF] status. Output relay
will not be allowed to close until phase lock has been established.
4. Close the output relay.
5. Clock/lock input frequency is continually compared to the target
frequency. If input frequency changed more than 10%, output
relay will open and an error message will be displayed.
When configured as an output, clock/lock signal outputs a square wave
of the same frequency as the power outputs. Select OFF to disable.
OFF will configure clock/lock as an input with the PLL mode disabled.
Table 33 System External Menu
Operation SW 5250ASW 3500ASW 1750A
3-28 Operation Manual
3.3.7.4 FAULT8 8ub-Menu
Pressing <Enter> to select FAULTS will display a list of logged fault conditions LOG1 -
LOG6. These are three 8-bit registers used to troubleshoot errors that cause the
system to shut down. Fault register definitions are available through the Help screens
of the SmartWave unit.
USER
COMM
EXTERNAL
FAULTS
CONFIG
FR1 FR2 FR3
LOG1: 00 00 00
LOG2: 00 00 00
LOG3: 00 00 00
LOG4: 00 00 00
LOG5: 00 00 00
LOG6: 00 00 00
SYSTEM OUTPUT [OFF]
Fault History
The FAULTS menu will display the last six shutdown faults in hexadecimal. All three
fault registers are displayed with LOG1 as the most recent fault and LOG6 the oldest.
Register bit definitions are shown in Table 34. Fault descriptions are shown in
Table 35 through Table 37.
FAULT REG8TER 1 FAULT REG8TER 2 FAULT REG8TER 3
8_ 8_ 8_ RLY156
4_ 4_ 4_ RLY312
2_ F3:OV 2_ F4:OT 2_
1_ DC_ERR 1_ F2:OV 1_
_8 ROV _8 UV _8
_4 OC _4 ROC _4
_2 48VLOW _2 RMS_OV _2
_1 GND_FLT _1 FLT_IN _1 RLY_FLT
Example: FR3 81: RLY156 and RLY_FLT
Table 34 Fault Register Bit Definitions
SW 5250ASW 3500ASW 1750A Operation
Operation Manual 3-29
Although the DFI Fault is not included in the fault history log, any of the faults will cause
a DFI error to occur. When a DFI fault is tripped, either because one of the faults above
or a fault external to the SW, the output relay is opened, VPROGRAM is set to zero,
and an error message is displayed.
FAULT REG8TER 1
F3:OV Indicates that the internal DC bus has pumped up to a hazardous
level (due to sinking power rather than sourcing, i.e., attempting to
discharge a very large capacitor into the SW output too quickly).
Action: SW system opens relays and shuts off.
DC_ERR Indicates that an unacceptable level of DC voltage is present in the
output when the AC coupling is selected (i.e., a waveform with DC
content was selected when AC coupling was active).
Action: SW system opens relays and shuts off.
ROV Redundant Over Voltage: The peak value of the output voltage has
exceeded the maximum capacity of the SW.
Action: Opens relays and sets voltage amplitudes to zero; displays
error message.
OC SW output current reached IPROGRAM value.
Action 1: Limit Mode: folds back voltage to maintain constant
current; displays i in right-hand corner.
Action 2: Shutdown Mode: opens relays and displays error
message.
Action 3: Time-Out Mode: Constant current for a specified time,
then opens relay and displays error message.
48V:LOW Monitors internal housekeeping supply.
Action: Shuts off SW system.
GND_FLT Monitors output neutral.
Action: If greater than 20 volts with reference to chassis ground,
opens relays and displays error message.
Table 35 Fault Description and Action (FR1)
Operation SW 5250ASW 3500ASW 1750A
3-30 Operation Manual
FAULT REG8TER 2
F4:OT Monitors power stage heat sink temperatures.
Action: Opens relays, displays error message for 30 seconds,
then shuts off SW system.
F2:OV Indicates that the internal DC bus has pumped up to an
unacceptable level (due to sinking power rather than sourcing,
i.e., attempting to discharge a very large capacitor into the SW
output too quickly).
