The Raman Effect
The Raman Effect
Introduction
While Raman spectroscopy has long been recognized as a valuable research technique in the years since the phenomenon was rst observed by Dr. C. V. Raman in 1928, it is only fairly recently that Raman has emerged as an important analytical tool across a number of industries and applications. No longer designed to appeal only to highly specialized and trained experts, the best of todays Raman instruments are fully integrated and come with built-in system intelligence that frees the user to focus on results and not on having to become an expert in the technology itself. Busy analytical laboratories are now able to adopt Raman spectroscopy without having to devote time to developing the expertise that used to be essential in order to be able to collect high quality data. Due to its sensitivity, high information content, and non-destructive nature, Raman is now used in many applications across the elds of chemistry, biology, geology, pharmacology, forensics, pharmaceuticals, materials science, and failure analysis. Spectral libraries in excess of 16,000 compounds are now available for direct compound identication. In many laboratories, infrared and Raman spectroscopy are used as complementary techniques, because each method looks at different aspects of a given sample. While IR is sensitive to functional groups and to highly polar bonds, Raman is more sensitive to backbone structures and symmetric bonds. Using both techniques provides twice the information about the vibrational structure than can be obtained by using either alone. In addition to providing unique information about a sample, Raman offers several additional benets, including: Minimal or no sample preparation Sampling directly through glass containers Non-destructive analysis, so the same sample can be used in other analyses Non-intrusive analysis, permitting study of more labile sample features, such as crystal structure Minimal water interference No interference from atmospheric CO2 or H2O The objective of this booklet is to provide an introduction to the technique of Raman spectroscopy and its relevance as an analytical tool. The rst section contains a simple description of the theory behind the Raman effect, followed by a discussion of the relevance of Raman spectroscopy and its complementarity with other vibrational techniques. This is followed by an overview of the two major Raman instrument designs: dispersive Raman and Fourier transform Raman (FT-Raman). Advantages and key applications are discussed for each of these instruments, followed by a list of desirable features in modern instruments.
As illustrated in the simplied energy level diagram, a molecule at rest resides in the ground vibrational and electronic states. The electric eld of the laser beam raises the energy of the system for an instant by inducing a polarization in the chemical species. The polarized condition is not a true energy state and is referred to as a virtual state. Relaxation from the virtual state occurs almost instantaneously and is predominantly to the initial ground state. This process results in Rayleigh scatter, which is scattered light of the same wavelength as the excitation laser. Relaxation to the rst excited vibrational level results in a Stokes-Raman shift. Stokes-Raman shift scattered light is of lower energy (longer wavelength) than that of the laser light. In addition, most systems have at least a small population of molecules that are initially in an excited vibrational state. When the Raman process initiates from the excited vibrational level, relaxation to the ground state is possible, producing scatter of higher energy (shorter wavelength) than that of the laser light. This type of scatter is called anti-Stokes Raman scatter and is not illustrated.
The vibrational states probed by Raman spectroscopy are similar to those involved in infrared spectroscopy. However, the two vibrational spectroscopy techniques are complementary, in that vibrations that are strong in an infrared spectrum (those involving strong dipole moments) are typically weak in a Raman spectrum. Likewise, non-polar functional group vibrations that give very strong Raman bands usually result in weak infrared signals. As an example, hydroxyl- or amine-stretching vibrations and the vibrations of carbonyl groups are usually very strong in an FT-IR spectrum and are weak in a Raman spectrum. The stretching vibrations of carbon double and triple bonds and the symmetric vibrations of aromatic groups give a very strong Raman signal. Raman spectroscopy provides key information about the structure of molecules. The position and intensity of features in the spectrum reect the molecular structure and can be used to determine the chemical identity of the sample. Spectra may also show subtle changes depending on the crystalline form. With the extensive spectral libraries that are now available, it is very straightforward to identify compounds by spectral library searching.
