Outside Micrometer Calibration Procedure: 1.0 Standards and Equipment
Outside Micrometer Calibration Procedure: 1.0 Standards and Equipment
NOTE: Standards and equipment used must have a valid calibration certificate
Step 1:
2.
Zero the Outside Micrometer at the start and adjust as required by the
manufacturers specifications. If you cannot zero it then use a gauge
block and adjust the micrometer to that value.
3.
When testing the Outside Micrometer, one of the points must be near
the lower limit that the instrument can measure, another somewhere in
the middle, and the third near the upper limit.
4.
Measure the ambient temperature and record it. If the temperature is <18oC or >24oC,
see the Division Manager for further instructions.
Page 2 of 2
Step 2:
Acceptable Limit:
Step 3:
Visual, Touch
Acceptable Limit:
Step 4:
Acceptable Limit:
Step 5:
Using gauge blocks and the smallest anvil, measure and record
5 different lengths. For micrometers with more than one anvil,
make a separate measurement with each anvil
+/- 0.0002 or 0.005 mm
Acceptable Limit: