Electrical Measuring: Unit 10
Electrical Measuring: Unit 10
Structure
10.1 Introduction
Objectives
10.6.2
10.6.3
10.6.4
10.9 Summary
10.10 Answers to SAQs
10.1 INTRODUCTION
In the preceding lessons, you have learnt that current, voltage, power and other electrical
variables are expressed in their respective units of A, V, W etc. In this chapter, we shall
study the principle of working and application of instruments which enable the
rneasurement of these variables. The presence of an electrical current or voltage can
ordinarily be felt only by some physical effect produced by it. For example, current through
a conductor causes thermal heating. The measurement of an electrical quantity is facilitated
by exploiting any one of such effects. Here, we shall confine ourselves to the study of
electrical instruments which use the electrortzagnetic effect.
Objectives
After going through this unit, you should be able to
explain the distinction between analog and digital instruments and their relative
merits,
outline the basic features of a general analog indicating instrument,
explain briefly the errors that could occur in a typical measurement task.
digits. Figure 10.1 shows the read-oul systems of the two types.
It is clear that the digital instrument will be free from 'reading error' which can be present
when judging the posilion of a pointer lying between two scale divisions on an analog
instrument. On the other hand, il often happens that one or more of the least significant
digits in a digilal instrument may be constantly fluctuating. This may cause discomfort to an
operator keeping a manual watch on the readings on different instruments. The position of
the pointer of an 'analog instrument on the other hand, would give the operator at a distance,
an idea of the approximate value of the measurand. With the input quantity varying over a
given interval of time, the conlinuous movement of the pointer would give the nh~erveran
t
idea of the trend of variat~on.
In this unit we shall learn about the iinporlant types of 'analog electrical instruments. Analog
instruments which employ electronic circuits will be treated in Unit 15. You will learn
about digital iilstruments in Unit 16.
SAQ 1
an indicating mechanism.
The deflection mechanism causes a deflecting torque to act on the moving system of the
instrument. One part of this mechanism will be usually located on the stationary member of
the instrument and the other part on the moving member. Under llie action of the deflecting
torque (Md)* which is a function of the quantity to be measured (qi), the moving system
starts to rotate from the initial position. The details of the deflection mechanism depend on
the measurand and the effect of electrical current or voltage being exploited for the purpose
of measurement. The control mechanism serves to ensure that the rotation of the moving
system is limited to a fraction of a revolution and that the final deflection is to a definite
position for a given measurand q,. A pair of phosphor-bronze helical springs are usually
employed for this purpose. They excrt on the moving system, a control torque (M,), which
acts in a direction opposite to that of Md and is proportional to the angular displacement (0)
from the initial position. A damping mechanism is also provided in an instrument to avoid
undue oscillations of the moving system about the rest position whenever the measurand
changes. This enables readings to be taken quickly. The damping mechanism brings about
a damping torque (M,), proportional to the angular velocity of the moving system. It is
therefore clear that damping mechanism does not affect the final steady deflection which is
essentially dependent on a balance between the deflection and cor~troltorques. A typical
way of achieving damping action is by the movement of an aluminium vane in a chamber
(dash pot) against air-friction. Figure 10.2 shows typical arrangements of the control springs
and dashpot-vane combination on the moving system of an indicating instrument.
As mentioned in Section 10.2, a pointer attached to the moving system and a graduated
scale constitute the indicating mechanism. In very sensitive instruments, a mirror may be
fixcd to the moving system. Light from a spot-source will get reflected by this minor.
Figure 10.2 : Arrangements of (a) control spr111gsand (b) air dashpot damping
Rotation of the minor along with the moving system will cause the image of the light spot
to move on a graduated scale. In this manner, minute angular deflections of the moving
system can be made to produce large deflections on the scale.
SAQ 2
What is the need for damping torque in an analog indicating instrument?
As a variant lo the notation in earlier chapters of this Block, we use here the symbol M
to denote torque.