Gratings are designed for optimum throughput over a relatively narrow wavelength range, so gratings should be selected for the desired resolution and for the correct laser wavelength. Using a single grating for more than one laser wavelength requires a compromise in instrument throughput and sensitivity. Ideally, gratings should be matched specically to each laser. The CCDs commonly used for dispersive Raman are highly sensitive silicon devices. The detecting surface of the CCD is a two-dimensional array of light-sensitive elements, called pixels (typically each pixel is < 30 m). Each pixel acts as an individual detector, so each dispersed wavelength is detected by a different pixel (or closely spaced group of pixels). CCD detectors commonly have a large wavelength response region, routinely extending from 400 nm up to approximately 1000 nm. Specialized detectors extend the response up to approximately 1100 nm and down into the UV range. This means that the longest excitation wavelength that can be used with a silicon CCD detector, without loss of the higher shifted wavenumber portions of the Raman spectrum, is about 780 to 785 nm. (A 3300 cm-1 Stokes shift from 780 nm corresponds to 1050 nm.) The low end cutoff of the Raman spectrum is determined by the ability of the Rayleigh lters to exclude Rayleigh scattering (light at the excitation laser wavelength) from the spectrograph. The performance of a Raman spectrometer at the low shifted wavenumber end of the spectrum also depends on the adequacy of the laser line lters. Good modern instruments should be able to attain at least a 100 cm-1 cutoff. Since inorganic compounds display important Raman bands below 100 cm-1, an instrument that can deliver a 50 cm-1 low end cutoff is ideal.
Raman Microscopy
Dispersive Raman microscopy is ideal for the analysis of very small samples. Since spatial resolution is diffraction-limited, the short excitation laser wavelengths used for dispersive Raman are optimal for analyzing small sample features. With 532 nm excitation, a modern Raman microscope can achieve sub-micron spatial resolution. The analysis of small defects in the polymer lms used in liquid crystal displays is a typical application. In order to achieve micron-level spatial resolution, it is critical that a Raman microscope is optimally aligned. In order to target and analyze a micron-sized particle, the visual light path, the excitation laser beam path and the Raman scatter beam path from the sample to the spectrograph detector must all be targeted precisely on the same spot. Since all instruments will become misaligned at the micron level as a result of exposure to thermal shifts and other environmental changes, the instrument design must make it easy for the user to perform precision alignment on a routine basis in order to ensure optimal performance. Placing a sufciently small aperture in the focal plane of the microscope makes it possible to perform confocal microscopy. Light rays from surrounding regions of the sample are blocked by the aperture and only rays from the optical focal point pass through the aperture to the detector. This is a powerful technique for nondestructively probing the depths of transparent samples without having to cross-section them. It is widely used to investigate inclusions and for analyzing polymer laminates.
FT-Raman Spectroscopy
In place of visible excitation lasers, an FT-Raman spectrometer uses a laser in the near infrared usually at 1064 nm. At this wavelength uorescence is almost completely absent, however because of the 1/4 relationship between Raman scattering intensity and wavelength, the Raman signal is weak. In addition, silicon CCD detectors cannot be used in this region of the spectrum. FT-Raman uses sensitive, single-element, near-infrared detectors such as indium gallium arsenide (InGaAs) or liquid nitrogen-cooled germanium (Ge) detectors. An interferometer converts the Raman signal into an interferogram, permitting the detector to collect the entire Raman spectrum simultaneously. Since at low signal levels the spectral noise is predominantly detector dark noise and is independent of the intensity of the Raman signal, delivering the entire spectrum at once onto the detector greatly improves the signalto-noise ratio. Application of the Fourier transform algorithm to the interferogram converts the results into a conventional Raman spectrum.
In addition to freedom from uorescence interference, another advantage of FT-Raman spectroscopy is its exceptionally good x-axis (shifted wavenumber) accuracy as a result of the internal interferometer calibration supplied by the built-in helium-neon laser. Both of these attributes make FT-Raman the ideal technique for collecting spectra for reference libraries.
FT-Raman spectroscopy is particularly well-suited for bulk sample analysis and can be congured to accept samples in most common formats, including vials, cuvettes, tubes, plastic bags, bottles, powders, lms and solids. The Thermo Scientic NXR FT-Raman MicroStage allows users to mount samples horizontally for easier handling and features an integrated video camera. With a minimum laser spot size of 50 m, the MicroStage provides microscope capabilities and sufcient spatial discrimination to be able to analyze heterogeneous samples.