Action: SW system tristates output amplifier until bus level
becomes normal.
UV Undervoltage: Indicates that the output voltage is less than the
programmed value, but not in the voltage foldback mode.
Action: Opens relays and displays error message.
ROC Redundant Over Current: The RMS value of the output current
has exceeded the maximum capacity of the SW.
Action: Opens relays and displays error message.
RMS_OV Monitors the external input in the External Direct mode to ensure
that it does not exceed 5.2 VRMS.
Action: Opens relays, displays error message, and sets
VPROGRAM to zero volts.
FLT_IN Fault-Input from Slave chassis: Represents the logical OR of
several faults.
Action: Opens relays, displays error message, and sets
VPROGRAM to zero volts. Clearing the error message
sends a fault reset to the Slave chassis. The particular
Slave chassis that had the fault will then extinguish its
fault indicator.
Table 36 Fault Description and Action (FR2)
FAULT REG8TER 3
RLY156,RLY312 Used for internal monitoring only.
RLY_FLT The open relay did not respond to the open/close command within
500 mSec.
Action: Programs the output to zero and attempts to change relay
state; displays error message and shuts off SW if output
relay hasnt changed state within 500 mSecs. If unable to
clear error message, a hazardous condition may exist. Use
caution; voltage may be present at the output terminals.
Table 37 Fault Description and Action (FR3)
SW 5250ASW 3500ASW 1750A Operation
Operation Manual 3-31
3.3.7.5 CONFG 8ub-Menu
Pressing <Enter> to select CONFIG will display the System Configuration options.
BOOST - Allows the Servo Boost circuit to be disabled during non-sequence operation.
This option is important when running a sequence at low voltage levels. See Section
1.12.2, Servo Boost, for a full explanation of the Boost function. The factory default
state is Boost ON. This option should be set to OFF when:
The RMS Servo is ON (Low Frequency = OFF) and
Sequence mode is being used with Exit mode set to Program Menu.
RESET CAL Resets calibration constants to default values. Enter a confirmation code
(2096) to activate this option. If you enter an incorrect code, the unit will beep and the
message ! INCORRECT RESET CODE will be displayed, the code will be reset to 0,
and no changes will occur to the calibration constants.
T&MB OPT Displays firmware configuration for the Test & Measurement option. ON
indicates the Test & Measurement board is installed; OFF indicates it is not. This field is
for display purposes only.
SWAE OPT Displays firmware configuration for the Low Impedance option. ON
indicates the Low Impedance hardware is installed; OFF indicates it is not. This field is
for display purposes only.
USER
COMM
EXTERNAL
FAULTS
CONFIG
BOOST ON
RESET CAL 0
T&MB OPT OFF
SWAE OPT OFF
SYSTEM OUTPUT [OFF]
CONFIG Sub-Menu Status and Information Data:
Menu Item Status and Information Line
BOOST [System boost enable/disable]
RESET CAL [Enter reset code]
T&MB OPT [T&MB board installed?]
SWAE OPT [Low impedance installed?]
Operation SW 5250ASW 3500ASW 1750A
3-32 Operation Manual
3.4 Front Panel Programming Conventions
When programming an amplitude in the Program Menu, Sequence Edit Menu,
or remotely over the GPIB, the Amplitude field is checked against the Function
(Waveform) field for positive or negative amplitude; that is, the function polarity is
used to validate the amplitude input value. For example, programming -100V
when a DC+ waveform is selected will result in an execution error.
The polarity check has the following rules:
Negative amplitudes cannot be input for positive functions.
Positive amplitudes cannot be input for negative functions.
If the amplitude of a phase is positive, selecting a negative polarity function
will set the amplitude to zero.
If the amplitude of a phase is negative, selecting a positive polarity function
will set the amplitude to zero.
Phases A, B, and C will be of the same frequency. A frequency change of any
phase will affect the output frequency of all phases. The frequency field is
available in all three programming screens as a convenience.
A sequence can be run on one, two, or all three phases. Two different
sequences cannot be run on two phases simultaneously.
A sequence setting will take priority over continuous settings.