\
Electrical Measuring
Instruments
Memaring Instruments
$\
Top pivot
i4
Cont ml spring
Cur rent in
Formr
carrying coil
( bl
where kd is a constant decided by the dimensions of the coil and the air-gap flux density.
The control torque M, due to the springs is given by
where 8 is the angular displacement of the moving system from the initial position and k, is
the spring constant. The moving system comes to rest, when M,. balances the average value
of M d . We then have the equation :
The dei'lection is directly proportional to the average current through the coil. The scale can
be graduated in terms of I. When the aluminium former moves along with the coil in the
air-gap field, it has an e.m.f. induced in it. As the former is in the form of a closed loop, the
e.m.f. causes a current to circulate through it. The interaction of this current (also called
eddy current) with the magnetic field produces the damping lorque. In these instruments
therefore, there is no need for an aluminium vane-cum-dashpot arrangement for this
purpose.
Figure 10.4 : Adaptation of a PMMC instrument as an ammeter to measure a large current nI,
If the circuit current to be measured is n times the allowable meter current, then
Eq. (10.5) tells us that a large number of shunts can be supplied with a given instrument, to
measure various currents. In these cases, the reading on the basic meter must be multiplied
by n which is called the multiplying factor of the shunt. The full-scale current and the
corresponding vqltage drop are usually marked on the body of the shunt. We thus have, for
example, shumlts of rating 60 A, 150 niV, or 20 A, 100 mV.
Example 10.1
The shunt supplied along with a 1 mA, 100 mV PMMC instrument does not have its
rating marked on it. Its resistance is separately measured and found to be 0.2004 SL.
What is the full-scale range provided by the shunt?
Solution
= 500(1- 0.002) =
Therefore
n - 1 =499
n = 500
499
Electrical Measuring
instruments
SAQ 3
The 1 mnA, 100 mTJinstrument of Exarnple 10.1 is lo he used as a vollmcter of
full-scale 100 V. Find the series rcsislancc required.
SAQ 4
Similar to the case of the shunt discussed in Section 10.4.2, define a multiplying
factor n for the series resistance of a voltmeter. Find R, in terms R,,, and n.
Example 10.2
Rs
Rm
99900
100
Eiedlical Measuring
Instruments
n is also equal to 1 + - = 1 + -
F~gure10.6 : Insertion of (a) an ammeter and (h) a voltmeter for the purpose of measurement
Load
Rsh
Figure 10.7 : Circuit of a voItammeter
10.4.5 Ohmmeter
We know from basic principles of circuit analysis that at constant supply voltage, the
current through a dc circuit is inversely proportional to the resistance of the circuit. If we rig
up a circuit as shown in Figure 10.8, we have the current given by
1, = V / (R, + R , + R , )
When
R x = m , Ix=I,=O
For
Rx=O, lx=Io=V/(Rn,+R,)
With lhe terminals a and 11 short-circuited, let R, be adjusted so that lois the full-scale
deflection of the PMMC-meter.
(10.7)
For any R, connected across the terminals a and b, the resulting current I, given by
Eq. (10.7) is a measure of the value of R,. The followillg worked example illustrates as to
how the scale of the PMMC-instrument can be calibrated in terms of resistance.
Example 10.3.
Fill up the following table showing scale markings in ohms, when the meter is
converted into an ohmmeter, employing a 3-V battery of negligible internal
resistance.
Scale div.
20
40
50
60
80
Ohms
(b)
Solution
1
1
(a)
100
Subtractiilg (R, + R,) from each of the above resistance values, the Rx-marking on
the scale can be arrived at and the table can be filled up as shown below:
Note that
(1)
the resistance-scale runs from right to left and NOT from left to right,
(2)
Electrical Measuring
Instruments
Let two such rectifiers I),, D, and a PMMC-animeterA be connected in an a.c. circuit as
shown In Figure 10.12.