Selecting an Instrument
Since most modern Raman instruments are purchased to be used as analytical tools, instrument design should focus on making Raman easy to use without compromising performance. A well-designed instrument lets its user focus on results and not on having to become a Raman expert. The following list includes some of the things to look for:
Easy to Use
Features such as autoexposure, smart backgrounds and autofocus rely on the instrument and software intelligence to optimize data collection parameters. This is analogous to the way that the expertise required to achieve professional-level photography has been embedded into digital cameras. Components such as lasers, gratings, Rayleigh lters and sampling accessories should be easy to exchange, either through automation and a mouse-click, or by employing SmartLock technology to ensure reproducible results. Smart components are readily recognized by the system software, their identities recorded for full traceability and the user is alerted if components are incompatible with each other.
Laser Safe
Instruments where the user cannot be exposed to a direct laser beam are classied as Class I laser safe and can be used in open laboratories. Class IIIb laser-safe instruments have to be used in restricted environments in order to prevent inadvertent exposure to the laser beam and are therefore much less convenient to use.
High Performance
High spatial resolution: one of the key reasons for purchasing a dispersive Raman microscope is to take advantage of its micron-level spatial resolution. In order to achieve this performance consistently, the instrument design must make it easy to maintain optimal alignment. The Thermo Scientic Nicolet Almega XR and Thermo Scientic DXR Raman microscope spectrometers employ a rapid, patented, software-driven alignment method that makes it easy to incorporate instrument alignment into routine laboratory maintenance procedures. Full spectral range: in order to take full advantage of the power and information content of Raman spectroscopy, spectrometers should cover the full spectral range; from 50 cm-1 at the low end to 3500 cm-1 at the high end of the spectrum.
Reproducibility: In the Past Raman had a Reputation for Poor Reproducibility, However this can be Addressed by Careful Design
The excitation laser power can vary considerably from one laser to the next and from one instrument to the next. Incorporation of a laser power regulator that permits the user to specify and regulate the laser power at the sample eliminates this cause for non-reproducible Raman scattering intensity. Automatic intensity correction compensates for the wavelength dependence of the silicon CCD detector response and results in reproducible peak intensities whatever the excitation wavelength. X-axis non-reproducibility is the result of inadequate spectrograph and laser calibration. Automated, software-driven multipoint laser and spectrograph calibrations are essential for reliable instrument-to-instrument comparisons. This is particularly important for successful sample identication using library searching. Heterogeneous samples, such as tablets or powders will give non-reproducible results if the excitation laser spot size is smaller than the heterogeneous components. Sampling devices for these applications should be designed to raster over the surface of the sample so that the excitation beam and the Raman scatter are collected over a representative portion of the sample surface. The Thermo Scientic DXR SmartRaman employs VDPS (Variable Dynamic Point Sampling) technology to automate this.
10
Conclusion
Raman spectroscopy is a technique which has evolved from what used to be regarded as an interesting research phenomenon into a valuable analytical tool. As a complementary technique to FT-IR, Raman offers benets that include minimal sample preparation, the ability to sample through containers, non-destructive analysis, easy analysis of inorganics, and minimal interference from water. Modern FT-Raman and dispersive Raman spectrometers are easy to use, freeing analytical chemists to focus on results and not on having to become an expert in the technique. Improvements in instrument design, particularly in instrument calibration and reproducibility, now make it easy to use Raman spectroscopy to identify unknown samples using spectral library searching. Raman is also used quantitatively, in QC environments, in forensics, in the analysis of cultural artifacts, gems and gemology, materials research, pharmaceuticals and an ever-expanding range of applications and industries.
11
Africa +43 1 333 5034 127 Australia +61 2 8844 9500 Austria +43 1 333 50340 Belgium +32 2 482 30 30 Canada +1 800 530 8447 China +86 10 8419 3588 Denmark +45 70 23 62 60 Europe -Other +43 1 333 5034 127 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9434 Italy +39 02 950 591 Japan +81 45 453 9100 Latin America +1 608 276 5659 Middle East +43 1 333 5034 127 Netherlands +31 76 579 55 55 South Africa +27 11 570 1840 Spain +34 914 845 965 Sweden / Norway / Finland +46 8 556 468 00 Switzerland +41 61 48784 00 UK +44 1442 233555 USA +1 800 532 4752
www.thermo.com
2003, 2008 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.