SW 5250ASW 3500ASW 1750A Operation
Operation Manual 3-33
3.5 Front Panel Programming Exercises
3.5.1 Current nrush Example
This example illustrates programming the SW to simulate a current inrush waveform
that is 0V between 0and 90, then instantaneously jumps from 0V to 120V at 90.
1. At the MAIN MENU, use either the Knob or the Up/Down Arrow keys to select
the WAVEFORM menu, then either press the Enter key or the Right Arrow key.
2. The WAVEFORM Menu will default to the LOAD sub-menu. If the cursor is not
pointing to LOAD, use the Knob or Up/Down Arrow keys to select LOAD.
3. Press the Enter key to enter the FUNC (function) mode.
4. Press <Enter> again to place a box around the waveshape selection.
5. Use the Knob to cycle through the waveshape selections until SINE appears in
the FUNC box.
LOAD
EDIT
VIEW
SAVE
DELETE
WAVEFORM OUTPUT [OFF]
FUNC Sine
Displays library of waveforms
6. Press <Enter> to load the sine waveshape into scratchpad memory. The
message, [waveform successfully loaded], will indicate that the waveform is now
in scratchpad memory.
7. Press <Menu> to return to the WAVEFORM Menu.
8. Use either the Knob or the Up/Down Arrow keys to select the EDIT sub-menu,
then either press <Enter> or the Right Arrow key. The EDIT menu will be
displayed.
Operation SW 5250ASW 3500ASW 1750A
3-34 Operation Manual
LOAD
EDIT
VIEW
SAVE
DELETE
FREQ Hz
Vrms V
START Deg
TIME ms
STOP Deg
AMPL V
60.00
120.00
0.00
4.17
90.00
0.00
WAVEFORM OUTPUT [OFF]
9. Use either the Knob or the Up/Down Arrow keys to select FREQ. The frequency
is normally defaulted at 60 Hz with all other values set to 0. On the keypad, enter
6, 0, decimal point, 0, 0, then press <Enter> to set the frequency to 60.00 Hz.
10. Use either the Knob or the Up/Down Arrow keys to select Vrms. Using the
keypad, enter 1, 2, 0, then press <Enter> to set the RMS voltage to 120 volts.
11. Use either the Knob or the Up/Down Arrow keys to select START. Using the
keypad, enter 0, then press <Enter> to set the start phase angle to 0.
12. Although the TIME can be calculated then entered, the STOP function is used in
this example. Select STOP, enter 90 for a 90phase angle, and press <Enter>.
Note that the unit automatically calculates the TIME (in this case, 4.17 ms).
13. Since the operator is not programming a transient (drop/spike), leave the AMPL
(amplitude) at 0V.
14. Press the Menu key or the Left Arrow key to return to the WAVEFORM Menu.
LOAD
EDIT
VIEW
SAVE
DELETE
WAVEFORM OUTPUT [OFF]
SW 5250ASW 3500ASW 1750A Operation
Operation Manual 3-35
15. Select VIEW, then press <Enter> to view the waveform. A waveform similar to
the one illustrated below should be displayed on the LCD.
2
2
1
2
2
1
0
90 180 270
WAVEFORM OUTPUT [OFF]
16. To display the waveform on the front panel phase A output:
a. Return to the MAIN MENU by pressing <Menu> twice.
b. Select the PROGRAM Menu, then press <Enter>.
17. In the PHASE A sub-menu:
a. Select FUNC with the Knob, then press <Enter> to place a box around the
waveshape selection.
b. Rotate the Knob until SCRATCH is displayed, then press <Enter>.
PROGRAM
MEASURE
WAVEFORM
SEQUENCE
INSTR
SYSTEM
PHASE A
AMPL 120.00 V
CURL 5.00 A
FREQ 60.00 Hz
ANG 0.00 Deg
FUNC SCRATCH
MAIN MENU OUTPUT [OFF]
Operation SW 5250ASW 3500ASW 1750A
3-36 Operation Manual
18. While still in the PHASE A sub-menu:
a. Select AMPL (amplitude) with the Knob or Arrow Keys.
b. Enter 120 with the keypad to set the amplitude at 120 volts, then press
<Enter>. The inrush waveform will now be displayed on Phase A output on
the front panel.