If the load current i,, is si~iusoidalas in the top waveform of Figure 10.1 2(b), one half-wave
in each cycle (lower waveform) flows through the meter, Ulc other half being diverted
through the diode 112. If I is the effective value of i,,. we see lliat the average value of ,i is
one-half of the absolutc averagc value of i, and lierice equal Lo
0.5 1 1 (for111factor of a sinusoid) = ( a / n ) 1.
Example 10.4
A nor~nal(0-1A) PMMC-anl~neteris used on n sinusoidal a.c. circuit with a rectifier
as iri Figurc 10.12. The a.c. load current is 0.5 A. Firid tlic average current tlxough
the meter.
Solution
RMS valuc
Absolute average value of the a.c. currerit =
form factor of a sinusoid
= 0.225 A
2Tt
As mentioned in Section 10.4.1, the deflection of the meter will correspond to the balancing
of the average value of M, by the spring. Thus the PMMC reads ( f i / n ) 1. This reading
can be nlultiplied by ( n / a ) to give the r.m.s. value of the lo;ld current.
Jn actual practice, such an a x . ammeter incorporates a scale in which the nlultiplication has
already been carried out at Ule manufacturing stilge.
I
Example 10.5
A 0- I niA, 100 mV PMMC-instrumelit is to be converted into a (0-0.5A) a.c.
anunetcr by using two ideal rcctitlers as in the circuit of Figure 10.12. Find the shunt
resistance required ;icross the iristrument.
Solution
Meter c u r r c ~corresponding
~t
to full scale = 0.5 x Tt
0.225
Multiplyi~igfactor oP shu111= -= 225
0.001
From Eq. (10.5)
1
100
Rs, = -R,,, = - = 0.446 52
( n - 1)
224
The circuit of a rectifier-type voltmeter is given in Figurc 10.13. As in thc case of the
PMMC-d.c. voltmeter (Section 10.4.3), a series-multiplier is to be uscd for dropping a large
portion of the load voltage. The instrument being a voltmeter, the points u and b in
Figure 10.13 are lo be connected across Ule load whose voltage is to be measured. Here
again, the meter scale is marked in terms of the r.1n.s. value of ,an a.c. sine wave input
voltage. Both the rectifier type ammeter of Figure 10.12 and voltmeter of Figure 10.13 will
Electricill Measuring
Instruments
~ceCrid~iaea&
show deflection when used on d.c also if the polarity of the d.c. is such that the diode in
Ml m t~
m m d eseries with themeter conducts. Since the scale has earlier been modified for a.c. use, the
meter will not indicate the correct value on d.c. The reading will also be erroneous on ax.
circuits where the waveform of current or voltage is not sinusoidal. Commercial
rectifier-type instruments should therefore be used only on sinusoidal a.c. circuits.
For servicing work on various types of electrical and electronic equipment, one needs a
compact instrument which has a large number of voltage and current ranges. A comon form
of such an instrument, called a multimeter, has provision for meaurement of a.c. and d.c.
voltages, d.c. currents and resistances. A single PMMC movement is emp!syed for
indication on all these ranges. You will now appreciate that such an instrument is likely to
incorporate a rectifier circuit for a.c. voltage measurement and a battery for resistance
measurement.
Moving im
addail C
M, = M d
The scale can be graduated in terms of I, the effective value of the current.
SAQ 5
A moving iron ammeter has a full-scale value of 1 A which causes a deflection of
100 scale divisions. Find the current that will cause a deflection of 50 divisions.
For the SAQ, you would have got an answer of 0.707 A, i.e., a current of about 71 percent
of full-scale deflection will cause a deflection of-50 divisions. If the same instrument were a
1 A, PMMC-meter, 50 scale divisions would have covered a range of 50 percent of
full-scale value of current. It is clear therefrom that if the scale is marked in amperes, the
markings of 0.1,0.2,0.3 A in a 1 - A moving iron ammeter would all lie quite near the zero
of the scale. This is called initial cramping.
