Cardiac Simulation: Technical Specifications

The Cardio ECG simulator can generate realistic cardiac waveforms for testing diagnostic equipment. It simulates 12-lead ECGs and allows testing of beat rate measurement, noise filtering, and other functions. The simulator produces various waveforms, has adjustable beat rates from 30-240 BPM with up to 20% deviation, multiple noise sources, amplitude adjustment, and 12 leads. It is compact at 150x90x26mm and 400g with connections for snap, banana, and pin cables.

Uploaded by

roekan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
51 views

Cardiac Simulation: Technical Specifications

The Cardio ECG simulator can generate realistic cardiac waveforms for testing diagnostic equipment. It simulates 12-lead ECGs and allows testing of beat rate measurement, noise filtering, and other functions. The simulator produces various waveforms, has adjustable beat rates from 30-240 BPM with up to 20% deviation, multiple noise sources, amplitude adjustment, and 12 leads. It is compact at 150x90x26mm and 400g with connections for snap, banana, and pin cables.

Uploaded by

roekan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Technical

Specifications
Waveform

Cardiac Simulation
The Cardio ECG simulator stands among its
competitors for generating realistic cardiac waveforms, an
advantage that gives you the certainty that any test conducted with this
simulator will be repeatable with a human patient.
Twelve leads ECG simulation is suitable to test modern
diagnostic equipment, including waveform recognition devices.

Advanced Analysis
Using the Cardio simulator you will also be able to
test the noise filtering systems and beat rate measurement functions of the
diagnostic device under test.
To perform noise testing, multiple real life noise sources are included. You can
select between line noise (related to poor ground connections), base line
wandering (related to poor electrode conductivity), muscle tremor, power supply
noise and more.
The beat rate setup function combines the selection of an arbitrary rate with
the addition of beat to beat variability, a key performance testing to make sure
diagnostic devices operate properly under critical conditions.

ECG, Sine, Triangular,


Square, Pulsed and DC

30 BPM to 240 BPM


Beat Rate0% to 20% deviation
Frequency

1/8Hz to 100Hz

Noise Sources

Line, Ripple, Random,


Baseline W., Tremor

Amplitude

0.5mV, 1mV, 2mV, 4mV


0% to 20%* deviation

Leads

12 (Snap / Banana / Pin)

Hi-Signal Output

200x Lead II plus DC

Output Impedance

22000 unipolar

Power Source

Standard 9V Battery

Product Size

150mm x 90mm x 26mm

Package Weight

400gr

Biopotential multi-connectors
The new multi-connectors give the simulator
enhanced compatibility, enabling its
connection to 4mm snap, 4mm banana and
2mm pin patient cables. Youll no longer need
expensive adapters that degrade reliability
and performance.

Neoprene travel holster


The new Neoprene travel holster protects the
simulators from shocks and moisture while
easing its transport. Its flexible structure
helps you carry cables or even a spare
battery.
Attach it to your belt to have your simulator at
hand every time you need it.

Retail box for global market


The new retail product box enables our
customers to resell the simulator in the US
and other global markets. It includes the UPC
product barcode, FCC exemption
declaration and an eye-catching graphic
design that will help you boos t your sales.