19. While still in the PROGRAM mode:
a. Select PHASE B by cursoring to Phase A, then press <Enter>.
b. Select FUNC with the Knob, then press <Enter> to place a box around the
waveshape selection.
c. Rotate the Knob to select TRIANGLE, then press <Enter>.
d. Select AMPL (amplitude) with the Knob, enter 120 with the keypad, then
press <Enter>. A triangular waveform will be displayed on Phase B output
on the front panel.
e. With the field select cursor still at AMPL, press <Enter> to place a box
around the waveshape selection then, using the Knob, vary the amplitude
of the triangular waveform on Phase B.
f. Select PHASE C by cursoring to Phase B, then press <Enter>.
g. Select FUNC with the Knob, then press <Enter> to place a box around the
waveshape selection.
h. Rotate the Knob to select SQUARE, then press <Enter>.
i. Select AMPL (amplitude) with the Knob, enter 120 with the keypad, then
press <Enter>. A square waveform will now be displayed on Phase C
output on the front panel.
j. With the field select cursor still at AMPL, press <Enter> to place a box
around the amplitude selection then, using the Knob, vary the amplitude of
the square waveform on Phase C.
20. Press <Menu> or the Left Arrow key to return to the MAIN MENU.
SW 5250ASW 3500ASW 1750A Operation
Operation Manual 3-37
3.5.2 Voltage 8pike Example
This example illustrates programming the SW for a voltage spike on a 150V sine
waveform that starts at 30and ends at 50with an amplitude of 220.6 volts.
PROGRAM
MEASURE
WAVEFORM
SEQUENCE
INSTR
SYSTEM
MAIN MENU OUTPUT [OFF]
1. At the MAIN MENU, select the WAVEFORM menu, then press <Enter>.
LOAD
EDIT
VIEW
SAVE
DELETE
WAVEFORM OUTPUT [OFF]
2. The WAVEFORM Menu will default to the LOAD sub-menu. If the cursor is not
pointing to LOAD, use the Knob or Up/Down Arrow keys to select LOAD.
3. Press the Enter key to enter the FUNC (function) mode.
4. Press <Enter> again to place a box around the waveshape selection.
5. Use the Knob to cycle through the waveshape selections until SINE appears in
the FUNC box.
Operation SW 5250ASW 3500ASW 1750A
3-38 Operation Manual
LOAD
EDIT
VIEW
SAVE
DELETE
WAVEFORM OUTPUT [OFF]
FUNC Sine
Displays library of waveforms
6. Press <Enter> to load the sine waveshape into scratchpad memory. The
message, [waveform successfully loaded], will indicate that the waveform is now
in scratchpad memory.
7. Press <Menu> to return to the WAVEFORM Menu.
8. Use either the Knob or the Up/Down Arrow keys to select the EDIT sub-menu,
then either press <Enter> or the Right Arrow key. The EDIT menu will be
displayed.
LOAD
EDIT
VIEW
SAVE
DELETE
FREQ Hz
Vrms V
START Deg
TIME ms
STOP Deg
AMPL V
WAVEFORM OUTPUT [OFF]
60.00
120.00
30.00
0.93
50.00
220.60
9. Select Vrms. Enter 150 using the keypad, then press <Enter> to set the RMS
voltage to 150 volts.
SW 5250ASW 3500ASW 1750A Operation
Operation Manual 3-39
10. Select START. Enter 30 using the keypad, then press <Enter> to set the
starting phase angle to 30.
11. Select STOP. Enter 50 using the keypad for an ending phase angle of 50, then
press <Enter>. Note that the instrument automatically calculates the TIME (in
this case, 0.93 ms).