The PMMC instrument is said to have a uniform scale, since 8 is proportional to I. Since 8,
is proportiond to z2 in Eq. (10.9), the moving iron instrument is said to possess a square law
scale. In practice, suitable steps are taken in the design of a moving iron ammeter to make
the scale nearly uniform over the top 80% of scale. But the initial crowding can not be
avoided.
SAQ 6
Shown in Figure 10.15(a) and (b) are the zero and full-scale markings of two 0-5 A
ammeters, one of the PMMC and the other of the moving-iron type. Mark on each
scale, the points 1, 2.3 and 4 A. Assume that the moving iron meter has a square
law scale.
A
Figure 10.15 (b) : Moving Iron meter
As shown in Figure 10.16, the load current is passed through the primary winding of a ,
10015 A C.T. The 5 A-ammeter is connected to the secondary. A C.T. is essentially a
transformer which is a step-down transformer for currents (vide Section 6.6.2). It has less
number of turns on the primary winding than on the secondary. The transformer is specially
designed so that the ratio I, / I , in Figure 10.16 is maintained constant from very low
magnitudes of primary current up to its rated value and slightly higher. The 5 A ammeter
connected on the secondary side can have its scale modified to read from 0 to rated value of
I, (0-100 A in the example considered).
Example 10.6
The primary of a 5015 A current transformer has 6 turns. Estimate the number of
turns to be provided on the secondary.
Solution
N2=N1Z1/Z2=50x6/5=60turns.
SAQ 7
A 0-5 A moving iron ammeter is to be used in turn at three different points in a
circuit, where the rated currents are 20 A, 50 A and 60 A respectively. The
manufacturer of a C.T. supplies a single-secondary,multiple-ratio C.T.
If the 5 A-secondary winding has 120 turns, find the number of turns required
(a)
on the primary winding for each ratio.
(b)
I-
Electrical Measuring
Instruments
E I e e b i c d ~ s c b k&
Mcppluing lmtlllmenta
measured from the voltmeter reading and the transformation ratio of the V.T. The
schematic of connections is given in Figure 10.17 for a 11 kV-circuit. You may recall that
the use of a transformer for this special applicaton has already been described in
Section 6.6.2
Example 10.7
The voltage across a 22 kV load is measured using 220001110V V.T. and a (0-150 V)
moving-iron voltmeter. The meter reads 108 V. Estimate thc voltage across the load.
Solution
V.T. ratio =
22000
110
-=
200
For measurement of voltage in a high voltage 3-phase circuit, the V.T. to be used will also
have to be of the three-phase type.
SAQ 8
Point out the differences between the PMMC and moving-iron type inslruments as
regards type of damping applied and scale shape.
100 x 20000 Q
= 2MQ
=
SAQ 9
A laboratory has a (0-10V) moving iron voltmeter of resistance 2000 C! and a
(0-500mV) PMMC-voltmeter of resistance 500 Q. Which of the two is the more
sensitive meter?
Field coils
Moving coil
WectIiesl Mcssaring
Instruments
EhdlicdMrbine3&
Mcudog ht~mncdr
where k, and k, are constants. i.e., the deflecting torque at any instant is proportional to the
instantaneous power in the load.
If
v, =
vLcos at
iL= dT~,cos (at- $1,
illid
the average value of the product vLiLover one complete cycle will be the true power,
namely,
P = VLILcos $
SAQ 10
Can you guess why an electrodynamic instrument is also referred to as a
dynamonieter type one?
Just as there are circuit symbols for ammeters and voltmeters (vide Fig. 10.6), there is also a
method of representing a wattmeter in a circuit. The same is shown in Figure 10.19. The
fixed coil, called the current coil, has terminals M and L. This coil has a fewer number of
turns lhan the moving coil which is called the voltage coil. A high series resistance which
limits the current Lo a few tens of milliamperes is connected in series with the voltage coil.
