How To Connect Two or More Signal Generators To Create A Multi-Channel Waveform Generator
How To Connect Two or More Signal Generators To Create A Multi-Channel Waveform Generator
Introduction
A single-channel function generator or arbitrary waveform generator meets the needs of most engineers most of the time. But occasionally you face a measurement challenge where you need to have two or more source channels locked together in frequency and phase. To deliver maximum performance at a reasonable price, test-equipment manufacturers design most function generators with a single channel. Singlechannel instruments can be simpler and easier to use than multi-channel instruments. In an environment where engineers tend to share test equipment, it makes sense to purchase single-channel sources and use two single-channel sources in tandem when you need multiple channels, rather than spending a limited budget on a single, dual-channel instrument. With single-channel sources you can disconnect the units and use them separately again when you are finished. This application note describes the steps you need to take to connect two or more signal generators to create a multi-channel waveform generator.
constant phase offset. If the two waveforms are not phase-locked, the oscilloscope will only be able to freeze (trigger on) one of the waveforms; the other waveform will appear to walk. To phase-lock two or more single-channel sources, you need a common clock signal, often called an external reference. Agilent Technologies offers two models of singlechannel function/arbitrary waveform generators: the 33120A, an economical 15-MHz source, and the 33250A, an 80-MHz higher-performance generator. If you want to combine multiple 33120As, you need to add optional external reference and BNC connectors (Option 001) to each generator so they will accept the external reference. If you already own a 33120A, you can look at the rear panel and determine if it has the optional BNC connectors below the labels Ext Ref In and Ref Out. If your 33120A does not have Option 001 installed, usually it
Figure 1: A 33120A (top) receiving an external reference signal from a 33250A (bottom).
can be retrofitted. You can read more about retrofitting the option on Agilent's Web site. Use the site search engine to find part number 33120-80001. The higher-performance 33250A includes external reference capabilities as a standard feature. External clock frequencies may vary slightly, and both the 33120A and 33250A can synchronize to a signal that is not perfectly calibrated. The table below shows the acceptable frequency range for the generators.
Lowest External Reference Frequency 10 MHz - 50 Hz 10 MHz - 35 kHz Highest External Reference Frequency 10 MHz + 50 Hz 10 MHz + 35 kHz
More on phase-locking
The term phase-lock may have several different meanings, depending upon how it is used with instruments. Analog generators tend to have poor stability and rely upon an external phase-lock loop to create additional stability. The function generator could be locked to any signal within its frequency range. Today, lab-quality function/arbitrary waveform generators use built-in clocks and digital circuitry for precise frequency control. These waveform generators can also be referenced to an external source, to which the generator will phase lock. It is important to note that digital function generators will lock only to cardinal frequencies (for example, 10 MHz, 1 MHz, 1 kHz, etc.). Most generators will be able to lock onto a range of frequencies near the cardinal frequency (50 Hz). Some will allow harmonics of the cardinal frequency or multiple frequencies. Once a function generator is phase-locked to another function generator, the two generators will have an unknown phase offset. Usually you can manually adjust the phase or use a common trigger to start the waveforms simultaneously. phase difference to the desired amount. A scope or universal frequency counter can be used to observe the phase relationship between the two instruments. The phase relationship can be changed via the front panel of the source or programmatically. Alternately, you could trigger both generators using a common trigger signal. With both the 33120A and the 33250A, there is a second BNC available so you can use the Trig In/Out BNC of one generator to trigger the other generators. Sharing a common trigger will create a phase offset of 25 ns or less when you trigger a 33120A and 1 ns when you trigger a 33250A. However, you can set the phase difference even more precisely with a scope or counter and manually adjusting the offset, as mentioned earlier.
33120A 33250A
The 33120A function generator's datasheet has more details about the external reference signal. The good news is that if you are using multiple 33120As or 33250As, you can use one unit to supply the external clock signal to the other units with a BNC cable. Once the generators are locked, you can set the actual phase difference between the two generators. Typically, you start by determining the phase difference between the signals. And, then adjust the
42V Max
42V Max
42V Max
42V Max
AM Modulation (5V=100%)
AM Modulation (5V=100%)
42V Max
LR33773-C LR33773-C
42V Max
Figure 2: Two phase-locked 33120A generators. The generators are using a common clock signal and a common trigger signal.
Glossary
BNC (Bayonet Neill Concelman) a type of connector commonly used with coaxial cables, often referred to as simply a bayonet connector External clock reference the reference signal used for timing purposes obtained from a source external to the system, or for the purposes of this article, external to the instrument Frequency lock The state that occurs when an oscillator is controlled by a frequency-correcting feedback loop for at least one cycle. Note: Phase locked oscillators are also in frequency lock, but oscillators controlled by a frequency lock do not necessarily maintain phase lock. Phase drift the change seen during the free run of two signals of slightly different periods. The phase of one signal will vary slightly but consistently in relation to a specific phase angle of the reference signal. Phase lock the state that occurs when an oscillator is controlled by a phase-correcting feedback loop for at least one cycle. A constant phase relationship is maintained between the phase of the oscillator and the reference signal of the feedback loop.
Summary
It is simple to connect two or more signal generators to create a multi-channel waveform generator. This technique works for all types of waveforms: sine waves, square waves, and arbitrary waveforms. When you need only a single channel you can disconnect the units and use them separately.
Phase offset the phase difference between two signals; may be defined in phase time (seconds) or phase angle (radians or degrees)
www.agilent.com
Agilent Email Updates www.agilent.com/find/emailupdates Get the latest information on the products and applications you select.
Agilent T&M Software and Connectivity Agilent's Test and Measurement software and connectivity products, solutions and developer network allows you to take time out of connecting your instruments to your computer with tools based on PC standards, so you can focus on your tasks, not on your connections. Visit www.agilent.com/find/connectivity for more information. By internet, phone, or fax, get assistance with all your test and measurement needs Online assistance: www.agilent.com/find/assist Phone or Fax United States: (tel) 800 452 4844 Canada: (tel) 877 894 4414 (fax) 905 282 6495 China: (tel) 800 810 0189 (fax) 800 820 2816 Europe: (tel) (31 20) 547 2323 (fax) (31 20) 547 2390 Japan: (tel) (81) 426 56 7832 (fax) (81) 426 56 7840 Korea: (tel) (82 2) 2004 5004 (fax) (82 2) 2004 5115 Latin America: (tel) (305) 269 7500 (fax) (305) 269 7599 Taiwan: (tel) 0800 047 866 (fax) 0800 286 331 Other Asia Pacific Countries: (tel) (65) 6375 8100 (fax) (65) 6836 0252 (e-mail) [email protected]