12. Select AMPL. Enter 220.6, then press <Enter> to set the amplitude to 220.6V.
13. Press <Menu> to return to the WAVEFORM Menu.
LOAD
EDIT
VIEW
SAVE
DELETE
WAVEFORM OUTPUT [OFF]
14. Select VIEW to view the scratchpad, then press <Enter>. A waveshape similar
to the one illustrated below should be displayed on the scratchpad.
2
2
1
2
2
1
0
90 180 270
WAVEFORM OUTPUT [OFF]
Operation SW 5250ASW 3500ASW 1750A
3-40 Operation Manual
15. To retain the waveshape and display it on Phase A output on the front panel:
a. Return to the MAIN MENU by pressing <Menu> twice.
b. Select the PROGRAM Menu, then press <Enter>.
PROGRAM
MEASURE
WAVEFORM
SEQUENCE
INSTR
SYSTEM
PHASE A
AMPL 120.00 V
CURL 5.00 A
FREQ 60.00 Hz
ANG 0.00 Deg
FUNC SCRATCH
MAIN MENU OUTPUT [OFF]
16. In the PHASE A sub-menu:
a. Select FUNC with the Knob, then press <Enter> to place a box around the
waveshape selection.
b. Rotate the Knob until SCRATCH is displayed, then press <Enter>. The
scratchpad waveform will now be displayed on Phase A of the oscilloscope.
17. Press <Menu> or the Left Arrow key to return to the MAIN MENU.
SW 5250ASW 3500ASW 1750A Operation
Operation Manual 3-41
3.5.3 Voltage Dropout Example
This example illustrates programming the SW for a voltage dropout on a 120V sine
waveform that starts at 45with a duration of 1 millisecond and an amplitude of 10
volts.
PROGRAM
MEASURE
WAVEFORM
SEQUENCE
INSTR
SYSTEM
MAIN MENU OUTPUT [OFF]
1. At the MAIN MENU, select the WAVEFORM menu, then press <Enter>.
LOAD
EDIT
VIEW
SAVE
DELETE
WAVEFORM OUTPUT [OFF]
2. The WAVEFORM Menu will default to the LOAD sub-menu. If the cursor is not
pointing to LOAD, use the Knob or Up/Down Arrow keys to select LOAD.
3. Press the Enter key to enter the FUNC (function) mode.
4. Press <Enter> again to place a box around the waveshape selection.
5. Use the Knob to cycle through the waveshape selections until SINE appears in
the FUNC box.
Operation SW 5250ASW 3500ASW 1750A
3-42 Operation Manual
LOAD
EDIT
VIEW
SAVE
DELETE
WAVEFORM OUTPUT [OFF]
FUNC Sine
Displays library of waveforms
6. Press <Enter> to load the sine waveshape into scratchpad memory. The
message, [waveform successfully loaded], will indicate that the waveform is now
in scratchpad memory.
7. Press <Menu> to return to the WAVEFORM Menu.
8. Use either the Knob or the Up/Down Arrow keys to select the EDIT sub-menu,
then either press <Enter> or the Right Arrow key. The EDIT menu will be
displayed.
LOAD
EDIT
VIEW
SAVE
DELETE
FREQ Hz
Vrms V
START Deg
TIME ms
STOP Deg
AMPL V
WAVEFORM OUTPUT [OFF]
60.00
120.00
45.00
1.00
66.60
10.00
9. Select Vrms. Enter 120, then press <Enter> to set the RMS voltage to 120
volts.
SW 5250ASW 3500ASW 1750A Operation
Operation Manual 3-43
10. Select START. Using the keypad, enter 45, then press <Enter> to set the
starting phase angle to 45.
11. Select TIME, enter 1 for a duration of 1 millisecond, then press <Enter>. Note
that the instrument automatically calculates the STOP phase angle value (in this
case, 66.6).
12. Select AMPL, enter 10 for 10 volts, then press <Enter>.
13. Press <Menu> to return to the WAVEFORM Menu.
LOAD
EDIT
VIEW
SAVE
DELETE
WAVEFORM OUTPUT [OFF]
14. Select VIEW to view the scratchpad, then press <Enter>. A waveshape similar
to the one illustrated below should be displayed on the scratchpad.