The terminals of the voltage circuit carry the markings "Com" and V or a number. say,
"300 V" representing the rated voltage.
Md
Com
Example 10.9
What are the ideal values of resistance for the fixed and moving coil circuits of a
wattmeter?
Solution
Ideally, the fixed coil, like a current-measuring element, should have zero resistance.
The voltage-coil circuit should likewise have infinite resistance in an ideal case,
since it is used to sense Uie voltage.
these are:
2.5 A, 150 V, 375 W:
5 A, 300 V, 1500 W;
10 A, 250 V. 2500 W etc.
Figure 10.20 shows a block representing a source on the left side and a block on the right
side standing for the load. The terminal 'M' of the current coil is connected to the mflinsor
the source and the terminal 'L' to the load. The reference load current iL then flows through
the instrument from M to L. The terminal "Com" of the voltage coil circuit is connected to
the reference positive terminal of the load, so that vLis applied as a voltage drop from the
terminal marked 'Com" to the other terminal '300 V' of thevoltage coil circuit.
,-.
A diagram, showing the method of connection is usually provided by the
wattmeter-manufacturer on the instrument cover. It is only with this type of connection that
the wattmeter will show a positive deflection to lndicate the power consumed by a load
supplied from a source. If the connections cither to the fixed coil or the moving coil (but not
both) is reversed, the pointer of the wattmeter will kick back.
Example 10.10
Two load impedances Z1 and & are connected in series across an a.c. mains. Show
the connections of a wattmeter to measure the power consumed in 2 2 .
Solution
The same current flows through Z1 and 2 2 . Hence the fixed coil can be placed in
series. But the voltage coil circuit should get across itself only the voltage across %.
The connections are as shown in Figure 10.21.
Electlid Merrsluimg
Iostrmnents
Electrid M~chiiws&
M e a s d g Indmneh
(10.12)
Consider Figure 10.21, which is the phasor diagram for a balanced star-connected load.
(Compare with Figure 4.16)
Figure 10.22 : Phasor diagram of voltages and currents in a balanced star-connected three-phase load
Example 10.11
What are the angles x and y in the phasor diagram of Figure 10.22 ?
Solution
(i)
x = (30" + Cp)
and
P2= VLILcos C
(p - 30),
P, + P, = 6 vLiL
cos $
so that
Thus, two wattmeters, properly connected, are sufficient to measure the power in a
three-phase, three-wire circuit. The method of connecting the wattmeters physically into the
circuit is shown Figure 10.23.
Source
Figure 10.23 : Insertion of two wattmeters for power measurement in a 3-phase, 3-wire load
Example 10.12
A balanced inductive 3-phase load of power factor 0.8 draws 10 A from a 400 V,
3-phase supply. Find the readings of the two wattmeters connected as in
Figure 10.23. Verify that their sum gives the total load power.
Solution
Load power PL =
6 VLILcos $
= 5 S 8 m = P,.
We know from Unit 8 that the power factor of operation of an induction motor would be
low under no-load conditions. (vide Example 8.6). Attempt the following SAQ.
SAQ 11
A 400 V, 3-phase induction motor runs on no load, drawing a current of 4 A from the
mains. If the power factor of operation is 0.3, what h e the readings of the two
wattmeters connected for power measurement?
rTo counter
electromagnets causes a torque to act on the moving system. The current coil circuit is
designed so that the field produced by the current magnet is in phase with the load current
Ik The voltage-circuit is made predominantly inductive so that the flux produced by the
by nearly 90". Under these conditions, it can be
voltage magnet lags the load voltage
shown that the driving torque M, acting on the moving system is proportional to the load
power PL = VLZLcos $. As mentioned earlier, there is no control spring. Under the action of
the driving torque, the moving system and the disc will accelerate. A permanent magnet
provided at one end of the disc causes an e.m.f. to be induced in the disc as soon as the
moving systemstxts from rest The currents induced in the disc as a result of this e.m.f.
cause a torque on t h b g system, akin to the damping torque in a PMMC instrument
(vide Section 10.4.1). This damping torque, called braking torque Mbin this case, is
proportional to the speed N of the disc. We have
vL
The disc accelerates and attains a steady speed at which the braking torque equals the
driving torque.
here PL is the power. Hence the speed of the disc is proportional to the power in the load.
The number of revolutions of the moving system or the aluminium disc in a given interval
of time fromt= t, t o t = t, is
where W is the energy consumed in the particular time interval. The counting of the
number of revolutions is carried out by a mechanical arrangement of wheels and pinions
and a final read-out, usually in digital form is provided against a window. Readings are
taken by the Utility Company at one or two-month intervals and the consumer is billed for
his energy consumption during the period.
The above energymeter can be used only on ac circuits, since the action of the meter
depends on induction principle.
Example 10.14
Mention the type of torque that is present in an indicating instrument but is absent in
an energymeter.
Solution
Control torque
SAQ 12
A 10 A, 230 V energymeter carries full-load current at 0.5 lag. The meter constant is
600 revolutions per kwh. Assuming the meter to be error-free, find the time taken by
the disc to make 10 revolutions.
and
In the above equations, the subscript eq stands for "equivalent value" referred to one side.
Because of inherent imperfections in manufacturing, the reading of any instrument is likely
to differ from the true value of the quantity being measured with it. The manufaciurer of an
instrument normally guarantees a certain accuracy for the instrument by specifying the
maximum margin of error that can be present in any reading taken on the instrument. He
expresses this magnitude as a percentage of full-scale reading. Anuneters, voltmeters and
wattmeters for industrial use are typically nlanufactured for a maximunl uncertainty of 1 or
1.5 per cent of full-scale. Thus, any reading taken on such a voltmeter of range (0-150 V)
can have a maximum error of 1.5 V or 2.25 V respecitvely. The sign indicates that the
error could lie on the positive or negative side. If a reading of 150 V is taken on a (0-1 50 V)
instrument having an uncertainty of 1 per cent of full-scale, the maximum percentage error
in the reading will be (+ 1.5 / 150) x 100 = 1 per cent. On the other hand, let us say that a
reading of 75 V is obtained on the same instrument when connected in a circuit. This
reading can have a maximum relative uncertainty of (+ 1.5 /75) x 100 = 2 percent.
10.8.2
In the light of the preceding discussion, one general rule that may be stated is :"Choose the
range of an instrument so tltat the expecterl reading will he as nearfull-scale vulue as
possible. " For example, if the current under measurement is likely to be 2 A, do not choost
a meter of full-scale 5 A. Try to use an ammeter of 2.5 A full scale value.
Turning to Eq. (10.14), Zeqis given by the ratio of V/I. Errors in thc measured values of V
and I could be of opposite nature i.e., Vcould have been measured as ( V + 6V) and I as
(I - 61), where 6V and 61 are the maximum possible errors in V and I. The measured value
of Zeqcan lie between (V + 6V) / ( I - 61) and (V- 6V) / ( I + 6)3. The point to bc appreciated
is that when a quantity is computed using an algebrac expression involving a number of
measured variables, the error in the result will depend on the errors occurring in the
measuremen1 of the different variables and the algebraic relation involved. Let us consider
the following example.
Example 10.14
0.1
Resistance before application of load = 600 -x 600
100
Electrical Measuring
Instruments
= 603
0.1
+x 603
100
Increase in resistance can be any value between 603.603 - 599.4 = 4.203 Q on the
high side and 602.397 - 600.6 = 1.797 L2 on h e low side.
While the nominal value of increase in resistance as obtained from the measurement
is 603 - 600 = 3 Q, the actual in value could have any value lying between 1.797 and
4.203. The maximum percentage uncertainties would be
= k 40.1 percent.