2
2
1
2
2
1
0
90 180 270
WAVEFORM OUTPUT [OFF]
Operation SW 5250ASW 3500ASW 1750A
3-44 Operation Manual
15. To retain the waveshape and display it on Phase A of the oscilloscope:
a. Return to the MAIN MENU by pressing <Menu> twice.
b. Select the PROGRAM Menu, then press <Enter>.
PROGRAM
MEASURE
WAVEFORM
SEQUENCE
INSTR
SYSTEM
PHASE A
AMPL 120.00 V
CURL 5.00 A
FREQ 60.00 Hz
ANG 0.00 Deg
FUNC SCRATCH
MAIN MENU OUTPUT [OFF]
16. In the PHASE A sub-menu:
a. Select FUNC with the Knob, then press <Enter> to place a box around the
waveshape selection.
b. Rotate the Knob until SCRATCH is displayed, then press <Enter>. The
scratchpad waveform will now be displayed on Phase A of the oscilloscope.
17. Press <Menu> or the Left Arrow key to return to the MAIN MENU.
3.6 8CP 8pecification
Refer to the SmartWave Switching Amplifier SCPI Programming Manual
(Elgar Document No. M162000-03).
Operation Manual 4-1
8ECTON 4
8TANDARD WAVEFORM8
The following standard waveforms are provided in the SW Series:
NAME DE8CRPTON
8CALE
FACTOR
FREOUENCY
RANGE
Clip0 A sine wave with the positive halfcycle clipped at 0V from
50 to 130
1.0 40-5000 Hz
ClipPos A sine wave with the positive halfcycle clipped from
50 to 130
1.0 40-5000 Hz
ClipNeg A sine wave with the positive halfcycle set to a 50%
negative value from 50 to 130
1.0 40-5000 Hz
DC+ Used to operate the SW as a positive DC source
(true rms)
0.7071 N/A
DC- Used to operate the SW as a negative DC source
(true rms)
0.7071 N/A
DcRip03+ +DC with 3% ripple. Frequency of the ripple is deter-
mined by the programmed output frequency (true rms)
0.7071 40-5000 Hz
DcRip03- -DC with 3% ripple. Frequency of the ripple is deter-
mined by the programmed output frequency (true rms)
0.7071 40-5000 Hz
DcRip10+ +DC with 10% ripple. Frequency of the ripple is deter-
mined by the programmed output frequency (true rms)
0.7071 40-5000 Hz
DcRip10- -DC with 10% ripple. Frequency of the ripple is deter-
mined by the programmed output frequency (true rms)
0.7071 40-5000 Hz
Drop45 Sine with a drop-out above 0V from 45 to 60 1.0 40-5000 Hz
DropNeg Sine with a drop-out to -Vpeak from 4 to 50 1.0 40-5000 Hz
DropPeak Sine wave with drop-out to 0V from 8 to 110 1.0 40-5000 Hz
ExtrnDir DC for use with System External direct input 1.0 40-5000 Hz
FlatTp05 Flat-top sine wave with 5% distortion (true rms) 0.9344 40-5000 Hz
FlatTp10 Flat-top sine wave with 10% distortion (true rms) 0.8894 40-5000 Hz
Standard Waveforms SW 5250ASW 3500ASW 1750A
4-2 Operation Manual
NAME DE8CRPTON
8CALE
FACTOR
FREOUENCY
RANGE
FlatTp15 Flat-top sine wave with 15% distortion (true rms) 0.8545 40-5000 Hz
FlatTp20 Flat-top sine wave with 20% distortion (true rms) 0.8251 40-5000 Hz
Four3 Fourier square wave with 1st and 3rd harmonics
(true rms)
0.8946 40-5000 Hz
Four5 Fourier square wave with 1st, 3rd, and 5th harmonics
(true rms)
0.8703 40-5000 Hz
Four7 Fourier square wave with 1st, 3rd, 5th, and 7th harmonics
(true rms)
0.8595 40-5000 Hz
Four9 Fourier square wave with 1st, 3rd, 5th, 7th, and 9th
harmonics (true rms)
0.8537 40-5000 Hz
FulRect+ Positive DC full rectified sine wave 1.0 40-5000 Hz
FulRect- Negative DC full rectified sine wave 1.0 40-5000 Hz
HalfCyc Sine wave with 90 to 180at 0Volts 1.0 40-5000 Hz
HlfRect+ Positive DC half rectified sine wave 1.0 40-5000 Hz
HlfRect- Negative DC half rectified sine wave 1.0 40-5000 Hz
Inrush Sine wave with first quarter (0 to 90) at 0Volts 1.0 40-5000 Hz
Noisz010 Sine with 10% noise at zero crossings (true rms) 1.0 40-5000 Hz
Noisz100 Sine with 100% noise at zero crossings (true rms) 1.0 40-5000 Hz
Pcntl110 Phase control at 110 1.0 40-5000 Hz
Pcntl170 Phase control at 170 1.0 40-5000 Hz
Scratch Waveform scratchpad. This is the waveform edit work
area. Waveforms being edited can be output for testing
purposes before they are saved to non-volatile memory.