The above example is revealing in the sense that the percentage uncertainty in a quantity,
which is estimated as the difference between two nearly equal quantities could be much
larger than the maximum percentage uncertainty in the two individual quantities. A second
general rule therefore is: A quantity which is expecteii to be of srrtall mrzgnitude should he
measured directly and not estimated as a clifference between two individually measured
Inrge qunntities.
A th~rdpoint to be remembered is that the tendency to go in for highly accurate instruments
for every measurement should be avoided. For example, if a particular method of
measurement can yield an overall result of, say, at most 5 per cent accuracy, there is no
need to use instruments which have a very high accuracy, say, of 0.01 per cent for one of
the component nleasurements.
10.9 SUMMARY
In this unit, we first learnt about Ule difference between analog and digilal indicaling
instrumncnls. We next studied briefly the constructional details and working of analog
electrical measuring instruments of the cleclrolnagnetic type. Ammeters and voltmeters for
use in d.c. and a.c. circuits were discussed. The electrodynamic instrument suitable for
power measurement was introduced and its application outlined. A brief idea has been
given abouut h e induction type energyn~eterused by any domeslic electricity consumer.
That every measurement is marred by unavoidable errors is pointed out and the need for the
proper choice of instruments for a specific measuring task is stressed.
It 1s true h a t conventional electrical measuring instruments for current and voltage that
havc been discussed in this unit are slowly giving way to electronic instruments. They are
slill however in common use. Principles of measurement and measuring techniques are
mostly similar for both electrical and electronic instruments. We shall be learning about the
laltcr in Units 15 and 16.
Should you get fascinated towards the pursuit of the topics in this subject, please go through
the books listed under "Further Reading" at the end of this Block.
t
I
SAQ 1
Advantages :
Disadvantages :
SAQ 2
The damping torque enables the moving system to come to the equilibrium position
quickly, without undue oscillations whenever the moving system goes from one
steady state position to another. ?he new reading can then be easily noted.
SAQ 3
In Eq. (10.6), we shall insert the values of V, Imand Rm.We then get
SAQ 4
Multiplying factor n =
SAQ 5
Let I be the current. From Eq. (10.9),
(a)
The PMMC ammeter has a uniform scale. Its graduations are clearly marked
out on the figure (a) below at equal distances.
(b)
0 .
Percentage
deflection
16
36
64
100
SAQ 7
(a)
(b)
SAQ 8
M.I.
Type of Damping
Air friction
Scale shape
Uniform
Square law
SAQ 9
1
-
$1
1
For the PMMC-voltmeter, -=
Ih
1
(10/2ooO)
1
(500 x 10-~/500)
The instrument is a meter working essentially on the principle of the dynamo. Hence
The name dynamometer.
SAQ 11
Notice that one of the wattmeter readings comes out as negative. In practice, the
pointer detlects to the left ol the zero-position. You should reverse the connections to
'Corn' and 'V' of the pressure coil circuit and note down the resulting reading as a
negative value.
Input Power Pi = P I + P2
As a check, we can see that 4 5 x 4 0 0 x 4 x 0.3 = 831.4 which agrees with the
algebraic sum of Y 1and P?. It ;nay therefore be stated with reference to Eq. (10.13)
that the algebraic sum P I + P2 of the two wattmeter readings gives the tolal power in
a three-phase load.
SAQ 12
kW
Therefore t =
60
1.15
- - 52.17 s.
2. A. E. ' ~ i t z ~ e r aand
l d others, "Electric Machinery", McGraw Hill, 1985.
3. M. R . Stoul, "Basic Electrical Measurements", Prcnlice-Hall Inc.. Englewcu,d Cliffs.
N.J. 1960.
4. WiD. Cooper and A.D. Hellrick, "ElecLronic lnslrumentalio~~
ant1 Mcilsurer~lenL
Techniques", Edn. 3, Prentice-Hall of Tndia, New Dclhi, 1985.