1.0 Maximum of
waveform
loaded
SinDc01+ Sine wave with +1% DC offset 1.0 40-5000 Hz
SinDc01- Sine wave with - 1% DC offset 1.0 40-5000 Hz
SinDc10+ Sine wave with +10% DC offset 1.0 40-5000 Hz
SinDc10- Sine wave with - 10% DC offset 1.0 40-5000 Hz
SinDc20+ Sine wave with +20% DC offset 1.0 40-5000 Hz
SinDc20- Sine wave with - 20% DC offset 1.0 40-5000 Hz
SinDc50+ Sine wave with +50% DC offset 1.0 40-5000 Hz
SinDc50- Sine wave with - 50% DC offset 1.0 40-5000 Hz
Sine Sine wave (true rms) 1.0 40-5000 Hz
Spike200 Sine with spikes at peaks. Spikes are from 85 to 95
and calibrated to be 200Vpeak when programmed to
120Vrms
1.1785 40-5000 Hz
SW 5250ASW 3500ASW 1750A Standard Waveforms
Operation Manual 4-3
NAME DE8CRPTON
8CALE
FACTOR
FREOUENCY
RANGE
Spike250 Sine with spikes at peaks. Spikes are from 85 to 95
and calibrated to be 250Vpeak when programmed to
120Vrms
1.4731 40-5000 Hz
Spike300 Sine with spikes at peaks. Spikes are from 85 to 95
and calibrated to be 300Vpeak when programmed to
120Vrms
1.7678 40-5000 Hz
Spike400 Sine with spikes at peaks. Spikes are from 85 to 95
and calibrated to be 400Vpeak when programmed to
120Vrms
2.3570 40-5000 Hz
Square Square wave with 50% duty cycle (true rms) 0.7071 40-5000 Hz
Step_06 Six voltage step sine wave 1.0 40-5000 Hz
Step_12 Twelve voltage step sine wave 1.0 40-5000 Hz
Step_24 Twenty-four voltage step sine wave 1.0 40-5000 Hz
Taylor 5th harmonic Taylor series wave (true rms) 0.9136 40-5000 Hz
Triangle Triangle wave (true rms) 1.2246 40-5000 Hz
Z1 3-phase output with a voltage spike occurring at the same
point in time. Spikes are from 45 to 90and calibrated to
be 200 Vpeak when programmed to 80 Vrms
1.0 40-5000 Hz
Z2 3-phase output with a voltage spike occurring at the same
point in time. Spikes are from 165 to 210and calibrated
to be 200 Vpeak when programmed to 80 Vrms
1.0 40-5000 Hz
Z3 3-phase output with a voltage spike occurring at the same
point in time. Spikes are from 285 to 330and calibrated
to be 200 Vpeak when programmed to 80 Vrms
1.0 40-5000 Hz
Standard Waveforms SW 5250ASW 3500ASW 1750A
4-4 Operation Manual
This page intentionally left blank.